Endotoxin Test Protocol For use of the Charles River Laboratories Endoscan-V Kinetic Turbidimetric Software and Reader I. Materials Needed: Charles River Laboratories Endoscan-V Software and PC BioTek ELx808 reader KTA Lysate 0.06EU/mL Sensitivity (Catalog #R15006) Control Standard Endotoxin -CSE (Catalog # E110) Lysate Reagent Water – LRW (Catalog #W110) Pyrogen-free 96 well test plates (Catalog #M905) Sterile, Pyrogen Free Pipette tips Pipettes for sample dispensing Pipettes for dilutions Pyrogen-free sample and dilution containers (Fisher Catalog#149592A) Gloves NOTE: Do not use polypropylene containers for storage or dilutions of samples as Endotoxin in the sample will stick to the container and lower recovery for the sample. II. Purpose and Background: The Charles River Laboratories Endoscan-V Kinetic Turbidimetric system consists of spectrophotometer, PC and software system using Kinetic Turbidimetric (KTA) Limulus Amebocyte Lysate (LAL) and Control Standard Endotoxin (CSE) for the creation of a known value standard curve. The purpose of the test is to provide quantitative Endotoxin results for in-process and end product samples. The assay sensitivity available for use is 0.06EU/mL using a standard curve of 5-0.05 EU/mL III. Making Standard Curve Solutions 1. Allow all reagents to warm to room temperature before being testing. Also remember to turn the reader on prior to testing to allow it to run through its start up process and warm the unit to 37°C. 2. Once the CSE has warmed to room temperature it can be used to make dilutions for your known standard curve. If the CSE has not been rehydrated, rehydrate it now per the instructions on the Certificate of Analysis that came with that lot of CSE (see CRL CofA folder located next to the Endoscan-V computer). a. NOTE: When it is first rehydrated the CSE vial must be vortexed for a minimum of 5minutes prior to initial use and vortexed day of use for 12minutes at minimum. 3. Depending on the concentration of the CSE once rehydrated as described in the CofA you will make dilutions to obtain at minimum 1mL of standard solutions of the following values: 5.0 EU/mL, 0.5EU/mL and 0.05 EU/mL a. EX: CSE labeled as – 1000EU/mL i. Perform a 1/20 dilution → 0.5mL/9.5mL → 50 EU/mL ↓ ii. Perform a 1/10 dilution → 0.5mL/4.5mL → 5.0 EU/mL ↓ iii. Perform a 1/10 dilution → 0.5mL/4.5mL → 0.5 EU/mL ↓ iv. Perform a 1/10 dilution → 0.5mL/4.5mL → 0.05 EU/mL IV. Setting up Reader Plate Template 1. Open the Endoscan-V software on the computer in NB1-10 by clicking on the Endoscan-V icon on the desktop and log on to the system. The screen will prompt you to choose to open an existing file or create a new file – choose new plate. 2. Next the Properties screen will pop up. On this screen you can see the default setting for the system (i.e. read times, wavelength setting, read intervals). You may also enter plate properties at this time (i.e. CSE, lysate and LRW lots and expirations) and well group properties. After closing this screen you will be able to begin setting up your plate template. Each 96 well plate will run one set of standards, one negative control and 20 samples. 3. First begin by clicking and dragging the 5.0; 0.5; 0.05 standard curves to the top left hand corner of the plate. Next type CTRL1 into the next set of wells below (this is your LRW negative control and the system will not provide an automatic spike for this well group; you only need to type it in one well – the other will populate automatically). 4. At this point you may start setting up the plates depending on the samples you have to run. If you are running standard stock product samples, use the product database (you will see this below your template) to “drag and drop” the samples you would like to test. You will need to add/enter the lot number of the product you are testing each time as well as check and confirm the dilution is correct. 5. If you are testing samples that are not in the database (i.e. customs or other samples) simply type the name of the sample in the appropriate well and hit enter – the system will automatically create one sets of replicates (this it’s default setting) and one set of automatic spike wells (positive product control). Continue adding samples until the plate is full a. NOTE: It is advisable to run a full plate of samples if at all possible to ensure cost efficiency b. NOTE: One bottle of lysate will allow you to run 10 samples + standards and negative control; you will need 2 full bottles of lysate to run a full plate of 20 samples, standards and control. c. NOTE: If you are running dilution series of one sample it is advisable for you to add these samples in succession on the plate as this will make reading and interpreting results easier. V. Pipetting the 96 well plate 1. 2. 3. 4. Once the template is complete on the computer you can begin to aliquot the samples onto the 96 well plate per your template. i. NOTE: Be cautious when pipetting so as not to splash into adjacent wells and to ensure the samples are dispensed into the proper wells as laid out in the template you created. Aliquot 100µL of all samples, standards and negative controls. Begin by aliquoting the LRW, then the standards from lowest to highest concentration (you may use the same pipette tip for this task). Next aliquot your samples as laid out on the computer template. i. NOTE: A minimum of 400µL of sample is required to run this test (100µL for each well, 2 sample replicates and 2 spike replicates). Next aliquot the spike into the designated spike wells. Use 10µL per spiked well of the 5.0 EU/mL standard to do this. Ensure you do not splash into adjacent wells when you do this. i. NOTE: The spike concentration should be the equivalent of the middle of your standard curve (i.e. 0.5 EU/mL). In order to achieve this concentration on the plate you must use the 5.0 EU/mL standard as 10µL of spike into 100µL of sample is effectively a 1/10 dilution providing you with your desired 0.5 EU/mL spike concentration. Lastly aliquot 100µL of rehydrated lysate (rehydrate lysate per instructions on the vials – do not vortex, simply rotate up and down until solution turns clear before use) to each well. Once the lysate is dispensed into the wells the test has effectively begun. i. Look over the plate quickly and check for bubbles and that you have dispensed lysate into all the wells prior to placing the plate in the reader. NOTE: Once lysate is rehydrated discard the rubber stopper and replace with parafilm. 5. VI. Immediately upon dispensing lysate place the plate in the reader with the top off and close the lid and hit the green arrow on the top just below the tool bar. Or click Tools and Start Collection. The unit will shake for a short period of time (this is normal and acts to remove bubbles from the wells and mix the samples) and then data collection will begin and the system will run the test automatically and shut off automatically. Reading Reports, Interpreting Results and Troubleshooting 1. Valid Test Parameters a. Spike Recovery: A valid test has a spike recovery between 50-200% i. ([Spike Well – Neg Product well)/Spike Value]*100 = Spike Percent recovery. If this value is out of range dilute the sample until it is brought within range – allowing for maximum valid dilution. b. CV: Coefficient of Variation = (Standard deviation/mean) of the test replicates. A valid test has a standard and control CV of <10% and a sample CV of <20%. Anything higher than this indicates a significant variation in the reaction times of the two replicates indicating an invalid assay. 2. Endotoxin Calculations: a. Endotoxin Limits: = K/M i. K = the threshold pyrogenic dose in endotoxin units/kg Experimentally determined in rabbits, confirmed in humans K = 5 EU/kg for drugs with a route of administration other than intrathecal K = 0.2 EU/kg for drugs administered intrathecally ii. M = maximum human dose of product/kg of body weight in a single hour period If the pediatric dose/kg is higher than the adult dose, then the pediatric dose/kg = M b. Maximum Valid Dilution: i. MVD provides an upper bound for dilution that still provides for endotoxin detection at the endotoxin limit ii. MVD = (Endotoxin Limit)(Concentration of sample solution) λ Where: Endotoxin Limit = K/M Concentration of sample solution = concentration in units/mL λ = confirmed label claim sensitivity of the GC lysate or lowest point on the referenced standard curve *If the limit is in EU/mL, the concentration of the sample solution is not needed *****MVD is unitless…it is a dilution factor***** VI. Media Lab Sample Acceptance Criteria Table Description Grace’s Insect Cell Media Cell Culture Media MilliQ Buffers L-Glutamine Penicillin/Streptomycin Cell Culture Media Concentrated Acceptance Criteria ≤20.0EU/mL ≤1.0EU/mL ≤0.05EU/mL ≤1.0EU/mL ≤0.5EU/mL ≤0.5EU/mL @ 5mL/500mL in DMEM ≤1.0EU/mL of a 1X Solution