University of Michigan Health System Program and Operations Analysis (POA) Analysis of Endoscope Reprocessing Final Project Report To: Lisa Sturm, Managing Director, Infection Control & Epidemiology Mary Duck, Industrial Engr. Expert & Lean Coach, POA Matthew Claysen, Industrial Engr. Intermediate, Operating Rooms Rama Mwenesi, Fellow, POA Mark Van Oyen, Instructor, Industrial and Operations Engineering From: IOE 481 Project Team 7 Danielle Hedden Alexandra Keyser Dominic Lefere Date: December 10, 2013 1 Table of Contents –FIX THIS Executive Summary Introduction 5 Background 3 6 Key Issues 7 Goals and Objectives Project Scope Methodology Findings 7 7 7 11 Conclusions 13 Recommendations 13 Expected Impact 14 References 15 Appendices 16 Appendix B: External Audit 19 Appendix B: Internal Audit 19 Appendix B: Matrix/Data sheets 19 2 Executive Summary Endoscopes are used in approximately 32 sites in the University of Michigan Health System (UMHS). These tools touch patients’ mucus membranes, and as such, have to be high-level disinfected, which is a process of chemical exposure that removes and/or renders inactive bacteria, viruses, and most spores. The process of removing viruses is especially important, so that viruses such as HIV and hepatitis B and C, are not transferred to a patient. High Level Disinfection is critical when reprocessing endoscopes. The Managing Director of Infection Control stated that The Joint Commission (TJC) cited UMHS on their scope reprocessing, thus elevating this to a high priority for the institution [1]. Infection Control needs better auditing tools to focus their efforts on problem areas in the reprocessing process. The Managing Director asked a student team from Industrial and Operations Engineering (IOE) 481 at the University of Michigan to trace the life cycle of the endoscopes at some of the operating sites and develop internal and external audit tools for the institution to use when evaluating its endoscope reprocessing. After examining the current situation, the team developed a comparison matrix for comparing the different sites visited to the institutional endoscope reprocessing standards put forth in a policy by Infection Control [2]. This matrix helped build the audit tools that the team is recommending. Background The Managing Director of Infection Control stated [1] that there have been occasional problems with unprocessed endoscopes being mistaken for clean and being present in the procedure room. Due to precautionary flushing done before procedures, this issue has never reached the patient. The Director believes there are many possible reasons why unprocessed endoscopes in the procedure room could occur: complicated endoscope reprocessing process, fast paced and crowded work environment, variations in training, and infrequently performed task at some sites. Methods The team performed six types of tasks to evaluate the reprocessing process and develop audit tool recommendations. • Selected key sites in UMHS to observe. The team worked with the Managing Director of Infection Control to determine which of the 32 sites that handle and reprocess endoscopes the team would visit. Sites with high volume (high volume is considered to be around 50-80 scopes a day, while low volume is considered to be only a couple a week) of endoscopes were selected, as well as sites that represent the clinic environment (low volume). The selected sites were Medical Procedures Unit (MPU), Medical Procedure Center (MPC), CS Mott’s Children’s Hospital Operating Room (OR), University Hospital’s (UH) OR, Taubman Otolaryngology Clinic, Taubman Urology Clinic, and the Cancer Center Urology Clinic. Out of these, the sites of main focus were the MPU, MPC, and Mott OR. UH OR, Taubman Otolaryngology Clinic, Taubman Urology Clinic, and the Cancer Center Urology Clinic were sites that did not have a high volume of endoscopes used each day, so they were a secondary focus, and would not have as many observations. 3 • Observed and interviewed staff at each selected key site. At the sites of main focus, the team observed at least three times. The sites that the team visited that were not high volume were only observed once or twice. The team completed a total of 40 hours of observation. The team observed the set up and take down of the endoscopes in the procedure rooms, and then the reprocessing of the endoscopes in decontamination. While observing, the team also asked the staff questions. • Conducted surveys at all UMHS endoscope sites. Since the team was unable to visit and observe at all 32 sites due to time constraints, the team developed and distributed a survey to all of the 32 sites that handle and reprocess endoscopes. The purpose of the survey was to provide baseline data for sites not observed, and provide additional data for observed sites. The survey questions were linked with the observation findings. • Performed literature and internet searches for endoscope reprocessing guidelines and audit tool formats. The team researched and found many useful references from journal articles and manufacturers instructions on endoscope reprocessing guidelines and typical audit tools off which to base their recommendations. • Compared observed sites’ endoscope reprocessing to best practices found in Infection Control’s endoscope reprocessing policy [2]. Using the detailed observation data gathered, the team compared each site to the best practices and determined a percent of best practice that each site was following. • Formulate internal and external audit tools. Based on the reprocessing areas that had the most deviations from the best practices, the team formulated an audit tool that focused on those problem areas. Findings After completing observations the team analyzed results from each site and compared these to Infection Control’s policy. The team observed that most of the high traffic scope reprocessing sites achieved percent compliance ratings of 87% or greater, however each still had at least one category of the process that needs improvement. Sites that reprocess scopes on a less frequent basis received, for the most part, much lower percent compliance ratings, as shown in Table E-1. Survey responses uncovered many of the same areas that deviated from the policy. Based on the surveys, inconsistencies exist across sites in the following areas: • Reprocessing Room: o Brush reprocessing o PPE requirements (especially glove use) • Transport of scopes from room to room • Set up procedures within the procedure room 4 Table E-1. Percent compliance to Infection Control’s policy across visited sites MPU Overall 91% Set Up 91% Pre-Cleaning 100% Transportation 100% Reprocessing 92% Storage 100% Miscellaneous 70% MPC 91% 55% 100% 75% 100% 100% 100% Mott-OR 86% 100% 100% 100% 84% 100% 83% UH-OR 97% 100% 50% 100% 100% 100% 100% Oto 73% 100% 100% 25% 88% 100% 50% Cancer Center 94% 75% 100% 75% 100% 100% 100% Overall 89% 87% 92% 79% 94% 100% 84% The team researched endoscope reprocessing guidelines from the Community and Hospital Infection Control Association (CHICA) – Canada. Documents from this source emphasized the importance of specific steps within the reprocessing process and helped the team better target potential sources of error. These findings were used to formulate better questions for use within the audit tools. Conclusions From these findings, the team concludes that most errors in the reprocessing occur in the reprocessing room itself during high-level disinfection. The team also noted that more errors tend to occur at locations that reprocess a smaller volume of endoscopes. Additional sources of common error include: Scope transportation Brush reprocessing Setup procedures Recommendations During observations, the team saw most deviations from best practice to occur during highlevel disinfection. The audit tools that the team developed focus on these areas and can be used to accurately pinpoint where deviations from best practice occur. By doing regular audits with the audit tool, Infection Control can identify errors in the process and focus their improvement efforts on the areas with the most errors. The internal audit will be used by sites to track their compliance with Infection Control’s policy on a more frequent basis. 5 Introduction Endoscopes are used in approximately 32 sites in the University of Michigan Health System (UMHS). These tools touch patients’ mucus membranes, which classifies them as semi-critical medical devices, according to the Spaulding Classification of Medical Devices. According to this classification system, all semi-critical medical devices must be high-level disinfected [3], which is a process of chemical exposure that removes and/or renders inactive bacteria, viruses, and most spores. The process of removing viruses is especially important, so that viruses such as HIV and hepatitis B and C, are not transferred to a patient. High Level Disinfection is critical when reprocessing endoscopes. The Managing Director of Infection Control stated that The Joint Commission (TJC) cited UMHS on their scope reprocessing, thus elevating this to a high priority for the institution [1]. The Managing Director stated [1] that there have been occasional problems with unprocessed endoscopes being mistaken for clean and being present in the operating room. Due to precautionary flushing done before procedures, this issue never reaches the patient. The director believes there are many reasons this could occur: endoscope reprocessing is a complicated process, the work environment is fast paced and crowded, there may be variations in training, and for some sites the task is not done very often. Infection Control needs better auditing tools to detect errors in the process, so that they know where to focus their improvement efforts. The Managing Director has asked the Industrial and Operations Engineering (IOE) 481 student team to trace the life cycle of the endoscopes, from the time they are taken out of the cabinet for use, to the time they are put back in the cabinet after being cleaned. The team has documented the reprocessing process for eight sites within UMHS, and has developed auditing tools to target deviations from Infection Control’s policy for the sites and Infection Control. This report presents the team’s data collection and analysis, findings, conclusions and recommendations and includes project deliverables. Background According to the Managing Director of Infection Control, endoscopes must be highlevel disinfected according to the Spaulding Classification scheme, which states anything that comes in contact with a mucus membrane must be high-level disinfected [1]. Failure to correctly follow endoscope reprocessing guidelines, or using damaged endoscopes accounts for all of the reported cases of bacterial and viral transmissions. The reprocessing of endoscopes is crucial to preventing infection. The transmission of bacteria and viruses from an endoscope procedure is rare; it happens in 1 in 1.8 million cases according to M. Juan Day, M.E. [2]. However, the types of diseases a person may contract can be dangerous, such as HIV and hepatitis B and C. UMHS has a policy for endoscope reprocessing which should be followed, but according to the Managing Director of Infection Control [1], there are occasional incidences where unprocessed endoscopes are mistaken for reprocessed. To prevent the unprocessed scope from being used in a procedure, the instrument is flushed with water before use, and at this step a dirty scope is identified. Based on the UMHS Infection Control Committee’s policy on endoscope reprocessing and high-level disinfection, “it must be clearly indicated that the endoscope is dirty” after it has been used; however, this step is not always followed [4]. Infection Control at UMHS is introducing a new policy that will use a visual cue to determine if an endoscope has been high-level disinfected. The cue will have to be removed before the endoscope can be used to error-proof the process. An endoscope will only be used if a visual cue is attached. Infection Control conducted a root-cause analysis [1] on why unprocessed endoscopes are being mistaken for reprocessed. They found in some instances departments were unclear on who was responsible for reprocessing the scopes and assumed another department was accountable. Variations in training could also account for discrepancies across sites. The Managing Director of Infection Control feels a tool to help identify where deviation from the standard procedure is needed. This project sought to decrease the number of dirty instruments being mistaken for clean by developing audit tools to be used internally by the sites, as well as externally by Infection Control. Key Issues The following issues were the primary reasons for this project: •Occasions have occurred where endoscopes that were not properly reprocessed are mistaken for clean •No visual audit system exists to differentiate between endoscopes that have been high-level disinfected and those that have not •Responsibility for cleaning endoscopes is unclear •Party responsible for cleaning endoscope is not fully trained Goals and Objectives To determine the best questions for an audit tool for endoscope reprocessing for UMHS, the team accomplished the following tasks: • Observed the steps in the current endoscope usage and reprocessing • Identified areas for improvement This information has helped the team develop recommendations to: • Have fewer instances of dirty scopes • Develop audit process Project Scope The main focus of this project was tracing the lifecycle of an endoscope including: selecting an endoscope for a procedure, transporting the scope, using in procedure, cleaning, high-level disinfecting, and returning the endoscope to storage. The team has determined best practices and observed endoscope reprocessing for eight sites throughout UMHS. This project did not focus on how the endoscopes are used or stored. The quality of the cleaning at the micro level was not focused on in this project. Any concerns with which chemicals should be used were also not a focus for this project. Due to the large number of sites, only eight were visited during this project. Methodology The team used two data collection techniques to ensure successful completion of the project. Observations served as the main type of data collection, but survey data was also used to document the endoscope reprocessing process, especially for those sites the team was not able to visit. Observations The team logged 40 hours of observation data at eight sites within UMHS. To perform observations, the team worked with the Managing Director of Infection Control, Intermediate Industrial Engineer, and Infection Control Practitioners (ICPs) to arrange the dates and times of observations. The team sent its availability for observations each week to its ICP contact, who distributed the availability to all the other ICPs so that observations could be arranged. The team observed the current endoscope lifecycle. Lifecycle is defined as the time from when the scope is taken from storage until it is returned to storage. The following activities were tracked with observations by using a standard collection form found in Appendix A: Transportation of scopes to procedure room Set up of the scope Pre-cleaning Transportation from procedure room to reprocessing room Reprocessing Transportation from reprocessing room to storage The main areas where the team collected observations are: Medical Procedures Unit East Ann Arbor Medical Procedures Center C.S. Mott OR These areas are high traffic endoscope reprocessing areas, and the Managing Director of Infection Control requested the team focus on these areas. The team also collected limited observations at a few other endoscope reprocessing locations at the request of the Managing Director of Infection Control. These areas do not do endoscopic procedures often, and therefore the team only visited these areas once or twice. These areas are: UH OR Cardiovascular Center (CVC) Taubman Center Otolaryngology Clinic Taubman Center Urology Clinic The Cancer Center The team planned on collecting 10 observations from each of the main areas, and one to two observations from each secondary site. Table 1 shows the number of observations collected from each visited site. Table 1. Observation sample size at visited sites Site Sample Size (n) 5 MPU 5 MPC 3 Mott-OR 1 Cancer Center 1 CVC 1 Oto 1 UH-OR 1 Urology Obtaining the desired number of observations was important for the team to better understand endoscope reprocessing for individual sites in the current state. The team used the observational data to determine where discrepancies exist between best practice and observed practice. The areas of discrepancies allowed the team to develop an audit to identify these issues in other sites, by determining what questions the team should ask in the audits to reveal the problems at other sites. To ensure that the team received unbiased data, the team would return to the main observational areas for multiple observations. The managers would then let the team observe unsupervised, which allowed for better data because with no managers in the room, the staff was more likely to reprocess the scopes as they normally would in an empty room. Surveys The team used survey data to supplement observations. Due to time constraints, the team was unable to visit all 32 sites, so surveys were used to extend the project reach to those sites the team did not physically visit. Infection Control facilitated distributing surveys and returning surveys to the team. Both paper and online surveys were used. Two different surveys were distributed, both electronically and manually, to all 32 sites, and each site was to complete the survey either electronically or on paper, which ever they preferred. One survey was for the employees that set up the endoscopes in the procedure room, and the other survey was for employees that reprocess the endoscopes. 50 surveys were returned from the medical assistants and reprocessing technicians. 31 of these surveys were procedure room surveys, and 19 surveys were reprocessing room surveys. SDS meetings The Sterilization/Disinfection Subcommittee meets every other week to discuss ways to improve infection control processes. The scope of these meetings goes beyond just endoscopes. By attending, the team gained more exposure and visibility throughout the departments they were observing. With this attention, staff throughout the departments had a better understanding of what the team was doing and were more willing to schedule observations. Literature search The team gathered research on endoscope reprocessing methods and audit tools. The team used the references to base its recommendations for audit tools. Based on the research of reprocessing method standards, the team formulated questions for the audits to target critical steps in reprocessing. The questions regarding the reprocessing steps were modeled after the audit tool research, so as to ask questions in a way that does not bias the answers of the staff to the correct answer. The team researched endoscope reprocessing practices from other institutions to compare UMHS’ policy to others and gain a deeper understanding of the expectations for this process. The team specifically looked at best practices of the Community and Hospital Infection Control Association (CHICA) – Canada, as resources were easily accessible. By comparing UMHS’s policy to that of CHICA, the team formulated pointed questions to ask while the team was observing. Before constructing the internal or external audit tools, the team researched similar tools at other institutions. One such tool designed by CHICA-Canada served as a basis for the development of the team’s audit tools [5]. Data analysis The team analyzed their observations by determining the percentage of best practice each observed site is following. From Infection Control’s policy on endoscope reprocessing, best practices were determined for each step of reprocessing. The team constructed an observation matrix as a tool to compare each visited site to Infection Control’s policy. Deviations from the policy were noted on the main page of the matrix document, as can been seen in Appendix B. The team also created individual tracking sheets for each of the visited sites that further track compliance with the policy. These tools revealed the most common sources of error to the team as well as which sites were farthest from meeting the standards set by Infection Control. Within each site tracking sheet, the team recorded observations for each visit. By using a binary scale, zero meaning an element of the policy that was not complied with, the team determined a site’s percent compliance by step in the reprocessing process and by trial. Use of these results yielded an overall percent compliance that was used as a comparison between sites. Findings From the team’s analysis of the observations and surveys, the team gathered the following findings. Policy Breakdown The team obtained a thorough understanding of Infection Control’s policy for reprocessing flexible endoscopes. These findings were used to create a structured approach to the observation data collection. The observation data collection sheet can be found in Appendix A. The policy can be broken up into five different sections: Setup, Pre-Cleaning, Reprocessing, Transportation/Storage, and miscellaneous. Setup Before a procedure takes place, scope setup should occur. The following steps are characteristic of this section: Obtaining scope (either from within room or cabinets outside rooms) Plugging in scope Connecting suction and drainage Checking the scope function (white balance, suction, flushing) Pre-Cleaning Immediately following a procedure pre-cleaning must take place. Skipping this step can cause organic tissue to dry and become fixated to the scope, causing a scope to be incorrectly reprocessed [6]. Pre-cleaning is made up of these steps: 1. Wiping scope exterior 2. Flushing channels with detergent or water Reprocessing The largest portion of time for endoscope reprocessing occurs in the reprocessing room. This room is separated from sites of patient care. Within the reprocessing room these steps occur: 1. Conducting a leak test 2. Suctioning detergent 3. Brushing all channels (even those not used during the procedure) 4. Attaching adapters 5. Flushing all channels with an enzymatic detergent solution 6. Immersing the scope in detergent solution 7. Wiping/washing scope with pad 8. Brushing ports and buttons with soft brushes 9. Purging all water and drying scope exterior 10. Disinfecting a. Manual i. Suctioning/injecting disinfectant into all channels ii. Immersing scope in disinfectant and cover iii. Soaking a minimum of 20 minutes using a timing device iv. Removing scope and purging disinfectant from all channels v. Rinsing with sterile, filtered, or tap water (in specified clean sink) vi. Purging water with forced air vii. Flushing channels with alcohol (30-60 cc) viii. Drying with 2 minutes forced air ix. Drying exterior with clean cloth x. Attaching green visual cue b. Automatic Endoscope Reprocesser (AER) i. Checking chemicals in machine to ensure correct pH ii. Entering patient and scope information into machine and/or record on log iii. Putting endoscope in machine with correct connections iv. Placing buttons and brushes in machine v. Running cycle vi. Injecting 30-60 cc alcohol after final rinse vii. Attaching green visual cue Transportation/Storage Scopes must be transported from room to room and stored according to the policy to prevent damage and the risk of infection. When scopes are transported they should be clearly indicated as “dirty-do not use” or “HLD-ready for patient use.” Following a procedure scopes should be transported to the reprocessing room in a covered tote or bin. After the dirty scope is removed from the bin, a disinfectant cleaner should be used to clean the container. Endoscopes should be stored in dry, clean, and well-ventilated cabinets. Scopes should not touch the bottom of the cabinets, however if this is unavoidable due to the length of the scope then the scope tip must be protected and the bottom of the cabinet must be regularly cleaned or be covered with a clean towel or pad. Miscellaneous Personal protective equipment, brush use, and chemical considerations are crucial to proper endoscope reprocessing. It is imperative that anyone that comes in contact with a scope be aware of the proper PPE requirements for handling it. Reprocessing technicians within the reprocessing room should wear a face shield or goggles and mask at all times during the process. Gowns are required both in the reprocessing and procedure room. Employees should be conscious of when changing gloves and/or hand washing is required. Brushes should be disinfected after each use (between scopes). Chemicals (Metricide) used in the AERs should be changed every 14 days, but strips should be used to check their pH once a day. Institutional Comparisons The team researched endoscope reprocessing practices from other institutions to compare UMHS’ policy to others and gain a deeper understanding of the expectations for this process. The team specifically looked at best practices of the Community and Hospital Infection Control Association (CHICA) – Canada, as resources were easily accessible. A variety of errors that can cause an endoscope to be improperly reprocessed including: policy breaches, issues with chemicals, improper drying, using damaged endoscopes, reusing rinse water, using contaminated germicide, and lack of scrubbing to kill biofilms [7]. The student team was able to use this knowledge to perform more pointed observations that focused on uncovering the mentioned issues within UMHS sites. This source stressed the importance of each part of the endoscope reprocessing policy; failing to comply with any part creates a risk of infection. Accreditation Canada’s updated Qmentum Infection Prevention and Control Standards [6] states: • The importance of reviewing the competencies of staff involved in the process of endoscope reprocessing • All endoscope reprocessing areas should be physically separate from client care areas and should be cleaned daily • Endoscope reprocessing areas should possess separate and clean areas for storage (cleaned/disinfected weekly) and should have dedicated plumbing and adequate air ventilation • Qualified staff members should reprocess endoscopes immediately after a procedure o Failing to immediately reprocess may allow bio material to harden to the scope and be resistant to the remaining reprocessing steps • Leak tests are required to ensure scopes are functioning properly • Manual cleaning is required before disinfection can occur o Accepted cleaning agent must be used o Correct concentration of cleaning agent to solution must be used for immersing endoscope o Correct brushes must be used and must be either disposable or cleaned between each scope o If brushes are not changed frequently enough, ineffective cleaning or channel damage can occur [8] • Scopes must be stored vertically, not coiled in a case • Endoscope reprocessing must be logged and kept o Includes: scope identification number, type of scope, AED identifier, date/time of reprocessing, client identifier, leak test results, and name of person responsible for reprocessing Compliance by Site Figure 1 shows the results of the analysis of the matrix. Sites that see a high scope volume, like MPU and MPC, appear to have higher percent compliance, while sites that do not regularly reprocess endoscopes, such as the clinics, have low percent compliance; the overall percent compliance. The Taubman Center Urology Clinic was not included in this analysis due to insufficient data. The CVC clinic was not included in this analysis because the collected observation targeted TEE probe reprocessing, which is not identical to flexible endoscope reprocessing. Figure 1. Overall percent pompliance to Infection Control’s endoscope reprocessing policy by site Dates collected: 9/25/13 – 11/13/13 Sample size located in Table 1 The compliance ratings can be broken down by reprocessing category. The categories are: setup, pre-cleaning, transportation, reprocessing, storage, and miscellaneous. Setup includes all steps that occur within the procedure room before the scope is used by the patient. Pre-cleaning looks at steps that occur in the procedure room immediately following the procedure. Transportation specifically looks at how scopes are transported from the procedure room to the reprocessing room. Reprocessing includes all steps that occur within the reprocessing room. Storage looks at how scopes are stored as well as how the green visual cues are used. The miscellaneous category includes PPE, brush use and reprocessing, and chemical maintenance. Figure 2 shows the overall percent compliance by reprocessing category. Specifically, transportation, miscellaneous, and set up should be the primary focus areas for improvement as they had the three lowest overall percent compliance across observed sites. Check about Oto** Figure 2. Overall percent compliance to Infection Control’s endoscope reprocessing policy by category Dates collected: 9/25/13 – 11/13/13 Sample size by category: Setup, n =15; Pre-cleaning, n = 13; Transportation, n = 15; Reprocessing, n = 16; Storage, n = 15; Miscellaneous, n = 16 Table 2 shows overall percent compliance for each visited site and the breakdown by category. Overall percent and percent by category are also included. Table 2 shows that even sites with high overall percent compliance have problem areas that need to be addressed. Overall, an 85% compliance was obtained by UMHS across the sites observed. Table 2. Percent compliance to Infection Control’s policy across visited sites Cancer MPU MPC Mott-OR UH-OR Oto Center Overall Overall 91% 91% 86% 97% 73% 94% 89% Set Up 91% 55% 100% 100% 100% 75% 87% Pre-Cleaning 100% 100% 100% 50% 100% 100% 92% Transportation 100% 75% 100% 100% 25% 75% 79% Reprocessing 92% 100% 84% 100% 88% 100% 94% Storage 100% 100% 100% 100% 100% 100% 100% Miscellaneous 70% 100% 83% 100% 50% 100% 84% Similarities Across Sites From the surveys, the team found that most sites are flushing their scopes at setup, as well as removing their green zip tie and white balancing the cameras, as seen in Table 3. However, only 61% are wiping down the scope as part of the cleaning that happens in the procedure room before transporting to reprocessing. And only 76% are doing a precautionary flush of water at setup. Table 3. Procedure Room Tasks Dates collected: 11/20/13 – 12/3/13, Sample size=33 technicians, 9 sites Practice Setup Flush scope with water 76% Wipe down scope with alcohol/enzyme soaked pads 3% Flush with aseptizyme 0% Label scope as dirty 12% Remove green zip tie 91% White balance camera on scope 91% PreNot cleaning done 48% 0% 61% 39% 94% 3% 73% 15% 6% 0% 3% 0% It is necessary for the endoscopes to be carried into the room in a bin if the endoscope is coming from a cabinet outside of the procedure room in which it will be used. 18% of sites are not conforming to this practice, as shown in Figure 2. Figure 2. Endoscope retrieval for procedure: “How is the scope brought into procedure room?” Dates collected: 11/20/13 – 12/3/13 Sample size=33 technicians, 9 sites Figure 3 below shows that most sites are transporting their dirty endoscopes into the reprocessing room in a bin, as they should. 27% of sites are not covering these bins. Figure 3. Endoscope transportation to reprocessing room: “How is the scope carried to the reprocessing room?” Dates collected: 11/20/13 – 12/3/13 Sample size=33 technicians, 9 sites Table 4. Reprocessing Room Best Practice Summary Dates collected: 11/20/13 – 12/3/13 Sample size=21 technicians, 9 sites Best Practice Scope labeled dirty Fresh water and detergent for each scope Disinfect brushes Disinfect brushes after every scope % follow 76% 43% 100% 14% In the reprocessing room, all reported sites are reprocessing their brushes; however, only 14% are high level disinfecting the brushes after every scope, which is the requirement. 71% of sites are only reprocessing their brushes daily. Conclusions Based on the team’s findings, problem areas were identified. A problem area is defined as a category in the endoscope reprocessing process that does not achieve a 100% compliance rating. Problem areas discussed are both those unique to certain sites and those needing attention across UMHS. These areas will provide the basis for the team’s conclusions. Common Areas for Improvement Based on the findings from the team’s observations, comparison matrix, and survey results, the following problem areas were identified, as all had overall percent compliance ratings less than 100%. Setup Pre-cleaning Transportation Reprocessing Miscellaneous Within these problem areas it was determined common deviations from best practice include: • Lack of flushing during scope setup • Failure to change water/detergent mix between scopes • Procedures for reprocessing buttons • Lack of proper brush reprocessing • Use of bins for transportation • Labeling of bins as dirty during transportation • Proper PPE/glove use Sites that do not reprocess endoscopes on a regular basis are of particular concern as the team observed large deviations from Infection Control’s policy throughout endoscope reprocessing. Both observation and survey data revealed common deviations between sites and Infection Control’s policy. Site-based Areas for Improvement - Mott: when scopes are set up and doctor wants different one, must reprocess scope even though they know is clean (WASTE) UH OR: scope sits in room after use until after procedure without flushing, maybe allow organic material to dry and harden MPU: maybe use comment from survey about “clean” bags/bins MPU: clean rack is easier to access than dirty rack, potential hazard Recommendations Each step in the endoscope reprocessing process is critical to affectively reprocessing a scope. Failure to comply with even one practice creates a risk of infection and a potentially dirty scope. Target Areas Specific problem areas were identified within the conclusions. These areas will be looked at in more detail below. Setup Many sites were performing the set up and take down (or pre-cleaning) of the scope differently. Set up itself does not have a documented policy for how it should be done, so it is being compared to the best practices the team saw at the majority of sites visited. The team recommends that all sites perform a flush test as standard practice for scope setup as the team is aware that this practice has caught potential hazards in the past. Pre-cleaning Immediately following a procedure the scope’s exterior should be wiped down with a cloth or sponge soaked in a fresh detergent solution [2]. All channels should then be flushed with detergent solution or water to prevent organic material from hardening to the endoscope. Failure to comply with this practice will allow the material to harden to the scope and prevent the endoscope from being properly reprocessed. Transportation Scopes should be transported into and out of both the procedure room and reprocessing room via a covered and clearly labeled bin [9]. As shown by survey data, this area varies greatly across sites and even within sites. To target this issue, pointed questions are used in the audit whenever the scope is moved from room to room. The team recommends that different colored bins be used for clean and dirty scopes as an extra safety check. This would create an easy and very visual way to distinguish between dirty and reprocessed endoscopes. Reprocessing Room Leak tests should be done immediately when the scopes are being reprocessed to ensure they are working properly before being sent to the procedure room. Each channel in the endoscope also must be suctioned and brushed with detergent, even if they were not used, because fluid or debris may be present in these channels [8]. Endoscopes must be soaked for 20 minutes in disinfectant to decrease the risk of transferring diseases. Soaking a scope for longer than 20 minutes can damage the endoscope. High level disinfection of buttons was a practice that varied across sites as well. Buttons should be high level disinfected along with the scope. AER machines are equip to handle the smaller accessories associated with scopes, including buttons. If a site reprocesses endoscopes manually, buttons should be soaked in disinfectant for 20 minutes. Miscellaneous A variety of current practices exist for the use of brushes to clean endoscopes. The team observed four different practices. The policy states that brushes must be disinfected between scopes [9], but the team found this to not be the case for seven out of eight sites. The audit asks specific questions about the type of brushes used (disposable vs. reused) as well as the reprocessing procedure for reusable brushes to reveal where the policy is not being followed. Audit Tools The team’s data collection and analysis greatly assisted in creating both an external and internal audit for endoscope reprocessing. Both audits focus on following a scope throughout its lifecycle. By analyzing both the observation and survey data, the team created two audit tools that target observed problem areas and known inconsistencies with Infection Control’s policy. The external audit should be performed by Infection Control at each site on an annual basis. Use of this tool will reveal potential risks in the endoscope reprocessing process at each site. Topics will cover all steps contained in Infection Control’s policy. Additionally, setup procedures, training, documentation of reprocessing steps, and other topics will be included. The internal audit was designed to be used by managers within each site. This tool should be used on a quarterly basis and should make sure to include all employees responsible for any portion of endoscope reprocessing. The internal audit is a simplified version of the external audit to allow for easy use and a short time commitment for managers. Expected Impact The team expects that the audit tool developed will assist Infection Control in evaluating the reprocessing capability at each of the 32 locations in the University of Michigan Health System. The audit tool will show areas in the reprocessing of endoscopes that could be improved, and this information can be used to help prevent future errors and citations from occurring. The audit tools developed target common errors seen by the team during observations, which should better uncover errors at future locations. References [1] L. Sturm, private communication. September 2013. [2] M.E. Day, M. Juan, et al. “Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes.” Society of Gastroenterology Nurses and Associates, Inc.: Chicago, Illinois. Report. 2012. [3] American Society for Gastrointestinal Endoscopy, “Multi-society Guideline on Reprocessing Flexible Gastrointestinal Endoscopes,” Gastrointestinal Endoscopy, vol. 73, no. 6, pp. 1075-1084, 2011. [4] Infection Control Committee. “Endoscopy: Reprocessing and High-Level Disinfection.” University of Michigan Hospitals and Health Centers: Ann Arbor, Michigan. Report. 15 Apr. 2013. [5] CHICA-Canada, “Infection Prevention and Control Audit for Medical Device Reprocessing in Ambulatory Clinic or Physician’s Office,” Audit Toolkit Version 2, September 2009. [6] Accreditation Canada, Qmentum Infection Prevention and Control Standards, CHICA-Canada, 2013 March 22. [Online]. Available: http://www.chica.org/pdf/13May02_Accred_Stds.pdf. [Accessed: 23 November, 2013]. [7] Truscott, Biofilms in Medicine: Patients Threatened by Highly Organized Militant Pathogens, CHICA-Canada. [Online]. Available: http://www.chica.org/CHICANS/2012_Truscott.pdf. [Accessed: 23 November, 2013]. [8] Bradley Catalone, Ph.D., and George Koos, Reprocessing Flexible Endoscopes. Olympus America, June 2005. [Online]. Available: http://www.olympusamerica.com/msg_section/files/mic0605p74.pdf. [Accessed: 1 November, 2013]. [9] M. Bachman. “Endoscope and Ancillary Equipment Reprocessing Competency.” University of Michigan Health System, Medical Procedures Center: Ann Arbor, Michigan. Audit. 4 Oct. 2013. Appendix A – Data collection tool Process Step Setup Pre-cleaning (after procedure) Wipe? Suction? Flush/rinse? Container placed in? Reprocessing Inspect/leak test? Suction detergent? Brush? Attach adapters and flush? Immerse in detergent? Wash with pad? Rinse? Clean valves? Purge water and wipe dry? Either connect to machine or manual Manual: Suction disinfectant Totally immerse in disinfectant Soak min 20 min? After done: flush with water? Flush with alcohol? Force air (2 min)? Dry external with clean cloth? Handle with clean hands? Attach visual cue? Transport? Notes Appendix B – Matrices and survey results, by site Appendix B.1 – Medical Procedures Units Matrix Summary Notes – MPU Steps Setup Best Practice Using clean hands or gloves, attach scope and irrigation Remove green tag, and put on buttons Do flush test White balance Coil scope and place to be ready for doctor Pre-cleaning 1. Wipe exterior surface 2. Flush channels with detergent OR water Transportation Needs to be in a bin to transport, with gloves. Bin must be clearly indicated that scope is dirty. Put bin on "dirty rack" Reprocessing 1. Leak test after each use 2. Suction detergent 3. Brush all channels 4. Attach adapters 5. Flush all channels 6. Immerse scope in detergent in sink 7. Wipe/wash scope with pad 8. Rinse with tap water 9. Brush ports and buttons with soft brushes 10. Purge all water and dry exterior 11. Disinfecting 11a. Manual -Suction/inject disinfectant into all channels -Totally immerse scope in disinfectant and cover -Soak minimum 20 minutes using a timing device -Remove scope and purge disinfectant from all channels -Rinse with sterile, filtered, or tap water (in a sink specified for clean scopes) -Purge water with forced air -Flush channels with alcohol (syringe of 30-60 cc) MPU: 9/25/13, 10/4/13, 10/17/13 Carried scope in from another room not in a bin (10/23) Didn't flush second time observing (maybe due to lack of time --are techs aware of importance of set up procedures?) Flushed with both aseptizyme AND water - suctioned both, put in high pressure flush button, and held button for 30 sec, then released button and flushed for 30 sec (10/23) 5 AERs, 4 soaking bins Sometimes uses same disinfectant for two scopes (not at same time) -Dry with 2 minutes forced air -Dry exterior with clean cloth -Handle with clean hands and attach visual cue 11b. AER -Check chemicals in machine -Enter patient and scope information and/or record on log -Put endoscope in machine -Wait for two endoscopes to one machine before connecting -Attach hoses, put buttons (and brushes) in machine -Run cycle -After final rinse, inject 30-60 cc alcohol Attach Green Custom Tags Transportation /Storage Storage areas should be labeled with "High Level Disinfected: Ready for Patient Use" on a rack specifically for clean endoscopes. Green Custom Tags still attached. Miscellaneous (PPE, brushes, buttons) Brushes should be disposable or high level disinfected. Buttons must be high level disinfected Metricide needs to be changed every 14 days, check with strips once/day. Easier to get to clean rack than dirty rack. Clean rack is outside room Bin not wiped down before next use Only throw away brushes when the bristles get bad (patchy). Only HLD brushes at the end of the day Weren't wearing face shield and/or goggles and mask (10/21) NOTE: Operators hate green tags. Matrix Individual Tracking Sheet – MPU Trial Steps Setup Pre-cleaning Transportation to reprocessing room Reprocessing Transportation /Storage Miscellaneous (PPE, brushes) Using clean hands/gloves to handle scope? Remove green tag Flush test White balance Coil scope Wipe exterior surface Flush channels with detergent OR water Scope in bin Wearing gloves? Bin marked "dirty"? Put bin on "dirty rack" Leak test Correct concentration chemical? Suction detergent brush all channels immerse scope in detergent in sink wipe/wash scope with pad rinse with tap water brush ports and buttons with soft brushes purge all water and dry exterior Disinfection (AER) check chemicals in machine enter patient info in machine and/or record in log buttons in machine? after cycle, 30-60 cc alcohol attach green tag Change water b/n scopes Separate clean/dirty sink Storage areas labeled "High Level Disinfected: Ready for Patient Use" Green tags still attached Scopes hung in cabinets Brushes disposable OR HLD between scopes % Compliance 1 2 3 4 5 1 1 1 1 1 1 0 0 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 0 1 0 0 0 1 0 1 0 1 1 1 1 1 1 1 1 1 100% 1 1 1 1 0 0 0 0 0 70% 91.3% 100% 100% 92% Wearing correct PPE (gown, facemask & goggles OR face shield, bouffant cap) in reprocessing room clean hands/gloves when needed Metricide changed every 14 days, check with strips once a day % Compliance 1 1 0 1 1 1 1 1 1 1 1 1 1 1 1 94% 87% 88% 94% 94% 91% Procedure Room Survey Results – MPU 3) For each task listed, check whether a task is either done at setup, after the procedure (pre-cleaning), both, or not at all (or not applicable). # Question 1 Flush scope with water Wipe down scope with 2 alcohol/enzyme soaked pads 3 Flush with aseptizyme 4 Label scope as dirty 5 Remove green zip tie 6 White balance camera on scope PreNot Total Setup cleaning done N/A Responses 12 9 0 0 21 0 0 4 13 13 14 14 12 1 1 4) How is the scope brought into procedure room/OR? # Answer 1 In an unlabeled bin 2 Carried in with hands 3 In a bin labeled "clean" 4 Gotten from room - no transportation Total Response % 4 29% 5 36% 0 0% 5 36% 14 100% 5) How is the scope carried to the reprocessing room? # Answer 1 In an unlabeled, uncovered bin 2 In an uncovered bin labeled "dirty" 3 Carried in with hands 4 In an unlabeled, covered bin 5 In a covered bin labeled "dirty" Total Response % 0 0% 7 50% 1 7% 0 0% 6 43% 14 100% 0 0 0 0 0 0 0 0 0 0 14 14 16 14 14 93 6) Does the scope have a green HLD zip tie attached to it when it is brought into the procedure room/OR? If no, why? # Answer 1 Yes 2 No Total Response % 13 93% 1 7% 14 100% Reprocessing Room Survey Results – MPU 3) Is scope labeled as dirty? # Answer 1 Yes 2 No Total 4) Is water/detergent reused for each scope? # Answer 1 Yes 2 No Total 5) Are brushes disposable? # Answer 1 Yes 2 No Total 6) If brushes are not disposable, are they disinfected? # Answer 1 Yes 2 No Total 7) How often are Response % 1 33% 2 67% 3 100% Response % 1 33% 2 67% 3 100% Response % 3 100% 0 0% 3 100% Response % 3 100% 0 0% 3 100% brushes disinfected? # Answer 1 Never 2 Daily 3 After each scope 4 Weekly 5 N/A Total Response % 0 0% 2 67% 0 0% 0 0% 1 33% 3 100% 8) For each task listed, indicate whether it is always, often, sometimes, rarely, or never done in the reprocessing room (or if not applicable, check "N/A") # Question Always Often Sometimes Rarely Never N/A Total 1 Leak test 3 0 0 0 0 0 3 Wipe down scope 2 exterior 3 0 0 0 0 0 3 3 Brush channels 3 0 0 0 0 0 3 4 Brush buttons 3 0 0 0 0 0 3 5 Flush channels 3 0 0 0 0 0 3 6 Suction aseptizyme 3 0 0 0 0 0 3 7 Rinse with water 3 0 0 0 0 0 3 8 Purge water 3 0 0 0 0 0 3 9 Dry exterior 2 0 0 1 0 0 3 10 Manual disinfection 2 0 0 0 1 0 3 Automatic 11 disinfection (AER) 3 0 0 0 0 0 3 12 Clean sink 3 0 0 0 0 0 3 Flush channels with 13 alcohol 3 0 0 0 0 0 3 14 Dry with forced air 3 0 0 0 0 0 3 Attach custom 15 green zip tie 3 0 0 0 0 0 3 Put scope in 16 unlabeled bin 2 0 0 0 1 0 3 Put scope in bin 17 labeled "clean" 1 0 0 0 2 0 3 51 Appendix B.2 – Medical Procedures Center Steps Setup Best Practice East Ann Arbor: 10/11/13, 11/1/13, 11/8/13, 11/13/13 Using clean hands or gloves, attach scope and irrigation Remove green tag, and put on buttons Do flush test White balance Coil scope and place to be ready for doctor Observed scope being hand carried to procedure room Not sure if they did flush test Upon second visit, scope was already set up upon arrival. No cup of water next on work station indicated flush test may not have been done. Flush test only observed 1/5 times 1. Wipe exterior surface 2. Flush channels with detergent OR water Needs to be in a bin to transport, with gloves. Bin must be clearly indicated that scope is dirty. Put bin on "dirty rack" Wiped with alcohol pads. Flushed using just water. This is where key card was attached. Bin taken to reprocessing room, places on unlabeled rack next to dirty sink Pre-cleaning Transportation Reprocessing 1. Leak test after each use 2. Suction detergent 3. Brush all channels 4. Attach adapters 5. Flush all channels 6. Immerse scope in detergent in sink 7. Wipe/wash scope with pad 8. Rinse with tap water 9. Brush ports and buttons with soft brushes 10. Purge all water and dry exterior 11. Disinfecting 11a. Manual -Suction/inject disinfectant into all channels -Totally immerse scope in disinfectant and cover -Soak minimum 20 minutes using a timing device -Remove scope and purge disinfectant from all channels -Rinse with sterile, filtered, or tap water (in a sink specified for clean scopes) -Purge water with forced air -Flush channels with alcohol (syringe of 30-60 cc) -Dry with 2 minutes forced air -Dry exterior with clean cloth -Handle with clean hands and attach visual cue 11b. AER -Check chemicals in machine -Enter patient and scope information and/or record on log -Put endoscope in machine -Wait for two endoscopes to one machine before connecting -Attach hoses, put buttons (and brushes) in machine -Run cycle -After final rinse, inject 30-60 cc alcohol Attach Green Custom Tags Transportation /Storage No indication how long scopes sit in bins in REPO room before being hung. Operator says people like to pull from this location because the scope is already in a bin. People will pull from different location if different type of scope or attachment is needed. Storage areas should be labeled with "High Level Disinfected: Ready for Patient Use" on a rack specifically for clean endoscopes. Green Custom Tags still attached. Miscellaneous (PPE, brushes, buttons) Cheaper green tags being used one week after custom tag launch. Brushes should be disposable or high level disinfected. Buttons must be high level disinfected NOTE: At end of weekday, scopes are partially cleaned and not hung up, except Fridays, where each scope is HLD and hung up at the end of the day. NOTE: Operators hate green tags Metricide needs to be changed every 14 days, check with strips once/day. Trial 1 2 3 4 % 5 Compliance Using clean hands/gloves to handle scope? 1 1 1 0 1 Remove green tag Flush test 0 0 0 1 0 Steps Setup 55% White balance 0 0 0 1 0 1 1 1 1 1 1 1 1 Flush channels with detergent OR water Transportation Scope in bin Wearing gloves? to reprocessing Bin marked "dirty"? room Put bin on "dirty rack" Leak test 1 1 1 1 0 1 1 1 1 1 0 1 1 1 1 1 0 1 1 1 1 0 1 1 1 1 0 1 Correct concentration chemical? Suction detergent brush all channels 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 Coil scope Wipe exterior surface 100% Pre-cleaning immerse scope in detergent in sink wipe/wash scope with pad rinse with tap water brush ports and buttons with soft brushes Reprocessing purge all water and dry exterior Disinfection (AER) check chemicals in machine enter patient info in machine and/or record in log buttons in machine? after cycle, 30-60 cc alcohol attach green tag Change water b/n scopes Separate clean/dirty sink? Storage areas labeled Transportation "High Level Disinfected: /Storage Ready for Patient Use" Green tags still attached 75% 100% 100% Scopes hung in cabinets Miscellaneous (PPE, brushes) Brushes disposable OR HLD between scopes Wearing correct PPE (gown, facemask & goggles OR face shield, bouffant cap) in reprocessing room clean hands/gloves when needed Metricide changed every 14 days, check with strips once a day % Compliance 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 90% 90% 90% 93% 90% 100% 91% Appendix B.3 – Mott OR Steps Setup Pre-cleaning Transportation Reprocessing Best Practice Using clean hands or gloves, attach scope and irrigation Remove green tag, and put on buttons Do flush test White balance Coil scope and place to be ready for doctor 1. Wipe exterior surface 2. Flush channels with detergent OR water Needs to be in a bin to transport, with gloves. Bin must be clearly indicated that scope is dirty. Put bin on "dirty rack" Mott OR: 11/4/13, 1. Leak test after each use 2. Suction detergent 3. Brush all channels 4. Attach adapters 5. Flush all channels 6. Immerse scope in detergent in sink 7. Wipe/wash scope with pad 8. Rinse with tap water 9. Brush ports and buttons with soft brushes 10. Purge all water and dry exterior 11. Disinfecting 1st Observation: Use the same gloves throughout shift. Use the same detergent and "clean" water until it looks dirty probably change 4-5 times a day 2nd Observation: Detergent solution changed between scopes/tools Very aware of when to change gloves (questioning if "double gloves" idea is okay) Brought up in "Soiled" elevator in a red "dirty" bin 11a. Manual -Suction/inject disinfectant into all channels -Totally immerse scope in disinfectant and cover -Soak minimum 20 minutes using a timing device -Remove scope and purge disinfectant from all channels -Rinse with sterile, filtered, or tap water (in a sink specified for clean scopes) -Purge water with forced air -Flush channels with alcohol (syringe of 30-60 cc) -Dry with 2 minutes forced air -Dry exterior with clean cloth -Handle with clean hands and attach visual cue 11b. AER -Check chemicals in machine -Enter patient and scope information and/or record on log -Put endoscope in machine -Wait for two endoscopes to one machine before connecting -Attach hoses, put buttons (and brushes) in machine -Run cycle -After final rinse, inject 30-60 cc alcohol Attach Green Custom Tags Transportation /Storage Storage areas should be labeled with "High Level Disinfected: Ready for Patient Use" on a rack specifically for clean endoscopes. Green Custom Tags still attached. Miscellaneous (PPE, brushes, buttons) Brushes should be disposable or high level disinfected. Buttons must be high level disinfected Metricide needs to be changed every 14 days, check with strips once/day. 1st Observation: did not check chemical levels in machine (policy for Mott says check every time?) 2nd Observation: did not inject alcohol, went straight to air 1st Observation: Take scope over to "clean" side with dirty gloves. Window between rooms is open 2nd Observation: Clean gloves to transport scope to window. Window was closed New brush at the beginning of the day. Supposed to run through washer but not confident that happens Brushes accumulate in bin until full enough to put in AER New brushes/rag for each scope Says check Metricide every scope, but observed skipping this step with one scope Trial 2 % 3 Compliance 1 1 1 1 1 1 Coil scope Wipe exterior surface 1 1 1 1 1 1 Flush channels with detergent OR water Scope in bin Wearing gloves? Bin marked "dirty"? Put bin on "dirty rack" Leak test 1 1 1 1 1 1 1 1 1 1 1 1 1 Correct concentration chemical? Suction detergent brush all channels 0 1 1 0 1 1 0 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 0 1 1 1 1 1 1 1 1 1 0 0 1 Steps 1 Using clean hands/gloves to handle scope? Setup Pre-cleaning Transportation to reprocessing room Remove green tag Flush test White balance immerse scope in detergent in sink wipe/wash scope with pad rinse with tap water brush ports and buttons with soft brushes Reprocessing purge all water and dry exterior Disinfection (AER) check chemicals in machine enter patient info in machine and/or record in log buttons in machine? after cycle, 30-60 cc alcohol attach green tag Change water b/n scopes 1 100% 100% 100% 84% Separate clean/dirty sink? Transportation /Storage Miscellaneous (PPE, brushes) 0 1 1 1 1 1 1 1 1 100% 0 1 1 83% 1 1 1 0 1 1 1 1 1 Storage areas labeled "High Level Disinfected: Ready for Patient Use" Green tags still attached Scopes hung in cabinets Brushes disposable OR HLD between scopes Wearing correct PPE (gown, facemask & goggles OR face shield, bouffant cap) in reprocessing room clean hands/gloves when needed Metricide changed every 14 days, check with strips once a day % Compliance 68% 94% 97% 86% 3) For each task listed, check whether a task is either done at setup, after the procedure (pre-cleaning), both, or not at all (or not applicable). PreNot Total # Question Setup cleaning done N/A Responses Flush scope with 1 water 4 2 0 0 6 Wipe down scope with alcohol/enzyme 2 soaked pads 0 3 1 0 4 Flush with 3 aseptizyme 0 4 0 0 4 Label scope as 4 dirty 0 3 1 0 4 Remove green zip 5 tie 4 0 0 0 4 White balance 6 camera on scope 3 0 0 1 4 26 4) How is the scope brought into procedure room/OR? # Answer Response % 1 2 3 4 In an unlabeled bin Carried in with hands In a bin labeled "clean" Gotten from room - no transportation Total 0 0% 1 25% 0 0% 3 4 75% 100% 5) How is the scope carried to the reprocessing room? # Answer Response % In an unlabeled, 1 uncovered bin 0 0% In an uncovered 2 bin labeled "dirty" 1 25% Carried in with 3 hands 0 0% In an unlabeled, 4 covered bin 0 0% In a covered bin 5 labeled "dirty" 3 75% Total 4 100% 6) Does the scope have a green HLD zip tie attached to it when it is brought into the procedure room/OR? If no, why? # Answer Response % 1 Yes 4 100% 2 No 0 0% Total 4 100% Appendix B.4 – UH-OR Steps Best Practice Setup Using clean hands or gloves, attach scope and irrigation Remove green tag, and put on buttons Do flush test White balance Coil scope and place to be ready for doctor UH OR: 10/15/13 Pre-cleaning Transportation 1. Wipe exterior surface 2. Flush channels with detergent OR water Needs to be in a bin to transport, with gloves. Bin must be clearly indicated that scope is dirty. Put bin on "dirty rack" Channels are not flushed until AFTER the procedure has been completed and cart is taken to reprocessing room 1. Leak test after each use 2. Suction detergent 3. Brush all channels 4. Attach adapters 5. Flush all channels 6. Immerse scope in detergent in sink 7. Wipe/wash scope with pad 8. Rinse with tap water 9. Brush ports and buttons with soft brushes 10. Purge all water and dry exterior 11. Disinfecting Reprocessing 11a. Manual -Suction/inject disinfectant into all channels -Totally immerse scope in disinfectant and cover -Soak minimum 20 minutes using a timing device -Remove scope and purge disinfectant from all channels -Rinse with sterile, filtered, or tap water (in a sink specified for clean scopes) -Purge water with forced air -Flush channels with alcohol (syringe of 30-60 cc) -Dry with 2 minutes forced air -Dry exterior with clean cloth -Handle with clean hands and attach visual cue Window between dirty and clean room open 11b. AER -Check chemicals in machine -Enter patient and scope information and/or record on log -Put endoscope in machine -Wait for two endoscopes to one machine before connecting -Attach hoses, put buttons (and brushes) in machine -Run cycle -After final rinse, inject 30-60 cc alcohol Attach Green Custom Tags Transportation /Storage Miscellaneous (PPE, brushes, buttons) Storage areas should be labeled with "High Level Disinfected: Ready for Patient Use" on a rack specifically for clean endoscopes. Green Custom Tags still attached. Brushes should be disposable or high level disinfected. Buttons must be high level disinfected Metricide needs to be changed every 14 days, check with strips once/day. Trial 1 % Compliance Steps Setup Buttons are not HLD, just passed through washer, BUT are sterilized Using clean hands/gloves to handle scope? Remove green tag Flush test White balance Coil scope Wipe exterior surface 1 1 1 1 1 1 100% 50% Pre-cleaning Flush channels with detergent OR water Transportation Scope in bin Wearing gloves? to reprocessing Bin marked "dirty"? room Put bin on "dirty rack" Leak test 0 1 1 1 1 1 Correct concentration chemical? Suction detergent 1 1 Reprocessing 100% 100% brush all channels immerse scope in detergent in sink wipe/wash scope with pad rinse with tap water brush ports and buttons with soft brushes purge all water and dry exterior Disinfection (AER) check chemicals in machine enter patient info in machine and/or record in log buttons in machine? after cycle, 30-60 cc alcohol attach green tag Separate clean/dirty sink? Change water b/n scopes 1 1 1 1 1 1 1 1 1 1 1 1 Storage areas labeled "High Level Disinfected: Transportation Ready for Patient Use" /Storage Green tags still attached Scopes hung in cabinets 1 1 1 100% Miscellaneous (PPE, brushes) 1 100% Brushes disposable OR HLD between scopes Wearing correct PPE (gown, facemask & goggles OR face shield, bouffant cap) in reprocessing room clean hands/gloves when needed Metricide changed every 14 days, check with strips once a day % Compliance 1 1 1 97% 97% Appendix B.5 – Taubman Otolaryngology Clinic Appendix B.6 – The Center Center Urology Clinic Appendix B.7 – Taubman Urology Clinic Appendix B.8 – CVC Appendix B.9 – Livonia Center for Specialty Care Appendix B.10 – Livonia Center for Specialty Care Urology Appendix B.11 – CW Peds Oto Appendix B.12 – OBGYN REE BW01 Appendix B.13 – Mott OBGYN 3) For each task listed, check whether a task is either done at setup, after the procedure (pre-cleaning), both, or not at all (or not applicable). PreNot Total # Question Setup cleaning done N/A Responses Flush scope with 1 water 0 1 0 0 Wipe down scope with alcohol/enzyme 2 soaked pads 0 1 0 0 3 Flush with aseptizyme 0 1 0 0 4 Label scope as dirty 0 1 0 0 5 Remove green zip tie 0 1 0 0 6 White balance camera 1 0 0 0 1 1 1 1 1 1 on scope 6 4) How is the scope brought into procedure room/OR? # Answer Response % 1 In an unlabeled bin 0 0% 2 Carried in with hands 0 0% 3 In a bin labeled "clean" 1 100% Gotten from room - no 4 transportation 0 0% Total 1 100% 5) How is the scope carried to the reprocessing room? # Answer Response % In an unlabeled, 1 uncovered bin 0 0% In an uncovered bin 2 labeled "dirty" 0 0% 3 Carried in with hands 0 0% In an unlabeled, 4 covered bin 0 0% In a covered bin 5 labeled "dirty" 1 100% Total 1 100% 6) Does the scope have a green HLD zip tie attached to it when it is brought into the procedure room/OR? If no, why? # Answer Response % 1 Yes 1 100% 2 No 0 0% Total 1 100% 3) Is scope labeled as dirty? # Answer 1 Yes 2 No Total Response % 1 100% 0 0% 1 100% 4) Is water/detergent reused for each scope? # Answer 1 Yes 2 No Total 5) Are brushes disposable? # Answer 1 Yes 2 No Total Response % 0 1 1 Response 0% 100% 100% % 0 1 1 0% 100% 100% 6) If brushes are not disposable, are they disinfected? # Answer Response % 1 Yes 1 100% 2 No 0 0% Total 1 100% 7) How often are brushes disinfected? # Answer 1 Never 2 Daily 3 After each scope 4 Weekly 5 N/A Total Response 8) For each task listed, indicate whether it is always, often, sometimes, rarely, or never done in the reprocessing room (or if not applicable, check "N/A") # Question Always 1 Leak test Wipe down scope 2 exterior 3 Brush channels 4 Brush buttons 5 Flush channels Suction 6 aseptizyme % 0 0 1 0 0 1 0% 0% 100% 0% 0% 100% Often Sometimes Rarely Never N/A Total 1 0 0 0 0 0 1 1 1 0 1 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 1 0 1 1 1 1 1 0 0 0 0 0 1 7 Rinse with water 8 Purge water 9 Dry exterior Manual 10 disinfection Automatic 11 disinfection (AER) 12 Clean sink Flush channels 13 with alcohol Dry with forced 14 air Attach custom 15 green zip tie Put scope in 16 unlabeled bin Put scope in bin 17 labeled "clean" Appendix C – Audit Tools External Audit Tool Appendix C continued External Audit Tool 1 1 1 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 1 1 1 1 0 0 0 0 0 1 0 0 0 0 0 0 0 0 0 0 1 1 1 1 1 0 0 0 0 0 1 1 0 0 0 0 0 1 1 0 0 0 0 0 1 1 0 0 0 0 0 1 1 0 0 0 0 0 1