Endoscope Reprocessing - TSICP Texas Society of Infection Control

advertisement

Prevention of Infection Due to Endoscopy

William A. Rutala, Ph.D., M.P.H.

University of North Carolina (UNC) Health Care

System and UNC at Chapel Hill

Disclosure

This educational activity is brought to you, in part, by

Advanced Sterilization Products (ASP) and Ethicon. The speaker receives an honorarium from ASP and Ethicon and must present information in compliance with FDA requirements applicable to ASP.

Endoscope Reprocessing

Lecture Goals

Background

Infections related to endoscopy

Reprocessing of endoscopes and accessories

Cleaning

High-level disinfection/sterilization

Automated endoscope reprocessing

Quality control

GI ENDOSCOPES

Widely used diagnostic and therapeutic procedure

Endoscope contamination during use (10 9 in/10 5 out)

Semicritical items require high-level disinfection minimally

Inappropriate cleaning and disinfection has lead to crosstransmission

In the inanimate environment, although the incidence remains very low, endoscopes represent a risk of disease transmission

TRANSMISSION OF INFECTION

Gastrointestinal endoscopy

>300 infections transmitted

70% agents Salmonella sp. and P. aeruginosa

Clinical spectrum ranged from colonization to death (~4%)

Bronchoscopy

90 infections transmitted

M. tuberculosis, atypical Mycobacteria, P. aeruginosa

Spach DH et al Ann Intern Med 1993: 118:117-128 and Weber DJ, Rutala WA Gastroint Dis 2002

Nosocomial Infections via GI Endoscopes

Observations

Number of reported infections is small, suggesting a very low incidence

Endemic transmission may go unrecognized (e.g., inadequate surveillance, low frequency, asymptomatic infections)

Spach DH. Ann Int Med 1993;118:117 and Weber DJ, Rutala, WA. Gastroint Dis 2002

Nosocomial Infections via GI Endoscopes

Infections traced to deficient practices

Inadequate cleaning (clean all channels)

Inappropriate/ineffective disinfection (time exposure, perfuse channels, test concentration, ineffective disinfectant, inappropriate disinfectant)

Failure to follow recommended disinfection practices (tapwater rinse)

Flaws is design of endoscopes or AERs

Endoscope Reprocessing: Current Status of

Cleaning and Disinfection

Guidelines

Multi-Society Guideline, 11 professional organizations, 2003

Society of Gastroenterology Nurses and Associates, 2000

European Society of Gastrointestinal Endoscopy, 2000

British Society of Gastroenterology Endoscopy, 1998

Gastroenterological Society of Australia, 1999

Gastroenterological Nurses Society of Australia, 1999

American Society for Gastrointestinal Endoscopy, 1996

Association for Professional in Infection Control and Epidemiology, 2000

Centers for Disease Control and Prevention, 2008 (in press)

Endoscope Reprocessing, Worldwide

Worldwide, endoscopy reprocessing varies greatly

India, of 133 endoscopy centers, only 1/3 performed even a minimum disinfection (1% glut for 2 min)

Brazil, “a high standard …occur only exceptionally”

Western Europe, >30% did not adequately disinfect

Japan, found “exceedingly poor” disinfection protocols

US, 25% of endoscopes revealed >100,000 bacteria

Schembre DB. Gastroint Endoscopy 2000;10:215

Endoscopes

ENDOSCOPE DISINFECTION

CLEAN-mechanically cleaned with water and enzymatic cleaner

HLD/STERILIZE-immerse scope and perfuse

HLD/sterilant through all channels for exposure time

RINSE-scope and channels rinsed with sterile water, filtered water, or tap water followed by alcohol

DRY-use forced air to dry insertion tube and channels

STORE-prevent recontamination

ENDOSCOPE REPROCESSING

Source of contamination for infections (36 outbreaks) transmitted by GI endoscopes from 1974-2001:

Cleaning-3 (12%)

Disinfection-19 (73%)

Rinse, Dry, Store-3 (12%)

Etiology unknown-11

ENDOSCOPE DISINFECTION

Cleaning (results in dramatic decrease in bioburden, 4-5 log

10 reduction)

No brushing biopsy channel. (Schousboe M. NZ Med J

1980;92:275)

No precleaning before AER. (Hawkey PM. J Hosp Inf

1981;2:373)

Biopsy-suction channel not cleaned with a brush.

(Bronowicki JP. NEJM 1997;337:237)

Bacterial Bioburden Associated with Endoscopes

After procedure

After cleaning

Gastroscope, log

10

CFU

6.7

6.8

2.0

4.8

2.3

4.3

5.1

Colonoscope, log

10

CFU

8.5

Gastro Nursing 1998;22:63

8.5

Am J Inf Cont 1999;27:392

9.8

Gastro Endosc 1997;48:137

Viral Bioburden from Endoscopes Used with AIDS Patients

Hanson et al. Lancet 1989;2:86; Hanson et al. Thorax 1991;46:410

Dirty

Gastroscopes

HIV (PCR) 7/20

HBsAg 1/20

Bronchoscopes

HIV (cDNA) 7/7

HBsAg 1/10

Cleaned

0/20

0/20

0/7

0/10

Disinfected

0/20

0/7

0/7

0/10

ENDOSCOPE REPROCESSING

Precleaning

After removal from patient, wipe the insertion tube with a wet cloth and alternate suctioning the enzymatic cleaner and air through the biopsy/suction channel until solution clean. The airwater channel is flushed or blown out per instructions.

Transport the endoscope to the reprocessing area.

Enyzmatic cleaner should be prepared per instructions. Some data suggest enzymes are more effective cleaners than detergents. Enyzmatic cleaners must be changed after use.

ENDOSCOPE REPROCESSING

Cleaning

Immerse in a compatible low-sudsing, enzymatic cleaner

Wash all debris from exterior by brushing and wiping

Remove all removal parts of the endoscope and clean each reusable part separately

After exterior cleaning, brush accessible channels with appropriate-sized cleaning brush

ENDOSCOPE REPROCESSING

Cleaning (continued)

After each passage, rinse the brush, remove debris before reinserting.

Continue until no visible debris on brush.

Attach cleaning adapters for each channel per manufacturer’s instructions and flush with enzymatic cleaner to remove debris.

After cleaning is complete, rinse the endoscope with clean water.

Purge water from channels using forced air. Dry exterior of the endoscope with a soft, lint-free cloth.

ENDOSCOPE DISINFECTION

CLEAN-mechanically cleaned with water and enzymatic detergent

HLD/STERILIZE-immerse scope and perfuse

HLD/sterilant through all channels for exposure time

RINSE-scope and channels rinsed with sterile water, filtered water, or tap water followed by alcohol

DRY-use forced air to dry insertion tube and channels

STORE-prevent recontamination

ENDOSCOPE REPROCESSING

Source of contaminations for infections (36 outbreaks) transmitted by GI endoscopes from 1974-2001:

Cleaning-3 (12%)

Disinfection-19 (73%)

Rinse, Dry, Store-3 (12%)

Etiology unknown-11

ENDOSCOPE REPROCESSING

Unacceptable Disinfectants for HLD

Benzalkonium chloride

Iodophor

Hexachlorophene

Alcohol

Chlorhexidine gluconate

Cetrimide

Quaternary ammonium compounds

Glutaraldehyde (0.13%) with phenol

ENDOSCOPE REPROCESSING

Inappropriate disinfectants

Benzalkonium chloride (Greene WH. Gastroenterol 1974;67:912)

70% alcohol (Elson CO. Gastroenterol 1975;69:507)

QUAT (Tuffnell PG. Canad J Publ Health 1976;67:141)

Hexachlorophene (Dean AG. Lancet 1977;2:134)

Hexachlorophene (Beecham HJ. JAMA 1979;1013)

70% alcohol (Parker HW. Gastro Endos 1979;25;102)

Povidone-iodine (Low DE. Arch Intern Med 1980;1076)

Cetrimonium bromide. (Schliessler KH. Lancet 1980;2:1246)

ENDOSCOPE REPROCESSING

Inappropriate disinfectants

3% hexachlorophene.

(Schousboe M. NZ Med J 1980;92:275)

0.5% CHG in alcohol, 0.015% CHG and 0.15% cetrimide; 87 s exposure to 2% glut.

(Hawkey PM. J Hosp Inf 1981;2:373)

1% Savlon (cetrimide and CHG)

.(O’Connor BH. Lancet 1982;2:864)

0.0075% iodophor.

(Dwyer DM. Gastroint Endosc 1987;33:84)

0.13% glut with phenol.

(Classen DC. Am J Med 1988;84:590)

70% ethanol for 3 min.

(Langenberg W. J Inf Dis 1990;161:507)

ENDOSCOPE REPROCESSING

Inappropriate disinfection

Air/water channel not exposed to glut. (Birnie GG. Gut

1983;24:171)

Air/water channel not exposed to glut. (Cryan EMJ. J Hosp Inf

1984;5:371)

No glut (water only) between patients. (Earnshaw JJ. J Hosp Inf

1985;6:95)

High Level Disinfection of

“Semicritical Objects”

Exposure Time > 12 m-30m (US), 20 o C

Germicide Concentration_____

Glutaraldehyde > 2.0%

Ortho-phthalaldehyde (12 m) 0.55%

Hydrogen peroxide* 7.5%

Hydrogen peroxide and peracetic acid* 1.0%/0.08%

Hydrogen peroxide and peracetic acid* 7.5%/0.23%

Hypochlorite (free chlorine)* 650-675 ppm

Glut and phenol/phenate** 1.21%/1.93%___

*

May cause cosmetic and functional damage; **efficacy not verified

New FDA-Cleared Sterilants/HLD

“Older”

> 2% Glut, 7.5% HP, 1.0% HP and 0.08% PA

Newer

0.55% ortho-phthalaldehyde (HLD- 5 m worldwide, 12 m in US)

0.95% glut and 1.64% phenol/phenate (HLD-20 m at 25 o C)

7.5% HP and 0.23% PA (HLD-15 m)

2.5% Glut (HLD-5 m at 35 o C)

Ensure antimicrobial activity and material compatibility

Ideal HLD/Chemical Sterilant

Rapid HLD (< 10 min)

No disinfectant residue after rinsing

Excellent material compatibility

Long shelf-life

Nontoxic (no odor or irritation issues)

No disposal problems

Monitor minimum effective concentration

Glutaraldehyde

Advantages

Numerous use studies published

Relatively inexpensive

Excellent materials compatibility

Disadvantages

Respiratory irritation from vapor (ACGIH 0.05 ppm)

Pungent and irritating odor

Relatively slow mycobactericidal activity

Coagulate blood and fix tissues to surfaces

Ortho-phthalaldehyde

Advantages

Fast acting HLD

No activation

Excellent materials compatibility

Not a known irritant to eyes and nasal passages

Weak odor

Disadvantages

Stains protein gray

Cost ($30/gal);but lower reprocessing costs-soak time, devices per gal)

Slow sporicidal activity

Eye irritation with contact

Exposure may result in hypersensitivity

Comparison of Glutaraldehyde and OPA

>2.0% Glutaraldehyde

HLD: 45 min at 25 o C

Needs activator

14 day use life

2 year shelf life

ACGIH ceiling limit, 0.05ppm

Strong odor

MEC, 1.5%

Cost - $12/gallon

0.55% Ortho-phthalaldehyde

HLD: 12 min at 20 o C

No activator needed

14 day use life

2 year shelf life

No ACGIH or OSHA limit

Weak odor

MEC, 0.3%

Cost - $30/gallon

OPA Research

Alfa and Sitter, 1994. OPA eliminated all microorganisms from 100 different endoscopes used in a clinical setting.

Gregory et al, 1999. OPA achieved a 6 log

10 reduction of

M. bovis in 5.5 min compared to 32 min for glutaraldehyde

Walsh et al, 1999. OPA effective against glutaraldehyderesistant M. chelonae strains

OPA Label Claims Worldwide

1. Europe, Asia, Latin America

5 min at 20 o C

2. Canada and Australia

10 min at 20 o C

3. United States

12 min at 20 o C

1. Antimicrobial tests support 5 min exposure time.

2. Canadian regulatory authority requires 6-log reduction in mycobacteria (5.5 m) and only 5 min intervals.

3. FDA requires 6-log reduction of mycobacteria suspended in organics and dried onto scope without cleaning

Ortho-phthalaldehyde (OPA)

Contraindication for OPA

Repeated exposure to OPA, following manual reprocessing of urological instruments, may have resulted in hypersensitivity in some patients with a history of bladder cancer undergoing repeated cystoscopy.

Out of approximately 1 million urological procedures, there have been reports of 24 patients who have experience ‘anaphylaxis-like’ reactions after repeated cystoscopy (typically after 4-9 treatments).

Risk control measures: residues of OPA minimized; and contraindicated for reprocessing of urological instruments used on patients with history of bladder cancer.

Peracetic Acid/Hydrogen Peroxide

Advantages

No activation required

No odor or irritation issues

Effective in the presence of organic matter

Disadvantages

Material compatibility issues for lead, brass, copper, zinc (both cosmetic and functional damage for 1% HP with 0.08% PA)

Limited clinical use

Minimum Effective Concentration (MEC)

High Level Disinfectant (HLD)

Dilution of HLD occurs during use

Test strips are available for monitoring MEC

For example, test strips for glutaraldehyde monitor 1.5%

Test strip not used to extend the use-life beyond the expiration date (date test strips when opened)

Testing frequency based on how frequently the solutions are used (used daily, test at least daily)

Record results

Endoscope Reprocessing

Disinfectant/Sterilant

Immerse the endoscope in HLD/sterilant (at least 12-20 minutes) and fill the channels with HLD/sterilant until no air bubbles are seen

Reusable endoscopic accessories that break the mucosal barrier (e.g., biopsy forceps) should be mechanically cleaned as described above and then sterilized between each patient use.

Rinsing

Rinse all surfaces and channels and removable parts with clean water to remove disinfectant. Inadequate rinsing of HLD has caused colitis.

Drying and Alcohol Flush

Purge channels with air; flush with alcohol; purge with air; dry

Storeprevent recontamination

ENDOSCOPE REPROCESSING

Rinse, Dry, Store

Irrigating water bottle. (Doherty DE. Dig Dis Sci

1982;27:169)

Inadequate drying (no alcohol). (Allen JI. Gastroenterol

1987;92:759)

Inadequate drying (no alcohol). (Classen DC. Am J Med

1988;84:590)

Nosocomial Outbreaks via GI Endoscopes

Infections Associated with Accessories

Biopsy forceps

Contaminated biopsy forceps. (Dwyer DM. Gastroint Endosc

1987;33:84)

Contaminated biopsy forceps (no cleaning between cases).

Graham DY. Am J Gastroenterol 1988;83:974)

Biopsy forceps not sterilized (glut exposed,? time) Bronowicki

JP. NEJM 1997;334:237)

ENDOSCOPE REPROCESSING

Manual/AER HLD

High level disinfection is the standard of care for reprocessing GI endoscopes and bronchoscopes

The process can be completed manually or with an AER

Until recently no automated endoscope reprocessor (AER) substitutes for manual cleaning

For manual disinfection, immerse completely in HLD and fill each channel with the HLD

Cover the basin to prevent vaporization and use timer

Flush channels with air before removing the scope from HLD

Automated Endoscope Reprocessors (AERs)

Advantages: automate and standardize reprocessing steps, reduce personnel exposure to chemicals, filtered tap water

Disadvantages: failure of AERs linked to outbreaks, does not eliminate precleaning, does not monitor HLD concentration

Problems: incompatible AER (side-viewing duodenoscope); biofilm buildup; contaminated AER; inadequate channel connectors

MMWR 1999;48:557. Used wrong set-up or connector

Must ensure exposure of internal surfaces with HLD/sterilant

EVOTECH w/Cleaning Claim

Product Definition:

Integrated double-bay AER

Eliminates manual cleaning

Uses New High-Level Disinfectant (HLD) with IP protection

Single-shot HLD

Automated testing of endoscope channels and minimum effective concentration of HLD

Incorporates additional features (LAN, LCD display)

Reliance ™ DG

Klenzyme

®

, CIP

®

200

Reliance ™ EPS

Endoscope Processing System

Endoscope Processing

Support

Reliance™ PI

Automatic Endoscope Reprocessors

EvoTech-integrates cleaning (FDA-cleared claim) and disinfection.

Automated cleaning comparable to manual cleaning. All residual data for cleaning of the internal channels as well as external insertion tube surfaces were below the limit of <8.5ug/cm 2

Reliance -requires a minimal number of connections to the endoscope channels and uses a control boot (housing apparatus the creates pressure differentials to ensure connectorless fluid flow through all channels that are accessible through the endoscope’s control handle channel ports). Data demonstrate that the soil and microbial removal effected by Reliance washing phase was equivalent to that achieved by optimal manual cleaning. Alfa, Olson, DeGagne. AJIC 2006;34:561.

HLD versus Sterilization for Endoscopes

Sterilization (S) or High-Level Disinfection (HLD)

Burns and colleagues compared S versus HLD with glut for arthroscopes and laparoscopes and found no difference (7.5/1000 procedures for S vs. 2.5/1000 procedures for HLD)

Burns et al Infect Control Hosp Epidemiol 1996; 17: suppl p42

Fuselier and Mason examined S (with peracetic acid) and glut and found no clinical difference (no clinical data). S about 10 more costly than HLD.

Fuselier and Mason Urology 1997; 50:337

Thus, no data S is superior to HLD

ENDOSCOPE REPROCESSING

Staff Safety

Personal Protective Equipment

Gloves

Eye protection

Impervious gown

Personnel who use chemicals should be educated about the biologic and chemical hazards present while performing procedures that use disinfectants

Reprocessing Room

Area designated for this function with: adequate space, proper airflow and ventilation (7-15 ACH), work flow patterns

Disinfection of Emerging Pathogens

Disinfection and Sterilization of

Emerging Pathogens

Hepatitis C virus

Clostridium difficile

Cryptosporidium

Helicobacter pylori

E.coli 0157:H7

SARS coronavirus

Antibiotic-resistant microbes (MDR-TB, VRE, MRSA)

Creutzfeldt-Jakob disease (no brain, eye, spinal cord contact)

Disinfection and Sterilization of

Emerging Pathogens

Standard disinfection and sterilization procedures for patient care equipment are adequate to sterilize or disinfect instruments or devices contaminated with blood and other body fluids from persons infected with emerging pathogens

Clostridium difficile

Disinfectants and Antiseptics

C. difficile spores at 10 and 20 min, Rutala et al, 2006

~4 log

10 reduction (3 C. difficile strains including BI-9 )

Clorox, 1:10, ~6,000 ppm chlorine (but not 1:50, ~1,200 ppm)

Clorox Clean-up, ~1,910 ppm chlorine

Tilex, ~25,000 ppm chlorine

Steris 20 sterilant, 0.35% peracetic acid

Cidex, 2.4% glutaraldehyde

Cidex-OPA, 0.55% OPA

Wavicide, 2.65% glutaraldehyde

Aldahol, 3.4% glutaraldehyde and 26% alcohol

Control Measures

C. difficile

Handwashing (soap and water) , contact precautions, and meticulous environmental cleaning (disinfect all surfaces) with an EPA-registered disinfectant should be effective in preventing the spread of the organism.

McFarland et al. NEJM 1989;320:204.

In units with high endemic C. difficile infection rates or in an outbreak setting, use dilute solutions of 5.25-6.15% sodium hypochlorite (e.g., 1:10 dilution of bleach) for routine disinfection. (Category II)

For semicritical equipment, glutaraldehyde (20m), OPA (12m) and peracetic acid (12m) reliably kills C. difficile spores using normal exposure times

ENDOSCOPE SAFETY

Quality Control

Ensure protocols equivalent to guidelines from professional organizations (APIC, SGNA, ASGE)

Are the staff who reprocess the endoscope specifically trained in that job?

Are the staff competency tested at least annually?

Conduct IC rounds to ensure compliance with policy

Consider microbiologic sampling of the endoscope

Conclusions

Endoscopes represent a nosocomial hazard

Proper cleaning and disinfection will prevent nosocomial transmission

Current guidelines should be strictly followed

Compliance must be monitored

Safety and efficacy of new technologies must be validated

Prevention of Infection Due to Endoscopy

Background

Infections related to endoscopy

Processing of endoscopes and accessories

Cleaning

High-level disinfection/sterilization

Automated endoscope reprocessing

Quality control

Thank you

References

Rutala WA, Weber DJ. Disinfection of endoscopes: Review of new chemical sterilants for high-level disinfection. Infect Control Hosp Epidemiol

1999;20:69-76.

Rutala WA, Weber DJ. Creutzfeldt-Jakob Disease: Recommendations for disinfection and sterilization. Clin Inf Dis 2001;32:1348-1356.

Society of Gastroenterology Nurses and Associates. Standards. 2000.

Weber DJ, WA Rutala, AJ DiMarino. Prevention of infection following gastrointestinal endoscopy. Gastro Dis. 2002;87-107

Rutala WA, Weber DJ, HICPAC. Disinfection and sterilization in healthcare facilities. MMWR. In press.

Download