The NC2100 soil analyzer is suited to measure major constituents such as carbon and nitrogen (also sulfur, hydrogen, and oxygen when properly configured). It does this by combusting a carefully weighed sample in a small tin foil wrapper. The combustion occurs in a stream of pure oxygen and helium. The resulting gases pass through a set of columns, which convert (oxidize/reduce) the gases into a single gas (e.g. CO2, N2, SO2). The gas stream is then passed into a chromatography column (See Instrument Layout below) and eventually arrives at a TCD (thermal conductivity detector). The peak areas are measured to get a sensitive analysis of the element in the original sample. Because of different responses for the different elements, a calibration must first be performed with at least one standard that contains the element(s) of interest. Usually this calibration is good for a number of samples, but quality control requires that a known sample or standard be run periodically. Known samples, which have a very similar composition (relative to your unknowns) will be the best check of proper instrument operation (e.g. Soil reference material, Peach Leaves, etc.). Elemental Analyzer (EA2100) 1. Before running: Check Instrument Maintenance via Eager software. If maintenance is required follow separate directions for such. Check uV level within the Eager window: this should be about 1000 uV. If it isn’t then there is probably a leak somewhere. If it has to be adjusted this is done via removing the panel on the right side of the instrument and adjusting the dial at the bottom. Left click on the Maintenance Program icon: Left click Edit; select Set Maintenance; Select Default. This will show you how many samples have been run: 2. Check anhydrous column (central module, small Mg Perch column). a. If wet it will look dull and clumpy. If the flow drops below 120 ml/min (measured with an external flow meter as necessary) the Mg Perch is probably wet. If there is any sign of wetness it must be changed or the chromatography column will be ruined! b. Add glass wool to the bottom of a clean column to cover the threads. Add Mg Perch up to the top threads then more glass wool. The Teflon section of the o-ring goes next to the glass. 3. Check for leaks: a. Turn machine on via the toggle at the back right of the machine. b. Take machine off standby (SPC FUN -> STBY -> YES) c. Helium tank should be turned on and flow set to 120 ml/min (SET UP-> “(arrow right)” -> SET FLOW. d. Screw on the cap to the carrier gas outlet (bottom of central module). If the flow doesn’t drop to below 5 ml/min then there is a leak somewhere in the system. Follow separate instructions for Leak Test. 4. Check purge: a. Disconnect at autosampler and connect tubing to external flow meter attached at the instrument top. This should read 60-90 ml/min (= 60-90 SCCM). b. If 60-90 ml/min flow rate is not achieved, open the central module by sliding the cover backwards (use some force to release the magnetic hold). c. Turn the needle valve counterclockwise if low, clockwise if high. 5. If all checks pass, set oven temperature to: a. LEFT (catalyst): 1000 C b. RIGHT (copper): 780 C c. GC (oven): 50 C d. Air: >62 psi (300-400 kPa needed to move the autosampler). NOTE: The current tank needs to run at 80 psi! e. Helium: 40 psi f. Oxygen: 40 psi 6. Record all problems, repairs, and maintenance actions (beyond the normal maintenance of the crucible, reaction columns, and the adsorption trap) in the CN Logbook. Record the gas levels and the number of samples being run. 7. Run a bunch of system blanks (no tin cup nor sample) through the instrument until the baseline is less than 0.2 mV differential at the nitrogen peak (about 65-68 seconds retention time). Program Files -> Thermo -> Eager XPerience for USB -> Methods -> Systemblanks.mth SAMPLE PREPARATION 1. Homogenizing the Sample a. A first coarse homogenizing on sample amounts of a few hundred grams is followed by finer homogenizing on a few dozen grams, until optimum granulometry (100-200 um) is reached. b. The resulting sample is dried in an oven: 105C 2. Weighing the sample a. First ensure that the work area and tools have been thoroughly cleaned with ethanol. Keep the door to Ag 48 closed to limit air flow that affects the balance. b. Place one tin in the balance and tare. c. Remove the tin from the balance and place it into one of the depressions in the wrapping block. d. Using a spatula place approximately 20 mg sample into the tin. Reweigh to double-check that you have that amount. e. Hold the tin above the sample (about half-way) with both forceps. The forceps should be holding the tin no more than ¼ of the way from the outside. f. Perform a “Z-fold”: move one hand slightly backwards while the other hand either stays stationary or moves slightly forward. You should see a “z” in the fold. Now push the forceps together to tighten the z-fold. g. Carefully squish the tin together above the sample using the forceps. Be extremely careful not to puncture the tin. This can be done by holding with one pair of forceps and squishing with the other. h. From the top of the tin carefully make one roll downwards to start closing the opening. Make about 3 rolls toward the sample. i. Hold the tin with one pair of forceps and, using the cylindrical tool, gently press down on the sample to remove all air. j. The sample should look like a cubical box. If it is too long and skinny it will not drop correctly into the autosampler. Pick up the tin and drop slightly onto the wrapping block to ensure that there are no spills. k. Reweigh and record the weight in milligrams. Place in the appropriate well in a sample plate. Sample Table: (A sample table is a “Method”) 1. RECALCULATION -> RESET CALIBRATION FACTORS 2. Clear sample table: EDIT -> SAMPLE TABLE -> EDIT SAMPLE -> CLEAR SAMPLE TABLE 3. Fill sample table (*.mth file; parameters are an elemental analyzer method or *.eam file) a. EDIT -> SAMPLE TABLE -> EDIT SAMPLE -> FILL SAMPLE TABLE b. Number of samples: use 200 (maximum amount) as your number of samples so that you can add later as needed. c. “Add” d. The system will run until it sees a blank line in the sample table. e. DON’T DELETE LINES!! Insert the blank as EDIT -> INSERT EMPTY LINE. f. SAMPLE 1 = System blank (no tin cup). g. BLANK = tin cup with no standard or sample i. Blank average (Nitrogen) should be close to zero (possibly as high as 4000). ii. Carbon blank should be <15000 area counts. h. BYPASS = a standard that is not weighed. It sets the instrument’s peak identification before other standards and samples are run. The retention times within the predicted range can be adjusted before running the calibration. i. STANDARD = chemical standard such as EDTA. This must be no more than 6 months old. Run 5 standards at the beginning of each test. j. CHECK STANDARD = a standard run as an unknown to check the calibration. Run at least one CHECK STANDARD every 10-20 samples to test the calibration. i. ii. iii. iv. v. 4. 5. 6. 7. 8. Setup Instructions for STD-CHECK Enter the component table. Enter a minimum and or maximum value for check standard (absolute %). A dialogue box will open. If you only want to apply minimum and maximum values to check standards then check first box on top of page otherwise will be applied to all unknown samples. vi. Determine action to be taken when check standard is out of range: 1. None (Set flag for custom report) or 2. Stop sequence or 3. Set instrument to standby. SAVE method before you run anything! Do not save over the original. Make sure you press “Enter” or move to a new cell before you close out of the sample table otherwise the last item entered will not be registered. RECALCULATION -> RESET CALIBRATION FACTORS START ANALYSIS VIEW SAMPLE BEING ACQUIRED a. STOP analysis: stops acquiring and saves up to that point b. ABORT analysis: stops acquiring and does not save SUMMARIZE RESULTS a. If something is wrong with the standard it won’t be on the curve. (View -> View Calibration Curve) i. Right click on a point to get information. ii. Perform a K-Fit analysis. Write down the numbers in the log book. iii. Carbon blank should be <15000 area counts; nitrogen blank average should be close to zero. b. CHECK STANDARDS – standards run as unknowns. These should read at +/- 0.2-0.5% of the tested values. c. Soil unknowns should read at +/- 10%. d. If standards, bypass, blanks, and checks look OK then load your samples. 9. ADD SAMPLES to sample table: a. This cannot be accomplished via a transfer from Excel. b. Rename any “sample name” and add the corresponding weights. c. “Enter” after entering a weight OR move off the cell. d. Add one blank line after every 6 samples. Add 2-3 blank lines and standard checks after 24 samples. e. If you have more than 200 samples/standards/blanks/checks you will need to save the excess as another sample table. f. SAVE METHOD 10. RUN/START analysis 11. NOTE: “Last Sample Calculated Results” and “Acquiring” windows are the only two that can be left open so as not to interfere with the sample acquisition. 12. NOTE: Sample Run is at the autosampler marked position when at the beginning of the run. At about 60 sec the autosampler advances to the next sample and holds it on top of the internal door ready to drop after the chromatogram is done. The instrument seems to run better when the NEXT sample after a blank line is a system blank. 13. NOTE: If an “error generating report” occurs, advance the sample table by one line manually and save method. If this is not done the next sample will rewrite over the previous one. 14. NOTE: “Stop Analysis” stops the sample acquisition and saves up to that point. “Abort Analysis” stop sample acquisition and does not save. 15. NOTE: With the K-factor analysis the carbon blank should be less than 15000 area counts. The blank averages should really be close to zero. 16. Summarize Results: if something is wrong with the standard it won’t be on the curve. Right click on the point on the curve to get information. a. In the summary table you can group like samples together by assigning the like sample a group number. (This is done if you have duplicate or triplicate of the same sample so you could see the elemental average, percent RSD, and Variance of the like samples.) b. To display the data for a specific group, place the cursor on the group and select the drop-down screen you are looking for the element that you want averaged. c. To print your data, select the “Print” drop down menu and select “Print All Groups”. This will bring up the print screen and select to “Print By Peaks”. 17. Recalculation: Save after integration. If there is a problem with a peak not saving after being added, go through the recalculation but for that one chromatogram do NOT reintegrate. 18. SHUT DOWN: a. Flip toggle switch on back of instrument. b. Turn off all gases. c. Turn off computer.