Limit Comparison Date Completed: ___5/5/2013______________ Performed By: ____Sam Remp______________ Specifications Tested Engr. Spec. # Unit of Measure Specification (description) The program developed must be able to log and compare input data to a standard. 6.1 Marginal Value +/-0.2, V/µs +/-0.05 Revision History Revision 1 Description Date Document created Equipment __X__ Fully assembled test fixture (with motors and tooth phantoms) __X__ Arduino Uno board and EasyDrivers __X__ Front Panel GUI __X__ DPR35E Pulser/Receiver Unit __X__ 10 MHz Ultrasound Transducer __X__ Tektronix TDS2012C Oscilloscope Sections Part I Data collection Part II Data comparison Comments/Status Standard is based on whether the delta taken of any two input data matches, and the tolerances to which they match. Limit Comparison Date Completed: ___5/5/2013______________ Performed By: ____Sam Remp______________ Part I Data collection __X__ 1. Ensure that the DPR35E unit is properly connected. __X__ 2. Ensure that the test fixture is properly assembled with all five motors. __X__ 3. Connect the motors to the proper pins on the EasyDrivers and Arduino Uno. __X__ 4. Ensure that the Arduino Uno has been configured for Labview integration, following the instructions shown in Configuring Test Station. __X__ 5. Turn on the DPR35E unit and Tektronix TDS2012C oscilloscope. __X__ 6. Open the following file from SVN: //source code/DPR35/front panel.vi __X__ 7. Ensure that the VISA resource name reflects the appropriate COM port. __X__ 8. Place a drop of coupling agent on the tooth phantom to be tested. __X__ 9. Flip the switch on the front panel VI to “ON” and run the program. __X__ 10. Export and save sampled data to a dedicated folder. __X__ 11. Repeat steps 9 to 11 for other teeth. Summarize data below Tooth Socket 0 Data Collection (Pass/Fail) Successfully able to determine materials and log data to file? PASS Sign off on section completion before continuing:______________________________ Limit Comparison Date Completed: ___5/5/2013______________ Performed By: ____Sam Remp______________ Part II Data comparison __X__ 1. Record the number of the tooth socket being tested. __X__ 2. Ensure that the DPR35E unit is properly connected. __X__ 3. Ensure that the test fixture is properly assembled with all five motors. __X__ 4. Connect the motors to the proper pins on the EasyDrivers and Arduino Uno. __X__ 5. Ensure that the Arduino Uno has been configured for Labview integration, following the instructions shown in Configuring Test Station. __X__ 6. Turn on the DPR35E unit and Tektronix TDS2012C oscilloscope. __X__ 7. Open the following file from SVN: __X__ 8. Ensure that the VISA resource name reflects the appropriate COM port. __X__ 9. Place a drop of coupling agent on the tooth phantom to be tested. __X__ 10. Flip the switch on the front panel VI to “ON” and run the program. __X__ 11. Export and save sampled data to a dedicated folder. __X__ 12. Repeat steps 9 to 11 for one more test. __X__ 13. Compare data obtained and determine if within comparison tolerances. __X__ 14. Repeat steps 9 to 13 for other teeth. Summarize data below Tooth Socket 0 35 Test Runs Data Comparison (Successfully Identifies Materials) Success Each Time Sign off on section completion before continuing:______________________________