Biological evaluation of medical devices 2Part 3: Tests for 10993Biocomp 03/16/2012 AAMI ANSI ISO 117 genotoxicity, 3:2003/(R)2009 carcinogenicity, and reproductive toxicity Recognition List Number: 028 Publication Date: 03/16/2012 Part B: SUPPLEMENTARY INFORMATION Recognition Number 2-117: AAMI / ANSI / ISO 10993-3:2003/(R)2009, Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity, and reproductive toxicity. (Biocompatibility) Date of Standard: 2003. Addresses of Standards Organizations: Association for the Advancement of Medical Instrumentation (AAMI) 4301 North Fairfax Drive Suite 301 Arlington, VA 22203 American National Standards Institute (ANSI) 25 West 43rd Street 4th Floor New York, NY 10036 International Organization for Standardization (ISO)* 1, Rue de Varembe Case Postale 56 CH 1211 Geneva 20, 0 SWITZERLAND CDRH Office and Division Associated with Recognized Standards: OFFICE OF DEVICE EVALUATION (ODE) Devices Affected: All permanent surface devices that contact mucosal membrane and breached or compromised surfaces Permanent external communicating devices that contact blood path indirectly Prolonged and permanent external communicating devices that contact tissue/bone/dentin All external communicating devices that contact the circulating blood All prolonged and permanent contact implant devices Processes Affected: 510(k), PMA, IDE, PDP, HDE Type of Standard: Horizontal Extent of Recognition: The genetic toxicity test battery for medical devices should include at least three assays: a test for gene mutation in bacteria, an in vitro test with cytogenetic evaluation of chromosomal damage with mammalian cells or an in vitro mouse lymphoma tk assay, and an in vivo test for chromosomal damage using rodent hematopoietic cells. Due to the diversity of medical devices, it is recognized that not all tests identified will be necessary or practical for any given device. It is indispensable for testing that each device be considered on its own merits: additional tests not indicated in the table may be necessary. Additionally, tests are suggestions and other comparable tests can be substituted for suggested tests with accompanying scientific justification. For devices with polymeric component(s), the test material should be prepared using an exaggerated sample extraction method. The standard recommends only two assays. Additionally, an assessment of risk in the standard includes only a partial list that should be considered. FDA recommends including the following factors when determining if carcinogenicity testing is warranted: 1. total duration of exposure, 2. exposure route, 3. extent of localized (at the site of implantation) and systemic exposure, 4. patient population, 5. device's indication for use, 6. cause of concern (e.g., structure-activity relationship, previous demonstration in the product class), 7. device's physical characteristics (e.g, device size, pore size, surface continuity, surface condition, device thickness), 8. metabolites of the device material 9. biologic components, and 10. genotoxicity results Complete standard, except as follows: Clause 2 - The OECD references should be modified to reflect medical device conditions for example, maximum tolerated dose testing may not be attainable or practical for device extracts. Clause 4.1 Exclude the following: ?Testing for genotoxicity, however, is not necessary for medical devices, and components thereof, made only from materials known to show no genotoxicity. Testing for genotoxicity is indicated where a review of the composition of the materials reveals the possible presence in the final medical device of compounds that might interact with genetic material, or when the chemical composition of the medical device is unknown. In such circumstances, the genotoxic potential of suspect chemical components should be assessed, bearing in mind the potential for synergy, in preference to carrying out genotoxicity tests on the material or medical device as a whole. When the genotoxicity of a medical device has to be experimentally assessed, a series of in vitro tests shall be used. This series shall include either two tests if 4.2.1.2 is performed, which uses the mouse lymphoma assay incorporating colony number and size determination, or three tests if 4.2.1.1 is performed. When tests are performed, at least two tests, investigating different endpoints, shall use mammalian cells.? Clause 4.2.2 Exclude this clause Clause 4.2.4 Exclude this clause Clause 4.2.5 Exclude this clause Clause 4.3.1 - For devices with cumulative contact duration greater than 30 days and involve the testing of polymeric or porous device materials, the extract (when applicable) should be prepared using an exhaustive method. Clause 4.3.2 - In addition to the finished device (ready-to-use state), testing may be warranted for additional device states such as, wear debris generated from implanted device or intermediate reaction states for in situ polymerizing devices. Clause 4.3.3 Exclude this clause Clause 5.1 Carcinogenicity testing may be recommended for new materials and old materials with new intended uses, patient populations or treatment sites. Clause 5.2.1 FDA interprets significant and adequate human-use history to mean the medical device, which undergoes a similar manufacturing process, is used to treat a similar patient population, with a similar intended use at a similar treatment site. Clause 5.3 Exclude this clause Clause 5.4.2 Exclude this clause Note: For in situ polymerizing materials, there may be additional sample preparation requirements, so the review division should be contacted prior to conducting these studies. For materials with submicron components, there may be additional sample preparation and/or test conduct requirements, so the FDA should be contacted prior to conducting these studies. Related CFR Citations and Product Codes: Relevant Guidance: There is no relevant guidance developed at this time.