Nikki DeLuca Lab 4: UV/VIS and Fluorescence Purpose To learn how to use the UV/VIS and Specrofluorometer instruments. Known samples will be tested along with unknown samples. The UV/VIS will be used to sample aspirin and the Spectrofluorometer used to sample tonic water. Introduction UV/VIS spectroscopy measures the transmittance and absorbance of solutions. It has an energy source, monochromater, reference cell with blank solution, detector, amplifier, and output. The instrument is most often used for quantitative information about the analyte in a solution. Spectrofluorometers measure emission and can be used for quantitative information. It is most effective for large elements. They consist of an energy source, a two wavelength selectors, a detector, amplifer, and output. They are not frequently used due to more simpler and less expensive instruments being available. Procedure (Thermo Evolution 600 UV/VIS) Turn on the instrument by flipping the power switch on the left side Open the “VisionPro” program on the computer Select the “Scan” icon on the tool bar Wavelength: 190nm-400nm Scanning for: Absorbance Lamp: Tungsten Set baseline correction for 100%T Prepare Blanks: o Wash a quartz cuvette with hexane multiple times, fill 3/4th full, and dry with Kim wipes o Place one blank in each sample holder o After setting the Baseline Correction the program will have had a request for blanks, and a simple approval of the request will start the baseline scan, if denied, simply press baseline/zero icon Set the results dropdown menu to peak pick Prepare sample/standard and place in front holder, leave blank in the back one Press the “Run” icon Run each sample/standard three times without opening the instrument Click on sample name on left and right click. Click “Plot Sample” to put all samples on one graph. Close program on computer, turn off instrument, turn off computer monitor. Leave actual computer on! Procedure (Schimadzu Spectrofluorometer) Turn on the instrument (switch on bottom right) and computer Open the PX-150x program Check lower right corner on the program screen to see if it says the instrument is on Clock “Configure” and go down to “PC Configuration.” Click the “First Portal” Set the instrument to “Quantitative Analysis” by going to “Acquire Mode” and select “Spectrum” Make a 100 ppm stock solution Make 0.2 ppm, 0.1 ppm, 1 ppm, 5 ppm, and 10 ppm standards of quinine in 0.1M H2SO4 Fill a cuvette with 0.2 ppm solution and put into instrument holder Scan under these settings: o Spectrum Type: Excitation o EM Wavelength: 400nm o Ex Wavelength Range: 350 nm to wavelength that contains peak o Sensitivity: Low o Scan Speed: Fast o Recording Range: -10.00 to an amount that shows entire peak o EX/EM Slit Width: 10/10 o Response Time: 0.02 o Repeat Scan/ Auto File: No setting needed/ disable Click “Search λ” and “Search.” Set the default ranges for the EX (Excitation) and EM (Emission) to 350 - 600 nm and 240 – 650 nm. Record the optimal wavelengths found for both. Perform a new run under these settings: o Spectrum Type: Emission o EX Wavelength: The ideal wavelength found in the optimal wavelength search o EM Wavelength Range: Make sure that the optimal EM wavelength is included in the range o Sensitivity: Low o Scan Speed: Fast o Recording Range: -10.00 (low); 500.00 (high) o EX/EM Slit Width: 10/10 o Response Time: 0.02 o Repeat Scan/ Auto File: No setting needed/disable Run each standard and record wavelengths and intensities of the largest peak. Run three trials. Make tonic water solution by diluting 5mL of tonic water with 95mL of H2SO4 Record wavelength and intensity of largest peak. Run three trials. Data UV/VIS and Fluorescence graphs are in lab notebook. UV/VIS Absorbances 0.02 ppm 0.1 ppm 0.032 0.062 0.059 0.068 0.083 0.073 0.5 ppm 0.139 0.119 0.116 1.0 ppm 0.056 0.059 0.055 Unknown 0.212 0.209 0.215 Fluorescence Intensity 0.2 ppm 1.0 ppm 30.706 177.849 30.789 178.256 30.802 178.482 5.0 ppm 838.670 840.650 840.495 10 ppm 1015.008 1015.008 1015.008 Unknown 387.763 386.985 386.410 Calculations Standard Preparation for UV/VIS: M1V1 = M2V2 0.02 ppm * 50 mL = 100 ppm * x mL 0.1 ppm * 50 mL = 100 ppm * x mL 1.0 ppm * 50 mL = 100 ppm * x mL 5.0 ppm * 50 mL = 100 ppm * x mL 10 ppm * 50 mL = 100 ppm * x mL x = 0.1 mL x = 0.05 mL x = 0.5 mL x = 2.5 mL x = 5.0 mL Standard Preparation for Fluorescence: M1V1 = M2V2 0.2 ppm * 50 mL = 100 ppm * x mL 1.0 ppm * 50 mL = 100 ppm * x mL 5.0 ppm * 50 mL = 100 ppm * x mL 10 ppm * 50 mL = 100 ppm * x mL x = 0.01 mL x = 0.05 mL x = 2.5 mL x = 5.0 mL UV/VIS Calibration Curve UV/VIS Calibration Curve using Salycylic Acid 0.2 0.15 y = 0.127x + 0.0558 R² = 0.9999 0.1 0.05 0 0 0.2 0.4 0.6 Concentration (ppm) 0.8 1 1.2 UV/VIS Calculations for Unknown y = 0.127x + 0.0558 0.212 = 0.127x + 0.0558 x = 1.23 ppm Fluorescence Calibration Curve Fluorescence Calibration Curve using Quinone 1200 1000 Intensity 800 y = 102.2x + 102.28 R² = 0.8966 600 400 200 0 0 2 4 6 8 10 12 Concentration (ppm) Fluorescence Calculations for Unknown y = 102.2x + 102.28 386.376 = 102.2x + 102.28 x = 2.78 ppm Conclusion Compared to other instruments used in lab, the UV/VIS and Spectrofluorometer were fairly simple to use and analyze data. There were no problems during out experiment that required troubleshooting, and the data all seemed reasonable. For the objective of learning how to use the instruments for the future and how to analyze the output, the experiment was very successful. If the objective were to accurately quantify the aspirin and tonic water unknowns, the procedures and sample preparations might have to be revised.