Bring reagents at Room Temperature before use. Safety precautions For in vitro diagnostic use only. Do not pipette by mouth. Exercise the normal precautions required for handling laboratory reagents. Glucose Enzymatic Colorimetric Method principle Enzymatic colorimetric determination of Glucose according to the following reactions: GOD Glucose + O2 Gluconic Acid + H2O2 POD 2 H2O2 + Phenol + 4 - Aminoantipyrine Red Quinone + 4 H2O Intensity of the colour is direct proportional to the Glucose concentration in the sample *Reagents composition Reagent 1 Glucose standard 100 mg/dl Reagent 2 Phosphate Buffer pH 7.0 100 mmol/l Phenol 10 mmol/l GOD (Glucose Oxidase) 10000 U/l POD (Peroxidase) 1000 U/l 4 - Aminoantipyrine 0.3 mmol/l Sodium Azide < 0.1% According to the present laws the kit does not contain substances classified as dangerous *Procedure 1. Wavelength Hg 510 nm (500 - 550) 2. Cuvette light path: 1 cm 3. Temperature 37°C 4. Adjust the instrument to zero against air or distilled water Pipette into cuvettes Blank Sample standard Distilled water 10 µl Sample 10 µl standard 10 µl Reagent 1 ml 1 ml 1 ml Mix, incubate for 10 minutes at 37°C or 15 minutes at Room Temperature. Read the Absorbance of Calibrator and Samples against the Blank at Wavelength of 510 nm. Final color is stable at least 30 minutes *Calculation Serum, plasma and CSF A Sample Glucose (mg/dl) = ------------ x standard Value A standard Urine A Sample Glucose (mg/dl) = ------------ x standard Value x 10 A standard Storage and Stability of Reagents Store the kit at 2 - 8°C All the Components are stable until the stated expiration date if stored tightly closed and refrigerated. Urine/24 h Preparation and Stability of Working solution Reagent liquid and ready to use Calibrator liquid and ready to use Once opened the bottle, the reagent is stable for at least 30 days if stored tightly closed, refrigerated and protected from light Serum, plasma: CSF: Urine: Glucose in urine (mg/dl) Glucose (g/24 h) = -------------------------------- x Urine Volume 24 h (lt) 100 Each laboratory should establish its Normal Reference Range Performance Characterics A. LINEARITY LIMIT The reaction is linear up to 500 mg/dl For higher value dilute sample 1:2 with normal saline, repeat the test and multiply the value by 2 B. DETECTION LIMIT Values less than 3 mg/dl give non - reproducible results C. INTRA - ASSAY PRECISION (N = 20 replicates) Mean (mg/dl) SD CV% Control 1 98 3.42 3.49 Control 2 247 6.49 2.63 D. INTER - ASSAY PRECISION (20 replicates for 3 days) Mean (mg/dl) SD CV% Control 1 98.4 2.34 2.38 Control 2 245.7 4.35 1.76 E. ACCURACY Comparation between this method (y) and another commercial one (x), gave the following results: N = 20 r = 0.986085 y = 1x - 5.5 F. INTERFERENCES 1. Haemoglobin up to 400 mg/dl does not interfere 2. Bilirubin up to 15 mg/dl does not interfere 3. Triglycerides up to 1000 mg/dl do not interfere 4. Ascorbic Acid up to 30 mg/dl does not interfere Quality Control Control sera are recommended to monitor the performance of manual and automated assay procedures. Bibliography 1. Henry R. J.: Clin. Chem, Hoeber N. Y. 625 (1964) 2. Trinder P.: Ann. Clin. Biochem 6, 24 (1969) 3. Sharp P.: Clin. Chem Acta 40, 115 (1972) For in vitro diagnostic use only. The following symbols are used on labels For in vitro diagnostic use Reference values Use by (last day of the month) 60 - 110 mg/dl 50 - 70 mg/dl < 0.5 g/24 h These values should only be used as a guideline. Temperature limitation Batch Code www.betalab-eg.com