Specification of ventilator for pediatric department of Hindu Rao

advertisement

Specification of ventilator for pediatric department of Hindu Rao Hospital.

S. No. Specifications

1. Advanced technology ventilator for use in PICU, suitable for ventilating pediatric patients, from neonates to pediatrics patients up to 50 Kg

2. Multi microprocessor controlled system with individual selection of various ventilation parameters in all modes for minimum work of breathing.

3.

4.

5.

Should be sturdy, user friendly and freely mobile, should be a single unit with compressor in built for easy portability and easy positioning design near the patient.

The system should have facility for both pressure triggering & flow triggering.

Should have large screen (> or equal to 10 inches) and simple logical menus, with touch or feather operated buttons

6.

7.

Should have the non invasive (CPAP) and invasive pressure and

Volume control (with pressure support), assist control, and pressure support modes of ventilation needed for neonates and children.

The system should have the following parameters.

8. a)Tidal volume : around 3-1500ml or more b)Freq. 4-120 breaths/mt or more c)Pressure level 0-75cm H20 d)PEEP 0-40/50cm of H20 e)FIO2 21-100% f)Trigger pressure 10-0cm below PEEP g)Trigger flow 0.5-3 its per min. h)I:E ratio 1:10-4-1

Should have the following displays

Vt(Both Inspiratory and Expiratory)

MV(Mode display and breaths)

Spont. MV (Triggered and spontaneous breath display)

Freq

Leaks

FIO

2

9.

Trigger

All setting display

Delivered parameter real time continues display

Should have adjustable audio visual alarms based on intelligent software warning settings. The color coded mild to servere pop up messages with alarms

10. Unit should be supplied with servo controlled humidifier, with auto feed provision. The humidifier should be compatible with the routinely used Fisher pickle and other circuits (both with and without heating wire). There should be extra heating cord supplied for the disposable circuit and the reusable circuit so that both can be used and nebulizer for effective uninterrupted

Nebulization during mechanical ventilation.

11. Should be possible to display various types of waveforms and of loops for each breaths on the same screen. Minimum of 2 loop and 3 scalar graphics with adjustable wave scales (should be real time)

12. Should be able to measure compliance and resistance of the patients

13. Should have Y sensors capability with necessary sensor.

14. Two sets each of reusable silicon patient tubes and circuits, 2 neonatal Y sensor resusable, autoclavable probes, Two test lungs one each for for pediatrics and the neonatal, Ten sets each of disposable circuits with auto feed chamber and heater wire for neonates and pediatrics should also be supplied.

15. Should have air compressor of same make complying with necessary certification with automatic switchover facility to central supply and vice versa.

16. Should be able to work independently on room air as well as on

100% oxygen. Sustenance with 100% Oxygen in case of compressor not working

17. Should have permanent oxygen cell and two autoclavable interchangeable expiratory cassettes.

18. Should have facility for ventilation data transfer & network connection.

19. Quoted model should be upgradeable for future software upgrades.

20. The firm must sign a tripartite agreement with the principal company and MCD.

21. The firm must give a warranty of five years at first instance. The inclusions and exclusions should be clearly mentioned. Rates of

CMC for next five years after expiry of warranty should also be given.

22. Rates of consumables to be quoted separately

23. The firm must certify that they have not supplied this equipment

at a lower rate to any Govt. hospital.

24. The firm must certify the availability of spares and consumables for next 10 years.

25. The firm must guarantee that any complaints would be attended to within one working day and be rectified within further 48 hours. If the complaints are not rectified within the stipulated time; a standby equipment would be provided by the firm.

26. All rates to be quoted in Indian Rupees. Breakup of the total cost may be given to enable comparison.

27. List of users to be given along with performance report from at least two government hospitals.

28. The firm must have a service base in Delhi with 24 hrs service backup.

29. Demonstration is must in Hindu Rao Hospital.

30. A performance guarantee of 5% of total cost would be with held and paid after successful completion of warranty period.

31. Technical and Financial bids must be given separately.