IEC SYSTEM FOR CONFIRMITY TESTING AND CERTIFICATION OF ELECTRICAL EQUIPMENT COMMITTEE OF TESTING LABORATORIES IEC 60601-2-45 Edition 3 TESTING AND MEASURING EQUIPMENT/ALLOWED SUBCONTRACTING R=Required by Lab S=May be subcontracted Clause 203.6.7.102 203.8.5.4.102.5. 201.8.8.3 Measurement/testing Motion of anti-scatter grid Dielectric strength 201.9.2.3.1 203.6.7.103.2 Unintended movement ARTEFACTS from grid lines 203.6.7.104.2 Maximum LOADING TIME 203.8.5.4.102.6 Compression force 201.9.2.101.3 Biopsy needle positioning accuracy Testing / measuring equipment / material needed - Adequate Sand-filled bags or soft rubber blocks - Aluminium plate of 2mm thickness - Densitometer (Range 0 to 3,5) - Required using of memory scope with measurement functions - High voltage dividers or kV meter - Force gauge - PHANTOM of 20 mm PMMA - aluminium plate of 2 mm thicknesswith sufficient dimensions - densitometer - 40 mm thick rectangular Polymethylmethacrylate (PMMA) PHANTOM with sides equal to or exceeding 150 mm. - Compression scale - soft rubber block 20 mm to 50 mm thick, and 100 mm to 120 mm long and wide - Test device for mammographic stereotactic devices 1(3) Subcontracting W W R W W W S IEC SYSTEM FOR CONFIRMITY TESTING AND CERTIFICATION OF ELECTRICAL EQUIPMENT Clause Measurement/testing COMMITTEE OF TESTING LABORATORIES Testing / measuring equipment / material needed - Dynamometer - Adequate ion chamber associated to Electrometer-source/ or adequate Radiation dosimeter. - Adequate ion chamber associated to Electrometer-source/ Test for Half-value layer or adequate Radiation dosimeter. - Aluminium filters (Purity >= 99,99%) - Adequate ion chamber associated to Electrometer-source/ Primary protective shielding or adequate Radiation dosimeter. - Fit shielding of ion chamber Protective barrier - Adequate X-ray generator (35kV; percentage ripple <=4%; Total filtration of 0,03mmMo) Reproducibility of radiation output - Adequate ion chamber associated to Electrometer-source/ without AEC actived or adequate Radiation dosimeter. ( High sensibility) Linearity and constancy - Same equipment than above Accuracy and reproducibility of X-ray - Required using of memory scope with measurement functions tube - High voltage dividers or kV meter voltage - Required using of memory scope with measurement Accuracy of X-ray tube current functions - Set of precision resistors - Differential amplifier AEC requirements - Appropriate phantom Subcontracting Stability in normal use Radiation output 203.10.1 203.7.6 203.11.101 203.13.101 203.6.4.3.102.2 203.6.4.3.102.3 203.6.5.2 Accuracy of Irradiation time - Required using of memory scope with measurement functions - High voltage dividers or kV meter 2(3) W W W W W W W IEC SYSTEM FOR CONFIRMITY TESTING AND CERTIFICATION OF ELECTRICAL EQUIPMENT Clause Measurement/testing Accuracy of Current time product COMMITTEE OF TESTING LABORATORIES Testing / measuring equipment / material needed - Required using of memory scope with measurement functions - High voltage dividers or kV meter - Set of precision resistors - Differential amplifier Subcontracting Note: The presence of equipment alone does not indicate a satisfactory situation. Assessors must evaluate the equipment design, calibration, uncertainty and documentation to ensure compliance with the directions of the standard. The requirements of ISO Guide 17025 regarding validation are applicable, as the tests of this standard are not standardised tests. 3(3)