Standard Operating Procedure Title: Oxytherm Respirometer Calibration and Operating Instructions Department: Agronomy Created by: Whitney Bouma Laboratory: Crop Production & Physiology Lab Suite Supervisor: Dr. Allen Knapp Lab Supervisor: Whitney Bouma Date approved: Procedure Overview: Oxygen uptake measurements of mitochondrial and cellular respiration can be achieved with the Oxytherm respirometer and associated Oxygraph Plus software. The following are instructions necessary for operating this equipment, including electrode preparation, system calibration, and electrode cleaning/maintenance. Equipment and reagents necessary: Electrode preparation: Equipment: Clark electrode Beaker Sharpie marker Weight boat Spatula Analytical balance Stir bar Stir plate Thermometer PTFE membrane Cigarette paper Membrane applicator Dropper O ring Squeeze bottle with DI water Plastic waste container Reagents: Potassium chloride Sodium hydrosulfite DI H2O at 25 C Oxytherm system: Equipment: Oxytherm control unit and power cord Oxytherm peltier electrode chamber w/ reaction vessel and plunger Oxygraph Plus software RS-232 cable fitted with USB adaptor Electrode connection cable revised 18 June 2008 me1 Calibration: Equipment: Vacuum aspirator fitted with rubber pipette tip Erlenmeyer flask Mini stir bars (flea) Spatula Reagents: Sodium dithionite DI H2O at 25 C Clean up & Maintenance: Cleaning kit (polish and cotton buds) Procedure: The set up procedure for operating the Oxytherm system can be broken into five parts: 1) electrode preparation, 2) system setup, 3) system calibration, 4) system operation, 5) storage/preventative maintenance. Please follow the instructions carefully and see Whitney Bouma if you have any questions (rm. 1523 or 294-3886). 1) Electrode Preparation: The high precision Clark type electrode sensors are very sensitive and require proper preparation for accurate measurement. Please refer to the manual under the help menu in the Oxygraph Plus software for more information on the electrode operating principles and references for the oxygen electrode theory. You will also use this manual for detailed instruction on how to properly prepare the electrode and calibrate the system. The electrode disc is prepared so that an electrolyte bridge is established between the anode and cathode. The will allow a current to flow in the presence of oxygen. You will first need to prepare the electrolyte which is a 50% saturate Potassium chloride solution. If there isn’t any fresh electrolyte already prepared, complete the following procedure: Using a spatula, weigh boat and analytical scale, weigh 17.5 g of Potassium chloride and set aside. In a glass beaker with a minimum of 100 ml capacity, measure out 100 ml of de-ionized H2O. Using a thermometer, check the temperature. You will want the H2O to be close to room temperature (25 C), so it may be necessary to slightly warm the water. Add a stir bar to the beaker containing the water and mix in the 17.5 g of KCl by placing the beaker on a stir plate and gently mixing the solution. You will want to make sure that the KCl is fully dissolved so continue mixing until this is achieved. Place the solution in a glass bottle with a lid and label appropriately with contents and date prepared (remember no short hand). This solution should be good to use for up to a month. revised 18 June 2008 me2 You are now ready to prepare the electrode for measurements. Please read the help manual on “Preparing the Electrode Disc” thoroughly before beginning. It may take multiple tried to prepare the electrode properly if this is your first time. It is vital that there are no air bubbles or creases directly over the platinum cathode on the top of the electrode dome. Creases along the sides of the dome however are sometimes unavoidable and will not interfere with the electrode signal. The electrode preparation is completed in 4 stages (visual aid is available in the help menu under “Preparing the Electrode Disc” of the online help manual). When applying the paper spacer and membrane, it is best to use foreceps as oils from your fingers may interfere with the electrode signal. Place a small drop of electrolyte on top of the dome of the electrode disc. This will be directly on top of the platinum cathode. Place a 1.5 sq. cm paper spacer (cigarette paper) over the electrolyte ensuring that at least one corner of the spacer is in the electrode well to act as a wick. Cover this with a slightly larger, 2.5 sq. cm, piece of PTFE membrane. Place the small electrode disc O-ring over the end of the applicator tool. Hold the applicator vertically over the dome and slide the applicator shaft down to push the Oring over the dome. Check that the membrane preparation is smooth and that there are no trapped air bubbles. Top the reservoir well up with several drops of electrolyte. It is vital to make sure the larger O-ring is placed in position on the recess around the electrolyte well. If this second O-ring is not in place when the disc is installed into the electrode chamber, the silver anode will not be sealed from ambient air and measurements may be affected. You will want to test the response of the electrode before putting the electrode disc into the chamber. To do this you will connect the electrode disc to the rear of the control unit at the electrode input. Open the Oxygraph Plus software and start a recording by clicking the GO button on the toolbar. The signal will take up to 20 minutes to stabilize but the disc should read between 2000mV to 1800mV in air. Make sure that the electrode has properly stabilize and doesn’t drift downward over time. Once this reading is stable, breathe or exhale across the disc and observe the reaction the screen. The signal should drop due to the decrease oxygen in the exhaled breathe. The signal should soon begin to return to the original level as ambient oxygen begins to equilibrate around the electrode disc. Once this test has been completed satisfactorily you can stop the recording, set up the rest of the equipment, and then mount the disc into the base of the Peltier chamber. 2) System Set Up: This step covers how to connect all the components of the Oxytherm system. On the back of the control unit from left to right are the following connections and revised 18 June 2008 me3 inputs: 25 A fuse, 12 V Power supply input, RS232 input and output, electrode disc connection, and the auxillary input connection. Before making any measurements, it is important that the system is set up correctly. The diagrams in the online manual show the various connections that must be made to the control units. Please ensure the following are done before continuing on with system calibration. Ensure that the 12V power supply is connected to the 12V input socket on the control unit. Confirmation that this connection has been made is shown by an illuminated green LED on the front panel of the control unit. Connect the RS232 serial cable to an available serial port on the PC. You will need to make sure to use the USB converter attachment for the PC input. Connect the other end of the RS232 cable to the INPUT RS232 connection on the rear of the control unit. Mount the electrode into the Peltier chamber using the following directions: Remove the black base ring from the Peltier electrode chamber and mount the prepared electrode disc into the recess at the top. The electrode cable connection slots into a keyway in the chamber base meaning that the disc can only fit one way. Fit the chamber base and the disc assembly onto the Peltier chamber itself so that the electrode cable connection fits with the keyway on the chamber base thread. Gently hold the base in position while threading the base ring on to the Peltier chamber. Just slightly more that finger tight will suffice. Connect the Peltier Electrode Chamber and the prepared electrode disc, if necessary, to the rear of the base unit. You are now ready to start the calibration part of the set up. 3) System Calibration – Before any measurements can take place, the electrode disc must be calibrated so that the electrical signal received from the disc can be presented as actual calibrated units (nmol/ml). The calibration procedure is performed so that the signal from the oxygen electrode can be referenced to two known oxygen concentrations (air line and zero air) to derive an offset and a calibration factor. Temperature (°C) [Oxygen (ppm)] [Oxygen (nmol/ml)] 0 14.16 442.5 5 12.37 386.6 10 10.92 341.3 15 9.76 305 20 8.84 276.3 25 8.11 253.4 30 7.52 235 35 7.02 219.4 At a given temperature and atmospheric pressure, air saturated deionized water contains a known concentration of dissolved oxygen which may be calculated mathematically. The following information in the figure above is used by the Oxygraph Plus software in order to accurately reference the electrical signal from the electrode for the air line stage of revised 18 June 2008 me4 calibration. The data was published by Truesdale & Downing (Nature 173: 1236, 1954). Additional formulas and information on these values can be referenced in the Oxygraph Plus Online help manual under “System Calibration – liquid phase”. Using the information derived from the figure 1 below we can begin the calibration process. The Oxygraph software will guide you through most of the process with easy to follow prompts. Please follow these guidelines: Ensure that the electrode disc is properly installed into the base of the Peltier Chamber. Using a P1000 pipette, pipette 2 ml of air saturated deionized water into the reaction vessel of the chamber. Air saturated water is obtained by vigorously shaking a small quantity of deionized water (~50 ml) in a large Erlenmeyer flask (~ 1L). Then drop one of the small magnetic stir bars (fleas) into the vessel. Next is to set the assay temperature through the Oxygraph software, by selecting HARDWARE > TEMPERATURE CONTROL from the menu bar. This will display the actual temperature and then the setup temperature you desire. Set the temperature you wish to run your assays, for example 25 C. It is important to remember that the sample of air saturated H2O should be equilibrated to the assay temperature before the calibration procedure actually begins. This may take several minutes to allow the H2O to reach temperature once added to the reaction vessel. Once the H2O temperature and assay temperatures have equilibrated, initiate the calibration sequence from the CALIBRATE > LIQUID PHASE CALIBRATION menu option. Enter the appropriate temperature for your assay. The pressure should be standard atmospheric, but make sure to check the weather to see what it is and make conversions accordingly. You will already have H2O in the chamber so proceed by pressing OK. The next window prompts you to enter the stir bar settings. This operation can be accessed by selecting HARDWARE>STIRRER SPEED. Enter a speed of 75. This is usually sufficient to ensure that the dissolved oxygen within the samples is kept evenly distributed. Then press OK. The next window will show the signal as it changes and stabilizes. Once it has reached a plateau the window will prompt you to press OK to continue on to the second stage of the calibration procedure, establishing zero oxygen line. To perform this stage, we will use the chemical Sodium Dithionite. Add a few crystals to the reaction vessel. It really only requires a small pinch of crystals to deplete the oxygen in the vessel so the tip of a spatula worth of crystals should be sufficient. Once they are added, press OK to continue. Watch the signal drop. You should expect to see a signal of less that 1% of the air line signal. For example, if your previous signal was 1802 mV, you would expect to see a signal less than 18.02 mV for the zero line. If you are satisfied with the signal, press OK to continue. The next window will give you the calibration factor and the calibration offset. If these values are adequate, click the save button to store the calibration details. You will need to remove all traces of sodium dithionite from the reaction vessel. Using the aspirator, fit with the rubber topped pipette tip, suck out the waste. Be mindful of the stir bar and the electrode. Fill with dionized H2O for rinsing and remove. Repeat the rinsing several times to ensure the vessel is clean. revised 18 June 2008 me5 For more details about the calibration process refer to the online help portion of the Oxygraph software manual You are now ready to begin measurements. 4) System Operation/Data Acquisition – Full details on the Oxygraph software functions and system configuration can be viewed on the online help manual in the Oxygraph software. You will need to thoroughly read this section before commencing with your measurements. This information will help you determine the appropriate software settings and actions depending on the type of experiment you will be running. The specific details of conduction your experiment should be included in the experiment SOP. 5) Cleaning& Storing Electrode/Preventative Maintenance - To ensure accurate measurement, it is imperative that proper electrode cleaning and preventative maintenance measure be taken. Deposits of silver chloride and/or silver oxide will build up on the silver ring of the disc. Again, detailed instruction and illustrations of well and poorly managed electrodes are available in the online help manual of the Oxygraph software. It is important that the electrode is never left to dry out while prepared with the KCl electrolyte. You will need to complete the electrode cleaning procedure at the end of each day and ensure it is properly stored. Please follow these instructions for cleaning the electrode: Using the polishing paste, provide by Hansatech, you will want to polish both the silver anode and platinum cathode. Take a cotton bud moistened with a bit of DI H2O. Apply a small amount of polishing paste on the tip. Insert the bud into the well of the electrode and gripping the bud just above the tip, apply a moderate amount of pressure and gentry rotate the bud around the electrode well in a circular motion, 6-10 times. Using the other end of the cotton bud, again moistened with a bit of DI H2O, apply a small amount of polishing paste. Paying attention to restrict the polishing to the small platinum cathode, use a circular motion to gently polish the cathode. Make sure to avoid as much of the epoxy resin dome as possible. Once adequately cleaned, rinse the electrode several times with DI H2O to ensure all the polishing paste has been removed. If necessary, gently brush the area with a soft bristled toothbrush to remove any extra paste. Dry the electrode with a kimwipe and store the electrode in the designated, dessicated box. Personal Protective Equipment / Engineering Controls: Lab coat, eye protection and hand protection must be selected as required by Section D of the EH&S Laboratory Safety Manual. Check the box(es) next to the item(s) which are appropriate for your procedure. x Nitrile gloves x Safety glasses Dust mask Lab coat Fume hood Latex gloves Splash goggles Neoprene gloves Vented goggles Apron Biosafety cabinet Eye wash station Safety shower Respirator Insulated gloves x x Face shield revised 18 June 2008 me6 Note: Open-toed and heeled shoes are NOT allowed. Other Control Measures: . Handling & Storage Precautions: Potassium chloride: May cause skin, eye and respiratory tract irritation. Use appropriate PPE as listed above. Keep containers tightly closed in a dry, cool and well ventilated place. Avoid dust formation. Sodium hydrosulfite: May cause skin, eye and respiratory tract irritation. Use appropriate PPE as listed above. Avoid dust formation Is.unstable and can be reactive, self-heating, may catch fire. Keep cool, dry, well ventilated area and protect from sunlight. Do not store near acids or combustible materials. Refer to MSDSs for full details. Waste Disposal Procedures: With the implementation of the new Lab Safety Manual, EH&S eliminated the ability to dispose of non-hazardous waste by flushing it down a drain. Instead, all waste needs to be collected for EH&S disposal regardless of what the MSDS indicates. The following should be all you need in this section unless you have been specifically instructed otherwise. Unless EH&S specifically instructs otherwise, all chemical/reagent waste (including excess solutions) must be placed in an appropriately labeled hazardous waste container for EH&S disposal. Compatible substances may be combined into one waste container. Potassium Chloride (and all reagents made with Potassium chloride): will need to be collected by EH&S. Please place them in the satellite accumulation area with proper labeling. Sodium hydrosulfite: Do not let this product enter drains. Collect in a suitable, properly labeled container for collection by EH&S. Spill/Release Containment and Clean Up/Decontamination Procedures: Potassium Chloride – collect any chemical that has spilled and keep contained in a suitable container or with a proper spill kit for collection by EH&S. Sodium hydrosulfite: collect any chemical that has spilled and keep contained in a suitable container or with a proper spill kit for collection by EH&S. Remember, all waste needs to be collected for EH&S disposal – including that from a spill clean up. Health & Safety Summary for Required Reagents: Potassium Chloride: Eye Contact: Rinse immediately with plenty of water, also under the eyelids for at least 15 minutes. Obtain medical attention. Skin Contact: Wash off immediately with plenty of water for at least 15 minutes. Get medical attention immediately if symptoms occur. Inhalation: Move to fresh air. If breathing becomes difficult, give oxygen. Get medical attention immediately if symptoms occur. Ingestion: Do not induce vomiting. Obtain medical attention. Sodium Hydrosulfite: revised 18 June 2008 me7 Eye Contact: May cause eye irritation. Rinse immediately with plenty of water, also under the eyelids for at least 15 minutes. Skin Contact: May be harmful if absorbed through skin. May cause irritation. Wash off immediately with plenty of water for at least 15 minutes. Inhalation: May be harmful if inhaled. May cause respiratory tract irritation. Move to fresh air. Ingestion: Toxic if swallowed. Rinse mouth with water and consult a physician. C a r c i n o g e n T e r a t o g e n Chemical name Sodium hydrosulfite Potassium chloride M u t a g e n R e p r o d u c t i v e S e n s i t i z e r I r r i t a n t T o x i c E f f e c t s H i g h l y C o r r o s i v e T o x i c C o m b u s t i b l e C o m p r e s s e d G a s E x p l o s i v e F l a m m a b l e O r g a n i c P e r o x i d e s O x i d i z e r P y r o p h o r i c U n s t a b l e Target Organ(s) x x x W a t e r H e a l t h R e a c t i v e R e a c t i v i t y Incompatibilities x Strong oxidizing agents, acids, water Not known The above summary consists of guidelines for proper handling & disposal of chemicals used in this procedure. You must read and understand the contents of the entire MSDS(s) before starting this procedure. References: Oxygraph Plus software online help manual. http://www.sigmaaldrich.com/MSDS/MSDS/DisplayMSDSPage.do http://www.fishersci.com/msds?productName=BP366500&productDescription=POTASSIUM+ CHLORIDE+500G&catNo=BP366-500&vendorId=VN00033897&storeId=10652 revised 18 June 2008 me8 F l a m m a b i l i t y 2 0 2 1 1 0