LABORATORY “2” Gel Permeation Chromatography (GPC) of Polystyrene Mohammed Alzayer Chris Clay Xinhang Shen Mat E 453 Lab Section 2 September 16, 2014 1 ABSTRACT In polymer science, polydispersity index (PDI) is an important measure of the sizes of molecules present in a given polymer. In order to get an estimate of this number, there are several size separation techniques used that allow PDI to be determined indirectly. Those size exclusion methods include gel permeation chromatography (GPC). In this lab, a sample of polystyrene (PS) was analyzed to determine the molecular weight averages, PDI, and degree of polymerization. 1. INTRODUCTION 1.1 GPC Background Gel permeation chromatography (GPC) is a separation technique that helps determine the molecular weight distribution of polymeric materials. The chromatographic method works by placing rigid gel beads, also known as the stationary phase, in a column (Figure 1). Even though the term GPC is commonly used by researchers, “size exclusion chromatography” Figure 1. Gel Permeation Chromatography columns. [1] is the more accurate term to describe the technique since the stationary phase is usually not a gel. These beads are non-ionic and contain pores in variety of sizes; they are usually made of highly cross-linked polystyrene or glass (Figure 2)[2]. 2 The polymeric sample, dissolved in an appropriate solvent, flows through the column along with the beads. The polymer molecules diffuse into the bead’s pores. The larger the molecule, the fewer pores it can enter and the Figure 2. Beads used for GPC [3] less time it takes to elute. The time a molecule spends inside the column is referred to as the elution time. Therefore, in GPC, elution time is the parameter that indirectly measures the molecular weight distribution of a polymer since it distinguishes the sizes of molecules in a given sample [2]. The elution time is easily recorded by a detector after the molecules are separated and eluted from the column. One of the most common detectors used is the differential refractive index detector (RI), which compares the refractive index of the sample solution to that of a reference solution. The difference in refractive index will be proportional to the concentration of eluent. Another common detector measures the absorption of UV radiation at a specific wavelength. The amount of radiation absorbed is proportional to the concentration of eluent. Light scattering detectors make use of the fact that the amount of light scatter by a polymer is proportional to the mass of the polymer squared. 1.2 Solvent Selection The stationary phase size falls between 3 to 20µm while its pores are in the order of 50Å to 100 nm. The number of pores in the beads, just like their sizes, is random [4]. In order to test a polymer using GPC, the sample must be in a diluted form in order to be able to get inside the pores of the stationary phase. The solvent used in the polymer solution 3 must be selected carefully depending on the type of the polymer molecules. There are several types of solvents. Polar protic solvents include water and alcohols while dipolar aprotic solvents include other solvents such as DMF, DMSO, and acetone. Finally, there are nonpolar solvents such as THF and benzene [4]. Following the ‘like dissolves like’ rule, a nonpolar polymer would dissolve in a nonpolar solvent. For example, in this lab, a nonpolar solvent (chloroform) was used to dissolve polystyrene. There are other factors that contribute to the process of solvent selection. Other than the ability to dissolve the sample, the solvent must be inert to the material used in the column. It must also be able to withstand the operating temperature range. Another important factor is the solvent’s ability to easily separate from solute [4]. 1.3 Molecular Weight Distribution There are number of parameters that can be used to characterize the size of a polymer chain. These include the number average molecular weight (Mn), the weight average molecular weight (Mw), polydispersity index (PDI), and the average degree of ̅̅̅̅ ). polymerization (𝐷𝑃 The number average molecular weight is the total weight of a sample divided by the number of molecules within the sample and is given by Equation 1: 𝑀𝑛 = ∑𝑖 𝑁𝑖 𝑀𝑖 ∑𝑖 𝑁𝑖 , (1) where Mn : number average molecular weight, Ni : number of polymeric chains, Mi : the molecular weight of chains Ni . 4 The weight average molecular weight uses the weight fraction of weight range and divides by the total weight of the sample. Mw is typically bigger than Mn. Mw can be calculated using Equation 2: 𝑀𝑤 = ∑𝑖 𝑁𝑖 𝑀𝑖 2 ∑𝑖 𝑁𝑖 𝑀𝑖 = ∑𝑖 𝑤𝑖 𝑀𝑖 ∑𝑖 𝑤𝑖 , (2) where 𝑀𝑤 : weight average molecular weight, 𝑁𝑖 : number of polymeric chains, 𝑀𝑖 : the molecular weight of chains 𝑁𝑖 . 𝑤𝑖 : Weight of chains 𝑁𝑖 which equals 𝑁𝑖 𝑀𝑖 . The PDI gives information about how the molecular weight varies and is calculated by taking the ratio of the Mw and Mn (Equation 3): 𝑃𝐷𝐼 = 𝑀𝑤 𝑀𝑛 , (3) where 𝑃𝐷𝐼: polydispersity index, 𝑀𝑤 : weight average molecular weight, 𝑀𝑛 : number average molecular weight. The degree of polymerization tells how many monomers are present in the average polymer and can be calculated using Equation 4: ̅̅̅̅ = 𝑀 , 𝐷𝑃 𝑀 (4) 0 where ̅̅̅̅ 𝐷𝑃 : average degree of polymerization, 𝑀: molecular weight of polymer (can be either 𝑀𝑤 or 𝑀𝑛 ), 𝑀0 : number average molecular weight. 5 2. EXPERIMENTAL PROCEDURES 2.1 Materials Rubber gloves, goggles, electric balance, extraction needle, measuring glass, beakers, GPC column, intensity detector, and computer. 2.2 Sample Preparation 1. 25 mg of polystyrene from the previous lab was added to 5 mL of HPLC grade chloroform to make a 5 mg/mL solution. 2. Five mL of solution were drawn into a 5 mL syringe and filtered with 0.45 µL filter. (Figure 3). 3. 1.5 mL of solution were placed in a 2 mL vial. 4. The carousel for the GPC was pulled out, and the vial was placed inside. Note the position of the vial (number slot). 5. Sample location and name were entered in the “Sample Queue” Figure 3. Filtering the polystyrene solution. [1] spreadsheet in the computer software. 6. The “Inject Broad Samples” function was used with the method of “201202013_PS”. 7. A Run Time of 30 minutes was used. 8. The Inj. Vol. used was 25 µL. 9. The number of injections was set to 1, with 1 minute being allowed per injection. 10. The method was run for 30 minutes. Data was collected in real time and saved. 11. The sample vial was removed from the carousel. 6 3. RESULTS 3.1 Raw Data The raw data from the GPC came in the format of a spreadsheet with time and intensity columns. These can be plotted to obtain Figure 4: GPC Curve 26 21 Intensity 16 11 6 1 -4 1 6 11 16 21 26 -9 -14 Elution Time (min) Figure 4. Entire GPC curve for PS sample (400 mg BPO) 7 31 As the relevant data occurs in the 15 to 27 minute range, a separate plot can be generated for this range so the details can be seen more clearly (Figure 5). GPC Curve for Analysis 2.5 Intensity 2 1.5 1 0.5 0 15 17 19 21 23 25 27 29 Elution Time (min) Figure 5. Zoomed in GPC curve for PS sample (400 BPO) (15.8 to 27 mins) 3.2 Calibration Curve Based on the elution time and molecular weight data for a polystyrene monodispersed standard (Appendix, Table 1,) a log MW vs elution time curve was plotted (Figure 6). A linear function was fit to this data, allowing the molecular weight of any polystyrene sample to be calculated from the elution time. For example, if the elution time of a PS segment is 15.8, its log molecular weight should be - 0.2859(15.8) + 9.8625= 5.35 and taking the inverse of log yields a molecular weight of 221452 g/mol. 8 log MW vs Elution Time 6.5 6 5.5 y = -0.2859x + 9.8625 R² = 0.9832 log MW 5 4.5 4 3.5 3 2.5 2 13 15 17 19 21 23 25 Elution Time (min) Figure 6. Standard plot log Mw vs elution time 3.3 Polydipersivity Index and Degree of Polymerization As explained in section 3.2, the molecular weights of the sample segments were calculated using the equation obtained from the standard monodispersed PS. Hence, it was possible to calculate N, the number of polymeric chains, by multiplying the molecular weight by the peak intensity. Only the peak between 15.8 min and 27 min was analyzed. Using the equations listed in section 1.2, it was possible to get an estimate of both the polydispersity index and the average degree of polymerization. To get an estimate of the PDI, 𝑀𝑛 and 𝑀𝑤 must be determined first. Having both M and N calculated as stated above, “N×M” and “N× 𝑀2 ” were calculated as well. In order to calculate 𝑀𝑛 , the numbers in the N column were summed up. The numbers in N×M column were summed up as well. 𝑀𝑛 was finally calculated by dividing the sum of N×M by the sum of N. This 9 is shown in equation (1). 𝑀𝑤 was calculated in similar fashion using equation (2). Furthermore, the degree of polymerization using both the number average molecular weight and the weight average molecular weight were calculated. The table below summarizes the results of this lab. Table 1. Summary of results obtained from GPC analysis. Parameter 𝑀𝑛 (g/mol) 𝑀𝑤 (g/mol) 𝑃𝐷𝐼 𝑀0 (g/mol) 𝑋𝑛 𝑋𝑤 Value 55,418 89,950 1.62313 104.15 532.10 863.66 4. DISCUSSION 4.1 GPC Curve Peaks Figure 4 shows the entire intensity vs elution time curve including parts that are insignificant to the analysis of PS distribution. The flat line at the beginning of the curve indicates that no polymer had eluted from the column yet. The peak between 15.8 and 27 minutes was the peak we analyzed. In the significant GPC curve between 15.8 and 27 min (Figure 5), there are more than one peak. There are two major peaks at 20.5 min and 26 min. The peaks indicate that there are more than one size distribution of molecular weight in the polymer. The segments with the molecular weights of 10036 g/mol and 269 g/mol are the most common in the polymer mixture (calculated by log Mw = - 0.2859 t + 9.8625) using the elution time of the major peaks. 10 The sharp positive peak after 27 minutes indicates the presence of a large amount of very low molecular weight polystyrene. This could be the result of unreacted monomer of low weight polystyrene radicals combining with each other, preventing further growth. The negative peak at the end of the GPC curve might be a result of the solution running out of solvent, so a smaller refractive index would be measured. 4.2 Effect of Initiator Concentration From the “polymer synthesis” experiment, the 400 mg BPO sample is expected to have the lowest molecular weight compared to the 100 mg and 250 mg samples. Increasing the amount of benzoyl peroxide relative to styrene shortens the polymeric chains and hence results in a lower molecular weight. However, since only the 400 mg sample was tested in GPC, no quantitative data of how much higher the molecular weights of the other samples are can be provided. 5. CONCLUSIONS 1. GPC is a useful technique to gain information about the size and dispersivity of a polymer sample. 2. If a larger concentration of radical initiator is used, the resulting polymer chains are expected to have a lower molecular weight. 3. Average molecular weight on the order of 104 – 105 g/mol were obtained, which are typical for polymers. 4. The 1.6 PDI of the polystyrene sample in this experiment falls in the range expected for free radical polymerization (1.5<PDI<2.0). 11 6. ACKNOWLEDGMENTS The data was taken from a Mat E 453 group from the previous year. 7. REFERENCES 1. Gel Permeation Chromatography. Iowa State Polymer Composites Research Group, 15 Jan. 2013. Web. 2. Mendoza, J. D., Lab 2: GPC, Iowa State University. 3. Size Exclusion Chromatography (SEC). Phenomenex, 7 May 2014. Web. 4. Mendoza, J. D., Mat E 453 Lecture, Iowa State University. 12 Appendix Table 1. Data used for generation of calibration curve. Table 2. Example of spreadsheet used for calculations. Entire spreadsheet is not shown for brevity. 13