Free Space Absorption Tubes General Info Sheet <blurb about molecular absorption> <blurb about our tubes and purpose of document> Our free space absorption tubes come in a wide variety of configurations to suit our customers’ needs across several industries including gas sensing, calibration and research. Due to the large number of configurations this info sheet was created to help our customers determine what absorption tube best fits their needs. Of course, if you have any questions, please don’t hesitate to contact us. Section 1: Tube Parameters Description Path Length (does not include windows) Length Tolerance Outer Diameter Clear Aperture Stem Tube Wedge Tilt Window Thickness Abbr. Standard Specs L 2.5cm, 5.0cm, 10cm OD CA T +/- 0.1cm 12mm, 25mm, 40mm 9mm, 21mm, 36mm < 8mm 0.5 degrees nominal 3 degrees nominal 1.5mm nominal Section 2: Window Material Window material is primarily determined by the application wavelength range: < 2.6 microns: AR-coated glass < 6 microns: Magnesium Fluoride < 8 microns: Calcium Fluoride < 14 microns : Zinc Selenide Section 3: Concentration and Pressure sales@wavelengthreferences.com Ordering Information (example) (if deemed necessary for info sheet) CH4-T(25x10)-100-ZnSe Carbo n isotop e H(3):3cm path H(5.5): 5.5cm path Pressu re (Torr) Connector type (include /PD for photodetec tor) <spectra with gas absorption regions marked- if I can get it to fit> sales@wavelengthreferences.com