Mehmet Emre SUMER Koc University-Molecular Biology and Genetics 17 DECEMBER 2010 Section 2- Group 5 MBGE 101 - Experiment 8: Measuring DNA and Protein Concentration Aim: In this experiment, our aim is to estimate concentration and purity of given DNA and protein samples. In order to determine concentrations, we use the UV absorbance of the samples at wavelength of 260 nm and 280 nm. Introduction: There are many methods in molecular biology for determination DNA, RNA and protein concentration of given samples. A common practice for obtaining qualitative and quantitative information about those samples is to perform spectrophotometric analysis which means using the UV absorbance at 260 nm and 280mnm. The UV method is very widely used even though it is not highly accurate because it is a quick way, easy to apply, and little DNA is required. Nucleic acids and proteins have different absorbance at wavelenght of 260 nm and 280 nm. While proteins have little absorbance at 260 nm, both proteins and nucleic acids absorb UV light at 280 nm. Therefore, if there is a sample which is mixture of protein an nucleic acid their spectra overlap with each other, so it is not possible to determine their concentration. Indeed, proteins have two absorbance peaks in the UV region. Because of the absorbtion by peptide bonds there is one peak between 215-230 nm, and another peak at about 280 nm because of aromatic amino acids (tyrosine, tryptophan, phenylalanine). Taking into account the nucleic acids, there are actually two absorbance peaks depending on the subunit composition of DNA. The purine subunits have an absorbance maximum slightly below 260 nm, and the pyrimidines slightly above 260 nm. Therefore it is said that the absorbance peak of nucleic acids is at 260 nm. Estimation of Concentration: There are two UV methods for estimating concentration of pure samples. First method is constructing a standard curve and the other is ‘short-cut’ method which is using the absorptivity constants of proteins and nucleic acids. These absorptivity constants are calculated by Beer-Lambert Law: A= e x l x c A = absorbance e = coefficient l = pathlength c = concentration c = A / (e x l) Absorptivity constant of a protein varies depending on percentage of aromatic amino acids it contains and absorptivity of DNA depending on its base composition and on whether it is single-stranded or double-stranded. However, analysts commonly use ‘average’ absorptivity constants . According to these average constants; If a sample containing pure double-stranded DNA has an absorbance of 1 at 260 nm, then it contains approximately 50 μg/mL of double-stranded DNA. If a sample containing pure single-stranded DNA has an absorbance of 1 at 260 nm, then it contains approximately 33 μg/mL of DNA. If a sample containing pure protein has an absorbance of 1 at 280 nm, then it contains approximately 1 mg/mL of protein. As a result; the concentration of a pure DNA c = A260 x 50 x dilution ratio the concentraiton of a pure protein c = A280 x 1 x dilution ratio Estimation of Purity To estimate the purity of a solution, the absorbance of the solution at two wavelengths, 260 nm and 280 nm, are measured and then the ratio of two absorbance calculated. A260/A280 ratio of 2.0 is characteristic of pure RNA A260/A280 ratio of 1.8 is characteristic of pure DNA A260/A280 ratio of about 0.6 is characteristic of pure protein. Materials: UV spectrophotometer Quartz cuvettes DNA sample (1:40 diluted) Bovine serum albumin (1:40 diluted) Methods: The absorbance of our samples 260 nm and 280 nm was meaured. A fter that the A260/A280 ratio was calculated based on our measuruing. Then, we determined whether the sample is DNA or protein based on the ratio. We calculated the concentration of each sample according to their absorptivity constants. Results: 1 A260 2 3 4 5 6 7 8 0.123 0.156 0.278 0.352 1.110 0.091 0.280 0.207 0.103 0.160 0.321 0.225 0.735 0.049 0.180 0.157 1.194 0.975 0.866 1.564 1.51 1.857 1.56 protein protein protein DNA DNA DNA DNA 1.318 Protein dominating concentration 0.103 mg/ml 0.160 mg/ml 0.321 Mg/ml 0.018 Mg/ml 0.056 Mg/ml 0.005 Mg/ml 0.014 Mg/ml 0.010 Mg/ml A280 A260/A280 Protein or DNA ε (for protein 280nm) 1.0 mg.cm/ml For first sample A= ε.l.c A=0.103 l=1 cm ε=1.0 mg.cm/ml A=1.0 l=1cm c=A/ ε.l ε(for DNA260 nm) 50μg.cm/Ml A= ε.c.l A=1.0 l=1cm c=0.103/1cm.1.0mg.cm/ml c=0.103 mg/ml 3.500 3.000 absorbation 2.500 2.000 quartz 1.500 plastic 1.000 0.500 0.000 -0.500 240 260 280 300 320 340 wavelength Graph 1: absorbation difference for distilled water with quartz and plastic cuvettes 0.600 absorbation 0.500 0.400 0.300 DNA protein 0.200 0.100 0.000 200 210 220 230 240 250 260 270 280 290 300 310 320 330 340 350 wavelength Graph 2: DNA and protein absorbance in different wavelength Discussion: To understand which kind of cuvettes more appropriate, we use plastic and quartz cuvettes with blank (distilled water) and we measured their absorbance. After measuring we obtained absorbance for plastic cuvettes 3.000 and for quartz cuvettes 0,000. Based on these results we can say quartz cuvettes are more appropriate to use. The reason that makes quartz cuvettes more appropriate is ability of transparence throughout the UV completely. When we try to decide which one is DNA or protein, we looked their absorbance ratio. Because we know that when the absorbance ratio is close to 0.6, it means that the sample is protein. On the other hand, when the absorbance ratio is close to 1.8, it means that the sample is DNA. Questions: 1.Examine your four absorbance spectra. Do the spectra overlap? If so, at what wavelength(s)? Compare the spectrum of the mixture of nucleic acids and proteins to the pure samples. What features of the spectra vary? Where do they overlap? Yes, the spectra do overlap in 268 wavelength. We find the overlap for mixture by summing up absorbance of DNA and protein. Higher value of this summing shows peak point and if there is a peak point , spectra do overlap . 2. Fill in this table. UV SPECTROPHOTOMETRY OF DNA AND PROTEINS DNA A260 : 0,358 A280 : 0,21 A260/A280 : 1,705 Purity? : impure Concentration : 0.0105 PROTEIN A260 : 0.231 A280 : 0,318 A260/A280 : 0,736 Purity? : impure Concentration :0,318 MIXTURE A260 :0.207 A280 :0,157 A260/A280 :1.318 Purity? :DNA dominating Concentration :0,010 3. These UV assays are very commonly used by molecular biologists. Why? What information is obtained from these simple assays? This method generally is used by molecular biologist because molecular biologist work with small molecules such as DNA, RNA, nucleic acids and this method is very good to ubderstand concentration of these small molecules. 4. What are the assumptions of these UV methods? UV methods is very good method to understand absorbance and concentrations of small molecules as standard absorbation method . Furthermore ,Different small molecules weight products and their absorbance can be calculated by this method. Another important asumption of these methods is wavelengths. this assumption is also very importan because with this assumption, purity and concentration of DNA, RNA, and protein are calculated . 5. What are the advantages and disadvantages of these UV methods? Consider their accuracy, precision, range, ease of use, cost, and any other factors you think are important. The UV methods is widely used because it is a quick and easy way to estimate purity and concentration of DNA and protein. In addition, it requires little amounts of samples and lower cost than the other methods, however it is not highly accurate. For example, at 280 nm both proteins and nucleic acids can absorb light. Therefore, if we have a sample which is mixture of nucleic acid and protein it is hard to determine purity and concentration of the sample. At 260 nm, there is another problem. While we are performing the UV method with a DNA sample, we ignore the absorption difference between purine and pyrimidine subunits.There are some assumptions used not to handle with these problems. The average absorptivity constant is one of these assumptions. The operations done by using these constants do not give the actual results. Taking into account the purity, the UV method do not actually seperate DNA and RNA form each other.On the other hand, to form a general idea about given samples the UV method is the simplest and the quickest way. REFERENCES: - Glasel, J.A. (1995) Validity of Nucleic Acid Purities Monitored by A260/A280 Absorbance Ratios, Biotechniques 18:62-63. http://www.biotek.com/resources/articles/nucleic-acid-purity-a260a280.html