The Applicability of Cone-beam CT-based 3D IGRT for Head and Neck Radiotherapy| Zoller The Applicability of Cone-beam CT-based 3D IGRT for Head and Neck Radiotherapy The Development of a Standard Protocol and Methods for Improvement on Existing Immobilization BY WESLEY ZOLLER THE OHIO STATE UNIVERSITY Abstract: This study explores the applicability of Cone-beam CT (CBCT) image guidance and its ability to detect/adjust translational and rotational discrepancies in patient set-ups. CBCT has shown merit for alignment use in areas such as the head and neck where many critical structures lie within close proximity to the PTV. However, this study also looks at the added dose accumulation from CBCT and the potential harm resulting from this additional dose. In this study, we will measure the set-up discrepancies recognized by CBCT for 20 patients taken throughout the course of treatment at the Arthur G. James Cancer Hospital & Richard J. Solove Research Institute as a means of establishing the primary sources of set-up error for the existing immobilization techniques. In addition, we suggest two immobilization techniques that may reduce this source of error—one involving an Alpha Cradle © foam headrest built inside an aquaplast mask and another involving the use of a mask-attachable bite block to reduce head tilt. Using CBCT, the set-up discrepancies will be measured using the same methodology to see if either reduces the set-up and internal variance throughout treatment. Improvements on immobilization could potentially allow for reduced CBCT exposures, reduced dose, a smaller PTV margin, and a higher dose to tumor volume. Introduction In external beam radiation therapy (EBRT), the primary goals remain the same. As a standard, the ideal plan sticks to a three prong premise. The plan will (1) have high conformity to the target volume allowing for higher tumor dose for ablative bio-effects on malignant cells, (2) have sharp dose fall-off to avoid irradiation to critical structures, and (3) ultimately produce the best possible outcome for the patient. To achieve this plan, many innovations have been made in external beam radiotherapy over the years. These delivery practices range anywhere from intensity-modulated radiation therapy with respiratory gating to linac-based stereotactic radiotherapy. However, one major player in the goal of producing a higher and more conformal dose is in the reliability of the set-up itself. Quite simply, tumor conformity serves to be useless if the treatment target varies in location from day to day due to set-up inconsistencies. To help alleviate this set-up The Ohio State University| Radiation Therapy Class of 2012 variability, the use of image guidance (IGRT) has become a standard method in the treatment process. Port films and on-board imaging (OBI) digital fluororadiographs provide the ability to align to bony anatomy each day in accordance with the plan. The capability to acquire MV or kV conebeam images from the linac even furthers the ability to look at internal structures and tumor volumes, in addition to anatomy of the bone. From this, it is possible to compare the set-up in the treatment room to the original planning CT as it pertains to the target of interest. Many studies have been conducted documenting the improvements of CBCT and its imaging capabilities compared to that of orthogonal imaging. These studies will be outlined in greater detail later in this paper. Nevertheless, areas of anatomical variance that possess multiple critical structures can benefit from additional alignment practices. Page | Abstract-1 The Applicability of Cone-beam CT-based 3D IGRT for Head and Neck Radiotherapy | Zoller The head and neck region is an anatomical area that tends to have the greatest range of variability as it pertains to the clustered presence of critical structures and day-to-day discrepancies. Many of these structures such as the spinal cord, parotid glands, brain stem, esophagus, mandible, and larynx tend to fall directly in the treatment field for tumors of this region. In 1995, a group from the New York University Medical Center detailed the potential late bio-effects for individuals treated with EBRT to the head and neck region.1 The group specifically evaluated the mucosal lining, salivary glands, teeth, and mandible. Permanent xerostomia, loss of taste, severe mucositis, radionecrosis of bone, and perforation and permanent narrowing of the esophagus were all included in the radio-effects when the dose became too high. The group evaluated the tolerance doses for these late effects and concluded by reiterating the importance of immobilization and avoidance of these structures whenever possible in radiotherapy.1 Due to the presence of these structures, as well as the increased motility of the neck curvature coupled with head tilt, the set-up of head-and-neck patients is the primary concentration of this project. Quite simply, it is an area with high propensity for variation. To this point, as stated, CBCT has become a major resource for correcting set-up variances for this treatment area. Resulting from conebeam capabilities, it has even been suggested that PTV margins could be reduced due to the fact that CBCT image guidance allows for internal realignment on a daily basis. However, it is important to keep image guidance in perspective. A fact that cannot be forgotten is that daily image guidance—especially that of CBCT—is excess dose to the patient. It is important to determine this level of additional dose before a proper risk-reward analysis can be performed. Determining the exact dose from CBCT has been a premise in multiple existing studies, and these specifically will be addressed in the Review of Literature. It is also important to note that many patients with malignancies have already had multiple diagnostic imaging procedures performed as a work-up, most of which involve ionizing radiation. Combined with this is the fact that follow-up scans will become regular practice post-completion of treatment. It is not known exactly what percentage of added bodily dose can lead to a secondary malignancy, but every portion of this dose increases the potential for stochastic effects. In addition to this, shifting daily based on conebeam images does not necessarily make reducing PTV margins a reasonable practice. The Ohio State University| Radiation Therapy Class of 2012 Often times, due to the motility of the anatomy and set-up variance in the head and neck region, the alignment from the CBCT image to the planning CT image has a degree of discrepancy. Due to this, it is often necessary to utilize every bit of the existing planned internal tumor margin in order to accommodate for the avoidance of the critical structures. “Splitting” between maintenance of the margin and keeping the spinal cord from lying in the 95% isodose, for example, can be a daily problem. The internal margin, after all, is there for a reason. Conebeam is a useful tool in measuring and adjusting for this variance, but if reducing PTV margins is the goal, as mentioned previously, the root of the situation all ties back to the initial set-up immobilization tactics. A plan can only be as accurate as its set-up is consistent. To improve the standard head-and-neck immobilization procedure, it is necessary to know where the source of the error is coming from. To do this, MV Conebeam image guidance will be used in this proposed study to measure the translational and rotational errors of the existing set-up—conformal board, custom headrest, and thermoplastic mask that covers the shoulder. Once the primary sources have been determined, two immobilization tactics that are hypothesized to improve on these weaknesses will be tested—one involving an Alpha Cradle © foam headrest built inside the thermoplastic mask and another involving the use of a mask-attachable bite block to reduce head tilt. The measurements will be performed using the same CBCT procedure as for the initial evaluation. The purpose is to determine if these methods alone or combined can improve immobilization reproducibility from day to day. In turn, this could lessen the need—and thus, the dose— of CBCT rather than require guidance daily. Potentially, improved immobilization tactics would make it possible to achieve reduced PTV margins and generate a plan that is highly conformal to tumor volume. With the ability to escalate tumor dose while still sparing critical structures, it is possible to achieve the ultimate goal of improving patient outcomes. Review of Literature This proposed study will use conebeam imaging as a means of measuring set-up error. Multiple studies have been performed to evaluate the usefulness of conebeam CT as a method for evaluating these rotation-based errors in addition to the translational variety. In addition, the studies provide rationale behind the selection of conebeam Page |1 The Applicability of Cone-beam CT-based 3D IGRT for Head and Neck Radiotherapy | Zoller imaging as a measuring tool when compared to its orthogonal counterpart. In order to assess the positioning accuracy of 3D Conebeam CT IGRT versus 2D orthogonal kV OBI IGRT, researchers from the University of Texas M.D. Anderson Cancer Center studied 21 head-and-neck cancer patients undergoing radiotherapy.2 The results showed that when physicians used the IGRT to make couch shifts, 4.1% of the CBCT group had shifts greater than or equal to 0.5 cm in at least one direction, and 18.7% were over 0.3 cm. For the kV OBI image guidance, only 1.7% had shifts greater than 0.5 cm and 11.2% were greater than or equal to 0.3 cm. Greater shifts on the 3D image indicated that the ability to visualize target tissue along with the bone anatomy could potentially be more accurate than the kV orthogonal images. These results were deemed statistically significant, but the team noted that the possibility remains that the differences could be caused by immobilization error and potential rotation of the head and spine due to set-up inconsistencies.2 This immobilization error is what this further proposed study will be designed to evaluate and address. In a similar study, researchers from the University of Wuerzberg, Germany used CBCT to determine the effectiveness of conebeam guidance versus orthogonal imaging in correcting translational and rotational setup errors.3 In the study, the researchers assessed the set-up accuracy of 24 patients—8 head and neck, 10 pelvic, and 6 thoracic—using a total of 209 CBCT studies and 148 electronic portal images. To do this, the team measured translational errors alone, rotational errors alone, and the effect of the combination of both translational and rotational error on target coverage and critical organ sparing. The results of the study showed that CBCT and portal imagining differed by less than 1 mm in translational error for 70.7% of the patients and less than 2 mm for 93.2% of the patients. However, with the use of CBCT, the researchers found rotational errors greater than 2° in 3.7% of pelvic tumors, 26.4% of thoracic tumors, and 12.4% of head-and-neck tumors. The maximum rotational errors for each body site were 5° in the pelvis, 8° in the head and neck, and 6° in the thorax. Based on comparing this to the initial treatment plans, the researchers concluded that for patients with longer target volumes and targets close to critical structures, the combination of translational and rotational errors significantly lowered target coverage and highly increased doses to the organs at risk. The team suggested that with the feasibility of daily conebeam guidance, the ability to successfully diminish significant rotational error is essential for highprecision external beam radiation therapy.3 If the goal The Ohio State University| Radiation Therapy Class of 2012 is to reduce PTV margins, which is typically around 5 mm, every millimeter counts. Although CBCT and orthogonal correction only differed by 2mm for 93.2% of patients in the correction of translational error, this 2mm can be significant to ensuring that the tumor volume falls within the PTV. In addition to the Wurzberg study, researchers at the Institute of Cancer Research in Surrey, UK evaluated the application of kV conebeam computed tomography image guidance when compared to electronic portal imaging (EPI) in correcting for set-up errors in oesophageal EBRT patients.4 In the study, the researchers measured variation in anterior–posterior, right–left, and craniocaudal directions as well as the rotational directions of pitch, roll, and yaw. In all, 20 patients with oesophageal tumors were implemented in the study, and set-ups included skin marks/tattoos, along with kV-CBCT or EPI prior to treatment which were compared to planning CT slices or corresponding DRR’s. A total of 122 EPI orthogonal images and 207 CBCT scans were collected, and the systematic and random errors were calculated. The authors found that the systematic and random errors for CBCT guidance were 1.3 mm, 1.7 mm, 1.4 mm, and 2.6 mm, 3.9 mm, 2.0 mm in right-left, cranio-caudal, and anterior-posterior directions, respectively. The results for the EPI were not significantly different. However, CBCT was able to identify rotational errors greater than 3o in 44 images, something EPI could not identify. Based on these results, the research team concluded that CBCT’s adequate 3D volumetric image quality improves the accuracy of treatment delivery, especially by catching rotational errors.4 Expanding the applicability of conebeam CT image guidance, researchers at the Cancer Research Center and the University of Heidelberg in Heidelberg, Germany also included tumors located in the chest and pelvis in their study.5 The group selected 6 patients with different tumor diagnoses— lung cancer, sacral chordoma, head-and-neck, paraspinal tumor, and two with prostate cancer—in order to assess the set-up correction applicability of kV CBCT. The head-and-neck and lung patients were immobilized using a vacuum pillow in combination with a head mask, while patients with prostate cancer or paraspinal tumors were immobilized using a wrap-around body cast and a head mask. The image quality of the kV CBCT was deemed sufficient in all cases, and the bony landmarks and anatomical target were visualized clearly, with the exception of poor soft tissue contrast for an obese prostate patient. The results showed a maximum set-up deviation of 3 mm for patients Page |2 The Applicability of Cone-beam CT-based 3D IGRT for Head and Neck Radiotherapy | Zoller immobilized with the body frame, and 6 mm for patients with the vacuum pillow. In 4 of the 6 cases, the target point was corrected. The authors predicted that the CBCT took an extra 10-12 minutes maximum, and was significant in the correction of set-up deviation—especially in areas with many critical structures such as the heart, spinal cord, and esophagus—for high dose stereotactic treatment.5 All of these studies seem to indicate the usefulness of CBCT as a guidance tool, especially to correct rotational errors where orthogonal imaging is inadequate. However, an error due to a patient being rotated for treatment set-up who was not rotated in the planning CT cannot be simply remedied by translational table shifts. This could only be the result of immobilization error, as the M.D. Anderson study seemed to indicate. Due to this fact, relying on CBCT image guidance to correct for inconsistency errors would only go so far. The immobilization techniques must improve. CBCT IGRT on a daily basis cannot fully correct the problem, especially if it is delivering excess radiation to the patient. As previously discussed, it is necessary to get an estimation of this excess radiation dose from CBCT image guidance before the trade-offs can be properly examined. Multiple studies have been performed to get this estimation. In a study from the Princess Margaret Hospital’s department of oncology (Hong Kong), a research group sought to determine the absorbed organ doses and effective tissue doses from using conebeam CT image guidance.6 Three different scan sites and 26 organs were specifically outlined for this study: the head-and-neck, the chest, and the pelvis. Using an on-board imager (OBI) kV conebeam system and numerous thermoluminescent dosimeters (TLD’s), effective body doses and absorbed organ doses were measured on a female anthropomorphic phantom. The study also utilized both the standard CBCT mode and the low-dose mode. The results showed single-scan skin doses from standard CBCT of 6.7 cGy for the head and neck, 6.4 cGy for the chest, and 5.4 cGy for the pelvis. The TLD’s showed single-scan effective body doses of 10.3 mSv for the head and neck, 23.7 mSv for the chest, and 22.7 mSv for the pelvis. When repeated for low-dose CBCT, the measurements were approximately 20% of the standard mode readings. The group estimated that this could total to over 2 Gy over the course of treatment for standard conebeam image guidance, and could potentially increase the risk of secondary cancer by 2-4%. The research team concluded by discussing the importance of reducing mAs whenever possible and using low-dose CBCT in areas where The Ohio State University| Radiation Therapy Class of 2012 the bony anatomy is the key structural set-up indicator.6 Along similar lines, a research team from the department of radiation physics of the University of Toronto also performed a study to determine the significance of the dose accumulation to normal tissue from daily kV-CBCT image guidance.7 The dose measurements were performed on an Elekta linear accelerator using a farmer ion chamber and MOSFET detectors at the surface, periphery (2 cm depth), and center of 2 cylindrical water phantoms, 16 and 30 cm in diameter. Along with this, the study utilized both 360° and half-rotation conebeam scans on the 30 cm diameter water phantom using a standard setting of 120 kVp and 660 mAs. The results showed that the maximum dose at the center measured to be 1.6 cGy and 2.3 cGy at the surface of the body for full rotation CBCT. The researchers estimated that this dose could account to over 1 Gy to normal tissue at the skin surface for patients receiving the image guidance daily over the course of external beam treatment. The team also noted that the dose cannot be accurately added into the dosimetry treatment plan due to suspicion that kV energy has different biological effects on cells than mV treatment. Due to this, the results show that the scanning parameters should be reduced whenever possible, and this can be accomplished using less kVp, smaller fields-of-view in the longitudinal direction, and by using only half-rotation conebeam scanning for image guidance.7 The results from these two studies indicate the significance dose delivery from CBCT imaging, especially in the terms of kV conebeam. The quality of kV imaging tends to produce better resolution, which gives a better comparison to the kV-acquired planning CT. Avoiding daily guidance would help to reduce this excess dose accumulation, but only as an effective trade-off if immobilization can be done sufficiently enough to merit reducing the number of CBCT’s. In 1995, Gunilla Bentel and a research group from Duke University Medical Center detailed the use of a customized set-up for head and neck cancer that included the use of a foaming agent customized headrest similar to the Alpha Cradle © material used in this study.8 Though dated, the team documented the conformity of the headrest to the back of the skull and stabilization of the neck. The study included a customized headrest contouring to the posterior head and extending to the inferior angle of the posterior neck. This also included a thermoplastic facemask contouring to the brow, nose, and chin. The custom Page |3 The Applicability of Cone-beam CT-based 3D IGRT for Head and Neck Radiotherapy | Zoller headrest was assembled using a 10 x 18 inch plastic bag and 100 cc of foaming agent, which the authors estimated of taking less than 10 minutes to create. The effectiveness of the set-up was then compared to conventional set-ups utilizing standard headrests. The team examined the custom head rests created for 20 head-and-neck patients, 2 being children. The height under the head ranged from 5.8- 9.8 cm and under the neck from 3.2 -6.2 cm, while the standard head supports ranged from 5.0-8.0 cm and 1.0- 5.0 cm, respectively. Also, the superior to inferior distance between the highest point under the neck and the lowest point under the head for the customized supports ranged from 5.0-14.0 cm, while only 9.0-11.0cm in the standard supports. The study concluded that the wide span of the measurements reflected the variability between the patients, and that the standard head supports do not adequately conform to the posterior contour. The authors also suggested that the shape of the customized headrest provides support to the most inferior aspect of the neck and the upper thorax, which is crucial for reproducibility of the cervical spine and is not addressed by standard head supports.8 This set-up was as improvement over standard Silverman headrests due to conformity to the patient’s head and spine. In the set-up of this proposed study, the Duke principle will be built upon. By allowing the foaming agent to expand to the brink of the thermoplastic shoulder mask, the goal would be to eliminate gaps, which, as a result, could improve reproducibility. In the area of the head and neck, the critical structures tend to shift with the patient based on the curvature of the neck and chin—a reason for being the primary area of interest in the proposed study. This shift tends to occur if the mask does not fit correctly or if the head rest does not properly support the entire neck. In a study performed by the Albert Einstein College of Medicine in Bronx, NY, the research team actually hypothesized that some headand-neck cancer patients tend to lose their lordotic cervical spine curvature throughout the course of external beam treatment.9 This spinal shifting can cause for anatomical marks to be varied from the original marks and cause divergence from the treatment field originally depicted on the planning CT. Using a 6 mV linear accelerator, the authors performed a study detailing the treatment of 50 headand-neck patients over the course of 40-42 elapsed days. Ensuring reproducibility, the therapists used shoulder straps, custom thermoplastic face masks, and multiple tattoos. To measure the change in this lordotic curvature, the investigation team defined a cervical spinal angle (CSA) to be calculated by The Ohio State University| Radiation Therapy Class of 2012 recording the initial lateral port field images and comparing these to port films in the final week. The borders of the CSA included a projected line running parallel to the posterior surface of the C2 vertebral body and another line lying parallel to the body of the C6 vertebrae. The results showed that on average, the CSA of the patients decreased by 2.26° over the course of the external beam treatment. For patients with the lowest isocenters, the results were higher showing a 3.8° change, which translates to a 1 cm divergence from a distance of 15 cm. Based on the results, the group suggested that orthogonal films be used to check spinal cord alignment prior to treatment as the elapsed days accumulate or perhaps using a more comfortable means of immobilization to keep the curvature from decreasing throughout the duration. 9 This is what the proposed immobilization tactics hope to account for, as well as expand further on the study to measure consistency of the curvature of the spine throughout treatment. In an effort similar to the proposed study, a group from the University of Heidelberg and Department of Oncology of the Mannheim Medical Center in Mannheim, Germany used 3D/3D matching, x-ray volume rendering, and conebeam CT image guidance to assess the accuracy of two different mask systems—the thermoplastic mask and the rigid Delta cast.10 In the study, 21 patients—14 had rigid masks and 7 had thermoplastic—with intracranial or head-and-neck tumors were assessed using the 3D/3D matching and CBCT, separating the skull and neck regions. Translational and rotational errors of the isocenter were recorded as well as an isolated analysis of the error based on cervical vertebra. To analyze the results, the researchers created a displacement vector, v, by taking the square root of the quantity, (x2 +y2 + z2), where x, y, and z represented shifts in the lateral, vertical, and longitudinal directions. This displacement vector quantity for the rigid masks calculated as 0.312 cm for intracranial tumors and 0.586 cm for the neck. For the thermoplastic mask, the researchers measured a vector of 0.472 cm for intracranial tumors, and 0.725 cm for the neck. In association with this, rigid masks accompanied with body tattoos had a v length of 0.35 cm in the neck region. In conclusion, the research team suggested that the Delta cast rigid masks were superior for intracranial precision, and that for the neck region, both thermoplastic and rigid should be accompanied by CBCT guidance.10 Building on this Heidelberg study, the proposed study entails the use of the thermoplastic mask in combination with the Alpha Cradle © foam custom headrest, along with a mask-attachable bite Page |4 The Applicability of Cone-beam CT-based 3D IGRT for Head and Neck Radiotherapy | Zoller block to reduce head tilt. The effort is to create a system that diminishes the need for the CBCT guidance—at least not daily or multiple times per week. Along similar lines for measurement, the proposed study will also utilize the calculation of displacement vector by taking the square root of the quantity (x2 +y2 + z2). Methods In this study, all images will be obtained from patients who have or will have undergone IMRT treatment at the Arthur G. James Cancer Hospital & Richard J. Solove Research Institute for head and neck tumors of varying stages and histologies. All conebeam images will be captured using a MV CBCT image of a 220° arc performed on a Siemens Oncor model treatment unit complete with EPID. The number of conebeam studies per patient will vary from 10-15 as the patients were imaged throughout the course of treatment 2-3 times per week. For the preliminary portion of the study, the CBCT images will be examined from 20 existing head and neck cancer patients of varying stages and histologies. All patients in the study will have received 35 total fractions—once daily for 5 treatments per week—for a total dose of 70 Gy. In addition, the patients will be non-boosted and never have been remarked throughout treatment. The standard set-up will include a conformal board, an aquaplastic mask that extends over the shoulders, a custom head rest that extends to the base of the neck, and a knee sponge. As a note, the patient will lie supine with arms placed by his or her sides. The translational shifts will be measured using the shifts performed, approved, and documented by the physician in the Lantis Record and Verify system. X denotes shifts in the cross-plane or horizontal direction, Y denotes shifts in the in-plane or vertical direction, and Z denotes shifts in the longitudinal position. Once these shifts are obtained, a standard deviation will be performed by simply taking the variance from the average of the values (X, Y, and Z each done independently). In addition, the average magnitude of the shifts will be obtained by taking the absolute value of the shifts, and taking an average based on directional independence. Once these are determined, two vector quantities will be calculated to better analyze the set-up error in three dimensions. To calculate an average shift displacement vector, Vshift, the square root of the quantity (Xshift2+Yshift2+Zshift2) will be obtained using the independent magnitude shift averages in each direction. Then, a deviation vector will be calculated The Ohio State University| Radiation Therapy Class of 2012 to determine the average variance of this vector by using the equation: Vdev = (Xstandard dev2+Ystandard 2 2 (1/2) . dev +Zstandard dev ) Once the translational sources of error have been measured, the next step is to quantify the rotational errors. This will be done by individual analysis of all conebeam images for these patients using the “Image J” program. This program is free software equipped with angle and distance measurement tools, complete with screen capture capabilities. The three rotation tilt planes examined in this study will be the axial, sagittal, and coronal directions. To measure the axial rotation, two lines will be projected on an axial slice. The first will be a vertically projected line, stopping at the tip of the posterior tubercle of the atlas vertebrae (C1), and the second line will start at this same tip of the posterior tubercle and extend through the lateral tip of the right transverse process of C1. The consistency of this angle will be measured throughout treatment based on the initial angle in the original planning CT. All measurements will be done using the angle tool. The measurement of sagittal tilt will be analyzed on a sagittal slice. An angle will be obtained between a vertically projected line (ending at the superior tip of the odontoid process) and a line starting at the superior tip of the odontoid process and extending through the tip of the external occipital protuberance. This consistency, too, will be measured from the planning CT as baseline. Using a coronal image, the coronal tilt determination will be performed by measuring the angle between a horizontally projected line and a line running parallel to the base of the skull at a level anterior to the cochlea. This, too, will use the planning CT as a baseline. Once the primary sources of immobilization inconsistencies have been quantified, this proposed study suggests two methods of immobilization improvement for a board reviewed study. In the first wing, 10 patients will be set-up using the same set-up as the preliminary analysis with the addition of a mask-attachable bite-block. It will be formed on the day of simulation and will be clipped into the mask as the aquaplast forms through cooling. This will allow for it to remain constant and permanent throughout treatment. The aim of this bite block is to prevent axial, sagittal, and coronal tilt variance as the upper maxillary molars and maxillary incisors are firmly and repetitively placed in the mouthpiece of the block. The central premise behind this is that the maxilla is much more stable than the mandible. In the existing mask system, the primary reliance is on Page |5 The Applicability of Cone-beam CT-based 3D IGRT for Head and Neck Radiotherapy | Zoller the conformity of the mask to the patient’s chin. Adding one more point of resistance, such as the bite block, is hypothesized to improve stability. In the second wing of the proposed study, 10 patients will be treated with a similar set-up as the preliminary study with the insertion of the Alpha Cradle © foam custom headrest in the place of the existing headrest. The headrest will be formed within the confines of a curved shell mimicking the maximum expansion of the posterior portion of the aquaplast mask frame. This study would allow this foaming agent to expand to its fullest within these confines, eliminating the gaps between the mask and the patient’s head as well as areas where the existing headrests do not conform. The headrest will be allowed to expand upward behind the ear and conforming to the superior portion of the patient’s skull. In addition, the foam will give stability down the patient’s neck extending to the upper back. The premise is that reduced gaps would hypothetically reduce variability. The aquaplast mask will not be made until after the foaming agent of the custom head rest has been allowed to completely harden (approximately 10 minutes). In the third and final wing, 10 patients will be set-up utilizing both suggested immobilization improvements. The set-up would include the Alpha Cradle © foam custom headrest in addition to the mask-attachable bite block. Again, the patient would be supine on a conformal board with arms by his or her side. All measurements and quantifications for these three wings will be determined using CBCT images and documented shifts, identical to the evaluation of the preliminary portion of the study. This will allow for an accurate comparison of set-up error. All patients will also receive 70 Gy over 35 fractions, in the same lines as the preliminary examination subjects. 3. Guckenberger M, Meyer J, Vordermark D, Baier K, Wilbert J, and Flent M. Magnitude and clinical relevance of translational and rotational patient setup errors: a cone-beam CT study. Int. J. Radiation Oncology Biol. Phys. 2006; 65(3):934–942. 4. Hawkins M, Aitken A, Hansen V, McNair H, and Tait D. Set-up errors in radiotherapy for oesophageal cancers – Is electronic portal imaging or conebeam more accurate? Radiotherapy and Oncology. 2011; 98:249–254. 5. Thilmann C, Nill S, Tücking T, et al. Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences. Radiation Oncology. 2006; 1(16):19. 6. Kan M, Leung L, Wong W, and Lam N. Radiation dose from cone beam computed tomography for image-guided radiation therapy. Int. J. Radiation Oncology Biol. Phys. 2008; 70(1): 272–279. 7. Islam M, Purdie T, Norrlinger B, et al. Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy. Med. Phys. 2006; 33(6):1573-1582. 8. Bentel G, Marks L, Sherouse G, and Spencer D. A customized head and neck support system. Int. J. Radiation Oncology Biol. Phys. 1995; 32(1):245-248. 9. Garg M, Yaparpalvi R, and Beitler J. Loss of cervical spine curvature during radiotherapy for head-and-neck cancers: the neck moves, too. Int. J. Radiation Oncology Biol. Phys. 2004; 58(1):185–188. References 1. 2. Cooper J, Fu K, Marks J, and Silverman S. Late effects of radiation therapy in the head and neck region. Int. J. Radiation Oncology Biol. Phys. 1995; 31(5):1141-1164. 10. Boda-Heggemann J, Walter C, Rahn A, et al. Repositioning accuracy systems—3D revisited: 3D/3D matching with Radiation Oncology 66(5):1568–1575. of two different mask comparison using true cone-beam CT. Int. J. Biol. Phys. 2006; Li H, Zhu XR, Zhang L, et al. Comparison of 2D radiographic images and 3D cone beam computed tomography for positioning head-andneck radiotherapy patients. Int. J. Radiation Oncology Biol. Phys. 2008; 71(3):916–925. The Ohio State University| Radiation Therapy Class of 2012 Page |6