MEC316 Lab#3 Mass-Flow Measurement Group #9: Kanchan Bhattacharyya – Writer; Synthesizing Paper Matthew Stevens – Graphs & Error Analysis Ting Zhang – Abstract and Introduction Xie Zheng – Principle of Operation and Experimental Procedure A. Abstract The general goal of this experiment is first to measure the flow rates of gases with Rotameter and Electronic Mass Flow Meter and estimate the accuracy of Rotameter compared with Electronic Mass Flow Meter. Basic principle of Rotameter is based on the variable area principle. The basic principle of Electronic mass flow meter is measuring the temperature of heat transfer in the liquid. (Variables introduction: Q represents flow rate measured with Rotameter. Q2 represents flow rate measured with Electronic Mass Flow Meter. U1 stands for voltage which is measured by DAQ. H stands for the float height in Rotameter.) Variables height H and voltage U1 are required to be measured in this experiment. As long as height H is known, we can calculate the flow rate Q1 which is measured with rotameter because Q1 is proportional height H. Flow rate Q2 which is measured with Electronic Mass Flow Meter can also be obtained as long as we know voltage U1. After find out Q1 and Q2 at same heights h, we can estimate the accuracy of Rotameter and Electronic Mass Flow Meter. The second goal of this experiment is to measure flow rate of liquids to estimate the accuracy of one method when compared to the other method. We use two equipments Vortexshedding flow meter and Ultrasonic flow meter in measuring liquid flow-rate. The operation of the vortex-shedding flow meter is sensing the frequency of vortex shedding behind a bluff-body which is directly proportional to the velocity of the fluid. The basic operation principle of Ultrasonic flow is depending on the ultrasonic pressure signals and their reflections to measure the velocity. (Variables introduction: Q3 represents flow rate measured with Vortex-shedding flow meter. Q4 represents flow rate measured with Ultrasonic flow meter. U2 stands for voltage which is measured by DAQ.). Liquid flow rate Q3 with can be directly measured with Ultrasonic flow meter. Q4 can also be obtained as long as we get U2 which is measured with DAQ because Q4 is proportional to U2. After mapping graph Q3-U and Q4-U, we can estimate the accuracy of Vortex-shedding flow meter compared to Ultrasonic flow meter. B. Introduction In gas flow-rate measurement, we use air as a representative gas. Both the rotameter and Electronic Mass Flow Meter are used to measure the gas flow rate under the same condition. We keep outlet pressure on the air tank at 10psig. Next, we begin to change the float height H in rotameter. Then, we started to measure the float height H in the Rotameter and voltage U1 with DAQ. After we get the H and U1, we can calculate the gas flow rate Q1 and Q2 respectively because Both H, Q1and U1, Q2 are proportional respectively. Let me mention something of Rotameter. Rotameter is an industrial flowmeter used to measure the flow rate of liquids and gases. Rotameter is popular because it has a linear scale, a relatively long measurement drop. It’s simple to install and maintain. It may be manufactured in a variety of materials of construction, and for a wide range of pressures and temperatures. The rotameter can be easily sized or converted from one particular service to another. In general, it owes it wide use to its versatility of construction and applications. In liquid flow-rate measurement, we use water as a representative fluid. Both the Vortexshedding flow meter and Ultrasonic flow meter are used to measure the fluid flow rate under the same condition. By adjusting Valve 1 (see Figure 3-3), we can change the liquid flow rate. Each change of Valve 1, we record two data. One of them is voltage U2 that is acquired by DAQ. The other data the liquid flow rate Q4 that is directly measured by Ultrasonic flow meter. We can obtain Q3 if we know U2 because they are proportional. C. Theory - Principles of Operation for Flow Meters Rotameter: Rotameter is an industrial flow meter used to measure the flow rate of gases and liquids. Its operation is based on the variable area principle, where the fluid flow raises a float in a tapered tube, increasing the area of passage of the fluid. The greater the flow, the higher the float is raised. The height of the float is directly proportional to the flow rate. The float moves up or down in the tube in proportion to the fluid flow rate and the annular area between the float and the tube wall. It reaches a stable position in the tube when the upward force exerted by the flowing fluid equals the downward gravitational force exerted by the weight of the float. A change in flow rate upsets this force balance. The float then moves up or down, changing the annular area until it again reaches a position where the forces are in equilibrium. With liquids, the float is raised by a combination of the buoyancy of the liquid and the velocity head of the fluid. With gases, the buoyancy is negligible, and the float responds to the velocity head alone. Since there are specific disadvantages when use the rotameter to read the liquid flows, the rotameter is mainly used for gases. Electronic mass flow meter: The gas flows through a precision sensing tube where a constant heat rate is applied by electrical resistance heating coils. The heat transfer to the liquid upstream and downstream of this heating coil is a function of the temperature of gases at these points as well as the mass flow rate. We can accurately determine the mass flow rate of the gas by knowing the temperature of the gas at these two points. Vortex-shedding flow meter: At low Reynolds numbers (velocity), the frequency of vortex shedding behind a bluff-body is directly proportional to the velocity of the fluid. The frequency of vortex shedding is sensed by a piezoelectric transducer which then gives us the velocity of the liquid. Ultrasonic flow meter: This is an advanced technique for measurement of flows in pipes and depends on the ultrasonic pressure signals and their reflections for the measurement of velocity D. List of Equipment and Experimental Procedure Air Mass Flow Rate Experiments 1) Electronic Mass Flowmeter 2) Rotameter (Omega Engineering, Inc) (Model FMA1728ST; Omega Engineering, Inc. FMA 1700/1800 Series) 3) Compressed Air Tank (Model DOT-3AA2400) 4) Harris Pressure Regulator (Model 35-500C-560) 5) Polyethylene Tubing & Connection Fittings 6) National Instruments DAQ Board 7) DC Power Supply 8) Laptop Computer Water Flow Rate Experiments 1) Ultrasonic Flowmeter Omega Engineering, Inc. FD-7000) 2) Vortex Shedding Flowmeter (Omega Engineering, Inc. A100F50) 2) Lightweight Laboratory Cylindrical Tank (Norton Plastics 54102-0015 HDPE 57L, with Cover and Spigot) 4) Leeson AC Motor (Model C6C34FK61A) 5) PVC Piping, Elbows, and Couplers. 6) Plastic Valves 6) National Instruments DAQ 7) DC Power Supply 8) Laptop Computer Experimental Procedure Gas flow-rate measurement (air) 1. Rotameter and the electronic mass flow meter are connected in series to measure the flow rate of air. We set the outlet pressure on the air tank to 10 psig. Flow through both meters is controlled by turning knob on the bottom front of the rotameter (By turning the knob counterclockwise, we can increase the flow rate). 2. Since we are using the duplicate sets of equipment, we do not need to setup the electrical flow meter except adjust the knob to make the excitation voltage of 17 volts. Also, we do not need to worry about the current which should not be exceeded 250mA in old equipment. Check all connections and then turn on the power supply. 3. Turn on the laptop and start the data acquisition program by loading WINDOWS, execute LABVIEW. Then open exp3a.vi in the \labview directory. 4. Keeping the outlet pressure on the air tank to 10 psig, we start to make measurement corresponding to each height of the float. First, we start with the height equals to 15mm (the top of the sphere lined up with the scalar on the ruler in rotameter). Record the setting of height and click on the run icon on the tool bar. We get a reading of voltage from the program and record that value. Then, we take increment of 15mm and take another 8 sets of data while repeating the recording. 5. Close the valve of the air tank and reset the height of the float in the rotameter back to the original position. Liquid flow-rate measurement (Water) 1. Since the equipment is configured already in the lab, we turn on the power supply and set the voltage to 24V. Configure the system as shown in Fig 3.3 and make sure the valve 1 is closed and valve 2 is open. 2. Turn on the pump and run the system for at least 5 minutes to purge the air out of the piping system. 3. Open file exp3b.vi in LABVIEW and then turn on the ultrasonic flow meter which is already configured to measure the flow velocity. 4. After approximately 6 minutes doing step 2, we start to make the sets of measurements by changing valve 1. Since we decide to take 10 sets of measurements, we start with turning the valve about 9 degree (it is hard to make sure the angle is close to 9, but we try to make the increments even). 5. Then, record the reading shows up on the ultrasonic flow meter and also click on the run icon on the tool bar to read the output current of the vortex meter which is measured using LABVIEW named exp3b.vi. 6. By adjusting valve 1, make the increment is approximately 9 degrees and repeat the step 5. After repeating the steps, we get 10 sets of data in total. 7. Close the valve 1 and turn off the power supply, pump and also the ultrasonic flow meter. E. Results and Discussion Part I: Air Flow Rate Experiments Table 1: Experimental Data & Corresponding Flow Rates (Air) Trial (i) Rotameter Reading (mm) EMFM Output Voltage (Volts) EMFM Flow Rate (L/min) [Reference] Rotameter Flow Rate (L/min) +/- Error in Absolute Rotameter Deviation of Measurements Rotameter Flow [2% Reading Rate from Acc. + ¼% Reference Repeatability] EMFM Flow Rate 1 20.000 0.567 5.670 5.025 0.1131 0.6450 2 35.000 1.250 12.500 11.970 0.2693 0.5300 3 50.000 1.800 18.000 18.915 0.4256 0.9150 4 65.000 2.492 24.920 25.861 0.5819 0.9410 5 80.000 3.284 32.840 32.806 0.7381 0.0340 6 95.000 3.997 39.970 39.752 0.8944 0.2180 7 110.000 4.614 46.140 46.697 1.051 0.5570 8 125.000 5.562 55.620 53.642 1.207 1.9780 9 135.000 5.728 57.280 58.272 1.311 0.992 Caption: This table shows the rotameter readings and the EMFM output voltages as water flow was changed, in Columns 2 & 3 respectively. In Column 4 we have the EMFM flow rates, which we assume to be our reference and in Column 5, we have the rotameter flow rates. Columns 6 & 7 show the error in rotameter measurements as per the manufacturer who gives a 2% reading accuracy and ¼% repeatability uncertainty range, as well as the absolute deviation of rotameter flow from the reference. These two can be compared to see the differences in the rotameter from the actual as determined by the manufacturer and the differences from the rotameter to the EMFM flow rate, which we assume as the reference and thus the actual value. Discussion & Analysis Part 1: Maximum Error of Rotameter Sample Standard Deviation: Maximum Error Formula: Maximum Error in Rotameter Flow Rate Measurement: 0.9422 (95% Confid. Int.) Maximum Error in Rotameter Reading (mm): 11.2 mm How Maximum Error Was Determined: The formula for calculating maximum error from probability & statistics is listed above, as the sample standard deviation (s) divided by the square root of the total sample space, multiplied by a t-value, which will vary depending on the number of samples (specifically the degree of freedoms (DOF), n-1) and the confidence interval desired. Using the electronic mass flow meter (EMFM) as the reference, the deviations between the EMFM and the rotameter for each measurement 1-9 were found. These were then squared and added, multiplied by 1/DOF, and then the square root of the result was taken. The last step was to multiply by the desired confidence interval which we took as 95%, multiplying our result by 2.896. Our final maximum error in the rotameter flow rate measurement came out to be 0.9422. This maximum error value covers the spread of deviation values in Column 7 of Table 1, aside from one outlier. In order to obtain a more useful maximum error to the user of the rotameter, we used the linear regression curve to calculate what change in rotameter reading (mm) would cause this maximum error value in flow rate. By backtracking and solving for x, we find that this comes out to 11.2 mm. Why The Maximum Error Seems So Large: (Sample Size & EMFM Reading Acc.) This may seem to be a staggering amount, but it is very important to keep in mind that this is the result of having to ensure such a large confidence interval of 95%, which requires multiplying by a larger t-value. If we had a larger sample size, then according to the t-score table, we would have had to use a smaller t-value as our multiplier and this would bring our maximum error for rotameter reading (mm) to a more sensible value. It is also important to point out, that the EMFM was treated as an absolute reference – that it was the exact value of the flow rate. However, the EMFM has it’s own reading accuracy error of 2%. Looking at Column 6 of Table 1, where the +/- error that the manufacturer says there is between the rotameter flow rate measurement and the actual flow rate (we assume to be from the EMFM) – which was calculated by taking 2.25% of each rotameter flow rate measurement (rotameter reading accuracy + repeatability), we see that this error is far less than the actual deviation of the rotameter flow rate measurement from the EMFM. Thus, it is not a good assumption to say that the EMFM is 100% exact. If we include the 2% from the EMFM and do a rough estimate of the manufacturer’s error assurance and we simply double the error values in Column 6, we see that they will come quite close to the actual deviation. This would result in smaller variances when calculating our sample standard deviation which would ultimately lead to a far smaller maximum error. Table 2: Rotameter Calibration: Linear Regression and Uncertainty Analysis Trial (i) x y xy x x, a+bx 1 20.000 5.670 113.400 400.000 5.025 0.416 2 35.000 12.500 437.500 1225.000 11.970 0.281 3 50.000 18.000 900.000 2500.000 18.915 0.838 4 65.000 24.920 1619.800 4225.000 25.861 0.885 5 80.000 32.840 2627.200 6400.000 32.806 0.001 6 95.000 39.970 3797.150 9025.000 39.752 0.048 7 110.000 46.140 5075.400 12100.000 46.697 0.310 i i i i i i (y - (a+bx )) i i 2 8 125.000 55.620 6952.500 15625.000 53.642 3.912 9 135.000 57.280 7732.800 18225.000 58.272 0.985 Sx Sy Sxx Sxy a b theta u 715.000 292.940 69725.000 29255.750 -4.236 0.463 7.676 0.600 a u b 0.007 Caption: This table shows the rotameter readings (mm) and rotameter flow rates in x and y, and how these are transformed into a linear regression curve that minimizes the objective error function and provides the closest linear fit to the data. This is assumed to be the curve that models the rotameter’s reading to the actual flow rate. Graph 1: Mass Flow Rate (Rotameter & EMFM) vs. Rotameter Position MASS FLOW RATE VS. ROTAMETER POSITION 70 MASS FLOW RATE (L/MIN) 60 y = 0.463x - 4.236 50 40 30 20 10 0 0 20 40 60 80 100 120 140 ROTAMETER POSITION (MM) Electronic Mass Flowmeter Flow Rate (L/min) -4.235+0.463x Rotameter Flow Rate (L/min) Discussion & Analysis Part II – Manufacturer’s Accuracy of Rotameter: Is the rotameter as accurate as the manufacturer claimed? As stated before in our analysis of maximum error, the 2.25% measurement assurance that the manufacturer gives, between the rotameter measurements and the actual value, results in error values far less than the actual deviations between the rotameter flow rate measurements and the reference (assumed to be actual) standard EMFM. So by only going by the “tolerance” of the rotameter measurement, it is not as accurate as it should be. However, we make a big assumption by saying that the EMFM is a viable reference standard since it has 2% reading accuracy uncertainty of its own and if added to the 2.25% measurement assurance from the rotameter manufacturer, it is enough to make the manufacturer’s error assurance equal the observed deviation. Part II: Water Flow Rate Experiments Table 3: Experimental Data & Corresponding Flow Rates (Water) +/- Error in Vortex Shedding Measurements [0.8% Rate Accuracy] Absolute Deviation of Vortex Shedding Flow Rate from Reference Trial (i) Output Voltage (Volts) Ultrasonic Meter Flow Rate [Reference] (G/min) Vortex Shedding Flow Meter Flow Rate (G/min) 1 1.846 9.070 8.964 0.0717 0.106 2 1.820 8.790 8.724 0.0698 0.066 3 1.724 7.830 7.840 0.0627 0.010 4 1.633 7.010 7.002 0.0560 0.008 5 1.580 6.340 6.513 0.0521 0.173 6 1.527 5.900 6.025 0.0482 0.125 7 1.468 5.420 5.482 0.0439 0.062 8 1.392 4.750 4.782 0.0383 0.032 9 1.334 4.370 4.247 0.0340 0.123 10 1.253 3.600 3.501 0.0280 0.099 Caption: This table shows the readings and the EMFM output voltages as air flow was changed, in Columns 1 & 2 respectively. In Column 3 we have the EMFM flow rates, which we assume to be our reference and in Column 4, we have the rotameter flow rates. Columns 6 & 7 show the error in rotameter measurements as per the manufacturer who gives a 2% reading accuracy and ¼% repeatability uncertainty range, as well as the absolute deviation of rotameter flow from the reference. These two can be compared to see the differences in the rotameter from the actual as determined by the manufacturer and the differences from the rotameter to the EMFM flow rate, which we assume as the reference and thus the actual value. Table 4: Vortex Shedding Flowmeter Calibration: Linear Regression and Uncertainty Analysis Trial (i) xi yi xiyi xixi a+bx (y - (a+bx )) 1 1.846 9.070 16.74322 3.407716 8.963795 0.011279604 2 1.820 8.790 15.9978 3.3124 8.724285 0.00431849 3 1.724 7.830 13.49892 2.972176 7.839941 9.88261E-05 4 1.633 7.010 11.44733 2.666689 7.001657 6.96048E-05 i i 2 5 1.580 6.340 10.0172 2.4964 6.513426 0.03007646 6 1.527 5.900 9.0093 2.331729 6.025194 0.015673605 7 1.468 5.420 7.95656 2.155024 5.481691 0.003805829 8 1.392 4.750 6.612 1.937664 4.781586 0.000997676 9 1.334 4.370 5.82958 1.779556 4.247295 0.015056502 10 1.253 3.600 4.5108 1.570009 3.50113 0.009775255 Sx Sy Sxx Sxy a b theta u 15.577 63.080 24.629 101.623 -8.041 9.212 0.091 0.277 a u b 0.177 Caption: This table shows the rotameter readings (mm) and rotameter flow rates in x and y, and how these are transformed into a linear regression curve that minimizes the objective error function and provides the closest linear fit to the data. This is assumed to be the curve that models the rotameter’s reading to the actual flow rate. Graph 2: Mass Flow Rate vs. Vortex Shedding Output Voltage MASS FLOW RATE VS. VORTEX SHEDDING OUTPUT VOLTAGE 10 FLOW RATE (GALLONS/MIN) 9 y = 9.2119x - 8.0414 8 7 6 5 4 3 2 1 0 1.125 1.25 1.375 1.5 1.625 1.75 1.875 2 VORTEX SHEDDING OUTPUT VOLTAGE (VOLTS) Ultrasonic Flow Rate (G/min) Vortex Shedding Flow Rate (G/min) -8.904+9.212x Caption: This plot shows both the vortex shedding flow rate and the ultrasound flow rate as a function of voltage. As can be seen the differences in terms of whole gallons of flow rate per minute, is marginally small. Discussion & Analysis Part III – Maximum Error for Vortex Shedding Flow Meter Sample Standard Deviation: Maximum Error Formula: Maximum Error in Vortex Shedding Flow Rate Measurement: 0.0971 G/min Maximum Error in Vortex Shedding Meter Reading: 0.2945 V The formula for calculating maximum error from probability & statistics is listed above, as the sample standard deviation (s) divided by the square root of the total sample space, multiplied by a t-value, which will vary depending on the number of samples (specifically the degree of freedoms (DOF), n-1) and the confidence interval desired. Using the ultrasound meter as the reference, the deviations between the ultrasound and the vortex shedder flow meter for each measurement 1-10, were found. These were then squared and added, multiplied by 1/DOF, and then the square root of the result was taken. The last step was to multiply by the desired confidence interval which we took as 95%, multiplying our result by 2.896. Our final maximum error in the vortex shedder flow rate measurement came out to be 0.0971. This maximum error value covers most of the spread of deviation values in Column 6 of Table 3. In order to obtain a more useful maximum error to the user of the vortex shedder, we used the linear regression curve to calculate what change in vortex shedder reading (V) would cause this maximum error value in flow rate. By backtracking and solving for x, we find that this comes out to 0.2945. Explaining Maximum Error: Again, like for the rotameter, the sample size is quite small, so to have a 95% confidence interval, means multiplying by a significant 2.896, almost 3x. If we were to have a larger sample size, such as more than 60 samples, then our t-score value multiplier would be far less and would result in the end, as a smaller maximum error for our vortex shedder flow meter. Evaluating Manufacturer Specifications Against Calibration Curve: The manufacturer of the vortex shedder ensures a rate accuracy of 0.8% while the manufacturer of the ultrasound meter ensures an accuracy of 1%. Taking a look at the last two columns in Table 3, which show the error in the vortex shedding flow meter and the absolute deviation between it from the reference ultrasound, respectively, we see that the deviations are extremely small and for the most part are covered by the manufacturer’s tolerance. F. Error Analysis Sources of error include improper calibration, inaccuracy in measurement, and malfunctions in the system. During calibration of the electronic flow meter we may have chosen a current value too far from 0.25 ampere before selecting the 24 V range, and similarly we may have chosen an inaccurate value when trying to achieve the 17 V required excitation voltage. This would directly correlate to inaccurate measurements for flow rate as any given position would always give a variant flow rate, depending on the degree to which the calibration current/voltage varies. For the liquid flow-measurements, we may have incorrectly calibrated the system by not allowing the pump to run for long enough before taking our measurements. In giving the system adequate time to circulate the previously stagnant water, we are ensuring that that the flow will be consistent throughout the system while simultaneously verifying the integrity of the system. Specifically, in allowing the liquid to flow for an extended period of time we are giving ourselves an opportunity to identify any leaks in the system or other mechanical failures. While we did not observe any of these failures during the trials, taking measurements prematurely may prevent us from identifying these sources. Sources for error during measurement include inaccurate readings of rotameter float position as well as inaccurate electronic mass/vortex shedding flow meter output voltage readings and ultrasonic flow meter flow rate readings. While we waited for the float to settle into position during each flow measurement, it is possible that we could have misjudged the actual position of the float. Observing actual flow rates which correspond to the various float positions we find that these flow rates depend linearly on the position, implying that any error in float position would give an inaccurate approximation for the flow rate. Similarly, we can consider our electronic mass flow meter output voltages to be potential sources for error. Like we did with the rotameter float, we waited for the output voltage to ‘settle’ before we recorded our final value at each given flow rate. However, it is possible that we recorded our values either prematurely or too late. Inaccuracies in these measurements could result in both high/low voltages for a given flow, which in turn would affect the calibration curve and subsequently our approximation for the accuracy of the rotameter. A similar assessment can be made for inaccurate ultrasonic meter flow rate readings and the calibration of the vortex shedding flow meter. We can also consider defects in the experimental setup; specifically those which exist within the respective tubing systems and flow meters themselves. While we thoroughly inspected all experiment materials before each respective trial (gas, liquid), we must consider leaks in the tubing/ducts assembled to transport the fluids, leaks in the joints connecting the tubing and ducts together, as well as failure in connections to the DAQ and power supplies. For the liquid water measurements our system was comprised of threaded ducts, couplers, and valves interconnected with a pump. We can consider losses due to losses in these joints, as well as those that may result from using basic stainless steel worm clamps to connect the reservoir to the pump. These losses have the propensity to propagate, and may affect other properties of the system in addition to the flow rate. Similarly, we can consider losses in the gas trials, as our meters were connected via polyethylene tubing. While the tubing is connected between the meters by threaded fittings, there may be losses which exist at these joints. Also, one of the tubing connections to the Rotameter was held in place by simple “Scotch” Tape, hardly a guaranteed method to keep the gas contained with the system. Finally, we can consider any inaccuracies that might have existed in the applied gage pressure (10 psig) during the gas trials. We assume that our estimation for an applied 10 psig pressure is accurate as measured by both our eye as well as the pressure gage indicating the output pressure. Before each individual trial we were sure to accurately set the pressure to 10 psig, however rapid fluctuation makes this very difficult to measure both accurately and precisely. Subsequently, any variance in pressure is going to have a direct correlation on the resulting flow rate through the system. Using Linear Regression Analysis; we found uncertainties of ua = 0.600/ub = 0.007 in the Rotameter and ua = 0.277/ub = 0.177. These values can be attributed to any of the above considered sources for error. G. Conclusion As we can see from our data plots, for both the gas flow measurements using air and for the liquid flow measurements using water, both instruments used in either type of measurement scale fairly well. However, absolute deviations exist and the manufacturing tolerance given is usually quite tight as we saw for the rotameter which was only 2.25% when you consider both reading accuracy and repeatability accuracy. Matching the data to this sort of tolerance works well when you have a large sample size. Unfortunately, since we have only 9-10 values for our data, any outliers can have a significant effect on the maximum error we calculate for the rotameter and for the vortex shedder. This is because, for smaller sample sizes, the t-score value grows enormously in order to ensure the same 95-99% confidence interval. To ensure an uncertainty to that confidence requires a larger breadth when you have smaller sample size, because you cannot be sure your sample data is enough to reflect the entire population. If we had say 60 samples, to the point where we could say our data is continuous, the t-score values would be far lower resulting in more reasonable maximum errors. But overall, as we can see from the graphs with respect to the scaling of L/min for the rotameter or G/min for the vortex shedders, the differences that these measurements are from our reference, is quite small all in all and would make these instruments a good test for various kinds of fluid flows. Reference: J.M. Kincaid, David Hwang, T.Y. Hsu. Laboratory Manual for MEC316