OPERATING MANUAL Philips XRG – X-Ray Goniometer System INDIANA UNIVERSITY Department of Geological Sciences Written by R. Ensman Contributions By: C. Shriner, P. Levy, B. Mertz & K. DeHart 1 OPERATING MANUAL FOR THE PHILIPS XRG-3100 COMPUTER-CONTROLLED X-RAY DIFFRACTION UNIT Rev. January 10, 2010 Table of Contents: page Principles of Crystallography Using X-Ray Diffraction 3 Safety Precautions for Using X-Ray Equipment 3 Sample Preparation 5 Operation of the X-Ray Generator and Goniometer Instrumentation Preliminary Checks Turn on the AC Line Power Turn on the X-Ray Generator Shut down the Generator 8 9 9 16 Startup Procedure for Databox / Computer Control 12 System – Operational Verification 13 Running Samples 14 Presentation of Results Using JADE Software 17 Optimization of Goniometer and Signal Conditioning Parameters Calibration of 2 Goniometer Readout Optimization of Signal Conditioning Parameters XRG Maintenance / Scheduling Condensed Operating Instructions Sample XRG Printout for Alpha Quartz 20 22 23 24 26 (See a separate Operating and Maintenance Log 3-ring binder for notes) Revision : 8/15/2011 file name: 2 XRG Manual Ha.docx Principles of Crystallography Using X-Ray Diffraction 1. A collimated beam of monochromatic x-ray photons directed at the surface of a finely divided mineral or kaolin sample will be diffracted at various angles that depend upon the mineral content of the sample. 2. The intensity of the diffracted beam and the angle between the incident and diffracted beam (2) can be used to identify a mineral or minerals present in the sample. 3. The relative intensity of peaks (usually two or more) and their corresponding 2 angles can be compared with spectra of known pure minerals to help determine the composition and relative percentages of complex mineral mixtures. Safety Precautions for Using Single Layer Shielded X-Ray Equipment 1. The Philips XRG-3100 X-Ray Diffractometer utilizes radiation that is readily absorbed by soft tissue. Severe damage to skin and/or longer term cancer risks is a possibility due to prolonged exposure or cumulative exposure over time. 2. The XRG-3100 system was designed many years ago and thus presents a higher risk source of radiation exposure than the double radiation shielding used in modern instrumentation. Additional lead sheet shielding has been added over the years to shield potential “hot spots” (very narrow beams of radiation). 3. However, extra precautions that are to be taken by operators are: a. All operators must wear a radiation badge (finger, film type) on their hand likely to be used close to the x-ray beam. Periodic badge testing will be responsibility of the machine supervisor. most the b. At the start of each machine use, with the beam ON, a sample in place and the shutter open, a radiation monitor (Scintillometer – Model 3 by Ludlum Co.) must be used to scan all areas where the beam will pass to look for leakage. 3 4. Proper use of the Ludlum Model 3 Scintillometer will include: a. Use the mode switch to select Battery Test – observe that the battery voltage is adequate. b. Select the Sensitivity mode at the “x1” position and be sure that the Audio feature is ON – to preclude having to watch the meter while moving the probe around the area. c. Normal background radiation readings will fall 200–400 Counts (beeps) Per Minute (CPM) Readings substantially above background are reported to qualified service personnel before continuing with instrument use. 4 in the range. to be Sample Preparation – Sample Carrier: Back Loading Method for Powdered Samples: 1. Materials Needed: a. Tape – masking or transparent, XRG sample holders, razor blade, glass slide (frosted), mortar and pestle, No. 150 mesh or equivalent sieve. 2. Procedure: a. Grind the sample to a fine powder and sift it through the sieve. b. Remove the metal backing plate from a XRG sample holder. c. Tape a glass slide (frosted is better) onto the front of the sample holder. 5 d. Load sifted sample into the open back of the holder. e. Use a razor blade to pack the sample into the holder and remove any excess. f. Snap on the backing plate and press down. 6 g. Turn the sample holder over and press down on the glass slide. h. Slide the glass plate off the sample /holder and hold it upside down to ensure the stability of the packing. i. Ready for insertion into the XRG Goniometer for analysis. (See section on Running Samples) 7 Operation of the X-Ray Generator and Goniometer Instrumentation Preliminary Checks: 1. Be sure the water pressure gages for water cooling system for the X-ray generator reads in the range of 40- 65 PSI. Check the gauges behind the X-ray generator – before (inlet) and after (outlet) the filter. A pressure greater than ~30 PSI is necessary for generator operation. a. Inlet and outlet valves handles are to be turned to a position parallel to the piping to permit water flow. With the X-Ray beam ON, if the input pressure is 10 PSI (or more) than the outlet pressure, the water filter probably needs to be changed. b. Note that generator interlocks measuring water flow will turn off the X-Rays, if the pressure falls below ~25 PSI. 2. Observe the setting of the Goniometer’s 2ϴ angle indicator to ensure that it reads +2.00 (degrees) or greater. Observe that the 2ϴ angle displayed is 40.00 degrees. a. Operating Less than 2 will increase the possibility of a direct x-ray beam impinging onto the detector – this may well damage the scintillator and/or photomultiplier detector. [NOTE: In an emergency, the Goniometer scan motor can be stopped immediately by shutting off the Data Box power switch.] b. If needed, engage the adjustment handle (push in) and manually adjust the 2ϴ indicator dial to a suitable starting angle. The Data Box Power Switch must be OFF to enable manual 2ϴ adjustment. 8 Turn On the AC Line Power: 1. Turn ON the main power for the X-ray generator and the power supply that services the X-ray signal processing electronics, the Goniometer angle controller (Databox) and the Goniometer stepping motor by pushing the handle up on the grey switch box mounted on the east wall. a. This enables AC power to the X-ray generator and powers a multi-voltage power supply, located in the lower section of the electronics cabinet, which powers the NIM Bin (holding the electronics signal modules and the DATABOX NIM module). Turn on the X-RAY Generator: 1. Be sure that the KV is set to 25 (minimum setting) and the MA to 10 (starting). 2. Turn on the Power key switch (clockwise) on the left side of the generator control panel and notice that the red X-RAYS OFF light comes ON. 9 3. Notice that none of the Overload or Fault lamps will have come on. However, if one or more does come on, try restarting (step 2). If unsuccessful, call a maintenance technician for help. 4. Be sure that the Current Control Switch (at the right end of the panel) is set to OPERATE (not to Minimum MA) and use the DISPLAY button switch to set the Panel Meter display to KV. . 5. Be sure that the sample compartment cover is on and to lock it in place and that the mechanical beam shutter has enabled the beam shutter – an audible click will be the cover is rotated. turned CW mechanism noticed as 5. Push the X-RAYS ON button (the second switch to the right of the key switch) and hold until the KV panel meter reading stabilizes at 25 KV. Observe that the yellow X_RAYS ON lamp sitting on top of the generator has turned ON. Know that, if the light bulb inside of the yellow shroud burns out, the beam shutter will not open. 10 Obser ON sa the ge Know of the beam 6. Bring up the KV (high voltage) SLOWLY (5 KV steps) to 45 KV maximum (for a copper, (CU), anode x-ray tube) by pushing the vertical stepping switch down and waiting for the Panel Meter to stabilize after each increment before proceeding. 7. Toggle the DISPLAY switch to indicate MA (Milliamperes) and use stepping switch to SLOWLY increase the current in one milliampere toggle both digit switches at once in going from 19 to 20 MA. Do not MA for a Cu x-ray tube. a. Using less than 20 MA may prolong the life of the x-ray tube, reduce the intensity of the X-Ray beam and diffraction peak the MA steps – exceed 20 but will heights.) Observe that the Shutter Light (orange plastic cover) at the output of the X-ray tube is OFF – indicating that the beam shutter remains Closed – though enabled. 8. Push the Shutter Open button on the right side of the Generator Panel. 9. Observe that the orange X-Ray beam light at the X-Ray source comes ON and know that an X-Ray beam is now passing into the sample compartment. 10. Refer to the Safety Precautions section in this manual – specifically part 3 and use the X-Ray beam detection meter / Scintillometer to check for radiation leaks before beginning a series of sample runs. This procedure completes the steps for start-up of the X-Ray Generator and attainment of beam. Next - go to: Running Samples, Goniometer 2 Calibration or Generator ShutDown Procedure. 11 Startup Procedure for DATABOX - Computer Control and Data Acquisition 1. Turn ON the power to the Computer, if it is not already ON. a. NOTE: The current computer in (9/2010) must be left ON at all to ensure proper startup communication with the DataBox module. If the computer power is to be OFF, communication with the DataBox can be established by following the instructions for “Restarting the communications between the DataBox and the Computer” – see below: use times found link 2. Turn ON the Monitor and the Printer. a. Note, that if the computer, monitor and printer are all that will be used to view or manipulate data already taken, bypass steps 3 & 4. 3. Turn ON the AC Line power at the breaker box which will turn on the NIM BIN power supply – the bottom unit in the relay rack cabinet. 4. Turn ON (UP) the NIM BIN signal electronics - Switch #1 (at the lower of the NIM BIN) and the power to the DATABOX – Switch #2 (on the – which also powers the Goniometer motor used to change 2 . left end Databox) stepper 5. Restarting the communications link between the DataBox and the Computer. If the Databox program will not communicate with the DataBox, this will likely be the result of having turned the computer off after its last use and communication between the DATABOX and the Computer has been lost. Go to the device manager on the computer (via: Start, Settings, Control Panel, System, double click and then to Device Manager) and delete both COM1 and COM2 ports. Then reboot the computer after being sure that the DataBox power is ON. The computer should automatically find the appropriate port and restore the link. 12 System – Operational Verification 1. Enter the Databox program using the computer cursor. (Note: The Databox NIM module must be ON at this point.) 2. Set Steps/Degree – Type “S” (if not on screen) and enter 200 (a “normal” selection). Type – RTN (Enter) 3. Determine the 2 angle as displayed on the Goniometer counter and dial – to the nearest degrees. (The 2 angle in the photo now 0.005 reads 40.00 degrees.) a. Notice that the counter display will “turn early as the dial is rotated in the CW direction; and may cause confusion as to actual 2 setting. For example: if the is 4.35 degrees, this is the correct angle. reading is 5.85, the angle is actually 4.85 5.xx has turned over early). over” the reading If the (the 4. Enter the Starting Angle – “A” and enter the 2 angle as read. (Note: following zeros need not be entered.) Do not set 2 at less than 2 or more than 50 degrees! Type – RTN 5. Enter the 2 angle at “G” The angle to which you want the Goniometer to GO (usually the starting angle for a run.) Do not set 2 at less than 2 or more than 50 degrees! Type – RTN (The stepper motor on the Goniometer should drive the detector to the selected 2 angle and stop precisely at that destination. If true, angle drive calibration is verified. a. For an Emergency Shutdown - If there is a malfunction or an erroneous 2 has been entered, it is possible that the X-ray detector or the Goniometer be damaged. Stop the slewing Goniometer motor at by turning off the Databox power switch - #2. 6. Enter a Title – at the “T” prompt. The Title may be as and descriptive as needed. 7. Type “E” for Enter and then RTN to go to the Edit Mode. 13 may once long RUNNING SAMPLES: 1. Close the Beam Shutter – using the white pushbutton on the right side of the generator panel. [photo 12] Notice that the orange beam shutter light is OFF and know that the X-Ray beam is OFF. 2. Open the Sample Compartment cover, release friction lock by turning the small knurled knob degrees CCW, then turn the cover CCW until mechanical shutter is released and pull the cover you. {This mechanical shutter lockout is a secondary mechanism to ensure that the x-ray blocked from entering the sample compartment.} the ~180 the toward beam is 3. Remove any sample holder in the chamber by pushing down on the spring clip with a spatula or similar tool and remove the sample holder. With a damp tissue wipe out any residual sample powder from the chamber. 4. Insert a new sample/holder using the reverse procedure above – being sure that the sample is centered on the holder pedestal. Release the spring clip carefully and replace the chamber cover with a CW rotation and lock it in place with modest CW tension on the locking screw. 14 5. Set Experimental Parameters: a. Type I (Initial 2) and enter the Initial/Starting 2 degrees for the run. RTN. Note: One can scan a desired range of angles beginning either at the highest or the lowest. b. Type F (Final 2) and enter the final/ending angle for the run. RTN. c. Type S, then enter the Step Size (see table below), then RTN. d. Type C, then enter the Count Time for each successive angle, then RTN. 2 STEP Size (Degrees) Count Time (seconds) Degrees per Minute .03 1 ~2 .06 2 ~2 .01 1 ~ 0.6 .02 2 ~ 0.6 .02 1 ~ 1.2 .10 1 ~5 .01 6 0.1 .03 / 1 is the default step / count parameter. e. Type E, then RTN to go back to the monitor mode. f. Type S to start the run. When prompted for a Title, enter a new title, then RTN or press RTN for the previously entered title. The title will display and data collection will begin. For a real-time display, press F7 (Press ESC to exit real-time). To change display parameters, press PGUP. g. When Data collection is complete, press F8 to quick dump the data into MDI format (this is an ASCII format). The file name for this format must be short and without “_” or spaces. No suffix is needed. With certain runs (multiple sequences) the quick dump command will give an error message. For this situation, use F4, then RTN to download the data (use the default file name supplied, since it is only a temporary assignment) then proceed to save it in the MDI format by pressing F8 and your own short and unique file name. Press ALT-F7 to retrieve the dump as a display, if desired. 15 h. To run additional samples, with the same parameters - Close the shutter, Replace the sample, Open the shutter, type T, then enter a new title, then RTN and Press S and RTN to start the run. i. To run additional samples with different parameters, Press E (Edit) then RTN and Press P (for previous session) to go back to step 5 above to enter the new parameter(s). To Shut Down the XRG System: 1. Close the Shutter and Remove the Sample – replace the sample compartment cover. 2. Reduce the HV current in one mA steps to 10 mA (starting setting). 3. Toggle the digital display selector to read KV and bring down the voltage in one KV steps to 25 kV (the minimum setting). 4. Push the X-RAYS OFF button. 5. Turn off the Generator with the Key Switch. 6. Turn off the Databox NIM module with switch #2. 7. Turn off the NIM Bin electronics rack with switch #1. 8. Turn off the main AC Line power switch box on the east wall. 9. Fill out the System Log Book noting the hours used, date, operator and enter any malfunctions noted in the Maintenance Notes section, etc. 16 Presentation of Results Using JADE Software JADE (program) COMPUTER PROTOCOLS For use with XRG Instrumentation 1. Opening files. a. To open a file, go to File -> Open. i. Files are automatically placed in the “Databox” file when saved out of Databox, they must be manually moved to other files (e.g. “Clay Lab”, etc) ii. After selecting a file, click on “New Plot” to open. iii. You may also choose to “overlay” one plot over another through this window. 2. Analyzing data a. Zoom window controls i. The zoom window is the window that opens automatically when a file is opened – it shows a graph of the XRD findings. ii. On the right side of the graph, there are several controls for the graph’s properties. 1. When the mouse is placed over these buttons, what they control will be shown on the top, blue, bar of Jade. Each of these buttons does at least two things – one for left-clicking, another for right-clicking. 2. The “Reset” button can only reset the screen back through one change, i.e., if you to a background elimination, then do a 2nd elimination, the reset will return you to the graph of the 1st elimination, not the original data. b. There is a data-edit horizontal toolbar located above the zoom window. i. The buttons are, left from right: 1. Cursor 2. Peak-edit I 3. Peak-edit II 4. Peak-paint 5. Background-edit 6. Data-edit 7. K-alpha2 edit 8. Profile fitting 9. Cubic indexing cursor c. Background-editing i. To filter out the background, click on either Analyze Fit Background, or left-click the “BG” button on the uppermost toolbar, or right-click on the Background-edit (BE) button. 1. By left-clicking on the “BE” button, you can manually fit a background curve. ii. Double-clicking on the “BG” button will re-fit the graph to correct for the background. 17 d. Peak list i. Jade can automatically find peaks in the data. To do this, press F2, or go to Analyze Find Peaks. ii. This will open the “Peak Search and Labeling Dialog,” which allows changing the search parameters, including filter length, and labeling options. Click “Find” for Jade to scan for peaks. iii. To read the peaks as a list, click View List Reports Peak List Report. 1. From the Peak List Report, you can delete peaks that are unnecessary or caused by background noise e. Comparing data with PDF files i. On the top toolbar, there is a space to type in either the mineral name (i.e. “QUARTZ”) or the mineral ID#. Press return, and a new toolbar will appear with the mineral name, and lines will appear on the graph corresponding to the mineral’s peaks. ii. You can then enter other minerals, to make multiple comparisons. 1. To keep a mineral’s peaks visible, you must “tag” it using the button that looks like a push-pin. Clicking on the ‘?” button will open a settings window. iii. If a mineral has been tagged, a printout will include that mineral’s peaks as a bar running beneath the graph of your sample’s data. iv. View List Reports Peak ID (Extended) or Peak ID (Compact) will show your data’s peaks with the corresponding mineral’s peaks and show the differences between them. f. Search/Match i. This allows for the computer to compare your data to its own database of minerals, and determine possible matches ii. Can be accessed either by clicking the “S/M” button or via the top toolbar. 1. Clicking on this will open a window with various setting that can be changed – since this computer has only one database of minerals, it is important not to change that setting. iii. After beginning the Search/Match run, a message may come up stating that the background noise must be eliminated, allow Jade to do this automatically. iv. The “hit list” will show various potential matches, and which peaks appear to correspond in the data. There are various filtering options you may use to alter the data. 18 g. Printing i. You can print the “zoom window” directly just by pressing the print button on the top of the screen. There are also print buttons for Peak List Report and Search/Match Report windows. Under the File button there is a Print Plot button which prints the current “zoom window” and a Print Report button that expands to who printable reports including the Peak List report which we usually print on the back of each printout. ii. File Print Setup allows you to make several changes to the page that will be printed. NOTE: 1. An important note: because of the printer’s settings, light colors do not print well. Once you are in print setup you can press the setup button and choose black and white under Plot Color Settings. 2. The buttons at the top of the print setup screen allow you to add arrows, shapes and text. For example to add text click the button all the way to the right with the “a and b” on it. Then click anywhere on the pattern and a text cursor will appear. Type the text you want to add and press enter. 3. There are five buttons on the bottom left that show small pictures of what the pattern printout will look like. Pushing these will change the style of the printout. Not all are always available. Some require stick bars from using the mineral identifying features and some require more than one pattern in a printout. a. The top button gives a standard printout. Any patterns and stick bars are printed in one band. b. The second button prints all patterns in a single band with a separate band on the bottom for stick bars. c. The third button is the same as the top button but splits the band into two sections. All patterns and stick bars are overlaid together. d. The fourth button prints each x-ray pattern in its own band with stick bars overlaying the pattern. e. The fifth button prints each x-ray pattern in its own band with a separate band at the bottom for stick bars. 19 Optimization of Goniometer and Signal Conditioning Parameters The procedures and parameter adjustments described in this section are normally left to an instrumentation specialist, electronics engineer and/or computer specialist, as applicable. Begin with the section on: Operation of the X-Ray Generator and Goniometer Instrumentation. Calibration of the 2 angle readout for the Goniometer: 1. Insert a known standard sample (-quartz or other standard desired) into the sample chamber (see section Running Samples - 1-4). 2. Note that the following NIM modules are in place and properly connected together – back panel: Detector Power Supply, Amplifier, Threshold and Window, Rate/Level Meter. The following parameter settings are those considered nominal. a. Detector Voltage – SCINTillation 0.6 KV + 50 Volts (FINE) KV – b. Amplifier Gain – x 128 ; Input (at bottom of module) – Scint. Select c. Baseline/Threshold – 0.45 ; Window - ~4/10 (not critical) d. Baseline Control – INTernal, INTEGrate e. Clock Signal - OFF– OFF Mode – f. Rate Meter Range – as needed (~ Time Constant – 1-2 seconds; %Supp. - ?, AUDio – ON or OFF desired. 10-5K); as 20 3. Start up the XRG System – see section Operation of the X-Ray Goniometer and Goniometer Instrumentation - through Startup Procedure for Databox – step 3. 4. Select two or three (better) major 2 diffraction peak angles listed in the literature for the calibration standard sample chosen. (For Alpha Quartz these are: 50.20, 39.52, 36.52, 26.68 and 20.85 degrees). and angle for the to 5. Turn the mechanical 2 angle counter dial vernier dial by hand to near the desired the first diffraction calibration peak. (As handle is rotated, push in the crank handle engage the vernier with the 2 gear train.) 6. Adjust the Amplifier Gain or the Rate Meter Range switch to obtain a working deflection of the Rate Meter and then manually rotate the vernier dial setting back and forth through the radiation peak (highest rate meter deflection. (Adjust the rate meter range and the time constant to obtain the highest possible meter deflection.) Note: The most accurate calibration procedure will result from approaching the final peak angle from the same direction (CW or CCW) as will be used to scan samples. This is because there is a small but finite back-lash in the gearing of the Goniometer – less than 0.02 deg. 7. Write down the 2 angle observed to the nearest 0.005 degree and compare this result with the literature value to determine the error and its sign. 8. In like manner, determine the magnitude and direction of the error for the other calibration peaks. 9. If the errors are found to be significant, mechanically adjust the vernier dial relative to the Goniometer gearing to minimize the net error by: a. Turn on the power to the Databox, which will of gear train. the b. Loosen the three screws in the face of the vernier and push in on the handle shaft until the friction between the shaft and the dial is broken. Rotate to the desired position and retighten the screws. dial fit the dial 10. Recheck the error at each of the calibration peaks. 21 Optimization of the Signal Conditioning Parameters: 1. Set the Voltage to the Detector: a photomultiplier tube that detects light pulses coming from a scintillator crystal as a function of X-Ray intensity. a. On the Scintillator NIM module, the SCINT. KV selector to 0 .6 and the vernier dial to a nominal Volts (0.50 of full scale). Settings to 1000 volts are possible, but the signal to noise ratio will fall near nominal setting of 650 Volts. adjust KV 50 of 450 best this b. Set the Gain of the Signal Amplifier Module (Nominal is x 128 to x 512). This module not substantially change the signal noise ratio as a function of gain. setting will to 2. Set the Baseline/Threshold dial to a level that exceeds the background noise from the photomultiplier and the previous signal conditioning modules – this setting will depend on the PM voltage and Gain settings. Nominally, a setting of 0.4 to 1 (full scale of 10.0) will work. If the Threshold is set too high, some of the weaker pulses from the diffracted x-ray beam will be rejected, too low and PM noise will dominate. (0.40 to 0.45 works well.) 3. Set the pulse height window to a value that will not exclude the strongest x-ray pulses, as detected by observing the rate meter. (Little effect will be noted if only diffracted X-Ray pulses are dominate.) 4. Adjust the Rate Meter Range and Time Constant settings to obtain sufficient pointer deflections to optimize signal levels. 5. Note that: The signal going to the Databox Module is derived from the output of the Baseline and Window Module and is not affected by any adjustments of the Rate Meter module. The Databox Input signal is applied at the back of the module and can be accessed (in parallel) at the front panel – marked Input. 22 XRG Maintenance / Scheduling: Daily: During each session of use – by the Operator. Each user should run a standard sample (Alpha Quartz) or a user selected standard similar to the series of samples being investigated. Prudently – run the standard first (and, perhaps, last). Any deviation from expected results or aberrant operation of the instrument should be reported in the separate Maintenance Logbook and/or directly to the Instrument Supervisor. Quarterly: Conducted by the Instrument Supervisor or a designee. Review all comments written by users in the XRG Operations Logbook and take action as necessary. Semi Annually: Conducted by the Instrument Supervisor or a designee. Run an Alpha Quartz Standard to confirm the 2 calibration of the Goniometer by comparing the results with previous data: angular and peak amplitude . The procedure and major useful diffraction angles are listed on Page 21, section 4. Check (and adjust) all electronic and mechanical aspects of machine operation – Lubrication of moving parts – Goniometer drive gearing; sample compartment loading mechanisms and mechanical shutter interlocks; cooling water filter and pressure; switches and lighted displays; standard settings of the signal conditioning electronics panels; Computer, Monitor and Recorder operation; JADE software functionality; 23 Philips XRG Instrument - Condensed Operating Instructions A. X-Ray Generator Start-Up: Reference to Manual - Page 1. Check cooling water pressure gages – normally 50-60 PSI 8 2. Mechanically and accurately set the Goniometer to a desired 2ϴ angle – greater than 2o 8 3. Check to see that the initial KV and MA settings are at 25 KV and 10 MA 9 4. Turn ON the AC Line power at the wall-mounted breaker box 9 5. Turn on the Generator using the Key Switch 9 6. Note that no overload or error lights come on 10 7. Set the Panel Meter Display modes to KV and Operate 10 8. Be sure that the sample compartment cover is on and rotated to enable the beam shutter 10 9. Push the X-Rays ON button and hold until the Panel Meter stabilizes at 25 KV 10 10. Did the X-Rays ON lamp sitting on top of the generator come ON? 10 11. Bring up the HV slowly to 45 KV in 5 KV steps. 11 12. Select MA with the Panel Meter Display push-button switch 11 13. Slowly Increase the MA from 10 to 20 (in 1 MA steps) using the MA rotary switch(s) 9 & 11 14. Push the Shutter Open switch and note that the orange “Shutter Open” lamp comes ON 15. Check for radiation leaks with the Scintillometer. 11 3, 4 & 11 16. Turn OFF the Shutter Open Switch and observe that the Orange lamp goes OFF 11 B. X-Ray Detector Instrumentation and Computer Start-Up: 1. Turn on the NIM BIN power switch (labeled #1) 12 2. Turn on the Computer, Monitor and Printer, if not already ON (the computer should be left ON) 12 3. Turn ON the Data Box NIM module using switch #2. The Goniometer should not be moving! 12 4. If the Goniometer stepper motor is running and 2ϴ is getting close to 2o or ≥ 50o, a. Immediately turn OFF the Data Box power with switch #2! Emergency Shut-Down! 5. Bring up the DataBox program on the computer. If the program does not work, see Restart. 12&13 24 12 C. Enter the DataBox Program to Verify Program Operation: 1. Type “S” to set the Steps / Degree - 200 is a nominal value – Type Return (RTN) /Enter 13 2. Type “A” to enter the Starting 2ϴ angle (least significant “0’s” need not be entered); a. Use the 2ϴ angle mechanically set in step A-2 above; but, not ≤ 2 or ≥50 degrees. 13 3. Type “G” (“Go To”) to enter any angle other than the Starting 2ϴ angle and type ENTER 13 a. The Goniometer should scan automatically to the “Go To” angle just entered. 4. Type “E” to exit to monitor. 5. Set Edit mode. D. Running Samples: 1. Close the X-ray beam Shutter with the white push button – the Orange Shutter light should go out 14 2. Open the Sample Compartment Cover and insert the sample carrier 14 3. Close the Sample Compartment Cover 14 4. Set the experimental program parameters using the DataBox program 15 5. Type “I” (Initial 2ϴ - Starting angle) and enter the desired angle – then type ENTER 15 6. Type “F” (Final 2ϴ angle desired) and type ENTER 15 7. Type “S” (Step Size) Type the angle desired – (Nominal value = 0.03o) – Type ENTER 15 8. Type “C” (Count Time) Type the time desired) – (Nominal 1 second) - Type ENTER 15 9. Type “E” (Edit Mode) Type ENTER to go back to the Monitor Mode. 15 10. Type “S” (Start ) to start the run. 11. When prompted for a Title, enter the Name of the run and then type ENTER to start the run. 15 12. For a real time display, Press F7 (ESC to exit real time) PGUP to change display parameters 15 13. When Data Collection is complete, Press F8 to dump the data – various options are available, see Operation Manual, section 5g, p. 15 15 14. To run additional samples - same parameters - Close Shutter, exchange sample, Open Shutter 16 15. Type “T” and enter a new title 16 and Type “S” and ENTER to start the next run. 16. For running different parameters, Press E (Edit) then ENTER; to go back to D4 25 Press P (Previous Session) 26