03. Shake the flask well to dissolve Potassium

advertisement

I.

PURPOSE:

To provide a procedure to operate and Calibrate the UV-VIS spectrophotometer ,

Model Shimadzu, UV1700 UV-VIS

II. SCOPE:

The procedure is applicable to the UV-VIS spectrophotometer, Model Shimadzu,

UV1700 UV-VIS being used for analytical purpose at the Quality Control

III. RESPONSIBILITY:

It is the responsibility of the Quality Control personnel involved in the

UV/Visible Spectrophotometer Calibration, to follow the procedure and schedule as mentioned in this Standard Operating Procedure.

IV. DOCUMENT REFERENCE:

SOPs : Core procedure: Calibration of Equipment

Forms : UV Visible Spectrophotometer Calibration Record

V.

OPERATION:

1. 0 PRELIMINARY CHECK:

1.1 Check and ensure that the instrument is clean and suitable for starting the operation. If it is not, clean and ensure the suitability.

2.0

BASIC OPERATION:

2.1 Put Power button `ON’ provided on left hand side Instrument.

2.2 Keep LCD unit in vertical position.

2.3 As soon as, power is kept ON, LCD will show, initialisation checking of different parts will start. Each test `PASS’ will be indicated by 

.

2.4

2.5

After every check is over & passed it will display mode.

When the power is turned `ON’ the spectrophotometer is checked and initialised. As each item enters it’s initialisation operation it is highlighted.

However, if any abnormality is detected, the initialisation process is interrupted at that item, without it’s star being highlighted.

2.6

UV

Wait for 15 minutes after automatic initialisation for optimum illumination of lamp.

3.0

3.1

3.2

BASIC MODE MENU:

Photometric :- Used to measure the ABSORBANCE or TRANSMITTANCE at desired wavelength.

Spectrum :- Used to scan the wavelength range to measure the absorbance or transmittance of range selected.

3.3 Quantitation :- Creates a calibration curve from a standard sample and quantitates an unknown sample.

3.4

Kinetics : - Used to measure the activity from the time dependent changes in absorbance

& % transmittance

3.5

Time scan: - Used to measure the Change in absorbance, transmittances or energy with

time in the fixed wavelength.

3.6 Multi-Component : - Used for multi - component determination.

3.7 Photometric (multi

) : To measured the absorbances or transmittances up to eight wavelength.

3.8 Optional program pack: Used for optional program pack.

3.9 Utilities: Used for setting and changing the basic operating parameters of the instrument.

D2 lamp off time should be .

On routine basis following modes are used with the given operations.

For measurement of Absorbance / Transmittance:

3.10 After initialisation of UV the basic mode menu will be displayed on the monitor. Press

‘1’ to go in the photometric mode.

3.11 Then press GOTO WL and give the required wavelength e.g. 277 nm & press ‘ENTER’ to get that wavelength.

3.12 Press F1 key to select Absorbance / Transmittance mode.

3.13 Fill the cuvettes with blank solution and place it in both the cell holders & press. Auto

Zero the current wavelength will automatically be set to `0’ Abs. (100% T)

3.14 Then remove front side cuvette, discard the blank solution. Rinse the cuvette with sample solution. Then fill the cuvette with sample solution & replace in the cell holder. The monitor will display the Absorbance of the sample solution at the wavelength displayed.

3.15 Then press ‘MODE’ to come back to the basic mode menu.

To get the spectrum in the desired range of wavelengths:

3.16 When in basic mode menu press ‘2’ to go in the spectrum mode. Then ‘6’ different

parameters will be displayed.

3.17 Press `1’ to select Absorbance / Transmittance / Energy mode.

3.18 Press ‘2’ & then give the higher wavelength of the desired range and press ‘ ENTER’.

Then give the lower wavelength & press ‘ENTER’.

3.19 Then press, ‘3’ and give higher value of Absorbance which is greater than the expected

Absorbance value & then press ‘ENTER’. Lower value should be kept zero only, so press ‘ENTER’.

3.20 The speed of the scanning can also be changed. Press ‘4’. Then select ‘1’, ‘2’,`3’`4’ and

`5’ for scanning with `Very Fast’, `Fast’, ‘Medium’, ‘Slow’ and `Very slow’ speed respectively.

3.21 Press `6’ to select display mode i.e. sequential or overlay.

3.22 Place the cells filled with blank solutions in both the cuvette holders and press ‘F1’ so that baseline correction will be done.

3.23 Rinse the cuvette with sample solution. Place the sample in the front side cuvette holder.

Then press ‘START’ key.

3.24 The scanning in the given wavelength range will be done.

3.25 To know the Absorbance value at a particular wavelength in the range given, make use of

‘<-----’ and ‘------> ‘ keys.

3.26 Once scanning is over. Press F2 and select 3 for peak detection and F4 for valley. Press graph to obtain scan with peak detection. Then press print key.

3.27 Press print key in each step to get required copy.

3.28 Then press ‘Return’ & ‘Mode’ to come back to the basic mode menu.

4.0 MAINTENANCE :

4.1 Do not keep cells above the outside body of the spectrophotometer. Clean the cells perfectly and wipe out sidewalls with the tissue paper in one direction only. Do not hold the cells with their optical surfaces between the fingers.

4.2

Press the keys and handle the instrument very carefully.

4.3 The extinction of the cells intended to contain the solution to be examined and the reference liquid when filled with the same solvent, should be identical. If this is not the case, an appropriate correction must be applied.. Cells should be cleaned and handled with great care.

4.4 When not in use, put off the power switch of the instrument and then the mains.

4.5 Check condition of Silica gel for effectiveness and replace once in a week (or) earlier if required.

4.6

Ensure that during spectrophotometer initialisation process spectrophotometer makes various checks with preset initialised parameters such as Detection of wavelength, lighting of WI lamp, optical alignment of D

2

lamp, Energy check of D

2

lamp, optical alignment of WI lamp, Energy check of WI lamp, Baseline correction etc. which were appropriately evaluated by the instrument with no error message. In case any abnormalities observed in the above parameters, contact service engineer.

4.7 SPECIFICATION OF CELLS:

In `Photometry mode ’ Select transmittance mode. Set the wavelength at 200 nm. Then press `autozero’ to get 100 % transmittance with air blank in both the cell holders. Then place the cell in the sample compartment filled with milli Q water using air blank and read the transmittance. Similarly repeat the operation for 220 nm and 240 nm. Repeat the test for second cell. The transmittance values should meet the requirement as given in the following table.

Note: Before adopting the new cells for laboratory analysis, check the cells using this procedure and qualify them.

Acceptance Criteria:

Wavelength

200 nm

% Transmittance

> 80 %

The tolerance on the path length of the cells used is +/- 0.005 cm

The tolerance for the absorbance between the two cells is 0.005 at each of the above nm.

220 nm

240 nm

> 80 %

> 80 %

VI.

CALIBRATION PROTOCOL:

Calibration frequency: Once in three months.

1.0

CONTROL OF WAVELENGTHS

1.1 Preparation of 1.4 M Perchloric acid solutions:

Take 11.5 mL of Perchloric acid (about 70%w/w strength) into a clean 100 mL volumetric flask. Dissolve and dilute to the volume with water

1.2 Preparation of Holmium per chlorate :

01. Weigh accurately about 400 mg Holmium oxide.

02. Transfer into a 10 mL volumetric flask containing 5 mL of 1.4 M Perchloric acid

solution.

03. Shake the solution well to dissolve Holmium oxide.

04. Make up the volume to the mark with 1.4 M Perchloric acid solution.

Procedure:

05. Take a pair of matched quartz cells having a path length of 1 cm.

06. Perform the base line correction with 1.4 M Perchloric acid solution from 200 to

600 nm.

07. Remove the cell from sample compartment.

08. Rinse the sample cell with Holmium perchlorate solution, fill the cell with the same, clean the smooth surfaces of the cell with tissue paper and place it in the sample compartment

Acceptance criteria:

Wavelength

241.15 nm

287.15 nm

361.50 nm

536. 30 nm

Maximum Tolerance

240.15 to 242 .15 nm

286.15 to 288.15 nm

360.50 to 362 .50 nm

533.30 to 539.30 nm

2.0 CONTROL OF ABSORBANCES:

2.1 Preparation of 0.005 M Sulphuric acid:

01.

02.

Take a clean 100 mL volumetric flask and containing 50 mL of water.

Cautiously add 2.8 mL Conc. Sulphuric acid to the above flask.

03. Dissolve and dilute to the volume with water.

04. Take 5 mL of the above solution into a 500 mL volumetric flask, dissolve and dilute to the volume with water.

Preparation of K

2

Cr

2

O

7

UV solution: and

01. Take about 200 mg of Analytical Reagent grade K

2

Cr

2

O

7

into a weighing bottle heat at 130 o

C in a laboratory oven for 2 hours.

02.

Cool it in a desiccator and weigh about 120mg in a 100 mL volumetric flask containing 50 mL of 0.005 M Sulfuric acid.

03. Shake the flask well to dissolve Potassium dichromate completely. Make up the volume with 0.005 M Sulfuric acid – Solution .1.

04.

Dilute 5mL of Solution .1.

to 100 mL with 0.005M Sulfuric acid –solution-2.

05.

Dilute 5mL of Solution .1.

to 10 mL with 0.005M Sulfuric acid solution-3.

2.2 Procedure:

01. Take a pair of matched quartz cells having a path length of 1 cm.

02. Perform base line correction with 0.005 M Sulphuric acid .

03. Remove the cell from sample compartment.

04. Rinse the sample cell with potassium dichromate UV solution-2, fill the cell with the

05. same, clean the smooth surfaces with tissue paper .

Place the cell in sample compartment, measure the absorbance at 235 nm, 257 nm,

313 nm and 350 nm. Calculate the specific absorbance using the following

equation.

Specific absorbance at x nm = Absorbance at x nm

Concentration of Potassium dichromate

06.

Perform base line correction with 0.005 M Sulphuric acid at about 430 nm.

07.

Remove the cell from sample compartment.

08.

Rinse the sample cell with potassium dichromate UV solution-3, fill the cell with the same, clean the smooth surfaces with tissue paper .

09.

Place the cell in sample compartment, measure the absorbance at 430 nm.

Calculate the specific absorbance using the following equation.

Specific absorbance at x nm = Absorbance at x nm

Concentration of Potassium dichromate

10. The specific absorbances against each wavelength should be as follows:

Wavelength (nm)

235

257

313

350

Specific absorbance

(1%, 1 cm)

124.5

144.5

48.6

107.3

Maximum tolerance

122.9 to 126.2

142.8 to 146.2

47.0 to 50.3

105.6 to 109.0

430 15.9 15.7 to 16.1

3.0

PHOTOMETRIC ACCURACY:

3.1

Preparation of Potassium nitrate stock solution (14.20 % w/v)

Weigh accurately about 14.2 g of Potassium nitrate in a 100 mL volumetric flask, dissolve

and dilute to volume with water.

3.2

Preparation of solution A (1.065% solution):

Pipette out 7.5 mL of the stock solution into a 100 mL volumetric flask, dissolve and

dilute to the volume with water.

3.3

Preparation of solution B (0.710% solution):

Pipette out 5.0 mL of the stock solution into a 100 mL volumetric flask, dissolve and

dilute to the volume with water.

3.4

Preparation of solution C (0.355% solution):

Pipette out 2.5 mL of the stock solution into a 100 mL volumetric flask, dissolve and

dilute to the volume with water.

3.5 Procedure:

Measure the absorbances of the above solutions at about 302 nm against water blank.

Check the absorbance of the above solutions at 302 nm using milli Q water as blank.

Record the absorbance at each wavelength in the Calibration Data Sheet.

Check the Absorbance for each solution against the acceptance criteria given below:

Calculate correlation coefficient using the absorbance of respective concentration.

3.6 Acceptance Criteria:

01. Solutions Absorbance

% w/v at 302 nm

1.065

0.710

0.355

0.751

0.500

0.250

Acceptance Criteria

± 0.010

0.741 to 0.761

0.490 to 0.510

0.240 to 0.260

02. Correlation coefficient: NLT 0.999.

4.0

LIMIT OF STRAY LIGHT:

4.1 Preparation of 1.2% W/V Potassium Chloride Solution:

01. Take about 3 gm of Potassium chloride dry at 250

O

C for about 1 hour.

02. Weigh accurately about 1.2 g of Potassium chloride.

03. Transfer into a 100 mL volumetric flask containing 75 mL water.

04. Shake the flask well to dissolve completely and make up to the volume with water.

4.2 Procedure:

01. Take a pair of matched quartz cells having a path length of 1 cm.

02. Perform the base line correction with water.

03. Remove the cell from sample compartment.

04. Rinse the cell in the potassium chloride solution, fill the cell with the same, clean the smooth surfaces with tissue paper and place it in sample compartment.

05. Measure the absorbance at 200 nm.

4.3 Acceptance Criteria: The absorbance should be more than 2.

5.0 RESOLUTION POWER:

5.1 Preparation of 0.020 % v/v of Toluene in Hexane solution: Dilute 2 mL of

Toluene to 100 mL with Hexane. Further dilute 1 mL of above solution to 100 mL with Hexane.

(Note: Use only spectroscopy grade Toluene and Hexane)

5.2 Record the spectrum in the range 260 to 275 nm of 0.020 % v/v Toluene in

Hexane

using Hexane in the reference cell.

5.3 the

Calculate the ratio of the absorbance at the maximum at about 269 nm to that at

minimum at about 266 nm.

Acceptance Criteria: Ratio is not less than 1.5

6.0

IO - LINE FLATNESS: a. With air blank in the sample and reference compartment, perform the test.

b. Scan in the wavelength range 220 to 700 nm. c. Obtain the spectrum in the % transmittance mode.

Acceptance Criteria: The Io line obtained should not deviate from a horizontal line at any point by more than 99.54 - 100.46 % transmittance.

Download