LHC Project Note XXX 2010-06-30 Per.Hagen@cern.ch Magnetic model of MQWA and MQWB P. Hagen for the FiDeL team CERN, Technology Department Keywords: Normal-conducting Magnets, Magnetic Field Model, Harmonics, LHC. 1. Introduction Function in the machine: The resistive, water-cooled MQW magnets are used together with the MQTLH magnets (Q6) to provide the optics used in IR3 and IR7, where collimators are installed to clean the beam. In IR3 the off-momentum particles are removed by using high dispersion and low beta functions. In IR7 the beam halo from drifting particles are “shaved off” by using high beta and low dispersion functions. Undesired particle showers are created in the collimation process. Normal-conducting magnets are therefore used instead of superconducting magnets to avoid the issue of particle showers from the collimators inducing quench. The MQW quadrupoles have twin 46 mm apertures. The unusual choice of common iron yoke is due to space constraints in the tunnel. The trade-off is that the two apertures must be powered in series. That is, the modulus of the main field must be the same in both apertures to achieve reasonable field quality. The nominal operational current for MQWA is defined as 710 A and 600 A for MQWB. The nominal gradients are 35 T/m and ~30 T/m, respectively [1-4]. Fig. 1: MQW cross-section, drawing (right) and picture (left). Numbers and variants: The difference between MQWA and MQWB is only the way the current flows through the two apertures (polarity). In the symmetric “A” mode they work like the lattice quadrupoles, i.e., “F/D” or “D/F” (if beam 1 is horizontally focused, then beam 2 is horizontally defocused, and vice versa). In the anti-symmetric “B” mode the apertures are This is an internal CERN publication and does not necessarily reflect the views of the LHC project management. “F/F” or “D/D” which means both beams sees either focus or defocus in a given plane. Six MQW quadrupole magnets constitute optical element Q4 or Q5, located on both sides of each IR. Among the six magnets, 5 operate in A mode, and one magnet operates in B mode. Both MQWA and MQWB are used to provide the nominal optics, see Figs. 2 and 3. Fig. 2: Optical functions in IR3 for beam 1, x-plane (V6.503). Fig. 3: Optical functions in IR7 for beam 1, x-plane (V6.503). An optical requirement is that both beams see the same sequence of focusing and defocusing fields in the insertion region. The “A” magnets are part of a regular FODO structure which has an antisymmetric constraint on their strengths (n = 3 or 4): πA (QnL ) = −πA (QnR ) The “B” magnets introduce extra symmetric focusing in around the IP: πB (QnL ) = πB (QnR ) The number of MQW in LHC is 2 IR × 2 sides of IR × 2 optical quadrupole functions × 6 magnets = 48 magnets. 48 + 4 spare magnets were produced. One has 48 x 5/6 = 40 MQWA and 48 x 1/6 = 8 MQWB. Naming convention: MQW magnets are identified by consecutive production numbers 1 to 55. The magnets numbers 2, 3 and 4 do not exist. The MTF naming scheme is -2- HCMQWm_001-AM0000nn where m is the operational mode (A or B), nn is the magnet number. AM is manufacturer code for Alstom Canada. Expected operational cycles, range of current and operational temperature: The injection current for MQWA is in the range of 35-38 A, corresponding to a gradient range of 1.9-2.0 T/m (see Table I). The MQWA inside Q4 on the left side of IR are powered in series with the Q4 right side. The same is true for Q5. The optical strength scales linearly with the particle momentum during ramp (as for all magnets). The gradient does not change once collision energy has been reached. That is, the optical functions remain constant during the LHC squeeze. Table I shows the scenario for a few selected energy levels. The values are based upon LHC optics V6.503 and on the revised FiDeL parameters for 2010. The LHC design optics assigns little strength to the MQWB, but each one has an individual power supply although the actual design uses symmetry around IP. The strength has not changed since LHC optics V6.500. The injection current is as low as 2 A for MQWB inside Q5.L7 and Q5.R7 which gives a gradient of 0.05 T/m. The operational range for the other MQWB is in the range 12-30 A with a gradient range of 0.5-1.5 T/m. Table II shows the scenario for a few selected energy levels. The values are based upon LHC optics V6.503 and on the revised FiDeL parameters for 2010. Table I: Operational currents and gradients for MQWA circuits based on FiDeL 2010 parameters. CIRCUIT RQ5.LR3 RQ4.LR3 RQ5.LR7 RQ4.LR7 E = 450 GeV E = 3.5 TeV E = 5 TeV E = 6 TeV E = 7 TeV I (A) G (T/m) I (A) G (T/m) I (A) G (T/m) I (A) G (T/m) I (A) G (T/m) 36.73 1.96 291.98 15.22 417.31 21.75 502.10 26.10 593.19 30.45 34.84 1.86 277.91 14.49 397.13 20.70 477.31 24.84 561.05 28.98 37.67 2.00 299.06 15.59 427.49 22.27 514.74 26.73 610.33 31.18 37.01 1.97 294.11 15.34 420.37 21.91 505.89 26.29 598.26 30.68 Table II: Operational currents and gradients for MQWB circuits based on FiDeL 2010 parameters. CIRCUIT RQT5.L3 RQT4.L3 RQT4.R3 RQT5.R3 RQT5.L7 RQT4.L7 RQT4.R7 RQT5.R7 E = 450 GeV E = 3.5 TeV E = 5 TeV E = 6 TeV E = 7 TeV I (A) G (T/m) I (A) G (T/m) I (A) G (T/m) I (A) G (T/m) I (A) G (T/m) -30.24 -1.46 -221.28 -11.35 -315.29 -16.21 -377.96 -19.46 -440.64 -22.70 22.04 1.03 157.35 8.04 223.94 11.49 268.34 13.78 312.75 16.08 -22.04 -1.03 -157.35 -8.04 -223.94 -11.49 -268.34 -13.78 -312.75 -16.08 30.24 1.46 221.28 11.35 315.29 16.21 377.96 19.46 440.64 22.70 2.11 0.05 9.50 0.38 12.63 0.54 14.73 0.65 16.82 0.76 11.74 0.50 76.82 3.87 108.88 5.53 130.26 6.64 151.63 7.74 -11.74 -0.50 -76.82 -3.87 -108.88 -5.53 -130.26 -6.64 -151.63 -7.74 -2.11 -0.05 -9.50 -0.38 -12.63 -0.54 -14.73 -0.65 -16.82 -0.76 Summary of manufacturing parameters: the MQW have been designed and produced in collaboration with TRIUMF, Canada [5]. They were built by Alstom, Canada. The series production took place in the period 2002-2003. All magnets were initially measured at CERN using rotating coils. Extended measurements were done in 2010 to characterise the transfer function. The main parameters are summarized in Table III. The design values are shown in () parenthesis if different. -3- Table II: Main parameters of MQW based upon measurements, design values within () if different. Magnet type Magnetic length Beam separation (mm) Aperture Operating temperature Nominal gradient Nominal current Resistance (both apertures) Inductance (both apertures) Power dissipation at Inom MQWA MQWB (m) 3.12 (3.108) 3.13 (3.108) 224 224 (mm) 46 46 (C) < 65 < 65 (T/m) 34.9 (35) 30.7 (30) (A) 710 600 (mohm) 37 37 (mH) 64 (28+28) 64 (28+28) (kW) 19 13 (14) 2. Layout Slots and positions: the 40 MQWA and 8 MQWB are located in IR3 and IR7 according to Table III and Figs. 4 and 5, which refer to the installation in 2008. In addition, there are 4 spares, i.e., magnets 5, 10, 38 and 51. Magnets 2, 3 and 4 do not exist. The field measurements have been done with longitudinal axis CS (connection side) to NCS (non connection side). Many of the MQW magnets on the left side of “IP” have been vertically ο°-rotated in the tunnel w. r. t. this orientation. The reason for the rotation is to better shield the power cables from radiation coming from the collimators. The transformation of harmonics when doing a vertical rotation is described in Appendix of [6]. For the normal MQW quadrupoles the equations become: π ′ n = (−1)n−2 πn π′ n = (−1)n−1 πn These equations are valid for both apertures as they have a common reference system. Care must be taken when using the data in MAD as there are several definitions of the coordinate system for beam 2 [7]. Circuits: The two apertures in a given magnet are connected in series. All 5 A magnets in one Q4 or Q5 quadrupole are connected in series. In addition, the Q4 on left side of IR is connected in series with Q4 on right side. The same holds for Q5. This minimises the number of power supplies for the A magnets, taking into account that the gradient is asymmetric (odd function) mirrored around the IR. In total one has 4 circuits for the A type. The “B” magnets have individual power supplies, for a total of 8 circuits. -4- IR3 MOMENTUM CLEANING Q5.L3 A A Q4.L3 A v B A A ο° ο° ο° ο° A A A B A A ο° ο° ο° ο° ο° A A ο° ο° Q4.R3 A A B A Q5.R3 A A A A B A ο° Fig. 4: Block diagram of IR3 showing magnet grouping and vertical rotations. IR7 BETATRON CLEANING Q5.L7 Q4.L7 A A A v B A A A A A B A A ο° ο° ο° ο° ο° ο° ο° ο° ο° ο° ο° A A ο° Q4.R7 A A B A Q5.R7 A A A A B A Fig. 5: Block diagram of IR7 showing magnet grouping and vertical rotations. -5- Table III: MQW slot allocation, position (s) in beam 1 direction and vertical rotation. SLOT MQWA.E5L3 MQWA.D5L3 MQWA.C5L3 MQWB.5L3 MQWA.B5L3 MQWA.A5L3 MQWA.E4L3 MQWA.D4L3 MQWA.C4L3 MQWB.4L3 MQWA.B4L3 MQWA.A4L3 MQWA.A4R3 MQWA.B4R3 MQWB.4R3 MQWA.C4R3 MQWA.D4R3 MQWA.E4R3 MQWA.A5R3 MQWA.B5R3 MQWB.5R3 MQWA.C5R3 MQWA.D5R3 MQWA.E5R3 MQWA.E5L7 MQWA.D5L7 MQWA.C5L7 MQWB.5L7 MQWA.B5L7 MQWA.A5L7 MQWA.E4L7 MQWA.D4L7 MQWA.C4L7 MQWB.4L7 MQWA.B4L7 MQWA.A4L7 MQWA.A4R7 MQWA.B4R7 MQWB.4R7 MQWA.C4R7 MQWA.D4R7 MQWA.E4R7 MQWA.A5R7 MQWA.B5R7 MQWB.5R7 MQWA.C5R7 MQWA.D5R7 MQWA.E5R7 S 6514.1728 6517.9728 6526.0128 6529.8128 6533.6128 6537.4128 6616.8458 6624.8858 6628.6858 6632.4858 6636.2858 6640.0858 6689.3558 6693.1558 6696.9558 6700.7558 6704.5558 6712.5958 6792.0288 6795.8288 6799.6288 6803.4288 6811.4688 6815.2688 19851.7964 19855.5964 19859.3964 19863.1964 19866.9964 19870.7964 19910.0864 19913.8864 19922.5864 19926.3864 19930.1864 19933.9864 20054.3384 20058.1384 20061.9384 20065.7384 20074.4384 20078.2384 20117.5284 20121.3284 20125.1284 20128.9284 20132.7284 20136.5284 MAGNET HCMQWA_001-AM000055 HCMQWA_001-AM000012 HCMQWA_001-AM000049 HCMQWB_001-AM000036 HCMQWA_001-AM000022 HCMQWA_001-AM000013 HCMQWA_001-AM000033 HCMQWA_001-AM000039 HCMQWA_001-AM000008 HCMQWB_001-AM000050 HCMQWA_001-AM000042 HCMQWA_001-AM000023 HCMQWA_001-AM000021 HCMQWA_001-AM000032 HCMQWB_001-AM000009 HCMQWA_001-AM000024 HCMQWA_001-AM000011 HCMQWA_001-AM000043 HCMQWA_001-AM000048 HCMQWA_001-AM000034 HCMQWB_001-AM000035 HCMQWA_001-AM000053 HCMQWA_001-AM000014 HCMQWA_001-AM000017 HCMQWA_001-AM000001 HCMQWA_001-AM000028 HCMQWA_001-AM000030 HCMQWB_001-AM000046 HCMQWA_001-AM000047 HCMQWA_001-AM000045 HCMQWA_001-AM000027 HCMQWA_001-AM000054 HCMQWA_001-AM000007 HCMQWB_001-AM000006 HCMQWA_001-AM000040 HCMQWA_001-AM000016 HCMQWA_001-AM000019 HCMQWA_001-AM000029 HCMQWB_001-AM000052 HCMQWA_001-AM000041 HCMQWA_001-AM000037 HCMQWA_001-AM000015 HCMQWA_001-AM000020 HCMQWA_001-AM000018 HCMQWB_001-AM000025 HCMQWA_001-AM000031 HCMQWA_001-AM000044 HCMQWA_001-AM000026 -6- V ROTATED yes yes yes yes yes yes yes yes yes yes yes yes yes yes yes yes yes yes yes yes yes yes yes yes 3. Series measurements All the MQW magnets were measured in CERN after delivery. Each magnet was measured both in A and B mode. The measurements are described in detail in [8-10]. Device: Measurements are done with a 750-mm-long mole having 5 rotating coils, 41 mm diameter, inserted into a guiding tube of stainless steel. Measurements were done at 5 positions to cover the integral. Repeated local measurements were done at position 3 to find the B versus I curves. The pre-cycle done to stabilize the magnet is shown in Fig. 6. It is the same for both “A” and “B” mode. Fig 7 shows the demagnetisation cycle which was performed in-between changing “A/B” modes.. 900 800 700 Current (A) 600 500 400 300 200 100 0 1 6 11 16 21 26 Time (s) Fig. 6: MQW pre-cycle where Imin=20A, Imax=810A, dI/dt=100A/s, 10s wait time between ramps, I = 0 at end of pre-cycle when measurement cycle starts. 600 400 Current (A) 200 0 0 20 40 60 80 100 120 140 160 180 -200 -400 -600 -800 Time (s) Fig. 7: MQW demagnetisation-cycle where Imin / Imax = -710, 500, -300, 120, -40, 12, -30A, dI/dt=100A/s, 10s wait time between ramps, I = 0 at end of demagnetisation cycle. -7- The precision in the transfer function measurements were estimated to 20 units at 40 A and 10 units for nominal current, the precision on b3 to 1 unit, and 10% accuracy for the other harmonics. The measurement sequence changed during production due to issues with reproducibility. At the end of the production all measurements were normalised to the cycles in Figs. 6-7. Available and missing measurements: Two types of measurements: local and integral. Type A: the integral measurements were done for 40, 200, 710 and 810 A. Magnet 10 (a spare) was not measured until 2009. The local measurements of position 3 have more currents: 30, 40, 100, 150, 200, 300, 400, 500, 600, 650, 710, 750 and 810 A. All A type magnets have local measurements. Type B: Integral measurements were done for: 40, 200 and 600 A. Magnet 40 was not measured and operates in “A” mode. The local measurements in position 3 were done with the currents 30, 40, 200 and 600 A, and repeated several times. All magnets except magnet 40 and magnet 5 were measured. Unfortunately measurements for 30 A is missing for magnet 36 and 51. They operate in “B” mode in LHC. All data has been normalised w. r. t. feed-down setting to zero ninth order harmonics. Rejected or faulty measurements: Local measurement of magnet 5 in B mode has been rejected during data processing. Use of the measurements in FiDeL: The geometric is computed through the integral measurements. The local measurements are used to obtain the saturation and residual magnetization parameters. The local and integral measurements are normalized to the geometric and cross-checked to have an estimate of the precision of the measuring system. 4. MQWA transfer function based upon series measurements 4.1 INITIAL DECISIONS Transfer function, ap1 + ap2 (T/A) Fig 8 shows the transfer function for all local measurements. There is a systematic difference between the apertures. Since the two apertures are connected in series, we decided to process them together, i.e., to use the average transfer function of local measurements for fitting rather than individual magnets. 0.00092 0.00090 0.00088 0.00086 0.00084 0.00082 0.00080 0.00078 0.00076 0 200 400 Current (A) 600 800 Fig. 8: TF for local measurements (both apertures plus dotted line for average) -8- ΔTransfer function, ap2-ap1 (units) 0.0 -5.0 -10.0 -15.0 -20.0 -25.0 -30.0 0 200 400 Current (A) 600 800 Fig. 9: Difference between TF in the aperture 1, and TF in aperture 2, in units. TF ap1 = 10 000 units 4.2 GEOMETRIC Fig. 10 shows the integral measurements for all magnets. We use the integral measurement at 200 A to define the geometric component, see Table IV. We use a different geometric for each circuit. Since the difference between each circuit is within 2 units we could have used the same value for all. 0.00285 Integrated Transfer function (T.m/A) ap1 + ap2 0.00280 0.00275 0.00270 0.00265 0.00260 0.00255 0.00250 0.00245 0.00240 0.00235 0 200 400 600 Current (A) Fig. 10: TF for integral measurements (both apertures) -9- 800 Table IV: TF geometric, average and spread for the MQWA circuits CIRCUIT RQ5.LR3 RQ4.LR3 RQ4.LR7 RQ5.LR7 TF (T.m/A) TF (units) 0.00275454 -1.1 0.00275492 0.3 0.00275534 1.8 0.00275452 -1.1 Average Stdev 0.00275483 1.4 The agreement between the local and integral measurements is shown in Fig. 11 and in Tables V and VI. Data are normalized, in units, using the geometric values at 200 A as reference. Local measurements have a lower transfer function at high currents of about 50-70 units w.r.t. integral measurements. Measurements at 40 A agree within 3 units, and match at 200 A by definition. Integrated Transfer function (units) 400 200 0 -200 -400 -600 -800 -1000 -1200 -1400 0 200 400 600 800 Current (A) Fig. 11: Average of local TF (circles) and integral TF (triangles), expressed in units of geometric Table V: TF average and spread in local measurements of MQWA I (A) TF ave (units) TF sigma (units) 30 175 32 40 93 27 100 -13 15 150 -6 14 200 0 13 300 1 14 400 -14 14 500 -44 17 600 -142 19 650 -292 20 710 -577 20 750 -806 20 810 -1184 20 - 10 - Table VI: TF average and spread in integral measurements of MQWA I (A) TF ave (units) TF sigma (units) 40 96 24 200 0 10 710 -631 18 810 -1249 16 5. Extra measurements in 2010 The measurements of beta-beating in LHC convinced us that the MQWA transfer function was wrong by 1.5 to 2.0% at injection. The spare magnets 10 and 51 were measured during spring 2010. The measurement campaign is described in [11]. Device: Measurements were done using SSW (single stretch wire) since we only needed to correct the integral transfer function. The pre-cycle with 4 different Imin to characterise the transfer function (“a” to “d”) is shown in Fig. 12 and Table VII. The pre-cycle type “b” is the one actually used in LHC since the power converters are not stable at lower currents. 800.0 700.0 Current (A) 600.0 500.0 400.0 300.0 200.0 100.0 0.0 0.0 500.0 1000.0 1500.0 2000.0 2500.0 3000.0 Time (s) Fig. 12: MQWA pre-cycle where Imin=20.1A, Imax=710A, dI/dt=2A/s, 10s wait time between ramps. Table VII: MQWA pre-cycle parameters (4 different Imin) Pre-cycle type No of periods ramp rate (A/s) I_MIN (A) t_MIN (s) I_FLAT_TOP (A) t_FLAT_TOP (s) a b 4 2 20 10 710 10 - 11 - c 4 2 25 10 710 10 d 4 2 30 10 710 10 4 2 35 10 710 10 6. MQWA transfer function based upon extra 2010 measurements 6.1 INITIAL DECISIONS Table VIII shows that the new measurements of MQWA 51 are consistent with the series measurements. The average discrepancy is about 10 units. The measurement reproducibility is in the range 2 to 15 units with largest spread at injection and nominal currents. The MQWA 10 was not measured during series so it cannot be used for comparison. Table VIII: Comparison of MQWA 51 measurements. (SSW – RC) / RC [units] I a b 200 710 0 -6 Aperture 1 c 0 -6 d 0 -6 a 0 -6 b -21 -15 Aperture 2 c d -21 -15 -14 -14 -18 -15 Fig. 13 and 14 show the shape of the transfer function with the new pre-cycles. We can see that it has a significant Imin dependency for ramp-up. The ramp-down TF curves follow very much the same path (no Imin dependency). Table IX summaries it numerically by taking the average value of the 4 apertures. The residual magnetisation is much stronger than in the series measurement. We find that the residual magnetisation is in reality 2% stronger at 30A and 1% at 40A compared to series measurements. 800 a TF (units) 600 b 400 c 200 d 0 -200 -400 -600 -800 0 200 400 600 Current (A) Fig. 13: MQWA 10 TF (units) after removing geometric at 200A, pre-cycles: a,b,c,d. - 12 - 800 700 a TF (units) 600 b 500 c 400 d 300 200 100 0 -100 0 50 100 150 200 Current (A) Fig. 14: Zoom of Fig 13. for the current range 0-200A. Table IX: Average MQWA TF (units) after removing geometric at 200A, pre-cycles: a,b,c,d. The column “b sigma” is spread (units) for pre-cycle b. I a 30 35 38 40 70 100 200 400 500 561 610 710 b 139 115 87 69 -1 -13 0 -16 -48 -94 -187 -626 c 366 267 220 186 36 5 0 -17 -48 -94 -189 -627 d 573 382 319 281 56 12 0 -18 -50 -98 -190 -629 b sigma 507 402 354 74 16 0 -17 -48 -94 -193 -627 38 26 28 28 7 4 0 8 13 18 20 24 Table X and Fig. 15 show the new synthesised transfer function. We have used the geometric component at 200A from the series integral measurements as well for the saturation. We have measurement points for the residual magnetisation and the beginning of the saturation. Table X: Synthesised MQWA TF average (units) and spread (units) after removing geometric at 200A. I TF ave TF sigma source 30 366 38 SSW 2010 35 267 26 SSW 2010 38 220 28 SSW 2010 40 186 28 SSW 2010 70 36 7 SSW 2010 200 0 0 all by def 300 0 14 RC ORIG 400 -17 8 SSW 2010 500 -48 13 SSW 2010 561 -94 18 SSW 2010 710 -631 18 RC ORIG 810 -1249 17 RC ORIG - 13 - why pre-cycle pre-cycle pre-cycle pre-cycle pre-cycle normalisation all magnets extra point extra point extra point all magnets all magnets 600 400 200 TF (units) 0 -200 -400 -600 -800 -1000 -1200 -1400 0 200 400 600 800 Current (A) Fig. 15: MQWA FiDeL 2010 TF. The encircled measurement points encircled are from the 2010 measurements (average of 2 magnets). The reminding are from the series measurements (average of all magnets) 6.2 GEOMETRIC The FiDeL geometric values at 200A from the series measurement are used. See 4.2. 6.3 SATURATION AND RESIDUAL MAGNETIZATION COMPONENTS We use the two components saturation and residual magnetization to fit the TF curve in Fig 15. There are no DC magnetization components for normal-conducting magnets, neither time-dependent components, and there is no beam screen in the bore. The fits are done according to the equations in [12]: ππΉ = πΎ πππ + ππΉ π ππ‘ + ππΉ πππ πππ’ππ ππΉ π ππ‘ = − ππΉ ππ πΌ − πΌ0 [1 + πππ (π )] 2 πΌπππ πππ πππ’ππ πΌπππ ππ = ππ ( ) |πΌ| Saturation: the fit is based on the average of the available measurements in Table X. We use the integral measurements from the series measurements (“RC ORIG” in Table X) and a few extra points from the 2010 SSW measurements. The fit is illustrated in Fig. 16. We have also validated the measurements against simulations with a ROXIE [13] model of the MQWA. The agreement is very good, with less than 50 units difference at 810 A. ROXIE computations relate to 2D cross-section while the integral measurements include the heads (end effects). - 14 - 600 400 200 TF (units) 0 -200 -400 -600 -800 -1000 -1200 -1400 0 200 400 600 800 Current (A) Fig. 16: Fit of the TF saturation (units), line = fit, triangles = measurements, diamonds = ROXIE simulation. Residual: the fit is based exclusively on 2010 SSW measurements in Table X. Only currents equal or less than 200 A are considered. The fit can be seen in Fig. 15. ROXIE cannot be used for validating the low current measurements. It cannot deal with residual magnetisation of materials, so it predicts that TF will be constant for low currents. Table XI shows the fit parameters and Fig. 17 illustrates the fit error as function of current. Table XI: FiDeL TF 2010 fit parameters for saturation and residual magnetisation. Component Saturation Parameter sigma I0 s Inomref Residual mag rho r Iinjref Value 0.00053814 766.2540 4.0882 710 0.00004855 2.7151 40 20 Residual error (units) 15 10 5 0 -5 -10 -15 -20 -25 0 200 400 600 800 Current (A) Fig. 17: Residual TF fit error, triangles = measurement-fit points, dotted line = artificial smoothing. - 15 - 7. MQWA field errors Given the unusual design of the twin apertures we check the field errors for all harmonics rather than assuming any symmetry. We need only to consider the integral data as the local measurements were done for the same currents. The available data has been corrected for feed-down using C9, i.e., the b9 a9 are set to zero. We consider modelling only harmonics where the mean value of the measurement series has a significant value compared to the corresponding spread. That is, we look at the ratio between the two. We need to process each aperture separately to avoid being fooled by odd functions. Table XII summarises the statistics. Similarly to the transfer function, we choose to give the FiDeL geometric value at 200 A. We conclude that we need to model saturation and residual magnetization for b1 b3 b4 b5 b6 b7. Table XIII shows the data after removing the geometric value. The b1 b3 b5 b7 are odd functions so we need to fit each aperture separately. We give only geometric values at 200 A for the other harmonics a1 ... a11, b8 ... b11. We use the same model parameters as for TF; saturation and residual magnetisation. We use a numeric method to find the local minima of the squared fit error. Since the number of free parameters is larger than the number of points to fit, we add an artificial, intermediary measurement point at 600 A. We use the data from the ROXIE simulation for this purpose. Figs. 18-27 show the FiDeL model curves, measurement points and the predictions from the ROXIE model. The error bars is ±1 sigma of the measured magnets at the given current. The ROXIE predictions match saturation with 20% error, systematically underestimating saturation. In the case of b4 there is no match, see Fig. 24, since ROXIE predicts no saturation and the measured one is about 4 units. In this case we used an extra point at 455 A for the fit, half-way between other measured points, assuming linear approximation. The fitting parameters are summarised in Table XIV. The error associated to the fit is 10 units for b1, 1 unit for b3, and less than 0.05 units for the other multipoles. Table XII: Average and standard deviation for the MQWA harmonics. Ap I (A) b1 a1 b3 a3 b4 a4 b5 a5 b6 a6 b7 a7 b8 a8 b9 a9 b10 a10 b11 a11 ave stdev 1 1 40.0 -174.8 9.1 0.0 72.6 67.6 -1.9 3.7 -0.2 4.8 0.2 -0.1 1.8 1.4 2.6 0.5 0.0 -1.4 0.0 -0.6 0.8 0.4 0.2 0.2 0.0 0.1 0.1 0.1 0.0 0.2 0.0 0.0 1.0 0.0 0.0 0.0 0.0 0.0 0.1 0.0 0.0 0.0 ave stdev 1 1 200.0 -138.9 5.0 0.1 56.7 61.7 3.5 3.0 0.0 5.2 0.7 -0.1 1.8 1.4 2.9 0.4 0.0 -1.1 0.0 -0.8 0.9 0.4 0.2 0.1 0.0 0.0 0.1 0.1 0.0 0.2 0.0 0.0 1.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 ave stdev 1 1 710.1 0.1 4.5 4.0 39.7 68.0 59.6 5.0 0.0 4.4 4.2 -0.1 1.8 1.4 5.8 0.5 0.0 -1.1 0.0 -1.8 0.8 0.4 0.2 0.2 0.0 0.1 0.0 0.1 0.0 0.2 0.0 0.0 1.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 ave stdev 1 1 810.1 0.1 -24.5 3.9 36.4 73.2 60.4 5.5 0.0 4.1 4.7 -0.1 1.8 1.4 5.7 0.6 0.0 -1.3 0.0 -1.8 0.8 0.4 0.2 0.2 0.0 0.1 0.0 0.1 0.0 0.2 0.0 0.0 0.9 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 ave stdev 2 2 40.0 155.8 20.9 0.0 64.8 68.4 7.3 4.6 0.9 5.5 0.7 1.6 0.0 1.3 -2.2 -0.1 -1.5 0.0 0.5 0.9 0.4 0.2 0.5 0.2 0.0 0.1 0.1 -0.1 0.1 0.2 0.0 0.0 1.0 0.0 0.1 0.0 0.0 0.0 0.1 0.0 0.0 0.0 ave stdev 2 2 200.0 129.4 6.9 0.1 55.6 64.1 -2.9 3.6 0.4 5.6 1.1 1.6 0.0 1.3 -2.8 -0.1 -1.1 0.0 0.5 0.9 0.4 0.2 0.8 0.1 0.0 0.1 0.0 0.1 0.0 0.2 0.0 0.0 1.0 0.0 0.1 0.0 0.0 0.0 0.0 0.0 0.0 0.0 ave stdev 2 2 710.1 0.1 -6.1 2.0 -38.9 67.4 58.8 5.2 -0.1 4.7 4.7 1.9 0.0 1.3 -5.6 0.6 0.0 -1.1 0.0 0.8 0.4 0.2 1.8 0.2 0.0 0.1 0.0 0.1 0.0 0.2 0.0 0.0 1.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 ave stdev 2 2 810.1 0.1 22.7 3.6 -35.6 72.6 58.5 5.6 -0.3 4.2 5.1 2.0 0.0 1.3 -5.6 0.6 0.0 -1.3 0.0 0.8 0.4 0.2 1.8 0.2 0.0 0.1 0.0 0.1 0.0 0.2 0.0 0.0 0.9 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 - 16 - Table XIII: Average and standard deviation for the MQWA harmonics after removing geometric at 200A. I (A) Ap b7 b6 b5 b4 b3 b1 b10 0.2 0.1 0.0 0.0 2.9 0.3 0.0 -1.1 0.1 0.1 -0.1 0.0 4.0 0.7 2.8 0.3 -0.2 -1.0 0.1 0.1 -0.1 0.0 26.4 10.2 19.4 3.9 -0.5 0.1 0.6 0.3 -0.4 -0.3 0.1 0.1 0.0 0.0 2 2 710.1 -135.5 -36.0 20.5 4.1 3.5 0.7 -2.9 0.3 0.0 0.1 1.1 0.1 -0.1 0.0 2 2 810.1 -106.6 -32.7 25.8 4.9 4.0 0.9 -2.8 0.3 -0.2 0.1 1.0 0.1 -0.1 0.0 -5.5 3.1 -0.5 0.1 -0.3 0.2 1 1 710.1 143.4 36.2 18.3 3.8 3.5 0.6 ave stdev 1 1 810.1 114.4 32.8 23.8 4.5 ave stdev 2 2 ave stdev ave stdev ave stdev 1 1 ave stdev 40.0 40.0 -41.0 18.7 -0.3 0.1 200 b1 (units) 150 100 50 0 -50 -100 0 200 400 600 800 Current (A) Fig. 18: Fit of b1 ap1 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. - 17 - 100 50 b1 (units) 0 -50 -100 -150 -200 0 200 400 600 800 Current (A) Fig. 19: Fit of b1 ap2 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. 50 40 b3 (units) 30 20 10 0 -10 -20 0 200 400 600 800 Current (A) Fig. 20: Fit of b3 ap1 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. - 18 - 20 10 b3 (units) 0 -10 -20 -30 -40 -50 0 200 400 600 800 Current (A) Fig. 21: Fit of b3 ap2 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. 5.0 4.0 b4 (units) 3.0 2.0 1.0 0.0 -1.0 -2.0 0 200 400 600 800 Current (A) Fig. 22: Fit of b4 (units)) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. - 19 - 3.5 3.0 b5 (units) 2.5 2.0 1.5 1.0 0.5 0.0 -0.5 -1.0 0 200 400 600 800 Current (A) Fig. 23: Fit of b5 ap1 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. 1.5 1.0 0.5 b5 (units) 0.0 -0.5 -1.0 -1.5 -2.0 -2.5 -3.0 -3.5 0 200 400 600 800 Current (A) Fig. 24: Fit of b5 ap2 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. - 20 - 0.2 0.1 b6 (units) 0.0 -0.1 -0.2 -0.3 -0.4 -0.5 -0.6 0 200 400 600 800 Current (A) Fig. 25: Fit of b6 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. 0.4 0.2 b7 (units) 0.0 -0.2 -0.4 -0.6 -0.8 -1.0 -1.2 -1.4 0 200 400 600 800 Current (A) Fig. 26: Fit of b7 ap1 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. - 21 - 1.4 1.2 1.0 b7 (units) 0.8 0.6 0.4 0.2 0.0 -0.2 -0.4 -0.6 0 200 400 600 800 Current (A) Fig. 27: Fit of b7 ap2 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. Table XIV: Harmonic fit parameters for saturation and residual magnetisation. Component Saturation Parameter sigma I0 s Inomref Residual mag rho r Iinjref Ave fit error b1 ap1 b1 ap2 b3 ap1 b3 ap2 b4 b5 ap1 b5 ap2 b6 b7 ap1 b7 ap2 -128.9058 121.0436 -34.4976 34.3321 -3.9984 -2.8514 2.8454 0.1762 1.0406 -1.0286 626.4983 625.9282 622.1215 622.0595 671.8697 625.6147 625.4454 727.6331 621.8350 622.4412 24.6573 24.5051 22.7959 22.7881 15.6639 20.2559 20.3618 30 21.6830 21.7802 710 710 710 710 710 710 710 710 710 710 -41.0205 26.4604 -5.4793 10.2438 -0.4690 -0.2542 0.5977 -0.3545 0.1982 -0.3065 6.9109 6.6607 5.7820 6.3439 0.7022 5.7820 6.3439 5.7819 4.0504 4.1166 40 40 40 40 40 40 40 40 40 40 9.16 9.12 1.07 1.05 0.08 0.01 0.01 0.01 0.02 0.02 8. MQWB transfer function based upon series measurements 8.1 INITIAL DECISIONS Given the fact that only 8 MQWB are installed in the LHC and that each magnet has its own power supply with the 2 apertures connected in series, and that they operate within 3 different ranges of low current we decide to model the transfer function individually for each magnet. We add the 4 spare magnets having integral measurements in “B” mode in case they one day will be needed. 8.2 GEOMETRIC As for the MQWA, we use the integral measurements at 200 A to define the geometric component, see Table XV. The spread between the magnets is not negligible (±15 units): this justifies the need of a different geometric value for each magnet. See overview in Figs. 28-29. - 22 - Table XV: TF geometric, average and spread for the MQWB circuits CIRCUIT MAGNET TF (T.m/A) TF (units) RQT5.L3 36 0.00275087 12.6 RQT4.L3 50 0.00274807 2.4 RQT4.R3 9 0.00274843 3.7 RQT5.R3 35 0.00274799 2.1 RQT5.L7 46 0.00275033 10.6 RQT4.L7 6 0.00274312 -15.6 RQT4.R7 52 0.00274457 -10.3 RQT5.R7 25 0.00274589 -5.5 Average Stdev 0.00274741 9.8 Integral data show a 1.5% larger TF at 40 A due to residual magnetisation, and 1.5% smaller at 600 A due to saturation (Table XVI). Local measurements are given in Table XVII: the spread at 600 A is much larger, thus suggesting that some data should be rejected. Table XVI: TF for integral measurements, expressed in units of geometric at 200 A. I (A) TF ave (units) TF sigma (units) 40 157 31 200 0 13 600 -151 15 Table XVII: TF for local measurements, before rejecting MQWB 5, expressed in units of geometric at 200 A. I (A) 30 40 200 600 TF ave (units) TF sigma (units) 256 45 150 34 0 17 -124 41 Table XVIII: TF for local measurements, after rejecting MQWB 5, expressed in units of geometric at 200 A. I (A) TF ave (units) TF sigma (units) 30 259 41 40 156 30 200 0 10 600 -129 14 Results after data rejection of local measurement of MQWB 5 are shown in Table XVIII. We have a very good match between integrals and locals at 40 A, and we have a systematic difference of about 20 units as in the MQWA case, with a similar spread. - 23 - Transfer function, ap1 + ap2 (T/A) 0.00092 0.00091 0.00091 0.00090 0.00090 0.00089 0.00089 0.00088 0.00088 0.00087 0.00087 0.00086 0 100 200 300 Current (A) 400 500 600 Fig. 28: TF for local measurements (both apertures). Integrated Transfer function (T.m/A) ap1 + ap2 0.00282 0.00280 0.00278 0.00276 0.00274 0.00272 0.00270 0.00268 0 100 200 300 400 500 600 Current (A) Fig. 29: TF for integral measurements (both apertures). 9. Extra measurements in 2010 The 2010 measurements of beta-beating in LHC convinced us that some of the MQWB transfer functions were wrong by more than 200% at injection! The spare magnets 10 and 51 were measured during spring 2010. The measurement campaign is described in [11]. Device: Measurements were done using SSW (single stretch wire) since we only needed to correct the integral transfer function. The pre-cycle is identical to the actual LHC pre-cycle which take advantage of the bipolar power supply in order make sure the magnet is at the right hysteresis branch at injection. See Fig. 30 and Table XIX. - 24 - I (A) 600 400 200 0 0 500 1000 1500 2000 -200 -400 -600 t (s) Fig. 30: MQWB pre-cycle where Imin=-600A, Imax=+600A, dI/dt=5A/s, 10s wait time between ramps. Table XIX: MQWB pre-cycle parameters Pre-cycle No of periods ramp rate (A/s) I_MIN (A) t_MIN (s) I_MAX (A) t_MAX (s) 4 5 -600 10 600 10 10. MQWB transfer function based upon 2010 measurements 10.1 INITIAL DECISIONS Table XX shows the comparison between the new SSW and the old RC measurements for the MQWB 51. MQWB 10 was not measured during series measurements. We have good agreement for 600 A, but the value at 200 A is surprisingly about 1% different. Table XX: Comparison of MQWB 51 measurements. (SSW – RC) / RC [units] I Ap 1 200 600 Ap 2 -87 -8 -93 -15 Fig. 31 and Table XXI show the shape of the transfer function, as usual based upon normalising the curve to the geometric component at 200A as usual. We can see from Table XXI that the sign of the transfer function is opposite from previous measurements due to the difference in pre-cycle. This was anticipated. But what comes a surprise is that the transfer function is not flat at 200 A. It reaches its flattest point at 437 A. So the whole transfer function has changed. This would then explain the difference in Table XX at 200A. We must re-express the transfer function with a geometric component at 437 A. See Table XXII for the new values. Unfortunately we do not have any more measurements close to 437 A to find more precisely the peak. We remove the measurement points at 1 and 2 A for making the FiDeL fit since these points give a negative field integral. See Fig. 32 for the shape of the transfer function. The only measurement point from the series we could have re-used is the one at 600 A. But since the spread is small at 600 A, we do not believe this is justified. - 25 - 25000 MQWB avg 20000 15000 b TF (units) 10000 5000 0 -5000 -10000 -15000 0 10 20 30 40 -20000 -25000 Current (A) Fig. 31: MQWB average TF (units) for current range 0-40A, after removing geometric at 200A. Table XXI: Average MQWB TF (units) and spread (units) after removing geometric at 200A. I TF avg TF sigma 1 -20915 -826 2 -10715 -6765 5 -4323 354 10 -2155 122 15 -1416 71 20 -992 48 27 -704 42 40 -437 28 100 -111 13 200 0 8 312 29 8 437 31 11 600 -70 17 437 138 9 Table XXII: Average MQWB TF (units) after removing geometric at 437A. I 5 10 15 20 27 40 100 200 312 437 600 TF ave -4340 -2179 -1442 -1019 -733 -467 -141 -31 -2 0 -101 - 26 - 0 MQWB 10 +51TF (units) -2000 -4000 -6000 -8000 -10000 0 100 200 300 400 500 600 -12000 Current (A) Fig. 32: MQWB magnets 10+51, average TF (units), after removing geometric at 437A. 10.2 GEOMETRIC The average FiDeL geometric value at 437A from the 2010 SSW measurements is used. This means we use the average value of 2 magnets to represent the 8 magnets installed. See Table XXIII. Table XXIII: FiDeL TF 2010 fit parameters for geometric, saturation and residual magnetisation. Component Geo Parameter Value gamma 0.00273293 Igeoref 437 Saturation sigma 0.00011784 I0 642.8172 s 11.0361 Inomref 600 Residual mag rho -0.00018512 r 1.0401 Iinjref 30 10.3 SATURATION AND RESIDUAL MAGNETIZATION COMPONENTS We use the two components saturation and residual magnetization to fit the TF curve in Fig 32. There are no DC magnetization components for normal-conducting magnets, neither time-dependent components, and there is no beam screen in the bore. The fits are done according to the equations in [12]: ππΉ = πΎ πππ + ππΉ π ππ‘ + ππΉ πππ πππ’ππ ππΉ π ππ‘ = − ππ πΌ − πΌ0 [1 + πππ (π )] 2 πΌπππ - 27 - ππΉ πΌπππ ππ = ππ ( ) |πΌ| πππ πππ’ππ Saturation: We have only two measure points from Table XXII (437A and 600A) to make the fit. The reason is that we did not anticipate that the entire TF should change as function of the pre-cycle so that we had to adjust the geometric to 437A. Fortunately we have confidence that the values at 600 A represent the MQWB family (Table XX). Fig. 33 shows the fit of the saturation. We use the ROXIE simulation of the MQWB 2D model to check. We have a very good agreement, within 10 units at 600A. ROXIE computations relate to 2D cross-section while the integral measurements include the heads (end effects). 0 MQWB 10 +51TF (units) -2000 -4000 -6000 -8000 -10000 0 100 200 300 400 500 600 -12000 Current (A) Fig. 33: Fit of the TF saturation (units), line = fit, triangles = measurements, diamonds = ROXIE simulation. Residual: We use the 2010 SSW measurements in Table XXII for the fit, where current is less geometric at 437A. The fit can be seen in left part of Fig. 32. ROXIE cannot be used for validating the low current measurements. It cannot deal with residual magnetisation of materials, so it predicts that TF will be constant for low currents. The fit error is much bigger than for MQWA. See Fig. 34. It is about 10% at lowest operational current 2A and 2% at 15A. There is too much spread and uncertainty in this nonlinear region to make any better estimates. The fit error undershoots with about 1% at 100A and then stays within 10 units as the current increases. - 28 - Residual error (units) 1000 800 600 400 200 0 -200 0 200 400 600 Current (A) Fig. 34: Residual TF fit error, triangles = measurement-fit points, dotted line = artificial smoothing. 11. MQWB field errors We process the MQWB field errors following the same considerations as for MQWA. As usual we use the FiDeL saturation and residual magnetisation components. We only have to consider the integral data as the local measurements were done for the same currents. We consider modelling only harmonics where the mean value of the measurement series has a significant value compared to the corresponding sigma. That is, we look at the ratio between the two. We need to process each aperture separately to avoid being fooled by systematic odd functions. Table XXIV shows the statistics of the integral measurements. We conclude that we need to model saturation and residual magnetization for b1 b3 b4 b5 b6 b7. The b1 b3 b5 b7 are odd functions so we need to fit each aperture separately. We give only geometric values at 200 A for the other harmonics a1 a3 ... a11 b8 ... b11. We should have expressed the field errors at 437 A as we did for the MQWB transfer function, but lack of measurements prevent us from doing so. Fortunately the MQWB are few and mostly to be regarded as corrector magnets. Therefore relative field errors will have small effects on LHC. We need to add an extra measurement point at higher current to be able to numerically fit the saturation in a more sensible way (it would otherwise be an “unphysical” step function). We use the value from ROXIE at 500 A. Table XXV shows the values used for the fitting curves after removing geometric at 200A. Figs. 35-43 show the measurements, ROXIE model and FiDeL fits. The fitting parameters are summarised in Table XXVI. - 29 - Table XXIV: Average and standard deviation for the MQWB integral harmonics. Ap I (A) b1 b3 40.01 -111.37 0.03 99.42 a3 b4 a4 b5 a5 b6 a6 b7 a7 b8 a8 b9 a9 b10 a10 b11 a11 ave stdev 1 1 3.19 -2.00 3.92 4.31 0.99 -0.17 2.04 1.59 2.88 0.34 -1.57 -0.03 -0.79 0.33 0.86 0.41 0.18 0.18 0.00 0.10 0.04 -0.01 0.00 0.00 0.11 0.20 0.00 0.00 1.02 -0.01 -0.04 0.05 0.05 0.04 -0.02 0.01 ave stdev 1 200.04 -218.84 -25.64 -2.09 1 0.08 96.87 3.12 5.56 1.46 -0.19 1.98 1.56 0.75 0.37 -1.30 -0.02 0.34 0.97 0.37 0.20 0.01 0.05 0.00 0.01 0.00 0.00 0.10 0.20 0.00 0.00 1.05 0.00 -0.11 0.02 0.05 0.02 -0.02 0.01 ave stdev 1 600.09 -160.33 -11.01 -1.34 1 0.13 98.67 3.70 5.57 3.57 -0.19 1.87 1.55 1.98 0.30 -1.13 -0.03 -0.31 -0.01 -0.02 0.02 0.00 0.00 0.40 1.00 0.47 0.19 0.19 0.05 0.10 0.19 0.00 0.00 1.00 0.00 -0.09 0.04 0.05 0.02 -0.01 0.01 ave stdev 2 2 40.00 0.03 31.75 -4.81 0.58 112.86 4.21 4.77 0.73 1.85 0.19 -2.93 -0.13 -1.59 1.36 0.63 0.73 0.43 0.05 -0.08 0.00 0.00 0.09 0.19 0.00 0.00 0.98 -0.01 0.14 0.05 0.08 0.03 -0.01 0.01 ave stdev 2 200.02 2 0.07 196.51 24.80 -0.72 97.43 4.67 4.75 1.06 1.87 0.20 -0.85 -0.04 -1.31 -0.01 -0.04 1.34 0.54 0.72 0.42 0.19 0.17 0.04 0.09 0.02 -0.05 0.00 0.01 0.08 0.18 0.00 0.01 1.05 -0.01 0.02 0.04 0.12 0.02 -0.01 0.01 ave stdev 2 600.09 2 0.13 146.16 10.42 -0.67 100.69 5.39 4.09 3.19 1.88 0.16 -2.09 -0.02 -1.15 -0.03 1.32 0.59 0.63 0.58 0.20 0.03 0.09 0.00 -0.04 0.00 0.00 0.08 0.18 0.00 0.00 1.00 -0.01 0.04 0.04 0.09 0.02 -0.01 0.01 0.00 0.19 0.07 0.17 0.84 -0.01 0.23 0.09 0.32 0.19 Table XXV: Average and standard deviation for the MQWB harmonics after removing geometric at 200A. Ap I (A) b1 b3 b4 ave stdev 1 1 40 107.47 28.83 -0.48 14.45 2.23 0.26 ave stdev 1 1 600 58.51 14.63 2.11 13.57 1.12 0.46 ave stdev 2 2 ave stdev 2 2 b5 b6 1.23 0.12 -50.35 -14.39 2.12 -1.24 12.01 1.00 0.31 0.09 - 30 - b10 2.13 -0.27 -0.86 -0.03 0.16 0.10 0.07 0.03 0.17 -0.38 -0.05 0.29 0.03 0.03 40 -164.76 -29.61 -0.33 -2.08 -0.28 91.97 3.16 0.46 0.43 0.11 600 b7 0.16 0.29 0.88 -0.07 0.15 0.14 0.36 -0.05 0.03 0.03 140 120 b1 (units) 100 80 60 40 20 0 -20 -40 0 100 200 300 400 500 600 Current (A) b1 (units) Fig. 35: Fit of b1 ap1 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. 40 20 0 -20 -40 -60 -80 -100 -120 -140 -160 -180 0 100 200 300 400 500 600 Current (A) Fig. 36: Fit of b1 ap2 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. - 31 - 35 30 b3 (units) 25 20 15 10 5 0 -5 0 100 200 300 400 500 600 Current (A) Fig. 37: Fit of b3 ap1 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. 5 0 b3 (units) -5 -10 -15 -20 -25 -30 -35 0 100 200 300 400 500 600 Current (A) Fig. 38: Fit of b3 ap2 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. - 32 - 3.0 2.5 b4 (units) 2.0 1.5 1.0 0.5 0.0 -0.5 -1.0 0 100 200 300 400 500 600 Current (A) Fig. 39: Fit of b4 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. 2.5 b5 (units) 2.0 1.5 1.0 0.5 0.0 -0.5 0 100 200 300 400 500 600 Current (A) Fig. 40: Fit of b5 ap1 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. - 33 - 0.5 b5 (units) 0.0 -0.5 -1.0 -1.5 -2.0 -2.5 0 100 200 300 400 500 600 Current (A) Fig. 41: Fit of b5 ap2 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. 0.2 b7 (units) 0.0 -0.2 -0.4 -0.6 -0.8 -1.0 0 100 200 300 400 500 600 Current (A) Fig. 42: Fit of b7 ap1 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. - 34 - 1.0 b7 (units) 0.8 0.6 0.4 0.2 0.0 -0.2 0 100 200 300 400 500 600 Current (A) Fig. 43: Fit of b7 ap2 (units) after removing geometric at 200A, line = fit, triangles = integral measurements, diamonds = ROXIE simulation. Table XXVI: MQWB harmonic fit parameters for saturation and residual magnetisation. Component Saturation Parameter b1 ap1 b1 ap2 b3 ap1 b3 ap2 b4 b5 ap1 b5 ap2 b6 b7 ap1 b7 ap2 sigma -58.509756 50.351598 -14.628818 14.385909 -2.116363 -1.228817 1.235635 -0.167727 0.375636 -0.364455 I0 538.33 538.26 529.17 529.17 527.03 523.22 523.22 526.96 523.16 523.21 s 33 33 36 36 34 34 34 381887770 447723571 447717906 Inomref 600 600 600 600 600 600 600 600 600 600 Residual mag rho 107.625114 -137.636805 28.868311 -29.619868 -0.402909 2.130273 -2.076697 -0.271532 -0.863979 0.884742 r 7.31 7.83 6.75 6.88 5.29 5.78 6.34 4.97 5.35 6.15 Iinjref 40 40 40 40 40 40 40 40 40 40 Min fit error 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 Ave fit error 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 Max fit error 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 12. Summary and critical points ο· The TF of the MQWA has a large saturation of about 1.5%, which is modelled. The spread between magnets is very low. At injection we have a strong magnetization component of about 3%. This effect is based upon extra 2010 measurements of 2 magnets and is due to the Imin=20A in the pre-cycle. Geometric components are modelled for each magnet. We have 3 different types of measurements: local and integral RC measurements, integral SSW measurements. ο· Normal multipoles have a systematic component of saturation and magnetization. We modelled an average b1, b3, b4, b5, b7 for the whole family. The ROXIE model is systematically underestimating the saturation by about 20%, but matches well the shape of the measured curve. The only exception is b4: measurements show a saturation giving 4 units at high field, against a model result of less than 0.1 units. For all other multipoles, a geometric component has been computed. ο· The TF of the MQWB does not fit the series measurements due to a different pre-cycle need for reproducibility in LHC operation. The saturation effect is about 1.5% at 600 A. The MQWB magnets operate in 3 different current ranges and the dominating - 35 - effect is always the residual magnetisation which is about 20% at 10 A. The MQWB TF is based upon the 2010 SSW measurements of 2 spare magnets. 13. Acknowledgements We wish to thank D. Cornuet and M. Silva for the original measurements and reports, X. Panagiota for making the original analysis and FiDeL REFPARM file for the 2008 LHC startup, G. de Rijk for the MQW design models and advices, B. Auchmann for help with ROXIE, E. Todesco for discussions and feedback, R. Tomas Garcia for the beta-beat measurements and analysis in LHC 2009-2010, the extra measurement campaign in 2009-2010 by M. Buzio, J. Garcia Perez and R. Chritin. References D. I. Kaltchev, et al., “Optics Solutions for the Collimation Insertions of LHC”, PAC 1999. O. Bruning, “Collimation Optics”, Presentation in LHC Machine Advisory Committee, meeting 14, 2003. O. Bruning, et al., CERN Report 2004-003 (2004). R. Assmann, et al., “The New Layout of the LHC Cleaning Insertions”, LHC Project Report 774, 2004 G. de Rijk, et al., “Construction and Measurement of the Pre-Series Twin Aperture Resistive Quadrupole Magnet for the LHC Beam Cleaning Insertions”, IEEE Trans. Appl. Superconduct., Vol 12, pp 55-58, 2002 [6] R. Wolf, “Corrector Memo”, internal department memo, 2006 [7] R. Wolf, “Field Error Naming Conventions For LHC magnets”, rev 3, EDMS 90250, 2001 [8] D. Cornuet, et al., “Field Quality Measurements of the LHC Warm Twin Aperture Quadrupole Magnets”, LHC Project Note 380, 2006, EDMS 703686, 2006. [9] M. Silva, “MQW magnets: results from the measurement campaign”, EDMS 609363, 2005 [10] D. Cornuet, “Status of data retrieval for normal-conducting magnets”, presentation in FiDeL WG 03 April 2007, web site: http://cern.ch/fidel [11] P. Hagen, “2010 measurement campaign and updated FiDeL models”, presentation in FiDeL WG 11 May 2010, web site: http://cern.ch/fidel [12] FiDeL Model specifications, EDMS 908232, 2008 [13] ROXIE web site: : http://cern.ch/roxie [1] [2] [3] [4] [5] - 36 -