Electronic Spectra of Change Transfer Complexes CHEM 366 Miaomiao Gu Experiment date: Feb 03rd, 2011 Student ID: 301100545 Submission date: Feb 17th, 2011 Abstract. The absorbance at wavelength of a series of o-xylene, mesitylene and organic complex solutions was measured by UV/Vis spectrometer and recorded. The equilibrium constant and molar extinction coefficients were then extrapolated and the values were: 0.075(0.003) M-1, 1250(20) M-1cm-1 and 0.08(0.002) M-1, 2000(13) M-1cm-1 for o-xylene and mesitylene complexes. Moreover, the free energy was extrapolated from equilibrium constant for each complex solution. Purpose: To determine the extinction coefficient and equilibrium constant corresponding to the charge transfer complex by analysis the charge transfer absorption spectra of a series of electron donor mixed with a same electron acceptor, teracyanoethylene. Introduction: In this experiment, the absorbance of each molecular complex in various concentrations was measured by UV-vis spectra. This instrument provides directly absorbance of each solution with no true unit. Further analysis between the same common electron acceptor and different electron donor, the equilibrium constant, molar extinction coefficient and free energy were determined. Therefore, in this experiment, the relationship between the electron acceptor and donor of the formed molecular complexes was studied and determined. Theory: The standard free energy is related to the equilibrium constant of the complex through the expression :( 1) ΔG0 = –RTlnK (1) At the desired wavelength, Beer-Lambert’s law relates the concentration to the absorbance at wavelength according to: OD = cl (2) A substitution of equations was re-arrangement leads to: ODmax/ (l*[A]o) = AB*K*[B] 0/ (1+K [B] 0) (3) Equilibrium constant was converted from a linear equation in the series of dilutions: U= s/ [B] 0 + i (4) The observable transition energy related to the ionization potential of the electron donor: ≈(Ib-P)/hc (5) Species A, B and the CT complex AB are in equilibrium with equilibrium constant: [AB] ≈K[A]o[B]0/(1+K[B]0) (6) Relate ΔGo to Ib according to : ΔGo=M-Q/(Ib-P) (7) Experimental procedure: Same procedures were as referenced in lab manual. (1) Result and Calculation: At room temperature, two series of o-xylene and Mesitylene solutions, Benzene, Toluene, and Hexa.MethylBenzene (HMB) solutions were prepared with electron acceptor, teracyanoethylene (TCNE) in different concentrations. Their uv/vis spectra were collected by Hp-8453 spectrophotometer. Each series was diluted by TCNE to four solutions with different concentrations. The concentration value of the acceptor and donor in each mixture was obtained and given in Table 1-3. Mixture ID [B]0/M ± [A]0/M ± OD ± max/cm 2.244 0.008 Benzene 0.00008 3E-07 0.13 0.002 0.372 1.882 0.007 Toluene 0.00008 3E-07 0.081 0.0007 0.398 o-Xylene 1.658 0.006 0.00008 3E-07 0.0981 0.0004 0.42 Mesitylene 1.437 0.005 0.00008 3E-07 0.1304 0.0003 0.448 0.01115 4E-05 HMB 0.0001 4E-07 0.1725 0.0001 0.527 Table 1. Donor and acceptor concentrations and measured data of mixture solutions. cm-1 2.688172043 2.512562814 2.380952381 2.232142857 1.897533207 Mixture ID [B]0/M ± [A]0/M o-xylene 1.658 0.006 0.00008 B2 0.663 0.002 0.000092 B3 0.332 0.001 0.000096 B4 0.1989 0.0007 0.0000976 B5 0.0663 0.0002 0.0000992 Table 2. Donor and acceptor concentrations different concentrations. ± OD ± max/cm cm-1 3E-07 0.0981 0.0004 0.42 2.380952381 3E-07 0.1091 0.0003 0.426 2.34741784 3E-07 0.1004 0.0004 0.426 2.34741784 3E-07 0.0846 0.0003 0.425 2.352941176 3E-07 0.0612 0.0003 0.424 2.358490566 and measured data of o-xylene at Mixture ID [B]0/M ± [A]0/M mesitylene 1.437 0.005 0.00008 B2 0.575 0.002 0.000092 B3 0.287 0.001 0.000096 B4 0.1725 0.0006 0.0000976 B5 0.0575 0.0002 0.0000992 Table 3. Donor and acceptor concentrations different concentrations. ± OD ± max/cm cm-1 3E-07 0.1304 0.0003 0.448 2.232142857 3E-07 0.1698 0.0003 0.449 2.227171 3E-07 0.1415 0.0003 0.45 2.222222 3E-07 0.1313 0.0003 0.448 2.232143 3E-07 0.0859 0.0002 0.449 2.227171 and measured data of mesitylene at The absorbance at wavelength of each mixture was obtained by uv/vis spectra. Equilibrium constant and free energy were computed and given in table 4. Mixture ID Benzene Toluene o-Xylene Mesitylene HMB [AB] 5.43E-05 6.43E-05 8.85E-06 7.93E-08 6.75E-05 ± b-P ΔG0/J 3E-07 4E-07 5E-08 4E-10 4E-07 9.66707E-38 1.07082E-37 1.39584E-38 1.17301E-40 8.48279E-38 149.7087 -1885.59 6267.202 6111.05 -12643.8 equilibrium K/M-1 AB/M-1cm 0.075(0.003) 0.08(0.002) 1250(20) 2000(13) -1 Table 4. Extrapolated values of complex solutions. The equilibrium constant K and molar extinction coefficient of o-xylene and mesitylene mixtures were calculated in turn of a calibration line which was plotted U versus 1/ [B]o. The K and values were extrapolated from the calibration equation, as shown in figure 1 and 2. U Vs 1/[B]0 0.0018 y = 6E-05x + 0.0008 0.0016 U 0.0014 0.0012 0.001 0.0008 0 2 4 6 8 1/[B]0 (M-1) 10 12 14 16 Figure 1. Calibration curve as U against 1/ [B] 0 of o-xylene mixture solution. U vs 1/[B]0 0.0012 y = 4E-05x + 0.0005 0.0011 0.001 U 0.0009 0.0008 0.0007 0.0006 0.0005 0 5 10 15 1/[B]0 (M-1) Figure 2. Calibration curve as U against 1/ [B] 0 of Mesitylene mixture solution. 20 By equation (4), i=1/AB where i=0.0005 in figure 2. Thus, AB= 2000M-1cm-1. Also, as s=1/ (AB*K) where s=4E-05 AB= 2000M-1cm-1, then K=0.08M-1 for mesitylene solutions. Therefore, the values of equilibrium constant and molar extinction for mesitylene solutions were 0.08M-1 and 2000M-1cm-1 as shown in Table 6. The equilibrium constant and molar extinction coefficients of o-xylene mixture were determined by the same method and the values were shown in Table 4. The free energy of mesitylene mixture was extrapolated from equation (1), where R is the gas constant 8.3145 J/mol/K, T is temperature at 291K and equilibrium constant K=0.08M-1. Therefore, the free energy ΔGo of mesitylene mixture was obtained with value of 6110.68J and shown in Table 6. Same as mesitylene mixture, the free energy ΔGo of o-xylene mixture was obtained and shown in Table 4. By equation (6) where K=0.08M-1, [A]0= 0.00008M, and [B]0= 1.437M, the mesitylene complex concentration was extrapolated as 8.248547033E-08M and shown in Table 6. The other concentration of each complex was obtained by the same method and given in Table 4. According to ΔGo related to Ib, the provided data complex solution could extrapolate an equation for the measuring complex solution. The relationship was ΔGo against Ib-P and shown in Figure 3. ΔGo Vs 1/Ib-p 0 -1000 0 2 4 6 8 10 12 -2000 -3000 y = 1369.9x - 14300 -4000 ΔGo /J -5000 -6000 -7000 -8000 -9000 -10000 -11000 -12000 -13000 1/Ib-P E+36 Figure 3. Calibration curve of ΔGo Vs 1/Ib-p. From equation (7), ΔGo value could be extrapolated from Ib where (Ib-P) was obtained from equation (5). Thus, ΔGo values were determined as 83777.14351J and 11656529.75J for o-xylene and mesitylene complexes. Error Analysis: The equilibrium constant and molar extinction coefficient of o-xylene were extrapolated from the calibration curve in figure 1 where y= 6E-05x+0.0008. AB was obtained from the intercept as 1250M-1cm-1. The standard deviation of AB was determined from σAB = 0.0008/√ (σy2Σ (xi2)/D). (2) Thus, σAB= 19.54268602, and σAB should recorded as 1250(20) M-1cm-1 and shown in Table 4. The error of σAB of mesitylene was determined by the same way and given in table 4. Same as AB, equilibrium constant K was determined from the slope of the calibration curve in figure 1 for mesitylene. By equation (4) where K= i/s, K was extrapolated as 0.075M-1. The uncertainty of equilibrium constant K was computed from standard deviation of the slope and intercept %e = √((%es)2+(%ei)2), where %es= 7.62 and %ei= 4.69. The uncertainty was obtained as 0.003 and the equilibrium constant K was recorded as 0.075(0.003) M-1 in Table 4. The uncertainty of equilibrium constant K of o-xylene was extrapolated and shown in table 4 as well. From equation (1), the uncertainty of free energy ΔGo is equal to the relative uncertainty in equilibrium constant K for mesitylene complex where ey= ex/x. Therefore, the uncertainty was extrapolated as 3 and the free energy ΔGo was recorded as 6267(3) J in table 4. And the obtained ΔGo uncertainty of o-xylene was shown in table 4. The uncertainty of calibration curve in figure 1 was determined from the vertical deviation di and the standard deviation of this slope was calculated from σy = √ (Σ di2/ (n-2)). By calculations, the uncertainty of the calibration curve was σy = 0.01. By the same method, the uncertainty for other figures was extrapolated. As the concentration of [A] 0, [B] 0 and [AB] were not treated as final values, their uncertainties were not calculated particularly, but they were still shown in Table 1-4. Discussion: For Benzene, Toluene and HMB complexes which did not measured equilibrium constants, their computed free energy and 1/ (Ib-P) values were consistent with the linear equation (7). By this linear equation, the computed free energy for o-xylene and mesitylene complexes were much bigger than the free energy which directed computed from equilibrium constant. Therefore, free energy determined from the experimental measurement was not consistent with the equation (7). This might because the equilibrium constant was determined from a series of dilutions. As the concentration of the complexes was already very small, the concentration of the dilutions might too small to close to the limit detection of the UV/vis spectra. Therefore, the noise of the spectra might cause the error in the extrapolation from the absorbance. As an electron acceptor, TCNE substituted benzenes, electron donor to form molecular complexes TCNE-Benzene. C6H5R + C3N2C3N2→ C6H5R-C3N2C3N2 Although, TCNE is the most powerful organic electron acceptors, being a prototype for other cyano-based electron acceptors, hexacyanobutandiene (HCBD) also can substitute benzenes to form molecular complexes. While TCNE is a basic constituent in charge transfer compounds with either magnetic or conducting properties, HCBD has been used in building hybrid organic-inorganic molecular magnets. Conclusion: By the measurement of absorbance at wavelength of benzene aromatic organics and dilution with various concentrations, equilibrium constant K and molar extinction coefficient of o-xylene and mesitylene complexes with TCNE was obtained and compared with literatures. Moreover, the free energy could obtain from equilibrium constant or the related Ib, the equilibrium constant method was more accurate. As a result, the UV/vis spectrum was a good instrument of determining equilibrium constant, but not good for measuring low concentration complexes. Reference: 1) Physical Chemistry lab manual. Pg.V-1-V-11. 2) Daniel C.Harris; Quantitative Chemical Analysis, 7thed; W.H.Freeman and Company, New York, 2007.P39-71. 3) Carl W. Garland; David P. Shoemaker; Experiments in Physical Chemistry, 8th edition; McGraw-Hill Higher Education, New York, 2009. P106-118 4) Daniel C. Harris; Quantitative Chemical Analysis, 6th edition; W. H. Freeman and Company, New York, 2003. P51-P56.