Binary Distillation University of Illinois at Chicago Binary Distillation–Pre Lab This schematic illustrates what happens in a distillation column. A liquid mixture is fed into the distillation column. On entering the column, the heated feed is partially vaporized and rises up the column. However, as it rises, it cools by contacting the descending cooler liquid and partially condenses so that, while part of vapor continues to flow upward, the condensed portion is enriched in the less volatile component(s) and flows downward. As the vapor continues to flow upward, it undergoes partial condensation a number of times and each time becomes richer in the more volatile component(s). Unit Operations Lab 4 March 11, 2010 Group 3 Shrikant Shah Brandon Farr Alex Georg Michael Ogiefo Mohammed Khatib Ghassan Alkhateeb Unit Operations ChE-382 Group No. 3 p. 1 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago Table of Contents 1. WP&C .................................................................................................................................................. 2 1. Introduction ......................................................................................................................................... 4 2. Theory .................................................................................................................................................. 5 3. Apparatus .......................................................................................................................................... 13 4. Materials and Supplies ..................................................................................................................... 18 5. Procedure ........................................................................................................................................... 19 8. Error Analysis ................................................................................................................................... 21 9. References .......................................................................................................................................... 22 1. WP&C What is the purpose of this experiment? Unit Operations ChE-382 Group No. 3 p. 2 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago The purpose of this experiment is separate a mixture of a 5% wt methanol-water solution in a binary distillation column. About 0.1-2 GPM of methanol-water mixture will be introduced into the round bottom flask of the column. The column will operate at temperatures ranging from 0-100oC. The column will be run and allowed to reach steady state before samples from the six stages will be obtained and tested with a Refractometer to eventually obtain the methanol concentration in each stage. What are the hazards associated with the experiment? 1. 2. 3. Methanol is relatively toxic fluid. It can cause eye, skin and respiratory tract irritation when carelessly exposed to lab personnel. The electrical wires for the thermocouples are carelessly exposed. If lab personnel were to accidentally touch them while conducting experiment, they run the risk of an electric shock. The persistent use of fluids like water and methanol could lead to fluid spillage. If lab personnel were to walk over an affected area, they run the risk of injuries due to a fall. How will the experiment be conducted in a safe manner? 1. 2. 3. Lab personnel should wear gloves, goggles, slip resistant shoes, and a facemask when conducting experiment especially when handling the methanol. Paper towels or task wipers should be in close proximity to clean up any fluid spills that may occur during experiment. The Distillation column should not be heated until the mixture has reached the Reboiler section of unit. What safety controls are in place? 1. There is a failsafe valve present that allows the removal of fluids incase the unit gets flooded. Describe safe and unsafe ranges of operations. 1. The operating flow rates of water should be between 0-2 GPM. All relevant data can be obtained in this range. An increase flow rate between 3-10 GPM could be hazardous due to increased pressure. Signatures: Shrikant Shah Brandon Farr Alex Georg Michael Ogiefo Mohammed Khatib Ghassan Alkhateeb Unit Operations ChE-382 Group No. 3 p. 3 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago 1. Introduction Distillation is a physical separation process that uses the differences in volatility between compounds in a liquid mixture. Binary distillation separates two liquid components from one another. All fluids possess some degree of volatility that is a measure of their tendency to vaporize. A higher volatile compound such as methanol will vaporize more quickly when compared to water under the same atmospheric conditions. Boiling the two components would also make the difference even more apparent, as the less volatile water would boil at a higher temperature than the methanol. More importantly, even when two components are mixed together, the unique physical properties of the individual component still causes the more volatile component to vaporize faster. This is what ultimately makes distillation possible. A single step of distillation uses these principles of volatility to achieve a separation of compounds at equilibrium. At the boiling point of a mixture, the more volatile component exists as a greater fraction in the vapor than it will in the liquid. In a methanol-water mixture, distillation would yield a greater mole fraction of methanol in the vapor than in the liquid. Condensing the vapor and adding more distillation steps or stages in series is the basis for continuous distillation. When a continuous system is arranged vertically, the system is defined as a distillation column. In such a column, falling liquids vaporize at lower stages and rising vapors condense at higher ones. A heat source at the bottom of the column enables this action, creating a decreasing temperature gradient up the column. With more distillation steps greater purity can be achieved in top and bottom of the column. Distillation columns are widely used for separations in industry, most notably in petroleum, natural gas, and chemical processing, as well as any other large-scale liquid Unit Operations ChE-382 Group No. 3 p. 4 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago production. In processing petroleum, different hydrocarbons can be separated according to their volatility. Medicinal herbs can be distilled from plant matter in the pharmaceutical industry. Methanol is distilled for high alcohol content products such as whiskey. In this experiment, we will distill a mixture of methanol and water in a column distiller. The apparatus has six trays that will act as the stages of the continuous distillation. Collection of the condensate at each of these stages will allow an analysis of the composition at each plate. The temperature will also be recorded at each stage. Ultimately, decreasing temperatures and increasing compositions of methanol will be observed up the column. Finally, vapor-liquid equilibrium data will allow a comparison between experimental results and theoretical trends. 2. Theory Distillation is the separation, by vaporization, of different components in a mixture because of the different volatilities that they possess at a given boiling point of the mixture. When a mixture reaches a specific temperature and pressure a certain amount of the mixture moves into the vapor phase until the vapor reaches the mixture’s vapor pressure. This point is known as the vapor-liquid equilibrium. Volatility is a measure of a pure component’s vapor pressure at a set pressure and temperature in a specific mixture. It is incorrectly assumed that the components of a mixture will separate based on their boiling points when pure. Rather, the boiling point of a mixture is based on the total vapor pressure of a mixture, which is a sum of the vapor pressures of each individual component in the mixture. This is known as Dalton’s law. π π ππ‘ = ∑(π£ππ + π£ππ … ) (1) Where, Unit Operations ChE-382 Group No. 3 p. 5 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago ο· π π ππ‘ [=] Vapor pressure of mixture [kPa] ο· π£ππ [=] Vapor pressure of component a [kPa] ο· π£ππ [=] Vapor pressure of component b [kPa] This means that a component will not boil off “cleanly” meaning it is impossible through distillation to obtain a pure substance. The vapor created above a mixture is also a mixture of components. The composition of the vapor is based on the volatility of each of the substances. Raoult’s law helps us to determine what the volatility, or “K value” of a substance. This in turn allows us to find the mole fraction of a component in the vapor phase. π¦π = π£ππ βπ₯π (2) ππ ππ‘ Where: ο· π¦π [=] mole fraction of component a in vapor phase (dimensionless) ο· π£ππ [=] vapor pressure of component a [kPa] ο· π₯π [=] mole fraction of component a in liquid phase (dimensionless) ο· π π ππ‘ [=] Vapor pressure of mixture [kPa] Not all mixtures obey Raoult’s law. Some components that have high solubility with each other form azeotrope. An azeotrope is a mixture that has either a higher or lower boiling point than the boiling point of any of the pure components. This means that when a mixture reaches an azeotrope, such as ethanol and water at 95.6 % water, it behaves as a pure substance. The simplest of all distillation techniques is called flash distillation. Flash distillation occurs when a mixture at a specific temperature and pressure is allowed to drop in pressure. This Unit Operations ChE-382 Group No. 3 p. 6 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago changes the vapor-liquid equilibrium of the mixture and creates a vapor rich in the most volatile component(s). This is also the crudest form of distillation and does not allow for refinement of the distillates. A more complicated form of distillation is batch distillation. Batch distillation uses both a boiler and a condenser, but only allows one separation, or cut, to be taken from the mixture. Figure 1: A schematic of a typical continuous distillation tower. A is the kettle that holds the liquid mixture, B is the Reboiler that adds heat to the mixture, C is the distillation column that holds the trays or packing, D is the condenser that turns the distillate to a liquid product. Continuous distillation is the most complicated and most common form. It has a boiler, condenser and multiple trays or packing which allows the vapor to condense as it moves up the column and cools. The trays or packing allows for a better separation of the components in the mixture which in the end gives purer products. It also adds a lot of complexity to the system. Unit Operations ChE-382 Group No. 3 p. 7 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago Figure 2: A typical tray in a distillation column. A is the tray itself, B are the holes in the tray that allow the vapor to pass through the tray, C and D are the tubes that allow liquid to pass from one tray to another. As the vapor moves up the column it gets progressively cooler which allows some of the mixture to condense and further concentrates the vapor with the lightest component(s). Figure 3: This shows a simple McCabe-Thiele diagram. The number of steps corresponds to the number of trays needed for a given mixture. Unit Operations ChE-382 Group No. 3 p. 8 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago Determining the size of the tower and the number of trays needed was historically a very difficult process involving many calculations. The McCabe-Thiele method greatly simplified that process. The method uses a graphical representation of the material balance equations as operating lines on a graph of the liquid composition (x-axis) and the vapor composition (y-axis). The bottom line in Figure 3 is the x-y line. This starts at the origin and ends where x and y both equal 1.This line would represent a distillation column that operated at total reflux and total boilup, meaning that all of the vapor and all of the liquid is recycled back into the system. The next line added is the vapor –liquid equilibrium line for a binary system which is found experimentally. By moving step-wise between the two lines we can find the number of theoretical plates needed for a specific separation of a binary mixture and the liquid and vapor composition at any point in the distillation column. These stage lines can be seen in Figure 3. Unit Operations ChE-382 Group No. 3 p. 9 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago Figure 4: McCabe-Thiele diagram with operating lines and feed line added. In practice, we want to draw a purified product out of the column in the form of either a distillate (top of the column) or the bottoms product (bottom of the column) or both. This requires a column that operates at a partial reflux and/or a partial boil-up ratio. This means that we cannot use the x-y line for such a column. In Figure 4 we see the addition of a line for reflux ratio (slope L/V) and a line for the boil-up (slope L’/V’). The boil-up line’s slope increases as we increase the amount of bottoms product that we remove from the system. Subsequently, as we remove more distillate as product we decrease the slope of the top operating line, which is the line for the reflux ratio. By changing the amount of liquid re-boiled, which is liquid returned to the column as a vapor, or by changing the amount of vapor refluxed, returned to the column as a Unit Operations ChE-382 Group No. 3 p. 10 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago liquid, we change the number of theoretical plates necessary for a given separation. The q (quality) line in Figure 4 is the feed line which is the composition of the stream entering the distillation column. We can see from the diagram in Figure 4 that where that line intersects with the two operating lines is the feed stage, or the tray at where the incoming stream enters. The quality q is defined as: π= πΏΜ −πΏ πΉ ≈ π»−βπ (3) π»−β Where: ο· π [=] quality of the feed (dimensionless) ο· πΏΜ [=] liquid flow rate below the feed (mol/hr) ο· L [=] liquid flow rate above the feed (kJ/kg) ο· F [=] feed flow rate (mol/hr) ο· H [=] saturated vapor enthalpy of feed (kJ/kg) ο· hf [=] enthalpy of feed (kJ/kg) ο· h [=] saturated liquid enthalpy of feed (kJ/kg) The feed line can then be defined as: π π π₯+ πΉ ) π−1 1−π π¦=( (4) Where: ο· y [=] vapor mole fraction of methanol (dimensionless) ο· q [=] quality of the feed (dimensionless) Unit Operations ChE-382 Group No. 3 p. 11 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago ο· x [= ] liquid mole fraction of methanol (dimensionless) ο· ZF [=] mole fraction of methanol in feed (dimensionless) The top operating line is defined as π¦= πΏ0 π· πΏ (1+ 0 ) π· π₯ + (1 − πΏ0 π· πΏ 1+ 0 π· ) π₯π· (5) Where: ο· y [=] vapor mole fraction of methanol (dimensionless) ο· L0 [=] liquid reflux rate into column (mol/hr) ο· D [=] distillate flow rate (mol/hr) ο· x [=] liquid mole fraction of methanol (dimensionless) ο· xD [=] mole fraction of methanol in distillate (dimensionless) The McCabe-Thiele method is widely used for binary mixtures. When dealing with multicomponent mixtures addition assumptions and calculations are necessary. There are a number of issues that need to be watched when sizing a continuous distillation column properly. They are: Foaming- foaming occurs when the gas passing upward causing the liquid to bubble excessively. While this normally depends on the properties of the liquid, it can also be caused by improper tray designs. Entrainment- Occurs when the vapor velocity is too high due to a column having too small of a diameter. This causes liquid to be held up at the trays and does not allow proper circulation of the liquid. Unit Operations ChE-382 Group No. 3 p. 12 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago Weeping- Occurs when the vapor velocity is too low. When the velocity drops too low, liquid starts to fall, or weep, through the holes in the plates. This does not allow enough liquid to get to the re-boiler which means that the entire column needs to be shut down and re-started. Flooding- This occurs when liquid entrainment become too severe. This causes a large pressure drop in the column and contaminates the distillate. 3. Apparatus Unit Operations ChE-382 Group No. 3 p. 13 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago Figure 5 represents the Binary Distillation apparatus without any of the support beams. The column is constructed of a round bottom flask (7) which holds a methanol- water solution. This solution is distilled by applying electrical energy to the heating jacket (24) which heats up the solution causing the volatile liquids and gases to rise. Total reflux occurs in the system by feeding cooling water (10) into a cooling coil (16) at the top of the column. Thermocouples (T1-T9) measure the temperatures at various points in the system and output them onto an electrical temperature monitor (21). Liquid and gaseous samples are taken at ever one of the 6 stages via sample ports (L1-L6 &G1-G6) and tested with a Refractometer to calculate the density and eventually the composition of the samples. Unit Operations ChE-382 Group No. 3 p. 14 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago Table 1: Binary Distillation Apparatus Summary (For Figure 5) Component No. Apparatus Component 1 Water Connection for Apparatus Manufacturer Description Notes/Safety ChE Unit Ops Water supplied by lab 2 Water Supply Valve MCD Used to load water into Storage Tank. 3 Funnel Na 4 13 Gallon Feed Tank Nalgene 5 Feed Supply Valve Nalgene 6 Three way valve Swagelok Used to ease filling of solutions Used to load 5 mol% ethanol solution before filling flask Only open when filling round bottom flask Valve is labeled to either drain, fill or prevent loss 7 Round Bottom Flask Na 8 Valves & Sample Ports Nalgene 9 Glass sections of column Na 10 Cooling water Connection for Apparatus. ChE Unit Ops Cooling water is supplied by lab 11 Cooling Water Supply Valve Cold Chicago Faucets Used to start flow of cooling water into system. 12 Cold Water Temperature Gauge Marsh Instrument Company Measures the temperature of the incoming cooling water. Range: 0-60 oF, Increments: 1 oF 13 Cooling Water Apparatus Valve Cold Chicago Faucets Used to start flow of cooling water into apparatus. Check for leaks and cracks. Always open when distillation column is in operation to supply ample cold water. 14 Cold Water Rotameter (Measures in liquid GPM) Schutte & Koerting Co. 15 Temperature Gauge Moeller 16 Cooling Coil Na 17 Temperature Gauge Weksler 18 Cold Water Drain Valve Cold Chicago Faucets 19 Heat Controller Na 20 Heat Supply for Heating Jacket Powerstat Used to hold the Methanol-water solution Located at different heights along the column to obtain experimental samples Each section at collection point is pieced together. The top piece contains the cooling water tubes and a narrowed opening so that total reflux can be assumed. Measures flow of cooling water into cooling coil. (Measures in liquid GPM) Measures temperature of cooling water before entering cooling coil in degrees Celsius. Coil where heat transfer takes place between distillate and cooling water to provide total reflux. Measures temperature of cooling water exiting column in degrees Celsius. Allows cooling water to exit system and drain. Used to turn on heaters individually including heating jacket. Conducts electricity and passes it into the heating jacket Unit Operations ChE-382 Group No. 3 p. 15 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Check for leaks and cracks. Clean up spills immediately. Check for leaks and cracks that could cause malfunctions. Do not overfill Check for cracks and overall structural integrity. Check for leaks and cracks that could cause malfunctions. Check for leaks and cracks that could cause malfunctions. Do not heat when empty!! Ensure flask is cooled before adding any fluid Look for any leaks or cracks before adding solution or heat. Check for cracks and leaks that could cause malfunctions. Examine entire column for any leaks, cracks or anything that could fracture structural integrity Always open when distillation column is in operation to supply ample cold water. Always open when distillation column is in operation to supply ample cold water. Check for leaks and cracks that could cause malfunction Check for any leaks or cracks that could cause malfunctions. Check for any leaks or cracks that could cause malfunctions. Check for any leaks or cracks that could cause malfunctions. Check for any leaks or cracks that could spill water Check for any leaks or cracks that could cause malfunctions. Examine for loose wires or malfunction Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago 21 Digital Temperature Monitor Monogram 22 Thermocouples Omega Power switch for heater Square D Safety Switch Electric Heating Jacket Na 23 24 Component No. T1 T2 T3 T4 T5 T6 T7 T8 T9 Check for loose wires or malfunctions that may be causedby leaks. Displays temperature taken at the different thermocouples. Measure temperature at different stages in distillation column. Reading outputs on Digital temperature display Controls the electricity passing into the electric heating jacket turning it either on or off. Heats the bottom of the distillation column electrically. Make sure probes are secure and allow no leaks to touch them. Do not turn on unless cold water is flowing. Monitor temperatures and adjust as needed. Very HOT. Exercise extreme caution during operation. Table 2: Binary Distillation Thermocouple Summary (For Figure 5) Apparatus Component Manufacturer Description Notes/Safety Round Bottom Flask Thermocouple Omega Measures temperature of solution in round bottom flask. Stage 1 Thermocouple Omega Measures the temperature at stage 1. Stage 2 Thermocouple Omega Measures the temperature at stage 2. Pay close attention to temperature so column does not overheat. Stage 3 Thermocouple Omega Measures the temperature at stage 3. Pay close attention to temperature so column does not overheat. Stage 4 Thermocouple Omega Measures the temperature at stage 4. Pay close attention to temperature so column does not overheat. Stage 5 Thermocouple Omega Measures the temperature at stage 5. Pay close attention to temperature so column does not overheat. Stage 6 Thermocouple Omega Measures the temperature at stage 6. Pay close attention to temperature so column does not overheat. Thermocouple above Stage 6 Omega Measures the temperature above stage 6. Pay close attention to temperature so column does not overheat. Thermocouple placed before cooling coil Omega Measures the temperature of the cooling water before it is introduced into the cooling coil. Pay close attention to temperature so column does not overheat. Unit Operations ChE-382 Group No. 3 p. 16 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Make sure probe is secure and tube is filled with white oil. Pay close attention to temperature so column does not overheat. Spring 2010 3/11/2010 Binary Distillation Component No. University of Illinois at Chicago Table 3: Binary Distillation Sample Port Summary (For Figure 5) Manufactur Apparatus Component Description er Notes/Safety Gas sample collected from the stage 1 sample plate via a sealed tube and release valve (collection port). Gas sample collected from the stage 2 sample plate via a sealed tube and release valve (collection port). Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. Na Gas sample collected from the stage 3 sample plate via a sealed tube and release valve (collection port). Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. G4 Gas Sample Collection Port for stage 4. Na Gas sample collected from the stage 4 sample plate via a sealed tube and release valve (collection port). Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. G5 Gas Sample Collection Port for stage 5. Na Gas sample collected from the stage5 sample plate via a sealed tube and release valve (collection port). Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. G6 Gas Sample Collection Port for stage 6. Na Gas sample collected from the stage 6 sample plate via a sealed tube and release valve (collection port). Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. L1 Liquid Sample Collection Port for stage 1. Na L2 Liquid Sample Collection Port for stage 2. Na Liquid sample collected from the stage 2 sample plate via a sealed tube and release valve (collection port). Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. L3 Liquid Sample Collection Port for stage 3. Na Liquid sample collected from the stage 3 sample plate via a sealed tube and release valve (collection port). Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. L4 Liquid Sample Collection Port for stage 4. Na Liquid sample collected from the stage 4 sample plate via a sealed tube and release valve (collection port). Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. L5 Liquid Sample Collection Port for stage 5. Na Liquid sample collected from the stage5 sample plate via a sealed tube and release valve (collection port). Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. L6 Liquid Sample Collection Port for stage 6. Na Liquid sample collected from the stage 6 sample plate via a sealed tube and release valve (collection port). Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. Gas Sample Collection Port for stage 1. Na G2 Gas Sample Collection Port for stage 2. Na G3 Gas Sample Collection Port for stage 3. G1 Liquid sample collected from the stage 1 sample plate via a sealed tube and release valve (collection port). Unit Operations ChE-382 Group No. 3 p. 17 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Keep watch for leaks. Close when not in use. Keep samples cold by collecting samples with ice. Clean up any spillage immediately. Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago 4. Materials and Supplies Table 4: Binary Distillation Materials and Supplies Summary Material Name Manufacturer Tap Water Lake Michigan White Oil Chevron Superla 5 Description/ Info Notes/Safety Used in preparation of Methanol-water solution. Used inside the temperature probe tube within the round bottom flask. Clean up any spillage immediately. Clean up any spillage immediately. Do not ingest Will vaporize at room temperature and is toxic. Do not allow to escape through top of distillation column by circulating ample cold water. Clean up any spillage immediately. Do not ingest Clean up any spillage immediately. Try to keep samples cold to ensure vaporization does not occur. Methanol Aldrich Chemical Company 99.8% Methanol used to prepare water-methanol solution that is then fed into the distillation column. Ice From Unit OPS Lab Used to cool test tubes to ensure vaporization does not occur in samples. Na Used to transfer samples from apparatus to test tubes. Clean up any glass breakage immediately if it occurs. Kimble Glass Company Used to hold test solutions to use in Refractometer for sample results. Clean up any glass breakage immediately if it occurs. Micropipette Test Tubes Refractometer Graduated Cylinder Task wipers Bausch & Lomb Nalgene Kimberly Clark Professional Used in the experiment to obtain the refractive index that can be used to calculate density once calibrated. Used to measure Methanol and H2O to make the solution Used to clean equipment before, during, and after operations. Unit Operations ChE-382 Group No. 3 p. 18 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Use caution when handling. Calibrate before operation. Clean up any spillage immediately. Clean any mess or spillage promptly. Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago 5. Procedure Distillation Column Operation: 1. Make a 5 mol% methanol solution by mixing it with water in the 13 gallon feed tank (4) using graduated cylinders for measurement. Note: make sure not to over fill the round bottom flask (7). It only needs to be filled halfway or until the thermocouple (T1) is in contact with the mixture. 2. Pour the methanol-water mixture into the feed container and turn ON the three-way valve in order to fill up the round bottom flask. 3. Open the cooling water supply valves (11 & 13) to supply cooling water to the distillation column. 4. Start up the distillation column by turning on the power supply switches (22) to supply heat to the distillation column. 5. Record the temperature of each of the thermocouples (T1-T9) located throughout the column by analyzing the digital temperature monitor (21). Also record the readings of the inlet cooling water temperature gauge (12), the outlet cooling water temperature gauge (17), the inlet cooling water pressure gauge (15), and the cooling water rotameter (14). 6. Make sure to collect the samples at each stage via the sample ports (G1-G6 &L1L6) of the distillation column and record the index of refraction by using the Refractometer. 7. After the distillation process is done, make sure to turn OFF the heater power supply, and the cooling water supply. Operating the Bausch & Lomb Abbe-3L Refractometer: 1. Make mixtures from 10% Methanol with water to 100% Methanol in small test tubes using a pipette. 2. Turn ON the Refractometer and the water heating system. 3. Open the prism assembly and remove the tissue. 4. Use a capillary tube to apply your liquid sample to the prism and close the prism assembly. Unit Operations ChE-382 Group No. 3 p. 19 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago NOTE: Be careful not to let the glass pipette tip touch the prism since this may scratch the prism glass. 5. Adjust the toric lens, so the light shines on the prism and look through the eyepiece. 6. Analyze the index of refraction of the sample: a. When the index of refraction of your sample is close enough, then you will see lighter region on the top and darker region on the bottom. b. If you cannot distinguish between these two regions then adjust the compensator scale dial of the machine and the toric lens until the dark and light region is completely separated. Note: it is an iteration process between adjusting the light and the focus wheel located on the front of the Refractometer. c. Once the clear distinguishable line between dark and light region is seen then press the momentary contact switch located on the left hand side of the machine until you see the scale. Then read off the index of refraction by looking at the top scale and the refined bottom scale. 7. After you have noted down the index of refraction, record the temperature by reading the temperature scale. 8. Make sure to clean the prism after testing each sample with a solvent and dabbing it with tissue. Unit Operations ChE-382 Group No. 3 p. 20 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago 8. Error Analysis Table 5: Sources of Uncertainty for Binary Distillation Lab Component Manufacturer Uncertainty Expected Description 100 mL Graduated Cylinder Nalgene ± 0.5mL The 100mL graduated cylinder measures liquid in increments of 1mL. The lines can be accurately read to within ± 0.5mL Electronic Balance Denver Instrument Co. ±0.01g Via the manufacture’s website, the scale used in this experiment has an associated uncertainty of ±0.01g ± 1 oF The temperature gauges measure in increments of 2 oF. By dividing the increments by 2 the expected uncertainty can be projected to be ± 1 oF. ± 0.1 oC The temperature monitor measures in increments of 0.1 oC. The readings are accurate when the column reaches steady state and according to the manufacturer’s website, are accurate to ± 0.1 oC. ±0.0001 The Refractometer measures the refractive index of a mixture in increments of 0.0002. However, the values can be read accurately to the nearest 0.0001. Weksler Temperature Gauges & Moeller Digital Temperature Monitor Monogram Refractometer Bausch & Lomb Listed above are components of the binary distillation lab whose specific uncertainties would directly affect our recoded data. If these uncertainties were to occur, they would affect our overall results by adding specific uncertainties to them. Unit Operations ChE-382 Group No. 3 p. 21 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010 Binary Distillation University of Illinois at Chicago 9. References 1. Wankat, Phillip C.; Separation Process Engineering, second edition, Prentice Hall, 2007 2. R. Bird, W. Stewart, E. Lightfoot. Transport Phenomena. Wiley, 2006 3. Towler, G. and Sinnott, R., Chemical Engineering Design: Principles. BH, 2008 4. W. McCabe, J. Smith, and P. Harriot, Unit Operations of Chemical Engineering. 7th ed. McGraw-Hill, 2005. 5. M. J. Moran, H. N. Shapiro, Fundamentals of Engineering Thermodynamics. 5th ed. Wiley, 2004 6. Edited by Don Green; John Perry’s Chemical engineering Handbook, seventh edition, McGraw-Hill, New York, 1997 7. Dean, John A., “Lange's Handbook of Chemistry,” 15th edition. New York, NY: John Wiley & Sons, Inc., 1998. 8. Felder, Richard M., and Ronald W. Rousseau. Elementary Principles of Chemical Processes. 3rd ed. New York, NY: John Wiley & Sons, Inc., 2000. 9. Bennett C.O., Myers J.E., Momentum, Heat, and Mass transfer. 3rd edition New York. McGraw-Hill, 1982. 10. Wikipedia: http://en.wikipedia.org/wiki/Distillation 11. Website: http://lorien.ncl.ac.uk/ming/distil/distil0.htm Unit Operations ChE-382 Group No. 3 p. 22 Shah, Farr, Georg, Ogiefo, Khatib, Alkhateeb Spring 2010 3/11/2010