Preparation of Piglet Incubator

advertisement
SOP Number: CP029.00
Title: Preparation of Piglet Incubator
Revision No: Replaces:
00
N/A
Author: Kristin Smith
I
Date in effect: Page:
7/8/2010
Page 1 of 9
MAL Director: Dr. Peggy Borum
FSHN Chair: Dr. Neil Shay
PURPOSE
Incubators serve to maintain stable temperature and humidity levels suitable to the survival and
health of the piglet.
II
SCOPE
This procedure provides instructions for the preparation of piglet incubators before an
experiment. The incubators need to be clean and maintained in an aseptic environment for the
experiment.
III
RESPONSIBILITIES
It is the responsibility of Metabolic Assessment Laboratory personnel to follow this procedure.
It is the responsibility of supervisory personnel to ensure compliance with this procedure and to
train employees and students responsible for performing this procedure. Students will report
accidents to the principal investigators immediately.
IV
REFERENCES
N/A
V
REAGENTS AND MATERIALS
V.A. Disinfectant cleaning solution of a mixture of 4 oz of disinfectant to 2 gallons of water is
used to clean incubator.
V.B. Autoclaved piglet floors.
VI
EQUIPMENT
VI.A. The piglet incubator is made of plexiglass with an inside base of 26.25 inches in length,
14.5 inches in width, and walls that are 13.5 inches in height. A 10 x 26 inch hinged
front door provides easy access to incubator.
VI.B. The piglet incubator solid door(s) can be replaced with a 10x26 inch bar grid door, which
is hinged to provide easy access to incubator.
VII
SAFETY PRECAUTIONS
VII.A. Members of the MAL have been trained extensively in the procedures described in this
SOP.
VII.B. Members of the MAL have completed animal contact training in order to work in the
PNICU.
 Uncontrolled Copy
Document1
CONTROLLED DOCUMENT-DO NOT DUPLICATE
 Controlled Copy No._______
SOP Number: CP029.00
Title: Preparation of Piglet Incubator
Revision No:
00
VIII
Replaces:
N/A
Date in effect:
7/8/2010
Page:
2 of 9
DEFINITIONS
VIII.A. Standard Operating Procedure (SOP) – Standard Operating Procedure is a document that
provides instructions for completing a specific task in the lab.
VIII.B. Metabolic Assessment Laboratory (MAL) – The Metabolic Assessment Laboratory is the
laboratory that will use this SOP.
VIII.C. Piglet Neonatal Intensive Care Unit (PNICU) – The PNICU is a unit where the piglet is
monitored by 24 hour care and routine check-up parameters using PNICU SOPs
conducted by the MAL.
VIII.D. Piglet Incubator—The Piglet Incubator is an infant incubator that is enclosed, lined with
piglet floors, and temperature and humidity regulated.
Piglet Incubator
VIII.E. Incubator Bedding Tray – The bedding tray is the bottom of the incubator where the
piglet and other items may be placed. See picture below for description.
 Uncontrolled Copy
Document1
CONTROLLED DOCUMENT-DO NOT DUPLICATE
 Controlled Copy No._______
SOP Number: CP029.00
Title: Preparation of Piglet Incubator
Revision No:
00
Replaces:
N/A
Date in effect:
7/8/2010
Page:
3 of 9
Incubator Bedding Tray
VIII.F. Incubator Deck Rest – The Hood Assembly rests on a deck gasket that fits on the outer
limits of the deck rest. The Hood Assembly must at all times rest on the deck gasket. See
picture below for description.
 Uncontrolled Copy
Document1
CONTROLLED DOCUMENT-DO NOT DUPLICATE
 Controlled Copy No._______
SOP Number: CP029.00
Title: Preparation of Piglet Incubator
Revision No:
00
Replaces:
N/A
Date in effect:
7/8/2010
Page:
4 of 9
VIII.G. Incubator Deck Gasket – The deck gasket fits on the outer limits of the deck rest. See
picture below for description.
Deck Gasket
VIII.H. Incubator Micro Oxygen Filter – The Micro Oxygen Filter filters all air drawn into the
piglet incubator. The Micro Oxygen Filter must be checked before each experiment. Clean
or change as needed. See picture below for description.
Micro Oxygen Filter
 Uncontrolled Copy
Document1
CONTROLLED DOCUMENT-DO NOT DUPLICATE
 Controlled Copy No._______
SOP Number: CP029.00
Title: Preparation of Piglet Incubator
Revision No:
00
Replaces:
N/A
Date in effect:
7/8/2010
Page:
5 of 9
VIII.I. Incubator Power Pack – The control power unit uses a heater to warm the piglet.
Temperature may be controlled manually using the air temperature control knob. See
picture below for description.
Power Pack
VIII.J. Incubator Humidity Fill Tube – The Humidity Fill Tube is used to pour water into the
humidity reservoir. It also serves as an indicator as to the level of water in reservoir. To
drain water from reservoir, gently turn Humidity Fill Tube upside down. See picture below
for description.
Humidity Fill Tube
 Uncontrolled Copy
Document1
CONTROLLED DOCUMENT-DO NOT DUPLICATE
 Controlled Copy No._______
SOP Number: CP029.00
Title: Preparation of Piglet Incubator
Revision No:
00
Replaces:
N/A
Date in effect:
7/8/2010
Page:
6 of 9
VIII.K. Incubator Humidity Reservoir Tank and Baffle – The Humidity Reservoir Tank holds
water to provide proper humidity in the incubator. A Baffle is used to stabilize water in
tank reservoir. See picture below for description.
Humidity Reservoir Tank and Baffle
VIII.L. Incubator Humidity Control Knob – The Humidity Control Knob on the left side of the
Hood Assembly controls the level of humidity inside the incubator. The knob must be
pushed in to come into contact with control place located on deck rest. See picture below
for description.
Humidity Valve (Control Plate)
 Uncontrolled Copy
Document1
CONTROLLED DOCUMENT-DO NOT DUPLICATE
 Controlled Copy No._______
SOP Number: CP029.00
Title: Preparation of Piglet Incubator
Revision No:
00
Replaces:
N/A
Date in effect:
7/8/2010
Page:
7 of 9
VIII.M. Incubator Entry Port Door(s) – The Entry Port Door(s) provide arm access to the
incubator from the front, side, and rear. Doors are latched to provide easy entry/exit. One
larger hinged entry door on front provides greater access. See picture below for description.
Entry Port Doors
VIII.N. Piglet Floor—The Piglet Floor is 7-8 sheets of pre-cut newspaper wrapped in butcher
paper and then sterilized to line the floor of the incubator.
VIII.O. Piglet Toys—Piglet Toys include balls that have been disinfected for piglet play.
VIII.P. Teddies—Teddies are sterilized towels that serve as blankets for the piglets.
IX
PROCEDURE
IX.A. Preparation of Incubator.
IX.A.1.
Before cleaning PNICU piglet incubator, unplug the power cord from wall receptacle
and remove micro air filter. Note: Filter must not get wet.
IX.A.2. Use handle on back of plexiglass hood to lift hood forward.
IX.A.3.
Remove incubator bedding tray.
IX.A.4.
Lift the deck rest and remove (deck must never rest on deck gasket).
IX.A.5.
Wash the plexiglass incubators.
IX.A.5.a.
Wash with disinfectant cleaning solution. Use a soft cloth to clean
because the plexiglass will scratch very easily.
IX.A.5.b.
Rinse the plexiglass incubators with tap water.
IX.A.5.c.
Never use alcohol on the incubators because it could result in cracks or
fissures to the plexiglass.
IX.A.5.d.
Clean piglet incubator humidifier air reservoir with disinfectant cleaning
solution.
IX.A.5.e.
Check the micro oxygen filter for cleanliness and change if necessary.
 Uncontrolled Copy
Document1
CONTROLLED DOCUMENT-DO NOT DUPLICATE
 Controlled Copy No._______
SOP Number: CP029.00
Title: Preparation of Piglet Incubator
Revision No:
00
Replaces:
N/A
Date in effect:
7/8/2010
Page:
8 of 9
IX.B. Clean and check power pack.
IX.B.1.
Unplug the power cord from the power pack on the right side of incubator, unscrew
knobs from front panel, and pull entire unit out using great care.
IX.B.2.
The heater must not be touched until it has cooled for at least 20 minutes.
IX.B.3.
Remove lid from heater coil and clean blades with air.
IX.B.4.
Remove oxygen intake tube (turn and pull forward) and clean with disinfectant and a
clean cloth.
IX.C. Clean the humidity fill tube located on left front below hood.
IX.C.1. Unscrew the humidity fill tube and clean tube. Note: Use caution because they are
easily broken.
IX.D. Clean the humidity reservoir tank and baffle with disinfectant cleaner and soft cloth.
IX.D.1.
Remove the humidity valve (control plate) from left inside hood assembly by
removing two retaining thumbscrews. Clean with disinfectant cleaning solution and
soft cloth.
IX.D.2.
Replace port sleeves on left side of incubator.
IX.D.3.
Carefully remove entry port assembly retaining rings on front of and back of incubator
and clean with a disinfectant cleaning solution.
IX.D.4.
Place a squirt bottle of saline and 70% alcohol, and box of KimWipes on top shelf of
incubator, for use during 24-hour care.
IX.D.5.
Place two autoclaved piglet floors on the bottom of each incubator so that they
overlap,f or use during 24-hour care.
IX.D.6.
Put an autoclaved teddie (towel or linen) in each incubator, for use during 24-hour
care.
IX.D.7.
If needed, set a baby bottle nipple (that has been autoclaved in water) in the bottle
holder on the outside of the incubator if available or tape the nipple to the rear inside
wall of the incubator to be used as a pacifier, for use during 24-hour care.
IX.D.8.
Place a wiffle ball that has been soaking in a disinfectant solution on the piglet floor,
for use during 24-hour care.
IX.E. Preparation of Cabinets on which the Incubator rests.
IX.E.1.
Wash the painted base cabinets with disinfectant solution.
IX.E.2.
Clean the plexiglass medical record holder with disinfectant solution. Use a soft cloth
to clean because plexiglass will scratch very easily. Place the clean plexiglass medical
record holder on the side of the cleaned incubator.
IX.E.3.
Clean the infusion pump if needed and place to the side of the incubator.
IX.E.4.
Place the medical record in the plexiglass medical record holder.
IX.E.5.
Tape an identification tag to the plexiglass medical record holder of each piglet.
IX.E.6.
Place one package of autoclaved piglet floors, one package of autoclaved teddies, one
box of KimWipes, and one roll of paper towels into the cabinet under the incubator.
 Uncontrolled Copy
Document1
CONTROLLED DOCUMENT-DO NOT DUPLICATE
 Controlled Copy No._______
SOP Number: CP029.00
Title: Preparation of Piglet Incubator
Revision No:
00
X
Replaces:
N/A
Date in effect:
7/8/2010
Page:
9 of 9
ATTACHMENTS
X.A. For a list of materials and their locations, refer to SOP# CG098.00.
 Uncontrolled Copy
Document1
CONTROLLED DOCUMENT-DO NOT DUPLICATE
 Controlled Copy No._______