Beam through entire system Align HFP chamber to beam Collect images of beam on diff pumping YAG Collect images of beam on beam viewing paddle Align Diagnostics chamber to beam Collect images of beam on upstream diff pumping YAG Collect images of beam on downstream HiRes beam screen Align laser to beam - rough FEL beam OFF Insert mirror Align beam, collecting images as necessary Remove mirror before beam back ON Pulse energy monitor Offline testing of pressure, magnet current, detector voltages, data acq Setup with appropriate pressure – observe signal Correlate signal with FEE pulse energy monitors Attenuate beam with gas attenuator & verify operation Vary pressure and tabulate response Shutter operation View beam at Diagnostics chamber downsream HiRes beam screen Beam OFF, shutter open Insert shutter to nominal position Beam ON, find beam on YAGs & center beam in shutter Close shutter – observe beam screen Pulse shutter – use pulse energy monitor to pick 1 Hz ? Slit operation View beam at Diagnostics chamber downstream HiRes beam screen Monitor pulse energy with pulse energy monitor Scan one slit blade at a time to find center of beam Mask beam to varying degrees, watch for intensity variations between FEE and AMO monitors Open slits back open KB optics Focus optics for HFP interaction point Insert KB optics into beam (nominal position) Align chambers to KB optics Heavily attenuated beam – check alignment & adjust Imprint beam into beam focus verification paddle & view with microscope Vary focal length between HFP and diagnostics chamber – dble check alignment Repeat beam imprinting after focal length variation Dynamic beam focus studies using Jacek’s set-up HFP iTOF Apply voltages to iTOF lenses Apply voltages to iTOF detector (dark counts, off-line) Acquire spectra using residual gas, either from background pressure or gas jet leakage – optimize voltages Vary acceleration, extraction voltages to define operational space Gas jet optimization & operation (see below) Pulsed extraction voltage operation – effects on spectra HFP gas jet st Optimize throughout of gas by varying position of gas jet (off-line 1 ) Optimize timing of gas nozzle valve to LCLS using iTOF Vary sample density via nozzle withdrawl, backing pressure variation to see effect on signal, alignment, etc HFP electron TOFs Apply voltages to lenses Apply voltages to eTOF detectors (dark counts, off-line) Measure low KE electrons (Ne photoelectrons) Optimize positions of all five TOFs Optimize voltages on detectors Optimize voltages on lenses (iterating these three) Measure high KE electrons (Ne Augers) Optimize voltages on lenses for maximum resolution, count rate Measure moderate KE electrons (Ar Augers ?) Optimize pressure for given focal volume – see broadening from charge ? iTOF pulsed extraction operation – timing & effect on KE resolution HFP Focus verification paddle Remove iTOF & top eTOF to install focus verification paddle Ensure limits are set to prevent damage to adjacent eTOFs Imprint unfocussed beam and view with the microscope Imprint focused beam & view with microscope Move out of focus (chamber & paddle motion) & view Optimize focusing HFP vTOF (velocity map imaging TOF) Replace iTOF with vTOF Acquire signal using atomic sample (low extraction voltage) Optimize position, voltages Change to molecular sample (N2?) & acquire signal (high extraction voltage) Optimize position, voltages Try pulsed extraction voltage mode – image ? Coincident measurements of N2 with vTOF Diagnostics Beam Screens Position downstream screen in beam Measure samples of beam screen info for statistical analysis Software available for position and extent determination? Position upstream screen in beam – still see it on downstream screen ? Measure samples of beam screen info from both screens Diagnostics Magnetic Bottle Current applied to solenoid – field ? Voltages on detector – dark counts & image ? Look for signal with high residual pressure from needle Optimize position of the permanent magnet – looking at image Optimize solenoid magnets Optimize gas delivery Spectra of low KE electrons Spectra of high KE electrons Test retarding field – tune solenoid magnets Measure KE of electrons versus photon energy Change focus to diagnostics position & repeat electron measurements Diagnostics gas delivery Optimize position of the needle in the beam Optimize backing pressure for different foci Effect of needle position on resolution of spectrometer Laser Bring laser beam into diagnostics chamber Focus into interaction region of magnetic bottle Overlap with FEL beam spatially/temporally Measure side-bands – varying intensity & timing nd 2 or third harmonic ? Bring laser beam into HFP chamber Focus into interaction region of electron spectrometers Overlap with FEL beam spatially/temporally Measure side-bands – varying intensity & timing nd rd 2 or 3 Harmonics & repeat Diagnostics gas detector studies (Moeller) Attach another gas detector to the instrument Measure response at variety of different attenuator settings HFP focusing studies (Jacek) Replace ion TOF with camera mount Install YAG screen on diagnostics paddle Attenuate LCLS beam Make measurements as function of position Tune focusing of optics Change focus to diagnostics chamber & measure Return focus to HFP chamber & verify repeatability HFP focusing studies (wavefront sensor) Attenuate beam to prevent damage Focus beam in HFP interaction region Measure wavefront Tune optics through iteraction Verify with YAG screen diagnostic HFP damage studies (Hau Riege) Install sample(s) on beam focus paddle Expose to single or multiple shots at various locations Change attenuation & repeat Change samples & repeat HFP mTOF Replace iTOF with mTOF Acquire signals using atomic sample (low extraction voltage) Optimize position, voltages Change to molecular sample (N2?) & acquire signal (high extraction voltage) Optimize position, voltages Try pulsed extraction voltage mode – image ? Coincident measurements of N2 with mTOF