Development of an Active Electron Polarized Target 1) Material properties Magnetic susceptibility is depicted on pic.1. Picture 1. Suscettivita’ magnetica a 100 Oe (unita’ emu per mole Gd2SiO5 Oe-1) Inverted magnetic susceptibility is given at pic.2. Picture 2. Fitting Curie. Misurati 8.02 B /at Gd (previsti 7.94 B /at Gd). Picture 3. Assenza di anisotropia magnetocristallina. Suscettivita’ a 100 Oe per H parallelo ai tre assi di un campione a forma cubica. M (emu/mol GSO) SQUID_0527 scheggia 0.0030 g M(H) 8.000 10 4 7.000 10 4 6.000 10 4 5.000 10 4 4.000 10 4 3.000 10 4 2.000 10 4 1.000 10 4 0.000 10 0 M teo M (emu/mole) 0 10000 20000 30000 40000 50000 60000 B (Oe) Picture 4. M(H) curva sperimentale confrontata con quella teorica corrispondente a momenti Gd liberi. M (emu/mol GSO) M teo con B SQUID_0521 18.36.00 07/03/2011 12000.00 M (emu/mol GSO) 10000.00 8000.00 6000.00 4000.00 2000.00 0.00 0 10000 20000 30000 40000 50000 B=H-0.26 4pai M+M (Oe) Picture 5. M(B) per GSO confrontata con la curva prevista per Gd liberi a 77 K. 2) GSO behavior at cold under magnetic field made by PMT Spectral measurements of 511 keV source (22Na) with PMT under different conditions were done. Measurements were held at room temperature (299 K) without and with magnetic field (B 1 Tl) and at cold temperature (13.3 K) with and without magnetic field. The scheme of this measurements is given at pic. 6. H.V. -1125 V Amplifier MCA PC Light pulser PMT Quartz bar Magnetic shields Fiber optics Quartz bar magnet 22 GSO Na Magnet Picture 6. The schematic view of experimental set-up (left) and it’s photo with opened metallic cover (right). All measurements were done with high voltage on PMT equal to -1125V, shaping time of the amplifier equal to 6 microseconds, fine gain and coarse gain of the amplifier are 1.5 and 2000 respectively. Also the pulse was given to PMT to check the gain under different conditions. Each time 2 measurements were done: the spectrum of 511 keV source and the PMT response to a light pulse. Room temperature measurements without and with magnetic field Results for measurements at room temperature without magnetic field are following: a Counts Counts 1340 30 1590 20 0 b 15 0 1000 2000 1000 2000 Channel Channel Picture 7. Measurements at 299 K without B-field; a) spectrum of 22Na, b) PMT response to a light pulse. Results for measurements at room temperature with magnetic field (1 Tl) are following: 20 250 a 1600 b 1460 Counts Counts 200 10 150 100 50 0 0 0 1000 2000 Channel 0 1000 2000 Channel Picture 8. Measurements at 300 K with B 1 Tl; a) spectrum of 22Na, b) PMT response to a light pulse. The position of the pulse has changed and this means that the gain of PMT has slightly changed too. If to take this change into account we can renormalize the position of the source. When this renormalization is done it turns that the spectrum of 22Na moved to lower channel number by factor 1.08. Result: Magnetic field makes no changes in the system and the light output of GSO scintillator doesn’t change. Cold temperature measurements without and with magnetic field Results for measurements at cold temperature without magnetic field are following: a 200 b 640 1390 Counts Counts 40 100 20 0 0 0 500 1000 1000 Channel 2000 Channel Picture 9. Measurements at 13.7 K without B-field; a) spectrum of 22Na, b) PMT response to a light pulse. Results for measurements at cold temperature with magnetic field (1 Tl) are following: 120 60 a 680 1790 b Counts Counts 90 30 60 30 0 0 0 500 1000 Channel 1500 0 1000 2000 Channel Picture 10. Measurements at 13.4 K with B 1 Tl; a) spectrum of 22Na, b) PMT response to a light pulse. When all renormalizations like for room temp measurements are done it turns that the spectrum of 22Na moved to lower channel number by factor 1.21. This is a minor movement so… Results: a) Magnetic field makes no changes in the system at room and at cold temperature and the light output of GSO scintillator doesn’t change; b) Spectrum moves because emission spectrum of GSO goes red [M. Sekita and others, Optical studies of Ce-doped Gd2SiO5 single crystals] PMT time response: at room temperature with B-field 70 60 Amp, mV 50 40 30 20 10 0 1 2 3 4 5 Time, s Picture A. Inverted PMT signal at 300 K. Approximate decay time observed at room temperature is 200 ns. But it is not a decay time of GSO scintillator. Really observed = 0.2µs = (2sci + 2system), sci 56ns. This means that we can calculate that system = 192 ns. at cold temperature with B-field 120 100 Amp, mV 80 60 40 20 0 0 1 2 3 4 5 Time, s Picture B. Inverted PMT signal at 13.4 K. Approximate decay time observed at room temperature is 900 ns. Now we can calculate that sci@cold = 880 ns. We also report no observable changes (in 10% level) in PMT response without magnetic field. Result: Response signal became longer because of decreasing temperature. 4) APD Measurements: A) 6 keV cold testing + B) GSO spectral measurements A) Spectral measurements of 6 keV source with APD under different conditions were done. Measurements were held at room temperature (300 K) without magnetic field and at cold temperature (13.3 K) without magnetic field. The scheme of measurements that were held at room temperature is given at pic. 11. 6keV source APD PC MCA H.V. 1800 V Pre Amp Selective Amp Fast filter Amp Picture 11. The schematic view of experimental set-up and it’s photo with opened metallic cover. Measurement at room temperature was done with APD voltage equal to 1800 V, integration time (ORTEC Fast filter amplifier) equal to 500 ns, fine gain and coarse gain of this amplifier are 1 and 250 respectively. Also the pulse was given to Preamplifier to check the gain of the system. Result for measurement at room temperature is following: Counts 60 440 Source (6 keV) Peak 440ch: height = 38.24 center = 440 HWHM = 67.64 3029 Pulse: f = 20 Hz t = 1 s A = 200 mV 30 0 0 1000 2000 3000 Channel Picture 12. Spectrum of 6 keV source at 300 K with APD. Measurements with APD at cold temperature (13.8 K) were held with a small integration device between selective amplifier and fast filter amplifier. It is shown on pic. 13. 6keV source APD PC MCA H.V. Pre Amp Selective Amp Fast filter Amp 1800 V Picture 13. The schematic view of experimental set-up used for measurements at 13.8 K. Spectral measurements of 6 keV source were done for different APD voltages: from 1250 V to 1325 V with a step in 25 V. Resulting curve (Channel number VS APD voltage) is given at pic. 14. Channel number 2000 1750 Data: Data1_B Model: ExpGro1 Equation: y = A1*exp(x/t1) + y0 Weighting: y No weighting 1500 Chi^2/DoF = 770.759 R^2 = 0.99899 1250 y0 A1 t1 162.2659 8.5093E-13 37.52909 ±30.68923 ±1.6362E-12 ±2.03496 1000 750 500 250 1200 1225 1250 1275 1300 1325 APD Voltage Picture 14. Channel number VS APD voltage at 13.8 K. Also measurements with a pulse were done. Pulse’s frequency was 100 Hz, it’s width was 5 ms and it’s amplitude was (10-35) mV with 5 mV step. Such measurements were done also for coarse gain of fast filter amplifier equal to 50. The resulting plot (Channel number VS Pulse amplitude) is given here: coarse gain 100 coarse gain 50 2250 2000 y = 56.3x - 4.7 y = 23.5x + 2.9 Channel number 1750 1500 1250 1000 750 500 250 5 10 15 20 25 30 35 40 Pulse amp, mV Picture 15. Channel number VS Pulse amplitude at 13.8 K for coarse gain = 100 (red one) and coarse gain = 50 (green one). This measurements with pulse allowed us to check the APD gain, so channel numbers could be renormalized to gain. Resulting curve (Gain VS APD voltage) is given at pic. 16. 800 700 6 keV source APD at cold: Tplate = 13.8 K Integr.time = 500 ns coarse gain = 100 fine gain = 1 Gain 600 Data: Data1_B Model: ExpGro1 Equation: y = A1*exp(x/t1) + y0 Weighting: y No weighting 500 400 Chi^2/DoF = 26408.65802 R^2 = 0.99339 300 y0 A1 t1 131.14148 1.3357E-14 33.4475 ±121.77834 ±5.794E-14 ±3.58888 200 100 1200 1225 1250 1275 1300 1325 1350 APD Voltage Picture 16. Gain VS APD voltage at 13.8 K. Also the Gain of the preamplifier was calculated. GPreAmp = 0.8 mV/fC Result: APD works reliably well at low temperature with Gain ≈ 1000 but at voltage higher than 1400 V starts to have some problem we did not understand yet. B) APD+GSO spectral Measurements Spectral measurements of 662 keV source with a GSO crystal which was on top of APD were done. No optical grease connection between GSO crystal and APD. source Coarse g = 5 Fine g = 0.84 Sh.time = 500ns GSO APD H.V. 1800 V Coarse g = 250 Fine g = 1 Int.time = 500ns Diff.time = 500ns Pre Amp Fast filter Amp 1 PC MCA Amplifier 2 Picture 17. The schematic view of experimental set-up for APD+GSO measurements. Measurement was done at room temperature with APD voltage equal to 1800 V, shaping time1 (ORTEC Fast filter amplifier 1) equal to 500 ns, fine gain and coarse gain of this amplifier are 1 and 250 respectively; integration time2 (Amplifier 2) equal to 500 ns, fine gain and coarse gain of this amplifier are 0.84 and 5 respectively. Also the pulse was given to Preamplifier to check the gain of the system. Result for measurement spectrum of 662 keV source at room temperature is following: 662 keV source is about 1190 channel APD Volt = 1800V sh.time1=500 ns 600 Counts sh.time2=500 ns 400 200 0 0 500 1000 1500 Channel Picture 18. Spectrum of 662 keV source at 298 K with APD+GSO. APD voltage = 1800V Measurements at cold temperature were done with the same settings as one at room temperature (except APD voltage). Result for measurement spectrum of 662 keV source at cold temperature is following: 30 Source is about 1680 channel APD Volt = 1325V, coarse gain1 = 250 25 Counts 20 15 10 5 0 0 1000 2000 Channel Picture 19. Spectrum of 662 keV source at 15 K with APD+GSO. APD voltage = 1325V If to make a necessary renormalization of the position of the peak it turns that peak moves almost 300 channels to the lower channel number! Result: At low temperature spectrum moves 300 channels (25%) to the lower channel number. This means that the APD Gain decreases. But we also have to take into account that GSO spectrum is going red at cold [M. Sekita and others, Optical studies of Ce-doped Gd2SiO5 single crystals]. The quantum efficiency of APD for 400 nm is about 80%, and for 800 nm it is about 60%. So this mean that we lose approximately 20% of quantum efficiency of APD when moving to red emission. This can explain why the gain of the APD decreases about 25% at cold. Magnetization signal estimation We couple an inductance around the GSO crystal to measure the flux variation caused by a laser light pulse onto the crystal at cold ( 4 Kelvin )under magnetic field ( 1 T ). Our attempts to estimate an electric current: L i , where L is an inductance of a coil and i is an electric current. At the same time 0 , where - changing of magnetization of GSO sample, H – height of the sample. Thus we can say that i 0 1 L Next we can write that . , where is an energy of laser pulse that we provide C to the sample and C is a heat capacitance of GSO sample. Further we can say that const . So for i we can assume that i 0 L C . 2 We use 1Tl , 1mJ , 2.5 10 m , C ~ 10 2 J (at temperature = 10K). But we do not know exactly the Debye temperature of this crystal mol K 1 2 emu at 10K we estimated from your data as 10 , mol Oe K 4 We have to measure an inductance L but we estimate it as 10 H .