Kartik Vadlamani LM4/Dr. Milor Energy Without Bounds Technical

advertisement
Kartik Vadlamani
LM4/Dr. Milor
Energy Without Bounds
Technical Review of Ultrasound Imaging Applications in Medicine
Introduction
Ultrasound is a growing area of research with applications expanding into various
industries, having uses in supplying energy or penetrating a medium to measure reflective
signatures. In medical diagnosis and monitoring, the latter use is more prevalent and inexpensive
in comparison to other imaging methods such as MRIs and CT scans. This paper presents a
technical review of ultrasound imaging applications in medicine, discussing commercially
available products, underlying technology including techniques and characteristics, and
requirements for their implementation.
Commercial Applications
Ultrasound imaging has uses in various facets of medicine including the imaging of
organs and fetal development during pregnancy through sonography. Focused ultrasound is also
used for cancer treatment among other diseases. A variety of products using ultrasound
technology are available from manufacturers, however their pricing depends on specific
applications. Ultrasound devices used for imaging in medical diagnosis are expensive due to
health and safety concerns, and can range from $32,000 to $150,000 depending on size,
portability, and features included [1]. The smaller portable devices such as GE’s Vivid i
($32,000) tend towards the lower pricing range, but have lower imaging resolution, and exclude
features such as 4D imaging, unlike GE’s Voluson E8 BT08 ($135,000) or Phillip’ IU22
($100,000) which include these features.
There is a continuous effort to make these devices smaller, more portable, and cheaper
through means of using different applications of ultrasound and improving on the algorithms
implemented to produce images. Research is on going in using higher frequency focused
ultrasound (20 MHz and above) to image soft tissue more rapidly while still maintaining the
ability to obtain a high-resolution image [2]. Other research topics include improving the material
applied to the human body to reduce the loss factor involved in signal propagation through human
tissue.
Underlying Technology
Ultrasound refers to frequencies larger than 20 kHz. In medical imaging applications,
shorter wavelengths corresponding to human tissue as a propagation medium are used, resulting
in frequencies ranging from 2 - 10 MHz, but most commonly between 3 - 5 MHz [3]. A gel-like
material used to match the permittivity of human tissue, reducing the loss factor of the signal
when changing mediums.
Images are formed using pulse-echo principles where short pulses of ultra-sound waves at
these frequencies, and lasting less than 1 µs [4], are output by a transducer that then receives
echo pulses from reflections. Using known constants such as the propagation speed in a medium
(1540 m/s in human tissue) [3], and arrival time/angles of reflected pulses, a two-dimensional
image can be created indicating the distance and shape of an object. Beamforming techniques
common in signal processing are used to enhance and layer images to increase resolution and
produce 3D/4D imaging [5] by directionally or spatially focusing echo signals received from
tissue structures. Weighted delays are applied to these signals to indicate different tissue regions
and are summed to form a 3D image. Further algorithms are also applied to estimate attenuation
and scattering [6] of pulses to prevent destructive interference from occurring and provide greater
resolution to images [7], which are limited by the ultrasound frequencies necessary to image
human tissue.
Another imaging technique, used specifically to monitor blood flow, is the Doppler
imaging. The Doppler effect, which states moving objects exhibit changing frequency,
characterizes Doppler imaging, and this changing frequency results in a shift that can be
continuously compared to a reference signal to represent blood flow. Doppler shifts are
commonly used to image blood flow from ultrasound pulses at 5 MHz [3].
Implementation
Ultrasound imaging is realized through the use of electro-acoustic transducer arrays to
send and receive ultrasound pulses from various locations. A full system connects the array to a
variable gain amplifier and an analog-to-digital converter (ADC) to generate a usable signal for
processors to interpret [8]. DSPs handle beamforming and spectral/color/power Doppler
processes implemented in software. These processors require higher bandwidth at 1 GHz or
above. An operating system (OS) is necessary for the front-end processing algorithms to increase
resolution and is usually paired with the user interface on a processor such as an ARM9. PCBs
cost can be lowered by adding an LVDS interface between the ADC and DSP for beamforming,
reducing the number of interface lines into the DSP [8]. A gel-like material with permittivity
close to that of human tissue is also necessary in obtaining high-resolution images.
References
[1] Absolute Medical Equipment, “Featured Products in Ultrasound Equipment,”
absolutemed.com, Sept. 6, 2011. [Online]. Available: http://www.absolutemed.com/MedicalEquipment/3D-4D-Ultrasound-Machines. [Accessed: Sep. 6, 2011].
[2] Q. Huang, M. Lu, and Y. Zheng, “A Novel Method to Obtain Modulus Image of Soft Tissues
Using Ultrasound Water Jet Indentation: A Phantom Study,” IEEE Transactions on Biomedical
Engineering, vol. 54, no. 1, Jan., pp. 114-121, 2007.
[3] A. Dhawan, “Medical Imaging Modalities: Ultrasound Imaging,” in Medical Image Analysis,
2nd ed., New York: Wiley-IEEE Press, 2011, pp- 157-172.
[4] J. Baun, “Pulse Echo Imaging,” March, 2009. [Online]. Available:
http://www.jimbaun.com/4_pulseecho_imaging.pdf. [Accessed: Sept. 3, 2011].
[5] D. Kreindler, “The Future of Beamforming in Ultrasound,” medicalelectronicsdesign.com,
para. 2, Sept. 28, 2010. [Online]. Available:
http://www.medicalelectronicsdesign.com/article/future-beamforming-ultrasound. [Accessed:
Sept. 3, 2011].
[6] N. Duric, K. Hanson, L. Huang, C. Li, and Y. Quan, “Ultrasound Pulse-Echo Imaging with an
Optimized Propogator,” In Proc. IEEE International Conference on BioMedical Engineering and
Informatics ’08, 2008, pp. 281-285.
[7] L. Leija, R. Martinez, and A. Vera, “Ultrasonic Attenuation in Pure Water: Comparison
Between Through-Transmission and Pulse-Echo Techniques,” In Proc. Pan American Heal Care
Exhanges Conference ’10, 2010, pp. 81-84
[8] Texas Instruments, “Ultrasound System,” ti.com, Sept. 27, 2010. [Online]. Available:
http://www.ti.com/solution/ultrasound_system. [Accessed: Sept. 3, 2011].