Advances in Trace Element Analysis 2013 ACS Spring Meeting Workshop Art Fitchett and Fergus Keenan The world leader in serving science 1 Agenda • Ion Chromatography (IC) • High Pressure Ion Chromatography (HPIC) • Inductively Coupled Plasma (ICP) • ICP-OES • ICP-MS • IC-ICP-MS • Speciation 2 Why High Pressure Ion Chromatography • Remember UHPLC? • As the particle size decreases from 8µm to 4µm, column efficiency doubles • This drop in particle size increases the column pressure by 4x • Like HPLC, IC is moving toward smaller particle column technology • HPIC Instrumentation can now handle the pressure of these smaller particle columns, even at higher flow rates. 3 HPIC Theory Influence of the Particle Diameter on Pressure and Efficiency 1200 Theoretical Plate Height [µm] 100 1000 Column pressure [bar] 10 µm particles 5 µm particles 3 µm particles 2 µm particles Optimal flow rate for maximum separation efficiency / resolution 800 600 400 200 0 0 0 2 4 6 Linear Velocity u [mm/s] 8 10 0 2 4 6 8 Linear Velocity u [mm/s] Faster Flows for Faster Separations generate Higher Pressure Smaller Particles for Higher Efficiency generate Higher Pressure 4 10 HPIC System Specifications Format Capillary Microbore Standard Bore Flow Rate Range 0.001-0.100 mL/min in 0.001 µL/min increments Typical range: 5-20 µL/min 0.001-10 mL/min in 0.001 µL/min increments Typical range: 0.2-0.5 mL/min 0.001-10 mL/min in 0.001 µL/min increments Typical range: 1-2 mL/min Max. Pressure 5 5000 psi (eluent generation) 5000 psi (eluent generation) 5000 psi (eluent generation) 6000 psi (pump pressure range) 6000 psi (pump pressure range) 6000 psi (pump pressure range) Column i.d.s Supported 0.2-0.6 mm 1-3 mm 3-7 mm Yearly Eluent Usage (continuous operation) 5.25 L (10 µL/min) 131L (0.25 mL/min) 525 L (1 mL/min) HPIC System Advantage • HPIC systems + 4 µm particle-size columns deliver significant performance advantages • Smaller resin particle columns • Produce more efficient peaks • Impact chromatographic speed and resolution • Easier integration – more accurate and reliable results • Increase sample throughput without compromising data quality • Improved quality of analytical results 6 New High Efficiency Dionex IonPac 4µm IC Columns in Analytical and Capillary Formats 4 µm Ion-exchange columns with 4 µm particle-size Benefits • Smaller particles provide better performance • Faster run times with higher flow rates using 150 mm columns • Better resolution with standard flow rates using 250 mm columns SEM Image of 4 µm Supermacroporous Bead Applications • Anions in environmental waters • Organic acids in foods and beverages • Amines in chemical process solutions 10 5.5 5 µS 10 µS Minutes 40 High Resolution using the Dionex IonPac AS11-HC-4µm µS 0 0 Minutes 40 High Resolution using the Dionex IonPac CS19-4µm -0.5 0 Minutes Fast Run using the Dionex IonPac AS18-4µm Improved Resolution Provides Faster Runs and Better Results 7 3 Improved Separations using 4 µm Particle Size Capillary Columns 25 Thermo Scientific™ Dionex™ IonPac™ AG11-HC-4µm/AS11-HC-4µm 3600 psi 21 22 20 13 µS 6 8 12 14 9 2 17 16 19 11 4 5 7 10 1 3 24 26 27 29 23 25 15 18 28 0 Dionex IonPac AG11-HC/AS11-HC 2200 psi µS -15 0 6 12 18 Minutes 8 Eluent Source: Thermo Scientific Dionex EGC-KOH Eluent Generator Cartridge (Capillary) Gradient: Potassium hydroxide: 1 mM from 0 to 5 min, 1–15 mM from 5 to 14 min, 15–30 mM from 14 to 23 min, 30–60 mM from 23 to 31 min Flow Rate: 15 µL/min Inj. Volume: 0.40 µL Temperature: 30 °C Detection: Suppressed conductivity, Thermo Scientific™ Dionex™ ACES™ 300 Anion Capillary Electrolytic Suppressor, recycle mode 24 30 Peaks: 1. Quinate 2. Fluoride 3. Lactate 4. Acetate 5. Propionate 6. Formate 7. Butyrate 8. Methylsulfonate 9. Pyruvate 10. Valerate 11. Monochloroacetate 12. Bromate 13. Chloride 36 14. Nitrite 15. Trifluoroacetate mg/L 5.0 1.5 5.0 5.0 5.0 5.0 5.0 5.0 5.0 5.0 5.0 5.0 2.5 5.0 5.0 16. Bromide 17. Nitrate 18. Carbonate 19. Malonate 20. Maleate 21. Sulfate 22. Oxalate 23. Tungstate 24. Phosphate 25. Phthalate 26. Citrate 27. Chromate 28. cis-Aconitate 29. trans-Aconitate mg/L 5.0 5.0 --7.5 7.5 7.5 7.5 10.0 10.0 10.0 10.0 10.0 --10.0 Faster Run Times without Sacrificing Resolution Inorganic anions separation using a 4 µm capillary column 20 1 2 34 Column: 56 Eluent Source: Eluent: Col. Temp.: Inj. Volume: Detection: 7 30 µL/min, 2820 psi 25 µL/min, 2430 psi 20 µL/min, 2030 psi µS Peaks: 15 µL/min, 1570 psi 10 µL/min, 1140 psi -15 0 9 Dionex IonPac AS18-4µm, 0.4 × 150 mm Dionex EGC-KOH (Capillary) 30 mM KOH 30 °C 0.4 µL Suppressed Conductivity, Dionex ACES 300 5 Minutes 10 1. Fluoride 2. Chloride 3. Nitrite 4. Sulfate 5. Bromide 6. Nitrate 7. Phosphate 0.2 1 1 1 1 1 2 mg/L Fast Run on the Dionex IonPac AS18-4µm Column 2 5.5 3 4 1 Dionex IonPac AS18-4µm, 0.4 × 250 Peaks: 1. 2. 3. 4. 5. 6. 7. 5 6 µS 7 -0.5 0 10 Column: mm Eluent Source: Eluent: Flow Rate: Inj. Volume: Col. Temp.: IC Cube Temp.: Detection: 1 2 3 Minutes 4 5 Dionex EGC-KOH Cartridge (Capillary) 35 mM KOH 30 µL/min 0.4 µL 30 °C 15 C Suppressed conductivity, Dionex ACES 300, recycle mode Fluoride Chloride Nitrite Sulfate Bromide Nitrate Phosphate 0.2 mg/L (ppm) 0.5 1.0 1.0 1.0 1.0 2.0 Faster Run Times without Sacrificing Resolution Inorganic anions separation using a 4 µm Microbore column 3 70 6 8 2 4 1 Column: Instrument: 7 5 9 0.50 mL/min, 4200 psi Eluent Source: Eluent: Flow Rate: Inj. Volume: Column Temp.: Detection: 0.45 mL/min,3800 psi µS Peaks: 0.40 mL/min, 3300 psi 0.25 mL/min, 2200 psi -20 0 1 2 3 4 5 Minutes 11 6 7 8 9 Dionex IonPac AS18-4µm, 2 150 mm Thermo Scientific™ Dionex™ ICS-5000+ HPIC™ System Dionex EGC 500 KOH 23 mM Potassium hydroxide 0.25, 0.40, 0.45, and 0.50 mL/min 5 µL 30 °C Thermo Scientific™ Dionex™ ASRS™ 300 Anion Self-Regenerating Suppressor™, 2 mm, recycle 1. Fluoride 2. Chlorite 3. Chloride 4. Nitrite 5. Carbonate 6. Bromide 7. Sulfate 8. Nitrate 9. Chlorate 0.5 5.0 3.0 5.0 20.0 10.0 10.0 10.0 10.0 mg/L Isocratic Separation of Common Anions Using the Dionex IonPac AS18-4µm Column (4 ×150 mm) at Various Flow Rates Inorganic anions separation using a 4 µm Standard bore 7 3 column4 1.0 mL/min 6 8 Column: Dionex IonPac AG18-4µm/AS18-4um 6 2574 psi 2 µS 9 1 5 0 0 2 4 Minutes 6 6 8 10 7 3 µS 4 1.25 mL/min 3332 psi 8 6 2 9 1 5 Peaks: 0 0 2 4 Minutes 6 6 8 10 7 3 µS 1.5 mL/min 3891 psi 8 6 4 2 9 5 1 0 0 12 2 (4 × 150 mm) Eluent: 23 mM KOH Eluent Source: Dionex EGC III KOH Cartridge Flow Rate: See chromatograms Inj. Volume: 10 µL Temperature: 30 °C Detection: Suppressed conductivity, Dionex ASRS 300, AutoSuppression, recycle mode 4 Minutes 6 8 10 1. Fluoride 2. Chlorite 3. Chloride 4. Nitrite 5. Carbonate 6. Bromide 7. Sulfate 8. Nitrate 9. Chlorate 0.5 5 3 5 20 10 10 10 10 mg/L Fast Analysis of Drinking Water Using High-Pressure IC Column: Instrument: Eluent Source: Eluent: Flow Rate: Inj. Volume: Column Temp.: Detection: 1.8 2 Sample: Sample Prep.: 5 µS Peaks: 4 1 3 6 7 0.8 0 1 3 2 Minutes 13 4 5 Dionex IonPac AS18-4µm, 2 150 mm Dionex ICS-5000+ HPIC system Dionex EGC 500 KOH 23 mM Potassium hydroxide 0.50 mL/min 5 µL 30 °C Dionex ASRS 300, 2 mm, 15 mA, recycle Municipal City A 5-fold dilution with deionized water 1. Fluoride 2. Chloride 3. Nitrite 4. Carbonate 5. Sulfate 6. Nitrate 7. Chlorate 0.4 2.3 < 0.1 --3.5 < 0.1 < 0.1 mg/L High Resolution Cation Analysis on IonPac CS16 at Different Flow Rates Column: 7 Eluant: Flow rate: 30 µL/min 3600 psi 20 µL/min 2400 psi 4 C 10 µL/min 1200 psi 5 µS Inj. volume: Temperature: Detection: 6 IonPac CS16, 2 x 250 mm x 0.5 mm ID 30 mmol/L MSA (EG) A: 10 µL/min B: 20 µL/min C: 30 µL/min 0.4 µL 40 °C Suppressed conductivity CCES 300, AutoSuppression, Recycle mode B Peaks: A 1 2 3 -1 0 14 20 Minutes 40 1. Lithium 2. Sodium 3. Ammonium 4. Potassium 5. Magnesium 6. Calcium 0.5 mg/L 2.0 2.5 5.0 2.5 5.0 Capabilities of HPIC in Capillary Format Increased Capabilities: • Faster separations with higher flow rates (left) • Higher resolution with longer columns (right) Thermo Scientific™ Dionex™ IonSwift™ MAX-100: 11 minutes 4.5 10000:1 Na : Ammonia 5 15,16 24 µL/min – 3900 psi 17 3 µS 67 1 2 45 19 11 12 8 1 13 14 9 10 Single Dionex IonPac CS16 B -0.5 0 15 5 Minutes 10 15 -2 2 Two Dionex IonPac CS16 in series µS 18 0 1 Minutes 2 16 Using HPIC to Identify Spoilage in Beverages 7 6 Column: Dionex IonPac AS11-HC-4µm Capillary (0.4 250 mm) Eluent Source: Dionex EGC-KOH (Capillary) Gradient: Potassium hydroxide, 1 mM from 0 to 8 min, 1-30 mM from 8-28 min, 30-60 mM from 28-38 min, 60 mM from 38-42 min Flow Rate: 15 µL/min Inj. Volume: 0.4 µL Column Temp.: 30 °C Detection: Suppressed conductivity Dionex ACES 300, recycle Mode Sample Prep.: 1:40 dilution with deionized water 15 µL/min, 3600 psi 15 9 3 µS 4 10 6 1 2 16 13 8 11 12 Peaks: 14 5 17 19 18 0 0 10 20 Minutes 16 30 42 1. Quinate 2. Fluoride 3. Lactate 4. Acetate 5. Formate 6. Unknown 7. Chloride 8. Unknown 9. Malate-Succinate 10. Carbonate 11. Maleate 12. Sulfate 13. Oxalate 14. Unknown* 15. Phosphate 16. Citrate 17. cis-Aconitate 18. trans-Aconitate 19. Unknown The Dionex ICS-5000+ HPIC High Pressure Ion Chromatography • High pressure capable with both capillary and standard flow rates • Continuous operation up to 5000 psi when configured as a Reagent-Free (RFIC™) system • Increased productivity with fast run times • Improved separations and higher resolution with 4 µm particle columns HPIC - High Resolution, Fast Analyses 17 Dionex ICS-4000 Capillary HPIC System Dedicated Capillary HPIC • • • • New level of resolution and speed Delivering best in class sensitivity Simplifies workflows Increases analytical efficiency and productivity • Small footprint • Electrochemical, Conductivity, or Charge detection Thermo Scientific™ Dionex™ IC Cube™ Cartridge HPIC - High Resolution, Fast Analyses 18 High-Pressure Ion Chromatography • HPIC systems provide better performance • HPIC systems allow for continuous operation up to 5000 psi • HPIC systems - High-pressure ion chromatography in an all PEEK™ plastic IC • High-pressure Reagent-Free ready • Smaller 4 µm particle-size ion-exchange columns in a variety formats 19 Advances in Trace Element Analysis Fergus Keenan Field Marketing Manager The world leader in serving science 20 Agenda • Advances in ICP-OES technology • High speed analysis • Advances in ICP-MS • Intelligent Auto-dilution • QCell technology • Trace element speciation by IC-ICP_MS 21 iCAP 7600 ICP-OES • Powerful analytical detection & resolution • Choice of plasma orientation to enable enhanced application suitability • Intelligent software for powerful autooptimization of the sample intro system • Advanced data acquisition including ‘Sprint’ modes for ultimate productivity & versatility • Comprehensive accessory compatibility for liquid & solid sampling Who’s it for • Labs requiring the extreme productivity • Labs who perform highly variable & demanding research-based applications • Labs who require solid sampling capability 22 Open Access Sample Introduction Compartment • Large fully opening outer door • Improved user access • Clear view of plasma source • Simplifies optimization • Easy access to sample introduction • Simple change of components • Peri-pump • 12 roller for smooth flow, micro tension control • Better stability allows shorter dwell times • Sprint Valve System • Highest Sample Throughput of any ICP • Drain Sensor • Monitors drain, detects leaks or blockages • Accessories • Easy connection of Argon Humidifier, Hydride Generation and Laser Ablation accessories 23 “Better user access, compatible with all accessories” Sprint valve system – How does it work? 24 Sprint valve system – How does it work? 25 Why segmented stream? Uptake / Washout Profile with Contiguous Flow Raised baseline Long transients Uptake / Washout Profile with Segmented Stream Discrete washout steps True baseline Sharp transients 26 Case-study – Wear Oil Analysis Typical Oil Method Sprint Valve Oil Method (already speed-optimized) Analysis Step 1.Autosampler Movement 2. Sample Uptake 5 sec. 15 sec. 3. Stabilization 20 sec. 4. Measurement 10 sec. 5. Rinse 30 sec. Total Time 27 Time Required 80 sec. Analysis Step Time Required 1. Autosampler Movement, Sample Uptake, Stabilization, and Rinse 17 sec. 2. Measurement 10 sec. Total Time 27 sec. Intelligently Monitored Wash 10000000 • Software automatically detects washout to baseline for selected analytes 1000000 100000 10000 • Non-productive time reduced; analysis time optimized 1000 100 10 • Washout completed sooner • Maybe no wash is needed? 28 1 0 20 40 60 80 100 120 140 160 180 200 CASE STUDY: Ultra-Fast Agricultural Soil Analysis • The soil samples were dried and ground 5 g of sample • 20 ml of the 1M ammonium acetate solution was added. • Samples shaken vigorously for at least 5 minutes and left to react overnight. • Samples were then shaken again and filtered before being made up to 250 ml with de-ionized water. • Sample extracts were analysed directly using the Sprint acquisition mode which further enhances the speed of the instrument. • A locally sourced soil sample was extracted 5 times & each extract was analysed 10 times • The total time required for these 50 repeats was 11 minutes and 35 seconds or 13.9 seconds per sample. 29 Ultra-Fast Agricultural Soil Analysis 30 Ultra-Fast Agricultural Soil Analysis 31 Ultra-Fast Agricultural Soil Analysis 32 Advances in Interference Removal in ICP-MS The world leader in serving science 33 iCAP Q - Dramatically Different ICP-MS 34 iCAP Q - Dramatically Different ICP-MS • Easy to use and learn • Reliability • New interface cone design giving less memory effects and less drift • Lower service costs and new longer life detector supplied as standard • Productivity • Single mode analysis capability for high throughput and quick flush times with the QCell • Cost of ownership • Lower gas consumption per analysis reduces running cost • Longer life components (cones, detector) reduces lifetime cost • Service contracts reduced by 30% over XSERIES2 • Performance • Best Signal /Noise of any Quadrupole ICP-MS on the Market • Best interference removal with unique QCell technology • New Leading Edge Design • Smallest bench space requirements by unique ion optics design • QCell Flatapole technology for the best in interference removal • The only quadrupole MS to offer singe mode analysis 35 Spectral Interferences Caused by molecular species formed in plasma overlapping with analyte isotope ArAr, ArO, ArN, ArC, ArH, ArCa, ArNa, ArK, ArMg, ArCl, ClO, NO, CO, CaO, NaO, etc Products 36 Ar, Air (O, N, C) H2O, Ca, Na, K, Mg, Cl, etc Reaction Reactants Collision/Reaction Cell Technology Collision/Reaction Cell Technology • • • • A multipole enclosed in a cylinder Controlled flow of gas into the cell Interaction of ions with the gas If reactive gas used, reactions occur • All cells are reaction cells M+ only out M+ and XnYn’+ 37 The Basis of KED Operation 38 51V+ 51[ClO]+ ~140 pm ~250 pm Collisional Energy Loss and Filtering: KED Energy Barrier Cell Post-Cell Pre-Cell Small collision crosssection M+ Larger collision crosssection MO+ X Increasing exit energy Bolder shades indicate higher energy for M+ and MO+ ions Key: He atom M+ ion MO+ ion 39 Improving Collision Cell Design •QCell with low mass cut-off •Flatapole technology for improved transmission •Non-consumable, zero-maintenance •50% smaller volume for faster mode switching, <10s •Single mode interference removal with He •Can also use reactive mode with O2, H2 or NH3 mixes 40 QCell – Low Mass Cut-Off KED mode QCell Mass Cut-Off Region (here all masses below 39) 2 Measuring 56Fe 3 41 1 QCell: Effect of Low Mass Cut-Off on in-cell Interference Formation 42 QCell Comparative Performance– He KED mode, No spike • 5%HNO3, 5%HCl, 1%IPA, 1%H2SO4 43 QCell Comparative Performance– He KED mode, 10ppb Spike • 5%HNO3, 5%HCl, 1%IPA, 1%H2SO4 + 10ppb Spike of Li, Be, B, Na, Mg, Al, K, Ca, Sc, Ti, V, Cr, Mn, Fe, Ni, Co, Ni, Cu, Zn, Ga, Ge, As, Se Note Co sensitivity 41,000cps/ppb 44 QCell Comparative Performance– He KED mode, No spike • 5%HNO3, 5%HCl, 1%IPA, 1%H2SO4, 200ppm Na, 200ppm Ca, 500ppm P 45 QCell Comparative Performance– He KED mode, 10ppb Spike • 5%HNO3, 5%HCl, 1%IPA, 1%H2SO4, 200ppm Na, 200ppm Ca, 500ppm P + 10ppb Spike of Li, Be, B, Na, Mg, Al, K, Ca, Sc, Ti, V, Cr, Mn, Fe, Ni, Co, Ni, Cu, Zn, Ga, Ge, As, Se 46 Analysis of Selenium 78Se Sensitivity 8441 cps/ppb IDL 5ppt 7%H2/He KED 47 Analysis of Vanadium without reactive gases Sensitivity 2,100 cps/ppb BEC 24ppt 0.5% HCl, He KED mode 48 Collisional Focusing for High Sensitivity Uranium Measurement Sensitivity 1223 cps/ppt IDL 16ppq Collisional focusing with 7.8mL/min He 49 IC-ICP-MS for Elemental Speciation The world leader in serving science 50 How can we Perform Speciation Analysis? Separation Thermo Scientific Dionex ICS-5000 IC 51 Detection Thermo Scientific iCAP Q ICP-MS Why use ICP-MS for Speciation Analysis? • It can detect most of the periodic table with sub ppt detection limits • It has >9 orders of magnitude linear dynamic range • The (atmospheric, ground potential) ICP ion source is easily connected to a wide range of coupled accessories: • Ion Chromatography (IC); Gas Chromatography (GC); High performance liquid chromatography (HPLC)… • ICP-MS is the ideal elemental detector for speciation analysis! 52 What are the Advantages of Ion Chromatography? • Metal-free systems • Powerful separation chemistries • Reagent-Free Ion Chromatography (RFIC) • Extensive IC product line for full flexibility 53 A Complete, Integrated IC-ICP-MS System • Very simple hardware connection: • Simple interchange between standard ICP-MS analysis and IC-ICP-MS • No need to turn off plasma Inert tubing 0.125 mm i.d. • A single software interface for both the IC and ICP-MS: • Thermo Scientific Chromeleon interface built into workflow • Fully integrated analysis • No trigger cable required • One sample list 54 Data System Speciation of As in Apple Juice • Differentiation between (toxic) inorganic As(III) & As(V) species and (non-toxic) organic species (MMA etc) • Requirements: • Single run anionic and cationic technique since both positive and negative charged species can be present in a sample • Good chromatographic resolution to separate out species • Sharp peaks for improved sensitivities 55 iCAP Qc with Dionex ICS-5000 • 0.45 ppb of each As standard • 6 species • ~8000 cps / ppb • ~15 minute analysis • Anion Exchange: • Dionex AS7 (2x250mm) • Gradient elution with 20-200 mM ammonium carbonate • Flow rate: 0.3 mL/min • Injection volume: 20 µL 56 As Species Detection Limits by IC-ICP-MS Compound 57 Detection limit pg g-1 AsB 2.3 DMA 3.8 As3+ 4.6 AsC 4.4 MMA 11.4 As5+ 1.2 As Species in Apple Juice 58 As Speciation in Apple Juice • iCAP Qc with Dionex ICS-5000: • Anion exchange chromatography AsB DMA As(III) AsC MMA As(V) Sum of Species Total As 0.002 0.004 0.005 0.004 0.011 0.001 - 0.005 Juice 3 ND ND 0.5 ± 0.01 ND ND 0.7 ± 0.01 1.2 1.7 ± 0.05 Juice 4 ND 0.4 ± 0.05 0.3 ± 0.01 ND 0.1 ± 0.05 0.7 ± 0.01 1.5 1.8 ± 0.05 MDL • iCAP Qc benefits: • Low method detection limits: 0.001 and 0.01 ng/g per species, 0.005 ng/g total As vs current EPA MCL (maximum contaminant level) is 10 ng/g in drinking water 59 IC-ICP-MS analysis of As in Organic Brown Rice Syrup • Media reports and scientific publications on the determination of arsenic (As) in foodstuffs have sparked renewed interest from consumer groups and politicians leading to responses from national regulatory bodies. • Following the publication of a report on high As levels in organic brown rise syrup the United State Food and Drug Administration (FDA) stated that it was carrying out a study on As in rice and rice products that is due to report later in 2012. 60 Analysis of As in rice syrup • Three different OBRS samples were sourced and prepared for analysis. • A closed microwave digestion method was used. • Preparation of the OBRS samples for As speciation analysis was achieved by taking 1.5 g of OBRS, adding 15 mL of 0.28 M HNO3 and refluxing for 90 minutes. 61 Speciation of As in OBRS • IC-ICP-MS speciation analysis showed that the predominant As species in the OBRS samples tested was the toxic inorganic As(III) with over 80% of the total arsenic concentration • (equivalent to 86 – 109 ng /g As (III)). 62 iCAP Q - Dramatically Different ICP-MS 63 Summary • iCAP 7600 is the fastest ICP-OES system available • With the iCAP Q ICP-MS is completely automated from standard prep to sample dilution and automated interference free analysis • The patented QCell combines low mass filtering with Collision Reaction Cell Technology for best-in-class interference removal • Ion Chromatography is for elemental speciation studies due to it inert metal free pathway and comprehensive method set for metal ion and organo-metallic separations 64