Particle Counter KS-41 for photoresist measurement 1. Problems of counting particles in photoresist 2. KS-41 Specifications 3. Adjustment and calibration method 4. Actual measurement example Principle of the Light scattering particle counter Interaction between Light and Particle Reflection Diffraction Incident Light (wave length=λ, Intensity=I0 ) Absorption Refraction φ Refractive index of media=m0 Particle (diameter=D, refractive index=m1) 1-1 Scattering (Intensity=Iφ ) Reyleigh scattering pattern Direction of electrolytic vibration Irradiating light Direction of magnetic vibration 1-2 Light scattering intensity pattern The structure of sensor for Airborne particle counter Inlet Scattering light reflection mirror Scattering light collection lenses Photodiode Beam forming lenses Outlet Light source 1-4 Pulse signal from PSL particle Pulse height Pulse height Pulse number (frequency) Time 1-5 Cumulative 累積分布 distribution Particle Count (standardization) 粒 子 数 (規 格 化 ) 1 0.8 Frequency 頻度分布 distribution 0.6 Calibration 校正 電圧 voltage 0.4 0.2 13 14 15 16 パルス 電圧 Pulse voltage 1-6 17 18 Problems of measuring particles in photoresist ・Scattering of irradiating light in resist becomes noise. ・Refractive index of the resist (approx. 1.4 to 1.6) affects the precision of particle diameter determination. ・Sample flow rate changes according to its viscosity. polymer molecule, polymer, etc 1.6 1.5 1.33 Sample Flow cell Irradiating light Photo detector Particle Particle Particle detector Scattering by polymer in photoresist (Top view) Photo Diode Flow cell Sample Laser beam Lens Aperture Collection light spot diameter depending on refractive index (Top View) Photo Diode Laser beam Flow cell Lens Sample Aperture Laser Spot diameter in DI-water Spot diameter in resist Noise increase due to photoresist measurement Scattered light pulses of particles 0.15μm 0.1μm 0.1μm 0.15μm Noise Noise Voltage Voltage Scattered light pulses of particles Time Time Scattered light of photo-resist and wall of flow-cell (DC voltage) Signal output of photo-detector Signal output of photo-detector KS-41 Specifications Outline • KS-41 is a sensor for measuring particles in photoresist. • By connecting the unit to the controller (KE-40 or KE-40A), power supply, measurement control and data output are carried out. • As the unit does not incorporate flow control circuit, the additional flow rate controller should be prepared. Specifications Optional system: Light source: 90°sideway light scattering method Laser diode (wave length 830nm, rated output 200 mW) Light detector: PIN type photodiodes Materials of component parts exposed to sample fluid Fused silica, PFA KS-41 Specifications Measurement size ranges: 0.15µm, 0.2µm, 0.3µm, 0. 5µm Counting efficiency: 50±10% (measuring PSL particles in the range of 0.3µm, using count of 0.2µm and above for comparison with reference unit) Sample fluid rate: 10mL/min Maximum particle concentration: 12,000 particles/mL (coincidence loss 5% for 0.15 µm particles) Ambient conditions for operation: 15 to 30ºC, 0 to 85% RH (no condensation) Allowable sample fluid pressure 300 kPa or less (gauge pressure) Particle detector system for photoresist measurement Scattered light ( from fringe of the flow cell) Photo Diode (Side View) Aperture ×1 Lenses Flow Cell Scattered light (from center of flow cell) KL-26 Optical System Laser Beam (Vertical to paper) Scattered light ( from fringe of the flow cell) 10ch Photo Diode L Shape Flow Cell ×8 Lenses KS-41 Optical System Scattered light (from center of flow cell) L Shape Flow Cell 10ch Photo Diode Photoresist measurement using sensor calibrated with PSL particles and DI-water Scattered light ( from fringe of the flow cell) L Shape Flow Cell 10ch Photo Diode ×8 Lenses Laser Beam (Vertical to paper) Proto-KS-41 Optical System Scattered light (from center of flow cell). Scattered light ( from fringe of the flow cell) 10ch Photo Diode L Shape Flow Cell ×8 Lenses KS-41 Optical System Scattered light (from center of flow cell) Standard PSL particles in DI-water Adjustment: Calibration: Standard PSL particles in photoresist Using a solution of glycerin (refractive index of 1.4) and PSL particles. With PSL particles (refractive index 1.6) in pure water KS-41 Sensor construction drawing Collection lenses Photodiode Laser diode Scattering light collection lens Beam forming lenses Noise increase when measuring photoresist Noise 64.5mv KL-24 DI-water Noise 113.5mv KS-41 DI-water Noise 538.6mv Noise increase rate 8.4 times KL-24 Photoresist Noise increase Noise rate 197mv 1.7 times KS-41 Photoresist Example of particle counts in photoresist ( I ) Noise(mV) 0.15< I-line positive 582 1052 Excimer A 243 740 235 620 Excimer B 339 236 Excimer C 415 256 Excimer D 494 584 Excimer E 0.2< 97 368 295 91 116 289 0.22< 74 278 195 65 79 213 0.25< 41 173 83 33 45 118 0.3< 28 107 54 21 35 77 (Number /10mL) Example of particle counts in photoresist ( II ) KL-20A C. E. 0.15≦ 0.20≦ 0.3≦ 0.5≦ 1.0≦ KS-41 875 512 150 72 22 1128 483 112 12 45% 31% 64% Resist A 0.15≦ 0.20≦ 0.3≦ 0.5≦ 1.0≦ KS-41 1209 67 10 4 1 KL-20A C.E. 116 34 8 1 58% 29% 50% Solvent Resist B 0.15≦ 0.20≦ 0.3≦ 0.5≦ 1.0≦ KS-41 574 76 22 9 4 KL-20A C.E. 142 56 16 3 54% 39% 56% Resist C 0.15≦ 0.20≦ 0.3≦ 0.5≦ 1.0≦ KS-41 845 322 102 42 12 KL-20A C.E. 814 243 50 6 40% 42% 84% (Number /10mL) Particle measurement in DI water KS-41 KL-20A 80 43 22 15 97 55 20 Unit: particle/10mL Counting efficiency 44.3% 40.0% 75.0% PSL particle measurement in Ethyllactate (particle size: 0.321 µm) First time KS-41 5657 4337 619 278 KL-20A 9575 2140 519 Second time Counting efficiency 45.3% 28.9% 53.6% KS-41 5832 4392 629 296 KL-20A 9682 2140 546 Counting efficiency 45.4% 29.4% 54.2% Reference: KS-41 (Counting efficiency 58%, measuring in the range of 0.2 µm, using count of 0.309 µm for comparison with reference unit) Viscous resistance of Particle counters Flow Rate(ml/min) 1000 KS-28(1×1mm) KL-22(0.5mmφ) KL-26(0.3×0.7mm) KL-20A(0.5×0.5mm) 100 10 1 10 100 Viscosity(cp) 1000 •When sample input pressure is 3 kg/cm2, flow rate limitations which depend on viscosity of sample are measured. •Absolute values depend on measurement error of sample viscosity and surface exposed to sample fluid. •The difference of the flow path shape varies the maximum flow rate widely. Measurement of babble and solid particles 1.0E+06 KL-26 Relative Intensity 1.0E+04 1.0E+02 Bubble PSL 1.0E+00 Media:H2O 1.0E-02 0.1 1 10 Diameter(μm) Light scattering intensity is almost the same with PSL particles and bubbles in water. Particle measurement example in APM cleaning tank 80 75 100 70 65 10 60 0.2μm 0.5μm temp Temperature (℃) Particle Numbers(N/10mL) 1000 55 1 50 0 10 20 30 TIME 40 50 •Particle counts which vary depending on the APM temperature are regarded as bubble counts. •Particle counts are not affected by the temperature 60℃ and below.