Simulator Theory and Operation Rev 3.09 13-1 Use and Theory EACH BREATH TEST MUST BE PRECEEDED BY A REFERENCE ANALYSIS REQUIRED BY THE CDPHE RULES AND REGULATIONS, 5 CRR 1005-2 THE DEVICE USED TO GENERATE THE REFERENCE VAPOR IS CALLED BREATH ALCOHOL SIMULATOR OR “SIMULATOR” H2O AND ETOH IS USED IN LUE OF BLOOD AND ETOH Rev 3.09 13-2 Use and Theory THE AMOUNT OF ETOH IN 2100 ml OF DEEP LUNG AIR AT 34.0C ETOH IN 1 ml OF ATERIAL BLOOD SIMULATOR SOLUTION 0.100 grams OF ETOH PER 2100 Liters OF AIR AT 34.0C PREPARED BY CERTIFICATION UNIT VALIDATES THE PERFORMANCE OF THE 5000EN Rev 3.09 13-3 Functional Structure AIR INLET TUBE HIGH INTENSITY L.E.D. DISPLAY ON/OFF-RESET SWITCH HEATER INDICATOR GLASS CONTAINER VAPOR PORT AGITATOR TEMPERATURE SENSOR BAFFLE PLATES HEATING ELEMENT DISPERSION TUBE Rev 3.09 13-4 Exclusive Features of the Guth Model 2100 Simulator MICROPROCESSOR ENHANCED: -PERFORMS INTERNAL DIAGNOSTICS -MONITERS TEMPERATURE SENSOR, RFI CIRCUITRY -CONTROLS AGITATOR MOTOR, DISPLAY DRIVER, COMMUNICATIONS AND A/V INFORMATION Rev 3.09 13-5 Exclusive Features of the Guth Model 2100 Simulator TEMPERATURE SENSOR: -RUGGED DESIGN: STAINLESS STEEL, BRASS -STABLE TEMPERATURE RANGE: MAINTAINS A PRECISE RANGE OF 34C .02C -ACCURACY OF INSTRUMENT TEMPERATURE READING IS .05C Rev 3.09 13-6 Exclusive Features of the Guth Model 2100 Simulator ELECTRONICALLY PROTECTED: -INTERNAL POWER TRANSFORMER AND OPTICALLY ISOLATED CONTROL COMPONENTS PREVENTS A.C. LEAKAGE TO ELECTRONICS -R.F.I. PROTECTED -HEATING ELEMENT PROTECTED FROM OVER HEATING WHEN TEMPERATURE SENS0R IS DEFECTIVE OR REMOVED FROM SOLUTION -EXTERNAL FUSE PROVIDES PROTECTION IN CASE OF ELECTRICAL MALFUNCTION Rev 3.09 13-7 Guth Simulator Error Codes MESSAGE PLATE MOUNTED ON SIMULATOR QUICK REFERENCE GUIDE Rev 3.09 13-8 Guth Simulator Error Codes CAUSE: CAUSE: CAUSE: THE ELECTRONICS THE TEMPERATURE NEED TO BE RESET. SENSOR IS OPEN OR SHORTED. COMPUTER IS RECEIVING NO SIGNAL TO ALLOW IT TO CONTROL THE TEMPERATURE. REMEDY: TURN OFF, WAIT 3-5 REMEDY: REMEDY: SECONDS, THEN TURN OFF THE THIS CAN’T BE TURN ON. IF THE SIMULATOR, FILL REMEDIED IN THE SIMULATOR WITH SOLUTION AND FIELD. REQUIRES REPEATS THIS REASSEMBLE. SERVICE * ERROR CODE. CALL FOR ASSISTANCE * * NOTIFY CDPHE NO SOLUTION IN CONTAINER OR SIMULATOR WAS POWERED UP WITHOUT TOP HOUSING ATTACHED TO CONTAINER. Rev 3.09 13-9 Guth Simulator Error Codes CAUSE: CAUSE: ELECTRONICS NEED TEMPERATURE OF TO BE RESET. THE SOLUTION IS ABOVE 34.2C OR RFI HAS BEEN DETECTED. REMEDY: REMOVE THE SOURCE OF THE INTERFERENCE OR CHANGE THE LOCATION OF THE SIMULATOR. RESET THE SIMULATOR. * NOTIFY CDPHE Rev 3.09 REMEDY: RESET THE SIMULATOR. CAUSE: EXCESSIVE AMOUNT OF TIME TO OBTAIN OPERATING TEMPERATURE (+15 MINUTES). THE HEATING ELEMENT MAYBE OPEN OR THE SOLUTION WAS TOO COLD. REMEDY: RESET THE SIMULATOR. IF THE SIMULATOR DOES NOT HEAT, THE HEATING ELEMENT MAY BE DEFECTIVE. REQUIRES SERVICE * 13-10 Guth Simulator Error Codes CAUSE: CAUSE: CAUSE: THE TEMPERATURE OF THE SOLUTION IS BELOW 33.8C AFTER IT HAS INITIALLY OBTAINED 34.0C. AN ERROR HAS OCCURRED WITH THE RS-232 COMMUNICATIONS EXTERNAL CONTROL OPTION. AN ERROR HAS OCCURRED WITH THE READ OR WRITE MEMORY OPTION COMMAND. REMEDY: RESET THE SIMULATOR. IF THE SIMULATOR DOES NOT HEAT, THE HEATING ELEMENT MAY BE DEFECTIVE. REQUIRES SERVICE* * NOTIFY CDPHE Rev 3.09 REMEDY: REMEDY: RESET THE RESET THE SIMULATOR. IF THE SIMULATOR. IF THE SIMULATOR REPEATS SIMULATOR REPEATS THE ERROR CODE, THIS ERROR CODE. CALL FOR CALL FOR ASSISTANCE * (THIS ASSISTANCE * ERROR SHOULDN"T OCCUR WITH COLORADO SIMULATORS!) 13-11 Correct Positioning of Simulator Tubing and Clamp #1 Shows the correct orientation of the Simulator Tubing Elbow #2 Shows the correct positioning of the Simulator Clamp •Make sure the clamp is tightened all the way closed •When sending the simulator in for repairs, remove the tubing at the clamp and disconnect the heated tube at the disconnect joint •Keep all tubing and the jar at your location Rev 3.09 13-12 Rev 3.09 13-13