Shunt CALIBRATION Calibration with the Shunt Resistor Attach a shunt resistor in parallel with one arm of the gauge to produce an output signal simulating strain. 1 K Rc Rg K There are two methods to attach the shunt cal resistor. Rc : Shunt cal resistor Rg : Gauge resistor K : Gauge factor : Reference strain (uSTR) Auto- balance : Cancel the initial offset value Shunt-CAL : Adjust the Gain modulus of the gauge Positive strain direction Negative strain direction Shunt CALIBRATION Shunt Calibration will be performed by linear scaling function. P1:Y set depends on the initial offset value but usually 0. P2:Y Reference strain (uSTR) After the above setting done, press the “Shunt Cal” key… P1: X DL sets the value using “get measure” function when releasing the shunt cal resistor 1 K Rc Rg K P2: X DL sets the value using “get measure” function when attaching the shunt cal resistor ex. Shunt cal resistor(Rc) and Reference strain( P2:Y) Rc (k Ω) Rg ( Rc Rg ) K 10 15 33 100 330 Reference strain (ε: uSTR) Rg = 120Ω Rg = 350Ω 5928.9 16908.2 3968.3 11400.7 1811.6 5247.4 599.3 1743.9 181.8 529.7 When K=2.00 Shunt CAL procedure Calibration procedure 1 1. Connect the strain module, 701271, the bridge head, 701957/701958 and the strain gauge. Attach the shunt resistor on the bridge head. 2. Set the range unit, the range, the excitation voltage and gauge factor 3. Implement the Auto-Balance function. Shunt CAL procedure Calibration procedure 2 4. Normally enter 0 for P1:Y. 5. Enter the reference strain value corresponding to the attached resistor for P2:Y. 6. Enter the “Units” 7. Press “Shunt Cal” to perform the calibration. Shunt CAL procedure After the calibration, P1:X and P2:X values have changed. Shunt CAL procedure After the calibration is done, the Upper Scale and Lower Scale are slightly changed. Both settings are adjusted for convenient display. Shunt CAL procedure When the negative strain direction is selected, the P2:Y value must be negative, and the calibration result, P2:X will also be negative.