MLC “Did not clear before slowpass” T-nuts T-nuts T-nuts From the Hyper Terminalen we can test the slop on each T-nuts with this command: diagLeafBacklashAll 1,5,2; diagLeafBacklashAll 2,5,2; 1: Select between bank A or B 5: Indicates how many times each leaf is tested 2: Tells how the results should be presented T-nuts Average of the 5 measurements = 103,8 counts 103,8 / 512 counts/mm = 0.203 mm T-nuts • A position or intensity af deviation indicates that the tolerance in the backlash test is overdue for both leaf. T nut skifte Positionsafvigele (mm) 0.4 0.3 0.2 0.1 Før Efter 0 -0.1 -0.2 -0.3 0 10 20 30 40 MLC-bladpar nummer 50 60 T-nuts T nut skifte Positionsafvigele (mm) 0.4 0.3 0.2 0.1 Før Efter 0 -0.1 -0.2 -0.3 0 10 20 30 40 MLC-bladpar nummer 50 60 T-nuts To large distance between leafs Here are both leafs out of position T-nuts Here are the distance between the leafs to small T-nuts T-nut slør