PPT

RadiAnalyzer

TM

Xpress

Set-up

2

Module Content

 Setup

 Overview

 1. Mounting RadiAnalyzer Xpress in the Cath Lab

 2. Connect Cables – Standard Set-up

 3. Connect Cables – Namic Adapter Cable

 4. Connect PressureWire

 5. Zero Cath Lab Pressure Channels – Manual

 6. Zero AO – Manual

 7. Calibrate PressureWire Manual

 Automatic Set-up Sequence

Set-up - Overview

 RadiAnalyzer Xpress interconnects between the patient and the cath lab’s existing hemodynamic recording system.

 For FFR calculation aortic pressure (AO) is required, in addition to PressureWire pressure.

 AO pressure is taken directly from the cath lab’s standard AO transducer and then passed on to the pressure ports on the hemodynamic recording system.

 Optionally, PressureWire pressure can also be passed on to the recording system but this is not mandatory.

AO - transducer

PressureWire

Hemodynamic Recording System

1. Mounting RadiAnalyzer Xpress in the Cath Lab

 It is recommended to mount the RadiAnalyzer Xpress unit permanently on the cath lab table and always route AO through the instrument, even when FFR is not used.

 The pressure channels on the hemodynamic recording system are zeroed once in the morning, while the AO pressure thereafter is zeroed only on RadiAnalyzer Xpress.

 In this way FFR measurement is always available and set-up time is completely eliminated.

 Use the multipurpose bracket to mount the RadiAnalyzer Xpress on the cath lab table rail

(separate order).

 For ambulatory use, the RadiAnalyzer Xpress may also be mounted on a mobile IV-pole or equivalent.

2. Connect Cables – Standard Set-up

1. Connect AO transducer to AO adapter cable

2. Connect AO adapter cable to AO IN port ( red )

3. Connect AO OUT port ( red ) to cath lab recording system

(P1)

4. Connect PressureWire OUT ( green ) to cath lab recording system (P2)

5

AO transducer

AO adapter cable cath lab pressure input P1 cath lab pressure input P2

6

3. Connect Cables – Namic

®

Adapter Cable

1. Disconnect the Namic AO transducer from its monitor interface cable

2. Connect the Namic AO transducer to the Namic adapter cable

3. Connect the monitor cable to the Namic Adapter cable

The Namic adapter cable should stay connected to RadiAnalyzer Xpress AO IN port ( red ) and AO OUT port ( red )

7

4. Connect PressureWire™ Certus™

8

5. Zero Cath Lab Channels – Manual

1. Select ZERO cath lab and press ENTER

2.

“Zero” both pressure channels on the cath lab recording system, press ENTER when done

9

6. Zero AO Pressure – Manual

1. Select CAL AO and press

ENTER

2. Open AO transducer to air and press ENTER

10

7. Calibrate PressureWire – Manual

1. Select CAL WIRE and press

ENTER

2. Flush the PressureWire packaging coil, place flat and press ENTER

11

Automatic Set-up Sequence

 The automatic setup sequence is an on-screen step-by-step guide that runs through the three main set-up steps.

 The sequence is enabled/disabled in the OPTIONS submenu.

 If enabled, the sequence starts automatically when the instrument is turned on.

 The sequence can be aborted by pressing ESC .

 It is possible to advance/go back using LEFT and RIGHT buttons.

 The sequence can always be started from the MAIN menu by keeping F1 pressed for more than 2 seconds.

12

Automatic Set-up Sequence

1.

“Zero” both cath lab pressure channels, press ENTER to continue

2. Open AO transducer to air and press ENTER

3. Flush the PressureWire packaging coil, place flat and press ENTER

Rx Only

Prior to using these devices, please review the Instructions for Use for a complete listing of indications, contraindications, warnings, precautions, potential adverse events, and directions for use.

Product referenced is approved for CE Mark

RadiAnalyzer Xpress is designed, developed and manufactured by St Jude Medical

Systems AB. Patent Pending. Unless otherwise noted, ™ indicates a registered or unregistered trademark or service mark owned by, or licensed to, St. Jude Medical, Inc. or one of its subsidiaries. RadiAnalyzer Xpress, RADI, ST. JUDE MEDICAL, the ninesquares symbol, and MORE CONTROL. LESS RISK. are registered and unregistered trademarks and service marks of St. Jude Medical, Inc. and its related companies.

©2010 St. Jude Medical, Inc. All rights reserved.

This is the last page of this module.

Use the home symbol to go back to the Content page.

Exit by pressing the cross symbol.