N-Dimensional Presentation State Use Cases Toshiba

advertisement
DICOM
nD Presentation State
Toshiba Cross-Sectional Imaging
Use Cases
Updated 15 Feb, 2011
Modalities Considered



CT
MR
PET
4/13/2015
nD Presentation State Use Cases
2
Clinical Specialties Considered



Radiology
Cardiology
Gastroenterology
4/13/2015
nD Presentation State Use Cases
3
CT Coronary Artery Stenosis
Input is a gated, contrast-enhanced CT cardiac exam









Device/user creating derived views and presentation states
CT technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Heart / myocardium

Coronary arteries
Automated / semi-automated coronary artery centreline finding
Views saved:

Three MPRs aligned with the axes of the heart intersecting at the point of stenosis

Curved MPR following the centreline of the stenotic vessel with stenosis marked

Cross-curved MPR at the point of stenosis with % stenosis measurement

SVR and / or volumetric MIP, oriented to show stenosis
Device/user using presentation state to replicate derived views
Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Centrelines

Markers / measurements

Position and orientation of MPRs

Position of cross-curved MPR

Orientation and VOI for SVR / MIP

W/L for MPR and MIP views

Opacity for SVR views – accuracy of colour mapping unimportant
4/13/2015
nD Presentation State Use Cases
4
CT Coronary Artery Plaque Evaluation
Input is a gated, contrast-enhanced CT cardiac exam









Device/user creating derived views and presentation states
CT technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Heart / myocardium

Coronary arteries
Automated / semi-automated coronary artery centreline finding
Views saved:

Slab MPR aligned with the affected vessel

3-4 MPR views perpendicular to the above view demonstrating the extent of the plaque

Curved MPR following the centreline of the affected vessel

Set of cross-curved MPRs with colour-coded plaque burden
Device/user using presentation state to replicate derived views
Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Centrelines

Markers / measurements

Position and orientation of MPRs

Width and Intensity Projection of MPRs

Position of cross-curved MPRs

W/L for MPR views
4/13/2015
nD Presentation State Use Cases
5
CT Coronary Artery Stent Evaluation
Input is a gated, contrast-enhanced CT cardiac exam









Device/user creating derived views and presentation states
CT technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Heart / myocardium

Coronary arteries
Automated / semi-automated coronary artery centreline finding
Views saved:

MPR aligned with the affected vessel

3-4 MPR views perpendicular to the above along the length of the stent

Curved MPR following the centreline of the affected vessel

Set of cross-curved MPRs along the length of the stent
Device/user using presentation state to replicate derived views
Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Centrelines

Markers / measurements

Position and orientation of MPRs

Position of cross-curved MPRs

W/L for MPR views
4/13/2015
nD Presentation State Use Cases
6
CT Coronary Artery Aneurysm Evaluation
Input is a gated, contrast-enhanced CT cardiac exam









Device/user creating derived views and presentation states
CT technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Heart / myocardium

Coronary arteries
Automated / semi-automated coronary artery centreline finding
Views saved:

Curved MPR following the centreline of the affected vessel

Cross-curved MPR at end-points of aneurysm and point of maximal diameter with measurements
Device/user using presentation state to replicate derived views
Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Centrelines

Markers / measurements

Position and orientation of MPRs

Position of cross-curved MPRs

W/L for MPR views
4/13/2015
nD Presentation State Use Cases
7
CT Coronary Artery Anomalous Vessel Evaluation
Input is a gated, contrast-enhanced CT cardiac exam









Device/user creating derived views and presentation states
CT technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Heart / myocardium

Coronary arteries
Automated / semi-automated coronary artery centreline finding
Views saved:

Three MPRs intersecting at the vessel origin

Slab MPR demonstrating the vessel anomaly
Device/user using presentation state to replicate derived views
Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Centrelines

Markers / measurements

Position and orientation of MPRs

Width and Intensity Projection of MPRs

W/L for MPR views
4/13/2015
nD Presentation State Use Cases
8
CT Coronary Artery CABG Evaluation
Input is a gated, contrast-enhanced CT cardiac exam









Device/user creating derived views and presentation states
CT technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Heart / myocardium

Coronary arteries
Automated / semi-automated coronary artery centreline finding
Views saved:

Curved MPR following the centreline of the bypass graft

SVR views oriented to show junction of bypass graft and coronary artery with different VOIs (survey and focused on junction)
Device/user using presentation state to replicate derived views
Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Centrelines

Markers / measurements

Position and orientation of MPRs

Orientation and VOI for SVR / MIP

W/L for MPR views

Opacity for SVR views – accuracy of colour mapping unimportant
4/13/2015
nD Presentation State Use Cases
9
CT Cardiac 4D Valve Evaluation
Input is a retrospectively gated, contrast-enhanced CT cardiac exam

Device/user creating derived views and presentation states

CT technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Left ventricle endocardium

Left ventricle epicardium
Views saved:

Three MPRs aligned with the axes of the heart intersecting at the aortic valve cycling through cardiac phases

SVR view oriented to show the aortic valve cycling through cardiac phases

Three MPRs aligned with the axes of the heart intersecting at the mitral valve cycling through cardiac phases

SVR view oriented to show the mitral valve cycling through cardiac phases

Device/user using presentation state to replicate derived views





Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Position and orientation of MPRs

Cine details?

Orientation and VOI for SVR

W/L for MPR views

Colour / opacity guideline – accuracy of colour and opacity mapping unimportant
4/13/2015
nD Presentation State Use Cases
10
CT Cardiac 4D Myocardial Contractility
Input is a retrospectively gated, contrast-enhanced CT cardiac exam








Device/user creating derived views and presentation states
CT technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Left ventricle endocardium

Left ventricle epicardium
Views saved:

Three MPRs aligned with the axes of the heart intersecting at the left ventricle cycling through cardiac phases

SVR view oriented to show the left ventricle valve cycling through cardiac phases
Device/user using presentation state to replicate derived views
Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Position and orientation of MPRs

Orientation and VOI for SVR

W/L for MPR views

Colour / opacity guideline – accuracy of colour and opacity mapping unimportant
4/13/2015
nD Presentation State Use Cases
11
CT Calcium Scoring
Input is a retrospectively gated, non-enhanced CT cardiac phase





Device/user creating derived views and presentation states
CT technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Coronary artery calcification
Output saved:

Assignment of calcification to coronary arteries
Device/user using presentation state to replicate derived views

Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Assignment of calcification to coronary arteries

Use case may be more appropriate to a CAD-like object than to an nD PS


4/13/2015
nD Presentation State Use Cases
12
CT Colonography
Input is a combination of prone, supine and decubitus CT volumes, with or without contrast








Device/user creating derived views and presentation states
CT technologist, radiologist or gastroenterologist using post-processing application
Automated / semi-automated segmentation of:

Luminal surface of colon

Colonic centerline

Air in small intestine

Polyps/flat lesions

Contrast tagged stool

Residual fluid
Views saved:

SVR of air in colon

MPR/slab views to illustrate individual lesions

Endoscopic (SVR/SSD) of colon

Dissected SVR view
Device/user using presentation state to replicate derived views
Radiologist / gastroenterologist performing primary read after tech workup
Radiologist / gastroenterologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Centrelines

Position and orientation of MPRs

Orientation and VOI for SVR

W/L for MPR views

Opacity /threshold for SVR/SSD views – accuracy of colour mapping unimportant
4/13/2015
nD Presentation State Use Cases
13
CT Angiography
Input is a contrast-enhanced CT angiography exam









Device/user creating derived views and presentation states
CT technologist or radiologist using post-processing application
Automated / semi-automated segmentation of:

Arteries to be analyzed

Bone adjacent to arteries of interest
Automated / semi-automated artery centreline finding
Views saved:

Three MPRs aligned with the artery intersecting at the point of stenosis

Curved MPR following the centreline of the stenotic vessel with stenosis marked

Cross-curved MPR at the point of stenosis with % stenosis measurement

SVR and / or boneless volumetric MIP, oriented to show stenosis
Device/user using presentation state to replicate derived views
Radiologist performing primary read after tech workup
Radiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Centrelines

Markers / measurements

Position and orientation of MPRs

Position of cross-curved MPR

Orientation and VOI for SVR / MIP

W/L for MPR and MIP views

Opacity for SVR views – accuracy of colour mapping unimportant
4/13/2015
nD Presentation State Use Cases
14
MR Cardiac LV Analysis
Input is a gated, contrast-enhanced MR cardiac exam








Device/user creating derived views and presentation states
MR technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Left ventricle endocardium

Left ventricle epicardium
Views saved:

Set of LV short axis views with overlaid segmentation
Device/user using presentation state to replicate derived views
Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

W/L for MPR views
4/13/2015
nD Presentation State Use Cases
15
MR Cardiac Perfusion
Input is a contrast-enhanced MR cardiac perfusion exam








Device/user creating derived views and presentation states
MR technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated segmentation of:

Left ventricle endocardium

Left ventricle epicardium
Views saved:

Set of LV short axis views with functional overlays – stress and rest acquisitions
Device/user using presentation state to replicate derived views
Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

W/L for MPR views
4/13/2015
nD Presentation State Use Cases
16
MR Cardiac Delayed Enhancement
Input is a delayed-enhancement MR cardiac exam









Device/user creating derived views and presentation states
MR technologist, radiologist or cardiologist using post-processing application
Automated / semi-automated determination of infarct enhancement threshold
Automated / semi-automated segmentation of infarct
Views saved:

Cross-sectional view highlighting infarct with overlays and measurements
Device/user using presentation state to replicate derived views
Radiologist / cardiologist performing primary read after tech workup
Radiologist / cardiologist reviewing priors alongside follow-up exams
Need to reproduce:

Segmentation

Signal intensity threshold used to perform infarct segmentation
4/13/2015
nD Presentation State Use Cases
17
PET / CT Fusion – Added 15 Feb
Input is a set of registered PET /CT exams, typically from a combined PET / CT scanner. A set of prior exams may also be loaded.

Device/user creating derived views and presentation states

Technologist or radiologist using post-processing application
If views are not already registered, registration is performed
Views saved:

Blended CT / PET MIP views demonstrating areas of interest with measurement overlays (point measurements or ROIs)

3D MIP view with an inverted W/L

Device/user using presentation state to replicate derived views





Radiologist performing primary read after tech workup
Radiologist reviewing priors alongside follow-up exams
Need to reproduce:

MPR positions

Blending function

W/L and or color maps

Measurement position and ROI details
4/13/2015
nD Presentation State Use Cases
18