EP SALES TRAINING Diagnostic Catheters Diagnostic Catheters Diagnostic catheters are used to obtain information about the morphology and timing of the electrical conduction of the heart. High Right Atrium – Surragate for SA Node – Timing relative to other locations Coronary Sinus – Surrogate for Left Atrial and Ventricular conduction – Morphology and timing of conduction from Left Arium to Left Ventricle HIS – Surrogate for AV Node – Morphology and timing of conduction from Right Atrium to Right Ventricle Right Ventricular Apex – Surrogate for Purkinje – Timing relative to other locations BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Agenda • Fixed Curve – – – – General Features of Catheter Type Features of BSC products Clinical Needs and BSC Catheter Solutions Brochure with positioning statement • Steerable Curve – – – – General Features of Catheter Type Features of BSC products Clinical Needs and BSC Catheter Solutions Brochure with positioning statement BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters General Characteristics CHARACTERISTICS FIXED CURVE Distal Tubing • French Size – 5f or 6f --- 5f for pediatrics or minimize vascular trauma • Construction – Braided or Durometer -- Enhances performance or minimizes trauma • Ring and Tip Electrode – Spacing and Number • • • • Wide interelectrode spacing = timing Narrow interelectrode spacing = morphology 2, 4, 6, 8 , 10 Electrodes = more information Curve – Access and Stability • • Access CS or cross the TV Stability at HRA, HIS, and RVA for sensing and/or capture for pacing Proximal Tubing • Construction – Braided or Durometer --- Enhanced performance Connector • Ease of use --- Signal fidelity and easy to connect/disconnect BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Fixed Curve -- Features EXPLORER EXPLORER 360/360jr 5f/6f EXPLORER ST Distal Tubing Continuous Braid Distal Tubing Continuous Braid Distal Tubing Polyurethane 50D Electode Number Quadripolar Electode Number Quadripolar Electode Number various (see catalog) Curve Conduction Curve various (see catalog) Curve various (see catalog) Proximal Tubing Continuous Braid Proximal Tubing Continuous Braid Proximal Tubing Continuous Braid Connector Rotary Connector Rotary Connector Quick Connect Positioning Cost Effective Positioning Enhanced Performance Positioning Less Traumatic BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Feature Summary Fixed Curve Catheters CURVE SHAPES DISTAL TUBING CONNECTORS BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Clinical Requirements – HRA HRA CLINICAL REQUIREMENTS CATHETER FEATURES Access from the femoral vein Pushablity and trackability of proximal tubing. Stability upper posterior of RA for signal fidelity and pacing capture Distal tubing material and braid reinforcement Timing of conduction at HRA 5mm or 10mm bipoalr or quadripolar spacing BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Clinical Requirements – CS CS CLINICAL REQUIREMENTS CATHETER FEATURES Access to the CS ostium via the IVC or SVC Cournand curve Trackability over the "lip" at the entrance to the CS Pushability and torqueability of proximal tubing Trackability to push the catheter to the distal posterior section of the CS Morphology and timing along the length of the CS, including the CS ostium Pushablity and trackability of proximal tubing. 2 - 5 - 2 or similar decapolar spacing BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Clinical Requirements – HIS HIS CLINICAL REQUIREMENTS CATHETER FEATURES Access from femoral vein Pushablity and trackability of proximal tubing. Access to the septal portion of the TV from the IVC Multipurpose curve Stablity at the TV for signal fidelity Distal tubing material and braid reinforcement Minimal trauma to the AV Node and HIS Non braided distal tubing Timing and Morphology of conduction from RA to RV, including HIS signal 2 mm quadripolar or hexapolar spacing BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Clinical Requirements – RVA RVA CLINICAL REQUIREMENTS CATHETER FEATURES Access from the femoral vein Pushablity and trackability of proximal tubing. Cross the TV from the IVC Josephson/Cournand Curve Stability at the Apex of the RV for signal fidelity and pacing capture Distal tubing material and braid reinforcement Minimize perforation of the RV Apex Non braided distal tubing Timing of conduction at RVA 5mm or 10mm bipoalr or quadripolar spacing BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Explorer 360 and 360Jr Brochure Clinical Benefits Technical Info POSITIONING Ordering Info BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Explorer 360 and 360Jr Brochure BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Explorer ST Brochure Clinical Benefits Technical Info POSITIONING Ordering Info BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Explorer ST Brochure BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Fixed Curve Diagnostic Catheters Competition COMPETITOR PRODUCT BSC SOLUTION BARD Viking™ Viking™ Soft Woven™ Woven Flexie™ Orbiter™ Explorer 360™ Explorer ST™ Explorer 360™ Explorer ST™ Explorer 360™ St Jude Medical / IBI Inquiry™ Fixed Curve Evaluator™ Response™ Supreme™ Explorer 360™ Explorer 360™ Explorer ST™ Explorer ST™ Biosense Webster Johnson & Johnson Standard Fixed Curve Standard Fixed Curve Soft Tip Small Anatomy Fixed Curve Explorer 360™ Explorer ST™ Explorer 360 JR™ Medtronic Soloist™ TORQR™ Explorer 360™ Explorer 360™ BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters General Characteristics STEERABLE Distal Tubing • French Size – 6f or 7f --- 6f for pediatrics and minimize vascular trauma • Construction – Tubing Durometer ---Enhanced performance or minimize trauma • Steering Mechanism – Tubing or Metal --- Reproduceability of curve, access, and stability • Ring and Tip Electrode – Spacing and Number • • • • Wide interelectrode spacing = timing Narrow interelectrode spacing = morphology 2, 4, 6, 8 , 10 Electrodes = more information Curve – Access and Stability • • Access CS or cross the TV Stability at HRA, HIS, and RVA = sensing and capture for pacing Proximal Tubing • Construction – Braided or Durometer --- Enhanced performance Handle • Piston or Lever – Ergonomics • Connector – Ease of use --- Signal fidelity and easy to connect/disconnect BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Steerable Curve -- Features MIGRATE OUR CUSTOMERS TO THE SteeroCath-Dx Polaris Dx Polaris X POLARIS X Distal Tubing 35D Pebax 6f Electode Number Decapolar Steering Mechanism Metal Unidirectional Steering Mechanism Tubing Unidirectional Steering Mechanism Metal Bidirectional Curve Standard Curve Various (see catalog) Curve Standard Proximal Tubing Continuous Braid 6f Proximal Tubing Continuous Braid 6f Proximal Tubing Continuous Braid 7f Handle Piston Handle Piston Handle Lever Connector Quick Connect Connector Connector Quick Connect Positioning Uni Directional CS Access Positioning Positioning Bi Directional CS Access Distal Tubing 50D Polyurethane 6f Distal Tubing Electode Number PLATFORM Electode Number Various (see catalog) Rotary & Quick Connect Enhanced Stability & CS Access 30D Polyurethane 7f Octopolar BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Feature Summary BSC Steerable Diagnostic Catheters POLARIS X STEEROCATH DX POLARIS Dx BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Clinical Requirements – CS CS CLINICAL REQUIREMENTS CATHETER FEATURES Access from the femoral vein Pushablity and trackability of proximal tubing. Access to the CS ostium from IVC Standard Steerable Curve Trackability over the "lip" at the entrance to the CS Pushability and torqueability of proximal tubing & steering of catheter. Trackability to push the catheter to the distal posterior section of the CS Morphology and timing along the length of the CS, including the CS ostium. Pushablity and trackability of proximal tubing. 2.5 - 5 - 2.5 or similar decapolar spacing BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Polaris X Brochure Clinical Benefits Technical Info POSITIONING Ordering Info BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Polaris X Brochure BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Polaris DX Brochure Clinical Benefits Technical Info POSITIONING Ordering Info BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Polaris DX Brochure BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Steerocath DX Brochure Clinical Benefits POSITIONING Technical Info Ordering Info BSC Confidential. For Internal Use Only. Do Not Copy or Distribute SteeroCath DX Brochure BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Polaris X Distal Tip Softness Comparison Excellent tool to use to show: Why is distal softness and lenght important to clinical performance What makes the Polaris X catheter superior Comparison to St Judes and Biosense Websters decapolar catheters BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Steerable Curve Diagnostic Catheters Competition COMPETITOR PRODUCT BSC Differentiation Bard Dynamic Tip™ Dynamic XT™ EP·XT™ Orbiter ST™ Polaris X™ / Polaris DX™ Polaris X™ / Polaris DX™ Polaris X™ / Polaris DX™ Steerocath DX™ St Jude Med./IBI Inquiry Steerable™ Livewire™ Reflexion™ IBI Inquiry™ Polaris Polaris Polaris Polaris X™ X™ X™ X™ / / / / Polaris Polaris Polaris Polaris DX™ DX™ DX™ DX™ Biosense Webster Std Deflectable Curve Johnson & Johnson Small Anatomy Deflectable Curve Parahissian Porterfield Polaris X™ / Polaris DX™ Medtronic Polaris X™ / Polaris DX™ Polaris X™ Stablemapr™ Marinr™ CS Polaris X™ / Polaris DX™ --------- BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Advanced Diagnostic Catheters Flutter Catheters used for mapping during flutter procedures. BSC Solution: • Polaris X™ • Map Freewall (1/2 Halo) • Pace from Polaris X and capture from Ablation catheter Biosense Webster Halo™ Decapolar along CT Data coming from Biosense Webster website – www.biosensewebster.com BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Advanced Diagnostic Catheters HALO vs Decapolar Mapping DECAPOLAR SIGNALS DECAPOLAR SIGNALS BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Advanced Diagnostic Catheters AFIB Circular Mapping Catheters Pulmonary Vein Mapping •10 or 20 (high density) electrode diagnostic catheter. •Shaped like a ring •Two versions •Steerable fixed diameter •Steerable with variable diameter St Jude Medical / IBI AFocus™ Biosense Webster Lasso™ Utilized to map the PV before, during and after RF Ablation Data coming from IBI website – www.ibiep.com Data coming from Biosense Webster website – www.biosensewebster.com BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Advanced Diagnostic Catheters AFIB Circular Mapping Catheters PVs showing fractionated potentials PVs showing no fractionated potentials BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Advanced Diagnostic Catheters AFIB PV Anatomical Mapping Pulmonary Vein Mapping BSC solution Constellation 32 electrodes (available 2008): • • • Map the ostium as well as the length of the PV 31mm size conforms to all PV diameters Under fluroscopy provides visual confirmation of PV Constellation™ BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Advanced Diagnostic Catheters AFIB PV Anatomical Mapping Constellation™ PVs with fractionated potentials prior to ablation PVs with no fractionated potentials after ablation BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Summary Clinical Applications HEART LOCATION CATHETER FAMILIES Fixed Explorer Explorer 360 Explorer 360jr Explorer 5f ST Explorer 6f ST Steerable Polaris Dx Polaris X SteeroCath-Dx CURVE SHAPES Fixed Josephson Cournand Multipurpose Conduction Steerable Standard (180º) Standard (270º) Longest HRA CS HIS RVA ----------------------------- -------- ---------------------- --------------- ---------------------- HRA CS HIS RVA -------- --------------- --------------- -------- --------------- --------------- --------------- BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Summary Steerable Curve Features STEERABLE FAMILIES Polaris Dx Polaris X Steerocath - Dx 7f -------- -------- 6f -------- -------- -------- -------- ---------------------- --------------- Octapolar Decapolar -------- -------- -------- --------------- --------------- ---------------------- --------------- QC - 8 pin QC - 10 pin -------- -------- TUBING FRENCH SIZE DISTAL TUBING FIRMNESS High Performance Less Traumatic RING ELECTRODE SPACING Quadripolar Hexapolar DISTAL FIXED CURVE SHAPES Standard (270º) Standard (180º) Longest PROXIMAL TUBING High Performance CONNECTOR Rotary - 4 pin QC - 6 pin BSC Confidential. For Internal Use Only. Do Not Copy or Distribute Diagnostic Catheters Summary Fixed Curve FIXED CURVE FAMILIES Features Explorer Explorer Explorer Explorer Explorer 360 360 Jr 5f ST 6f ST TUBING FRENCH SIZE 6f -------- -------- 5f -------- -------- -------- -------- -------- -------- -------- -------- -------- -------- -------- -------- --------------- -------- -------- -------- DISTAL TUBING FIRMNESS High Performance Less Traumatic RING ELECTRODE SPACING Bipolar Quadripolar Decapolar DISTAL FIXED CURVE SHAPES ---------------------- -------- --------------- --------------- -------- High Performance Pins -------- -------- ---------------------- --------------- --------------- --------------- --------------- Josephson Cournand Multipurpose Conduction PROXIMAL TUBING CONNECTOR Rotary - 4 pin QC - 4 pin QC - 10 pin -------- BSC Confidential. For Internal Use Only. Do Not Copy or Distribute