The Radiological Physics Center`s Anthropomorphic Quality

advertisement
The Radiological Physics Center’s Anthropomorphic Quality Assurance Phantom Program
Carrie F. Amador, Nadia Hernandez, Andrea Molineu, Paola Alvarez, and David Followill
Department of Radiation Physics
The University of Texas, M.D. Anderson Cancer Center, Houston, Texas
d.
b.
Esophagus
Anterior
Right
Left
Methods and Materials:
The RPC has developed an extensive phantom credentialing program consisting of
four different phantom designs: H&N, pelvis, lung and spine. These QA phantoms are
water-filled plastic shells, with imageable targets, avoidance structures, and
heterogeneities that contain thermoluminescence dosimeters (TLD) and radiochromic
film dosimeters. We currently have an inventory of 31 H&N, 10 pelvis, 9 lung, and 8
spine phantoms that are mailed to institutions nationally and internationally.
Spine
Bone
a.
c.
Posterior
Cord
e.
Figure 4. a. Spine phantom, b. axial film that was used for gamma analysis, c.
axial plane from the treatment planning system, d. gamma analysis, e. CT image.
The H&N phantom has two planning target volumes (PTV), the primary, a half lunar
shaped, and the secondary, a cylindrical shape, both composed of solid water
material, as shown in Figure 1. This phantom also has an acrylic organ at risk. The
primary PTV is prescribed 6.6 Gy and the secondary PTV is prescribed 5.4 Gy. The
organ at risk is limited to 4.5 Gy. The criteria for the H&N are 0.93 – 1.07 for the
TLD’s and greater than or equal to 85% of the pixels analysis from each the film
should pass a 7%/4 mm gamma test.
The pelvis has a polystyrene prostate, a nylon bladder and a wax rectum, as shown in
Figure 2. This phantom also includes both the left and right femoral heads. The
prescription dose is 6 Gy to at least 98% of the PTV. There are volume dose
constraints for the bladder and rectum. The criteria for the pelvis is 0.93 – 1.07 for the
TLD’s and greater than or equal to 85% of the pixels analysis from each the film
should pass a 7%/4 mm gamma test.
d.
b.
Anterior
Primary PTV
Right
Left
Phantoms Mailed
Introduction:
This study describes the Radiological Physics Center’s (RPC) anthropomorphic QA
phantom program for credentialing institutions for NCI-sponsored clinical trials using
advanced technologies. We introduce the phantoms, describe the program logistics
and report current results of four different phantom designs: head and neck (H&N),
pelvis, lung and spine.
Results:
Methods and Materials:
500
400
Spine
Pelvis
Lung
H&N
300
200
100
0
Figure 7. Phantom mail out per year.
Figure 5. Motion table available thru
the RPC website when using gating,
tracking, or breath hold.
Figure 6. Phantom request form,
available thru the RPC website.
The lung phantom has an elliptical shaped high impact polystyrene PTV that is
embedded in cork lung material, as shown in Figure 3. Institutions are instructed to
prescribe 6 Gy to at least 95% of the PTV with at least 99% of the PTV receiving 5.4 Gy.
The heart and spinal cord are limited to less than or equal to 4 Gy and at least 5 Gy,
respectively. A motion table is available with our lung phantom. The table is provided
when the institution uses gating, tracking or voluntary breath hold motion control
methods, as shown in Figure 5. The criteria for the lung phantom is 0.92 – 1.02 for the
TLD’s and greater than or equal to 85% of the pixels analysis from each the film should
pass a 7%/5 mm gamma test.
Year
Phantom Type
Number of
phantoms
Average time at
institution (days)
Average time from receipt of
inst. data to report sent (days)
H&N
31
27
20
PELVIS
10
21
17
LUNG
9
23
24
SPINE
8
35
23
Table 1. Number of phantoms and time logistics after phantom was mailed.
Phantom
H&N
Pelvis
Spine
Lung
Irradiations (all years)
1139
313
120
458
Pass (all years)
928 (81%)
265 (85%)
78 (65%)
361 (79%)
Pass (2011)
101 (93%)
45 (80%)
31 (78%)
68 (85%)
Criteria
7%/4mm
7%/4mm
5%/3mm
5%/5mm
Secondary PTV
c.
Posterior
a.
Organ at risk
e.
Figure 1. a. H&N phantom, b. axial film that was used for gamma analysis,
c. axial plane from the treatment planning system, d. gamma analysis, e. CT
image.
Primary PTV
Femoral Head
Femoral Head
Rectum
e.
a.
Rectum
Superior
Left
Right
b.
c.
Inferior
d.
Figure 2. a. Prostates phantom, b. coronal film that was used for gamma analysis,
c. coronal plane from the treatment planning system, d. gamma analysis, e. CT
image.
The spine phantom has an acrylic PTV, a solid water spinal cord surrounded by
polyester bone, a nylon heart, a solid water esophagus and cork lung, as shown in
Figure 4. The spine is prescribed a dose of 6 Gy with a minimum dose of 5.25 Gy. The
criteria for the spine is 0.93 – 1.07 for the TLD’s and greater than or equal to 85% of the
pixels analysis from each the film should pass a 7%/3 mm gamma test.
All of our phantoms are used to credential institutions for advanced technology clinical
trials. Techniques include IMRT, SBRT and 3D-CRT with motion management.
Institutions wishing to be credentialed request the phantom thru the RPC website. A
copy of the form is shown in Figure 6. Institutions are then prioritized based on a
multitude of factors including internal review board (IRB) approval of the protocol and the
image guide radiotherapy (IGRT) approval when relevant. Once the requested phantom
is available the physicist will be contacted either thru email or by phone to arrange the
logistics of the process. At this time the physicist at the institution must agree to irradiate
and return the phantom within 14 days.
Once at the institution, the phantom is assembled, filled with water and imaged. For
phantoms that do not have both imaging and the dosimeter inserts, external TLD’s are
placed on the shell of the phantom during imaging and are removed before treatment is
delivered. This imaging dose is removed from the target dose during analysis.
Next, a treatment plan is developed according to the irradiation instructions, using the
constraints that are based on advanced technology clinical trials. The institution is asked
to submit its treatment plan through the secure submission on the Advanced Technology
Consortium (ATC) website. The treatment plans for the lung and spine phantom are
analyzed for adherence to the prescription guidelines. All plans are used in phantom
analysis.
Table 2. Phantom historical and 2011 pass/fail rate.
RPC phantoms have been mailed out over 2100 times. In 2011, 444 of the phantoms
were mailed. On average it has taken institutions 26 days to return the phantom to the
RPC though for the spine it has taken an average 35 days, as shown in Table 1. The
report is sent on average 21 days after receipt of the electronic submission.
The overall pass rates form 2001 to 2010 for each of the phantoms are 81% for the
H&N, 85% for the pelvis, 79% for the lung, and 65% for the spine. In 2011 the percent
of phantoms meeting the acceptance criteria increased from overall rates by 12, 13
and 6 percentage points for the H&N, spine and lung phantoms, respectively. It fell by
5 percentage points for the pelvis phantom, as shown in Table 2. This increase may be
due to better treatment planning systems and algorithms. It may also be to better IMRT
planning and more experience with these phantoms.
In the course of ten years the phantom program has grown in phantom types and
amount of each. This is correlated with the number of advanced technology clinical
trials groups requiring the phantom for one of their credentialed requirements.
d.
b.
Heart
Anterior
Right
a.
Left
Primary PTV
e.
c.
After the plan is created the institution is asked to perform their customary patient
specific QA measurements. Next the phantom is positioned on the treatment couch and
the treatment is delivered. After delivery, the water is drained and the phantom is
returned to the RPC.
Cord
Posterior
Figure 3. a. Lung phantom, b. axial film that was used for gamma analysis, c.
coronal plane from the treatment planning system, d. gamma analysis, e. CT
image.
Once the phantom is returned, TLD are allowed to fade 14 days before being read. After
the TLD and the film are read, dose distributions are compared to the institution’s
electronically submitted treatment plan dosimetry data. TLD results, profiles through the
PTV, and gamma analysis are report to the institution.
Conclusions:
The RPC’s QA phantom program has been an effective and responsive QA tool for
assessing the use of advanced technologies in NCI sponsored clinical trials. The RPC
has been efficient in its mailing of phantoms, and analyzing and reporting results.
Support:
Work supported by PHS grant CA10953 and CA081647 (NCI, DHHS).
Download