Convection Temperature Device - CTD 600

advertisement
Convection Temperature Device
CTD 600
www.anton-paar.com
Convection Temperature Device - CTD 600
Gas Convection heating and cooling
2
Typical working range:
With LN2:
ambient – 600°C
-150°C – 600°C
Advantages:
- Large temperature range
- Fast heating and cooling
- Enough working space for special accessories
like solid torsion bar fixture etc.
Disadvantages:
- LN2 needed for below ambient temperatures
Convection Temperature Device - CTD 600
Gas convection and radiation heating and cooling
3

-150°C – 600°C

Symmetrical design

Extremely small
temperature gradients

Controller temperature
equals sample temperature

No temperature overshoots

Easy sample access

DMTA with solid torsion
bar fixture etc.
Convection Temperature Device- CTD 600
Solid torsion bar fixture (SRF)
UXF
4
SER 2
Convection Temperature Device- CTD 600
 Temperature
Standard:
Low temperature option:
Temperature sensor:
ambient to 600°C
-150°C to 300°C (600°C)
Pt100 Class B
 Supply pressure
Gas:
Liquid nitrogen:
max. 5bar
up to 1.2 bar
(0.8 to 1.0 recommended)
 Consumption rate
Gas heating:
Air shaft cooling:
Liquid nitrogen:
5
1.5 Nm³/h up to 400°C
0.9 Nm³/h from 400 to 500°C
0.5 Nm³/h from 500 to 600°C
1.2 Nm³/h
2-5 L/h
Convection Temperature Device - CTD 600
Chamber cooling
with liquid or air
Gas
or cooled
Nitrogen
gas
Heater
The Sample is heated by
convection and radiation
Pt100 (Heater)
6
Pt100 (Heater)
Pt100 (Sample)
Convection Temperature Device- CTD 600
Heater
Actual
Temp.
Set value
Power
Ready
Set value
Heater
7
Actual
Temp.
Set value
Set value
Actual
Temp.
Controlled
Power
EVU10
Convection Temperature Device - CTD 600
Schematic diagram
P
3
1
2
7
-150.0
P
3
5
6
TC30
Temperature Controll Unit
1
25.0°C
4
8
1....GAS
2....LSVE
3....Precision pressure controller
4....Dewar
8
5....EVU10
6....CTD600
7....TC30
8....water bath or air cooling
Convection Temperature Device - CTD 600
Lower measuring plates
3
4
2
5
1
1....Holder for lower plates
2....lower measuring plate 8mm
3....lower measuring plate 25mm
4....lower measuring plate 50mm
5....lower measuring plate with groove
9
Convection Temperature Device - CTD 600
Lower disposable systems
4
6
3
2
2
5
1
10
1....Holder for lower plates
2....Fixture for disposable plates
3....Disposable lower measuring plate 50 mm
4....Disposable lower measuring plate25 mm
5....Fixture for disposable dishes
6....Disposable dishes
Convection Temperature Device - CTD 600
UV-Option for Photo DMTA
 Based on the CTD 600
 Temperatures up to 200°C
 Measuring plate up to
25 mm diameter recommended
11
Convection Temperature Device - CTD 600
UV accessory CTD 600 – lower part for Photo DMTA
Glass plate
diameter: 40 mm
height: 4 mm
Coupling ring
with O-ring
Light guide
diameter 3 mm
12
Light guide
diameter 3 mm
Convection Temperature Device - CTD 600
Features and Benefits
The CTD 600 has mayor advantage:
• Extremely small temperature gradients
• No temperature overshoots
• PT100 signal reflects true sample temperature
• No icing at low temperatures
• Compact, robust design
• Automatic adjustment of LN2 consumption as evaporation unit
actively controls continuous flow of liquid nitrogen
• Easy to open, good sample access
• Sample observation during measurement
• Temperature-isolated jacket (hand-warm for safe use)
• Actively cooled jacket according to EN61010-1:2001
• Modular configuration (DMTA torsion/tension, Photo DMTA/UV,
reaction kinetics, SER extensional viscosity, disposable plates, etc.
13
Convection Temperature Device - CTD 600
Temperature Calibration - Software supported Automatic Temperature Calibration
with Certified Temperature Sensor
2 Sensors available for different temperature ranges
Working range:
-40°C - 200°C
Calibration Temperatures:
10°C and 70°C
Measurement Uncertainty:
100mK
Working range:
-150°C – 650°C
Calibration Temperatures:
Eight temperatures between
-40°C and 500°C
Measurement Uncertainty:
< 200°C: 100mK; > 200°C 200mK
14
Download