Cardiotocography
1
Cardiotocography (CTG) is defined as
the graphic recording of fetal heart
rate and uterine contractions by the
use of electronic devices indicated for
the assessment of fetal condition.
2
This can be deciphered from the name itself –
Cardio (Heart) Toco (Uterine Contractions) and
Graph (Recording). A CTG Test during pregnancy is
usually performed during the third trimester as
well as during labour. The correct utilization of this
test can help prevent the baby from dying due to a
shortage of oxygen.
3
A normal fetal heart rate is between 110 to 160
beats a minute. The sound of the beat is usually
strong and regular. It is normal to have some
changes in the fetal heart rate during labor, but
drastic changes in heart rate before or after a
contraction may indicate that the fetus is in
distress. A fetus with large changes in
heart rate may need to be removed from
the womb immediately, by Caesarean section.
4
CTG Recording Parameters
❑ Fetal Heart Rate
➢ External Cardiotocography
➢ Internal Cardiotocography
❑ Uterine Contractions
5
➢ External Cardiotocography
▪ This is the most common form of testing and is done by placing Two
Transducers on the abdomen of the mother. One Transducer emits
an ultrasound frequency which is used to detect the baby’s
heartbeat. The other Transducer is used to record the pressure on
the mother’s abdomen along with the contractions. Sometimes a gel
may be applied to get a stronger signal. The reading by the device is
given in the form of a loud beating sound.
▪ Prior to going into labour, the mother
would be asked to press a button on the
machine each time she feels the baby
move.
6
➢ Internal Cardiotocography
▪ Internal fetal heart rate monitoring uses an electronic transducer
connected directly to the fetal skin. A wire electrode is attached to the
fetal scalp or other body part through the cervical opening and is
connected to the monitor. This type of electrode is sometimes called a
spiral or scalp electrode. Internal monitoring provides a more accurate
and consistent transmission of the fetal heart rate than external
monitoring because factors such as movement do not affect it.
▪ Internal monitoring may be used when
external monitoring of the fetal heart rate is
inadequate, or closer surveillance is needed.
7
➢ Uterine Contraction Measurement
▪ Uterine contractions are measured with a Strain Gage Transducer or
Pressure Sensor usually strapped with a belt around the abdomen
positioned over the fundus (top) of the uterus. This indirectly records
contractions through sensing changes in skin tension arising from the
uterine muscular activity.
8
➢ Uterine Contraction Measurement
▪ During labour, it is important to have some indication of how labour is
progressing, and to monitor the build-up of contractions, both in terms
of frequency and strength. Such contractions have a powerful impact
on the fetus. Therefore, fetal heart rate is measured simultaneously.
9
➢ Relationship Between Uterine Contractions & Fetal Heart Rate
▪ During Uterine contractions the heart rate will decrease and revert to
previous levels after the end of the contraction. If there is a delay in the
heart rate returning to its previous level it can indicate that there is fetal
distress.
Fetal Heart Rate
Uterine Contractions
10
11
12
13
14
➢ Ultrasound Transducer-(Noninvasive)
15
➢ Fetal Scalp Electrode-(Invasive)
16
➢ TOCO Transducer
17
➢ Event Marker
18
➢ Recorder & Paper
19
➢ Transducer Belts & US Gel
20
▪ Comprehensive perinatal monitoring and the flexibility
and convenience that give you instant access to the
detailed information you need during labor and
delivery. Having that information readily available helps
you provide a positive birthing experience for mother
and baby.
❑ Fetal Heart Rate
➢
➢
❑
❑
❑
❑
❑
External Cardiotocography
Internal Cardiotocography
Uterine Contractions
ECG of Mother
Spo2 of Mother
NIBP of Mother
Temperature of Mother
21
❑ User errors are common with fetal monitoring, e.g.
✓ Incorrect Connections of the transducers
✓ Incorrect Loading of the Paper or loading of the Wrong paper
✓ Lack of Gel – gel can dry out over a period of time in long labour
monitoring sessions; gels may flow away from where it’s needed.
✓ Loose Belt – it is important to apply enough belt tension to ensure that
firm contact with the skin is maintained.
✓ Transducer Movement – it may have slipped due to maternal movement,
pulling the cable, etc.
✓ Fetal Movement – the fetus may have moved out of the ultrasound beam
✓ Poor transducer Positioning. This is an acquired skill and should include
palpation of the fetus to determine the best position for the transducer.
22
▪ Power Supply Problems (SMPS circuit or Fuses).
▪ Be sure that Gel is being used between the US Transducer and the patient.
▪ The US Transducer is the most sensitive part of CTG. When it Breaks no output is heard, even You can
check the Transducer operation by gently tapping the Transducer surface about once per second. If this is
not detected, there is certainly a Transducer problem.
▪ The US probe consists of the Transducer assembly, the Cable, and a multi-pin fixable connector. The
cable contains between 5 and 20 separate conductors. The most frequent malfunction occurs as a result
of a break in one or more of the cable conductors. Such malfunctions are usually the result of
mishandling of the cable or of soaking it with Gel.
▪ Transducer are Expensive and typically cannot be Repair only replaced with new.
▪ Using of Gel with TOCO Transducer cause Malfunctions & wrong results.
▪ Applying little pressure on TOCO Transducer & check the display for 0 to Increasing reading to check the
transducer proper operation.
▪ The face of the probe is usually an acoustic lens (Crystal). It must be handled with care. Do not drop the
probe, and avoid scratching the face with sharp objects. Keep the probe assembly clean of oil and gel.
Always clean the probe and cable with a tissue or damp cloth, after finishing work.
▪ Recorder Problem. Thermal Printer Head need to be clean or replace.
▪ Wrong paper or paper direction cases malfunctions.
23
24
✓Electrical Patient Safety Testing.
❑ Protective Earth Continuity Test
❑ Insulation Resistance Test
❑ Earth Leakage Current
❑ Enclosure Leakage Current
❑ Patient Leakage Current
❑ Patient Auxiliary Current
❑ Mains On Applied Parts (Patient Leakage)
25