CONDITION APPRAISAL/TECHNICAL REPORT TO
PICU WARD
ON
INFUSION PUMP
MODEL:MEDCAPTAIN SYS-6010A
OF
HOSPITAL WANITA DAN KANAK-KANAK SABAH
NO SIRI PENDAFTARAN:SBH915/001036
Date Prepared: 10/1/2023
TECHNICAL REPORT
………………………..
Prepared By:
Sylvester Junir
Technician, BEMS
CONDITION APPRAISAL/TECHNICAL REPORT OF
Infusion Pump
MODEL: Medcaptain Sys-6010A
1.0
INTRODUCTION
The purpose of this report is to provide a condition appraisal/technical report infusion pump
Medcaptain sys-6010A.
2.0
SYSTEM OVERVIEW
2.1
TECHNICAL DATA
Type Description
: Infusion pump
Manufacturer
: Medcaptain Medical Technology Co Ltd
Model
: Medcaptain SYS-6010A
Type code
: 13-215
Serial No
: 91420120031122
Asset No
: Sbh915/001036
Department
: Picu
Purchased Date
: 14-Apr-2020
Purchased Price
: RM3,150.00
3.0
PROBLEM DESCRIPTION
As per request, user reported that after the end of infusion rate change from Rate(ml/h)
to dots/min.
4.0
FINDINGS & ACTION TAKEN
We set infusion as :
Setting
Rate = 120ml/h
Volume = 12.07ml
Picture result :
Test using IV drip general 20
Result (using infusion pump analyzer)
Flow average ml/h = 100.87
Volume (ml)= 10.07
5.0 CONCLUSION
In conclusion, infusion was deliver within range. To ensure the infusion accuracy,
recommend using B.Braun and Medcaptain IV set. After several test, we don’t experience
infusion mode was changing after infusion was finished. This unit was in good condition and
ready to use.