Sedafiat Sdn Bhd CONDITION APPRAISAL/TECHNICAL REPORT TO PICU WARD ON INFUSION PUMP MODEL:MEDCAPTAIN SYS-6010A OF HOSPITAL WANITA DAN KANAK-KANAK SABAH NO SIRI PENDAFTARAN:SBH915/001036 Date Prepared: 10/1/2023 TECHNICAL REPORT ……………………….. Prepared By: Sylvester Junir Technician, BEMS CONDITION APPRAISAL/TECHNICAL REPORT OF Infusion Pump MODEL: Medcaptain Sys-6010A 1.0 INTRODUCTION The purpose of this report is to provide a condition appraisal/technical report infusion pump Medcaptain sys-6010A. 2.0 SYSTEM OVERVIEW 2.1 TECHNICAL DATA Type Description : Infusion pump Manufacturer : Medcaptain Medical Technology Co Ltd Model : Medcaptain SYS-6010A Type code : 13-215 Serial No : 91420120031122 Asset No : Sbh915/001036 Department : Picu Purchased Date : 14-Apr-2020 Purchased Price : RM3,150.00 3.0 PROBLEM DESCRIPTION As per request, user reported that after the end of infusion rate change from Rate(ml/h) to dots/min. 4.0 FINDINGS & ACTION TAKEN We set infusion as : Setting Rate = 120ml/h Volume = 12.07ml Picture result : Test using IV drip general 20 Result (using infusion pump analyzer) Flow average ml/h = 100.87 Volume (ml)= 10.07 5.0 CONCLUSION In conclusion, infusion was deliver within range. To ensure the infusion accuracy, recommend using B.Braun and Medcaptain IV set. After several test, we don’t experience infusion mode was changing after infusion was finished. This unit was in good condition and ready to use.