A5 Anesthesia System Service Manual Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rent, adaption and translation of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. 、 and are the registered trademarks or trademarks owned by Mindray in China and other countries. All other trademarks that appear in this manual are used only for editorial purposes without the intention of improperly using them. They are the property of their respective owners. Contents of this manual are subject to changes without prior notice. Revision History This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical specification change. Contents of this manual are subject to change without prior notice. Revision 1.0 is the initial release of the document. Revision number: 9.0 Release time: 2022-11 © Copyright 2011-2022 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. I Foreword This Service Manual is intended as a guide for technically qualified personnel performing repair and calibration procedures. Warnings, Cautions, and Notes Warnings, Cautions and Notes Please read and adhere to all warnings, cautions and notes listed here and in the appropriate areas throughout this manual. A WARNING is provided to alert the user to potential serious outcomes (death, injury, or serious adverse events) to the patient or the user. A CAUTION is provided to alert the user to use special care necessary for the safe and effective use of the device. They may include actions to be taken to avoid effects on patients or users that may not be potentially life threatening or result in serious injury, but about which the user should be aware. Cautions are also provided to alert the user to adverse effects on this device of use or misuse and the care necessary to avoid such effects. A NOTE is provided when additional general information is applicable. Warnings WARNING: Whenever using anesthetic gases, nitrous oxide, oxygen, or any hospital gas always follow the appropriate agent evacuation/collection procedures. Use the hospital gas evacuation system. WARNING: Use only an approved lubricant on any O-ring in contact with oxygen. Krytox® is the recommended oxygen service lubricant. WARNING: For continued protection against fire hazard, replace all fuses with the specified type and rating. WARNING: In order to prevent an electric shock, the machine (protection class I) may only be connected to a correctly grounded mains connection (socket outlet with grounding contact). WARNING: Remove all accessory equipment from the shelf before moving the anesthesia machine over bumps or on any inclined surface. Heavy top loading can cause the machine to tip over causing injury. WARNING: Possible explosion hazard. Do not operate machine near flammable anesthetic agents or other flammable substances. Do not use flammable anesthetic agents (i.e. ether or cyclopropane.) WARNING: The use of anti-static or electrically conductive respiration tubes, when utilizing high frequency electric surgery equipment, may cause burns and is therefore not recommended in any application of this machine. WARNING: Possible electric shock hazard. The machine may only be opened by authorized service personnel. WARNING: Avoid exposure to respiratory gases by always directing the fresh gas flow from the fresh gas outlet to the waste gas scavenger. II Cautions CAUTION: This device uses high pressure compressed gas. When attaching or disconnecting backup gas cylinders, always turn the cylinder valves slowly. Use the A5 flow meters to bleed down the pressure, watching the cylinder gauge indicate the depleting cylinder pressure, before disconnecting the cylinder from the yoke. Always open and close cylinder valves fully. CAUTION: This device operates using compressed gas at high pressures from the hospital central supply. When connecting gas supply lines attach the hose connection to the machine before connecting the quick disconnect fitting to the hospital source. Disconnect the supply hose from the hospital source connection prior to disconnecting it from the A5 gas connection fittings. CAUTION: Refer to the 3.3Equipment Maintenance Schedule for assistance when performing scheduled periodic maintenance. CAUTION: Do not leave gas cylinder valves open if the pipeline supply is in use and the system master switch is turned to 'ON'. If used simultaneously, cylinder supplies could be depleted, leaving an insufficient reserve supply in the event of pipeline failure. CAUTION: Use cleaning agent sparingly. Excess fluid could enter the machine, causing damage. CAUTION: This machine must only be operated by trained, skilled medical staff. CAUTION: Perform the electrical safety inspection as the last step after completing a repair or after routine maintenance. Perform this inspection with all covers, panels, and screws installed. CAUTION: After changing the CO2 absorbent, carry out a system leak test. CAUTION: Only Selectatec™ compatible vaporizers with Interlock-System may be used with the A5 unit. CAUTION: After each exchange of a vaporizer, carry out a system Leak test. CAUTION: Do not clean the machine while it is on and/or plugged in. CAUTION: Pressing “cancel” at any time during the procedure will cancel the session's settings and reload the previously-stored calibration coefficients. CAUTION: Depleted soda lime changes color. Replace the soda lime if approximately 2/3 of the absorber content is discolored. CO2 absorbent can be safely changed without stopping mechanical ventilation. CAUTION: This equipment contains parts which are easily damaged due to electrostatic discharge (ESD). Follow ESD prevention program when touching, taking out, or inserting parts or components. Notes NOTE: Unauthorized servicing may void the remainder of the warranty. Check with the factory or with a local authorized distributor to determine the warranty status of a particular instrument. III FOR YOUR NOTES IV Contents Warnings, Cautions, and Notes .............................................................................................II Warnings, Cautions and Notes ................................................................................................. II Warnings .................................................................................................................................. II Cautions .................................................................................................................................. III Notes ....................................................................................................................................... III Contents ................................................................................................................................... 1 1 Theory of Operation ........................................................................................................ 1-1 1.1 Introduction ..................................................................................................................... 1-1 1.2 Electrical and Pneumatic Connections ............................................................................ 1-3 1.2.1 Electrical Connection ......................................................................................... 1-3 1.2.2 Pneumatic Connection ....................................................................................... 1-5 1.2.3 Connections between Pneumatic Circuit, Breathing System and Ventilator Control Board ............................................................................................................ 1-10 1.3 Pneumatic Circuit .......................................................................................................... 1-13 1.3.1 Pneumatic Circuit Diagram .............................................................................. 1-13 1.3.2 Parts List .......................................................................................................... 1-14 1.3.3 Key to Symbols ................................................................................................ 1-15 1.3.4 Descriptions ..................................................................................................... 1-15 1.4 Components of the Anesthesia System ......................................................................... 1-29 1.4.1 Auxiliary Electrical Outlets .............................................................................. 1-29 1.4.2 Lighting Board ................................................................................................. 1-30 1.5 Breathing System .......................................................................................................... 1-32 1.5.1 Introduction ...................................................................................................... 1-32 1.5.2 Mechanical Mode, Inspiration.......................................................................... 1-32 1.5.3 Mechanical Mode, Expiration .......................................................................... 1-33 1.5.4 Manual Mode, Inspiration ................................................................................ 1-34 1.5.5 Manual Mode, Expiration ................................................................................ 1-35 1.5.6 Pneumatic PEEP............................................................................................... 1-35 1.5.7 Ventilator in Standby ........................................................................................ 1-35 1.5.8 Components of the Breathing System .............................................................. 1-35 1.6 User Interface ................................................................................................................ 1-36 1.6.1 Display ............................................................................................................. 1-36 1.6.2 Main Control Board ......................................................................................... 1-42 1.7 Ventilation Control and Drive ....................................................................................... 1-44 1.7.1 Motherboard ..................................................................................................... 1-44 1.7.2 Monitoring Module .......................................................................................... 1-54 1.7.3 Battery .............................................................................................................. 1-59 1.7.4 Heater of the Breathing System ....................................................................... 1-60 1.7.5 Anesthesia Calculation and Infrared Module Rack .......................................... 1-60 1.7.6 Power Board ..................................................................................................... 1-62 1.8 O2-driven Ventilator...................................................................................................... 1-64 1.8.1 O2-driven Ventilator......................................................................................... 1-64 1.8.2 High-pressure Drive Gas Regulator (200kPa, 29psi) ....................................... 1-64 1.8.3 Drive Gas Assembly ......................................................................................... 1-64 1.8.4 Tube Color Coding ........................................................................................... 1-64 2 Installation Guide ............................................................................................................. 2-1 2.1 Preparation ...................................................................................................................... 2-1 2.2 Assembly ......................................................................................................................... 2-1 2.2.1 Unpacking and Setup ......................................................................................... 2-1 2.2.2 Breathing System, Breathing System Accessories and Checkout Procedures ..2-11 2.2.3 AGSS Connections ............................................................................................2-11 1 2.2.4 Vaporizer .......................................................................................................... 2-12 2.2.5 Backup Cylinder............................................................................................... 2-14 2.2.6 Monitoring Products Mounting and Electrical Connection (if available) ........ 2-14 2.3 Factory Setup ................................................................................................................ 2-15 2.3.1 Function Activation .......................................................................................... 2-16 2.3.2 Drive Gas ......................................................................................................... 2-16 2.3.3 Flowmeter Standard ......................................................................................... 2-17 2.3.4 Flowmeter Pipeline .......................................................................................... 2-17 2.3.5 ACGO............................................................................................................... 2-17 2.3.6 Module Rack .................................................................................................... 2-17 2.3.7 AG Module ....................................................................................................... 2-17 2.3.8 BIS Module ...................................................................................................... 2-17 2.3.9 CO2 Module ..................................................................................................... 2-17 2.3.10 NMT Module ................................................................................................. 2-18 2.4 Functional Tests ............................................................................................................ 2-18 2.4.1 Gas Delivery System Tests ............................................................................... 2-18 2.4.2 Breathing System Leak Test ............................................................................. 2-18 2.4.3 O2 Sensor Calibration ...................................................................................... 2-24 2.5 Gas Supply Leak Tests .................................................................................................. 2-30 2.5.1 N2O Cylinder Leak Test................................................................................... 2-30 2.5.2 O2 Cylinder Leak Test ..................................................................................... 2-30 2.5.3 AIR Cylinder Leak Test.................................................................................... 2-30 2.5.4 N2O Pipeline Supply Pressure Leak Test ......................................................... 2-31 2.5.5 O2 Pipeline Supply Pressure Leak Test............................................................ 2-31 2.5.6 Air Pipeline Supply Pressure Leak Test ........................................................... 2-31 2.6 Breathing System Checks ............................................................................................. 2-32 2.6.1 AGSS Test (if available) ................................................................................... 2-32 2.6.2 Internal Gas Connections Test .......................................................................... 2-32 2.6.3 Alarm of Drive Gas Pressure Low, O2&N2O Cutoff Functional Test ............. 2-32 2.7 Performance Verification............................................................................................... 2-33 2.7.1 Standby Mode Ventilation Test......................................................................... 2-33 2.7.2 Manual Mode Ventilation Test ......................................................................... 2-33 2.7.3 Apnea Alarm Test ............................................................................................. 2-33 2.7.4 Alarm Silence Test ........................................................................................... 2-33 2.7.5 VCV Adult Ventilation Mode Test ................................................................... 2-33 2.7.6 VCV Pediatric Ventilation Mode Test .............................................................. 2-34 2.7.7 Airway Disconnection Alarm Test ................................................................... 2-35 2.7.8 PCV Adult Ventilation Mode Test .................................................................... 2-35 2.7.9 Pressure Support (PS) Ventilation Mode Test .................................................. 2-35 2.8 Alarm Function ............................................................................................................. 2-36 2.8.1 Setup................................................................................................................. 2-36 2.8.2 FiO2 Too Low Alarm Test ................................................................................ 2-37 2.8.3 FiO2 Too High Alarm Test ............................................................................... 2-37 2.8.4 Peak Pressure Alarm Test ................................................................................. 2-37 2.8.5 MV Alarm Test ................................................................................................. 2-38 2.9 Vaporizer Test ................................................................................................................ 2-38 2.9.1 Vaporizer Interlock Test ................................................................................... 2-38 2.9.2 Vaporizer Accuracy Test ................................................................................... 2-39 2.10 Other Tests .................................................................................................................. 2-40 2.10.1 Power Failure Alarm Test ............................................................................... 2-40 2.10.2 Work Light and Auxiliary Gas Supply Backlight Test ................................... 2-40 2.11 Electric Safety Test ...................................................................................................... 2-40 2.11.1 Test on Auxiliary Electrical Outlet ................................................................. 2-40 2.11.2 Electrical Safety Inspection ............................................................................ 2-41 2.11.3 Electrical Safety Inspection Form .................................................................. 2-41 3 Equipment Maintenance ................................................................................................. 3-1 2 3.1 Repair Policy ................................................................................................................... 3-1 3.2 Periodical Maintenance Kit ............................................................................................. 3-1 3.3 Equipment Maintenance Schedule .................................................................................. 3-1 3.4 Pre-operation test ............................................................................................................ 3-2 3.5 Maintenance Kit (12-month) Replacement ..................................................................... 3-2 3.6 Functional Tests .............................................................................................................3-11 3.6.1 Gas Delivery System Tests ................................................................................3-11 3.6.2 Breathing System Leak Test ..............................................................................3-11 3.6.3 Check the Sensor Zero Point ............................................................................ 3-15 3.6.4 Check the Accuracy of Flow Sensors ............................................................... 3-17 3.6.5 Check the Accuracy of Pressure Sensors ......................................................... 3-20 3.7 High Pressure Tests ....................................................................................................... 3-25 3.7.1 N2O Cylinder Leak Test................................................................................... 3-25 3.7.2 O2 Cylinder Leak Test ..................................................................................... 3-25 3.7.3 AIR Cylinder Leak Test.................................................................................... 3-26 3.7.4 N2O Pipeline Supply Test ................................................................................ 3-26 3.7.5 O2 Pipeline Supply Test ................................................................................... 3-27 3.7.6 AIR Pipeline Supply Test ................................................................................. 3-27 3.8 Breathing System Tests ................................................................................................. 3-28 3.8.1 Sample Gas Emission Test/AGSS Test (optional configuration) ..................... 3-28 3.8.2 Internal Pneumatic Connection Test................................................................. 3-28 3.8.3 Alarm of Drive Gas Pressure Low, N2O Cutoff Test ....................................... 3-29 3.9 Performance Verification............................................................................................... 3-29 3.9.1 Standby Mode Ventilation Test......................................................................... 3-29 3.9.2 Manual Mode Ventilation Test ......................................................................... 3-30 3.9.3 Apnea Alarm Test ............................................................................................. 3-30 3.9.4 Alarm Silence Test ........................................................................................... 3-30 3.9.5 VCV Adult Ventilation Mode Test ................................................................... 3-30 3.9.6 VCV Pediatric Ventilation Mode Test .............................................................. 3-31 3.9.7 Airway Disconnection Alarm Test ................................................................... 3-32 3.9.8 PCV Adult Ventilation Mode Test .................................................................... 3-32 3.9.9 Pressure Support (PS) Ventilation Mode Test .................................................. 3-33 3.10 Alarm and Automatic Fault Protection Function ......................................................... 3-34 3.10.1 Setup............................................................................................................... 3-34 3.10.2 FiO2 Too Low Alarm Test .............................................................................. 3-34 3.10.3 FiO2 Too High Alarm Test ............................................................................. 3-35 3.10.4 Peak Pressure Alarm Test ............................................................................... 3-35 3.10.5 MV Alarm Test ............................................................................................... 3-36 3.11 Vaporizer Test .............................................................................................................. 3-36 3.11.1 Vaporizer Interlock Test.................................................................................. 3-36 3.11.2 Vaporizer Accuracy Test ................................................................................. 3-37 3.11.3 Vaporizer Leak Test ........................................................................................ 3-38 3.13 Other Tests .................................................................................................................. 3-39 3.13.1 Power Failure Test .......................................................................................... 3-39 1.13.2 Work Light and Auxiliary Gas Supply Backlight Test ................................... 3-39 3.14 Auxiliary AC Electrical Outlets Test ........................................................................... 3-39 3.14.1 Anesthesia System Installation Checklist ....................................................... 3-39 3.15 Electric Safety Test...................................................................................................... 3-40 3.15.1 Test on Auxiliary Electrical Outlet ................................................................. 3-40 3.15.2 Electrical Safety Inspection ............................................................................ 3-40 3.15.3 Electrical Safety Inspection Form .................................................................. 3-42 4 Calibration ........................................................................................................................ 4-1 4.1 Introduction ..................................................................................................................... 4-1 4.2 Warnings, Precautions, and Notes ................................................................................... 4-1 4.2.1 Warnings............................................................................................................. 4-1 4.2.2 Cautions ............................................................................................................. 4-1 3 4.2.3 Notes .................................................................................................................. 4-2 4.3 System Calibration .......................................................................................................... 4-2 4.3.1 Flow Calibration (User) ..................................................................................... 4-3 4.3.2 Flow Calibration (Factory) ................................................................................. 4-6 4.3.3 Pressure Calibration (Factory) ......................................................................... 4-28 4.3.4 Pressure and Flow Zeroing (Factory) ............................................................... 4-41 4.3.5 Electronic Flowmeter Zeroing (User) .............................................................. 4-44 4.3.6 O2 Sensor Calibration ...................................................................................... 4-46 4.3.7 External AG Calibration ................................................................................... 4-52 4.3.8 CO2 Module Calibration .................................................................................. 4-52 4.3.9 ORC Calibration............................................................................................... 4-53 4.3.10 High-Pressure Gas Cylinder Regulator Calibration ....................................... 4-58 5 Repair and Troubleshooting ............................................................................................ 5-1 5.1 Troubleshooting Guidelines ............................................................................................ 5-1 5.1.1 Identify Problems ............................................................................................... 5-1 5.1.2 Avoid Shorting Component Leads ..................................................................... 5-1 5.1.3 Use Proper Tools ................................................................................................ 5-1 5.1.4 Clean up the Repair Area ................................................................................... 5-1 5.2 Technical Alarms ............................................................................................................. 5-1 5.2.1 Startup Alarm Messages ..................................................................................... 5-2 5.2.2 CPU Board Runtime Alarms .............................................................................. 5-6 5.2.3 Power Board Runtime Alarms............................................................................ 5-6 5.2.4 Electronic Flowmeter Board Runtime Alarms ................................................... 5-7 5.2.5 VCM Runtime Alarms ..................................................................................... 5-10 5.2.6 External AG Module Runtime Alarms ............................................................. 5-14 5.2.7 BIS Module Runtime Alarms ........................................................................... 5-16 5.2.8 CO2 Module Runtime Alarms.......................................................................... 5-19 5.2.9 NMT Module Runtime Alarm .......................................................................... 5-21 5.3 Pneumatic System Problems ......................................................................................... 5-22 5.3.1 Tools for on-site Maintenance .......................................................................... 5-22 5.3.2 Gas Supplies and Drive Gas ............................................................................. 5-31 5.3.3 Anesthetic Gas Delivery System ...................................................................... 5-38 5.3.4 Breathing System ............................................................................................. 5-50 5.3.5 Tidal Volume .................................................................................................... 5-63 5.4 Sensor and Valve Related Failures ................................................................................ 5-65 5.4.1 Preparations before Using Diagnostic Tests ..................................................... 5-66 5.4.2 Zero Point Problems of Flow and Pressure Sensors ......................................... 5-66 5.4.3 Connections and Measurement Problems of the Flow Sensors........................ 5-67 5.4.4 Connections and Measurement Problems of the Pressure Sensors .................. 5-68 5.4.5 Opening State Problems of the Inspiratory Valve ............................................ 5-69 5.4.6 Opening State Problems of the PEEP Safety Valve.......................................... 5-69 5.4.7 Opening State Problems of the PEEP Valve ..................................................... 5-70 5.5 Hardware and Electrical Problems ................................................................................ 5-70 5.6 Software Upgrade and Software Function Activation ................................................... 5-72 5.6.1 1.6.1 Software Upgrade.................................................................................... 5-72 5.6.2 Software Function Activation........................................................................... 5-75 6 Repair and Disassembly .................................................................................................. 6-1 6.1 Prepare for Disassembly ................................................................................................. 6-1 6.1.1 Tools ................................................................................................................... 6-1 6.1.2 Preparations ........................................................................................................ 6-1 6.1.3 Bleed Gas Pressure ............................................................................................. 6-1 6.2 Disassemble the Assemblies ........................................................................................... 6-2 6.2.1 Disassemble the Internal Assemblies of the Machine Upper Half ..................... 6-2 6.2.2 Disassemble Hardware Box Assembly............................................................... 6-7 6.2.3 Disassemble the Work Surface ......................................................................... 6-17 4 6.2.4 Disassemble the Display Assembly.................................................................. 6-24 6.2.5 Remove the Module Rack Box Assembly ........................................................ 6-30 6.2.6 Remove the Module Rack Fan ......................................................................... 6-31 6.2.7 Remove the Instrument Panel Assembly .......................................................... 6-32 6.2.8 Remove the Auxiliary Gas Supply Assembly .................................................. 6-35 6.2.9 Remove the Circuit Arm Cover ........................................................................ 6-36 6.2.10 Remove the AGSS Assembly ......................................................................... 6-36 6.2.11 Remove the Built-in AG Module (when configured with AG module) ......... 6-37 6.2.12 Remove the AG Module Inlet Tube Assembly (when configured with AG module) ..................................................................................................................... 6-38 6.2.13 Disassemble the Base Assembly .................................................................... 6-39 6.3 Disassemble the Breathing System ............................................................................... 6-44 6.3.1 Remove the O2 Sensor and O2 Cell Unit......................................................... 6-44 6.3.2 Remove the Breathing Tubes ........................................................................... 6-45 6.3.3 Remove the Flow Sensor ................................................................................. 6-46 6.3.4 Remove the Manual Bag .................................................................................. 6-47 6.3.5 Remove the Absorber Canister Assembly ........................................................ 6-47 6.3.6 Remove the CO2 Bypass Assembly ................................................................. 6-49 6.3.7 Remove the Circuit Main Body Assembly ....................................................... 6-50 6.3.8 Remove the Bellows Assembly ........................................................................ 6-50 6.3.9 Remove the Pop-off Valve Assembly ............................................................... 6-51 6.3.10 Remove the Expiratory/Inspiratory Check Valve Assembly .......................... 6-51 6.3.11 Remove the Water Collection Cup ................................................................. 6-52 6.3.12 Remove the Airway Pressure Gauge .............................................................. 6-53 6.3.13 Remove the Bag Arm Assembly .................................................................... 6-53 6.3.14 Remove the Upper Rear Cover Assembly and Lower Rear Cover Assembly 6-54 6.3.15 Remove the Upper Front Cover Assembly, Median Plate Assembly, and Lower Front Cover Assembly............................................................................................... 6-55 6.3.16 Remove the Auto/Manual Switch Assembly .................................................. 6-57 6.3.17 Remove the APL Valve Assembly .................................................................. 6-58 7 Replacement Parts ........................................................................................................... 7-1 7.1 Introduction ..................................................................................................................... 7-1 7.2 Ordering Replaceable Parts ............................................................................................. 7-1 7.3 Diagrams and Tables ....................................................................................................... 7-2 7.3.1 A5 Structure ....................................................................................................... 7-2 7.3.2 A5 Upper Half .................................................................................................... 7-5 7.3.3 A5 Hardware Box ............................................................................................... 7-6 7.3.4 A5 Work Surface ................................................................................................ 7-7 7.3.5 A5 Patient Circuit Main Body ............................................................................ 7-8 7.3.6 A5 Pre-pak Absorber Canister Assembly ........................................................... 7-9 7.3.7 A5 Valve Assembly .......................................................................................... 7-10 7.3.8 A5 O2 Cell Assembly ........................................................................................7-11 7.3.9 A5 Display Assembly ....................................................................................... 7-12 7.3.10 A5 Display Assembly Main Body .................................................................. 7-13 7.3.11 A5 Vaporizer Manifold ................................................................................... 7-14 7.3.12 A5 Auxiliary Gas Outlet Assembly ................................................................ 7-15 7.3.13 Base Assembly ............................................................................................... 7-16 7.3.14 Cart Assembly ................................................................................................ 7-17 7.3.15 Other............................................................................................................... 7-17 7.3.16 Tubes .............................................................................................................. 7-18 7.3.17 O-ring ............................................................................................................. 7-19 A Test Items ......................................................................................................................... A-1 A.1 Post-Installation Test Items............................................................................................. A-1 A.2 Post-Maintenance Test Items ......................................................................................... A-2 A.3 Post-Repair Test Items.................................................................................................... A-3 5 FOR YOUR NOTES 6 1 Theory of Operation 1.1 Introduction A5 is a compact, easy-to-use anesthetic delivery system. It transports anesthetic gas which is produced and controlled by the configured vaporizer, supports manual ventilation and multiple automatic vent modes, and monitors various parameters such as patient’s airway pressure, inspiratory tidal volume, and expiratory tidal volume in real time. A5 is configured with the following vent modes: Volume Control Ventilation (VCV) with the Pressure Limit Ventilation (PLV) function Pressure Control Ventilation (PCV) with/without volume guarantee (VG) ventilation mode Synchronized Intermittent Mandatory Ventilation (SIMV) with VC mode Synchronized Intermittent Mandatory Ventilation (SIMV) with PC mode Pressure Support (PS) ventilation mode Spontaneous mode in manual mode with APL fully open Manual vent mode using manual bag. Electronic PEEP is available in all mechanical modes. The user can control inspiratory flow (Tslope) in PCV, SIMV and PS modes. The function of automatic fresh gas compensation reduces the effect upon patient ventilation due to change of fresh gas flow by the operator. The traditional bellows system is driven by gas, allowing the operator to view ventilation clearly. A5 fresh gas electronic flowmeter system makes the traditional anesthesia system easier to use and offers the characteristics of traditional anesthesia system as well. Dual-tube electronic flowmeter displays readings more precisely. A knob guard prevents inadvertent movement of the flow control knobs. Supply gas pressure gauges indicate the realtime pressures of pipeline supply gas and cylinder gas. The auxiliary O2 supply is conveniently located in the upper left. The O2 flush button is in the traditional location near the front left corner of the work surface. The safety system of A5 prevents low-O2 mixed gas from entering the patient. The system delivers N2O only when O2 pressure is present. The mechanical safety system ORC ensures 21% of O2 in the mixed gas. The patient breathing circuit heating system of A5 minimizes water condensation and delivers gas back to the patient. The pressure gauge, APL valve, and manual bag in the breathing circuit can be plugged quickly for the convenience of installation and maintenance. A rotary knob on the APL valve provides clear view of setting value. The sodalime absorber canister assembly can be quickly opened and closed via a handle. It can use standard Pre-pak or loose fill sodalime absorbent. The sodalime absorber canister assembly also provides a drain valve. Two flow sensors are configured in the patient breathing circuit to monitor inspiratory/expiratory gas flow and airway pressure. The O2 cell monitors FiO2. The patient breathing circuit can be rotated and fixed as required by the operator. Besides, a test plug is available in the circuit to do leak test. The anesthetic gas scavenging system (AGSS) connector is located at the back of A5. 1-1 When external AC power is connected, A5 power system powers the main system and charges the internal batteries as well. In case of AC power failure, A5 is powered by batteries. The two new batteries can provide power supply for at least 150 minutes. The system power can be turned on and off by the main system switch. The four auxiliary electrical outlets of A5 are located at the back of the machine and are independent of the main switch position. Note: The heater in the patient breathing circuit is disabled when A5 is battery powered. Note: O2 output is cut off if the main switch is in OFF position. 1-2 1.2 Electrical and Pneumatic Connections 1.2.1 Electrical Connection 1-3 No. Description P/N B1 B2 B3 B4 B5 B6 B7 B8 B9, B11 B13 Power board Battery adapter board Mother board Alarm light board Indicator light board Touchscreen control board Display adapter board Main control board 801-0631-00025-00 801-0631-00109-00 801-0631-00108-00 801-0631-00019-00 801-0631-00004-00 801-0631-00018-00 801-0631-00017-00 115-035003-00 Monitoring module 115-051206-00 Top light board Electronic flowmeter board Electronic flow sensor adapter board Auxiliary gas supply flowmeter backlight board Filter Breaker (5.0A) Brazilian-standard auxiliary electrical outlet British-standard auxiliary electrical outlet B-specification 220V auxiliary electrical outlet and connection line Australian-standard auxiliary electrical outlet Indian-standard auxiliary electrical outlet European-standard auxiliary electrical outlet Big South African auxiliary electrical outlet Breaker (3.0A) Lithium battery Speaker Fan Touchpad Expiration valve assembly Display replacement material package Circuit heater Three-way valve assembly Touchscreen Module rack fan Module rack assembly 801-0631-00039-00 051-002834-00 115-044851-00 801-0631-00002-00 801-0631-00029-00 801-0631-00130-00 008-000229-00 008-000027-00 B14 B15 P1 P2 P3-6 P7-10 P11-12 P13 P14 P15 P16 P17 P18 P19 P21 P22 P23 1-4 009-002044-00 M39-000207--008-000060-00 M39-000208--008-000487-00 801-0631-00031-00 115-062081-00 801-0631-00038-00 801-0631-00028-00 801-0631-00052-00 115-036693-00 115-036373-00 115-034450-00 801-0631-00046-00 801-0631-00014-00 024-000407-00 115-016480-00 1.2.2 Pneumatic Connection 1-5 1-6 1-7 Note: The materials listed in the table are not sold separately. Please consult the engineers if any questions. No. From To P/N 25 Total flowmeter Vaporizer manifold assembly M6G-020045--- 26 Back pressure valve assembly Total flowmeter 30 ORC assembly 31 ORC assembly Pneumatic resistor block assembly Pneumatic resistor block assembly Air flow sensor 33 ORC assembly Y2 Y1 34 N2O cylinder yoke assembly N2O large cylinder supply inlet assembly 30 35 Y1 36 Air cylinder yoke assembly 37 O2 flow sensor M6G-020045--M6G-020026--M6G-020026--M6G-020026--M6G-020026--- Y1 Pressure regulator assembly M6G-020045--M6G-020045--M6G-020045--- Y1 Y1 Pressure regulator assembly M6G-020045--M6G-020045--- 38 Y1 Y1 M6G-020046--- 39 O2 pipeline pressure gauge O2 supply inlet assembly M6G-020046--- 40 N2O pipeline pressure gauge N2O supply inlet assembly M6G-020046--- 41 Air pipeline pressure gauge Air supply inlet assembly M6G-020026--- 42 Pressure regulator assembly Y2 M6G-020026--- 43 Y2 System switch M6G-020026--- 44 Y2 O2 flush assembly M6G-020026--- 45 System switch O2 needle valve assembly M6G-020026--- 46 Y2 M6G-020045--- 47 Y1 ORC assembly Pressure regulator assembly M6G-020046--- 48 Pressure regulator assembly ORC assembly M6G-020026--- 49 ORC assembly N2O needle valve assembly M6G-020026--- 50 Pressure regulator assembly Y2 M6G-020026--- 51 Y2 M6G-020026--- 54 Air supply inlet assembly Air needle valve assembly ACGO assembly or CGO assembly ACGO assembly or CGO assembly Y1 or pressure regulator assembly 55 Expiration valve assembly Gas reservoir assembly M6G-020026--- 56 Y2 Y2 M6G-020026--- 57 Y2 M6G-020026--- 58 Electronic flowmeter Auxiliary O2 supply Back pressure valve assembly 52 53 O2 flush assembly Vaporizer manifold assembly 1-8 M6G-020026--M6G-020045--M6G-020045--- M6G-020045--- No. From To P/N 59 O2 supply inlet assembly Y1 M6G-020045--- O2 cylinder yoke assembly O2 large cylinder supply inlet assembly Y1 60 61 Y1 62 M6G-020045--Y1 M6G-020045--- N2O supply inlet assembly Y1 Y1 or pressure regulator assembly 63 Y1 Expiration valve assembly M6G-020045--- 64 Auxiliary Air supply Y2 M6G-020026--- 65 Auxiliary O2 supply Y2 M6G-020026--- 66 Y2 Auxiliary gas outlet M6G-020026--- 67 Y2 Auxiliary Air supply M6G-020026--- 69 Y1 Auxiliary gas outlet M6G-020045--- 73 O2 needle valve assembly Y2 M6G-020026--- 74 Y2 Y2 M6G-020026--- 75 Poppet valve assembly Y2 M6G-020026--- 76 Poppet valve assembly Y2 M6G-020026--- 77 Poppet valve assembly Y2 M6G-020026--- 78 Y2 O2 flow sensor M6G-020026--- 79 Y1 O2 supply inlet assembly M6G-020045--- 81 Y1 Y1 M6G-020045--- 83 N2O large cylinder supply inlet assembly N2O supply pressure gauge M6G-020026--- 85 O2 large cylinder supply inlet assembly O2 supply pressure gauge M6G-020026--- 90 Air needle valve assembly Pneumatic resistor block assembly M6G-020026--- 90 Air needle valve assembly Air flow sensor M6G-020026--- 91 N2O needle valve assembly M6Q-030025--- 91 O2 needle valve assembly N2O flow sensor Pneumatic resistor block assembly M6Q-030024--- 92 Y1 Y1 M6G-020045--- 93 Y1 Expiration valve assembly M6G-020045--- Y1 Φ8 three-way connector M6Q-030025--- Y2 Φ6 three-way connector M6Q-030024--- 1-9 M6G-020045--- 1.2.3 Connections between Pneumatic Circuit, Breathing System and Ventilator Control Board 1-10 Note: The materials listed in the table are sold separately. Please consult the engineers if any questions. No. From To P/N 1 Expiration valve assembly A21-000007--- 2 Expiration valve assembly Monitoring board connector Hardware box tube connector A21-000007--- 3 Expiration valve assembly Monitoring board connector A21-000007--- 4 Three-way valve Monitoring board connector A21-000007--- 5 Three-way valve Y2 A21-000007--- 6 Three-way valve Monitoring board connector A21-000007--- 7 Y2 Y2 Monitoring board connector A21-000007--A21-000007--A21-000007--A21-000007--M6G-020046--- 8 9 10 11 Y2 Hardware box tube connector Y1 Monitoring board connector Monitoring board connector Y1 Circuit adapter block 12 Hardware box tube connector Y1 A21-000007--- 13 Y1 Circuit adapter block M6G-020046--- 14 15 Hardware box tube connector Y1 Y1 Circuit adapter block A21-000007--M6G-020046--- 16 Hardware box tube connector Y1 A21-000007--- 17 Y1 Circuit adapter block M6G-020046--- 18 Expiration valve assembly M6G-020018--- ACGO assembly Gas reservoir assembly Circuit adapter block ACGO assembly ACGO separate outlet ACGO assembly Circuit adapter block Common gas outlet Circuit adapter block 23 Circuit adapter block Expiration valve assembly 115-034137-00 24 Circuit adapter block Hardware box tube connector or three-way connector Gas reservoir assembly 115-027258-00 Monitoring board connector A21-000007--- 72 Three-way valve Monitoring board connector 80 Hardware box tube connector Three-way valve A21-000007--- 82 Hardware box tube connector Three-way valve A21-000007--- 84 Hardware box tube connector Three-way valve A21-000007--- 86 Hardware box tube connector Y2 A21-000007--- 88 Y2 A21-000007--- 94 Y1 Three-way valve Hardware box tube connector 95 ACGO separate outlet Y1 96 Hardware box tube connector Three-way valve M6G-020046--A21-000007--- 98 Hardware box tube connector Three-way valve A21-000007--- 116 Waste gas scavenging connector CPC connector M6G-020005--- 127 AG calculation module Plug 9200-10-10556 128 AG calculation module AG calculation module 9200-10-10556 21 22 1-11 082-000519-00 082-000519-00 A21-000007--- 129 ACGO assembly(with AG) 9200-10-10557 115-015829-00 M6G-020046--- 130 AG calculation module FittingΦ4 141 Gas reservoir assembly 143 ACGO assembly(with AG) AG calculation module Waste gas scavenging connector FittingΦ4 156 Y1 Y1 Y1 Two-way connector M90-100027--- Y2 Three-way connector M90-100030--- 1-12 M6G-020018--082-001198-00 1.3 Pneumatic Circuit 1.3.1 Pneumatic Circuit Diagram 1-13 1.3.2 Parts List Several parts are sorted according to the design scheme of production systems. The whole pneumatic diagram is shown below: 1 O2 pipeline supply connector 32 Sodalime canister 2 3 4 5 6 7 8 9 O2 cylinder supply connector N2O pipeline supply connector N2O cylinder supply connector Air pipeline supply connector Air cylinder supply connector Cylinder pressure regulator(400kPa) Pressure relief valve(758kPa) Drive gas inlet filter 33 34 35 36 37 38 39 40 Bypass O2 sensor Paw gauge Inspiratory flow sensor Patient connector Expiratory flow sensor Water collection cup Expiratory check valve 10 11 12 13 14 Pneumatic block pressure regulator(200kPa) Inspiratory flow control valve Flow sensor Pneumatic block pressure relief valve(110cmH2O) Pop-Off valve Auto/Manual switch 42 Manual bag 43 APL valve High-pressure O2 output 44 45 Bellows 16 17 18 19 20 21 22 23 24 Drive gas pressure switch(140kPa) Proportional PEEP valve Expiration valve Pneumatic block resistor O2 flush button Flow restrictor Gas supply pressure switch(220kPa) Pressure regulator(200kPa) System switch 47 48 49 50 51 52 53 54 55 Mechanical pressure relief valve(1kPa,10cmH2O) Negative check valve(1cmH2O) Gas reservoir 2 Pressure sensor Total flowmeter Auxiliary O2 supply Negative pressure valve Pressure sensor AGSS Check valve2 25 26 27 28 29 30 31 / ORC Electronic flow sensor Dual-vaporizer manifold Check valve3 Pressure relief valve(37.9kPa) Auxiliary Air supply Inspiratory check valve / 56 57 58 59 60 61 62 63 Back pressure valve Poppet valve Needle valve Pneumatic resistor ACGO ACGO separate outlet O2 supply three-way connector N2O pneumatic resistor of ORC 15 PEEP safety valve 41 46 1-14 1.3.3 Key to Symbols Filter Pressure regulator Pressure gauge Check valve Gas supply connector Pressure relief valve Flowmeter Flow control valve Pressure switch Flow restrictor 1.3.4 Descriptions 1.3.4.1 Anesthetic Gas Delivery System The anesthetic gas delivery system is connected to gas source, anesthetic gas delivery device (vaporizer), and breathing system. Its inputs are N2O, O2, and Air from the respective gas supply assembly. Its outputs include mixed gas of these three gases and anesthetic agent (namely, fresh gas), pure O2 (high-pressure O2 output, auxiliary O2, flushing O2), and mixed gas of Air and O2. The pneumatic circuit of anesthetic gas delivery system is as shown below. Drive gas Fresh gas The following picture shows how the gas supply system is connected. O2 is divided into four limbs, which are system switch(24), O2 flush valve(20), auxiliary O2 supply(51), and highpressure O2 exhaust port(44). Air goes to two limbs. One is to needle valve(58) and the other to auxiliary Air supply(30). N2O goes to ORC(25) directly. When system switch(24) is turned on, O2 enters needle valve(58). When O2 flow is greater than 300ml, N2O can enter ORC(25) through needle valve(58). After passing through needle valve(58), pre-set ORC O2 pneumatic resistor(59) and ORC N2O pneumatic resistor(63) control the ratio of O2 and N2O and ensure the minimum O2 concentration. 1-15 To auxiliary Air supply 25. ORC N2O tube ORC N2O pneumatic resistor ORC O2 pneumatic resistor Needle valve Air tube O2 tube To auxiliary O2 supply 24. System switch To high-pressure O2 output port 60. ACGO 20. O2 flush valve 1-16 O2 and N2O enter electronic flow sensor (26) after passing through the ORC O2 pneumatic resistor (59) and ORC N2O pneumatic resistor (63) respectively. Air enters electronic flow sensor(26) after passing through needle valve(58) and the electronic flowmeter monitors and displays the gas flow. 56. Back pressure valve 26. Flow sensor 50. Total flowmeter The converged gas goes from float flowmeter(50) to anesthetic gas delivery device (vaporizer), forming fresh gas after mixing with anesthetic agent. The fresh gas then goes from check valve(28) through CGO(29) assembly to the breathing system. The flushing O2 also enters the breathing system through the CGO(29) assembly. 28. Check valve Outlet of anesthetic gas delivery device 1-17 Inlet of anesthetic gas delivery device Gas Supply 8. Pressure relief valve 22. Pressure switch Inlet of O2 pipeline supply Pressure sampling tube of O2 supply gas pressure gauge The above picture shows the O2 pipeline supply assembly. The pneumatic circuit of the gas source is the starting point of the pneumatic circuit of the anesthesia machine, which functions to introduce the external pipeline gas or cylinder gas into the machine. Since the pressure of external gas is generally very high and the external gas contains foreign substance, there is pressure reducing valves, filters and pressure relief valves available in the supply gas circuit. Also, check valves are available in the supply gas circuit to prevent gas from flowing back into the pipeline or cylinder. The anesthesia machine has three pipeline gas supplies available: O2, N2O, and Air. They enter the “Pipeline gas supply inlet assembly” of the anesthesia machine through pipeline gas supply connectors 1, 3 and 5 respectively. Generally, the pipeline gas pressure is 280~600kPa. The anesthesia machine has three cylinder gas supplies available: O2, Air and N2O. They enter the system through cylinder gas supply connectors 2, 4 and 6 respectively. Their working pressure ranges 6.9~15.5Mpa, 6.9~15.5Mpa and 4.2~6Mpa respectively, which are reduced to around 440kPa through pressure regulator(7). Each connector is clearly marked and designed to prevent misconnection. All connectors have filters and check valves. Color coded pressure gauges show the pipeline and cylinder pressures. Pressure relief valve(8) functions to prevent the supply gas pressure from being too high. It releases excess gas when gas pressure exceeds 758kPa. Each supply gas is outputted after gas pressure is reduced to 200kPa through pressure regulator(24). Pressure switch(23) monitors the O2 supply pressure. When O2 supply pressure is less than around 200kPa, the ventilator gives the alarm of O2 supply failure. 1-18 System Switch Assembly O2 goes into system switch(24) and then to the needle valve. System switch is configured with a socket to manage the power-on status of the system. When the system switch is turned on, O2 enter the needle valve and the system is powered on simultaneously. The anesthetic ventilator starts to monitor the system status. When the system switch is turned off, O2 cannot enter the needle valve and the system is powered off. Flow Adjustment and Control Assembly 58. Flow adjustment needle valve Needle valve(58) controls gas flow. ORC(25) controls the ratio of O2 and N2O to ensure that O2 concentration is not less than 25%. Turning needle valve knob counterclockwise increases the flow and clockwise decreases the flow. 1-19 Flow Display Assembly The flow display assembly is made up of electronic flowmeter(26) and total flowmeter(50). Its outputs are gas from the flow adjustment and control assembly and mixed gas going through the anesthetic gas delivery device (vaporizer). Electronicflowmeter(26) measures and displays the flow of each gas. Total flowmeter(50) displays the total gas flow. 26. Flow sensor 50. Total flowmeter O2 Flush Button Assembly The above picture shows the O2 flush button assembly. When O2 flush valve(20) is depressed, O2 flow rushes to the pneumatic circuit, which is cut off when this valve is released. Its input is O2 of 0.2MPa, which comes from O2 supply gas and is regulated. Its output is O2 which flows through the ACGO assembly and enters the breathing system. The O2 flush button assembly is not controlled by the system switch. O2 flushing can be performed as long as O2 supply is normal. The O2 flush button assembly has sliding valve structure inside, which ensures automatic reset via the spring each time the valve is depressed and released. 1-20 Vaporizer Manifold The above picture shows the vaporizer manifold assembly. The anesthetic gas delivery device (vaporizer) is connected to the anesthetic gas delivery system. The mixed gas of N2O, O2 and Air go into the device; and the fresh gas containing these three gases and anesthetic agent is finally outputted to the ACGO assembly. Dual-vaporizer manifold (27) is integrated with check valve(28) to prevent flushing O2 and fresh gas from flowing back to the vaporizer, so as to affect the concentration of outputted anesthetic gas. When using dual-vaporizer manifold, Selectatec mounting with interlocking function can prevent the user from turning on two vaporizers simultaneously. Triple-vaporizer manifold is provided for optional configuration. Gas inlet Gas outlet Check valve Three canisters position The above picture shows the triple-vaporizer manifold assembly. The anesthetic gas delivery device (vaporizer) is connected to the anesthetic gas delivery system. The mixed gas of N2O, O2 and Air go into the device; and the fresh gas containing these three gases and anesthetic agent is finally outputted to the ACGO assembly. 1-21 Triple-vaporizer manifold is integrated with the check valve to prevent flushing O2 and fresh gas from flowing back to the vaporizer, so as to affect the concentration of outputted anesthetic gas. When using triple-vaporizer manifold, Selectatec mounting with interlocking function can prevent the user from turning on two vaporizers simultaneously. ACGO Assembly The above picture shows the ACGO assembly. This assembly contains flow restrictor(21), pressure relief valve(29), and pressure relief valve(13). The delivered O2 enters ACGO after mixing with the fresh gas. Pressure relief valve(29) in the front restricts the pressure of delivered O2 and prevents the pressure of fresh gas from exceeding 37.9kPa. Pressure relief valve(13) prevents the pressure of gas coming out of the ACGO from exceeding 12.5kPa (ACGO has two types of configuration: separate outlet and breathing system outlet). CGO Assembly The following picture shows the pressure relief valve (29) and flow restrictor (21) of CGO module. 1-22 Auxiliary O2/Air Supply Assembly Auxiliary O2 supply assembly(51) and auxiliary Air supply assembly(30) control flows by the needle valve. Gas is displayed by a glass tube flowmeter and delivered to the patient. The gas flow can be adjusted within the range of 0 to 15L/min. Turning the flow control counterclockwise increases the flow and clockwise decreases the flow. High- pressure O2 Output Assembly The high-pressure O2 output assembly is directly connected to the O2 supply inlet assembly. It can be used for the external ventilation device (jet ventilation device) without going through pressure regulator. When no external device is connected, the high-pressure O2 output is closed. The maximum output flow is 90L/min. Auxiliary gas supply output High-pressure O2 output 1-23 1.3.4.2 Anesthetic Ventilator Pneumatically-controlled Module The anesthetic ventilator pneumatically-controlled module provides drive force for the patient’s breathing process. Its input is O2 (or Air) from the gas supply inlet assembly. Its outputs include drive gas to the breathing system and PEEP exhaust gas to the built-in gas reservoir and muffling part. The ventilator can control the drive gas flow to realize various vent modes and prevent too high pressure inside the pneumatic circuit from injuring the patient. The following picture shows the gas flow directions and components of the anesthetic ventilator pneumatically-controlled module. Pneumaticallycontrolled module Output gas discharged through PEEP outlet Input tube of the pneumaticallycontrolled module Exhaust tube Output drive gas The pneumatic circuit of the anesthetic ventilator pneumatically-controlled module is as shown below: 45 52 Drive gas 9 10 11 12 Drive Gas 13 14 Flow Breathing Breathing system System 46 18 AGSS 15 17 16 1-24 P 49 19 Filter(9) filters the drive gas. Pressure regulator(10) adjusts the pressure in the pneumatic circuit (about 0.2MPa). Proportional solenoid valve(11) controls gas entering flow. Differential pressure flow sensor(12) monitors gas flow in the drive gas circuit. Mechanical overpressure valve(13) ensures that the pressure in the drive gas circuit does not exceed safe pressure. It releases excess gas when gas pressure exceeds 11kPa (110cmH2O). “18” is expiration valve. During expiration, gas inside the bellows is discharged from this valve. PEEP function is implemented through expiration valve control limb. “15” is low-flow proportional solenoid valve. When it opens, gas is discharged from pneumatic resistor(19), forming relatively stable pressure in the pneumatic circuit between “15” and “19”. Such pressure is exerted onto the membrane of expiration valve(18) to form PEEP. To prevent too high pressure in the pneumatic circuit from injuring the patient and damaging the equipment, safety valve(15), which is a solenoid on-off valve, is necessary to be placed before the expiration valve control limb. “16” is a pressure switch. When drive gas pressure is less than 125kPa, an alarm is triggered. “49” is a pressure sensor which monitors the expiration valve closing pressure. Pressure relief valve(46) ensures that the tube pressure after the expiration valve is less than 10cmH2O during expiration phase. 1.3.4.3 Breathing System The breathing system provides a closed loop for the anesthetic gas. CO2 in the patinet’s expired gas can be absorbed in the inspiration phase, so as to maintain the gas temperature and humidity conditions. During inspiration, the drive gas compresses the bellows to force the gas to enter the patient’s lungs. During expiration, the patient’s pulmonary gas enters the bellows again through the chest muscular contractions. Sodalime absorber canister(32) in the circuit absorbs the patient’s expired CO2. The pneumatic circuit of the breathing system is as shown below. 43 45 41 AGSS NO NC 42 Drive Drive gas Gas 14 38 gas Flowflow 40 33 32 39 33 37 53 P 31 Flow gas 36 O2 34 1-25 35 Manual ventilation and mechanical ventilation are selected through the Auto/Manual switch. When manual ventilation is selected, the doctor presses the manual bag(42), replacing mechanical ventilation, to provide gas for the breathing system. APL valve(43) adjusts the pressure in the pneumatic circuit in manual ventilation mode. When mechanical ventilation is selected, the ventilator starts to work, driving the drive gas to compress the bellows inside the bellows housing(45) and providing gas for the breathing system based on the selected vent mode. The breathing system is connected to the anesthesia machine main unit through the circuit adapter. The breathing system is highly integrated. Its tubes are all built-in except the tubes connected to the patient and to the O2 cell cable, as shown below. 41. Auto/Manual switch 45. Bellows assembly 43. APL valve Bag arm 40. Expiration valve 35. Paw gauge 1-26 31. Inspiration Patient connection (built-in inspiratory and expiratory flow sensors) 33. Built-in bypass valve 32. CO2 absorber canister Operating handle In mechanical ventilation mode, during inspiration, the gas flows through Auto/Manual switch(41), BYPASS valve(33) or sodalime canister(32), inspiration valve(31), O2 sensor(34), Paw gauge(35), and inspiratory flow sensor(36) to the patient. During expiration, the gas flows through expiratory flow sensor(38), expiration valve(40), and Auto/Manual switch(41) to the bellows. The airway pressure is monitored by pressure sensor(53). The water produced by condensation is accumulated in the water collection cup at the bottom of the breathing system. The breathing system is conveniently disassembled and is autoclavable at 134℃. 1.3.4.4 Anesthetic Gas Scavenging System The anesthetic gas scavenging system (AGSS) is composed of AGSS transfer system, AGSS receiving system, and AGSS disposal system. Waste gas goes from the exhaust port of the anesthesia machine, through the AGSS transfer system and the AGSS receiving system, and to the hospital’s waste gas disposal system (AGSS disposal system), as shown below. Waste gas AGSS transfer system AGSS receiving system AGSS disposal system (hospital’s disposal system) 1-27 The following figure shows the operational theory of the AGSS. The throttling holes reduce the effect of negative pressure at the AGSS outlet upon the flow at the entrance. The float helps the user to determine if the disposal system meets the minimum extract rate requirement. The filter filters foreign substance to prevent the disposal system from being occluded. The gas storage device is connected to the external air through pressure compensation openings. When there is positive or negative pressure inside the gas storage device, gas is inputted or outputted to ensure pressure balance inside the system. The both ends of the AGSS transfer tube are 30mm conical connectors. The inlet is female 30mm conical connector and the outlet is male 30mm conical connector. The transfer tube is connected to the AGSS main body through 30mm connector. The following picture shows the structure of AGSS and how the AGSS transfer system, receiving system, and disposal system are connected. Receiving system Transfer system The AGSS three-way connector assembly used for exhausting the open circuit waste gas is also provided for optional configuration. 1-28 Transfer system Negative pressure supply provided by other hospital Blocked at ordinary time. But Connect to the outlet of open circuit waste gas once the open circuit is initiated. Receiver system 1.4 Components of the Anesthesia System 1.4.1 Auxiliary Electrical Outlets A5 anesthesia system has 4 auxiliary electrical outlets (3 for big South African standard). Each outlet has one 250V3A breaker. Besides, the total breaker restricts the total current of the 4 outlets to 5A. Aux. electrical outlet1 Aux.Auxiliary electrical Auxiliary Outlet 1 Outlet 2 l 2 Breaker 1 Breaker Breaker 2 Breaker Aux. electrical Auxiliary Outlet 3 l Breaker 3 Breaker 1-29 Aux. electrical outlet4 Auxiliary Outlet 4 Breaker 4 Breaker 1.4.2 Lighting Board 1.4.2.1 Top Light Board Top light board (Top view) Top light board (Bottom view) Auxiliary gas supply flowmeter backlight board interface, J1 Pin Name Function 1 12V 2 3 4 GND GND LIGHT_OUT Auxiliary gas supply flowmeter backlight board 12V power supply Grounding Grounding Auxiliary gas supply flowmeter backlight control Top light switch interface, J2 Pin Name Function 1 2 3 4 5 6 HIGH_BRIGHTNESS HIGH_BRIGHTNESS OFF OFF LOW_BRIGHTNESS LOW_BRIGHTNESS High brightness High brightness Off Off Low brightness Low brightness Power supply interface, J3 Pin Name Function 1 2 3 12V GND 12V_AUX Top light board 12V power supply Grounding Auxiliary gas supply flowmeter backlight board 12V power supply 1-30 1.4.2.2 Auxiliary Gas Supply Flowmeter Backlight Board Auxiliary gas supply flowmeter backlight board (Top view) Auxiliary gas supply flowmeter backlight board (Bottom view) Auxiliary gas supply flowmeter backlight board interface, J1 Pin Name Function 1 12V 2 3 4 GND GND LIGHT_IN Auxiliary gas supply flowmeter backlight board 12V power supply Grounding Grounding Auxiliary gas supply flowmeter backlight control signal 1-31 1.5 Breathing System 1.5.1 Introduction A5 breathing system supports three types of operating modes: mechanical ventilation, manual ventilation, and standby, allowing the operator to apply proper ventilation strategy based on the patient’s needs. The types of flow path through the breathing system vary with the operating mode or status. 1.5.2 Mechanical Mode, Inspiration When the Auto/Manual switch is set to Auto position, the system closes the manual ventilation path. The drive gas depresses the bellows inside the bellows housing. Gas flows out of the bellows, through the CO2 canister and inspiratory check valve, and to the patient. During inspiration, the fresh gas flows into the inspiratory limb, upstream of the inspiratory check valve. In volume mode, tidal volume is compensated for variation in gas flow, to ensure that the tidal volume delivered to the patient complies with the setting value. In pressure mode, the inspiratory pressure is adjusted by adjusting the gas flow and airway pressure, to ensure that the airway pressure is held at the set inspiratory pressure level during the patient’s inspiration. APL Valve Drive Gas Pop-Off valve NO NC Breathing Bag Auto/Manu Switch Bellows Exhaust valve AGSS Exp. Valve Exp. Flow Sensor Water collecti on cup Water trap cup Absorber Canister Insp. Valve Insp. Flow Sensor Fresh Gas O2 Cell 1-32 1.5.3 Mechanical Mode, Expiration When the Auto/Manual switch is set to Auto position, the system closes the manual ventilation path. The drive gas flow stops and the expiration valve opens. The patient expires gas. The expired gas flows through the expiratory check valve and into the bellows. Residual drive gas flows out of the bellows housing through the exhaust valve to the scavenging system (AGSS). If PEEP setting is selected, the pressure at the exhaust valve port is held at the PEEP level. During expiration, the fresh gas flows backwards through the CO2 canister into the expiratory limb, downstream of the expiratory check valve. APL Valve Drive Gas Pop-Off valve NO NC Breathing Bag Auto/Manu Switch Bellows Exhaust valve AGSS Exp. Valve Exp. Flow Sensor Water collecti on cup Water trap cup Absorber Canister Insp. Valve Insp. Flow Sensor Fresh Gas O2 Cell 1-33 1.5.4 Manual Mode, Inspiration When the Auto/Manual switch is set to Manual position, the system closes the mechanical ventilation path. When the manual bag is depressed, the gas flows out of the manual bag through the CO2 canister into the breathing circuit, and through inspiratory check valve to the patient’s lungs. During inspiration, the fresh gas flows from the machine into the inspiratory limb, upstream of the inspiratory check valve. If the airway pressure exceeds the setting value of the APL Valve, the excess gas will pass through the APL valve to the scavenging system (AGSS). APL Valve Auto/Manu Switch Bellows Breathing Bag Pop-Off valve NO NC Exhaust valve AGSS Exp. Valve Exp. Flow Sensor Water collecti on cup Water trap cup Absorber Canister Fresh Gas In Insp.Ins Valve s p. O2 Cell 1-34 Insp. Flow Sensor 1.5.5 Manual Mode, Expiration When the Auto/Manual switch is set to Manual position, the system closes the mechanical ventilation path. The patient’s expired gas enters the manual bag after passing through the expiratory check valve. During expiration, the fresh gas enters the breathing system. The residual fresh gas passes through the APL valve to the scavenging system (AGSS). . APL Valve Auto/Manu Switch Bellows Exhaust valve Breathing Bag Pop-Off valve NO NC AGSS Exp. Valve Exp. Flow Sensor Water trap cup Water collecti on cup Absorber Canister Insp. Valve Insp. Flow Sensor Fresh Gas O2 Cell 1.5.6 Pneumatic PEEP Exhaust valve opening pressure decides PEEP. Therefore, if PEEP is selected, during Auto ventilation, the ventilator adjusts exhaust valve opening pressure based on the set PEEP value. 1.5.7 Ventilator in Standby If the anesthesia system is in standby mode, monitoring function is disabled and mechanical ventilation is unavailable. When the system is standby, the patient can not be ventilated. 1.5.8 Components of the Breathing System 1.5.8.1 Ventilation Bellows Assembly The ventilator’s drive system is drive gas. Drive gas inflates the space outside the bellows so as to compress the bellows. The bellows press the gas into the patient’s breathing circuit. When the patient expires, the bellows are re-inflated with gas. 1.5.8.2 Manual Bag In manual mode, manual bag helps the user to ventilate the patient manually. 1-35 1.5.8.3 CO2 Canister The sodalime inside the CO2 canister absorbs the CO2 in the expired gas. The CO2 canister can be equipped with standard sized Pre-paks or loose-fill CO2 absorbent. 1.5.8.4 Inspiration Valve and Expiration Valve To ensure correct gas flow direction to and from the patient’s lungs, check valves are integrated into both the inspiratory and expiratory limb of the breathing system. 1.5.8.5 APL (Airway Pressure Limit) Valve In manual mode, APL valve, as gas pressure relief valve, limits the airway pressure in the breathing system. 1.6 User Interface 1.6.1 Display 1.6.1.1 Display Adapter Board J1 J5 J2 20 2 19 1 Display adapter board (Top view) Display adapter board (Bottom view) 1-36 J4 Inverter interface, J1 Pin Name Function 1 2 3 4 5 12V 12V GND GND LCD_EN Backlight control board 12V power supply Backlight control board 12V power supply Grounding Grounding LCD backlight switch 6 LCD_BR LCD backlight brightness control Alarm light board interface, J2 Pin Name Function 1 2 3 4 5 6 Alarm light board 12V power supply Grounding Main control board I2C data signal Main control board I2C clock signal 3.3V power supply Grounding 12V GND SDA_CPU SCL_CPU 3_3V GND Touchscreen control board interface, J4 Pin Name Function 1 2 3 4 Pin 5V RXD_TOUCH_PANEL TXD_TOUCH_PANEL GND 5V power supply Touchscreen control board serial port receiving signal Touchscreen control board serial port transmitting signal Grounding Display control signal interface, J5 Name Function 1 2 3 4 LCD_EN LCD_BR RXD_TOUCH_PANEL TXD_TOUCH_PANEL 5 6 7 8 9 10 11 12 13 14 15 16 17 18 GND LED_AC LED_BAT SDA_CPU SCL_CPU RSVD RSVD RSVD RSVD RSVD RSVD 12V 12V GND LCD backlight switch LCD backlight brightness control Touchscreen control board serial port receiving signal Touchscreen control board serial port transmitting signal Grounding AC indicator light drive signal Battery indicator light drive signal Main control board I2C signal Main control board I2C signal Reserved Reserved Reserved Reserved Reserved Reserved 12V power supply 12V power supply Grounding 1-37 Pin Name Function 19 20 GND GND Grounding Grounding 1.6.1.2 Alarm Light Board J1 Alarm light board (Top view) Alarm light board (Bottom view) Alarm light board interface, J1 Pin Name Function 1 2 3 4 5 12V power supply Grounding Main control board I2C data signal Main control board I2C clock signal 3.3V power supply 12V GND MAIN_BRD_SDA MAIN_BRD_SCL 3_3V 1-38 1.6.1.3 Display A5 anesthesia machine has a 15-inch, 24-bit display color, 1024x768, LVDS display with touchscreen of the same size. The backlight brightness of the display can be adjusted through the inverter. 1.6.1.4 Electronic Flowmeter Board Electronic flowmeter board (Top view) Electronic flowmeter board (Bottom view) 1-39 Electronic flowmeter board interface, J2 Pin Name Function 1 2 3 4 5 6 DVCC DVCC GND GND DVDD DVDD Electronic flowmeter board 5 V power supply Electronic flowmeter board 5 V power supply Grounding Grounding Electronic flowmeter board 3.3 V power supply Electronic flowmeter board 3.3 V power supply 7 8 9 10 11 12 13 14 15 16 17 18 19 20 SCL_AIR SCL_O2 GND GND SDA_AIR SDA_O2 / / DVCC SCL_N20 GND GND DVDD SDA_N20 II2C clock signal II2C clock signal Grounding Grounding II2C data signal II2C data signal Electronic flowmeter board 5 V power supply II2C clock signal Grounding Grounding Electronic flowmeter board 3.3 V power supply II2C data signal Electronic flowmeter board interface, J3 Pin Name Function 1 TXD_FLOW_BRD 2 3 4 RXD_FLOW_BRD GND 12V Electronic flowmeter board serial port transmitting signal Electronic flowmeter board serial port receiving signal Grounding Electronic flowmeter board 12V power supply (range: 10.8–13.2 V) 1.6.1.5 O2/N2O/Air Flow Sensor Transfer Board Air J1 O2/N2O/Air flow sensor transfer board (top view) Flow sensor transfer board interface, J1 1-40 Pin 1 2 3 4 5 6 Name VCC GND VDD SCL GND SDA Function 5 V power supply Grounding 3.3 V power supply Flow sensor IIC clock signal Grounding Flow sensor IIC data signal Flow sensor transfer board interface, J2 Pin Name Function 1 VCC 5 V power supply 2 NC Ais suspension 3 VDD 3.3 V power supply 4 GND Grounding 5 SCL Flow sensor IIC clock signal 6 GND Grounding 7 SDA Flow sensor IIC data signal 8 GND Grounding 1.6.1.6 Indicator Light Board J1 Indicator light board (Top view) 1-41 Indicator light board (Bottom view) Indicator light board interface, J1 Pin Name Function 1 2 3 AC indicator light drive signal Battery indicator light drive signal Grounding LED_AC LED_BAT GND 1.6.2 Main Control Board J8 J4 J9 Main control board (Top view) 1-42 Main control board (Bottom view) Network interface, J9 Pin Name Function 1 2 3 4 5 6 7 8 TX+ TXRX+ CT1 CT1 RXCT2 CT2 Positive end of transmitting signal Negative end of transmitting signal Positive end of receiving signal Not specified Not specified Negative end of receiving signal Not specified Not specified USB interface, J8 Pin Name Function 1 2 3 4 5 6 7 8 USB power supply USB data signal – (negative) USBdata signal + (positive) Grounding USB power supply USB data signal – (negative) USB data signal + (positive) Grounding VCC DM0 DP0 GND VCC DM1 DP1 GND RS-232 interface, J4 Pin Name Function 1 Not connected NC 1-43 2 3 4 5 6 7 8 9 RXD TXD NC GND NC NC NC NC RS-232 receiving signal RS-232 transmitting signal Not connected Grounding Not connected Not connected Not connected Not connected 1.7 Ventilation Control and Drive 1.7.1 Motherboard J7 J12 J13 J10 J1 J8 J9 1 5 J11 4 8 2 1 1 2 65 33 1 49 50 J5 Motherboard (Top view) 1-44 34 18 1 50 33 17 37 19 20 1 25 13 14 1 J4 J3 J2 Motherboard (Bottom view) Monitoring module interface, J1 Pin Name Function 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 Safety valve drive signal Inspiration valve drive signal Inspiration valve drive signal PEEP valve drive signal PEEP valve drive signal Three-way valve 1 Three-way valve 2 Three-way valve 3 Three-way valve 4 Pneumatic block pressure switch signal Not connected Not connected Gas supply inlet O2 pressure switch signal Auto/Manual switch signal Grounding CO2 canister in-position signal O2 concentration signal O2 concentration signal SAFE_VALVE 7VIN INSP_VALVE 7VIN PEEP_VALVE SOLENOID_VALVE1 SOLENOID_VALVE2 SOLENOID_VALVE3 SOLENOID_VALVE4 PNEUM_PRES_SW NC NC O2_PRE_SW MANU_AUTO_SW GND CO2_BYPASS_SW O2+ O2- 1-45 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 TXD_AUX_BRD RXD_AUX_BRD GND RSVD RSVD RSVD VF GND TXD_MON_BRD RXD_MON_BRD 12V GND 5V 5V TXD_CALIBRATE RXD_CALIBRATE GND 12V Auxiliary monitoring board serial port transmitting signal Auxiliary monitoring board serial port receiving signal Grounding Reserved Reserved Reserved Expiratory flow sensor differential pressure output signal Grounding VCM serial port transmitting signal VCM serial port receiving signal 12V power supply Grounding VCM control safety valve signal 5V power supply Calibration serial port transmitting signal Calibration serial port receiving signal Grounding 12V power supply Pneumatic assembly interface, J2 Pin Name Function 1 2 3 7Vout SAFE_VALVE 7Vout Safety valve power supply Safety valve drive signal Inspiration valve power supply 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 INSP_VALVE 7Vout PEEP_VALVE 12V SOLENOID_VALVE1 12V SOLENOID_VALVE2 12V SOLENOID_VALVE3 12V SOLENOID_VALVE4 RSVD RSVD RSVD VF 12V GND NC NC NC GND O2_PRE_SW Inspiration valve drive signal PEEP valve power supply PEEP valve drive signal Three-way valve power supply Three-way valve drive signal 1 Three-way valve power supply Three-way valve drive signal 2 Three-way valve power supply Three-way valve drive signal 3 Three-way valve power supply Three-way valve drive signal 4 Reserved Reserved Reserved Expiratory flow sensor differential pressure output signal Sensor adapter board 12V power supply Grounding Not connected Not connected Not connected Grounding Gas supply inlet O2 pressure switch signal 1-46 Pin Name Function 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 GND PNEUM_PRES_SW GND MANU_AUTO_SW GND CO2_BYPASS_SW GND NC GND O2+ O2LOOP_SW GND NTC_R11 NTC_R12 NTC_R21 NTC_R22 HEA_PWR_15V HEA_PWR_15V GND GND 12V GND RXD_FLOW_BRD TXD_FLOW_BRD Grounding Pneumatic block pressure switch signal Grounding Auto/Manual switch signal Grounding CO2 canister in-position signal Grounding Not connected Grounding O2 concentration signal Grounding Circuit switch Grounding Thermistor signal 1 Thermistor signal 2 Thermistor signal 1 Thermistor signal 2 Heater drive voltage signal Heater drive voltage signal Grounding Grounding Electronic flowmeter board power supply Grounding Electronic flowmeter board serial port receiving signal Electronic flowmeter board serial port transmitting signal Display interface, J3 Pin Name Function 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 LCD backlight switch signal LCD brightness adjustment signal Touchscreen control board serial port receiving signal Touchscreen control board serial port transmitting signal Not connected AC indicator light drive signal Battery indicator light drive signal Main control board I2C signal Main control board I2C signal Not connected Not connected Not connected Not connected Not connected Not connected 12V power supply Grounding LCD_EN LCD_BR RXD_TOUCH_PANEL TXD_TOUCH_PANEL NC LED_AC LED_BAT MAIN_BRD_SDA MAIN_BRD_SCL NC NC NC NC NC NC 12V GND 1-47 Pin Name Function 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 GND VCC_LCD VCC_LCD GND GND LVDS_DATA0+ LVDS_DATA0GND LVDS_DATA1+ LVDS_DATA1GND LVDS_DATA2+ LVDS_DATA2GND LVDS_DATA3+ LVDS_DATA3GND LVDS_CLK+ LVDS_CLKGND Grounding LCD power supply LCD power supply Grounding Grounding LVDS data signal LVDS data signal Grounding LVDS data signal LVDS data signal Grounding LVDS data signal LVDS data signal Grounding LVDS data signal LVDS data signal Grounding LVDS clock signal LVDS clock signal Grounding Patient monitor module rack interface, J4 Pin Name Function 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 LED_AC 3V3 12V 12V GND GND NC NC 12V RSVD RSVD GND LED_BAT PCON-_NON GND PCON+ PCONTOUCHPAD_5V TOUCHPAD_GND TOUCHPAD_USB+ TOUCHPAD_USB12V AC indicator light drive signal 3.3V power supply 12V power supply 12V power supply Grounding Grounding Not connected Not connected 12V power supply Reserved Reserved Grounding Battery indicator light drive signal Power on/off inverse phase signal Grounding Power on/off signal + Power on/off signal 5V power supply Grounding TOUCHPAD_USB differential signal + TOUCHPAD_USB differential signal 12V power supply 1-48 23 24 25 GND NC NC Grounding Not connected Not connected Calibration interface, J5 Pin Name Function 1 2 3 4 5 6 7 8 9 Not connected Not connected Not connected Not connected Not connected 12V power supply Calibration serial port receiving signal Calibration serial port transmitting signal Grounding NC NC NC NC NC 12V RXD_CALIBRATE TXD_CALIBRATE GND Fan interface, J7 Pin Name Function 1 2 3 4 12V power supply Reserved Fan status signal Grounding 12V3 RSVD FAN1_STATE GND Speaker interface, J8 Pin Name Function 1 2 Speaker positive end Speaker negative end Speak+ Speak- Battery adapter board interface, J9 Pin Name Function 1 2 3 4 5 6 7 8 BAT1+ NTC1 BC1 GND BAT2+ NTC2 BC2 GND Battery voltage Battery internal thermistor Battery in-position signal Grounding Battery voltage Battery internal thermistor Battery in-position signal Grounding Power board interface, J10 Pin Name Function 1 2 3 4 5 6 Buzzer drive signal (drive the buzzer directly) Power board serial port receiving signal Circuit switch (indicates if the circuit is in position) Power board serial port transmitting signal Not connected Grounding PLAM RXD_PWR_BRD LOOP_SW TXD_PWR_BRD NC GND 1-49 Pin Name Function 7 8 9 10 11 12 LED-BAT LCD_EN LED-AC LCD_BR PCON3.3VBF 13 14 15 BAT2+ GND BC2F 16 17 18 19 NTC2 BAT1+ GND BC1F 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 NTC1 GND GND HEA_PWR_15V HEA_PWR_15V NTC_R12 NTC_R22 NTC_R11 NTC_R21 GND GND 3V3 3V3 5V GND GND GND NC GND NC NC GND GND 12V GND 12V 12V NC Battery status indicator light drive output LCD backlight switch signal AC status indicator light drive output Backlight brightness adjustment signal Power on/off signal 3.3V power supply, only used for power on/off the machine No.2 battery input (connect to the battery positive end) Grounding No.2 battery in-position signal. Low level indicates that battery is in position and high level indicates that battery is not in position. No.2 battery internal thermistor signal No.1 battery input (connect to the battery positive end) Grounding No.1 battery in-position signal. Low level indicates that battery is in position and high level indicates that battery is not in position. No.1 battery internal thermistor signal Grounding Grounding Heater drive voltage output Heater drive voltage output Thermistor (for controlling heater) pin 1 Thermistor (for controlling heater) pin 2 Thermistor (for controlling heater) pin 1 Thermistor (for controlling heater) pin 2 Grounding Grounding 3.3V voltage output 3.3V voltage output 5V voltage output Grounding Grounding Grounding Not connected Grounding Not connected Not connected Grounding Grounding 12V power supply output Grounding 12V power supply output 12V power supply output Not connected 1-50 Pin Name Function 48 49 50 15V2 NC 15V2 15.2V voltage output Not connected 15.2V voltage output Main control board interface, J11 Pin Name Function A1 LCD_VDD LCD power supply A2 GND Grounding A3 NC Not connected A4 NC Not connected A5 RSVD Reserved A6 RSVD Reserved A7 GND Grounding A8 RSVD Reserved A9 RSVD Reserved A10 Grounding A11 GND UIVCC_USB A12 TOUCHPAD_USB+ TOUCHPAD_USB data signal + A13 TOUCHPAD_USB- TOUCHPAD_USB data signal - A14 GND Grounding A15 SCL CPU board I2C clock A16 SDA CPU board I2C data A17 GND Grounding A18 NC Not connected A19 A20 NC NC NC Not connected Not connected Fan status signal A23 FAN1_STATE NC Not connected A24 GND Grounding A25 3V3 Main control board power supply A26 3V3 Main control board power supply A27 GND Grounding A28 GND Grounding A29 5V Main control board power supply A30 5V Main control board power supply A31 GND Grounding A32 GND Grounding B1 NC Not connected B2 NC Not connected B3 NC Not connected B4 SPK_OUT+ Speaker drive signal + A21 A22 USB power supply Not connected 1-51 Pin Name Function B5 SPK_OUT- Speaker drive signal - B6 GND Grounding B7 NC Not connected B8 NC Not connected B9 NC Not connected B10 NC Not connected B11 NC Not connected B12 GND Grounding B13 TXD_PWR_BRD Power board serial port transmitting signal B14 RXD_PWR_BRD Power board serial port receiving signal B15 GND Grounding B16 TXD_FLOW_BRD Electronic flowmeter board serial port transmitting signal B17 RXD_FLOW_BRD Electronic flowmeter board serial port receiving signal B18 GND Grounding B19 TXD_TOUCH_PANEL Touchscreen control board serial port transmitting signal B20 RXD_TOUCH_PANEL Touchscreen control board serial port receiving signal B21 RSVD Reserved B22 RSVD Reserved B23 GND Grounding B24 NC Not connected B25 NC Not connected B26 NC Not connected B27 GND Grounding B28 NC Not connected B29 NC Not connected B30 NC Not connected B31 NC Not connected B32 Not connected C1 NC GND C2 LVDS-TO0+ LVDS data signal C3 LVDS data signal C4 LVDS-TO0GND C5 LVDS-TO1+ LVDS data signal C6 LVDS data signal C7 LVDS-TO1GND C8 LVDS-TO2+ LVDS data signal C9 LVDS data signal C10 LVDS-TO2GND C11 LVDS-TO3+ LVDS data signal C12 LVDS-TO3GND LVDS data signal C13 Grounding Grounding Grounding Grounding Grounding 1-52 Pin Name Function C14 LVDS-TOC+ LVDS clock signal C15 LVDS-TOC- LVDS clock signal C16 GND Grounding C17 TXD_MON_BRD Monitoring board serial port transmitting signal C18 RXD_MON_BRD Monitoring board serial port receiving signal C19 GND Grounding C20 TXD_AUX_BRD Auxiliary monitoring board serial port transmitting signal C21 RXD_AUX_BRD Auxiliary monitoring board serial port receiving signal C22 GND Grounding C23 NC Not connected C24 NC Not connected C25 GND Grounding C26 TP_PWR_CTRL Touchpad power control signal C27 Reserved C28 RSVD RSVD Reserved C29 RSVD Reserved C30 RSVD Reserved C31 NC NC Not connected C32 Not connected Debugging power on/off interface, J12 Pin Name Function 1 2 PCON+ PCON- Power on/off signal Power on/off signal Test point definition Label Name Function Range (unit:V) T1 BAT1 Battery voltage 1 Fully charged 12.6±5% T2 T3 BAT2 LED_BAT Battery voltage 2 Battery indicator light drive signal T4 TXD_7024 Auxiliary monitoring board serial port transmitting signal T5 TXD_TOUCH Touchscreen control board serial port transmitting signal T6 RXD_7024 Auxiliary monitoring board serial port receiving signal T7 RXD_TOUCH Touchscreen control board serial port receiving signal T8 TXD_33209 Auxiliary monitoring board serial port transmitting signal T9 TXD_FLOW Electronic flowmeter board serial port transmitting signal Fully charged 12.6±5% Battery in-position 2.5~3.5; Battery not in-position 0~0.4 High level 2.4~5; Low level 0~0.4 High level 2.4~5; Low level 0~0.4 High level 2.4~5; Low level 0~0.4 High level 2.4~5; Low level 0~0.4 High level 2.4~5; Low level 0~0.4 High level 2.4~5; Low level 0~0.4 1-53 Label Name Function Range (unit:V) T10 RXD_33209 VPM serial port receiving signal T11 RXD_FLOW Fresh flow sensor board serial port receiving signal T12 TXD_POWER Power board serial port transmitting signal T13 RXD_POWER Power board serial port receiving signal T14 T15 T16 T17 T18 T19 5V 3.3V 15V2 12V P15V LED_AC 5V power supply 3.3V power supply 15.2V power supply 12V power supply Heater power supply AC indicator light drive signal T20 T21 T22 T23 T24 12V1 12V2 12V3 7V LCD_EN 12V power supply 12V power supply 12V power supply 7V power supply LCD backlight enabled signal T25 LCD_BR LCD backlight brightness adjustment signal High level 2.4~5; Low level 0~0.4 High level 2.4~5; Low level 0~0.4 High level 2.4~5; Low level 0~0.4 High level 2.4~5; Low level 0~0.4 4.75~5.25 3.135~3.465 14.44~15.96 11.4~12.6 0~15 With AC: 5~3.5; Without AC: 0~0.4 11.4~12.6 11.4~12.6 11.4~12.6 6.65~7.35 High level 3.145~3.465; Low level 0~0.3 Brightest 0~1.5; least bright 4.75~5.25 1.7.2 Monitoring Module The monitoring module performs pressure and flow detection of anesthetic ventilator and anesthetic breathing system, valve control, status monitoring collection, O2 concentration reading, accurate pressure and flow monitoring within the circuit, and accurate control of tidal volume. The monitoring module is composed of two boards, which are monitoring signal detection board and valve drive board. 1-54 1.7.2.1 Monitoring Signal Detection Board J1 J5 J12 J6 J3 Monitoring signal detection board (Top view) Monitoring signal detection board communication interface, J1 Pin Name Function 1 TXD Serial port transmitting 2 3 4 RXD 12V GND Serial port receiving 12V power supply (range: 10.8V~13.2V) Grounding 5 6 7 8 GND 12V CLOSE_SAFE 5V Grounding 12V power supply (range: 10.8V~13.2V) Safety valve control signal 5V power supply (range: 4.75V~5.25V) VT calibration communication interface, J7 Pin Name Function 1 TXD_CALIBRATE Calibration serial port transmitting signal 2 3 4 RXD_CALIBRATE GND 12V Calibration serial port receiving signal Grounding 12V power supply (range: 10.8V~13.2V) Three-way valve interface, J6 Pin Name Function 1-55 J7 1 12V Three-way valve power supply (range: 10.8V~13.2V) 2 3 4 SOLENOID_VALVE1 12V SOLENOID_VALVE2 Three-way valve control 1 Three-way valve power supply (range: 10.8V~13.2V) Three-way valve control 2 5 6 7 12V SOLENOID_VALVE3 12V Three-way valve power supply (range: 10.8V~13.2V) Three-way valve control 3 Three-way valve power supply (range: 10.8V~13.2V) 8 SOLENOID_VALVE4 Three-way valve control 4 Valve drive board communication interface, J5 Pin Name Function 1 TXD_VPM Ventilation protection board serial port transmitting signal 2 3 4 RXD_VPM GND 12V Ventilation protection board serial port receiving signal Grounding 12V power supply (range: 10.8V~13.2V ) Pin O2 cell interface, J12 Name Function 1 O2+ O2 cell + 2 3 O2GND O2 cell Grounding Test point definition Label Function Range (unit:V) 10V5 R5V 7V A5V 3V3 1V3 2V5 R5V 12V VH1~VH2 VF1~VF3 VP1~VP2 SAN1 SAN2 SAN3 SAN4 10.5V Power supply 5V Power supply 7V Power supply 5V Power supply 3.3V Power supply 1.3V Power supply 2.5V Power supply 5V Power supply 12V Power supply 1.136V Power supply 0.278V Power supply 0.877V Power supply Three-way valve signal 1 Three-way valve signal 2 Three-way valve signal 3 Three-way valve signal 4 9.957~11.025 4.75~5.25 6.65~7.35 4.75~5.25 3.15~3.45 1.235~1.365 2.375~2.625 4.75~5.25 10.8~13.2 1.079~1.193 0.264~0.292 0.833~0.921 0~12 0~12 0~12 0~12 1-56 1.7.2.2 Valve Drive Board Valve drive board (Top view) Valve drive board (Bottom view) Proportional valve and safety valve drive interface, J4 Pin Name Function 1 7V Safety valve power supply (Range: 6.65V~7.37V) 1-57 Pin Name Function 2 3 4 5 6 SAFE_VALVE 7V INSP_VALVE 7V PEEP_VALVE Safety valve control signal Inspiration valve power supply (Range: 6.65V~7.37V) Inspiration valve control signal PEEP valve power supply (Range: 6.65V~7.37V) PEEP valve control signal Power supply interface, J2 Pin Name Function 1 2 3 Grounding Grounding 12V power supply (Range: 10.8V~13.2V) GND GND 12V Status detection interface, J8 Pin Name 1 2 GND PNEUM_PRE_SW 3 4 5 6 7 8 GND NC GND NC GND ACGO_PRE_SW 9 10 GND O2_PRE_SW 11 12 GND MANU_AUTO_SW Function Grounding Pneumatic block pressure switch signal: high level indicates the pneumatically-driven supply is not in position and the low level indicates it in the position; The high level ranges from 2.4V~3.5V, and the low level from 0V~0.8V. Grounding Not connected Grounding Not connected Grounding ACGO switch signal: A high level indicates ACGO is off and a low level indicates ACGO is on; the high level ranges from 2.4V~3.5V and the low level ranges from 0V~0.8V. Grounding Gas supply inlet O2 pressure switch signal: High level indicates O2 supply is not in the position and the low level indicates O2 supply is in the position. the high level ranges from 2.4V~3.5V and the low level ranges from 0V~0.8V. Grounding Auto/Manual switch signal: High level indicates the auto switch and the low level indicates the manual switch. the high level ranges from 2.4V~3.5V and the low level ranges from 0V~0.8V. Test point definition Label Function Range (unit: V) 3V3-VCM 1238mV V7V1 V5V1 SAFE1 7V 3.15~3.45 1.176~1.300 6.65~7.35 6.65~7.35 0~12 6.65~7.35 3.3V Power supply 1.238V Power supply 7V Power supply 7V Power supply Safety valve signal 7V Power supply 1-58 D5V 1V3 D3V3 2V5 R5V 5V Power supply 1.3V Power supply 3.3V Power supply 2.5V Power supply 5 V Power supply 4.75~5.25 1.235~1.365 3.15~3.45 2.375~2.625 4.75~5.25 1.7.3 Battery 1.7.3.1 Battery For A5 anesthesia system: Battery: 11.1V4.5Ah×2 lithium-ion battery (sealed) Battery running time: 150 min (new battery) Battery charge time: maximum 8 hours calculated from initial capacity of 10% 1.7.3.2 Battery Adapter Board J3 Battery adapter board (Top view) J1 J2 Battery adapter board (Bottom view) Battery interface, J1 and J2 Pin Name Function 1 2 3 4 5 6 7 Battery voltage + Battery voltage + Battery in-position signal Battery voltage Battery internal thermistor Battery voltage Battery voltage - BAT+ BAT+ BC BATNTC BATBAT- Battery connection line interface, J3 1-59 Pin Name Function 1 2 3 4 5 6 7 8 VBAT1 NTC1 BC1 GND VBAT2 NTC2 BC2 GND Battery voltage Battery internal thermistor Battery in-position signal Grounding Battery voltage Battery internal thermistor Battery in-position signal Grounding 1.7.4 Heater of the Breathing System The breathing system heater of A5 anesthesia machine provides software and hardware temperature protection. It switches over its operating mode automatically according to the change in ambient temperature. 1.7.5 Anesthesia Calculation and Infrared Module Rack Anesthesia calculation and infrared module rack involve two boards—infrared communication board and anesthesia signal transfer board 1.7.5.1 Infrared Communication Board Infrared communication board (Top view) 1-60 Infrared communication board (Bottom view) Infrared communication board interface, J2 Pin Name Function 1 VPP 12V power supply 2 VPP 12V power supply 3 GND Grounding 4 GND Grounding 5 VDD 3.3V power supply 6 VDD 3.3V power supply (IO pull-up power supply) Pin 1 2 3 Infrared communication board interface, J4 Name Function INTERFACE_TX0 Infrared communication board transmitting signal INTERFACE_RX0 Infrared communication board receiving signal GND Grounding 1.7.5.2 Anesthesia Signal Transfer Board Anesthesia signal transfer board (Top view) 1-61 Anesthesia signal transfer board (Bottom view) Anesthesia signal transfer board power supply and communication interface, P1 Pin Signal Function 1 RXD_AG_TTL AG module receiving signal 2 TXD_AG_TTL AG module transmitting signal 3 GND Grounding 4 VPP 12 V power supply (range: 10.8–13.2 V) AG module power supply and communication interface, P2 Pin Signal Function 1 RXD_AG_232 AG module receiving signal 2 TXD_AG_232 AG module transmitting signal 3 GND Grounding 4 NC No internal connection 5 VPP 12 V power supply (range: 10.8–13.2 V) 6 VPP 12 V power supply (range: 10.8–13.2 V) 7 GND Grounding 8 GND Grounding 1.7.6 Power Board The power board is divided into AC-DC and DC-DC. AC-DC converts the electrical grid power input to 15.2 V DC power supply, and DC-DC converts the 15.2 V power supply output by ACDC or the lithium battery power supply to the following DC voltages required by the integrated system: 12 VA, 12 VB, 5.0 V, and 3.3 V. The power board implements lithium battery charging management and provides functions such as power-on/off control and driving of the AC indicator, battery indicator, and heating module. When the system is started, the power board CPU detects the startup signal and controls the power-on of 3.3 V, 5 V, and 12 VA (12 VB) in sequence, of which 12 VA and 12 VB are powered on simultaneously. When the system is closed, 12 VA (12 VB), 5 V, and 3.3 V are powered off in sequence, of which 12 VA and 12 VB are powered off simultaneously. Interface to the motherboard, J3 Pin Signal Function 1 2 3 Power board serial port communication receiving Power board serial port communication transmitting Grounding RXD TXD GND 1-62 4 LCD-EN LCD backlight enabling signal 5 LCD-BR LCD backlight brightness control voltage 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 3.3VBF GND NTC2 GND NTC1 GND P15V HRT22 HRT21 GND 3.3V GND GND GND NC GND GND 12V1 3.3 V only used for power-on/off (range: 3.135–3.465 V) Grounding Thermoelectric signal connected inside lithium battery 2 Grounding Thermoelectric signal connected inside lithium battery 1 Grounding Heating wire drive voltage output (range: 12.42–15.18 V) Heating wire thermoelectric pin 1 Heating wire thermoelectric pin 2 Grounding 3.3 V voltage output (range: 3.135–3.465 V) Grounding Grounding Grounding No internal connection Grounding Grounding First 12 V voltage output (range: 10.8–13.2 V) 24 15.2VB 15.2 V voltage output (range: 14.44–15.96 V) 25 26 15.2VB PLAM 27 SWITCH 28 29 30 P-FAN+ LED-BAT LED-AC 15.2 V voltage output (range: 14.44–15.96 V) Buzzer drive signal used to drive the buzzer directly Circuit switching, which indicates whether the circuit is in position Power board cooling fan drive Battery status indicator drive output 31 PCON 32 BAT2+ 33 BC2 34 BAT1+ 35 BC1 36 37 38 39 40 41 42 GND P15V HRT12 HRT11 GND 3.3V 5.0V AC status indicator drive output Power-on/off signal, LVTTL pulse signal. When the signal is at high level, the system is powered on. When the signal is at low level, the system is powered off. Lithium battery 2 input, connected to the battery anode Signal indicating the existence of lithium battery 2. When the signal is at low level, lithium battery 2 exists. When the signal is at high level, lithium battery does not exist. Lithium battery 1 input, connected to the battery anode Signal indicating the existence of lithium battery 1. When the signal is at low level, lithium battery 2 exists. When the signal is at high level, lithium battery does not exist. Grounding Heating wire drive voltage output (range: 12.42–15.18 V) Heating wire thermoelectric pin 1 Heating wire thermoelectric pin 2 Grounding 3.3 V voltage output (range: 3.135–3.465 V) 5.0 V voltage output (range: 4.75–5.25 V) 1-63 43 44 45 46 47 48 49 50 GND NC NC GND 12V1 12V1 NC NC Grounding No internal connection No internal connection Grounding First 12 V voltage output (range: 10.8–13.2 V) First 12 V voltage output (range: 10.8–13.2 V) No internal connection No internal connection 1.8 O2-driven Ventilator 1.8.1 O2-driven Ventilator O2 is the drive gas of the ventilator. In addition to the flowmeter module, a high-pressure regulator reduces the supply gas pressure to 200kPa (29psi). This pressure is the pressure of ventilator drive gas. The drive gas pressure regulator is located before the proportional valve. The proportional valve produces drive gas flow during inspiration phase. The gas flow fills the space between the bellows and bellows housing. 1.8.2 High-pressure Drive Gas Regulator (200kPa, 29psi) The drive gas pressure regulator functions to stabilize the supply gas pressure provided for the proportional valve. The gas flow is controlled by the electronic proportional valve and is independent of pressure level. Setting the drive gas pressure regulator to 200kPa (29psi) can cause the ventilator’s maximum inspiratory flow to reach 110L/min. 1.8.3 Drive Gas Assembly The drive gas assembly mainly consists of inspiratory circuit and PEEP circuit. The gas flow in inspiratory circuit goes through the inspiration valve and produces flow of 0 to 110L/min. The gas flow in PEEP circuit goes through the PEEP valve and produces pressure of 3 to 30cmH2O. Both PEEP valve and inspiration valve are electronic proportional valves and are controlled by valve drive board. 1.8.4 Tube Color Coding All the pneumatic tubes used in the anesthesia system are transparent. 1-64 2 Installation Guide 2.1 Preparation The following additional materials are required before installation. The customer is responsible for supplying these materials. Missing items may result in delays, incomplete installations, and/or additional service visits. O2 cylinder, N2O cylinder, and Air cylinder in compliance with the configuration standard. Anesthetic vaporizer and adapter (if not purchased together with the anesthesia system) Anesthetic agent Loose fill or Pre-pak CO2 absorbent Effective O2, N2O, and Air gas sources (280 to 600kPa (40 to 87psi)) Dropdown hoses for ceiling mounted medical gas equipment, compatible with quickdisconnect hoses (if not purchased together with the anesthesia system) 2.2 Assembly NOTE: after the flow sensors are calibrated, the anesthesia machine must be used together with the breathing system module. If the breathing system module is removed, make sure that it is re-installed onto the original anesthesia machine. If other breathing system module is to be installed, the flow sensor must be re-calibrated (please refer to 4.3.2 Flow Calibration (Factory)). 2.2.1 Unpacking and Setup 1. a. b. After receiving A5, check the packing box for damage immediately. If there is no damage and the anti-tilt label on the packing box is in good condition, sign the name and date on the bill of lading or air waybill to indicate that the anesthesia machine has been received safely. If there is damage or the anti-tilt label on the packing box is not in good condition, describe the damage on the bill of lading or air waybill. Both the carrier and the receiver must sign his name and date on the bill of lading or air waybill. Save all damaged delivery package until further information is given by Mindray. The receiver must contact Mindray Customer Service Department immediately. NOTE: When unpacking the equipment, keep as much of the plastic covering on the equipment. When all parts are unpacked, place the packing materials into the original packing box. Then, place the smaller boxes inside the large box. 2-1 2. Cut, remove, and discard the white packing straps on the packing box. 3. Remove the packing box cover and put it on the floor beside the equipment. Later, the cover can be used as slope for the convenience of pushing A5 onto the floor. 4. Lift the packing box and remove it from the top of the equipment. 2-2 5. Remove the foam from the top of A5. 6. Cut the plastic straps as shown below. Pull down the plastic bag from the equipment. 7. Cut the plastic straps close to the back of A5. Take care not to scratch or damage the equipment. Remove and discard the plastic straps. Remove the empty box on the equipment bench. 2-3 8. Remove the foam from the display and bench. 9. After removing the plastic straps and foam, check if there is a box available at both sides of the equipment, as shown below. Remove the package. 10. Use a pair of scissors to cut the two fiber straps on the base of packing box platform. 11. Remove the wood from the front of the A5. Remove the foam and packing material from around the front the equipment. 2-4 12. Remove the foam and packing material from around the back the equipment. 13. Place the packing box cover beside the wooden box base, using as the slope for pushing the equipment onto the floor, as shown below. Place the smooth side of the wood upward. A supporting foot is available at the other side of the wood to support the slope. Fix the slope onto the wooden box with buckle strap. 14. Rotate the casters for 90°. Then push A5 gently to roll it down the slope. Remove the packing box on the equipment. Keep the packing box in case that A5 needs re-packing. Note: Please check the packing details when performing the following installations. 15. Open the bottom drawer and take out the breathing system assembly and bag arm assembly. 2-5 16. Install the breathing system assembly onto A5. Take care to align the assembly with A5 and then push it to A5 directly until it is stuck at the proper position. 17. Open the small box containing bellows assembly and bellows housing gently. Remove the plastic bag from the assembly and put the foam and plastic bag into the box. 18. Install the bellows onto the breathing system. After installation, make sure that the bellows are fully extended and surround the tabs of breathing system. 19. Place the bellows housing downward onto the breathing system. Then rotate it clockwise to fix it at proper position (the scale marks on the bellows housing shall be placed forward for the convenience of being seen by the operator). 2-6 20. Align with the slots to install the bag arm assembly as shown below. 21. Push the bag arm assembly into the breathing system and tighten the fixing nut as shown below. 2-7 22. Open the middle and bottom drawers and take out the detachable absorber canister assembly, waste gas scavenging system hose, and waste gas scavenging assembly gently. 23. Install the white absorber canister hose onto the absorber canister assembly. Install the absorber canister assembly when it is not stably placed. First install its lower part. Align the pins with the holes. Then align the upper part. Push in the upper part upward until the front and back latches are stuck at the proper position. 2-8 24. Fill Pre-pack or loose fill CO2 absorbent into the canister. Slide the canister into the canister assembly. Rotate the lock bar anticlockwise for 90° to fix the canister at the proper position. 25. Slide the scavenging assembly into the rail at the left bottom of A5 (namely, the same side with the breathing system assembly). Then tighten the thumbscrew to fix the assembly at the proper position. Install one end of the scavenging hose onto the scavenging assembly and the other end onto A5, as shown below. 26. Open the top drawer and check if it contains the following: Auxiliary O2/Air reference card Preoperative checklist Inspiratory flow sensor and expiratory flow sensor O2 cell unit (optional configuration) Cylinder wrench (optional configuration) Drawer key 2-9 27. Loosen the two fast locking screws to open the service door. Remove the two hand screws to open the battery box. Install one or two batteries and make sure that the batteries are installed with correct polarity. Close the battery box and tighten the hand screws. 28. Before installing the cylinders onto the back of A5, make sure that washer is already installed. Remove the tape fixing the washer. Then install the cylinders. Make sure that the cylinders are fixed onto the respective cylinder connectors (marked with O2, Air, and N2O). 29. Connect the gas supplies. Rotate the hose connectors clockwise to tighten them onto the gas supply connectors. Make sure that the supply gas pressure is within the specified range. 30. Connect the manual bag (supplied by the user) to the bag arm. 31. Connect the patient’s breathing circuit (supplied by the user) to the inspiratory and expiratory connectors. WARNING: Use breathing circuit and manual bag according to ASTMF1208. Standard male 22mm conical fittings can be used for breathing circuit and manual bag according to ASTM F1054. 32. Rotate the knob at the top of AGSS tank until the float is between the minimum and maximum marks. Connect the hose from the gas scavenger to the EVAC connector of operating room. 33. Install the O2 sensor inside the stainless housing. Tighten the O2 sensor properly. 34. Connect the O2 sensor external cable between the O2 sensor and the side of A5. Align the cable with the yellow mark on the connector. 2-10 35. Insert the power cord into the grounded socket. Set the main unit power switch (located on the front of A5) to ON position to turn on A5. Wait until the LCD display displays leak test related information. View if power-on selftest is successful. Do not connect, disconnect, or move the breathing circuit or bag during selftest. 36. Install the patient monitor and arm according to the instructions in the monitoring kit. WARNING: Use only Mindray approved patient monitor and arm for A5. 37. Connect the sample gas return hose (P/N: 115-015457-00) to the gas module outlet and the sample gas return port at the back of A5 (above the EVAC connector). 38. Put the following parts into the bottom drawer: A5 operator’s manual Sealing washer 39. Hang the preoperative checklist and auxiliary O2/Air reference card onto the handle of A5. 2.2.2 Breathing System, Breathing System Accessories and Checkout Procedures 1. Patient circuit 3.AGSS 2. AGSS transfer tube 4. CO2 absorbent 5. O2 sensor and cable 2.2.3 AGSS Connections 1. 2. 3. Take out the AGSS from the foam and plastic bag. Install the AGSS onto A5 via the GCX compatible rail system. Connect the AGSS transfer tube between the AGSS inlet port and A5 exhaust port. 2-11 2.2.3.1 Cylinder Wrench and Preoperative Checklist 1. 2. Fix the cylinder wrench onto the back of A5 so that it can be used to open or close cylinders without disconnecting from the machine. Put the preoperative checklist at the location of A5 easily accessible for the operator. 2.2.4 Vaporizer WARNING: If the vaporizer is incompatible with the A5 Anesthesia System, the vaporizer will not work at all. Use vaporizers with Selectatec mounting system that are compliant to ISO8835-4. Refer to the vaporizer manufacturer’s Instructions For Use for filling or draining the vaporizer and other information. WARNING: For the A5 Anesthesia System, using or turning on more than one vaporizer simultaneously is mechanically prevented by the Selectatec mount system. Do not attempt to override this safety feature. WARNING: Use care in lifting and manipulating vaporizers during the mounting process as their weight may be greater than expected, based on their size and shape. NOTE: The barometric pressure at the customer side may be different from the factory calibration pressure of anesthetic vaporizer. This may cause inaccurate concentration output of anesthetic agent. The operator should continuously monitor the concentration of anesthetic agent during system use. 1. Align the vaporizer over the valve cartridges of the mounting bar. Hang the vaporizer on the mounting bar as shown below. Note that the locking handle is in the unlocked position. Locking handle in the unlocked position 2-12 2. Rotate the locking handle clockwise to the locked position as shown below. Locking handle in the locked position 2.2.4.1 Assemble the Vaporizer 1. 2. 3. 4. 5. Mount the vaporizer onto the manifold. Push and turn the locking lever clockwise to lock the vaporizer in position. Ensure that the top of the vaporizer is horizontal. If not, remove the vaporizer and reinstall it. When reinstalling the vaporizer, lift each vaporizer straight up off the manifold rather than pulling forward. Do not rotate the vaporizer on the manifold. If a vaporizer unintentionally lifts off the manifold, install it again and complete steps 1 through 3. If the vaporizer lifts off a second time, do not use the system. NOTE: A Desflurane vaporizer may be mounted similarly as other vaporizers, but may require a power cord. For more detailed instructions on installation and proper use, refer to the specific manufacturer’s Instructions for Use of the Desflurane vaporizer. 2.2.4.2 Fill the Vaporizer NOTE: The A5 should use vaporizers with Selectatec mounting system that are compliant to ISO8835-4. Refer to the vaporizer manufacturer’s Instructions For Use for filling or draining the vaporizer and other information. WARNING: Ensure that the correct anesthetic agent is used. The vaporizer is designed with the specific anesthetic agent name on it and further indicated by color coded label. The concentration of the anesthetic agent actually outputted will vary if the vaporizer is filled with the wrong agent. 2.2.4.3 Drain the Vaporizer WARNING: Do not reuse the agent drained from the vaporizer. Treat as a hazardous chemical and follow local regulations for proper disposal. NOTE: The A5 should use vaporizers with Selectatec mounting system that are compliant to ISO8835-4. Refer to the vaporizer manufacturer’s Instructions For Use for filling or draining the vaporizer and other information. 2-13 2.2.5 Backup Cylinder 1. 2. 3. 4. 5. 6. 7. Remove the cover from a new O2, N2O, and AIR cylinder. Mount one cylinder at a time onto the rear of the anesthesia machine. Discard the cylinder's tank washer. Always use the approved tank washer provided with the A5. Open the bail of each yoke and mount the cylinder over the tank washer. Ensure the O2 cylinder mates to the O2 Pin Index Safety System (PISS) connection on the O2 yoke. Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke port. Ensure that the N2O cylinder mates to the N2O PISS connection on the N2O yoke. Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke port. Ensure that the AIR cylinder mates to the AIR PISS connection on the AIR yoke. Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke port. 2.2.6 Monitoring Products Mounting and Electrical Connection (if available) 1. Mount the monitor (if available) according to the manufacturer’s monitor assembly instructions. NOTE: Use of other monitor and mounting hardware is performed by the installer. 2. After mounting a monitor to the A5, connect it to one of the AC outlets located on the rear of the A5. 3. Turn on each monitor one at a time. Verify that the circuit breaker holds without tripping. 4. Dress each line cord neatly along the side of the anesthesia machine or tucked inside the monitor arm. An optional cable routing kit is available. The cable routing kit contains three (3) clips, screws, and two (2) Ethernet cables. The clips attach to three (3) sets of holes on the rear door of the A5. Ethernet and power cables can be routed through the clips. 2-14 2.3 Factory Setup The password set by the factory is 558188. Choose [Main] > [Service] and then enter the password set by the factory to access the factory setting menu. The factory setting menu includes function activation, drive gas setting, flowmeter pipeline setting, ACGO setting, module rack setting, AG module setting, BIS module setting, NMT module setting and CO2 module setting. After you modify some factory configurations, you have to restart the machine for the modification to take effect. After you enter the password set by the factory, a screen as shown in the following figure is displayed. NOTE: All setting items under factory setting menu are enabled only in standby mode. 2-15 2.3.1 Function Activation The system supports activation code-based activation and activation code file-based activation. The activation code file must be placed in the root directory of USB. The newly activated function can take effect only after you restart the machine. 2.3.2 Drive Gas Drive gas has two options: O2 and AIR. Maintenance personnel shall configure this function as required. 2-16 2.3.3 Flowmeter Standard Flowmeter standard has five options: American, European, Chinese, Canadian, and Custom. Maintenance personnel shall configure this function as required. NOTE: Canadian-standard flowmeters are in the order of American-standard flowmeters but with the colors of European-standard flowmeters. Custom-standard flowmeters are in the order of AIR, N2O and O2 from left to right with the colors of European-standard flowmeters. 2.3.4 Flowmeter Pipeline A Flowmeter pipeline has three options: O2+N2O+AIR, O2+AIR, and O2+N2O. Maintenance personnel should configure this function according to the practical requirements. 2.3.5 ACGO ACGO has three options: Off, ACGO with three-way valve, and electronic ACGO. Among them, Off applies to the machine that does not need to be configured with ACGO, machine that needs to be configured with a non-independent ACGO, and machine that needs to be configured with independent ACGO without three-way valve. Independent ACGO with threeway valve applies to the machine that needs to be configured with ACGO with three-way valve. Electronically controlled ACGO applies to the machine that needs to be configured with electronically controlled ACGO. 2.3.6 Module Rack The setting of module rack provides two options: Open and Close. When module rack is set to Open, the machine can identify module plugging and monitor the module rack fan speed. When module rack is set to Close, the machine cannot identify module plugging or monitor the module rack fan speed. 2.3.7 AG Module The setting of AG module provides two options: Open and Close. When AG module is set to Open, the machine can identify the external AG module. When AG module is set to Close, the machine cannot identify the external AG module. 2.3.8 BIS Module The setting of BIS module provides two options: Open and Close. When BIS module is set to Open, the machine can identify the external BIS module. When BIS module is set to Close, the machine cannot identify the external BIS module. 2.3.9 CO2 Module The setting of CO2 module provides two options: Open and Close. When CO2 module is set to Open, the machine can identify the external CO2 module. When CO2 module is set to Close, the machine cannot identify the external CO2 module. 2-17 2.3.10 NMT Module The setting of NMT module provides two options: Open and Close. When NMT module is set to Open, the machine can identify the external NMT module. When NMT module is set to Close, the machine cannot identify the external NMT module. 2.4 Functional Tests Refer to 4 Calibration if any values are out of specification. NOTE: The A5 system must be under power-up state (AC power instead of battery) and the breathing system heating device set to ON position at least an hour before performing the ventilation test (starting from 2.7.5.) (version 2.01.00 and higher). 2.4.1 Gas Delivery System Tests 2.4.1.1 O2 Flush Verification 1. 2. 3. 4. 5. Connect the bag arm to the expiratory port via a breathing hose. Set the Auto/Manual switch to the Manual position. Set the APL valve to 75. Connect the flowmeter to the inspiratory port. Verify that the O2 flush flow is between 35 and 50 L/min when the O2 flush valve is pressed. 2.4.1.2 O2:N2O Ratio System 1. 2. 3. 4. 5. 6. Set the O2 and N2O flow control valves to minimum. Rotate the O2 flow control valve until the flowmeter displays 1L/min. Rotate the N2O flow control valve and adjust N2O flow to maximum. Verify that the N2O flow is between 2.5L/min and 3.7L/min. Decrease the N2O flow to minimum. Decrease the O2 flow to minimum. 2.4.2 Breathing System Leak Test NOTE: Perform leak test again each time after servicing the anesthesia machine, replacing the components, or re-connecting the tubes. 2-18 2.4.2.1 Breathing System Leak Test in Mechanical Ventilation Mode The test aims to check if the pneumatic circuit has leaks in mechanical ventilation mode. Test items include bellows, drive gas circuit, CO2 canister, patient tubes, flow sensors and their connectors. To do the breathing system leak test in mechanical ventilation mode: 1. You can access the automatic circuit leak & compliance test screen after passing poweron selftest. You can also access the automatic circuit leak & compliance test screen by opening the Main Menu and select Test Leak/Compliance from General. The automatic circuit leak & compliance test screen is shown below. Set up the machine as per the instructions on the screen. Then select Continue to execute automatic circuit leak and compliance test. You can select Skip to access standby screen directly. 2. Undertake the related operations according to the notes in the screen. Results Automatic circuit leak volume: Pass Compliance test: XX.X mL/cmH2O Indications/Options Leak volume ≤ 200 mL/min Compliance test result displays Green. Choose [Continue] and access to the manual circuit leak test screen. 2-19 Automatic circuit leak: XXX mL/min Compliance test: Pass 200 mL/min < Leak volume ≤ 1000 mL/min Compliance test went on successfully. Choose [Continue] and access to manual circuit leak test screen. Make use of the previous compliance value. Or Choose [Retry] and repeat the automatic circuit leak and compliance test. Automatic circuit leak: XXX mL/min Compliance test: Fail Treatment measures: 1.Check if all the pipes are connected well. 2.Check if the Y-piece port is sealed well. 3.Check if the water collection cup is installed and the drain valve is closed well. 4.Check if the sampling pipe is connected well. Leak volume >200 mL/min and ≤ 1000 mL/min Choose [Continue] and access the manual circuit leak test screen and make use of the previous compliance value.(Use the fresh air to compensate the leak volume) Or Choose [Retry] and repeat the automatic circuit leak and compliance test. Automatic circuit leak: Fail Compliance test: Fail Treatment measures: 1.Check if all the pipes are connected well. 2.Check if the Y-piece port is sealed well. 3.Check if the water collection cup is installed and the drain valve is closed well. 4.Check if the sampling pipe is connected well. Leak volume > 1000 mL/min. No fresh air is detected. Eliminate the troubles according to the tips in the screen. Or Select [Manual Only] to enable only manual ventilation for the equipment. The low-priority alarm "Automatic Ventilation Disabled–Leak Test Failed" will be displayed on the main screen. WARNING Selecting Manual Only will disable mechanical ventilation. Treatment measures: 1.Check if all the pipes are connected well. 2.Check if the Y-piece port is sealed well. 3.Check if the water collection cup is installed and the drain valve is closed well. 4.Check if the sampling pipe is connected well. 2-20 MACHINE NON-FUNCTIONAL Automatic Circuit Leakage: Fail Compliance Test: XX.X mL/cmH2O Safety Valve Control: Fail The safety valve control test or pressure verification test failed. Select Retry to repeat the automatic circuit leak and compliance tests. Or Contact the customer service center if this error condition persists. NOTE Service Access button: This button is available only to Mindray-authorized service personnel and a service password is required. Treatment measure: Check if the safety valve connection cables are connected well. NOTE: In case of automatic circuit leak test failure or compliance test failure, make use of the visual tips and the helpful information to check if the drain valve and the sampling port have been sealed completely. Besides, the connecting error or interface damage of bellows, respiratory system pipes or CO2 canister also can cause the test failure. 2-21 2.4.2.2 Breathing System Leak Test in Manual Ventilation Mode The test aims to check if the pneumatic circuit has leaks in manual ventilation mode. Test items include APL valve, check valve, CO2 canister, patient tubes, flow sensors and their connectors. To do the breathing system leak test in manual ventilation mode 1. You can access the manual circuit leak test screen after passing automatic circuit leak & compliance test. You can also access the manual circuit leak test screen after opening the Main Menu and select Test Leak/Compliance from Calibrate/General to complete automatic circuit leak & compliance test. The manual circuit leak test screen is shown below. 2. Perform the related operations according to the directions in the screen. Results Pass Indications/Options Leak volume <= 200ml/min. Fail: XXX mL/min Adjust APL valve to SP. Choose Continue and access to the main menu. Leak volume >200 ml/min & <=1000 ml/min Eliminate troubles according to the tips in the screen. Or Adjust APL valve to SP and choose Accept and then access to the main menu. Treatment measures: 1. Check if all the pipes are connected well. 2. Check if the Y-piece port is sealed well. 2-22 Results Fail Indications/Options 3. Check if the water collection cup is installed and the drain valve is closed well. 4. Check if the sampling pipe is connected well. Leak volume > 1000 ml/min Eliminate troubles according to the tips in the screen. Or Adjust APL valve to SP and choose Accept and then access to the main menu. Fail: Fresh gas flow detected Time out Treatment measures: 1. Check if all the pipes are connected well. 2. Check if the Y-piece port is sealed well. 3. Check if the water collection cup is installed and the drain valve is closed well. 4. Check if the sampling pipe is connected well. The manual circuit leak test failed. Fresh gas is detected. Zero all flowmeters. Select Retry to repeat the test. Treatment measure: Cut off the fresh air. Test results cannot be displayed due to an internal communication error. Select Retry to repeat the automatic circuit leak and compliance tests. Or Select Override to skip the test. Treatment measures: 1. Check if all the pipes are connected well. 2. Check if the Y-piece port is sealed well. 3. Check if the water collection cup is installed and the drain valve is closed well. 4. Check if the sampling pipe is connected well. NOTE: If there is a leak, check the pneumatic circuit system for leakage and troubleshoot the problems as described in 5.3 Pneumatic System Problems. Do leak test again after failure source is fixed, NOTE: If there is a leak, check the pneumatic circuit system for leakage and troubleshoot the problems as described in 5.3.4 Breathing System. Do leak test again after leak failure is fixed. 2-23 2.4.2.3 Commonly-encountered Problems and Recommended Actions The following table lists the commonly-encountered problems and recommends actions for 2.4.2.2Breathing System Leak Test in Manual Ventilation Mode. Failure description Possible cause Recommended action Leak test failure is prompted immediately after [Begin] is selected (typically, the leak test requires at least 3 minutes). The Auto/Manual switch is set to the Manual position and the message [Manual Vent.] is prompted. The reading on the drive gas (O2) pressure gauge indicates drive gas pressure low (lower than 200kPa) and the alarm of [Drive Gas Pressure Low] is produced. During leak test, the pressure indicated by the airway pressure gauge fails to reach 30cmH2O. 1. Before the leak test, the bellows is not fully inflated. 2. The Y piece on the breathing tube is not connected to the test plug. 3. The bellows housing is not properly installed. Set the Auto/Manual switch to the Auto position Replace or connect gas supplies and make sure that the drive gas pressure is at 350 to 450kPa. Check the pneumatic circuit connections and re-install the pneumatic circuit. 2.4.3 O2 Sensor Calibration NOTE: Perform 21% and 100% O2 calibrations before using A5 for the first time. The O2 sensor is not calibrated before delivery. NOTE: Calibrate the O2 sensor again when a great deviation of O2 concentration monitored value occurs, or when the O2 sensor or valve drive board is replaced, or when the anesthesia system gives a prompt message. NOTE: Before calibration, check if the O2 sensor displays numerics on the measure screen. If not, check the O2 sensor connection line or replace the O2 sensor until measure numerics are displayed. 2.4.3.1 21% O2 Calibration Follow these steps to calibrate O2 sensor at 21% O2. 1. Select [Main]> [General]> [Calibrate O2 Sensor], or select [Main]> [System]> [Calibrate / Check O2 Sensor], or select [Main]> [Service]> [Calibration]> [O2 Sensor] to enter the screen shown below. The [General] tab displays only the O2 sensor calibration at 21% O2. The [System] and [Service] tabs display the O2 sensor calibration at 21% O2 and 100% O2, and need passwords. Follow the instructions on the screen to set the machine. Select [Begin] to start calibration. 2-24 2. The calibration screen shown below is displayed after [Begin] is selected. During the calibration, you can select [Cancel] to abort. 2-25 3. The screen shown below is displayed if the ongoing calibration is aborted. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 4. The screen shown below is displayed if the calibration fails. The fonts in red indicate the fault code. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 5. The screen shown below is displayed after a successful calibration. Select [Done] to exit the calibration screen. 2-26 2.4.3.2 100% O2 Calibration NOTE 100% O2 calibration must be performed in standby mode. NOTE 100% O2 calibration can be performed only after a successful 21% O2 calibration. Follow these steps to calibrate O2 sensor at 100% O2. 1. Enter the standby mode. 2. Select [Main] > [System] > [Calibrate / Check O2 Sensor] or select [Setup] > [Service] > [Calibration] > [O2 Sensor]. The [System] and [Service] tabs display O2 sensor calibration at 21% O2 and 100% O2 and need passwords. The calibration screen shown below is displayed after [100%] is selected. Follow the instructions on the screen to set the machine and select [Next]. Set the Auto/Manual switch to the Auto position. 3. After [Next] is selected, the calibration screen shown below is displayed. Follow the instructions on the screen to set the machine. 2-27 4. After [Next] is selected, the calibration screen shown below is displayed. Follow the instructions on the screen to set the machine. Wait 2 minutes and ensure that the O2 cell voltage has stabilized at the maximum value for at least 30s. Select Begin. 5 The calibration screen shown below is displayed after [Begin] is selected. During the calibration, you can select [Cancel] to abort. 6. The screen shown below is displayed if the ongoing calibration is aborted. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 2-28 7. The screen shown below is displayed if the calibration fails. The fonts in red indicate the fault code. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 8. The screen shown below is displayed after a successful calibration. Select [Done] to exit the calibration screen. 2-29 2.4.3.3 Commonly-encountered Problems and Recommended Actions Failure description After [Begin] is selected, calibration failure is prompted very soon. Possible Cause Recommended Action If the alarm [O2 Sensor Unconnected] is displayed, it indicates that O2 sensor is not connected. Connect the O2 sensor. O2 supply pressure is insufficient (lower than 200kPa). 21% O2 calibration is not completed before 100% O2 calibration. Calibration failure is prompted about 3 minutes after calibration is started. O2% count value is not within the normal range (450 to 2700). Change or connect the gas supply and make sure that O2 supply pressure is sufficient. Perform 100% O2 calibration after 21% O2 calibration is completed. Replace the O2 sensor. 2.5 Gas Supply Leak Tests Fully close the three needle valves before doing the following tests. 2.5.1 N2O Cylinder Leak Test 1. Remove the N2O line pressure hose from the line pressure inlet on the A5. 2. Mount a full N2O cylinder to the rear panel yoke. If necessary, place a new clean tank washer between the cylinder and the yoke to minimize any leaks at the yoke connection. 3. Open the N2O cylinder until its pressure gauge indicates cylinder pressure. 4. Close the N2O cylinder. 5. The N2O cylinder pressure gauge should not drop more than 10% of its initial pressure over one minute. 2.5.2 O2 Cylinder Leak Test 1. Remove the O2 line pressure hose from the line pressure inlet on the A5. 2. Mount a full O2 cylinder to the rear panel yoke. If necessary, place a new clean tank washer between the cylinder and the yoke to minimize any leaks at the yoke connection. 3. Open the O2 cylinder until its pressure gauge indicates cylinder pressure. 4. Close the O2 cylinder. 5. The O2 cylinder pressure gauge should not drop more than 10% of its initial pressure over one minute. 2.5.3 AIR Cylinder Leak Test 1. Remove the AIR line pressure hose from the line pressure inlet on the A5. 2. Mount a full AIR cylinder to the rear panel yoke. If necessary, place a new clean tank washer between the cylinder and the yoke to minimize any leaks at the yoke connection. 3. Open the AIR cylinder until its pressure gauge indicates cylinder pressure. 4. Close the AIR cylinder. 5. The AIR cylinder pressure gauge should not drop more than 10% of its initial pressure over one minute. 2-30 2.5.4 N2O Pipeline Supply Pressure Leak Test NOTE Remove the N2O cylinder from the N2O cylinder yoke when doing this test. 1. Connect the machine to the N2O supply connector and O2 supply connector on the wall via N2O supply hose and O2 supply hose. Check that the N2O supply pressure gauge displays 280~600Kpa. Turn on the system switch. Open the O2 needle valve. Adjust the flow to 1L/min to fully open the N2O needle valve. Observe the float of total flowmeter, which should fall between 3L and 5L. Disconnect the N2O pipeline supply. Check that the N2O supply pressure gauge decreases to zero. 2. 3. 4. 5. 6. 7. 2.5.5 O2 Pipeline Supply Pressure Leak Test NOTE Remove the O2 cylinder from the O2 cylinder yoke when doing this test. 1. 2. 3. 4. 5. 6. 7. Connect the machine to the O2 supply connector on the wall via O2 supply hose. Check that the O2 supply pressure gauge displays 280~600Kpa. Turn on the system switch. Fully open the O2 needle valve. Observe the float of total flowmeter, which should be at the top of the flowmeter. Disconnect the O2 pipeline supply. Make sure that the alarms of [O2 Supply Failure] and [Drive Gas Pressure Low] occur with the decrease of O2 pressure. Check that the O2 supply pressure gauge decreases to zero. 8. 2.5.6 Air Pipeline Supply Pressure Leak Test NOTE Remove the Air cylinder from the Air cylinder yoke when doing this test. 1. 2. 3. 4. 5. 6. 7. Connect the machine to the Air supply connector on the wall via Air supply hose. Check that the Air supply pressure gauge displays 280~600Kpa. Turn on the system switch. Fully open the Air needle valve. Observe the float of total flowmeter, which should be at the top of the flowmeter. Disconnect the Air pipeline supply. Check that the Air supply pressure gauge decreases to zero. 2-31 2.6 Breathing System Checks 2.6.1 AGSS Test (if available) 1. Connect one end of the waste gas exhaust hose to the port on the waste gas scavenging assembly and the other end to the EVAC port. NOTE: If other type of waste gas scavenging system is used for the anesthesia system, make sure that the waste gas enters this scavenging system after emitting from the EVAC port. 2. Connect the respiratory gas monitor exhaust port to the sample gas return port. 3. Make sure that the flow of waste gas scavenging system can be adjusted between the minimum and maximum scales. 2.6.2 Internal Gas Connections Test 1. Close and remove all gas cylinders from the A5 2. Connect only the AIR pipeline supply hose to the A5 from the wall supply. Leave all other pipeline supply hoses disconnected. 3. With the A5 powered on, rotate the AIR flow control knob to ensure a continuous flow increase throughout its full range 4. Fully rotate the N2O flow control knob and verify that there is no flow. 5. Fully rotate the O2 flow control knob and verify that there is no flow. 6. Disconnect the AIR pipeline supply hose from the A5, and connect the O2 pipeline supply hose from the wall supply, rotate the O2 flow control knob to ensure a continuous flow increase throughout its full range. 7. Fully rotate the N2O flow control knob and verify that there is no flow. 8. Fully rotate the AIR flow control knob and verify that there is no flow. 9. Connect the N2O pipeline supply hose from the wall supply. With the O2 flow control knob fully opened, rotate the N2O flow control knob to ensure a continuous flow increase throughout its full range. NOTE: The maximum N2O flow must be between 10 and 12L/min. 10. Fully rotate the AIR flow control knob and verify that there is no flow. 2.6.3 Alarm of Drive Gas Pressure Low, O2&N2O Cutoff Functional Test 1. Set the O2 flow to 2L/min using the flow control valve. 2. Set the N2O flow to 2L/min using the flow control valve. 3. Set the AIR flow to 2L/min using the flow control valve. 4. Interrupt the O2 supply for the A5. 5. Verify that N2O and O2 stop flowing within two minutes and that AIR (if available) continues to flow at 2L/min. 6. Verify the following alarms are activated: O2 Supply Failure appears on the screen An alarm tone sounds. 2-32 2.7 Performance Verification 2.7.1 Standby Mode Ventilation Test 1. Ensure that the gas pressures of O2, N2O, and AIR are within the range of specifications. 2. Power on the anesthesia system. 3. Perform the startup test as per the on-screen instructions. Ensure successful completion. 4. Set the Auto/Manual switch to the Manual position. 5. Attach a breathing circuit and test lung to the Y piece of the breathing circuit. NOTE: For testing purpose always use a reusable breathing circuit. 6. Set the APL valve to approximately 15cmH2O. 7. Set the AIR flow to 5L/min using the flow control valve. Use O2 if AIR is not available. 8. Squeeze the breathing bag once every 10 seconds to inflate and deflate the test lung so as to produce pressure of approximately 20cmH2O. 9. Verify the inflation and deflation of the test lung. 2.7.2 Manual Mode Ventilation Test 1. Set the Auto/Manual switch to the Manual position to switch to manual mode 2. Set the APL valve to approximately 25cmH2O. Push the O2 flush button to fill the breathing bag. 3. Set the AIR flow to 1L/min using the flow control valve. 4. Squeeze the breathing bag once every 3 seconds. 5. Verify the inflation and deflation of the test lung. 6. Verify that an airway pressure waveform and all numeric values appear on screen during bag compressions. 7. Stop squeezing the breathing bag and set the APL Valve to the open position (SP). 2.7.3 Apnea Alarm Test 1. While in the manual ventilation mode, stop ventilating the test lung. 2. Verify that the following apnea alarm signals are activated at approximately 30 seconds from the last bag compression (for enterprise standard, it is 20 seconds for SFDA machine). Apnea appears on the screen. An alarm tone sounds. 2.7.4 Alarm Silence Test 1. While the apnea alarm is sounding, press the Silence soft key. 2. Make sure that the audio alarm stops and resumes after two minutes. 2.7.5 VCV Adult Ventilation Mode Test 1. Attach a breathing circuit and breathing bag. NOTE: For testing purpose always use a reusable breathing circuit. 2. Attach an adult test lung to the Y piece of the breathing circuit. 3. Attach a ventilation test device between the expiration port and the expiratory hose. 4. Set the O2 flow to 2L/min and set the N2O and AIR flows to the minimum.. 5. Set the Auto/Manual switch to the Auto position. 6. Set the ventilating control parameters to: 2-33 Ventilation control parameter Setting of ventilation control parameter Patient type Vent mode Vt Rate I:E Tpause PEEP Plimit Adult VCV 600 8 1:2 10 Off 50 7. Press Set Mode button to begin ventilation. 8. Verify that pressure waveform, Vt, MEAN or PLAT, Rate and MV appear on the screen. 9. Verify the tidal volume displayed on the ventilation test device is within 7% (±42mL) of the setting value within approximately one minute from the start of ventilation. 10. Verify the tidal volume displayed is within 9% (±54mL) of the setting value within approximately one minute from the start of ventilation. 11. Verify the measured O2 concentration is at least 97% after 5 minutes. 12. Set the AIR flow to 3L/min and set the N2O and O2 flows to the minimum. 13. Verify the measured O2 concentration is 21% ±3% vol. % after 5 minutes. 2.7.6 VCV Pediatric Ventilation Mode Test 1. Attach a breathing circuit and breathing bag. NOTE: For testing purpose always use a reusable breathing circuit. 2. Attach a pediatric test lung to the Y piece of the breathing circuit. NOTE: Limit the volume in the test lung to provide sufficient airway pressure to satisfy the Low Peak Pressure alarm. Or reduce the Peak Pressure alarm limit to a lower value to prevent the alarm when using an adult test lung. 3. Attach a ventilation test device between the expiration port and the expiratory hose. 4. Set the O2 flow to 2L/min and set the N2O and AIR flows to the minimum. 5. Set the ventilating control parameters to: Ventilation control parameters Patient type Vent mode Vt Rate I:E Tpause PEEP Plimit Setting of ventilation control parameter Pediatric VCV 120 15 1:2 10 Off 40 6. Press Set Mode button to begin ventilation. 7. Verify that pressure waveform, Vt, MEAN or PLAT, Rate and MV appear on the screen. 8. Verify the tidal volume displayed is within ±18ml of the setting value within approximately one minute from the start of ventilation. 2-34 9. Verify the delivered volume as measured by ventilation test device at the expiration port, is within ±15ml of the setting value within approximately one minute from the start of ventilation. 2.7.7 Airway Disconnection Alarm Test 1. While the ventilator is running, disconnect the expiratory limb from the expiration port on the breathing system. 2. Verify the following airway pressure disconnection alarm signals are activated: Paw Too Low message appears on the screen. An alarm tone sounds. 2.7.8 PCV Adult Ventilation Mode Test 1. Attach a breathing circuit and breathing bag. NOTE: For testing purpose always use a reusable breathing circuit. 2. Attach an adult test lung to the Y piece of the breathing circuit. 3. Attach a ventilation test device between the expiration port and the expiratory hose. 4. Set the O2 flow to 3L/min and set the N2O and AIR flows to the minimum. 5. Set the ventilator controls to: Ventilator Controls Ventilator Settings Patient type Vent mode VtG Pinsp Rate I:E PEEP Tslope PlimVG Adult PCV Off 15 8 1:2 Off 0.2 NA 6. Press Set Mode button to begin ventilation. 7. Verify the Peak Pressure reading on the display is within ±2cmH2O of the set Pinsp. 8. Verify that the pressure waveform, Vt, MEAN or PLAT, Rate and MV appear on the screen. 9. Verify that the PEAK value measured by the ventilation test device reaches 15±2.5cmH2O within five breaths from the start of ventilation. 2.7.9 Pressure Support (PS) Ventilation Mode Test 1. Attach a breathing circuit and breathing bag. NOTE: For testing purpose always use a reusable breathing circuit. 2. Attach an adult test lung to the Y piece of the breathing circuit. 3. Attach a ventilation test device between the expiration port and the expiratory hose. 4. Set the O2 flow to 1L/min and set the N2O and AIR flows to the minimum. 5. Set the ventilator controls to: Ventilator Controls Ventilator Settings Patient type Vent mode Adult PS 2-35 Ventilator Controls Ventilator Settings Min Rate ΔP Trigger PEEP Tslope 4 20 3 Off 0.2 6. Press Set Mode button to begin ventilation. 7. Begin triggering breaths by slightly squeezing the test lung and releasing. Maintain a continuous breath rate. 8. Verify that a pressure waveform and all ventilation parameters appear on the screen. 9. Verify that the Peak Pressure reading on the display is ±2 of ΔP+PEEP. 10. Stop triggering breaths. 11. Verify that after 15 seconds the ventilator delivers a breath and displays the message Apnea Ventilation. 12. Verify the system ventilates at a frequency of 4bpm. 2.8 Alarm Function 2.8.1 Setup 1. Ensure that the gas pressures of O2, N2O, and AIR are within the range of specifications. 2. Power on the anesthesia system. 3. Perform the startup test as per the on-screen instructions. Ensure successful completion. 4. Attach a breathing circuit and breathing bag. NOTE: For testing purpose always use a reusable breathing circuit. 5. Attach an adult test lung to the Y piece of the breathing circuit. 6. Set the O2 flow to 2L/min and set the N2O and AIR flows to the minimum. 7. Set the ventilating control parameters to: Ventilation control parameter Setting of ventilation control parameter Patient type Vent mode Vt Rate I:E Tpause PEEP Plimit Adult VCV 600 8 1:2 10 Off 50 8. Press Set Mode button to begin ventilation. 2-36 2.8.2 FiO2 Too Low Alarm Test 1. Set the FiO2 Too Low alarm limit to 50%. 2. Set the AIR flow control valve to 5L/min. 3. Set the O2 flow control to the minimum. 4. Verify the following FiO2 too low alarm signals are activated within three ventilation cycles: FiO2 Too Low message appears on the screen. An alarm tone sounds. 5. Set FiO2 Too Low alarm limit to 18%. 6. Verify the alarm signals cease. 2.8.3 FiO2 Too High Alarm Test 1. Set the FiO2 Too High alarm limit to 50%. 2. Set the O2 flow control valve to 5L/min. 3. Set the AIR flow control to the minimum. 4. Verify the following FiO2 Too High alarm signals are activated: FiO2 Too High message appears on the screen. An alarm tone sounds. 5. Set the FiO2 Too High alarm limit to the maximum value. 6. Verify the alarm signals cease. 2.8.4 Peak Pressure Alarm Test 1. Set the PEAK too low alarm to the minimum value. 2. Set the PEAK too high alarm limit to about 5 to 8 digits below the Peak Pressure displayed on the screen. 3. Verify the following (high) peak pressure alarms are activated: a. Paw Too High message appears on the screen. b. An alarm tone sounds. c. Inspiration ends and expiration begins as the pressure meets the high alarm limit. 4. Set the PAW too high alarm limit to 65(cmH2O). 5. Verify the alarms signals cease. 6. Set the PAW too low alarm limit to 50(cmH2O). 7. Verify the following (low) peak pressure alarms are activated: a. Paw Too Low message appears on the screen. b. An alarm tone sounds. 8. Set the PAW too low alarm limit to 12(cmH2O). 9. Verify the alarm signals cease. 2-37 2.8.5 MV Alarm Test 1. Set the MV Too Low alarm limit to the maximum value. 2. Verify the following alarms are activated: MV Too Low message appears on the screen. An alarm tone sounds. 3. Set the MV Too Low alarm limit to the minimum value. 4. Verify the alarm signals cease. 5. Set the MV Too High alarm limit to the minimum value. 6. Verify the following alarms are activated: MV Too High message appears on the screen. An alarm tone sounds. 7. Set the MV Too High alarm limit to the maximum value. 8. Verify that the alarm signals cease. 2.9 Vaporizer Test 2.9.1 Vaporizer Interlock Test 1. Attach two vaporizers to the vaporizer manifold and lock them in place. 2. Rotate one vaporizer dial to 3%. 3. Verify that the other vaporizer dial is disabled. 4. Set both vaporizer dials to 0. 5. Rotate the other vaporizer dial to 3%. 6. Verify that the first vaporizer dial cannot be rotated. 7. Rotate both vaporizer dials to T and remove both vaporizers. 8. Verify that the locking spring is in good condition. 9. Reconnect both vaporizers to the vaporizer manifold. 2-38 2.9.2 Vaporizer Accuracy Test 1. 2. 3. 4. 5. Set the APL valve to 70cmH2O. Put the Auto/Manual switch to the Manual position. Connect one end of the breathing hose to the expiration port and the other end to the bag arm. Connect the sampling tee of the gas analyzer to the inspiration port. Use a breathing hose to connect the output end of the sampling tee to the scavenging system. 6. Verify that the scavenging system is connected to the wall and the float is between MIN and MAX scales. 7. Mount the vaporizers and fill with anesthetic agent (if necessary). NOTE: Do not overfill by filling past the indicator line on the vaporizer. 8. Turn on the equipment. 9. Test the vaporizer accuracy per the manufacturer’s instructions. 10. Test each vaporizer in turn. 11. Test all the vaporizers on the vaporizer storage mount. 12. Remove the measuring equipment. 13. Disconnect the waste gas scavenging hose. NOTE: The vaporizer has deviations due to change of barometric pressure (high altitude). And the Riken F-211 gas analyzer also has deviations. When testing the vaporizers using the Riken F-211 gas analyzer, the altitude can be ignored as the deviations cancel each other out. If using a different gas analyzer, check the effect of change of barometric pressure before using it in high altitude area. 2-39 2.10 Other Tests 2.10.1 Power Failure Alarm Test 1. Interrupt AC line voltage. 2. Verify that the following alarms are activated: An alarm tone sounds. Battery in Use message appears on the screen. 3. Connect the anesthesia system to AC line voltage. 4. Verify that the alarm signals cease. 5. Verify the presence of the battery charging icon in the upper right corner of the screen. 2.10.2 Work Light and Auxiliary Gas Supply Backlight Test 1. 2. 3. 4. Turn on the top-mounted lamps at the bottom of the top plate. Verify that both lamps are turned on. Verify that the work light works at the low brightness level when the switch is set to I, works at the high brightness level when the switch is set to II, and is turned off when the switch is set to O. Verify that the auxiliary gas supply backlight can be lit on when the switch is set to both I and II positions. 2.11 Electric Safety Test NOTE: Perform electrical safety inspection after servicing or routine maintenance. Before the electrical safety inspection, make sure that all the covers, panels, and screws are correctly installed. NOTE: It is recommended to do electrical safety test once a year. 2.11.1 Test on Auxiliary Electrical Outlet Verify that there is mains power voltage on every auxiliary electrical outlet when the anesthesia machine is connected to the mains power supply. 2-40 2.11.2 Electrical Safety Inspection 1. 2. 3. Perform protective earth resistance test: a. Plug the probes of the analyzer into the protective earth terminal and equipotential terminal of the AC power cord. b. Test the earth resistance with a current of 25A. c. Verify the resistance is less than 0.1ohms (100 mohms). d. Plug the probes of the analyzer into the protective earth terminal of the AC power cord and the protective earth terminal of any auxiliary outlet. Repeat steps b and c. e. If the resistance is greater than 0.1ohms (100 mohms) but less than 0.2ohms (200 mohms), disconnect the AC power cord and plug the probe that is previously plugged in the protective earth terminal of the AC power cord into the protective earth contact of the power outlet. Repeat steps a to d. Perform the following earth leakage current tests: normal polarity reverse polarity normal polarity with open neutral reverse polarity with open neutral Verify that the maximum leakage current does not exceed 500 μA (0.5 mA) in the first two tests. While for the last two tests, verify that the maximum leakage current does not exceed 1000 μA (1 mA). NOTE: Make sure the safety analyzer is authorized by certificate organizations (UL, CSA, or AMAI etc.). Follow the instructions of the analyzer manufacturer. 2.11.3 Electrical Safety Inspection Form Location: Technician: Equipment: Control Number: Manufacturer: Model: SN: Measurement equipment /SN: Date of Calibration: Inspection and Testing Pass/F ail 1 2 3 Auxiliary mains socket outlets Protective Earth Resistance Ω Normal condition(NC) ____μA Single Fault condition(SFC) ____μA Limit Max 0.1 Ω Max: NC: 500μA SFC: 1000μA Earth Leakage 2-41 For routine maintenance, all the test items included in the ELECTRICAL SAFETY INSPECTION FORM shall be performed. The following table specifies test items to be performed after the equipment is repaired after disassembled. When power board, transformer, or patient circuit board is not repaired or replaced. When power board or transformer is repaired or replaced. Test items: 1, 2 Test items: 1, 2, 3 2-42 3 Equipment Maintenance 3.1 Repair Policy The following table lists equipment maintenance operations of the A5 anesthesia system. Physical check, consumables replacement, and performance check shall be conducted at regular intervals according to the schedule listed in the following table. The factory will bear no liability if some parts are damaged or lost because the consumables are not replaced within the suggested schedule. The preceding contents are described in Equipment Maintenance Schedule (see section 3.3). 3.2 Periodical Maintenance Kit The periodical maintenance kit contains: Basic maintenance kit (12-month), P/N: 115-036367-00. Expand maintenance kit (12-month), P/N: 115-036368-00. Complete maintenance kit (12-month), P/N : 115-036369-00 Remarks: For details about the differences of maintenance kits (12-month), see section 3.5 "Maintenance Kit (12-month) Replacement". 3.3 Equipment Maintenance Schedule Maintenance operation After each maintenance Preoperative checklist: 1. Check the mechanical ventilation state 2. Breathing system leak test under mechanical ventilation state 3. Breathing system leak test under manual ventilation state 4. Check the correctness of sensor zero point 5. Check the flow sensor measurement accuracy 6. Check the pressure sensor measurement accuracy 7. Check the electronic flowmeter measurement accuracy System check Consumables replacement Functional test Preoperative test 3-1 Every 12 months ● ● ● ● ● ● 3.4 Pre-operation test Cosmetic inspection 1. Make sure that the equipment is in good condition. 2. Make sure that the breathing system and Pre-pak absorbent canister are connected correctly. 3. Make sure that there is appropriate amount of anesthetic agent inside the vaporizer. 4. Make sure that the preoperative checklist is available. 5. Make sure that the cylinder wrench is available. 6. Make sure that the AGSS transfer tube is not damaged. Drain the water build-up. 7. Make sure that the AC power cord is not damaged. 3.5 Maintenance Kit (12-month) Replacement The following table lists the parts to be replaced periodically inside the periodical maintenance kit. The replacement date starts from the date when the machine is assembled. Remark: The materials in the form below are not sold separately. Please contact the engineer if any questions. P/N Material Description Quantity 0348-00-0185 Special seal Sealing O-ring, 15.54X2.62, fluorubber, A70 Pre-pak absorbent canister sealing Sealing O-ring, 16X2, fluorubber A50, black Sealing O-ring, 8.5X2, fluorubber, A70 Sealing O-ring, 4.7X1.8, fluorubber, A70 Pre-pak absorbent canister sealing Bypass trigger board sealing Pop-off rubber cushion AGSS filter Pre-pak absorbent canister tube Bellow sealing (0631) O-ring, 14X2.65 Sealing O-ring, 27X1.5, fluorubber, A50, transparent Sealing O-ring, 20X1.5 3 Basic Mainten ance Kit (12Month) Replace 3 Replace Replace Replace 1 Replace Replace Replace 2 Replace Replace Replace 2 Replace Replace Replace 4 Replace Replace Replace 1 Replace Replace Replace 1 1 1 Replace Replace Replace Replace Replace Replace Replace Replace Replace 1 Replace Replace Replace 1 4 Replace Replace Replace Replace Replace Replace 2 Replace Replace Replace 2 Replace Replace Replace 082-000673-00 049-000143-00 M6M-010058--082-000665-00 082-000667-00 049-000145-00 049-000142-00 049-000240-00 082-000506-00 049-000146-00 049-000243-00 082-000934-00 082-001501-00 082-001503-00 3-2 Expand Mainten ance Kit (12Month) Replace Complete Maintenan ce kit (12Month) Replace fluorubber, A50 transparent Sealing O-ring, 8.5X2, 082-001525-00 1 Replace Replace Replace fluorubber, A50, transparent Sealing O-ring, 23.47X2.95, 082-001504-00 3 Replace Replace Replace fluorubber, A50, transparent Sealing O-ring, 10.72X1.78, 082-001524-00 1 Replace Replace Replace fluorubber A50, transparent Filter 0.45 um, 082-000423-00 1 Replace Replace Replace TP026AATV004AD01 Sealing O-ring, 20.29X2.62, 082-001508-00 1 Replace Replace Replace fluorubber, A50, transparent 045-000241-00 Air filter (0631) 1 Replace Replace Replace Folding bag, neoprene, 040-000358-00 1 / Replace Replace black, 8 folds 0611-10-45654 O2 sensor, MedicelMOX-2 1 / / Replace Expiratory flow sensor 115-008263-00 1 / / Replace assembly Inspiratory flow sensor 115-008262-00 1 / / Replace assembly Replace the maintenance materials inside the maintenance kit. Replacement may not be required if the machine is not configured. (1) Drive gas and AGSS outlet sealing ring (M6M-010058---), quantity: 2 Drive gas and AGSS outlet sealing ring (2) Fresh gas inlet and ACGO sealing ring (082-000665-00), quantity: 2 Fresh gas inlet and ACGO sealing ring 3-3 (3) Sampled gas interface sealing ring (082-000667-00), located inside the interface, quantity: 4. Sampled gas interface sealing ring (4) Bellow cover sealing (049-000243-00), quantity: 1; folding bag (040-000358-00), quantity: 1 Folding bag Bellow cover sealing (5-1) Manual support arm assembly sealing ring (082-000673-00), quantity: 2 Manual support arm assembly sealing ring (5-2) O2 battery sealing ring (082-000673-00), quantity: 1 O2 battery sealing ring 3-4 (6) Airway pressure gauge (082-001524-00), quantity: 1 Airway pressure gauge (7) POP-OFF valve, quantity: 1 POP-OFF valve (8) Folding bag base sealing ring (082-001508-00), quantity: 1 Folding bag base sealing ring 3-5 (9) Valve cover sealing ring (082-001501-00), quantity: 2 Valve cover sealing ring (10) Valve sealing ring (082-001503-00), quantity: 2 Valve sealing ring (11) O2 battery port sealing ring (082-001525-00), located on the lower cover assembly, requiring patient circuit dismantling, quantity: 1 O2 battery port sealing ring 3-6 (12) External absorbent canister sealing (049-000143-00), quantity: 1 External absorbent canister sealing (13) Internal absorbent canister sealing (049-000145-00) (cut the redundant part after replacement), quantity: 1 Internal absorbent canister sealing (14) Bypass trigger board sealing (049-000142-00), quantity: 1 Bypass trigger board sealing 3-7 (15-1) Bypass sealing ring (082-001504-00), quantity: 2 Bypass sealing ring (15-2) Water collection cup sealing ring (082-001504-00), quantity: 1 Water collection cup sealing ring (16) Absorbent canister tube (049-000146-00), quantity: 1 Absorbent canister tube 3-8 (17) Vaporizer support sealing ring (082-000934-00) (After the sealing rings are replaced, install the vaporizer vertically in the horizontal direction. Fasten the vaporizer when it is in position.), quantity: 4 Vaporizer support sealing ring (18) Cylinder support sealing (0348-00-0185), quantity: 3 Cylinder support sealing (19) Filter (082-000423-00) (Internal AG module filter. After replacement, fix the filter with 2.5 mm x 10 mm ties.), quantity: 1 Filter 3-9 (20) AGSS filter (082-000506-00), quantity: 1 AGSS filter (21) Air filter (045-000241-00), quantity: 1 Air filter (22) Expiratory flow sensor assembly (115-008263-00), quantity: 1; inspiratory flow sensor assembly (115-008262-00), quantity : 1 Expiratory flow sensor assembly Inspiratory flow sensor assembly 3-10 (23) O2 sensor MedicelMOX-2 (0611-10-45654), quantity: 1 MOX-2 O2 sensor 3.6 Functional Tests Refer to 4 Calibration if any values are out of specification. NOTE: The anesthesia system (use AC power instead of battery) must be turned on for at least one hour before performing ventilation test. 3.6.1 Gas Delivery System Tests 3.6.1.1 O2 Flush Verification 1. Connect the bag arm to the expiratory port via a breathing hose. 2. Set the Auto/Manual switch to the Manual position. Set the APL valve to 75. 3. Connect the flowmeter to the inspiratory port. 4. Verify that the O2 flush flow is between 35 and 50 L/min when the O2 flush valve is pressed. 3.6.1.2 O2:N2O Ratio System 1. Set the O2 and N2O flow control valves to minimum. 2. Rotate the O2 flow control valve until the flowmeter displays 1L/min. 3. Rotate the N2O flow control valve and adjust N2O flow to maximum. 4. Verify that the N2O flow is not greater than 3.7L/min. 5. Decrease the N2O flow to minimum. 6. Decrease the O2 flow to minimum. 3.6.2 Breathing System Leak Test NOTE: Perform leak test again each time after servicing the anesthesia machine, replacing the components, or re-connecting the tubes. 3-11 3.6.2.1 Automatic Circuit Leak and Compliance Test This test checks if the pneumatic circuit has leakage in mechanical ventilation mode. The test items include bellows, drive gas circuit, CO2 absorber canister, patient tube, flow sensor, and flow sensor connector. The following figure shows the Automatic Circuit Leak & Compliance Test screen. To perform an automatic circuit leak test, NOTE: The system records the result of the last automatic circuit leak test in the General tab, including if the test result is passed or failed, or the test is skipped. To view the last test result, select [Main] > [General] on the main screen. NOTE: If fresh gas is detected by the system before undergoing the automatic circuit leak and compliance test, a message is displayed on the screen for zeroing all flowmeters. 1. From system power-up: If the system is being powered on, the system automatically initiates a system self test and then enters the Automatic Circuit Leak Test screen. Or From the main screen: Select Main > General > Test Leak/Compliance. 2. Follow the directions on the screen. NOTE: The Continue button can be selected only when the Auto/Manual switch is set to the Auto position and no fresh gas is detected. 3-12 NOTE: The result of automatic circuit leak and compliance test is recorded in Service Log. To view the test result, select [Main] > [Service] > [Review Logs] > [Service] on the main screen. 3.6.2.2 Manual Circuit Leak Test This test checks if the pneumatic circuit has leakage in manual ventilation mode. The test items include APL valve, check valve, CO2 absorber canister, patient tube, flow sensor, and flow sensor connector. The following figure shows the Manual Circuit Leak Test screen. To perform a manual circuit leak test, NOTE: The Continue button can be selected only when the Auto/Manual switch is set to the Manual position and no fresh gas is detected. 1. From system power-up: If the system is being powered on, the system automatically initiates a system self test followed by automatic circuit leak and compliance tests. Or From the main screen: Select [Main] > [General] > [Test Leak/Compliance]. 2. Follow the directions on the screen. 3-13 NOTE: The Continue button can be selected only when the Auto/Manual switch is set to the Manual position and no fresh gas is detected. NOTE: The result of manual circuit leak test is recorded in Service Log. To view the test result, select [Main] > [Service] > [Review Logs] > [Service] on the main screen. NOTE: If there is a leak, check the pneumatic circuit system for leakage and troubleshoot the problems as described in 5.3 Pneumatic System Problems. Do leak test again after failure source is fixed, NOTE: If there is a leak, check the pneumatic circuit system for leakage and troubleshoot the problems as described in 5.3.4 Breathing System. Do leak test again after leak failure is fixed. 3.6.2.3 Commonly-encountered Problems and Recommended Actions The following table lists the commonly-encountered problems and recommends actions for 3.6.2.1. Failure description Possible cause Recommended action Leak test failure is prompted immediately after [Begin] is selected (typically, the leak test requires at least 3 minutes). The Auto/Manual switch is set to the Manual position and the message [Manual Vent.] is prompted. The reading on the drive gas (O2) pressure gauge indicates drive gas pressure low (lower than 200kPa) and the alarm of [Drive Gas Pressure Low] is produced. During leak test, the pressure indicated by the airway pressure gauge fails to reach 30cmH2O. 1. Before the leak test, the bellows is not fully inflated. 2. The Y piece on the breathing tube is not connected to the test plug. 3. The bellows housing is not properly installed. Set the Auto/Manual switch to the Auto position Replace or connect gas supplies and make sure that the drive gas pressure is at 350 to 450kPa. Check the pneumatic circuit connections and re-install the pneumatic circuit. 3-14 3.6.3 Check the Sensor Zero Point To check the sensor zero point: 1. Turn off all fresh gases and position the Y piece in the patient circuit to the air. 2. Make sure that the system is in Standby mode. 3. Select Main -> Service-> Data Monitors-> Component-> Zero Sensor to access the following menu. 4. The second column is the current sensor zero point and the third column is the zero point saved before delivery. Zero Point Range of DSP Plateau: The following table lists the normal range of the zero point of some of A5 pressure and flow sensors. 3-15 Sensor name Airway pressure sensor PEEP pressure sensor Inspiratory flow sensor Expiratory flow sensor Internal flow sensor Normal range of zero point (AD Counts) 7432~16206 7432~16206 554~26457 554~26457 554~26457 The zero point A/D value of the airway pressure sensor and PEEP pressure sensor should fall within the normal range of 7432 to 16206. The zero point A/D value of internal flow sensor, the inspiratory flow sensor and expiratory flow sensor should fall within the normal range of 554 to 26457. If there is a great deviation between the current zero point and the zero point in case of factory calibration, it indicates that the sensor is ageing but it does not mean that normal measurement cannot be performed. If the current zero point is found to exceed the specified normal range, normal measurement is affected and you need to calibrate the zero point again. If the zero point of the flow sensor is not within 554 to 26457 and that of the pressure sensor not within 7432 to 16206, replacing the VCM. For internal flow sensor, replace the flow sensor interface board. Zero Point Range of EPSON Plateau: The following table lists the normal range of the zero point of some of A5 pressure and flow sensors. Sensor name Airway pressure sensor PEEP pressure sensor Inspiratory flow sensor Expiratory flow sensor Internal flow sensor Normal range of zero point (AD Counts) 200~800 200~800 50~1800 50~1800 100~400 The zero point A/D value of the airway pressure sensor and PEEP pressure sensor should fall within the normal range of 200 to 800. The zero point A/D value of the inspiratory flow sensor and expiratory flow sensor should fall within the normal range of 50 to 1800. The zero point A/D value of the internal flow sensor should fall within the normal range of 100 to 400. 3-16 If there is a great deviation between the current zero point and the zero point in case of factory calibration, it indicates that the sensor is ageing but it does not mean that normal measurement cannot be performed. If the current zero point is found to exceed the specified normal range, normal measurement is affected and you need to calibrate the zero point again. If the zero point of the flow sensor is not within 0 to 2000 and that of the pressure sensor not within 0 to 1200, replacing the VCM. For internal flow sensor, replace the flow sensor interface board. NOTE: If the zero point of the pressure sensor has an error, in ventilation status, the baseline of the Paw waveform is not at the zero point and a great deviation occurs between pressure control and measurement. NOTE: If the zero point of the inspiratory/expiratory flow sensor has an error, in ventilation status, the baseline of the flow waveform is not at the zero point and a great deviation occurs between TV control and measurement. NOTE: If the zero point A/D value of any sensor is outside of the normal range, it cannot be corrected. The valve drive board must be replaced. 3.6.4 Check the Accuracy of Flow Sensors NOTE: If a great deviation of Vt measured value occurs, test the measurement accuracy of flow sensors to determine whether to perform flow calibration again. NOTE: You can use any flowmeter that has an accuracy of at least ±2% for the accuracy measurement of flow sensors. To check the measurement accuracy of flow sensors: 1. Remove the bellows and water collection cup. 2 The pneumatic connections between the anesthesia machine and calibration device are as shown below. You can connect the tube to the high-flow connector or low-flow connector based on the requirement. High-flow connector Low-flow connector Calibration Device (VT Plus) 3-17 3. Set up the calibration device as described below. a. Flow Setting: Press the Flow button on the front control panel of the calibration device. You can set Range to High Flow or Low Flow as required. b. Gas Setting: Press the Setup button, select Setting->ENTER->Gas Settings>MODIFY->Gas Type->O2. 3-18 c. Select BACK→BACK→BACK. 4. When the system is Standby, select Main -> Service-> Diagnostic Tests->Valves to access the following menu. 5. Set safety valve to ON. 6. Set the PEEP valve pressure to 30cmH2O. 7. Set the Insp Valve Flow to the following values: (3±0.5)L/min, (10±1)L/min, (20±1)L/min, (30±2)L/min, (60±3)L/min. 3-19 8. Make sure that the deviation between the measured data of inspiratory flow sensor, expiratory flow sensor, and internal flow sensor and that of the anesthesia machine calibration device cannot exceed 1L/min or 5% of the measured value of the calibration device, whichever is greater. Otherwise, refer to 4.3.2 Flow Calibration (Factory) to perform flow calibration again. When testing the flow accuracy of (60±3)L/min, only compare the accuracy with the inspiratory flow sensor and expiratory flow sensor. It is no need to compare with the internal flow sensor. After these steps are completed, re-install the bellows and water collection cup. 3.6.5 Check the Accuracy of Pressure Sensors NOTE: Generally, measurement deviations do not easily occur to pressure sensors. However, in case of replacing the valve drive board, solenoid valve assembly, or expiration valve assembly, you need to perform pressure calibration and check the flow sensor accuracy so as to confirm the effectiveness of calibration. NOTE: You can use any flowmeter that has an accuracy of at least ±2% for the accuracy measurement of pressure sensors. To check the measurement accuracy of pressure sensors: 1. Perform pneumatic connections as follows: The following picture shows a four-way device. A four-way device connecting the sampling lines for pressure calibration 3-20 Remove the top cover (3 captive screws). Remove tube #72 and #9 from the pressure sensor (as shown below). Connect the four-way tubes to pressure sensor P1 of the monitor board, PEEP pressure sensor P2, tube marked as #72, and low pressure port of Fluke VT-Plus. Do not connect tube marked as #9. A D C B VT Plus Low Press + Calibration Device (VT Plus) 3-21 2. Set up the calibration device as described below. a. Pressure Setting: Press the PRESSURE button on the front control panel of the calibration device, select Range and then set it to Low Press. 3-22 b. Gas Setting: Press the Setup button, select Setting->ENTER->Gas Settings->MODIFY>Gas Type->O2. c. Select BACK→BACK→BACK. 3-23 3. When the system is Standby, select Main ->Service-> Diagnostic Tests-> Valves to access the following menu. 4. Set safety valve to ON. 5. Set the PEEP valve pressure to the following values: (5±1)cmH2O, (20±1)cmH2O, (50±1)cmH2O (70±2)cmH2O, (90±2)cmH2O. 6. Make sure that the deviation between the measured data of the airway pressure sensor, and PEEP pressure sensor and that of the anesthesia machine calibration device cannot exceed 1cmH2O or 2% of the measured value of the calibration device, whichever is greater. Otherwise, refer to 4.3.3 Pressure Calibration (Factory) to perform pressure calibration again. 3-24 3.7 High Pressure Tests Close the three needle valves and turn all fresh gas flows to 0 before doing the following tests. 3.7.1 N2O Cylinder Leak Test 1. Remove the N2O pipeline supply from A5 gas supply inlet. 2. Make sure that the N2O cylinder has sufficient pressure. Mount the cylinder onto the cylinder yoke at the rear panel. If necessary, place a new clean sealing cushion between the cylinder and the yoke to minimize any leak at the yoke connection. 3. Open the N2O cylinder valve. Read and record the value on the cylinder pressure gauge. 4. Close the N2O cylinder. 5. Read and record the value on the cylinder pressure gauge one minute later. The pressure value displayed on the N2O cylinder pressure gauge cannot decrease by more than 10% of the initial value. 3.7.2 O2 Cylinder Leak Test 1. Remove the O2 pipeline supply from A5 gas supply inlet. 2. Make sure that the O2 cylinder has sufficient pressure. Mount the cylinder onto the cylinder yoke at the rear panel. If necessary, place a new clean sealing cushion between the cylinder and the yoke to minimize any leak at the yoke connection. 3. Open the O2 cylinder valve. Read and record the value on the cylinder pressure gauge. 4. Close the O2 cylinder. 5. Read and record the value on the cylinder pressure gauge one minute later. The pressure value displayed on the O2 cylinder pressure gauge cannot decrease by more than 10% of the initial value. 3-25 3.7.3 AIR Cylinder Leak Test 1. Remove the AIR pipeline supply from A5 gas supply inlet. 2. Make sure that the AIR cylinder has sufficient pressure. Mount the cylinder onto the cylinder yoke at the rear panel. If necessary, place a new clean sealing cushion between the cylinder and the yoke to minimize any leak at the yoke connection. 3. Open the AIR cylinder valve. Read and record the value on the cylinder pressure gauge. 4. Close the AIR cylinder. 5. Read and record the value on the cylinder pressure gauge one minute later. The pressure value displayed on the AIR cylinder pressure gauge cannot decrease by more than 10% of the initial value. 3.7.4 N2O Pipeline Supply Test NOTE Remove the N2O cylinder from the N2O cylinder yoke when doing this test. 1. Connect the machine to the N2O supply connector and O2 supply connector on the wall via N2O supply hose and O2 supply hose. 2. Check that the N2O supply pressure gauge displays 280~600Kpa. 3. Turn on the system switch. 4. Open the O2 needle valve. Adjust the flow to 1L/min to fully open the N2O needle valve. 5. Observe the float of total flowmeter, which should fall between 3L and 5L. 6. Disconnect the N2O pipeline supply. 7. Check that the N2O supply pressure gauge decreases to zero. 3-26 3.7.5 O2 Pipeline Supply Test NOTE Remove the O2 cylinder from the O2 cylinder yoke when doing this test. 1. Connect the machine to the O2 supply connector on the wall via O2 supply hose. 2. Check that the O2 supply pressure gauge displays 280~600Kpa. 3. Turn on the system switch. 4. Fully open the O2 needle valve. 5. Observe the float of total flowmeter, which should be at the top of the flowmeter. 6. Disconnect the O2 pipeline supply. 7. Make sure that the alarms of [O2 Supply Failure] and [Drive Gas Pressure Low] occur with the decrease of O2 pressure. 8. Check that the O2 supply pressure gauge decreases to zero. 3.7.6 AIR Pipeline Supply Test NOTE Remove the Air cylinder from the Air cylinder yoke when doing this test. 1. Connect the machine to the Air supply connector on the wall via Air supply hose. 2. Check that the Air supply pressure gauge displays 280~600Kpa. 3. Turn on the system switch. 4. Fully open the Air needle valve. 5. Observe the float of total flowmeter, which should be at the top of the flowmeter. 6. Disconnect the Air pipeline supply. 7. Check that the Air supply pressure gauge decreases to zero. 3-27 3.8 Breathing System Tests 3.8.1 Sample Gas Emission Test/AGSS Test (optional configuration) 1. Connect one end of the sample gas exhaust emission hose to the port on the waste gas disposal system. Connect the other end of the hose to the EVAC port on the module. NOTE: If other type of waste gas scavenging system is used for A5, make sure that sample waste gas is exhausted from the EVAC port to enter the scavenging system. 2. Connect the module waste gas exhaust port to the fitting port on the waste gas disposal system. 3. Cover all openings on the waste gas disposal system. 4. Make sure that the flow of waste gas disposal system can be adjusted between the minimum and maximum scales. 3.8.2 Internal Pneumatic Connection Test 1. Close and remove all gas cylinders from the anesthesia system. 2. Connect only the AIR pipeline supply hose to the anesthesia system. Disconnect all the other gas supply hoses. 3. Connect A5 to the power supply. Rotate the AIR flow control knob to ensure a continuous flow increase throughout its full range 4. Fully rotate the N2O flow control knob and verify that there is no N2O flow. 5. Fully rotate the O2 flow control knob and verify that there is no O2 flow. 6. Disconnect the AIR pipeline supply hose from the anesthesia system, and connect the O2 pipeline supply hose. Rotate the O2 flow control knob to ensure a continuous flow increase throughout its full range. 7. Fully rotate the N2O flow control knob and verify that there is no N2O flow. 8. Fully rotate the AIR flow control knob and verify that there is no AIR flow. 9. Connect the N2O pipeline supply hose. With the O2 flow control knob not fully opened, rotate the N2O flow control knob to ensure a continuous flow increase throughout its full range. NOTE: The maximum N2O flow must be between 10 and 12 L/min. 10. Fully rotate the AIR flow control knob and verify that there is no AIR flow. 3-28 3.8.3 Alarm of Drive Gas Pressure Low, N2O Cutoff Test 1. Set the O2 flow to 2 L/min using the flow control valve. 2. Set the N2O flow to 2 L/min using the flow control valve. 3. Set the AIR flow to 2 L/min using the flow control valve. 4. Interrupt the O2 supply for the A5. 5. Verify that the N2O and O2 stop flowing within 2 minutes and that AIR (if available) continues to flow at 2 L/min. 6. Verify the following alarms are activated: O2 Supply Failure appears on the screen An alarm tone sounds. 3.9 Performance Verification 3.9.1 Standby Mode Ventilation Test 1. Ensure that the gas pressures of O2, N2O, and AIR are within the range of specifications. 2. Power on the anesthesia system. 3. Perform the startup test as per the on-screen instructions. Ensure successful completion. 4. Set the Auto/Manual switch to the Manual position. 5. Attach a breathing circuit and test lung to the Y piece of the breathing circuit. NOTE: For testing purpose always use a reusable breathing circuit. 6. Set the APL valve to approximately 15cmH2O. 7. Set the AIR flow to 5L/min using the flow control valve. Use O2 if AIR is not available. 8. Squeeze the breathing bag once every 10 seconds to inflate and deflate the test lung so as to produce pressure of approximately 20cmH2O. 9. Verify the inflation and deflation of the test lung. 3-29 3.9.2 Manual Mode Ventilation Test 1. Set the Auto/Manual switch to the Manual position to switch to manual mode 2. Set the APL valve to approximately 25cmH2O. Push the O2 flush button to fill the breathing bag. 3. Set the AIR flow to 1L/min using the flow control valve. 4. Squeeze the breathing bag once every 3 seconds. 5. Verify the inflation and deflation of the test lung. 6. Verify that an airway pressure waveform and all numeric values appear on screen during bag compressions. 7. Stop squeezing the breathing bag and set the APL Valve to the open position (SP). 3.9.3 Apnea Alarm Test 1. While in the manual ventilation mode, stop ventilating the test lung. 2. Verify that the following apnea alarm signals are activated at approximately 30 seconds from the last bag compression (for enterprise standard, it is 20 seconds for SFDA machine). Apnea appears on the screen. An alarm tone sounds. 3.9.4 Alarm Silence Test 1. While the apnea alarm is sounding, press the Silence soft key. 2. Make sure that the audio alarm stops and resumes after two minutes. 3.9.5 VCV Adult Ventilation Mode Test 1. Attach a breathing circuit and breathing bag. NOTE: For testing purpose always use a reusable breathing circuit. 2. Attach an adult test lung to the Y piece of the breathing circuit. 3. Attach a ventilation test device between the expiration port and the expiratory hose. 4. Set the O2 flow to 2L/min and set the N2O and AIR flows to the minimum. 5. Set the Auto/Manual switch to the Auto position. 3-30 6. Set the ventilating control parameters to: Ventilation control parameter Setting of ventilation control parameter Patient type Vent mode Vt Rate I:E Tpause PEEP Plimit Adult VCV 600 8 1:2 10 Off 50 7. Press Set Mode button to begin ventilation. 8. Verify that pressure waveform, Vt, MEAN or PLAT, Rate and MV appear on the screen. 9. Verify the tidal volume displayed on the ventilation test device is within 7% (±42mL) of the setting value within approximately one minute from the start of ventilation. 10. Verify the tidal volume displayed is within 9% (±54mL) of the setting value within approximately one minute from the start of ventilation. 11. Verify the measured O2 concentration is at least 97% after 5 minutes. 12. Set the AIR flow to 3L/min and set the N2O and O2 flows to the minimum. 13. Verify the measured O2 concentration is 21% ±3% vol. % after 5 minutes. 3.9.6 VCV Pediatric Ventilation Mode Test 1. Attach a breathing circuit and breathing bag. NOTE: For testing purpose always use a reusable breathing circuit. 2. Attach a pediatric test lung to the Y piece of the breathing circuit. NOTE: Limit the volume in the test lung to provide sufficient airway pressure to satisfy the Low Peak Pressure alarm. Or reduce the Peak Pressure alarm limit to a lower value to prevent the alarm when using an adult test lung. 3. Attach a ventilation test device between the expiration port and the expiratory hose. 4. Set the O2 flow to 2L/min and set the N2O and AIR flows to the minimum. 3-31 5. Set the ventilating control parameters to: Ventilation control parameters Patient type Vent mode Vt Rate I:E Tpause PEEP Plimit Setting of ventilation control parameter Pediatric VCV 120 15 1:2 10 Off 40 6. Press Set Mode button to begin ventilation. 7. Verify that pressure waveform, Vt, MEAN or PLAT, Rate and MV appear on the screen. 8. Verify the tidal volume displayed is within 18ml of the setting value within approximately one minute from the start of ventilation. 9. Verify the delivered volume as measured by ventilation test device at the expiration port, is within 15ml of the setting value within approximately one minute from the start of ventilation. 3.9.7 Airway Disconnection Alarm Test 1. While the ventilator is running, disconnect the expiratory limb from the expiration port on the breathing system. 2. Verify the following airway pressure disconnection alarm signals are activated: Paw Too Low message appears on the screen. An alarm tone sounds. 3.9.8 PCV Adult Ventilation Mode Test 1. Attach a breathing circuit and breathing bag. NOTE: For testing purpose always use a reusable breathing circuit. 2. Attach an adult test lung to the Y piece of the breathing circuit. 3. Attach a ventilation test device between the expiration port and the expiratory hose. 4. Set the O2 flow to 3L/min and set the N2O and AIR flows to the minimum. 5. Set the ventilator controls to: Ventilator Controls Ventilator Settings Patient type Vent mode Adult PCV 3-32 VtG Pinsp Rate I:E PEEP Tslope PlimVG Off 15 8 1:2 Off 0.2 NA 6. Press Set Mode button to begin ventilation. 7. Verify the Peak Pressure reading on the display is within ±2cmH2O of the set Pinsp. 8. Verify that the pressure waveform, Vt, MEAN or PLAT, Rate and MV appear on the screen. 9. Verify that the PEAK value measured by the ventilation test device reaches 15±2.5cmH2O within five breaths from the start of ventilation. 3.9.9 Pressure Support (PS) Ventilation Mode Test 1. Attach a breathing circuit and breathing bag. NOTE: For testing purpose always use a reusable breathing circuit. 2. Attach an adult test lung to the Y piece of the breathing circuit. 3. Attach a ventilation test device between the expiration port and the expiratory hose. 4. Set the O2 flow to 1L/min and set the N2O and AIR flows to the minimum. 5. Set the ventilator controls to: Ventilator Controls Ventilator Settings Patient type Vent mode Min Rate ΔP Trigger PEEP Tslope Adult PS 4 20 3 Off 0.2 6. Press Set Mode button to begin ventilation. 7. Begin triggering breaths by slightly squeezing the test lung and releasing. Maintain a continuous breath rate. 8. Verify that a pressure waveform and all ventilation parameters appear on the screen. 9. Verify that the Peak Pressure reading on the display is ±2 of ΔP+PEEP. 10. Stop triggering breaths. 11. Verify that after 15 seconds the ventilator delivers a breath and displays the message Apnea Ventilation. 12. Verify the system ventilates at a frequency of 4bpm. 3-33 3.10 Alarm and Automatic Fault Protection Function 3.10.1 Setup 1. Ensure that the pressures of O2, N2O, and AIR are within the specified range. 2. Power on the anesthesia system. 3. Perform the Startup Tests per the on-screen instructions. Ensure successful completion. 4. Attach a breathing circuit and breathing bag. NOTE: For testing purposes always use a reusable breathing circuit. 5. Attach an adult test lung to the Y-fitting of the breathing circuit. 6. Set the O2 flow to 2 L/min and set the N2O and AIR flows to the minimum. 7. Set the ventilation control parameters to: Ventilation control parameters Setting of ventilation control parameter Patient Type Ventilation Mode Vt Rate I:E Tpause PEEP Plimit Adult VCV 600 8 1:2 10 Off 50 8. Press Set Mode button to begin ventilation. 3.10.2 FiO2 Too Low Alarm Test 1. Set the FiO2 Too Low alarm limit to 50%. 2. Set the AIR flow control valve to 5L/min. 3. Set the O2 flow control to the minimum. 4. Verify the following FiO2 too low alarm signals are activated within three ventilation cycles: FiO2 Too Low message appears on the screen. An alarm tone sounds. 5. Set FiO2 Too Low alarm limit to 18%. 6. Verify the alarm signals cease. 3-34 3.10.3 FiO2 Too High Alarm Test 1. Set the FiO2 Too High alarm limit to 50%. 2. Set the O2 flow control valve to 5L/min. 3. Set the AIR flow control to the minimum. 4. Verify the following FiO2 Too High alarm signals are activated: FiO2 Too High message appears on the screen. An alarm tone sounds. 5. Set the FiO2 Too High alarm limit to the maximum value. 6. Verify the alarm signals cease. 3.10.4 Peak Pressure Alarm Test 1. Set the PEAK too low alarm to the minimum value. 2. Set the PEAK too high alarm limit to about 5 to 8 digits below the Peak Pressure displayed on the screen. 3. Verify the following (high) peak pressure alarms are activated: Paw Too High message appears on the screen. An alarm tone sounds. Inspiration ends and expiration begins as the pressure meets the high alarm limit. 4. Set the PAW too high alarm limit to 65(cmH2O). 5. Verify the alarms signals cease. 6. Set the PAW too low alarm limit to 50(cmH2O). 7. Verify the following (low) peak pressure alarms are activated: Paw Too Low message appears on the screen. An alarm tone sounds. 8. Set the PAW too low alarm limit to 12(cmH2O). 9. Verify the alarm signals cease. 3-35 3.10.5 MV Alarm Test 1. Set the MV Too Low alarm limit to the maximum value. 2. Verify the following alarms are activated: MV Too Low message appears on the screen. An alarm tone sounds. 3. Set the MV Too Low alarm limit to the minimum value. 4. Verify the alarm signals cease. 5. Set the MV Too High alarm limit to the minimum value. 6. Verify the following alarms are activated: MV Too High message appears on the screen. An alarm tone sounds. 7. Set the MV Too High alarm limit to the maximum value. 8. Verify that the alarm signals cease. 3.11 Vaporizer Test 3.11.1 Vaporizer Interlock Test 1. Attach two vaporizers to the vaporizer manifold and lock them in place. 2. Rotate one vaporizer dial to 3%. 3. Verify that the other vaporizer dial is disabled. 4. Set both vaporizer dials to 0. 5. Rotate the other vaporizer dial to 3%. 6. Verify that the first vaporizer dial cannot be rotated. 7. Rotate both vaporizer dials to T and remove both vaporizers. 8. Verify that the locking spring is in good condition. 9. Reconnect both vaporizers to the vaporizer manifold. 3-36 3.11.2 Vaporizer Accuracy Test 1. Set the APL valve to 70cmH2O. 2. Put the Auto/Manual switch to the Manual position. 3. Connect one end of the breathing hose to the expiration port and the other end to the bag arm. 4. Connect the sampling tee of the gas analysis device to the inspiration port. 5. Use a breathing hose to connect the output end of the sampling tee to the scavenging system. 6. Verify that the scavenging system is connected to the wall and the float is between MIN and MAX scales. 7. Mount the vaporizers and fill with anesthetic agent (if necessary). NOTE: Do not overfill by filling past the indicator line on the vaporizer. 8. Turn on the equipment. 9. Test the vaporizer accuracy per the manufacturer’s instructions. 10. Test each vaporizer in turn. 11. Test all the vaporizers on the vaporizer storage mount. 12. Remove the measuring equipment. 13. Disconnect the waste gas scavenging hose. NOTE: The vaporizer has deviations due to change of barometric pressure (high altitude). And the Riken F-211 gas analyzer also has deviations. When testing the vaporizers using the Riken F-211 gas analyzer, the altitude can be ignored as the deviations cancel each other out. If using a different gas analyzer, check the effect of change of barometric pressure before using it in high altitude area. 3-37 3.11.3 Vaporizer Leak Test 1. Set the bag/mechanical ventilation switch into bag ventilation switch. 2. Set ACGO to patient circuit (if ACGO is configured). 3. Set APL valve to SP. 4. Connect the bag bellows to the bag arm port. 5. Connect the Vent Y-piece of the respiratory circuit to the leak test port and connect another two ports to the expiration port and the inspiration port of the circuit. Note: When the AG module is configured, the sampling line of the module sample gas should not be connected to the Y-piece sampling port and ensure that the Y-piece sampling port is blocked. 6. Install the vaporizer and lock it in the vaporizer manifold. (In order to ensure the accuracy of the test, Some vaporizer needs to be set at 1%. Please refer to the factory’s guide line for specific situation) 7. Set the fresh air speed at 0.2 L/min. 8. Set the APL valve to 75 cm H2O. 9. Press the flushing O2 button till the airway pressure gauge falls at about 30 cm H2O. 10. Loose the flushing O2 button and observe the airway pressure gauge to make sure the numerical reading would not reduce. 3-38 11. Turn off the vaporizer 12. Repeat step 6 to step 11 on the other vaporizer. 3.13 Other Tests 3.13.1 Power Failure Test 1. Disconnect the power supply when the machine is under power-on state. 2. Verify that the following alarms are activated: 3. An alarm tone sounds. Battery in Use message appears on the screen. Re-connect the power supply. 4. Verify that the alarm signals cease. 5. Verify the presence of the battery charging icon in the upper right corner of the screen. 1.13.2 Work Light and Auxiliary Gas Supply Backlight Test 1. Turn on the top light at the bottom of the top plate. 2. Verify that this lamp can display two level of brightness normally. 3. Verify that the work light works at the low brightness level when the switch is set to I, works at the high brightness level when the switch is set to II, and is turned off when the switch is set to O. 4. Verify that the auxiliary gas supply backlight can be lit on when the switch is set to both I and II positions. 3.14 Auxiliary AC Electrical Outlets Test CAUTION: Perform the electrical safety inspection as the last step after completing a repair or after routine maintenance. Perform this inspection with all covers, panels, and screws installed. Verify that AC voltage is present at each AC electrical outlet with a 3A breaker. 3.14.1 Anesthesia System Installation Checklist Refer to 2 Installation Guide for the post-installation check items. Complete each step to check the functions of the anaesthesia machine prior to clinical use. Also, perform the installation checkout procedure after installation, reinstallation, servicing or after any periodic maintenance activity. This checklist cannot replace periodic maintenance activities necessary for maintaining optimal performance. 3-39 3.15 Electric Safety Test NOTE: Perform electrical safety inspection after servicing or routine maintenance. Before the electrical safety inspection, make sure that all the covers, panels, and screws are correctly installed. NOTE: It is recommended to do electrical safety test once a year. 3.15.1 Test on Auxiliary Electrical Outlet Verify that there is mains power voltage on every auxiliary electrical outlet when the anesthesia machine is connected to the mains power supply. 3.15.2 Electrical Safety Inspection CAUTION: It is recommended to do electrical safety test once a year. 1. Perform protective earth resistance test: a. Plug the probes of the analyzer into the protective earth terminal and equipotential terminal of the AC power cord. b. Test the earth resistance with a current of 25A. c. Verify the resistance is less than 0.1ohms (100 mohms). d. Plug the probes of the analyzer into the protective earth terminal of the AC power cord and the protective earth terminal of any auxiliary outlet. Repeat steps b and c. e. If the resistance is greater than 0.1ohms (100 mohms) but less than 0.2ohms (200 mohms), disconnect the AC power cord and plug the probe that is previously plugged in the protective earth terminal of the AC power cord into the protective earth contact of the power outlet. Repeat steps a to d. 2. Perform the following earth leakage current tests: normal polarity reverse polarity normal polarity with open neutral reverse polarity with open neutral 3. Verify that the maximum leakage current does not exceed 500 μA (0.5 mA) in the first two tests. While for the last two tests, verify that the maximum leakage current does not exceed 1000 μA (1 mA). 4. If BIS module is configured, perform the following patient leakage current tests: normal polarity reverse polarity normal polarity with open neutral 3-40 reverse polarity with open neutral normal polarity with open earth reverse polarity with open earth normal polarity with mains on AP reverse polarity with mains on AP 5. Verify that the maximum leakage current does not exceed 100 μA (0.1 mA) in the first two tests, does not exceed 500μA (0.5mA) in the middle 4 tests, and does not exceed 5000μA (5mA) in the last two tests. 6. If BIS module is configured, perform the following patient auxiliary current tests between each electrode and other electrodes. 7. normal polarity reverse polarity normal polarity with open neutral reverse polarity with open neutral normal polarity with open earth reverse polarity with open earth Verify that the maximum leakage current does not exceed 100 μA (0.1 mA) in the first two tests. While for the last two tests, verify that the maximum leakage current does not exceed 500μA (0.5mA). NOTE: Make sure the safety analyzer is authorized by certificate organizations (UL, CSA, or AMAI etc.). Follow the instructions of the analyzer manufacturer. 3-41 3.15.3 Electrical Safety Inspection Form Location: Technician: Equipment: Control Number: Manufacturer: Model: SN: Measurement equipment /SN: Date of Calibration: Inspection and Testing Pass/F ail 1 2 3 4 Auxiliary mains supply output voltage Earthing impedance Ω Maximum: 0.1 Ω Normal condition ____μA Single fault condition ____μA Normal condition ____μA Single fault condition ____μA Maximum: Normal condition: 100μA Single fault condition:500μA ____μA Maximum: 5000μA Normal condition ____μA Single fault condition ____μA Maximum: Normal condition: 100μA Single fault condition:500μA Earth leakage current Patient leakage current Limit 5 Applied part plus mains supply voltage 6 Patient auxiliary leakage current 3-42 Maximum: Normal condition: 500μA Single fault condition: 1000μA For routine maintenance, all the test items included in the ELECTRICAL SAFETY INSPECTION FORM shall be performed. The following table specifies test items to be performed after the equipment is repaired after disassembled. When power board, transformer, or patient circuit board is not repaired or replaced. When power board or transformer is repaired or replaced. When patient circuit board is repaired or replaced. When power board, transformer, and patient circuit board are all repaired or replaced. Test items: 1, 2 Test items: 1, 2, 3 Test items: 1, 2, 4, 5, 6 Test items: 1, 2, 3, 4, 5, 6 3-43 FOR YOUR NOTES 3-44 4 Calibration 4.1 Introduction This section provides detailed information required to test and calibrate the A5 anesthesia system. Calibration consists of making mechanical and electrical adjustments with the test device. The equipment should be tested and calibrated after repairs have been completed or at regular intervals as part of a periodic maintenance procedure. NOTE: Functional test must be performed after calibration to verify correct operation. Before calibration, make sure that all testing materials (including drive gas, breathing circuit, test fixture, tools and documents) are the latest, calibrated and in good condition. 4.2 Warnings, Precautions, and Notes 4.2.1 Warnings WARNING: Replace all fuses with the specified type and rating to prevent fire. WARNING: The machine (protection class I) can be connected only to a correctly grounded power supply (including the socket with grounding contact) to prevent electric shocks. WARNING: Remove all auxiliary devices from the shelf before moving the anesthesia machine on the rough road or any slope. The anesthesia machine with a weight on its top is likely to tip over, causing personal injuries. WARNING: To avoid the possible explosion hazard, do not operate the machine near flammable anaesthetic agents or other flammable substances. Do not use flammable anesthetic agents such as ether or cyclopropane. WARNING: When high-frequency electric surgery equipment is used, anti-static or electrically conductive breathing tubes may cause burns. Therefore, they are recommended at no time. WARNING: The electric shock hazard may exist. The machine can be opened only by authorized service personnel. WARNING: According to IATA and DOT regulations, compressed gases are dangerous. Therefore, the packages or transport packages of dangerous goods must be properly identified, packed, marked, classified, labeled, and documented in accordance with the DOT and IATA regulations. For details, see the International Air Transport Association – Dangerous Goods Regulations or Part 171-180 "Transportation" in Title 49 of the Code of Federal Regulations. 4.2.2 Cautions CAUTION: Refer to Equipment Maintenance Schedule of 3 Equipment Maintenance when performing scheduled periodic maintenance. CAUTION: Do not leave gas cylinder valves open if the pipeline gas supply is in use and the system main switch is at ON position. Otherwise, cylinder supplies could be depleted, resulting in an insufficient reserve supply in the event of pipeline gas supply failure. CAUTION: Use cleaning agent sparingly. Excess fluid could enter the machine, causing damage. CAUTION: This machine can be operated only by trained and skilled medical personnel 4-1 4.2.3 Notes NOTE: Only bacterial filters with a low flow resistance can be connected to the patient module and/or the patient connector. NOTE: Wear surgical gloves when touching or disassembling valves or other internal components of the breathing system. NOTE: Ensure that the gas supply of the machine always complies with technical specifications. NOTE: The APL valve and the PAW gauge marker are for reference only. The anesthesia machine displays the calibrated airway pressure. NOTE: If the machine malfunctions during initial calibration or testing, do not use it until the fault is rectified by a professional service technician. NOTE: After servicing, functional tests, sensor tests, and system tests must be carried out prior to clinical application. NOTE: The machine supports up to two vertical mounting brackets to accommodate additional monitors and other devices. Unauthorized mounting accessories are not recommended. NOTE: Ensure that all devices on the top plate of the machine are secured. 4.3 System Calibration NOTE: The drive gas of the anesthesia machine and the drive gas set in the calibration device must be consistent with the actual drive gas configured for the machine. NOTE: If a test item concerning measurement accuracy fails in the system test, perform relevant calibration. NOTE: Anesthesia machine calibration device (VT) or Fluke VT Plus: the zero point (baseline) of pressure measurements may slightly drift with the temperature and time. Users can perform zeroing manually to eliminate zero offset. In general, zeroing needs to be performed if the monitor displays a non-zero reading when there is no pressure. Perform zeroing on the calibration device before making any measurements. NOTE: You can use anesthesia machine calibration device (VT) or Fluke VT Plus to perform automatic calibration on pressure sensors or flow sensors, or use other calibration devices meeting the accuracy requirement to perform manual calibration. The anesthesia machine provides the functions of monitoring the volume, pressure, inspiratory O2 concentration, CO2 concentration, and gas concentration inside the machine. Great measurement deviations of the monitored values are very likely to be caused by the offset in measured values of relevant measurement components, and calibration needs to be performed. After the machine is serviced, such as replacement of VCM or expiratory valve assembly, the flow sensors or pressure sensors of the machine need to be calibrated. The following table lists the possible calibration items and calibration time. S/N Calibration item Functional description Calibration time 1 Flow calibration (user) Calibrate the flow sensors of the breathing system. 2 Flow calibration (service) Calibrates the flow sensors and inspiration valve of the anesthesia machine. 1. The TV measurement deviation is great (more than 9% compared with the setting value) after the flow sensors in the patient circuit have been used for a long time. 2. The flow sensor in the patient circuit has been replaced. 1. The expiratory valve assembly is replaced. 2. The VCM is replaced. 3. The measured value of the built-in flow sensor deviates from that of the flow measuring device by 5% of the reading or 1 L/min, whichever is larger. 4-2 S/N 3 4 5 6 Calibration item Pressure calibration (service) Functional description Calibrates the pressure sensors and PEEP valve of the anesthesia machine. Pressure and flow zeroing (service) Electronic flowmeter zeroing (user) O2 sensor calibration (user) Calibrates zero offset in the VCM and VPM. Calibrate the deviation from zero point of the fresh flow sensor board. Calibrate the accuracy of O2 sensor at 21% and 100% O2. Calibration time 1. The VCM is replaced. 2. The expiratory valve assembly is replaced. 3. The measured value of the machine's pressure sensor deviates from that of the standard pressure gauge by 5% of the reading or 2cmH2O, whichever is larger. The flow or pressure waveform deviates from the baseline. The electronic flowmeter has a zero point error. The electronic flowmeter still displays flow readings when fresh gases are all turned off. 1. The measured value of the O2 sensor has a great deviation. The deviation exceeds 3% both in Air (21%) and pure O2 (100%). 2. The O2 sensor is replaced. 3. The VCM is replaced. 4.3.1 Flow Calibration (User) NOTE: Flow sensor calibration must be performed again after a flow sensor is replaced. NOTE: The measurement accuracy of flow sensors may be affected by the operating environment, especially when they have been used for a long time, and the tidal volume control may also experience a great deviation correspondingly. After-sales engineers can call users to help them resolve the deviation problem through calibration. NOTE: Before calibration, perform automatic circuit leak test and make sure that the test is passed. NOTE: During calibration, make sure that the drive gas pressure is kept within specifications. Failure to do so may lead to calibration failure. This calibration is only intended for the flow sensors in the breathing circuit. The inspiratory flow sensor and expiratory flow sensor in the breathing system are calibrated through the machine built-in flow measurement base source After the inspiratory flow sensor and expiratory flow sensor have been used for several months, for example, three months after calibration, great deviations (more than 9% compared with the setting value) may occur to tidal volume measurement due to sensor ageing or environmental factors. Or, the user replaces flow sensors. In this case, flow calibration (user) can be adopted. Follow these steps to calibrate flow sensors. 1. Enter the standby mode. 4-3 2. Select [Setup] > [Calibrate] > [Calibrate Flow Sensors] (version 1.03.02 and earlier) or select [Main] > [General]> [Calibrate Flow Sensors] (version 2.01.00 and later) to enter the screen shown below. 3. Follow the instructions on the screen to set the machine and select [Next] to enter the screen shown below. 4. Select [Begin] to calibrate the flow sensor. During the calibration, you can select [Cancel] to abort. 4-4 5. The screen shown below is displayed if the ongoing calibration is aborted. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 6. The screen shown below is displayed if the flow sensor calibration fails. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 7. The screen shown below is displayed upon a successful flow sensor calibration. Select [Done] to exit the calibration screen. NOTE: If measurement deviations are not corrected after multiple flow sensor calibrations, it is recommended that a flow sensor in the circuit be replaced and then a calibration be performed for the new flow sensor. If the problem persists, send the machine to the manufacturer for maintenance. After the problem is fixed, perform the calibration and system tests. 4-5 4.3.2 Flow Calibration (Factory) NOTE: Flow calibration (factory) is necessary in case of replacing the VCM and expiration valve assembly. NOTE: When a great deviation is detected between the measured value of the machine builtin flow sensor and that of the standard flow measurement device, perform flow calibration (factory). This calibration is intended for the flow sensors in the breathing circuit, machine built-in flow sensor, and also inspiration valve. The standard flow measurement device is used to calibrate the flow sensors and inspiration valve. 4.3.2.1 Calibration Procedures NOTE: Make sure that the tubes are not leaking when connected. NOTE: Do not move or press the tubes during calibration. NOTE: When connecting calibration tubes, make sure that gas flows in the correct direction, which is from the inspiration connector of the breathing system, through high-flow inlet of the anesthesia machine calibration device, anesthesia machine calibration device, high-flow outlet of the anesthesia machine calibration device, and to the expiration connector of the breathing system. NOTE: Before calibration, make sure that no sensor or valve related technical alarms are generated. NOTE: During calibration, make sure that the drive gas pressure is sufficient. Failure to do so may lead to calibration failure. NOTE: You can select anesthesia machine calibration device (VT) or Fluke VT Plus for automatic calibration. You can also select flow calibration device which satisfies the accuracy requirement for manual calibration. NOTE: For calibration device with high and low flow channels, flow channel switchover is required during automatic or manual calibration. Manual calibration is available for version 1.03.00 and later. NOTE: When the altitude changes significantly, if the TV precision is out of range,perform the Flow Calibration (Factory). 4-6 Automatic Calibration Follow these steps to calibrate flow sensors. 1. Enter the standby mode. 2. Select [Main] > [Service] > [Calibration] > [Flow Sensors] to enter the screen shown below. 3. Select [Calibrate Automatically] to enter the screen shown below. Select the required calibration device. 4-7 4. Connect the calibration device with the anesthesia machine by using a communication cable. The calibration device can be anesthesia machine calibration device (VT), Fluke VT Plus or FPM. Connectors and settings related to anesthesia machine calibration device (VT) are shown below. Calibration communication connector Power connector High flow outlet port (0 ~ 120 L/min) High-flow inlet port (0 ~ 120 L/min) Connect the communication connector of the anesthesia machine calibration device with the anesthesia machine calibration communication connector using the special communication cable. The anesthesia machine calibration device can communicate with the anesthesia machine through two types of connection: The following picture shows A and B, the two special communication cables for anesthesia machine calibration device. 4-8 Connection 1: Remove the top cover of the anesthesia machine to expose the monitor board. Use communication cable A to connect the calibration communication connector of the calibration device with that of the monitor board, as shown below. Communication cable A for Calibration communication connector of anesthesia machine calibration the monitor board device Connection 2: Use communication cable B to connect the calibration communication connector of the calibration device with that of the anesthesia machine (also the CIS power connector), as shown below. Communication cable B for anesthesia machine calibration device 4-9 Connectors and settings related to Fluke VTPlus are shown below. (1).Connect Fluke VTPlus to the power source. The following pictures show the connectors on Fluke VTPlus. RS232 connection line to the calibration device RS232 connection line to the anesthesia machine Calibration fixture board. Except connection shown here, the fixture board requires additional USB power supply. Calibration device (VT Plus) end RS232 connecting wire 4-10 (2).Connect Fluke VTPlus to anesthesia machine. The related connectors are shown below. RS232 connection line to the anesthesia machine Connection line at the end of anesthesia machine The interfaces and settings of FPM are shown as follows: Anesthesia machine end connection mode is same as Fluke VT Plus. FPM dedicated calibration line FPM interface and connection mode 5 Set the calibration device as follows. Set VT as follows if VT is selected as the calibration device: (1) Start the calibration device to enter the startup screen followed by sensor heating screen (waiting for approximately 5 minutes as required by the prompt message) and then zeroing screen. Press the “ZERO” key on the panel. After the new screen pops up, press the “OK” key on the panel to complete zeroing. (2) Press the “MODE” key on the panel of the calibration device. Select “Calibration Mode” 4-11 from the pop-up menu and then press the “OK” key on the panel to enter calibration screen, as shown below. Set Fluke VT Plus as follows if VT Plus is selected as the calibration device: (1) Gas settings: Press the Setup button, and select Setting->ENTER->Gas Settings>MODIFY->Gas Type->O2->BACK->BACK. (2)Correction mode: Press the Setup button, and select Setting ->ENTER->Correction Mode->BTPS->BACK->BACK. 4-12 (3) Zeroing mode settings: Press the Setup button, and select Setting ->ENTER->Zero Mode>Manual->BACK->BACK. (4) Serial mode settings: Press the Setup button, and select Setting->System->Enter->Serial Mode ->OTIS Ctrl->BACK->BACK. (5) The serial mode screen appears after setting VT Plus is completed. Set FPM as follows if FPM is selected as the calibration device: (1) Start the machine and wait until the sensor is pre-heated. (2)Zero the sensor without connecting any tubes. 4-13 (3)User Setup->Gas Type->O2 (4)User Setup->Measure Mode->BTPS 4-14 6. Select [Next] to enter the screen shown below. If VT Plus is selected as the calibration device, select low-flow channel first, as shown below. Pneumatic connection with VT Plus If VT is selected, connect to the large-flow port (0~120 L/min) of the anesthesia machine calibration device, as shown below. 4-15 Gas flow direction Connector for water collection cup Pneumatic connection with VT 7. If FPM is selected, connect to the inspiratory flow sensor port (0~250 L/min) of the anesthesia machine, as shown below. Pneumatic connection with FPM Select [Continue] to enter the screen shown below. 4-16 8. The screen shown below is displayed after the low-flow channel calibration is completed. 9. If VT Plus is selected as the calibration device, connect the high-flow channel of the calibration device, as shown below. Pneumatic connection with VT Plus If VT is selected as the calibration device, there is no need to replace range. Just select [Continue]. 10. Select [Continue] to enter the screen shown below. 4-17 11. The screens shown below are displayed after calibration is completed. The screen shown below is displayed if the flow sensor calibration fails. Read details displayed on the calibration device, locate and rectify the fault. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. The screen shown below is displayed upon a successful flow sensor calibration. Select [Done] to exit the calibration screen. 12. The screen shown below is displayed if the ongoing calibration is aborted. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 4-18 Manual Calibration (01.03.00 and higher) Follow these steps to calibrate flow sensors. 1. Enter the standby mode. 2. Select [Main] > [Service] > [Calibration] > [Flow Sensors] to enter the screen shown below. 3. Select [Calibrate Manually] to enter the screen shown below. 4-19 4. 5. Connect the inspiration port and expiration port of the anesthesia machine following the on-screen instruction by referring to step 7 of section Automatic Calibration. Select [Begin] to enter the screen shown below. 6. The screen shown below is displayed after the first step of manual calibration is completed. 7. Connect the low-flow channel (If available) of calibration device with the pneumatic circuit of anesthesia machine following the on-screen instructions by referring to step 10 of section Automatic Calibration. When the calibration device needs to be switched to the high-flow channel, refer to step 13 in "Automatic Calibration" for connection. 8. NOTE: If the calibration device has multiple channels, refer to the manufacturer's specifications to know the time for channel switchover 4-20 9. Select [Continue] to enter the screen shown below. The system calibrates 32 calibration points one by one. If [Waiting] is displayed in the cell, wait for the system to implement flow control. If [Input Cal Value] is displayed in the cell, enter the standard flow value displayed on the calibration device. You can select to re-calibrate any calibration point during calibration. After entering standard flow values for all the 32 calibration points, select Accept to check and save the calibration data. 10. Select [Accept]. The screens shown below are displayed. The screen shown below is displayed upon a successful flow sensor calibration. 4-21 The screen shown below is displayed if the flow sensor calibration fails. 11. Select [Cancel] The screen shown below is displayed. NOTE: After flow calibration, check the accuracy of flow sensors by referring to 3.6.4Check the Accuracy of Flow Sensors. NOTE: In case of calibration failure, first fix the problem and then perform flow calibration again. 4.3.2.2 Common Failures and Recommended Actions Failure Description Possible Cause Recommended Action After Begin is selected, no ventilation sound is heard. The prompt message "Calibration Failure! Please try again." is displayed very soon. The prompt message "Manual Vent." is displayed, indicating that the Auto/Manual switch is set to the Manual position. The alarm "Drive Gas Pressure Low" is generated. The pressure indicated by the drive gas (O2) pressure gauge is lower than 200 kPa. Zero point error occurs in the inspiratory/expiratory flow sensor. Set the Auto/Manual switch to the Auto position. 4-22 Change or connect the gas supply to ensure that the drive gas pressure is within the specified range. Replace the VCM. Failure Description Possible Cause Recommended Action After Begin is selected, the ventilation sound is heard. The prompt message "Calibration Failure! Please try again." is displayed very soon. The sampling line of at least one of the inspiratory flow sensor, expiratory flow sensor, and ventilator flow sensor is not connected or connected reversely. The maximum flow is less than 90 L/min when the inspiration valve is opened. 1. An error occurs in the pneumatic circuit connection between the anesthesia machine calibration device and the VCM. 2. An error occurs in the communication connection between the anesthesia machine calibration device and the anesthesia machine. 3. The anesthesia machine calibration device is set incorrectly. Re-connect the sensor sampling line. The prompt message "Calibration Failure! Please try again." is displayed 15 minutes after calibration is started. Calibration data is incorrect. [00 00 00 02] is displayed. When the flow reaches 90 L/min, the collected AD value of the inspiratory flow sensor or expiratory flow sensor is greater than 3900, which is out of the normal range. The drive gas pressure is too low. [00 00 00 04] is displayed. The Auto/Manual switch is set to the Manual position. 4-23 Replace the expiration valve assembly. 1. Check the pneumatic circuit connection between the anesthesia machine calibration device and the VCM. Reconnect the pneumatic circuit if necessary. 2. Check the communication connection between the anesthesia machine calibration device and the anesthesia machine, or reconnect them to ensure normal communication. Replace the communication cable if the problem persists. 3. Check the settings of the anesthesia machine calibration device. Set the anesthesia machine again if necessary. Replace the inspiratory and expiratory flow sensors and conduct calibration again. If calibration still fails, replace the VCM. 1. Replace the flow sensor in the circuit. 2. Replace the VCM. 1. Check the drive gas supply. 2. If the gas supply is normal, check the gas supply pressure switch. 1. Check whether the Auto/Manual switch is set to the Auto position. 2. Check whether the Auto/Manual switch can switch normally. Failure Description Possible Cause Recommended Action [00 00 00 08] is displayed. A zero point error occurs in the inspiratory flow sensor (the AD value at zero point is larger than or equal to 2000, or smaller than 0). [00 00 00 10] is displayed. A zero point error occurs in the expiratory flow sensor (the AD value at zero point is larger than or equal to 2000, or smaller than 0). [00 00 00 20] is displayed. A zero point error occurs in the internal flow sensor (the AD value at zero point is larger than or equal to 2000, or smaller than 0). 1. Check whether the fresh gas is turned off. 2. Check whether the inspiration valve can be closed tightly. According to the valve diagnosis tool, after the valve is closed and the DA value is zero, if the gas supply is disconnected and then connected, the AD values of the ventilator sensor basically remain unchanged (the reading change does not exceed 1%), indicating that the valve is closed tightly. 3. Check the zero point correctness of the sensor. 4. Replace the VCM. 1. Check whether the fresh gas is turned off. 2. Check whether the inspiration valve can be closed tightly. According to the valve diagnosis tool, after the valve is closed and the DA value is zero, if the gas supply is disconnected and then connected, the AD values of the ventilator sensor basically remain unchanged (the reading change does not exceed 1%), indicating that the valve is closed tightly. 3. Check the zero point correctness of the sensor. 4. Replace the VCM. 1. Check the zero point of the sensor. 2. Check whether the inspiration valve can be closed tightly. According to the valve diagnosis tool, after the valve is closed and the DA value is zero, if the gas supply is disconnected and then connected, the AD values of the ventilator sensor basically remain unchanged (the reading change does not exceed 1%), indicating that the valve is closed tightly. 3. Replace the VCM. 4-24 Failure Description Possible Cause Recommended Action [00 00 00 40] is displayed. The measurement range of the inspiratory flow sensor is abnormal. [00 00 00 80] is displayed. The measurement range of the expiratory flow sensor is abnormal. [00 00 01 00] is displayed. The measurement range of the internal flow sensor is abnormal. 1. Check whether the sampling line is connected correctly. 2. Start the valve diagnosis tool in the calibrated pneumatic connection environment, close the expiratory valve based on 4000 DA, open the inspiration valve gradually, and observe the measured value of the calibration device under each DA value. If the collected AD value of the inspiratory flow sensor at the time point with the measured value of the calibration device close to (smaller than) 90 L/min is larger than 3900, the measurement range of the inspiratory flow sensor is abnormal. Replace the inspiratory flow sensor. 3. Replace the VCM. 1. Check whether the sampling line is connected correctly. 2. Start the valve diagnosis tool in the calibrated pneumatic connection environment, close the expiratory valve based on 4000 DA, open the inspiration valve gradually, and observe the measured value of the calibration device under each DA value. If the collected AD value of the expiratory flow sensor at the time point with the measured value of the calibration device close to (smaller than) 90 L/min is larger than 3900, the measurement range of the expiratory flow sensor is abnormal. Replace the expiratory flow sensor. 3. Replace the VCM. 1. Check whether the sampling line is connected correctly. 2. Start the valve diagnosis tool in the calibrated pneumatic connection environment, close the expiratory valve based on 4000 DA, open the inspiration valve gradually, and observe the measured value of the calibration device under each DA value. If the collected AD value of the inspiratory flow sensor at the time point with the measured value of the calibration device close to (smaller than) 45 L/min is larger than 3900, the measurement range of the ventilator flow sensor is abnormal. Replace the ventilator flow sensor. 4-25 Failure Description Possible Cause Recommended Action [00 00 02 00] is displayed. The inspiratory flow sensor does not conform to monotonicity. [00 00 04 00] is displayed. The expiratory flow sensor does not conform to monotonicity. [00 00 08 00] is displayed. The internal flow sensor does not conform to monotonicity. [00 00 10 00] is displayed. The resolution of the inspiratory flow sensor is incorrect. [00 00 20 00] is displayed. The resolution of the expiratory flow sensor is incorrect. [00 00 40 00] is displayed. The resolution of the internal flow sensor is incorrect. [00 00 80 00] is displayed. The output flow of the valve is low. 1. Check whether the unidirectional valve is connected correctly. 2. Check whether the sampling line is connected correctly. 3. Replace the inspiratory flow sensor. 4. Replace the VCM. 1. Check whether the unidirectional valve is connected correctly. 2. Check whether the sampling line is connected correctly. 3. Replace the expiratory flow sensor. 4. Replace the VCM. 1. Check whether the sampling line is connected correctly. 2. Replace the internal flow sensor. 3. Replace the VCM. 1. Check the connection of the sampling line as well as the gas tightness. 2. Check the gas supply pressure. 3. Check the settings of the calibration device. 4. Refer to 5.4 to troubleshoot sensor and valve failures by using valve diagnostic tools. 5. Replace the relevant flow sensor in the circuit. 6. Replace the VCM. 1. Check the connection of the sampling line as well as the gas tightness. 2. Check the gas supply pressure. 3. Check the settings of the calibration device. 4. Refer to 5.4 to troubleshoot sensor and valve failures by using valve diagnostic tools. 5. Replace the relevant flow sensor in the circuit. 6. Replace the VCM. 1. Check whether the sampling line is connected correctly. 2. Replace the internal flow sensor. 3. Replace the VCM. 1. Check whether the gas supply is sufficient for the whole calibration process. 2. Check whether the maximum output flow of the valve is greater than 90 L/min. If not, replace the inspiration valve. 4-26 Failure Description Possible Cause Recommended Action [00 01 00 00] is displayed. The valve resolution is low. [00 02 00 00] is displayed. The flow change is not unidirectional. [00 04 00 00] is displayed. The communication with the calibration device is interrupted. [00 08 00 00] is displayed. The system fails to write data into EEPROM. [00 10 00 00] is displayed. The ACGO switch is in the ON position. [00 20 00 00] is displayed. The maximum value is not found. [00 40 00 00] is displayed. The minimum value is not found. 4-27 1. Check whether the gas supply is sufficient for the whole calibration process. 2. Check whether the calibration device works properly. 3. Replace the inspiration valve. 1. Check whether tubes are connected as instructed. 2. Check whether the gas supply is sufficient for the whole calibration process. 3. Check whether the calibration device works properly. 1. Check the communication connection between the calibration device and the anesthesia machine or reconnect them to ensure normal communication. Otherwise, replace the communication cable. 2. Check the settings of the anesthesia machine calibration device. Do the settings again if required. 1. Perform the calibration again. 2. Replace the VCM. 1. Check whether the ACGO switch is set to OFF. 2. Check the monitoring status of the ACGO switch on the screen of the anesthesia machine. diagnosis. (1) Open the inspiration valve based on 4000 DA. The flow measured by the calibration device is larger than or equal to 90 L/min. (2) Close the inspiration valve, increase the opening of the inspiration valve based on a certain DA value. When the flow measured by the calibration device is about 80 L/min, increase 10 DAs to open the valve. If the flow increase does not exceed 5 L/min, the maximum point (the software does not find) may exist. Recalibration is recommended. 2. Replace the inspiration valve. 1. Use the valve diagnosis tool for diagnosis. Open the inspiration valve within the 0-2000 DA. If the AD value collected by the inspiratory flow sensor increases gradually, recalibration is recommended. 2. Replace the inspiration valve. Failure Description Possible Cause Recommended Action [FF FF FF FF] is displayed. A communication error occurs. 1. Restart the machine. 2. Check the communication cable. 3. Check whether a communication error alarm is generated. Replace the VCM. 4.3.3 Pressure Calibration (Factory) NOTE: Pressure calibration (factory) is necessary in case of replacing the VCM and expiration valve assembly. NOTE: When a great deviation is detected between the measured value of the built-in pressure sensor and that of the standard pressure measurement device, you need to perform pressure calibration (factory). This calibration is intended for the airway pressure sensor in the breathing circuit, PEEP pressure sensor and PEEP proportional valve of the expiratory valve assembly. The standard pressure measurement device is used to calibrate the pressure sensors and PEEP proportional valve. 4.3.3.1 Calibration Procedures NOTE: Before pressure calibration, make sure that the tubes are not leaky when connected. NOTE: Do not move or press the tubes during calibration. NOTE: You can select anesthesia machine calibration device (VT), Fluke VT Plus or FPM for automatic calibration. You can also select pressure calibration device which satisfies the accuracy requirement for manual calibration. Manual calibration is available for version 1.03.00 and later. Automatic Calibration Follow these steps to calibrate pressure sensors and PEEP proportional valve. 1. Make sure that the anesthesia machine is in standby mode. 2. Select [Main] > [Service] > [Calibration] > [Pressure Sensors] to enter the screen shown below. 4-28 3. Select [Calibrate Automatically] to enter the screen shown below. 4. Use a communication cable to connect the calibration device and the anesthesia machine by referring to step 4 in 4.3.2.1Calibration Procedures. Select [Next] to enter the screen shown below. 5. A four-way device is required to connect sampling tubes for pressure calibration. The following picture shows the four-way device, connectors on the calibration device, and VCM involved in the pressure calibration. Four-way device connecting sampling tubes for pressure calibration 4-29 Remove two tubes marked with #72 and #9 from the pressure sensors (see the following figure). Connect the four-way device to the pressure sensor P1 of the monitor board, PEEP pressure sensor P2, sampling tube marked with #72, and low-pressure port of Fluke VTPlus (or high pressure sample port of VT, or sample port of FPM). The sampling tube marked with #9 is not connected for this calibration. A D To the low pressure port of Fluke VT Plus or high pressure port of VT C B Pneumatic connection with VT Plus High pressure sampling port of VT 4-30 For FPM as a calibration device, prepare the following plugs and sampling ports. Connect the plugs, sampling ports and FPM inspiratory flow sensor ports as shown in the following figure. Note that the tube length must not exceed 0.6m. 6. 7. 8. Power on the anesthesia machine calibration device and manually zero the calibration device. Set the calibration device. Refer to step 5 in 4.3.2.1Calibration Procedures. Select [Next] to enter the screen shown below. 4-31 9. After selecting the required calibration device, select [Begin] to enter the screen shown below. During the calibration, you can select [Cancel] to abort. 10. The screens shown below are displayed after calibration is completed. The screen shown below is displayed if the pressure sensor calibration fails. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 4-32 The screen shown below is displayed upon a successful pressure sensor calibration. Select [Done] to exit the calibration screen. 11. The screen shown below is displayed if the ongoing calibration is aborted. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 4-33 Manual Calibration NOTE: Manual calibration can be performed for software version 01.03.00 and higher. Follow these steps to calibrate pressure sensors and PEEP proportional valve. 1. Make sure that the anesthesia machine is in standby mode 2. Select [Main] > [Service] > [Calibration] > [Pressure Sensors] to enter the screen shown below. 3. Select [Calibrate Manually] to enter the screen shown below. 4-34 4. 5. 6. 7. Use a communication cable to connect the calibration device and the anesthesia machine by referring to step 4 in 4.3.2.1Calibration Procedures. Complete the pneumatic connection by referring to step 5 in Automatic Calibration. Select [Begin] to enter the screen shown below. The screen shown below is displayed after the first step of manual calibration is completed. 4-35 8. Select [Continue] to enter the screen shown below. The system calibrates 32 calibration points one by one. Of the 32 calibration points, points 1 to 16 map to the rising curve while points 17 to 32 map to the falling curve. You can select to recalibrate any calibration point during calibration. When a calibration point mapped to the falling curve is calibrated, you cannot change the points mapping to the rising curve. If [Waiting] is displayed in the cell, wait for the system to implement pressure control. If [Input Cal Value] is displayed in the cell, enter the standard pressure displayed on the calibration device. After entering standard pressures of all the 32 calibration points, select [Accept] to check and save the calibration data. 9. Select [Accept]. The screens shown below are displayed. The screen shown below is displayed upon a successful pressure sensor calibration. 4-36 The screen shown below is displayed if the pressure sensor calibration fails. 10. Select [Cancel]. The screen shown below is displayed. NOTE: After pressure calibration, test the accuracy of pressure sensors by referring to 3.6.5Check the Accuracy of Pressure Sensors. NOTE: In case of calibration failure, first fix the problem and then perform pressure calibration again. 4-37 4.3.3.2 Common Failures and Recommended Actions Failure Description After [Begin] is selected, no ventilation sound is heard. The prompt message "Calibration Failure! Please try again." is displayed very soon. After [Begin] is selected, ventilation sound is heard. The prompt message "Calibration Failure! Please try again." is displayed very soon. The prompt message "Calibration Failure! Please try again." is displayed 15 minutes after calibration is started. [00 00 00 02] is displayed. Possible Cause Recommended Action The alarm "Drive Gas Pressure Low" is generated. The pressure indicated by the drive gas (O2) pressure gauge is lower than 200 kPa. Zero point error occurs in the Paw sensor or PEEP branch pressure sensor. Refer to“3.6.3 Check the Sensor Zero Point”. The sampling line of at least one of the Paw sensor or PEEP branch pressure sensor is not connected or connected incorrectly. Refer to 5.4 Sensor and Valve Related Failures. The maximum pressure generated by the PEEP valve is smaller than 95 cmH2O. Refer to 5.4 Sensor and Valve Related Failures. 1. An error occurs in the pneumatic circuit connection between the anesthesia machine calibration device and the VCM. 2. An error occurs in the communication connection between the anesthesia machine calibration device and the anesthesia machine. 3. The anesthesia machine calibration device is set incorrectly. Change or connect the gas supply to ensure sufficient drive gas pressure. The calibration data is incorrect: The measurement range of sensors is abnormal; the calibration data does not conform to monotonicity; the sensor resolution is incorrect. Refer to 3.6.5 Check the Accuracy of Pressure Sensors. The drive gas pressure is too low. 4-38 Replace the VCM. Reconnect the sensor sampling line. Replace the expiration valve assembly. 1. Check the pneumatic circuit connection between the anesthesia machine calibration device and the VCM. Reconnect the pneumatic circuit if necessary. 2. Check the communication connection between the anesthesia machine calibration device and the anesthesia machine or reconnect them to ensure normal communication. Replace the communication cable if the problem persists. 3. Check the settings of the anesthesia machine calibration device. Set the anesthesia machine again if necessary. Replace the VCM. 1. Change or connect the gas supply to ensure that the gas pressure ranges from 350 kPa to 450 kPa. 2. If the gas supply is normal, check the gas supply pressure switch. Failure Description Possible Cause Recommended Action [00 00 00 04] is displayed. The Auto/Manual switch is set to the Manual position. [00 00 00 08] is displayed. Zero point error occurs in the Paw sensor. [00 00 00 10] is displayed. Zero point error occurs in the PEEP pressure sensor. [00 00 00 20] is displayed. The measurement range of the Paw sensor is abnormal [00 00 00 40] is displayed. The measurement range of the PEEP pressure sensor is abnormal. [00 00 00 80] is displayed. The calibration data of the Paw sensor does not conform to monotonicity. [00 00 01 00] is displayed. The calibration data of the PEEP pressure sensor does not conform to monotonicity. [00 00 02 00] is displayed. The resolution of the Paw sensor is incorrect. The difference between the maximum and minimum AD values is smaller than or equal to 500AD. The resolution is smaller than 1 cmH20 ~2AD when the difference between two adjacent pressures is larger than or equal to 1 cmH2O. 1. Check whether the Auto/Manual switch is set to the Auto position. 2. Check whether the Auto/Manual switch can switch normally. 1. Check the zero point. 2. Replace the VCM. 1. Check the zero point. 2. Replace the VCM. 1. Check the pipeline connection and gas tightness. 2. Check the gas supply pressure. 3. Check the settings of the calibration device. 4. Refer to 5.4 to troubleshoot sensor and valve failures by using valve diagnostic tools. 5. Replace the VCM. 1. Check the pipeline connection and gas tightness. 2. Check the gas supply pressure. 3. Check the settings of the calibration device. 4. Refer to 5.4 to troubleshoot sensor and valve failures by using valve diagnostic tools. 5. Replace the VCM. 1. Check the pipeline connection and gas tightness. 2. Check the gas supply pressure. 3. Refer to 5.4 to troubleshoot sensor and valve failures by using valve diagnostic tools. 4. Replace the VCM. 1. Check the pipeline connection and gas tightness. 2. Check the gas supply pressure. 3. Refer to 5.4 to troubleshoot sensor and valve failures by using valve diagnostic tools. 4. Replace the VCM. 1. Check the pipeline connection and gas tightness. 2. Check the gas supply pressure. 3. Check the settings of the calibration device. 4. Refer to 5.4 to troubleshoot sensor and valve failures by using valve diagnostic tools. 5. Replace the VCM. 4-39 Failure Description Possible Cause Recommended Action 1. Check the pipeline connection and gas tightness. 2. Check the gas supply pressure. 3. Check the settings of the calibration device. 4. Refer to 5.4 to troubleshoot sensor and valve failures by using valve diagnostic tools. 5. Replace the VCM. [00 00 08 00] is displayed. The resolution of the PEEP pressure sensor is incorrect. The difference between the maximum and minimum AD values is smaller than or equal to 500AD. The resolution is smaller than 1 cmH20 ~2AD when the difference between two adjacent pressures is larger than or equal to 1 cmH2O. The output pressure of the valve is low. [00 00 10 00] is displayed. The pressure change is not unidirectional. [00 00 20 00] is displayed. The communication with the calibration device is interrupted. [00 00 40 00] is displayed. The system fails to write data into EEPROM. [00 00 80 00] is displayed. The ACGO switch is in the ON position. [00 01 00 00] is displayed. The valve resolution is low. [00 00 04 00] is displayed. [00 02 00 00] is displayed. The maximum value is not found. 4-40 1. Check whether the gas supply is sufficient for the whole calibration process. 2. Use the valve diagnosis tool to check whether the maximum output pressure of the PEEP valve is larger than 90 cmH20. If not, replace the airway module. 1. Check if the sampling line is correctly connected. 2. Replace the VCM. 1. Check the connection between the calibration device and the communication cable. 2. Replace the calibration device and perform the calibration again. 1. Perform the calibration again. 2. Replace the VCM. 1. Check whether the ACGO switch is set to OFF. 2. Check the monitoring status of the ACGO switch on the screen of the anesthesia machine. 1. Check whether the gas supply is sufficient for the whole calibration process. 2. Check whether the calibration device works properly. 3. Replace the PEEP valve. 1. Use the valve diagnosis tool for diagnosis. (1) Open the PEEP valve based on 4000DA. The pressure measured by VT is greater than or equal to 90 cmH2O. (2) Close the PEEP valve and increase the opening of the PEEP valve based on a certain DA value. The AD values collected by the Paw sensor increase gradually. Recalibration is recommended if the preceding conditions are met. 2. Replace the PEEP valve. Failure Description [00 04 00 00] is displayed. [FF FF FF FF] is displayed. Possible Cause Recommended Action The minimum value is not found. A communication error occurs. 1. Use the valve diagnosis tool for diagnosis. Open the PEEP valve within the 0-2000DA. If the AD values collected by the inspiratory pressure sensor increase gradually, recalibration is recommended. 2. Replace the PEEP valve. 1. Restart the anesthesia machine and/or calibration device. 2. Check the communication cable. 3. Check whether a communication error alarm is generated. 4. Replace the VCM or calibration device. 4.3.4 Pressure and Flow Zeroing (Factory) The anesthesia machine automatically performs pressure and flow zeroing at regular intervals during operation. You can also perform pressure and flow zeroing manually in the factory maintenance menu. Manual zeroing can immediately eliminate measurement deviations caused by zero offset of sensors. The anesthesia machine supports automatic and periodic pressure and flow zeroing. The first four zeroing operations are done automatically 5, 15, 30, and 60 minutes after ventilation starts. Then automatic zeroing takes place every 120 minutes. The three-way valve is opened and closed for valve flushing during mechanical ventilation and before automatic zeroing. A dent appears in the waveform during zeroing, or opening and closing of the three-way valve. 4.3.4.1 Zeroing Procedures Follow these steps to zero pressure and flow sensors. 1. Select [Main] > [Service] > [Calibration] > [Zero Sensors] to enter the screen shown below. 4-41 2. Select [Begin] to enter the screen shown below. During the zeroing, you can select [Cancel] to abort. 3. The screen shown below is displayed if the ongoing zeroing is aborted. Select [Try Again] to do the zeroing again or select [Done] to exit the zeroing screen. 4. The screen shown below is displayed if the zeroing fails. Select Try Again to do the zeroing again or select Done to exit the zeroing screen. 4-42 5. The screen shown below is displayed upon a successful zeroing. Select [Done] to exit the zeroing screen. NOTE: In case of zeroing failure, other faults may exist. You must isolate and eliminate the problem. 4.3.4.2 Troubleshoot Pressure and Flow Zeroing Failures If the zeroing fails, perform the following steps to locate and rectify the fault. 1. Set the anesthesia machine to manual or standby mode. Turn off the fresh gas. Remove breathing tubes from the breathing system so that the inspiration and expiration connectors are exposed in the air. Bleed the residual gas inside the bellows. Ensure that no flow or pressure enters the flow or pressure sensor inside the machine. 2. Check whether the zero points of sensors are normal by referring to 3.6.3 Check the Sensor Zero Point. 3. If a zero point error is detected, remove the sampling line from the sensor to eliminate effects caused by the sampling line occlusion or three-way valve. If the zero point is still incorrect, the VCM is faulty. Replace the VCM. 4. If the zero points of sensors are correct but zeroing still fails, the three-way valve assembly is faulty. Replace the three-way valve assembly. 4-43 4.3.5 Electronic Flowmeter Zeroing (User) After the gas supply is disconnected, zero offset may occur to the sensor of the electronic flowmeter if the pointer of the pressure gauge returns to zero but the electronic flowmeter still displays a non-zero flow value. You can zero the flowmeter manually to immediately eliminate measurement deviations caused by zero offset. 4.3.5.1 Zeroing Procedures Follow these steps to zero the electronic flowmeter. 1. Select [Main] > [General] > [Zero Flow Meters] to enter the screen shown below. Set up the machine as per the instructions on the screen and select [Begin] to start zeroing. 2. The zeroing screen shown below is displayed. During zeroing, you can select [Cancel] to abort. 4-44 3. The screen shown below is displayed if the ongoing zeroing is aborted. Select [Try Again] to do the zeroing again or select [Done] to exit the zeroing screen. 4. The screen shown below is displayed if the zeroing fails. Select [Try Again] to do the zeroing again or select [Done] to exit the zeroing screen. 5. The screen shown below is displayed upon a successful zeroing. Select [Done] to exit the zeroing screen. NOTE: In case of zeroing failure, the electronic flowmeter may have other faults. You must isolate and eliminate the problem. 4-45 4.3.5.2 Troubleshoot Electronic Flowmeter Zeroing Failures If the zeroing fails, perform the following steps to locate and rectify the fault. 1. Disconnect the gas supplies. Bleed the gas inside the machine (or adjust the flowmeter for the pointer of the pressure gauge to return to zero) and then perform zeroing again. 2. If the zeroing is successful, it is concluded that the previous zeroing failure is caused by the mechanical failure in the three-way valve. Replace the three-way valve assembly. 3. If zeroing is still failed, it is concluded that zeroing is caused by the three-way valve hardware circuit fault or electronic flowmeter board fault. Replace the three-way valve assembly or electronic flowmeter board. 4.3.6 O2 Sensor Calibration NOTE: The O2 sensor needs to be calibrated when a great deviation occurs in the O2 monitored value or the O2 sensor or VCM is replaced. Calibrate the O2 sensor again when a great deviation of O2 concentration monitored value occurs or when the O2 sensor or the VCM is replaced. NOTE: Before calibration, observe if the O2 sensor displays numerics on the measure screen. If not, confirm that the O2 measure switch is turned on, check the O2 sensor connection line, or replace the O2 sensor until measure numerics are displayed. 4.3.6.1 21% O2 Calibration Follow these steps to calibrate the O2 sensor at 21% O2. 1. Select [Main] > [General] > [Calibrate O2 Sensor], or select [Main] > [System] > [Calibrate / Check Sensor], or select [Main] > [Service] > [Calibration] > [O2 Sensor] to enter the screen shown below. The [General] tab displays only the O2 sensor calibration at 21% O2. The [System] and [Service] tabs display the O2 sensor calibration at 21% O2 and 100% O2, and need passwords. Follow the instructions on the screen to set the machine. Select [Begin] to start calibration 4-46 2. The calibration screen shown below is displayed after [Begin] is selected. During the calibration, you can select [Cancel] to abort. 3. The screen shown below is displayed if the ongoing calibration is aborted. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 4. The screen shown below is displayed if the calibration fails. The red font indicates failure code. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 4-47 5. The screen shown below is displayed upon a successful calibration (in standby mode). Select [Done] to exit the calibration screen. 4.3.6.2 100% O2 Calibration NOTE: 100% O2 calibration shall be performed in standby mode. NOTE: 100% O2 calibration shall be performed after a successful 21% O2 calibration. Follow these steps to calibrate the O2 sensor at 100% O2. 1. Enter the standby mode. 2. Select [Main] > [System] > [Calibrate / Check Sensor] or select [Main] > [Service] > [Calibration] > [O2 Sensor]. The [System] and [Service] tabs display O2 sensor calibration at 21% O2 and 100% O2 and need passwords. The calibration screen shown below is displayed after [100%] is selected. Follow the instructions on the screen to set the machine and select [Next]. Set the Auto/Manual switch to the Auto position. 4-48 3. After Next is selected, the system turns off N2O and Air flows automatically, turns on O2 flow which is greater than 8L/min automatically, and enters the calibration screen as shown below. Follow the instructions on the screen to set the machine. Wait 2 minutes and ensure that the O2 cell voltage has stabilized at the maximum value for at least 30s. Select Begin. 4. The calibration screen shown below is displayed after [Begin] is selected. During the calibration, you can select [Cancel] to abort. 5. The screen shown below is displayed if the ongoing calibration is aborted. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 4-49 6. The screen shown below is displayed if the calibration fails. The red font indicates failure code. Select [Try Again] to do the calibration again or select [Done] to exit the calibration screen. 7. The screen shown below is displayed upon a successful calibration. Select [Done] to exit the calibration screen. 4-50 4.3.6.3 Common Failures and Recommended Actions Failure Description Possible Cause Recommended Action After Begin is selected, a calibration failure is prompted. If the alarm "O2 Sensor Disconnected" is displayed, the O2 sensor is not connected. The O2 supply pressure is insufficient (less than 200 kPa). Connect the O2 sensor. A calibration failure is prompted three minutes after the calibration is started. 21% O2 calibration is not completed before 100% O2 calibration. The O2% sampling value is not within the normal range. That is, The 21% O2 sampling value is out of the range of 150 to 560 and the 100% O2 sampling value is out of the range of 523 to 2123. Select Main > Service > Data Monitors > VCM to view the O2% sampling value Error Code Description 00000002 The O2 supply pressure is low during the calibration at 100% O2. 00000004 The O2 sensor is not connected and the collected data is in the range of (41960 to 63165) (AD value). 00000008 The calibration value at 21% O2 is out of the normal range (3586 to 9140) (AD value). 00000010 The calibration value at 100% O2 is out of the normal range (9602 to 35727) (AD value). 4-51 Change or connect the gas supply to ensure sufficient O2 supply pressure. Perform 21% O2 calibration followed by 100% O2 calibration. Replace the O2 sensor. Recommended Action 1. Check whether cables of the O2 sensor are connected correctly. 2. Check the O2 supply pressure. 3. Check whether the output voltage of the O2 sensor is stable in the calibration menu. 4. Replace the O2 sensor. 1. Check whether the alarm "O2 Sensor Disconnected" is displayed. If yes, check whether the cables of the O2 sensor are connected correctly. 2. Check whether the output voltage of the O2 sensor is sable in the calibration menu. 3. Replace the O2 sensor. 1. Check whether cables of the O2 sensor are connected correctly. 2. Check whether the O2 sensor is in the 21% O2 environment. 3. Check whether the output voltage of the O2 sensor is stable in the calibration menu. 4. Replace the O2 sensor. 1. Check whether cables of the O2 sensor are connected correctly. 2. Check whether the O2 sensor is in the 100% O2 environment. 3. Check whether the output voltage of the O2 sensor is stable in the calibration menu. 4. Replace the O2 sensor. Error Code Description 00000020 An error occurs when data is written into EEPROM. Recommended Action 1. Perform the calibration again. 2. Replace the O2 sensor. 3. Replace the CPU board. 4.3.7 External AG Calibration Use the sampling pipe to connect the standard gas (its concentration: 5% CO2, 45% O2; 45% N2O; 5% DES), cylinder, pressure reduce valve, water trap and anaesthesia machine. Perform the following tests after the module is preheated: Open the cylinder valve and access to the gas calibration-related screen of the integrated equipment (see the following details). Observe and check if the concentration of CO2 falls within 5.0%±0.2% and O2 within 45% ±2%. Notice that the O2 concentration will not be displayed for the module without O2. Notices: a. The method of accessing to calibration screen: under the Standby mode, access to Main→System→Input the password (789789)→Accept. Click Calibrate and choose External AG and then enter into the gas calibration screen of the integrated equipment. b. If the gas concentration displayed does not within the given range, set the object value of the gas type corresponded to the standard gas to be the standard value, then click Calibrate. 4.3.8 CO2 Module Calibration Pressure Relief Valve Three-way connectors Cylinder 4-52 CO2 module Make sure the module has been running for over 15 minutes under the test status when start to calibrate. First, click Zero Calibration and then choose Continue. Set the “concentration option” to be 5% in the CO2 module calibration screen. Then, Input the standard CO2 at a concentration of 5%. After the value staying unchanged for 20 seconds, click Calibration Confirmation button. 4.3.9 ORC Calibration NOTE: Adjust N2O and Air resistance of resistance block assembly to the minimum before adjusting the initial volume to ensure that the maximum flow can be satisfied when adjusting the O2 and N2O ratio. NOTE: Perform this calibration if the ORC is replaced. Follow these steps to perform ORC calibration. 1. Connect the anesthesia machine to the O2, AIR and N2O supplies. Close the O2, AIR and N2O needle valves. Switch the screen to standby mode. 2. Turn the N2O needle valve to the maximum position. Slowly turn the O2 needle valve to produce O2 flow of 0.25 L/min. If there is N2O flow displayed, turn the O2 resistor on the ORC anti-clockwise with the flathead screwdriver until there is no N2O flow. If there is no NO2 flow displayed, turn the O2 resistor on the ORC clockwise with the flathead screwdriver until the N2O flow marginal value appears, which is not “0”. 4-53 4-54 O2 resistor 3. 4. Keep the N2O needle valve at the maximum position. Close the O2 needle valve knob and then slowly open the O2 needle valve. Observe if the O2 flow falls within 0.20 to 0.25 L/min when the N2O flow begins. If not, repeat step 2 until the requirement is satisfied. Keep the N2O needle valve at the maximum position. Turn O2 flow to 3 L/min and observe N2O flow. If the N2O flow is less than 8.1±0.4 L/min, turn the N2O resistor on the ORC anti-clockwise with the flathead screwdriver until the N2O flow reaches 9 L/min. If the N2O flow is greater than 8.1±0.4 L/min, turn the N2O resistor on the ORC clockwise with the flathead screwdriver until the N2O flow reaches 8.1±0.4 L/min. 4-55 N2O resistor 4-56 5. Turn the O2, AIR and N2O needle valves to the maximum positions. Turn the AIR resistor on the pneumatic resistor block assembly to cause the AIR flow to fall within 15 to 15.5 L/min. AIR resistor 4-57 4.3.10 High-Pressure Gas Cylinder Regulator Calibration Perform the following steps to calibrate the high-pressure gas cylinder regulator. The following uses the N2O high-pressure gas cylinder regulator as an example. The calibration procedure of O2 and Air high-pressure gas cylinder regulators is the same as that of the N2O high-pressure gas cylinder regulator. The pressure in the cylinder must be 1000 psi at least for O2 and atmospheric air and 500 psi at least for N2O. For O2 and air, set the output pressure as per the following table. For N2O, set the output pressure to 58 psi. Cylinder Input Pressure (psi) 1000 1150 1400 1650 1900 2150 2400 2650 2900 1. Regulator Output Pressure (psi) 59.2 58.5 57.2 56.0 55.1 54.3 52.9 51.8 50.7 Turn off the power supply and all gas supplies. Open the service door. 4-58 2. Remove the rear cover plate. 3. Open the rear cover plate of the integrated equipment and take off the rear panel of the cart. After extracting the pipe connected with check valve of cylinder assembly from one port of the check valve, connect it to the Φ8 three-way valve. And then connect another two ports of the three-way valve to the check valve and the pressure monitoring device separately. 4. Initiate the integrated equipment and turn on the gas supply of the cylinder (The gas supply of pipes should be off now). Adjust the fresh gas flow to 1 L/min. Remove the cap type self-locking nut at the head of the press reducer. 5. 4-59 6. Rotate the regulating screw at the head slowly with a 4mm hexagon head spanner to adjust the output pressure range (rotate clockwise to increase the pressure and counterclockwise to reduce the pressure) until the adjusted pressure is within the range listed in the preceding table (±5%). After adjusting the pressure, reinstall and tighten the self-locking acorn nut. 7. Turn off the gas supply and restore the machine tube connections. 4-60 5 Repair and Troubleshooting 5.1 Troubleshooting Guidelines 5.1.1 Identify Problems Due to the wide variety of potential symptoms, certain problems may be more subtle than others. Following the guidelines of the tests will help determine the problem, if one exists. 5.1.2 Avoid Shorting Component Leads During repair procedures, it can be tempting to make a quick series of measurements. Always turn the power off before connecting and disconnecting the test leads and probes. The accidental shorting of leads can easily damage the components and cause a second failure (aside from the safety risk). 5.1.3 Use Proper Tools During repair procedures, the following tools may be required: Metric Allen wrench es(2.5, 3, 4, 5, 8mm) Phillips screwdriver(#1 and #2) Diagonal pliers Flathead screwdriver Metric M3 and M4 socket screwdriver Adjustable wrench Tweezers Krytox Lubricant (P/N:0510-00-0020) It is imperative to use the designated equipment in order to ensure proper results of any and all test procedures. 5.1.4 Clean up the Repair Area Clean up the repair area after any repair. 5.2 Technical Alarms A technical alarm, as apposed to a parameter alarm, is an alarm condition that exists no matter whether a patient is connected to the machine. Technical alarms include: Startup alarm messages CPU board runtime alarms Power board runtime alarms Electronic flowmeter board runtime alarms VCM runtime alarms AG module runtime alarms A technical alarm, as apposed to a parameter alarm, is an alarm condition that exists no matter whether a patient is connected to the machine. Technical alarms include: Startup alarm messages CPU board runtime alarms Power board runtime alarms Runtime alarms of the flow sensor board VCM runtime alarms AG module runtime alarms 5-1 5.2.1 Startup Alarm Messages Message Priority Cause Solution Bundle Version Error Bundle Version: Time out High Software versions are incompatible. 1. Upgrade the software to a compatible version. Flowmeter Voltage Error High DVCC, AVDD or VC voltage error High 1. CPU, Flash, or WTD error. 2. After restart, the CPU board cannot communicate with the electronic flowmeter board. High 1. CPU, Flash, WTD error. 2. After restart, the CPU board cannot communicate with the auxiliary VCM. Ventilator Selftest Error Ventilator Self Test: Time out High 1. CPU, timer, RAM, WTD, EEPROM, or AD error. 2. After restart, the CPU board cannot communicate with the VCM. Ventilator Voltage Error High 5 V or 12 V voltage error Flowmeter Selftest Error Flowmeter Self Test: Time out Aux Control Module Selftest Error Aux Control Module Self Test: Time out 5-2 1. Restart the machine. 2. Measure the 12V input voltage (voltage on the power cable) of the electronic flowmeter to see if it is within specifications. Check if the cable is defective. If the input voltage is out of specification, replace the power board or cable. 3. If the problem persists, replace the electronic flowmeter board. 1. Restart the machine. 2. Re-plug or replace the communication cable between the CPU board and the electronic flowmeter board. 3. If the problem persists, replace the electronic flowmeter board. 4. If the problem persists, replace the CPU board. 1. Restart the machine. 2. Re-plug or replace the communication cable between the CPU board and the auxiliary VCM. 3. If the problem persists, replace the auxiliary VCM. 4. If the problem persists, replace the CPU board. 1. Restart the machine. 2. Re-plug or replace the communication cable between the CPU board and the VCM. 3. If the problem persists, replace the VCM. 4. If the problem persists, replace the CPU board. 1. Restart the machine. 2. Measure the input voltage of the VCM to check whether it is within the normal range. 3. If the input voltage is out of Message Priority Cause PEEP Valve Failure Med 1. PEEP valve voltage error. 2. PEEP valve pressure error. Insp Valve Failure Med 1. Inspiratory valve voltage error. 2. Inspiratory valve flow error. PEEP Safety Valve Failure Med PEEP safety valve fault. 5-3 Solution range, replace the power board or cable. 4. If the problem persists, replace the VCM. 1. Check whether the pressure sensor on the PEEP circuit complies with the specifications. Calibrate the pressure sensor in the service menu or replace the sensor on the PEEP circuit when a pressure error occurs. 2. Measure the voltage at the corresponding test point. 3. Check the connection between the power supply and the expiratory valve assembly. 4. Replace the VCM when necessary. 5. Replace the expiratory valve assembly when necessary. 6. Replace the power board when necessary. 1. Check whether the inspiratory flow sensor complies with the specifications. Calibrate the flow sensor in the service menu or replace the flow sensor when a flow error occurs. 2. Measure the voltage at the corresponding test point. 3. Check the connection between the power supply and the expiratory valve assembly. 4. Replace the VCM when necessary. 5. Replace the expiratory valve assembly when necessary. 6. Replace the power board when necessary. 1. Measure the voltage at the corresponding test point. 2. Check the connection between the power supply and the expiratory valve assembly. 3. Replace the VCM when necessary. 4. Replace the expiratory valve assembly when necessary. 5. Replace the power board when necessary. Message Flow Sensor Failure Calibrate Flow Sensor and Insp Valve Calibrate Pressure Sensor and PEEp Valve Calibrate O2 Sensor Priority Cause Solution Low The measurement of the flow sensor is out of normal range. Low 1. No calibration table is found in the EEPROM. 2. The checksum in the calibration table is incorrect. 1. Check whether the zero point of the flow sensor complies with the specifications. 2. Check whether the values measured by the flow sensor are within the normal range. 3. Replace the flow sensor and calibrate it. 4. Replace the VCM and calibrate it. Perform service calibration. For details about how to calibrate the flow sensor, see related section. Low 1. No calibration table is found in the EEPROM. 2. The checksum in the calibration table is incorrect. Perform service calibration. For details about how to calibrate the pressure sensor, see related section. Low 1. No calibration table is found in the EEPROM. 2. The checksum in the calibration table is incorrect. 1. Calibrate the O2 sensor again. 2. If the problem persists, replace the O2 sensor. Ventilator Initialization Error Ventilator Initialization: Time out High After power-on, the CPU board cannot send the parameter settings to the VCM. Drive Gas Pressure Low High The drive gas pressure is low. O2 Supply Failure High O2 supply failure. Power Supply High 3.3 V, 5 V, 12 V voltage errors. 5-4 1. Restart the machine. 2. Re-plug or replace the communication cable between the CPU board and the VCM. 3. If the problem persists, replace the VCM. 4. If the problem persists, replace the CPU board. 1. Check whether gas is supplied normally. 2. Short-circuit the pressure switch and the alarm should disappear. Otherwise, it indicates that the pressure switch is defective. Replace the pressure switch. If the pressure switch is not defective, check the connection between the pressure switch and the VCM. 3. If the problem persists, replace the VCM. Use the same method as that of Drive Gas Pressure Low to check the O2 pressure switch. 1. Measure the voltage at the Message Voltage Error Priority Cause AC power failure Low AC power failure RT Clock Failure High The clock chip malfunctions. Keyboard Selftest Error Keyboard Selftest: Timeout High The keyboard malfunctions. External AG Selftest Error External AG: Time out Low The external AG module malfunctions. BIS Selftest Error BIS Selftest Error: Time out Low The BIS module malfunctions. CO2 Selftest Error CO2 Selftest Error: Time out Low The CO2 module malfunctions. NMT Selftest Error NMT Selftest Error: Time out Low The NMT module malfunctions. 5-5 Solution corresponding test point. 2. If the problem persists, replace the power board. 1. Check AC power supply 2.Check power connection 1. Restart the machine. 2. If the problem persists, replace the VCM. 1. Check the cable connection between the keyboard and the CPU board. 2. Restart the machine. 3. If the problem persists, replace the keyboard. 1. Re-plug the external AG module, and then start it. 2. Check the cable connection between the gas bench and the CPU board. 3. Check whether the gas bench works properly. 4. Replace the external AG module. 1. Re-plug the BIS module, and then start it. 2. Check the cable connection between the gas bench and the CPU board. 3. Check the main cables connection. 4. Check whether the gas bench works properly. 5. Replace the BIS module. 1. Re-plug the CO2 module, and restart the machine for selftest. 2. Check the cable connection between the gas bench and the CPU board. 3. Check whether the gas bench works properly. 4. Replace the CO2 module. 1. Re-plug the NMT module, and restart the machine for selftest. 2. Check the cable connection between the gas bench and the CPU board. 3. Check whether the gas bench works properly. 4. Replace the NMT module. 5.2.2 CPU Board Runtime Alarms Message IP Address Conflict Fan Failure Fan Failure 02 Priority Cause Solution Med The IP address is the same as that of another machine on the local network. 1. Set the IP address again. 2. If the problem persists, upgrade the system software or replace the CPU board. Med The fan speed is 20% less than the rated value. Med The fan speed of the gas bench is lower than 3640 rounds/min. 1. Check whether the fan stops or runs slowly (about 4000 rounds/min normally). 2. Re-plug the fan cable. 3. If the problem persists, check whether the 12 V power supply for fans is within the normal range. If not, check the power board. 4. If the problem persists, replace the fan. 5. If the problem persists, replace the CPU board. 1. Check whether the fan stops or runs slowly (about 4000 rounds/min normally). 2. Re-plug the fan cable. 3. If the problem persists, check whether the 12 V power supply for fans is within the normal range. If not, check the power board. 4. If the problem persists, replace the fan. 5. If the problem persists, replace the CPU board. 5.2.3 Power Board Runtime Alarms Message Priority Cause Power System Comm Stop High The CPU board fails to communicate with the power board for 10 seconds. Power Supply Voltage Error High 3.3V, 5V, 12V voltage errors Solution 1. Restart the machine. 2. Re-plug the communication cable. 3. Remove the internal battery. Power on the power board again 5 minutes after it is powered off. 4. Replace the communication cable. 5. Check whether the power board software is correct. Upgrade the software when necessary. 6. If the problem persists, replace the power board. 7. If the problem persists, replace the CPU board. 1. Measure the voltage at the corresponding test point. 2. Remove the internal battery. Power on the power board again 5 minutes after it is powered off. Repeat step 1. 3. If the problem persists, replace the power board. 4. If the problem persists, contact the technical support personnel. 5-6 Message Priority Cause Low Battery Voltage! High The battery voltage is less than 10.6 V for a consecutive of 5 seconds. System going DOWN, battery depleted! High The battery voltage is less than 10.2 V. Battery Undetected Med No battery is detected. Battery in Use Low AC power failure. High The power board temperature exceeds 95ºC. Low 1. Both resistance temperatures exceed 106ºC for 20 seconds. 2. Either resistance temperature exceeds 110ºC for 15 seconds. 1. Restart the machine. 2. If the problem persists, check whether the heating temperature and voltage are within the normal ranges. If not, replace the power board. 3. If the problem persists, replace the CPU board and upgrade the system software. High No breathing circuit is mounted. 1. Check that the circuit is installed correctly. 2. Test the connection between the connection line and the connector. 3. Replace the power board. Power Board High Temp Heating Module Failure Breathing Circuit Not Mounted Solution 1. Check the connection to the AC mains. Reconnect the AC mains immediately. 2. Check whether the battery voltage is within the normal range. 3. Check whether the charging circuit works properly. If not, replace the power board. 1. Restart the machine. 2. If the problem persists, check the battery voltage in the service menu. If the battery voltage is less than 10.2 V, replace the battery. 3. If the problem persists, connect to the normal mains supply. Make sure that the AC indicator is on and charge the battery for 20 minutes. 4. If the problem persists, replace the battery. 5. If the problem persists, replace the power board. 1. Check whether the battery voltage is within the normal range. 2. Check whether the cable is connected correctly. 3. Replace the battery. 4. If the problem persists, replace the power board. 1. Check the connection to the AC mains. 2. If the AC mains is connected correctly and the voltage is within the normal range, check the connection between the AC mains and the power board. 3. If the problem persists, replace the power board. 1. Check whether the fan on the power module operates properly. 2. Stop using the machine for a period of time. If the problem persists after the machine is restarted, replace the power board. 5.2.4 Electronic Flowmeter Board Runtime Alarms 5-7 Message Flowmeter Voltage Error N2O Flow Too High Priority High DVCC, AVDD or VC voltage error Low N2O flow is greater than 15L/min for 1 second. O2 Flow Too High Low Air Flow Too High Low O2-N2O Ratio Error Flowmeter Comm Stop No Fresh Gas Internal N2O Flow Failure Cause O2 flow is greater than 25L/min for 1 second. Air flow is greater than 20L/min for 1 second. Solution 1. Restart the machine. 2. Measure the 12V input voltage (voltage on the power cable) of the electronic flowmeter to see if it is within specifications. Check if the cable is defective. If the input voltage is out of specification, replace the power board or cable. 3. If the problem persists, replace the electronic flowmeter board. 1. Turn off other gas flow. Compare the concerned gas flow with the measurement result displayed on the total flowmeter. 2. Use test tools. Turn Auto/Manual to Bag, Set APL valve at Max value. Measure fresh gas flow at the inspiration connector. Check the measurement error of electronic flowmeter. 3. If the problem persists, replace the flowmeter relative subassembly. Use the same method with “N2O Flow Too High” to “O2 Flow Too High” to check the O2 flowmeter. Use the same method with “N2O Flow Too High” to “Air Flow Too High” to check the Air flowmeter. High N2O flow is greater than 0.5 L/min and greater than 4 times O2 flow, this condition last for 1.6 seconds. 1. Restart the machine. 2. Check the measurement correctness of O2 and N2O flow sensors. If measurement error occurs, replace the flow sensor. 3. Check the ORC for leakage. Connect the tubes again. If an error occurs, replace the ORC. 4. If the problem persists, replace the electronic flowmeter board. High The electronic flowmeter board cannot establish normal communication with the CPU board for 10 seconds. 1. Restart the machine. 2. Re-plug or replace the communication cable between the CPU board and the electronic flowmeter board. 3. If the problem persists, replace the electronic flowmeter board. 4. If the problem persists, replace the CPU board. Med Gas flow of Air, O2, or N2O is all less than 50mL/min. Low The communication between the electronic 1. Check if the fresh gas knob is turned on. 2. Check the measurement correctness of flow sensors. If measurement error occurs, replace the flow sensor. 3. Check the ORC for leakage. Connect the tubes again. If an error occurs, replace the ORC. 4. If the problem persists, replace the electronic flowmeter board. 1. Restart the machine. 2. Check the measurement correctness of flow sensors. If measurement error occurs, replace the flow sensor. 5-8 Message Priority Internal O2 Flow Failure Low Internal Air Flow Failure Low Cause flowmeter CPU and the N2O flow sensor is not successful for continuous 3s. The communication between the electronic flowmeter CPU and the O2 flow sensor is not successful for continuous 3s. The communication between the electronic flowmeter CPU and the N2O flow sensor is not successful for continuous 3s. Solution 3. If the problem persists, replace the electronic flowmeter board. Handle in the similar way to handling “Internal N2O Flow Failure”. Handle in the similar way to handling “Internal N2O Flow Failure”. 5-9 5.2.5 VCM Runtime Alarms Message Priority Cause Aux Control Module Comm Stop High The CPU board fails to communicate with the auxiliary VCM for 10 seconds. Ventilator Voltage Error High 5V or 12V voltage error. Med 1. PEEP valve voltage error. 2. PEEP valve pressure error. Insp Valve Failure Med 1. Inspiratory valve voltage error. 2. Inspiratory valve flow error. PEEP Safety Valve Failure Med PEEP safety valve voltage error. PEEP Valve Failure Solution 5-10 1. Restart the machine. 2. Re-plug the communication cable. 3. If the problem persists, replace the auxiliary VCM. 4. If the problem persists, replace the CPU board. 1. Restart the machine. 2. Measure the input voltage of the VCM to check whether the voltage is within the normal range. Check whether the cable is defective. 3. If the input voltage is out of range, replace the power board or cable. 4. If the problem persists, replace the VCM. 1. Check whether the pressure sensor on the PEEP circuit complies with the specifications. Calibrate the pressure sensor in the service menu or replace the sensor on the PEEP circuit when a pressure error occurs. 2. Measure the voltage at the corresponding test point. 3. Check the connection between the power supply and the expiratory valve assembly. 4. Replace the VCM when necessary. 5. Replace the expiratory valve assembly when necessary. 6. Replace the power board when necessary. 1. Check whether the inspiratory flow sensor complies with the specifications. Calibrate the flow sensor in the service menu or replace the flow sensor when a flow error occurs. 2. Measure the voltage at the corresponding test point. 3. Check the connection between the power supply and the expiratory valve assembly. 4. Replace the VCM when necessary. 5. Replace the expiratory valve assembly when necessary. 6. Replace the power board when necessary. 1. Measure the voltage at the corresponding test point. 2. Check the connection between the power supply and the expiratory valve assembly. 3. Replace the VCM when necessary. 4. Replace the expiratory valve assembly when necessary. 5. Replace the power board when necessary. Message Flow Sensor Failure Check Flow Sensors Pinsp Not Achieved Vt Not Achieved Patient Circuit Leak Priority Cause Solution 1. Check whether the zero point of the flow sensor is within the normal range. 2. Check whether the values measurement by the flow sensor is within the normal range. 3. Replace the flow sensor and calibrate it. 4. Check the cable connection for the internal flow sensor. Re-plug the cables. Replace the sensor when necessary. 5. Replace the VCM and calibrate it. 1. Check the one-way valve. 2. Check whether the sampling lines of the sensor are connected in correct order. 3. Test the measurement status of the sensor in the diagnostic test tool. 1. Check for breathing circuit leakage. 2. Check the measurement accuracy of the pressure sensor. 3. Calibrate the pressure sensor in case of measurement failure. 4. Replace the VCM and calibrate it. Low 1. The inspiratory flow is out of range. 2. The expiratory flow is out of range. 3. The internal flow sensor is disconnected. High 1. Inspiratory reverse flow. 2. Expiratory reverse flow. Low In pressure mode, Pinsp is less than 2/3 of its setting value (setting value is greater than 9cmH2O) or less than setting value-3cmH2O(setting value is less than 9cmH2O) for continuous 6 breath cycles. Low TVi is less than 20% of the setting TV or 50ml for continuous 6 cycles, whichever is greater. 1. Check for breathing circuit leakage. 2. Check the measurement accuracy of the flow sensor. 3. Calibrate the flow sensor in case of measurement failure. Med 1. During mechanical ventilation, the pressure is less than 2cmH2O for continuous 30s. 2. The patient is not connected. 1. Check whether the circuit and flow sensor are correctly connected. 2. Check whether the tidal volume measured by the flow sensor meets the accuracy requirements. 3. Check for breathing circuit leakage. CO2 Absorber Canister Not Locked High The CO2 absorber canister is not mounted. O2 Sensor Disconnected Low The O2 sensor is disconnected. Replace O2 Sensor Med The O2 concentration value sampling AD is less than 173 AD. 5-11 1. Re-mount the Co2 absorber canister. 2. Check the cable connected between the CO2 absorber canister and the VCM. Replace the cable when necessary. 3. If the problem persists, replace the switch on the CO2 absorber canister. 4. If the problem persists, replace the VCM. 1. Make sure that the cable is correctly connected for the O2 sensor. 2. Check the voltage O2 sensor output in the calibration menu. 3. Replace the O2 sensor. 1. Check the output voltage of the O2 sensor in the calibration menu. Message Priority Cause Low The O2 concentration value exceeds 110% or ranges from 5% to 15% for 3 seconds. Ventilator Comm Stop High The CPU board fails to communicate with the VCM for 10 seconds. Drive Gas Pressure Low High The drive gas pressure is low. O2 Supply Failure High The O2 supply pressure is low. Calibrate O2 Sensor 3-way Valve Failure Low The status of the 3way valve is incorrect. ACGO 3-way Valve Failure Med The status of the 3way valve is incorrect. Auto Ventilation Disabled Low The machine fails the startup self-test and only manual ventilation is applicable. Auto Ventilation Disabled Leak Test Failed Low The automatic circuit leak test fails. Auto Ventilation Disabled High The machine can apply only manual ventilation, but the manual ventilation switch is turned to the auto position. ACGO Failure Med The status of the ACGO switch is Solution 2. Calibrate the O2 sensor again. 3. Replace the O2 sensor. 1. Calibrate the O2 sensor again. 2. Replace the O2 sensor. 1. Restart the machine. 2. Re-plug the communication cable. 3. If the problem persists, replace the VCM. 4. If the problem persists, replace the CPU board. 1. Check whether gas is supplied normally. 2. Short-circuit the pressure switch and the alarm should disappear. Otherwise, it indicates that the pressure switch is defective. Replace the pressure switch. If the pressure switch is not defective, check the connection between the pressure switch and the VCM. 3. If the problem persists, replace the VCM. Use the same method as that of Drive Gas Pressure Low to check the O2 pressure switch. 1. Test the connection of the 3-way valve. 2. Replace 3-way valve. 3. Replace the VCM. Note: this alarm can only be viewed in alarm log and is not displayed on the screen. 1. Test the connection of the 3-way valve. 2. Replace 3-way valve. 3. Replace the VCM. 1. Restart the machine. 2. If the problem persists, check related modules based on the startup self-test results. 3. If the problem persists, replace related modules based on the startup self-test results. 1. Check that the circuit and pipeline are installed correctly. Check whether the sampling port is occluded and whether condensate drain is closed. 2. Perform the leak test again. 1. Set the manual ventilation switch to the manual position. 1. Check whether the drive gas pressure is within the normal range. If not, ensure that 5-12 Message Priority Cause incorrect. Electronic ACGO Undetected Low No electronic ACGO is detected. Aux Control Board Voltage failure Low Protective module 1.3V power supply Pressure monitoring channel failure Med 1.PEEP pressure sensor contraconnection. 2.The measured value of airway pressure sensor or PEEP pressure sensor is beyond the range. 3.Airway pressure sensor or PEEP pressure sensor zero point abnormality. 5-13 Solution the drive gas pressure is within the normal range. Then turn on or off the ACGO switch. 2. If the problem persists, replace the ACGO switch. 3. If the problem persists, replace the CPU board. 4. If the problem persists, replace the VCM. 1. Check whether an electronic ACGO module is configured for the machine. If not, enter password 558188 to access the special service menu, and configure a correct ACGO type based on actual conditions of the machine. 2. If so, check the connections between the ACGO module and the CPU board, and between the ACGO module and the VCM. 3. Replace the electronic ACGO module when necessary. 4. Replace the CPU board or VCM when necessary. 1.Restart the anaesthesia machine. 2.Measure the supply voltage of the aux control board to make sure the supply voltage is within the specific given range. 3.Check if the cables are invalid. If the supply voltage is out of the specific voltage, replace the power supply board or cables. 4.If the problem persists, replace the VCM. 1.Cut off the gas supply and disconnect the gas connection with patients. Observe the sampling value of the pressure sensor. 2.If the zero deviation is large, check if the pressure sampling pipes were contraconnected, blocked or oppressed. 3.Zero calibrating manually. 4.Restart the machine. 5. If the problem persists, replace the VCM. 5.2.6 External AG Module Runtime Alarms Message Priority AG Hardware Error Med O2 Sensor Error Med External AG Selftest Error AG Hardware Malfunction AG Init Error Cause Solution The AG hardware malfunctions. The O2 sensor malfunctions. Replace the AG module. Replace the AG module. Low An error occurs during AG module self-test. 1. Re-plug the AG module. 2. If the problem persists or the alarm occurs repeatedly, replace the AG module. High The AG module is not properly installed or malfunctions. Replace the AG module. High An error occurs during AG initialization. 1. Re-plug the AG module. 2. If the problem persists or the alarm occurs repeatedly, replace the AG module. AG No Watertrap Low AG Watertrap Type Wrong Low The AG watertrap is disconnected from the anesthesia machine. The patient type is infant/neonate, and watertrap type is adult. AG Change Watertrap Med The watertrap needs to be replaced. AG Comm Stop High The AG module malfunctions or the communication fails. AG Airway Occluded High The pump rate of the AG module is below 20 ml/min for more than 1 second. AG Data Limit Error Med AG Zero Failed Low AG Cal. Failed High O2 Accuracy Unspecified Low N2O Accuracy Unspecified Low CO2 Accuracy Unspecified Low The AG module malfunctions. Zeroing for the AG module fails. Calibration for the AG module fails. The measured values exceed the normal range. The measured values exceed the normal range. The measured values exceed the normal range. 5-14 1. Check the AG watertrap. 2. Replace the AG watertrap. 3. Replace the AG module. Change the watertrap type. 1. Check the AG watertrap. 2. Replace the AG watertrap. 3. Replace the AG module. 1. Replace the communication cable of the AG module. 2. Replace the AG module. 1. Check the sampling of the AG module. 2. Replace the AG watertrap. 3. Replace the AG module. Replace the AG module. 1. Zero the AG module again. 2. Replace the AG module. 1. Calibrate the AG module again. 2. Replace the AG module. 1. Calibrate the AG module again. 2. Replace the AG module. 1. Calibrate the AG module again. 2. Replace the AG module. 1. Calibrate the AG module again. 2. Replace the AG module. Message Priority Enf Accuracy Unspecified Low Iso Accuracy Unspecified Low Sev Accuracy Unspecified Low Hal Accuracy Unspecified Low Des Accuracy Unspecified Low Mixed Agent Low Mixed Agent Med Mixed Agent and MAC ≥ 3 Med Cause Solution The measured values exceed the normal range. The measured values exceed the normal range. The measured values exceed the normal range. The measured values exceed the normal range. The measured values exceed the normal range. The AG module detected two and more kinds of halogenated agents. The measured MAC is less than 3. The AG module detected two and more kinds of halogenated agents. It cannot detect the specific MAC value (invalid value). The AG module detected two and more kinds of halogenated agents. The measured MAC is greater than or equal to 3. iCO2 Overrange Low The monitored value exceeds the measurement range. O2 Overrange Low The monitored value exceeds the measurement range. N2O Overrange Low The monitored value exceeds the measurement range. Hal Overrange Low The monitored value exceeds the measurement range. 5-15 1. Calibrate the AG module again. 2. Replace the AG module. 1. Calibrate the AG module again. 2. Replace the AG module. 1. Calibrate the AG module again. 2. Replace the AG module. 1. Calibrate the AG module again. 2. Replace the AG module. 1. Calibrate the AG module again. 2. Replace the AG module. Use only one type of halogenated agent. Use only one type of halogenated agent. Use only one type of halogenated agent. 1. Decrease the concentration of the monitored gas to normal range. 2. Calibrate the AG module again. 3. Replace the AG module. 1. Decrease the concentration of the monitored gas to normal range. 2. Calibrate the AG module again. 3. Replace the AG module. 1. Decrease the concentration of the monitored gas to normal range. 2. Calibrate the AG module again. 3. Replace the AG module. 1. Decrease the concentration of the monitored gas to normal range. 2. Calibrate the AG module again. 3. Replace the AG module. Message Priority Cause Solution Enf Overrange Low The monitored value exceeds the measurement range. Iso Overrange Low The monitored value exceeds the measurement range. Sev Overrange Low The monitored value exceeds the measurement range. Des Overrange Low The monitored value exceeds the measurement range. Rate Overrange Low The monitored value exceeds the measurement range. 1. Decrease the concentration of the monitored gas to normal range. 2. Calibrate the AG module again. 3. Replace the AG module. 1. Decrease the concentration of the monitored gas to normal range. 2. Calibrate the AG module again. 3. Replace the AG module. 1. Decrease the concentration of the monitored gas to normal range. 2. Calibrate the AG module again. 3. Replace the AG module. 1. Decrease the concentration of the monitored gas to normal range. 2. Calibrate the AG module again. 3. Replace the AG module. 1. Decrease the respiratory rate to normal range. 2. Calibrate the AG module again. 3. Replace the AG module. 5.2.7 BIS Module Runtime Alarms Message BIS Init Error BIS Comm Abnormal Priority High High BIS Over Range BIS L Over Range Low BIS R Over Range Cause Solution The system fails to receive information about BIS module initialization. Waveform data packet is not received for continuous 2s; or, data packet of BIS monitored parameters is not received for continuous 4s. The monitored value exceeds the measurement range of the module. The sensor is not properly adhered to the patient skin. The impedance is too high. BIS High Imped. Low BIS Sensor Off Low The sensor falls off from the patient. BIS Electrode 1 High Imped. Low The impedance of BIS electrode 1 is too high. BIS Electrode 1 Lead Off Low BIS electrode 1 falls off from the patient. 5-16 1. Re-plug the BIS module. 2. Replace the BIS module. 1. Check the communication cable between BIS module and main system. 2. Replace the BIS module. Replace the BIS module. 1. Check the contact between BIS sensor and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode 1 and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode 1 and patient skin. 2. Replace with a new BIS sensor. Message Priority Cause Solution BIS Electrode 2 High Imped. Low The impedance of BIS electrode 2 is too high. BIS Electrode 2 Lead Off Low BIS electrode 2 falls off from the patient. BIS Electrode 3 High Imped. Low The impedance of BIS electrode 3 is too high.。 BIS Electrode 3 Lead Off Low BIS electrode 3 falls off from the patient. BIS Electrode 4 High Imped. Low The impedance of BIS electrode 4 is too high.。 BIS Electrode 4 Lead Off Low BIS electrode 41 falls off from the patient. BIS Electrode G High Imped. Low The impedance of BIS electrode G is too high. 1. Check the contact between BIS sensor electrode 2 and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode 2 and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode 3 and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode 3 and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode 4 and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode 4 and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode G and patient skin. 2. Replace with a new BIS sensor. BIS Electrode G Lead Off Low BIS electrode G falls off from the patient. BIS Electrode C High Imped. Low The impedance of BIS electrode C is too high. BIS Electrode C Lead Off Low BIS electrode C falls off from the patient. BIS Electrode LE High Imped. Low The impedance of BIS electrode LE is too high. BIS Electrode LE Lead Off Low BIS electrode LE falls off from the patient. BIS Electrode LT High Imped. Low The impedance of BIS electrodeLT is too high. BIS Electrode LT Lead Off Low BIS electrode LT falls off from the patient. BIS Electrode RE High Imped. Low The impedance of BIS electrode RE is too high. BIS Electrode RE Lead Off Low BIS electrode RE falls off from the patient. 5-17 1. Check the contact between BIS sensor electrode G and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode C and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode C and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode LE and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode LE and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode LT and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode LT and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode RE and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode RE and patient skin. 2. Replace with a new BIS sensor. Message Priority Cause Solution BIS Electrode RT High Imped. Low The impedance of BIS electrode RT is too high. BIS Electrode RT Lead Off Low BIS electrode RT falls off from the patient. BIS DSC Error Low BIS DSC Malf Low BIS No Cable Low BIS No Sensor Low BIS Sensor Too Many Uses Low BIS sensor is overused BIS SQI<50% Low SQI is low. BIS DSC communication error. BIS DSC malfunctions and turns off automatically. BIS cable is not connected. BIS sensor is not connected. SQI is very low, prompting the user that BIS value is not reliable. SQI L is very low, prompting the user that BIS value is not reliable. 1. Check the contact between BIS sensor electrode RT and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor electrode RT and patient skin. 2. Replace with a new BIS sensor. Replace the BIS module. Replace the BIS module. 1. Check the BIS cable. 2. Replace the BIS cable. 1. Check the BIS sensor. 2. Re-install the BIS sensor. 1. Check if the use times of BIS sensor exceeds the restricted times (100 times). 2. Replace the BIS sensor. 1. Check the contact between BIS sensor and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor and patient skin. 2. Replace with a new BIS sensor. BIS SQI<15% Low BIS SQI L<15% Low BIS SQI L<50% Low SQI L is lower. 1. Check the contact between BIS sensor and patient skin. 2. Replace with a new BIS sensor. BIS SQI R<15% Low SQI R is very low, prompting the user that BIS value is not reliable. 1. Check the contact between BIS sensor and patient skin. 2. Replace with a new BIS sensor. BIS SQI R<50% Low SQI R is lower. BIS Sensor Expired Low BIS sensor is expired. BIS Electrode Unconnected Low BISx Disconnected Low BIS Wrong Sensor Type Low BIS sensor electrode is not connected. BIS cable is not connected to the plugin module or communication malfunctions. Mindray undeclared BIS sensor is used. 5-18 1. Check the contact between BIS sensor and patient skin. 2. Replace with a new BIS sensor. 1. Check the contact between BIS sensor and patient skin. 2. Replace with a new BIS sensor. 1. Check the validity period of BIS sensor. 2. Replace the BIS sensor. 1. Check the interface between BIS cable and plug-in module. 2. Replace the BIS cable. 3. Replace the plug-in module of BIS module. 1. Check the type of BIS sensor. 2. Replace the BIS sensor. Message Priority BIS Sensor Failure Low Disconnect/Reconnect BIS Low BIS Self Test Error Low Cause Solution Sensor overcurrent, or positive electrode error. Overcurrent error accumulates to three times and positive electrode error accumulates to twice. It is necessary to power off and then power on the BIS module. The use shall re-plug the BIS module or BIS plug-in box. An error occurs to BIS module selftest. 1. Check the BIS sensor. 2. Replace the BIS sensor. Re-plug the BIS sensor. 1. Re-plug the BIS module. 2. If the problem persists, replace the BIS module. 5.2.8 CO2 Module Runtime Alarms Message Priority Cause CO2 Comm Stop High CO2 module error or Comm error. CO2 Sensor High Temp Low CO2 Sensor Low Temp Low CO2 High Airway Press. CO2 Low Airway Press. CO2 High Barometric CO2 Low Barometric Solution CO2 Sensor Temp too high(greater than 63° C). CO2 Sensor Temp too low(lower than 5 °C) Low Airway Press too high(greater than 790mmHg) Low Airway Press too low(lower than 428mmHg) Low Low Barometric is greater than 790mmHg Barometric is lower than 790mmHg 1.External AD 2.5V Error 2.12V Voltage Error 3.Internal AD 2.5V Error 4.Air pump Error 5.3-Way Valve Error CO2 Hardware Error High CO2 Sampleline Occluded Low Sampleline Occluded CO2 System Error Low multi-System error 5-19 1. Replace the communication cable for CO2 module. 2. Replace the CO2 module. Replace the CO2 module. Replace the CO2 module. 1. Check the airway pressure. 2. Re-install the sampling line of CO2 module. 3. Replace the CO2 module. 1. Check the airway pressure. 2. Re-install the sampling line of CO2 module. 3. Replace the CO2 module. Replace the CO2 module. Replace the CO2 module. Replace the CO2 module. 1. Check the sampling line of CO2 module. 2. Replace the CO2 module. Replace the CO2 module. Message Priority CO2 No Watertrap Low EtCO2 Overrange Low FiCO2 Overrange Low CO2 Zero Failed Low CO2 Init Error High Incompatible CO2 Software Version High CO2 Sensor Error Low CO2 No Sensor Low CO2 Check Airway Low CO2 No Sample line CO2 Main Board Error CO2 Check Sensor or Main Board CO2 Replace Scrubber&Pump Low High Low Low CO2 Replace Sensor Low CO2 15V Voltage Over Range High CO2 Temp Over range Low CO2 Check Cal. Low Cause Solution CO2 No Watertrap or Watertrap disconnected. 1. Check the CO2 watertrap. 2. Re-install the CO2 watertrap. 3. If the problem is not eliminated, replace the CO2 watertrap. 4. If the problem persists, replace the CO2 module. The monitoring value exceeds the measurable range. The monitoring value exceeds the measurable range. Mindray CO2 module Error CO2 Init Error The CO2 module detected is not compatible. The mainstream CO2 module sensor Error The mainstream CO2 module sensor disconnection or communication error MiniMedi CO2 airway abnormal or block MiniMedi CO2 disconnected with sampling pipes. MiniMedi CO2 module Error MiniMedi CO2 module Error MiniMedi CO2 module Error MiniMedi CO2 module Error MiniMedi CO2 module Error MiniMedi CO2 module working temperature is over than that in regular working status. MiniMedi CO2 module Error 5-20 1. Re-calibrate the CO2 module. 2. Replace the CO2 module. 1. Re-calibrate the CO2 module. 2. Replace the CO2 module. 1. Re-zero the CO2 module. 2. Replace the CO2 module. Replace the CO2 module. Replace the CO2 module(not applicable to M02A module). Replace the CO2 module. 1.Check the sensor of the mainstream CO2 module 2. Replace the mainstream CO2 module sensor Replace the CO2 module. 1.Check CO2 module sampling pipe 2.Replace CO2 module sampling pipe 3.Replace the CO2 module. Replace the CO2 module. Replace the CO2 module. Replace the CO2 module. Replace the CO2 module. Replace the CO2 module. Replace the CO2 module. Recalibrate the CO2 module 5.2.9 NMT Module Runtime Alarm Message NMT No Main Cables NMT No Sensor Priority Low NMT Stimulation Electrode off Low NMT stimulating electrode fell off. NMT Sensor Comm. Error Low NMT sensor comm. error NMT Stimulation current over Limit Low NMT Comm. Abnormal High Abnormal NMT communication NMT Comm. Stop High NMT communication stops NMT Init Error High NMT initializes incorrectly. NMT Self Test Error Low NMT Selftest Error Low Cause NMT main cables disconnected. NMT sensor disconnected The stimulating current value of NMT module output overanges the specific value. 5-21 Solution Please check the connection between the main cables and NMT module. 1 Check the connection between the sensor and NMT module to ensure the normal connection. 2 If the alarm persists, please replace the sensor. 1 Check the connection between the stimulating cables and the main cable to ensure the normal normal connection. 2 If the alarm persists, please check the connection between the patient and the stimulating electro. 1 Check the connection between the sensor and the main cable of NMT module to ensure the normal connection. 2 If the alarm persists, please replace the sensor or NMT module 1 Check the connection between the stimulating cable and the main cable to make sure the normal connection between the patient and the stimulating eletro. 2 If the alarm persists, please replace the NMT module or related component. 1 Re-plug the NMT module, main cable, sensor and the stimulating electro, or plug the NMT to the other machine. Eliminate the error situations one by one. 2 If the alarm persists, please replace the NMT module or related component. 1 Re-plug the NMT module, main cable, sensor and the stimulating electro, or plug the NMT to the other machine. Eliminate the error situations one by one. 2 If the alarm persists, please replace the NMT module or related component. 1 Re-plug the NMT module or restart the anaesthesia machine 2 If the alarm persists, please replace the NMT module. 1 Re-plug the NMT module or restart the anaesthesia machine 2 If the alarm persists, please replace the NMT module. NMT Power Error High NMT mainstream disconnected or abnormal communication ST-Ratio Over Range Low TOF-Ratio Over Range Low DBS-Ratio Over Range Low NMT power supply error. ST-Ratio measurement overrange. TOF-Ratio measurement overrange 1 Please check the connection between the main cable and NMT module. 2 If the alarm persists, please replace the NMT module. Replace the NMT module. 1 Recalibrate the sensor. 2 Replace the sensor or NMT module. 1 Recalibrate the sensor. 2 Replace the sensor or NMT module. 5.3 Pneumatic System Problems The pneumatic system is mainly composed of anesthetic gas delivery system, anesthetic agent delivery device (vaporizer), anesthetic ventilator, breathing and anesthetic gas scavenging system. This chapter details possible failures regarding the pneumatic system and how to troubleshoot them. 5.3.1 Tools for on-site Maintenance The tools required for troubleshooting are listed below. The materials listed in the form are not sold separately. Please consult the engineer if any questions. Name Quantity P/N Negative pressure ball Injector(100ml) Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture Vaporizer manifold test fixture 1 MPa(10 bar) Test pressure gauge T-shaped Allen wrench(4*100) 3106-04-06 adapter connector 3106-06-08 adapter connector 3106-10-00 adapter connector 3106-06-00 adapter connector Breathing tube adapter connector 3126-04-00 tube plug 3126-06-00 tube plug 3126-08-00 tube plug 3126-10-00 tube plug Y piece Breathing tube Y piece 3140-08-00 Y piece PU tube (4X200) PU tube (6X100) PU tube (6X200) PU tube (6X300) PU tube (8X200) Breathing tube 5-22 1 1 1 1 1 1 1 1 1 2 1 1 2 3 4 3 2 1 1 1 1 1 1 2 4 040-000814-00* 040-000040-00* 115-002452-00* 115-002456-00* 115-002453-00* 0611-30-67602* M90-100111---* M6Q-030068---* M6Q-030051---* 082-000021-00* M6Q-030059---* 115-002454-00* 082-000023-00* M6Q-120001---* M6Q-120002---* 082-000022-00* M90-100030---* M6Q-030028---* M6Q-030025---* M6G-020046---* M6G-020026---* M6G-020014---* M6G-020017---* Name Quantity P/N Φ6 silicone tube A5 Service Manual Adult test lung Wrench Y-Fitting 15 mm connection Respiration Tube, 0.6 meter silicone, 15mm Breathing Bag 2.3 L silicone Regulator Calibration Hose A5 troubleshooting kit Vaporizer Instruction Manual Safety Analyzer Dempsey 430 or equivalent Digital Volt Meter 3 1/2 digit Agent (and NO2) Analyzer ±0.3 V/V%+5% of reading Digital Pressure Meter BC Biomedical DPM-2301751 NMC Digital Pressure Meter or equivalent Central supplied O2,NO2,AIR Minimum of 35 psi, DISS connections. Cylinder gases O2,NO2,AIR Full PISS yoke connections Gas flow analyzer with accuracy of 2% Communication box (this box is required if FlukeVT Plus gas flow analyzer is used) Lucer adapter connector Ethernet Crossover Cable USB flash drive * = is part of the 115-009450-00 A5 troubleshooting kit. 3 / / / / 2 / / / / / / / / A21-000007---* 046-001140-00 0138-00-0012 0367-00-0080 0103-00-0508 0004-00-0076 0992-00-0139 0453-00-1216 115-009450-00 / / / / / / / / / / / / / / / / / 0012-00-1392-06 0992-00-0297-01 The following pictures show the tools listed above. (1) (2) (3) (4) (5) 5-23 (6) (7) (8) (9) (10) (11) (12) (13) (14) (15) (16) (17) (18) (19) (1)PU tube (4X200);(2)PU tube (6X100);(3)PU tube (6X200);(4)PU tube (6X300);(5)PU tube (8X200);(6)Φ6 silicone tube;(7)Breathing tube; (8)310604-06 adapter connector;(9)3106-06-00 adapter connector;(10)3106-06-08 adapter connector;(11)3106-10-00 adapter connector;(12) Breathing tube adapter connector; (13)3126-04-00 tube plug;(14)3126-06-00 tube plug;(15)3126-08-00 tube plug; (16)3126-10-00 tube plug;(17)Y piece;(18)3140-08-00 Y piece;(19)Breathing tube Y piece; Negative pressure ball: Circuit adapter test fixture: 5-24 Flow sensor pressure sampling pipeline test fixture: Vaporizer manifold test fixture: Anesthesia machine calibration device: 5-25 1 MPa (10bar) test pressure gauge: 5.3.1.1 Precautions for Use of Circuit Adapter Test Fixture There are four connectors for pressure sampling lines and four 6 quick plug-in connectors with number marked on the circuit adapter test fixture, as shown below. Connector for pressure sampling line 6 quick plug-in connector 6 quick plug-in connector The connectors for pressure sampling lines can be connected with Φ6 silicone tubes and the 6 quick plug-in connectors with PU tube (6X100), PU tube (6X200) and PU tube (6X300), as shown below. The circuit adapter test fixture can be mounted either onto the circuit adapter of the anesthesia machine main unit or onto the removed patient circuit. The following pictures show the test fixture mounted in position. 5-26 If it is hard to install and remove the test fixture, apply a layer of lubricant (Krytox highperformance fluorine grease) to the seals (as shown below). Seal (M6M-010058---) Seal (082-000667-00) Seal (082-000665-00) 5.3.1.2 Precautions for Use of Flow Sensor Pressure Sampling Pipeline Test Fixture There are two connectors for pressure sampling lines on the flow sensor pressure sampling pipeline test fixture, as shown below. Connector for pressure sampling line 5-27 The connector for pressure sampling line can be connected with Φ6 silicone tubes. When using the flow sensor pressure sampling pipeline test fixture, remove the expiratory or inspiratory flow sensor from the patient circuit first. Then mount the flow sensor pressure sampling pipeline test fixture onto the position where the expiratory or inspiratory flow sensor was originally mounted and tighten the Inspiration/Expiration Connector Coupling, as shown below. Perform test after connecting the Φ6 silicone tube to the connector for pressure sampling line. 5-28 5.3.1.3 Precautions for Use of Vaporizer Manifold Test Fixture When using the vaporizer manifold test fixture, remove the o-ring seal on the vaporizer manifold assembly. Then slide the test fixture onto the connector, as shown below. After the test fixture is slid on Before the test fixture is slid on Turn the knob clockwise until the bottom surface of the pressure head is in contact with the top surface of the connector, as shown below. Contact here when mounted 5-29 5.3.1.4 Precautions for Use of Negative Pressure Ball Besides one sealing cover, the negative pressure ball also has two one-way valves at its front end, as shown below. The built-in one is connected with the gas inlet of the ball which permits the gas come in only, and the exterior one only permits the gas come out. If the front sealing cover is removed or loosened, the sealing performance of the negative pressure ball will compromise. In this case, you need to tighten the sealing cover. Sealing cover Tighten the sealing cover Gas inlet External one-way valve (gas outlet) When the negative pressure ball is connected with the tested component, the ball permits the gas in only, but meanwhile it’s free to release air when it’s pressed. Before using the negative pressure ball, make sure that it is not leaky. Check if the front sealing cover is tightened. Then flatten the negative pressure ball to remove the gas inside. Install the gas outlet plug properly. Block the front gas inlet with your finger then release the ball. Ball should not visibly inflate for at least 30 seconds. If it does, replace the ball. 5-30 5.3.2 Gas Supplies and Drive Gas The following table lists gas supplies and drive gas related failures. Failure description Possible cause Recommended action Leak The gas supply tube is damaged or the seal at the connection is damaged. The quick plug-in connector leaks. Replace the gas supply tube or the seal at the connection. The pipeline gas supply inlet assembly leaks. The drive gas pipeline leaks. Pipeline pressure gauge shows inaccurate readings or no readings. The readings on the pipeline pressure gauge fluctuate greatly. The pipeline pressure gauge is damaged. No “O2 Supply Failure” alarm occurs when the O2 pressure is low or this alarm occurs when the O2 supply pressure is within specifications. No “Drive Gas Pressure Low” alarm occurs when the drive gas pressure is low or this alarm occurs when the drive gas pressure is within specifications. The gas pressure switch of the O2 supply inlet assembly is ineffective. The filter of pipeline gas supply inlet assembly or the PU tube of the pipeline pressure gauge is occluded or the pressure gauge is damaged. The pressure switch on the integrated pneumatic circuit of the expiratory valve assembly or the PEEP safety valve is ineffective. Or, the filter on the integrated pneumatic circuit of the expiratory valve assembly is occluded. 5-31 Replace the quick plug-in connector or PU tube (when the PU tube is not damaged, if the tube is long enough, cut off a small segment of the tube where the quick plug-in connector is met, and then insert the tube into position). Check if the check valve of the pipeline gas supplies inlet assembly leaks in the reverse direction. Replace it if necessary. Check and replace the damaged seal of the pipeline gas supplies inlet assembly. If the problem persists, replace the pipeline gas supplies inlet assembly. Check and repair the expiratory valve assembly as per the procedures described in 5.3.4.2 Leak Test of Low-pressure Pneumatic System. Replace the pipeline pressure gauge. 1. After confirming that the pipeline gas pressure is stable, check the PU tube of the pipeline pressure gauge and filter of the pipeline gas supply inlet assembly. If the tube or the filer is occluded, replace it. 2. If the problem persists, replace the pipeline pressure gauge. Adjust the pressure switch of the O2 supply inlet assembly to cause O2 supply pressure to approach 0.2 MPa as much as possible within the range of 0.15 to 0.25 MPa when this alarm occurs. If the adjustment fails, replace the pressure switch (refer to 5.3.2.3 Adjust the Pressure Switch). Adjust the pressure switch on the integrated pneumatic circuit of the expiratory valve assembly to cause drive gas pressure to approach 0.14 MPa as much as possible within the range of 0.05 to 0.2 MPa when this alarm occurs. If the adjustment fails, replace the pressure switch. If the problem persists after the pressure switch is replaced, replace the integrated pneumatic circuit of the expiratory valve assembly (refer to 5.3.2.3 Adjust the Pressure Switch)。 5.3.2.1 Test the Pipeline Pressure Gauge and Correct the Regulator Use the following tools to test the pipeline pressure gauge and regulator of the pipeline gas supply inlet assembly: 1 MPa (10bar) test pressure gauge (before the test, make sure that the 1 MPa (10bar) test pressure gauge is in good condition) (quantity: 1) 3106-04-06 adapter connector (quantity: 1) PU tube (4X200) (quantity: 1) PU tube (6X200) (quantity: 1) Test procedures: I O2 supply inlet assembly: 1. Turn off the pipeline gas supply and bleed the residual pressure through O2 flushing. 2. Disconnect tube 57. The end of the tube which connects the auxiliary O2 supply is not pulled out but the end to Y piece is pulled out. 3. Connect 1MPa test pressure gauge to the above Y piece through “3106-04-06 adapter connector”. 4. 5. 6. 7. Turn on O2 pipeline supply and record the reading on the O2 pipeline pressure gauge. Observe the test pressure gauge. If the reading on the test pressure gauge is not within the range of 0.15 to 0.25 MPa (namely 1.5 to 2.5bar), adjust the regulator of the O2 supply inlet assembly to cause the reading on the test pressure gauge to reach 0.2 MPa (namely, 2bar). For operations of the regulator, refer to 5.3.2.4 Adjust the Regulator of the Pipeline Gas Supply Inlet Assembly. Turn off the pipeline gas supply and bleed the residual pressure through O2 flushing. Reconnect tube 57. Disconnect tube 39 which connects the O2 supply inlet assembly to the O2 pipeline pressure gauge. Remove the end of tube to the O2 supply inlet assembly. 5-32 8. Connect 1MPa test pressure gauge to the outlet of O2 supply inlet assembly through “3106-04-06 adapter connector”. 9. Turn on the pipeline gas supply and record the reading on the test pressure gauge. If the difference between this reading and the reading on the O2 pipeline pressure gauge is more than 0.1 MPa (1bar), it indicates that the O2 pipeline pressure gauge is damaged. Handle this problem as described in the troubleshooting table. 10. Reconnect tube 39. Note 1: For numbers of all PU tubes, refer to 1.2.2 Pneumatic Connection and 1.2.3 Connections between Pneumatic Circuit, Breathing System and Ventilator Control Board. II N2O supply inlet assembly: 1. Turn off the pipeline gas supply. Open the needle valve to bleed the residual pressure and close the needle valve. Disconnect tube 48. The end of the tube which connects the ORC is pulled out but the other end is not pulled out. 2. Connect 1MPa test pressure gauge, the pulled-out end of tube 48, and ORC N2O outlet through “3106-04-06 adapter connector” and Y piece (17). 3. 4. 5. Turn on N2O and O2 pipeline supplies. Adjust the regulator of the N2O supply inlet assembly to the same value as measured/set for the O2 supply inlet assembly (see step 4 of O2 supply inlet assembly) Record the reading on the N2O pipeline pressure gauge. Turn off N2O pipeline supply and bleed the residual pressure by opening the N2O flow regulator. Reconnect tube 48. 5-33 6. 7. 8. 9. Pull out No.40 PU tube which connects the N2O supply inlet assembly to the N2O pipeline pressure gauge. Remove the tube end which connects N2O supply inlet assembly. Connect 1MPa test pressure gauge to the outlet of N2O supply inlet assembly through “3106-04-06 adapter connector”. Turn on the N2O pipeline supply and record the reading on the test pressure gauge. If the difference between this reading and the reading on the N2O pipeline pressure gauge is more than 0.1 MPa (1bar), it indicates that the N2O pipeline pressure gauge is damaged. Handle this problem as described in the troubleshooting table. Reconnect tubing to pressure gauge. Reconnect tube 40. III AIR supply inlet assembly: 1. Turn off the pipeline gas supply. Disconnect tube 67. The end of the tube which connects auxiliary gas supply is not pulled out but the other end which connects the Y piece is pulled out. 2. Connect 1MPa test pressure gauge to the above Y piece through “3106-04-06 adapter connector”. 3. 4. Turn on AIR pipeline supply. If the reading on the test pressure gauge is not within the range of 0.2±0.05 MPa (2.0±0. 5bar), adjust the regulator to cause the reading on the test pressure gauge to reach 0.2 MPa (2.0bar). Record the reading on the AIR pipeline pressure gauge. Turn off AIR pipeline supply and bleed the residual pressure by opening the AIR flow regulator. 5-34 5. 6. 7. 8. 9. Re-connect tube 67. Pull out No.41 PU tube which connects the AIR supply inlet assembly to the AIR pipeline pressure gauge. Remove the tube end which connects AIR supply inlet assembly. Connect 1MPa test pressure gauge to the outlet of AIR supply inlet assembly through “3106-04-06 adapter connector”. Turn on the AIR pipeline supply and record the reading on the test pressure gauge. If the difference between this reading and the reading on the AIR pipeline pressure gauge is more than 0.1 MPa (1.0 bar), it indicates that the AIR pipeline pressure gauge is damaged. Handle this problem as described in the troubleshooting table. Reconnect PU tube No.41 to the pressure gauge. 5.3.2.2 Test the Pressure Switch Use the following tools to verify the pressure switches of the O2 supply inlet assembly and the expiratory valve assembly are within specifications: 1 MPa (1.0 bar) test pressure gauge (quantity:1) 3106-04-06 adapter connector (quantity:1) 3106-06-08 adapter connector (quantity:1) 3140-08-00 Y piece (quantity:1) PU tube (8X200) (quantity:2) PU tube (6X200) (quantity:1) PU tube (4X200) (quantity:1) Test procedures: 1. Turn off the pipeline gas supply and bleed the residual pressure by pushing the O2 flush button. 2. Disconnect tube 47. The end of the tube which connects pressure regulator assembly is pulled out but the other end is not pulled out. 3. Connect one PU tube (8X200) to the O2 inlet of pressure regulator assembly and connect the other end of the PU tube and also the pulled-out end of tube 47 to the two connectors of “3140-08-00 Y piece” respectively. 4. Connect the test pressure gauge to the third connector of “3140-08-00 Y piece” through “3106-06-08 adapter connector” and “3106-04-06 adapter connector”. 5-35 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. Turn on the O2 pipeline supply. Turn on the machine to enter Standby. Turn off all flow regulators. Turn off the pipeline gas supply (if the reading on the test pressure gauge begins to fall dramatically and continuously after the gas supply is turned off, it indicates that there are one or more leaks in the O2 supply inlet assembly, expiratory valve assembly, O2 flush button assembly, system switch assembly, and/or the O2 flow regulator. Perform the subsequent operations after the leaks are serviced. Failures can be located by using the methods described in section 5.3.3 Anesthetic Gas Delivery System and 5.3.4 Breathing System except O2 supply inlet assembly related failures). Manually adjust the O2 flow regulator until O2 flow is approximately 1 L/min, causing the reading on the test pressure gauge to fall gradually to 0.25 MPa (2.5bar). Turn off O2 flow to cause the reading on the test pressure gauge not to fall. If the “O2 Supply Failure” alarm occurs 10 seconds later, it indicates that the pressure switch of the O2 supply inlet assembly is defective. Troubleshoot this problem as described in the relevant failure table. Adjust the O2 flow regulator until O2 flow is approximately 0.5 L/min, causing the reading on the test pressure gauge to fall gradually to 0.2 MPa (2bar). Turn off O2 flow to cause the reading on the test pressure gauge not to fall. If the “Drive Gas Pressure Low” alarm occurs 10 seconds later, it indicates that the pressure switch on the integrated pneumatic circuit of the expiratory valve assembly is defective. Troubleshoot this problem as described in the relevant failure table. Adjust the O2 flow regulator until O2 flow is approximately 0.3 L/min, causing the reading on the test pressure gauge to fall gradually to 0.15 MPa (1.5bar). Turn off O2 flow to cause the reading on the test pressure gauge not to fall. If the “O2 Supply Failure” alarm does not occur 10 seconds later, it indicates that the pressure switch of the O2 supply inlet assembly is defective. Troubleshoot this problem as described in the relevant failure table. Adjust the O2 flow regulator until O2 flow is approximately 0.3 L/min, causing the reading on the test pressure gauge to fall gradually to 0.05 MPa (0.5bar). Turn off O2 flow to cause the reading on the test pressure gauge not to fall. If the “Drive Gas Pressure Low” alarm does not occur 10 seconds later, it indicates that the pressure switch on the integrated pneumatic circuit of the expiratory valve assembly is defective. Troubleshoot this problem as described in the relevant failure table. 5-36 5.3.2.3 Adjust the Pressure Switch Adjust the O2 supply pressure switch and drive gas pressure switch as described below. Use a flathead screwdriver to adjust the O2 supply pressure switch as shown below. Turn for small degrees each time such as 30 degrees. Note that turning the pressure switch clockwise will decrease its alarm limits and counterclockwise increase its alarm limits. Test the assembly after each pressure adjustment is made. Repeat until the pressure switch is properly adjusted and is within specification (nominal value 220kPa+/-10kPa). 5.3.2.4 Adjust the Regulator of the Pipeline Gas Supply Inlet Assembly Pull up the knob cover of the regulator. Turn the cover clockwise to increase pressure or counterclockwise to decrease pressure, as shown below. Bleed the inside pressure of the pipeline gas supply inlet assembly after each pressure adjustment is made. Then, turn on the pipeline gas supply again. Observe the adjusted pressure through the test pressure gauge. Adjust it to 200kPa. 5-37 5.3.3 Anesthetic Gas Delivery System The following table lists anaesthetic gas delivery system related failures. Failure description Possible cause Recommended action Leak The O2 flush button assembly leaks. The system switch assembly leaks. The vaporizer is installed improperly, which results in leak. The seal between the vaporizer manifold assembly and the vaporizer is damaged. The seal between the vaporizer manifold inside and the connection or the rubber plain washer between the vaporizer manifold inside and the spring is damaged or dirty. The vaporizer manifold assembly is damaged. The total flowmeter leaks. The ORC assembly leaks. The flow regulator leaks. The restrictor leaks. The pressure relief valve at the breathing connection leaks. The ACGO assembly leaks. The fresh gas connections of the circuit adapter assembly leak. The gas supplies cannot be turned off after the machine is turned off. The machine cannot be powered on after turned on. The flowmeter float indicates inaccurate value or remains unmoved. The knob of the flow regulator gets loose. N2O supply cannot be cut off in case of O2 supply failure. The seal inside the system switch assembly is damaged. Replace the seal on the O2 flush button assembly or replace the O2 flush button assembly. Replace the seal on the system switch assembly or replace the systems switch assembly. Re-install the vaporizer. Clean or replace the seal. The seal should be replaced at least once per year as required. Clean the sealing part or replace the damaged seal and rubber plain washer. Replace the vaporizer manifold assembly. Replace the total flowmeter. Replace the ORC assembly. Replace the flow regulator. Re-calibrate after the restrictor is replaced (for calibration, refer to “Instructions of Use for FPM-65 Flow and Pressure Detection Device” Check and replace the defective pressure relief valve. Replace the ACGO assembly. Check the seals and tubes at the fresh gas connections. Replace the defective parts and re-install the parts. Replace the system switch. The contact switch is ineffective. The float of total flowmeter is damaged. Replace the contact switch of the system switch assembly. Replace the float of total flowmeter. The flow regulator is damaged. The ORC assembly is damaged. Replace the flow regulator. 5-38 Replace the ORC assembly. 5.3.3.1 Leak Test of the O2 Flush Button Assembly Perform a leak test of the O2 flush button assembly by using the following tools: Negative pressure ball (quantity:1) 3106-06-00 adapter connector (quantity:1) PU tube (6X100) (quantity:1) Test procedures: 1. Turn off the pipeline gas supplies and bleed the residual pressure through O2 flushing. 2. Remove the work surface. Pull out No.52 PU tube which connects the O2 flush button assembly to the ACGO assembly. Disconnect at the ACGO end. 3. Connect the inlet of the negative pressure ball to PU No.52 through 3106-06-00 adapter connector and then flatten the negative pressure ball to remove the gas inside. 4. Release the negative pressure ball as shown below. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the O2 flush button assembly is damaged. 5.3.3.2 Leak Test of the Flowmeter Related Assembly Perform a leak test of the flowmeter related assembly (from flow regulator to total flowmeter) by using the following tools: Negative pressure ball (quantity:1) 3106-06-00 adapter connector (quantity:1) 3106-06-08 adapter connector (quantity:1) 3126-06-00 tube plug (quantity:3) 3126-08-00 tube plug (quantity:1) PU tube (6X100) (quantity:1) Test procedures: 1. Turn off the pipeline gas supplies and turn on the system switch. Bleed the residual pressure by opening the flow regulators. 2. Turn off the system switch. Turn on the flow regulators and turn them counterclockwise for more than half a circle. 3. Pull out No.25 PU tube which connects the total flowmeter to the vaporizer manifold assembly. Disconnect at the vaporizer manifold end. 5-39 4. 5. Pull out No.45, 49 and 51 PU tubes which connect with the flow regulator. Disconnect at flow regulator end. Occlude the pulled-out tube end on the flow regulator by using three 3126-06-00 tube plugs. 6. Connect the other end of the negative pressure ball to the pulled-out end of No.25 PU tube through 3106-06-08 adapter connector, as shown below, and then flatten the negative pressure ball to remove the gas inside. 7. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the total flowmeter assembly is damaged. In this case, pull out the tube plug at the inlet of needle valve and perform the following operations. Pull out No.26 PU tube which connects the back pressure regulator to the total flowmeter. Disconnect at the total flowmeter end. 8. 5-40 9. Occlude the pulled-out tube end on the total flowmeter by using 3126-08-00 tube plug. 10. Compress the negative pressure ball still connected to tube 25 to bleed the air inside. 11. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the total flowmeter leaks. 12. Re-connect tube 25. Disconnect tube 58 and pull out its end connecting the back pressure valve. Disconnect tube 26 and pull out its end connecting the total flowmeter. 13. Occlude the pulled-out tube end on the back pressure regulator by using one 3126-08-00 tube plug and connect the negative pressure ball to the pulled-out end of No.26 PU tube through 3106-06-08 adapter connector. 5-41 14. Compress the negative pressure ball to bleed the air inside. 15. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the back pressure regulator leaks. 16. Pull out No.27, 28 and 29 PU tubes which connect the flow sensor to the gas mixer. Disconnect at the gas mixer end. 17. Occlude the pulled-out tube end on the gas mixer by using three 3126-06-00 tube plugs. 18. Disconnect tube 58 and plug out its end connecting the back pressure valve. Connect the negative pressure ball to the pulled-out end of tube 58 through “3106-06-08 adapter connector”. 19. Flatten the negative pressure ball to remove the gas inside. 20. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the gas mixer leaks. 21. For N2O branch, disconnect tube 32 and pull out its end connecting the sensor (disconnect tube 31 for AIR branch and tube 78 for O2 branch. 22. Occlude the pulled-out tube end on the flow sensor by using 3126-06-00 tube plug. 5-42 23. For N2O branch, disconnect tube 27, pull out its end connecting gas mixer and connect it to the gas inlet of the negative pressure ball through “3106-06-00 adapter connector” (disconnect tube 28 for AIR branch and tube 29 for O2 branch). Compress the negative pressure ball to bleed the gas inside. 24. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the flow sensor leaks. 25. For N2O branch, disconnect tube 49 and pull out its end connecting the needle valve (disconnect tube 51 for AIR branch and tube 45 for O2 branch). 26. Occlude the pulled-out tube end on the needle valve by using 3126-06-00 tube plug. 5-43 27. For N2O branch, disconnect tube 91, pull out its end connecting the needle valve, and connect the gas inlet of the negative pressure ball to the gas outlet of the needle valve through “3106-06-00 adapter connector” and PU tube (6X200) (disconnect tube 90 for AIR branch and tube 73 for O2 branch). Compress the negative pressure ball to bleed the gas inside. 28. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the needle valve leaks. 5-44 5.3.3.3 Leak Test of the System Switch Assembly Perform a leak test of the system switch assembly by using the following tools: Negative pressure ball (quantity:1) 3106-06-00 adapter connector (quantity:1) 3106-06-08 adapter connector (quantity:1) 3126-08-00 tube plug (quantity:1) PU tube (6X100) (quantity:1) Test procedures: 1. Turn off the pipeline gas supplies and turn on the system switch. Bleed the residual pressure by opening the flow regulators. 2. Pull out No.45 which connects the system switch assembly to the flow regulator. Disconnect at the flow regulator end and connect the pulled-out tube end to the negative ball through one 3106-06-00 adapter connector. 3. Pull out No.43 which connects the system switch assembly to the Y piece. Disconnect at the Y piece end and occlude the pulled-out tube end through one 3106-06-08 adapter connector and one 3126-08-00 tube plug. 4. Flatten the negative pressure ball to remove the gas inside. 5. 6. 7. 8. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that tube connected with the system switch assembly is damaged. Turn off the system switch. Pull out the 3126-08-00 tube plug which was used to occlude tube No.43 before. Flatten the negative pressure ball to remove the gas inside. 5-45 9. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds during one of the two tests, it indicates that the system switch assembly is damaged. 5.3.3.4 Leak Test of the Oxygen Ratio Controller (ORC) Perform a leak test of the ORC assembly by using the following tools: Negative pressure ball (quantity:1) 3106-06-00 adapter connector (quantity:1) 3126-06-00 tube plug (quantity:2) 3126-04-00 tube plug (quantity:1) PU tube (6X100) (quantity:1) Test procedures: 1. Disconnect tube 3 and pull out the end of the tube which connects to the Y piece. Occlude the pulled-out end with “3106-06-00 adapter connector and 3126-06-00 tube plug. 2. Disconnect tube 46 and pull out the end of the tube which connects to the Y piece. Connect the pulled- out end to the negative pressure ball directly. 3. Flatten the negative pressure ball to remove the gas inside. PU tube No.33 PU tube No.46 4. 5. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the Oxygen Ratio Controller is damaged. Handle this problem as described in the troubleshooting table. If not, continue the following test. Re-connect the tube. 5-46 6. Disconnect tubes 48, 49, and 32. Pull out the tube ends which connect the ORC. Occlude the pulled-out ends with two 3126-06-00 tube plugs and one 3126-04-00 tube plug. Disconnect tube 30. The end of the tube which connects the ORC is not pulled out but the other end is pulled out and is connected to the negative pressure ball through “3106-06-00 adapter connector”. 7. 8. Flatten the negative pressure ball to remove the gas inside. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the Oxygen Ratio Controller is damaged. Handle this problem as described in the troubleshooting table. 5-47 5.3.3.5 Leak Test of the Vaporizer Manifold Assembly Perform a leak test of the vaporizer manifold assembly by using the following tools: Negative pressure ball (quantity:1) 3106-06-08 adapter connector (quantity:1) 3126-06-00 tube plug (quantity:1) PU tube (6X100) (quantity:1) PU tube (8X200) (quantity:1) Vaporizer manifold test fixture (quantity:1) Test procedures: 1. Turn off the system switch. 2. Remove the vaporizer. 3. 4. 5. 6. Pull out No.25 PU tube which connects the total flowmeter to the vaporizer manifold assembly. Disconnect at the vaporizer manifold end and occlude it with 3126-08-00 tube plug Pull out No.55 PU tube which connects the vaporizer manifold assembly to the ACGO assembly. The end of the tube which connects the vaporizer manifold assembly is pulled out, and connected with the negative ball through one 3106-06-08 adapter connector Flatten the negative pressure ball to remove the gas inside. Release the negative pressure ball. If the negative pressure ball is fully expanded within 30s, it indicates that the rubber plain washers or its upper surface contacted mechanical surface are damaged. Handle this problem as described in the troubleshooting table. If not, continue the following tests. 5-48 7. Remove the seal ring, and mount the vaporizer manifold test fixture onto the connector of the vaporizer manifold assembly (remove the seal between the connector and the vaporizer when mounting the test fixture) 8. Repeat step 5 and step 6 every time when the vaporizer manifold test fixture is transferred to the next position. Once the negative pressure ball is fully expanded within 30s, it indicates that the rubber plain washers or its lower surface contacted mechanical surface are damaged. Handle this problem as described in the troubleshooting table. If the four tests are all past, then continue the following tests. Put the seal ring back, mount the vaporizer and turn it on. 9. 10. Repeat step 5 and step 6 every time when the vaporizer manifold test fixture is transferred to the next position. Once the negative pressure ball is fully expanded within 30s, it indicates that the seal rings are damaged. If the two tests are both past, it indicated that the vaporizer manifold assembly and the four seal rings are in good condition. 5-49 5.3.4 Breathing System The following table lists breathing system related failures. Failure description Leak Possible cause Recommended action The sodalime canister is not installed properly. Re-install the sodalime canister. Remove the sodalime at the sealing connection. Ensure the correct installation of sadalime canister. Replace the sealing component of the sodalime canister assembly. It is required to replace the seal once a year. The sealing piece for the absorbent canister assembly is damaged, including the two sealing cushions (049-000142-00 and 049-000145-00)which are in direct contact with the absorbent canister and the two sealing rings (082-000629-00) on the bypass upper cover which are in contact with the circuit bottom housing. The seal for the bag arm is damaged. The water collection cup gets loose. The seal for the water collection cup assembly is damaged. The seal for the circuit adapter assembly is damaged. The bellows housing or bellows is not installed properly. The bellows sealing cushion falls off or is damaged. The valve cover of the breathing valve assembly is not installed properly. The seal for the valve cover of the breathing valve assembly is damaged. The O2 sensor is not installed properly. The seal for the O2 sensor or the seal for the O2 sensor plug is damaged. The breathing tube connecting the patient is damaged. The bellows is damaged. 5-50 Replace the seal for the bag arm. It is required to replace the seal once a year. Check and tighten the water collection cup. Replace the seal for the water collection cup assembly. It is required to replace the seal once a year. Replace the seal, which is required to be replaced once a year. Re-install the bellows housing or bellows. Ensure their correct installation. Replace the bellows sealing cushion, which is required to be replaced once a year. Re-install the valve cover and ensure its correct installation. Replace the seal. Re-install the O2 sensor and ensure its correct installation. Replace the seal. Replace the breathing tube. Replace the bellows, which is required to be replaced once a year. Failure description O2 concentration measurement fails or has great deviations. The airway pressure gauge shows inaccurate reading or its pointer cannot move. Irregular flow wave is displayed. Possible cause Recommended action The sealing connection of other parts of the breathing system is damaged. Repair or replace the sealing connection as per the procedures described in 5.3.2.4. The condensate valve of the canister assembly is not installed properly or the seal inside is damaged. There is water built up on the measurement surface of O2 sensor. The O2 sensor is not calibrated. Re-install the condensate valve or replace the damaged seal inside. The O2 sensor is damaged. Remove the built-up water and allow the O2 sensor to air dry. Calibrate the O2 sensor as per 4.3.6O2 Sensor Calibration. Replace the O2 sensor. The airway pressure gauge is damaged. Replace the airway pressure gauge. The flow sensor assembly is not installed properly. There is water built up inside the flow sensor assembly. The membrane of the flow sensor assembly is distorted, dirty or its inside resistance changes. Zero drift occurs to the pressure sensor of the fresh flow sensor board. The flow sensor is damaged. The pressure sensor on the fresh gas flow sensor board is defective. The flow sensor pressure sampling pipeline leaks. Re-install the flow sensor assembly. Remove the flow sensor assembly and clear its inside water build-up. Enter the service mode and calibrate the flow sensor as per 4.3.2. Replace the flow sensor assembly. Replace the fresh gas flow sensor board. Repair the flow sensor pressure sampling pipeline after checking as per the procedures described in 5.3.4.1. 5.3.4.1 Leak Test of Flow Sensor Pressure Sampling Pipeline If irregular flow waveform is displayed, the flow sensor pressure sampling pipeline may be leaky. Perform the leak test by using the following tools: Anesthesia machine calibration device (quantity:1) Flow sensor pressure sampling pipeline test fixture (quantity:1) Circuit adapter test fixture (quantity:1) Injector (quantity:1) Φ6 silicone tube (quantity:3) Y piece (quantity:1) 5-51 Test procedures: I Leak test of the flow sensor pressure sampling pipeline (the four sampling pipelines of the expiratory and inspiratory flow sensors are all tested) 1. Turn off the system switch. 2. Install the breathing circuit properly. 3. Remove the flow sensor assembly. 4. Mount the flow sensor pressure sampling pipeline test fixture onto the position where the flow sensor assembly was originally mounted. Tighten the breathing connector rotary cap. 5. Connect the Φ6 silicone tubes to the pressure sensor connector (positive pressure end) on the anesthesia machine calibration device), injector (before mounting, pull out the push rod of the injector) connector and the connector for the flow sensor pressure sampling pipeline test fixture by using a Y piece, as shown below. 6. Push in the push rod of the injector to let the pressure reading on the anesthesia machine calibration device rise to 70 to 90 cmH2O and then stop pushing. Keep the relative position between the push rod and the injector unchanged. If the pressure reading on the anesthesia machine calibration device does not fall more than 5cmH2O within 15 s, this test is passed. 5-52 II Leak test of the flow sensor pressure sampling pipeline inside the main unit (perform this test if test “I” fails) 1. Mount the circuit adapter test fixture onto the circuit adapter assembly. 2. Connect the Φ6 silicone tubes to the pressure sensor connector (positive pressure end) on the anesthesia machine calibration device), injector (before mounting, pull out the push rod of the injector) connector and the connector (one connector out of No.3 through 6 connectors on the test fixture) for the circuit adapter test fixture by using a Y piece, as shown below. 3. Push in the push rod of the injector to let the pressure reading on the anesthesia machine calibration device rise to 70 to 90 cmH2O and then stop pushing. Keep the relative position between the push rod and the injector unchanged. If the pressure reading on the anesthesia machine calibration device does not fall more than 5cmH2O within 15s, this test is passed. If test “I” is failed and “II” passed, it indicates that the flow sensor pressure sampling pipeline on the breathing system is damaged. In this case, replace the breathing system. If both tests “I” and “II” are failed, check the sampling lines and connectors inside the main unit, seals and solenoid valve of the circuit adapter assembly until test “II” is passed. Then perform test “I”. If test “I” is still failed, it indicates the flow sensor pressure sampling pipeline on the breathing system is damaged. In this case, replace the breathing system. 5.3.4.2 Leak Test of Low-pressure Pneumatic System After making sure that the flow sensor pressure sampling pipeline is not leaky, perform leak tests of the low-pressure pneumatic system as shown in the following figures (figures a through d). 1. Leak test of the breathing system in the mechanical ventilation mode Pass Figure b Leak test of the part only working in the manual ventilation mode Figure d Leak test of the breathing system Fail Pass Re-install the bellows housing and sodalime canister (clean the canister mouth and sealing cushion) and check the folding bag. Completed 1. Leak test of the breathing system in the mechanical ventilation mode 5-53 Pass Fail 4.Leak test of all pipelines on the circuit adapter Fail Figure c Leak test of the pipelines inside the main unit Figure a System leak test Pass 2. Leak test of the breathing system in the manual ventilation mode Fail Check the manual bag and the seal for bag arm Completed 7. Check the APL valve. support tube on the medium plate, and Completed sealing components of the Auto/ Manual ventilation switch assembly Figure b Leak test of the part only working in the manual ventilation mode Pass Fail Note: After repairing the faulty part, repeat the latest failed test prior to the current step. If it is still failed, continue leak test following the flow chart. But you can skip the step to which the passed test or repaired part corresponds. For example, after the manual bag is checked, found damaged and replaced, you should return to “Leak test of the breathing system in the manual ventilation mode” at the very beginning. If“2”is still failed, you can skip “Check the manual bag and the seal for bag arm”and directly go to the next step. Replace the patient circuit End Figure b Leak test of the part only working in the manual ventilation mode Fail to pass the leak test of fresh gas and ACGO pipelines Fail to pass the leak test of drive gas pipelines Figure b Leak test of the part only working in the manual ventilation mode 8. Check the ACGO assembly Pass Completed Check the expiratory valve assembly 4.Leak test of all pipelines on the circuit adapter Pass Fail Check the sealing components of the circuit adapter assembly Figure c Leak test of the pipelines inside the main unit 5-54 1. Leak test of the breathing system in the mechanical ventilation mode Fail Figure d Leak test of the breathing system Check the lifting device and sealing components of the sodalime canister Fail 2. Leak test of the breathing system in the manual ventilation mode Fail 6. Check the accessories and circuit connectors Completed 2. Leak test of the breathing system in the manual ventilation mode Fail 5. Check the lifting device and sodalime canister Pass Completed Pass Pass 1. Leak test of the breathing system in the auto ventilation mode End 7. Check the APL valve. support tube on the medium plate, and sealing components of the Auto/ Manual ventilation switch assembly Pass 2. Leak test of the breathing system in the manual ventilation mode Fail Replace the patient circuit Fail 3. Check the bellows housing and bag position of the Auto/Manual ventilation switch Fail Check the bellows housing, bellows sealing cushion, and seal for lower cover 2 assembly Completed 1. Leak test of the breathing system in the auto ventilation mode Pass End Fail Check the Auto/Manual ventilation switch 1. Leak test of the breathing system in the auto ventilation mode Pass Pass Fail End Note: After repairing the faulty part, repeat the latest failed test prior to the current step. If it is still failed, continue leak test following the flow chart. But you can skip the step so which the passed test or repaired part corresponds. Replace the patient circuit Figure d Leak test of the breathing system 1. 2. Leak test in the mechanical ventilation mode Perform leak test as described in 3.6.2. Leak test of the breathing system in the manual ventilation mode Tools required: Breathing tube (quantity: 3) Breathing tube Y piece (quantity: 1) Test procedures: (1) Let the system enter Standby. (2) Mount the breathing circuit properly. (3) Set the Auto/Manual switch to the Manual position. (4) Set the pressure value of the APL valve to maximum. 5-55 (5) Occlude the patient port and bag arm port by using three breathing tubes and one Y piece as shown below. (6) Turn on the O2 flow regulator and adjust O2 flow to 0.2L/min, (7) Push the O2 flush button to let the reading on the Paw pressure gauge rise to 30cmH2O. (8) Stop O2 flushing. If the reading on the Paw pressure gauge falls under 30cmH2O, this test is failed. (9) If the reading on the Paw pressure gauge rises rapidly, to prevent defective APL valve from damaging the Paw pressure gauge, note to turn off the O2 flow regulator timely to prevent the overrange of the Paw pressure gauge (The test which involves O2 flow regulator turned off due to this reason is considered to be passed). 3. Check the bellows housing and the Manual position of the Auto/Manual switch Tools required: Anesthesia machine calibration device (quantity: 1) Circuit adapter test fixture (quantity: 1) Injector (quantity: 1) Φ6 silicone tube (quantity: 2) PU tube (6X300) (quantity: 1) Y piece (quantity: 1) Test procedures: (1) Remove the bellows. (2) Mount the bellows housing properly. (3) Set the Auto/Manual switch to the Manual position. (4) Remove the breathing circuit (5) Mount the circuit adapter test fixture onto the breathing circuit. 5-56 (6) Connect the Φ6 silicone tubes and PU tube (6X300) to the injector connector, pressure sensor (of the anesthesia machine calibration device) connector (positive pressure end), and No.2 connector to which drive gas corresponds on the circuit adapter test fixture by using a Y piece, as shown below. (7) Push in the push rod of the injector to let the pressure reading on the anesthesia machine calibration device rise to 30 to 35 cmH2O and then stop pushing. Keep the relative position between the push rod and the injector unchanged. If the pressure reading on the anesthesia machine calibration device falls more than 10cmH2O within 30s, this test is failed. It indicates that the bellows housing or the Manual position of the Auto/Manual switch is leaky. (Removing the bag arm indicated in the picture has no impact upon the test because the drive gas does not pass through the bag arm.) 4. Leak test of all pipelines on the circuit adapter Tools required: Negative pressure ball (quantity: 1) Circuit adapter test fixture (quantity: 1) PU tube (6X100) (quantity: 1) Test procedures: (1) Turn off the system switch. (2) Turn off the flow regulators. (3) Remove the breathing system. (4) Mount the circuit adapter test fixture onto the circuit adapter. 5-57 (5) Flatten the negative pressure ball to remove the gas inside. Then re-install the plug to seal the ball. Connect the other end of the negative pressure ball to the No.7 connector (on the circuit adapter test fixture) fresh gas pipeline of the circuit adapter test fixture, as shown below. (6) Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the test of fresh gas pipeline has failed. Locate the leak inside the main unit as per the method described in 1.3.4.1 Anesthetic Gas Delivery System. (7) Turn on the system switch and let the systems enter Standby. (8) Select [Main] → [Service] → [Diagnostic] → [Valves] to set the A/D value of the PEEP valve to make PEEP exceed 50 cmH2O. Set the A/D value of the inspiratory valve to “0” to produce 0 L/min of flow. Set PEEP safety valve to ON, as shown below. 5-58 (9) Flatten the negative pressure ball to remove the gas inside. Then re-install the plug to seal the ball. Connect the other end of the negative pressure ball to No.1 connector to which drive gas pipeline of the circuit adapter test fixture corresponds, as shown below. (10) Release the negative pressure ball. If the negative pressure ball is fully expanded within 30 seconds, it indicates that the test of the drive gas pipeline has failed. Check the expiratory valve assembly and the drive gas related pipeline inside the main unit. 5. Check the absorb canister assembly Tools required: VT PLUS (quantity: 1) Lucer adapter connector (quantity: 1) Injector (quantity: 1) Φ6 silicone tube (quantity: 2) PU tube (6X300) (quantity: 1) Breathing tube (quantity: 3) Y piece (quantity: 1) Breathing tube Y piece (quantity: 1) Breathing tube adapter connector (quantity: 1) T-shaped Allen wrench (quantity: 1) Test procedures: (1) Turn off the system switch. (2) Disassemble the pre-pak assembly and remove the breathing circuit. (3) Mount the pre-pak assembly. 5-59 (4) Remove the seals on the two connectors of the absorb canister assembly. Connect the two connectors of the lifting device by using two breathing tubes and one breathing tube Y piece. The other end of the breathing tube Y piece is connected to the breathing tube adapter connector through another breathing tube. Connect the injector connector, pressure sensor (of the anesthesia machine calibration device) connector (positive pressure end), and the breathing tube adapter connector to a Y piece, as shown below. (5) Push in the push rod of the injector to cause the pressure reading on the anesthesia machine calibration device rise to 30 to 35 cmH2O and then stop pushing. Keep the relative position between the push rod and the injector unchanged. If the pressure reading on the anesthesia machine calibration device falls more than 10cmH2O within 30 seconds, it indicates that absorb canister assembly are leaky. The test is failed. This step is required when the absorbent canister assembly is in bypass on or bypass off status. Connected to the Φ6 silicone tube Connected to the VT plus Turn the blue part to lock the connector to the VT plus Lucer adapter connector Lucer adapter connector Bypass off 5-60 Lucer adapter connector Bypass on (6) Check the seals on the two connections of the lifting device. It they are damaged, replace the seal and then re-mount the lifting device onto the breathing system. 6. Check the accessories and circuit inspiratory/expiratory part Test procedures: (1) Turn off the system switch. (2) Check the manual bag and replace if damaged. (3) Check the breathing tube and replace if damaged. (4) Remove the Paw pressure gauge. Check the seal and replace if found damaged. (5) Remove the water collection cup. Check the seal and replace if found damaged. (6) Remove the O2 sensor (if there is no O2 sensor, remove the plug where the O2 sensor should be installed). Check the seal and replace if damaged. (7) Remove the check valve dome. Check the seal and replace if damaged. (8) Remove the bag arm. Check the seal and replace if damaged. (9) Remove the prepak assembly as shown below. Check the seal and replace if damaged. Seal to be checked 5-61 7. Check the APL valve, support tube on the median plate, and sealing components of the Auto/Manual switch assembly The test requires a T-shaped Allen wrench. Test procedures: (1) Turn off the system switch. (2) Remove the APL valve. Check all seals and replace the defective ones. (3) Remove the support tube on the median plate. Check the seals and replace the defective ones. (4) Remove the Auto/Manual switch. Check the seals and replace the defective ones. 8. Check the ACGO assembly Tools required: Negative pressure ball (quantity: 1) Circuit adapter test fixture (quantity: 1) PU tube (6X100) (quantity: 1) 3106-08-10 adapter connector (quantity: 2) 3126-06-00 tube plug (quantity: 1) 3126-08-00 tube plug (quantity: 1) 3126-10-00tube plug (quantity: 2) Test procedures: (1) Turn off the system switch. (2) Pull out No.21 and 22 PU tubes which connect the ACGO assembly to the circuit adapter. The ends of the tubes which connect the ACGO assembly are pulled out but the other ends are not, as shown below. (3) Occlude the pulled-out tube ends by using two 3126-10-00 tube plugs. (4) Repeat steps 3 through 7 in “4 Leak test of all pipelines on the circuit adapter”. If the test is failed, it indicates that the connectors of the circuit adapter or seals are damaged. If there is leak, insert the pulled-out tubes into the ACGO assembly. Note that the positions of black lines on PU tubes No.21 and 22 are same with those before the tubes are pulled out. (5) Pull out No.52 and 53 PU tubes which connect the O2 flush button assembly and the vaporizer manifold assembly to the ACGO assembly. The ends of the tubes which connect the ACGO assembly are pulled out but the other ends are not 5-62 (6) Occlude the pulled-out tube ends by using 3126-06-00 and 3126-08-00 tube plugs, as shown below. (7) Repeat steps 3 through 7 in “4 Leak test of all pipelines on the circuit adapter”. If the test fails, it indicates the ACGO assembly is damaged. Check the seals in the ACGO assembly and replace any damaged seals. 5.3.5 Tidal Volume The following table lists tidal volume inaccuracy related failures. Failure Possible cause Recommended action description Inaccurate tidal volume The flow sensor is not installed properly. The setting of fresh gas flow is inappropriate. There are significant leaks in the breathing system and the fresh gas flow is too low. * There is water build-up inside the flow sensor. *The membrane of the flow sensor assembly is distorted, dirty or its inside resistance changes. Zero drift occurs to the pressure sensor on the ventilator control board. *The flow sensor pressure sampling pipeline is leaky. *The flow sensor is damaged. *The pressure sensor on the ventilator control board is defective. The inlet gas flow regulator on the integrated pneumatic circuit of the expiratory valve assembly is defective. The current Plimit is set too low, which causes expiration to start in advance. The displayed TVe and TVi are not the same. 5-63 Re-install the flow sensor. Adjust the fresh gas flow. Repair the leaking points after checking as per the procedures described in sections 1.3.4.1 Anesthetic Gas Delivery System and 1.3.4.3 Breathing System. Remove the flow sensor and clear its inside water build-up. Enter the service mode and calibrate the flow sensor as per 4.3.2 Flow Calibration (Factory). Repair the leaking points after checking as per the procedures described in 5.3.4.1 Leak Test of Flow Sensor Pressure Sampling Pipeline. Replace the flow sensor. Replace the ventilator control board. Replace the integrated pneumatic circuit of the expiratory valve assembly or replace the expiratory valve assembly. Set Plimit to a higher value to cause Paw not to exceed the limit. In the valves test tool, compare the measurement error made by three sensors and judge whether to perform calibration as per 4.3.2 Flow Calibration (Factory) . In the above table, possible causes marked “*” are related to inaccurate measured values by flow sensors. Do the following to detect if tidal volume inaccuracy results from “*” marked causes. 1. Turn off the flow regulators. 2. Make sure that the patient is disconnected from the system and that the Auto/Manual switch is set to the mechanical ventilation position. 3. Remove the bellows and then install the bellows housing properly. 4. Remove the water collection cup. 5. Connect the inspiration and expiration connectors together by using a breathing tube, as shown below. The water collection cup is already removed. 6. Turn on gas supplies and enter Standby. 7. Select [Main] → [Service] → [Diagnostic] → [Valves] to set the A/D value of the PEEP valve to make PEEP exceed 40 cmH2O. Set PEEP safety valve to ON, as shown below. 8. Set the A/D value of the inspiratory valve to cause the flow of inspiratory valve to reach a certain value. In this case, the flows measured by the ventilator flow sensor, 5-64 inspiratory flow sensor, and expiratory flow sensor should be the same. Test multiple points by setting the A/D value of the inspiratory valve. For each point, the flows measured by the three sensors should be the same. If not, the measured value by the flow sensor is inaccurate. Troubleshoot the possible causes marked “*” in the above table. 5.4 Sensor and Valve Related Failures To use Diagnostic Tests to troubleshoot sensors or valves related failures, you must be familiar with the one-to-one correspondence between the menu options on the Diagnostic Tests screen and the actual pneumatic circuit and hardware components. Correspondence between valve diagnostic tests screen and pneumatic circuit The following figure shows the one-to-one correspondence between the sensors & valves on the valves-test tool screen and the actual components in the pneumatic circuit diagram. 5-65 Correspondence between valve diagnostic tests screen and hardware The following figure shows how the sampling lines of the sensors are actually connected on the VCM. 5.4.1 Preparations before Using Diagnostic Tests Make the following preparations before using the valves-test tool to locate the valves or sensors related failures: 1 Connect the pneumatic circuit according to the type of sensor or valve to be checked. Before using the Diagnostic Tests Menu: connect the tubes of the anesthesia machine following the constant-flow connection method to check the flow sensors and inspiratory valve. For details, refer to 4.3.2 Flow Calibration (Factory). 2. Make sure that the supply gas pressure is within specifications. 3. When the system is in Standby, select the [Main] shortcut key → [Service] → [Diagnostic] to access the [Diagnostic] menu. 5.4.2 Zero Point Problems of Flow and Pressure Sensors By using the [Diagnostic Tests], you can easily detect if the zero points of all the pressure and flow sensors are within specifications. To diagnose the zero points of the sensors: 1. Disconnect all gas supplies and make sure that the actual values of the sensors are “0”. 2. Check the A/D counts of the sensors in the valve-test tool menu, which are the zero points of the sensors. 3. If the zero point of one sensor is outside of specifications, it indicates that the ventilator control board is defective. You need to replace the board. You can also detect the zero points of the sensors by referring to 3.6.3 Check the Sensor Zero Point. NOTE: For the normal range of sensors’ zero points, refer to 3.6.3 Check the Sensor Zero Point. 5-66 5.4.3 Connections and Measurement Problems of the Flow Sensors The flow sensor has two sampling lines. Connection errors include: Two sampling lines are connected reversely. One sampling line is not connected. Both sampling lines are not connected. By using the Diagnostic Tests tool, you can detect if the sampling lines are connected correctly. To diagnose the sampling line connection of the flow sensor: 1. Connect the tubes of the anesthesia machine following the constant-flow connection method. Refer to 5.4.1. 2. Make sure that gas supplies are within specifications. In the [Diagnostic Tests ], PEEP safety valve to ON and the D/A value of the PEEP valve to more than “1500”, making sure that the PEEP valve closes at above 30 cmH2O. 3. Increase the D/A value of the inspiratory valve gradually and the A/D value of the flow sensor should also increase. With the gradual increase of gas supplied, If the sampling A/D value of one flow sensor decreases gradually, it is possible that the two sampling lines of the flow sensor are connected reversely. If the sampling A/D value of one flow sensor stays unchanged, it is possible that the two sampling lines of the flow sensor are broken or not connected. If the sampling A/D value of one flow sensor nears saturation (above “4000”) quickly, it is possible that the sampling line at the low pressure end (gas outlet end) of the flow sensor is not connected. 4. If sampling line connection errors are detected, re-connect all sampling lines and verify proper connection of lines. To diagnose the measurement error of the flow sensors: After confirming that both the zero points of the sensors and the sampling line connections of the sensors are correct, check the flow sensor accuracy:. With the gradual increase of actual flow, the measured value of the flow sensor should also increase. Otherwise, the calibration data have errors. You need to calibrate the flow sensor again. Compared with the measured value of the standard flow measurement device (anesthesia machine calibration device), the measured value of the flow sensor should be accurate. Otherwise, the calibration data have errors. You need to calibrate the flow sensor again. For details, refer to 3.6.4 Check the Accuracy of Flow Sensors. 5-67 5.4.4 Connections and Measurement Problems of the Pressure Sensors The pressure sensor has one sampling line. Connection errors include: The sampling line is not connected. The sampling line is connected incorrectly. By using the Diagnostic Tests tool, you can detect if the sampling lines are connected correctly. To diagnose the sampling line connection of the pressure sensor: During normal ventilation, if a sampling line connection error occurs, it is easily detected through the Paw waveform and technical alarms. If with the increase of actual pressure, pressure waveform data decreases and the alarm of “Paw Too Low” or “Patient Circuit Leak” occurs simultaneously, it is possible that the sampling line of the airway pressure sensor is connected incorrectly. You can enter the [Diagnostic Tests] menu to set the PEEP safety valve to ON. Gradually increase the D/A value of the PEEP valve and observe if the A/D value of the PEEP pressure sensor also increases gradually. If not, it further indicates that the PEEP pressure sensor may be connected incorrectly. To diagnose the sampling line connection of the pressure sensor in case of pressure calibration failure: 1. Connect the tubes of the anesthesia machine just like 5.4.1. 2. Make sure to mount the water collection cup again 3. Make sure that gas supplies are within specifications. In the [Diagnostic Tests] menu, set the PEEP safety valve to ON. Set the Insp. Valve at 5 L/min. 4. Increase the D/A value of the PEEP valve gradually and the sampling A/D value of the pressure sensor should also increase due to the gradual increase of actual pressure, If the sampling A/D value of one pressure sensor decreases gradually, it is possible that the sampling line of the pressure sensor is connected incorrectly. If the sampling A/D value of one pressuer sensor remains unchanged, it is possible that the sampling line of the pressure sensor is not connected. The pressure of the airway pressure gauge should increase. If not, it is possible that the airway pressure gauge is defective.。 5. If sampling line connection errors are detected, re-connect the sampling lines and verify proper connection of all lines. To diagnose the measurement error of the flow sensors: With the gradual increase of actual pressure, the measured value of the pressure sensor should also increase. Otherwise, the calibration data have errors. You need to calibrate the pressure sensor again. Compared with the measured value of the standard pressure measurement device (anesthesia machine calibration device), the measured value of the pressure sensor should be accurate. Otherwise, the calibration data have errors. You need to calibrate the pressure sensor again. For details, refer to 3.6.5 Check the Accuracy of Pressure Sensors。 5-68 5.4.5 Opening State Problems of the Inspiratory Valve By using [Diagnostic Tests], you can detect if the opening state of the inspiratory valve is correct. 1. The methods for tube connections and settings of the anesthesia machine are the same as those for sampling line connections of the flow sensors. For details, refer to 5.4.1. 2. In the [Diagnostic Tests] menu, gradually increase the D/A value of the inspiratory valve. If the measured values of the ventilator flow sensor, inspiratory flow sensor, and expiratory flow sensor change very little and low gas flow is felt at the connector of water collection cup, it indicates that the inspiratory valve or the D/A on the Ventilator Control Board is defective. 3. Normally, when the D/A value of the inspiratory valve is set to “2500”, the flow measured by the standard flow measurement device can reach 90 L/min. 4. If when the D/A value of the inspiratory valve is set to more than “4000”, the flow measured by the standard flow measurement device fails to reach 90 L/min, flow calibration will fail. In this case, you need to replace the expiratory valve assembly or the ventilator control board. 5. To locate if the DA on the Ventilator Control Board is defective, you can use a multimeter to measure the output of DA on the Ventilator Control Board corresponding to the inspiratory valve. If voltage also increases with the increase of D/A value, and voltage nears 6V when D/A value is set to more than “4000”, it indicates that the DA on the Ventilator Control Board corresponding to the inspiratory valve may be correct. 6. After the expiratory valve assembly or the Ventilator Control Board is replaced, you can use the similar method to check if the problem is fixed. 5.4.6 Opening State Problems of the PEEP Safety Valve When the PEEP safety valve is permanently OFF and the gas supplies are within specifications, the [Drive Gas Pressure Low] is alarmed. By using Diagnostic Tests, you can detect if the opening states of the PEEP safety valve and PEEP valve are correct. To diagnose the opening state of the PEEP safety valve: 1. Make sure that gas supplies are within specifications. 2. In the [Diagnostic Tests] menu, when the PEEP safety valve is switched on, a subtle click is heard. 3. Adjust the D/A value of the PEEP valve to cause the pressure measured by the PEEP pressure sensor to exceed 0 cmH2O. 4. Switch off the PEEP safety valve. The pressure measured by the PEEP pressure sensor should drop to 0 cmH2O immediately. Switch on the PEEP safety valve again. The measured value of the PEEP pressure sensor rapidly restores almost the same value to that before PEEP safety valve is switched off. During this period, gas flow and also change of gas flow when the PEEP safety valve is switched on or off can be felt at the PEEP outlet, which helps to judge if the PEEP safety valve can be switched on or off correctly. 5. If an error is detected, it is possible that the PEEP safety valve or the safety valve drive voltage on the ventilator control board is defective. You can use a multimeter to measure the drive signals on the ventilator control board corresponding to the PEEP safety valve (measurement can be performed at the corresponding socket). When the PEEP safety valve is turned on, the drive voltage should near 6V. When the PEEP safety valve is turned off, the drive voltage should near 0V. If these two conditions are met simultaneously, the ventilator control board is normal. 6. If the PEEP safety valve is defective, replace the expiratory valve assembly. After replacement, you can use the similar method to check if the problem is fixed. 5-69 5.4.7 Opening State Problems of the PEEP Valve When the PEEP valve is defective, pressure related alarms occur in mechanical ventilation modes. By using Diagnostic Tests, you can detect if the opening states of the PEEP valve is correct. To diagnose the opening state of the PEEP valve: 1. Make sure that gas supplies are within specifications. In the [Diagnostic Tests] menu, set the PEEP safety valve to ON. 2. With the increase of D/A value of the PEEP valve, the measured value of the PEEP pressure sensor (or the anesthesia machine calibration device) should also rise. Note that there is a non-response area for the PEEP valve when the D/A value is relatively small. When the D/A value is less than this range, the PEEP valve cannot be opened and the output is “0” continuously. When the D/A value is greater than this range, the pressure output will increase with the increase of D/A value. This phenomenon also exists for the inspiratory valve. 3. For subsequent diagnosis rules, refer to 5.4.5. 5.5 Hardware and Electrical Problems Failure description Possible cause Recommended action When switch ON, AC and battery indicator lamps are not lit, the machine cannot ventilate and the screen is not lit. AC power supply is not connected and the batty capacity is insufficient. The breaker of AC input is tripped and the battery capacity is insufficient. The auxiliary A/C electrical outlet has no voltage output. The breaker of auxiliary outlet is tripped. The anesthesia machine cannot be started. The system switch is damaged. The cable connected to the system switch is disconnected or is not connected reliably. The power board failure results in no power output of 15.2V, 3.3V, 5V, and 12V. The power board software failure results in no power output of 15.2V, 3.3V, 5V, and 12V. Check and make sure that the AC power supply is connected properly. Reset the breaker. If the problem persists after breaker is reset, it indicates the machine inside is shortcircuited. Replace the fuse. If the problem persists after the fuse is replaced, it indicates the machine inside is shortcircuited. Reset the breaker. If there is still no voltage output, it indicates the load equipment to which the auxiliary A/C electrical outlet is connected is shorted or overcurrent. Check if the load equipment is normal and make sure that its working current is within the specified specification of auxiliary outlet. Replace the system switch. Check and make sure that the cable is reliably connected. The fuse of power board is burned and the battery capacity is insufficient. 5-70 Replace the power board. Update the power board software. Failure description Possible cause Recommended action The screen of anesthesia machine cannot be lit. The cable connected to the inverter is disconnected or is not connected reliably. The inverter is damaged. The power board failure causes power output error. The power board software failure causes power output error. The screen power supply fuse is burned out, which results in no 3.3V output. The main control board failure results in no display output. Check and make sure that the cable is reliably connected. Replace the inverter. Replace the power board. The power board failure causes instable supply voltage of the display. The time sequence of main control board LVDS is abnormal. Replace the power board. The screen of anesthesia machine can be lit, but without any contents. The screen of anesthesia machine can be lit and shows contents, but the screen flashes. The heater is ineffective. The touchscreen is ineffective. During the operation of the anesthesia machine, ventilation stops all of a sudden but the display and buttons work normally. Exiting Standby fails. Alarm messages are displayed on the screen but without alarm sound. Update the power board software. Replace the fuse. Replace the main control board. Update the software of main control board. If the screen continues flashing, replace the main control board. The heating drive circuit of the power board is faulty. The heater is damaged. The internal sensor of heater is ineffective. The cable connected to the heater is disconnected or is not connected reliably. The touchscreen is damaged. The touchscreen control board is damaged. The cable connected to the touchscreen or touchscreen control board is disconnected or is not connected reliably. The monitoring module or valve is damaged. Check and make sure that the cable is connected reliably. Replace the touchscreen. Replace the touchscreen control board The monitoring module selftest is failed. Replace the monitoring module. The speaker is damaged. The speaker cable is not connected reliably. Replace the speaker. Check and make sure that the cable is connected reliably. 5-71 Replace the power board. Replace the heater. Check and make sure that the cable is connected reliably. Select [Main] → [Service] → [Diagnostic] → [Valves] Test the status of each valve and reference power supply in the valves-test tool window. If valve malfunction or reference power supply error is detected, replace the monitoring module. Failure description Possible cause Recommended action Network connection is failed. The network cable is disconnected or is not connected reliably. The network cable is too long. Check and make sure that the network cable is connected reliably. Shorten the network cable. Recommended cable length is approximately 1.5 m. The network cable has two linear orderings. Make sure that network cable with correct linear ordering is used. Check the screen to see if the anesthesia machine is in mechanical ventilation mode and if there is an alarm triggered. 1. Set tidal volume to maximum. 2. Switch between standby and mechanical statuses or between manual and mechanical statuses repeatedly. 3. Replace the pneumatic block. The network cable is used incorrectly. No gas is outputted through the valve in mechanical ventilation mode. The Auto/Manual switch is damaged. The valve cannot be opened. 5.6 Software Upgrade and Software Function Activation 5.6.1 1.6.1 Software Upgrade NOTE: Update the software when replacing the main control board, monitoring module, power board or electronic flowmeter board. NOTE: Read the update package accompanying document before updating A series. 1. Connect the Ethernet port of the PC to the Ethernet port of the A5 using the Ethernet crossover cable. 2. Before running the Mindray anesthesia machine and ventilator software upgrade tool, verify that the IP address of the PC is set to 192.168.23.1., and the subnet mask is set to 255.255.255.0. Check and set the IP address on the PC by following these instructions. (1). Ensure that the IP address of PC and the anaesthesia machine network falls within the same network segment. For example, if the IP address of the PC is 192.168.23.1, the IP address of the anaesthesia machine should be 192.168.23.xxx. (2) On the PC, click Start, Control Panel and then Network Connections. (3).Right-click Local Area Connection and then left click Properties. (4).Scroll down to Internet Protocol (TCP/IP), click to Internet Protocol (TCP/IP and then click Properties. (5). Click the radio button for "Use the following IP address". Set the IP address and Subnet mask and then click OK. 5-72 3. Make sure that the software upgrade tool in the software package has been installed to the PC. If it has not been installed then follow these steps: (1) Run the SystemUpdateTool.exe file. (2). When prompted to setup a language select “English” and then select “OK”. (3). When the Welcome dialog window is displayed select “Next”. (4). When the Customer Information dialog window is displayed enter the following: (5). User Name: Mindray (6). Company Name: Mindray (7). Serial Number: 26582640 (8). Select “Next”. (9). When the Destination Location dialog window is displayed select “Next” to accept the default Destination Folders. (10). When the Select Program Folders dialog window is displayed select “Next” to accept the default Program Folder. (11) When the Install Shield Wizard dialog window is displayed select “Finish” to complete the installation. 4. Upgrade Software: (1) From the Desktop, run the icon for Mindray anesthesia machine and ventilator software upgrade tool. When the “Select Product Series” dialog is displayed select A Series(Non-FDA) followed by OK. Once the system update tool starts to work perform the following to update the software. (2). Select “Select Package” from the top tool bar. (3). When the “Select Package” dialog is displayed, select “>>>” (4). When the “Open dialog” is displayed, select the directory where the software package is located for viewing the details. Select the software “01_BIOS V01.16.00.00.pkg”. 5-73 Item 1 2 3 Software Upgraded Package BIOS(01.16.00.00) Bundle (V02.16.00) Power(V1.7) Modules Included Version Guide program United version number V01.16.00.00 V02.16.00 System program V03.16.00 FPGA display program FPGA voice program Aux monitor module software Monitor module software Language Electronic flowmeter software Power supply card V1.4 V1.5 V01.02.00 V01.04.00 V4.6 V1.10 V1.7 Note: This is only an example of latest software version in the form above. Check for Technical bulletins to find the correct table for the software version you are installing. Note: Perform upgrade in accordance with the sequence numbers in the software package. Note: In the software package 02.16.00 and later versions, BIOS is displayed and upgraded independently. (5). Select “Open”. (6) A dialog box will be displayed. Click OK.(7) Click “Start(Single)” on the Mindray anesthesia machine and ventilator software upgrade tool. (8) Turn off the A5 and then turn it on. (9) You will see that the PC network icon indicates that it is connected. The anesthesia machine displays the upgrade screen. Note: If the screen as shown above is not displayed after turning on the equipment, and the equipment starts up normally and enters system check, please check the IP setting and network cable. (10). While updating the software the A5 will show text which explains the progress of the software update. (11). When upgrade is completed, the anesthesia machine will display “Upgrade Completed!” and Mindray anesthesia machine and ventilator software upgrade tool will display “update system successfully”. (12). Repeat the steps 3 through 11 to update the "02_A5 Bundle_V02.16.00.mpkg" and "03_Power V1.7.pkg" software. Note: The power board software upgrade usually fails the first time but succeeds the second time. (13). When the software “03_Power V1.7.pkg” is upgraded turn off the A5 unit and then restart it. Note: If the equipment fails to start up normally after completing upgrade, please find the corresponding guide document in the upgrade package and operate by following the instructions. 5-74 (14). Select Main (hot key) > Service > enter the service password "789789" > System Info, you can check the Bundle Version Number on the Service screen. Verify that the Bundle Version Number on the software keeps accordance with that of the Bundle. 5.6.2 Software Function Activation The manufacturer can activate and upgrade the following functions. If you need to activate a function, buy the corresponding activation code from the manufacturer. Configuration Item Supporting Function Description Activation SIMV-VG Synchronous intermittent instructionbased ventilation – pressure adjustment and volume control PS Pressure-supported ventilation mode CPAS/PS Pressure-supported ventilation/Continual recording of positive pressure ventilation Recruitment Lung recruitment AA Calculation Anesthesia calculation 5.6.2.1 Apply for a Software Function Activation Code Note: When applying for an activation code, provide the following information about the anesthesia machine whose configurations will be activated: machine ID, existing configurations, and configurations to be activated. When you need to add the preceding billed function configurations, apply from our company's user service department for a software function configuration activation code. The application procedure is as follows: 1. 2. 3. 4. Record the sequence number of the anesthesia machine whose configurations will be activated (see the label on the left side of the machine). Do the following to record the ID and existing configurations of the anesthesia machine: Choose Main (hot key) > System >> > Configuration Info >> to open the Configuration Information menu. Record the configurations to be activated. Send the preceding records to our company's user service department to apply for an activation code. 5-75 5.6.2.2 Software Function Activation Process Note: Before activation, check and record the existing function configurations and those to be activated. Note: During activation, ensure that the input activation code is consistent with that provided by the manufacturer; otherwise, activation will fail. Take the following steps to activate software function configurations: 1 Open the Function Activation menu, and choose Main (hot key) > Service (enter the correct password) > Factory Setup> Activate Functions, as shown in the following figure. 2. Input the correct activation code in the Activation Code dialog box. 5-76 3. 4. Configuration activation: Click Activate. If the input activation code is correct, the system displays the message "Activation completed. Restart the machine for the activated configurations to take effect." below the activation code input box. Click Accept and restart the anesthesia machine so that the activated configurations take effect. Note: Maintain the anesthesia machine in the powered-on state before activation is complete or an error is returned; otherwise, the BIOS program of the main control board of the anesthesia machine will be corrupted. Note: When the system indicates that activation is successful, restart the anesthesia machine to make the activated configurations take effect. 5-77 FOR YOUR NOTES 5-78 6 Repair and Disassembly 6.1 Prepare for Disassembly 6.1.1 Tools During parts disassembly and replacement, the following tools may be required: Metric Allen wrench (1.5, 2, 2.5, 3, 4, 5, 6, 8, 10mm) Phillips screwdriver(#1 and #2) Diagonal pliers Flathead screwdriver Metric M3 and M4 socket screwdriver Adjustable wrench Tweezers Krytox lubricant (P/N:099-000107-00) 6.1.2 Preparations Before disassembly, Make sure that the anesthesia machine is turned off and disconnected from the AC power source. Bleed down the gas pressure inside the anesthesia machine as described below. Disconnect all pipeline and cylinder gas supplies. Prepare the tools required for disassembly. Maneuver the anesthesia machine to an appropriate location and then apply the brake. CAUTION: The internal parts may be contaminated. Wear special gloves during disassembly and inspection. 6.1.3 Bleed Gas Pressure Make sure to bleed down the gas pressure inside the anesthesia machine before disassembling pneumatic fittings to avoid personal injury or equipment damage. To bleed gas pressure: 1. Close other cylinder valves and disconnect pipeline gas supplies. Do not disconnect the O2 pipeline. If O2 pipeline is not available, connect O2 cylinder and open the O2 cylinder valve. 2. Set the system switch to ON. 3. Turn on all the flow controls (except O2). 4. Make sure that N2O and AIR pipeline pressure gauges read zero. 5. Disconnect the O2 pipeline supply (or close the O2 cylinder valve). Push the O2 flush button to bleed O2 from the system. 6. Set the system switch to OFF. 6-1 6.2 Disassemble the Assemblies 6.2.1 Disassemble the Internal Assemblies of the Machine Upper Half 6.2.1.1 Open the Service Door Turn the two screws on the service door counter-clockwise for a half turn to open the door. 6.2.1.2 Remove the Gas Supply Inlet Assembly 1. 2. 3. 4. Open the service door. Refer to 6.2.1.1 Open the Service Door. Disconnect the tubes from the N2O, AIR and O2 supply inlet assemblies. Remove the two screws from the N2O and AIR supply inlet assemblies. Remove the related cables from O2 supply inlet assembly and Remove the two screws to remove the assembly. 6-2 6.2.1.3 Remove the Pressure Regulator Assembly 1. 2. 3. Open the service door. Refer to 6.2.1.1 Open the Service Door. Remove the tubes from the pressure regulator assembly. Remove the three screws on the pressure regulator assembly and remove the assembly. 6.2.1.4 Remove the Electronic Flowmeter Assembly 1. 2. 3. Open the service door. Refer to 6.2.1.1 Open the Service Door. Remove the related cables and tubes from the electronic flowmeter. Remove the four screws and remove the electronic flowmeter. Machine before ECL128 takes effect Machine after ECL128 takes effect 6-3 6.2.1.5 Remove the Poppet Valve Assembly 1. 2. 3. Open the service door. Refer to 6.2.1.1 Open the Service Door. Remove the tubes from the poppet valve assembly. Remove the two screws on the pressure regulator supporting plate to remove the poppet valve. (Machine before ECL128 takes effect) 4. Unfasten the four screws to remove the poppet valve assembly. (Machine after ECL128 takes effect) 6.2.1.6 Remove the Gas Mixer Assembly (the machine has an independent gas mixer assembly before ECL128 takes effect) 1. 2. 3. Open the service door. Refer to 6.2.1.1 Open the Service Door. Remove the tubes from the gas mixer assembly and the poppet valve. Remove the four screws on the gas mixer assembly to remove the gas mixer assembly and the poppet valve assembly. 6-4 4. Remove the two screws on the pressure regulator supporting plate to remove the gas mixer assembly. 6.2.1.7 Remove the ORC Assembly 1. 2. 3. Open the service door. Refer to 6.2.1.1 Open the Service Door. Remove the tubes from the ORC assembly. Remove the two screws on the ORC assembly to remove the assembly. 6.2.1.8 Remove the Back Pressure Valve Assembly 1. 2. 3. Open the service door. Refer to 6.2.1.1 Open the Service Door. Remove the tubes from the back pressure valve assembly. Remove the four screws on the back pressure valve assembly to remove the assembly. 6-5 6.2.1.9 Replace the Lithium-ion Battery 1. 2. 3. Open the service door. Refer to 6.2.1.1 Open the Service Door. Remove the two screws on the battery box cover and then take out the lithium-ion battery. Place the new lithium-ion battery into the battery box in the original direction to complete battery replacement. 6.2.1.10 Remove the Vaporizer Manifold Assembly 1. 2. 3. Open the service door. Refer to 6.2.1.1 Open the Service Door. Remove the tubes from the vaporizer manifold assembly. Remove the four screws and remove the vaporizer manifold assembly. 6-6 6.2.1.11 Remove the Pneumatic Resistor Block Assembly (For the Anesthesia System Released Before ECL115) 1. 2. 3. Open the service door. Refer to 6.2.1.1 Open the Service Door. Remove the tubes from the pneumatic resistor block assembly. Remove the two screws and remove the pneumatic resistor block assembly. 6.2.2 Disassemble Hardware Box Assembly 6.2.2.1 Remove the Top Plate Assembly 1. Fully loosen the 3 captive screws fixing the top plate assembly to remove the assembly. 2. Lift the top plate to remove it. 6-7 6.2.2.2 Remove the Main Control Board NOTE: To replace the main control board, you may need to re-install the software. NOTE: Before the replacement of the main control board, please record the settings of the service menu. After the replacement of the main control board, please restore the settings of the service menu. 1. Remove the top plate assembly. Refer to 6.2.2.1 Remove the Top Plate Assembly. 2. Remove the three screws fixing the main control board to remove the board. 6.2.2.3 Remove the Monitoring Module NOTE: To replace the monitoring module, you may need to re-install the software. 1. Remove the top plate assembly. Refer to 6.2.2.1 Remove the Top Plate Assembly. 2. Remove the related cables and tubes from the monitoring module. 3. Remove the four screws fixing the monitoring module to remove the module. 6-8 4. Refer to 1.2.3Connections between Pneumatic Circuit, Breathing System and Ventilator Control Board for tube connections between the pneumatic circuit, breathing system, and modules. 6.2.2.4 Remove the Power Board 1. 2. Remove the top plate assembly. Refer to 6.2.2.1 Remove the Top Plate Assembly. Remove the two screws fixing the high-voltage protective cover to remove the protective cover. 3. Remove the filter cable connector. Remove the screw on the power board supporting sheet, the three screws and one nut fixing the power board to remove the power board. 4. Remove the screw on the power board supporting sheet to remove the supporting sheet. 6-9 6.2.2.5 Remove the Motherboard 1. 2. Open the service door. Refer to 6.2.2.1 Remove the Top Plate Assembly. Remove the three cables from the motherboard. 3. Remove the four stud bolts on the two front sockets. 4. Remove the top plate assembly. Refer to 6.2.2.1 Remove the Top Plate Assembly. 5. Remove the two screws fixing the main control board output frame and one screw fixing the main control board to remove the main control board together with the sheet metal parts. 6. Remove the power board. Refer to 6.2.2.4 Remove the Power Board. 6-10 7. Remove the three screws to remove the air deflector plate. 8. Remove the power board insulation overlay. 9. Remove the cables from the motherboard. 10. Loosen the two screws on the motherboard output bracket and Remove the 12 screws fixing the motherboard to remove the motherboard. 6-11 6.2.2.6 Remove the Three-Way Valve Assembly 1. 2. Remove the top plate assembly. Refer to 6.2.2.1 Remove the Top Plate Assembly. Remove the related tubes and cables from the three-way valve assembly. 3. Remove the three screws on the three-way valve assembly to remove the assembly. Please refer to 1.2.3 Connections between Pneumatic Circuit, Breathing System and Ventilator Control Board. 6.2.2.7 Remove the Speaker 1. 2. 3. Remove the top plate assembly. Refer to 6.2.2.1 Remove the Top Plate Assembly. Remove the speaker cable. Remove the two screws to remove the speaker. 6-12 6.2.2.8 Remove the Hardware Box Rear Cover Plate Assembly 1. 2. Remove the top plate assembly. Refer to 6.2.2.1 Remove the Top Plate Assembly. Open the service door. Refer to 6.2.1.1 Open the Service Door. Remove the cables and tubes from the hardware box assembly. 3. Remove the three screws fixing the power plug retaining clip to remove the power board and the power plug retaining clip. 4. Remove the two screws fixing the high-voltage protective cover to remove the protective cover. 6-13 5. Remove the 7 screws fixing the hardware box assembly and lift the hardware box assembly to remove it. 6. 7. Remove the AC filer and fan cables. Remove the 11 screws on the hardware box rear cover plate assembly to remove the assembly. 6.2.2.9 Remove the Fan 1. 2. Remove the hardware box rear cover plate assembly. Remove the four screws on the fan to remove the fan. 6-14 6.2.2.10 Remove the Filter 1. 2. Remove the hardware box rear cover plate assembly. Remove the four screws and the nut on the filter to remove the filter. 6-15 6.2.2.11 Remove the Auxiliary Electrical Outlet 1. 2. Remove the hardware box rear cover plate assembly. Remove the cables from the auxiliary electrical outlets and remove the auxiliary electrical outlets. A5 has four auxiliary electrical outlets, as shown below. 6.2.2.12 Remove the Breaker Assembly 1. 2. Remove the hardware box rear cover plate assembly. Remove the breaker nuts to remove the breakers. A5 has five breakers, as shown below. 6-16 6.2.3 Disassemble the Work Surface 6.2.3.1 Remove the Drawer Assembly 1. 2. Pull out the drawer until the black locking piece on the rail can be seen. Depress the locking piece on the right rail of drawer, and lift the locking piece on the left rail at the same time. 3. Take out the drawer. 6-17 6.2.3.2 Remove the Work Surface Cover Plate Assembly 1. 2. Remove the first drawer. Fully loosen the five captive screws fixing the work surface cover plate assembly. 3. Lift off the work surface cover plate assembly from the work surface. When the work surface cover plate assembly is removed, the internal structure of the work surface is as shown below. 6-18 6.2.3.3 Remove the Work Surface Front Panel Supporting Plate 1. 2. Remove the work surface cover plate assembly. Remove the 8 screws on the work surface front panel supporting plate to remove the front panel supporting plate. 6.2.3.4 Remove the Rear Cover Plate Assembly 1. Remove the knob on the cylinder yoke assembly (only applicable to configuration with high-pressure cylinder yoke assembly). 2. 3. Remove the AGSS transfer tube connecting to the exhaust tube and refer to 6.2.10. Remove the 6 screws on the cart rear cover plate assembly to remove the assembly (the picture on the right shows the configuration with high-pressure cylinder yoke assembly). 6-19 3. Remove the four screws on the main unit rear cover plate to remove the main unit rear cover plate (the picture on the right shows the configuration with high-pressure cylinder yoke assembly). 6.2.3.5 Remove the Expiratory Valve Assembly 1. 2. Remove the work surface cover plate assembly. Remove the tubes and cables from the expiratory valve assembly. 3. Remove the pressure switch cable. Remove the switch jacket before Removing the cable to avoid damaging the cable. Switch jacket 6-20 4. Remove the four screws on the expiratory valve assembly to remove the assembly. 6.2.3.6 Remove the O2 Flush Assembly 1. 2. 3. 4. Remove the work surface cover plate assembly. Refer to 6.2.3.2. Remove the work surface front panel supporting plate. Refer to 6.2.3.3. Remove the tubes from the O2 flush assembly. Remove the two screws on the bracket and remove it. 5. 6. Remove the metal post on the right side of the O2 flush assembly. Remove the four screws around the O2 flush assembly to remove the assembly. 6-21 6.2.3.7 Remove the Touchpad Assembly 1. 2. 3. 4. Remove the work surface cover plate assembly. Refer to 6.2.3.2. Remove the work surface front panel supporting plate. Refer to 6.2.3.3. Remove the data line of the Touchpad assembly. Remove the 8 screws on the Touchpad to remove the Touchpad bracket and Touchpad assembly. 6.2.3.8 Remove the ACGO Assembly 1. 2. 2. 3. Remove the work surface cover plate assembly. Refer to 6.2.3.2. Remove the work surface front panel supporting plate. Refer to 6.2.3.3. Remove the tubes and cable from the ACGO assembly. Remove the 7 screws on the Touchpad bracket and ACGO assembly to remove the Touchpad bracket and ACGO assembly. 6.2.3.9 Remove the Cylinder Yoke Assembly (only applicable to configuration with cylinder yoke assembly) 1. 2. 3. Open the service door and refer to 6.2.1.1. Remove the rear cover plate and refer to 6.2.3.4. Remove the Module Rack Box Assembly and refer to 6.2.5 (if configured with module rack box). Remove the Module Rack Fan and refer to 6.2.6 (if configured with module rack box). 5. Remove the tubes on the cylinder yoke assembly. 6. Fully loosen the brass nut on the cylinder yoke assembly. 4. 6-22 7. Remove the four screws on the cylinder yoke mounting plate to remove the assembly. 8. Remove the four screws on the cylinder yoke to remove the O2, N2O, and Air cylinder yoke assemblies respectively. The following picture shows the cylinder yoke assemblies before the ECL 112 changes. The following picture shows the cylinder yoke assemblies after the ECL 112 changes. 6-23 6.2.3.10 Remove and Limit the Spring tube Assembly 1. Remove the work surface cover plate assembly. For details, see section 6.2.3.2 "Remove the Work Surface Cover Plate Assembly". 2. Remove the circuit arm cover. For details, see section 6.2.9 "Remove the Circuit Arm Cover". 3. Cut the ties binding the spring tube and disconnect the tube from the connector at the #2 end to remove the spring tube assembly. 6.2.4 Disassemble the Display Assembly 1. 2. Open the service door. Refer to 6.2.1.1. Remove the display connection line from the hardware box and the tubes from the needle valve assembly. Fully loosen the four captive screws to remove the display assembly. 6-24 6.2.4.1 Remove the Alarm Lamp Board 1. 2. 3. Remove the display assembly. Refer to 6.2.4. Remove the cables from the alarm lamp board. Remove the two screws on the alarm lamp board to remove the board. 6-25 6.2.4.2 Remove the Display Adapter Board 1. 2. Remove the display assembly. Refer to 6.2.4. Remove the five screws on the cover plate of display adapter board. 3. 4. Remove the cables from the display adapter board. Remove the four screws on the display adapter board and remove the board. 6.2.4.3 Remove the Touch Screen Control Board 1. 2. Remove the display assembly. Refer to 6.2.4. Remove the four screws on the cover plate of the touch screen control board to remove the cover plate. 6-26 3. 4. Remove the cables from the touch screen control board. Remove the two screws on the touch screen control board to remove the board. 6.2.4.4 Remove the Backlight Inverter Board 1. 2. Remove the display assembly. Refer to 6.2.4. Remove the four screws on the cover plate of the backlight inverter board to remove the cover plate. 3. 4. Remove the cables from the backlight inverter board. Remove the two screws on the backlight inverter board to remove the board. 6-27 6.2.4.5 Remove the Needle Valve Assembly 1. 2. Remove the display assembly. Refer to 6.2.4. Remove the two hexagon set screws on the needle valve knob to remove the knob. 3. Remove the two screws on the needle valve assembly to remove the assembly. 6.2.4.6 Remove the Touch Screen 1. 2. Remove the display assembly. Refer to 6.2.4. Remove the four screws on the cover plate of the touch screen control board to remove the cover plate. 6-28 3. Remove the cables from the touch screen control board. 4. Remove the 11 screws around the display backplane to remove the backplane. 6-29 5. Slightly lift the touch screen off the display front cover to remove the touch screen. 6.2.4.7 Remove the Display Replacement Package 1. 2. Remove the touch screen. After removing the above mentioned components from the display assembly, the remaining part is the display replacement package. 6.2.5 Remove the Module Rack Box Assembly 1. 2. 3. Open the service door. Refer to 6.2.1.1. Remove the rear cover plate assembly. Refer to 6.2.3.4. Remove the cables from the module rack box assembly. 6-30 4. Remove the four screws to remove the module rack box assembly. 5. 6. Remove the module rack fan cable. Remove the four screws to remove the module rack assembly. 6.2.6 Remove the Module Rack Fan 1. 2. 3. Remove the module rack box assembly. Refer to 6.2.5. Remove the module rack fan cable. Remove the four screws to remove the module rack assembly. 4. Remove the two screws to remove the fan together with the bracket. 6-31 5. Remove the four screws to remove the module rack fan. 6.2.7 Remove the Instrument Panel Assembly 1. 2. 3. 3. 4. Open the service door. Refer to 6.2.1.1. Remove the rear cover plate assembly. Refer to 6.2.3.4. Remove the module rack box assembly. Refer to 6.2.5. Remove the related tubes from the gas supply pressure gauge, system switch assembly, and total flowmeter assembly. Remove the three copper pipes on the high pressure gauge from the cylinder yoke assembly. Refer to 6.2.3.9。 6-32 5. Remove the four screws on the pressure gauge panel to remove the panel. 6.2.7.1 Remove the Pressure Gauge 1. 2. Remove the instrument panel assembly. Remove the two screws on the pressure gauge to remove the gas supply pressure gauge and high pressure gauge. 6-33 3. If it is necessary to replace the pressure gauge cover, take out the white gasket from the pressure gauge cover. Remove the pressure gauge cover from the instrument panel. Before replacement, clean up the residual self-adhesive from the instrument panel. White gasket inside the pressure gauge cover 6.2.7.2 Remove the Total Flowmeter 1. 2. Remove the instrument panel assembly. Remove the four screws on the total flowmeter and to remove the total flowmeter. 6.2.7.3 Remove the System Switch Assembly 1. 2. Remove the instrument panel assembly. Remove the four screws on the system switch to remove the switch. 6-34 6.2.7.4 Remove the Indicator Light Board 1. 2. Remove the instrument panel assembly. Remove the two screws on the indicator light board to remove the board. 6.2.8 Remove the Auxiliary Gas Supply Assembly 1. 2. 3. Open the service door and refer to 6.2.1.1. Remove the related tubes from the auxiliary gas supply assembly. Remove the two screws on the auxiliary gas supply assembly. Pull the auxiliary gas supply assembly to the side for some distance. Remove the internal cables to remove this assembly. 6-35 6.2.9 Remove the Circuit Arm Cover 1. 2. 3. Remove the work surface cover plate assembly and refer to 6.2.3.2. Remove the breathing system assembly on the circuit arm cover. Remove the four screws. Lift the circuit arm cover, remove the internal cables, and remove the circuit arm cover. 6.2.10 Remove the AGSS Assembly 1. 2. Remove the transfer tube from the AGSS assembly. Loosen the AGSS groove locking knob on the AGSS assembly. Lift the AGSS assembly along the fixing slider to remove the assembly. 6-36 6.2.11 Remove the Built-in AG Module (when configured with AG module) 1. Remove the 6 screws to remove the cart rear cover plate assembly. 2. Remove the four screws to remove the AG module together with sheet-metal bracket. 3. Remove the four screws to remove the AG module 6-37 6.2.12 Remove the AG Module Inlet Tube Assembly (when configured with AG module) 1. 2. 3. Open the service door. Refer to 6.2.1.1. Remove the rear cover plate assembly. Refer to 6.2.3.4. Remove the AG module and two-way connector from the tubes 4. Remove the two screws to remove the AG module inlet tube assembly. 5. Cut the cable strap fixing the filter. Remove the filter after pulling out the tube connecting the filter. 6-38 6.2.13 Disassemble the Base Assembly 6.2.13.1 Remove the Caster Assembly (The Old base assembly: Four casters are locked synchronously at the brake status) 1. Remove the breathing system and tilt the machine backward. 2. Remove the five screws and remove the caster assembly. 6-39 6.2.13.2 Remove the casters (New base assemblies: three casters are locked and another one can rotate freely at the brake status.) 1. Remove the drawer assemblies according to 6.2.3.1 part. 2. Disassemble the rear panel referring to the third step of 6.2.3.4 part and remove the real board of the cart. 3. Loose the ten screws fixed at two sides of the cart and carry the whole top part down. 4 Loose the four screws fixed at the base board and remove the base cover. 6-40 5 Loose the two screws used for fixing the decoration board at the sides of base. 6. Turn over the base assembly, loose the four screws used for fixing the two sides decoration board at the base and remove the decoration board. 7. According to the replacement of the castors, loose the screws used for fixing the front and rear housing of the base and remove the front and rear housing. 6-41 8. Loose the two screws used for fixing the ram and remove the ram. 9. Loose the two screws used for fixing the installation board of castors shaft and remove the installation board and the drive shaft. 10. Loose the two screws used for fixing the castors and remove the castors. 6-42 6.2.13.3 Disassemble the front and rear housing of the base and the decoration board. Refer to the step 1 to 7 of 6.2.13.2 part. 6.2.13.4 Remove the Brake Indicator Drive Plate I and II 1. 2. Tilt the machine backward. Remove the two screws on the brake indicator drive plate I to remove the plate. 3. Remove the screw on the brake indicator drive plate II to remove the plate. 6.2.13.5 Remove the Brake Assembly 1. 2. Tilt the machine backward. Remove the 6 screws on the brake assembly to remove the assembly. 6-43 6.2.13.6 Remove the Brake Spindle 1. 2. 3. Tilt the machine backward. Remove the brake assembly Remove the 6 screws and remove the brake spindle. (10 screws are needed to be removed for the new base assemblies) 6.2.13.7 Remove the Brake Linkage 1. 2. 3. Tilt the machine backward. Remove the brake indicator drive plate II. Remove the three screws on the brake linkage and remove the brake linkage. 6.3 Disassemble the Breathing System 6.3.1 Remove the O2 Sensor and O2 Cell Unit 1. Remove one end of the O2 cell unit from the circuit arm cover connection. Pull out the O2 cell unit from the port on the breathing system directly. 6-44 2. Rotate the O2 cell shell counter-clockwise to take it out from the O2 cell cable. Then rotate the O2 sensor counter-clockwise to take it out from the O2 cell shell. 6.3.2 Remove the Breathing Tubes NOTE: When disassembling the breathing tube, hold the tube connectors at both ends of the tube to avoid damaging the tube. 1. Remove the filter from the Y piece. 2. Disconnect the breathing tubes from the inspiratory/expiratory connectors on the circuit. 6-45 6.3.3 Remove the Flow Sensor 1. Rotate the rotary cover of the breathing connector counter-clockwise. 2. Pull out the breathing connector together with the rotary cover. Then pull out the flow sensor horizontally. 3. The following pictures show the appearance of inspiratory flow sensor assembly and expiratory flow sensor assembly. Inspiratory flow sensor assembly Expiratory flow sensor assembly 6-46 6.3.4 Remove the Manual Bag Remove the manual bag from the bag arm assembly, as shown below. 6.3.5 Remove the Absorber Canister Assembly 1. Hold and turn the rotary handle clockwise for 90 degrees. 2. Take out the absorber canister assembly horizontally. 6-47 If it is necessary to replace or re-install the Pre-pak absorber sealing cushion (049000143-00), set the internal groove of the Pre-pak absorber sealing cushion in the side flange of the Pre-pak absorber, as shown below. Side flange of the Pre-pak absorber Flat surface of the sealing cushion faces upward. Make sure that the sealing cushion is correctly installed. The comparison between correct installation and incorrect installation is shown below. Correct installation Incorrect installation WARNING: Sodalime is a caustic substance and is a strong irritant to eyes, skin and respiratory system. Affected areas should be flushed with water. If irritation continues after flushing with water, seek medical assistance immediately. 6-48 6.3.6 Remove the CO2 Bypass Assembly 1. 2. Remove the absorber canister assembly. Refer to 6.3.5. Press inward the fasteners on both sides and pull downward to remove the upper half of the CO2 bypass assembly. Then raise the chassis to remove the entire CO2 bypass assembly. If it is necessary to replace or re-install the bypass trigger plate sealing cushion (049000142-00), pay attention to the matching between the interior convex rib of the bypass trigger plate sealing cushion and the groove of the bypass trigger plate when assembling the bypass trigger plate sealing cushion onto the bypass trigger plate, as shown below. Groove of the trigger plate bypass trigger plate sealing cushion Make sure that the sealing cushion is correctly installed. The comparison between correct installation and incorrect installation is shown below. Correct installation Incorrect installation 6-49 6.3.7 Remove the Circuit Main Body Assembly 1. Refer to 6.3.6to remove the CO2 bypass assembly. 2. Pull out the circuit main body assembly from the circuit adapter block assembly. 6.3.8 Remove the Bellows Assembly 1. Turn the bellows housing counter-clockwise and lift off to remove it. 2. Remove the bellows from the bellows base. 6-50 6.3.9 Remove the Pop-off Valve Assembly 1. 2. Refer to 6.3.8to remove the bellows assembly. Remove the four locking screws to remove the Pop-off cover plate assembly. 3. Remove the Pop-off valve assembly. 6.3.10 Remove the Expiratory/Inspiratory Check Valve Assembly 1. Turn the valve cover clockwise to remove it. 6-51 2. Take out the check valve, as shown below. 6.3.11 Remove the Water Collection Cup 1. Turn the water collection cup clockwise to remove the cup. 2. The removed water collection cup is as shown below. 6-52 6.3.12 Remove the Airway Pressure Gauge Lift the airway pressure gauge straight up to remove it 6.3.13 Remove the Bag Arm Assembly 1. Turn the locking nut counter-clockwise and pull it out vertically. 2. Lift straight up to remove the bag arm assembly. 6-53 6.3.14 Remove the Upper Rear Cover Assembly and Lower Rear Cover Assembly 1. 2. Refer to 6.3.1,6.3.2,6.3.3,6.3.4,6.3.5,6.3.6,6.3.7,6.3.8,6.3.11,6.3.12, 6.3.13 to remove the O2 sensor, breathing tubes, manual bag, absorber canister assembly, CO2 bypass assembly, circuit main body assembly, bellows assembly, water collection cup, airway pressure gauge, and bar arm assembly. Fully loosen the 6 screws, as shown below. 3. Fully loosen the hand screws on the upper cover, as shown below. 4. Turn over the circuit main body. Pull upward to remove the upper rear cover assembly. 6-54 5. Pull left to remove the lower rear cover assembly. 6.3.15 Remove the Upper Front Cover Assembly, Median Plate Assembly, and Lower Front Cover Assembly 1. 2. Refer to 6.3.14to remove the upper rear cover assembly and lower rear cover assembly. Remove the two screws on the lower rear cover. 3. Fully loosen the 6 screw on the upper cover. 6-55 4. Fully loosen the captive screws on the upper cover. 5. Tightly hold the upper cover assembly and pull it upward to remove. 7. Pull the median plate assembly upward to remove it. Median plate assembly Lower cover assembly 6-56 6.3.16 Remove the Auto/Manual Switch Assembly 1. 2. Refer to 6.3.15 to remove the upper cover assembly. Turn over the upper cover assembly and remove the three screws to remove the Auto/Manual switch assembly, as shown below. 3. Remove the three screws to remove the Auto/Manual switch bracket, as shown below. 4. Remove the O-ring. Pull out the connection shaft to pull out the Auto/Manual switch axis. Remove the O-ring and pull out the connection shaft. Auto/Manual switch axis Auto/Manual switch axis compress spring 6-57 5. Remove the Auto/Manual switch compress spring to replace the two sealing rings. Sealing ring 6.3.17 Remove the APL Valve Assembly Turn the locking nut counter-clockwise and pull out the APL valve assembly vertically. 6-58 7 Replacement Parts 7.1 Introduction The A5 anesthesia system can be divided into 18 parts according to its structure and functions. Each part contains several replaceable parts. Table 8-1 through Table 8-18 provide the information about the replaceable parts and Diagram 8-1 through Diagram 8-14 give the positions of replaceable parts. When selecting replaceable parts, consideration should be given to the characteristics of the parts, cost of replacement, and maintenance efficiency. When the parts whose sub components are not convenient to replace (such as the electrical component on the board) are faulty, replacing the board can improve the maintenance efficiency. For example, if a pressure gauge on the instrument panel is faulty, replacing the pressure gauge can reduce the cost. 7.2 Ordering Replaceable Parts Provide the following information to order replaceable parts: FRU code of the parts Number of the parts in the document table Description of the characteristics of the parts For example: P/N: 801-0631-00001-00 Auxiliary gas supply, No.1 7-1 7.3 Diagrams and Tables 7.3.1 A5 Structure 20 19 1 18 2 3 13 4 12 5 11 6 10 9 7 8 7a 7-2 No. 1 2 3 4 5 6 7 7a 8 9 10 11 12 13 18 19 20 Description P/N Auxiliary gas supply assembly (American standard) service parts 115-034401-00 Auxiliary gas supply assembly (Europe standard) service parts Breathing system CO2 bypass assembly O2 supply pressure gauge assembly N2O supply pressure gauge assembly Air supply pressure gauge assembly Spare maintenance kit of drawer assembly Drawer handle Drawer key Total flowmeter The pressure gauge assembly of O2 high pressure The pressure gauge assembly of N2O high pressure The pressure gauge assembly of Air high pressure System switch knob Backup vaporizer manifold Vaporizer manifold Panel of gauge(FRU/US) Panel of gauge(FRU/EU) Panel of gauge(FRU/EU/big tank) Panel of gauge(FRU/reverse EU) 115-034400-00 115-025569-00 115-036378-00 115-024841-00 115-024842-00 115-024843-00 115-034451-00 801-0631-00140-00 115-023320-00 801-0631-00008-00 115-024844-00 115-024845-00 115-024846-00 801-0631-00012-00 801-0631-00076-00 801-0631-00024-00 115-023177-00 115-023179-00 115-023180-00 115-023181-00 7-3 21 22 23 29 24 25 28 27 26 29 30 No. Description P/N 20 21 maintenance cover Air cylinder yoke assembly (with copper pipe) Air cylinder yoke assembly (with copper pipe/air driven) N2O cylinder yoke assembly (with copper pipe) O2 cylinder yoke assembly (with copper pipe) O2 cylinder yoke assembly (with copper pipe/air driven) Caster 115-026809-00 115-034396-00 115-040986-00 AGSS transfer tube AGSS assembly (high flow) AGSS assembly (low flow) Work surface cover plate Cylinder lock block Tank knob 801-0631-00074-00 115-017377-00 115-017378-00 801-0631-00120-00 041-002678-00 043-001266-00 22 23 24 26 27 28 29 30 7-4 115-034397-00 115-034395-00 115-040984-00 034-000285-00 7.3.2 A5 Upper Half 1 2 17 3 16 4 5 15 6 14 17 7 No. 1 2 3 4 5 6 7 8 9 10 8 9 10 11 12 Description Pneumatic resistor module assembly ORC assembly N2O supply inlet assembly(NIST) Air supply inlet assembly(NIST) O2 supply inlet assembly(NIST) N2O pressure regulator assembly O2 and Air pressure regulator assembly O2 system switch Indicator light board PCBA Pressure test tube (left) 11 Pressure test tube (middle) 12 Pressure test tube (right) 13 14 15 16 17 Electro flowmeter assembly Electro flowmeter board (FRU/0634/without N2O) PCBA Back pressure valve Poppet valve Lithium battery (11.1V, 4500mAh) N2O needle valve 7-5 13 P/N 801-0631-00119-00 115-051809-00 801-0631-00128-00 801-0631-00129-00 115-015435-00 801-0631-00086-00 801-0631-00087-00 801-0631-00003-00 801-0631-00004-00 801-0631-00123-00 115-034389-00 801-0631-00124-00 115-034390-00 801-0631-00125-00 115-034391-00 115-044266-00 (O2/N2O/air) 115-044267-00 (O2/air) 115-024621-00 115-033941-00 115-036458-01 115-062081-00 082-002943-00 7.3.3 A5 Hardware Box 9 8 7 6 11 5 4 12 3 13 2 14 1 15 No. Description P/N 1 2 3 4 5 Top light board PCBA Speaker and connection line Power socket 3C certificate (black) Breaker (3.0A) Breaker (10.0A) Breaker (5.0A) Filter Fan Power board PCBA Air deflector plate Main control board Battery adaptation board PCBA Monitoring module Motherboard PCBA Three-way valve assembly (applicable for system without ACGO configuration) Three-way valve assembly (ACGO) FRU 801-0631-00039-00 801-0631-00038-00 801-0631-00132-00 801-0631-00031-00 801-0631-00030-00 801-0631-00130-00 801-0631-00029-00 801-0631-00028-00 801-0631-00025-00 / 115-035003-00 801-0631-00109-00 115-051206-00 801-0631-00108-00 6 7 8 9 11 12 13 14 15 7-6 801-0631-00046-00 801-0631-00146-00 7.3.4 A5 Work Surface 2 3 1 4 11 5 6 7 8 9 10 No. Description P/N 1 2 3 4 5 6 7 Limit switch Auto/Manual switch spring Drive shaft Spare maintenance kit of circuit heater Circuit adapter block cover O2 sensor connector O2 flush assembly 8 ACGO assembly 9 10 11 Track pad FRU Touch pad Spare maintenance kit of expiratory valve assembly (DSP) Gas reservoir assembly 801-0631-00101-00 801-0631-00114-00 801-0631-00113-00 115-034450-00 801-0631-00068-00 801-0631-00067-00 801-0631-00044-00 801-0631-00045-00 (Common Gas Outlet assembly) 115-043752-00 (ACGO) 115-043753-00 (independent ACGO/anesthesia calculation) 115-017032-00 801-0631-00052-00 13 7-7 115-036693-00 801-0616-00007-00 7.3.5 A5 Patient Circuit Main Body 1 14 2 13 3 12 11 4 5 10 9 6 7 8 No. Description P/N 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Bellows housing Bellows assembly Airway pressure gauge Expiratory flow sensor assembly Inspiratory/expiratory interface Breathing circuit unit Rotary cap for inspiratory/expiratory interface Water collection cup Inspiratory flow sensor assembly O2 cell cover and O2 cell cable APL valve assembly Auto/Manual switch assembly Check valve Bag arm assembly 801-0631-00054-00 0601-30-78968 115-051819-00 801-0631-00056-00 801-0631-00057-00 115-025565-00 801-0631-00059-00 801-0631-00058-00 801-0631-00060-00 801-0631-00102-00 801-0631-00062-00 801-0631-00065-00 801-0631-00104-00 115-048600-00 7-8 7.3.6 A5 Pre-pak Absorber Canister Assembly 4 3 1 2 No. Description P/N 1 2 3 4 5 Absorber hose CO2 absorber base Bottom plate CO2 absorber canister CO2 bypass assembly 801-0631-00092-00 801-0631-00100-00 801-0631-00100-00 801-0631-00066-00 115-036378-00 7-9 7.3.7 A5 Valve Assembly 1 2 3 No. Description P/N 1 2 3 Valve cover Valve shield Valve 801-0631-00061-00 801-0631-00110-00 801-0631-00111-00 7-10 7.3.8 A5 O2 Cell Assembly 1 3 2 No. Description P/N 1 2 3 O2 cell cover MedicelMOX-2 O2 sensor (O2 sensor) O2 cell cable 801-0631-00090-00 0611-10-45654 801-0631-00091-00 7-11 7.3.9 A5 Display Assembly 1 3 2 4 No. Description P/N 1 / N2O needle valve (Applicable for device with O2, N2O and Air configuration) Air needle valve (Applicable for device with O2, N2O and Air configuration) O2 needle valve (Applicable for device with O2, N2O and Air configuration) O2 needle valve (Applicable for device with O2 and Air configuration) / 2 3 4 3, 4 7-12 115-036762-00 115-036457-00 115-036761-00 115-024292-00 7.3.10 A5 Display Assembly Main Body 1 2 3 4 5 No. Description P/N 1 Alarm lamp board PCBA Touch screen replacement material package FRU (0634) Display screen Touch screen control board Display interface board PCBA 801-0631-00019-00 2 3 4 5 7-13 801-0631-00014-00 115-036373-00 801-0631-00018-00 801-0631-00017-00 7.3.11 A5 Vaporizer Manifold 1 2 3 No. Description P/N 1 2 3 Connector (vaporizer fixing) Vaporizer manifold valve Vaporizer manifold spring 801-0631-00117-00 801-0631-00106-00 801-0631-00107-00 7-14 7.3.12 A5 Auxiliary Gas Outlet Assembly 1 2 No. Description P/N 1 2 Auxiliary gas supply flowmeter lighting board PCBA Auxiliary gas outlet fitting (FRU) 801-0631-00002-00 801-0631-00122-00 7-15 7.3.13 Base Assembly 1 2 3 4 5 No. Description P/N 1 2a 2b 3 4 Brake assembly (FRU) Long ram of the brake (0634) Short ram of the brake (0634) Indicator light drive board Brake principal axis Chassis assembly FRU (CE) Chassis assembly FRU (CFDA) 115-016692-00 042-016810-00 042-017262-00 801-0631-00072-00 041-022463-00 115-044426-00 115-044424-00 5 7-16 7.3.14 Cart Assembly 2 3 1 4 No. 1 2 3 4 Description M01A AION AG module(03-31 type) AG calculation inlet tube assembly (0632) Filter.0.45um,TP026AATV004AD01 Anesthetic signal adapter board (service parts) P/N 115-002504-00 115-017879-00 082-000423-00 801-0613-00033-00 7.3.15 Other 1 2 3 7-17 No. Description P/N 1 2 3 – – – Cylinder wrench A5 preoperative checklist A5 auxiliary O2/Air reference card Basic maintenance kit (12-month) (0635) Expand maintenance kit (36-month) (0635) Complete maintenance kit (12-month) (0635) 115-033063-00 801-0631-00081-00 801-0631-00082-00 115-036367-00 115-036368-00 115-036369-00 7.3.16 Tubes No. Description P/N 1 2 3 4 5 6 7 8 9 10 11 12 13 14 7mmX10mm transparent PU tube (polyether) 3/32"X7/32"X100ft silicone 8mmx5mm transparent PU precision hose 20X25mm silica gel 4*6 transparent PU tube Tubing. 4mmx2.5mm transparent PU precision hose Tubing.MPF tube 2mmX3.5mm (1feet) PE Tubing.ID2XOD4 Tubing.ID3/32''XOD5/32'',Silicon System Tubing PUR.AION,1.4/2.8mm,60-12110-00 System Tubing PUR.AION, 2.2/4.4mm,60-12120-00 Moisture Exchanger Assembly Driver gas spring tube assembly APL valve spring tube assembly 082-000519-00 A21-000007--M6G-020014--M6G-020018--M6G-020026--M6G-020046--M6G-020005--082-001198-00 60-12140-00 9200-10-10556 9200-10-10557 082-001461-00 115-034448-00 115-034449-00 7-18 7.3.17 O-ring No. Description P/N Notes 1 Valve port sealant Cushion of CO2 bypass assembly Absorber canister external cushion Absorber canister internal cushion CO2 absorber hose pop-off rubber feet Bellows absorber bottom cushion AGSS filter screen 14X2.65 O-ring (used for vaporizer manifold) 30X2 O-ring 25X2 O-ring 27X1.5 O-ring 18X2.5 O-ring 20X1.5 O-ring 049-000140-00 Used for CO2 bypass shaft 049-000142-00 / 049-000143-00 / 049-000145-00 / 049-000146-00 049-000240-00 / / 049-000243-00 / 082-000506-00 082-000624-00 082-000625-00 082-001501-00 082-000627-00 082-001503-00 23.47X2.95 O-ring 082-001504-00 52X2 O-ring 29X2.62 O-ring 082-000630-00 082-000633-00 6.07X1.78 O-ring 082-000641-00 29.82X2.62 O-ring 40X2.2 O-ring 8.5X2.0 O-ring 082-000642-00 082-000648-00 082-001525-00 8.5X2 O-ring 082-000665-00 4.7X1.8 O-ring 082-000667-00 6X1 O-ring 082-000669-00 15.54X2.62 O-ring 082-000673-00 4.47X1.78 O-ring O-ring (used for airway pressure gauge) 082-000679-00 / Used for vaporizer fixing connector Base of bag arm APL valve Used for check valve dome Base of breathing system Used for check valve Used for water collection cup and CO2 bypass assembly Auto/Manual switch Bellows base Bottom of the breathing system cover screw APL valve Auto/Manual switch O2 cell port Used for rotary module of breathing circuit Used for rotary module of breathing circuit Used for Auto/Manual switch Used for bag arm and O2 cell cover Used for CO2 bypass shaft 082-001524-00 Used for airway pressure gauge 16X2 O-ring M6M-010058--- O-ring maintenance kit 801-0631-00141-00 2 3 4 5 6 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 082-000934-00 7-19 Used for rotary module of breathing circuit / FOR YOUR NOTES 7-20 A Test Items A.1 Test items after installation Hospital name: Instrument SN: Test Item After Installation Reference Section in the Service Manual Test Completed □ Breathing System Leak Test in Mechanical Ventilation Mode 2.4.2.1 Breathing System Leak Test in Manual Ventilation Mode 2.4.2.2 □ N2O Cylinder Leak Test 2.5.1 □ O2 Cylinder Leak Test 2.5.2 □ Air Cylinder Leak Test 2.5.3 □ N2O Pipeline Supply Pressure Leak Test 2.5.4 □ O2 Pipeline Supply Pressure Leak Test 2.5.5 □ Air Pipeline Supply Pressure Leak Test 2.5.6 □ Vaporizer Leak Test 3.11.3 □ O2 Flush Verification 2.4.1.1 □ O2 Sensor Calibration 2.4.3 □ Manual Mode Ventilation Test 3.9.2 □ VCV Adult Ventilation Mode Test 3.9.5 □ PCV Adult Ventilation Mode Test 3.9.8 □ A-1 A.2 Test items after maintenance Hospital name: Instrument SN.: Test Item After Maintenance Reference Section in the Service Manual Test Completed □ Breathing System Leak Test in Mechanical Ventilation Mode 2.4.2.1 Breathing System Leak Test in Manual Ventilation Mode 2.4.2.2 □ Vaporizer Leak Test 3.11.3 □ O2 Flush Verification 2.4.1.1 □ O2 Sensor Calibration 2.4.3 □ Manual Mode Ventilation Test 3.9.2 □ Power Failure Test 3.13.1 □ Performance Verification 3.9 □ Check the Sensor Zero Point 3.6.3 □ Check the Accuracy of Flow Sensors 3.6.4 □ Check the Accuracy of Pressure Sensors 3.6.5 □ A-2 A.3 Test items after service Hospital name: Instrument SN.: Reference Section in the Service Manual Test Item After Service Test Completed □ System self-test □ Breathing System Leak Test in Mechanical Ventilation Mode 2.4.2.1 Breathing System Leak Test in Manual Ventilation Mode 2.4.2.2 □ N2O Cylinder Leak Test 3.7.1 □ O2 Cylinder Leak Test 3.7.2 □ Air Cylinder Leak Test 3.7.3 □ N2O Pipeline Supply Pressure Leak Test 3.7.4 □ O2 Pipeline Supply Pressure Leak Test 3.7.5 □ Air Pipeline Supply Pressure Leak Test 3.7.6 □ Needle valve correspondence test □ Needle valve microswitch test □ Alarm of Drive Gas Pressure Low, N2O Cutoff Test 3.8.3 □ O2:N2O Ratio System 2.4.1.2 □ Vaporizer Interlock Test 3.11.1 □ Vaporizer Accuracy Test 3.11.2 □ Vaporizer Leak Test 3.11.3 □ O2 Flush Verification 2.4.1.1 □ O2 Sensor Calibration 2.4.3 □ Manual Mode Ventilation Test 3.9.2 □ Sample Gas Emission Test/AGSS Test (optional configuration) 3.8.1 Power Failure Test 3.12.1 □ Alarm of Drive Gas Pressure Low, N2O Cutoff Test 3.8.3 □ Performance Verification 3.9 □ Check the Sensor Zero Point 3.6.3 □ Check the Accuracy of Flow Sensors 3.6.4 □ Check the Accuracy of Pressure Sensors 3.6.5 □ Auxiliary AC Electrical Outlets Test 3.14 □ Work Light and Auxiliary Gas Supply Backlight Test 3.13.2 □ Electrical Safety Inspection 3.15.2 □ A-3 □ FOR YOUR NOTES A-4 P/N: 046-002301-00(9.0)