Whole Body X-ray CT System Presto(V) TECHNICAL GUIDE -- Installation & Adjustment --- Check & Service -- Tokyo, Japan L2E-BW1444-7 CONFIDENTIALITY : This material contains proprietary information of Hitachi Medical Corporation and it shall not be disclosed to the third party or reproduced without the prior written permission of Hitachi Medical Corporation. Copyright © Hitachi Medical Corporation. 2004, 2005,2006. All rights reserved. NOTE The tightening torque of bolts, screws, nuts, and double nuts in the work specified in this manual shall conform to the values shown in Table 1 and 2. Table 1 Tightening torque of bolts and screws Kinds Tightening torque N・cm (kgf・cm) Sizes Max. Min. M3 98 (10) 69 (7) Flat head screw M4 216 (22) 147 (15) Pan head screw M5 392 (40) 275 (28) M6 540 (55) 373 (38) M3 108 (11) 79 (8) M4 265 (27) 187 (19) M5 539 (55) 392 (40) M6 833 (85) 588 (60) M8 1990 (203) 1392 (142) M10 3528 (360) 2450 (250) M12 5586 (570) 3920 (400) M3 138 (14) 98 (10) M4 343 (35) 245 (25) M5 696 (71) 510 (52) M6 1078 (110) 765 (78) M8 2588 (264) 1804 (184) M10 4587 (468) 3185 (325) M12 7262 (741) 5096 (520) M16 15925 (1625) 11143 (1137) Hex. head bolt Hex. socket head bolt (1) 1 L2E-BW1444 Table 2 Tightening torque of nuts * Nut M3 108 (11) 79 (8) M4 265 (27) 187 (19) M5 539 (55) 392 (40) M6 833 (85) 588 (60) M8 1990 (203) 1392 (142) M10 3528 (360) 2450 (250) M12 5586 (570) 3920 (400) * Tightening method of double nuts (1) Fix the low nut by a spanner. (2) Tighten the normal height nut by the other spanner. The tightening torque shall conform to the values specified in the above table. Normal height nut Low nut (2) 2 L2E-BW1444 Important safety work for installation of X-ray tube(3.5MHU, 5MHU) installation engineer Customer installation date NO. Check point Criteria 1 Four hexagon socket head bolts (M8, 50mm long) for trunnion rings shall be tighten by 19.6± 2N・m (200±20kgf・cm) of torque. Check that the fixing bolts are tightly fixed that tightening torque is 19.6±2N・m (200±20kgf・cm). Pass ・ Fail 2 Four hexagon socket head bolts (M8, 30mm long) for the base shall be tighten by 19.6±2N・m (200±20kgf・cm) of torque. Check that the fixing bolts are tightly fixed that tightening torque is 19.6±2N・m (200±20kgf・cm). Pass ・ Fail 3 Seven hexagon socket head bolts (M6) for the mounting plates 1 and 2 shall be tighten by 8± 0.8N・m (82±8.2kgf・cm) of torque. Check that the fixing bolts are tightly fixed that tightening torque is 8±0.8N・m (82±8.2kgf・cm). Pass ・ Fail No.1 4-M8X50L Hexagon socket head bolt Trunnions ring + - H9 CS H7 H8 No.2 4-M8X30L Hexagon socket head bolt Mounting bracket 1 Base Mounting bracket 2 No.3 7-M6 nut, spring, and plain washer (3) 3 L2E-BW1444 Contents Page Section 1 CARRYING -IN, MOUNTING, AND ASSEMBLING.......................................................................... 1-1 1.1 Carrying in and mounting of scanner unit .................................................................................... 1-1 1.2 Leveling of scanner unit ............................................................................................................... 1-6 1.3 Carrying in the patient table ......................................................................................................... 1-7 1.4 Leveling and fixing of patient table.............................................................................................. 1-8 1.5 Mounting of other units .............................................................................................................. 1-12 1.6 1.7 1.8 Section 2 Operation unit ............................................................................................................... 1-12 1.5.2 X-ray control unit.......................................................................................................... 1-13 Mounting of detector and detector side collimator...................................................................1-14A 1.6.1 Mounting of detector ..................................................................................................1-14A 1.6.2 Mounting of detector side collimator..........................................................................1-17A Mounting of X-ray tube unit and cooler ..................................................................................... 1-22 1.7.1 Mounting procedure of X-ray tube assembly (Presto 3.5MHU) ................................... 1-22 1.7.2 Mounting procedure of cooler ...................................................................................... 1-23 1.7.3 Mounting procedure of X-ray tube assembly (Presto 5 MHU) ..................................... 1-24 1.7.4 Mounting procedure of cooler ...................................................................................... 1-25 Removal of protective materials for transport ............................................................................ 1-26 1.9 Internal inspection of patient table ............................................................................................. 1-27 1.10 Mounting of covers .................................................................................................................... 1-29 WIRING .................................................................................................................................................. 2-1 2.1 Wiring of cables between units..................................................................................................... 2-1 2.2 Wiring of cables in each unit ........................................................................................................ 2-6 2.3 Section 3 1.5.1 2.2.1 Scanner ........................................................................................................................... 2-6 2.2.2 Table wiring .................................................................................................................... 2-8 2.2.3 Mounting of foot switch, table electric unit cover, and table control unit....................... 2-9 2.2.4 Image processing unit ................................................................................................... 2-13 2.2.5 X-ray control unit.......................................................................................................... 2-23 2.2.6 Mounting the counter .................................................................................................... 2-27 Internal wiring of system ............................................................................................................ 2-28 2.3.1 Wiring on scanner rotary plate...................................................................................... 2-28 2.3.2 Detector cabling work................................................................................................... 2-30 ADJUSTMENT....................................................................................................................................... 3-1 3.1 Turning on the power supply........................................................................................................ 3-1 3.2 Operation check of image processing unit in operation Unit........................................................ 3-1 3.3 Operation check and adjustment of scanner unit .......................................................................... 3-2 3.4 Operation check of patient table and adjustment of scanner gantry tilting ................................. 3-20 3.5 Laser marker adjustment ............................................................................................................ 3-24 3.6 Adjustment of X-ray system ....................................................................................................... 3-27 3.6.1 X-ray tube assembly check ........................................................................................... 3-27 3.6.2 Anode rotation drive unit (starter) adjustment. ............................................................. 3-27 3.6.2.1 Starter CPU-CONT PCB adjusting method .................................................................. 3-31 (4) (4A) 4 L2E-BW1444 3.6.2.2 Adjustment of control knobs on PC board and dial time (VR adjustment while checking the LED display) ........................................................ 3-32 3.6.2.3 Voltage adjustment (Measure H7-H9 voltage by a true RMS meter)............................ 3-32 3.6.2.4 CPU-CONT PC board adjustment ................................................................................ 3-33 3.6.2.5 Status display of anode rotation drive unit.................................................................... 3-34 3.7 3.8 3.6.3 Adjustment of mA waveform and checking of kVmA .................................................. 3-35 3.6.4 X-ray tube aging ........................................................................................................... 3-37 3.6.5 Pre-Heating Parameter of RTC Board .......................................................................... 3-39 Adjustment of X-ray field........................................................................................................... 3-40 3.7.1 Adjustment in X-axis direction ..................................................................................... 3-40 3.7.2 R-direction adjustment (Common to each X-ray tube assembly).................................. 3-41 3.7.3 Normal condition of X-ray tube.................................................................................... 3-42 3.7.4 X-ray beam exposure field adjustment (slice thickness direction) ................................ 3-43 Preamplifier check...................................................................................................................... 3-47 3.8.1 Bus line check............................................................................................................... 3-47 3.8.2 Output sequence check ................................................................................................. 3-48 3.9 Checking the influences of vibration noises on measuring data (Vibration check) .................... 3-51 3.10 Center adjustment....................................................................................................................... 3-55 3.11 Compensator position adjustment .............................................................................................. 3-58 3.12A Focus shift detector (FSD) adjustment *To be applied to PB1 board Rev.1(Display VR3**)................................................................. 3-61 3.12B Focus shift detector (FSD) adjustment *To be applied to PB2 board Rev.2(Dsiplay VR*)..................................................................... 3-72 3.13 Detector side collimator (Post Patient Collimator - PPC) position adjustment .......................... 3-83 3.14 Phantom calibration.................................................................................................................... 3-88 3.14.1 Procedure ...................................................................................................................... 3-88 3.14.2 Mounting method of phantom mounting bracket .......................................................... 3-89 3.14.3 X-ray detection system warming up.............................................................................. 3-94 3.14.4 X-ray tube warm up and air calibration ........................................................................ 3-94 3.14.5 Phantom position adjustment and phantom calibration................................................. 3-96 3.15 CT number adjustment ............................................................................................................. 3-104 3.15.1 Procedure .................................................................................................................... 3-104 3.15.2 CT number adjustment scan........................................................................................ 3-105 3.15.3 Perform CT number adjustment calculation ............................................................... 3-108 3.16 Interphone (intercommunication) check....................................................................................3-111 3.17 Institution info. settings ............................................................................................................ 3-112 3.18 Computer name change ............................................................................................................ 3-113 3.19 Communication setting............................................................................................................. 3-115 3.19.1 LAN card setting......................................................................................................... 3-115 3.19.2 Deletion of DICOM transmission software................................................................. 3-115 3.19.3 Entry of DICOM transmission software...................................................................... 3-116 3.19.4 Setting of DICOM transmission software ................................................................... 3-118 3.19.5 Setting of storage commitment software ..................................................................... 3-119 3.19.6 Setting of filming software.......................................................................................... 3-120 3.19.7 Entry of SMPTE pattern ............................................................................................. 3-121 (5) 5 L2E-BW1444 Section 4 Reference data ......................................................................................................................................... 4-1 4.1 Stationary scan ............................................................................................................................. 4-1 4.2 Detector - preamplifier connectors correspondence table ............................................................ 4-2 4.3 Section 5 Details of service information ...................................................................................................... 4-5 Maintenance software.............................................................................................................................. 5-1 5.1 Maintenance mode ON/OFF ........................................................................................................ 5-1 5.1.1 Maintenance mode ON ................................................................................................... 5-1 5.1.2 Maintenance mode OFF.................................................................................................. 5-1 5.2 Maintenance settings .................................................................................................................... 5-2 5.3 Date / time setting......................................................................................................................... 5-2 5.4 Study number setting .................................................................................................................... 5-2 5.5 System configuration settings....................................................................................................... 5-3 5.6 Maintenance service ..................................................................................................................... 5-4 5.7 Raw Data Analysis ....................................................................................................................... 5-4 5.7.1 Introduction..................................................................................................................... 5-4 5.7.2 Raw Data Analysis start .................................................................................................. 5-5 5.7.3 Opening data................................................................................................................... 5-5 5.7.3.1 Raw Data ........................................................................................................................ 5-5 5.7.3.2 Air calibration data ......................................................................................................... 5-6 5.7.3.3 Phantom profile............................................................................................................... 5-6 5.7.3.4 Phantom calibration data................................................................................................. 5-7 5.7.4 Closing data .................................................................................................................... 5-8 5.7.5 Raw Data Analysis end ................................................................................................... 5-9 5.7.6 Raw Data Analysis screen function................................................................................. 5-9 5.7.6.1 A change of window width and window level ................................................................ 5-9 5.7.6.2 Slice data selection.......................................................................................................... 5-9 5.8 5.7.7 Analysis menu................................................................................................................. 5-9 5.7.8 CACK ........................................................................................................................... 5-17 Check program ........................................................................................................................... 5-19 5.8.1 Basic window structure................................................................................................. 5-19 5.8.2 Preamplifier check ........................................................................................................ 5-21 5.8.3 Scanner check ............................................................................................................... 5-22 5.8.3.1 Scanner system check ................................................................................................... 5-22 5.8.3.2 Table system check ....................................................................................................... 5-23 5.8.3.3 X-ray control system check........................................................................................... 5-25 5.8.3.4 PreAmplifier signal ReTransmission system check ...................................................... 5-25 5.8.3.5 DSP system check......................................................................................................... 5-26 5.9 Control Error List ..................................................................................................................... 5-28 5.9.1 Control error list start/end............................................................................................. 5-28 5.9.2 Setting of the order of date and time............................................................................. 5-28 5.9.3 Display of a specified process only............................................................................... 5-29 5.9.4 Display of a specified message only ............................................................................. 5-29 5.9.5 Setting of the error log information Directory .............................................................. 5-29 5.9.6 Saving of the error log information to MO ................................................................... 5-31 (6) 6 L2E-BW1444 5.10 Service backup ........................................................................................................................... 5-32 5.10.1 Start / end...................................................................................................................... 5-32 5.10.2 Initialization of saving MO ........................................................................................... 5-33 5.10.3 Saving ........................................................................................................................... 5-34 5.10.4 Loading ......................................................................................................................... 5-35 5.10.5 Deletion ........................................................................................................................ 5-36 5.10.6 Other data ..................................................................................................................... 5-37 5.11 System maintenance mode ......................................................................................................... 5-38 5.11.1 System maintenance mode / release mode selection ..................................................... 5-38 5.12 Error codes list ........................................................................................................................... 5-39 5.12.1 Control error codes list ................................................................................................. 5-39 5.12.2 System error log codes table ......................................................................................... 5-56 5.12.3 Details of DSP error codes............................................................................................ 5-70 5.13 Event viewer............................................................................................................................... 5-71 5.13.1 Introduction................................................................................................................... 5-71 5.13.2 Method of use ............................................................................................................... 5-71 5.14 Touch panel calibration .............................................................................................................. 5-73 5.14.1 Locating touch panels ................................................................................................... 5-73 5.14.2 Calibration of touch panel 1.......................................................................................... 5-74 5.14.3 Calibration of touch panel 2.......................................................................................... 5-75 5.15 Cache off setting of MO drive.................................................................................................... 5-77 5.16 Tube current graph ..................................................................................................................... 5-78 5.16.1 Start/Exit ....................................................................................................................... 5-78 5.16.2 Opening Raw data......................................................................................................... 5-79 5.16.3 Viewing chart ................................................................................................................ 5-80 Section 6 Section 7 Software installation ................................................................................................................................ 6-1 6.1 Summary of software installation ................................................................................................. 6-1 6.2 System installation........................................................................................................................ 6-3 6.2.1 CPU1 system installation ................................................................................................ 6-4 6.2.2 CPU0 system installation ................................................................................................ 6-6 6.3 Application installation ................................................................................................................ 6-7 6.4 Setting of System Serial No. (only when the system has been installed)...................................... 6-8 RAID disk................................................................................................................................................ 7-1 7.1 Decision of RAID disk ................................................................................................................. 7-1 7.2 High Point RAID.......................................................................................................................... 7-2 7.2.1 Initialization of RAID ..................................................................................................... 7-2 7.2.2 Remedial measures against the occurrence of troubles ................................................... 7-8 7.2.2.1 In case of Critical Menu.................................................................................................. 7-8 7.2.2.2 In case of Broken Menu.................................................................................................. 7-9 7.3 Promise RAID ............................................................................................................................ 7-10 7.3.1 Initializing the RAID disk............................................................................................. 7-10 7.3.2 Remedial measures against the occurrence of troubles ................................................. 7-14 (7) 7 L2E-BW1444 Section 8 Scanner firmware version up method....................................................................................................... 8-1 8.1 Preparation for version up ............................................................................................................ 8-1 8.1.1 Preparation for version up of VMCPU, VSCPU, and VTCPU/VTCPU2....................... 8-1 8.1.2 Preparation for version up of RXCPU and VATCPU/ VTCPU2 .................................... 8-2 8.1.3 Preparation for version up of XCPU............................................................................... 8-2 8.2. Down loading method .................................................................................................................. 8-3 8.3. Version up end processing............................................................................................................ 8-5 8.4 8.3.1 Version up end processing of VMCPU, VSCPU, and VTCPU/ VTCPU2 ..................... 8-5 8.3.2 Version up end processing of RXCPU and VATCPU ..................................................... 8-6 8.3.3 Version up end processing of XCPU............................................................................... 8-6 Version check ............................................................................................................................... 8-6 Section 9 Scanner FPGA version up method ........................................................................................................... 9-1 9.1 9.2 9.3 RXCPU, XCPU version up method ............................................................................................. 9-1 Version up method of MFPGA, TFPGA, TXRFPGA and TXSFPGA......................................... 9-5 Version-up procedure of MFPGA, TFPGA and RFPGA (for the system installed with the MTCPU2)................................................................................ 9-6 Section 10 Various setting methods of table .......................................................................................................... 10-1 10.1 10.2 Preset height setting method....................................................................................................... 10-1 Stop position accuracy setting method ....................................................................................... 10-2 Appendix 1 System configuration setting items ............................................................................................Appendix -1 Appendix 2 How to initialize HU..................................................................................................................Appendix -2 (8) 8 L2E-BW1444 Section 1 CARRYING -IN, MOUNTING, AND ASSEMBLING 1.1 Carrying in and mounting of scanner unit The scanner unit is packaged together with its transport cart. After unpacking, carry in and mount the scanner unit according to the following procedure. (1) Scribing the mounting position Determine the mounting position, referring to the mounting layout of the scanner unit and table shown in Fig.1.1.1, and scribe the outside dimensions on the floor referring to the dimensions in Fig.1.1.2. Foot rest and table top are used:2775 Foot rest and table top are used:4180 ※3950 (3750) 1405 (1205) ※2545 830 690 In case of forward tilting In case of forward tilting 800 300 (100) 740 470 ※1175 ※475 1000 500 500 1980 250 (100) (1090) 1430 1010 640 520 (300) 2750 (2380) 1035 1240 In case of backward tilting 365 1560 1915 (1715) Scale 1:50 Unit mm Fig. 1.1.1 Mounting layout of scanner unit and table Caution 1) Don’t put any obstacle within the two-point dot-dash line when mounting the scanner unit. Numeric values show the required sizes. 2) Parenthesized dimensions in ( ) show the mounting limit sizes. 3) The dotted line portion shows the required area for opening the front panel and rear cover of the scanner unit during servicing. 4) The shadowed portion shows the movable range of the patient table. 5) For asterisked sizes (※), their required sizes depend upon the option attachments of the patient table. 6) The sizes (in case of forward tilting) or (backward tilting) shows the cover position sizes during tilting. 1-1 L2E-BW1444 7250N (740Kgf) 4-M10 anchor bolt 455 122 1414 633 672 1286 64 500 640 359 7250N (740Kgf) 2940N (300Kgf) 4-M10 anchor bolt 3585 Fig. 1.1.2 Dimensions of scanner table (2) Bring the scanner unit to the mounting position. For carrying in the scanner unit, the route should have the dimensions specified in Fig.1.1.3. Check if the route size is as specified. Fig. 1.1.3 Required space for carrying in the scanner unit 1-2 L2E-BW1444 Dimensions and weight when carrying in the scanner and patient table 1) Standard carrying in Scanner : 1980mm wide×990mm deep×1870mm high, 1540kg (with cart) (Weight becomes 1490kg when the rear cover and cone cover are removed at site.) Patient table : 650mm wide×2120mm long×740mm high, 300kg (with cart having a width of 650mm) (3) The patient table can be positioned by using the attached templates. After lowering the scanner unit to the specified position, mount the templates according to the scanner position and scribe the patient table position. (See Fig.1.1.4) The patient table position shows an approximate reference position when using the templates. For precise positioning and anchor bolt fixing, observer [1.4 Leveling and fixing of patient X-ray center table]. Stand Template ① Scanner base plate 75(Notched size) Scanner base plate Scribe table anchor holes 4 positions Scribe L-holes 4 positions 75 (Notch) 500 Align the notch of the scanner base plate with the notch of the template ① ① Stand 130 ② 1055 1000 Fig. 1.1.4 Mounting of templates 1-3 L2E-BW1444 (4) Lower the scanner unit onto the floor. 1) After removing the rear cover once and then, opening the lift cover, loosen the caster adjusting screws alternately at four right, left, forward, and backward positions, and lower the scanner unit horizontally until its bottom cover touches the floor. 2) Remove two cart brackets (front),two cart brackets (rear), and connecting members 1,2,3 connecting the right and left stands from the scanner unit. Fig. 1.1.5 Installation of scanner unit 1-4 L2E-BW1444 (5) Remove two suspension jigs, two rotation stop jigs, and two binding jigs from the scanner unit. Fig. 1.1.6 Removal of transportation jigs (6) Remove the rotation stop jigs (lower A), (lower B), and tilt preventive jigs. The corn cover can be mounted as it is during this work. Fig. 1.1.7 Removal of transportation jigs 1-5 L2E-BW1444 1.2 Leveling of scanner unit Level the scanner unit according to the following procedure before fixing it to the floor with anchor bolts. (1) Check the level of the scanner gantry by using a level gauge on both front and rear connecting members provided between the right and left stands and also on both right and left stand bases as shown in Fig.1.2.1. (2) Level the scanner unit by adjusting the leveling screws of the scanner stand until the bubble of the level comes to the center. Place an adjusting plate below the stand bottom cover. Fig. 1.2.1 Leveling of scanner unit during installation (3) Fixing of scanner unit to the floor After leveling the scanner unit, fix the stand bases to the floor with anchor bolts at three places of the right and left stand bases respectively. Use the following accessory anchor bolts selectively depending upon the floor conditions. Article name Floor with mortar layer of 20mm or less in thickness ..................................... SANPICK ANCHOR (SB-1080) 80mm (By Sanyo Shoki) Floor with mortar layer of 20mm or more in thickness ..................................... SANPICK ANCHOR (SB-1010) 100mm 1) (By Sanyo Shoki) By using a vibration drill, make holes of 14.5mm in diameter in positions on the floor for fixing the stand bases. The hole depth should be 70mm for anchor bolts SB1080 or 90mm for SB1010. 2) Insert anchor bolts into these holes and drive them in lightly. (See Fig.1.2.2) 3) Tighten the bolts until dome washers are flattened and closely attached to the stand base surface. Fig. 1.2.2 Anchoring procedure 1-6 L2E-BW1444 1.3 Carrying in the patient table The patient table is packaged as mounted on a transport cart. After unpacking, carry in and mount it according to the following procedure. (1) Bring the patient table to the mounting position. (2) Mount the patient table on the floor face. 1) Lift the rear portion of patient table by a hydraulic truck. 2) Remove the cart mounted at the rear portion of the patient table from the table base. (a) Loosen the set screws of the cart. (b) Remove the fixing screws from L-bracket. (c) Pull the pin of L-bracket out of the table base and remove the L-bracket. (d) Draw out the pin on the opposite side of the cart from the table base. 3) Mount the installing jig to the table base. 4) Tighten the supporting bolts in installing jigs. 5) Put the receiving plate on the floor. 6) Lower the patient table in such a way as the supporting bolts meet the groove in the receiving plate to support the rear portion of the patient table. 7) Lift the front portion of the patient table by a hydraulic cart. 8) Support the front portion of the table according to steps (2) ~ (6). 9) Remove the hydraulic cart from the patient table. 10) After loosening four supporting bolts alternately to lower the patient table horizontally until the table base touches the floor. 11) After the table base has landed on the floor, loosen the supporting bolts furthermore and remove the receiving plates. 12) Remove the installing jig from the table base. (Caution) Tools from *1 to *3 are servicing tools. Return them to the works soon after the end of work. Supporting bolt Installing jig (*1) Hydraulic cart Pin Cart Cart Installing jig (*1) Supporting bolt (*2) Switchboard cover Receiving plate (*3) L-bracket Cart push screw Installing jig (*1) Supporting bolt (*2) Receiving plate (*3) Installing jig (*1) Fig. 1.3.1 Installation of table 1-7 L2E-BW1444 1.4 Leveling and fixing of patient table Fix the foot switch shown in 2.2.3 temporarily before starting this work, and make sure that the foot switch can be mounted without any trouble. If the table unit is too close to the scanner side, it is possible that the foot switch cover cannot be mounted later. (1) Lift the table to a height of 750mm (display height 0mm) above the floor and turn off the power supply. (2) Move the tabletop forward up to the extreme front position by turning the manual retreating handle for the tabletop mounted at the rear of the tabletop. (See Fig.1.4.1) Tabletop manual retreating handle Fig. 1.4.1 (3) Remove the rear base cover. (See Fig.1.4.5) 1) Remove screws A. 2) Remove the rear base cover. (4) Remove the power distribution board cover. (See Fig.1.4.5) 1) Remove screws B and then, remove the power distribution board cover. (5) Put the leveling jig on the rail, and check if the horizontal level of X-axis turning and Z-axis turning leveled properly by putting the level on the rail. (See Fig. 1.4.2) X-axis turning Level Z-axis turning Level Side cover Leveling jig Y X Leveling jig Z Fig. 1.4.2 (6) If the tabletop is not leveled properly, put adjusting liners below the base after lifting the base by turning the leveling screw, and then, level the patient table. (See Fig.1.4.6) Perform adjustment (7) ~ (12) in and after the next paragraph after scanogram has been set to be ready for radiography. 1-8 L2E-BW1444 (7) Adjust the verticality of the table to the scanner unit by using the scanogram. 1) Scan the tabletop in the scanogram mode. Set the scano mode to “P.A”. 2) Display a scanogram image of the table top as grid display. (See Fig.1.4.3) 3) Make sure that the scanogram image of the tabletop satisfies the following requirements (a), (b) with reference to the grid. (a) Measure the distances (A1-A2, B1-B2) between grid center line and the right and left edges of the tabletop and make sure that these distances are equal to each other. (b) Make sure that the right and left edges of the tabletop are parallel to the grid. 4) Adjust the lateral position of the table until the requirements (a) (b) are satisfied, and confirm it by the scanogram image again. Distance measuring icon Grid display icon Grid center line Make sure that A1-A2 = B1-B2 Displayed by the distance measuring icon. Fig. 1.4.3 1-9 L2E-BW1444 (8) Turn off the table power supply. (9) Fix the table body by driving in anchor bolts on the floor. (See Fig.1.4.4) anchor bolt Dome washer Base Fig. 1.4.4 (10) Mount the removed switchboard cover and rear base cover to the table. (11) Cover the installing jig screw holes on the side face of the base cover with an installing seal. (12) Turn on the table power supply. 1-10 L2E-BW1444 A (For the system installed with the MTCPU2 PCB, refer to Fig. 1.4.6.) Switchboard cover Screw B Bar Leveling screw Anchor Installing jig M10 washer Ratchet handle Base Anchor bolt Washer Adjusting plate Leveling screw M10 x 35L hexagon socket bolt Anchor tongue Base cover (rear) Base cover (side face) Seal (attachment) Screw A Arrow view A Fig. 1.4.5 1-11 L2E-BW1444 A Switchboard cover Screw B Bar Leveling screw Anchor Installing jig M10 washer Ratchet handle base Anchor bolt Washer Adjusting plate Leveling screw M10 x 35L hexagon socket bolt Anchor tongue Screw A Base cover (side face) a b Arrow view A Seal (attachment) Base cover (rear) Fig. 1.4.6 1-11-1 L2E-BW1444 1.5 Mounting of other units 1.5.1 Operation unit The operation unit is composed of the following units. (1) IPU (Image Processing Unit) (2) LCD monitor (3) Touch panel unit (4) Keyboard (5) Mouse (6) Interphone box (Note ; The desk is not included in the standard configuration) Monitor Touch panel unit Mouse Keyboard Image Processing Unit Interphone box 1-12 L2E-BW1444 X-ray control unit (1) Since cooling fans are provided at the back, place the X-ray control unit more than 15cm away from the wall. (2) Secure a space for wiring and maintenance work at the front and right side face. X-ray control unit 15cm or over (3) After placing the X-ray control unit in position, insert the attached holders to fix the casters. Work space 1.5.2 Work space Caster Caster holder 1-13 L2E-BW1444 1.6 Mounting of detector and detector side collimator 1.6.1 Mounting of detector Perform the following work by two persons for securing safety. Before mounting the detector, see the revision label ⑩. If the revision label shows “0”, you have to connect the 552CN connector before detector fixed. [ reference → appendix “How to plug in the Temperature Controller” ] An optimum directive position depends upon the detector. Mount the detector at an optimum directive position by using the designated direction adjusting pin out of five directive positions. The optimum directive position is indicated on the direction nameplate ⑧. Mount the detector, referring to Fig.1.6.1 shown below. 3 2 10 5 1 2 6 6 4 6 1 8 1 7 6 9 ① Grip bar ⑥ Plain washer ② Guide pin ⑦ Direction adjusting pin ③ Rotary plate ⑧ Direction nameplate ④ Detector ⑨ Shield cover (Front) ⑤ Shaft ⑩ Revision label Fig. 1.6.1 1-14A L2E-BW1444 (1) Fix grip bar ① to four detector fixing flanges respectively. (2) Mount two guide pins ② to rotary plate ③. (3) Push the detector ④ against rotary plate ③ by inserting the center hole of detector ④ onto the shaft ⑤, and fix the inner flange temporarily to the rotary plate ③ by using M8 x 35L hexagon socket bolts (A,B in the figure) with plain washer ⑥ each to such an extent as the detector ④ can rotate slightly. (4) Remove guide pins ② and grip bars ①. (5) Fix the outer left flange (C) temporarily to rotary plate ③ by using M8 x 35L hexagon socket bolts with plain washer ⑥ each to such an extent as the detector ④ can rotate slightly. (6) Screw in a direction adjusting pin ⑦ to rotary plate ③ through the hole of the outer right flange (D) (Select the direction adjusting pin ⑦ employed, referring to the direction nameplate ⑧). (7) Apply the inside of the hole of outer right flange (D) to the direction adjusting pin ⑦ by lightly turning the detector ④ clockwise or counterclockwise around the axis ⑤ as the center according to Table 1.6.1. (8) Fasten three M8 x 35L hexagon socket head bolts being fixed temporarily in step ③ and ⑤ with tightening torque of 2700N-cm (270kgf-cm), while keeping the angle so as not to allow the detector ④ to get out of shaft ⑤. (9) Remove the direction adjusting pin ⑦ from rotary plate ③, and fix the outer right flange (D) with tightening torque of 2700N-cm (270kg-cm) by using M8 x 35L hexagon socket head bolts with plain washer ⑥ each. Table 1.6.1 Setting positions Direction adjusting (Indicated on direction pins used for nameplate) positioning CENTER φ20 φ18LD φ18 φ18RD φ18 φ16LD φ16 φ16RD φ16 Detector fixing methods Fix the detector at the center as viewed from the front Fix the detector by lowering its left side as viewed from the front Fix the detector by lowering its right side as viewed from the front Fix the detector by lowering its left side as viewed from the front Fix the detector by lowering its right side as viewed from the front 1-15 L2E-BW1444 After mounting the detector, connect 552CN of the detector to the temperature controller and slice selection cable (SD-1CN ~ WA-7CN), referring to Fig.1.6.2. Temperature controller Slice selection cable SD-1CN 552CN Detector WA-7CN Preamplifire unit Preamplifire unit Fig. 1.6.2 It is necessary for stabilizing the performance of the detector (temperature controller) and preamplifier to apply power to these units for about 4 hours before using them. Continue feeding power to these units normally. For the connection between the detector and the preamplifier, refer to 2.3.2. 1-16 L2E-BW1444 Appendix “How to plug in the Temperature Controller (connect 552CN)” 1. Remove COVER and temperature controller (lower side only). COVER Temperature Controller (lower side) Fig. 1.6.3 How to remove COVER → remove four screws (fig.1.6.3 dot circle), then take the COVER off. Temperature Controller → unlock the hook (fig.1.6.4 dot circle), then pull the controller body out. 1-16-1 L2E-BW1444 Fig. 1.6.4 2. Connect the 552CN with the detector hanging. Fig. 1.6.5 1-16-2 L2E-BW1444 After the detector mount on the guide pin, keep the detector hanging. (don’t push to the rotary plate) One person grasps the grip rod so as not to drop it. (fig.1.6.5) Another person plugs in the 552CN, and tightens two screws. (fig.1.6.6) Fig. 1.6.6 3. Fix the detector . ( Refer to 1.6.1 Mounting of detector ) 4. Put the Temperature Controller and the COVER back. 1-16-3 L2E-BW1444 (10) Mount the shield cover (front) ⑨ onto the detector with M4×10L pan-head screws. (11) Using the shield metal and plate, clip the sheet copper of the shield cover (front) ⑨ between them, and fix it with M4×10L pan-head screws. (The plate is attached to the shield cover.) (Fig. 1.6.7, 1.6.8) shield cover (Front) ⑨ plate sheet copper plate shield cover Fig. 1.6.7 Fig. 1.6.8 mounting completed 1-16-4 L2E-BW1444 1.6.2 Mounting of detector side collimator (1) Mounting of drive unit Apply the end face of the left mechanical section base to the peak of the spring pin, and fix the left mechanical section base by three M8×20L hexagon socket head bolts. Fix the right mechanical section base in the same way. (See Fig.1.6.9 and 1.6.10) (2) Mounting of collimator plate Apply the end face of the left mounting bracket to the peak of the spring pin, and fix the left mounting bracket by two M8×20L hexagon socket head bolts. Make sure that the mechanical section is engaged with the gear of the collimator plate. Fix the right mounting bracket in the same way. (See Fig.1.6.9 and 1.6.10) Mechanical section base (right) Spring pin M8 x 20L hexagon socket head bolt Mounting bracket (right) Spring pin Collimator plate Mounting bracket (left) Mechanical section base (left) Fig. 1.6.9 1-17A L2E-BW1444 (3) Connection of connector Connect the cable connector 230CN being wound to the lower part of the transformer mounting bracket to connector 230CN of the right mechanical section. (See Fig.1.6.10) Wiring differs at the right and left terminals of the collimator mechanical section between 50Hz districts and 60Hz districts. Check the wiring without fail before turning on the power supply. (See Fig. 1.6.11). Transformer bracket Mounting bracket (right) Mechanical section base (left) Spring pin Mounting bracket (left) Dial Spring pin Fig 1.6.10 Mechanical section base (right) 230CN Right terminal (Same on the opposite side, too) 50Hz district 60Hz district A cable for shorting No. 5 and 8 is connected. Fig.1.6.11 1-18A A cable for shorting No.5 and 8 is not connected. L2E-BW1444 (4) Connect cables. (See Fig. 1.6.12) Make sure that cables are not caught tight nor stretched when the collimator operates. Collimate plate of detector side collimator Detector side collimator cable Drive unit of detector side collimator 230CN Detector Fig.1.6.12 1-19A L2E-BW1444 (5) Adjustment of detector side collimator (coarse adjustment) 1) Turn off the scanner power supply. 2) Mount the adjusting jig to the collimator plate by M4×12L pan-head screws (with springs and plain washers). 3) Press the jig tip to the rotary plate face by pushing the collimator plate slowly by hand. (See Fig.1.6.13) Position the stopper to enable the stopper to touch the L bracket under this condition. (See Fig.1.6.14) Adjust both right and left stoppers. Don’t push the center of the collimator plate, but slowly push the right and left ribs at the end of the collimator plate by hand, otherwise the slit accuracy deteriorate. Stopper (right) L bracket Stopper mounting screw Rib (Move the collimator plate by pushing here) Stopper (left) Adjusting jig Apply the stopper to L bracket Collimator plate Detector Don’t push this portion. Adjusting jig M4×12Lpan-head screw (with spring and plain washer) Rib Adjusting jig Move the collimator plate forward and lightly apply the adjusting jig tip face against the rotary plate. The stopper is positioned correctly when the stopper and L bracket touch under this condition. Mount the jig to the collimator plate in such a way as both side faces of the collimator plate are lightly caught by the jig. Fig. 1.6.13 1-20A L2E-BW1444 M4 x 40L hexagon socket head bolt (*2) M4 nut (*3) L bracket 2- M4 x 12L hexagon socket head bolt (*1) (spring, plain washer) Stopper After loosening the bolts and nuts marked (*1) (*2) (*3), adjust the stopper position. Then, fasten the bolts (*1). Screw in the bolt (*2) until the bearing surface of bolt (*2) exactly touches “L bracket”, and lock the nut (*3). Perform this work on both right and left sides. Fig. 1.6.14 Coarse adjusting method of stopper position 1-21A L2E-BW1444 1.7 Mounting of X-ray tube unit and cooler In Presto, two of X-ray tube assemblies (3.5MHU and 5MHU) can be mounted. Since the X-ray tube assembly and cooler are connected to each other with hoses, etc., be careful not to scratch cooler hoses or cables. 1.7.1 Mounting procedure of X-ray tube assembly (Presto 3.5MHU) The following mounting procedure applies to the X-ray tube assembly having the tube shift mechanism. For details, refer to Fig.1.7.1. (1) Turn the rotary plate until collimator case ① comes up to the highest position. (2) Screw the holder ③ to the rotary plate. (Bring the key way of holder ③ into contact with the key of rotary plate.) (3) Fit insulation ring ⑩ to X-ray tube assembly ⑤, and fix it with Mylar tape. (4) Mount X-ray tube assembly ⑤ to holder ③. (5) Mount trunnion ⑥ to holder ③ by using four M8×50 long hexagon socket head bolts. (Tightening torque 1960N-cm (200kgf-cm).) (6) Connect the anode and cathode high-voltage cables to the X-ray tube assembly. (7) Connect the starter cable from the X-ray tube assembly to the terminal ⑦ (7, 8, 9, CS and EX) and the cable to terminal ⑧ (ED, E, LO, L100, H, and HO). (8) For wiring and piping, refer to [2.3 Wiring inside the system], too. (9) Mount pedestal ⑯ and gauge mount ⑰. (10) Mount the retaining band ⑱ to holder ③ by M8×16L hexagon socket head bolt. (11) Mount guard bracket ⑲ to the rotary plate by M8×20L hexagon socket head bolts. (1) Be careful not to forget to use the insulation Mylar sheets (insulation ring) when mounting the X-ray tube assembly. If they are not applied, secondary electronic parts may be damaged due to discharge. Since holders are used to fix the X-ray tube assembly during delivery, check if the X-ray tube is correctly insulated by using a circuit tester. (2) At the delivery time from the works, ⑯ and ⑰ are removed. Mount these parts securely. (3) Parts ⑯ and ⑰ must be mounted before centering in chapter 3. 1-22 L2E-BW1444 A : In case of 3.5 MHU X-ray tube assembly ⑱ 4-M8X20L Hexagon socket head bolt (with spring and plain washer) ⑥ 2-M8X20L Hexagon socket head bolt (with spring and plain washer) 3-M4X10L Hexagon socket head bolt (with plain washer) a ⑤ ⑲ 21 ⑩ + - 3-M4X10L Hexagon socket head bolt (with plain washer) H9 ③ CS H7 H8 ⑭ Power cord (L0,L110,H,H0,E) a ⑫ ⑬ ⑰ ⑨ ⑯ ① 7-M6 nut, spring, and plain washer H,H0,L0,L100,E,ED ⑧ ⑪ ⑦ 7,8,9,CS,EX ④ ① ③ ④ ⑤ ⑥ ⑦ ⑧ Collimator case Holder (base plate) Mount plate X-ray tube assembly Trunnion Terminal Terminal 1.7.2 ⑨ ⑩ ⑪ ⑫ ⑬ ⑭ ⑯ Cooler ⑰ Gauge mount Insulating ring ⑱ Push band Oil container ⑲ Guard bracket 1 Mounting bracket 1 Guard plate Mounting bracket 2 Hexagon socket hesd bolt M16×14L with spring and plain washer Pedestal Fig. 1.7.1 Mounting procedure of cooler (1) Mounting bracket 1 ⑫ and mounting bracket 2 ⑬ of cooler ⑨ by hexagon bolts ⑭. (2) Put the cooler ⑨ on the mount plate ④, and mounting bracket 1 ⑫ and mounting bracket 2 ⑬ with M6 nuts, spring washers, and plain washers. (3) Connect the power cord from the cooler to terminal ⑧. (4) Mount guard plate 1. to guard brackets ⑲ ⑳ by M4×10L hexagon socket head bolts. Mount guard plate sure that after fixing the oil hose and HV cables using tie lap. Make does not touch any hose or cables. If not, cleat the hose and cables again so as not to allow them to touch 2. . The bending radius of the oil hose is R80mm or over (on the hose center line) as a reference. Cleat the hose so as not to crush it flat. 1-23 L2E-BW1444 1.7.3 Mounting procedure of X-ray tube assembly (Presto 5 MHU) The following mounting procedure applies to the X-ray tube assembly having the tube shift mechanism. For details, refer to Fig.1.7.2. (1) Turn the rotary plate until collimator case ① comes up to the highest position. (2) Screw the holder ③ to the rotary plate. (Bring the key way of holder ③ into contact with the key of rotary plate.) (3) Fit insulation ring ⑩ to X-ray tube assembly ⑤, and fix it with Mylar tape. (4) Mount X-ray tube assembly ⑤ to holder ③. (5) Mount trunnion ⑥ to holder ③ by using four M8×50 long hexagon socket head bolts. (Tightening torque 1960N-cm (200kgf-cm).) (6) Connect the anode and cathode high-voltage cables to the X-ray tube assembly. (7) Connect the starter cable from the X-ray tube assembly to the terminal ⑦ (7, 8, 9, CS and EX) and the cable to terminal ⑧ (ED, E, LO, L100, H, and HO). (8) For wiring and piping, refer to [2.3 Wiring inside the system], too. (9) Mount pedestal ⑯ and gauge mount ⑰. (10) Mount the retaining band ⑱ to holder ③ by M8×16L hexagon socket head bolt. (11) Mount guard bracket ⑲ to the rotary plate by M8×20L hexagon scoket head bolts. (1) Be careful not to forget to use the insulation Mylar sheets (insulation ring) when mounting the X-ray tube assembly. If they are not applied, secondary electronic parts may be damaged due to discharge. Since holders are used to fix the X-ray tube assembly during delivery, check if the X-ray tube is correctly insulated by using a circuit tester. (2) At the delivery time from the works, ⑯ and ⑰ are removed. Mount these parts securely. (3) Parts ⑯ and ⑰ must be mounted before centering in chapter 3. 1-24 L2E-BW1444 B : In case of 5MHU X-ray tube assembly ⑱ 4-M8X20L Hexagon socket head bolt (with spring and plain washer) ⑥ 2-M8X20L Hexagon socket head bolt (with spring and plain washer) 3-M4X10L Hexagon socket head bolt (with plain washer) a ⑤ ⑲ 21 ⑩ + - H9 ③ 3-M4X10L Hexagon socket head bolt (with plain washer) CS H7 H8 ⑭ Power cord (L0,L110,H,H0,E) a ⑫ ⑬ ⑰ ⑨ ⑯ ① 7-M6 nut, spring, and plain washer H,H0,L0,L100,E,ED ⑧ ⑪ ⑦ 7,8,9,CS,EX ④ ① ③ ④ ⑤ ⑥ ⑦ ⑧ Collimator case Holder (base plate) Mount plate X-ray tube assembly Trunnion Terminal Terminal 1.7.4 ⑨ ⑩ ⑪ ⑫ ⑬ ⑭ ⑯ Cooler ⑰ Gauge mount Oil container ⑱ Push band Insulating ring ⑲ Guard bracket 1 Plate (1) Guard plate Plate (2) Hexagon socket hesd bolt M16×20L with spring and plain washer Pedestal Fig. 1.7.2 Mounting procedure of cooler (1) Mount the mounting brackets 1 and 2 to the upper face of cooler ⑨ by hexagon bolts ⑭. (2) Put the cooler ⑨ on the mount plate ④, and fix plate A ⑫ and plate B ⑬ with M6 nuts, spring washers, and plain washers. (3) Connect the power cord from the cooler to terminal ⑧. (4) Mount guard plate to guard brackets ⑲ ⑫ by M4×10L hexagon socket head bolts (with springs and plain washers). 1. 2. after fixing the oil hose and HV cables using tie lap Make sure Mount guard plate that does not touch any hose or cables. If not, cleat the hose and cables again so as not to allow them to touch . The bending radius of the oil hose is R80mm or over (on the hose center line) as a reference. Cleat the hose so as not to crush it flat. 1-25 L2E-BW1444 1.8 Removal of protective materials for transport Remove the vinyl bag covering the electromagnetic brake of the scan drive unit of scanner. 1-26 L2E-BW1444 1.9 Internal inspection of patient table Make sure that no oil leaks from the hydraulic system and check the oil level in the oil tank according to the procedure at the end of wiring of patient table. Observe the following procedure before moving the table. (1) Remove the link supporting bracket. 1) Remove screws. 2) After lifting the table, remove fixtures A and B. Return the link supporting bracket (fixtures A, B) to the works. (2) Set the patient table to 900mm above the floor. (3) Make sure that no oil leaks from the hydraulic system. (4) Make sure that the oil level of the oil tank gauge is 60mm ±10mm. (5) Set the patient table to 700mm above the floor and turn off the power supply. (6) Lift the upper edge pipe of the bellows cover and fit it to the bellows holders. (Three bellows holders are mounted below the right and left side covers, respectively.) 1-27 L2E-BW1444 Fig. 1.9.1 (7) Fix the front and rear parts of the upper edge pipe of the bellows cover with the bellows fixtures (Attachments of patient table) at 2 positions respectively. 1-28 L2E-BW1444 1.10 Mounting of covers The scanner covers consist of those illustrated in Fig.1.10.1. When the system is shipped in the standard style of packaging, the scanner unit is delivered with the top cover, both side covers, stand covers (right, left, front, rear), opening covers (right, left, front, and rear), and front lower cover detached. Top cover Side cover Rear cover Opening cover (rear) Cone cover Lift cover Front lower cover Stand cover Opening cover (front) Side cover Stand cover (rear) (Front stand cover on the opposite side) Fig. 1.10.1 Scanner covers After completion of wiring and adjustment of the scanner, mount the covers to the scanner referring to Fig. 1.10.2 and Fig. 1.10.3. (1) Remove four thumb screws and open the lift cover. (2) Mount the top cover receiver with four sems screws (M4×10L) and then, mount the top cover with four M4 truss head screws and toothed washers. Connect the connector 250CN of the top cooling fan. Connect the grounding wire. (3) Insert two hooks mounted at the upper part of the side cover into the U-receiving recesses on either end of the top cover. Fix the side covers to the support base of the scanner gantry with four M4×8L truss screws and toothed washers. Cover the screws with attached screw caps. (4) Mount the front opening cover to the lift cover. 1-29 L2E-BW1444 Lift cover Hinge M6 Nut (Spring and plain washers) NOTE 1) Hinge Large brass spacer Large washer M8 nut (Spring and plain washer) Lift cover NOTE 2) Lift cover cable Gas spring NOTE: 1) When mounting the lift cover, firstly put the large brass spacer into the opening center shaft and then put the hinge into the shaft. After that, put the large washer into the shaft and mount the M8 nut (with spring and plain washers). Work for both left and right sides. 2) Put the lift cover cable over the opening center shaft as shown above. Wire the cable so as not to be caught by the ceiling cover when closing the lift cover. Wire the cable so as not to apply any excessive load to it when opening/closing the lift cover,. Fig. 1.10.1A Mounting Lift Cover 1-29-1 L2E-BW1444 Top cover Side cover (left) Lift cover Side cover (right) Front lower cover Opening cover (front) Screw cap Screw cap Supporting base Bottom cover Fig. 1.10.2 Mounting of scanner front covers (5) Mount the stand covers. Fig. 1.10.3 Mounting of stand covers 1-30 L2E-BW1444 (6) Check the cone cover position after mounting the cone cover to meet the scribed line of the gantry side mounting bracket in such a way as its center is aligned with the lift cover center. Connect the traffic light, microphone and mat switch connectors 32CN, 312CN, and 88CN. Connect mat switch connector 88CN without fail before starting the scan, tilt, and table operation. They are not operated unless 88CN connector is connected. (7) Mount the front rod and cable holder assembly to the stand. The length of front and rear rods shall be as shown in Fig. 1.10.4. After mounting the covers, tilt the system (±25°)and make sure that no excessive load is applied to the hinge of the opening cover. If an excessive load is applied, adjust the length of the front and rear rods so as not to apply any excessive load. (8) Connect cable 163CN of the rear cover X-ray exposure indicator lamp. (9) Close the rear cover and fix it with truss head screws (M4×8L) at 5 places. Cover these screws with attached screw caps. (Mount the attached rear cover cart to the rear cover in advance. After mounting the rear cover, remove the rear cover cart and store it.) (10) Fix the side face of the rear cover to the side cover by using four sems screws M14×12L, and cover these screws with the attached screw caps. (11) Close the lifting cover and fix it with thumb screws. (12) Mount the front lower cover to the lift cover. (13) Fix the rear opening cover with truss screws (M4×8L). Fig. 1.10.4 Mounting of scanner rear cover 1-31 L2E-BW1444 (14) Mount the X-ray transmission window to the lift cover and cone cover. 1) Label the convex side of the velcro fastening to the concave side of the magic tape of the cone cover and lift cover. (Be careful with the laser projector window position when labeling the tape.) (1) Mount the X-ray transmission window closely so as not to produce any clearance from the cover. (2) Overlap both ends of the X-ray transmission window with each other as illustrated below so as not to allow any liquid like a vomited substance to enter the scanner interior even if it flows down from the upper part. 2) Insert the collar on the belt into the long hole of the cone cover (lift cover). 3) Fix one end of the belt by M3×6L flat countersunk head screws. 4) Butt both ends of the belt to each other, and fix the end on the unscrewed side by M3×6L flat countersunk head screw, and push the belt to closely attach to the cone cover (lift cover). 5) Cover the M3 flat counter-sunk head screw portions fixing the collar on the belt with the attached white seal. X-ray transmission window Magic tape of cover (concave part) Belt A Belt B Table side Table side Long hole Cone cover Magic tape 4-M3×6Lsems screw Seal (white) Lift cover Detail of part B Seal (white) M3 flat countersunk head screw Belt X-ray transmission window X-ray transmission window Arrow view A Butt each other Collar (Fit in arrow direction) 7×20 long hole (2 positions on lift cover (cone cover) ) Insert 1-32 46 L2E-BW1444 Section 2 WIRING 2.1 Wiring of cables between units Refer to the wiring diagram and cable list on pages 2-2, 2-3, 2-4 and 2-5. Cautions on wiring ・ Cables are classified roughly into four kinds; high voltage (HV) cables, power (POW) cables, signal (SIG) cables, and grounding (GND) cables. ・ For preventing malfunction of the CT system, isolate the cables from each other by using wiring pits as shown below. Ground each pit without fail. If they cannot be isolated from each other by using pits, isolate them by using a shielded zipper tube Ground the seal of the zipper tube without fail. HV POW SIG GND 50 100 100 150 An example of wiring pits SIG GND 100 POW 100 HV 50 Unit mm 150 2-1 L2E-BW1444 2-2 L2E-BW1444 Fig.2.1.1 External cable connection diagram (without step-down transformer) OPTICAL 2-3 L2E-BW1444 Fig.2.1.2 External cable connection diagram (with step-down transformer) External Cables CABLE INTERNAL OUTER LENGTH IN CONNECTOR SIZE CABLE NO. OF CONDUCTORS STD DESTINATION MAX CABLE TYP (H×W×D) RUN CASE OF STANDARD USE O. Dia AND WIRE THICKNESS LENGTH LENGTH No. LENGTH POW. SIG. GND FROM TO FROM TO FROM TO DRAWING No. 003 HP-PB WS 15 12 0 1.2 10.8 3×20 UL1683 6AWG * KA340700-E 031 HP-PB WS 15 12 0 1.1 10.9 3×20 UL1683 6AWG * KA340700-F 061 HP-PB OC 15 10 0 0.3 9.7 3×20 UL1683 6AWG * KA340700-H 100 HP-PB WB 15 10 0 1.4 8.6 41.2 105 WB WS 15 15 0.4 1.2 13.4 35.0 106 WB WS 15 12 1.7 0.7 9.6 18.6 107 WB WL 15 10 1.3 0.0 8.7 6.6 131 WB WS 15 12 1.5 0.7 9.8 161 WB OC 15 10 1.6 0.3 302 WB WS 15 12 1.5 0.7 UL. CSA-ST-SB * KB315896-A * KB315894-A UL-ST-SB 3×10AWG * KB327035-C UL2464SB 2×20AWG * KA318151-E 18.6 UL-ST-SB 3×10AWG * KB327035-D 8.1 13.5 ST-SB 2×14AWG * KA338123-E 9.8 11.5 UL2464SB 17P×28AWG 4C×2AWG UL. CSA-ST-SB 2C×2AWG 20 × 20×61.7 61.7 * ×51.5 KB211638-C ×51.5 303 WB DS 15 10 1.4 0 8.6 6.6 UL2464SB 2×20AWG KA318151-G * 20 331 WS OC 15 12 0.7 0.5 10.8 11.5 UL2464SB 25P×28AWG × 20×76.8 76.8 * ×51.5 KB211639-F SC KB322808-A ×51.5 333 WS OC 22 1.5 0.5 20.0 7.0 Optical FIBER SC * Internal Cables (Scanner) INTERNAL OUTER LENGTH IN CONNECTOR SIZE CABLE NO. OF CONDUCTORS CABLE DESTINATION MAX STD CABLE TYP (H×W×D) USE CASE OF STANDARD O. Dia AND WIRE THICKNESS No. LENGTH LENGTH LENGTH POW. SIG. GND FROM TO FROM TO FROM TO *1 WS WT 1.0 1.0 - - - 3×20 *2 WS WT 3.1 3.1 - - - 13.5 UL1683 6AWG ST-SB 2×14AWG DRAWING No. KB325041-A * KB325527-A * *3 WS WT 4.7 4.7 - - - 11.5 UL2464SB 25P×28AWG * *4 WS SP 6.0 6.0 - - - 5.6 UL2464SB 2×22AWG * 7.5×68 20×76.8 ×15.6 ×51.5 10.5 × KB219492-B KB325031-A 7.4 ×11.7 Table 2.1.1 Cable list (without stepdown transformer) 2-4 L2E-BW1444 External Cables CABLE INTERNAL OUTER LENGTH IN CONNECTOR SIZE CABLE NO. OF CONDUCTORS STD DESTINATION MAX CABLE TYP (H×W×D) RUN CASE OF STANDARD USE O. Dia AND WIRE THICKNESS LENGTH LENGTH No. LENGTH POW. SIG. GND FROM TO FROM TO FROM TO DRAWING No. 003 TR WS 15 12 0 1.2 10.8 3×20 UL1683 6AWG * KA340700-E 031 TR WS 15 12 0 1.1 10.9 3×20 UL1683 6AWG * KA340700-F 061 TR OC 15 10 0 0.3 9.7 3×20 UL1683 6AWG * KA340700-H 100 TR WB 15 10 0.5 1.4 8.1 41.2 105 WB WS 15 15 0.4 1.2 13.4 35.0 106 WB WS 15 12 1.7 0.7 9.6 18.6 UL-ST-SB 3×10AWG * KB327035-C 107 WB WL 15 10 1.3 0.0 8.7 6.6 UL2464SB 2×20AWG * KA318151-E 131 WB WS 15 12 1.5 0.7 9.8 18.6 UL-ST-SB 3×10AWG * KB327035-D 161 WB OC 15 10 1.6 0.3 8.1 13.5 ST-SB 2×14AWG * KA338123-E 110 HP-PB TR 5 3 0 0.5 2.5 41.2 * KB315896-D 302 WB WS 15 12 1.5 0.7 9.8 11.5 UL2464SB 17P×28AWG * 303 WB DS 15 10 1.4 0 8.6 6.6 UL2464SB 2×20AWG * 331 WS OC 333 WS OC 15 UL. CSA-ST-SB 4C×2AWG UL. CSA-ST-SB 2C×2AWG UL. CSA-ST-SB 4C×2AWG * KB315896-C * KB315894-A 12 0.7 0.5 10.8 11.5 UL2464SB 25P×28AWG * 22 1.5 0.5 20.0 7.0 Optical FIBER * 20×61.7 20×61.7 ×51.5 ×51.5 KB211638-C KA318151-G 20×76.8 20×76.8 ×51.5 ×51.5 SC SC KB211639-F KB322808-A Internal Cables (Scanner) OUTER LENGTH IN INTERNAL CASE OF STANDARD CABLE NO. OF CONDUCTORS CABLE DESTINATION MAX STD CABLE TYP CONNECTOR SIZE USE (H×W×D) O. Dia AND WIRE THICKNESS LENGTH No. LENGTH LENGTH POW. SIG. GND FROM TO FROM TO FROM TO *1 WS WT 1.0 1.0 - - - 3×20 *2 WS WT 3.1 3.1 - - - 13.5 UL1683 6AWG ST-SB 2×14AWG DRAWING No. KB325041-A * KB325527-A * *3 WS WT 4.7 4.7 - - - 11.5 UL2464SB 25P×28AWG * *4 WS SP 6.0 6.0 - - - 5.6 UL2464SB 2×22AWG * 7.5×68 20×76.8 ×15.6 ×51.5 10.5 × KB219492-B KB325031-A 7.4 ×11.7 Table 2.1.2 Cable list (with stepdown transformer) 2-5 L2E-BW1444 2.2 Wiring of cables in each unit 2.2.1 Scanner All the cables for connection to external units should be laid via the power distribution board (joint base unit) at the lower rear part of the scanner. For the block diagram of wiring routes, refer to Page 2-2. For the wiring routes in detail, refer to Page 2-3. Connect each cable after removing the joint base unit cover. Fig. 2.2.1 Fig. 2.2.2 2-6 L2E-BW1444 Wiring of RUN No. 333 cable inside the scanner (For the system installed with the MTCPU1 PCB) TXR5 PCB SWBD PCB TXR5 PCB SBPCB UNIT 333 Be careful with thin protective sheath-free cables so as not to allow them to come out to the rotation panel side during wiring. Connect to the rear Side of the stand clamp Left side of the scanner Caution: The bending radius of RUN No.333 cable (optical fiber cable) shall be more than 100mm. The optical cable is connected through the rear of the stand. Fix the optical cable with tie laps with due care so as not to cause any cable contact failure during tilt operation. 2-7 L2E-BW1444 Connection of the RUN No. 333 cable inside the scanner (installed with the MTCPU2 PCB) SBPCB SWB Dプリ 板 unit SBPCB S B P unit CB ユニッ ト 333 Route the cable スタンド の behind the stand 裏側に配線する MT CPPCB Uプリ 板 MTCPU ク ランプする。 Clamp the cable スキャナ側面左側 Left side of the scanner The smaller-gauge wire without the protective sheath should not be fixed with Tyrap. Because the RUN No. 333 cable is optical fiber cable, the curvature R of the part of the cable with the protective sheath (larger-gauge wire) should be 100mm or more and R of the part without protective sheath (smaller-gauge wire) should be 12mm or more. When routing the cable, as it is routed through behind the stand, care must be taken in fixing the cable with Tyrap not to allow the cable to touch with the tilting scanner unit. 2-7-1 L2E-BW1444 2.2.2 Table wiring All table cables are internally connected from the scanner. (These cables are enclosed in the scanner and table without any external protrusion.) Table Base *3(SIG.) *1(GND.) *2(POW.) WT-2CN WT-1TN Cable cleat 1TN R S Terminal mounting plate Fig. 2.2.3 Table wiring 1. Be careful not to apply any excessive force to the connectors or terminals when holding the cables, since the cable cleat is mounted to be close to the connectors and terminals. 2. Connect the Power cable to 1TN through the ferrite core of the foot switch base. 2-8 L2E-BW1444 2.2.3 Mounting of foot switch, table electric unit cover, and table control unit (1) Mount the foot switch base on the power distribution board at the front of the table. (2) Fix the foot switch base by driving grip anchors (M6) on the floor. 1) Prepare each hole of φ10.5×30 deep corresponding to the anchor hole on the floor. 2) Drive the grip anchor (M6) into theφ10.5 hole. 3) Fix the foot switch base to the floor by M6×30L hexagon socket head bolts with washers. Scanner side Power distribution board cover M6x30L hexagon socket head bolt Washer Grip anchor(M6) Anchor 8mm Stand base Be careful not to allow the foot switch base to touch the stand base or stand cover. The clearances between the foot switch base and the stand base or stand cover are 8mm and 2mm respectively. (3) Connect cables to the power distribution board. (4) Connect he foot switch cable connector. (5) Fix the cables to the foot switch base suitably. Arrange the cables regularly on the stand base at the lower part of the gantry without overlapping or coiling them as shown in the following figure. (All cables are internally connected from the scanner.) 2-9 L2E-BW1444 M4x10L screw with spring and plain washer WS-5CN 121G Signal cable(WT-2CN) Foot switch base Table side Power distribution board 120G Power cable(WT-1TN) (Connect through the ferrite core) Ferrite core Earth cable(WT-E) Part D Details of part D (6) Mount the power distribution board cover to the base cover. Foot switch cover (7) Cover the screw hole on the side face of the table base cover with the seal (attachment of table). Seal 2-10 L2E-BW1444 (8) Mount the table control unit to the rear of the table by using the screws mounted in advance. Sems screw (9) Connect cables to the table control unit and table unit. ① PCB140 PCB 16CN ② 113CN ③ 121CN ④ 128CN ⑤ 129CN ⑥ PCB140 PCB 13CN ⑦ PCB140 PCB 14CN ⑧ PCB140 PCB 17CN 2-11 L2E-BW1444 (10) Mount the rear base cover. Base cover (rear) Base cover (side) Screw 2-12 L2E-BW1444 2.2.4 Image processing unit (1) Method of removing the image processing unit covers. 1) Remove the rear cover by removing four screws. Rear cover 2) Remove the top cover by removing two screws. Top cover 2-13 L2E-BW1444 (2) Cabling work procedure 1) Connect the external cables from the scanner unit to the image processing unit. Connect the external cables from the rear panel of the scanner unit to connectors 56CN and 79CN. Cleat these cables as shown by arrow view A. 2-14 L2E-BW1444 Arrow view A Cleat the optical cable (RUN#333) with the power cable and tie lap so as not to allow its bending radius to be less than 70mm. Perform cleating from the Image Processing unit to the duct at intervals of about 50cm. FR-MB4 PC board 79CN 56CN Fix with cleat Fix cables so that their shields are securely attached to the fixtures 56CN Fix with tie lap 79CN RUN# RUN# 333 331 To gantry unit 2-15 L2E-BW1444 (2) As shown in Diagrams (A) and (B), connect the cable to the connector on top of the image processing unit. Following the procedure below, insert the cable into the gap on the rear of the unit to connect it as illustrated in the diagram below (the diagram shows the top of the unit): Figure (A) Front 28CN 6FC 21CN 79CN 9CN 4FC 2FC FG4 shielding part Rear 82CN 83CN Front 前 Figure (B) 5FC 28CN 22CN 21CN (33CN) 6FC 4FC 2FC 79CN 9CN 40CN 83CN FG4 shielding part シールド処理部 AI B H CJ G F 後 Rear D ① Connect the cable ‘A’ (intercom box – 9CN) to 9CN of the unit. At this time, be sure to fix the shielding part with a shielding metal so that it is firmly joined (see diagram (C) or (E) on page 2-17-1). ② Pass cable ‘C’ (touch panel – 22CN) through the core (5FC), and connect it to 22CN (diagram (B) only). 2-16 L2E-BW1444 ③ Pass cable ‘D’ (monitor – 21CN) through the cores (2FC, 4FC, and 6FC), and connect it to 21CN of the unit. When the LCD monitor is connected, connect cable ‘D’ to 33 CN (diagram (B) only). (4) Pass cable ‘F’ (keyboard – 83CN) through the cores (2FC, 4FC, and 6FC), and connect it to 83CN of the unit. (5) Pass cable ‘G’ (mouse – 40CN or 82CN) through the cores (2FC, 4FC, and 6FC), and connect it to 40CN or 82CN of the unit. (6) Fix cable ’B’ (shielded wire from the intercom box) to FG4 with a screw. (7) Connect cable ‘H’ (optical cable from the gantry unit) to 79CN of the unit. (8) Fix cable ‘I’ (shielded wire from the touch panel unit) to FG4 with a screw. (diagram (B) only) (9) Connect cable ‘J’ (ETHER (10/100Base T)) to 28CN of the unit. (* Only when DICOM or Hyper Q-Net is used) (3) Cleat the cables wired in (2) above as shown in diagrams (C) or (D), and (E). (1) Make sure that each pc board in the PCB unit is firmly inserted in place. (2) Do not bend cable ‘H’ (optical cable) at a bending radius below 70mm. (3) If cable ‘A’ has no shielding part, cleat it as shown in diagram (D). 2-17 L2E-BW1444 Figure (C) Front Fix F.G cables by using a wiring band Shield bracket (Mount to the shield processing part of cable A) Rear A IB H Monitor power cable (Take it out to the rear face of IPU from a clearance of rubber in the same way as in other cables) CJ G F D Fix cables B ~ J connected to IPU collectively by using the cable clamp together with the monitor power cable. Figure (D) Front Fix F.G cables by using a wiring band Rear Monitor power cable AI (Take it out to the rear face of IPU from a clearance of rubber in the same way as in other cables) B H CJ G F D 2-17-1 Fix cables A ~ J connected to IPU collectively by using the cable clamp together with the monitor power cable. L2E-BW1444 Figure (B) Front Fix F and G with a wiring band. Shield plate (Mount it on the shielding part of cable ‘A’.) Rear Bundle the cables ‘B’ – ‘J’, monitor power cable, touch panel signal cable, and touch panel power cable together, and fix them while holding them down with a metal. Monitor power cable (Like other cables, take it out of the rubber gap towards the rear of the unit.) Touch panel signal cable Touch panel power cable 2-17-2 L2E-BW1444 (4) Magnetic disk units mounting procedure. 1) Remove the cover of the image processing unit. Front cover 2) Connect the cables to the magnetic disks. Mount the signal cable to face the label face of the connector upward (in the same way as in the power cable). 3) Mount the magnetic disks. Magnetic disk units 2-18 L2E-BW1444 (5) Connect cables to the monitor. Connect the monitor power cable and monitor signal cable to the specified connectors on the rear panel of the monitor as illustrated below. (The following figure shows the rear panel of the monitor.) Connect the power cable only to the monitor whose signal cable directly comes out of the rear panel of the monitor. Monitor power cable Image processing unit Monitor signal cable 2-19 L2E-BW1444 (6) Remove the cable cover on the rear of the touch panel. Touch panel (rear view) Hook and loop fastener Cable cover (7) Connect cable to the touch panel (only in the case of diagram (A)). As shown in the diagram below, connect touch panel signal cable and touch panel power cable to the appropriate connectors located on the bottom of the touch panel, respectively. Rear view of the membrane panel Touch panel power cable Image processing unit Touch panel signal cable 2-20-1 L2E-BW1444 (8) Replace the cable cover in such a way that it closes over two signal cables. Cable cover Signal cable 2-20-2 L2E-BW1444 (9) Connect the power cable to the image processing unit. 1) Fix the power cable with clamp A. In this case, fix the cable shield by the shield clamp. 2) Connect the GND cable as illustrated below and fix it by clamp B. Connect the GND cable securely, otherwise the image processing unit may be broken. Shield clamp Clamp A Power cable 3) RUN# RUN# 161 061 Clamp B GND cable Mounting of image processing unit cover. Mount the cover by reversing the removal work procedure. (Turn on 1FFB.) 2-21 L2E-BW1444 Fixing of image processing unit After turning the front casters sideways, fix them by inserting the attached caster holders. Caster Caster holder 2-22 L2E-BW1444 2.2.5 X-ray control unit RUN# 100 RUN# 303 RUN# 106 107 161 131 105 RUN# RUN# 131 161 RUN# 106 100 Fig. 2.2.4 Front view of X-ray control unit 2-23 L2E-BW1444 Power supply unit RUN# 100 RUN# 302 RUN# 161 RUN# 131 Starter RUN# 106 Fig. 2.2.5 Left side face of X-ray control unit Fig. 2.2.6 Right side face of X-ray control unit 2-24 L2E-BW1444 Terminal 1TN Power supply unit Terminal 5TN RUN# 100 RUN# 131 161 Fig. 2.2.7 Power supply unit Starter Terminal TB1 (H7,H8,H9) RUN# 106 Fig. 2.2.8 Starter 2-25 L2E-BW1444 For cabling to the X-ray control unit, remove the front, right, and left covers, and observe the following procedure. (1) For the cables being connected from the front lower part of the X-ray control unit, separate these cables from the front of the inverter unit and the front of the rectifier circuit as much as possible for the purpose of preventing ingress of noises. (2) Cleat Run #100 after cleating Run #131, 161, and 106. (3) Short 2TN DI terminals if the door switch is not used, otherwise a “DOOR OPEN” error occurs. (4) AC 100V is output to 2TN - WL terminals in the following cases. ・ The voltage is output by turning on the X-ray when No. 1 of DIP SW DSW1 of XCONT5 PCB is turned on. ・ The voltage is output by turning on the gantry power supply when No. 1 of DIP SW DSW1 of XCONT5 PCB is turned off. For setting the mode where the AC 100V is output when turning on X-rays, don’t connect any fluorescent lamp between WL terminals, otherwise the system does not function normally. Connect a fluorescent lamp whose power consumption is lower than 60W. (5) RUN#100 cable is provided with a crimp style terminal as an attachment. Terminate this cable as occasion demands after adjusting its length. * Treat the shield securely. (6) RUN #105 cable is a high-voltage transformer input side cable. It is not a high voltage cable, but it produces noises. Separate this cable from the other cables. A cable length change may cause X-ray output fluctuations. Fix the cable length to a range of 15m. 2-26 L2E-BW1444 2.2.6 Mounting the counter To avoid malfunction caused by vibration during transportation, the scanner counter is shipped as dismantled. To install it, follow these steps: 1. Press the separately packed counter onto the metal plate on the pc board of the X-ray control unit and fit into it. 2. Connect the lead wire of the counter to 501CNT cable. Then get on a tube over the terminal so that the bare terminal is not exposed. 2-27 L2E-BW1444 2.3 Internal wiring of system 2.3.1 Wiring on scanner rotary plate Connect the following cables after assembling the X-ray tube assembly and detector unit if these units are delivered separately without being assembled to the system. (1) Connect the anode and cathode high voltage cables to the X-ray tube assembly. (Remove dirt from the high voltage cable plugs by using gauze or the like and coat them with silicon compound before connecting these cables.) (2) Connect X-ray tube assembly and cooler cables to terminal boards 91TN and 92TN. (See Fig.2.3.1) (3) Connect the oil container ⑧ and relief valve ⑨ of the X-ray tube assembly to each other by using vent hose ⑦, and fix the hose by hose clip ⑩. (4) Fix high voltage cables ②, ③, oil hose ⑥, and vent hose ⑦ securely by using tie laps so as not to allow them to come outside. (5) Turn the rotary plate clockwise (as viewed from the gantry unit front) slowly by hand, and make sure that neither high voltage cables nor vent hose touches any other part. Refer to 1.7 Mounting of X-ray unit and cooler. 2-28 L2E-BW1444 Front of scanner ① ② ⑪ ⑫ ④ 92TN Rear of scanner 91TN ⑥ ③ ⑤ ⑨ ⑧ ⑩ ⑦ Note : This figure shows X-ray tube assembly GS4570. ⑤ Starter cable ① X-ray tube assembly (91TN:7, 8, 9, CS, EX) ② High voltage cable (cathode side) ⑥ Cooling oil hose ③ High voltage cable (anode side) ⑦ Vent hose ④ Cooler power cable ⑧ Oil container (92TN:ED,E, L0, L110, H, H0) ⑨ ⑩ ⑪ ⑫ Relief valve of X-ray tube assembly Hose clip Guard bracket 1 Guard plate Fig. 2.3.1 X-ray tube assembly wiring 2-29 L2E-BW1444 2.3.2 Detector cabling work (1) Remove the cover from the detector connector unit. (2) Connect one slice width switching signal cable (10-signal flat ribbon cable) and 38 detector signal cables (100 signal flexible PC board cables) between the detector and the preamplifier unit, referring to Fig.2.3.2. (3) Mount the cover of the detector connector. Fig. 2.3.2 Signal connections between detector and preamplifier 2-30 80 L2E-BW1444 Section 3 ADJUSTMENT 3.1 Turning on the power supply Check the following items before turning on the power supply (1) Check if cable connection destinations are correct. Check if terminal screws are fastened securely. (2) Check the insulation between each power line and the grounding point in all units. (3) Check if PCB and connectors are securely connected. 3.2 Operation check of image processing unit in operation Unit POWER ON (It takes about 2 minutes after turning on the power supply) Is the unit initialized? (Screen display) NO YES Check if PCB are inserted securely and also cables are connected securely Check DC power supply Image display NO YES Normal? Is the unit initialized? NO YES Check by the check program OK Fig. 3.2.1 3-1 L2E-BW1444 3.3 Operation check and adjustment of scanner unit (1) Turning on the power supply Open the front cover and side covers of the scanner unit to make sure that circuit breakers 21FFB , 22FFB, 31FFB are turned on (set to the upper position) and switches 21S and 22S are set to NORMAL (lower position). When the front cover is opened, front cover limit switch 3LS is turned off. Turn on the limit switch by using front cover limit switch jig. When front cover limit switch is detected please turn off the scanner at once. Refer to the enlarged view of SWBD PCB periphery (2) 3LS:Front cover switch Control unit MTCPU1 PCB Or MTCPU2 PCB Inverter unit 31FFB 22FFB Left side view of scanner 21FFB Front view of scanner Fig. 3.3.1 Mounting layout of scanner switch and PCB 3SW Manual/Normal selector switch 14SW Manual rotation switch 15SW Manual rotation switch 4SW Emergency switch 22SW Brake free selector switch 21SW Stationary/Normal selector switch Fig.3.3.2 Enlarged view of SWBD PCB periphery 3-2 L2E-BW1444 (2) By turning on the power switch on the scanner gantry side on the central operator console, the STATUS indicator lamp on MTCPU1 (or MTCPU2) PCB changes since the CPU is being checked first. Then, power is applied to the inverter. When no abnormal symptom occurs, STATUS indicator lamp LD1 lights V-TRIG2 (Not used) X-TRIG indicator LED/check pin V-TRIG indicator LED/check pin ZEBRA indicator LED/check pin ABORT indicator LED/check pin to be ready for receiving the communication from the operator console. DEADMAN check pin 7 ~ 0 STATUS indicator LED Fig.3.3.3 Mounting layout of indicator LED and check pins on MTCPU1 PCB 3-3 L2E-BW1444 ABORT check pin ZEBRA check pin V-TRIG check pin X-TRIG check pin 7~0 STATUS display LED X-TRIG check pin V-TRIG check pin ZEBRA check pin ABORT check pin DEADMAN check pin Fig.3.3.3.1 Wiring Diagram of MTCPU2 PCB Display LED and Check Pins 3-3-1 L2E-BW1444 (3) Check parameter values according to the following procedure after power supply has been applied to the motor controller (inverter). Change the parameter values according to the same procedure when the motor controller has been replaced or when the values have been changed. Data 024, 025, 026, 027, 028, and 031 have been finely adjusted by scan speed adjustment and scanner stationary position adjustment. Don’t change them, except for readjustment. (Adjusted values remain held even after the power supply has been turned off) * In case of inverter type CIMR-V7AA22P2 The following figure shows the names of keys and their functions on the inverter operator panel. By turning on the power supply, FREF lamp of handy run lamps lights. By pressing DSPL key, the lamp indications change in the order of FREF →FOUT → IOUT → MNTR → F/R→ LO/RE → PRGM → FREF. For running the inverter, reset the lamp to the FREF lighting condition. Indicator Data indicator DIGITAL OPERATOR JVOP-140 Handy run lamps Lighting of the lamps of these items moves one by one each time DSPL key is pressed once. Also, the upper data display contents change correspondingly. FREF FOUT F/R IOUT MNTR Changes the frequency setting continuously LO/RE PRGM Changes the handy run contents DSPL RUN Increases constant No/ data value DATA ENTER STOP RESET MIN constant No. by PRGM key MAX Issues a run command Enters constant data Data are displayed after selecting Frequency setting control knob Decreases constant No /data value Issues a run stop command Error reset key in case of the occurrence of an error Status indicator lamp (same function as in RUN lamp of main unit) Names of handy run lamps FREF Frequency command setting/monitor(green) Parenthesized colors in ( ) indicate the lighting colors of lamps FOUT Output frequency monitor (green) Forward/reverse selection of operator run command (green) IOUT Output current monitor (green) MNTR Multifunction monitor (green) LO/RE Local/remote selection (red) PRGM Constant No/data (green) Fig. 3.3.4 Inverter operator panel 3-4 L2E-BW1444 For referring to setting parameters or changing the setting, light PRGM lamp, select a parameter number according to the following procedure, and refer to or change data. After the end, reset to FREF lighting condition by pressing DSPL key. Remote/local selection Method of changing data by selecting parameter No. LO/RE DSPL PRGM Parameter No /data DSPL ・Setting of parameter n003 (run command selection) PRGM ENTER Parameter n003 run command selection PRGM Initialization contents [0] operator command PRGM FREF PRGM ENTER Setting selection [1] Control circuit terminal command (Data flickers during change) PRGM Returns to constant Data setting No. display after 1 second Don’t change to [Lo], but keep [RE] unchanged as preset at the delivery time from the works in remote/local selection. Change it to [RE] if the initial value is set to [Lo]. If [Lo] remains set, the scanner does not rotate even if the mode is returned to FREF mode. Fig. 3.3.5 Inverter operation procedure 3-5 L2E-BW1444 List of inverter setting parameters (For Presto) No. Parameter names Parameter Maker’s Adjusting No(n) set values values 001 1 4 002 0 0 003 0 1 1 2 3 Constant write Control mode selection Run command selection 4 Frequency command selection 004 0 1 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 Reverse run disable selection Maximum output frequency Maximum voltage Maximum voltage output frequency Intermediate output frequency Intermediate output frequency voltage Minimum output frequency Minimum output frequency voltage Acceleration/deceleration time unit selection Acceleration time 1 Deceleration time 1 Frequency command 1 Frequency command 2 Frequency command 3 Frequency command 4 Frequency command 5 Frequency command 6 Frequency command 7 Frequency command 8 Motor rated current Multi- function input selection Input terminal 7 function selection Carrier frequency DC braking current Stall prevention function selection during deceleration Stall prevention operation level during acceleration Stall prevention operation level during run Moor rating slip Motor phase component resistance Motor no-load current Frequency command 9 006 011 012 013 014 015 016 017 018 019 020 024 025 026 027 028 029 030 031 036 051 056 080 089 0 60.0 200.0 60.0 1.5 12.0 1.5 12.0 0 10.0 10.0 6.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 8.5 2 10 3 50 1 60.0 200.0 60.0 5.0 15.0 0.1 18.0 1 5.5 5.5 53.42 42.85 28.65 21.50 14.37 10.00 0.60 0.80 7.0 9 8 13 100 092 0 1 093 170 200 094 106 107 110 120 160 2.9 0.800 35 0.00 200 2.4 1.175 52 47.61 29 30 31 32 33 34 35 Remarks Settable up to 4: n001 ~ n179 1: Input terminals are effective 1: Frequency command 1 is effective 1: Reverse run is disabled 1: 0.01s unit *Note 2 *Note 2 *Note 2 *Note 2 *Note 2 *Note 3 8: Multi-stage speed command 3 13:15kHz 1: Not provided (1) This table shows initial set values when replacing the inverter. Don’t change any parameter values other than indicated here. Don’t change any parameter during motor rotation. (2) Parameters n024, n025, n026, n027, and n028 are finally adjusted in case of speed check described later. (3) Parameter n031 is finally adjusted during stop adjustment described later. (4) If the contents of a parameter number other than specified in the table was changed by mistake, set parameter number n001 to 8 once. All parameter values are reset to the setting at the delivery time from the works of the inverter maker. 3-6 L2E-BW1444 (4) Select 3SW switch to the manual side. Perform manual scan operation by pressing 14SW and 15SW switches at the same time to make sure that the rotating part does not contact any fixed part. (5) Perform a series of scan operation at a scan speed of 0.8sec, 1sec,1.5sec, 2sec, and 3sec in NORMAL SCAN mode to make sure that the operation ends normally without being interrupted halfway. If the X-ray system adjustment and check is not completed yet, perform scanning under the turn-off condition of X-rays (by turning off 2FFB of the X-ray control unit). Perform scan once each. (6) Check the scanner speed from the operator console after checking the manual scan operation in (4). 1) Display the basic window by the following method. ・ Enter the maintenance mode by selecting [System/Exit] →[Maintenance] → [Maintenance Mode ON/OFF] from the menu. ・ Select [System/Exit] →[Maintenance] → [Services]→ [Check] again from the menu. ・ By starting [Check], a message is displayed. Check its contents and click [Yes] when no problem exists. Check dialog Check program window 3-7 L2E-BW1444 2) Click a desired check time out of 0.8sec / 1.0sec, 1.5sec, 2.0 sec., and 3.0sec. from <SCANNER SUBSYSTEM> items in <CHECK MENU> in this window, and put a check mark. After clicking a selection item, click [Check start] button, and the check operation is started. 3) When check ends, the required CW scan time is displayed in Check Result window. Make sure that respective speed values conform to the specified values. 4) If the speed does not conform to the specified values, set it to the value, whichever smaller in the adjusting value range by adjusting the motor controller parameters according to the following table. 3-8 L2E-BW1444 Speed Adjusting values 0.8 1.0 1.5 2.0 3.0 0.80±0.02 1.00±0.03 1.50±0.05 2.00±0.06 3.00±0.10 Parameter names Inverter (V7AA22P2) n024 n025 n026 n027 n028 As the parameter value increases, the speed value decreases. As the parameter value decreases, the speed value increases. (7) Adjust the parameters so that the stop position detection sensor center is closest to the striker center when the scanner has stopped after rotating it. If the distance between these centers is deviated from each other more than 2mm, adjust the inverter parameter n031 every 0.05 so that the deviation between the sensor center and the striker center does not exceed 2mm. As n031 increases as 0.80 → 0.85 →0.90, the striker position shifts in the plus direction. As n031 decreases as 0.80 → 0.75 →0.70, the striker position shifts in the minus direction. If the striker does not stop at the center completely, set it to stop after being deviated toward minus direction. If the scan rotation is started and stopped repeatedly, the motor and regenerative resistor are heated to decrease the braking force, causing the striker position to be gradually deviated toward the plus direction. Don’t repeat the stop operation at the sensor position more than 5 times. If it is desired to repeat the stop operation more than 5 times, pause it for longer than 3 minutes before restarting the operation. Striker (-) direction Sensor (+) direction Sensor mounting plate Rear view Fig. 3.3.6 Mounting layout of sensor and striker 3-9 L2E-BW1444 (8) Setting of interlocking buzzer tone The buzzer tone can be turned on or off by selecting the DIP switches (DSW5 - 6 bit) on MTCPU1 PCB. Buzzer tone ON : DSW5- bit 6 ON (standard) Buzzer tone OFF : DSW5- bit 6 OFF (9) 7-segment LED display selection By selecting the SW16 [rotary switch (RSW)] on SWBD PCB, the following contents can be displayed on the 7-segment LED on SWBD PCB. The adjustment mode may be set by selecting the rotary switch. Set the rotary switch to [0] whenever it is not used. RSW Contents Remarks 0 MCPU status display Note 1 1 SCPU status display Note 1 2 TCPU status display Note 1 3 Table position display 4 Table height display 5 Tilt angle (display/adjustment) 6 Tilt speed display 7 Limit switch status display Note 3 8 Switch status display Note 3 9 TXR error display Note 3 A MFPGA Revision display Note .5 B TFPGA Revision display Note .5 C RFPGA Revision display Note .5 D MCPU firmware (Revision display/write mode) Note 4 E SCPU firmware (Revision display/write mode) Note 4 F TCPU firmware (Revision display/write mode) Note 4 Note 2 Note 1: For the status contents, refer to the subsequent items. Note 2: For the tilt angle adjusting method, refer to “Tilt angle adjustment” item. Note 3: For details of display, refer to the subsequent items. Note 4: For the firmware writing method, refer to the version up method item. Note 5: These are displayed only for the system installed with the MTCPU2 PCB. 3-10 L2E-BW1444 Details of MCPU status By setting RSW to 0, MCPU status are displayed on the 7-segment LED The status contents are described below. MCPU status (Displayed on LD3) Display No. Status contents 1 Idle status 2 Preparation for measurement in progress 3 Preparation for measurement ends 4 Measurement in progress 5 Error status 6 Emergency status 7 Check program is operating MCPU sequence (Displayed on LD4 ~ 5) Display No. Sequence contents 0 Other than measurement in progress 11 Waiting for offset measurement start by warm up 12 Offset measurement is being executed by warm up 13 Waiting for normal measurement by warm up 14 Normal measurement is in progress by warm up 15 Communication status after the end of normal measurement by warm up 21 Waiting for the offset measurement start by scanogram 22 Offset measurement is being executed by scanogram 23 Table move start after the end of offset measurement by scanogram 24 Waiting for normal measurement by scanogram 25 Normal measurement is in progress by scanogrm 26 Communication status after the end of normal measurement by scanogram 31 Waiting for the offset measurement start by normal scan 32 Waiting for offset measurement/table feed end by normal scan 33 Waiting for the normal measurement start from OC by normal scan 34 Waiting for the normal measurement start from SCPU by normal scan 35 Normal measurement is being executed by normal scan 41 Waiting for offset measurement start by volume scan 42 Offset measurement is being executed by volume scan 43 Waiting for offset measurement/table operation start by volume scan 44 Waiting for normal measurement start after starting the table operation by volume scan 45 Normal measurement is being executed by volume scan 46 Communication status after the end of normal measurement by volume scan 3-11 L2E-BW1444 Details of SCPU status By setting RSW to 1, SCPU status are displayed on the 7-segment LED. The status contents are as described below. LD6/7 flicker alternately to indicate that CPU is operating. SCPU status (Displayed on LD4~ 5) Display No. 0 4 9 16 17 18 25 30 32 80 21 22 23 24 26 27 28 29 30 39 40 48 96 97 Status contents Measurement ready Preliminary exposure ready Stationary scan ready Not ready (Tube up) Not ready (Other than Tube up) Not ready (Rotating) Stationary switch ON Manual switch ON Brake reset switch ON Check sequence in progress Inverter power control relay OFF Inverter unit power control relay OFF Tilt power supply OFF General-purpose inverter FFB OFF X-ray tube overheat Servo pack power supply OFF Tilt power cylinder thermal relay L OFF Tilt power cylinder thermal relay R OFF Emergency Tape switch ON Scanner error (For error status, refer to the following table) Command error Communication error (time out) Communication error (sum value error) SCPU status (Displayed on LD1 ~ 2) Display No. 01 02 03 04 05 06 07 08 Note: Status contents Idle status Preparation for measurement in progress Preparation for measurement ends Measurement in progress Error status Emergency Check in progress Manual When no scanner error occurs, the above values are displayed on LD1 ~ 2. If an error occurs, the display is switched to the error status display shown in the next item. 3-12 L2E-BW1444 Error status (Displayed on LD1 ~ 2) Display No. 8 29 30 37 67 68 69 84 87 88 89 90 Error status contents Upside sensor miss Time over every 90o +12% Time shortage every 90o -12% Collimator error operation General-purpose inverter thermal relay OFF General purpose inverter trip OFF Scanner front cover open DPRAM communication initialization error Measurement end signal time-out error Timer 0/4 asynchronization Rotary panel PCB cover open Zebra counter abnormal Details of TCPU status By setting RSW to 2, TCPU status are displayed on the 7-segment LED The status contents are as described below. Display No. 0 2 3 4 10 11 14 15 16 18 19 20 25 31 32 33 34 35 36 37 38 39 64 112 Note: Status contents Ready Index sequence in progress Tilt set sequence in progress Measurement sequence (scano /volume) in progress Table - direction (forward move) limit Table + direction (backward move) limit Operation inhibition (Measurement in progress) Guide shot measurement in progress Manual operation/PAT sequence in progress Waiting for the end of index sequence Waiting for the end of tilt set sequence Waiting for the end of table operation during measurement sequence STOP switch ON Emergency Table speed error Servo ready time out Servo alarm error Servo motor stop time out Deadman relay OFF time out A/D conversion end time out SPULSE cycle abnormal (±5%) Tape switch ON Table tabletop overrun Angio mode The above status are displayed on LD3 ~ 5. Flickering of [0] of LD1 shows that the CPU is operating. 3-13 L2E-BW1444 Details of status display By setting RSW to 7/8, the status of the limit switch (LSW) and switch (SW) being read by CPU are displayed on the 7-sgment LED as shown below. If inputs overlap each other, a value obtained by adding the overlapped signals is displayed. <Display example> If [6] is displayed on LD4 when 7 is set to RSW, table forward move deceleration LSW and table forward move stop LSW are input. Details of limit switch status display RSW Display digits Display No. 7 LD5 1 Operation contents Table highest digit LSW 2 Table lowest digit LSW 4 Table backward move deceleration LSW 1 Table backward move stop LSW 2 Table forward move deceleration LSW 4 Table forward move stop LSW 1 Table forward/backward overrun LSW 2 Tilt - (backward tilt) LSW 4 Tilt + (forward tilt) LSW LD2 1 Servo alarm LD1 1 Deadman (IPU) 2 Deadman circuit is operating 4 Delay relay is operating LD4 LD3 Switch status display RSW Display digits Display No. 8 LD5 1 LD4 LD3 LD2 LD1 Operation contents LIGHT SW 2 PAT 4 RESET 1 HOME SW 2 PRESET 4 TAPE SW 1 SET 2 EMG SW 4 TILT- SW 1 TILT+ SW 2 HISPEED SW 4 LOWFORWARD SW 1 LOWBACKWARD SW 2 TABLEDOWN SW 4 TABLEUP SW 3-14 SW SW SW SW L2E-BW1444 By setting RSW to 9, the number of data retransmittion errors if image data is displayed on the 7-segment LED. Kinds of errors are displayed on LD1, and No. of errors is displayed on LD3 ~ LD5. (Maximum displayable number of errors is 255 times) TXR errors display RSW Display digits Display No. 9 LD1 0 Operation contents Number of resend errors (VN non-coincidence error) 1 Number of resend errors (LRC error) 2 Number of resend errors (start/end bit error) 3 Number of resend errors (No. of data abnormal error) Note: For the error contents, refer to the operation principle volume. 3-15 L2E-BW1444 (10) RXCPU・7-segment LED display selection The following contents can be displayed by 7-segment LED on RXCPU PCB by selecting the rotary switch (SW1/STATUS) on RXCPU PCB. Select this rotary switch as occasion demands. Set values of the rotary switch are displayed by 7-segment LED/LED1. 0.............. CPU status (LED2, 3…ATMA status) (LED4, 5…RCPU status) 1.............. TXS5 status 2.............. ATMA CPU status 3, 4.......... Error status 5.............. Collimator sensor status 6.............. RXCPU DIP switch (CPU1-DIP1) set value 7.............. FPGA DIP switch (CPU-DIP2/FPGA) set value 8.............. RXCPU CPU firmware version 9.............. RXCPU FPGA revision A ............. ATMA CPU firmware version B ............. Auxiliary heating reference data value C ~ F....... Not used RXCPU PCB 7-segment LED SW1 3-16 L2E-BW1444 (11) Optical module laser output check After confirming the rotary action, measure the optical module (hereinafter referred to as OPTM) laser output, and make sure that the measured value conforms to the reference range. 1) Turn off the scanner power supply. 2) Remove the lift cover, rear cover, and cone cover of the scanner. 3) Remove the OPTM protective cover on the rotary panel side at the position shown in the photo. Stationary side OPTM Rotary panel side OPTM and protective cover Stationary side (Lower right side on the rear panel of gantry) Rotary panel side (Lower part of rotary panel CPU unit) 2KYSW 5LSW Enlarged view of rotary panel Fig. 3.3.7 OPTM position and limit switch and key switch positions For removing the rotary panel side OPTM protective cover, remove the key switch (2KYSW) first. After removing the cover, insert the key again, and lock it by turning it. (Bring the OPTM cover to the lowest position by turning the rotary panel for easy work) Caution : The black cable around the OPTM is the optical cable. Be careful not to flaw, pinch, or bend it to be less than 30mm in radius. 4) After removing the cover, fix the limit switch (rotary panel side 5LSW) which detects the removal of the cover to the ON condition by using tie lap or the like. 3-17 L2E-BW1444 5) Set the LASER TEST switch on the rotary panel side OPTM (FO6508-1) to TEST. (The selecting direction of the switch is described on the nameplate on the side.) LASER TEST switch Rotary side OPTM (FO6508-1) FO6508-1 switch position Fig. 3.3.8 Rotary side OPTM 6) Make sure that the TEST PHOTODIODE switch of the stationary side OPTM (FO6509-1) is set to LOW, and select it to LOW if it is set to HIGH. TP2CN TP3CN TEST PHOTODIODE switch Stationary side OPTM (FO6509-1) F06509-1 switch, connector position Fig. 3.3.9 Stationary side OPTM (The selecting direction of the switch is described on the upper nameplate.) Prepare a circuit tester or a measuring unit that can measure the voltage, and two crocodile clips, and connect the plus side of the probe to TP2CN and also connect the minus side to TP3CN. Probe minus side Probe plus side An example of the crocodile clip connection 3-18 L2E-BW1444 7) Turn on the scanner power supply, and measure and record the voltage of the measuring unit connected to the stationary side OPTM. (Caution : Don’t directly see the semiconductor laser beam after turning on the power supply.) The value becomes about 0mV. This is an OFFSET voltage and it serves as the reference of the voltage to be measured hereafter. Make sure that the entire FORJ on the rear side of the rotary panel illuminates red. 8) Remove the stationary side OPTM cover, and change the stationary side OPTM optical fiber connection from RX connector to TEST connector. If a black protective cap is mounted to the TEST connector, remove it before changing the connection. OPTM cover TEST connector (Connected during test) RX connector (Normally connected) Fig. 3.3.10 Stationary side OPTM test connector connection face 9) Clean the FORJ groove by using alcohol and a cotton rod (Prepare alcohol. For the cleaning method, refer to [Maintenance and Inspection] volume.) 10) Release the scanner brake. Turn the rotary panel side at a ratio of one turn/minute manually, observe the maximum voltage and minimum voltage of the connected voltage measuring unit, and make sure that these values are within the following range. (This voltage is called optical output.) Record the maximum value and minimum value as reference values for periodical check. Table 3.3.10 Optical output range Measuring channel Reference range High-speed light transmission 50mV~400mV 11) After the end of measurement, reset the unit to the original condition by reversing the above procedure. Be careful with the following items. ・ Make sure that the TEST PHOTODIODE switch of stationary OPTM is set to LOW. ・ Make sure that the fiber on the stationary OPTM is connected to RX connector. ・ Make sure that the LASER TEST switch of rotary panel side OPTM is set to RUN. ・ Remove the tie lap which fixes the limit switch (5LSW). ・ Make sure that the key switch (2KYSW) is inserted and locked by rotating to the ON side. 3-19 L2E-BW1444 3.4 Operation check of patient table and adjustment of scanner gantry tilting (1) Vertical move of patient table For the system installed with the MTCPU2 PCB, refer to (5). (a) Lift the tabletop to 750 +0/-3mm above the floor level, and make sure that the height display is 000. (b) Make sure that the patient table height is 1000±3mm above the floor and the display indicates 250 when lifting the tabletop up to 1000mm. Make sure that the lift switch LED goes out. (c) Make sure that the patient table height is 420±3mm above the floor when the table lowers to the lowest position. Make sure that the lowering switch LED goes out. If deviated, adjust TCB140 PCB in the table according to the following procedure. (d) Adjust R41 of TCB140 PCB so that the table height indication becomes 079 when the patient table height above the floor level is 829+0/-1mm. (e) Adjust R42 so that the table height indication becomes 250mm when the patient table height above the floor level is 1000±3mm. Caution 1) No display appears if the patient table is lower than 750mm above the floor level. Mounting position of R41 of TCB140 PCB See Fig.3.4.1 Display decreases R41 Display Mounting position of R42 increases of TCB140 PCB See Fig.3.4.1 Display increases Display decreases Reference adjusting value Change quantity in 180° rotation = approx. 5mm Reference adjusting value Change quantity in 3 turns = approx. 5mm R42 Update the height indication by pressing table up/down switch momentarily after adjusting R41/42, and check the height. 3-20 L2E-BW1444 Rear of table Table control unit TCB 140 PCB DC power supply 1DPS (lower part of PCB) TCB140 PCB R41 (offset adjustment) R42 (gain adjustment) Fig. 3.4.1 Mounting layout of component parts of patient table 3-21 L2E-BW1444 (2) Gantry tilting (a) Set an angle gauge on the gantry. (b) Make sure that the gantry angle indicating LED goes out at 0o gantry angle position on the angle meter and also the tilt angle display LED becomes +25.0 and -25.0 when the gantry angle is +25o and -25o positions respectively. For tilting the scanner gantry before mounting the lift cover, fix the gas spring to the frame (square pipe) with tie lap or the like, and be careful not to allow the gas spring to collide against the scanner gantry to be broken. (c) If the gantry angle on the angle gauge does not coincide with the value of the tilt angle indicating LED , perform the tilt adjustment according to the following procedure. 1. Press RESET switch first without fail when pressing plural switches at the same time in the following procedure. 2. If the work was mistaken in the following procedure, continue pressing the RESET switch until the buzzer sounds (for longer than 15 seconds) ,and then, repeat the procedure from the beginning. 3. When the buzzer sounds once in the following procedure, it shows that the work has been succeeded. If it sounds three times, it shows that the work was failed. Survey the PCB in case of the latter. (d) Move the table to the most backward move position, and make sure that the buzzer sounds at this time. If the buzzer does not sound, open the rear cover of the scanner, and set No. 6 of DIP switch 5 (DSW5) of MTCPU1 (or MTCPU2) PCB to ON (1) [Buzzer warning tone ON setting]. (e) Open the small cover on the side panel of the scanner (left side as viewed from the front). (f) Set the rotary switch (SW16) of SWBD PCB to [5 Tilt angle display]. DIP switch 5 (DSW5) DIP switch 5 (DSW5) Fig.3.4.2 MTCPU1 PCB Fig.3.4.3 MTCPU2 PCB 3-22 L2E-BW1444 Rotary switch (SW16) Fig.3.4.4 SWBD PCB (g) Continue pressing the RESET switch until the buzzer sounds (for longer than 15 seconds) until the buzzer sounds. When the buzzer has sounded, release the switch, and make sure that the tilt angle indicating LED flickers at intervals of one second. (h) Continue pressing the tilt (+) switch and tilt (-) switch while pressing RESET switch at the same time until the buzzer sounds (for longer than 5 seconds) until the buzzer sounds at 0o gantry angle position on the angle gauge. When the buzzer has sounded, release the switch. (i) Continue pressing the tilt (+) switch while pressing RESET switch at the same time until the buzzer sounds (for longer than 5 seconds) until the buzzer sounds at +25.0o gantry angle position on the angle gauge. When the buzzer has sounded, release the switch. (j) Continue pressing the tilt (-) switch while pressing. RESET switch at the same time until the buzzer sounds (for longer than 5 seconds) until the buzzer sounds at -25o gantry angle position on the angle gauge. When the buzzer has sounded, release the switch. (k) After restarting the power supply, make sure that the tilt angle indicating LED goes out at 0o position and also the tilt angle meets the LED value at ±25o positions. If they don’t coincide with each other, repeat the above procedure starting with step (g). (l) Set the rotary switch (SW16) of SWBD PCB to [0]. (m) If No. 6 of DIP switch 5 (DSW5) was changed to [ON (1)] in step (d), reset it to [OFF (0)]. (n) End (3) Longitudinal move of tabletop (a) Measure the time when moving the table top 300mm at high speed, and make sure that this time is within a range of 5±0.2s. (Type CT-WT-12L, within a range of 3±0.2s) (b) Measure the time when moving the table top 25mm at low speed, and make sure that this time is within a range of 5±1s. (4) Table stop position check (a) Move the table top several times by index 10mm feed from the console side, and make sure by indications that the table top move precisely by a set value. (b) If the table top does not move by the set value, perform adjustment, referring to [10.2 Stop position accuracy setting method]. 3-23 L2E-BW1444 (5) Table vertical movement (the system with the MTCPU2 PCB installed) (a) Make sure that the height indicator LED shows a value of 000 when the tabletop is set at the height 750+1/-3mm above the floor level. (b) Make sure that when the tabletop is moved up to the height 1000mm the height above the floor level is 1000±3mm and the height indicator LED shows a value of 250. (c) Make sure that when the tabletop is moved down to the lowest position the height is 420±3mm. Also, make sure that the move-down switch LED goes out. If the height is not correctly indicated, make adjustment in the following procedure. 1. The interlock for vertical movement does not function while the adjustment for table height being made, and the movement is stopped only with the limit switch. The interlock for tilt movement does not accurately function. Therefore, full care must be exercised in carrying out the adjustment. 2. When more than one button are simultaneously pressed in the following steps, be sure to press the PRESET button first. 3. If the work is incorrectly done in the following steps, after pressing and holding down the PRESET button until the buzzer sounds (more than 15sec.) with the rotary switch (SW16) on the SWBD PCB set to [4], retry the steps from the beginning. 4. The buzzer sounding one beep means successful work in the following steps, while three beeps means unsuccessful work. If the buzzer sounds three beeps, the PCB must be checked. (d) Move back the table to the backmost position, and make sure that the buzzer sounds at this position. If the buzzer does not sound, open the rear cover of the scanner unit, and set No. 6 of the dip switch 5 (DSW5) on the MTCPU2 to [ON(1): “Buzzer warning sound set to ON”]. (e) Open the small cover of the side cover of the scanner unit (left side in view from the front). (f) Set the rotary switch (SW16) on the SWBD PCB to [4: “Table height display”]. (g) Press and hold down the RESET button until the buzzer sounds (more than 15sec.). When the buzzer sounds, release the button, and make sure that the table height indicator LED flashes at [999]. (h) At the table height 745mm above the floor level measured with a scale, as pressing and holding down the RESET button press and hold down the move-down switch all the two buttons together until the buzzer sounds (more than 5 sec.). When the buzzer sounds, release the buttons. (i) At the table height 829mm above the floor level measured with a scale, as pressing and holding down the RESET button press and hold down the move-up button and move-down button all the three buttons together until the buzzer sounds (more than 5 sec.). When the buzzer sounds, release the buttons. (j) At the table height 1005mm above the floor level measured with a scale, as pressing and holding down the RESET button press and hold the move-up switch all the two buttons together until the buzzer sounds (more than 5 sec.). When the buzzer sounds, release the buttons. 3-23-1 L2E-BW1444 (k) After the system power is turned on again, make sure that the table height indicator LED does not flash and the table height indicator LED shows 0 and 250 respectively at the table height 750+1/-3mm and 1000±3mm measured with a scale. ※ If the measured table height and LED indicator height do not agree, after pressing and holding down the RESET button until the buzzer sounds (more than 15 sec), retry the steps from (4). (l) Set the rotary switch (SW16) on the SWBD PCB to [0]. (m) If No. 6 of the dip switch 5 (DSW5) was changed to [ON(1)] in the step (3), reset it to [OFF(0)] and finish the work. 3-23-2 L2E-BW1444 3.5 Laser marker adjustment Fig. 3.5.1 shows the mounting layout of the laser makers. Fig.3.5.4 ~ Fig.3.5.7 show the mounting method of each laser marker. Mount the laser marker adjusting jig to the rotary plate and adjust the laser marker beam to meet the black line of the jig. Laser marker (3) (Patient positioning) Laser marker (1) (slice side) Laser marker (4) (median line) Laser marker (2) (Scan center height) Adjusting jig Fig.3.5.1 Black line A Black line C Black line B 4-M6×25L Hexagon socket head bolt Black line D Fig.3.5.2 Basic method of using the adjusting jig ... Common to each laser marker (i) Coincide the laser marker beam with the black line on Black line of rear plate the front plate (plate close to the laser marker in case of the right laser marker). (ii) Coincide the beam passing through the front plate hole with the black line of the rear plate. Black line of front plate Laser beam passing hole of front plate Fig. 3.5.3 3-24 L2E-BW1444 (1) Laser marker ① adjustment This laser marker projects the slice face. Coincide the laser marker beam with black line A shown in Fig.3.5.2 by moving the laser marker after loosening the screws shown in Fig.3.5.4. Coincide the beam passing the hole with the black line of the rear plate. M4 hexagon socket head screw These screws are used for largely rotating the laser marker itself. Don’t adjust these screws normally, but keep them fixed. M4 pan head screw Laser marker M4 pan head screw Set to 36mm Bracket Fig. 3.5.4 (2) Laser marker ② adjustment This laser marker projects the horizontal line showing the scan center height. Coincide the laser marker beam with black line B shown in Fig.3.5.2 by moving the laser marker after loosening the screws shown in Fig.3.5.5. Coincide the beam passing the hole with the black line of the rear plate. M4 pan head screw M4 hexagon socket head screw These screws are used for largely rotating the laser marker itself. Don’t adjust these screws normally, but keep them fixed. M4 pan head screw Laser marker Set to 36mm Bracket Fig. 3.5.5 3-25 L2E-BW1444 (3) Laser marker ③ adjustment This laser marker projects the beam perpendicular to the patient body axis. Coincide the laser marker beam with black line D shown in Fig.3.5.2 by moving the laser marker after loosening the screws shown in Fig.3.5.6. Coincide the beam passing the hole with the black line of the lower plate. M4 pan head screw M4 hexagon socket head screw These screws are used for largely rotating the laser marker itself. Don’t adjust these screws normally, but keep them fixed. M4 pan head screw Laser marker Fig. 3.5.6 Set to 36mm Bracket (4) Laser marker ④ adjustment when the system is provided with a tube shift device. This laser marker projects the median line. Coincide the laser marker beam with black line C shown in Fig.3.5.2 by moving the laser marker after loosening the screws shown in Fig.3.5.7. Coincide the beam passing the hole with the black line of the lower plate. M4 pan head screw M4 hexagon socket head screw These screws are used for largely rotating the laser marker itself. Don’t adjust these screws normally, but keep them fixed. M4 pan head screw Laser marker Set to 36mm Bracket Fig. 3.5.7 3-26 L2E-BW1444 3.6 Adjustment of X-ray system 3.6.1 X-ray tube assembly check Check the following operation. (1) The cooler fan is rotating. (2) The cooler pump operates and coolant is circulating. 3.6.2 Anode rotation drive unit (starter) adjustment. Connect a digital voltmeter across H7 and H9 of terminals TB1 of the built-in starter of the X-ray control unit. Turn off 2FFB of the X-ray control unit. Select Scan or Scanogram in the measuring main menu on the operation panel, and also select 120kV, 200mA, slice thickness 1.25mm×4, 2-second scan on the operator console, and press Confirm . As a result of this operation, the anode rotation drive unit operates in the following sequence. Perform the adjustment and check as shown in the following table. Start Normal run Note : Approx. 290 seconds Braking Fig. 3.6.1 In this system, this time is set to starter unit time plus 40 seconds on the CPU side. (Hold on time 250 seconds) 3-27 L2E-BW1444 X-ray tube GS3070 (3.5MHU) Table 3.6.1 Low speed mode Items Adjusting value Adjusting position (CPU CONT PCB) Normal voltage +10 230 -0 Vac rms +0.5 12 -0 sec. +5 88 -0 Vac rms 4RSW Anode rotation speed 2700 rotations/min. or over ――― Braking voltage 90±5Vdc 5RSW Braking time 12.7±0.5sec. 14VR Hold on time 250±1sec. 12VR Start voltage Start time 16VR 10VR Select 120kV, 175mA, 0.75mm×2 slice, 2-second scan on the operator console, and press Confirm . As a result of this operation, the anode rotation drive unit operates according to the following sequence. Perform adjustment and check according to the following table. Table 3.6.2 High speed mode Items Adjusting value Adjusting position (CPU CONT PCB) Normal voltage +10 470 -0 Vac rms +0.5 12 -0 sec. +5 168 -0 Vac rms 2RSW Anode rotation speed 8000 rotations/min. or over ――― Braking voltage 135±5Vdc 3RSW Braking time 12.7±0.5sec. 13VR Hold on time 250±1sec. 11VR Start voltage Start time 15VR 9VR 3-28 L2E-BW1444 X-ray tube GS4570 (5MHU) Table 3.6.3 Low speed mode Items Adjusting value Adjusting position (CPU CONT PCB) Normal voltage +10 244 -0 Vac +0.5 12 -0 sec. +5 104 -0 Vac Anode rotation speed 3000 rotations/min. or over Braking voltage 100 -0 Vdc 5RSW Braking time 12.7±0.5sec. 14VR Hold on time 250±1sec. 12VR Start voltage Start time rms 16VR 10VR rms 4RSW ――― +10 Select 120kV, 175mA, 0.75mm×2 slice, 2-second scan on the operator console, and press Confirm . As a result of this operation, the anode rotation drive unit operates according to the following sequence. Perform adjustment and check according to the following table. Table 3.6.4 High speed mode Items Adjusting value Adjusting position (CPU CONT PCB) Normal voltage +10 470 -0 Vac rms +0.5 12 -0 sec. +5 150 -0 Vac rms 2RSW Anode rotation speed 6000 rotations/min. or over ――― Braking voltage +10 135 -0 3RSW Braking time 12.7±0.5sec. 13VR Hold on time 250±1sec. 11VR Start voltage Start time 15VR 9VR Vdc 3-29 L2E-BW1444 (1) Output voltage adjustment (CPU CONT PCB) (1) For adjusting VR and VSW, check the symbols sufficiently so as not to mistake them. Refer to the figure (Mounting positions of VR and RSW on CPU CONT PC board). (2) Connect a true RMS meter across H7 and H9 securely and measure the output voltage. (3) An accurate value is not always be displayed in auto mode depending upon the digital voltmeter types. Select the manual mode for measurement. (4) Adjust VR and RSW while turning them bit by bit. VR can be adjusted continuously, but RSW changes stepwise form 0 to F. (5) The start voltage cannot be adjusted in real time during starting. After adjusting VR first and starting, check the output voltage. Starting must be done about 3 times until the feedback control is stabilized. Accordingly, check the output voltage after starting three times. (6) The steady-state voltage and braking voltage can be adjusted in real time. However, since the output voltage changes stepwise, set RSW to the position where is the closest to the specification. (7) Select the correct drive frequency. For details, refer to 3.6.2.1 and subsequent paragraphs. (8) Start the starter once more after braking, and check the anode rotation and braking by applying your hand to the X-ray tube container. Use the following true RMS digital voltmeter for the starter combined with 5Mhu ・ FLUKE87 ・ FLUKE179 ・ TR6845 ADVANTEST 3-30 L2E-BW1444 3.6.2.1 Starter CPU-CONT PCB adjusting method CPU-CONT PC board (in anode rotation drive unit) Fig. 3.6.2 Entire CPU-CONT PC board High speed braking voltage adjustment Low speed normal voltage adjustment Low speed braking voltage adjustment High-speed normal voltage adjustment Numeric display LED 1RSW display setting change Low speed braking time adjustment Low speed starting voltage adjustment Low speed hold ON time adjustment High speed hold ON time adjustment Low speed starting time adjustment High speed braking time adjustment High speed starting voltage adjustment High speed starting time adjustment Fig. 3.6.3 Adjusting points of CPU-CONT PC board 3-31 L2E-BW1444 3.6.2.2 Adjustment of control knobs on PC board and dial time (VR adjustment while checking the LED display) Set IRSW (16-position code switch) on CPU CONT PC board to the position corresponding to the time to be adjusted, and display the set time. Adjust each time, referring to the following table. Set the starter to the idle state. IRSW position 0 * Contents of numerical display LED Control knob Standby state Numeric LED display Setting method - - Set an actual adjusting value 1 High speed starting time 9VR 1H☆☆☆ 2 High speed hold ON time 11VR 2H☆☆☆ Set steady-state rotation time - 40 sec. 3 High speed braking time 13VR 3H☆☆☆ Set an actual adjusting value 4 Low speed starting time 10VR 4L☆☆☆ Set an actual adjusting value 5 Low speed hold ON time 12VR 5L☆☆☆ Set steady-state rotation time - 40 sec. 6 Low speed braking time 14VR 6L☆☆☆ Set an actual adjusting value For the hold ON time, ROM of the PC board has 40 seconds as its fixed value. Accordingly, the set value becomes 290 seconds (total) - 40 seconds (ROM) = 250 seconds when it is desired to set the normal rotation time to 290 seconds. 3.6.2.3 * ☆☆☆ becomes an adjusting value (a desired adjusting time is displayed). * IRSW 7 ~ F positions are used for adjustment in the works. * After the end of adjustment, reset IRSW to the 0 position. Voltage adjustment (Measure H7-H9 voltage by a true RMS meter) Voltage setting Mode Adjusting VR, RSW High speed starting voltage AC 15VR Display does not change in real time. High speed normal voltage AC 2RSW Adjustable while monitoring the display. High speed braking voltage DC 3RSW Adjustable while monitoring the display. (Check the voltage in DC mode) Low speed starting voltage AC 16VR Display does not change in real time. Low speed normal voltage AC 4RSW Adjustable while monitoring the display. Low speed braking voltage DC 5RSW Adjustable while monitoring the display. (Check the voltage in DC mode) 3-32 Remarks L2E-BW1444 3.6.2.4 No. 1 CPU-CONT PC board adjustment Adjustment and check Low speed starting time check and adjustment Adjusting value As specified in adjustment instruction Adjusting VR, RSW Measuring point mode 10VR 1RSW – 4 2 Low speed hold ON time change 0 12VR 1RSW – 5 3 Low speed braking time As specified in adjustment instruction Low speed starting voltage adjustment and check 14VR 1RSW – 6 As specified in adjustment instruction Low speed starting voltage adjustment and check 16VR H7-H9 AC voltage ① ② ③ ④ 4 ⑤ ⑥ ⑦ ⑧ Remarks For shortening the adjusting time Caution Select 120kV, 200mA, 1.25mm x 4 slice, 2-sec scan and press Confirm . Record H7-H9 voltage while observing it. Don’t touch 16VR. Proceed to the next item if no adjustment is necessary. Change 16VT by one scale division if voltage adjustment is necessary. Normal voltage turns off in 40 seconds at present After confirming that the braking voltage ends, select 120kV, 200mA, 1.25mm x 4 slice, 2-sec scan and press Confirm . Record H7-H9 voltage while observing it. Don’t touch 16VR. Proceed to the next item if no adjustment is necessary. Check a difference between the first starting voltage and the present starting voltage. Decide the next adjusting quantity by comparing the change quantity of one scale division of 16VR with the starting voltage change quantity, and change the adjusting quantity. Normal voltage turns off in 40 seconds at present. After confirming that the braking voltage ends, select 120kV, 200mA, 1.25mm x 4 slice, 2-sec scan and press Confirm . Repeat steps ⑤~⑦ and terminate the adjustment. 5 Low voltage normal voltage adjustment and check As specified in adjustment instruction 4RSW H7-H9 AC voltage 6 Low speed braking voltage adjustment and check As specified in adjustment instruction 5RSW H7-H9 DC voltage 7 Low speed starting voltage adjustment and check As specified in adjustment instruction 16VR H7-H9 AC voltage Caution 8 Low speed hold ON time setting As specified in adjustment instruction 12VR 1RSW – 5 Normal rotation time - 40 seconds 9 High speed starting time check and adjustment As specified in adjustment instruction 9VR 1RSW – 1 10 High speed hold ON time change 0 11VR 1RSW – 2 11 High speed braking time As specified in adjustment instruction 13VR 1RSW – 3 As specified in adjustment instruction 15VR H7-H9 AC voltage High speed starting voltage adjustment and check ① ② ③ ④ 12 ⑤ ⑥ ⑦ ⑧ For shortening the adjusting time Caution Select 120kV, 175mA, 0.75mm x 2 slice, 2-sec scan and press Confirm . Record H7-H9 voltage while observing it. Don’t touch 15VR. Proceed to the next item if no adjustment is necessary. Change 15VR by one scale division if voltage adjustment is necessary. Normal voltage turns off in 40 seconds at present. After confirming that the braking voltage ends, select 120kV, 175mA, 0.75mm x 2 slice, 2-sec scan and press Confirm . Record H7-H9 voltage while observing it. Don’t touch 15VR. Proceed to the next item if no adjustment is necessary. Check a difference between the first starting voltage and the present starting voltage. Decide the next adjusting quantity by comparing the change quantity of one scale division of 15VR with the starting voltage change quantity, and change the adjusting quantity. Normal voltage turns off in 40 seconds at present. After confirming that the braking voltage ends, select 120kV, 175mA, 0.75mm x 2 slice, 2-sec scan and press Confirm . Repeat steps ⑤~⑦ and terminate the adjustment. 13 High voltage normal voltage adjustment and check As specified in adjustment instruction 2RSW H7-H9 AC voltage 14 High peed braking voltage adjustment and check As specified in adjustment instruction 3RSW H7-H9 DC voltage 15 High speed starting voltage adjustment and As specified in adjustment check instruction 15VR H7-H9 AC voltage Caution 16 High speed hold ON time setting As specified in adjustment instruction 11VR 1RSW – 2 Normal rotation time - 40 seconds 17 Measurement of each setting time As specified in adjustment instruction Stop watch Measure 3-33 L2E-BW1444 3 4 5 6 B C D E F 0 1 2 7 8 9 A 1DSW 1 2 3 4 1RSW 3RSW 5RSW 2RSW 4RSW Numerical LED CPU CONT PC board Fig. 3.6.4 3.6.2.5 Status display of anode rotation drive unit (a) is displayed in numeric LED under an idle condition. (b) is displayed in numeric LED in low speed starting mode. (c) is displayed in numeric LED in high speed starting mode. (d) is displayed in numeric LED in the low speed normal run mode. (e) is displayed in numeric LED in the high speed normal run mode. (f) is displayed in numeric LED in the low speed braking mode. (g) is displayed in numeric LED in the high speed braking mode. 3-34 L2E-BW1444 3.6.3 Adjustment of mA waveform and checking of kVmA The following check must be carried out in stationary scan mode. Connect CH1 of the oscilloscope probe to TP4 (coordinates:1D) on the RTC board of the scanner rotor plate. Connect CH2 of the probe to TP10 (coordinates:1F) of the RTC board. Connect GND of the probe to GND terminal of the board (at coordinates:6A) (to avoid radiation exposure, be sure to use an oscilloscope which has STORAGE mode). Set the mode, time range, and voltage range of the oscilloscope to CHOP mode, 20[ms]/div, and CH1:2V/div and CH2:1V/div, respectively. Through the operation on the central operator console, perform X-ray irradiation and check the waveform as shown in the diagrams below. Adjustment criteria using waveform (with RSW9): Make sure that the time from kV rise to steady mA waveform (at [A]) be adjusted close to within 30 [ms], using the following waveforms as targets. It is absolutely necessary to avoid rise overshoot (otherwise the tube is liable to be damaged). RSW for adjustment (common to large and small focuses): RSWs (shown below) are mounted on RXCPU-1 board: RSW8 (for steady-state value adjustment) on front left and RSW9 (for rise adjustment) on front right. Be sure not to mess up. S W8 : mA-T B L E 0 : ±0 1 -7 : +3 mA/b i t F -8 : -3 mA/b i t SW9 position Waveform S W9 : mA-Ma c 3 1 0 : ±0 1 -7 : +1 0 mA/b i t F -8 : -1 0 mA/b i t 8 9 A B C D E F 0 1 2 3 4 5 6 7 Undershoot ・・・・・・・・・・・・Overshoot Fig. 3.5.5.1 RSW for adjustment (1) Perform stationary scan in maintenance scan of four 1.25mm slices at 120kV and 200mA for two seconds, and refer to the diagrams below to make adjustment. (1) : Basic waveform (2) : 1-bit lowered (3) : NG waveform Fig.3.5.5.2 Example of adjustment of large focus waveform 3-35 L2E-BW1444 (2) Like (1) above, perform stationary scan of two 0.75mm slices at 120kV and 150mA for two seconds to check that the waveform is as shown in Fig. 3.5.5.3. (1) : Basic waveform (2) : 1-bit lowered (3) : NG waveform Fig.3.5.5.3 Example of adjustment of small focus waveform If overshoot is observed also in small focus, lower SW9 by one bit and check the small focus waveform again whether it is as shown in Fig. 5.1. If overshoot is still observed, lower SW9 by another one bit to check if the waveform is as shown in Fig. 3.1. Combination of large and small focuses: In adjustments (1) and (2) above, set RSW so that either of these focuses will not make overshoot. Where any one of these focuses makes waveform with overshoot, it may cause the tube to be damaged. (3) kV and mA adjustment In all combinations of kV and mA that are capable of irradiating X-ray, check that tube voltage and tube current fall within the following range: kV : ±7% mA : ±10% 3-36 L2E-BW1444 3.6.4 X-ray tube aging Perform the X-ray tube aging according to the following procedure. 1. In case of GS3070 (3.5 MHU ) (1) Perform scanning under the following conditions. Perform serial scan 19 times under the conditions of 100kV, 250mA, 1.25mm x 4 slice, and 3.0 sec. scan at 10 seconds internal without scanning. (2) Wait for 30 seconds without scanning. (3) Perform scanning 11 times under the X-ray conditions specified in (1) at 35 seconds interval without scanning. (4) Wait for 30 minutes without scanning while keeping the power supply turned on. (5) Perform scanning under the following conditions. Perform serial scan 4 times under the conditions of 120kV, 250mA, 1.25mm x 4 slice, and 2.0 sec. scan at 10 seconds interval without scanning. (6) Wait for 30 seconds without scanning. (7) Perform scanning under the following conditions. Perform serial scan 4 times under the conditions of 130kV, 250mA, 1.25mm x 4 slice, and 2.0 sec. scan at 10 seconds interval without scanning. (8) Wait for 30 seconds without scanning. (9) Perform scanning under the following conditions. Perform serial scan 4 times under the conditions of 100kV, 150mA, 0.75mm x 2 slice, and 2.0 sec. scan at 10 seconds interval without scanning. (10) Wait for 30 seconds without scanning. (11) Perform scanning under the following conditions. Perform serial scan 4 times under the conditions of 120kV, 150mA, 0.75mm x 2 slice, and 2.0 sec. scan at 10 seconds interval without scanning. (12) Wait for 30 seconds without scanning. (13) Perform scanning under the following conditions. Perform serial scan 4 times under the conditions of 130kV, 150mA, 0.75mm x 2 slice, and 2.0 sec. scan at 10 seconds interval without scanning. (14) Wait for 30 seconds without scanning. 3-37 L2E-BW1444 2. In case of GS4570 (5 MHU ) (1) Perform scanning under the following conditions. Perform serial scan 22 times under the conditions of 100kV, 250mA, 1.25mm x 4 slice, and 3.0 sec. scan at 5 seconds internal without scanning. (2) Wait for 30 seconds without scanning. (3) Perform scanning 8 times under the X-ray conditions specified in (1) at 30 seconds interval without scanning. (4) Wait for 30 minutes without scanning while keeping the power supply turned on. (5) Perform scanning under the following conditions. Perform serial scan 4 times under the conditions of 120kV, 250mA, 1.25mm x 4 slice, and 2.0 sec. scan at 10 seconds interval without scanning. (6) Wait for 30 seconds without scanning. (7) Perform scanning under the following conditions. Perform serial scan 4 times under the conditions of 130kV, 250mA, 1.25mm x 4 slice, and 2.0 sec. scan at 10 seconds interval without scanning. (8) Wait for 30 seconds without scanning. (9) Perform scanning under the following conditions. Perform serial scan 4 times under the conditions of 100kV, 150mA, 0.75mm x 2 slice, and 2.0 sec. scan at 10 seconds interval without scanning. (10) Wait for 30 seconds without scanning.. (11) Perform scanning under the following conditions. Perform serial scan 4 times under the conditions of 120kV, 150mA, 0.75mm x 2 slice, and 2.0 sec. scan at 10 seconds interval without scanning. (12) Wait for 30 seconds without scanning. (13) Perform scanning under the following conditions. Perform serial scan 4 times under the conditions of 130kV, 150mA, 0.75mm x 2 slice, and 2.0 sec. scan at 10 seconds interval without scanning. (14) Wait for 30 seconds without scanning. 3-38 L2E-BW1444 3.6.5 Pre-Heating Parameter of RTC Board After installation and adjustment, this parameter should be checked when: (1) The RTC board is replaced with an RTC (B/1 or later); (in this case follow steps from 1 below), and/or (2) X-4A/4C error occurs sporadically; (in this case check the parameter first, then carry out adjustment 3 or 4) For ordinary installation other than the above cases, no particular adjustment is required. To carry out the following adjustment, be sure that HU = 20[%]or less. 10 to 20 minutes after turning on the power, check the parameter value, displayed when the RSW of RXCPU-1 is set to ‘b’, to see if the value falls within 5A-60. If it does not, carry out the following adjustment: 1. Where the 7-segment LED of RXCPU-1 indicates “2B” Turn off the power, rotate VR16 counterclockwise by three to five turns, and then turn on the power again. If the LED still indicates “2B”, repeat rotating VR16 counterclockwise by further three to five turns. After making sure that LD8 (yellow) is blinking, carry out adjustment 3 or 4.. 2. Where the 7-segment LED of RXCPU-1 indicates “24” Turn off the power, rotate VR16 clockwise by three to five turns, and then turn on the power again. If the LED still indicates “X-24”, repeat rotating VR16 clockwise by further three to five turns. After making sure that LD8 (yellow) is blinking, carry out adjustment 3 or 4.. 3. Where the figure of the 7-segment LED is decreasing while LD8 is blinking (be careful not to rotate VR16 excessively) The figure of the 7-segment LED will not change when VR is rotated by one to 1.5 turns. Further turns will make the 7-segment LED change slightly. Do not attempt to rotate it impatiently. With the power being as it is, turn VR16 clockwise until the 7-segment LED indicates “5A-60”. If it indicates “X-2B” in the course of adjustment, turn the power on again, and continue to make the above-mentioned adjustment.. 4. Where the figure of the 7-segment LED is increasing while LD8 is blinking (be careful not to rotate VR16 excessively) The figure of the 7-segment LED will not change when VR is rotated by one to 1.5 turns. Further turns will make the 7-segment LED change slightly. Do not attempt to rotate it impatiently. While observing the 7-segment LED, (1) If the figure exceeds “60”, turn VR16 counterclockwise to adjust the figure until it becomes “5A-60”. (2) If the figure stops increasing at “5A”, turn VR16 clockwise to adjust the figure until it becomes “5A-60”. (3) If the figure indicates “5A-60”, no adjustment is needed. If the figure indicates “X-2B” during adjustment, turn on the power again, and continue the above-mentioned adjustment. 3-39 L2E-BW1444 3.7 Adjustment of X-ray field 3.7.1 Adjustment in X-axis direction (1) Loosen M8 hexagon socket head bolts ①. (2) Loosen nuts ②. (3) Adjust the X-axis direction position of X-ray tube by using bolts 3A 3B . (For this adjustment, mount a dial gauge as shown in the following figure, and check the moving quantity of the X-ray tube. Remove the dial gauge after adjustment) (4) Fasten four M8 hexagon socket bolts ①. (5) Tighten nuts ② to lock bolts 3A 3B securely. M8 hexagon socket head bolt ① Nut ② R direction A X direction Fig. 3.7.1 Positioning of X-ray tube assembly Dial gauge M8 hexagon socket head bolt (1) Bolt 3B (for pulling) Bolt 3A (for pushing) Fig. 3.7.2 Enlarged view of part A 3-40 L2E-BW1444 3.7.2 R-direction adjustment (Common to each X-ray tube assembly) (1) Loosen four M8 hexagon socket head bolts ⑤ [or six M6 hexagon socket head bolts in case of 5MHU] and remove the retaining band ④. (2) Remove four M8 hexagon socket head bolts ⑥, and remove X-ray tube clamp ⑦. (3) Align the center mark line of the X-ray tube housing with the end face of the X-ray tube holder exactly. (4) Fasten the X-ray tube clamp ⑦, and then, fasten four M8 hexagon socket head bolts ⑥. Fastening torque 1960N-cm (200kg-cm). (5) After mounting the retaining band ④, fasten four M8 hexagon socket head bolts ⑤ [or six M6 hexagon socket head bolts in case of 5MHU]. Retaining band ④ X-ray tube clamp ⑦ M8 hexagon socket head bolt ⑥ M8 hexagon socket head bolt ⑤ Fig.3.7.3 Fig.3.7.4 3-41 L2E-BW1444 3.7.3 Normal condition of X-ray tube Certain adjustment work of the system must be done under the stable condition of the elongation of the anode axis of the X-ray tube. In this manual, the stable condition of the elongation of the anode axis of X-ray tube is called normal condition. For setting the X-ray tube to be normal condition, the X-ray tube must be cooled down suitably after heating it sufficiently. Warm it up according to 3.1.4.3 and interrupt it just before starting the air calibration to set the accumulated heat quantity of the anode to be 43 ~58%. This condition is called normal condition. This normal condition must be maintained in 3.7.5 and 3.12. If the normal condition is already maintained, it is not necessary to repeat the warm up in the adjustment work requiring the [normal condition]. 3-42 L2E-BW1444 3.7.4 X-ray beam exposure field adjustment (slice thickness direction) (1) Stop the focus shift detector control by turning off SW10 of RXCPU PCB (LD26 (FSD_ON) goes out). SW6 BACK LED15 SW7 MOVE LED17 FRWD LED16 FSD_ON LED26 SW10 (RXCPU PCB) Goes out OFF SW10 : Focus shift detector ON/OFF selector toggle switch LED26 (FSD_ON) : Lights when the focus shift detector turns on. SW6 : Press this switch when moving the X-ray tube toward the rear of the gantry. SW7 : Press this switch when moving the X-ray tube toward the front of the gantry. LED15 (BCK) : Lights when pressing SW6 LED16 (FWRD) : Lights when pressing SW7 LD17 (MOVE) : Lights when the X-ray tube moves (2) Set the X-ray tube to the normal condition. Keep the X-ray tube under normal condition hereafter by using the [stationary scan] as occasion demand during the X-ray beam exposure position adjustment in the slice thickness direction. The anode accumulated heat quantity increases when executing 100kV, 200mA, 3-second scan. Anode accumulated heat quantity increases about 3% in case of 3.5MHU X-ray tube Anode accumulated heat quantity increases about 2% in case of 5MHU X-ray tube 3-43 L2E-BW1444 (3) Set the switch 21SW at the front right of the scanner unit to the STATIONARY SCAN side so that the scanner does not rotate (See Fig.3.7.5). Front view of scanner Fig. 3.7.5 Stationary method of scanner (5) Apply the beam adjusting jig ① to beam adjusting jig collision plate ② (See Fig.3.7.6). The 1mm groove of beam adjusting jig ① serves as the normal X-ray beam center. (6) Set the film ③ to be caught between the detector and beam adjusting jig ①. (7) Select the maintenance scan out of the main menu on the operation panel. (If the maintenance scan is not displayed, display it according to 5.1.1). (7.1) Set the following conditions and press Items No. of scans Table index (mm) Gantry angle (°) Tube current (mA) Tube voltage (kV) Scan time (s) Collimation (mm) Stationary scan Other items Confirm switch. Set values 1 0 0 50 100 1 1.25mm×4slice ON Any 3-44 L2E-BW1444 (7.2) START Press switch flickers when the system has been set to be ready for measurement. START switch. (7.3) After the end of stationary scan, develop the film. (8) Measure the deviation between the narrow X-ray tube part and the broad X-ray tube part. The move quantity of the X-ray tube corresponds to about 40% of the above quantity. Move the X-ray tube backward by pressing SW6 of RXCPU PCB if the narrow X-ray tube part is displaced toward the front of gantry unit. Move the X-ray tube forward by pressing SW7 if the narrow x-ray tube part is displaced toward the rear of gantry tube. Repeat steps (2) ~ (5) until the deviation between the narrow X-ray tube part and the broad X-ray tube part becomes within 0.5mm. (9) After the end of adjustment, perform (9.1) ~ (9.2) (9.1) Reset switch 21SW changed in (4) above to NORMAL. (See Fig.3.7.5) (9.2) Turn off No.1 of 3SW of FSD PC board. If the X-ray beam exposure position was adjusted in the system having the focus move detector, the adjustment (3.12) about the focus move detector must be done even if the other adjustment is not necessary. 3-45 L2E-BW1444 ② ③ ① ① ③ ② Detector Detector Broad X-ray part Narrow X-ray part Photosensitive film Less than 0.5mm ① Beam adjusting jig ② Beam adjusting jig collision plate ③ Film Fig. 3.7.6 X-ray beam adjustment 3-46 L2E-BW1444 3.8 Preamplifier check 3.8.1 Bus line check Execute the bus-line check of the check program referring to the bus line check method in 5.8.2 (1). In the bus-line check, whether data bus line between the preamplifier and the image processor is normal or not by generating a simulated signal from the preamplifier. In this case, the preamplifier control signal functions in the same way as in scan measurement. For confirming the control signal, set the switch 21SW at the front upper right of the scanner unit to STATIONARY so that the scanner does not rotate. 3-47 L2E-BW1444 3.8.2 Output sequence check Check the data sequence from the detector channel to the image processing unit via the preamplifier according to the following procedure. (1) Execute the calibration according to the following procedure. (1.1) Select the modes in the order of warm up→calibration select mode from the main menu on the operation panel. (1.2) Select 120kV, 5mm x 4 slice thickness on the operation panel, and press Confirm button. (1.3) After the system has been set to be ready for measurement, Press START START switch flickers switch. (2) After the end of calibration, fix either Cu wire solder wire having a size of about 1mm to the right or left end of the tabletop using an adhesive tape, and adjust the patient table height to obtain the maximum visual field. (3) Select the maintenance scan from the main menu of the operation panel. (3.1) Set the scan conditions on the operation panel as shown below and press Confirm button. Items No. of scans Table index (mm) Gantry angle Tube current Tube voltage Scan time Collimation Inter-scan time Stationary scan Delay Start mode FOV (mm) Filter Variation code Phantom correction Bias ADPP (measurement) ADPP (reconstruction) Scattered ray correction CT number adjustment Large attenuation correction Set values 1 0 0 50 120 1 5mm×4slice Any OFF Any Any Maximum 1 OFF OFF 0 ON ON OFF OFF OFF (Note) For the items indicated as [option], their set values at that time may be used as they are. For the item indicated as [Maximum], input a settable maximum value. (3.2) After the system has been set to be ready for measurement, 3-48 START switch flickers. L2E-BW1444 (4) After the end of scan, start RAW DATA analysis of the service program to indicate raw data. (4.1) Select items in the order of [File] - [Open] - [Raw Data]. (Click [File] by the left key of mouse, point [Open], and click [Raw Data] by the left key of mouse.) (4.2) Select the Raw Data scanned in (3) (Click the corresponding patient number and slice number data by the left key of mouse), and click OK by the left key of mouse to display Raw Data. (4.3) Set a window width (W/W) suitable for observing the wire trajectory. (Be careful since the input is not completed unless ENTER key is pressed when inputting them from the keyboard.) 3-49 L2E-BW1444 (4.4) Make sure that the wire trajectory changes smooth without any difference in level or other defects. Short-circuit Disconnection ch disconnection of detector cable (4.5) For terminating the raw Data, select items in the order of [File]-[Exit]. (Click [File] by the left key of mouse, and then, click [Exit] by the left key of mouse.) 3-50 L2E-BW1444 3.9 Checking the influences of vibration noises on measuring data (Vibration check) Check the vibration noises of detector, detector cable, and preamplifier according to the following procedure. (1) Turn off 2FFB of the X-ray control unit to turn off X-rays. (2) Select the maintenance scan from the main menu of the operation panel. (If the maintenance scan is not displayed, display it according to 5.1.1) (2.1) Set the scan conditions on the operation panel as shown below and press Items button. Set values No. of scans Table index (mm) Gantry angle (°) Tube current (mA) Tube voltage (kV) Scan time (s) Collimation (mm) Inter-scan time (s) Stationary scan Delay (s) Start mode F.O.V. (mm) Filter B.H.C. B.G.C. Variation code Phantom correction Bias ADPP (measurement) ADPP (reconfiguration) Scattered ray correction CT number adjustment Large attenuation correction (Note) Confirm 1 0 0 175 120 0.8 0.75mm×2slice Any OFF Any Any Maximum 4 OFF OFF 3 OFF 320 OFF OFF OFF OFF OFF For the items indicated as [option], their set values at that time may be used as they are. For the item indicated as [Maximum], input a settable maximum value. (2.2) After the system has been set to be ready for measurement, START switch flickers. Press it. (3) At the end of scan, start RAW DATA analysis of the service program to indicate Raw Data. (3.1) Select items in the order of [File] - [Open]- [Raw Data] from the menu bar of RAW DATA analysis, click [File] by the left key of mouse, point [Open], and click [Raw Data] by the left key of mouse. 3-51 L2E-BW1444 (3.2) Select the Raw Data scanned in (2) (Click the corresponding patient number and slice number data by the left key of mouse), and click OK by the left key of mouse to display Raw Data. (3.3) Set a window width (W/W) to 500 and window level (W/L) to 21430 (Be careful since the input is not completed unless ENTER key is pressed when inputting them from the keyboard). (3.4) Make sure that the displayed Raw Data are free of extreme longitudinal streaks, lateral streaks, black lines, and white lines, and execute VCK processing as described in and after (3.5). 3-52 L2E-BW1444 (3.5) Select items in the order of [Analysis] - [VCK] from the menu bar of RAW DATA analysis (Click [Analysis] by the left key of mouse and then, click [VCK] by the left key of mouse). (3.6) Set the Diff. threshold value to 2000 and set the SD threshold value to 400 (No input is required). Click Check ENTER key by the left key of mouse. Threshold value of Diff. Threshold value of SD 2000 400 (3.7) If no channel exceeds the threshold values being set in (3.6), a message [All channels are OK] is displayed. If a channel exceeds these threshold values, data of the channel are displayed. If none of channels exceeds Diff. threshold values, check results are acceptable (OK). (3.8) The ordinate range of Diff. graph or SD graph can be changed by clicking Max./Min. , or Key Input as desired using the left key of mouse. Switches Default Default , Processing contents Set ordinate higher-limit value to 65535 and ordinate lower-limit value to 0. Max./Min. Set ordinate higher-limit value to the maximum value of graph and the ordinate lower-limit value to the minimum value of graph. Key Input Display the key input window, and input the higher-limit value and lower-limit value of ordinate. 3-53 L2E-BW1444 (3.9) For returning to the RAW DATA display after terminating VCK processing, select items in the order of [Analysis] - [VCK] from the menu bar of RAW DATA analysis (Click [Analysis] by the left key of mouse and then, click [VCK] by the left key of mouse). (3.10) For terminating Raw Data analysis, select items in the order of [File]-[Exit] (Click [File] by the left key of mouse and then, click [Exit] by the left key of mouse). (4) Reset 2FFB of the X-ray control unit to ON. 3-54 L2E-BW1444 3.10 Center adjustment Adjust the center for correcting the positional relations of X-ray focus, scanner rotation center, and detector center according to the following procedure. (1) Turn off SW10 of RXCPU PCB. (LED26 (FSD_ON) goes out). (2) Execute air calibration according to the following procedure. (2.1) Select the modes in the order of warm up →calibration select mode from the main menu on the operation panel. (2.2) Select 120kV, 1.25mm×4 slice and 0.75mm×2 slice on the operation panel and press Confirm button. (2.3) After the system has been set to be ready for measurement, START switch flickers. Press it. (3) After the end of air calibration, set the center adjustment rod to 20 ~ 25mm deviated from the scanner rotation center. (4) Select the maintenance scan from the main menu of the operation panel. (4.1) Set the scan conditions on the operation panel as shown below and press Items button. Set values 1 0 0 50 120 3 1.25mm×4slice Auto OFF Any Any 160 1 OFF OFF 0 OFF 0 OFF OFF OFF OFF OFF No. of scans Table index (mm) Gantry angle (°) Tube current (mA) Tube voltage (kV) Scan time (s) Collimation (mm) Inter-scan time (s) Stationary scan Delay (s) Start mode F.O.V. (mm) Filter B.H.C. B.G.C. Variation code Phantom correction Bias ADPP (measurement) ADPP (reconfiguration) Scattered ray correction CT number adjustment Large attenuation correction (Note) Confirm For the items indicated as [option], their set values at that time may be used as they are. (4.2) After the system has been set to be ready for measurement, START switch flickers. Press it. (4.3) At the end of scan under the conditions specified in (4.1) above, perform scan at 0.75mm×2 slice (The other conditions remain unchanged from (4.1)). 3-55 L2E-BW1444 (5) At the end of scan, start RAW DATA analysis of the service program to indicate Raw Data. (5.1) Select items in the order of [Analysis] - [CACK] from the menu bar of Raw Data analysis. (5.2) Select the Raw Data scanned in (4) (Click the corresponding patient number and slice number data by left key of the mouse), and then, click OK by the left key of mouse. Now, the X-ray tube move quantity is displayed. 3-56 L2E-BW1444 (6) Move the X-ray tube according to the D calculated in (5) above. (See 3.6) Perform the work while monitoring the move quantity on the dial gauge. (a) Don’t move the X-ray tube when -0.03 ≦D ≦0.03. (b) Fasten fixing bolts at 4 positions without fail after moving the X-ray tube (See 3.6). (7) Repeat (2)~(6) until the D calculated in (5) satisfies -0.03 ≦D ≦0.03. Be careful since the air calibration must be repeated without fail after moving the X-ray tube. (8) Perform (8.1)~(8.3) after the end of center adjustment. (8.1) Fasten and check the X-ray tube fixing bolts. (See 3.7) (8.2) Loosen the X-ray moving bolts, and fix them by lock nuts at a position where they don’t touch the X-ray tube assembly. (See 3.7) (8.3) Turn on SW10 on the RXCPU pc board (LED26 (FSD_ON) is lighted). After center adjustment, Compensator adjustment in 3.11 and focus shift detector (FSD) adjustment in 3.12 must be done. 3-57 L2E-BW1444 3.11 Compensator position adjustment Adjust the compensator position according to the following procedure. (1) After making sure that the phantom, tabletop of patient table, etc. are not put in the gantry opening to be ready for air scan, select the maintenance scan from the main menu of the operation panel. (Display the maintenance scan according to 5.1.1, if not displayed.) (1.1) Set the scan conditions on the operation panel as shown below and then, press Items button. Set values 1 0 0 150 120 1 2.5mm×4slice Auto OFF Any Any Any Any OFF OFF 4 OFF 0 OFF OFF OFF OFF OFF No. of scans Table index (mm) Gantry angle (°) Tube current (mA) Tube voltage (kV) Scan time (s) Collimation (mm) Inter-scan time (s) Stationary scan Delay (s) Start mode F.O.V. (mm) Filter B.H.C. B.G.C. Variation code Phantom correction Bias ADPP (measurement) ADPP (reconfiguration) Scattered ray correction CT number adjustment Large attenuation correction (Note) Confirm For the items indicated as [option], their set values at that time may be used as they are. (1.2) After the system has been set to be ready for measurement, START switch flickers. Press it. (2) At the end of scan, start RAW DATA analysis of the service program to indicate Raw Data. (2.1) Select items in the order of [File] - [Open] - [Raw Data] from the menu bar of RAW DATA analysis, click [File] by the left key of mouse, point [Open] and click [Raw Data] by the left key of mouse. 3-58 L2E-BW1444 (2.2) Select the Raw Data scanned in (1) (Click the corresponding patient number and slice number data by the left key of mouse), and click OK by the left key of mouse to display Raw Data. (2.3) Select items in the order of [Analysis] - [CLB] from the menu bar of Raw Data analysis (Click [Analysis] by the left key of mouse, and then, click [CLB] by the left key of mouse). (2.4) Check the values being displayed at the lower end of the CLB window. These values are Left, Right, Center, and Diff. in 4 kinds. A “Diff” of less than 5% is acceptable. If “Diff” exceeds 5%, compare Left and Right values with each other, and move the X-ray tube assembly as shown below. If Left >Right, move the compensator rightward as viewed from the table side. If Left < Right, move the compensator leftward as viewed from the table side. Move the compensator according to the following procedure (See Fig.3.11) 1) Remove the collimator box cover. 2) Loosen the right and left sems screws. 3) Loosen the M4 hexagon socket head bolts in the moving direction to produce a suitable clearance at the bolt tips. 4) Move the compensator by fastening M4 hexagon socket head bolts on the opposite side. 5) Fix the compensator by fastening sems screws. 6) Loosen the right and left M4 hexagon socket head bolts to produce a suitable clearance at the bolt tips. Then, fix the bolts with nuts. 3-59 L2E-BW1444 7) After confirming that Diff in CLB processing is less than 5%, mount the collimator box cover. Collimator box cover Sems screw Hexagon socket head bolt Nut Fig. 3.11 (2.5) For returning to the Raw Data display at the end of CLB processing, select items in the order of [Analysis] - [CLB] from the menu bar of Raw Data analysis in the same way as in (2.3) (Click [Analysis] by the left key of mouse and then, click [CLB] by the left key of mouse). (2.6) For returning to the Raw Data display, select items in the order of [File] - [Analysis] (Click [File] by the left key of mouse and then, click [Exit] by the left key of mouse). (2.7) Repeat steps (1) and (2) until the criterion in (2.4) is satisfied. The Center value in CLB processing should be within a range of 21500+4300/-5375 (16125 ~ 25800) when the criterion in (2.4) has been satisfied. After compensator adjustment, focus shift detector (FSD) adjustment in 3.12 must be done. 3-60 L2E-BW1444 3.12A Focus shift detector (FSD) adjustment Rev.1(Display VR3**) *To be applied to RTC PCB Perform all FSD adjustment to all KV conditions, otherwise a ring artifact occurs. Set the X-ray tube to the normal condition in advance. If an oscillation waveform is observed during the measurement of each TP, this symptom may be avoided by turning the corresponding VR clockwise or counterclockwise. If the system was installed with the X-ray tube mounted at the installation time, no adjustment is necessary any more. Perform steps (15) ~ (17) in this chapter to make sure that FSD functions normally. Adjustment is necessary in all cases other than described above. For this adjustment, observe the following flows as required. Remove the PCB upper cover and PCB side cover. (1) Refer to Flow 1 when X-ray tube has been mounted at the installation time, when the X-ray tube has been replaced, or when the image detector has been replaced (2) Refer to Flow 2 when FSD has been replaced (3) Refer to Flow 3 when RTC board has been replaced (4) Refer to Flow 4 in the cases other than specified above PCB Upeer Cover RTC PCB PCB Side Cover TP27(GND) 3-61 L2E-BW1444 [Flow 1 ] when the X-ray tube or the image detector has been replaced [Flow 2 ] when FSD has been replaced Normal condition of X-ray tube (See 3.7.3) Normal condition of X-ray tube (See 3.7.3) X-ray tube positioning FSD offset adjustment FSD positioning X-ray tube positioning RTC board VR adjustment FSD positioning Correction to X-ray tube Correction to X-ray tube Correction to rotating operation RTC board VR adjustment Correction to X-ray tube Correction to rotating operation 3-62 L2E-BW1444 [Flow 3 ] when RTC board has been replaced [Flow 4 ] in the cases other than specified above Normal condition of X-ray tube (See 3.7.3) Normal condition of X-ray tube (See 3.7.3) FSD offset adjustment FSD offset adjustment FSD reference adjustment FSD reference adjustment X-ray tube positioning X-ray tube positioning RTC board VR adjustment width check FSD positioning RTC board VR adjustment RTC board VR adjustment width check Correction to X-ray tube RTC board VR adjustment Correction to rotating operation Correction to X-ray tube Correction to rotating operation Indications of adjusting parts of FSD The adjusting methods are described every flow Corresponding flow numbers are marked O below respective adjusting items, and also, whether HU is restricted or not and whether rotary scan is done or stationary scan is done are also described. Proceed the work while confirming that the present work is correct or not. (Example) Detector positioning Rotary ・ Stationary HU during measurement =50~ 55% 1 2 3 4 Corresponding flow No. In this case, the work is necessary for flow No. 2 and No. 4, the work is done by stationary scan, and measurement is done at HU = 50 ~ 55%. 3-63 L2E-BW1444 1. Normal condition of X-ray tube 1 2 3 Rotary ・ Stationary HU during measurement =Normal condition 4 (1) Make sure that SW10 of RXCPU PCB is turned off ( LED26 (FSDON) goes out) and the collimator case cover is mounted. (2) Set the X-ray tube to normal condition. (See 3.7.4) Perform step (3) and (4) during stop of scan. For (3) and (4), use a storage oscilloscope. 2. FSD offset adjustment 1 2 3 Rotary ・ Stationary HU during measurement = Not restricted 4 (3) Adjust VR363 so that the TP22 output becomes 0 ±10mV when X-ray tube is turned off. (A digital voltmeter can be used ) Adjusting point Measuring point Adjusting value TP22 0±10mV VR363 Table (3) Rotary ・ Stationary HU during measurement = Not restricted 3. FSD reference adjustment 1 2 3 The output increases when turning VR363 clockwise. 4 (4) Adjust RTC board VR according to the following table. Measuring conditions Slice kV mA 1.25mm×4 100 0.75mm×2 150 Adjusting Measuring Adjusting Scan time point point value 3 sec VR430 1 sec VR428 3 sec VR355 TP21 1~3V 1 sec VR429 Table (4) 4 .X-ray tube adjustment 1 2 3 The output increases when turning each VR clockwise. Rotary ・ Stationary HU during measurement = 50 ~ 55% 4 The OPR change quantity per 10µm is dispersed depending upon systems. The dial gauge point may be deviated when the scanner is set to the rotary condition. Perform adjustment by increasing or decreasing the move quantity of the X-ray tube to (the output ratio) as occasion demands. (5) Perform “FSD adjustment – small focus 1” of maintenance scan by rotary scan,check its raw data by OPR of raw data analysis , and then, adjust the X-ray tube position by SW6 and SW7 of RXCPU PCB so that the output ratio in 444 ~ 453 channels becomes 0±3%. ( Move the X-ray tube backward by pressing SW6 when the output ratio is plus or move the X-ray tube forward by pressing SW7 when the output ratio is minus. ) The output ratio changes about 3% when moving the X-ray tube 10µm in case of small focus. 3-64 L2E-BW1444 5. FSD positioning (lateral direction) Rotary ・ Stationary HU during measurement = Not restricted 1 2 3 4 (6) Set the X-ray tube at a position where it is rotated clockwise by 90°as viewed from the top as illustrated below and set 21SW to “STATIONARY SCAN” side. 90° Figure (6) (7) Perform scan under the following conditions and measure the sum of the outputs at TP19 and TP20. Set the oscilloscope trigger to TP21. Slice 1.25mm×4 Measuring conditions kV mA Scan time 120 150 Stationary 3sec. Measuring point TP19 TP20 Table (7) Figure (7) (8) Adjust the lateral position of FSD so that the sum of the outputs of TP19 and TP20 becomes as shown below. -3.0V ≦TP19+TP20≦-1.5V Perform adjustment as specified in the following table by tube. FS D検 出器 FSD Lateral 囲 動範range 横移move A Adjusting固screw 定ね じ (a) Loosen four fixing screws mounted on the lower face of FSD. (b) Position the FSD , and then, fix it by fastening the fixing screws. X 線管 X-ray tube コ リメ ータ ボッ クス Collimator box Figure (8) 3-65 L2E-BW1444 TP19+TP20 1 FSD adjustment -3.0V≦TP19+TP20≦-1.5V No adjustment is necessary Move the FSD rightward ( in the inserting direction 2 -1.5V<TP19+TP20<0V into the collimator box) , and fix it at the position where -3.0V ≦TP19+ TP20 ≦-1.5V. Move the FSD leftward ( in the direction coming 3 -3.0V>TP19+TP20 out of the collimator box) , and fix it at the position where -3.0V ≦TP19+ TP20 ≦-1.5V Table (8) In (8), if, as a result of FSD movement, the sum of the outputs at TP19 and TP20 does not fall within the following range: -3.0≦TP19 + TP20 < -1.5V make fine adjustment of the position of the Compensator according to the following instruction. ・where -1.5V < TP19 + TP20 < 0 Move the Compensator rightward, viewed from the table side, within the adjustment value in 3.11 “Compensator position adjustment”, and check again the sum of the outputs at TP19 and TP20. ・where -3.0V > TP19 + TP20 Move the Compensator leftward, viewed from the table side, within the adjustment value in 3.11 “Compensator position adjustment”, and check again the sum of the outputs at TP19 and TP20. 6. FSD positioning ( longitudinal direction) Rotary ・ Stationary HU during adjustment = 45 ~ 65% 1 2 3 4 (9) After turning VR368 ( white mark for 120KV) (click sound is heard by turning it about 30 turns ) , turn it counterclockwise by 12 turns from the above condition, and fix FSD at the position where TP22 output (point A) becomes 0±2V after 1.5 seconds from turning on the X-ray in the scan under the following conditions. Measuring conditions Measuring Reference Slice kV mA Scan time point value 1.25mm×4 120 150 3 sec Table (9) TP22 0±2V (Oscilloscope Screen) 2[V] TP22 Point[A] 0[V] *Set TP21 as trigger when checking the TP21 TP22 waveform. -2[V] 1.5sec 3sec X-Ray OFF X-Ray ON Figure (9)-a 3-66 L2E-BW1444 C:Scanner スキャside ナ側 D:Table テーside ブル側 Longitudinal direction 前後方 向 Collimator コ リメーbox タボックス 引 ねじ Pullき screw Push 押 しscrew ねじ Fixing 固定screw ねじ Fixing 固定screw ねじ Figure (9)-b (a) Loosen the fixing screws on both right and left sides of the FSD. (b) Increase the output voltage by moving the FSD in the C direction by the push screw when the TP22 output voltage is lower than -2V. Decrease the output voltage by moving the FSD in the D direction by the pull screw when the TP22 output voltage exceeds +2V. (c) Fix FSD at the position where TP22= 0±2V. 7. RTC board VR adjustment width check Rotary ・ Stationary HU during adjustment = 45~ 65% 1 2 3 4 (10) (The VR to be adjusted are VR364 ~ 366 (yellowish green) and VR367~ 369 (white). ) Make sure that the TP22 output after 1.5 seconds from turning on the X-ray in scan under the following conditions fluctuates when turning the corresponding VR. Measuring conditions Slice 1.25mm×4 kV Scan time 100 VR369 120 VR368 130 VR367 100 0.75mm×2 mA Check position 150 3 sec VR366 120 VR365 130 VR364 VR clockwise and it becomes lower than -1V when turning VR counterclockwise. Table (10) If the output does not become higher than 1Vunder certain KV condition in (10), adjust the FSD position again to increase the TP22 output and check (10) again. If the output does not become lower than -1Vunder certain KV condition in (11), adjust the FSD position again to decrease the TP22 output and check (10) again. 3-67 L2E-BW1444 8. RTC board VR adjustment 1 2 3 Rotary ・ Stationary HU during adjustment = 45~ 65% 4 (11) (The VR to be adjusted are VR364 ~ 366 (yellowish green) and VR367~ 369 (white). ) Adjust respective VR so that the TP22 output during scan shown in the following table becomes 0±0.3V at 100kV or 0±0.5V at 120kV and 130kV after about 1.5 seconds from turning on the X-ray when HU is 45 ~ 65%. (The output decreases when turning VR364~ VR369 counterclockwise. ) Measuring conditions Slice kV 1.25mm×4 mA 2 3 Adjusting point point value VR369 120 VR368 130 VR367 150 3sec TP22 VR366 120 VR365 130 VR364 Table (11) tolerance 0V ±0.3V 0V ±0.5V 0V ±0.3V 0V ±0.5V Rotary ・ Stationary HU during adjustment = 45~ 65% 9. Correction to X-ray tube 1 Measuring 100 100 0.75mm×2 Scan time Adjusting 4 (12) Measure [a ~ d] under the following table for adjusting the FSD operation when the X-ray tube s set to 0°and 180°positions. ※Measurement of a,b,c,d (Measure the output after 1.5 seconds from X-ray irradiation) Measuring conditions X-ray tube position a 0° b 180° c 0° d 180° Slice kV mA Scan time 120 150 3sec Measuring point 1.25mm×4 TP22 0.75mm×2 Table (12) (13) Perform calculation & set the value as adjusting value by substituting[a~d] obtained by measurement into the table(13). Perform scan with each measuring condition on the table(13). Check the output of measuring point & adjust the value to be “Adjustment Value”. Perform scan again & check if the value becomes “adjustment value”. ※If the calculated value is within “0~0.2[V]”、set to the “Adjustment Value” 0.2[V]. 3-68 L2E-BW1444 Measuring conditions Slice kV mA Adjusting Measuring Calculated Adjusting Scan time point point value value 3sec VR430 1sec VR428 (a-b) + 0.3 V 2 1.25mm×4 120 (a-b) +0.3 ×0.6 V 2 150 TP21 3sec VR355 1sec VR429 (c-d) +0.3 V 2 0.75mm×2 (c-d) +0.3 ×0.6 V 2 (Example) tolerance +0.2V -0V +0.2V -0V +0.2V -0V +0.2V -0V Table (13) TP22 output T P22 の出力 0° 180° Calculation 算value 出値 = 1.2V ← ”a” -1.8V ← ”b” 1.2-(-1.8) 1 2 3 +0.3V =1.8 2 10. Correction to rotary operation Adjust VR430 so that the TP21 output becomes 1.8 [V] ~ 2.0[V] in this case. Rotary ・ Stationary HU during adjustment = 45~ 65% 4 This adjustment is done under the rotary operation. Make sure that the probe ,etc. are removed without any skip. Start scanning when HU is 45 ~ 65% . (14) Turn on SW10 of RXCPUPCB ( LED26 (FSD_ON) lights to be turned on) and set 21SW to “NORMAL”. (15) Check the OPR value under the following conditions If the OPR value does not meet the specified value, adjust each VR shown in (11) again. Perform this scan by switching the kV and slice thickness from the protocol of FSD adjustmentlarge focus 1 of maintenance scan. Turn off PPC in case of 0.75mm x 2 scan. Measuring conditions Slice kV OPR value check mA Scan time No. of scans 150 3sec 3 times 150 3sec 3 times 100 1.25mm×4 120 130 100 0.75mm×2 120 130 OPR value in the 3rd scan shall be within a range of -0.6~ -1.8% OPR value in the 3rd scan shall be within ±6% *OPR : show P5-15(11) Table (15) 3-69 L2E-BW1444 (16) Measure the OPR value by executing maintenance scan under the following. Protocol conditions and assume [f~i] be the obtained values. (All these values shall be the OPR values in the final scan. ) プロトコname ル名 Protocol スscan キャンtime タイム No. スキofャンscans 回数 f FSD adjustment-large focus2 1.0[s] 5 5 回 times g FSD adjustment-large focus3 0.8[s] 5 5 回 times h FSD adjustment-small focus2 times 1.0[s] 5 5 回 i FSD adjustment-small focus3 times 0.8[s] 5 5 回 Table (16) (17) Adjust relevant VR so that [f~ i] obtained in (16) conform to the following specified values respectively. The DSW setting is OFF-Defalut. When adjustment is impossible even if you turn volume, turn on the DSW. ① [f] → Adjust VR385 so that -0.6% ≧f ≧ -1.8%. (See the following table) OPR value -0.6%≧f≧-1.8% VR385 adjustment No adjustment is necessary When No. 6 of DSW2 is turned off Adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment -1.8%>f cannot be done even when turning VR fully When No. 6 of DSW2 is turned on clockwise, turn off No.6 of DSW2 , and adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment f>-0.6% cannot be done even when turning VR fully When No. 6 of DSW2 is turned off clockwise, turn off No.6 of DSW2 , and adjust VR by turning it counterclockwise. When No. 6 of DSW2 is turned on Adjust VR by turning it counterclockwise. Table (17)-① ② 「g」→ Adjust VR383 so that -0.6% ≧ g ≧ -1.8% (See the following table) OPR value -0.6%≧g≧-1.8% VR383 adjustment No adjustment is necessary When No. 8 of DSW2 is turned off Adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment -1.8%>g When No. 8 of DSW2 is turned on cannot be done even when turning VR fully clockwise, turn off No.8 of DSW2 , and adjust VR by turning it counterclockwise. Adjust VR by turning it counterclockwise. If g>-0.6% When No. 8 of DSW2 is turned off adjustment cannot be done even when turning VR fully clockwise, turn off No.8 of DSW2 , and adjust VR by turning it counterclockwise. When No. 8 of DSW2 is turned on Adjust VR by turning it counterclockwise. Table (17)-② 3-70 L2E-BW1444 ③ 「h」→ Adjust VR401 so that 6% ≧ h ≧ -6% (See the following table) OPR value 6%≧h≧-6% VR401 adjustment No adjustment is necessary When No. 5 of DSW2 is turned off Adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment -6%>h cannot be done even when turning VR fully When No. 5 of DSW2 is turned on clockwise, turn off No.5 of DSW2 , and adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment h>6% cannot be done even when turning VR fully When No. 5 of DSW2 is turned off clockwise, turn off No.5 of DSW2 , and adjust VR by turning it counterclockwise. When No. 5 of DSW2 is turned on Adjust VR by turning it counterclockwise. Table (17)-③ ④ 「i」→Adjust VR384 so that 6% ≧ i ≧ -6%. (See the following table) OPR value 6%≧i≧-6% VR384 adjustment No adjustment is necessary When No. 7 of DSW2 is turned off Adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment -6%>i When No. 7 of DSW2 is turned on cannot be done even when turning VR fully clockwise, turn off No.5 of DSW2 , and adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment i>6% When No. 7 of DSW2 is turned off cannot be done even when turning VR fully clockwise, turn off No.5 of DSW2 , and adjust VR by turning it counterclockwise. When No. 7 of DSW2 is turned on Adjust VR by turning it counterclockwise. Table (17)-④ Perform scan in (16) again when VR383, VR384, VR385, or VR401 was adjusted , and make sure that the values in f ~ i are within a range where no adjustment is necessary. (18) Make sure that SW10 of RXCPU PCB is turned on again ( LED26 (FSD_ON) lights). 3-71 L2E-BW1444 3.12B Focus shift detector (FSD) adjustment Rev.2(Dsiplay VR*) *To be applied to RTC2 PCB Perform all FSD adjustment to all KV conditions, otherwise a ring artifact occurs. Set the X-ray tube to the normal condition in advance. If an oscillation waveform is observed during the measurement of each TP, this symptom may be avoided by turning the corresponding VR clockwise or counterclockwise. If the system was installed with the X-ray tube mounted at the installation time, no adjustment is necessary any more. Perform steps (15) ~ (17) in this chapter to make sure that FSD functions normally. Adjustment is necessary in all cases other than described above. For this adjustment, observe the following flows as required. Remove the PCB upper cover and PCB side cover. (1) Refer to Flow 1 when X-ray tube has been mounted at the installation time, when the X-ray tube has been replaced, or when the image detector has been replaced (2) Refer to Flow 2 when FSD has been replaced (3) Refer to Flow 3 when RTC2 board has been replaced (4) Refer to Flow 4 in the cases other than specified above PCB Upeer Cover RTC2 PCB PCB Side Cover 3-72 L2E-BW1444 [Flow 1 ] when the X-ray tube or the image detector has been replaced [Flow 2 ] when FSD has been replaced Normal condition of X-ray tube (See 3.7.4) Normal condition of X-ray tube (See 3.7.4) X-ray tube positioning FSD offset adjustment FSD positioning X-ray tube positioning RTC board VR adjustment FSD positioning Correction to X-ray tube Correction to X-ray tube Correction to rotating operation RTC board VR adjustment Correction to X-ray tube Correction to rotating operation 3-73 L2E-BW1444 [Flow 3 ] when RTC board has been replaced [Flow 4 ] in the cases other than specified above Normal condition of X-ray tube (See 3.7.4) Normal condition of X-ray tube (See 3.7.4) FSD offset adjustment FSD offset adjustment FSD reference adjustment FSD reference adjustment X-ray tube positioning X-ray tube positioning RTC board VR adjustment width check FSD positioning RTC board VR adjustment RTC board VR adjustment width check Correction to X-ray tube RTC board VR adjustment Correction to rotating operation Correction to X-ray tube Correction to rotating operation Indications of adjusting parts of FSD The adjusting methods are described every flow Corresponding flow numbers are marked O below respective adjusting items, and also, whether HU is restricted or not and whether rotary scan is done or stationary scan is done are also described. Proceed the work while confirming that the present work is correct or not. (Example) Detector positioning Rotary ・ Stationary HU during measurement =50~ 55% 1 2 3 4 Corresponding flow No. In this case, the work is necessary for flow No. 2 and No. 4, the work is done by stationary scan, and measurement is done at HU = 50 ~ 55%. 3-74 L2E-BW1444 1. Normal condition of X-ray tube 1 2 3 Rotary ・ Stationary HU during measurement =Normal condition 4 (1) Make sure that SW10 of RXCPU PCB is turned off ( LED26 (FSDON) goes out) and the collimator case cover is mounted. (2) Set the X-ray tube to normal condition. (See 3.7.4) Perform step (3) and (4) during stop of scan. For (3) and (4), use a storage oscilloscope. 2. FSD offset adjustment 1 2 3 Rotary ・ Stationary HU during measurement = Not restricted 4 (3) Adjust VR1 so that the TP22 output becomes 0 ±10mV when X-ray tube is turned off. (A digital voltmeter can be used ) Adjusting point Measuring point Adjusting value VR1 TP22 Table (3) 0±10mV Rotary ・ Stationary HU during measurement = Not restricted 3. FSD reference adjustment 1 2 3 The output increases when turning VR1 clockwise. 4 (4) Adjust RTC board VR according to the following table. Measuring conditions Slice kV mA 1.25mm×4 100 0.75mm×2 4 .X-ray tube adjustment 1 2 3 150 Adjusting Measuring Adjusting Scan time point point value 3 sec VR2 1 sec VR3 3 sec VR4 TP21 1~3V 1 sec VR5 Table (4) The output increases when turning each VR clockwise. Rotary ・ Stationary HU during measurement = 50 ~ 55% 4 The OPR change quantity per 10µm is dispersed depending upon systems. The dial gauge point may be deviated when the scanner is set to the rotary condition. Perform adjustment by increasing or decreasing the move quantity of the X-ray tube to (the output ratio) as occasion demands. (5) Perform “FSD adjustment – small focus 1” of maintenance scan by rotary scan,check its raw data by OPR of raw data analysis , and then, adjust the X-ray tube position by SW6 and SW7 of RXCPU PCB so that the output ratio in 444 ~ 453 channels becomes 0±3%. ( Move the X-ray tube backward by pressing SW6 when the output ratio is plus or move the X-ray tube forward by pressing SW7 when the output ratio is minus. ) The output ratio changes about 3% when moving the X-ray tube 10µm in case of small focus. 3-75 L2E-BW1444 5. FSD positioning (lateral direction) Rotary ・ Stationary HU during measurement = Not restricted 1 2 3 4 (6) Set the X-ray tube at a position where it is rotated clockwise by 90°as viewed from the top as illustrated below and set 21SW to “STATIONARY SCAN” side. 90° Figure (6) (7) Perform scan under the following conditions and measure the sum of the outputs at TP19 and TP20. Set the oscilloscope trigger to TP21. (about 0.5 - 1.0V) Measuring conditions Slice kV mA 1.25mm×4 120 150 Scan time Measuring point Stationary TP19 3sec. TP20 Table (7) Figure (7) (8) Adjust the lateral position of FSD so that the sum of the outputs of TP19 and TP20 becomes as shown below. -3.0V ≦TP19+TP20≦-1.5V Perform adjustment as specified in the following table by tube. FSFSD D検出器 Lateral move 囲 動範range 横移 A Adjusting固定ねじ (c) Loosen four fixing screws mounted on the lower face of FSD. (d) Position the FSD , and then, fix it by fastening the fixing screws. X線管tube X-ray コ リメータボッbox クス Collimator Figure (8) 3-76 L2E-BW1444 TP19+TP20 1 FSD adjustment -3.0V≦TP19+TP20≦-1.5V No adjustment is necessary Move the FSD rightward ( in the inserting direction 2 -1.5V<TP19+TP20<0V into the collimator box) , and fix it at the position where -3.0V ≦TP19+ TP20 ≦-1.5V. Move the FSD leftward ( in the direction coming 3 -3.0V>TP19+TP20 out of the collimator box) , and fix it at the position where -3.0V ≦TP19+ TP20 ≦-1.5V Table (8) In (8), if, as a result of FSD movement, the sum of the outputs at TP19 and TP20 does not fall within the following range: -3.0V≦TP19 + TP20 < -1.5V make fine adjustment of the position of the Collimator according to the following instruction. ・where -1.5V < TP19 + TP20 < 0 Move the Collimator rightward, viewed from the table side, within the adjustment value in 3.11 “Collimator position adjustment”, and check again the sum of the outputs at TP19 and TP20. ・where -3.0V > TP19 + TP20 Move the Collimator leftward, viewed from the table side, within the adjustment value in 3.11 “Collimator position adjustment”, and check again the sum of the outputs at TP19 and TP20. 6. FSD positioning ( longitudinal direction) Rotary ・ Stationary HU during adjustment = 45 ~ 65% 1 2 3 4 (9) After turning VR7 ( white mark for 120KV) (click sound is heard by turning it about 30 turns ) , turn it counterclockwise by 12 turns from the above condition, and fix FSD at the position where TP22 output (point A) becomes 0±2V after 1.5 seconds from turning on the X-ray in the scan under the following conditions. Measuring conditions Measuring Reference Slice kV mA Scan time point point 1.25mm×4 120 150 3 sec Table (9) TP22 0±2V (Oscilloscope Screen) 2[V] TP22 Point[A] 0[V] Set TP21 as trigger when checking the TP22 waveform. TP21 -2[V] 1.5sec 3sec X-Ray OFF X-Ray ON Figure (9)-a 3-77 L2E-BW1444 C:Scanner スキャside ナ側 D:Table テーside ブル側 Collimator コリメーbox タボックス Longitudinal direction 前後方 向 引 ねじ Pullき screw Push 押 しscrew ねじ Fixing 固定screw ねじ Fixing 固定screw ねじ Figure (9)-b (a) Loosen the fixing screws on both right and left sides of the FSD. (b) Increase the output voltage by moving the FSD in the C direction by the push screw when the TP22 output voltage is lower than -2V. Decrease the output voltage by moving the FSD in the D direction by the pull screw when the TP22 output voltage exceeds +2V. (c) Fix FSD at the position where TP22= 0±2V. 7. RTC board VR adjustment width check Rotary ・ Stationary HU during adjustment = 45~ 65% 1 2 3 4 (10) (The VR to be adjusted are VR9 ~ 11 (yellowish green) and VR6 ~ 8 (white). ) Make sure that the TP22 output after 1.5 seconds from turning on the X-ray in scan under the following conditions fluctuates when turning the corresponding VR. Measuring conditions Slice 1.25mm×4 kV Scan time 100 VR6 120 VR7 130 VR8 100 0.75mm×2 mA Check position 150 3 sec VR9 120 130 VR10 VR clockwise and it becomes lower than -1V when turning VR counterclockwise. VR11 Table (10) If the output does not become higher than 1Vunder certain KV condition in (10), adjust the FSD position again to increase the TP22 output and check (10) again. If the output does not become lower than -1Vunder certain KV condition in (11), adjust the FSD position again to decrease the TP22 output and check (10) again. 3-78 L2E-BW1444 8. RTC board VR adjustment 1 2 3 Rotary ・ Stationary HU during adjustment = 45~ 65% 4 (11) (The VR to be adjusted are VR9 ~ 11 (yellowish green) and VR6~ 8 (white). ) Adjust respective VR so that the TP22 output during scan shown in the following table becomes 0±0.3V at 100kV or 0±0.5V at 120kV and 130kV after about 1.5 seconds from turning on the X-ray when HU is 45 ~ 65%. (The output decreases when turning VR6~ VR11 counterclockwise. ) Measuring conditions Slice kV 1.25mm×4 2 3 point point Scan time VR6 120 VR7 130 VR8 150 3sec VR9 120 VR10 130 VR11 Table (11) 9. Correction to X-ray tube 1 Measuring 100 100 0.75mm×2 mA Adjusting tolerance ±0.3V ±0.5V TP22 ±0.3V ±0.5V Rotary ・ Stationary HU during adjustment = 45~ 65% 4 (12) Measure [a ~ d] under the following table for adjusting the FSD operation when the X-ray tube s set to 0°and 180°positions. ※Measurement of a,b,c,d (Measure the output after 1.5 seconds from X-ray irradiation) Measuring conditions X-ray tube position a 0° b 180° c 0° d 180° Slice kV mA Scan time 120 150 3sec Measuring point 1.25mm×4 TP22 0.75mm×2 Table (12) (13) Perform calculation & set the value as adjusting value by substituting[a~d] obtained by measurement into the table(13). Perform scan with each measuring condition on the table(13). Check the output of measuring point & adjust the value to be “Adjustment Value”. Perform scan again & check if the value becomes “adjustment value”. ※If the calculated value is within “0~0.2[V]”、set to the “Adjustment Value” 0.2[V]. 3-79 L2E-BW1444 Measuring conditions Slice kV mA Adjusting Measuring Calculated Adjusting Specified Scan time point point value value value 3sec VR2 1sec VR3 (a-b) +0 .3 V 2 1.25mm×4 120 (a-b) +0.3 ×0.6 V 2 150 TP21 3sec VR4 1sec VR5 (c-d) +0.3 V 2 0.75mm×2 (c-d) +0.3 ×0.6 V 2 +0.2V -0V +0.2V -0V +0.2V -0V +0.2V -0V Table (13) (Example) TP22 output T P22 の出力 0° 180° Calculation 算value 出値 = 1.2V ← ”a” -1.8V ← ”b” 1.2-(-1.8) 2 10. Correction to rotary operation 1 2 3 Adjust VR2 so that the TP21 output becomes 1.8 [V] ~ 2.0[V] in this case. +0.3V =1.8 Rotary ・ Stationary HU during adjustment = 45~ 65% 4 This adjustment is done under the rotary operation. Make sure that the probe ,etc. are removed without any skip. Start scanning when HU is 45 ~ 65%. (14) Turn on SW10 of RXCPUPCB ( LED26 (FSD_ON) lights to be turned on) and set 21SW to “NORMAL”. (15) Check the OPR value under the following conditions If the OPR value does not meet the specified value, adjust each VR shown in (11) again. (turning VR clockwise causes ‘%’ to alter in ‘+’ direction) Perform this scan by switching the kV and slice thickness from the protocol of FSD adjustmentlarge focus 1 of maintenance scan. Turn off PPC in case of 0.75mm x 2 scan. Measuring conditions Slice kV OPR value check mA Scan time No. of scans 150 3sec 3 times 150 3sec 3 times 100 1.25mm×4 120 130 100 0.75mm×2 120 130 OPR value in the 3rd scan shall be within a range of -0.6~ -1.8% OPR value in the 3rd scan shall be within ±6% *OPR : show P5-15(11) Table (15) 3-80 L2E-BW1444 (16) Measure the OPR value by executing maintenance scan under the following. Protocol conditions and assume [f~i] be the obtained values. (All these values shall be the OPR values in the final scan. ) プロトコname ル名 Protocol スscan キャンtime タイム No.スキofャンscans 回数 f FSD adjustment-large focus2 1.0[s] 5 5 回 times g FSD adjustment-large focus3 0.8[s] 5 5 回 times h FSD adjustment-small focus2 times 1.0[s] 5 5 回 i FSD adjustment-small focus3 times 0.8[s] 5 5 回 Table (16) (17) Adjust relevant VR so that [f~ i] obtained in (16) conform to the following specified values respectively. The DSW setting is OFF-Defalut. When adjustment is impossible even if you turn volume, turn on the DSW. ① [f] → Adjust VR12 so that -0.6% ≧f ≧ -1.8%. (See the following table) OPR value -0.6%≧f≧-1.8% VR12 adjustment No adjustment is necessary When No. 6 of DSW2 is turned off Adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment -1.8%>f cannot be done even when turning VR fully When No. 6 of DSW2 is turned on clockwise, turn off No.6 of DSW2 , and adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment f>-0.6% cannot be done even when turning VR fully When No. 6 of DSW2 is turned off clockwise, turn off No.6 of DSW2 , and adjust VR by turning it counterclockwise. When No. 6 of DSW2 is turned on Adjust VR by turning it counterclockwise. Table (17)-① ② 「g」→ Adjust VR13 so that -0.6% ≧ g ≧ -1.8% (See the following table) OPR value -0.6%≧g≧-1.8% VR13 adjustment No adjustment is necessary When No. 8 of DSW2 is turned off Adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment -1.8%>g When No. 8 of DSW2 is turned on cannot be done even when turning VR fully clockwise, turn off No.8 of DSW2 , and adjust VR by turning it counterclockwise. Adjust VR by turning it counterclockwise. If g>-0.6% When No. 8 of DSW2 is turned off adjustment cannot be done even when turning VR fully clockwise, turn off No.8 of DSW2 , and adjust VR by turning it counterclockwise. When No. 8 of DSW2 is turned on Adjust VR by turning it counterclockwise. Table (17)-② 3-81 L2E-BW1444 ③ 「h」→ Adjust VR14 so that 6% ≧ h ≧ -6% (See the following table) OPR value 6%≧h≧-6% VR14 adjustment No adjustment is necessary When No. 5 of DSW2 is turned off Adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment -6%>h cannot be done even when turning VR fully When No. 5 of DSW2 is turned on clockwise, turn off No.5 of DSW2, and adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment h>6% cannot be done even when turning VR fully When No. 5 of DSW2 is turned off clockwise, turn off No.5 of DSW2, and adjust VR by turning it counterclockwise. When No. 5 of DSW2 is turned on Adjust VR by turning it counterclockwise. Table (17)-③ ④ 「i」→Adjust VR15 so that 6% ≧ i ≧ -6%. (See the following table) OPR value 6%≧i≧-6% VR15 adjustment No adjustment is necessary When No. 7 of DSW2 is turned off Adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment -6%>i When No. 7 of DSW2 is turned on cannot be done even when turning VR fully clockwise, turn off No.5 of DSW2, and adjust VR by turning it counterclockwise. Adjust VR by turning it clockwise. If adjustment i>6% When No. 7 of DSW2 is turned off cannot be done even when turning VR fully clockwise, turn off No.5 of DSW2, and adjust VR by turning it counterclockwise. When No. 7 of DSW2 is turned on Adjust VR by turning it counterclockwise. Table (17)-④ Perform scan in (16) again when VR12 ~ VR15 was adjusted, and make sure that the values in f ~ i are within a range where no adjustment is necessary. (18) Make sure that SW10 of RXCPU PCB is turned on again (LED26 (FSD_ON) lights). 3-82 L2E-BW1444 3.13 Detector side collimator (Post Patient Collimator - PPC) position adjustment The positioning method of the detector side collimator (Post Patient Collimator hereafter referred to as PPC) used for sub-milli slice measurement is described below. Execute this item without fail in the following cases ・ When the collimator was removed in special disassembly, etc., before delivering the system ; ・ When the PPC was mounted again after removing it once and mounting the detector in case of the system where the detector was removed in special disassembly, etc. before delivering the system Before starting this work, it is necessary that PPC has been mounted in 1.6.2 and especially, coarse adjustment of PPC has been done in 1.6.2 (5). (1) Set the scan conditions on the maintenance scan conditions setting panel as shown below and then, press Confirm switch. Items Set values 1 75 120 3 0.75mm 2i STAY 0 Any Auto Any 2 Manual First Any Any OFF Any OFF OFF OFF OFF OFF OFF 3 0 OFF OFF OFF ON No. of scans Tube current (mA) Tube voltage (kV) Scan time (s) Image slice thickness (mm) Image mode Direction Table index (mm) F.O.V. (mm) Inter-scan time (s) Inter-sequence delay (s) No. of images Start mode Reconstruction mode Upper WW / WL Lower WW / WL Stationary scan Filter/image filter B.H.C. B.G.C. Phantom correction ADPP (measurement) ADPP (reconfiguration) Scattered ray correction Variation code Bias CT number adjustment Large attenuation correction Image ring correction P.P.C. (Note) For the items indicated as [option], their set values at that time may be used as they are. 3-83 L2E-BW1444 (2) After the system has been set to be ready for measurement, pressing START START key flickers. Perform scanning by key. (3) Display the Raw Data scanned in (2) above by Raw Data analysis, and select OPR out of the analysis menu (See the left figure shown below) Another window opens to indicate the OPR calculation results (See the right figure shown below). Set the OPR values (%) of channels 11 ~ 20 as OPR (11-20), the OPR values (%) of channels 444 ~ 453 as OPR (444 ~ 453) and the OPR values (%) of channels 877 ~ 886 as OPR (877 ~ 886) (For the definition of OPR values, refer to 3.12 Focus shift detector (FSD) adjustment (7)). (4) PPC position adjustment ends when each OPR value is as specified below -5% ≦OPR (444 - 453) ≦+5% and |OPR (11~ 20) - OPR ( 877 ~ 886)|≦5% 3-84 L2E-BW1444 (5) Adjust the PPC position finely according to Table 3.13.1 and < Fine adjusting method of PPC position>. OPR (444 - 453) changes about 5% when moving the stopper by 0.1mm in Fig.3.13.1. Table 3.13.1 OPR value codes and stoppers to be adjusted OPR value (%) codes OPR(11-20) OPR(444-453) OPR(877-886) Positive Positive Positive Positive Positive Negative Positive Negative Positive Magnitude relation between Stoppers to be adjusted and moving directions |OPR (11-20)|and|OPR (877-886)| (as viewed from the table side) |OPR(11-20)| ≧ |OPR(877-886)| Move the left stopper toward the table |OPR(11-20)| < |OPR(877-886)| Move the right stopper toward the table - Move the left stopper toward the table |OPR(11-20)| ≧ |OPR(877-886)| Move the right stopper toward the rotary panel |OPR(11-20)| < |OPR(877-886)| Move the left stopper toward the rotary panel Positive Negative Negative - Move the right stopper toward the rotary panel Negative Positive Positive - Move the right stopper toward the table |OPR(11-20)| ≧ |OPR(877-886)| Move the right stopper toward the table Negative Positive Negative |OPR(11-20)| < |OPR(877-886)| Move the left stopper toward the table - Move the left stopper toward the rotary panel |OPR(11-20)| ≧ |OPR(877-886)| Move the left stopper toward the rotary panel |OPR(11-20)| < |OPR(877-886)| Move the right stopper toward the rotary panel Negative Negative Positive Negative Negative Negative 3-85 L2E-BW1444 <Fine adjusting method of PPC position> (1) Make sure that the PPC is set to the rotary panel side (position just above the X-ray incident window of detector). If the PPC is not positioned on the rotary panel side, move the PPC toward the rotary panel electrically by scanning under the condition in step (1). (2) Mount the dial gauge (attachment to scanner) by M6 x 14L sems screw. Set the dial gauge scale division to zero. (See Fig.3.13.1) (3) Adjust the collimator finely up to the aimed position according to the following procedure, while reading the dial gauge. 1) Loosen bolts (*1) and nut (*2). 2) Position the collimator by moving the stopper while turning the bolt (*3) clockwise or counterclockwise. For the relation between the turning direction of the bolt and the moving direction of the collimator, see Table 3.13.2. 3) Fasten bolt (*1). Then, fasten nut (*2) to fix the collimator position. 4) Make sure that the dial gauge value indicates an aimed value. If deviated, repeat steps (1) ~ (3) again. Dial gauge (attached to scanner) (Mounted by M6 x 14L sems screw) Rotary panel Rotary panel direction Table direction M4 x 40L hexagon socket head bolt (*3) Stopper M4 nut (*2) 2-M4 x 12L hexagon socket head bolt (*1) (spring, plain washer) Fig. 3.13.1 Table 3.13.2 Moving direction of collimator 1 Table direction 2 Rotary panel direction (*3) Turning direction of bolt Rotating direction of pointer of dial gauge Clockwise Counterclockwise rotation (Decreasing direction of numeric value) Counterclockwise Clockwise rotation (Increasing direction of numeric value) 3-86 L2E-BW1444 (6) Set the scan conditions on the maintenance scan conditions setting panel as shown below and then, press Confirm switch. Items No. of scans Tube current (mA) Tube voltage (kV) Scan time (s) Image slice thickness (mm) Image mode Direction Table index (mm) F.O.V. (mm) Inter-scan time (s) Inter-sequence delay (s) No. of images Start mode Reconstruction mode Upper WW / WL Lower WW / WL Stationary scan Filter/image filter B.H.C. B.G.C. Phantom correction ADPP (measurement) ADPP (reconfiguration) Scattered ray correction Variation code Bias CT number adjustment Large attenuation correction Image ring correction P.P.C. Set values 1 75 120 3 0.75mm 2i STAY 0 Any Auto Any 2 Manual First Any Any OFF Any OFF OFF OFF OFF OFF OFF 3 0 OFF OFF OFF ON (Note) For the items indicated as [option], their set values at that time may be used as they are. (7) After the system has been set to be ready for measurement, pressing START START key flickers. Perform scanning by key. (8) Display the Raw Data scanned in (2) above by Raw Data analysis, and select OPR out of the analysis menu (See the left figure). (9) Make sure that each OPR value is as specified below -5% ≦OPR (444 - 453) ≦+5% and|OPR (11~ 20) - OPR ( 877 ~ 886)|≦5% If the OPR value is not within the above range, return to step (5). Be sure to remove the dial gauge mounted in step (5) after the end of this work. 3-87 L2E-BW1444 3.14 Phantom calibration 3.14.1 Procedure Perform X-ray warm up (or air calibration) and phantom calibration measurement under the mounted conditions of the scanner corn cover, lift cover, and X-ray transmission window for executing calibration under the X-ray conditions during user’s use. 3-88 L2E-BW1444 3.14.2 Mounting method of phantom mounting bracket (1) Mounting (a) Turn the knob counterclockwise and loosen fixing plate L under the condition shown below. Knob Fixing plate Fixing plate L (b) Lift up the phantom mounting bracket and insert the phantom mounting in the direction until the end bracket touches the tabletop end as illustrated below. Tabletop End bracket 3-89 L2E-BW1444 (c) Adjust the fixing plate R to the right edge of tabletop and clip the tabletop by sliding fixing plate L in the direction. Turn the knob to fix the phantom mounting bracket. If the bracket is not fixed properly, the bracket may possibly fall down during the calibration resulting in the damage of scanner cover. Knob Fixing plate Fixing plate L Table top Fixing plate R 3-90 L2E-BW1444 (2) Mounting of phantom Insert the hole portion of the phantom onto the shaft of the phantom mounting bracket. Shaft Phantom Hole (3) Adjustment of phantom position Adjust the phantom position by moving the fine move mounting base. The fine move mounting base moves toward D, U, L, or R when turning the DU adjusting dial or LR adjusting dial in the indicating directions on the DU nameplate and LR nameplate. Dial indication Moving quantity Between bold lines (between numeric values) 0.1mm Between lines (minimum scale division) 0.02mm DU nameplate DU adjusting dial Fine move mounting base LR adjusting dial U L R D LR nameplate 3-91 L2E-BW1444 (4) Phantom angle adjustment (a) Adjust the phantom angle by turning the knob clockwise to tilt the phantom as illustrated below. Knob (b) Keep the phantom perpendicular to the scan face by tilting the phantom with the knob by a deflection component of the table top as illustrated below. Horizontal direction Tabletop Angle adjustment Deflection of table top 3-92 L2E-BW1444 (5) Removal Loosen the knob to make the fixing plate L free to move, and slide the fixing plate L outward. Remove the phantom mounting bracket from the tabletop. Knob Fixing plate L Tabletop Fixing plate L For storing the phantom mounting bracket, store it after grounding the insertion part of the phantom as illustrated below. 3-93 L2E-BW1444 3.14.3 X-ray detection system warming up Table 3.14.1 shows the required time from turning the power supply for temperature controller of Detector and preamplifier power supply to be ready for calibration. With a longer warm-up time, better results may be expected. When it is necessary for any work to turn off the power supply, the off time should be as short as possible. Table 3.14.1 Required time from turning on the power supplies to be ready for calibration Power supplies Required time to be ready for calibration Power supply for temperature controller of Detector Preamplifier power supply 3.14.4 Longer than 4 hours after power supply has been turned on. (Keep power been fed normally.) Longer than 2 hours after turning on the power supply. X-ray tube warm up and air calibration If it has passed longer than 2 hours after the latest X-ray exposure, perform continuous operation [(1) described later] of X-ray tube warm up and air calibration before starting the other measurement. If the phantom calibration is done within 2 hours after the latest X-ray exposure, observe the “caution” in (1.5). The X-ray tube warm up and air calibration can be operated systematically by Warm up in the measurement main menu on the operation panel. Since the air calibration is executed in protocol mode during continuous operation of warm up and air calibration, the air calibration conforming to the [combinations of X-ray tube voltage and slice thickness] being entered in the measurement protocol. If air calibration is necessary under the other conditions, execute it in the calibration select mode ((2) described later). 3-94 L2E-BW1444 (1) Continuous operation of X-ray tube warm up and air calibration. (1.1) Make sure that none of tabletop, phantoms, etc. is present in the gantry aperture. (1.2) If the maintenance scan is not displayed on the measurement main menu on the operation panel, display it according to 5.1. and enter the patient newly. Be careful since the maintenance warm up cannot be selected in (1.4) without this display. (1.3) Select Warm up in the measurement main menu on the operation panel, and display the mode selection menu. (1.4) Select the maintenance warm up and click Confirm button. Air calibration before phantom calibration must be executed under the stable condition (normal condition) of the elongation of the anode axis of X-ray tube. Maintenance warm up is executed to realize the normal condition for executing air calibration. Keep the anode accumulated heating quantity to be within a range of 43 ~ 75% after executing the maintenance warm up until all necessary phantom calibration measurement ends. (1.5) When the system is ready for measurement, switch. START switch flickers. Press the START Now, the X-ray tube warm up and air calibration in protocol mode are executed automatically. If air calibration under the conditions other than protocol mode is necessary, perform necessary air calibration in the calibration select mode. If it has passed shorter than 2 hours after the latest X-ray exposure, execute the following procedure. (1) Carry out 120kV, 200mA, 2s, 1.25mm×4 slice thickness continuous scan until anode accumulated heating quantity ≧90%. (2) Continue scanning after (1) to continue the condition of anode accumulated heating quantity ≧85% for longer than 12 minutes before executing (3). (3) Start air calibration when the anode accumulated heating quantity has become within a range of 43 ~ 58% after about 10 minutes pause. Keep the anode accumulated heating quantity within a range of 43~ 75% hereafter until all necessary phantom calibration ends. 3-95 L2E-BW1444 (2) Air calibration (select mode) (2.1) Make sure that none of tabletop, phantoms, etc. is present in the gantry aperture. (2.2) Select Warm up in the measurement main menu on the operation panel, and display the mode selection menu. (2.3) Select a desired tube voltage and slice thickness, and click (2.4) After the system has become ready for measurement, START 3.14.5 Confirm START button. switch flickers. Press the switch. Now, the air calibration is done under the selected conditions. Phantom position adjustment and phantom calibration Be sure to carry out the air calibration under the relevant conditions before starting the phantom calibration. Perform the air calibration after setting the X-ray tube to the normal condition according to 3.14.4 (1). Keep the tube rating to be within a range of 43 ~ 75% until all necessary phantom calibration measurement ends. (1) If [Adjust / Calib.] is not displayed in the measurement main menu, display [Adjust / Calib.] button and newly enter the patient according to 5.1. (2) Adjust the calibration phantom position. (2.1) Click [Adjust / Calib.] button of the main menu to display the phantom calibration CT value adjustment screen. (2.2) Select [Phantom calibration] tab of the phantom calibration CT value adjustment screen, and display the [phantom position adjustment] face. 3-96 L2E-BW1444 (2.3) Set the phantom to be measured to the center of the gantry opening (Use phantoms sequentially starting with smaller ones). Set the projection line (for slice face) of the projector to be in parallel with the phantom by the angle adjusting knob of the phantom mounting bracket. Four kinds of phantoms are provided as shown in Table 3.14.1. Table 3.14.2 Kinds of phantoms used for phantom calibration Phantom Measuring F.O.V of phantom position adjusting scan (mm) (Not changeable) φ230 polyethylene 250 φ260 polyethylene 300 φ350 polyethylene 420 φ410 polyethylene 420 (2.4) Select a phantom size to be measured out of the pull down list and click (2.5) After the system has become ready for measurement, START START Confirm button. switch flickers. Press the switch Now, the phantom position adjustment scan of the selected phantom is executed. (2.6) When the phantom position adjustment scan is done, a phantom moving indicating value is displayed in the moving quantity indicating column. (2.7) Move the phantom by using the dial of the phantom mounting bracket according to the displayed phantom moving indicating value. (2.8) Repeat steps (2.4) ~ (2.7) until the phantom moving indicating value becomes less than 0.1mm in the vertical and lateral directions. When this value becomes less than 0.1mm in the vertical and lateral directions, the [Measurement] button is displayed as shown in the following figure, and “within a specification” is displayed below the moving quantity indicating column. 3-97 L2E-BW1444 (3) Measure the phantom calibration. (3.1) After the end of phantom position adjustment, click [Measurement] button to display the [phantom calibration measurement] face as shown in the following figure. (3.2) Select the measuring condition by clicking the cell of phantom calibration measurement executing condition (combination of X-ray tube voltage and collimation), and click Confirm button. A check mark is indicated in the clicked measuring condition cell. (3.3) After the system has been ready for measurement, START switch flickers. Press START switch, and the phantom calibration is executed under the selected condition. Perform the phantom calibration calculation after measuring the phantoms of all diameters (See Table 3.14.2). Calculation cannot be executed every phantom. 3-98 L2E-BW1444 (4) Perform phantom calibration calculation. (4.1) Display the following [phantom calibration calculation] by clicking [Calculation] button. (4.2) Click the cell of executing conditions of calculation (combination of X-ray tube voltage and collimation). A check mark is displayed in the clicked cell. If a non-measurement condition is present, the calculation condition is displayed gray and conditions are not selectable. (4.3) Calculate the phantom calibration data by clicking [Calculate] button. If an error occurs halfway in the course of calculation, the following window appears and the phantom calibration data calculation is stopped. An error display example If an error occurs halfway in the course of calculation, measure the phantom calibration under the relevant condition from the beginning, and calculate again. For erasing the error display window, click 3-99 OK button by the left key of mouse. L2E-BW1444 (4.4) At the end of calculation, make sure that the data selected in (4.3) are displayed as “Calculated ”in the [Remarks] column of the date and time information list, and terminate the measurement and calculation of phantom calibration by clicking [Exam. End] button by the left click of mouse. Date and time information list The following contents are displayed in the remarks column of the day and time information list depending upon the calculation date and time and measuring date and time. “Un-measuring” : Data not measured yet “Calculation required” : Data measured but not calculated yet “Calculated” : Data whose calculation date and time are later than the measurement date and time “Completed” : Calculated data (5) Check phantom calibration data (5.1) After clicking [System / Exit] button of the main menu, select [Maintenance] →[Services]→[Raw Data Analysis] out of the displayed menu, and start Raw Data analysis. 3-100 L2E-BW1444 (5.2) Select [Open]→[Cal. Data] out of the [File] menu on the Raw Data analysis window, and display the Cal. Data Selection dialog. (5.3) Select Phantom Cal. Data of Cal. Data Selection dialog, select the X-ray tube voltage condition of Cal. Data to be displayed, and display the PLP setting dialog by clicking [OK] button. 3-101 L2E-BW1444 (5.4) Select a desired display condition in PLP Setting screen, and click [OK] button. Selected phantom calibration data are displayed in PLP window. (5.5) After the selected data graph has been displayed on the PLP window, click Max./Min. of PLP window by the left key of mouse to set the graph to the Max./Min. display condition. A display example of phantom calibration data alpha 3-102 L2E-BW1444 A display example of phantom calibration data beta (5.6) Make sure that no abnormal irregularity is detected in the calibration data graph. If any, repeat the procedure starting with air calibration (3.14.4). If abnormality cannot be eliminated yet, check the hardware. An example of abnormal data (5.7) Select [Exit] from [File] menu to close the Raw Data analysis window. (6) Execute all necessary phantom calibration. 3-103 L2E-BW1444 3.15 CT number adjustment 3.15.1 Procedure (1) Scan all water phantoms (See table 3.15.1) by using the CT number adjustment scan under the combined conditions of X-ray tube voltage to be adjusted and the collimation. (2) Calculate CT number adjusting parameter by inputting the combined conditions of X-ray tube voltage measured in (1) and collimation in [Calculate] of CT number adjustment calculation program. (3) Check the CT number at the center of water phantoms by executing the image reconstruction in the post reconstruction using Raw Data of the water phantoms measured in CT number adjustment scan, and check the CT number at the water phantom center (Radius of measuring R.O.I is 75 pixels in X and Y). If the CT number is deviated from 0±4, repeat the procedure from the CT number adjustment scan of the corresponding water phantom. Start Scan water phantom by “CT number adjustment scan” Calculate CT number adjustment parameter by “CT number adjustment calculation” Check CT number by reconstruction N.G. O.K. End Fig.3.15.1 Execute CT number adjustment scan under the mounted condition of scanner cone cover, lift cover, and X-ray transmission window for the purpose of adjusting the CT number under the X-ray conditions during user’s use. 3-104 L2E-BW1444 3.15.2 CT number adjustment scan (1) If [CT number adjustment scan] is not displayed in the measurement main menu on the operation panel, display [CT number adjustment scan] according to 5.1 and newly enter the patient. (2) Adjust the water phantom position. (2.1) Click [Adjust / Calib.] button of the main menu to display the phantom calibration CT number adjustment screen. (2.2) Select [CT No. Adjust] tab of the phantom calibration / CT number adjustment screen, and display the [phantom position adjustment] face. (2.3) Set the phantom to be measured to the center of the gantry opening (Use phantoms sequentially starting with smaller ones). Set the projection line (for slice face) of the projector to be in parallel with the phantom by the angle adjusting knob of the phantom mounting bracket. Four kinds of phantoms are provided as shown in Table 3.15.1. Table 3.15.1 Kinds of phantoms used for phantom calibration Phantom Measuring F.O.V of phantom position adjusting scan (mm) (Not changeable) φ210 water 250 φ230 water (*1) 300 φ305 water 420 φ380 water (*2) 420 (*1) Used only for 0.75mm x 2 and 1.25mm x 4 collimation. (*2) Not used for 0.75mm x 2 and 1.25mm x 4 collimation. (2.4) Select a phantom size to be measured out of the pull down list and click Confirm button. 3-105 L2E-BW1444 (2.5) After the system has become ready for measurement, the START START switch flickers. Press switch. Now, the phantom position adjustment scan of the selected phantom is executed. (2.6) When the phantom position adjustment scan is done, a phantom moving indicating value is displayed in the moving quantity indicating column. (2.7) Move the phantom by using the dial of the phantom mounting bracket according to the displayed phantom moving indicating value. (2.8) Repeat steps (2.4) ~ (2.7) until the phantom moving indicating value becomes less than 5mm in the vertical and lateral directions. When this value becomes less than 5mm in the vertical and lateral directions, the [Measurement] button is displayed as shown in the following figure and “within a specification” is displayed below the phantom moving indicating column. 3-106 L2E-BW1444 (3) Perform CT number adjustment scan. (3.1) After the end of phantom position adjustment, click [Measurement] button to display the [CT number adjustment measurement] face as shown in the following figure. (3.2) Select the measuring condition by clicking the cell of CT number adjusting scan executing condition (combination of X-ray tube voltage and collimation), and click [Confirm] button. A check mark is indicated in the clicked measuring condition cell. Non-required measuring conditions are displayed light to be not selectable. (3.3) After the system has been ready for measurement, START switch flickers. Press START switch, and the CT number adjustment scan is executed under the selected condition. Perform the CT number adjustment calculation after calculating all necessary water phantoms by CT number adjustment scan in the X-ray tube voltage and collimation to which the execution of CT number adjustment is desired. 3-107 L2E-BW1444 3.15.3 Perform CT number adjustment calculation (1) By clicking [Calculation], the following window is displayed for selecting the X-ray tube voltage and collimation. (2) Click the cell of executing conditions of calculation (combination of X-ray tube voltage and collimation). A check mark is displayed in the clicked cell. If a non-measurement phantom is present, the calculation condition is displayed gray to be not selectable. (3) Calculate the CT number adjusting parameter by clicking [Calculate] button. If an error occurs halfway in the course of calculation, the following window appears and the CT number adjusting parameter calculation is stopped. An error display example If an error occurs halfway in the course of calculation, measure the CT number adjustment scan under the relevant condition from the beginning, and calculate again. For erasing the error display window, click 3-108 OK key by the left switch of the mouse. L2E-BW1444 (4) At the end of calculation, make sure that the data selected in (3) are displayed as “Completed ”in the [Remarks] column of the day and time information list, and terminate the CT number measurement calculation by clicking [Exam. End] button by clicking the left key of mouse. The following contents are displayed in the remarks column of the day and time information list depending upon the calculation date and time and measuring date and time. “Un-measuring” : Data not measured yet “Calculation required” : Data measured but not calculated yet “Calculated” : Data whose calculation date and time are later than the measurement date and time “Completed” : Calculated data (5) Check the CT number. Check the water CT number after CT number adjustment calculation processing by using Raw Data of the water phantom measured by CT number adjustment scan. (5.1) Perform post reconstruction by using Raw Data (measured by CT number adjustment scan in 3.15.2) of the water phantom requiring the water CT number check, and a CT number -adjusted water phantom image is obtained. Set the post reconstruction conditions as specified below. Items Set values FOV Mag FOV.X, FOV.Y Filter BGC, BHC Recon. Type See Table 3.15.1 1 0, 0 4 (Abdomen Standard) OFF Normal 3-109 L2E-BW1444 (5.2) Measure a mean CT number (MEAN) in ROI by setting the circular R.O.I of 75 pixels in radius to the center of the post-reconstructed water phantom image, and make sure that the mean value is 0.0±4.0. If this mean value does not enter the 0.0±4.0 range, repeat the above operation starting from 3.15.2 [CT number adjustment scan]. If the mean value does not enter the 0.0±4.0 range even if the operation is repeated from the CT number adjustment scan, execute the air calibration under the relevant condition (combination of X-ray tube voltage and collimation), and then, perform the CT number adjustment scan, CT number adjustment calculation, and CT number check. (6) Perform the CT number adjustment scan, CT number adjustment calculation, and CT number check about all necessary conditions (combinations of X-ray tube voltage and collimation). To make it possible to recover the system at an earlier time in case of software crash, be sure to make service backup after phantom calibration or CT number adjustment. 3-110 L2E-BW1444 3.16 Interphone (intercommunication) check (1) Operation room → scanner room (1-1) Microphone volume adjustment The sound in the operation room is sent to the scanner room by pressing TALK Adjust the sound volume by “TALK” control knob on the operation panel of the interphone box or by “OC→ GANTRY” control knob on the rear panel, if “TALK” control knob is not mounted on the operation panel. (1-2) Auto voice volume adjustment Adjust the auto volume by “A VOICE” control knob on the operation panel or by “A.V OC→GANTRY” control knob on the rear panel, if “A. VOICE ” control knob is not mounted on the operation panel. Adjust the control knob to an-easy-use position. By pressing TALK , the sound from the scanner room is automatically reduced to a low tone. . (2) Scanner room → operation room Adjust the sound volume by “AUDIO” or “VOL” control knob on the operation panel of the interphone box Make sure that the sound volume is adjustable. 3-111 L2E-BW1444 3.17 Institution info. settings Set the institution information according to the following procedure. (1) A institution info. settings window is displayed by selecting [System / Exit] – [Settings / Registration] – [Institution Info. Settings]. (2) Input the hospital name, department name, and station name, and then, click [Complete] button. 3-112 L2E-BW1444 3.18 Computer name change If Presto system is connected to two or more LAN, change the computer name of the 2nd and subsequent systems according to the following procedure. Respective systems can be identified from each other by changing the computer name. Settings including LAN card setting for DICOM transmission and filming, and LAN card setting for remote service will not be restored. Be sure to reconfigure them after renaming the computer name. (1) Disconnect the LAN cable , if it is connected to the 28CN connector on the upper face of the computer. (2) Turn on the system power supply. (3) Switch the mode to the system maintenance mode. (See 5.11) (4) Select [Computer name change] from the start menu. (5) Select a desired computer name to be changed from the computer name selection area on the computer name and security identifier change screen. Be careful not to overlap the computer name with that of the other Presto system. Computer name selection area Present computer name 3-113 L2E-BW1444 (6) A confirmation screen is displayed. Click [Yes] button if you want to continue the computer name change processing. By clicking [Yes], the computer name change is started automatically. It takes about 10 minutes to change the computer name. (7) At the end of the computer name change, an information dialog is displayed Click [OK] button. (8) Connect LAN cable to 28CN connector. (9) Start the CT system. (Refer to 5.11 <Reference>) 3-114 L2E-BW1444 3.19 Communication setting Perform the following entry and setting for executing DICOM transmission and filming. 3.19.1 LAN card setting (1) Turn on the maintenance mode. (See 5.1) (2) Restart the CT unit in the system maintenance mode. (See 5.11) (3) Select [My Network Places] icon on the screen, and right-click the icon. Then, click [Properties] of pop-up menu. (4) A network connection screen is displayed. Select local area connection n(*) (device name “corega Fether PCI-TXS...” , “corega Gether PCI-T32....” or “CG-LAPCIGT2…” , right-click the connection, and click [Properties] of pop-up menu. (5) A local area connection n (*) property screen is displayed. Select Internet Protocol (TCP/IP) of the General tab. (6) Click [Properties] button on the property screen. (7) An Internet Protocol (TCP/IP) property screen is displayed. Select “Next IP address is used (S)”, and set the IP address, subnet mask, and default gateway. * For these values, ask the network administrator of this facility. * Do not use “172.18.1.*” for the IP address, which is already used between the image processor and scanner. * Don’t change any setting of DNS, WIND, DHCP routing. (8) Click [OK] button to close the property window of Internet Protocol (TCP/IP). (9) Click [Close] button to close the property window of the local area connection n(*). (10) Click [Start] button to select [Start the CT system] from the Start menu, and restart the CT unit. * ‘n’ denotes numerical value. This value varies, depending on the LAN card which is built in the CT unit. 3.19.2 Deletion of DICOM transmission software This operation is not required when no entry of DICOM transmission software exists. (1) Set the DICOM transmission software CD-ROM in the CD drive. (2) Click [Image list] button of the main menu and start the image explorer. (3) Select [Tools] [Option Register…] from the menu of image explorer to display the option entry dialog box. (4) Input “e:\Setup” to the name column of option entry dialog box, and click [OK] button while pressing Ctrl key. 3-115 L2E-BW1444 (5) A dialog box appears, asking to exit the image explorer. Click [Yes] button. (6) When the image explorer ends, a DICOM transmission software entry screen is displayed. Click [Next] button. (7) A file deletion check dialog screen is displayed. Click [OK] button. 3.19.3 Entry of DICOM transmission software (1) Set the DICOM transmission software CD-ROM in the CD drive. (2) Click [Image list] button of the main menu and start the image explorer. (3) Select [Tools] [Option Register…] from the menu of image explorer to display the option entry dialog box. (4) Input “e: \Setup” to the name column of option entry dialog box, and click [OK] button while pressing Ctrl key. 3-116 L2E-BW1444 (5) A dialog box appears, asking to exit the image explorer. Click [Yes] button. (6) When the image explorer ends, a DICOM transmission software entry screen is displayed. Click [Next] button. (7) After the entry of DICOM transmission software has been entered, click [Finish] button. (8) Start the image explorer by clicking [Image list] button of the main menu, and make sure that the [DICOM Trans.] icon is displayed. 3-117 L2E-BW1444 3.19.4 Setting of DICOM transmission software (1) Select an optional patient in the image explorer. (2) Display the DICOM transmission screen by clicking [DICOM Trans.] icon on the image explorer or by selecting [File] [Send], and [DICOM Trans.] from the menu. (3) Click [Settings] button to display the DICOM transmission setting screen. (4) Click [Destination] tab. (5) Click [Add] button in the transmission destination setting column. (6) After setting the following items, click [OK] button, and the transmission destination name is added to the transmission destination list as illustrated below. Item names Dest. Name AE Title IP Address Port No. Description Set the transmission destination name. This set value is either the destination host name or optional name. Set an AE title of the transmission destination. For this set value, ask the network administrator of the facilities. Set the IP address of the transmission destination. For this set value, ask the network administrator of the facilities. Set a port number of the transmission destination application. For this set value, ask the person in charge of the transmission destination application maker. (7) Set the following item in local machine setting. Item names AE Title Description Set the AE title of CT unit Presto or Hyper Quick Net. For this set value, ask the network administrator of the facilities. (8) Click [OK] button. * Answer “104” if the transmission destination maker asks you the transmission source (CT unit side) application port number. In practice, however, OS assigns an optional port number. 3-118 L2E-BW1444 3.19.5 Setting of storage commitment software The following shows the steps required to execute storage commitment (function to ensure that images transferred by DICOM transmission are stored normally at the destination) (1) Choose any one of the patients on the image explorer. (2) Choose a menu, [File], [Send], and [Storage Commit] on the image explorer. Or choose [File], [Send], and [Storage Commit] to display storage commitment screen. (3) Click [Settings] to display storage commitment setting screen. (4) Click [Destination] tab. (5) Click [Add] button in the destination setting column. (6) Set the following items, and click [OK] button. Item names Dest. Name AE Title IP Address Port No. Description Set the AE title of the destination. This set value is either the host name, to which storage commitment is to be transmitted, or optional name. Set the IP address of the destination. For this set value, ask the network administrator of the facilities. Set the IP address of the destination. For this set value, ask the network administrator of the facilities. Set the port number of destination application. For the set value, ask a person in charge of destination application maker. (7) Set the following item in local machine setting. Item names AE Title Port No. Description Set the AE title of CT unit Presto or Hyper Quick Net. For this set value, ask the network administrator of the facilities. Set the port number of CT unit Presto or Hyper Quick Net. Set this value to 104. (8) Click [OK] button. This is the end of setting storage commitment. 3-119 L2E-BW1444 3.19.6 Setting of filming software (1) Select any one of the patients in the image box. (2) Display the image viewer by clicking [Image viewer] icon of the image explorer. (3) Display the filming screen by clicking [Filming] button on the image viewer. (4) In the menu on the filming screen, choose [Tools/Option] to display filming option screen. (5) Click [Printer] tab. (6) Click [Add] button in the printer setting column. (7) After setting the following items, click [OK] button, and the print name is added to the printer list as shown in the following figure. Item names Printer Name AE Title IP Address Port No. Description Set a printer name. This set value is either printer host name or optional name. Set an AE title of the printer. For this set value, ask the network administrator of the facilities. Set the IP address of the printer. For this set value, ask the network administrator of the facilities. Set a port number of the printer application. For this set value, ask the person in charge of the printer maker. (8) Set the following item in local machine setting. Item names AE Title Description Set the AE title of CT unit Presto or Hyper Quick Net. For this set value, ask the network administrator of the facilities. (9) Click [OK] button. 3-120 L2E-BW1444 3.19.7 Entry of SMPTE pattern (1) Set the application install disk/SMPTE pattern CD in the CD drive. (2) Open the image explorer by clicking [Image list] button. (3) Open the MO drive selection dialog by clicking [Tools | Select MO Drive] of the image explorer. (4) After clicking the MO drive 2 text box, select “e:” drive from the drive selection box. (5) Click [OK] button of the select MO drive dialog → “e:” is displayed on the image explorer. (6) Select “e-image-SMPTE TEST PATTERN” from the image explorer. (7) Click the right mouse button and select [Send | HD] from the pop up menu. SMPTE TEST PATTERN image is sent to HD. (8) Open the select MO drive dialog by clicking [Tools | Select MO Drive] of the image explorer. (9) Blank the MO drive 2 text box, and click [OK] button. →“e :”drive is deleted form the image explorer. (10) Take the application install disk/SMPTE pattern CD out of the CD drive. 3-121 204 L2E-BW1444 Section 4 Reference data 4.1 Stationary scan (1) If [Maintenance scan] is not displayed in the measurement main menu on the operation panel, display it according to 5.1.1. (2) Set 21SW mounted at the front right of the scanner unit to STATIONARY SCAN to set the scanner to be stationary (See Fig.4.1.1). Fig. 4.1.1 Stationary method of scanner (3) Select maintenance scan from the main menu on the operation panel. (4) After selecting a desired protocol on the operation panel, set the scan conditions to desired values, and click Confirm button. Turn on [Stationary scan] item. No. of scans Patient table feed quantity Gantry angle Stationary scan Delay Start mode Variation code Phantom correction Bias Tube current Tube voltage Scan time Slice thickness Inter-scan time Reconfiguration moment Large ADPP ADPP Scattered ray CT number attenuation (measurement) (reconfiguration) correction adjustment correction FOV Filter (5) After the system has been set to be ready for measurement, B.H.C. START B.G.C. switch flickers Press it. (6) When the stationary scan is not necessary, reset 21SW changed in (2) to NORMAL. (See Fig.4.1.1) (7) For deleting [Maintenance scan]] from the measurement main menu on the operation panel, observe 5.1.2. [Maintenance scan] only cannot be deleted, but [phantom calibration], [maintenance scan], and [CT number adjustment scan] belonging to the service scan are deleted collectively. 4-1 L2E-BW1444 4.2 Detector - preamplifier connectors correspondence table Channel No. 1~8 Detector Preamplifier PC board Device block No. PC board No. PC board No. 1 1 1CN Remarks 17 ~ 24 channels only are used 9~32 2 1 2CN 33~56 3 2 3CN 57~80 4 2 4CN 81~104 5 3 5CN 105~128 6 3 6CN 129~152 7 4 7CN 153~176 8 4 8CN 176~200 9 5 9CN 201~224 10 5 10CN 225~248 11 6 11CN 249~272 12 6 12CN 273~296 13 7 13CN 297~320 14 7 14CN 321~344 15 8 15CN 345~368 16 8 16CN 369~392 17 9 17CN 393~416 18 9 18CN Left preamplifier termination 417~440 19 10 19CN 1st right preamplifier 441~464 20 10 20CN 448-449ch is the measuring data center 465~488 21 11 21CN 489~512 22 11 22CN 513~536 23 12 23CN 537~560 24 12 24CN 561~584 25 13 25CN 584~608 26 13 26CN 609~632 27 14 27CN 633~656 28 14 28CN 657~680 29 15 29CN 681~704 30 15 30CN 705~728 31 16 31CN 729~752 32 16 32CN 753~776 33 17 33CN 777~800 34 17 34CN 801~824 35 18 35CN 825~848 36 18 36CN 849~872 37 19 37CN 873~896 38 19 38CN 4-2 L2E-BW1444 Combinations of segment outputs and external take-out signals Mode Switching signal sel 1 sel 2 sel 3 1 2 3 4 5 6 7 8 L L L L L H L H H H L L H L H Use 1.25mm×4 2.5mm×4 3.75mm×4 5mm×4 For check Seg. e' Seg. d' Seg. c' Seg. b' Seg. a' Seg. a Seg. b Seg. c Seg. d Seg. e A-GND A-GND A-GND D slice C slice B slice A slice A-GND A-GND A-GND A-GND D slice D slice C slice C slice B slice B slice A slice A slice A-GND L H L Special measurement D slice D slice D slice C slice C slice B slice B slice A slice A slice A slice D slice D slice C slice C slice C slice B slice B slice B slice A slice A slice D slice C slice C slice C slice C slice B slice B slice B slice B slice A slice C slice C slice C slice C slice C slice B slice B slice B slice B slice B slice H H L Special measurement A-GND A-GND D slice D slice C slice B slice A slice A slice A-GND A-GND H H H Special measurement D slice D slice D slice D slice C slice B slice A slice A slice A slice A slice Output signal connector pins assignment Pin No. Output signal Pin No. Output signal Pin No. Output signal Pin No. Output signal a1 A-GND a26 12ch-D b1 A-GND b26 12ch-B a2 24ch-D a27 12ch-C b2 24ch-B b27 12ch-A a3 24ch-C a28 11ch-D b3 24ch-A b28 11ch-B a4 23ch-D a29 11ch-C b4 23ch-B b29 11ch-A a5 23ch-C a30 10ch-D b5 23ch-A b30 10ch-B a6 22ch-D a31 10ch-C b6 22ch-B b31 10ch-A a7 22ch-C a32 9ch-D b7 22ch-A b32 9ch-B a8 21ch-D a33 9ch-C b8 21ch-B b33 9ch-A a9 21ch-C a34 8ch-D b9 21ch-A b34 8ch-B a10 20ch-D a35 8ch-C b10 20ch-B b35 8ch-A a11 20ch-C a36 7ch-D b11 20ch-A b36 7ch-B a12 19ch-D a37 7ch-C b12 19ch-B b37 7ch-A a13 19ch-C a38 6ch-D b13 19ch-A b38 6ch-B a14 18ch-D a39 6ch-C b14 18ch-B b39 6ch-A a15 18ch-C a40 5ch-D b15 18ch-A b40 5ch-B a16 17ch-D a41 5ch-C b16 17ch-B b41 5ch-A a17 17ch-C a42 4ch-D b17 17ch-A b42 4ch-B a18 16ch-D a43 4ch-C b18 16ch-B b43 4ch-A a19 16ch-C a44 3ch-D b19 16ch-A b44 3ch-B a20 15ch-D a45 3ch-C b20 15ch-B b45 3ch-A a21 15ch-C a46 2ch-D b21 15ch-A b46 2ch-B a22 14ch-D a47 2ch-C b22 14ch-B b47 2ch-A a23 14ch-C a48 1ch-D b23 14ch-A b48 1ch-B a24 13ch-D a49 1ch-C b24 13ch-B b49 1ch-A a25 13ch-C a50 A-GND b25 13ch-A b50 A-GND 4-3 L2E-BW1444 Device block (Top View) Measurement at each slice thickness 1.25mm×4 2.5mm×4 3.75mm×4* 5.0mm×4 D slice C slice B slice 1CH A slice 24CH Sel 1 Switch selection signal (Common to the right and left) D-GND Sel 1 Sel 2 Sel 3 Vcc Switch selection signal connector (Surface mounting) b1 b2 b3 b4 Sel 3 D slice SW’ Outputs are connected collectively by 4 segments C slice Seg. e' Seg. d' a1 a2 a3 a4 Seg. c' Output signal connector (Mounted on the rear panel) Output signal connector (Surface mounting) Sel 2 Pin No. Pin No. Seg. b' Seg. a' Seg. a Seg. b Seg. c Seg. d Seg. e b47 b48 b49 b50 a47a48 a49 a50 Outputs are connected collectively by 4 segments B slice SW A slice Sel 1 Sel 2 Sel 3 Segment divisions in channels and external take-out signals * With the X-ray beam collimation combined, A,D slices of 3.75mm×4 slice measurement execute 3.75mm measurement by collecting the outputs of inner 2 segments out of Seg. e and Seg. d. 4-4 L2E-BW1444 4.3 Details of service information The following table shows the service information being displayed on the image taken by Presto. Items BG :Bowel Gas Correction BH :Beam Hardening Correction CA :Phantom Cal. No AC :Air calibration elapsed time RC :Reference Covering Correction PI :Part Info. (PGA Gain) OD :Over Dose LA :Large Att. Correction (ADNR) IC :Image Ring Correction AM :Adaptive-mA Contents Indicates bowel gas correction ON/OFF of an image. 0: Bowel gas correction is turned off 1: Bowel gas correction is turned on. Indicates beam hardening correction ON-OFF of an image. 0: Beam hardening correction is turned off 1: Beam hardening correction is turned on Indicates the phantom calibration data employed when preparing raw data. 0: Phantom calibration data are not used 1: 230mm phantom calibration data are used 2: 260mm phantom calibration data are used 3: 350mm phantom calibration data are used 4: 410mm phantom calibration data are used Indicates the elapsed time of air calibration data employed when preparing raw data. 0: Shorter than 30 minutes 1: Longer than 30 minutes but shorter than 1 hour 2: Longer than 1 hour but shorter than 1 hour and 30 minutes 3: Longer than 1 hour and 30 minutes but shorter than 2 hours 4: Longer than 2 hours but shorter than 2 hours and 30 minutes 5: Longer than 2 hours and 30 minutes but shorter than 3 hours 6: Longer than 3 hours but shorter than 3 hours and 30 minutes 7: Longer than 3 hours and 30 minutes but shorter than 4 hours 8: Longer than 4 hours but shorter than 4 hours and 30 minutes 9: Longer than 4 hours and 30 minutes or air calibration is not executed E: Air calibration data are abnormal The elapsed time is calculated every conditions (kV, collimation). For an image taken at 120kV, 2.5mm x 4 slices , the elapsed time of 120kV, 2.5mm x 4 slice air calibration data is displayed.。 Indicates whether the reference covering correction is executed or not. 0: Reference covering correction is not executed 1: Reference covering correction is executed Not used in Presto Indicates whether over X-ray dose was detected or not when preparing raw data. 0: Over dose X-rays were not detected 1: Over dose X-ays were detected Indicates whether large attenuation correction (ANDR) was executed or not when preparing raw data. 0: Large attenuation correction (ADNR) was not executed 1: Large attenuation correction (ADNR) was executed ON/OFF state of the image ring correction: 0: Image ring correction is OFF. 1: Image ring correction is ON. ON/OFF state of Adaptive-mA 0: Adaptive-mA is OFF. 1: Adaptive-mA is ON. 4-5 L2E-BW1444 Appendix How to remove the side cover (Remove 3 screws) The upper part of the cover is inserted. Draw out the cover downward while pulling the lower part this side. (This also applies to the opposite side) Side cover 4-6 210 L2E-BW1444 Section 5 Maintenance software 5.1 Maintenance mode ON/OFF 5.1.1 Maintenance mode ON (1) Click [System/Exit] button of main menu, and select [Maintenance] - [Maintenance Mode ON/OFF] from the displayed menu. (2) A confirmation message is displayed as shown in the following figure. Click [Yes] button. (3) When the maintenance mode is turned on, [Mainte. Scan] button and [Adjust / Calib.] button are displayed in the main menu. Also, [Settings] and [Services] are displayed in the sub-menu of [Maintenance]. 5.1.2 Maintenance mode OFF After clicking [System / Exit] button of the main menu, select [Maintenance] - [Maintenance Mode ON/OFF] from the displayed menu. 5-1 L2E-BW1444 5.2 Maintenance settings [Settings] is displayed as a sub-menu of [System / Exit]-[Maintenance] when the maintenance mode is turned on. From [Settings] menu, the date / time setting, study number setting and system configuration settings can be started. 5.3 Date / time setting A date and time change may affect the HU value. Change it after making sure that the HU is set to 0%. If a time zone is changed, HU may keep 100% more than in 10 hours. In that case, initialize HU to 0% (see Appendix 2). (1) A date and time property window is displayed by selecting [System / Exit] - [Maintenance] - [Settings] [Date / Time Setting]. (2) Set the Date & Time and Time Zone, and click [OK] button. 5.4 Study number setting (1) A study number setting screen is displayed by selecting [System / Exit] - [Maintenance] - [Settings] - [Study No. Setting]. (2) Input a study number and click [Save & Exit] button. 5-2 L2E-BW1444 5.5 System configuration settings REFERENCE When service backup data is returned to the system, system configuration setting data which was set during service backup is automatically selected. “Service Backup” is also added to System Type. When “Presto” is selected, default data of Presto is selected; when “Service Backup” is selected, system configuration setting data which was set during service backup is selected. When closing the screen after executing [Apply], “Service Backup” is removed from the menu. (1) A system configuration settings window is displayed by selecting [System / Exit] - [Maintenance] [Settings] - [System configuration Settings]. Present set data are displayed with a green check mark on the window. (2) For setting items, see Appendix 1. Items of System Configuration Setting. (3) A setting execution check message is displayed by clicking [Apply] button. (4) Setting is executed by clicking [Yes] button on the setting execution check message, and a green check mark showing the present set condition is displayed to the set data. (5) Click [Close] button. (6) When setting is executed, a message “Please shut the system down, and then restart.” is displayed. Click [OK] button and then, start the system again. 5-3 L2E-BW1444 5.6 Maintenance service When the maintenance mode is turned on, [Service] is displayed as a sub-menu of [System / Exit] [Maintenance]. From the [Service] menu, Raw Data Analysis, check, error list display, service backup, and system maintenance mode can be started. 5.7 Raw Data Analysis 5.7.1 Introduction Raw Data Analysis function is provided to analysis raw data picked up by the CT unit and execute the adjustment and maintenance of the CT unit. Raw Data Analysis comprises the following analytical items. (1) PTD Displays data value at an optional point designated on the displayed raw data. (2) PLD Displays numerically the data corresponding to an optional projection designated on the displayed raw data. (3) CHD Displays numerically the data corresponding to an optional channel designated on the displayed raw data. (4) PLP Displays profile data corresponding to an optional projection designated on the displayed raw data. (5) CHP Displays profile data corresponding to an optional channel designated on the displayed raw data. (6) PLC Calculates and displays the maximum value, minimum value, (maximum value - minimum value), mean value, variance, and standard deviation every projection (data of all channels) over a range of all projections of displayed raw data. (7) CHC Calculates and displays the maximum value, minimum value, (maximum value - minimum value), mean value, variance, and standard deviation every channel (displayed all projection data)) over a range of an optional channel of displayed raw data. (8) CLB This is used for centering of the collimator. It indicates a lateral deviation of the collimator as an output difference between the first half channels and the latter half channels. 5-4 L2E-BW1444 (9) VCK Calculates the maximum value- minimum value and standard deviation every all channels (all measuring projection channels) after setting threshold values about maximum value - minimum value and standard deviation, and displays channel numbers and calculation results if these calculated values exceed preset threshold values. This is normally used for vibration check. (10) FCE Enters and deletes defective channels. (11) OPR Indicates a deviation in the slice direction of beams by an output difference of AB slices and CD slices. This is used for beam adjustment. (12) ADP Displays profile of tube current value of Raw data. (13) CACK Starts this processing after indicating raw data obtained by measuring the attached adjusting rod of the system. This is used for adjusting the X-ray tube position. 5.7.2 Raw Data Analysis start A Raw Data Analysis screen is displayed by selecting [System / Exit] - [Maintenance] - [Services] [RawData Analysis]. 5.7.3 5.7.3.1 Opening data Raw data (1) A raw data selection screen is displayed by selecting [File|Open|Raw Data] from the menu. A patient information list is displayed at the upper part, while a raw data list of the patient is displayed at the lower part. (2) By clicking [OK] button after selecting desired raw data to be displayed, the selected raw data are displayed. 5-5 L2E-BW1444 5.7.3.2 Air calibration data (1) By selecting [File|Open|Cal. Data] from the menu, a Cal. Data Selection screen is displayed. (2) Selected air calibration data are displayed by clicking [OK] button after selecting the air calibration data and tube voltage of air calibration data to be displayed on the Cal. Data Selection screen. 5.7.3.3 Phantom profile (1) A Cal .Data Selection screen is displayed by selecting [File|Open|Cal. Data] from the menu. (2) A PLP setting screen is displayed by clicking [OK] button after selecting the phantom profile and the tube voltage of the phantom profile to be displayed on the Cal. Data Selection screen. 5-6 L2E-BW1444 (3) By clicking [OK] button after selecting the collimation and phantom size on the PLP setting screen, a red line is displayed on the selected phantom profile and a profile in the view direction is displayed on the PLP screen. 5.7.3.4 Phantom calibration data (1) A Cal. Data Selection screen is displayed by selecting [File|Open|Cal. Data] from the menu. (2) A PLP setting screen is displayed by clicking [OK] button after selecting phantom profiles and phantom calibration data to be displayed on the Cal. Data Selection screen. 5-7 L2E-BW1444 (3) By clicking [OK] button after selecting the collimation, alpha/beta, and phantom size on the PLP Setting screen, a red line is displayed to the selected phantom calibration data and a profile in the view direction is displayed on the PLP screen. 5.7.4 Closing data The present displayed Raw Data / Cal. Data are closed by selecting [File|Close] from the menu. 5-8 L2E-BW1444 5.7.5 Raw Data Analysis end The Raw Data analysis screen is closed by selecting [File|Exit] from the menu. 5.7.6 Raw Data Analysis screen function 5.7.6.1 A change of window width and window level 5.7.6.2 For changing the window width and window level, input values to W/W, W/L text box. Slice data selection 5.7.7 For selecting slice data, click a radio switch of a desired slice (A,B,C, or D) to be displayed. Analysis menu Selected analysis can be done when selecting [Analysis] from the menu and displaying a check mark by clicking the sub-menu after displaying raw data or calibration data. Analysis ends when deleting the check mark by clicking the sub-menu again. * For CACK, refer to 5.8.8. (1) PTD By selecting PTD, a task bar is displayed at the lower part of the Raw Data Analysis screen. By clicking at an optional position, a red cross caret flickers at the clicked position and coordinates (channel, view) and data value are displayed in the status bar. 5-9 L2E-BW1444 (2) PLD A PLD screen is displayed by selecting PLD. By clicking an optional position on the Raw Data Analysis screen, a red line is displayed on the selected view and data in the channel direction to the selected view are displayed. (3) CHD A CHD screen is displayed by selecting CHD. By clicking an optional position on the Raw Data Analysis screen, a red line is displayed in the selected channel and data in the view direction to the selected channel are displayed. 5-10 L2E-BW1444 (4) PLP A PLP screen is displayed by selecting PLP. By clicking an optional position on the Raw Data Analysis screen, a red line is displayed on the selected view and a profile in the view direction is displayed on the PLP screen. By pressing the left button of the mouse on the desired data to be observed on the PLP screen, a red line is displayed on the selected channel, and channel number and data are displayed in the status bar. 5-11 L2E-BW1444 (5) CHP A CHP screen is displayed by selecting CHP. By clicking an optional position on the Raw Data Analysis screen, a red line is displayed in the selected channel, and data in the channel direction are displayed on the CHP screen. By pressing the left button of the mouse on the desired data to be observed on the CHP screen, a red line is displayed on the selected view, and view and data are displayed in the status bar. (6) PLC By selecting PLC, the maximum value, minimum value, (maximum value - minimum value), mean value, variance, and standard deviation every projection of displayed raw data are displayed on the PLC screen. 5-12 L2E-BW1444 (7) CHC By selecting CHC, the maximum value, minimum value, (maximum value - minimum value), mean value, variance, and standard deviation of the displayed raw data are displayed on the CHC screen. (8) CLB By selecting CLB, a profile obtained by averaging all views of displayed raw data is displayed on the CLB screen. Averages of 11 channels at the right side region and 10 channels at the center region and errors of the left side region and right side region are displayed on the Compensator Balance screen. The Compensator Balance screen is closed by clicking [Close] button. 5-13 L2E-BW1444 (9) VCK By selecting VCK, a list of the maximum value- minimum value profile (Diff. profile) and standard deviation profile (SD profile) of all channels of displayed raw data, and the maximum value, minimum value, (maximum value - minimum value), mean value, variance, and standard deviation of the channel number exceeding the default set threshold value are displayed on the VCK screen. A set threshold value is displayed to respective profiles by a dotted line. By clicking [Diff. Profile], [SD Profile], or [Error List] button, whether Diff. Profile, SD Profile, and Error List are displayed or not can be clicked. A dotted line showing a threshold value on the error list and profiles is updated by clicking [Check] button after inputting a desired threshold value to be changed to the Diff or SD text box and then, selecting the slice (A, B, C, or D) radio button. By clicking [All Check] button. all slices are checked to update the error list. Also, whether an error exists in each slice or not is displayed in the Vibration Influence screen. 5-14 L2E-BW1444 (10) FCE By selecting FCE, entered defective channels are displayed on the FCE screen. For entering a defective channel, input the channel number to be entered into the text box, select the slice (A, B, C, D or All) radio button, and then, click [Entry] button. The number of entry-enable defective channels is max. 8 channels in the total of A~ D slices and maximum 8 channels in all slices. For deleting a defective channel, input the channel number to be deleted to the text box, select the slice (A, B, C, D or All) radio button, and click [Delete] button. When clicking [Close] button while the defective channel is entered, the following message is displayed, asking to turn off Adaptive-mA function in system configuration setting. If FCE registration is made in the range from 17CH to 88CH, Adaptive-mA control may not normally be carried out. In this case, turn OFF Adaptive-mA in the system configuration setting. Please be requested to repair it immediately. In case of defective preamplifier, you may temporarily replace the relevant pc board with the both edge. However, special care should be paid to Ref. CH, (11) OPR By selecting OPR, a difference between a mean output of AB slices of individual 10 channels in the left region, right region, and center region and a mean output of CD slices and a mean output of each slice are displayed in the OPR screen. 5-15 L2E-BW1444 (12) ADP By selecting ADP, tube current values of each view are displayed on ADP screen. When left-clicking on the data you wish to observe on the ADP screen, a red line is displayed on the selected view, and the view and data are displayed on the status bar. (13) [Dfault], [Max./Min.], [Key Input], and [Redisplay] button. [Dfault], [Max./Min.], [Key Input], and [Redisplay] button are displayed on the profile display screen. Each switch functions as follows. * Default Sets the data display range to -32768 ~ 32767. * Max./Min. Sets the data display range to the minimum value and maximum value in the displayed profile. * Key Input Sets the data display range optionally. By clicking [Key Input] button, the minimum /maximum value input dialog opens. After inputting a numeric value, click [OK] button, and the set value is reflected. * Redisplay Redisplays a profile by adjusting the scale of the abscissa to meet the lateral length of the window when changing the profile screen size. 5-16 L2E-BW1444 5.7.8 CACK (1) By selecting [Analysis | CACK] from the menu after setting by displacing the attached adjusting rod of the system within the slice face from the 20 ~ 25 scanner center, a raw data selection screen is displayed. (CACK cannot be selected when raw data are open.) 5-17 L2E-BW1444 (2) By clicking [Set as Small Focus Data] button after selecting small focus raw data, selected data are set to small focus data. By clicking [Clear] button, the set data are deleted. (3) By clicking [Set as Large Focus Data] button after selecting large focus raw data, selected data are set to large focus data. By clicking [Clear] button, the set data are deleted. (4) By clicking [OK] button, a trajectory trace of the adjusting rod is displayed. The moving direction and moving distance of the X-ray tube are also displayed on the Center Alignment screen. Adjust the X-ray tube position according to these direction and distance. It is not always necessary to set both small focus and large focus raw data in (2) and (3). By setting either small focus data or large focus data, the moving direction and moving distance of X-ray tube can be calculated. 5-18 L2E-BW1444 5.8 Check program Start the check program by selecting in the order of [System], [Maintenance], [Service], and [Check]. 5.8.1 Basic window structure ④ Looping object mark (*) ① Kind of check ③ Check box ② Check menu ⑤ Threshold value ⑥Single unit check start button ⑧Check mark clear button ⑦Looping check start button ⑨End button ① Kind of check Indicates the kind of check menu. ② Check menu Indicates the contents of check menu. ③ Check box Checked menu becomes an object of processing. ④ Looping object mark Indicates a menu to be able to execute looping check that repeats checking until an error occurs. ⑤ Threshold value Indicates a threshold value of offset data check. By double- clicking the offset data check menu column by the mouse, a threshold value setting window is displayed to be ready for setting a threshold value. 5-19 L2E-BW1444 Threshold value setting screen ⑥ Single unit check start button Checks the menu being checked to the check box sequentially once from the upper part. A check progress window is displayed during check. ⑩ Next check start button ⑪ All check end button Check progress window When the check ends, a result display window is displayed. When plural check are selected to execute checking, the present check menu is terminated to start the next check by ⑩ next check start button. By pressing all check end button ⑪, the present executing check ends and a result display window is displayed. ⑦ Looping check start button This button is used to check the menu being checked to the check box sequentially from the upper part until an error occurs. The check progress window is displayed during check. This button is the same as in the single unit check basically. However, this switch is not effective if a menu having no looping object mark ④ is not selected. The result display window indicates the result to only the check where an error occurs. ⑧ Check mark clear button Removes check marks in all the check boxes. ⑨ End button Terminates the check program. 5-20 L2E-BW1444 5.8.2 Preamplifier check (1) Bus-line data check Check the test pattern data are read from ROM in preamplifier and the data transfer route to see if it functions normally. <Checking method> Check the check box of [Execute Bus-Line data check] item. Click single unit check start button ⑥. The check progress window is displayed. A result is displayed on the check result window at the end of check. (2) Offset data check Check if the leak current from the detector and the offset voltage of the integrator in the preamplifier are within a fixed value, respectively. <Checking method> Check [Execute Offset data check] check box. For changing a threshold value, double-click the [Execute Offset data check] column. A threshold value setting window is displayed. Input a threshold value and click OK button. Click single unit check start button ⑥. After the end of check, a result is displayed in the check result window. 5-21 L2E-BW1444 5.8.3 5.8.3.1 Scanner check Scanner system check (1) Communication check Perform communication check between the scanner and the table as well as between the scanner and X-ray control unit. <Checking method> Check the check box of [Execute Scanner subsystem communication check] item. Click single unit check start button ⑥. The check progress window is displayed. After the end of check, a result is displayed in the check result window. (2) Timing pulse generation Generate and stop the timing pulse <Checking method> Check the check box of [Execute Scanner timing pulse generation check] item. Click single unit check start button ⑥. The check progress window is displayed. A view trigger signal and other control signals are generated. This check is not terminated automatically. Terminate it by pressing all check end button ⑩. After the end of check, a result is displayed in the check result window. (3) Scanner rotation time check Rotate the scanner clockwise at the designated scan time and measure the required time. <Checking method> Check the check box of [Execute Scanner rotation time (*.*s) check box. Click single unit check start button ⑥. The check progress window is displayed. After the end of check, a result is displayed in the check result window. 5-22 L2E-BW1444 5.8.3.2 Table system check (1) LED check Light LED on the scanner operation panel sequentially. <Checking method> Check the check box of [Execute Scanner LED lighting check] item. Click single unit check start button ⑥. The check progress window is displayed. Switches and LED on the scanner operation panel lights sequentially. Switches or LED light every 400 mm sec. in the following sequence (800 mm sec during home →lifting switch lighting). After upper and lower [200] light, this operation returns to PAT lighting. →PAT→PRESET→HOME→Lifting.→Lowering→Forward move→Backward move→Forward tilting(TILT+)→Backward tilting(TILT-)→ (TILT)+00.0→+01.1→+02.2→+03.3→+04.4→+05.5→+06.6→+07.7→+08.8→+09.9→+10.0 →+20.0→-20.0→(longitudinal)+001.1→+002.2→+003.3→+004.4→+005.5→+006.6→+007.7 →+008.8→+009.9→+110.0→+220.0→+330.0→+440.0→+550.0→+660.0→+770.0→+880.0→ +990.0→+1000.0→‐1000.0→(height)011→022→033→044→055→066→077→088→099 →100→200→・・・ This check is not terminated automatically. Terminate it by pressing all check end button ⑩. After the end of check, a result is displayed in the check result window. (2) Scanner and table operation switch check Read the scanner operation panel switches and indicate the switch numbers on the operation panel. <Checking method> Check the check box of [Execute Scanner and table switch read check ] item. Click single unit check start button ⑥. The check progress window is displayed. A switch number is displayed on the scanner operation panel, each time an operation switch is pressed once. Respective switch numbers are as specified below. 1. 1. PAT 8. Forward move 2. RESET 9. High speed 3. PRESET 10. Backward move 4. HOME 11. Forward tilting (Tilt + ) 5. Projector 12. Backward tilting (Tilt - ) 6. Lifting. 13. MOVE (OC side) 7. Lowering 14. START (OC side) If two or more switches are pressed at the same time, a smaller value switch takes preference of the other. 2. Don’t operate EMERGENCY switch. If operated by mistake, terminate the table operation switch check, and restart the system after releasing the emergency once. 5-23 L2E-BW1444 This check is not terminated automatically. Terminate it by pressing all check end button ⑩. After the end of check, a result is displayed in the check result window. (3) Table moving speed check Measure the time required by moving the table. <Checking method> Check the check box of [Execute Table moving speed check] item. Click single unit check start button ⑥. The check progress window is displayed. Perform the following table move check. For table move check, operate the table on the gantry side as follows. Press RESET switch several times to indicate “6” in HEIGHT and “5.0” in TILT. Press the table backward move switch or table forward move switch and high-speed move switch together to move the table forward or backward. Speed measurement is started 1.5 seconds after starting the move until the speed is stabilized. Actual moving speed is displayed in POSITION. When this value is within the range of the following table, it is normal. If it is out of the range, judge it as an error and adjust the table. Check 3 kinds of speed shown in the table by the above procedure. Operating speed Height display Tilt display Tolerance 5.0mm/s 6 5.0 5.0±1.0mm/s 20.0mm/s 12 20.0 20.0±1.0mm/s 60.0mm/s 13 60.0 60.0±3.0mm/s This check is not terminated automatically. Terminate it by pressing all check end button ⑩. After the end of check, a result is displayed in the check result window. 5-24 L2E-BW1444 5.8.3.3 X-ray control system check (1) X-ray main circuit rectification operation check Operate the rectification circuit only at a DC output voltage of 200V. <Checking method> Check the check box of [Execute rectifier operation check] item. Click single unit check start button ⑥. The check progress window is displayed. Only the rectification circuit operates for 20 seconds under such condition as the rectification circuit output becomes 215Vdc. After the end of check, a result is displayed in the check result window. 5.8.3.4 PreAmplifier signal ReTransmission system check (1) Scanner restransmission circuit error display Indicate an error information detected by the resend circuit of the scanner. <Checking method> Check the check box of [Display preamplifier signal retransmission circuit] item. Click single unit check start button ⑥. The check progress window is displayed. After the end of check, a result is displayed in the check result window. (2) Scanner restransmission circuit error counter clear Clear the error information detected by the resend circuit of the scanner and error frequency <Checking method> Check the check box of [Clear preamplifier signal retransmission circuit] item. Click single unit check start button ⑥. The check progress window is displayed. After the end of check, a result is displayed in the check result window. 5-25 L2E-BW1444 5.8.3.5 DSP system check (1) Ideal reconstruction check Displayed image data are checked by executing the reconstruction using exclusive raw data Whether B.G.C is executed or not can be selected. <Checking method> When B.G.C is executed, check the check box of [Execute ideal reconstruction check (BHC ON)]. When B.G.C is not executed, check the check box of [Execute ideal reconstruction check (BHC OFF)]. If the number of reconstruction image is changed, double-click the check marked column, and the number of reconstruction image setting window is displayed. Input a number of reconstruction image, and click OK button. Click unit check start button ⑥. A checking window is displayed. A reconstructed image is displayed together with the image display window. When the check ends, results are displayed in the check results window. <Number> indicates a slice number in the ideal reconstruction check program. It shows that what number of image an error occurred. After the end of check, close the image display window by clicking [×] button at the upper right of the window. (2) Scan DSP check Check the external memory (scan side) on the board by using DSP of DSP board. Either full mode or easy mode can be selected. In the full mode, the external memory is checked by using 6 kinds of test patterns. In the easy mode, the external memory is checked by using 2 kinds of test patterns. <Checking method> In case of the full mode, check the check box of [Execute Scan DSP check (Full Mode)]. In case of the easy mode, check the check box of [Execute Scan DSP check (Easy Mode)]. Click unit check start button ⑥. A checking window is displayed. The progressive condition is displayed by a progress bar on the checking window. After the end of check, results are displayed in the check results window. <Number> indicates a DSP number referring to the memory in error in the DSP check program. For the correspondence of the DSP number, refer to the following inter-DSP board connection diagram. 5-26 L2E-BW1444 (3) Reconstruction DSP check Check the external memory ( reconstruction side) on the board by using DSP of DSP board. Either full mode or easy mode can be selected. In the full mode, the external memory is checked by using 6 kinds of test patterns. In the easy mode, the external memory is checked by using 2 kinds of test patterns. <Checking method> In case of the full mode, check the check box of [Execute ReconDSP check (Full Mode)]. In case of the easy mode, check the check box of [Execute ReconDSP check (Easy Mode)]. Click unit check start button ⑥. A checking window is displayed. The progressive condition is displayed by a progress bar on the checking window. After the end of check, results are displayed in the check results window. <Number > indicates a DSP number referring to the memory in error in the DSP check program. For the correspondence of the DSP number, refer to the following inter-DSP board connection diagram. CPU1 CPU0 Slot 10 HR-BP-V8 Slot 2 A Raw Data HR-PCI 1 2 S 14 I Slot 9 C Image HR-PCI 13 15 Slot 4 Slot 12 3 5 7 9 11 4 6 8 10 12 Scan HR-DSP Recon HR-DSP Scan DSP check side Reconstruction DSP check side Inter-DSP board connection diagram 5-27 L2E-BW1444 5.9 Control Error List 5.9.1 Control error list start/end (1) An error list window is displayed by selecting [System / Exit] - [Maintenance] - [Services] [Error List Display]. (2) The error list window is closed by clicking [Exit] button. 5.9.2 Setting of the order of date and time (1) Click [Date / Time] of the list view or select [Settings (S) |[Date / Time (D)] from the menu, and a displayed items settings dialog is displayed. (2) Select the order (descending / ascending) of date and time from the sort order combo box. (3) For displaying time down to milli-second unit, click “ Display to milliseconds” and put a check mark. (4) Click [OK] button. 5-28 L2E-BW1444 5.9.3 Display of a specified process only (1) Click [Process] of the list view or select [Settings (S)|Process (P)] from the menu, and a displayed items settings dialog is displayed. (2) Select a desired process to be displayed to the list from the displayed items combo box and click [OK] button. 5.9.4 Display of a specified message only (1) Click [Message] of the list view or select [Settings (S)|Message (M)] from the menu, and a displayed items settings dialog is displayed. (2) Select a desired message to be displayed to the list from the displayed items combo box and click [OK] button. 5.9.5 Setting of the error log information Directory (1) Click [Tools (T)|Error Log Info. Settings (S)] from the menu, and an setting for error log information dialog is displayed. 5-29 L2E-BW1444 (2) Check the check box of the information to be changed and change the aimed contents. <CRT side information> * Error Log File Directory : Set a folder for saving an error log file by the absolute path. * Max. Size of Error Log File (bytes) : Set the maximum size of error log file in units of byte. * Touch Panel Side Error Log File Directory Viewed from CRT Side : Set the error log file to be saved to the CPU side by the path when referring from the CPU side. <Touch panel side information> * Error Log File Directory : Set a folder for saving an error log file by the absolute path. * Error log time max. size (bytes) : Set the maximum size of error log file in units of byte. (3) Click [OK] The error log file is controlled in units of processing name and if a file size of a certain processing name exceeds this size, the present file is saved as a file before 1 generation and the present file is initialized. A file before 2 generations is not saved. If an error occurs frequently or it is desired to save old error information, increase this size. If initial value 16384 (bytes) is set, about 300 items information can be saved in the present file plus file before 1 generation in total. 5-30 L2E-BW1444 5.9.6 Saving of the error log information to MO (1) An error log information order dialog is displayed by clicking [Tools (T)|Error Log Info. Order (O)] from the menu. (2) Insert an MO into the MO drive, select an order method, and click [Start (S)] button. “Clear error log” becomes selectable when an error log is copied to an MO. (3) Click [Close (C)] REFERENCE Error log information can also be saved into an MO by using service backup (See 5.10). 5-31 L2E-BW1444 5.10 Service backup This processing serves to save or read the setting data of various options on the magnetic disk, such as the CT number adjustment data, phantom calibration data, date/time calculated phantom data, system serial number, latest study number/patient ID/accession number, series number management file, system configuration status, faulty channel, filming and DICOM transmission destinations settings, and MWM option setting, to or from an MO cartridge, and to enter them into a magnetic disk. Data can be saved more than once so far as the MO capacity permits. During service backup, special care must be ensured in the following operation. (1) Do not read out service data or user data to a different system than the system in which they were saved. (2) Do not read out service data or user data to a system while an entry of a patient is registered. (3) After the CT value adjustment data and phantom calibration data were read out to the system, be sure to restart the system before making scanning (including CT value adjustment and phantom calibration). 5.10.1 Start / end (1) Click [System/Exit] of the main menu and select [Maintenance], [Service], and [Service Backup] from the displayed menu. (2) A Backup/Restore window is displayed. (3) Click [Close] button to exit. 5-32 L2E-BW1444 5.10.2 Initialization of saving MO To save service data into an MO, initialize the MO for service data. (1) Set an MO to the drive Set an MO to the drive. if it is not initialized for service data, the following message is displayed. Click [Yes] button, and a format window is displayed. (2) Click [Start] button. (File system is fixed to FAT32. Format time can be shortened by checking the quick format.) A message warning the deletion of data is displayed. (3) After confirming the contents of the message, click [OK] button. Formatting is started. After the end of formatting, a message is displayed to inform of the end. (4) Click [Close] button. Now, the service backup MO has been formatted. 5-33 L2E-BW1444 5.10.3 Saving (1) To save service data, set an MO initialized for service data into the drive after the Backup/Restore dialog is displayed. (2) If the service data saving MO has already been formatted, [Save…] button can be clicked. Click [Save…] button. (3) A backup dialog is displayed. An item checked in the selection column is saved. (A check mark is attached to all default items) Save all data. (4) Click [Save] button. Service data saving is started. After the end of saving, the display returns to the Backup / Restore window. 5-34 L2E-BW1444 5.10.4 Loading (1) For loading service data, set the MO saving service data. To load service data, set the MO saving service data after the Backup/Restore dialog is displayed. The service data being saved in the MO are displayed on the list. Select data to be loaded by clicking the selection column. By selecting data set, clicking [Load…] button is enabled. A Restore dialog is displayed by clicking [Load…] button. (2) Data being checked in the selection column are loaded (A check mark is attached to all default items). Load all data. (3) Click [Load] button. Service data loading is started. After the end of loading, the display returns to the Backup/Restore window. 5-35 L2E-BW1444 5.10.5 Deletion Unnecessary service data can be deleted. (1) For deleting service data, set the MO saving service data after Backup / Restore window has been displayed. The service data being saved into the MO are displayed in a list. Select data to be deleted by clicking the selection column. By selecting data set, [Delete] button can be clicked. A confirmation dialog is displayed by clicking [Delete] button. (2) Click [Yes], and service data are deleted. 5-36 L2E-BW1444 5.10.6 Other data In service backup, user data and log as well as service data can be saved in the same MO. Log and data base are saved as compressed by using a freeware, UNLHA32.DLL, the use of which is permitted by the owner of the copyright, Mr Micco. (1) User data saving Select “User Data” from the data type list of Backup / Restore window. Saving, loading, and deletion methods of user data are the same as in starting the [Save User Data] from [System/Exit]. (2) Log saving Select “Log” from the data type list of Backup/Restore window. A log can be saved or deleted, but it cannot be loaded. (3) Database saving Select “Database” from the data type list of Backup/Restore dialog. Database can only be saved and deleted, but cannot be loaded. Database is provided for images and RawData, and saving database is enabled for the purpose of investigating causes of troubles. 5-37 L2E-BW1444 5.11 System maintenance mode The system maintenance mode is provided to execute application installation and other system settings after switching to normal Windows environment where the application software is not started. 5.11.1 System maintenance mode / release mode selection (1) Select [Maintenance], [Service], and [System Maintenance mode] from the displayed menu after clicking [System/Exit] button of the main menu. (2) A password input window is displayed. Input rs0471314185 (initial value) and click [OK]. (3) A system maintenance mode switching confirmation dialog is displayed. Click [Yes]. The system is restarted and switched to the system maintenance mode (Windows environment where the application is not started). REFERENCE For returning from the system maintenance mode to normal application start mode (release mode), double-click the “Start the CT system” short cut at the left of the screen, and the system is restarted and the application is started. 5-38 L2E-BW1444 5.12 Error codes list 5.12.1 Control error codes list Error codes Error list messages SE08 "Invalid Message Command " "Invalid Message Command : CheckPro" "Invalid Message Command : MScanDriV" "Invalid Message Command : DSPDriV" "Invalid Message Command : SysMoni" SE09 "Invalid Message SCode " "Invalid Message Scode:Cmd ***" SE0A SE0B SE0C SE10 SE18 SE20 SE26 SE27 Dialog messages A communication error has occurred in the scan control process. Please retry. Error contents Communication header command error ① An undefined instruction was received from MC Server ② Commands received by MC Server from respective tasks were undefined Communication header send source name error Sender name of received message is not MC Server "Invalid Message DCode " Communication header send "Invalid Message Dcode: destination name error Cmd ***" Receiver name of received message is in error Communication header parameter "Invalid Message Length " A communication error has length error "Invalid Message Length: occurred in the scan control Length of a received message is in Cmd ***" process. error Please retry. Socket error during communication "Invalid Socket:Cmd **" A communication error has An attempt was made to transfer the occurred in the scan control communication command by TCP process. socket, but a transfer error occurs Please retry. "Parameter Error " A parameter error was detected. Parameter error "Illegal Scan Data Type : **" Change parameters, and then retry. Contents of communication parameter unit are not correct ① Scan processing command parameter is undefined ② Check processing command parameter is undefined ③ Table/gantry processing command parameter is undefined The system is busy. Exclusive error Please retry. Other processing is being executed Initialization error The scan control process startup "Config Setting Error " Task start error failed. "DSP Boot Error " Please shut the system down, and ① Definition file setting error "Event Initialize Error" ② DSP boot error then restart. "Process Boot Error" ③ Communication event initialization error ④ Process initialization error "File Open Error " An error has occurred in the scan File open error control process. Please retry. "File Close Error " An error has occurred in the scan File close error control process. Please retry. A communication error has occurred in the scan control process. Please retry. A communication error has occurred in the scan control process. 5-39 L2E-BW1444 Error codes Error list messages SE28 "File Read Error " SE29 "File Write Error " SE2A "DataBase Access Error" "RawDisk DataBase Error" "DataBase Create Error" "DataBase Parameter Error" "Memory Error " SE2B SE2C SE2D "Communication Error " "Invalid Socket" "Internal Communication Error " "Process Timeout " SE2E "DICOM Format Error " SE40 "DSP Driver Error " SE41 "DSP reported Error " "DSP Busy" "DSP Start Process Error " SE44 "Insufficient Disk Space" SE45 "Recon Timeout" "PostRecon Timeout" SE48 "RawData Read Error " SE49 "RawData Write Error " SE50 "Offset Scan error" SE51 "Windows API Error" Dialog messages An error has occurred in the scan control process. Please retry. An error has occurred in the scan control process. Please retry. An error has occurred in the scan control process. Please retry. Error contents File read error An error has occurred in the scan control process. Please retry. A communication error has occurred in the scan control process. Please retry. Memory error An error occurred during memory access Communication error An error occurred during inter -control task communication An error has occurred in the scan control process. Please shut the system down, and then restart. An error has occurred in the scan control process. Please retry. An error has occurred in the scan control process. Please shut the system down, and then restart. An error has occurred in the scan control process. Please shut the system down, and then restart. The disk has insufficient free space. Delete images or RawData, and then retry. An error has occurred in the scan control process. Please shut the system down, and then restart. An error has occurred in the scan control process. Please retry. An error has occurred in the scan control process. Please retry. An error has occurred in the scan control process. Please retry. An error has occurred in the scan control process. Please retry. Processing time out 5-40 File write error Data base access error An error occurred during data base access Image format error. An error occurred during DICOM image format conversion. DSP driver error DSP error Shortage of file space Reconstruction time over Reconstruction was started, but an image is not completed Raw Data read error Raw Data write error Offset error Shortage of offset number Windows API error L2E-BW1444 Error codes Error list messages SE60 "Release File Space Error" SE61 "Viewer Error" SE62 "Warning:DSP Flag ON at Start" "Warning:Abnormal DSP PCode" "Warning:Faulty CH Out of Range" "Warning:Requested DSP not found" "Unexpect Error" SE73 SE74 SE75 "Internal communication error" "Event Communication Error" "Unexpect Error" OE 02 S 04 "Shutter Scan is in progress" S 09 "Stationary Switch is ON" S 10 "Scanner is not Ready" S 11 "Scanner is not Ready" Dialog messages An error has occurred in the scan control process. Please retry. An error has occurred in the scan control process. Please retry. An error has occurred in the scan control process. Please retry. Error contents File space reserve cancel error An error occurred when canceling the file space reserve Viewer error ① An error occurred when starting the viewer ② An error occurred when sending a message to the viewer DSP warning error An error has occurred in the scan control process. Please retry. An error has occurred in the scan control process. Please retry. Unexpected error Program internal error Images were not stored nor displayed within predetermined time. For safety, please suspend the guideShot, and try again after restarting application. The table height is too low. Please raise the table above zero. Image file writing timeout error Internal communication error Internal communication error of task Table height is not higher than 0 ① Table height is not higher than 0 during scanogram ② Table height is not higher than 0 when checking the table moving speed An abnormality has occurred in the An error during shutter scan scanner. Please retry. The scanner is in an abnormal Stationary switch (21SW) turns on condition. during normal scan Please contact your service representative. The scanner is in an abnormal Scanner not ready condition. ① Scanner is not set to be ready Please retry. ② Stationary switch (21SW) is turned off during stationary scan The scanner is in an abnormal Scanner not ready condition. ① Scanner is not set to be ready Please retry. ② Stationary switch (21SW) is turned off during stationary scan 5-41 L2E-BW1444 Error codes Error list messages S 12 "Scanner is not Ready : timeout" S 15 "Relay OFF of inverter" S16 "Relay OFF of inverter" S 17 "FFB OFF for Tilt" S 18 "FFB OFF for inverter" S 19 "Stationary Switch is ON" S 1C "L-Tilt Over Heat" S 1D "R-Tilt Over Heat" S 1E "Manual Switch is ON" S 1F "Emergency Switch is ON" S 20 "Brake Release is ON" S 27 "Tape Switch is ON" S 28 "Scanner Error" Dialog messages The scanner is in an abnormal condition. Please retry. An abnormality has occurred in the scanner. Please shut the system down, and then restart. An abnormality has occurred in the scanner. Please retry. The scanner is in an abnormal condition. Please contact your service representative. The scanner is in an abnormal condition. Please contact your service representative. The scanner is in an abnormal condition. Please contact your service representative. An abnormality has occurred in the scanner. Leave the system power on and allow to cool. An abnormality has occurred in the scanner. Leave the system power on and allow to cool. The scanner is in an abnormal condition. Please contact your service representative. The system has stopped due to an Emergency Switch being activated. Confirm safety, and then click [OK] button to release the Emergency status. The scanner is in an abnormal condition. Please contact your service representative. The touch sensor has activated. Please confirm safety, and then click [OK] button to release the stop status. The touch sensor has activated. Please confirm safety, and then click [OK] button to release the stop status. 5-42 Error contents Scanner not ready Scanner is not set to be ready Inverter power control relay off Inverter unit power control relay off Tilt power supply FFB (22FFB) turns off Scanner rotation inverter control power supply FFB (21FFB) off Stationary switch (21SW) turns on during normal scan Tilt hydraulic pump is overheated (left side) Tilt hydraulic pump is overheated (right side) Selector switch (3SW) is set to MANUAL during normal scan Emergency switch was pressed Brake release switch (22SW) is turned on Touch sensor turns on Scanner error L2E-BW1444 Error codes Error list messages S 28(08) "Miss Sense" S 28(1D) S 28(1E) S 28(43) S 28(44) S 28(45) S 28(54) S 28(57) S 28(58) Dialog messages An abnormality has occurred in the scanner. Please retry. "Over Speed Error" An abnormality has occurred in the scanner. Please retry. "Under Speed Error" An abnormality has occurred in the scanner. Please retry. "Inverter Thermal Relay ON" An abnormality has occurred in the scanner. Leave the system power on and allow to cool. "Inverter Trip" An abnormality has occurred in the scanner. Please shut the system down, and then restart. "Front Cover is Open" The front cover of the scanner is open. Please contact your service representative. "DPRAM Commu. Init Error" An error has occurred in the scanner. Please shut the system down, and then restart. "Scan Seq. Timeout" An error has occurred in the scanner. Please retry. "Asynchronized Timer An error has occurred in the Counter" scanner. Please retry. S 28(5A) "Rotation Encoder Error" S 28(71) "Communication Error to IPU" S 28(72) "Illegal Command" S 28(73) "Measurement seq. Abnormal" S 28(74) "Measurement Parameter Error" S 28(76) "X-Ray Abort" An error has occurred in the scanner control. Please shut the system down, and then restart. An error has occurred in the scanner. Please shut the system down, and then restart. An error has occurred in the scanner. Please retry. An error has occurred in the scanner. Please retry. An error has occurred in the scanner. Please retry. [SUSPEND] button was pressed. 5-43 Error contents Upper side sensor of X-ray tube does not turn on at the specified position Rotation over speed (+12%) Rotation under speed (-12%) General-purpose inverter thermal relay turns on Inverter trips (Wrong operation of old motor) Front cover of scanner is open DPRAM communication initialization error Measurement end signal time out Timer 0 and 4 counters are not synchronized with each other during stable rotation speed of scanner Abnormal pulse from the rotary encoder of the scanner Communication cannot be done between main CPU and image processing Command receive under improper condition Scanner sequence number change is abnormal A parameter other than ready command (81H) parameter was received X-ray abort instruction was received L2E-BW1444 Error codes Error list messages S 28(77) "1 Scan Timeout" S 28(78) "Table motion Time out" S 28(79) "Parameter Error from IPU" S 30 "Command Error to Scanner CPU" S 50 "Scanner Check Seq. Now" S 60 "Communication Error MCPU & SCPU" X 05 "Arcing Recovery is activated" X 11 "Rotor is not ready" X 14 "Initialization error at WB" X 17 "Door Open" X 1A "Tube Over Heat" X 1F "WB is emergency " X 20 "Arcing Recovery counts Over" Dialog messages An error has occurred in the scanner. Please retry. An error has occurred in the scanner. Please shut the system down, and then restart. An error has occurred in the scanner. Please shut the system down, and then restart. An error has occurred in the scanner. Please shut the system down, and then restart. An error has occurred in the scanner. Performing check sequence. Please do not perform other operations. An error has occurred in the scanner. Please shut the system down, and then restart. Error contents 1-scan time out Table does not move for 3 seconds after table move instruction An invalid parameter was received from image processing Communication command error Communication can be done between main CPU and scanner CPU, but a code is abnormal Check sequence in progress A communication error between MCPU and SCPU Communication is impossible between main CPU and scanner CPU Arcing occurred Arcing recovery was executed An arcing has occurred at the X-ray tube in scan. When an image is created, artifact may have occurred. The rotor of the X-ray tube startup Rotor stop condition failed. (during standby operation) ① Rotor stop Please retry. ② Abort An error has occurred in the X-ray X-ray unit initialization error control system. Please shut the system down, and then restart. The door of the CT room is ajar. Scanner room door is open Close the door. The X-ray tube has overheated. Overheat ① Oil temperature rises in X-ray Leave the system power on and tube housing allow to cool. ② Cooling pump stops operating The X-ray control system is in an Emergency switch was pressed emergency condition. Please contact your service representative. The number of arcing recoveries Arcing recovery counts over has been exceeded. Please retry. 5-44 L2E-BW1444 Error codes Error list messages X 21 "Rotor Error " X 22 X 23 X 24 X 25 X 26 X 27 X 28 X 29 X 2A X 2B X 2C X 2D X 2E X 30 Dialog messages An abnormality has occurred at the rotor of the X-ray tube. Please retry. "Primary Error or Secondary An abnormality has occurred in the Abnormal (WB)" X-ray generator main circuit system. Please retry. "Focus Error Miss Selection" The X-ray tube focus switch failed. Please retry. "Breaker OFF Control A breaker tripped in the X-ray Circuit" control unit. Please contact your service representative. "CPU Memory Error (WB)" An abnormality has occurred in the X-ray control unit. Please shut the system down, and then restart. "Arcing Error" Arcing has occurred during calibration. Please retry. "Filament Power is OFF" The power of the filament drive circuit is OFF. Please contact your service representative. "HV Tank Over Heat" The high voltage transformer has overheated. Leave the system power on and allow to cool. "Abunormal KV & MA An abnormality has occurred in (WB)" tube voltage and tube current. Please retry. "Tube Voltage Error An abnormality has occurred in (WB)" tube voltage. Please retry. "Filament Cut" The filament or the heating circuit is in an abnormal condition. Please contact your service representative. "Backup Timer Error (WB)" A backup timer error has occurred. (WB) Please retry. "A/D Converter Error An abnormality has occurred in the (WB)" X-ray control unit. Please retry. "Miss Radiation (WB)" An abnormality has occurred in the X-ray control unit. Please retry. "Command Error unresolved An error has occurred in the X-ray (WB)" control unit. Please retry. 5-45 Error contents Rotor stop (during exposure) Rectification circuit or inverter is abnormal Focus selection error Control circuit power supply is turned off Memory error Arcing occurred during calibration Filament power off HV transformer bellows switch turns on Center section current is abnormal KV error Filament or heating circuit error Exposure time over ADC error X-ray is generated at timing other than specified Communication system error (An unacceptable command was received) L2E-BW1444 Error codes Error list messages X 32 "Command Error (WB)" X 3C X 3D X 3E X 40 X 41 X 42 X 43 X 44 X 45 X 46 X 47 X 48 X 49 X 4A X 4B Dialog messages An error has occurred in the X-ray control unit. Please retry. "Collimator Error" An error has occurred during movement of the collimator. Please retry. "A/D Converter Error An abnormality has occurred in the (Filament)" filament control. Please retry. "MA Reference Setting An error has occurred in the Error" filament control. Please retry. "+KV Over voltage" An abnormality has occurred in tube voltage. Please retry. "-KV Over voltage" An abnormality has occurred in tube voltage. Please retry. "KV Unbalance" An abnormality has occurred in tube voltage. Please retry. "KV Over Voltage" An abnormality has occurred in tube voltage. Please retry. "Anode over current" An abnormality has occurred in tube current. Please retry. "Cathode over current" An abnormality has occurred in tube current. Please retry. "Center section over current" An abnormality has occurred in tube current. Please retry. "Filament Over Voltage" An abnormality has occurred in the filament. Please retry. "Filament Over Current" An abnormality has occurred in the filament. Please retry. "CPU WatchDog Timer An error has occurred in the Error" filament control. Please shut the system down, and then restart. "Filament Preheat Error " An abnormality has occurred in the filament. Please retry. "Backup Timer Error (Turn A backup timer error has occurred. Plate)" (Turn Plate) Please retry. 5-46 Error contents Communication system error Collimator error ADC error Setting error of tube current reference value Tube voltage (+) overvoltage error Tube voltage (-) overvoltage error Unbalance of tube voltage (+) and tube voltage (-) Tube voltage overvoltage error Anode overcurrent error Cathode overcurrent error Center section overcurrent error Filament overvoltage error Filament overcurrent error Watch dog error of tube current control CPU Auxiliary heating error Backup time over (Tube current control circuit detection) L2E-BW1444 Error codes Error list messages X 4C "Filament Preheat FeedBack Error" X 60 X 61 X 62 X 70 T 00 T 02 T 03 T 04 T 0A T 0B T 0E T 0F T 10 Dialog messages An abnormality has occurred in the filament. Please retry. "Communication error to A communication error has Turn plate" occurred in the turn plate board. Please retry. "Communication error to A communication error has WB" occurred in the X-ray control system. Please shut the system down, and then restart. "Communication error to A communication error occurred in ATMA" the Adaptive mA control circuit. Please shut the system down, and then restart. "Retry Over (Turn Plate)" A error has occurred in the data communications of a preamplifier signal. Please retry. "Mismatch Status:Table [RESET] button on the gantry side Ready" may have been pressed. Please retry. "Mismatch Status:Index An error has occurred in the table. Move Now" Please retry. "Mismatch Status:Gantry Tilt An error has occurred in the table. Now" Please retry. "Mismatch Status:Scan Seq. An error has occurred in the table. Now" Please retry. "Table Limit Inward" "Table Limit Outward" "Table Error :Inhibit" Table limit. (Inward) Table limit. (Outward) The table is in an operation-prohibited condition. Please retry. "Mismatch Status:Fluoro is in An error has occurred in the table. Progress" Please retry. "Irregular OP" Manually operating at the gantry. Please retry after checking that the operations at the gantry have been completed. T 14 "Manual OP done in Scanning" T 1F "Emergency" T 20 "Table Speed Error" Error contents Auxiliary heating feedback voltage is not within the reference value range Communication error (Turn plate) Communication error (WB) Communication error between adaptive mA control circuit and tube current control Data retransmission count over Table ready condition Table moving (Index) Gantry tilting Scan measurement in progress ① Scangram in progress ② Volume scan in progress Table position in direction limit Table position out direction limit Operation is disabled during measurement Fluoroscopic measurement in progress Manual operation in progress ① Manual operation in progress ② PAT operation in progress ③ Table height is lower than 0mm Waiting for the end of scanogram The table is operating at reduced speed. Please retry. The system has stopped due to an Emergency switch was pressed Emergency switch being activated. Confirm safety, and then click [OK] button to release the Emergency status. An error has occurred in the table. Table speed error Please retry. 5-47 L2E-BW1444 Error codes Error list messages T 21 "Servo Ready Timeout" T 22 T 23 T 24 T 25 T 26 T 27 T 40 T 60 T 70 SD00 SD01 SD05 SD06 Dialog messages The servo driver did not become ready within the standard time. Please shut the system down, and then restart. "Servo Alarm Error" An error has occurred in the servo driver. Please shut the system down, and then restart. "Servo-motor Stop Timeout" The servo motor did not stop within the standard time. Please shut the system down, and then restart. "DeadMan Relay OFF The deadman relay was not Timeout" released within the stipulated time. Please shut the system down, and then restart. "A/D Convert Timeout" The A/D conversion did not finish within the standard time. Please shut the system down, and then restart. "S,Pulse Cycle Error(5%)" The S.Pulse Signal exceeded the stipulated cycles Please shut the system down, and then restart. "Tape Switch is ON" The touch sensor has activated. Please confirm safety, and then click [OK] button to release the stop status. "Table Over Run" The table is located outside the operational range. Move the table manually to the operational range. "Table Error not A communication error has Cummunicate(Main1)" occurred in the table. Please retry. "CT mode Error" The angio system is not position of the CT mode. Move the angio system to position of the CT mode. "Scanner Error" An error has occurred in the scanner control. "Scanner Driver Open Error" An error has occurred in the scanner control. Please retry. "Status Read Timeout" An error has occurred in the scanner control. Please retry. "Status(Table) Read An error has occurred in the Timeout" scanner control. Please retry. 5-48 Error contents Time out during servo preparation Servo alarm error Servo motor stop time out Deadman relay off time out A/D conversion end time out S pulse cycle error (±5%) Touch sensor ON Table run over Communication error CT mode error Scanner driver open error Shared communication memory to scanner cannot open Scanner status read time out Scanner status read (table area) time out L2E-BW1444 Error codes Error list messages SD07 "Scanner Driver Init Error" Dialog messages An error has occurred in the scanner control. Please retry. "Scanner Driver Close Error" An error has occurred in the scanner control. Please retry. "Scanner Abnomal" An error has occurred in the scanner control. Please retry. "Check Sum Error" An error has occurred in the scanner control. Please retry. "Scanner Invalid Code" An error has occurred in the scanner control. Please retry. Error contents Scanner driver cannot be initialized SD39 "In Scanner Wakeup" Scanner control process is starting up. Please wait... SD42 "Set LED Error" An error has occurred in the scanner control. Please retry. After scanner power supply has been turned on during scanner starting, the scanner is not standing by Set LED cannot be controlled (lighting and extinction) SD08 SD0A SD30 SD31 SD32 SD34 SD35 SD36 SD37 SD38 Scanner driver cannot be closed Scanner abnormal answer (Status is normal) Check sum error when receiving a command from scanner ① No scanner answer in 97H ・ Scan execution time (except scanogram) ・ Status read time ・ Check program ② Improper code as a result of scanner communication check “Gantry power is OFF” Gantry power supply is OFF. Gantry power off ・ When sending a command to Press [GANTRY] button. scanner ・ When receiving a command from scanner "Abnormal Table Position" An error has occurred in the Table position error scanner control. ① After clearing the table position, table position is not zero Please retry. ② After resetting the table position, the table position is different from its set value "Table Pos. is not expected" Table did not move to the specified Table position does not conform to position. the set value after moving the table Please retry. "Gantry Ang. is not expected" Gantry did not tilt to the specified After gantry tilting, the gantry angle angle. is different from the set value Please retry. "Table not Move" An error has occurred in the Table does not move scanner control. Please retry. TapeSW ON Error" An error has occurred in the After sending touch sensor off scanner control. command, touch sensor does not turn off Please retry. 5-49 L2E-BW1444 Error codes Error list messages SD44 "Start LED Error" Dialog messages An error has occurred in the scanner control. Please retry. An error has occurred in the scanner control. Please retry. SD46 "DeadMan Set Error" SD47 "DeadMan Unset Error" An error has occurred in the scanner control. Please retry. SD48 "X-Ray Soft Switch On Error" An error has occurred in the scanner control. Please retry. SD49 "X-Ray Soft Switch Off Error" An error has occurred in the scanner control. Please retry. SD4A "LED Flick Error" SD4C "CPU Ready On Error" SD4D "CPU Ready Off Error" SD4E "Scanner ResetOn Error" SD4F "Scanner ResetOff Error" SD61 "Send Command Parameter Error" SD62 "Send Command Device Erro" An error has occurred in the scanner control. Please retry. An error has occurred in the scanner control. Please retry. An error has occurred in the scanner control. Please retry. An error has occurred in the scanner control. Please retry. An error has occurred in the scanner control. Please retry. An error has occurred in the scanner control. Please retry. An error has occurred in the scanner control. Please retry. 5-50 Error contents Start LED cannot be controlled (lighting and extinction) Deadman cannot be to manual mode ・ During scanner initialization ・ During table move ・ During gantry control ・ During scan ・ During table speed check Deadman cannot be to auto mode ・ During table move ・ During gantry control ・ During scan ・ During table speed check Soft switch cannot be turned on ・ During view trigger signal check ・ During preamplifier check ・ During scan start Soft switch cannot be turned off ・ During scanner initialization ・ During view trigger signal check ・ During preamplifier check ・ Just before starting the preamplifier check ・ At the end of scan LED cannot flicker CPU ready does not turn on (when starting the scanner) CPU ready does not turn off (when starting the scanner) Scanner reset cannot be turned on Scanner reset cannot be turned off Transmitting command parameter error Transmitting command error ①Error during transmission ②Ring buffer full L2E-BW1444 Error codes Error list messages SD63 "Send Command Scanner Ready Error" SD64 SD65 SD66 SD67 SD68 SD69 SD6A SD6B SD6C SD6D OT02 OT03 Dialog messages An error has occurred in the scanner control. Please retry. "Send Command Other Error" An error has occurred in the scanner control. Please retry. "Recv Command Device An error has occurred in the Error" scanner control. Please retry. "Recv Command NoData" An error has occurred in the scanner control. Please retry. "Recv Command Other An error has occurred in the Error" scanner control. Please retry. "Recv Command Open Event An error has occurred in the Error" scanner control. Please retry. "Recv Command Clear Event An error has occurred in the Error" scanner control. Please retry. "Recv Command WaitObject An error has occurred in the Timeout" scanner control. Please retry. "Send Command Timeout An error has occurred in the Error" scanner control. Please retry. "Invalid TableData Code" An error has occurred in the scanner control. Please retry. "Invalid TableData Data" An error has occurred in the scanner control. Please retry. "Offset Scan Timeout" An error has occurred in images and RawData acquisition system. Please retry. "Scan Start Timeout" An error has occurred in images and RawData acquisition system. Please retry. 5-51 Error contents Transmitting command acceptance disable error Receiving is impossible against send-to-request Transmitting command error Receiving command error ①Error during receiving ②Ring buffer full Receiving command error Error occurs when no information is given to receiving data Receiving command error Receiving command open error Receiving clear processing error Receiving time out from scanner Transmitting command cannot be sent to cause a time out error Transmitting command parameter error Command error in table area Transmitting command parameter error Data error in table area Offset time out Offset measurement end information does not come from DSP ・Scanogram (3sec) ・Normal scan (4sec) Scan start time out Scan start information does not come from DSP ・Scanogram (4sec) ・Volume scan (3sec) L2E-BW1444 Error codes Error list messages OT04 "Fast Scan end wait timeout:Sc" "Scan End Timeout:Sc" Dialog messages An error has occurred in images and RawData acquisition system. Please retry. OT05 "HU Exceeded" An error has occurred in the scanner control. Please retry. An error has occurred in the scanner control. Please retry. An error has occurred in the scanner control. Please retry. OT07 "HU calc Error" OT08 "Scanner Commu. Check Timeout" OT09 "Scanner Rot. Time Check (CW) Timeout" An error has occurred in the scanner control. Please retry. OT0B "Table Moving Speed Check Timeout" An error has occurred in the scanner control. Please retry. OT0C "Status Read Timeout" An error has occurred in the scanner control. Please retry. 5-52 Error contents Scan end time out ① Table status does not change from 04 to 14 even if waiting for 2 seconds from the predetermined scanogram end time ② Answer ( 97H) does not come from scanner even if waiting for 2 seconds during normal scan ③ Answer (97H) does not come from the scanner even if waiting for 1 second after all data collection end during volume scan ④ Next data collection end information does not come even if waiting for one rotation time +1sec from the last data collection end during volume scan ⑤ Answer (97H) does not come from the scanner even if waiting for 112% + 3seconds of all scan time during volume scan HU excess condition was input in HU estimate calculation HU estimate calculation error Check processing time out Answer (97H) does not come from scanner even if waiting for 10 seconds during scanner communication check Check processing time out Answer (97H) does not come from scanner even if waiting for 20 seconds during scanner rotation time check Check processing time out Answer (97H) does not come from scanner even if waiting for 2 seconds at the table moving speed check end Status read time out Answer (97H) does not come from scanner even if waiting for 5 seconds during status read L2E-BW1444 Error codes Error list messages OT0D "Move Table Timeout" Dialog messages An error has occurred in the scanner control. Please retry. OT0E "Tilt Gantry Timeout" An error has occurred in the scanner control. Please retry. OT0F "Emergency Release Timeout" An error has occurred in the scanner control. Please retry. OT10 "Tube Return Timeout" An error has occurred in the scanner control. Please retry. OT11 "Standby Timeout" An error has occurred in the scanner control. Please retry. OT12 "Fast Scan end wait An error has occurred in the timeout:DSP" scanner control. "Scan end wait timeout:DSP" Please retry. OT13 "Image Recon Timeout" An error has occurred in the scanner control. Please retry. OT14 "RotStop Timeout" An error has occurred in the scanner control. Please retry. OT15 "SeqStart Timeout:DSP" An error has occurred in the scanner control. Please retry. OT60 ”FileRead Error” OT70 "Pause Interval TimeOut" A file read error has occurred. Please retry. The pause limit time has been exceeded. The scan has ended. 5-53 Error contents Table move time out Table move does not end (OEH) even if waiting for 30 seconds during table move Gantry tilting time out Tilting does not end (OEH) even if waiting for 30 seconds during gantry tilting Emergency processing time out Reset does not end (other than 1FH) even if waiting for 80 seconds during emergency reset processing Tube recovery processing time out Tube does not return upward even if waiting for 20 seconds during tube recovery processing Measurement preparation time out ・Scanogram (40 seconds) ・Normal scan (40 seconds) Raw Data waiting time out Last Raw Data end information does not come from DSP even if waiting for 1 second during Raw Data waiting time out scan Image reconstruction time out end information does not come from DSP even if waiting for 3 seconds during normal scan Scanner rotation stop processing time out Scanner does not does not stop rotating even if waiting for 40 seconds Sequence start answer waiting time out DSP ready information does not come from DSP even if waiting for 2 seconds after standby Temporary stop time out Time setting file cannot be read Time out from pause to restart Scanner stopped running due to a long time from pause to restart in normal scan L2E-BW1444 Error codes Error list messages Dialog messages Error contents OT20 "Av Pattern DB Open Error " Could not open the auto voice data AV DB open error file. ① Auto voice generation pattern table is not present or it is broken ② Auto voice generation pattern table is being used at the present application ③ Auto voice generation pattern table cannot open OT26 "AvData DB Open Error " Could not open the auto voice data AV DB open error file. ① Voice data table is not present or it is broken ② Voice data table is being used at the present application ③ Voice data table cannot open OT29 "Av Pattern No. Not Existed" An error has occurred in the auto AV pattern No. error voice data. Designated auto voice No. does not exist in the generation pattern table OT30 "AvPh No. Error" An audio data file access error has AV standby phrase error occurred. Designated phrase No. does not exist in the voice data table An audio data file access error has AV standby phrase error OT32 "AvPh not Exist" occurred. Designated phrase file (*** wav) does not exist An audio data file access error has AV standby phrase error OT33 "AvPh Open Error" occurred. AV phrase cannot open Please shut the system down, and then restart. An audio data file access error has AV standby phase error OT34 "AvPh Play Error" occurred. AV phrase cannot play Please shut the system down, and then restart. An audio data file access error has AV standby phrase error OT35 "AvPh Close Error" occurred. AV phrase cannot be closed Please shut the system down, and then restart. An audio data file access error has AV standby phrase error OT36 "AvPh Wait Timeout" occurred. AV phrase end information does Please shut the system down, and not come then restart. TX01 "VN Error (TXR)" View No. non-conformity error An error occurred in preamplifier signal data transfer system and data was sent again TX02 "LRC Error (TXR)" Horizontal parity check error An error occurred in preamplifier signal data transfer system and data was sent again TX03 "Start/End Bit Error (TXR)" Start end bit error An error occurred in preamplifier signal data transfer system and data was sent again 5-54 L2E-BW1444 Error codes Error list messages TX04 "Data Number Error (TXR)" TX05 TX06 TX07 TX08 TX09 Dialog messages Error contents Data number error An error occurred in preamplifier signal data transfer system and data was sent again An error has occurred in the signal DAS view non-conformity error "DAS VN Error (TXS)" from the DAS. Please shut the system down, and then restart. "DAS Data Clk Error (TXS)" An error has occurred in the signal DAS data lock out from the DAS. Please shut the system down, and then restart. An error has occurred in the data "Retry Over (TXS)" Retry over transmission of a preamplifier signal. Please contact your service representative. "Data Retransmit (TXS)" Preamplifier signal resend "Reply Signal Error(TXS)" Reply signal error 5-55 L2E-BW1444 5.12.2 Error codes System error log codes table Application names 80000000 80000001 80000002 80000003 80000004 80000005 80000006 80000007 80000008 80000009 80000020 80000021 Main Menu Main Menu Main Menu Main Menu Main Menu Main Menu Main Menu Main Menu Main Menu Main Menu Main Menu Main Menu 80000022 80000030 Main Menu Main Menu 80000031 Main Menu 80000032 Main Menu 80000033 Main Menu 80000034 Main Menu 80000035 Main Menu 80000036 Main Menu 80000037 Main Menu 80000050 80000051 Main Menu Main Menu 80000052 80000053 Main Menu Main Menu 80000060 80000070 Main Menu Main Menu 80000071 Main Menu 80000072 Main Menu Dialog messages The Patient Reservation startup failed. The Patient Registration startup failed. The Scanogram startup failed. The Scan startup failed. The Recon. Status startup failed. The Scan Viewer startup failed. The Table Move Setting startup failed. The Warm-up startup failed. One of the system task startup failed. The Image Explorer startup failed. Loading of the patient information file (Pt_data.df) failed. Loading of the scanning operation preference settings file (UserCustom.df) failed. Loading of the table information file (tblmove.dat) failed. Could not establish communication with the System Monitor. Please shut the system down, and then restart. Failed to transmit the patient information because of communication error with the System Monitor. Failed to transmit the hard button information because of communication error with the System Monitor. Failed to transmit the table information because of communication error with the System Monitor. Failed to transmit the post recon. queue information because of communication error with the System Monitor. Failed to transmit the date/time information because of communication error with the System Monitor. Failed to transmit the number of remaining images that can be stored because of communication error with the System Monitor. Failed to transmit the online recon. information because of communication error with the System Monitor. Perform short Warm-up. Short Warm-up is required. Do you want to do it, now? Perform Warm-up. Warm-up is required. Do you want to do it, now? Do you want to shut the system down? Are you sure you want to enter Maintenance Mode? Keep in mind if you change any parameter in the Maintenance Mode, it may affect the system. Change to the System Maintenance Mode. Are you sure you want to reboot the system to change the mode? The following conditions are required to run the check program: -Examination has finished. -Post recon. is not running. -Image Explorer is closed. Do you want to start the check program? 5-56 L2E-BW1444 Error codes Application names 80000073 Main Menu 8000007A Main Menu 80000080 80000090 Main Menu Main Menu 80000100 Main Menu 80000101 Main Menu 80000102 Main Menu 80000103 Main Menu 80000110 Main Menu 80000111 Main Menu 82000001 82000002 82000003 82000004 82000005 82000006 83000000 Table Move Settings Table Move Settings Table Move Settings Table Move Settings Table Move Settings Table Move Settings Post Recon. Queue Management Post Recon. Queue Management Post Recon. Queue Management RawData Post Recon. Registration Task 83000001 83000002 84007000 84007001 84007002 84007003 84007004 84007005 84007006 RawData Post Recon. Registration Task RawData Post Recon. Registration Task RawData Post Recon. Registration Task RawData Post Recon. Registration Task RawData Post Recon. Registration Task RawData Post Recon. Registration Task Dialog messages The password is incorrect. Passwords are case sensitive. It is necessary to close the Image Explorer in order to start up the Image Table Reset. Are you sure you want to close it? (If the Recon. Status window is displayed, it will be closed.) Do you want to exit from current patient examination? Releasing the system lock... Please wait... RawData that does not correspond to Keep RawData exists. Are you sure you want to delete now? Deleting RawData... Please wait... System was shut down abnormally. An error check will be performed. Checking if there is any error... Please wait... Suspending the recon... Please wait... Stopping transfer... Please wait... There is an error in the INI file settings. There is no configuration file. There is no image file. There is an error in the input data. An unexpected error has occurred. Input the value of *****. There is a recon. queue in process. Are you sure you want to stop the recon.? All of information for recon. queue is deleted because the previous shutdown was done abnormally. All of information for recon. queue will be deleted. Are you sure you want to continue? Could not acquire the RawData information. Click the list item again after completing processes such as scanning and post recon. Could not acquire the RawData information. Perform the Image Table Reset. The image storage capacity is insufficient. Could not acquire the RawData patient information. Perform the Image Table Reset. Could not acquire the RawData series information. Perform the Image Table Reset. Could not acquire the RawData slice information. Perform the Image Table Reset. A communication time out error has occurred. Please shut the system down, and then restart. 5-57 L2E-BW1444 Error codes 84009007 84005008 84007009 84007010 Application names RawData Post Recon. Registration Task RawData Post Recon. Registration Task RawData Post Recon. Registration Task RawData Post Recon. Registration Task 84007011 RawData Post Recon. Registration Task 85000000 90001011 Auto-Transfer Queue Management Auto-Transfer Queue Management System Monitor 90001015 System Monitor 90001016 System Monitor 90002001 System Monitor 90002002 System Monitor 90002003 System Monitor 90002004 System Monitor 90002005 System Monitor 90002006 System Monitor 90002007 System Monitor 90002008 System Monitor 85000001 Dialog messages The registration to the post recon. queue was completed. Do you want to run the registered queue? Could not run the post recon. queue. Open the Recon. Status window and try to run the post recon. queue. RawData of the following slice numbers does not exist. ***** Please do not include the displayed slice number in the image reconstruction range. Since RawData of the following slice numbers do not exist, it could not register them to the post recon. queue. ***** Please do not include the displayed slice number in the image reconstruction range. Transfer is being performed. Are you sure you want to stop transfer? Stopping *****... Please wait... A communication error between applications has occurred. Please shut the system down, and then restart. An error has occurred while checking the Raw Data management file. There is a problem with the Raw Data disk. Please contact your service representative. An error has occurred while recreating the Raw Data management file. The Raw Data disk has some problems. Please contact your service representative. RawData could not be acquired. Either there is no RawData, or there is a problem with the RawData disk. Please contact your service representative. RawData could not be written. There is a problem with the RawData disk. Please contact your service representative. RawData could not be deleted. Either there is no RawData, or there is a problem with the RawData disk. Please contact your service representative. RawData disk could not be initialized. Please contact your service representative. Could not acquire the RawData information. Either there is no RawData, or there is a problem with the RawData disk. Please contact your service representative. Could not protect the RawData. Either there is no RawData, or there is a problem with the RawData disk. Please contact your service representative. Could not unprotect the RawData. Either there is no RawData, or there is a problem with the RawData disk. Please contact your service representative. Could not sort the RawData. Please contact your service representative. 5-58 L2E-BW1444 Error codes Application names 90002009 System Monitor 90002010 System Monitor 90002011 System Monitor 90002012 System Monitor 90002013 System Monitor 90008012 System Monitor 90008013 System Monitor 90008014 System Monitor 91000001 CPU0 Startup Process 91000002 CPU0 Startup Process A0000001 A0000002 A0000003 A0000004 A0000005 A0000006 Patient Reservation Patient Reservation Patient Reservation Patient Reservation Patient Reservation Patient Reservation A0000007 Patient Reservation A0000010 A0000011 A0000012 A0000013 Patient Reservation Patient Reservation Patient Reservation Patient Reservation A0000014 A0000016 A0000017 A1000001 A1000002 A1000003 A1000004 A1000005 Patient Reservation Patient Reservation Patient Reservation Patient Registration Patient Registration Patient Registration Patient Registration Patient Registration A1000006 Patient Registration Dialog messages Could not acquire the RawData information. Please contact your service representative. Could not acquire the RawData information. Please contact your service representative. Could not acquire the RawData information. Please contact your service representative. Could not acquire the RawData information. Please contact your service representative. Could not acquire the RawData information. Please contact your service representative. The system has stopped because of Emergency switch activation. After checking for safety, click [OK] button to release the Emergency status. The system has stopped because of the touch sensor activation. After checking for safety, click [OK] button to release the stop status. The system has stopped because of the occurrence of a backup timer error. Click [OK] button to release the stop status. Could not connect with the CRT side computer. Please shut the system down, and then restart. The startup procedure failed. Please contact your service representative. There is an error in the INI file settings. There is no preference settings file. There is no image file. There is no applicable table. There is no Japanese era file. The patient name must be entered. Input the patient name. The following characters are not available in the patient ID. \/:*? <>|. Input a different character. ***** is not a numerical value. There is an error in the input data for *****. An unexpected error has occurred. Are you sure you want to delete the reserved data whose patient name is *****? There is no *****. Set the scan protocol. Do you want to register the input information? There is an error in the INI file settings. There is no preference settings file (*****). There is no image file. There is no applicable table. The patient ID must be entered. Input the patient ID. The following characters cannot be used in the patient ID. \/:*? <>|. Input a different character. 5-59 L2E-BW1444 Error codes Application names A1000009 A1000010 Patient Registration Patient Registration A1000011 Patient Registration A1000012 A1000014 Patient Registration Patient Registration A1000015 A1000017 A1000018 A1000019 A1000020 A1000021 Patient Registration Patient Registration Patient Registration Patient Registration Patient Registration Patient Registration A1000030 A1000031 Patient Registration Patient Registration A1000032 Patient Registration A1000033 Patient Registration A1000034 Patient Registration A1000035 Patient Registration A2000000 A2000001 A2000002 A2000003 A2000004 A3000000 A3000001 A3000002 A3000003 A3000004 B0001000 Work List Work List Work List Work List Work List MPPS MPPS MPPS MPPS MPPS SCServer B0001001 SCServer B0001002 SCServer B0007003 SCServer B0007004 SCServer B0002005 SCServer Dialog messages ***** is not a numerical value. The patient name has not been input. Do you want to set \None\" to the patient name and proceed to scanning?" There is an error in the input data for *****. An unexpected error has occurred. Data with different information already exists for the same patient ID. ***** There is no patient registration file (*****). Input the name of the contrast media. Input a value from ***** to ***** for the Rate. There is no *****. Set scan protocol. Since patient information was updated as the same patient registration, another folder will be created. The current scan plan will be invalid. Are you sure you want to continue? ***** There is no work list message file. Are you sure you want to proceed to scanning with these settings? Feet First is set with ***** protocol. Are you sure you want to proceed to scanning with these settings? An orientation that differs from the scan protocol orientation has been set. Are you sure you want to proceed to scanning with these settings? Data with the same patient information exist in the Reservation List. Are you sure you want to overwrite the input patient information? Pat. Info. was not saved because data with different information already exists for the same patient ID. ***** ***** ***** ***** There is no work list message file. ***** ***** ***** ***** ***** A communication error has occurred with the CPU0. Please shut the system down, and then restart.. Failed to initialize SCServer. Please shut the system down, and then restart. Failed to access memory. Please shut the system down, and then restart. Failed to load the following file. ***** Failed to write the following file. ***** An unknown error has been received. 5-60 L2E-BW1444 Error codes Application names B0012000 B0012001 B0012002 B0022000 SCServer SCServer SCServer SCServer B0025001 SCServer B0044000 SCServer B0045001 SCServer B0042002 B0042003 B0042004 SCServer SCServer SCServer B0043005 B0043006 SCServer SCServer B0043007 SCServer B0043008 SCServer B0043009 SCServer B004200A SCServer B004200B SCServer B004200C SCServer B004200D SCServer B004200E SCServer B004200F SCServer B0047010 SCServer B0047011 SCServer B0047012 SCServer B0047013 SCServer Dialog messages An error occurred while correcting the table position. An error occurred while correcting the scan timing. An error occurred during the image disk access procedure. HU will exceed the limit for scan with current parameters. Correct the scan parameters. HU wait has occurred. ***** seconds remaining until start. Do you want to continue scanning? The following parameters were out of the limits, and have therefore been changed automatically. ***** The sequence will be deleted. Are you sure you want to continue? The image storage capacity is insufficient. The RawData storage capacity is insufficient. The scanner is busy with another process. Please wait... The maximum number of sequences has been reached. The protocol for air calibration (Protocol Mode) has not been registered. Set the scan protocol. Cannot add a sequence because it will be over the maximum number of sequences. Delete the unscanned sequence, and then click [Re-use. Seq.] button again. The number of scans / rotations exceed the limit. Set the No. of Scans / Rotations less than or equal to *****. Multiple scans will be run at the same position. Are you sure you want to continue? Generator overload. Correct the scan parameters. Delay error. Correct the scan parameters. Scanning range has been exceeded. Correct the scan parameters. The maximum number of images exceed for one sequence. Correct the scan parameters. The maximum number of scans / rotations exceed for one examination. Correct the scan parameters. Due to a large number of images, the reconstruction process will not be completed before the next scan begins. Correct the scan parameters. Moving the table or gantry... Please wait... Scan in progress... Please wait... Check program in progress... Please wait... Locking the system... Please wait... 5-61 L2E-BW1444 Error codes Application names B0047014 SCServer B0047015 SCServer B0048016 SCServer B0045017 SCServer B0048018 B0047019 SCServer SCServer B004701A SCServer B004701B B004701C SCServer SCServer B004501D SCServer B004501E SCServer B004501F SCServer B0064000 SCServer B0064001 SCServer B0064002 SCServer B0064003 SCServer Dialog messages Gantry power supply is OFF. Press [GANTRY] button. Starting up the gantry... Please wait... The number of scans / rotations exceed the limit. Register as same Patient again. The current table height differs from the table height of the last scanogram. Height of Scano.: ***** mm Current Height : ***** mm Do you want to continue? The multi recon. parameter was changed for ***** change. There is an error in the multi recon. process. Perform a post recon. using the Image Explorer. The maximum number of images exceeded for multi recon. Correct the scan parameters. Add this scanning sequence. Because no reference image exists, Predict Scan cannot be performed. Change to other than Predict protocol, scan a reference image, and then perform Predict Scan. Air Calibration for these scan parameters (Tube Voltage, Collimation) cannot be performed with Warm-up / Air Calibration normally used. Select this Tube Voltage and Collimation in Air Calibration (Select) of Warm-up, and then perform Air Calibration. (This message will also be displayed after the Air Calibration has been performed.) Did you perform Air Calibration? Yes >> Perform scanning. No >> Return to preparation for scanning, and then perform Air Calibration according to the above method. The orientation for the selected protocol is different from the orientation set in the Patient Registration. If the protocol is changed, check the scanning direction. Are you sure you want to make changes? Monitoring may not be performed correctly on the set ROI position because the gantry angle of the reference image is different from that of the actual predict scan. Move the ROI position on the monitoring image as necessary. ReScan cannot be performed because there is no normal scan image. The scan including the selected image has already been registered. Included in the scan of Series No.*****, Image No.*****. The Multi-recon. image was selected. Registers the main recon. image of this image. (Series No.*****, Image No.*****) Multi-recon. processing is not complete.Imcomplete multi-recon. conditions cannot be registered for ReScan. If the multi-recon. image is needed, clear the registration and register it again when multi-recon. processing is complete. 5-62 L2E-BW1444 Error codes Application names B0064004 SCServer B0074000 B00A5000 SCServer SCServer B00A4001 B00A4002 SCServer SCServer C0000000 IViewer C0000001 IViewer D0001000 Check Program D0001001 Check Program D1000001 D1000002 Error List Display Error List Display D1000003 Error List Display D1000004 Error List Display D1000005 D1000007 Error List Display Error List Display D1000008 D1000009 D100000A Error List Display Error List Display Error List Display D2000001 E0000001 Phantom Calibration /CT Number Adjustment Phantom Calibration /CT Number Adjustment Phantom Calibration /CT Number Adjustment Auto Voice Settings E0000002 Auto Voice Settings E0000003 Auto Voice Settings E0000004 E0000005 Auto Voice Settings Auto Voice Settings D2000002 D2000003 Dialog messages Images other than normal scan exist. Images other than normal scan are not targeted for ReScan, and not displayed during the ReScan process. Auto Zoom cannot be performed because there is no target image. If the protocol is deleted once, it will not be retrieved. Are you sure you want to delete *****? There is no previous scan protocol. There is no region for the last scan protocol. The region for the selected protocol is set. Another process caused the timeout. Please shut the system down, and then restart. Image Layout operation is not complete. Complete Image Layout. There is no response from the CPU0. Check Program is terminated. Please shut the system down, and then restart. The communication with the CPU0 was disconnected. Check Program is terminated. Please shut the system down, and then restart. ***** drive is disabled. There is no MO disk in the drive. Insert a MO disk into drive *****:. The MO disk in drive *****: is write protected. Unlock its write protection and try to copy again. The MO disk in drive *****: is not formatted correctly. Check the MO disk, and then format it if necessary. I/O error (*****) Failed to copy the error log. File Name: ***** Failed to clear the error log. ***** was set to *****. The error log could not be copied due to insufficient space on the MO disk. Insert another MO disk and try to copy again. A data error has occurred. Calculations after ***** are canceled. (*****) Perform a scan again. There is insufficient memory. Please shut the system down, and then restart. Could not acquire the phantom image. Set an appropriate phantom and scan again. Are you sure you want to delete the '*****' pattern? Pattern name '*****' already exists. Set other pattern name. The pattern name is blank. The pattern name should be entered. Are you sure you want to delete the '*****' phrase? Phrase comment '*****' already exists. Add a different phrase comment. 5-63 L2E-BW1444 Error codes Application names Dialog messages E0000006 Auto Voice Settings E0000007 Auto Voice Settings E0000008 Auto Voice Settings E0000009 Auto Voice Settings E0000010 Auto Voice Settings E0000011 Auto Voice Settings E1000001 E1000002 E1000003 E1000004 E1000005 E1000006 E1000007 E1000008 E1000009 E1000010 E1000011 E1000012 E2000001 E2000002 E2000003 E2000004 E2000005 E2000006 E2000007 E2000008 E2000009 E2000010 E2000011 E2000012 Series Comment Registration Series Comment Registration Series Comment Registration Series Comment Registration Series Comment Registration Series Comment Registration Series Comment Registration Series Comment Registration Series Comment Registration Series Comment Registration Series Comment Registration Series Comment Registration Protocol Group Registration Protocol Group Registration Protocol Group Registration Protocol Group Registration Protocol Group Registration Protocol Group Registration Protocol Group Registration Protocol Group Registration Protocol Group Registration Protocol Group Registration Protocol Group Registration Protocol Group Registration E2000013 E2000014 E3000001 Protocol Group Registration Protocol Group Registration Patient/Study Comment Registration Patient/Study Failed to back up the table. Comment Registration Patient/Study Failed to update the table. Comment Registration Patient/Study Failed to delete the backup file. Comment Registration E3000002 E3000003 E3000004 The phrase comment is blank. The phrase comment should be input. A recording/playback error has occurred. Check if recording or playback is in progress. The Auto Voice Settings cannot be started. Please contact your service representative. The Auto Voice Settings startup failed. If you want to perform the Auto Voice Settings, shut the system down, and then restart. The Auto Voice Settings caused an error. If you want to perform the Auto Voice Settings again, shut the system down, and then restart. Register some phrases. If any phrase is not registered, auto voice pattern is not ready to set. There is an error in the INI file settings. Failed to back up the table. Failed to update the table. Failed to delete the backup file. A DB exception error has occurred. Input a comment. The same comment has already been registered. There is no applicable table. An unexpected error has occurred. Are you sure you want to delete *****? There is no preference settings file. Do you want to register this comment? There is an error in the INI file settings. Failed to delete the backup file. Failed to update the table. Failed to delete the backup file. A DB exception error has occurred. Input the protocol group name. The same name is already registered. There is no applicable table. The basic group has not been registered. The maximum number of the protocol group is 5. An unexpected error has occurred. Are you sure you want to delete all scanogram and scan protocols classified in the ***** group? There is no preference settings file. Do you want to register this protocol group? There is an error in the INI file settings. 5-64 L2E-BW1444 Error codes E3000005 E3000006 E3000007 E3000008 E3000009 E3000010 E3000011 E3000012 E4000001 E4000002 E4000003 E4000004 E4000005 E4000006 E4000007 E4000008 E4000009 E4000010 E4000011 E4000012 E4000013 E5000001 E5000002 E5000003 E5000004 E5000005 E5000006 E5000007 Application names Patient/Study Comment Registration Patient/Study Comment Registration Patient/Study Comment Registration Patient/Study Comment Registration Patient/Study Comment Registration Patient/Study Comment Registration Patient/Study Comment Registration Patient/Study Comment Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Physician/Operator Name Registration Contrast Media Settings Contrast Media Settings Contrast Media Settings Contrast Media Settings Contrast Media Settings Contrast Media Settings Contrast Media Settings Dialog messages A DB exception error has occurred. Input a comment. The same comment has already been registered. There is no applicable table. An unexpected error has occurred. Are you sure you want to delete *****? There is no preference settings file. Do you want to register this comment? There is an error in the INI file settings. Failed to back up the table. Failed to update the table. Failed to delete the backup file. A DB exception error has occurred. Input a name. The same name has already been registered. There is no applicable table. An unexpected error has occurred. Are you sure you want to delete *****? There is an error in the character code for *****. There is no preference settings file. Do you want to register this *****? There is an error in the INI file settings. Failed to back up the table. Failed to update the table. Failed to delete the backup file. A DB exception error has occurred. Input a contrast media name. The same information has already been registered. 5-65 L2E-BW1444 Error codes E5000008 E5000009 E5000010 E5000011 E5000012 E5000013 E5000014 E7000001 Application names E8000001 E8000002 Contrast Media Settings Contrast Media Settings Contrast Media Settings Contrast Media Settings Contrast Media Settings Contrast Media Settings Contrast Media Settings Institution Information Settings Institution Information Settings Study Number Setting Study Number Setting E8000003 E8000004 E8000005 EA000001 EA000002 EA000003 EA000004 EA000005 EA000006 EA000007 EA000008 EA000009 EA000010 EB000001 Study Number Setting Study Number Setting Study Number Setting Accord Mode Settings Accord Mode Settings Accord Mode Settings Accord Mode Settings Accord Mode Settings Accord Mode Settings Accord Mode Settings Accord Mode Settings Accord Mode Settings Accord Mode Settings Preference Settings EB000002 Preference Settings EB000003 Preference Settings EB000004 Preference Settings EB000005 Preference Settings EB000007 Preference Settings EB000008 Preference Settings EB000009 Preference Settings EB000010 Preference Settings E7000003 Dialog messages There is no applicable table. There is an error in the input data. An unexpected error has occurred. Are you sure you want to delete *****? Input a value from ***** to ***** for the injection rate. There is no environment settings file. Do you want to register this contrast media information? Input the data. An unexpected error has occurred. There is an error in the INI file settings. The Study No. Setting cannot be started. Register the patient, and then start up the Study No. Setting again. An unexpected error has occurred. Input the data. The input data is not a numerical value. There is an error in the INI file settings. Failed to back up the table. Failed to update the table. Failed to delete the backup file. A DB exception error has occurred. An unexpected error has occurred. There is no configuration file. The accord mode has not been registered. There is no applicable table. Input a value from *****. The Preference Settings cannot be started, because there is no region data file. The Preference Settings cannot be started, because the region data file cannot be accessed. Failed to write the preference data. These data cannot be saved. Preference Settings will exit. Failed to write the region data. These data cannot be saved. Environment Settings will exit. The following characters are invalid for fixed. /:*?"<>|. Input the valid characters. Input the number of days/studies that RawData is to be kept. Input a value from 1 to ***** for the number of days that RawData is to be kept. Input a value from 1 to ***** for the number of studies of RawData to be kept. The Preference Settings cannot be started, because there is no protocol group data file. 5-66 L2E-BW1444 Error codes Application names EB000011 Preference Settings EB000012 EB000013 EC000001 Preference Settings Preference Settings Region Registration EC000002 EC000003 EC000004 EC000005 EC000006 EC000007 EC000008 EC000009 EC000010 EC000011 EC000012 EC000013 EC000014 EC000015 EC000016 EC000017 Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration Region Registration EC000018 Region Registration F0000001 Calculator F1000001 CPU1 Startup Process F1000002 CPU1 Startup Process F2000001 Image Table Reset F2000002 Image Table Reset F2000003 Image Table Reset F2000004 Image Table Reset F2000005 Image Table Reset F2000006 Image Table Reset Dialog messages The Preference Settings cannot be started, because the protocol group data file cannot be accessed. Enter No. of Fixed Digits. Enter a value from ***** to ***** for No. of Fixed Digits. Since another region is already registered on the same position, it cannot be updated. Input the region name. Are you sure you want to delete the region name *****? A communication error has occurred. A data error has occurred. Set the DICOM region name. There is an error in the INI file settings. Failed to back up the table. Failed to update the table. Failed to delete the backup file. There is no applicable table. There is no image file. An unexpected error has occurred. Do you want to register this region information? The original DICOM region cannot be changed or deleted. Cannot connect to the database. Are you sure you want to delete the DICOM region information of *****? The region information cannot be changed because ***** is write-protected. The calculator cannot be started. Please contact your service representative. Could not connect with the submonitor side computer. Please shut the system down, and then restart. The startup procedure failed. Please contact your service representative. If the Image Table Reset is cancelled, images and RawData may not be displayed in the image list, and the post recon. may not be executed. Are you sure you want to cancel? Initial process failed. Please contact your service representative. Not all lock files could be deleted. Please shut the system down, and then restart. And try the Image Table Reset again. Not all database could be deleted. Please shut the system down, and then restart. And try the Image Table Reset again. Failed to re-create the image table. Please shut the system down, and then restart. And try the Image Table Reset again. Failed to acquire the number of images. Please shut the system down, and then restart. And try the Image Table Reset again. 5-67 L2E-BW1444 Error codes Application names F2000007 Image Table Reset F2000008 F2000009 Image Table Reset Image Table Reset F200000A F200000B Image Table Reset Image Table Reset F3000001 Image Table Reset F3000002 Image Table Reset F3000003 F3000004 Image Table Reset Image Table Reset F3000005 Image Table Reset F3000006 Image Table Reset F3000007 Image Table Reset F3000008 Image Table Reset F3000009 Image Table Reset F300000A Image Table Reset F300000B Image Table Reset F4000000 Image Table Reset F4000001 Save User Data / Service Backup Save User Data / Service Backup Save User Data / Service Backup Save User Data / Service Backup Save User Data / Service Backup F4000002 F4000003 F4000004 F4000005 Dialog messages Failed to acquire the number of Raw Data. Please shut the system down, and then restart. And try the Image Table Reset again. ***** image files cannot be re-created. These files may be damaged. ***** RawData files cannot be re-created. These files may be damaged. Are you sure you want to delete these files? ***** image files could not be re-created. These files may be damaged. Failed to acquire the last patient's information. 'The Last patient' is not available for the Image Table Reset. Select one of 'All', 'Image', or 'RawData', and perform Image Table Reset. ***** has been changed. Do you want to save this image? ***** already exists. Do you want to overwrite this image? Input a file name. The directory name is invalid. The following symbols cannot be used for directory name. /:*?"<>| Also \ can be used only as a directory separator. The file name is invalid. The following characters are invalid for a file name. / : * ? "< > | There is no MO disk in the drive. Insert a MO disk into drive *****. The MO disk in drive *****: is write protected. Unlock its write protection and try to save again. The MO disk in drive *****: is not formatted correctly. Check the MO disk, and then format it if necessary. The image could not be saved because the disk in the destination drive is full. Insert another disk to continue. An error has occurred. Error Code: ***** Failed to paste the image. Select [Edit - Paste] from the main menu to paste the image. The disk in drive ***** has not been initialized to save data. Are you sure you want to initialize now? Warning: Initializing the disk will erase all data on the MO disk. Are you sure you want to delete selected items? The startup process failed. Could not create the folder. The disk is write protected. Check the write protection setting. Could not complete the processing. 5-68 L2E-BW1444 Error codes F4000006 F4000007 F4000008 F4000009 F400000A F400000B F5000001 F5000002 F5000003 F5000004 F5000005 F5000006 F5000007 F5000008 F5000009 F5000010 F5000011 F5000012 Application names Save User Data / Service Backup Save User Data / Service Backup Save User Data / Service Backup Save User Data / Service Backup Save User Data / Service Backup Save User Data / Service Backup System Configuration Settings System Configuration Settings System Configuration Settings System Configuration Settings System Configuration Settings System Configuration Settings System Configuration Settings System Configuration Settings System Configuration Settings System Configuration Settings System Configuration Settings System Configuration Settings Dialog messages The disk in drive ***** is insufficient. Insert another disk. Cannot create a temporary file. The disk in drive ***** has been initialized for service data. It cannot be initialized for saving data. Could not format the disk. Formatting disk was cancelled. Error formatting the disk. Cannot find the system configuration type file. Cannot execute the system configuration setting. Cannot find the system configuration file. Cannot execute the setting. Cannot find the copy source file. Cannot execute the setting. Cannot find the edit destination file. Cannot execute the setting. The edit destination file is for reading only. Cannot execute the setting. A file access error has occurred. Cannot execute the setting. Cannot find the copy source folder. Cannot execute the setting. The copy destination file name is incorrect. Cannot execute the setting. The copy destination file is for reading only. Cannot execute the setting. Do you want to apply the settings? The setting has not been applied. Do you want to exit without change? Please shut the system down, and then restart. 5-69 L2E-BW1444 5.12.3 Details of DSP error codes Error codes Contents Remarks 0x00000000 Normal end 0x00000001 Suspended end (Suspension command from application program) Set by scan DSP 0x00000002 Suspended end (EOS from scanner) Set by scan DSP 0x00010001 Parameter status (parameter start) error 0x00010002 Parameter read error 0x00010003 Parameter transfer error to scan DSP 0x00010004 Calibration data transfer error (including the waiting for parameter transfer end) 0x00010010 Scan DSP start (demand transfer) error 0x0002**** Parameter error (Lower significant 4 digits show an error Set by scan DSP parameter address) 0x0003*1*2*3*4 Measure buffer data error (See the following figure) *1 *2 # # # # # # 0 0 Set by scan DSP *3 *4 bit15 bit0 0 0 0 0 0 0 0 0 View header error Measure buffer overflow Link port error Channel number error Data higher-significant 16 bits error View number error Details of major buffer data errors 0x00050000 Inter-DSP communication error (Scan DSP answer time-out, etc.) 5-70 L2E-BW1444 5.13 Event viewer 5.13.1 Introduction The event viewer is a system event monitoring program assembled in Windows Xp Embedded as standard program. By using the event viewer, detailed information such as system start and error, etc. occurred during Presto system operation can be known. 5.13.2 Method of use (1) Start the system in system maintenance mode. (See 5.11) (2) An event viewer window is displayed by clicking [Start] button and selecting [Programs] [Administrative Tools] - [Event Viewer] from the start menu. 5-71 L2E-BW1444 (3) An event properties window is displayed by double-clicking a desired detailed information event. The event properties window is closed by clicking [OK] or [Cancel] button. (4) The event viewer is closed by selecting [File|Exit] from the event viewer screen menu. 5-72 L2E-BW1444 5.14 Touch panel calibration When the touch panel is replaced with another or the coordinates on the touch panel is out of alignment, make calibration of the touch panel. In this system, more than one touch panels are used. Locate the devices being connected and make calibration of an appropriate device. 5.14.1 Locating touch panels Touch panels available can be located by corresponding device drivers installed. If “Fujitsu Takamisawa Touch Panel driver (Windows2000/XP)” device is registered, it is considered to be touch panel 2 (see 5.14.3). If not, it is regarded as touch panel 1 (see 5.14.2). Click Device Manager on the Computer Management window, and click “Mice and other pointing devices” on the right side window to determine whether “Fujitsu Takamisawa Touch Panel driver (Windows2000/XP)” device is registered. To display Computer Management, click [Start] button, choose [Programs], [Administrative Tools], and [Computer Management] in this order, and click “Device Manager” on the left side window. 5-73 L2E-BW1444 5.14.2 Calibration of touch panel 1 (1) Start the system in system maintenance mode (see 5.11). (2) Replace the keyboard cable from CPU1 cable connector (83CN) to CPU0 cable connector (63CN). (3) Press to display Start menu. Using ↑ ↓ ← or →, choose [Programs], [Fujitsu Touch Panel], and [Touch Panel Calibration] in this order, and then press Enter key. (4) Keep depressing the ‘+’ mark on the upper left corner with a fingertip for five seconds, and release it. The ‘+’ will move to the next. Repeat this for nine points, upper left, upper center, upper right, middle left, center, lower left, lower center, and lower right, respectively, then press Enter . (5) At the bottom of screen, the following message appears. Press [Enter] to apply the calibration. (6) After replacing the keyboard cable from CPU0 cable connector (63CN) to CPU1 cable connector (83CN), choose [Start] and [Start the CT System] to start CT system. 5-74 L2E-BW1444 5.14.3 Calibration of touch panel 2 (1) Start the system in system maintenance mode (see 5.11). (2) Replace the keyboard cable from CPU1 cable connector (83CN) to CPU0 cable connector (63CN). (3) Press to display Start menu. Using ↑ ↓ ← or →, choose [Programs], [Fujitsu Touch Panel], and [Touch Panel Calibration] in this order, and then press Enter key. (4) Choose [Calibration]. 5-75 L2E-BW1444 (5) Click [Calibration Now] button. (6) Keep depressing the ‘+’ mark on the upper left corner with a fingertip for five seconds, and release it. The ‘+’ will move to the next. Repeat this for nine points, upper left, upper center, upper right, middle left, center, lower left, lower center, and lower right, respectively, then press 5-76 Enter . L2E-BW1444 (7) At the top of screen, the following message appears. Press [Update] to apply the calibration. (8) After replacing the keyboard cable from CPU0 cable connector (63CN) to CPU1 cable connector (83CN), choose [Start] and [Start the CT System] to start CT system. 5.15 Cache off setting of MO drive When replacing the MO drive with another, set the write cache setting to OFF. REFERENCE Cache off setting is reflected after the CT system is restarted. (1) Switch the mode to system maintenance mode (see 5.11). (2) Set the application installation disk (V1.32 or later) in the CD drive. (3) On the Windows Explorer, open E: drive and double-click “DisableMOWriteCache.exe”. (4) When Confirm dialog box appears, click [Yes] button. (5) Choose [Start] and [Start the CT System] to start the CT system. 5-77 L2E-BW1444 5.16 Tube current graph Tube current value graph is a function which chronologically displays tube current values recorded in the Raw data that was taken by the CT unit. 5.16.1 Start/Exit (1) Click [System/Exit] button on the main menu, and select [Maintenance], [Services], and [Tube Current Graph] on the menu displayed. (2) Tube Current Graph dialog is displayed. (3) Click ‘x’ button on the upper right, or click [Exit] of [File] menu to exit the dialog. 5-78 L2E-BW1444 5.16.2 Opening Raw data (1) Click [Open] from [File] menu. (2) The Open RawData dialog is displayed. The Open RawData dialog is the one that lists patient information and series information tables vertically. When clicking any one line on the patient information table, series information just below the line is switched to the information of the selected patient. By clicking the series you wish to view on the series table and then clinking [OK] button, tube current values recorded in the Raw data of the selected series are displayed. Click [Cancel] button to close the dialog box without doing anything. Note: The dialog illustrated above is an example, which may differ from that actually displayed. (3) Press [OK] to display Reading Data dialog. While reading RawData, reading progress can be seen on the dialog. This dialog is displayed until [Cancel] button is clicked to abort reading RawData, and disappears automatically when the processing is completed. 5-79 L2E-BW1444 (4) Graph of tube current values is displayed. In addition to the graph, user-set tube current value, and maximum, minimum, and average tube current values are displayed at the bottom of the screen. Note: The dialog illustrated above is an example, which may differ from that actually displayed. 5.16.3 Viewing chart Immediately after reading RawData, allowable length of the horizontal axis on the screen is up to ten seconds. Where entire scanning time of the series selected exceeds ten seconds, entire graph can be displayed by using the scroll bar located at the bottom of the graph to scroll the time axis sideward. (1) Scaling along time axis To take a closer look of the graph, time axis can be scaled up or down by using up/down buttons of the time axis scaling list box. In the list box, length of time axis is displayed on the left hand side of the buttons. Pressing the up arrow button will scale down the graph; while pressing the down arrow button will scale up the graph. The minimum width of the time axis is 0.1 seconds, and the maximum 100 seconds. Note: If total scanning time of the series which was read is below 100 seconds, length of time axis may not set to 100 seconds. 5-80 L2E-BW1444 (2) Scaling along tube current value axis To take a closer look of the graph, tube current axis can be scaled up or down by using up/down buttons of the tube current axis scaling list box. In the list box, length of tube axis is displayed on the left hand side of the buttons. Pressing the up arrow button will scale down the graph; while pressing the down arrow button will scale up the graph. The minimum width of the tube curent axis is 1 mA, and the maximum 500 mA. Note: If the maximum tube current value of the series which was read is below 500 mA, length of tube current value axis may not set to 500 mA. 5-81 291 L2E-BW1444 Section 6 Software installation 6.1 Summary of software installation Software installation comprises [system installation] and [application installation]. In [system installation], CPU0 and CPU1 systems are installed on DOS base. This applies to Windows Xp Embedded system and system environment of various drivers, etc. In the application installation, software is installed to CPU0 system and CPU1 system after system installation. This applies to the development application and various short cuts. The install flow of software is shown below. (1) Perform service backup ( saving of user data and service data) without fail before installing software. *1 (2) After installing software, restore the data that have been saved in (1), and restart the system. Set the system configuration. *2 (3) Perform both CPU0 and CPU1 without fail when installing the system. (4) Data base of image/raw data must be reconstructed when the system installation mode is SYSTEM INSTALL (See chapter 6.2) (when image/raw data are not initialized). Perform the image table resetting after installing the application. *3 (5) If the application installation only was done in the system where DICOM transmission software is installed, delete the DICOM transmission software. *4 6-1 L2E-BW1444 START *1 User data and service data saving (Chapter 5.10) Not executed System installation Executed CPU1 installation (Chapter 6.2.1) CPU0 installation (Chapter 6.2.2) Computer name change (3.18 ) Application installation (Chapter 6.3) Not installed Whether system is installed or not *4 Deletion of DICOM transmission software (3.19.2) Installed Setting of system serial number (Chapter 6.4) Setting of LAN card (Chapter 3.19.1) Entry of DICOM transmission software (Chapter 3.19.3) *2 Restore of user data and service data (Chapter 5.10) and system configuration setting (Chapter 5.5) *3 Is the system installation mode “SYSTEM INSTALL” ? NO YES Image table reset ( [System / Exit] - [Image Table Resetting] ) END 6-2 L2E-BW1444 6.2 System installation The system install work flow differs according to whether HD is replaced or not and whether data ( images/Raw Data ) being saved in HD are initialized or not as shown below. START Not replaced Whether CPU1 side HD is replaced or not Replaced RAID disk initialization (Chapter 7.1) Not initialized Whether Images are initialized or not Initialized CPU1 system installation in DISK INITIALIZE AND SYSTEM INSTALL mode CPU1 system installation in SYSTEM INSTALL mode Not replaced Whether CPU0 side HD is replaced or not Replaced Initialized Whether raw data are initialized or not Not initialized CPU0 system installation in DISK INITIALIZE AND SYSTEM INSTALL mode CPU0 system installation in system install mode END <Important:Difference of install modes (DISK INITIALIZE AND SYSTEM INSTALL and SYSTEM INSTALL> * DISK INITIALIZE AND SYSTEM INSTALL Prepare disk partition and install the system. Select this mode when a unused disk or image/raw data is initialized or when the disk partition is broken. * SYSTEM INSTALL Leave the disk partition other than basic partition as it is, and perform the system installation only. Select this mode when the disk partition is free of trouble and image/raw data are not initialized. * It is not necessary to equalize the install mode on CPU0 side and CPU1 side. 6-3 L2E-BW1444 6.2.1 CPU1 system installation When CPU1 side disk has been replaced, initialize RAID disk (Chapter 7.1) before starting the system installation. (1) Insert a NO BOOT FD into the FD drive. (2) Turn on the OC side power supply, and soon, insert a “CPU1 SYSTEM INSTALL DISK” into the CD drive. (3) After CD has been booted and the install menu of CPU1 system has been displayed, perform Step 1 of install menu Input ‘1’ ( when the disk has been replaced or images are deleted) or ‘2’ ( when images are left undeleted), and press Enter key. If CD is not booted, press CTRL+ALT and DEL keys to restart the system. *** CPU1 Install Menu *** ** Step1 ** 1. DISK INITIALIZE AND SYSTEM INSTALL 2. SYSTEM INSTALL *** Step2 *** 3. TEMP PARTITION INSTALL *** Other *** 4. QUIT Enter your choice (1, 2, 3 or 4)= (4) After installing CPU1 system, restart the system by pressing CTRL + ALT+ DEL keys. (5) After CD has been booted and the install menu of CPU1 system has been displayed, perform Step 2 of install menu. Input ‘3’ and press Enter key. REFERENCE The following error may be displayed in Step 2. In such a case, press Enter key and continue processing. Error #1822 Error:No patitions selected. OK 6-4 L2E-BW1444 (6) After the end of installation, take out the installed CD. Restart the system by pressing CTRL + ALT + DEL keys and make sure that CPU1 side is started normally. REFERENCE 1. Detection wizard of new hardware The following dialog to install the software necessary for the hardware may be displayed by restarting the system. These message may be displayed in certain instances a few minutes after the system startup. In the event that a message is displayed after starting the CPU0 installation, be sure to install the driver after the CPU0 is installed. (1) In the case of Ethernet Controller: In this case, set the supplied “CG-LAPCIGT2 Gigabit Ethernet PCI Adapter” CD. The hardware data is read from the CD and installation is started. The dialog indicating completion of the installation is displayed. Click the [Finish] button and take out the CD. (2) In the case of Multi-media audio controller: In this case, press Enter key. 6-5 L2E-BW1444 Then, the following dialog is displayed. press Enter key. 2. System setting change dialog The following figure may be displayed when restarting the system. In such a case, focus [Yes] button by ← → keys and press Enter The system is restarted automatically. If key. Enter key was pressed when [No] button is focused, the system does not restart automatically. After shutting down the power supply, turn it on again. 6.2.2 CPU0 system installation (1) After installing CPU1 system, insert the CPU0 SYSTEM INSTALLATION DISK into the CD drive. (2) After taking out NO BOOT FD, insert boot FD “DISK INITIALIZE AND SYSTEM INSTALL BOOT FD” (when an unused disk and raw data are deleted) or “SYSTEM INSTALL BOOT FD”( when raw data are left undeleted). (3) By pressing the CPU0 side reset switch (3SW) mounted on the left side of the hard disk, the installation is started automatically after booting. (4) After the installation, take out the boot FD and install CD, and restart the system by pressing the CPU0 side reset switch. REFERENCE The following dialog may be displayed when restarting the system. In such a case, touch [Yes] button. The system is restarted automatically. If [NO] button was touched , turn on the power supply again after shutting it down once. 6-6 L2E-BW1444 6.3 Application installation (1) Turn on the maintenance mode (See Chapter 5.1) and restart the CT unit in the system maintenance mode (See Chapter 5.11). (After the system installation, the system is already set to the system maintenance mode.) (2) By clicking the [Start] button and selecting [Application Installation] from the start menu, a confirmation message dialog is displayed. (3) Insert “Application Install Disk” into the CD drive, and select [Yes]. The application installation is started. (4) After the installation a confirmation dialog is displayed. Then take out the application install disk. (5) When the application installation only was done, select [Yes] to start the CT system. When the system installation was done, select [No] to proceed to the setting of system serial number (Chapter 6.4). 6-7 L2E-BW1444 6.4 Setting of System Serial No. (only when the system has been installed) (1) By clicking [Start] button and selecting [System Serial No. Setting] from the start menu, the machine serial number setting window is displayed. (2) Input the serial number (4 digit numeric numbers) described in the nameplate on the OC rear panel or the scanner rear cover, and click [Complate] button. 6-8 299 L2E-BW1444 Section 7 RAID disk 7.1 Decision of RAID disk Either High Point RAID utility or Promise RAID utility composes the array disk as a BIOS setting utility. Judge which utility is assembled according to the following procedure. (1) Insert NO BOOT FD into the FD drive. (2) By turning on the OC side power supply , the following message is displayed in the course of start-up. Judge which is assembled. If the display could not be read out due to its switching, restart the system by pressing Ctrl + Alt + Delete keys. In case of High Point RAID HighPoint Technologies, Inc. HPT370/372 BIOS Setting Utility v2.31 (c) 2000-2002 HighPoint Technologies, Inc. All rights reserved. Press <CTRL><H> to run BIOS Setting Utility Scan Devices. Please Wait... In case of Promise RAID Fast Trak TX2000 (tm) BIOS Version 2.00.0.33 (c) 1995-2002 Promise Technology, Inc. All rights reserved. No Array is defined.................... Press <Ctrl-F> to enter FastBuild (tm) Utility or Press <ESC> to continue booting ... 7-1 L2E-BW1444 7.2 High Point RAID 7.2.1 Initialization of RAID When a hard disk has been replaced, create an array disk according to the following procedure before installing software. (1) Insert NO BOOT FD into FD drive. (2) By turning on the OC side power supply, the following message is displayed in the course of start-up. Press <Ctrl><H> to run BIOS Setting Utility Scan Devices. Please wait ... Display a BIOS setting screen by pressing Ctrl + H keys. Help Menu Channel Status (3) Proceed to (4) when the Main Menu has been displayed in the Menu area or proceed to (27) in other cases. Main Menu 1. Create Array 2. Delete Array 3. Create/Delete Spare 4. Select Boot Disk 7-2 L2E-BW1444 (4) When “Channel Status” of the Channel Status area is BOOT in the disk, the Main Menu is displayed in the Menu area. Select [4.Select Boot Disk] by ↑ ↓ keys and press Enter key. Channel Status Channel Drive Name Array Name Mode Size(GB) Primary Master: xxxxxxxxxx ATA/133 xx.xx BOOT Primary Slave: No Drive ATA/133 xx.xx HDD1 or Hidden Secondary Master: xxxxxxxxxx Secondary Slave: Status No Drive (5) Make sure that “Status” of Channel Status of Channel Status area is displayed as shown below. Channel Status Channel Drive Name Array Name Mode Size(GB) Primary Master: xxxxxxxxxx ATA/133 xx.xx HDD0 Primary Slave: No Drive ATA/133 xx.xx HDD1 or Hidden Secondary Master: xxxxxxxxxx Secondary Slave: Status No Drive (6) Proceed to (7) when Array Name of Channel Status in Channel Status area has been set to either “Primary Master” or “Secondary Master”. Proceed to (9), if not set. Channel Status Channel Drive Name Array Name Mode Size(GB) Status Primary Master: xxxxxxxxxx ATA/133 xx.xx HDD0 Primary Slave: No Drive Secondary Master: xxxxxxxxxx Secondary Slave: RAID_1_0 ATA/133 xx.xx HDD1 or Hidden No Drive (7) The Main Menu is displayed on the Menu area. Select [2. Delete Array] by ↑ ↓ keys and press Enter key. Main Menu 1. Create Array 2. Delete Array 3. Create/Delete Spare 4. Select Boot Disk (8) The following message is displayed. Press Y key. WARNING: All data on the Array will be deleted. Continue? Press Y/N 7-3 L2E-BW1444 (9) Make sure that “Array Name” of Channel Status of Channel Status area has not been set and “Status” is “HDD0” and “HDD1”. Regarding “HDD0” and “HDD1”in “Status”, “Primary Master” may become “HDD1” and “Secondary Master” may become “HDD0”. Channel Status Channel Drive Name Primary Master: xxxxxxxxxx Primary Slave: No Drive Array Name Mode Size(GB) Secondary Master: xxxxxxxxxx Secondary Slave: Status ATA/133 xx.xx HDD0 ATA/133 xx.xx HDD1 No Drive (10) Select [1. Create Array] from the Main Menu of the Menu area by ↑ ↓ keys and press Enter key. Main Menu 1. Create Array 2. Delete Array 3. Create/Delete Spare 4. Select Boot Disk (11) Menu area changes to Create RAID menu. Select [1. Array Mode RAID 0 ( Stripping) ] by ↑ ↓ keys and press Enter key. Create RAID 1. Array Mode RAID 0 (Striping) 2. Array Name RAID_0_0 3. Select Disk Drives 4. Block Size 64K 5. Start Creation Process (12) The following message is displayed. and press Enter Select [ RAID 1 (Mirroring) for data protection] by ↑ ↓ keys key. RAID 0 (Striping) for performance RAID 1 (Mirroring) for data protection RAID 0/1 (Mirrored Striping) JBOD(Volume) (13) The following message is displayed. Select [Create only ] by ↑ ↓ keys and press Enter key. Create only Create and duplicate 7-4 L2E-BW1444 (14) Menu area changes to RAID 1 (Mirroring) menu. Select [2. Array Name RAID_1_0] by ↑ ↓ keys and press Enter key. RAID 1 (Mirroring) 1. Array Mode RAID 1 (Mirroring) 2. Array Name RAID_1_0 3. Select Disk Drives 4. Start Creation Process (15) The following message is displayed. Default is set by pressing Enter key. Please input the name of the array. Name: (16) Select [3. Select Disk Drives] by ↑ ↓ keys in RAID1 (Mirroring) menu, and press Enter key. RAID 1 (Mirroring) 1. Array Mode RAID 1 (Mirroring) 2. Array Name RAID_1_0 3. Select Disk Drives 4. Start Creation Process (17) Select [Primary Master] in Channel Status area and press (18) Select [Secondary Master] in Channel Status area and press (19) If the following message is displayed, press Y Enter Enter key. key. key. WARNING: A valid partition is found at the Primary Master disk you have selected. Continue? Press Y/N or WARNING: A valid partition is found at the Secondary Master disk you have selected. Continue? Press Y/N 7-5 L2E-BW1444 (20) Select [4.Status Creation Process] by ↑ ↓ keys in RAID1 (Mirroring) menu in Menu area, and press Enter key. RAID 1 (Mirroring) 1. Array Mode RAID 1 (Mirroring) 2. Array Name RAID_1_0 3. Select Disk Drives 4. Start Creation Process (21) The following message is displayed. Press Y key. WARNING: All data on the selected disk will be deleted. Continue? Press Y/N (22) Select [4.Select Boot Disk] from Main Manu of Menu area by ↑ ↓ keys and press Enter key. Main Menu 1. Create Array 2. Delete Array 3. Create/Delete Spare 4. Select Boot Disk (23) Select [Primary Master] of Channel Status area by ↑ ↓ keys and press Enter key. (24) Press [ESC] key after making sure that the following is displayed in Channel Status area. Channel Status Channel Drive Name Array Name Mode Size(GB) Status Primary Master: xxxxxxxxxx RAID_1_0 ATA/133 xx.xx BOOT Primary Slave: No Drive RAID_1_0 ATA/133 xx.xx Hidden Secondary Master: xxxxxxxxxx Secondary Slave: No Drive (25) The following message is displayed. Press Y key. Are you sure you want to exit the BIOS utility? (Yes/No) (26) Set CPU1 system install disk to the CD drive at once, and install the CPU1 system (See 6.2.1). 7-6 L2E-BW1444 (27) When Broken RAID 1 has been displayed in Menu area in starting the system after replacing the HD, select [2. Destroy the Mirroring Relationship] by ↑ ↓ keys and press displayed, repeat the procedure starting with (1) by press Ctrl + Enter Alt + key. If Critical RAID1 is Delete keys. Broken RAID 1 1. Power off and replace the failed drive 2. Destroy the Mirroring Relationship 3. Choose replacement drive and rebuild 4. Continue to boot Remaining members of the failed array Channel Drive Name Array Name Mode Size(GB) Primary Master: xxxxxxxxxx RAID_1_0 ATA/133 xx.xx (28) Perform the procedure in and after (4) when Main Menu has been displayed in Menu area. Repeat the procedure starting with (1), if not displayed. Main Menu 1. Create Array 2. Delete Array 3. Create/Delete Spare 4. Select Boot Disk 7-7 L2E-BW1444 7.2.2 Remedial measures against the occurrence of troubles If RAID disk is in trouble, Critical Menu or Broken Menu is displayed when turning on the OC side power supply. The critical menu is displayed when non-conformance occurs between array disks, while the broken menu is displayed when a hardware trouble occurs. If either critical or broken menu has been displayed, replace CPU1 side disk according to the following procedure. Critical status may be displayed due to a connection failure or a contact failure of IDE cable and / or power cable. Particularly be careful when replacing the disk. 7.2.2.1 In case of Critical Menu (1) The following message is displayed. Select [2. Continue to boot] by ↑ ↓ keys and start up the CT system by pressing Enter key. Critical RAID 1 1. Duplicate now 2. Continue to boot Array members Channel Drive Name Array Name Mode Size(GB) Primary Master: xxxxxxxxxx RAID_1_0 ATA/133 xx.xx Secondary Master: xxxxxxxxxx RAID_1_0 ATA/133 xx.xx (2) Save user data and service data into MO by service backup (Section 5.10). (3) Back up necessary images /raw data. (4) Shut down the system. (5) Replace two CPU1 side hard disks. (6) After initializing the RAID disk (Section 7.2.1), install the software (Section 6 ). 7-8 L2E-BW1444 7.2.2.2 In case of Broken Menu (1) The following message is displayed. the CT system by pressing Enter Select [4. Continue to boot] by ↑ ↓ keys and start up key. Broken RAID 1 1. Power off and replace the failed drive 2. Destroy the Mirroring Relationship 3. Choose replacement drive and rebuild 4. Continue to boot Remaining members of the failed array Channel Drive Name Array Name Mode Size(GB) Primary Master: xxxxxxxxxx RAID_1_0 ATA/133 xx.xx (2) Save user data and service data into MO by service backup (Section 5.10). (3) Back up necessary images /raw data. (4) Shut down the system. (5) Replace two CPU1 side hard disks. (6) After initializing the RAID disk (Section 7.2.1), install the software (Section 6 ). 7-9 L2E-BW1444 7.3 Promise RAID 7.3.1 Initializing the RAID disk When a hard disk has been replaced, create an array disk according to the following procedure before installing software. (1) Insert NO BOOT FD into FD drive. (2) Proceed to (3) by pressing Ctrl + F keys when the following message is displayed by turning on the OC side power supply in the course of start-up. Fast Trak TX2000 (tm) BIOS Version 2.00.0.33 (c) 1995-2002 Promise Technology, Inc. All rights reserved. No Array is defined.................... Press <Ctrl-F> to enter FastBuild (tm) Utility or Press <ESC> to continue booting ... Proceed to (3) by pressing Ctrl + F keys when the following message is displayed. Fast Trak TX2000 (tm) BIOS Version 2.00.0.33 (c) 1995-2002 Promise Technology, Inc. All rights reserved. ID MODE 1 * 1x2 Mirror SIZE xxxxxM TRACK-MAPPING xxxx/xxx/xx STATUS Functiona Press <Ctrl-F> to enter FastBuild (tm) Utility … Proceed to (5) by pressing ID MODE 1 * 1x2 Mirror Ctrl + SIZE xxxxxM F keys when the following message is displayed. TRACK-MAPPING xxxx/xxx/xx STATUS Critical Problem is detected with Array : 1 Critical Status: A disk member of a mirrored array has failed or is not responding. The array is still functional, but fault tolerance is disabled. Before continuing, power off system and confirm that the drives and cables are properly attached before replace the failed drive and rebuilding the array. 1) Identify which drive has failed eith <3> Define Array menu option. 2) Power off the system, replace the failed drive. 3) Restart the system and enter the FastBuild (tm) setup menu. 4) Choose the <5> option to rebuild the array with replacement drive. Press <Ctrl-F> to enter FastBuild (tm) Utility or Press <ESC> to continue booting... 7-10 L2E-BW1444 (3) Main Menu is displayed. Select [Auto Setup] by 1 [ Main Menu ] Auto Setup ...………........... View Drive Assignments .... Define Array .........……….. Delete Array ........………… Rebuild Array ...........……… Controller Configuration ….. key. [1] [2] [3] [4] [5] [6] [ Keys Available ] Press 1..6 to Select Option [ESC] Exit (4) Proceed to (5) by pressing Space key when the following message is displayed or proceed to (11) if not displayed. No Available Disks! Any Key to Continue ……. (5) Select [Delete Array] from the Main Menu by 4 key. [ Main Menu ] Auto Setup ...………........... View Drive Assignments .... Define Array .........……….. Delete Array ........………… Rebuild Array ...........……… Controller Configuration ….. [1] [2] [3] [4] [5] [6] [ Keys Available ] Press 1..6 to Select Option (6) Delete Array Menu is displayed. Press [ESC] Exit Delete key. [ Delete Array Menu ] Array No RAID Mode Array 1 Array 2 Array 3 Array 4 Total Drv Mirror ------------- Capacity(MB) 2 ------------- 40000 ------------------- Status Functional ---------------- [ Keys Available ] [↑] Up [↓] Down [ESC] Exit [Del] Delete 7-11 L2E-BW1444 (7) View Array Definition Menu and Drives Assignments are displayed. When “Status” is “Functional” [ View Array Definition Menu ] Array No RAID Mode Array 1 Total Drv Capacity(MB) 2 40000 Mirror Stripe Block : Not Available Status Functional Gigabyte Boundary : ON [ Drives Assignments ] Channel : ID Drive Model 1 : Mas WDC WD400EB-00CPF0 2 : Mas WDC WD400EB-00CPF0 Capacity(MB) 40020 40020 When “Status” is “Critical” [ View Array Definition Menu ] Array No RAID Mode Array 1 Total Drv Capacity(MB) 2 40000 Mirror Stripe Block : Not Available Status Critical Gigabyte Boundary : ON [ Drives Assignments ] Channel : ID Drive Model Capacity(MB) 1 : Mas WDC WD400EB-00CPF0 40020 ? :? failed or disconnected … (8) The following message is displayed. Press Ctrl + Y keys. Are you sure you want to delete this array? Press Ctrl-Y to Delete, or others to abort .. (9) following message is displayed. Press Y keys. Would you like to clean the MBR and reserved sector? Doing so will result in a total loss of all data. < Press Y to continue or N to return to menu > (10) The following message is displayed. Press Space key. Reserved sectors wiped …. <Press Any Key to REBOOT> The system reboot and then return to (2) 7-12 L2E-BW1444 (11) Auto Setup Options Menu is displayed. Press Space key repeatedly until “Mode” becomes “Mirror”. Proceed to (12) after “Mirror” has been selected. If “Mirror” cannot be selected, turn off the power supply and check the connecting condition of the disk. Then, repeat the operation again starting with (2). [ Auto Setup Options Menu ] Optimize Array for: Typical Application to use: Security Not Available [ Array Setup Configuration ] Mode .....................................…………….. Spare Drive ..........................…………….... Drive(s) Used in Array .................………... Array Disk Capacity (Size in MB) ........….. Mirror 0 2 40000 Please make sure if the setting is conect. [ Keys Available ] [↑] Up [↓] Down [←,→,Space] Change Option [ESC] Exit [CTRL-Y]Save (12) Press Ctrl + Y keys. (13) The following message is displayed. Press N key. Warning: 64K of Disk Space is reserved at the end of the Source Disk. If Source Disk is at or near full capacity, Data Loss can occur! All data on Destination Disk will be LOST! Press the "Y" key to continue with Duplication or "N" key to abort. (14) The following message is displayed. Press Space key. Array has been created.... <Press Any Key to REBOOT> (15) Set CPU1 system install disk into the CD drive at once and install the CPU1 system (See Section 6.2.1). 7-13 L2E-BW1444 7.3.2 Remedial measures against the occurrence of troubles If RAID disk is in trouble, Critical Status is displayed when turning on the OC side power supply. Critical Status is displayed when non-conformance occurs between array disks or a hardware trouble occurs. If Critical Status has been displayed, replace CPU1 side disk according to the following procedure. Critical Status may be displayed due to a connection failure or a contact failure of IDE cable and/or power cable. Particularly be careful when replacing the disk. (1) The following is displayed. Initialize the CT system by pressing [ESC] key. ID MODE 1 * 1x2 Mirror SIZE xxxxxM TRACK-MAPPING xxxx/xxx/xx STATUS Critical Problem is detected with Array : 1 Critical Status: A disk member of a mirrored array has failed or is not responding. The array is still functional, but fault tolerance is disabled. Before continuing, power off system and confirm that the drives and cables are properly attached before replace the failed drive and rebuilding the array. 1) Identify which drive has failed eith <3> Define Array menu option. 2) Power off the system, replace the failed drive. 3) Restart the system and enter the FastBuild (tm) setup menu. 4) Choose the <5> option to rebuild the array with replacement drive. Press <Ctrl-F> to enter FastBuild (tm) Utility or Press <ESC> to continue booting... (2) Save user data and service data into MO by service backup (Section 5.10). (3) Back up necessary images /raw data. (4) Shut down the system. (5) Replace the two hard disks on CPU side. (6) After initializing RAID Disk , install software (Section 6). 7-14 313 L2E-BW1444 Section 8 1. Scanner firmware version up method VMCPU, VSCPU, VTCPU, VTCPU2, RXCPU, XCPU and VATCPU mean programs for controls of the CT system. 2. The following items are sent in case of version up. (1) CD-ROM containing the firmware down load application and version up file (2) RS-232C cable (straight) 8.1 Preparation for version up 1. Prepare a PC where RS-232C port (USB serial conversion adapter is allowable) is employable. 2. If the PC is not provided with a CD-ROM drive, copy the contents of the distributed CD-ROM into a folder of the PC. 8.1.1 Preparation for version up of VMCPU, VSCPU, and VTCPU/VTCPU2 (1) Open the small cover on the left side of the scanner as viewed from the front. (2) Turn on the scanner power supply, and connect each connector of PC and SWBD PCB (See Fig.8.1 ) by an RS-232C cable according to the version up firmware , referring to Table 8-1. (3) Set the rotary switch (SW16) to the designated position, and press CPU RESET (SW17) of SWBD PCB twice. Make sure that a beep sounds at an interval of one second. Table 8-1 Connector and rotary switch setting Firmware VMCPU VSCPU VTCPU/VTCPU2 Connector CNSW2 MCPU CNSW3 SCPU CNSW4 TCPU Switch position D (13) E (14) F (15) CPU RESET (SW17) CNSW2 MCPU Rotary switch (SW16) CNSW3 SCPU CNSW4 TCPU Fig. 8-1 8-1 SWBD PCB L2E-BW1444 8.1.2 Preparation for version up of RXCPU and VATCPU (1) Open the front cover of the scanner. (2) Turn on the scanner power supply, and connect PC and the connector of RXCPU PCB (See Fig.8.2 ) by an RS-232C cable according to the version up firmware , referring to Table 8-2. (3) Press CPU_LD (SW3) while pressing CPU_RST (SW2) of RXCPU PCB , and make sure that CPUMODE (LD6) lights. (4) Release CPU_LD (SW3) (5) Release CPU_RST (SW2) Table 8-2 Connector setting Firmware RXCPU VATCPU Connector CN_RCPU CN_ATMA CN_RCPU CN_ATMA CPUMODE (LD6) CPU_LD (SW3) Fig. 8-2 8.1.3 CPU_RST (SW2) RXCPU PCB Preparation for version up of XCPU. (1) Open the left cover of X-ray control unit. (2) Turn on the scanner power supply, and connect PC and CN_PCSUB connector of XCONT5 PCB ( Fig. 8-3) by an RS-232C cable. (3) Press CPU_LOAD SW (SW2) while pressing RESET SW (SW1) of XCONT5 PCB, and make sure that CPU_LOAD SW (SW2) flickers. (4) Release CPU_LOAD SW (SW2) (5) Release RESET SW (SW1) 8-2 L2E-BW1444 CN_PCSUB CPU_LOAD FPGA_READY (LD13) RESET SW (SW1) CPU_LOAD SW (SW2) Fig. 8-3 XCONT5 PCB 8.2. Down loading method (1) Insert a CD-ROM containing the following contents into the PC. ・ MainteTool_V***.exe (firmware down load application) ・ ***.mot (firmware version up file) ・ Other DLL files (2) Copy the contents of the distributed CD-ROM into a folder of the PC. (3) Start MainteTool_V***.exe (hereinafter referred to as MainteTool) of folder. Fig. 8-4 MainteTool start screen 8-3 L2E-BW1444 (4) Designate the communication port number being used at present in the Port No. column by clicking [SERIAL PORT (O)] button of MainteTool (The port number can confirm in device manager of PC). Fig. 8-5 Serial Port Setup screen (5) Start PROGRAM LOADER tool by clicking [PROGRAM LOADER (P)] button of MainteTool. Fig. 8-6 PROGRAM LOADER start screen (6) Designate the firmware version up file ( ***mot) of folder by clicking [WRITE FILE(W)] button of PROGRAM LODER tool. Fig. 8-7 Version up file designation screen 8-4 L2E-BW1444 (7) [Please reset in loader mode] dialog box is displayed by clicking [START (S)] button of PROGRAM LODER tool, and click [OK] button. Fig. 8.8 Download start screen When any error occurs before download is completed, reconfirm connection of the RS-232C cable and retry step (4). (8) After the end of down load, [Write finish] dialog is displayed. Click [OK] button. Fig. 8.9 Download end screen 8.3. Version up end processing 8.3.1 Version up end processing of VMCPU, VSCPU, and VTCPU/VTCPU2 (1) Set the rotary switch (SW16) of SWBD PCB to [0]. (2) Press CPU RESET (SW17) of SWBD PCB. 8-5 L2E-BW1444 8.3.2 Version up end processing of RXCPU and VATCPU (1) Press CPU_LD (SW3) while pressing CPU_RST (SW2) of RXCPU PCB, and make sure that CPUMODE (LD6) goes out. (2) Release CPU_LD (SW3). (3) Release CPU_RST (SW2). 8.3.3 Version up end processing of XCPU (1) Press CPU_LOAD SW (SW2) while pressing RESET switch (SW1) of XCONT5 PCB, and make sure that FPGA_RADY (LD13) lights. (2) Release CPU_LOAD switch (SW2). (3) Release RESET switch (SW1). 8.4 Version check (1) Check if the version is indicated by setting the rotary switch of the PCB shown in Table 8-3 to the designated position using the version up firmware. (except for XCPU) Table 8-3 Firmware VMCPU VSCPU VTCPU/VTCPU2 RXCPU VATCPU After check, set each switch to [0]. Rotary switch setting for version check PCB SWBD SWBD SWBD RXCPU RXCPU Rotary switch SW16 SW16 SW16 SW1 SW1 Switch position D (13) E (14) F (15) 8 (8) A (10) *After version check, set each rotary switch to [0] without fail. (2) After turning off the scanner power supply, disconnect the RS-232C cable, reset the cover as before, and check the operation by executing the normal scan, scanogram, and volume scan. If the Version-up program is completed unsuccessfully, retry the steps from “8.1 Preparation for Version-up”. 8-6 319 L2E-BW1444 Section 9 Scanner FPGA version up method 9.1 RXCPU, XCPU version up method The following items are sent in case of version up. (1) CD-ROM containing the FPGA down load application and version up file (2) RS-232C cable (straight) 1. Prepare a PC where RS-232C port (USB serial conversion adapter is applicable) is employable. 2. If the PC is not provided with a CD-ROM drive, copy the contents of the distributed CD-ROM into a folder of the PC. 3. Prepare RS-232C cable (straight). (1) Open the front cover of the scanner in case of RXCPU or the left cover of X-ray control unit in case of XCPU. (2) Turn on the scanner power supply. (3) Connect PC and the connector of each PCB by an RS-232C cable according to the version up FPGA, referring to Table 9-1. Table 9-1 Connector setting FPGA PCB Connector RXCPU RXCPU CN_RCPU XCPU XCONT5 CN_PCSUB CN_RCPU CPU_RST (SW2) Fig. 9-1 RXCPU PCB Connector 9-1 L2E-BW1444 CN_PCSUB RESET SW (SW1) CPU_LOAD SW (SW2) Fig. 9-2 XCONT5 PCB Connector (4) Insert the CD-ROM containing the following contents. * MainteTool_V***.exe (FPGA down load application) * ***.hexout (FPGA version up file) * Other DLL files (5) Copy the contents of the distributed CD-ROM into a folder of the PC. (6) Start the MainteTool_V***.exe (hereinafter referred to as MainteTool) of folder. Fig. 9-3 MainteTool start screen 9-2 L2E-BW1444 (7) Select the communication port number of PC being used in [Port No] column at present by clicking [SERIAL PORT (O)] of the MainteTool (The port number can confirm in device manager of PC). Fig. 9-4 Serial port setup screen (8) Start the FPGA LOADER TOOL by clicking [FPGA LOADER (F)] button of the MainteTool. Fig. 9-5 FPGA LOADER start screen (9) Designate the FPGA version up file ( ***.hexout) of folder by clicking [WRITE FILE (W)] button of the FPGA LOADER tool. Fig. 9-6 Version up designation screen 9-3 L2E-BW1444 (10) In case of RXCPU, release CPU RST (SW2) of RXCPU_PCB. In case of XCPU, press CPU_LOAD switch (SW2) twice while pressing RESET SW (SW1) of XCONT_PCB, and then, release RESET SW (SW1). Wait for longer than 40 seconds after leaving. (11) Start down loading by clicking [START (S)] button of the FPGA LOADER tool. Fig. 9-7 Download start screen (12) After down loading, [Write finish!] dialog is displayed. Click [OK] button. Fig. 9-8 Download end screen When any error occurs before download is completed, never turn the power OFF. Reconfirm connection of the RS-232C cable and retry from setp (9). (13) Turn off the main power supply (1FFB) of the X-ray control unit (WB) after shut-down operation on the operation panel, and disconnect RS-232C cable. After download is completed normally, be sure to turn the power OFF. (14) After starting the scanner power supply again, check the operation by executing the normal scan, scanogram, and volume scan, and then, reset the cover as before. 9-4 L2E-BW1444 9.2 Version up method of MFPGA, TFPGA, TXRFPGA and TXSFPGA For the version-up procedure of MFPGA, TFPGA and TXRFPGA (RFPGA) of the system with the MTCPU2 PCB installed, refer to 9.3. (1) Open the rear cover of the scanner in case of MFPGA and TFPGA or open the front cover and side cover (on the left side as viewed from the front) in case of TXRFPGA and TXSFPGA. (2) Make sure that the power supply is not turned on in the scanner and version up PCB. (3) Take out FPGA ROM of each PCB, and insert the sent ROM. Table 9-2 shows the addresses of FPGA ROM of each PCB. Table 9-2 FPGA ROM addresses Change ROM PCB ROM address ROM type MFPGA MTCPU1 16F DIP 8Pin TFPGA MTCPU1 15G DIP 8Pin TXRFPGA TXR5 6D DIP 8Pin TXSFPGA RXCPU 7D DIP 8Pin (4) After replacing, reset the covers, the cover, and check the operation by executing the normal scan, scanogram, and volume scan. 9-5 L2E-BW1444 9.3 Version-up procedure of MFPGA, TFPGA and RFPGA (for the system installed with the MTCPU2) The following parts are provided for this version-up. (1) CD-ROM containing firmware down-load application software and version-up file (2) RS-232C cable (straight) (1) Prepare a PC with an usable RS-232C port (USB serial conversion adaptor is acceptable). (2) In the case of the PC without CD-ROM drive, make copy of the contents in the provided CD-ROM in a folder of the PC in advance. (1) Open a small cover in the side cover of the scanner (left side in view from the front). (2) Turn on the power for the scanner system. (3) Connect the PC and the connector CNSW2 MCPU on the SWBD PCB with the RS-232C cable. CPU RESET (SW17) ) CNSW2 MCPU Fig. 9-9 SWBD PCB (4) Insert the CD-ROM containing the following programs in the PC. • MainteTool_V***.exe (FPGA down-load application) • ***.rpd (Version-up file for MFPGA, TFPGA and RFPGA) • Other files 9-6 L2E-BW1444 (5) Start MainteTool_V***.exe (hereafter called MainteTool) in the CD-ROM. Fig. 9-10 MainteTool Start-up Window (6) Click the [SERIAL PORT (O)] button in MainteTool and select the communication port No. being used in the PC in Port No. selection box. Fig. 9-11 Serial Port Set Up Window (7) Click the [FPGA LOADER (F)] button in MainteTool to start FPGA LOADER tool. Fig. 9-12 FPGA LOADER Start Window 9-7 L2E-BW1444 (8) Click the [WRITE FILE (W)] button in FPGA LOADER tool and select FPGA Version-up file (***.rpd) in the CD-ROM. Fig. 9-13 Version-up File Selection Window (9) Press the CPU RESET (SW17) button in the SWBD PCB once and release it, and then wait more than one minute. (10) Click the [START (S)] button in EFGA LOADER tool to start the download. Fig. 9-14 Download Start Window 9-8 L2E-BW1444 (11) As the [Write finish!] dialog is displayed after the download is completed, click [OK] button. Fig. 9-15 Download End Window If any error occurs before completion of the download, never shutdown the scanner power. Retry the procedures from the step (6). (12) After executing the shutdown process on the control console, turn of the main power supply of the X-ray control unit and disconnect the connected RS-232C cable. After successful completion of the download, be sure to shut down the main power. (13) After the system power is turned on again, perform normal scan, scanogram scan and volume scan to make sure of normal operation of the system, and then put back the cover. 9-9 328 L2E-BW1444 Section 10 Various setting methods of table 10.1 Preset height setting method 1. When the versions of various firmware have been updated (version up) or the table has been replaced, the preset height is held. Accordingly, no resetting is necessary. 2. When MTCPU1 (or MTCPU2) PCB has been replaced, the preset height is not held. Accordingly, resetting is necessary. Set the PRESET height in the exclusive service mode. The setting procedure is shown below. 1. Turn on the power supply during this work. 2. Table height indicating LED value should be more than 0. 3. If plural switches are pressed at the same time, press [RESET] button first without fail. 4. A buzzer does not always sound depending upon the system setting in the following procedure. In such a case, check LED only. (1) Set the table to a desired preset height (Table height indicating LED value is within a range of 0 ~ 250) by the vertical move switch. (2) Continue pressing PRESET switch while pressing RESET switch concurrently for longer than 5 seconds. A buzzer sounds once. Leave the switches when the RESET switch and table height LED start flickering. If the buzzer sounds three times or if LED does not start flickering after longer than 8 seconds, check the above cautions and repeat the above procedure starting with step (1). (3) After restarting the power supply, move the table to the HOME position and make sure that the table stops within a range of [-1 ~ +5mm] of the value being set by PRESET key. If the table does not stop as specified above, repeat the above procedure starting with step (1). (4) End 10-1 L2E-BW1444 10.2 Stop position accuracy setting method 1. When the versions of various firmware have been updated (version up), the preset stop position accuracy is maintained. Accordingly, no resetting is necessary. 2. When MTCPU1 (or MTCPU2) PCB or table has been replaced, the preset stop position accuracy is not held. Accordingly, resetting is necessary. Set and check the stop position accuracy only when the table stop position is deviated in the longitudinal operation of the table. The setting procedure is shown below. (1) Open the small cover and rear cover of the scanner side face ( left side as viewed from the front ). (2) After turning on the scanner power supply, set the rotary switch (SW16) of SWBD PCB (Fig.10-1 ) to [0], and No.4 of DIP switch 4 (DSW4) of MTCPU1 (or MTCPU2) PCB (Fig. 10-2,10-3) to [ON (1)]. CPU RESET (SW17) 7SEG LED RESET SW ROTARY SW (SW16) TABLE FORWARD SW PAT SW HI-SPEED SW TABLE BACK SW Fig. 10-1 SWBD PCB 10-2 L2E-BW1444 DIP SW4 (DSW4) DIP SW4 (DSW4) Fig. 10-2 MTCPU1 PCB Fig. 10-3 MTCPU2 PCB (3) Press CPU RSET (SW17) of SWBD PCB. (4) Set the rotary switch (SW16) of SWBD PCB to [F]. (5) If the work is done by one person only, put the phantom or the like on the table as a weight. If the work is done by two persons, one should ride on the table. (6) Stop the table at an optional forward position (reference: a position retreated from the longitudinal limit by 50cm) by TABLE FORWARD switch and TABLE BACK switch of SWBD PCB. Make sure that the 7-SEG LED value of SWBD PCB changes. (Be careful since a value exceeding 999.99 is displayed as a blank.) If the above value does not change, confirm the setting of DIP switch 4 (DSW4) and rotary switch (SW16) in step (2), and repeat step (5) by pressing CPU RESET (SW17) of SWBD PCB. (7) Clear the 7-segment LED value by pressing RESET switch of SWBD PCB once. (8) Continue pressing PAT switch of SWBD PCB until the table stops running. Release the switch when the table has stopped running. (9) Record a deviation of the stop position from [10.00] as a reference by 7-segment LED. (10) Repeat steps (7) ~ (9) ten times from the position. (11) Obtain an average value of ten times and calculate the stop position accuracy. Set [7] and [8] of DIP switch 4 (DSW4) of MTCPU1 (or MTCPU2) PCB as shown below according to the calculated stop position accuracy. 10-3 L2E-BW1444 Table 10-1 DIP switch 4 (DSW4) setting Stop position accuracy [mm] DSW4 No.7 DSW4 No.8 Index correction value ~+0.025 OFF(0) OFF(0) Specified value -0.00mm +0.026~+0.075 ON(1) OFF(0) Specified value -0.05mm +0.076~+0.125 OFF(0) ON(1) Specified value -0.10mm +0.126~ ON(1) ON(1) Specified value -0.15mm (12) After setting DIP switch 4 (DSW4), set the rotary switch (SW16) of SWBD PCB to [0], and then, press CPU RESET (SW17). If setting of DIP switch 4 (DSW4) has been changed by (11), repeat steps (4) ~ (10) again and make sure that the stop position accuracy has been improved. If the stop position accuracy is not improved, set both No.7 and No. 8 of DIP switch 4 (DSW4) to [OFF (0)], and repeat the above procedure starting with step (4) by pressing CPU RESET (SW17) of SWBD PCB. (13) Reset No. 4 of Dip switch 4 (DSW4) changed in step (2) to [OFF (0)] and rotary switch (SW16) to [0]. (14) After turning off the scanner power supply, reset the cover as before, and check the operation by executing the normal scan, scanogram, and volume scan. 10-4 332 L2E-BW1444 Appendix 1 System configuration setting items System Config Type System Item Set Data X-Ray Tube Presto Default 5.0M(Dual Focus) 3.5M(Dual Focus) Tube Shift ○ None Arcing Recovery Install ○ OFF ○ ON Table Standard ○ Extended Adaptive-mA* None ○ OFF ON Bitmap Wall Paper None ○ Wave Orientation Diagram Gantry to left ○ Gantry to right Date Format Date Format MM/dd/yyyy ○ dd/MM/yyyy yyyy/MM/dd * Adaptive-mA will can be used by installing Adaptive-mA Function Software in case Application software version is VI.32. Words in column “Set Data” are defined as follows: None: Set this parameter on the system which does not have Adaptive-mA-compatible ROM. OFF: Set this parameter on the system which has Adaptive-mA-compatible ROM, to set the Adaptive-mA function to OFF temporarily. ON: Set this parameter on the system which has Adaptive-mA-compatible ROM, to set the Adaptive-mA function to ON. Appendix-1 L2E-BW1444 Appendix 2 How to initialize HU If a time zone is changed at the Date/Time Setting, HU capacity may keep 100% more than in 10 hours. In that case, please initialize HU capacity to 0% by the following procedures: NOTE: This HU initialization decreases HU capacity intensively, and it may damage tube. Please do them under X-ray tube cool fully(i.e. HU capacity is 0). (1) Turn on the maintenance mode (see 5.1) and restart the CT system in "system maintenance mode" (see 5.11). (2) Insert "application install disk"(V1.32 or later) into CD drive. (3) Open up E: drive from "My Computer" on the desk top at CRT side. (4) Double click the "InitializeHU.bat" file in the "InitializeHU" folder on E: drive to initialize HU. (5) Take out "application install disk" from CD drive. (6) Double click "Start the CT system" shortcut icon on the desk top to return to the release mode(user mode). (7) Make sure that HU capacity becomes 0% on the touch panel. End. E Appendix-2 334 L2E-BW1444