Instruction manual Geratehandbuch Mode d'emploi 35 MHz Dual channel oscilloscope 35 MHz Zweikanal - Oszilloskop Oscilloscope 35 MHz a double trace PM 3218 9444 032 1 8.. 1 PHILIPS 9499 443 00102 )90625 IMPORTANT In correspondence concerning this instrument, please quote the type number and serial number as given on the type plate. WICHTIG Bei Schriftwechsel uber dieses Gerat wird gebeten, die genaue anzugeben. Diese befinden sich auf dem Typenbezeichnung und Geratenummer die Leistungsschild. IMPORTANT RECHANGE DES PIECES DETACHEES Dans votre correspondance le numero de type Note: et !e dans vos reclamations numero de The design of Consequently, contained et this serie qui sont instrument this in this (Reparations) is instrument se rapportant a cet appareil, veuillez marques sur la subject to continuous development may TOUJOURS indiquer plaquette de caracteristiques. and improvement. incorporate minor changes in detail from the information manual. Bemerkung: Die Konstruktion und Schaltung dieses Gerats wird standig weiterentwickelt und verbessert. Deswegen kann dieses Gerat von den in dieser Anieitung stehenden Angaben abweichen. Remarques. Cet appareil est details / objet de developpements et ameliorations continuels. mmeurs peuvent differer des informations donnees dans la En consequence certains , presente notice d'emploi et d'entretien. N.V. PHILIPS' GLOEILAMPENFABRIEKEN - EINDHOVEN - THE NETHERLANDS PRINTED IN 1979 THE NETHERLANDS 3 Contents 1. GENERAL INFORMATION 9 1.1. INTRODUCTION 9 1.2. CHARACTERISTICS 10 1.2.1. C.R.T. 10 1.2.2. Vertical or Y-axis 10 1.2.3. Horizontal or X-axis 11 Main time base 11 1.2.5. Delayed time base 11 1.2.6. X 12 1.2.7. Triggering of the main time base 12 Triggering of the delayed time base 12 1.2.9. Calibration generator 12 1.2.10. Power supply 13 1.2.11. Environmental characteristics 13 1.2.12. Mechanical data 13 1.2.13. Z-mod input 13 1.3. ACCESSORIES 14 1.3.1. Supplied with the instrument 14 1.3.2. Optional 14 1.4. BLOCK DIAGRAM DESCRIPTION Y Channel 15 1.4.1. 1.4.2. Main time-base triggering 15 1 .4.3. Main time-base 15 1 .4.4. Hold-off circuit 15 1.4.5. Z-axis 15 1 .4.6. Delayed time-base triggering 16 1.4.7. Delayed time-base circuit 16 1.4.8. Alternate time-base logic 16 1.4.9. Power supply 16 2. DIRECTIONS FOR USE 17 2.1. INSTALLATION 17 2.1.1. Safety regulations 17 2.1.2. Local mains connection and fuse protection 17 2.1.3. Connection to an external supply 18 2.1.4. Front cover and instrument positions 18 2.2. 2.2.1. CONTROLS AND SOCKETS Cathode-ray tube and POWER 2.2.2. Vertical channels 19 2.2.3. Horizontal channel 19 1 1 .2.4. .2.8. Deflection circuit 15 18 controls 18 2.2.4. Main time-base generator 20 2.2.5. Delayed time base generator 21 2.2.6. Miscellaneous 21 2,3. OPERATING INSTRUCTIONS 22 2.3.1. Switching on the Instrument 22 2.3.2. Preliminary settings of the controls 22 2.3.3. Input coupling (AC/DC, 0) 22 4 2.3.4. Use of probes 22 2.3.5. 23 2.3.6. Adjustment of attenuator probes Selection of chopped or alternate modes 2.3.7. Differential 2.3.8. Selection of trigger 2.3.9. Triggering source 2.3.10. Time-base magnifier 26 2.3.11. Hold-off 26 2.3.12. XY-measurements 26 2.3.13. Using the delayed time base 26 2.4. PROCEDURES REQUIRED FOR THE REMOVAL OF FUSES, BEZEL AND CONTRAST PLATE 27 24 mode 24 25 2.4.4. Removing the instrument covers Removing the mains transformer Replacing the thermal fuse F101 Replacing the fuse F201 2.4.5. Removing the 3. SERVICE DATA 2.4.1. 2.4.2. 2.4.3. 23 mode 27 28 29 29 bezel and the contrast plate 29 75 FIGURES 1.1. 1 .2. 1.3. 2.1 . 35MHz dual-channel oscilloscope PM3218 9 Dimensions 1 Block diagram 14 Rear view of the oscilloscope showing the voltage adapter set to 220 V 1 7 2.2. Front view of the oscilloscope showing controls and sockets 18 2.3. Adjusting an attenuator probe 23 2.4. Suppression of 2.5. Block diagram of composite trigger circuit 2 2.6. Removing the instrument covers and the carrying handle 27 2.7. Replacing the thermal fuse 2.8. Replacing the thermal fuse 28 28 2.9. Removing the 29 common mode signals bezel and the contrast plate 24 5 5 Inhait 1. ALLGEMEINES 31 1.1. EINLEITUNG 31 1.2. TECHNISCHE DATEN 1.2.1. Elektronenstrahlrohre 1.2.2. Vertikale Oder Y-Achse 1.2.3. Horizontale Oder X-Achse 32 32 32 33 33 33 34 34 34 34 35 35 35 35 1.2.4. Hauptzeitablenkung 1.2.5. Verzogerte Zeitablenkung 1.2.6. X-Ablenkung 1.2.7. Triggerung der Hauptzeitablenkung 1.2.8. Triggerung der verzogertenzeitablenkung Kalibriergenerator 1.2.9. 1 .2.1 0. Stromversorgung 1.2.11. Einflussgrossen 1.2.12. Mechanische Oaten 1.2.13. Z-achsen steuerung 1.3. ZUBEHOR 1.3.1. Standardzubehor 1.3.2. Wahizubehor 1.4. BLOCKSCHALTBILDBESCHREIBUNG 1.4.1. Y-Kanal 1.4.2. Triggerung der Hauptzeitablenkung 1.4.3. Haupzeitablenkschaltung 1.4.4. Sperrschaltung 1.4.5. Z-Achse 1 37 ^"7 Triggerung der verzogerten Zeitablenkung .4.6. 1.4.8. Alternierende Zeitablenk-Logik 1.4.9. Speisung 38 38 38 2. GEBRAUCHSANLEITUNG 39 2.1. INBETRIEBNAHME 2.1.1. Sicherheitsvorschriften 2.1.2. Ortlicher Netzanschluss 39 39 39 2.1.3. Anschluss an eine externe Stromversorgung 2.1.4. Abdeckhaube und 2.2. BEDIENUNGSELEMENTE UND BUCHSEN Schaltung der verzogerten Zeitablenkung 1.4.7. und Sicherung Betriebslage Elektronenstrahlrohre und POWER-Einstellelemente 40 40 2.2.2. Vertikale Kanale 41 2.2.3. Horizontaler Kanal 41 2.2.4. Hauptzeitablenkgenerator 2.2.5. Verzogerter zeitablenkgenerator 2.2.6. Verschiedenes 2.3. BEDIENUNGSANLEITUNG 2.2.1 . 2.3.1. Einschalten des Gerats 2.3.2. Vorbereitende Einstellungen 2.3.3. Eingangskopplung 42 42 43 44 44 44 44 6 2.3.4. Anwendung von Messkopfen 44 2.3.5. Einstellen der Abschwacher-Messkopfe 2.3.6. Einstellen der geschoppten oder der alternierenden Darstellungsart 2.3.7. Differentielie betriebsart 2.3.8. Einstellen der Triggerart 45 45 46 46 2.3.9. Triggerquelle 47 2.3.10. Dehnung der Zeitablenkung 48 48 48 48 2.3.11. Einstellen der Sperrzeit 2.3.12. XY 2.3.13. messungen Gebrauch der verzogerten zeitablenkung 2.3.11. XY 48 48 2.3.12. messungen Gebrauch der verzogerten zeitablenkung 2.4. VERFAHREN ZUM ENTFERNEN VON SICHERUNGEN, BILDROHRENRAHMEN UNO KONTRAST- 2.4.1. Abnehmen 2.4.2. Ausbau des Netztransformators 49 49 50 2.4.3. Ersetzen der Schmelzsicherung FI 01 51 2.4.4. Ersetzen der Sicherung F201 51 2.4.5. Abnehmen 3. SERVICE OATEN PLATTE der Abdeckhauben des Bildrohrenrahmens und der Kontrastplatte 51 75 (nur aufs Englisch) ABBILDUNGEN 1.1. 1.2. 2.1 . 2.2. 35 MHz Zweikanal-Oszillograf PM3218 Abmessungen Riickansicht des Oszillografen mit Spannungsumschalter in 220 V Stellung Vorderansicht des Oszillografen mit Bedienungsorganen und Buchsen Abschwacher-Messkopfs 2.3. Einstellen des 2.4. Gleichtaktunterdruckung 2.5. Blockschaltbild der zusammengesetzte (composite) Triggerschaltung 2.6. Abnehmen 2.7. Ersetzen der Schmelzsicherung der Geratabdeckhauben und des Tragbiigels 2.8. Ersetzen der Schmelzsicherung 2.9. Abnehmen des Bildrohrenrahmens 49 50 50 und der Kontrastplatte 51 7 Table des matieres 1. GENERALITES 53 1.1. INTRODUCTION 53 1.2. CARACTERISTIQUES 54 1.2.1. 54 1.2.3. Tube cathodique Axe vertical ou Y Axe X ou horizontal 1.2.4. Base de temps principale 55 1.2.2. 54 55 1.2.5. Base de temps retardee 55 1.2.6. Deviation X 56 1.2.7. 1.2.8. Declenchement de Declenchement de la base de temps principale 56 la base de temps retardee 56 1.2.9. Generateur d'etalonnage 1.2.10. Alimentation 57 1.2.11. Conditions ambiantes 57 1.2.12. Caracteristiques mecaniques 57 1.2.13. Entree modulation-Z 57 1.3. ACCESSOIRES 58 1.3.1. Fournis avec Tinstrument 58 1.3.2. En option 58 DESCRIPTION DU SCHEMA SYNOPTIQUE Y 59 1.4. 1.4.1. 57 Voie 1.4.2. Declenchement de 1.4.3. Circuit de base de 1.4.4. Circuit de blocage 1.4.5. Axe Z 1.4.6. Declenchement de la base de temps principale temps principale 59 59 59 59 59 la base de temps retardee temps retardee 60 60 1.4.7. Circuit de base de 1.4.8. Logique de base de temps alternee 60 1.4.9. Alimentation 60 2. MODE D'EMPLOl 61 2.1. INSTALLATION 61 2.1.1. Reglements de securite 2.1.2. Branchement sur le 2.1.3. Branchement sur une alimentation exterieure 2.1.4. Couvercle avant et position de ^instrument 62 62 61 secteur et fusibles 61 62 2.2. COMMANDES ET PRISES 2.2.1. Tube cathodique 2.2.2. Deviation verticale 63 2.2.3. Deviation horizontale 63 2.2.4. Generateur de base de temps principale 64 2.2.5. Generateur de base de temps retardee 2.2.6. Divers 65 65 et commandes de puissance 62 2.3. INSTRUCTIONS D'UTl LISATION 66 2.3.1. MIse de Tinstrument en circuit 66 2.3.2. Reglage preliminaire des 2.3.3. Couplage d'entree commandes 66 66 8 66 2.3.4. Emplol des sondes 2.3.5. Reglage des sondes attenuatrices 2.3.6. Choix entre modes commute 2.3.7. Mode et alterne 67 67 68 differentiel 2.3.8. Choix du mode de declenchement 2.3.9. Source de declenchement 2.3.10. Agrandisseur de base de temps 70 2.3.11. Reglage de blocage 70 2.3.12. Messures 2.3.13. Utilisation de XY 68 69 70 la base de temps retardee 70 2.4. PROCESSUS REQUIS A LA DEPOSE DE FUSIBLES, VISIERE ET PLAQUE DE CONTRASTE 71 2.4.1. Depose des couvercles 71 2.4.2. Depose du transformateur secteur 72 2.4.3. 73 2.4.4. Remplacement d'un fusible thermique F101 Remplacement du fusible F201 2.4.5. Depose de 3. NOTICE DE SERVICE la vIsiSre et de la plaque de contraste (seulement en Anglais) 73 73 75 FIGURES MHz double trace PM321 8 1.1. Oscilloscope 35 1.2. Dimensions 58 2.1. Vue arriere de I'oscilloscope avec carrousel en position 220 V Vue avant de Toscilloscope montrant les commandes et douilles 61 2.2. 2.3. Ajustage d'une sonde attenuatrice 67 2.4. Rejection en 2.5. Schema synoptique du 2.6. Depose des couvercles 2.7. Remplacement d'un fusible thermique Remplacement d'un fusible thermique Depose de la visiere et de la plaque de contraste 2.8. 2.9. a mode commun de 62 68 circuit et 53 la de declenchement mixte 69 poignee de I'appareil 71 72 72 73 General information INTRODUCTION The 35 MHz dual-channel facilitate its extensive oscilloscope PM 3218 is a compact, portable instrument, ergonomically designed to measuring capabilities. The instrument provides both comprehensive triggering a delayed timebase with provision for alternate timebase displays, including peak-to-peak Auto, DC coupling and automatic TV waveform main and facilities a display. A large 8x10 cm screen with illuminated internal graticule lines makes for easier viewing, and a 10 kV accelerating potential gives a high intensity trace with a well-defined spot. power supply allows the frame ground to be directly connected to floating ground circuits ground does not carry live potentials. By this means, interference by ground currents, as is A double-insulated provided that this frequently experienced with grounded oscilloscopes, is also substantially reduced. The wide range of applications enabled by the above features is further extended by a versatile power supply that enables the instrument to be operated from different line voltages as well as from d.c. For field operation an optional battery version is also available. Warning: The frame ground (and the ground lead of the probe) must not be connected to Fig. 1.1. 35 MHz duaFchannel oscilloscope PM32 18 live potentials. CHARACTERISTICS This instrument has been designed and tested according to has been supplied in a which shall EC I Publication 348 for Class ... instruments and be followed by the purchaser to ensure safe operation and to retain the instrument Properties expressed in numerical values with stated tolerances are guaranteed for +5 ^C II safe condition. The present Instruction Manual contains information and warnings +40 ®C In a safe condition. ambient temperatures of unless stated otherwise. Numerical values without tolerances are typical and represent the characteristics of an average instrument. The data apply after a warming-up period of 30 minutes. Additional Information Specification Designation C.R.T. D14-125 GH/08 Type Rectangular tube face, mesh type, post accelerator, metal backed phosphor. Measuring area 8x10 Screen type P31 (GH) Total acceleration 10 kV Graticule Internal Engravings Centimetre divisions with divisions 1 div. equals cm 1 P7 (GM) optional Cont. variable illumination mm subdivisions of 2 along the central axes. Dotted 10% and 90% lines indicate of measuring lattice for measurement of rise time. Vertical or Y-axis Display modes Channel A only Channel B only A A A Channel B polarity and B chopped and B alternating and B added Normal or inverted Response: Frequency range Rise time Pulse aberrations DC AC < < : 0 .... : 2 Hz 35 MHz (-3dB) .... 35 MHz (-3dB) 10ns ± 3% (< 4% Measured pp) applied Deflection coefficients 2mV/DIV Continuous control range 1 Deflection accuracy ± 3 Input impedance 1 Input RC Maximum time > : div. amplitude and > 1 ns. 1-2-5 sequence % pF Coupling switch to s AC permissible input voltage 400 V, Chopping frequency % d.c. + ax. peak 500 kHz Vertical positioning range 1 Dynamic range 24 divisions Visible signal delay > 2 divisions >40 dB at MHz C.M.R.R. 6 time of 2,5 MW20 0,1 10V/DIV .... at rise in A-B mode 6 divisions At 10ns After adjustment at dx. or low 1 frequencies Cross talk between channels -40 dB or better at 10 Instability of the spot position: Temperature drift < 0,3 div/hour MHz Both attenuators in the same setting 11 1.2.3 Horizontal or X-axis Horizontal deflection can be obtained from either the Main time base or the Delayed time base or a combination of the two, or from the signal source selected for X-deflection. In this case be displayed using A, B, the Ext input connector, or Line Display modes — Main — Main as a signal X-Y diagrams can source for horizontal deflection. time base time base intensified by delayed time base — Main time base and delayed time base alternately displayed — Delayed time base — XY or XY/Y operation X deflection by: — — — Channel A signal Channel B signal Signal applied to EXT connector of main time base — 1.2.4 Line frequency Main time base Automatic Operation Possibility of in automatic free-running the absence of triggering signals Triggered Time coefficients 0,5s/DIV 1 Coefficient error ± Magnification lOx Max. ak/DIV 1 -2-5 sequence > 2,5 Continuous control range : 0,1 3% ± 5% including effective time 10 ns/DIV coefficient Delayed time base Operation Delayed time base either starts Immediately after delay time or is triggerable after the delay time, by the selected delayed time base trigger source Time 1 ms/DIV-0,1 Continuous control range 1 : Coefficient error ± coefficients Delay time jLts/DlV >2,5 3% In steps variable with main time base. Continuously variable with 10-turn potentiometer between 0 x and 10 x the time coefficient of the main time base Incremental delay time accuracy Delay time 0,5% jitter 1 : > 20.000 1 -2-5 sequence xIO magnifier 12 Designation 1.2.6 X Specification Additional information Deflection Source EXT^ EXT, A, B, As selected by 10 or LINE If Deflection coefficients A or As B: push-button trigger source switch. X DEFL. is depressed by selected AMPL/DIV EXTERNAL 0,2 DIV EXT-^ 10 2V/DIV LINE 8 divisions at nominal : : line voltage. ± 10 % Deflection accuracy Frequency range DC: 0 .... MHz 1 (-3 dB) over 6 divisions 1.2.7 Phase shift <3° Dynamic range 24 divisions Triggering of the 100 kHz at < 100 kHz main time base Source Ch. A, Ch. B, Composite, External Trigger For frequencies mode 10 and line Automatic, normal AC normal DC, TV-line and Trigger sensitivity TV frame (DC (DC External 150 mV (DC 200 mV (DC Ext. ^10 :1,5V (DC 2V (DC Internal: 0,5 div div 1 : Triggering frequency range MHz) 50 MHz) 5MHz) 50MHz) 5MHz) 50MHz) > 50 MHz >50 MHz >50 MHz AUTO: 20 Hz AC: 5 Hz DC: 0 Hz Level range 5 AUTO: Proportional to peak-to-peak value of trigger signal. AC, DC: 16 3,2 trigg., div. at Internal at external and 32V at ext. - 10 Triggering slope Positive or negative going Input impedance 1 Maximum 1.2.8 V MW/20 + or -i-or + or —8 div and —1,6V referenced to centre of screen —16V referenced to centre of screen pF permissible input voltage 400 V, Hold-off time variable d.c. + a.c. peak Triggering of the delayed time base Source chA, chB, Composite, External, MTB. Other trigger specifications are identical to "triggering of the main time base" with the exception of the modes EXT. ^ 10, TV and AUTO. trigger 1.2.9 Calibration generator Output voltage 1 Accuracy ± Frequency ,2 Vpp 1% 2 kHz Square wave 13 Designation Specification Additionai Information Double insulated Safety class Power supply 1.2.10 AC supply: Nominal voltage range (on 110, 127, 220 or 240 Vac line-mains voltage adaptor) ± 10 % Nominal frequency range 50 Power consumption 30 DC II, 400 Hz ± 10% W max. At nominal mains voltage 20-28 V dc V A max. 22-27 Current consumption 1,1 Capacity to earth 185 pF Floating input with relaxed specifications Measured with rubber feet on grounded metal plate of Measured 30 27 pF of . 2 11 . . EC 348 supply: Voltage range 1 I 1 1 cm m^ above grounded plate m2 Environmental characteristics The environmental data are valid only if the instrument is checked in accordance with the offical checking procedure. Details on these procedures and failure criteria are supplied on request by the PHILIPS GLOEILAMPENFABRIEKEN, TEST AND MEASURING DEPARTMENT, EINDHOVEN, THE NETHERLANDS. organisation in your country, or by N.V. PHILIPS' Ambient temperatures +40°C Rated range of use + 5°C Operating -10°C ...+55°C Storage and transport -40°C ... ... +70°C Altitude; Non-operating to m 15000 m Humidity 21 days cyclic Shock 30 Operating to 5000 (15000 ft) (45000 ft) heat direction for a Vibration 25°C — 40°C, R.H. 95% sinewave shock of half g: damp Vibrations in 1 1 ms duration: 3 shocks per total of 18 shocks maximum of 15 min. 55 Hz and amplitude of 0.7mmpp and 49 max. three directions with a per direction, 5 — acceleration. Unit mounted on vibrafion table without shock absorbing material. Electromagnetic interference Meets Safety The VDE 0871 and isolation VDE 0875 Grenzwertkiasse between the oscilloscopes and line fulfills requirements of lEC 348 for metal encased class 1.2.12 B. II the safety instruments. Mechanical data Dimensions: 445 335 Height 137 Weight 8,4 kg (18,5 Length 1.2.13. mm mm mm Width Z-mod Input TTL compatible "0" blanks display Handle and controls excluded Handle excluded Feet excluded lb) approx. "1" normal intensity see chapter 3.6. AMPL/DIV. VAR Fig. 1.3. Block diagram POS, 14 Fig. Dimensions 1.2. 1.3. ACCESSORIES 1.3.1. Supplied with the instrument Front cover 2 BNC 4 mm adaptor 2 Probes Instruction manual 1.3.2. Optional PM PM PM PM PM PM PM PM PM PM PM PM PM PM PM PM 8925 8925L Passive probe set 10 8921 Passive probe set 1 : 1 (1 .5 8921 L Passive probe set 1 : 1 (2.5 8940 8943 9355 8910 9380 Isolation amplifier Oscilloscope camera 8971 Camera adaptor 8962 9366 8980 Collapsible viewing Passive probe set 10 1:1,10:1 1 (1 .5 : 1 (2.5 : 1 650 MHz FET-probe Current probe Polaroid anti-glare filter 19" Rack mount adaptor hood Closed long type viewing hood 8901 Rechargeable battery pack 330 8991 Trolley 8992/01 Trimming V d,c./24 Accessory pouch too! kit See also Chapter 3.5. MAT 323 and 100 m) 40 MHz m) 40 MHz m) 1 5 MHz m) 15 MHz : "INFORMATION CONCERNING ACCESSORIES". V d.c. 15 1.4. BLOCK DIAGRAM DESCRIPTION 1.4.1. Y (Fig. 1.3.) Channel The vertical channels A signals to be displayed are identical, each and B for the comprising an input coupling switch, an input step attenuator, an impedance converter and a preamplifier with trigger pick-off. A channel multivibrator, controlled by the display mode pushbuttons, switches either channel operated by A or channel B the end of the Y amplifier via the delay line. The channel multivibrator is sweep, and offers an uninterrupted display of the A and B waveforms in the ALT mode. In the ADD position, both switching amplifiers couple the signals through, thus adding channels A and B. By inverting the B channel to the final amplifier (PULL TO INVERT B) the The AMPL/DIV switches provide x with the step attenuator a 1.4.2. full A- 1 B mode is a pulse at obtained. or x 10 gain control of the preamplifier, which offers in conjunction range of deflection coefficients In a 1-2-5 sequence. Main time base triggering To sweeps, trigger signals can be derived from the initiate A and B vertical channel preamplifiers, from an external source, or internally from the mains supply (LINE triggering) as selected by the trigger source switch. Composite triggering (A and B depressed) is derived from the delay -line driver stage. The polarity of the trigger signal, negative or positive-going, on which the display will start is determined by changing the output polarity of the impedance converter. With the AUTO switch depressed, the peak-to-peak of the signal then determines the range of the With AC In the or DC depressed, the range of the TVL and TVF modes circuit, 1 .4.3. the LEVEL thus initiating sweeps with Main time base level LEVEL LEVEL control is detector comes into operation. The peak-to-peak control is fixed. TV inoperative and the or frame pulses depending line level control. on sync separator the setting of the is TVL switched into and TVF switches. circuit For normal internal time base operation the horizontal amplifier is fed by sweeps from the time base circuit. AUTO depressed, in the absence of trigger signals, the output of the sweep generator is fed back via the With hold-off circuit and gate to its input. This causes sweeps to free-run and a resultant trace is displayed on the logic) As soon as the AUTO control circuit detects a trigger (i.e. a change in the output of the sweep-gating the sweep is fed back to the sweep-gating logic. This causes the circuit to revert to the normal triggering mode in screen. With AC which sweeps are or DC trigger signal is initiated only depressed, AUTO present and the by control LEVEL is trigger pulses at the input of the sweep-gating logic. made inoperative. The display can be magnified in the horizontal direction by factor of xlO (also the X DEFL mode). When the X DEFL pushbutton the final amplifier way, the is Sweeps produced provided a increasing the gain of the final amplifier by a of the horizontal selection switch inhibited and the are then only control appropriately set. impedance converter is is depressed, the sweep generator output to connected directly to the signals normally selected for triggering, or an external source, can now final amplifier. In this be used for horizontal deflection. 1.4.4. Hold-off circuit The hold-off stage, as its name trace has completely returned and the time base circuits decreased by turning the 1.4.5. from the input of the time base circuit until the are completely reset. The hold-off time can be implies, ”holds-off" triggers HOLD-OFF control clockwise. ZAxIs The Z amplifier provides for the blanking of the trace during the fly-back and hold-off time. In addition, it blanks the sweep in the CHOP mode during the switching transients. More over the trace can be blanked by a signal applied to the external Z-mod input. The l.f. components of the blanking signal are modulated and demodulated before they are applied to the Wehnelt cylinder together with the a.c. coupled h.f. components. 16 1.4.6. Delayed time base triggering To sweeps, trigger signals can be derived from the initiate A and B vertical channel preamplifiers, or from an external source as selected by the trigger source push button switch. A and WFth both the line driver stage B pushbuttons depressed simultaneously, composite triggering Y of the amplifier channel. AC and DC polarity of the trigger signal, negative or positive-going, coupling MTB is derived from the delay- provided to the Impedance converter. The on which the display SLOPE changing the output polarity of the impedance converter by the With is will start, is determined by switch. selected, the delayed time base starts directly after the delay time. The DELAY TIME control in conjunction with the comparator determines the delay time for the delayed time base generator. 1.4.7. Delayed time base circuit The delayed time base operative unless Is after the delay time, or upon its TIME/DIV switch in is the OFF position. It starts immediately receipt of the first trigger pulse after the delay time. The sawtooth signal derived from the main time base sweep generator is passed to a comparator where compared with an accurately adjustable d.c. voltage, controlled by the DELAY TIME control. The comparator output pulse-shaped and provides the required delay pulse for the sweep-gating logic of the Is delayed time base generator. The delayed sweep is the main time base. It it is A sawtooth voltage by the hold-off reset is then initiated. circuit of the delayed time base (end of the sweep detection) or by can be started again by the output signal of the comparator after the Initiation of the next main time base sweep. 1.4.8. When pushbutton MTB of the horizontal deflection with the delayed sweep is mode controls is selected, the part of the trace coinciding intensified. Alternate time base logic In ALT TB mode alternately traced an electronic switch enables main time base display and delayed time base display to be on the screen. The two displays can be separated by varying the voltage applied driving circuits of the electronic switch. This separation SEPARATION In the to the vertical amplifier, derived from the TRACE is symmetrically variable by means of the Is controlled by a signal derived from the electronic control on the front panel. ALT TB mode the vertical channel multivibrator switch. In the vertical 1.4.9. ALT modes, successively are displayed on the screen, Channel A and main and delayed time base, Channel B and main time base, Channel B and delayed time base. and horizontal time base, Channel A Power supply The mains When (line) supply the instrument is is transformed and rectified before being applied to operated from a battery supply, the battery output a d.c. to a.c. converter. is connected directly to the d.c. to a.c. converter. The output — 1.5 kV of the regulator is coupled to a transformer and rectifier which, after rectification, provides the potential and the circuit supply voltages. required total accelerating voltage of ^ 10 kV. The -1.5 kV is also multiplied to 8.5 kV to supply the 17 2. Directions for use 2.1 INSTALLATION 2.1.1 Safety regulations (in accordance with lEC 348) Before connecting the instrument to the mains to ascertain whether any damage has occurred instrument to the mains check the cabinet, controls and connectors any defects are apparent, do not connect the (line), visually in transit. If etc. (line). The instrument must be disconnected from all voltage sources and any high voltage points discharged before any maintenance or repair work is carried out. If adjustments or maintenance of the operating instrument with covers removed is inevitable, it must be carried out only by a skilled person who is aware of the hazards involved. In normal operation the double-insulated power supply obviates the need of a safety ground. Warning: must be borne in mind that in all measurements the frame ground of the oscilloscope is raised to the same potential as that of the measuring ground probe connection. Neither the It probe's ground lead nor the frame ground shall be connected to 2.1 .2 live potentials. Local mains (line) connection and fuse protection Before connecting the instrument to the mains delivery the instrument is set to 220 V. If (line) ensure that the instrument is it is set to the local to be used with 1 10 V, 1 mains 27 V (line) voltage. or On 240 V supply, the appropriate voltage should be selected by turning the adaptor on the rear panel to indicate the voltage required (see Fig. 2.1 The instrument windings. It is ) protected from overloads by a thermal fuse fitted between the mains can be replaced after having removed the instrument rear panel (line) (see section 2.4.) MAT 315 Fig. 2. 7. Rear view of the oscilloscope showing the voltage adapter set to 220 V. transformer 18 2.1.3 Connection to an external supply An external supply or battery of 22 socket on the rear panel. V to 27 V (DC Power input cord capable of delivering at least set: 1 A can be connected to the 4822 321 20125). The inner conductor must be connected to the negative pole and the outer conductor to the positive pole, as indicated on the rear panel. The instrument is protected against overloads and reversed polarity by an interna! fuse and diode. This fuse can be replaced after having removed the instrument rear panel 2.1.4 (see section 2.4.). Front cover and instrument position The front cover can be simply removed by pulling it from the front. The instrument may be used horizontally or in several sloping positions by using the carrying handle as a tilting bracket. To unlock the handle, simultaneously push 2.2 CONTROLS AND SOCKETS 2.2.1 Cathode-ray tube and ILLUM POWER ON POWER in both pivot centre knobs. (Refer to Fig. 2.2) controls Continuously variable control of the graticule illumination; incorporates mains (line) switch. POWER ON pilot lamp indicates the ON state. INTENS Continuously variable control of the trace FOCUS Allows beam to be focused for minimum spot TRACE ROTATION Screwdriver adjustment to align the trace with the horizontal brilliance. size. graticule lines. MAT Fig. 2.2, Front view of the oscilloscope showing controls and sockets. 313 19 2.2.2 Vertical channels Display Function mode switch A - ALT - CHOP - ADD - B 5-way pushbutton switch selecting the With A all ALT mode buttons released, the Vertical deflection is vertical display is in mode. operation. achieved by the signal connected to the input of channel A. The display end of every cycle of the timebase at the CHOP The switched over from one vertical channel to the other is display signal. switched over from one vertical channel to the other is at a fixed frequency, (f ^ 500 kHz) sum A and B, ADD Vertical deflection is achieved by the B Vertical deflection is achieved by the signal connected to the input channels signal of of channel B. PULL TO INVERT B (outer-knob) pulled, channel B signal Step control of the (centre-knob) is inverted. vertical deflection coefficients, ranging 2 mV/div up to 10 V/div AMPL/DIV in a 1 Continuously variable control of the centre-knob switched to the from -2-5 sequence. Note that the deflection coefficient Input coupling switch of the display. combined with the channel B POSITION control. Push-pull switch When AMPL/DIV vertical shift Continuously variable controls giving POSITION CAL vertical deflection coefficients. is calibrated only with the position (fully-clockwise). 2-way pushbutton switch Signal coupling; AC/DC - 0 AC (depressed) Coupling DC (released) Direct coupling blocking capacitor via a Connection between input 0 (depressed) the input circuit and input socket circuit is Interrupted and grounded. A (1M^2//20pF) BNC socket for channel B (1Mn//20pF) BNC socket for channel B input A 2.2.3 is input Horizontal channel X deflection source switch MTB-XDEFL DTB ^ALT TB— Function DTB The horizontal deflection voltage Horizontal-deflection controls; 3-way pushbutton switch is supplied by the delayed time- base generator. MTB The horizontal generator. timebase A Is deflection voltage is supplied by the main timebase intensified when the delayed running. The delayed timebase generator DELD TIME/DIV If is portion of the trace no buttons switch is is in the switched off OFF are depressed the effect depressed (only the MTB LEVEL is when the position. the same as the control is MTB button not operating in this situation). DTB I- ALT MTB TB— When both the DTB and MTB pushbuttons are selected simultaneously, the horizontal deflection voltage is supplied i>y the main and delayed timebases alternately. X DEFL Horizontal deflection signal, the is achieved by the channel mainsfrequency A signal, the channel B signal or an external signal applied to the external input socket of the main time base. 20 X POS/X MAGN Continuously variable control giving horizontal incorporates a push-pull switch, PULL FOR shift of the display; x 10, which increases The the horizontal deflection coefficient by a factor of 10. magnifier TRACE SEP is also operative deflection signal is used. between the two vertical space ALT.TB mode. the in X an external Continuously variable control of the time-base displays 2.2.4 If Main time base generator LEVEL SLOPE (IN +,OUT-) Continuously variable control to select the which the timebase generator signal at This control incorporates a level of the triggering starts. push-pull switch, which enables choice of triggering on the positive or negative-going edge of the triggering signal. For TV triggering, select + for positive video Trigger mode switch AC - AUTO - DC i- Function With all A trace LEVEL selecting the trigger pushbuttons released a fixed range of the AUTO and for negative video signals signals. 3-way pushbutton switch tvl-jltvf-j — is LEVEL AUTO mode. sweep mode is operation at in control. displayed in the absence of trigger signals. The range of the control is proportional to the peak-to-peak value of the triggering signal. AC Normal DC Normal LEVEL triggering and fixed range of the component of the trigger signal triggering component is LEVEL and fixed range of the of the trigger signal AUTO AC '-tvl-' Line synchronisation AUTO DC Frame synchronisation is is control. The DC control. The DC blocked. passed. obtained. is obtained. 1-TVF-I HOLD OFF Continuously variable control of hold-off time Trigger source switch Function A - B-EXT L-COMP--J - EXT^IO LINE-J 4-way pushbutton switch of horizontal deflection for if selects the trigger source, (or the X deflection source switch is source depressed X DEFL). With all pushbuttons released, the effect is A the same as the button depressed. A A Signal derived from channel B Signal derived from channel B A B Composite i-COMP-l signal, derived after the electronic switch. Triggering occurs on the displayed waveforms (not usable with EXT and EXT^IO External signal derived via the adjacent LINE (MAINS) Signal derived Instrument TIME/DIV (outer switch) is from the 1 Mn//20pF line (mains) voltage. X DEFL). socket. (Inoperable when battery-powered). Selects the time coefficient from .1 /Lis/div to .5 s/div in a 1-2-5 sequence. TIME/DIV (inner knob) Continuously variable control of the time coefficients. Must be switched to CAL position (i.e. fully clockwise) for the be calibrated according to the indication of the 1 MO - 20 pF BNC time axis to TIME/DIV socket for external triggering or horizontal deflection. switch. 21 Delayed time base generator LEVEL SLOPE (IN +, OUT-) Continuously variable control to select over a fixed range the level of the triggering signal at which the time base generator starts. This control incorporates a push-pull switch, which enables choice of triggering on the positive or negative going edge of the triggering signal. Trigger mode switch Function mode AC - DC 2-way pushbutton switch selecting the AC Normal triggering and fixed range of the LEVEL control. The DC component of the trigger signal is blocked. DC Normal triggering and fixed range component of the trigger signal is of trigger LEVEL control. The DC passed. DELAY TIME Continuously variable vernier control of the delay time, together with the TIME/DIV controls of the main time base generator. Trigger source switch Function A - B - EXT - MTB 4-way pushbutton switch selects the trigger source point of delayed time base. LcOMpJ effect as the A MTB and No pushbutton depressed starting has the same button depressed. Internal triggering Signal derived from channel A. Internal triggering B Signal derived A Composite B from channel B. signal, derived after the electronic channel switch. Triggering occurs on the displayed waveform, after selected delay *-COMP>J time. Triggering on an external signal connected to the adjacent EXT 1Mr2//20pF socket. MTB Internal triggering signal derived from the main time base to start the delayed time base immediately after the selected delay time. TIME/DIV (outer switch) Selects the time coefficient sequence. Incorporates an time base generator TIME/DIV (inner knob) is from OFF switched .1 /is/div to ms/div in a 1-2-5 off. Continuously variable control of the time coefficients. Must be the CAL position (i.e. fully clockwise) for the time axis to be calibrated according to the indication of the BNC 1Mn-20 pF 1 position by which the delayed TIME/DIV in switch. socket for external triggering signal. Miscellaneous Output socket supplying squarewave voltage ~ 2 kHz at an amplitude of 1.2 V p-p ± 1% To be used for probe compensation and/or CAL . checking DC POWER vertical deflection accuracy. input socket at the rear of the instrument allows operation by an V to 27 V, current IN external d.c. supply. Rated supply voltage 22 capability LINE (MAINS) VOLTAGE Must be > lA. set according to section 2.1.2 before the ADAPTOR connected to the Z-MOD See chapter 3.6. local mains voltage. instrument is 22 2.3 OPERATING INSTRUCTIONS 2.3.1 Switching on the instrument Before connecting the instrument to any supply, the instructions given in section 2.1 Installation, should be carefully carried out. The oscilloscope will meet specifications (see section 1.2) normally after a warming-up period of approximately 15 minutes. However, if the instrument has been subjected to an extremely cold environment (e.g. left in a car overnight in freezing conditions) and is then brought in for use in a warm room, a warming-up period of sufficient length should be allowed (see 1.2.11). 2.3.2 Preliminary settings of the controls This procedure a general indication of is able starting point before any whether the oscilloscope measurements is functioning correctly and provides a suit- made. are Refer to Fig. 2.2 for location of controls. INTENS Set Depress AUTO switch. With FOCUS and all controls in mid position. and select an average time coefficient between 10 ^s/div and 10 ms/div with the TIME/DIV other pushbuttons normal (not depressed) channel A and channel B traces can be positioned on the screen with the relevant POSITION controls. Set the ness 2.3.3 and adjust FOCUS INTENS control for a display of medium bright- control for well focused traces. Input coupling (AC/DC, 0) AC coupling (pushbutton depressed) is useful to block the d.c. component of a signal. Choice of AC limits the lower frequencies, causing low repetition rate sinewave signals to be attenuated and low repetition rate square- waves to be distorted. The degree of attenuation extended by 10 When if is determined by the input RC time (0.1s). Input RC time is 10:1 passive probes are employed. switching to AC coupling it will take approximately five input RC times before the trace is stabilised to the average value of the input signal. AC position measurements cannot be made with respect to ground. 0 position disconnects input source and short-circuits input of amplifier to provide zero signal check. DC 2.3.4 coupling (pushbutton released) provides for full range frequency input, i.e. down to d.c. Use of probes 1 :1 passive probes should only be used for d.c. and low frequencies. Capacitive loading attenuates high frequencies or increases the rise-time of measurement signals (dependent on source Impedance). 10:1 passive probes have especially limit of the oscilloscope 10:1 passive probes or less capacitive loading; usually about when measurements amplitude errors are to be taken lOpF from high impedance to 20pF. FET probes are superior, test points or at the upper frequency bandwidth. must be properly compensated before use. Incorrect compensation leads to pulse distortion at high frequencies. For correct adjustment, the CAL output connection can be used (see Fig. 2.3.) 23 2.3.5. Adjustment of attenuator probes — - Connect the compensation box to socket A and place the tip of the probe on socket CAL. Insert a small screwdriver through the hole in the compensation box and adjust the trimmer to obtain correct display as shown a in Fig. 2.3. under compensated over compensated correct MA8329 Fig. 2.3. 2.3.6. Adjusting an attenuator probe Selection of (A In chopped or alternate modes ALT CHOP ADD B) dual channel operations (CHOP or ALT depressed) the chopped mode (depress CHOP) must be selected for relatively slow sweep speeds (from .1 ms/div to .5 s/div) or at low repetition rates of sweeps occuring, even at high sweep rates. Selection of the ALTernate mode under these circumstances would make comparisons between waveforms difficult because traces would actually appear to be written one by one. However, when is fast enough to be interrupted by the chopping frequency the alternate mode must be selected ALT), usually at sweep rates faster than .1 ms/div. the display (depress 24 2.3.7. mode Differential The In A — B mode measurements where signal lines carry substantial common mode To obtain the degree of common mode signals (e.g. hum) the differential mode — B). The capability of the oscilloscope to signals is CMR by the given common mode factor (see Fig. 2.4). rejection as specified, channel done by connecting both channels to the This can be tinuous controls with the When and pulling the channel B POSITION control. out these signals and leave the remainder of interest (A will cancel suppress ADD can be selected by depressing AMPL/DIV CAL minimum switch for minimum and B gains must first be equalised. deflection on the screen. passive 10:1 probes are used a similar equalisation process sating controls for A output connector, and adjusting one of the con- recommended by is adjusting their compen- deflection. As drawn here CMR :: 10 M A 9397 Suppress! on of Fig. 2. 4. 2.3.8. Selection of trigger common m ode signafs mode AUTO DC) LtVL-II- TVF-l (AC The AUTO mode Furthermore, for is most useful because of the position of the provides trace(s) on the screen even in the absence of trigger signals, it amplitude larger than a signal LEVEL control; its division, this 1 range is mode provides stable triggering independently automatically adjusted to the peak-to-peak value of the signal selected for triggering. In this The way the setting of the AUTO mode In DC control is facilitated at small amplitudes of the trigger signal. cannot be employed for signals with low repetition would be able to free run used (AC or LEVEL in between triggers. rates (10 Hz or lower) because the sweeps Therefore, for low repetition signals normal triggering must be depressed). norma! triggering, sweeps are only initiated with a trigger signal applied and the LEVEL control set appropriately. With AC or DC depressed the range of the LEVEL control LEVEL control with respect to mid screen). fixed is (-i- -S or divisions or more at the when triggering extremes of the The DC component take place With AC on the trigger signal can be blocked by depressing AC. This in a.c. signals superimposed on useful coupling, the level at which the display starts will change with alterations trigger signal. The trigger level waveforms that vary Slope selection is in is thus no longer referenced to signal ground. This made with pushbutton +/-. No buttons depressed range is time interval from cycle to cycle. Normally + for positive video signals. LEVEL is fixed. The LEVEL offers an extra must a large d.c. level. In control mode TV mode - must Is Inoperable of use, a trace is in it is in the may average value of the cause instability with preferable to use the DC position. be selected for negative video signals and the TV mode. on screen in the absence of a trigger signal, but the 25 2 ^.9 . Trigger sources trigger sources The main time base A > i-COMP-J — — LINE Internal triggering will be TRIG or X DEFL pushbuttons. ext^io ext B can be selected by the front-panel * most commonly used because it requires only one signal (the input signal) to obtain stable triggering. advantageous to use an external signal for triggering. There Is no need to set and reset the trigger controls (LEVEL, SLOPE and SOURCE) on changing the input waveforms. signal. Furthermore the two A and B Inputs remain free for examining — External triggering. — frequency, always Selection of trigger source. In comparing waveforms that are a multiple of each other's lead to double may doing so Not select the signal, that has the lowest repetition rate as the trigger source. — Composite pictures (chopped When mode) triggering. Composite 1. when COMP is signals are mode, it is When composite there is selected by depressing both it is A A channel, the B channel and B pushbuttons, from the delay Line (mains) triggering is compare possible to triggering trigger useful triggering is derived from the differential signal. Triggering not necessary to switch trigger sources from is source when B or vice versa. signals that are not related in time. is A or B input), stable also switched from the signal input Channel B triggering Block diagram of composite A to is employed in dual channel operation (chopped or alternate), and only one signal applied (to normal since the Fig. 2.5. mode). obtained from either the follows the electronic channel switch. For one channel operation 3. In the alternate — it is or untrue time-shifts (alternate mode (A - B measurements) by common mode signals. In differential Note: signals triggering offers three advantages: not disturbed 2. many With Internal triggering preamplifier stages or, line driver stage that tracing trigger circuit is A mains triggering to B (line) cannot be obtained. This (see Fig. 2.5). frequency related. Composite triggering is only 26 2.3.10. Time^base magnifier The magnifier When operated by pulling the switch incorporated is in the X position control. sweep speed is increased 10 times. The sweep time therefore determined by dividing the indicated TIME/D IV value by 10. 2.3.11. this switch is the in xIO position, the time-base is Hold-off. This control can be used to increase the sweep hold-off time. 2.3.12. XY Measurements XY measurements are made with the TIME/DIV switch being selected by the XY TRIG measurements provide or X DEFL pushbutton at EXT X DEFL, the source of horizontal EXT, EXT H- 10 or LINE) means of making frequency or phase a useful deflection switch (A, B, shift comparisons by displaying Lissa- jous patterns. Measurements can be made up to 100 kHz with The sensitivity for the different X 2.3.13. deflection XY modes is shown than 3° phase error between horizontal and vertical channels. in the following table: Sensitivity A AMPL/DIV A ± 10% B AMPL/DIV B EXT 0.2 EXT less 10 2 LINE ± 10% V/DIV V/DIV 8 divisions Using the Delayed time-base The delayed time base can be used for the accurate study of complex signals. The delayed time base generator starts (TIME/DIV switch not at OFF) after the selected delay time and the delayed signal is intensified when the The MTB pushbutton of the horizontal deflection controls has been selected. DELAY TIME potentiometer control enables the intensified portion to be duration of the intensified portion, TIME/DIV controls In the is its controls of the delayed time base generator. When pushbutton DTB depressed, the intensified portion occupies the DTB DELAY TIME control. is full The by means of the of the horizontal deflection width of the screen. the interval between the starting points of the main time base and determined by the setting of the main time base TIME/DIV controls and the position, the delay time that of the delayed time base) shifted along the time axis. length, can be controlled in steps and continuously The PM3218 (i.e. is equipped with display switching. This offers the instrument user a simultaneous display of the signal on the two time scales provided by the main time base and by the delayed time base. By selecting ALT TB, detailed expanding the time interval of Expansion is examination of portion of the main time base display The part of the signal DELAY TIME also serves as a visual indication of the between MTB as facilitates the location of the required detail TIME/DIV and DTB. control an intensified portion during dialling, but portion of the overall trace being examined. Selection of enables immediate correlation of the detail with the overall signal, which the need to switch enabled by potentiometer. under detailed observation by the delayed time base also remains main time base display. This not only is delayed time base. achieved by selecting a correspondingly faster sweep for the delayed time base and positioning the time interval by the of the a certain interest, using the may ALT TB thus be extremely complex, without 27 2.4. PROCEDURES REQUIRED FOR THE REMOVAL OF FUSES, BEZEL AND CONTRAST PLATE 2.4.1. Removing the instrument covers Always ensure that the mains supply is disconnected before removing any instrument cover plates. The instrument is protected by three covers: a front panel protection cover, a wrap-around cover with carrying handle, and a rear panel. To facilitate removal of the wrap-around cover and the rear panel, first ensure that the front cover is in position. Then proceed — — — — — - as follows: Hinge the carrying handle clear of the front cover; to this end, push both pivot centre buttons Stand the instrument on its protective front cover on a flat surface. Slacken the two coin-slot screws located on the rear panel. Lift the rear panel Lift off the and unplug the connector on the power supply board. wrap-around cover. For access to the front-panel, stand the instrument horizontally and snap off the front cover. Fig, 2.6. Removing the instrument covers (Fig. 2.6.). 28 2.4.2. Removing the mains transformer — Remove wrap-around cover and rear panel (section 2.4.1,). — Take the lid off the voltage adapter compartment after removing the 4 cross-slotted screws. — Remove the 4 cross-slotted screws that hold the lid of the transformer compartment. — Lift the lid with the attached transformer, simultaneously sliding the wire form between transformer voltage adapter out of the — The slit in the transformer compartment. transformer and thermal fuse are then accessible for replacement THERMAL FUSE THERMOSiCHERUNG FUSIBLE THERMIQUE Fig. 2. 7. Fig. 2.8. Replacing the thermal fuse Replacing the thermal fuse and 29 2.4.3. Replacing the thermal fuse F101 - Remove the mains transformer (section 2.4.2.). — Unsolder fuse terminals and 2 (Fig. 2.7. and 2.8.). - Only the fuse wire of the old fuse is replaced and not 1 — the fuse slightly outwards, disengage the locking pin and pull out the wire. Take the new fuse and remove the fuse wire out of its housing in the same Push the new fuse wire into the housing of the old one in — 2.4.4. 2.4.5. the complete fuse; to this end, bend the housing of the fuse wire must point to terminal 1. and 2. Solder the fuse wire to terminals 1 way until the locking pin as described above. snaps into the hole. The loop Replacing the fuse F201 — Remove wrap-around cover and rear panel (section 2.4.1.). — Fuse F201, which is located on the power supply printed circuit Removing the board, is accessible for replacement. bezel and the contrast plate - Take hold of the bezel's bottom corners and gently pull it from the front - The contrast filter can be removed by pressing it gently out of the bezel. Fig. 2.9. now Removing the bezel and the contrast plate panel (Fig. 2.9.). 113 3.3. INFORMATION FOR ASSISTENCE 3.3.1. Mains transformer data IN FAULT FINDING BALANCE CHANNEL A D.C. The available unloaded voltage tappings diagram 3.3.2. (Fig. 3.44) in the Voltages and waveforms in form and the number of turns per winding are listed in gain GAIN X10 R504 of a table. C502 CHANNEL A the instrument BALANCE CHANNEL A GAIN The d.c. voltage levels at generators are The at the relevant points on the circuit oscilloscope under test must be set in the following in Fig. — — — — — “ — — — — — — — — — — shown the electrodes of the transistors and the voltage waveforms 3.44 and Display X mode diagrams (Fig. 3.44, way Fig. in the time-base 3.45 and to measure the voltage Fig. 3.46). wave forms as shown Fig. 3.46. switch SI to position S3 to position "MTB''. deflection selector switch MTB trigger mode switch S4 to position ''AUTO'". A POSITION potentiometer R2 at mid-range. A AMPL/DIV switch S9 to V/div. and potentiometer R9 to CAL. Input signal on A Input socket X2: 2,5 kHz sine-wave voltage for 8 div. Z3B =*1 D.C. BALA deflection. =IM GAIN BALANCE CHANNEL for measuring the diagrams 11-12-1 5-1 6-17-18c-18e and 19c. SHIFT BAL CHANNE DTB TRIGGER Remark In case When of a defect it the instrument is is always possible to apply to the world wide PHI LIPS Service Organization. to be sent to a PHILIPS Service Workshop for repair, the following points should be observed: Attach a label with your name and address to the Instrument. Give a complete description of the faults found, or the service required. Use the original packing, or, if this is no longer available, carefully pack the instrument In a wooden crate MTB TRIGGER or box. — Send the instrument to the address obtained after consultation with the local PHILIPS Organization. B R674 SENSITIVITY — — — R6 CHANT 19a en 19b. 3.3.3. BALANCE CHANNEL B MTB LEVEL potentiometer R7 at mid-range. DTB LEVEL potentiometer R5 at mid-range. MTB SLOPE switch S8 in position DTB SLOPE switch S6 in position MTB TRIGGER source selector switch S22 to position "A". DTB TRIGGER source selector switch S21 to position "A". MTB TIME/DI V switch SI 5 to 0,2 ms/div. and potentiometer R 12 to CAL. DTB Tl ME/DI V switch SI 3 to OFF for measuring the diagrams 1-2-3-4-5-6-7-8-9-10-1 3-1 4-1 8a-18b-18d/Lis/div. E NORMAL/ X POSITION potentiometer R6at mid-range. X MAGN switch S7 to position "XI”. switch SI 3 to 50 R SHIFT CHAr 1 — DTB TIME/DI V XI CHANNEI the circuit SENSITIVITY 113 CAL. BALANCE CHANNEL A D.C. L802 R519 R1607 ^ GAIN X10 R504 C502 R 848 CHANNEL A GAIN XI VERTICAL FINAL AMPLIFIER R514> R 547 BALANCE CHANNEL A GAIN m m. SHIFT BALANCE CHANNEL A D.C. L801 \GAIN X1 A y CHANNEL *^^C 818 BALANCE CHANNEL B R 647 R 619 NORMAL/ INVERT R604 C602 GAIN X10 BALANCE CHANNEL CHANNEL B B C1407 C 1406 R614 GAIN BALANCE CHANNEL B mi "^21 R674 BALANCE CHANNEL B SHIFT COMENSATION R1107 DTB SWEEP TIME X GAIN XI DTB TRIGGER SENSITIVITY mi R 1417 R1336 R1537 R >e X mi' te tm BRILLIANCE R 1216 ^ MTB GAIN X10 MTB SWEEP R1507 BRILLIANCE RATIO TIME 1016 TRIGGER SENSITIVITY — 1419 MULTIPLIER SETTINGS mi[ mi_ im' MAT 353 Fig. 3. 19. Adjusting elements amplifier board 114 34 . 34 . . . 1 . 114 f 3.4. DISMANTLINGTHE INSTRUMENT r 3.4.1. General information This section provides the dismantling procedures required for the removal of components during repair and f routine maintenance operations. All circuit boards removed from the oscilloscope should be adequately protected against damage, and all During dismantling procedures, normal precautions regarding the use of tools must be observed. a careful note of all leads disconnected must be made so that they may be reconnected to their correct terminals during assembly. f I Always ensure that the mains supply Is disconnected before removing any instrument cover Damage may is switched on plates. ^ board is result if the instrument removed within one minute of switching when a circuit board has been removed, or if a circuit off the instrument. f 3.4.2. Removing the instrument covers The Instrument is protected by three covers: a front panel protection cover, a wrap-around cover with carrying handle, and a rear panel. To facilitate removal of the wrap-around cover and the rear panel, Then proceed — — — — — — first ensure that the front cover is in position. as follows: hinge the carrying handle clear of the front cover; to this end, push both pivot centre buttons (Fig. 3.20). stand the Instrument on its protective front cover on a flat surface slacken the two coin-slot screws located on tfie rear panel lift the rear panel and unplug the connector on the power supply board lift off the wrap-around cover for access to the front-panel, stand the instrument horizontally and snap off the front cover. Fig. 3.20. Removing the instrument covers. COMPONENT LOCATION Item Grid loc. C101 rear panel C200 C201 C202 C203 C204 C206 C207 C208 C209 C211 ^ power supply power supply power supply power supply power supply power supply power power power power C212 C213 C214 C216 supply C221 C222 C223 C224 C226 C227 C228 C229 C231 C251 C252 C253 C254 C255 C256 C257 C258 C259 Item Grid loc. Item Grid loc. Item C285 C286 D-3 Cl 007 C-5 Cl 509 G-4 E-2 Cl 008 C-5/D-5 att. E-2 Cl 009 C-5 C305 C307 C308 C309 C310 att, unit C-2 C1011 C-5 att. unit C521 E-2 C-5 att. unit F-2 att. unit C522 C523 C-2 att. unit C601 C-3 C1012 C1101 C1102 C1103 B4 C4 C-3 Cl 104 C-4 C-3 E-2 C1201 Cl 202 Cl 203 C1204 D-5 att. unit C602 C603 C604 C607 C609 D-5 att. unit C611 D-3 Cl 205 D-5 att. unit E-3 Cl 206 E-5 C1511 C1512 C1601 Cl 602 C1651 C1652 Cl 653 C1654 C1655 C1656 C1657 C1658 Cl 659 Cl 660 G-5 C301 C517 C518 C519 C520 E-2 G-3 E-3 D4 R1 E-3 C1207 Cl 208 E-5 E-3 Cl 209 E-5 F-3 C1210 E4 C-3 E-5 E4 E-3 C1301 C1302 C1303 Cl 304 C1305 Cl 306 Cl 307 R2 R3 R4 R5 R6 R7 R8 R9 RIO D-5/E-5 R11 R267 C-4 R268 D-4 R269 B-2/C-2 R271 B-3 D-3 R272 R273 att. unit R274 B4 R276 C-3 D-4 R277 F4 R278 R279 G-3 R281 G-3 R302 att. unit RSOSl on switch R304 ^ S9 R306/ R307 att. unit R308 att. unit R309 att. unit R311 att. unit R312 att. unit on switch R313' R314 J S9 R316^ att. unit R317 att. unit R318 att. unit R319 att. unit unit att. unit supply unit supply supply C321 att. unit supply att. unit att. unit C621 E-3 att. unit F-3 att. unit C622 C623 C-3 att. unit C701 E-3 supply C322 C324 C351 C352 C353 C356 C613 C616 C617 C618 C619 C620 att. unit E-2 supply C401 att. unit supply C405 C407 C408 C409 C410 att. unit att. unit att. unit C702 C703 C704 C705 C706 C707 att. unit att. unit supply supply supply supply supply supply C-2 D-2 C-3 D-3 att. unit att. unit att. unit att. unit att. att. unit unit D-3 D-3 D4 F-2 C1310 C1311 C801 F-3 C1321 D-4 C802 C803 C804 C805 C806 C807 C808 C809 C810 F-3 E-4 R201 R202 C811 F-3 att. unit G-3 C422 C424 att. unit G-3 Cl 409 att. unit G-2 C451 att. unit att. unit B-3 C452 C453 att. unit D-3 C501 C-2 C812 C813 C814 C815 C816 C817 C818 C1411 C1412 C1413 C1414 C1416 C1417 C1501 E4 B-3 D4 D4 C4 C4 D4 C-2/C-3 J277 C278 C279 C280 att. C281 D4 D4 unit B4/B-5 B-3 att. F4 G-2 unit C502 C503 C504 C507 C509 unit att. unit att. unit att. unit att. unit att. unit C-2 C-2 D-2 D-2 E-2 C511 D-2 C513 E-2 > E-2 att. unit F-5 unit E-5 R12 R13 R14 R15 R200 C421 D4 att. E-4 C821 C1001 Cl 002 Cl 003 Cl 004 C1005 Cl 006 F-3 C1314 C1315 C1316 C1402 D-5 B-2 B-2 F-4 F-4 F-5 F-5 F-4 F-5 F-5 F-5 F-5 E-3 front panel power power power power power power power power power power R251 B-2/C-2 G-2 R252 R253 R254 R256 R257 R258 R259 B-2/C-2 C-4 Cl 503 G-5 B-5 Cl 504 G-5 C-4/D-4 C1506 G4 C-4 Cl 507 G-5 Cl 508 G-5 ! R261 R262 R263 R264 R266 C-3 C-3 R701 C-3 E-3 F-2/G-2 E-2 R861 F-2 D-2 R611 C-3 E-2 D-2 R612 R613 R614 R616 R617 R618 R619 C-3 E-3 D-3 R702 R703 R704 R706 R707 R708 R709 C-3/D-3 R711 F-3 C-3/D-3 F-2 C-2/D-2 R621 C-3 C-2/D-2 D-3 D-3 D-3 R801 F-2 C-5 D-3 F-3 R1011 C-5 D-2 R631 D-3 D-2 R632 R633 R634 R635 R636 R637 R638 R639 D-3 R802 R803 R804 R806 R807 R809 F-3 D-2 R622 R623 R624 R626 R627 R628 R629 R712 R713 R714 R716 R717 R862 R863 R864 R866 R1001 R1002 R1003 R1004 R1006 R1007 R1008 R1009 D-3 R811 F-3 D-3 R812 R813 R814 R816 R817 R818 R819 F-3 R1012 R1013 R1014 R1016 R1017 R1018 R1019 R501 R502 C-2 C-2 C-2 D-2 G-3 G-5 C-3 C-2 R602 R603 R604 R606 R607 R608 R609 D-2 G-2 Cl 502 att. unit R852 R853 R854 R856 R857 R858 R859 D-2 high t.u. C-4 E-3/F-3 D-2 power supply supply B-3/C-3 B-3/C-3 C-3 D-3 att. R851 R672 R673 R674 R676 R677 R521 R212 R227 G-5 E-3/F-3 C-3 unit R211 G-2 R671 R601 att. G-3 F-2 C-3 att. unit unit R403"] 1 E-3/E-4 R404 R406 R407 R408 R409 B-3 D-4 B-4 B-3 E-3/E-4 F-5 unit unit att. unit att. unit att. unit att. unit att. unit att. unit att. unit R522 R523 R524 R526 R527 R528 R529 att. unit R531 on switch ' r S11 J att. unit att. unit att. unit R411 att. unit R412 att. unit unit att. unit unit G-3 B-5 F-2 att. unit R572 R573 R577 R600 att. att. att. G-3 G-3 att. unit att. R362 R363 R364 R365 R366 R367 R368 R369 R374 R402 unit E-3 G-3 R662 R663 R664 R668 R669 supply att. E-2 G-3 R571 R361 supply D-3 G-3 att. unit supply Cl 407 G-2/G-3 E-3 E-3 att. unit supply C1408 G-2 E-3 F-2 D-2 E-3 R462 R463 R464 R466 R467 R468 R469 R500 att. unit att. unit C-2 G-3 F-2 E-2 E-2 F-3/G-3 G-3 R461 unit R512 R513 R514 R516 R517 R518 R519 Cl 406 F-2 E-3 att. unit F-2 G-2/G-3 R661 unit att. G-3 F-2 E-2 att. C-2 Cl 405 G-3 E-2 R652 R653 R654 R658 R659 R511 F-3 F-3 D-2 E-2 loc. R552 R553 R554 R558 R559 R568 R569 att. unit unit G-3 F-5 F-3 R822 R823 R824 R825 R826 R827 R828 R829 R831 R832 R833 R837 R838 R839 R843 R847 R848 R849 att. unit unit unit unit R821 R651 att. att. unit E-3 E-2 R646 R647 R648 R649 R650 D-3 D-3/E-3 E-2 unit att. att. E-2 Grid R551 att. unit att. R641 Item Grid loc. unit att. unit att. supply R640 D-2 Item att. att. unit unit supply D-2 loc. R452 R453 R454 R456 R457 R458 R459 R416 R417 R418 R419 att. supply Grid att. unit S11 R414Jr R352 R353 R354 R355 R356 R357 R358 R359 R360 supply Item R451 on switch . unit supply loc. R542 R543 R546 R547 R548 R549 R550 R413^ att. Cl 404 D4 Grid R351 F-3 F-3 D-4 Item R503 R504 R506 R507 R508 R509 R203 R204 R206 R207 R208 R209 R210 F-3 G-2 Grid loc. t E-5 C412 C413 C414 C415 C416 C417 C418 C419 C420 att. D4 D-5 unit B-4/B-5 D-4 on tube 'I Cl 309 att. B-3 D-5 C1308 C41 att. unit E-5 E-3 unit 1 B-4 E-2 att. E4 C282 C283 C284 Grid loc. att. unit C262 C263 C266 C267 C268 C269 C272 C273 C274 Item C311 tension C261 C271 Grid loc. C312 C313 C314 C315 C316 C317 C318 C319 C320 high power power power power power power power power power power power Item supply C2I7J C218 C219 1 145 LIST C-2 C-2 D-2 C-2 D-2 D-2 C-2/D-2 D-2 D-2 R532 R533 R534 R535 R536 R537 R538 R539 R540 D-2 R541 E-2 D-2 D-2 D-2 D-2 E-2 E-2 D-2/E-2 E-2 D-2 D-2 E-2 F-2 E-2 C-3 C-3 C-3 C-3 D-3 C-3/D-3 D-3 D-3 D-3 D-3 E-3 D-3/E-3 E-3 D-3/E-3 D-3 D-3 D-3 E-3 E-3 E-3 E-3 E-3 E-3 E-3 E-3 F-3 E-2/E-2 E-2/E-3 E-2/E-3 F-2 E-3 E-2 F-3 F-3 F-2 F-3 F-3 F-2/F-3 F-3 F-2 F-2 G-3 F-3 G-2 F-2 F-2 F-2 F-2 F-2 G-2 F-2 F-2 F-2 F-2 F-2 F-2 F-2/G^2 F-2 G-2 F-2/F-3 B-5 C-4 C-4 C-4 C-4 C-5 C-5 B-4 B-5 C-5 B4 C-5 C-4 C-5 G-3 R1021 C-5 F-3 R1022 R1023 R1024 C-4 F-3 F-3 F-2 F-3 C-4 C-4 146 Item Grid loc. R1026 C-5 R1027 C-5 R1028 C-5 R1029 C4 R1031 C4 R1032 C4 R1034 C-4 R1036 C4 R1037 C-5 R1038 C4 R1039 C4 R1041 C-4 R1042 C-5 R1043 C-5 R1044 C-5 R1046 C-5 R1047 C-5 R1048 C-5 R1049 C-5 R1051 R1052 R1101 R1102 R1103 R1104 R1106 R1107 R1108 R1109 Rim C-5 C-5 B4 C-4 B4 B4 B4 B4 B4 C-4 C-4 R1112 R1113 R1114 R1116 R1117 R1118 R1119 C-4 R1121 C-4 R1122 C-4 C-4 C-4 C4 C4 C-4 C4 R1123 R1124 C-4 R1201 D-5 C-4 R1202 D-5 R1203 D-4 R1204 D-4 R1205 D-5 R1206 D-5 R1207 D-5 R1208 D-5 R1209 G-4 R1210 D-5 R1211 R1212 R1213 R1214 R1216 R1217 E-5 E-5 E-4 E-4 E-4 D-5 Item Grid loc. Item Grid R1218 R1219 D-5 R1344 R1346 R1347 R1376 R1377 R1378 R1379 E-4 R1221 R1222 R1223 R1224 R1226 R1227 E-b D-5 E-5 D-5 D-5 D-5 E4 R1276^ R1277 R1278 R1279 R1281 R1282 R1283 > on switch S15 R1284 R1286 R1287 R1288 R1289 R1291 / R1301 E-5 R1302 E-4 R1303 E-5 R1304 E-5 R1305 E-4 R1306 E-5 R1307 E-5 R1308 E-5 R1309 E-5 R1311 R1312 R1313 R1314 R1316 R1317 R1318 R1319 R1321 R1322 R1324 R1326 R1327 R1328 R1329 R1330 R1331 R1332 R1333 R1334 R1336 R1337 R1338 R1339 R1340 R1341 R1342 R1343 E-5 E-5 D4 E-5 E4 E-5 E4 E4 D4 E4 D4 D4 E4 D4 E4 D-4 'I ^ on switch SI o R138’ R1382 R1383 > R1401 F-4 R1402 F4 R1403 F-4 R1404 F-5 R1406 F4/F-5 R1407 F-5 R1408 F-5 R1409 F-5 R1411 F-5 R1412 R1413 R1414 R1416 R1417 R1418 R1419 R1421 R1422 R1423 R1424 R1425 R1426 R1427 R1428 R1429 R1431 R1432 R1433 R1434 R1436 F-5 R143:^ G-3 R1438 R1439 G4 F-5 F4 F4 F4 F-4 F4 F-4 F-4 G-3 R 1 544 R1546 R1547 R1548 R1549 G-2 G-3 G-3 E-4 R1501 R1502 R1503 E-4 R 1 504 E-4 E-4 E-4 G4 G-4 G-5 G-5 G-5 G-5 G-5 G4 G-5 G-4 G-4 G-4 G-4 G-5 G-4 G-5 G-4 G-4 G-5 on tube G-4 G-4 power power power power power power power power power power G-2 G-3 G-2 G-3 G-2 G-2 ^ on R6 1 ^ F-4 G4 G4 G4 supply supply supply supply supply supply supply supply supply supply power supply on tube R1552 R1553 G-5 R1601 B-2 R1602 B-2 R1603 B-2 R1604 B-2 R1606 B-2 R1607 B-2 R1608 B-2 R1609 B-2 R1611 B-2 R1612 B-2 R1613 B-2 R1614 B-2 R1616 B-2 R1617 B-2 R1618 B-2 R1619 B-2 R1651 F-5 I E-4 G-4 G-3 •N R1451 G4 R1551 G-3 loc. G-4 G-3 R 1443 R1449 R1522 R1524 R1526 R1527 R1528 R1529 R1531 R1532 R1533 R1534 R1535 R1536 R1537 R1538 R1539 G-3 E-4 E-4 R1521 G-3 E-4 E-4 F-5 R1542 R1543 G-3 E-4 R1506 R1507 R1508 R1509 R1511 R1512 R1513 R1514 R1516 R1517 R1518 R1519 R1541 R1444 R1446 R1447 E-4 Grid F-5 G-2 D-4/E-4 Item F4 R1441 R1442 R1443 E4 loc. Item Grid loc. Item Grid loc. Item Grid loc. Item Grid loc. Item Grid loc. R1652 R1653 R1654 R1656 R1657 R1658 R1659 R1661 R1662 R1663 R1664 R1666 R1667 R1668 R1669 R1671 R1672 R1673 R1674 R1676 R1677 R1678 R1679 R1681 R1682 R1683 R1684 R1686 R1687 R1688 F-5 V232 V233 V234 V236 V237 V238 V239 V241 V242 V243 V244 V246 V247 V351 V352 V353 V354 power supply V701 E-2 E-3 VI 302 VI 303 VI 304 E-5 V1516 V1517 V1518 V1519 V1521 G-4 V702 V703 V704 V801 V802 V803 V804 V806 V807 V808 V809 V1218 V1219 V1221 V1301 E-5 power supply power supply power supply G-4/G-5 E-5 VI 527 VI 528 V1601 B-2 VI 001 VI 002 VI 003 VI 004 VI 006 VI 007 VI 008 VI 009 VI oil C-4 B-2 E4 V1602 VI 603 D-4 VI 604 B-2 C-5 V1305 V1306 V1307 V1308 V1314 V1316 V1318 V1319 V1321 VI 522 VI 523 VI 524 VI 526 E-4 B-5 VI 322 E-4 C-4/D-4 V1323 E-4 C-5 VI 324 VI 326 V1401 D-4 V1012 V1013 V1014 V1016 V1017 V1018 V1019 C-5 V1402 V1403 V1404 V1406 V1407 V1408 F-5 VI 409 V1411 F-5 F-4/F-5 C-4 V1412 V1413 C-4 VMM F-4 C-5/D-5 VM16 G-2 C-5/D-5 VMM G-2 G-3 V1651 V1652 V1653 V1654 V1655 V1656 V1657 V1658 V1659 V1661 V1662 V1663 VI 664 VI 666 V1667 VI 668 D501 D601 D801 D1001 D1101 D1201 D1202 F-5 F-5 F-5 F-5 F-5 F-5 F-5 F-5 F-5 F4 F-4 F-4 F-4 F-5 F-4 F-4 E4 F4 F-5 F-4 F-4 power supply power supply power supply power supply power supply power supply power supply power supply power supply att. unit att. unit att. unit att. unit V451 att. unit V452 V453 att. unit att. unit V501 V504 C-2 C-2 F-5 F-5 E-2 F-4 V508 V509 F-4 V511 D-2 F-5 F-4 F4 F-4 VI V201 tube power supply power supply V202 V203 power supply power supply V204 power supply V206 power supply V207 power supply V208 power supply V209 power supply V211 power supply V212 power supply V213 power supply V214 power supply V216 power supply V217 power supply V218 power supply V219 power supply V221 power supply V222 power supply V223 power supply V224 V226 high V227 V228 ^ tension unit V229 V231 / "" E-2 V512 V513 V514 V518 V519 D-2 V521 F-2 E-2 E-2 E-2 E-2 V522 V523 V524 V526 F-2 V601 C-3 V604 V608 V609 C-3 veil D-3 V612 V613 V614 V616 V617 V618 V619 V621 V622 V623 V624 V626 F-2 F-2 F-2 E-3 E-3 D-3 E-3 E-3 E-3 E-3 E-3 E-3 F-3 F-3 F-3 F-3 F-3 VI 020 VI 021 VI 022 VI 023 VI 024 VI 026 VI 027 E-2 E-3 F-2/F-3 F-3 F-3 G-2 G-2 F-2 F-2 F-2 C-4 C-4 C-5 c-4 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 VI 028 C-5 V1 101 B-4 V1 102 B-4 V1 103 C-4 V1 104 C-4 V1 106 C-4 V1 107 C-4 V1 108 C-4 V1 109 C-4/D-4 VI 201 D-5 VI 202 D-5 VI 206 D-4 VI 207 D-5 VI 208 D-5 VI 209 E-5 V1211 D-5 V1212 V1213 V1214 V1216 V1217 D-5 E-5 E4/E-5 E-5 E-5 E-5 E-5 D4 E-4 E-4 E-4 E-4 F-5 F-4 F4 F4 F-5 F-5 F-4 VM18 VM19 G-3 VM21 VM22 VM23 VM24 VM26 VM27 VM28 VM29 VM31 G-3 G-3 G-4 G-5 G-5 G4 G4/G-5 G4 G-5 B-2 B-2 B-2 F-5 B4/B-5 F-5 F-5 F-5 F-5 F-5 F-5 F-5 F4 E4 F-4/F-5 E-4/F-4 F-4 F-4/F-5 F-4 C-2/D-2 C-3/D-3 F-3 C-5 B-4/C-4 D-5 D-5 G-3 D1203 D-4 G-2 D1204 G-3 D-4/D-5 D-4 G-2 D1301 D1302 G-2 B1 LED G-2 rear panel G-2 T101 T201 G-4 G-4 G-4 F201 K501 K601 K1401 G-4 L201 G-4 G-4 L202 L203 power supply power supply power supply E-4 VI 501 VI 502 VI 503 VI 504 VI 506 VI 507 VI 508 VI 509 VI 511 T202 G-4 L801 G-2 D-5 V1512 G-4 L802 F-2 E-4 V15M G-4 LI 501 trace rot. coil D-5 D-5 G-4 G4 D4 power supply power supply power supply C-2 C-3 F4 147 ^S9 3a 2a 2^5^20 y O 2 0»2 q:> CD w(/) < n t 2 % 2 »- m 2 CD »- (S •“ *- Q Q < tr^ CD i 4R539V mOR862}*n t\^- .— C519 ' -Eim-' si I Ic 82 i £>*+ KS01 CS03 9 •r== I EEWsRg 8® HRSja h t^Clcieor :> I rr \^viV1DUJ 603 S ^ iS S ^ IsieI I ; «—*---. i — » nHHH - . (R52^ "'''^-•f-; ' " wt. — "» R519 =fE«g70rir-. ^JgH K601 ' i o «> ea ea —S'- B3 - T/ “ ^ I I — I ' ‘ HR626 V - h* T— >T C't »- 00 OI CO 00 CO (O to to to crkclcel(ri R6U ) « ^ tv608 HR646 I-V608 CD l^^lV rnc) C61 1 - «1R636 ? jJT^ /CD C601 fT\V61€ LZJceisTnj^® S S S T...n L.«' V613 I I I JEM£ 4R659 i> HR673 c II m\ 1-^%. ) IR821 W9 , t- HR 661 h 1^HR663 h ii y 0Mi^^T3£Jr H ftl43^ (vu^ .uu jfvi^ ZZHsSOi, I / •m £gs mp2Q^ 14 fysuD] I fS" /f'-" ! J-{Hll7> - _ »vt | TO DELAY LINE {R647 untune 4R64Sp-*^ 4 S1E i#' oc cell 2 5^17^ :< gs y\M HSi “ \ I ^ s » CE :--s, StO HR613 55 mmm fyi4i7 aC7D2/f^iv701 . ' ii Is V804 ] V703 \i " »- f ' gn„ V1528 V1527 r ol ' : ; aorj^S? —^ KStMSJ « fN KV :S1B-!' '>';f '^-•*— ' ; ; iiii ^^HRsTlh * yyi$02 VTOOi tDl^L S. R848 I=lc^ ' -^-tsUllMvtepi i L 801 •V61 8 S ^ KC jgc ^ y 8I2I IT -^§02 fvK3) HR823 V623 > ^ ^HY62ti> s ^R829 U grvi^ C804p n=Li-ii»i FROM DELAY jSm ^ R828X C806 -R827 k g?QW ‘ I k c.-M :r. c:= LINE ^ RJLS ^ACli"j 25 a (V-^}P£Zlfi te V1422 E >N«iiX ti C286 HR1427h- HRU26H y*i I ; . if tsES si H ' I 1662 K3 :z= CM H meeSV- ^ ™f -n'- qs^eg hp- -i.RU22> 1 -isSwh I n S-^kTSlV-^ V014 ]JL £g i^lgl C1655 Px-I** VV.06 (rui*; V jg -m3- yJS|g -tami! ^viss4 rin Sii-i S3A 0 Vr- JpL a HRUtTV /laf |51 * irg^ 'ki4o,, Hri668I^L i;;~:-d? ^l vuil( V1501 visi4HRisn Tr C3CSL« ^ {rkw) y V140A /t:xi V V HR1403 g i Hr 1401 h\HR1402 1} * 7 d 7 ? ' ': S4B feS4C-tsa> S13(R1383) S13 n I F7 R6 TIME/DIV.(R11) CAL(S14) 513 HF9 TIME/.DIV.(R11) CAL(S14) ri . . 2 tn m S ce “J ^ ^2 XMAGN.(S7) “w ^ “ y ^ if> iZ S u. tl £ tl in to in Fig. 3.40. Vertical amplifier unit with rear side tracks 148 XI CAL EZD=D= TO DELAY LINE ALT TB MAT350 tr_'J Fig. 3.43. Vertical amplifier unit with upper side tracks 149 FOR THE SETTINGS OF THE OSCILLOSCOPE UNDER TEST SEE CHAPTER 3.3.2. FOR MEASURING THESE WAVE FORMS A PROBE 10:1 MUST BE USED MEASURED IN 1V/div. V1501 I— INTENSIFIED V1504 i BAW62 , ! ^f I I ! RELATION MINIMUM TO POINT A C*' MEASURED maxFmui. 20Ais/div. IN I RELATION TO POINT A R15 3IG1 9IG2 G4 G5 G6 6 C7 t TRACE iL1501 2 R1522 \ V1519 J.C150t 7BC548CTlOn ROTATION COIL + 180V IV/div. 20yus/div, TrtACE Irotation Lc 1503 rip C V1521 BC548C. ZIZLp IP -- n1 S _ r mr Ylffl I ^R 256 jr 0 Nr 257 H C2 - 2E 22 -OH n,HH R jjR 261 gR 262 T258 Tc 259 1 c 262 -OH ' -OH 259 15/u 15aj ^ 22 n R 258 |C257 ^R1602 n HM 22 n 15/u J Q^BC548C i 0 @R1606 R1603 ' ' + 12 G + 12 H 22 n O.SV/div. 0, 0,5V/div. O.Sms/div. Sms/di + 12 K R1604B + 12 F 0RI6OI /0\V1601 Nv + 12 J -C1601 JT^ *330n HIGH TENSION UNIT I , CAL. jvi602 VOLTAGE R1611 / ’2'' \ yZ 0 , \bC54«C ^R264 ^R2K |r267 r [ j C267 -OH C266 C269 -OH -OH 15/u 15m HH C268 Rieoe Y id2\V1603 T 12 V ' % R1619 V C271 22 n ^ V XI \/yBC548C -IH 22n 15 /J ^1616 Sr»^1607 ^R1612 L 4 I h §R1617: . .C1602 P47p ' 1 VOLTAGE ADAPTOR 3R227 POWER ON TO TRANSRDRMER POINTS 2 0 4-7 3-6 1 5 ISl CQY24A-1 I S23 •T220/U J3: L202 i,C227 J,c: J470fi J6l B R200 » W3 ^ BY225/2C J220AJ J3: Xc231 1 -F EXT. D.C. SUPPLY X7'|^ 22...27 VOLT MAX.t.tA J4. SI SI STa Sl"b S2 S3 CONNECTIONS 1-2 3-4 6-7 5-6 8-9 10-11 TURNS RMS. VOLTAGE 530 82 82 530 115 115 110V 17V 17V 110 V 24V 29sn 4,50 4.50 29,50 10 24V COIL R20*C± 10’C jer 0.224 10.224 0.224 0.224 0.63 10 0,63 Fig. 3.44. Circuit diagram of power supply , Z-ampiifier and C.R.T. circuit 150 Fig. 3.45. Circuit diagram of the vertical amplifiers FOR THE SETTINGS OF THE OSCILLOSCOPE UNDER TEST SEE CHAPTER 3.3.2. FOR MEASURING THESE WAVE -FORMS A PROBE 10:1 MUST BE USED V V MAT Fig. 3.46. Circuit diagram of the main 310 and delayed time-b^ses