Uploaded by Durjoy Chanda

IAL Physics Unit 3 Practical Skills

advertisement
UNIT – 3
Uncertainty
Every measuring instrument has some uncertainty. This uncertainty is always added or subtracted
from the reading that we receive. It means if the measuring instrument has an uncertainty of 0.1
units, and the reading which is found is 10 units, then the absolute uncertainty is 10±0.1.
The length of the an object AB is measured by a measuring instrument which has smallest division of
1cm. The object is placed close to the measuring instrument and parallel to it to minimize
experimental error.
Maximum
Minimum
𝑅! = 4 + 1
𝑅" = 20 + 1
𝑅! = 4 − 1
𝑅" = 20 − 1
There are two types of readings, the maximum reading and minimum reading.
𝐿#$% = 𝑅" – 𝑅! = (20 + 1) − (4 − 1) = 18
𝐿#&' = 𝑅" – 𝑅! = (20 − 1) − (4 − 1) = 14
The smallest division of a measuring instrument is called precision. If a physical quantity is measured
by using a measuring instrument, it gives a range of values of measurement rather than a particular
magnitude. This range depends on the precision of the measuring instrument. Usually, this range is
equal to twice the smallest division of the measuring instrument. A fraction of the smallest division
of the measuring instrument cannot be determined using the particular instrument. For a single
measurement, uncertainty is equal to its precision. As the number of measurement increases,
uncertainty also increases.
Uncertainty of measuring instruments can be represented in three different ways:
1. Absolute uncertainty
2. Fractional uncertainty
3. Percentage uncertainty
Absolute uncertainty
Absolute uncertainty is represented with unit. For example,
(𝑥 + 𝛥𝑥) cm
or
𝑥 cm ±𝛥𝑥 cm
Fractional uncertainty
The ratio of uncertainty and the measured value is called fractional uncertainty. For example,
2𝑥 ±
∆%
%
3 cm
Percentage uncertainty
Fractional uncertainty multiplied by 100 gives the percentage uncertainty.
Some Common Instruments and Their Precision
Measuring Instrument
Metre Rule
Vernier Caliper
Micrometer Screw Gauge
Stopwatch
Protractor
Thermometer
Description
It is suitable for measuring lengths from 0 to
1m. For large length, measuring tape is used.
It is suitable for measuring diameters, and
lengths from 0 to 2.5cm.
It is suitable for measuring diameters, and
lengths from 0 to 2.5cm.
It is used to measure time. A large
uncertainty is produced due to human
reaction. The average human reaction time
is 0.25s.
It is used to measure angles.
It is used to measure temperature.
Precision
1 mm
0.1 mm
0.01 mm
0.01 s
10
1 0C
Uncertainty of Multiple Instruments/Readings
For Length,
𝐿! = 𝑎 ± 𝛥𝑎
𝐿" = 𝑏 ± 𝛥𝑏
Total Length,
𝐿!)" = (𝑎 ± 𝛥𝑎) + (𝑏 ± 𝛥𝑏)
𝐿!)" = (𝑎 + 𝑏) ± (∆𝑎 + ∆𝑏)
𝐿!)" = 𝐿* ± ∆𝐿
Difference between lengths,
𝐿!+" = (𝑎 ± 𝛥𝑎) − (𝑏 ± 𝛥𝑏)
𝐿!+" = (𝑎 − 𝑏) ± (∆𝑎 + ∆𝑏)
𝐿!+" = 𝐿* ± ∆𝐿
For Area,
Area,
𝐴 = (𝑥 ± 𝛥𝑥)(𝑦 ± 𝛥𝑦)
𝐴 = 𝑥𝑦 ± 𝑥𝛥𝑦 ± 𝑦𝛥𝑥 ± 𝛥𝑥𝛥𝑦
𝐴 = 𝑥𝑦 ± (𝑥𝛥𝑦 + 𝑦𝛥𝑥 + 𝛥𝑥𝛥𝑦)
𝐴 = 𝐴 ± 𝛥𝐴
Errors
Experimental Errors
Reliability of an experiment decreases due to experimental errors. There are two types of
experimental error.
Random Error:
These errors are produced due to variation in performance of instruments or the operator of the
experiment. During this error, readings are shifted along both side of the actual value.
Some sources of random error are:
•
•
•
Variations of temperature
Readings taken under lack of concentration
Parallax error
Random errors can be minimized by:
•
•
Taking large number of measurements.
Producing a best fit curve with graphical method.
Systematic Error:
These errors are produced due to measuring apparatuses. For this error, the reading may be shifted
to a particular side only. The reading will either be increased or decreased from the actual value.
This error cannot be minimized by taking several readings and calculating average. It can only be
reduced by using more precise instruments. Thus, systematic error is also known as instrumental
error.
Some sources of systematic errors are:
•
•
•
Zero error of instrument
Poor calibration of instrument
Instrumental parallax error
Some Common Techniques to Improve an Experiment
•
•
•
•
•
•
•
•
•
•
Check zero error of instrument.
Repeat the measurement at appropriate position.
Different method of measurement should be used.
Readings should be taken from eye level to avoid parallax error.
Use set square to ensure vertical position where necessary.
Using trigonometric calculations to measure angle.
All readings must have same precision.
Sufficient number of readings should be taken for graph plotting purposes (at least 6 for a
straight line and 9 for a curve).
Each reading should be repeated and the average should be taken.
There must be consistent gap between readings.
Download