M7 Series
Diagnostic Ultrasound
System
Service Manual
Revision 9.0
Contents
Contents ..............................................................................................................................................i
Revision History ................................................................................................................................. I
Intellectual Property Statement .......................................................................................................... I
Responsibility on the Manufacturer Party ......................................................................................... II
Warranty ........................................................................................................................................... III
Important Information ........................................................................................................................ V
Safety Precautions ........................................................................................................................... VI
1
2
System Overview ........................................................................................................ 1-1
1.1
Intended Use ........................................................................................................................ 1-1
1.2
Contraindication ................................................................................................................... 1-1
1.3
Product and Model Code ..................................................................................................... 1-1
System Structure ........................................................................................................ 2-1
2.1
Introduction of Each Unit ...................................................................................................... 2-1
2.2
Peripherals Supported ......................................................................................................... 2-3
2.3
Extend Modules.................................................................................................................... 2-3
2.4
3
4
2.3.1
Transducer Extend Module........................................................................................... 2-3
2.3.2
I/O Extend Module. ....................................................................................................... 2-5
2.3.3
V/A Extend Module ....................................................................................................... 2-6
2.3.4
ECG Module ................................................................................................................. 2-8
Control Panel........................................................................................................................ 2-9
Principle Description .................................................................................................. 3-1
3.1
Electric Principle of the System............................................................................................ 3-1
3.2
Principle of Boards ............................................................................................................... 3-2
3.2.1
Power System............................................................................................................... 3-2
3.2.2
Principle of the Main Board ........................................................................................ 3-17
3.2.3
Principle of Transmitting Board................................................................................... 3-47
3.2.4
4D Driving Board ........................................................................................................ 3-52
3.2.5
Principle of Keyboard Board ....................................................................................... 3-56
3.2.6
Display Unit................................................................................................................. 3-65
3.2.7
Principle of ECG Board .............................................................................................. 3-69
3.2.8
Principle of Video/Audio Capture Card ....................................................................... 3-73
3.2.9
Principle of the CW Board .......................................................................................... 3-74
Structure and Assembly/Disassembly...................................................................... 4-1
4.1
Explosive Figure of the Structure ......................................................................................... 4-1
4.1.1
Explosive Figure of the Complete System ................................................................... 4-1
i
4.1.2
Explosive Figure of the Modules .................................................................................. 4-3
4.2
Field Replaceable Unit ......................................................................................................... 4-6
4.3
Assembly/Disassembly ...................................................................................................... 4-14
4.3.1
Removing Batteries .................................................................................................... 4-14
4.3.2
Removing Display Assembly ...................................................................................... 4-14
4.3.3
Removing Top Cover of Main Unit .............................................................................. 4-19
4.3.4
Removing the Power Module ..................................................................................... 4-21
4.3.5
Removing CW Board and Installation......................................................................... 4-24
4.3.6
Removing CPU Fan .................................................................................................... 4-26
4.3.7
Removing CPU Board and Radiator Module .............................................................. 4-27
4.3.8
Removing the System Fan ......................................................................................... 4-29
4.3.9
Remove the Speakers ................................................................................................ 4-30
4.3.10 Removing Transmission Board................................................................................... 4-31
4.3.11 Removing Main Board and Transducer Board Assembly ........................................... 4-32
4.3.12 Remove the Transducer Board................................................................................... 4-34
4.3.13 Removing 4D Board and Installation .......................................................................... 4-35
4.3.14 Removing Hard Disk ................................................................................................... 4-37
4.3.15 Removing Transducer Extend Module PEM-11 ......................................................... 4-38
4.3.16 Removing ECG Module .............................................................................................. 4-41
4.3.17 Removing Video and Audio Capture Module ............................................................. 4-43
4.3.18 Removing I/O Extend Module..................................................................................... 4-44
5
System and Software ................................................................................................. 5-1
5.1
XPE System Introduction ..................................................................................................... 5-1
5.2
Introduction of Doppler ......................................................................................................... 5-2
5.2.1
Resume Package ......................................................................................................... 5-2
5.2.2
Data Processing ........................................................................................................... 5-2
5.3
Run the Resume Package in Doppler .................................................................................. 5-4
5.4
Power ON Process ............................................................................................................... 5-5
5.5
Software Maintenance.......................................................................................................... 5-6
5.6
5.5.1
Entering Preset ............................................................................................................. 5-6
5.5.2
Viewing System Information ......................................................................................... 5-7
5.5.3
Manage Settings ........................................................................................................... 5-7
5.5.4
Maintenance Status ...................................................................................................... 5-9
5.5.5
Model Setup................................................................................................................ 5-10
5.5.6
Installing Optional Software ........................................................................................ 5-10
Warning .............................................................................................................................. 5-11
5.6.1
Battery warning ........................................................................................................... 5-11
5.6.2
PHV Exception Warning Verification .......................................................................... 5-15
5.6.3
Fan Warning ............................................................................................................... 5-16
ii
6
Volt Warning................................................................................................................ 5-17
Note ...................................................................................................................................... 6-1
6.2
System Running Status ........................................................................................................ 6-1
6.5
6.2.1
Running Status ............................................................................................................. 6-1
6.2.2
Working Condition ........................................................................................................ 6-1
General exam ....................................................................................................................... 6-2
6.3.1
Check Flow ................................................................................................................... 6-2
6.3.2
Checking Content ......................................................................................................... 6-2
Function Checking................................................................................................................ 6-4
6.4.1
Checking Flow .............................................................................................................. 6-5
6.4.2
Content ......................................................................................................................... 6-5
Performance Test ............................................................................................................... 6-11
6.5.1
Test Process ............................................................................................................... 6-11
6.5.2
Test Content................................................................................................................ 6-11
System Self-diagnosis ............................................................................................... 7-1
7.1
Screen Introduction .............................................................................................................. 7-1
7.2
Region of Self-diagnosis Items and Testing Results ............................................................ 7-4
7.3
7.2.1
Info Display Area........................................................................................................... 7-5
7.2.2
Status Bar ..................................................................................................................... 7-6
7.2.3
Test Report ................................................................................................................... 7-7
Description of Self-diagnosis Test Items .............................................................................. 7-9
Care and Maintenance................................................................................................ 8-1
8.1
8.2
8.3
9
5.6.5
6.1
6.4
8
Temperature Warning ................................................................................................. 5-17
Function and Performance Checking Method ......................................................... 6-1
6.3
7
5.6.4
Overview .............................................................................................................................. 8-1
8.1.1
Tools, Measurement Devices and Consumables ......................................................... 8-1
8.1.2
Care and Maintenance Items ....................................................................................... 8-2
Cleaning ............................................................................................................................... 8-3
8.2.1
Clean the System ......................................................................................................... 8-3
8.2.2
Content ......................................................................................................................... 8-3
8.2.3
Clean the Peripherals ................................................................................................... 8-5
Checking .............................................................................................................................. 8-6
8.3.1
General check............................................................................................................... 8-6
8.3.2
System Function Check ................................................................................................ 8-7
8.3.3
Peripherals and Options Check .................................................................................... 8-7
8.3.4
Mechanical Safety Inspection ....................................................................................... 8-8
8.3.5
Electrical Safety Inspection .......................................................................................... 8-9
Troubleshooting of Regular Malfunctions................................................................ 9-1
iii
9.1
Inspection and Repair Work Flow of Power System Failure ................................................ 9-1
9.1.1
Inspection and Repair Work Flow When Powered by Adapter..................................... 9-1
9.1.2
Inspection and Repair Work Flow When Powered by Batteries ................................... 9-2
9.1.3
Inspection and Repair Work Flow of the Charging Unit of the Power System ............. 9-3
9.2
Location and Repair of System Boot Failure Malfunctions .................................................. 9-3
9.3
Location and Repair of Non-echo Failure in Image Region ................................................. 9-9
9.4
Location and Repair of the Problem of Image with Black Area .......................................... 9-11
9.5
Abnormalities in Image Region .......................................................................................... 9-11
9.6
Inspection and Repair Work Flow of USB Port Failure ...................................................... 9-12
9.7
Inspection and Repair Work Flow of Network Failure ........................................................ 9-13
9.8
Troubleshooting for 4D Board ............................................................................................ 9-13
9.9
Troubleshooting for the Keyboard Board ........................................................................... 9-15
9.10 Troubleshooting for ECG Board ......................................................................................... 9-16
9.11 Troubleshooting for Video & Audio Capture Card .............................................................. 9-18
Appendix A
Electrical Safety Inspection .................................................................. A-1
Appendix B
Phantom Usage Illustration................................................................... B-1
iv
Revision History
Mindray may revise this publication from time to time without written notice.
Revision
Date
Reason for Change
1. Increase “4.2 Field Replaceable Unit” list with pictures and
important information.
2. Change “6 Maintenance Requirement” to “6 Function and
Performance Checking Method”, increase detailed information
of checking.
5.0
2012.3.30
3. Change “7 Checking” to “6 Care and Maintenance”, increase
detailed information of maintaining.
4. Change “8 System Maintenance” to “7 System Self-diagnosis”,
move the content of cleaning to the chapter 8.
5. Increase “Electrical Safety Inspection” normative description
and delete original content of electrical safety.
6. Increase CW installation method in 4.3.5.
7. Modify several clerical errors in the initial release.
1. Add FRU part number for HDD of different regions/types
(CE/FDA/human/vet) in 4.2
6.0
2013.4.17
7.0
2013.6.24
Add “The attentions to the assembly/disassembly, otherwise the
hard disk will be damaged” to Chapter 4.3.14.
8.0
2015.2.15
Add SONY UP-D898MD, UP-X898MD printers in chapter 2.2
along with compatibility description.
9.0
2016.1.14
Change the order number of the display assembly in chapter 4.2
2. Add warning that human resume package is not compatible with
vet machine, and vet resume package is not compatible with
human machine.
© 2016 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All Rights Reserved.
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray)
owns the intellectual property rights to this Mindray product and this manual. This manual may refer
to information protected by copyright or patents and does not convey any license under the patent
rights or copyright of Mindray, or of others.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure of
the information in this manual in any manner whatsoever without the written permission of Mindray
is strictly forbidden.
I
Release, amendment, reproduction, distribution, rental, adaptation, translation or any other
derivative work of this manual in any manner whatsoever without the written permission of Mindray
is strictly forbidden.
,
,
,
,
BeneView, WATO, BeneHeart,
are the trademarks, registered or otherwise, of Mindray in China and other countries. All other
trademarks that appear in this manual are used only for informational or editorial purposes. They
are the property of their respective owners.
Responsibility on the Manufacturer Party
Contents of this manual are subject to change without prior notice.
All information contained in this manual is believed to be correct. Mindray shall not be liable for
errors contained herein or for incidental or consequential damages in connection with the furnishing,
performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product, only if:
z
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel;
z
the electrical installation of the relevant room complies with the applicable national and
local requirements; and
z
the product is used in accordance with the instructions for use.
z
MindrayVNC ver. 1.0 is free open source software, the performance of MindrayVNC ver.
1.0 is not guaranteed by MINDRAY.
Note
This equipment must be operated by skilled/trained clinical professionals.
Warning
It is important for the hospital or organization that employs this equipment to carry out a reasonable
service/maintenance plan. Neglect of this may result in machine breakdown or personal injury.
II
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED
OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY
PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other
charges or liability for direct, indirect or consequential damages or delay resulting from the improper
use or application of the product or the use of parts or accessories not approved by Mindray or
repairs by people other than Mindray authorized personnel.
III
This warranty shall not extend to:
„
Malfunction or damage caused by improper use or man-made failure.
„
Malfunction or damage caused by unstable or out-of-range power input.
„
Malfunction or damage caused by force majeure such as fire and earthquake.
„
Malfunction or damage caused by improper operation or repair by unqualified or
unauthorized service people.
„
Malfunction of the instrument or part whose serial number is not legible enough.
„
Others not caused by instrument or part itself.
Company Contact
Manufacturer:
E-mail Address:
Shenzhen Mindray Bio-medical Electronics Co., Ltd.
service@mindray.com.cn
Tel:
+86 755 26582479 26582888
Fax:
+86 755 26582934 26582500
EC-Representative:
Address:
Shanghai International Holding Corp. GmbH(Europe)
Eiffestraβe 80, Hamburg 20537, Germany
Tel:
0049-40-2513175
Fax:
0049-40-255726
IV
Important Information
1. It is the customer’s responsibility to maintain and manage the system after delivery.
2. The warranty does not cover the following items, even during the warranty period:
(1) Damage or loss due to misuse or abuse.
(2) Damage or loss caused by Acts of God such as fires, earthquakes, floods, lightning, etc.
(3) Damage or loss caused by failure to meet the specified conditions for this system, such as
inadequate power supply, improper installation or environmental conditions.
(4) Damage or loss due to use of the system outside the region where the system was
originally sold.
(5) Damage or loss involving the system purchased from a source other than Mindray or its
authorized agents.
3. This system shall not be used by persons other than fully qualified and certified medical
personnel.
4. DO NOT make changes or modifications to the software or hardware of this system.
5. In no event shall Mindray be liable for problems, damage, or loss caused by relocation,
modification, or repair performed by personnel other than those designated by Mindray.
6. The purpose of this system is to provide physicians with data for clinical diagnosis. It is the
physician’s responsibility for diagnostic procedures. Mindray shall not be liable for the results of
diagnostic procedures.
7. Important data must be backed up on external memory media.
8. Mindray shall not be liable for loss of data stored in the memory of this system caused by
operator error or accidents.
9. This manual contains warnings regarding foreseeable potential dangers, but you shall always
be alert to dangers other than those indicated as well. Mindray shall not be liable for damage or
loss that results from negligence or from ignoring the precautions and operating instructions
described in this operator’s manual.
10. If the manager for this system is changed, be sure to hand over this operator’s manual to the
new manager.
V
Safety Precautions
1. Meaning of Signal Words
In this manual, the signal words
“
DANGER”, “
WARNING”, “
CAUTION”,
“NOTE” are used regarding safety and other important instructions. The signal words and their
meanings are defined as follows. Please understand their meanings clearly before reading this
manual.
Signal Word
DANGER
Meaning
Indicates an imminently hazardous situation that, if not avoided, will
result in death or serious injury.
WARNING
Indicates a potentially hazardous situation that, if not avoided, could
result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that, if not avoided, may
result in minor or moderate injury.
NOTE
Indicates a potentially hazardous situation that, if not avoided, may result in
property damage.
2. Meaning of Safety Symbols
Symbol
Description
Type-BF applied part.
The ultrasound probes connected to this system are type-BF applied parts.
The ECG leads applied on this system are type-BF applied parts, too.
“Attention” indicates the points that you should pay attention to. Be sure to
read the operator’s manual concerning these points before using the system.
VI
3. Safety Precautions
Please observe the following precautions to ensure patient’s and operator’s safety when using this
system.
Do not use flammable gasses, such as anesthetic gas, oxygen or
DANGER:
hydrogen, or flammable liquids such as ethanol, near this system,
because there is danger of explosion.
WARNING:
1. Do connect the power plug of this system and power plugs of the
peripherals to wall receptacles that meet the ratings indicated on
the rating nameplate. Using an adapter or multifunctional
receptacle may affect the system’s grounding performance, and
cause the leakage current to exceed safety requirements. You
must use the power cord provided with the system.
2. Use the cable provided with this system to connect the printer.
Other cables may result in electric shock.
3. Before cleaning the system, disconnect the power cord from the
outlet. System failure and electric shock may result.
4. DO NOT use a probe that has a damaged, scratched surface, or
exposed wiring of any kind. Immediately stop using the
transducer and contact Mindray Customer Service Department or
sales representative. There is risk of electric shock if using a
damaged or scratched transducer.
5. This system is not water-proof. Do not use this system in any
place where water leakage may occur. If any water is sprayed on
or into the system, electric shock may result. If water is
accidentally sprayed on or into the system, contact Mindray
Customer Service Department or sales representative.
6. Do not allow the patient to contact the live parts of the ultrasound
system or other devices, e.g. signal I / O ports. Electric shock
may occur.
7. Do not use an aftermarket transducer other than those specified
by Mindray. The transducers may damage the system causing a
profound failure, e.g. a fire in the worst case.
VII
8. Do not subject the transducers to knocks or drops. Use of a
defective transducer may cause an electric shock.
9. Do not open the covers and front panel of the system. Otherwise,
short circuit or electric shock may result.
10. Do not use this system when any digital device such as a
high-frequency electrotome, high-frequency therapeutic device
or defibrillator is applied already. Otherwise, there is a risk of
electric shock to the patient.
11. When moving the system, you should first fold the LCD display,
disconnect the system from other devices (including
transducers) and disconnect the system from the power supply.
Separate the system from the mobile trolley before
transportation.
12. Prolonged and repeated use of keyboards may result in hand or
arm nerve disorders for some individuals. Observe the local
safety or health regulations concerning the use of keyboards.
13. Accessory equipments connected to the analog and digital ports
must comply with the relevant IEC standards (e.g., IEC 60950
information technology equipment safety standard and IEC
60601-1 medical equipment standard).Furthermore, all
configurations must comply with the standard IEC 60601-1-1.It is
the responsibility of the person, who connects additional
equipment to the signal input or output ports and configures a
medical system, to verify that the system complies with the
requirements of IEC 60601-1-1.If you have any questions
regarding these requirements, consult your sales representative.
1. Precautions concerning clinical examination techniques:
This system must be used only by qualified medical professionals.
CAUTION:
This operator’s manual does not describe clinical examination
techniques. The clinician should select the proper examination
techniques based on specialized training and clinical experience.
VIII
2. Malfunctions due to radio wave:
If a radio wave emitting device is used in the proximity of this system,
it may interfere with operations. Do not bring or use devices that
generate radio waves, such as cellular telephones, transceivers, and
radio controlled toys, in the room where the system is installed.
If a person brings a device that generates radio waves near the
system, ask him / her to immediately turn OFF the device.
3. Precautions concerning movement of the system:
When you place the system on the mobile trolley and move them
together, you must secure all objects on the mobile trolley to prevent
them from falling. Otherwise you should separate the system from the
mobile trolley and move them individually.
When you have to move the system with the mobile trolley upward or
downward the stairs, you must separate them first and then move
them individually.
Object placed on the monitor may fall and injure an individual when
moving.
Fasten and fully secure any peripheral device before moving the
system. A loose peripheral device may fall and injure an individual.
4. If the circuit protector is tripped, it indicates that the system or a
peripheral device was improperly shut down and the system is
unstable. You cannot repair the system under this circumstance and
must call the Mindray Customer Service Department or sales
representative.
5. There is no risk of high-temperature burns during normal
ultrasound examinations. It is possible for the surface temperature
of the transducer to exceed the body temperature of a patient due to
environmental temperature and exam type combinations. Do not
apply the transducer to the same region on the patient for a long
time. Apply the transducer only for a period of time required for the
purpose of diagnosis.
6. Do not use the system to examine a fetus for a long period of time.
IX
7. The system and its accessories are not disinfected or sterilized
prior to delivery. The operator is responsible for the cleaning and
disinfection of transducers and sterilization of biopsy brackets
according to the manuals, prior to the use. All
items
must
be
thoroughly processed to completely remove harmful residual
chemicals, which will not only harmful to the human body, but also
damage the accessory.
8. It is necessary to press <End Exam> to end the current scan that is
in progress and clear the current Patient Information field.
Otherwise, new patient data may be combined with the previous
patient data.
9. Do not connect or disconnect the system’s power cord or its
accessories (e.g., a printer or a recorder) without turning OFF the
power first. This may damage the system and its accessories or
cause electric shock.
10. Read the Acoustic Output Principle in the operation manual
carefully before operate this system on clinical examination.
11. The cover contains natural rubber that can cause allergic reactions
in some individuals.
12. Please use the ultrasound gel compliant with the relevant local
regulations.
13. Do not connect this system to outlets with the same circuit breakers
and fuses that control the current of devices such as life-support
systems. If malfunctions or over-current appear on this system, or
when there is an instantaneous current at power ON, the breakers
and fuses of the building’s supply circuit may be tripped.
NOTE:
1. Do not use the system in the vicinity of strong electromagnetic field (such as a
transformer), which may affect the performance of the system.
2. Do not use the system in the vicinity of high-frequency radiation source, which may
affect the performance of the system or even lead to failure.
X
3. To avoid damaging the system, DO NOT use it in following environment:
z
Locations exposed to direct sunlight;
z
Locations subject to sudden changes in environmental temperature;
z
Dusty locations;
z
Locations subject to vibration;
z
Locations near heat generators;
z
Locations with high humidity.
4. Turn ON the system only after the power has been turned OFF for more than 20
seconds. If the system is turned ON immediately after being turned OFF, the system
may not be rebooted properly and could malfunction.
5. Press <Freeze> or turn off the power of the system before connecting or
disconnecting a transducer. Otherwise, the system and / or transducer may be
damaged.
6. Remove ultrasound gel from the face of a probe when the examination is complete.
Water in the gel may enter the acoustic lens and adversely affect the performance
and safety of the transducer.
7. You should properly back up the system to a secure external storage media,
including system configuration and settings and patient data. Data stored to the
system’s hard drive may be lost due to system failure, improper operation or
accident.
8. Do not apply external force to the control panel. Otherwise, the system may be
damaged.
9. If the system is used in a small room, the room temperature may rise. Please provide
proper ventilation and free air exchange.
10. When using or placing the system, keep the system horizontal to avoid disbalance.
11. To dispose of the system or any part, contact Mindray Customer Service Department
or sales representative. Mindray is not responsible for any system content or
accessories that have been discarded improperly. Mindray is not responsible for any
system content or accessories that have been discarded improperly.
12. Electrical and mechanical performance may be degraded due to long usage (such as
current leakage or distortion and abrasion), the image sensitivity and precision may
become worse too. To ensure optimal system operations, it is recommended that you
maintain the system under a Mindray service agreement.
XI
13. Do not use gel, disinfectant, transducers or needle-guided brackets that are not
compatible with the system.
14. The replaceable fuse is inside the chassis. Refer replacing job to Mindray service
engineers or engineers authorized by Mindray only.
15. Do not turn OFF the power supply of the system during printing, file storage or
invoking other system operations. An interrupted process may not be completed, and
can become lost or corrupted.
16. If the system is powered off improperly during operation, it may result in data
damage of the system’s hard disk or system failure.
17. Do not use a USB memory device (e.g., an USB flash drive, removable hard disk)
which has unsafe data. Otherwise, system damage may result.
18. It is recommended to only use the video devices specified in this manual.
19. When using this system with portable package, be careful not to block the ventilation
slots.
20. DO NOT directly remove a USB memory device during file being copied; otherwise,
the USB memory device and/or the system may be damaged.
21. Do not connect a removable hard disk of many partitions to the system.
22. Do not press any key on the control panel when perform DICOM file sending or file
duplication from the USB memory devices.
4. Warning Labels
The warning labels are attached to this system in order to call your attention to potential hazards.
The symbol
on the warning labels indicates safety precautions. The warning labels use the
same signal words as those used in the operator’s manual.
Refer to the operator’s manual for detailed information about the warning labels. Read operator’s
manual carefully before using the system.
XII
5. Symbol Explanation
Symbol
Name
Function
Power off
Power button
│
Power on
Remote control port
Video print control
Video out
VGA out
I/O panel
Footswitch port
USB
ETHERNET
A
Transducer port A
Transducer port
B
Transducer port B
XIII
1
1.1
System Overview
Intended Use
This product is applicable in ultrasound examining room, ward, emergency, anesthesia room and
operating room, etc.
M7, M7T, M7 Expert, M7 Super
Applicable for examinations to adults, pregnant women, pediatric patients and neonates, and it is
intended for use in abdominal, cardiac, small parts (breast, testes, thyroid, etc.), peripheral vascular,
fetal, transesophagel, transrectal, transvaginal, pediatric, neonatal cephalic, musculoskeletal
(general and superficial), intraoperative, interventional, anesthesia (nerve), emergency, ICU/CCU,
urology, sports health and transcranial exams.
M7Vet
M7Vet is a veterinary use color Doppler diagnostic ultrasound system, which is intended for use in
clinical ultrasonic exams, such as abdominal, cardiac, small parts, peripheral vascular, transrectal,
musculoskeletal (general and superficial), and intraoperative exams.
1.2
Contraindication
None.
1.3
Product and Model Code
M–□
Model code
Product code
1-1
2
System Structure
2.1
Introduction of Each Unit
Fig 2-1 External Appearance of the System (1)
Fig 2-2 External Appearance of the System (2)
2-1
System Structure
Fig 2-3 External Appearance of the System (3)
No.
Name
Function
Displays the images and parameters during
1
Display
2
Control panel
Operator-system interface or control
3
Handle
Used for carrying the system
4
Transducer port
scanning.
Connects a probe to the main unit; or
connects a probe extend module
Locks or unlocks the probe connected with
the main unit
5
Transducer locking lever
: Locked symbol
: Unlocked symbol
6
IO extend port
Connects the IO extend module
Power input port
Connects the power adapter
7
8
USB port
Connects USB devices
2-2
System Structure
No.
9
Name
Network port
S-Video separate video output
10
Function
Connects the network
Connects with image signal output devices
like the video recorder, video printer,
text/graph workstation and so on.
2.2
Peripherals Supported
„
B/W video printer: MITSUBISHI: P93DC, SONY:UP-D897, UP-D898MD, UP-X898MD
„
Color video printer: SONY: UP-D23MD
„
DVD Recorder: MVR-11
„
Graph/text printer
NOTE: printer UP-D898MD and UP-X898MD should be used in compatible with 1.00.7 2015-1-27
OS version or above versions. You can also select “DVR: 897” in the DIGITAL -> - DRIVER path on
the printer (use UPD897 driver) if OS is not updated.
2.3
Extend Modules
Note:
Make sure VGA cable is not too long .
There are four extend modules available for the system: probe extend module, IO extend module,
V/A extend module, and ECG module. Interfaces of each extend module is described as follows:
2.3.1
Transducer Extend Module
The system can be configured with 2 models of probe extend modules: PEM-11 and PEM-21.
„
PEM-11
z
Introduction of each part
2-3
System Structure
<1>
<2>
Fig 2-4 PEM-11 Transducer Extend Module
No.
<1>
Name
Connector
Function
Connects to the transducer port of the main unit, three extend
ports are available.
<2>
Probe port
Extended ports for connecting two live transducers; the middle one
is a parking port
z
Connection
Turn off the ultrasound system, insert the module connector into the probe port with its bulge facing
to the right side of the system (when you are facing to the front of the system).
Fig 2-5 PEM-11 Connect the Transducer Extend Module
2-4
System Structure
„
PEM-21
Fig 2-6 PEM-21 Transducer Extend Module
PEM-21 transducer extend module can be used only if the mobile trolley is configured, for the
figures, please refer to the service manual of UMT-300.
CAUTION:
You must turn off the ultrasound system before connecting or
disconnecting the probe extend module. Otherwise the system
may be damaged.
NOTE:
If you use the probe extend module to connect a probe, the image quality may
be degraded.
2.3.2
„
I/O Extend Module.
Interfaces
Fig 2-7 I/O Extend Module
2-5
System Structure
No.
Symbol
Function
<1>、<2>
USB port
Connects USB devices.
<3>
ECG Port
Connects the ECG Module
<4>
Serial port
Connects serial port devices
<5>、<6>
Used for audio signals of D mode sound from
Audio output port
DVD output or audio comments
Reserved. (Connects a microphone used for
<7>
Mic In port
receiving audio comments when a recorder is
used to record images)
<8>
Remote control port
Composite video output port
<9>
Connects the control port of the video printer
Connects with image signal output devices like
the video recorder, video printer, text/graph
workstation and so on.
<10>
„
DVI-I output port
Connects a display or projector, etc.
Connection
Connect the I/O extend module to the main unit via the I/O extend port, thus the data port is
extended. As shown in the following figure.
I/O extend module
Fig 2-8 PEM-11 Connect the I/O Extend Module
2.3.3
„
V/A Extend Module
About the Interfaces
The module is connected to the USB port of the main unit via a USB cable.
2-6
System Structure
< 1>
< 3>
<2>
<4>
Fig 2-9 V/A Extend Module
No.
„
Name
Function
<1>
Audio input port
Used for audio signal input
<2>
Audio input port
Used for audio signal input
<3>
Composite video input port
Used for composite signal video input
<4>
Separate video input port
Used for separate signal video input
Connection
Connect the V/A extend module to the main unit via an USB port, after that, you can see USB 2861
Device and USB Audio Device in the device manager. V/A extend module connection is shown in
the following figure.
Fig 2-10 PEM-11 Connect the V/A Extend Module
2-7
System Structure
2.3.4
„
ECG Module
Interfaces
Fig 2-11 ECG Module
Name
ECG lead port
„
Function
Used for ECG signal input
ECG Lead
Two kinds of ECG leads are available for M7: AHA standard and IEC standard. For FDA region, only
the ECG leads of AHA standard can be selected, and for other regions, both kinds can be used.
„
Connection
To use the ECG module, you need to configure with the I/O extend module at the same time.
First connect the I/O extend module, and then connect the ECG module to the I/O extend module,
see the figure below:
Fig 2-12 Connect the ECG Module
2-8
System Structure
2.4
Control Panel
<1>
<2>
<8>
<9>
<10>
<3>
<11>
<4>
<12>
<5>
<13>
<14>
<15>
<6>
<7>
<16>
<18>
<17>
<19>
<20>
<21>
<22>
<35>
<25>
<23>
<24>
<36>
4>
<3
<26>
<38>
<37>
<29>
<28>
<31>
<33>
<30>
<32>
<27>
<40>
<39>
<41>
<43>
<50>
<42>
<44>
<49>
<47>
<45>
<48>
<46>
<51><52>
Fig 2-13 Control Panel
No.
<1>
<2>
<3>
Silk Print
/
/
/
Name
Multifunctional
key 1
Multifunctional
key 2
Multifunctional
key 3
Function
Press to select the soft menu items displayed on the
bottom of the screen. Refer to the subsequent
contents for specific functions.
Press to select the soft menu items displayed on the
bottom of the screen. Refer to the subsequent
contents for specific functions.
Press to select the soft menu items displayed on the
bottom of the screen. Refer to the subsequent
contents for specific functions.
2-9
System Structure
No.
Silk Print
Name
Function
Up / down controls are used to turn pages up / down
<4>
/
Multifunctional
key 4
when there are more-than-one pages for the soft
menu;
Left / right keys are used to switch between the
different modes.
<5>
<6>
/
/
Multifunctional
key 5
Multifunctional
key 6
Press to select the soft menu items displayed on the
bottom of the screen. Refer to the subsequent
contents for specific functions.
Press to select the soft menu items displayed on the
bottom of the screen. Refer to the subsequent
contents for specific functions.
Power button
It does not illuminate when the system is turned off;
Press the button to turn on the system, the system
<7>
/
Power button
enters the work status and the indicator lights on in
green.
When the system enters the standby status, the
indicator turns green.
<8>
Esc
Exit
Press to exit the current status to the previous status.
<9>
Help
Help
<10>
Report
Report
<11>
iStation
/
<12>
F1-F4
User-defined key
You can assign a function to the key.
<13>
Quad
Quad-split screen
Press to enter or exit the quad-split screen display.
<14>
Biopsy
Biopsy
Press to show or hide the biopsy guide line.
<15>
Setup
Setup
Press to show or hide the Setup menu.
<16>
Del
/
Press to delete the comment, etc.
<17>
TGC
/
Move to adjust time gain compensation.
Press to open or close the accompanying help
documents.
Press to open or close the diagnosis reports.
Press to enter or exit the patient information
management system.
2-10
System Structure
No.
Silk Print
Name
Alphanumeric
Function
<18>
/
<19>
Menu
Main menu
<20>
Comment
ABC (Comments)
Press to enter or exit the character comment status.
<21>
Arrow
Arrow
Press to enter or exit the arrow comment status.
<22>
Clear
Delete
keys
Same as those of PC.
Press to display or hide a mode-specific parameter
menu.
Press to clear the comments or measurement
calipers on the screen.
Up/down,
<23>
left/right
Fn+/: to adjust volume or monitor brightness
keys
Rotate to increase or decrease the image gain; press
<24>
iTouch
/
to optimize the image, serving as a one-key
optimization.
<25>
Patient
Patient Info
To enter into patient information input interface.
<26>
Exam
Exam Mode
Switch the exam mod and probe
<27>
Review
Review
To review the stored images.
End Exam
Press to end an exam.
<28>
End
Exam
Body
Body Marks
Mark
(Pictograms)
<30>
Cine
Cine Review
Press to enter or exit the Cine Review status.
<31>
Zoom
Zoom
Press to enter or exit the Zoom status.
<32>
Cursor
Cursor
Press to show the cursor.
<33>
F5
User-defined key
You can assign a function to the key.
<34>
Measure
Measurement
<29>
Press to enter or exit the Body Mark status.
Press to enter or exit the application measurement
mode.
2-11
System Structure
No.
Silk Print
Name
Function
Switching key: Press to change the currently active
<35>
Update
/
window.
Press to start or end image capture in 3D/4D or
iScape mode.
<36>
Set
Set
<37>
B
/
Press to confirm an operation, same as the
left-button of the mouse.
Press to enter the B mode
Press to enter the Dual mode from another mode;
<38>
Dual
/
Press to switch between the two display windows in
the Dual mode.
<39>
Color
/
Press to enter the Color mode.
<40>
Power
/
Press to enter Power mode.
<41>
CW
/
Press to enter the CW mode.
<42>
M
/
Press to enter the M mode.
<43>
PW
/
Press to enter the PW mode.
<44>
F6
User-defined key
You can assign a function to the key.
<45>
Print
Print
Press to print: user-defined key.
<46>
Save
Storage
Press to save; user-defined key.
<47>
Depth
Depth
<48>
Freeze
Freeze
Press to freeze or unfreeze the image.
<49>
/
Trackball
Roll the trackball to change the cursor position.
Press to increase or decrease the imaging depth in
scanning mode.
Rotate to adjust the menu items, parameters and
<50>
/
Multifunction knob
direction of the comment arrow, etc.
Press to show the menu, and rotate the view the
items.
2-12
System Structure
No.
Silk Print
Name
Function
Indicates if the main unit is connected to the power
supply.
<51>
/
Indicator 1
If not connected, the indicator does not illuminate;
If connected, the indicator illuminates in green color.
Indicates the current status of the batteries.
When the system is supplied with power by the
batteries and the power capacity is lower than 30%,
the indicator is yellow and flashes.
<52>
/
Indicator 2
When the batteries are being charged, the indicator
light is on and in yellow color.
When the battery capacity is charged to the full
capacity, the indicator color turns green.
In other statuses, the indicator light is off.
Tip: “/” means the key / knob has no silk-printed name.
Keys which can work with the Fn key:
Key
Fn+
Fn+
Fn+S
Fn+
Fn+
Function
Increase the brightness of the LCD display.
Decrease the brightness of the LCD display.
Press the two keys to mute the speaker.
Press the two keys to increase display contrast.
Press the two keys to decrease display contrast.
2-13
3
Principle Description
3.1
Electric Principle of the System
The schematic diagram of M7 diagnostic ultrasound system is show in the following figure:
Fig 3-1 Schematic Diagram of System
According to functions of hardware, the block diagram of the hardware system is shown below:
3-1
Principle Description
Fig 3-2 Block Diagram of the Hardware System
The main board unit is the core of the whole system, including the transducer port, transmitting,
receiving, beam forming, signal processing, system monitor and management, audio, video,
interfaces and so on, and these will be introduced later. Main logic unit completes the main logic
processing functions, including transmitting control logic, beam former logic, signal processing logic.
The functions of the other modules can be explained by their names, and will not be explained here.
Besides, M7 also supports ECG module and video capture module. Accompanied with the trolley,
the transducer extend module can be used.
3.2
Principle of Boards
3.2.1
Power System
3.2.1.1
Basic Functions of Power System
As the power supply of the system, the power system consists of adapter, batteries, main power
board, auxiliary power board, flexible boards for connecting batteries and flexible boards of analog
power, etc. The power is supplied through two methods, external power adapter or batteries. The
system is first powered by the adapter, and the batteries are provided for back-up. The main power
board and auxiliary power board are mainly responsible for supplying the system with the DC output
listed in the following table, charging management of the batteries, monitoring the power, and
communicating with the main board.
Table 3-1 DC Output of Power Supply System
Board of the Function
Project
Output
1
3V3STB
Main power board
2
5VSTB_CPU
Main power board
Controlled by 5Vstb_CPU_EN signal
3
+12V
Main power board
Controlled by POWER_ON# signal
4
+95V
Main power board
Controlled by POWER_ON# signal
5
-95V
Main power board
Controlled by POWER_ON# signal
6
PHV1P
Main power board
Controlled by POWER_ON# signal
7
PHV1N
Main power board
Controlled by POWER_ON# signal
Unit
Notes
Available when there is an adapter or dual
batteries
3-2
Principle Description
8
PHV2P
Main power board
Controlled by POWER_ON# signal
9
PHV2N
Main power board
Controlled by POWER_ON# signal
10
+5V
Auxiliary power board
Controlled by POWER_ON# signal
11
+3.6V
Auxiliary power board
Controlled by POWER_ON# signal
12
+3.3V
Auxiliary power board
Controlled by POWER_ON# signal
13
+2.8V
Auxiliary power board
Controlled by POWER_ON# signal
14
+1.8V
Main power board
Controlled by POWER_ON# signal
15
+1.5V
Auxiliary power board
Controlled by POWER_ON# signal
16
-5V
Auxiliary power board
Controlled by POWER_ON# signal
17
-12V
Auxiliary power board
Controlled by POWER_ON# signal
3.2.1.2
Basic Principle of Power System
The principle diagram of the power system is shown in figure below. The power is supplied through
two methods, external power adapter or batteries. The adapter is connected to the power board via
the system main board when the system is powered by adapter. The batteries are connected to the
power board via the flexible boards when the system is powered by batteries. The main power
board generates 4 PHV outputs: 3V3STB, 5VSTB_CPU, +12V and ±95V, monitors and manages
the power output. The auxiliary power board generates +5V, -5V, -12V, +3.6V, +3.3V, +2.8V, +1.8V
and +1.5V outputs, and manages the charging of the batteries. The main power board and the
auxiliary power board are connected via two 26 PIN connectors.
3-3
Principle Description
Fig 3-3 Schematic Diagram of Power System
The system main board and the digital ports of the power board are connected together via a 64
PIN connector, and the digital power ports and signals are defined in table below; the system
main board and the auxiliary power board are connected together via an analog flexible board,
and the analog power ports and signals are defined in
Table 3-3; the main power board and the batteries are connected together via the battery flexible
board, and the signals are defined in Table 3-4; the main power board and the auxiliary power board
are connected together via two 26 PIN connectors, and the signals are defined in Table 3-5 and
Table 3-6.
Table 3-2 Signal definition of connectors between main power board and system main board
digital power
No.
Signal name
No.
Signal name
1
EDC_PWR
2
EDC_PWR
3
EDC_PWR
4
EDC_PWR
5
EDC_PWR
6
EDC_PWR
7
GND
8
GND
9
GND
10
GND
11
GND
12
GND
13
N12V
14
STS_EDC#
15
GND
16
GND
17
POWER_ON#
18
CPU12V_OK#
19
T_D+
20
T_D-
21
GND
22
GND
23
GND
24
GND
25
D12V
26
D12V
27
D12V
28
D12V
29
GND
30
GND
31
D5V1
32
D5V1
3-4
Principle Description
No.
Signal name
No.
Signal name
33
GND
34
GND
35
5VSTB_OK#
36
5VSTB_EN#
37
GND
38
GND
39
5VSTB_CPU
40
3V3STB
41
GND
42
GND
43
D3V3
44
D3V3
45
GND
46
GND
47
D2V8
48
D1V5
49
GND
50
GND
51
D1V8
52
D1V8
53
GND
54
GND
55
PWR_RX
56
PWR_CONF_N
57
PWR_TX
58
PWR_RST
59
GND
60
GND
61
BATA_PRT#
62
BATB_PRT#
63
BATSTS_O
64
BATSTS_G
Signal definition
EDC_PWR: adapter power input;
GND: ground;
N12V: analog -12V output;
STS_EDC#: adapter normal working signal, low level means the output of adapter is normal;
POWER_ON#: power on signal from the system main board, low level effective;
CPU12V_OK#: +12V output signal, lower level means normal output;
T_D+, T_D-: power board temperature detecting signal;
D12V: digital power +12V output;
D5V1: digital power +5V output;
5VSTB_OK#: 5VSTB output signal, lower level means normal output;
5VSTB_EN#: 5VSTB enable signal from the system main board, low level effective;
3-5
Principle Description
5VSTB_CPU: standby power 5VSTB_CPU output, the voltage is +5V;
3V3STB: standby power 3V3STB output, the voltage is +3.3V;
D3V3: digital power +3.3V output;
D2V8: digital power +2.8V output;
D1V5: digital power +1.5V output;
D1V8: digital power +1.8V output;
PWR_RX: serial port communication receiving port of the the power and system mother board;
PWR_TX: serial port communication transmitting port of the power and system mother board;
PWR_CONF_N: power board chip configuration signal from the system main board;
PWR_CONF_N: power board chip reset signal from the system main board;
BATA _ PRT#: battery A presenting signal, low level means the battery is presenting;
BATA _ PRT#: battery B presenting signal, low level means the battery is presenting;
BATSTS_O: battery status indicating light, orange;
BATSTS_G: battery status indicating light, green.
Table 3-3 Signal and analog power definition between auxiliary power board and system main
board
No.
Signal name
No.
Signal name
1
N95V
2
GND
3
95V
4
GND
5
PHV2N
6
GND
7
PHV1N
8
GND
9
PHV2P
10
GND
11
PHV1P
12
GND
13
A12V
14
GND
15
A2V8
16
GND
17
GND
18
GND
19
A3V6
20
A3V6
21
GND
22
GND
3-6
Principle Description
23
A1V5
24
A1V5
25
A1V5
26
A1V5
27
GND
28
GND
29
NA12V
30
GND
31
NA5V1
32
GND
33
GND
34
GND
35
A5V1
36
A5V1
37
A5V1
38
A5V1
39
GND
40
GND
Signal definition
N95V: -95V fixed high voltage output;
GND: ground;
95V: +95V fixed high voltage output;
PHV2N: PHV2 negative voltage output, -5~-85V;
PHV1N: PHV1 negative voltage output, PW outputs -13~-85V, CW outputs -2~-5V;
PHV2P: PHV2 positive voltage output, 5~85V;
PHV1P: PHV1 positive voltage output, PW outputs 13~85V, CW outputs 2~5V;
A12V: analog +12V output;
A2V8: analog 2.8V output;
A3V6: analog 3.6V output;
A1V5: analog 1.5V output;
NA12V: analog -12V output;
NA5V: analog -5V output;
A5V1: analog +5V output.
Table 3-4 Signal definition between the battery and main power board
No.
Signal name
No.
Signal name
1
BATB+
2
BATB+
3
BATB+
4
BATB+
5
BATB_PRT#
6
GND
3-7
Principle Description
No.
Signal name
No.
Signal name
7
NTC_BATB
8
GND
9
BATB_FB
10
GND
11
BATB_SCL
12
GND
13
BATB_SDA
14
GND
15
BATA_PRT#
16
GND
17
NTC_BATA
18
GND
19
BATA_FB
20
GND
21
BATA_SCL
22
GND
23
BATA_SDA
24
GND
25
GND
26
GND
27
BATA+
28
BATA+
29
BATA+
30
BATA+
Signal definition
BATB+: battery B input;
GND: ground;
BATB_PRT#: battery B presenting signal, low level is effective;
NTC_BATB: battery B temperature signal;
BATB_FB: battery B far end sampling signal;
BATA_SCL: battery B SMBUS communication clock signal;
BATA_SDA: battery B SMBUS communication data signal;
BATA_PRT#: battery A presenting signal, low level is effective;
NTC_BATA: battery A temperature signal;
BATA_FB: battery B far end sampling signal;
BATA_SCL: battery A SMBUS communication clock signal;
BATA_SDA: battery A SMBUS communication data signal;
BATA+: battery A input;
3-8
Principle Description
Table 3-5 Signal definition between main power board and auxiliary power board
No.
Signal name
No.
Signal name
1
VBUS+
2
VBUS+
3
VBUS+
4
GND
5
GND
6
GND
7
5V_CW
8
D5V1
9
START
10
D2V8
11
GND
12
GND
13
N5V_CW
14
A12V
15
GND
16
N12V
17
PHV2P
18
D1V5
19
PHV1P
20
GND
21
PHV2N
22
GND
23
PHV1N
24
GND
25
95V
26
N95V
Signal definition
VBUS+: input bus voltage on the power board;
GND: ground;
5V_CW: LDO input of PHV1P CW voltage;
D5V1: digital power +5V output;
START: auxiliary power board power on signal from the system main board, low level effective;
D2V8: digital power +2.8V output;
5V_CW: LDO input of PHV1N CW voltage;
A12V: analog +12V output;
N12V: digital power -12V output;
PHV2P: PHV2 positive voltage output, 5~85V;
D1V5: digital power +1.5V output;
PHV1P: PHV1 positive voltage output, PW output 13~85V, CW output 2~5V;
PHV2N: PHV2 negative voltage output, -5~85V;
3-9
Principle Description
PHV1N: PHV1 negative voltage output, PW output -13~85V, CW output -2~5V;
95V: +95V fixed high voltage output;
N95V: -95V fixed high voltage output.
Table 3-6 Signal definition between main power board and auxiliary power board (2)
No.
Signal name
No.
Signal name
1
EDC_POWER
2
EDC_POWER
3
ISENSE
4
EDC_POWER
5
GND
6
GND
7
CHARGE_EN
8
GND
9
BATA+
10
BATA+
11
BATA_FB
12
NTC_BATA
13
GND
14
GND
15
BATB_FB
16
NTC_BATB
17
BATB+
18
BATB+
19
GND
20
GND
21
D1V8
22
D1V8
23
GND
24
GND
25
D3V3
26
D3V3
Signal definition
EDC_PWR: adapter power input;
ISENSE: main system load current signal from the power main board to the auxiliary power board;
GND: ground;
CHARGE_EN: battery charge enable signal, low level effective;
BATA+: battery A input;
BATA_FB: battery A far end sampling signal;
NTC_BATA: battery A temperature signal;
BATB_FB: battery B far end sampling signal;
NTC_BATB: battery B temperature signal;
BATB+: battery B input;
3-10
Principle Description
D1V8: digital power +1.8V output;
D3V3: digital power +3.3V output.
3.2.1.3
3V3STB Circuit Unit
Basic Principle of 3V3STB Circuit
The 3V3STB takes EDC_POWER from the adapter or the BATTERY_BUS from the batteries as the
input and generates 3V3STB via linear voltage-stabilizing chip TPS715A01.
TPS715A01 is equipped with current-limiting function itself.
The Maintenance Points of 3V3STB Circuit
If the 3V3STB unit is faulty, first check whether the input voltage of U12 PIN1 is normal; if it is
normal, it indicates that the U12 is damaged or the fault is caused by the over-current or short
circuit of the load.
3.2.1.4
5VSTB_CPU Circuit Unit
Basic Principle of 5VSTB_CPU Circuit
The 5VSTB_CPU takes the BATTERY_BUS from the batteries as input and generates the output
via the synchronous rectifying step-down chip TPS62110 through BUCK transformation.
The TPS62110 chip itself is equipped with MOSFET, current-limiting function and over-temperature
protection.
The Maintenance Points of 5VSTB_CPU Circuit
If the 5VSTB_CPU unit is faulty, first check whether the input voltage of U35 PIN2 is normal and
whether the enabling signal of PIN4 is 3.3V; if they are all normal, it indicates that the U35 is
damaged or the fault is caused by the over-current or short circuit of the load; if PIN4 is low level,
check whether the system main board provides 5VSTB_EN# signal and whether the feedback
connection or overvoltage circuit is normal.
3.2.1.5
+12V Circuit Unit
Basic Principle of +12V
Voltage rise-and-fall IC LTC3780 of LINEAR Company is applied in +12V circuit unit, and whether
the input voltage is higher or lower than 12V, the output voltage will be kept at 12V.
The +12V over-current and overvoltage protection is not locked, and the over-current is over 8A.
The output voltage returns to normal after over-current disappears.
3-11
Principle Description
The Maintenance Points of +12V
Before testing, ensure that the POWER_ON# signal is of low level and the input voltage VBUS+ is
normal.
The work flow of maintenance is shown as follows:
Test the voltage of PIN8 of U14, which should be higher than 3V;
Test the voltage of PIN19 of U14, which should be higher than 6V;
Use an oscilloscope to measure G poles of Q6, Q7, Q8 and Q9 to check whether there is driving
pulse; at least two MOSFETs’ G poles should have drive pulse;
Check if there is any damage to the four MOSFETs Q6, Q7, Q8 and Q9 and diode D5 and D7;
If +12V hiccups, check whether the output is overloaded, and check the feedback and overvoltage
protection circuits.
3.2.1.6
PHV Circuit Unit
Basic Principle of PHV
The PHV circuit consists of PHV1P, PHV1N, HV2P and PHV2N, of which, the PHV1P and PHV1N
consist of PW and CW output. The PW output range of PHV1 is ±13V~±85V, and the CW output
range is ±2V~±4V. Only PW output of PHV2 is available, and the output range is ±5V~±85V.
The input of the PW circuit of the 4 PHVs is controlled +12V, which is under the control of PMOS
Q5.The PW output voltage of PHV1P and PHV2P is generated by BOOST circuit. The controlling IC
is TL594, which is frequently adopted by our company. The range of output voltage of PHV2P is
+5V~+85V, so when the output voltage is over than 15V, the input is +12V; and when the output
voltage is lower than 15V, the input is +5V. Both the two inputs are under the control of output
voltage controlling signal. The PW output voltage of PHV1N and PHV2N is generated by
BUCK-BOOST circuit. The controlling IC is MAX1847, which is frequently adopted by our company.
The CW output of PHV1P and PHV1N is generated through linear voltage-stabilizing. The input of
CW circuit of PHV1P is +5V and the chip LT1963 LDO is adopted for the linear voltage-stabilizing;
the input of CW circuit of PHV1N is -5V, and the chip LT1185 LDO is adopted for the linear
voltage-stabilizing. The voltage of CW output is regulated by the output voltage controlling signal.
When the controlling signal is 0.45~2.25V, the corresponding output voltage is ±2V~±4V.
The 4 PHV are all equipped with overvoltage protection, and the overvoltage protection of each
circuit is independent. The overvoltage will not be locked, and the voltage is about 93V when
overvoltage occurs.
The 4 PHV are all equipped with over-current protection. In PW mode, once over-current occurs, all
the 4 PHV PW outputs and ±95V will be switched off, while the CW output is not influenced. The
3-12
Principle Description
over current protection in CW mode is implemented by current-limiting of LDO chip itself. Only the
output is cut off and other outputs are not influenced.
The Maintenance Points of PHVP PW Circuit Unit
Make sure the +12V output is normal, and PHV_CW is low level before testing.
Test the grid electrode of Q5, and the voltage should be around 3V and VCC2 is 12V.
The work flow of troubleshooting for U17 and U18 and their peripheral circuits are shown as follows:
Test the voltage of PIN8 and PIN11, which should be 12V;
Test the voltage of PIN14, which should be +5V;
Test PIN5, and there should be oscillation wave form, the frequency of which is around 100kHz;
Test the voltage of PIN15, which should be around 5V;
Test PIN9 and PIN10, and there should be driving pulses.
If the test shows the grid electrode of Q5 is 12V, check the +5V output and whether over-current or
short circuit occurs.
The Maintenance Points of PHVN PW Circuit Unit
Make sure the +12V output is normal, and PHV_CW is low level before testing.
Test the grid electrode of Q5, and the voltage should be around 3V and VCC2 is 12V.
The work flow of troubleshooting for U24 and U25 and their peripheral circuits are shown as follows:
Test the voltage of PIN15, which should be 12V;
Test the voltage of PIN5, which should be +1.25V;
Test the voltage of PIN2, which should be around 4.25V;
Test PIN14, and there should be driving pulses.
If the test shows the grid electrode of Q5 is 12V, check the +5V output and whether over-current or
short circuit occurs.
The Maintenance Points of PHV CW Circuit Unit
Make sure the +5V and -5V outputs are normal, and PHV_CW is high level before testing;
The work flow of troubleshooting for U13 and its peripheral circuit is shown as follows:
Test the voltage of PIN12, which should be +5V;
Test the voltage of PIN6, which should be +1.21V;
If the output voltage is not consistent with the controlling signal, check R66, R203, R260, R150 and
U2 related circuits.
3-13
Principle Description
The work flow of troubleshooting for U31 and its peripheral circuits are shown as follows:
Test the voltage of PIN3, which should be -5V;
Test the voltage of PIN2, which should be -2.37V;
If the output voltage is not consistent with the controlling signal, check R67, R204, R198 and U3
related circuits.
3.2.1.7
+5V Circuit Unit
Basic Principle of +5V Circuit
The input of +5V circuit is VBUS+, and stable +5V output is generated through BUCK
transformation step-down. The controlling IC is LM5642 from NS, which is a biphase step-down
synchronously rectifying controlling IC.
+5V circuit is equipped with overvoltage and over-current protection and is locked during protection.
The Maintenance Points of +5V Circuit
Check by sight whether LM5642 and its peripheral devices are well welded before power on, and
check whether the MOSFETs Q16~Q19 and diodes D3 and D6 are broken down using a
multimeter.
Before testing, make sure that the STRAT signal is free or of low level, and the input voltage VBUS+
is normal.
The work flow of maintenance after power on is shown as follows:
Test the voltage of PIN22, which should be VBUS+;
Test the voltage of PIN7, PIN19 and PIN25, which should be around +5V;
Test PIN9 and PIN10, which should be over than 1V;
Test PIN17, PIN20, PIN23 and PIN26, and there should be driving pulses at the frequency of
around 200kHz;
If there is no output with the circuit, check whether the output is over-current and check the
feedback circuit.
3.2.1.8
+1.5V Circuit Unit
Basic Principle of +1.5V Circuit
+1.5V output adopts the synchronously rectifying step-down chip EL7566, which is frequently
adopted by our company. This chip is integrated with over-current and over-temperature protection
internally. The over-current protection protects from hiccup.
3-14
Principle Description
The Maintenance Points of +1.5V Circuit Unit
Check by sight whether EL7566 and its peripheral devices are well welded before power on. The
work flow after power on is:
Test the voltage of PIN19~PIN21, which should be +5V;
Test the voltage of PIN22, which should be +5V;
Test PIN27, which should have oscillatory wave output;
Test PIN8~PIN13, which should have pulse voltage waveform;
If there is no output with the circuit or the if circuit is hiccupping, check if the output is over-current
and check the feedback circuit.
3.2.1.9
+3.6V/3.3V/2.8V/1.8V Circuit Unit
Basic Principle of +3.6V/3.3V/2.8V/1.8V Circuit
The output adopts the synchronously rectifying step-down chip LM2854 of NS Corporation. This
chip is integrated with MOSFET and over-current protection internally. The over-current protection
protects from hiccup.
The Maintenance Points of +3.6V/3.3V/2.8V/1.8V Circuit Unit
Check by sight whether LM2854 and its peripheral devices are well welded before power on. The
work flow after power on is:
Test the voltage of PIN5~PIN7 and PIN10~11, which should be +5V;
Test PIN12~PIN13, which should have pulse voltage waveform;
If there is no output with the circuit or if the circuit is hiccupping, check if the output is over-current
and check the feedback circuit.
3.2.1.10 -5V and -12V Circuit Unit
Basic Principle of -5V and -12V
The implementations of -5V and -12V are the same. The negative voltage IC MAX1847 is adopted
for both circuits. The -5V and -12V are both equipped with overvoltage and over-current protection
circuits, so once either overvoltage or over-current occurs, both circuits will be cut off and locked.
The Maintenance Points of -5V and -12V Circuits
Make sure that VBUS+ for -5V and -12V is within the normal range.
After power on, refer to the work flow shown as follows:
3-15
Principle Description
Test the voltage of PIN15 of the power supply chip U15~16, which should be the same as VBUS+;
Test the voltage of PIN5 (the reference voltage pin), which should be 1.25V;
Test PIN8, the overvoltage and over-current protection controlling pin, which should be higher than
2V;
Test PIN14 (drive pin), which should output drive pulse.
If there is no output with the circuit, check if the output is over-current and check the feedback
circuit.
3.2.1.11 Charging Unit
Basic Principle of Charging Unit
The rise-and-fall charging chip MAX1870A made by MAXIM is adopted for both charging units A
and B.
The MAX1870A enabling is controlled by the system’s main board and batteries’ temperature, and
eventually controlled by level of PIN15. When PIN15 is high level (about 5V), the charging is
enabled; when PIN15 is low level, the charging is disenabled. The charging current is affected by
the status of the batteries and the load current of the main unit. The charging circuit regulates the
charging current dynamically according to the load current of the main unit to ensure fast charging
and the load current of the adapter is not overloaded when the system is on.
The Maintenance Points of Charging Unit
Before testing, ensure that the charging chip is enabled, that is, PIN15 is high level.
The maintenance work flow of charging IC U7 and U8 and their peripheral circuits.
Check by sight if the soldering position departs from normal position, if so, it will cause short circuit
of pins.
Test power supplying voltage PIN32, which should be equal to adapter’s output voltage of
EDC_POWER.
Test reference end’s PIN2, which should be about 4.0V.
Test linear stabilizing output PIN1, which should be 5.4V.
Test the voltage difference between PIN28 and PIN30, which should be less than 90mV.
Test driving pin PIN27 and PIN22, and at least one PIN should have driving pulse output.
Test if there is any damage to the power device MOS semiconductor and diode.
±95V circuit.
3-16
Principle Description
Basic Principle of ±95V Circuit
±95V is generated from +12V through the non-isolated reverse converter, the chip is UC3843B of
ONSEMI.±95V output is figured with overvoltage, over-current and short circuit protection, it will not
be locked in the case of overvoltage, and be locked during over-current and short circuit protection.
The Maintenance Points of ±95V Circuit
First check if +12V output voltage is within the normal range, and if START signal is low level.
After power on, refer to the work flow shown as follows:
Test PIN8 of U1, the voltage should be +5V;
Test PIN4 of U1, there should have triangular wave of about 130 kHz;
Test PIN4 of U1, there should have drive pulse output;
Test PIN3 of U1, there should have triangular wave with the amplitude less than 1V.
If there is no output with the circuit, check if the output is over-current, check the feedback circuit,
meanwhile, check if PHV_OCP is high level.
3.2.2
Principle of the Main Board
The schematic diagram of the main board is shown in figure below.
3-17
Principle Description
Fig 3-4 Principle Diagram of Main Board
As shown in figure above, the main board unit is mainly configured with the following functional
modules:
Transducer sockets;
Ultrasound receiving;
Ultrasound transmission;
Beam forming;
System clock;
Signal processing;
System monitoring and managing;
Audio;
3-18
Principle Description
Video;
IO ports;
Where, the transducer sockets, ultrasound transmission and receiving, beam forming and system
clock are categorized as analog power region. The signal processing, signal monitoring and
managing, audio, video and I/O ports are categorized as digital power region.
The main board provides the interfaces for:
Transducer board;
Transmission board;
CW board;
4D board;
Keyboard board;
Flexible board for connecting the batteries;
CPU module;
LVDS display ports;
Power supply module;
Hard disk;
Reserved power interfaces;
Reserved signal interfaces;
Fan;
Speaker.
External interfaces include:
S_Video;
Network;
USB (2);
Extend interface.
The layout of the above-mentioned module interfaces in the main board is shown in figure below.
3-19
Inlet
fan
SVIDEO
Net port
Dual
USB
port
Adapter
port
)
Notebook PC HDD
3-20
Battery (11.1V,4400mAH)
Power module
Digital power socket
CPU module
Display port
Extension port
Power
connec
t
socket
Analog
power
socket
Battery (11.1V,4400mAH)
4D boardSignal process
Main
board
Fan
CW board
Transmitting
Receiving
Probe board
Probe port
Inlet
fan
Pencil
probe
port
Principle Description
Fig 3-5 Layout of Each Module and Socket on the Main Board
Principle Description
3.2.2.1
Power Distribution
The power distribution of M7 is an enormous network, refer to the following two figures for the
detailed distribution.
PHV1P
Filter
Transmitting: transmitting, +high volt
PHV1N
Filter
Transmitting: transmitting, - high volt
PHV2P
Filter
Transmitting: transmitting, + high volt
PHV2N
Filter
Transmitting: transmitting, - high volt
+95V (10mA)
Filter
-95V (10mA)
Filter
+12V (400mA)
Filter
-12V (400mA)
Filter
A+95V
Receiving: high volt switch, + high volt
A+95V
Probe port: high volt switch, + high volt
A-95V
Receiving: high volt switch, - high volt
A-95V
Probe port: high volt switch, - high volt
A+12V
Transmitting: transmitting drive volt
A+12V
CW: pencil transmitting drive volt
A-12V
Transmitting: transmitting drive volt
Receiving: high volt isolation power
Transmitting: transmitting gate drive volt
Receiving: +5V_VCA
Volt
IN
+5V1 (3260mA)
CW: CW filter power
Filter
Receiving: ATGC power
Receiving: channel select logic devices power supply
Probe port: extension module, 192-elemented probe power supply, etc. (control power off
and short circuit)
Receiving: high volt isolation power
-5V1 (550mA)
Filter
Transmitting: transmitting gate drive volt
CW: CW filter power
Receiving: VCA power 3V3
Receiving: ADS power 3V3
+3V6 (2100mA)
Filter
BF: logic IO power 3V3
Clock module: power
Probe port: probe flash power
BF: LVDS power
+2V8 (1000mA)
Filter
Transmitting: transmitting gate drive volt
Receiving: ADS power 1V8
+1V5 (3000mA)
BF: FPGA kernel
Filter
Transmitting: FPGA kernel
Fig 3-6 Front-end Power Tree of the Main Board
3-21
Principle Description
Fig 3-7 Back-end Power Tree of the Main Board
3-22
Principle Description
3.2.2.2
Ultrasound Receiving
The basic structure of the front-end circuit is shown in figure below:
Clock circuit
Clock
allocator
64-channel 5-level transmitting circuit
(MOS driver + high-voltage MOS)
Transducer
POUT[1:128]
2 to 1 channel switching
circuit
128<->64 channel
switching
(high-voltage switch
16pcs)
TX[1:64]
High-voltage
transmission
module
DSP FPGA
Transmission
controlling
bus
PCIE bus
Beam forming
data bus
40MHz*8ch
LVDS
64-channel receiving circuit
RX[1:64]
ATGC
High-voltage
isolation
Beam forming
controlling bus
160MHz
LVDS
Ultrasou
nd backend
AD data
BF FPGA
AFE chip
CW circuit
160MHz crystal
oscillator
40MHz
LVDS
160MHz
LVDS
Transmission
FPGA
160MHz
LVPECL
CW
data
Fig 3-8 Diagram of Front-end Circuit
The receiving circuit of the main board consists of channel switching and receiving circuits.
Channel switching circuit
Transmitting and receiving circuits are 64 channels, but the actual transducer is 128 channels, so
channel switching circuit is needed to switch the 64 channels of the transmitting and receiving into
128 channels. 8 channel conversion circuit is completed through 16-channel high voltage switch
chip. Refer to the principle diagram for the principle.
Receiving circuit
The receiving circuit consists of high-voltage isolation circuit, voltage-controlled gain amplification
circuit and high-speed ADC.
The high-voltage isolation circuit protects the receiving amplification circuit during transmission from
high-voltage transmission pulse damage.
Voltage-controlled gain amplification circuit amplifies weak ultrasound echoes and transmits them
to ADC for analog-to-digital conversion. It converts ultrasound echoes to digital signals for
succeeding digital processing.
Voltage-controlled amplification circuit and high-speed ADC are all integrated in one AFE chip in
M7.An AFE chip consists of 8-channel voltage-controlled gain amplification circuit and high-speed
ADC. There are 8 AFEs in M7, composing 64 receiving channels.
3.2.2.3
Clock Circuit
The clock circuit provides low phase noise clock to the above-mentioned modules.
3-23
Principle Description
The clock circuit consists of crystal oscillator and clock allocator. The crystal oscillator is a 160Mhz
oscillator with good phase noises. It distributes the clock signals via the clock allocator. The clock
allocator has 12 LVDS outputs, and each one can split the frequency independently.
The outputs of the clock circuit are shown in figure below:
Fig 3-9 Diagram of Clock Circuit
3.2.2.4
Signal Processing
The signal processing is mainly implemented through two FPGAs: FPGA1 and FPGA2.
The FPGA1 mainly responsible for controlling on the front-end chips (such as channel selection
CPLD, VGA gain control DAC, ADC, transducer and transducer extend module), beam forming, and
orthogonal demodulation, generating transmission sequence pulse, etc.
The FPGA2 implements signal processing, data uploading, real time scan parameters loading, 4D
controlling, etc. Data is uploaded through PCIE interface.
3.2.2.5
Power Management
Power management is basically implemented by the power management FPGA.
System power can be sorted as STANDBY power, CPU STANDBY power and normal working
power in terms of power-on sequence.
STANDBY power is provided in the status of system POWERDOWN, including 3V3STB and
1V5_FPGA.These power supplies are mainly for powering the power managing FPGA devices for
power management. The power of power managing FPGA reset signal, 10k crystal oscillator, power
managing FPGA, CPU12V_OK and CPU5VSTB_OK, etc. As long as the adapter is connected or
batteries are loaded, they will be powered.
CPU STANDBY power is required in the POWER DOWN status of PC module, including 5VSBY
and 3V3_STB_CPU.In the system is POWER DOWN, this power is turned off. Turn on this power
before starting the PC system.
3-24
Principle Description
The working power is provided when the system is working normally. Except the two types of
powers mentioned above, other powers all belong to working powers. The following indicators are
used for the working powers:
Table 3-7 Power indicators on main board
No.
LED Number
Meaning
1
D227
12V
2
D18
VCC
3
D17
VDD
There are some indicators on the control panel for indicating power status. The signals are
described in the following table:
Table 3-8 Indicators on the control panel
No.
Indicator
Drive signal
Definition
1
Work Status
WORKSTS_O
Dual-color, indicating the current status
LED
of the main unit:
It remains off when the system is turned
off.
WORKSTS_G
Press POWERBTN, the light flashes in
green when S3# is disenabled.
When S3# is disenabled after system
power on, the light turns into green.
2
EDC Status
EDCSTS_G
LED
Single color, indicating if the main unit
is connected with the mains power:
When disconnected, the light is off,
otherwise, in green.
3
Battery
Status LED
BATSTS_O
Dual-color, indicating the current
battery status:
It illuminates in orange color when
battery is charging;
BATSTS_G
It illuminates in green color when
battery is charging to full capacity;
It flashes in orange color if battery
3-25
Notes
Principle Description
No.
Indicator
Drive signal
Definition
Notes
capacity is less than 30%, the battery is
discharging, and the system is working;
It flashes in green color if battery
capacity is more than 30%, the battery
is discharging, and the system is at
standby status;
It flashes in green and orange
alternately if battery capacity is less
than 30%, the battery is discharging,
and the system is at standby status;
The light flashes in orange at the cycle
of 0.5S, that is to say, 0.25S on, and
0.25S off.
The light flashes in green at the cycle of
1S, that is to say, 0.5S on, and 0.5S off.
When the light flashes in in green and
orange alternately, the green light is
0.25S on and 0.25S off; the orange light
is 0.25S on and 0.25S off.
3.2.2.6
System Monitoring
The system monitoring includes voltage monitoring, temperature monitoring and fan rotating
control.
Monitor on the voltage: direct measurement, including measurement of 12V, N12V, VCC, VDD, 1.5V,
2.8V, 1.8V and 1.2V;
It detects power board and front-end temperature.
Monitors and controls on fans and ventilation unit.
3.2.2.7
Video Processing
There are two versions for M7 video processing:
The first version:
On the PCBA of the main board 051-000098-00 VA, 051-000098-01 V1-V4, the following method
is adopted:
3-26
Principle Description
Fig 3-10 Block diagram of video module (the first version)
Of which, the video signals of LVDS, VGA, and DVI are all from LVDS of PC module.
Both S-VIDEO and VIDEO are generated by SVIDEO of PC module.
The display switch is on the control panel.
The display parameters are saved in EEPROM of LCD parameter board, they are accessed by
SMBUS of the PC module.
The second version:
On the PCBA of the main board 051-000098-02 V1, the following method is adopted:
External
CRT
VGA
1024x768@60Hz
SVideo
COME
Module
CEGM4
5
Frame Buffer
MT47H64M16HR
S-Video
CVBS
SDVO
B
SDVO
C
LVDSA
LVDSB
1024x768@60Hz
LPC
33MHz
MF
FPGA
EP3C16
I2C
0-400KHz
RGB888
1024x768@60Hz
Video
Encoder
CH7026B
DVI Transmitter
TFP410
PAL
NTSC
S-Video
PAL
NTSC
CVBS
DVI
1024x768@60Hz
External
Display
LVDS
1024x768@60Hz
Main LCD
SSC
CY25562
Fig 3-11 Block diagram of video module (the second version)
External display video signal is generated from VGA of PC module.
S-Video, CVBS, DVI, and LVDS video signals are all generated from LVDS of PC module.
The display switch is on the control panel.
3-27
Principle Description
The display parameters are saved in EEPROM of LCD parameter board, they are accessed by
SMBUS of the PC module.
3.2.2.8
Interfaces of Main Board
Interfaces of Main Board and Transducer Board
Signals of the main board and transducer board are defined in table below.
Table 3-9 Interfaces of the main board and transducer board
CON1
CON2
PI
PIN
PIN
NAME
N
NAME
PIN
PIN NAME
PIN
PIN NAME
PIN
PIN NAME
PIN
PIN
1
GND
2
POUT1
3
GND
1
POUT69
2
POUT70
3
POUT71
4
POUT2
5
GND
6
POUT3
4
POUT72
5
POUT73
6
POUT74
7
GND
8
POUT4
9
POUT5
7
POUT75
8
GND
9
POUT76
10
POUT6
11
POUT7
12
GND
10
GND
11
POUT77
12
POUT78
13
POUT8
14
GND
15
POUT9
13
POUT79
14
POUT80
15
GND
16
GND
17
POUT10
18
POUT11
16
POUT81
17
GND
18
POUT82
19
POUT12
20
POUT13
21
GND
19
GND
20
POUT83
21
POUT84
22
POUT14
23
GND
24
POUT15
22
POUT85
23
POUT86
24
GND
25
GND
26
POUT16
27
POUT17
25
POUT87
26
GND
27
POUT88
28
POUT18
29
POUT19
30
GND
28
GND
29
POUT89
30
POUT90
31
POUT20
32
GND
33
POUT21
31
POUT91
32
POUT92
33
GND
34
GND
35
POUT22
36
POUT23
34
POUT93
35
GND
36
POUT94
37
POUT24
38
POUT25
39
GND
37
GND
38
POUT95
39
POUT96
40
POUT26
41
GND
42
POUT27
40
POUT97
41
POUT98
42
GND
43
GND
44
POUT28
45
POUT29
43
POUT99
44
GND
45
POUT100
46
POUT30
47
POUT31
48
GND
46
POUT101
47
POUT102
48
POUT103
49
POUT32
50
GND
51
POUT33
49
POUT104
50
POUT105
51
GND
52
GND
53
POUT34
54
POUT35
52
POUT106
53
GND
54
POUT107
55
POUT36
56
POUT37
57
GND
55
GND
56
POUT108
57
POUT109
58
POUT38
59
GND
60
POUT39
58
POUT110
59
POUT111
60
POUT112
61
GND
62
POUT40
63
POUT41
61
POUT113
62
GND
63
POUT114
3-28
NAME
PIN
Principle Description
CON1
CON2
64
POUT42
65
POUT43
66
GND
64
POUT115
65
POUT116
66
POUT117
67
POUT44
68
GND
69
POUT45
67
POUT118
68
POUT119
69
GND
70
GND
71
POUT46
72
POUT47
70
POUT120
71
GND
72
POUT121
73
POUT48
74
POUT49
75
GND
73
GND
74
POUT122
75
POUT123
76
POUT50
77
GND
78
POUT51
76
POUT124
77
POUT125
78
POUT126
79
GND
80
POUT52
81
POUT53
79
POUT127
80
GND
81
POUT128
POUT54
82
POUT55
83
GND
87
GND
POUT58
89
92
POUT64
98
POUT67
101
GND
103
POUT65
99
POUT66
100
94
GND
FLASH_
FREEZE_
86
POUT68
UG
ROBE
CLK_192
92
_PROBE
2_PROBE
ROBE
DIN1_192
93
LE_N_19
95
N
A+85V_P
90
_PROBE
A-85V_PR
96
OBE
EXPRESS_
PRB_DP_
PROBE_P
PRESENT_
SWITCH_
RESENT_
N
98
N
99
N
GND
10
GND
101
E
104
2
A+5V_PR
10
A+5V_PR
OBE
5
OBE
The signals are described in the following table:
Table 3-10 Interface & signals of main board and transducer board
SIGNAL NAME
DESCRIPTION
POUT[1:128]
Transmitting signals and receiving echo signals
SPIDIN_PROBE
Probe Flash SPI Data IN
SPIDOUT_PROBE
Probe Flash SPI Data OUT
SPICS_N_PROBE
Probe Flash SPI CS_N
SPICLK_PROBE
Probe Flash SPI CLOCK
3-29
87
GND
A+5V_PROB
103
WP_N
89
100
105
PIDIN
DOUT
GND
102
GND
104
97
84
PRB_DP_
DIN2_192_P
96
GND
97
91
SPICS_N
PROBE_
PROBE_PL
POUT63
95
83
PROBE_S
PROBE_SPI
88
GND
GND
PROBE_
A+3V3_PRB
93
POUT62
94
POUT59
POUT61
CLK
85
90
POUT60
91
82
POUT57
86
88
PROBE_SPI
84
POUT56
85
GND
Principle Description
SIGNAL NAME
DESCRIPTION
FLASH_WP_N_PROBE
Probe Flash Write Protect
FLASHSPIOE_N_PROB
Probe Flash SPI enable
E
CTRL_A+3V3_PROBE_
Enable Flash power of transducer
N
CTRL_A+5V_PROBE
Enable transducer extend power
PRB_PRESENT
Transducer present signal
EXP_PRESENT
Transducer extend module present signal
192_CLK_PROBE
High-voltage switch clock of 192-elemented transducer
192_LE_N_PROBE
Enabling signal of 192-elemented transducer high-voltage switch
192_DIN1_PROBE
High-voltage switch data 1 of 192-elemented transducer
192_DIN2_PROBE
High-voltage switch data 2 of 192-elemented transducer
192_SPIOE_N_PROBE
Enabling SPI of 192-elemented transducer
CTRL_192HV _PROBE
Control high-voltage output of 192 transducer
PRB_DP_FRZ
FREEZE signal of real-time bi-planar transducer
PRB_DP_SW
Switching signal of real-time bi-planar transducer
PRB_PLUG_N
Detection of connection and disconnection with power on of the
transducer
Interfaces of Main Board and Transmission Board
Signals of the main board and transmission board are defined in table below.
Table 3-11 Interfaces of main board and transmission board
CON1(J6)
PIN
SIG
PIN
SIG
PIN
SIG
PIN
SIG
1
GND
2
GND
61
EMIT33
62
GND
3
EMIT1
4
EMIT2
63
GND
64
EMIT36
5
EMIT3
6
EMIT4
65
EMIT35
66
GND
3-30
Principle Description
7
GND
8
GND
67
GND
68
EMIT38
9
EMIT5
10
EMIT6
69
EMIT37
70
GND
11
EMIT7
12
EMIT8
71
GND
72
EMIT40
13
GND
14
GND
73
EMIT39
74
GND
15
EMIT9
16
EMIT10
75
GND
76
EMIT42
17
EMIT11
18
EMIT12
77
EMIT41
78
GND
19
GND
20
GND
79
GND
80
EMIT44
21
EMIT13
22
EMIT14
81
EMIT43
82
GND
23
GND
24
EMIT16
83
GND
84
EMIT46
25
EMIT15
26
GND
85
EMIT45
86
GND
27
GND
28
EMIT18
87
GND
88
EMIT48
29
EMIT17
30
GND
89
EMIT47
90
GND
31
GND
32
EMIT20
91
GND
92
EMIT50
33
EMIT19
34
GND
93
EMIT49
94
GND
35
GND
36
EMIT22
95
GND
96
EMIT52
37
EMIT21
38
GND
97
EMIT51
98
GND
39
GND
40
EMIT24
99
GND
100
EMIT54
41
EMIT23
42
GND
101
EMIT53
102
GND
43
GND
44
EMIT26
103
GND
104
EMIT56
45
EMIT25
46
GND
105
EMIT55
106
GND
47
GND
48
EMIT28
107
EMIT57
108
EMIT58
49
EMIT27
50
GND
109
GND
110
EMIT60
51
GND
52
EMIT30
111
EMIT59
112
GND
53
EMIT29
54
GND
113
EMIT61
114
EMIT62
55
GND
56
EMIT32
115
GND
116
EMIT64
57
EMIT31
58
GND
117
EMIT63
118
GND
59
GND
60
EMIT34
119
GND
120
GND
3-31
Principle Description
CON2(J5)
PIN
SIG
PIN
SIG
PIN
SIG
PIN
SIG
1
PHV1_N
2
PHV1_P
31
GND
32
NC
PHV1_N
3
PHV1_P
4
NC
5
7
NC
8
PHV2_N
9
10
11
15
14
20
_EMIT
24
26
28
IT
43
GND
HSSB_TX0RN
HSSB_TX0RP
A+1V5
49
51
HSSB_TX1RP
T
53
GND
N_EMIT
55
HSSB_TX0TN
DO_BF
57
TCK
30
GND
HSSB_TX1RN
TDI_EMIT_T
TDO_EM
29
CKP
CLOCKEN_
TMS
27
41
RESET_EMI
RATE_N
25
HSSB_TX0RB
A+1V5
RATE_P
CKN
47
22
_EMIT
HSSB_TX0RB
A+5V1
A+2V8
21
GND
45
18
A+2V8
37
A+12V
A-5V1
23
A+12V
16
17
CKP
39
12
A-12V
35
NC
A-12V
19
PHV2_P
CKN
HSSB_TX0RF
PHV2_P
NC
13
33
6
PHV2_N
HSSB_TX0RF
HSSB_TX0TP
59
3-32
34
36
38
40
42
44
46
48
50
52
54
56
58
60
A+75V_sub
NC
GND
CLK_N_EMIT
CLK_P_EMIT
GND
HSSB_TX0TFCK
N
HSSB_TX0TFCK
P
GND
HSSB_TX0TBCK
N
HSSB_TX0TBCK
P
GND
HSSB_TX1TN
HSSB_TX1TP
Principle Description
Interface of Main Board and CW Board
Signals between the main board and CW board interfaces are defined as Table 3-12 and Table
3-14 .
Table 3-12 Definition of Signals of CW CON1
PIN
SIGNAL
PIN
SIGNAL
1
GND
2
GND
3
CW1
4
CW2
5
GND
6
GND
7
CW3
8
CW4
9
GND
10
GND
11
CW5
12
CW6
13
GND
14
GND
15
CW7
16
CW8
17
GND
18
GND
19
CW9
20
CW10
21
GND
22
GND
23
CW11
24
CW12
25
GND
26
GND
27
CW13
28
CW14
29
GND
30
GND
31
CW15
32
CW16
33
GND
34
GND
35
CW17
36
CW18
37
GND
38
GND
39
CW19
40
CW20
3-33
Principle Description
Wherein the signals of CW CON1 are defined as follows:
Table 3-13 Signals Definition of CW CON1 Socket
Signal name
Description
CW[20:1]
CW echo signal
Table 3-14 Signals Definition of CW CON2 Socket
1
A-5V1
2
A+5V1
21
TPU1_PENCIL_CW
22
AD_SDI0
3
A-5V1
4
A+5V1
23
GND
24
GND
5
GND
6
GND
25
TPU0_PENCIL_CW
26
AD_SCK0
7
BORADID_C
8
PPID_CW
GND
28
GND
10
AD8333_EBL_CW
GND
30
GND
12
GND
31
AD8333_4LO0_CW
32
AD_SDO1
14
SW_CTRL1
33
AD8333_4LO1_CW
34
AD_SDI1
16
SW_CTRL0
35
GND
36
GND
18
GND
37
VOL_ADJ_SDA_CW
38
AD_SCK1
20
AD_SDO0
39
A+12V
40
W
9
CWPPPRESE
NT
11
13
15
17
19
27
29
Interfaces of Main Board and 4D Board
The interfaces between the main board and the 4D board are defined in table below.
Table 3-15 Signal Definition of main board and 4D board
PIN
Signal name
PIN
Signal name
1
12V
2
12V
3
12V
4
12V
5
12V
6
12V
3-34
AD_CNV1_
CW
GND
AD_CNV0_
CW
GND
VOL_ADJ_
SCK_CW
A+3V6
Principle Description
PIN
Signal name
PIN
Signal name
7
12V
8
12V
9
GND
10
GND
11
GND
12
GND
13
GND
14
GND
15
GND
16
GND
17
-12V
18
-12V
19
-12V
20
-12V
21
-12V
22
-12V
23
GND
24
GND
25
GND
26
GND
27
GND
28
GND
29
VDD
30
VDD
31
GND
32
GND
33
VCC
34
VCC
35
GND
36
GND
37
NC
38
NC
39
NC
40
NC
41
NC
42
NC
43
NC
44
NC
45
NC
46
NC
47
NC
48
NC
49
GND
50
GND
51
POWEREN_N
52
GND
53
HALL
54
GND
55
NC
56
GND
57
SPI_CLK
58
GND
3-35
Principle Description
PIN
Signal name
PIN
Signal name
59
SPI_DAT
60
GND
61
SPI_SYNC
62
GND
63
FD_ID
64
GND
65
AD_SCLK
66
GND
67
AD_DIN
68
GND
69
AD_DOUT
70
AD_NCS
Interfaces of Main Board and Display Module
There are 3 sockets in all for the main board and display module, refer to the following 3 tables for
the definitions:
Table 3-16 Definition of signals of the main board end (J25) of the LVDS socket
NO.
NAME
NO.
NAME
1
VDD
2
VDD
3
GND
4
GND
5
RXE0-
6
RXE1-
7
RXE0+
8
RXE1+
9
GND
10
GND
11
RXE2-
12
RXEC-
13
RXE2+
14
RXEC+
15
GND
16
GND
17
RXE3-
18
GND
19
RXE3+
20
GND
Table 3-17 Socket definition of display power supply (J26)
No:
Name
1
+12V
2
+12V
3
+12V
3-36
Principle Description
4
BL_On/Off
5
BL_ADJ
6
GND
7
GND
8
GND
Table 3-18 Definition of Color Temperature Correction Socket (J24)
PIN
Signal name
1
VDD
2
GND
3
SCL
4
SDA
Interfaces of Main Board and Speakers
The interfaces of the main board and speakers are defined in the following table:
Table 3-19 Interfaces of Main Board and Speakers (J27)
PIN
SIG
1
ROUTP
2
ROUTN
3
NC
4
LOUTP
5
LOUTN
Socket between Fan and Main Board
The interfaces of the main board and fan are defined in the following table:
Table 3-20 Socket Definition of Fan Power Supply (J29)
No.
Signal name
No.
Signal name
1
FANLF_SP
2
FANLF_PWR
3
FANLFM_CTRL
4
FANLMC_PWR
5
FANLM_SP
6
FANC_CTRL
3-37
Principle Description
7
FANLB_SP
8
FANC_SP
9
FANLBR_CTRL
10
FANLB_PWR
11
FANR_SP
12
FANR_PWR
FPGA JTAG Sockets of Main Board Power Management
The FPGA JTAG sockets of the main board are defined in the following table:
Table 3-21 Signal Definition of FPGA JTAG sockets (J23) of Power Management
No.
Signal name
1
VPUMP
2
VJTAG
3
GND
4
TDO
5
TDI
6
TCK
7
TMS
8
TRST
Reserved Signal Interfaces
The reserved signal interfaces are defined in the following table:
Table 3-22 Definition of Reserved Signal Interfaces (J17)
No.
Signal name
No.
Signal name
1
NC
2
GND
3
NC
4
GND
5
GND
6
GND
7
NC
8
ATA_ACT#
9
SYS_RESET#
10
GND
11
CPRT_PRINT
12
BK_STS
13
CPRT_BUSY
14
GND
3-38
Principle Description
15
GND
16
SVIDEO_Y2
17
PWR_SSW
18
SVIDEO_C2
19
GND
20
GND
Reserved Power Interfaces;
The reserved power interfaces are defined in the following table:
Table 3-23 Definition of Reserved Power Interfaces (J20)
No.
Name
No.
Name
1
+12V
2
+12V
3
+12V
4
+12V
5
Gnd
6
Gnd
7
Gnd
8
Gnd
9
Gnd
10
Gnd
11
+5V
12
+5V
13
+5V
14
+5V
15
+5V
16
+5V
17
+5V
18
+5V
19
Gnd
20
Gnd
21
Gnd
22
Gnd
23
+3.3V
24
+3.3V
The Interface between the Main Board and Adapter
The interfaces between the main board and the adapter are defined as follows:
Table 3-24 Interface (J21) of Main Board and Speakers
No.
Name
Direction
1
Gnd
\
6
Gnd
\
3-39
Principle Description
No.
Name
Direction
2
Gnd
\
4
Gnd
\
3
+12V
In
7
+12V
In
5
+12V
In
8
+12V
In
The Interface between the Main Board and Power Module
The interfaces between the main board and the power module consist of digital socket and analog
power socket.
The digital socket is defined in the following table:
Table 3-25 Socket Definition of igital Socket (J22)
No.
Signal name
No.
Signal name
1
EDC_PWR
2
EDC_PWR
3
EDC_PWR
4
EDC_PWR
5
EDC_PWR
6
EDC_PWR
7
EDC_GND
8
EDC_GND
9
EDC_GND
10
EDC_GND
11
EDC_GND
13
-12V
14
STS_EDC#
15
GND
16
GND
17
POWER_ON#
18
CPU12V_OK#
19
T_D+
20
T_D-
21
GND
22
GND
23
GND
24
GND
25
12V+
26
12V+
27
12V+
28
12V+
EDC_GND
3-40
Principle Description
No.
Signal name
No.
Signal name
29
GND
30
GND
31
5V+
32
5V+
33
GND
34
GND
35
5VSTB_OK#
36
5VSTB_EN#
37
GND
38
GND
39
5VSTB_CPU
40
3V3STB
41
GND
42
GND
43
3V3
44
3V3
45
GND
46
GND
47
2V8
48
1V5
49
GND
50
GND
51
1V8
52
1V8
53
GND
54
GND
55
PWR_RX
56
PWR_CONF_N
57
PWR_TX
58
PWR_RST
59
GND
60
GND
61
PRT_BATA
62
PRT_BATB
63
BATSTS_O
64
BATSTS_G
Table 3-26 Signal Definition of Socket (J9) between Analog Power and Main Board
No.
Signal name
No.
Signal name
1
GND
2
GND
3
A5V1
4
A5V1
5
A5V1
6
A5V1
7
GND
8
GND
9
-A5V1
10
GND
11
-A12V
12
GND
3-41
Principle Description
No.
Signal name
No.
Signal name
13
GND
14
GND
15
A1V5
16
A1V5
17
A1V5
18
A1V5
19
GND
20
GND
21
A3V6
22
A3V6
23
GND
24
GND
25
A2V8
26
GND
27
A12V
28
GND
29
PHV1+
30
PHV2+
31
GND
32
GND
33
PHV1-
34
PHV2-
35
GND
36
GND
37
85V+
38
GND
39
GND
40
85V-
Interfaces of Main Board and I/O Extend Board
Signal definition between master board and I/O extended board is shown as table below.
Table 3-27 Signal Definition between Mater Board and I/O Extended Socket-J1
PIN
Definition of
PIN
Signal
Definition of
PIN
Signal
Definition of
PIN
Signal
Definition of
Signal
A1
USB+4
B1
USB-4
C1
USB+5
D1
USB-5
A2
VBUS_SYS45
B2
GND
C2
VBUS_SYS45
D2
GND
A3
GND
B3
VGA_HS
C3
VGA_RED
D3
VGA_GREEN
A4
UART_RX0
B4
VGA_VS
C4
GND
D4
SDISP_PRT_N
A5
UART_TX0
B5
GND
C5
VGA_BLUE
D5
A6
GND
B6
AUDIO_L
C6
GND
D6
A7
BWPRT _PRINT
B7
AUDIO_R
C7
3-42
D7
GND
Principle Description
PIN
Definition of
PIN
Signal
PIN
Definition of
Signal
Definition of
PIN
Signal
Definition of
Signal
A8
BWPRT_BUSY
B8
GND
C8
DVI_D0+
D8
DVI_D0-
A9
GND
B9
ECG_PWR
C9
DVI_D1+
D9
DVI_D1-
A10
MIC_L
B10
DVI_HPD
C10
GND
D10
GND
A11
MIC_R
B11
C11
DVI_CLK+
D11
DVI_CLK-
A12
VIDEO
B12
ECG_TX
C12
DVI_D2+
D12
DVI_D2-
\
\
B13
ECG_RX
\
\
D13
ECG_PRT
The Interfaces between the Main Board and the Keyboard
The interfaces between the main board and the keyboard are defined in table below.
Table 3-28 Signal Definition between Master Board and Keyboard Board-J19
No.
Name
No.
Name
1
+3.3V
2
+3.3V
3
+3.3V
4
+3.3V
5
Gnd
6
Gnd
7
Gnd
8
Gnd
9
+5V
10
+5V
11
+5V
12
+5V
13
Gnd
14
Gnd
15
Gnd
16
Audio Out
17
Gnd
18
Gnd
19
Work_Status_O
20
Work_Status_G
21
Battery_Status_O
22
Battery_Status_G
23
Gnd
24
EDC_Status_G
25
USB_DN
26
LCD_SW
27
USB_DP
28
Power_BTN
29
Gnd
30
Gnd
3-43
Principle Description
3.2.2.9
Testing Points of Main Board and LED Indicator Light
The testing points of the main board are shown in the following figures:
Fig 3-12 Testing Points of Main Board
Table 3-29 List of Testing Points of Main Board
No.
Label of the Testing Point
Network of the Testing Point
1
TP206-GND
Ground network
2
TP248-VAA0
Voltage of video DA chip (3.3V)
3
TP228-1V5STB
Power management standby voltage (1.5V)
4
TP249-SV-VDD
Voltage of video driving chip (3.3V)
5
TP202-12V
12V voltage (12V)
6
TP246-3V3S_CPU
PC module standby (3.3V)
7
TP7-PWROK
Power OK signal of PC module (3.3V)
Initialization status indicating signal S3 of PC module
8
TP9-S3
(3.3V)
3-44
Principle Description
No.
Label of the Testing Point
Network of the Testing Point
Initialization status indicating signal S4 of PC module
9
TP8-S4
(3.3V)
Initialization status indicating signal S5 of PC module
10
TP10-S5
(3.3V)
11
TP224-EDC_PWR
12V input of the adapter (12V)
12
TP218-N12V
-12V voltage (-12V)
13
TP215-5VSBY
Standby voltage (5V)
14
TP203-VDD
VDD voltage (3.3V)
15
TP217-D1V5
D1V5 voltage (1.6V)
16
TP216-D2V8
D2V8 voltage (3.0V)
17
TP6-CPUBTN
Button start-up signal of PC module (3.3V)
18
TP214-D1V8
D1V8 voltage (1.8V)
19
TP201-VCC
VCC voltage (5V)
20
TP20-IQ_VTT
VTT voltage of DDR2 (0.9V)
21
TP236-VCCINT_DSP_MF
FPGA core voltage (1.2V)
22
TP226-12V_4D
12V voltage of 4D module (12V)
23
TP235-D2V5
D2V5 voltage (2.5V)
24
TP19-UP_VTT
VTT voltage of DDR2 (0.9V)
25
TP227-3V3STB
Standby voltage (3.3V)
26
TP146-A+5V_PROBE
Transducer port 5V
27
TP148-A+3V3_PRB
Transducer port 3.3V
28
TP147-A+5V1
Analog 5V
29
TP155-A-5V_HVISO2
High-voltage isolation power
30
TP241-A-85V_HVSW
High-voltage switch power
31
TP151-A+85V_HVSW
High-voltage switch power
32
TP153-A+5V_HVISO2
High-voltage isolation power
33
TP152-A+5V_HVISO1
High-voltage isolation power
3-45
Principle Description
No.
Label of the Testing Point
Network of the Testing Point
34
TP193-A-12V
-12V power
35
TP165-A+1V8_ADS
AFE power
36
TP194--12V_A
-12V power
37
TP163-A+3V3_VCA
AFE power
38
TP164-A+3V3_ADS
AFE power
39
TP178-A+5V_ATGC
ATGC power
40
TP162-A+5V_VCA
AFE power
41
TP144-A+3V3_CLOCK
Clock power
42
TP240-VBAT
Voltage of the button cell (3V)
43
TP190-A+1V5
Analog 1.5V power
44
TP189-+1V5_A
Analog 1.5V power
45
TP184-+5V1_A
Analog 5V power
46
TP137-A+5V1
Analog 5V power
47
TP196-+12V_A
Analog 12V power
48
TP195-A+12V
Analog 12V power
49
TP187-+2V8_A
Analog 2.8V power
50
TP185-+3V6_A
Analog 3.6V power
51
TP186-A+3V6
Analog 3.6V power
52
TP180--95V_A
Analog -95V power
53
TP188-A+2V8
Analog 2.8V power
54
TP192--5V1_A
Analog -5V power
55
TP36-+95V_A
Analog 95V power
56
TP191-A-5V1
Analog -5V power
57
TP38-A-95V
Analog -95V power
58
TP179-A-5V_ATGC
ATGC-5V power
59
TP37-A+95V
Analog 95V power
3-46
Principle Description
Table 3-30 Main board LED indicator light
No.
LED
Indication
1
D227
12V
2
D18
VCC
3
D17
VDD
Multifunctional FPGA
4
D220
configuration indicator light
DSP FPGA configuration
5
D11
indicator light
Beam FPGA configuration
6
3.2.3
D15
indicator light
Principle of Transmitting Board
The main board sends commands to transmitting board, the transmitting board calculates the
parameters in real time and controls the high-voltage pulses necessary for the output of the
transmitting circuit. The schematic diagram is shown below:
Fig 3-13 Structure Diagram of Ultrasound Transmission
3-47
Principle Description
3.2.3.1
Power Structure of Transmission Board
A+1V5
LDO
VF+1V2
100mA
A+5V1
LDO
P+3v3
100mA
A+2V8
LDO
LC
A+2V5
VCCAUX
LC
500mA
LC
VF+2V5
VLL+2V5
LC
LDO
VDD+11V
A+12V
AVDD1+11V
LC
1600mA
LC
DVDD1+11V
DVDD2+11V
A-12V
LDO
LC
VSS-11V-HIGH
AVSS-11V-HIGH
LC
360mA
DVSS-11V-HIGH
LDO
LC
VSS-11V-LOW
AVSS-11V-LOW
LC
DVSS-11V-LOW
Capacitor filter
PHV1_P
PHV1_N
PHV2_P
PHV2_N
300mA
Capacitor filter
300mA
Capacitor filter
300mA
Capacitor filter
PHV1_P
PHV1_N
PHV2_P
PHV2_N
300mA
Fig 3-14 Structure Diagram of Transmitting Board Power Supply
3.2.3.2
Definition of Transmission Board Interface
Detailed definition of external interface of transmission board is shown below:
3-48
Principle Description
Table 3-31 Detailed Definition of External Interface of Transmission Board
(DSP_FPGA)HSSB_TX0TP/N,HSSB_TX1TP/N,
HSSB_TX0TBCKP/N,HSSB_TX0TFCKP/N,HSSB_TX0RP/N,
Interface with the
Main Board
HSSB_TX1RP/N, HSSB_TX0RBCKP/,
HSSB_TX0RFCKP/NCLOCKEN_N_EMIT
(BF_FPGA) RESET_EMIT,RATE_P/N_EMIT
(clock) CLK_P/N_EMIT
Interface with analog
64-channel transmission pulse/echo POUT[EMIT64..1]
part of main board
Interface with power
A+12V,A-12V,A+5V1,A-5V1,A+2V8,
of main board
A+1V5,PHV1_P,PHV1_N,PHV2_P,PHV2_N
JTAG signal
TCK,TDO_EMIT,TMS,TDI_EMIT_TDO_BF
The instrumentation diagram of transmission board is shown in the following figure. The figure
marks the sockets and their names.
3-49
Principle Description
Fig 3-15 The instrumentation diagram of transmission board
Table 3-32 Definition of J1 of JTAG Socket
1
JTAG_TCK
2
JTAG_ TDO_EMIT
3
JTAG_ TMS
4
JTAG_ TDI_EMIT_TDO_BF
5
P+3V3
6
GND
7
NC
8
NC
Table 3-33 Detailed Definition of Bus Socket J2
1
GND
2
GND
61
EMIT33
62
GND
3
EMIT1
4
EMIT2
63
GND
64
EMIT36
5
EMIT3
6
EMIT4
65
EMIT35
66
GND
7
GND
8
GND
67
GND
68
EMIT38
9
EMIT5
10
EMIT6
69
EMIT37
70
GND
11
EMIT7
12
EMIT8
71
GND
72
EMIT40
13
GND
14
GND
73
EMIT39
74
GND
15
EMIT9
16
EMIT10
75
GND
76
EMIT42
17
EMIT11
18
EMIT12
77
EMIT41
78
GND
19
GND
20
GND
79
GND
80
EMIT44
21
EMIT13
22
EMIT14
81
EMIT43
82
GND
23
GND
24
EMIT16
83
GND
84
EMIT46
25
EMIT15
26
GND
85
EMIT45
86
GND
27
GND
28
EMIT18
87
GND
88
EMIT48
29
EMIT17
30
GND
89
EMIT47
90
GND
31
GND
32
EMIT20
91
GND
92
EMIT50
3-50
Principle Description
33
EMIT19
34
GND
93
EMIT49
94
GND
35
GND
36
EMIT22
95
GND
96
EMIT52
37
EMIT21
38
GND
97
EMIT51
98
GND
39
GND
40
EMIT24
99
GND
100
EMIT54
41
EMIT23
42
GND
101
EMIT53
102
GND
43
GND
44
EMIT26
103
GND
104
EMIT56
45
EMIT25
46
GND
105
EMIT55
106
GND
47
GND
48
EMIT28
107
EMIT57
108
EMIT58
49
EMIT27
50
GND
109
GND
110
EMIT60
51
GND
52
EMIT30
111
EMIT59
112
GND
53
EMIT29
54
GND
113
EMIT61
114
EMIT62
55
GND
56
EMIT32
115
GND
116
EMIT64
57
EMIT31
58
GND
117
EMIT63
118
GND
59
GND
60
EMIT34
119
GND
120
GND
Table 3-34 Detailed Definition of Bus Socket J3
1
PHV1_N
3
PHV1_N
5
NC
7
PHV2_N
9
PHV2_N
11
NC
13
A-12V
15
A-12V
2
4
6
8
10
12
14
16
PHV1_P
PHV1_P
NC
PHV2_P
PHV2_P
NC
A+12V
A+12V
31
33
35
37
39
41
43
45
3-51
GND
HSSB_TX0RFC
KN
HSSB_TX0RFC
KP
GND
HSSB_TX0RBC
KN
HSSB_TX0RBC
KP
GND
HSSB_TX0RN
32
34
36
38
40
42
44
46
NC
A+75V_sub
NC
GND
CLK_N_EMIT
CLK_P_EMIT
GND
HSSB_TX0TFC
KN
Principle Description
17
A-5V1
18
19
A+2V8
21
A+2V8
23
25
27
29
RATE_P_EMIT
RATE_N_EMIT
TMS
22
24
26
28
TDO_EMIT
3.2.4
20
30
A+5V1
47
A+1V5
49
A+1V5
51
RESET_EMIT
53
CLOCKEN_N_E
MIT
TDI_EMIT_TDO
_BF
TCK
55
57
59
HSSB_TX0RP
GND
HSSB_TX1RN
HSSB_TX1RP
GND
HSSB_TX0TN
HSSB_TX0TP
48
50
52
54
56
58
60
HSSB_TX0TFC
KP
GND
HSSB_TX0TBC
KN
HSSB_TX0TBC
KP
GND
HSSB_TX1TN
HSSB_TX1TP
4D Driving Board
The principle diagram of 4D driving board is shown in figure below. The interface signals of 4D
driving board and main board consist of power, SPI interface signals of ADC and DAC, and the
enabling signal of power module. The interface signals of 4D driving board and transducer
connecting board are: two driving signals A and B, HALL signal and its 5V power supply, and
temperature and angle signals of TEE transducer. 4D driving board mainly implements the following
two functions:
z
Amplifies the command-driven current and generates signals of appropriate power to drive
the transducer to the desired position.
z
As a reserved design, it amplifies the temperature and angle signals of the TEE transducer
and provides data collection channel.
3-52
Principle Description
Fig 3-16 Block diagram of 4D driving board
3.2.4.1
Definition of the Interface between the 4D Driving Board and Main
Board
Signals of the socket are defined as follows:
Table 3-35 Definition of Socket between 4D Driving Board and Main Board
PIN
Signal name
Direction
PIN
Signal name
Direction
1
12V
Input
2
12V
Input
3
12V
Input
4
12V
Input
5
12V
Input
6
12V
Input
7
12V
Input
8
12V
Input
9
GND
/
10
GND
/
11
GND
/
12
GND
/
13
GND
/
14
GND
/
15
GND
/
16
GND
/
17
-12V
Input
18
-12V
Input
3-53
Principle Description
PIN
Signal name
Direction
PIN
Signal name
Direction
19
-12V
Input
20
-12V
Input
21
-12V
Input
22
-12V
Input
23
GND
/
24
GND
/
25
GND
/
26
GND
/
27
GND
/
28
GND
/
29
VDD
Input
30
VDD
Input
31
GND
/
32
GND
/
33
VCC
Input
34
VCC
Input
35
GND
/
36
GND
/
37
NC
/
38
NC
/
39
NC
/
40
NC
/
41
NC
/
42
NC
/
43
NC
/
44
NC
/
45
NC
/
46
NC
/
47
NC
/
48
NC
/
49
GND
/
50
GND
/
51
PWR_EN
Input
52
GND
/
53
HALL
Output
54
GND
/
55
NC
/
56
GND
/
57
SPI_CLK
Input
58
GND
/
59
SPI_DAT
Bidirectional
60
GND
/
61
SPI_SYNC
Input
62
GND
/
63
FD_ID
Bidirectional
64
GND
/
65
AD_SCLK
Input
66
GND
/
67
AD_DIN
Input
68
GND
/
69
AD_DOUT
Output
70
AD_NCS
Input
3-54
Principle Description
Note: wherein the direction of signal is defined as: the signal from main board to 4D driving board is
input signal, the signal from 4D driving board to main board is output signal.
3.2.4.2
Definition of the Interface between the 4D Driving Board and
Transducer Connecting Board
The wire-to-board socket (WTB) of 4D driving board signals are defined as follows:
Table 3-36 Definition of Socket between 4D Driving Board and Transducer Connecting Board
PIN
Signal name
Direction
1
AGND
/
2
5V
Output
3
HALL
Input
4
ANGLE
Input
5
AGND
/
6
T+
Input
7
T-
Input
8
PHASE_B_NEG
Output
9
PHASE_B_NEG
Output
10
PHASE_B_POS
Output
11
PHASE_B_POS
Output
12
PHASE_A_NEG
Output
13
PHASE_A_NEG
Output
14
PHASE_A_POS
Output
15
PHASE_A_POS
Output
Note: wherein the direction of signal is defined as: the signal from transducer connecting board to
4D driving board is input signal, the signal from 4D driving board to transducer connecting board is
output signal.
3.2.4.3
Testing Points of 4D Driving Board
The testing points of 4D driving board are listed in the following table:
3-55
Principle Description
Table 3-37 List of Testing Points of 4D Driving Board
No.
Label of the Testing
Point
Signal definition of testing points
1
TP3
Command current bias voltage from DAC (DAC_BIAS)
2
TP4
Output of command current error amplification circuit of phase B
3
TP5
Positive output of driving current of phase B (PHASE_B_POS)
4
TP6
Command current of phase B from DAC (DAC_B_SET)
5
TP7
Output of command current signal amplification circuit of phase B
6
TP8
Command current of phase A from DAC (DAC_A_SET)
7
TP9
Output of command current signal amplification circuit of phase A
8
TP10
Output of command current error amplification circuit of phase A
9
TP11
Positive output of driving current of phase A (PHASE_A_POS)
3.2.5
Principle of Keyboard Board
3.2.5.1
General Description
The control panel of 2118 implements the following functions:
Input from buttons: input by general button, combination button, repeat button and encoder are
supported;
Backlight LED of buttons, control of status indicating LEDs
Control of beeps of buttons; larger adjustable range of volume; more than 3 grades are provided;
TGC input, full-scale and 0 available
Input from encoder
Input from optical trackball; control of backlight of optical trackball
3.2.5.2
Block Diagram of Structure
The block diagram of control panel is shown in below:
3-56
Principle Description
Fig 3-17 Block Diagram of the Control Panel
3.2.5.3
Scanning Circuit of Buttons
The keyboard array is controlled by row-and-column controlling method. FPGA outputs tests signals
line-by-line. When a key is pressed, the corresponding row and line signals shall be consistent.
According to the numbers of the row and line, the position of the key pressed can be defined.
74LS0
7
F
P
G
A
L0
R
L1
R
L2
R
R0
R1
R2
GND
74HC14
Fig 3-18 Principle Diagram of Key Scanning
3.2.5.4
Driving of LED
Constant-current source driving dedicated chip is adopted to drive the LEDs.
3-57
Principle Description
Fig 3-19 Connection of LED Driving Circuit
There are 16 constant-current sources for each TLC5922, and the brightness can be adjusted
respectively.SPI interface control is adopted for TLC5922, and plural chips can be connected in
series for control. The power supply voltage is 5V.
3.2.5.5
Processing of Signals from Trackball
A newly designed optical trackball and a mechanical trackball are supported by 2118. The
advantages of the new trackball are:
z
Higher resolution and performance;
z
Full color backlight;
z
Optical device, no need for daily clean;
z
Larger in size;
z
Digital interface, more flexible in control;
The data returned from the trackball interface is square waveform data. There are a pair of signal
cables in X and Y directions respectively, and four signal cables in all. The trackball is powered by
5V, supplied by the control panel.
The basic scheme of the optical trackball is shown below. It consists of trackball IC and backlight
LED.
Fig 3-20 Basic Scheme of the Optical Trackball
3-58
Principle Description
3.2.5.6
Buzzer
The speaker is piezoelectric ceramic. It is driven by 74LS07. The basic drive circuit is shown below:
+5V
Piezoelectric
ceramic
piece
Control
logic
74LS07
Fig 3-21 Speaker Control Circuit
The piezoelectric ceramic has no positive / negative pole. As long as driving voltage with
appropriate frequency is put to its two ends, it will make a sound with the same frequency as that of
the driving voltage.
Whether it sounds or not, this is under the control of control logic. To generate sounds of different
frequencies, just change the frequency of the driving circuit.
3.2.5.7
Encoder Circuit
There are two encoders on the control panel, and each encoder has two signal cables. The output
signal is in the form of square waveform and the electrical level is 5V.
3.2.5.8
TGC Interface Circuit
TGC is actually a group of sliding rheostats, the output voltage of which is in linear relationship with
its slider displacement. The output analog voltage is converted to digital signal by AD converter and
sent to FPGA, thus obtaining the slider position.AD is powered by 3.3V, externally connected to
2.5V reference level. To keep good precision of the AD converter, the output of the sliding
potentiometer is applied with LM358 for follow-up. The principle of the interface is shown in figure
below.
Fig 3-22 TGC Interface Circuit
3-59
Principle Description
3.2.5.9
FPGA and Peripheral Circuits
FPGA is the core of boards and cards, externally connected to SPI flash and SDRAM. Its logic and
software codes are stored in SPI Flash, and they are loaded to FPGA and initiate the software when
powered on.
NiosII soft core is a 32bit CPU. It can read and write SPI Flash on line and support on-line code
upgrading.
FPGA, Flash and SDRAM are all 3.3V devices. The core voltage of FPGA is 1.2V.
3.2.5.10 USB Communication Interfaces
FPGA is externally connected with a USB interface chip U4 (3.3V device), supporting USB
communication protocol 1.1. The frequency of external crystal oscillator X1 is 6MHz.USB interface
is shown in figure below. In the figure, L8 is common mode suppression inductor, and U5 is
anti-static chip.
VBUS
DM
USB chip
U4
L8
DP
GND
F
P
G
A
U5
Fig 3-23 Schematic of USB Port
3.2.5.11 Socket Definition
The components and sockets positions of the control panel are shown in the following figures. The
socket definition is shown in Table 3-38.
3-60
Principle Description
Display switch
Encoder
Encoder
Fig 3-24 Components Layout of the Control Panel Board (Frontal)
3-61
Principle Description
Fig 3-25 Components Layout of the Control Panel Board (Bottom)
Table 3-38 Definition Table of Keyboard Board Socket
Socket
Name
number
Definition of Pin
Pin No
on PCB
Notes
Definition of
Description of PIN
Pin
1
TCK
TCK signal,
pull-down
JTAG
J1
2
TDO
TDO
3
TMS
TMS signal, pull up
3.3V
debugging
socket
4
TDI
TDI signal, pull up
3.3V
J2
TGC socket
5
VDD
3.3V power
6
GND
System ground
1
TGC VDD
3.3V power
3-62
Principle Description
2
GND
System ground
3
ATGC0
Slide bar 0 output
4
ATGC1
Slide bar 1 output
5
ATGC2
Slide bar 2 output
6
ATGC3
Slide bar 3 output
7
ATGC4
Slide bar 4 output
8
ATGC5
Slide bar 5 output
9
ATGC6
Slide bar 6 output
10
ATGC7
Slide bar 7 output
3
DP
USB-DP
4
POWER_STB
Power status
Power
connecting
switch K93
5
DN
USB-DN
6
LCD_SW
Display position
Connect
status
display
position
switch S3
Main board
J3
8
connecting
EDC_STATU
EDC status
S_G
Connectio
n indicator
socket
light D4
9
BATTERY_ST Power status O
Connectio
ATUS_O
n indicator
light D5
10
BATTERY_ST Power status G
Connectio
ATUS_G
n indicator
light D5
11
WORK_STAT
US_O
Working status O
Connectio
n indicator
light D6-7
3-63
Principle Description
12
WORK_STAT
Working status G
US_G
Connectio
n indicator
light D6-7
1,2,7,1
GND
System ground
Audio Out
Keyboard board
3~15,1
7~18,
23~26
16
Reserved
audio power
amplification output
19~22
+5
External input 5V
27~30
+3V3
External input 3.3V
1
GND
System ground
2
VCC
5V
3
TRACK_HBA
Trackball signal 2
LL2
and 1 is a
Mechanical
J5
trackball
4
TRACK_HBA
Trackball signal 3
LL3
socket
5
Signal 0
match;
signal 2
and 3 is a
TRACK_HBA
Trackball signal 0
match
LL0
6
TRACK_HBA
Trackball signal 1
LL1
J4
Mechanical
1,9,
trackball
11,13,
socket
15
GND
2
VDD
3
MISO
System ground
Serial Data Output
Output
(Master In/Slave
Out)
4
MOSI
Serial Data
Input( Master
Out/Slave In)
3-64
Input
Principle Description
5
SCLK
Serial Clock Input
Input
6
NCS
Chip Select(Active
Input
Low Input)
7
MOTION
Input
Motion Detect
(Active Low
Output)
8
SHTDWN
Shutdown(Active
Input
High Input)
10
LED_DRIVER
LED driver 1
Input
LED driver 2
Input
LED driver 3
Input
1
12
LED_DRIVER
2
14
LED_DRIVER
3
J8
3.2.6
1
BUZ0
Buzzer signal 0
2
BUZ1
Buzzer signal 1
Buzzer socket
Display Unit
The LCD interface consists of LVDS digital signal input interface and CCFL strip lamps interface;
wherein LVDS signals are low-voltage difference digital signals, and VDD is 3.3V; the definition of
its socket connecting to main board is shown as follows:
Table 3-39 Interfaces of Main Board and Speakers
PIN
Definition
PIN
Definition
1
VDD
2
VDD
3
GND
4
GND
5
RXE0-
6
RXE1-
7
RXE0+
8
RXE1+
9
GND
10
GND
11
RXE2-
12
RXEC3-65
Principle Description
PIN
Definition
PIN
Definition
13
RXE2+
14
RXEC+
15
GND
16
GND
17
RXE3-
18
GND
CCFL strip lamps are supplied with high voltage from the inverter, producing different brightness
according to different currents. The principle of the inverter is:
The 12V input of the inverter comes from the main board. The main IC of the inverter drives 4
MOSFETs opening/closing alternatively and puts AC square wave to the primary winding the
transformer. The high-voltage square wave induced by the secondary winding is transformed into
sine wave through LC filtering and drives CCFL strip lamp shining. The switch signal controls the
on/off of the strip lamp by controlling the enabling of main IC. The brightness signal adjusts the
brightness of the strip lamp by controlling the duty cycle of square wave of main IC.
The inverter board is equipped with protection circuits of strip lamp open circuit, overvoltage and
over-current. Protection of strip lamp: whenever any of the strip lamps is open circuit, the protection
circuit will act; over-current protection: when the output current is over-current, the feedback signals
from the strip lamp to main IC will restrain the duty ratio of the square wave to reduce the current for
protection; overvoltage protection: the IC has been configured with the voltage-limiting function to
ensure the output voltage is not too high. The socket J26 connecting to main board is defined as
follows:
Table 3-40 Interfaces of Main Board and Inverter
PIN
Definition
1
12V
2
12V
3
12V
4
ON/OFF
5
ADJ
6
GND
7
GND
8
GND
3-66
Principle Description
Wherein ON/OFF is the switch signal of the inverter; ADJ controls the brightness of the LCD by
inputting different voltages which are then converted to currents by the inverter;
EEPROM board storages one-to-one related information of LCD (such as serial number and color
temperature version, etc) and gamma correction parameters; the socket connecting to main board
is defined as follows:
Table 3-41 Interfaces of Main Board and EEPROM
PIN
Definition
1
VDD
2
GND
3
SCL
4
SDA
The logic will output different RGB mapping tables according to the storage contents to ensure the
consistence of image output effect for different LCDs;
Note: To ensure the display effect, when changing LCD assembly or main board assembly,
enter Maintenance Menu to reload the LCD parameters. This is display parameter
synchronization. The synchronization procedure is described as follows:
1. Restart the Doppler system:
2. Log onto the system as service engineer:
3. Press <F10> to open the Setup menu, see the figure below:
Fig 3-26 Setup Menu
4. Click [Maintenance] in the menu to open the menu in the figure below:
3-67
Principle Description
Fig 3-27 Maintenance Menu
5. Click [Monitor Test] →[Main Monitor] to open the following screen:
Fig 3-28 Main Screen Maintenance Screen
6. See the figure above, click the top icon on the screen (update) to start the maintenance, when
completed, a dialogue box showing successful maintenance will be displayed, restart the
system to make the new setting take effect. If the maintenance is failed, please return the
system to development center. The following two figures show the two situations respectively.
3-68
Principle Description
Fig 3-29 Successful Update
Fig 3-30 Update Failed
3.2.7
Principle of ECG Board
The function of the ECG board contains: ECG signal detection; ECG waveform display; ultrasound
image reference signal; producing real-time 2-D image and color flow image synchronous signal;
displaying ECG signals from external patient monitor.
After being amplified, filtered and sampled, the ECG signal is sent to host PC, and R-waveform
detection is performed simultaneously. After being detected, the ECG trigger signal is sent to host
PC via serial port. Block diagram is shown as follows:
3-69
Principle Description
Fig 3-31 Block Diagram of the ECG Board
3.2.7.1
Power Principle of ECG Board
ECG board is powered by main unit, and the voltage is 5V. 5V is transformed to 12V after going
through the step-up circuit. T1 is the isolated transformer. The 12V power is isolated and then sent
to ECG board to power the analog and digital circuits. As shown in figure below.
Fig 3-32 Power Functional Block Diagram of ECG Board
3.2.7.2
Instrumentation Diagram and Component Description
The instrumentation diagram of ECG board is shown in the following figure. The figure marks the
sockets and their names. The blue dotted line indicates the component isolation boundary. The right
isolation area indicates the power circuit and communication port, and the left isolation area
indicates ECG signal processing circuit
3-70
Principle Description
Fig 3-33 The instrumentation diagram of ECG board
The interfaces are shown in the figure, and the interfaces are defined in the following table:
Table 3-42 ECG Board Instrumentation Diagram Description
Socket
Name
number
on PCB
J1
J2
DC_IN
Lead
cable
Definition of Pin
Notes
Pin No
Definition of Pin
Description of PIN
1
FGND
ECG
module floating ground
2
FGND
ECG
module floating ground
3
DC_IN
DC IN
4
FGND
ECG
module floating ground
5
FGND
ECG
module floating ground
6
FGND
ECG
module floating ground
1
Lead_F_IN
Lead cable 1
2
Lead_R_IN
Lead cable 2
3-71
Principle Description
interface
3
Lead_N_DR
Lead cable 3 (connected with
socket
J4
Main
system
interface
the right ) (to the right leg)
4
NC
No connection
5
C_SHIELD
Cable shield drive
6
NC
No connection
1
GND
Main system ground
2
ECG_RX
ECG signal receiving (relative to
the main board )
socket
3
ECG_TX
ECG signal transmitting (relative
to the main board )
J6
JTAG
debugging
socket
J7
ARM
4
ECG_PWR
Power for ECG board
5
GND
Earth terminal of main system
6
NC
Earth terminal of main system
7
GND
Earth terminal of main system
8
GND
Earth terminal of main system
1
VCC
3.3V
2
GND
ECG module floating ground
3
TRST
JTAG——TRST
4
TCK
JTAG——TCK
5
TDI
JTAG——TDI
6
TDO
JTAG——TDO
7
TMS
JTAG——TMS
8
RTCK
JTAG——RTCK
9
NRST
JTAG——NRST
10
NC
No connection
1
/
Connect to LPC213X P0.14 pin
BOOT
of ARM chip
2
GND
ECG module floating ground
3-72
Debugging
Service
Principle Description
3.2.8
Principle of Video/Audio Capture Card
The diagram of video capture card is shown in the figure below:
Reset control
EMP202 syn. Clock
Syn. signal
Audio_L
Audio_R
Control feedback
EMP202
Control
USB
TVP5150 Synchronization clock
VIDEO/S
VIDEO
SVIDEO
ADG774
VIDEO
TVP5150
CH select
EM2860
8-bit parallel
digital signal
I2C
Line-field syn. signal
Fig 3-34 System Block Diagram of Video/Audio Capture Card
The power relation of video audio capture card is shown in the figure below:
3.3V
Fig 3-35 Power Relation Diagram of Video/Audio Capture Card
Principle of the video capture card is described as follows:
After the video capture card is powered on, EM2860 first resets and communicates with the main
unit, and TVP5150 resets simultaneously. At this moment the system displays the newly found
hardware, namely EM2860.After EM2860 resets, its GPIO1 signal will change and control the reset
of EMP202 through PMOSFET.
After the process is completed, both the BCLK and VCLK have clock signals, and SYNC has cyclic
pulses.
3-73
Principle Description
When the capture software is started, LED pin of EM2860 is pulled high, and green LED in the
board is lit up. The video capture card transmits the video capture configuration in the software to
TVP5150 via I2C bus. The TVP5150 sends channel selecting command to analog switch ADG774
according to the received command, and selects encoding method corresponding to configuration.
TVP5150 converts the input video signals to 8-bit digital signals, and transmits them to EM2860. At
the same time, EMP202 transmits the captured audio signals to EM2860 via serial channel.
EM2860 integrates the received signals and then sends them to PC via USB port.
3.2.9
Principle of the CW Board
The block diagram of CW board is shown as figure below:
Fig 3-36 Principle Diagram of CW board
The hardware block diagram of CW board is shown as the figure above. Echo signals that received
by the phased probe and pencil probe will first be amplified through low-noise-amplifying (LNA)
circuit, then pass through the DEMODULATOR to obtain I, Q signals, and I, Q signals will be
sampled through ADC after being filtered out the useless low-frequency and high-frequency signals.
The sampled signals then will be sent through the interface for beam-forming and so on.
Low-phase noise demodulation clock is provided by the clock distributor circuit.
In addition to echo receiving, CW board is also responsible for pencil probe signal transmitting.
The control signals, as well as the data signals, are communicated with the external through
interfaces
Definition of CW board interfaces is shown as follows:
3-74
Principle Description
Table 3-43 Pin definition of J1
1
GND
2
GND
21
GND
22
GND
3
CW19
4
CW20
23
CW9
24
CW10
5
GND
6
GND
25
GND
26
GND
7
CW17
8
CW18
27
CW7
28
CW8
9
GND
10
GND
29
GND
30
GND
11
CW15
12
CW16
31
CW5
32
CW6
13
GND
14
GND
33
GND
34
GND
15
CW13
16
CW14
35
CW3
36
CW4
17
GND
18
GND
37
GND
38
GND
19
CW11
20
CW12
39
CW1
40
CW2
Table 3-44 Pin definition of J2
1
A-5V1
2
A+5V1
21
SHUTDOWN
22
AD8339_CSB
3
A-5V1
4
A+5V1
23
GND
24
GND
5
GND
6
GND
25
TPU0_PENCIL_CW
26
AD_SCK
7
BOARDID_CW
8
PPID_CW
27
GND
28
GND
9
CWPPPRESENT
10
TUNING_SW_CW
29
GND
30
AD_CNV_CW
11
GND
12
GND
31
AD8339_4LO0_CW
32
GND
13
AD_SDO1
14
AD8339_RSTS
33
AD8339_4LO1_CW
34
AD8339_RSET
15
AD8339_SDI
16
AD8339_SCLK
35
GND
36
GND
17
GND
18
GND
37
VOL_ADJ_SDA
38
VOL_ADJ_SCK
19
AD8339_SDO
20
AD_SDO0
39
A+12V
40
A+3V6
Table 3-45 Pin definition of J3
1
CWPPPRESENT
4
PEN_EMIT
2
PPID_CW
5
GND
3
GND
6
PEN_SIGNAL
Table 3-46 Connector signal description of CW board
3-75
Principle Description
Name
Meaning
Direction
Type
Notes
CW[10:1]
CW IN
I(sink) Single-end
current
CW[20:11]
CW IN
I(sink) Single-end
Current (reserved)
A-5V1
A-5.1V power
I
Power
A+5V1
A+5.1V power
I
Power
A+12V
A+12V power
I
Power
A+3V6
A+3.6V power
I
Power
BOARDID_CW
CW board ID
I/O
J1
J2
CWPPPRESENT
AD_SDO[1:0]
Pencil probe
present
AD serial port
output data
O
O
chain serial port
One signal device
Single-end
Single-end
Single-end
AD8339SPI daisy
AD8339_SDI
Single-end
I
input data
Single-end
AD8339SPI daisy
AD8339_SDO
chain serial port
O
output data
PPID_CW
TUNING_SW_CW
AD8339_RSTS
AD8339_SCLK
TPU0_PENCIL_CW
SHUTDOWN
Pencil probe ID
Tuning inductive
relay switch
AD8339 internal
register reset
AD8339 SPI data
clock
CW pencil probe
transmitting input
ENB of AD8432
O
I
I
I
I
I
3-76
Single-end
One signal device
Single-end
Single-end
Single-end
Single-end
Single-end
Low-phase noise
Principle Description
AD8339_4LO[1:0]_CW
AD8339_CSB
AD_SCK
AD_CNV_CW
AD8339_RSET
Input demodulation
clock
AD8339 SPI CS
(chip selection)
ADC data clock
ADC sample clock
input
AD8339 internal LO
reset
I
I
I
I
I
digital potentiometer
Variable-frequency,
low-phase noise
Single-end
Single-end
I/Q share
Single-end
I/Q share,
low-phase noise
Single-end
Single-end
I2C data line of
VOL_ADJ_SDA
difference
I/O
I2C
AD5259
Single-end
I2C clock line of
VOL_ADJ_SCK
digital potentiometer
I/O
I2C
AD5259
J3
PEN_SIGNAL
PEN_EMIT
PEN_ID
PEN_PRESENT
Pencil probe input
signal
Pencil probe
transmit signal
Pencil probe ID
Pencil probe
present signal
I
O
I
I
3-77
Single-end
Single-end
Single-end
Single-end
4
Structure and
Assembly/Disassembly
4.1
Explosive Figure of the Structure
4.1.1
Explosive Figure of the Complete System
4-1
Structure and Assembly/Disassembly
1
4
2
6
3
7
9
5
12
10
8
Fig 4-1 Overall exploded view
No.
Name
No.
Name
No.
Name
<1>
Display assembly
<2>
Main board
<3>
Transducer board
115-002254-00
assembly
051-000098-01
115-002943-00
<4>
CPU board and
<5>
Rear cover
radiator module
assembly of main
115-003754-00
unit
034-000041-00
4-2
<6>
Damping axis cover
043-000453-00
Structure and Assembly/Disassembly
<7>
Power supply
<8>
Top cover assembly
<9>
Fan mounting
module;
of main unit
bracket of CPU
115-002256-00
115-002942-00
module
042-000415-00
<10> Battery
<11>
2108-20-65792
CW board PCBA
<12>
051-000170-00
4D board PCBA
051-000102-01
4.1.2
Explosive Figure of the Modules
4.1.2.1
Explosive Figure of the Power Module
1
3
2
4
5
6
Fig 4-2 Explosive figure of the power system
No.
Name
No.
<1>
Top cover of power box <2>
Name
No.
Name
Philips panhead screw
<3>
Power supply main
M3×6
board
4-3
Structure and Assembly/Disassembly
<4>
Separating plate.
<5>
Power supply auxiliary
<6>
Bottom cover of
board
4.1.2.2
power box
Explosive Figure of Main Unit Top Cover
Fig 4-3 Explosive figure of the main unit top cover
No.
Name
No.
Name
No.
Name
<1>
Philips panhead screw
<4>
Buzzer
<7>
Control panel
<5>
Top cover of main unit
<8>
Trackball
M3×6
<2>
TGC adjusting board
connection
wire
<3>
Philips panhead screw
M2×10
<6>
Philips panhead screw
M3×6
<9>
STC
connection
wire
4-4
Structure and Assembly/Disassembly
3
1
2
4
Fig 4-4 Explosive figure of the main unit
No.
Name
No.
Name
<1>
Main board
<2>
4D board
<3>
HDD
<4>
Transmitting board
shielding cover
4-5
4.2
NO.
Field Replaceable Unit
Classification
Description
Order Number
Photo
Model
1.1
Main Board
801-2118-00006-00
M7、M7T
1.2
CPU module
801-2118-00011-00
M7、M7T
Compatibility
Includes button battery and jumper
Assembly/
Disassembly
Refer to
4.3.11
Refer to
4.3.7
Main unit
related
Mark the following when apply: software
version (Human use or Vet), machine
115-018485-00
M7、M7T
type: human;
Configuration type: FDA.
Not include metal support
1.3
HDD
Mark the following when apply: software
version (Human use or Vet), machine
115-018486-00
M7Vet
type: Vet;
Configuration type: FDA.
Not include metal support
4-6
Refer to
4.3.14
Structure and Assembly/Disassembly
NO.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
Mark the following when apply: software
version (Human use or Vet), machine
115-018487-00
M7、M7T
type: human;
Configuration type: CE.
1.3
HDD
Not include metal support
Refer to
Mark the following when apply: software
4.3.14
version (Human use or Vet), machine
115-018488-00
M7Vet
type: Vet;
Configuration type: CE.
Not include metal support
1.4
1.5
Transmission
Board
Probe board
assembly
801-2118-00004-00
M7、M7T
801-2118-00013-00
M7、M7T
4-7
Mark software version when apply
Refer to
4.3.10
Refer to
4.3.12
Structure and Assembly/Disassembly
NO.
Classification
1.6
CW board
4D and TEE
1.7.1
board assembly
4D board
1.7.2
2.1
Description
assembly
Power
Power supply
related
module
Order Number
Photo
Model
801-2118-00030-00
M7、M7T
801-2118-00036-00
M7、M7T
801-2118-00031-00
M7、M7T
801-2118-00001-00
M7、M7T
4-8
Compatibility
Assembly/
Disassembly
Refer to
4.3.5
Support TEE probe and 4D probe
Not support TEE probe
Mark software version when apply
Refer to
4.3.13
Refer to
4.3.13
Refer to
4.3.4
Structure and Assembly/Disassembly
NO.
Classification
Description
Adapter
2.2
(ADP1210-01)
Order Number
Photo
801-2108-00049-01
Model
Li-ion Battery
2.3.2
For CE
2108-30-66176
M7、M7T
together
Rechargeable
For FDA
115-010085-00
M7、M7T
Pack
3.1
Control
panel related
Two batteries should be replaced
Pack
Li-ion Battery
Control panel
assembly
Two batteries should be replaced
together
801-2118-00035-00
M7、M7T
4-9
Assembly/
Disassembly
/
M7、M7T
Rechargeable
2.3.1
Compatibility
Not included KeyBoard signal cable
Refer to
4.3.1
Refer to
4.3.1
Refer to
4.3.3
Structure and Assembly/Disassembly
NO.
3.2.1
3.2.2
3.3.1
3.3.2
3.4
Classification
Description
Control panel
PCBA (EAF196)
Control panel
PCBA
Silicon
keypad(EAF196)
Silicon keypad
Optical trackball
unit
Order Number
Photo
Model
Compatibility
If the Silicon keypad is
801-2118-00027-00
M7、M7T
801-2118-00028-00(049-000113-00
), you should apply this part.
801-2118-00007-00
M7、M7T
M7、M7T
M7、M7T
801-2118-00009-00
M7、M7T
4-10
Refer to
4.3.3.2
Refer to
Control panel assembly
4.3.3.2
801-2118-00027-00(051-000095-02
), you should apply this part.
801-2118-00010-00
Disassembly
Stop production, plese replace
If the Control panel PCBA is
801-2118-00028-00
Assembly/
Refer to
4.3.3.2
Stop production, please replace
Refer to
Control panel assembly
4.3.3.2
Not include Trackball connecting
Refer to
cable
4.3.3.1
Structure and Assembly/Disassembly
NO.
Classification
3.5
4.1
Description
TGC Board
Display
related
5.1
LCD assembly
Fan
Order Number
Photo
801-2108-00014-00
Model
M7、M7T
115-037592-00
Compatibility
Not include connecting cable
Assembly/
Disassembly
Refer to
4.3.3.2
Refer to
M7、M7T
4.3.2
801-2118-00015-00
M7、M7T
CPU fun
801-2108-00052-00
M7、M7T
Right fun
Refer to
4.3.6
Others
5.2
Fan & Connect
Wire
4-11
Refer to
4.3.8
Structure and Assembly/Disassembly
NO.
Classification
Description
Fan & Connect
5.3
Wire
5.4
speaker
6.1
IO Extend
Order Number
Photo
Model
Compatibility
115-020423-00
M7、M7T
Left fan (3 as a group )
801-2108-00019-00
M7、M7T
115-003485-00
M7、M7T
For CE
115-003486-00
M7、M7T
For FDA
2108-30-66045
M7、M7T
For CE
M7、M7T
For FDA
Module
6.2
6.3
6.4
Accessories
Video Extend
related
Module
Assembly/
Disassembly
Refer to
4.3.8
Refer to
4.3.18
Refer to
4.3.17
2108-30-66211
This blank was left intentionally.
4-12
Structure and Assembly/Disassembly
NO.
Classification
Description
6.5
6.6
ECG module
6.7
6.8
Wireless card
6.9
Order Number
Photo
Model
Compatibility
115-003488-00
M7、M7T
For CE, AHA ECG lead
115-003489-00
M7、M7T
For CE, IEC ECG lead
6.10
Disassembly
Refer to
4.3.16
115-003490-00
M7、M7T
115-005370-00
M7、M7T
115-006962-00
M7、M7T
For FDA, AHA ECG lead
/
For CE
M7 Probe extend
module
Assembly/
Refer to
UMT-300
mobile trolley
115-006966-00
M7、M7T
For FDA
service
manual
6.11
M7 Probe extend
115-009385-00
M7、M7T
For CE
module
6.12
(PEM-21) for
UMT-200
Refer to M7
Probe
Expander
115-009386-00
M7、M7T
For FDA
UMT-200
installation
manual
4-13
4.3
Assembly/Disassembly
This section describes the disassembling and assembling of the main modules and boards. The
assembling is the inverse process of disassembling if not mentioned in particular.
4.3.1
Removing Batteries
2
1
Fig 4-5 Removing the battery
Move the two battery buckles in the direction as shown in the figure above, then take out the
battery.
No.
Name
No.
Name
<1>
Battery Buckle
<2>
Battery
4.3.2
Removing Display Assembly
As shown in below, remove the screws and the bottom cover and cable cover;
Disconnect the display signal cable and power cord connectors.
Fig 4-6 Removing Cables Cover at the Bottom Side
No.
Name
No.
Name
<1> Screws M3×8
<2>
Cable cover
<3> Display signal cable connector
<4>
Display power cord connector
4-14
Structure and Assembly/Disassembly
1. As shown in figure below, open the display to the biggest angle, press down the damping axis
cover to pop up the damping axis cover; take out the damping axis cover upward.
There are left and right damping axes, and they can be removed with the same method.
Fig 4-7 Removing the Damping Axis Cover
No.
Name
No.
Name
<1>
Main unit assembly
<2>
Damping axis cover
2. As shown in figure below, remove the six M3×8 screws; take out the display assembly upward
from the assembly. Pay attention not to damage the display signal cable, power cord, and
connectors during removal.
Fig 4-8 Disassembling Display Assembly
4-15
Structure and Assembly/Disassembly
No.
Name
No.
Name
<1>
Display assembly
<2>
Screw M3×8
3. As shown in figure below, take out the the screw caps securing the display, remove the four M3×8
screws, and then take out the display front cover from the back cover.
1
2
3
5
4
Fig 4-9 Removing Display Front Cover
No.
Name
No.
Name
<1>
Display screw cap
<2>
Screw M3×8
<3>
Display front cover
<4>
LCD assembly
<5>
Display back cover
Note: A hook is used to connect the front cover with back cover. It is recommended that you
first pry the lower-left corner and lower-right corner.
4. Remove the 2 screws fixing the inverter board of the LCD, and remove the 1 screw from the
cable clip, then remove the power cable of the backlight of the LCD, and then remove the
inverter board.
1
2
4-16
Structure and Assembly/Disassembly
Fig 4-10 Removing the Inverter
No.
Name
No.
Name
<1>
Screw M3×8
<2>
LCD display backlight inverter
As shown in figure below, remove one screw to loose the clip fixing LCD display cable, and then
take out the cable.
Remove the 4 screws and remove the LCD.
2
1
3
Fig 4-11 Removing the LCD
No.
Name
No.
Name
<1>
Screw M3×8
<2>
LCD
<3>
Display back cover
5. As shown in figure below, remove the six screws, and take out the left and right damping axes.
1
2
3
Fig 4-12 Removing the Left/right Damping Axis
No.
Name
No.
Name
<1>
Screw M3×8
<2>
Left damping axis
<3>
Right damping axis
4-17
Structure and Assembly/Disassembly
6. Take off the LCD screen, the back side view is as shown in figure 4-13.
LCD signal cable
LCD parameter board
Fig 4-13 Backside of LCD Screen
NOTE:
1. The whole LCD module should be replaced in the maintenance. The
corresponding LCD parameter board should be replaced once the LCD is
replaced, since the two is bonded together.
2. Before replacing, make sure the module has passed the test and the
related parameters have been burnt on the LCD parameter board.
7. According to figure below, first press the spring in axial direction, wait until the shorter axis comes
out, and then take out the hook upward.
Fig 4-14 Removing Hooks of Display
No.
Name
No.
Name
No.
Name
<1>
Right display hook
<2>
Left display hook
<3>
Display hook spring
NOTE:
After replacing the LCD module, start up the system and updates the LCD by
entering [Maintenance] -> [Main Monitor], and clicking the update button on the
top of the screen.
4-18
Structure and Assembly/Disassembly
4.3.3
Removing Top Cover of Main Unit
1. As shown in figure below, remove the 9 screws.
Fig 4-15 Removing the Top Cover Assembly of Main Unit (1)
No.
Name
<1>
Screw M3×8
No.
Name
2. Lift up the main unit top cover for 30 degrees, disconnect the wire between the control panel and
main board; pull it out in the direction as per the figure, and then take out the main unit top
cover.
Fig 4-16 Removing the Top Cover Assembly of Main Unit (2)
No.
Name
No.
Name
<1>
Top cover assembly of main unit
<2>
Connective cable between
control panel and main board
<3>
Display
4-19
Structure and Assembly/Disassembly
4.3.3.1
Removing the Trackball
As shown in figure below, pull out the socket of the trackball and remove the screws fixing the
trackball, and then remove the optical trackball.
1
2
Fig 4-17 Removing the Trackball
4.3.3.2
No.
Name
No.
<1>
Philips panhead screw M3×6 <2>
Name
Trackball
Removing Control Panel and TGC Board
1. As shown in figure below, remove TGC sliders, big encoder knob, small encoder knob, big
encoder button, small encoder button. When removing the encoder knob, please pry it at one
end of the knob.).
Fig 4-18 Removing the TGC Sliders, Encoder Knob and Button
4-20
Structure and Assembly/Disassembly
No.
Name
No.
Name
<1>
TGC sliders
<2>
Big encoder button
<3>
Small encoder button
<4>
Big encoder button
<5>
Small encoder button
2. As shown in figure below, disconnect the STC cable, trackball cable and buzzer cable; remove
the screws fixing the buzzer and then remove the buzzer; remove the screws fixing the TGC
adjusting board and the control panel and then remove the TGC adjusting board and the
control panel.
Fig 4-19 Removing TGC Adjusting Board and Control Panel
No
Name
No.
Name
No.
Name
<1>
Philips panhead screw
<4>
Buzzer
<7> Control panel
<5>
Top cover of
<8> Trackball connection
M3×6
<2>
TGC adjusting board
main unit
<3>
Philips panhead screw
M2×10
4.3.4
<6>
Philips panhead
screw M3×6
Removing the Power Module
1. According to 4.3.3, remove the top cover of the main unit;
4-21
wire
<9> STC cables
Structure and Assembly/Disassembly
2. As shown in figure below, remove the four M3×6 Philips panhead screws to take out the power
module from the main board.
1
2
Fig 4-20 Removing the Power Module
No.
Name
No.
Name
<1>
Philips panhead screw
<2>
Power supply module;
M3×6
Note: When removing the power module, keep the power board horizontally to prevent
damaging the connector pins.
4.3.4.1
Remove the Battery Socket Pressing Board
See figure below, remove the 1 Philips panhead screw M3×6, and then remove the battery socket
pressing board from the flexible board for connecting the batteries.
4-22
Structure and Assembly/Disassembly
2
1
Fig 4-21 Remove the Battery Socket Pressing Board
No.
Name
No.
Name
<1>
Philips panhead screw
<2>
Battery socket pressing
M3×6
4.3.4.2
board
Disassembling Power Module
1. As shown in figure below, remove the 2 screws 5 to remove the top cover of the power module;
remove the 4 screws 2 to take out the power board, separating board and auxiliary board
assembly.
1
5
2
3
4
Fig 4-22 Disassembling the Power Module
No.
Name
No.
Name
No.
4-23
Name
Structure and Assembly/Disassembly
<1>
Top cover of power <2>
Philips panhead screw
box
M3×6
<3>
Power main board,
separating board,
and auxiliary board
<4>
Bottom cover of
<5>
power box
Philips panhead screw
M3×6
2. As shown in figure below, remove the 8 screws to disconnect the connector between the main
power board and auxiliary power board.
1
2
5
3
4
Fig 4-23 Disassembling Main Power Board, Separating Board and Auxiliary Board
No.
Name
No.
Name
No.
Name
<1>
Philips panhead screw
<2>
Main power board
<3>
Power
M3×6
separating
board
<4>
Auxiliary power board
<5>
Connector of power main
board
4.3.5
Removing CW Board and Installation
‹Removing
Remove the two Philips panhead screws fixing the CW board, and take out the CW board upward.
Note: If a pencil probe is configured with the system, you shall first disconnect the pencil probe
connector before removing the continuous wave Doppler board.
4-24
Structure and Assembly/Disassembly
1
2
Fig 4-24 Removing CW Board
No:
Name
No.
Name
<1>
Philips panhead screw M2.5×8
<2>
CW board
‹Installation
1. If pencil probe is optional configured, remove the top of pencil probe hole in the rear cover
assembly of main unit, then install the pencil probe socket (The gap of socket is up),as shown in
the figure.
Pencil probe socket
2. Use two M2.5X8 panhead screws to fix the CW board. (If pencil probe is optional configured,
plug the pencil probe cable into the socket on the CW board and then tidy up the cable).
4-25
Structure and Assembly/Disassembly
CW Board
Pencil probe cable
Two M2.5X8 screws
4.3.6
Removing CPU Fan
As shown in 4.3.3, remove the top cover assembly of main unit;
As shown in figure below, pull out the CPU fan connector,
Remove the 4 screws of fan fixing bracket and 2 insulation grains on the left, and remove fan fixing
bracket and CPU fan and connection wire from the main board;
Take away the 4 CPU fan screws, and then remove the CPU fan.
Note: when install the lateral side fan, make sure the fan direction is the same as shown in
the figure below.
4-26
Structure and Assembly/Disassembly
3
1
2
4
5
Fig 4-25 Removing CPU Fan
No.
Name
No.
Name
<1>
Philips panhead screw
<2>
Insulation particles
<4>
CPU fan and cable
M3×6
<3>
Fan fixing bracket
<5>
Philips panhead screw
M4×10
4.3.7
Removing CPU Board and Radiator Module
1. According to 4.3.3, remove the top cover of the main unit;
2. As shown in figure below, first remove the four M3×6 Philips panhead screws, and then
disconnect CPU board radiator and industrial control board assembly from the main board;
Note: there is a jumper cap on the industrial control board, if it is forgotten to install, then the
system time will return to the default time each time you restart the system. So make sure to fix
the jumper cap at the right position.
4-27
Structure and Assembly/Disassembly
1
2
Fig 4-26 Removing CPU Board and Radiator Module
No.
Name
No.
Name
<1>
Philips panhead screw
<2>
CPU board and radiator
M3×6
module
3. As shown in figure below, remove the 5 M2.5×8 Philips panhead screws fixing the CPU board
radiator and industrial control board, separate the CPU board radiator and industrial control
board.
Prop-up
hole
2
1
3
Fig 4-27 Removing CPU Board
4-28
Structure and Assembly/Disassembly
No.
Name
No.
Name
<1>
CPU board and
<2>
Shielding cover <3>
Philips panhead screw
of CPU board
M2.5×8
memory
4.3.7.1
No.
Name
Removing Memory Stick
Move the buckles of the memory stick outside, the memory stick will automatically pop up, then take
out the memory stick.
Buckle
Fig 4-28 Removing the Memory Stick
4.3.8
Removing the System Fan
As shown in 4.3.3, remove the top cover assembly of main unit.
As shown in figure below, pull out connectors between the system fan, main board, and CPU fan;
Pull out the fan from the main unit bottom cover in the vertical direction.
4-29
Structure and Assembly/Disassembly
2
3
1
Fig 4-29 Removing CPU Fan
No.
Name
No.
Name
No.
<1>
Bottom cover of
<2>
Fan (3 as a group) <3>
Name
Fan
main unit
Note: When you reinstall the fans, note that the three system fans shall blow wind outside
(the brand shall face to the outside); the single fan shall blow wind outside (the brand shall
face to the inside),then verify that all fans can rotate normally.
4.3.9
Remove the Speakers
As shown in 4.3.3, remove the top cover assembly of main unit;
As shown in figure below, pull out the connector of the speaker;
Remove the four (2×2) M2.5×8 screws, and remove the speakers.
4-30
Structure and Assembly/Disassembly
2
1
Fig 4-30 Removing the Speakers
No.
Name
No.
Name
<1>
Speaker
<2>
Philips panhead screw
M2.5×8
4.3.10 Removing Transmission Board
As shown in 4.3.3, remove the top cover assembly of main unit;
As shown in figure below, remove the nine M2.5×8 Philips panhead screws;
Remove the top shield cover of the main board;
Remove the 4 screws fixing the transmission board, and disconnect the transmission board with
two hands. Note that the transmission board and main board are connected by four connectors, so
use force evenly.
4-31
Structure and Assembly/Disassembly
1
2
4
3
Fig 4-31 Removing Transmission Board
No.
Name
No.
Name
<1>
Philips panhead screw
<2>
Top shield cover of main
<3>
M2.5×8
board
Philips panhead screw M3×6 <4>
Transmission board
4.3.11 Removing Main Board and Transducer Board
Assembly
Remove the M3X8 screws as shown in Fig 4-32.
As shown in 4.3.3, remove the top cover assembly of main unit;
According to steps 1) and 2) in 4.3.1, remove bottom cover cable cover, disconnect display signal
cable and power cord;
According to 4.3.4, remove power module;
According to , remove CPU fan;
According to 4.3.7, remove CPU board and radiator module;
According to step 1) in 4.3.8, pull out connectors between the system an and main board;
According to step 1) in 4.3.9, pull out connectors between speaker and main board;
4-32
Structure and Assembly/Disassembly
As shown in Fig 4-33, remove the four M3×6 screws, and tilt the main board assembly up for about
30 degrees;
Pull the main board and transducer board assembly out slightly to remove them.
1
Fig 4-32 Removing the M3X8 Screw
No.
Name
<1>
M3×8 screw
No.
Name
Fig 4-33 Remove the Main board and Transducer Board
No.
Name
No.
<1>
Philips panhead screw M3×6
4-33
Name
Structure and Assembly/Disassembly
4.3.12 Remove the Transducer Board
1. According to 4.3.11, remove the main board and transducer board;
2. As shown in figure below, remove the 9 screws fixing transmission board shield cover; then
remove main board top shield cover and main board bottom shield cover.
1
2
3
Fig 4-34 Remove the Transducer Board (1)
No.
Name
No.
Name
<1>
Philips panhead screw M2.5×8
<2> Top shield cover of main
board
<3>
Bottom shield cover of main board
3. As shown in Fig 4-35 Remove the Transducer Board (2), pull out the transducer board from the
main board with two hands, use force evenly to prevent damaging the connector pins.
1
2
Fig 4-35 Remove the Transducer Board (2)
4-34
Structure and Assembly/Disassembly
No.
Name
No.
Name
<1>
Transducer board assembly
<2>
The connector between
main board and transducer
4.3.13 Removing 4D Board and Installation
‹ Removing:
1. According to 4.3.11, remove the main board and transducer board;
2. As shown in figure below, remove the 3 M2.5×8 screws, and then take out the 4D driving board.
1
2
3
Fig 4-36 Removing 4D Board
No.
Name
No.
Name
<1>
Main board assembly
<2>
Philips panhead screw
M2.5×8
<3>
4D Driving Board
‹ Installation:
1、 Affix the 4D heat outlet washer in the bottom board of the main unit cover as shown in
figure 4-37.
2、 Plug the 4D board PCBA to the main board, fix the board with 3 M2.5X8 cross panhead
screws, as shown in figure 4-38.
3、 Install the cable:
A. Plug one end of the 4D signal line to the 4D board PCBA socket, and affix the cables
tightly with fiber tape, as shown in figure 4-39.
4-35
Structure and Assembly/Disassembly
B. Plug the other end of the 4D signal line to the probe board socket across the hole in the
main board, and fix the cable using fiber cape and cable tie, as shown in figure 4-40.
4、 Install the mother board and probe board assembly into the main unit cover.
Next to stud hole
<1>
28~30mm to bottom
Figure 4-38
Board (left)
<3>
<4>
Figure 4-39
Fiber tape
<2>
Figure 4-40
4-36
Structure and Assembly/Disassembly
Figure 4-41
Serial
No.
<1>
Name
4D heat outlet
washer
No.
Serial
No.
<2>
1
Name
No.
4D signal line
009-000188-00
1
Philips panhead screw
M2.5×8
3
045-000164-00
<3>
4D board PCBA
051-000102-01
<4>
1
M04-051121---
Caution:
Note:
Use screw driver to manually screw the M2.5 screw tightly and
appropriately, be careful not to damage the board.
1. Discard the films at the both sides of the 4D heat outlet washer.
2. Press and hold the left board as in figure 2 when screwing the 4D board
PCBA screws.
3. Use adhesive tape to affix the 4D signal line, or other insulting tapes.
4.3.14 Removing Hard Disk
NOTE
please pay attention to the following matters during disassembling/assembling, otherwise
the hard disk will be damaged:
z Hold the side of the hard disk, and please do not touch the board of the hard disk.
z Fasten the screws with the handy screw driver, and do not with the electric screw driver.
z The torsion value of the screw lock is:
M3: 4 to 6 kgf.cm
M4: 6 to 8 kgf.cm
1. According to 4.3.11, remove the main board and transducer board;
2. Shown as figure below, remove the M3×6 screws, horizontally pull out the HDD and HDD fixing
part, then remove the four M3×4 screws.
4-37
Structure and Assembly/Disassembly
4
3
1
2
5
Fig 4-42 Removing the HDD
No.
Name
No.
Name
<1>
Main board assembly
<2>
Philips panhead screw
M3×4
<3>
Philips panhead screw M3×6 <4>
<5>
Hard disk
HDD fixing part,
4.3.15 Removing Transducer Extend Module PEM-11
1. Remove four M3×6 Philips panhead screws, and then remove the fixing board of the transducer
extend module.
2. Remove four M3×6 Philips panhead screws, then remove the bottom cover of the transducer
extend module.
4-38
Structure and Assembly/Disassembly
1
2
3
4
Fig 4-43 Remove the Transducer Extend Module (1)
3. Remove three M3×6 Philips panhead screws.
4. Pull out B transducer extend board vertically upward, and separate it from socket on the A
transducer extend board.
5. Disconnect the connective board on the transducer extend module socket of the B transducer
extend board, and remove B transducer extend board.
6. Hold the transducer wire sheath and pull it out vertically, and remove the extend module socket.
1
4
2
3
5
Fig 4-44 Remove the Transducer Extend Module (2)
4-39
Structure and Assembly/Disassembly
7. Remove two M2×4 Philips sunk screws, and pull out the transducer lever.
1
2
Fig 4-45 Remove the Transducer Extend Module (3)
No.
Name
No.
Name
<1>
Transducer lever
<2>
Philips sunk screws M2×4
8. Remove the 8 M3×6Philips panhead screws, and carefully pull out the A transducer extend
board.
1
2
Fig 4-46 Remove the Transducer Extend Module (4)
No.
Name
No.
4-40
Name
Structure and Assembly/Disassembly
<1>
Philips sunk screws M3×6
<2>
A transducer extend board
9. As shown in step 7), remove the transducer lever.
10. Remove the 8 M3×6 Philips panhead screws, remove the female base of transducer socket.
1
2
Fig 4-47 Remove the Transducer Extend Module (5)
No.
Name
No.
<1>
Philips panhead screw M3×6 <2>
Name
Transducer bracket
Note: PEM-21 transducer extend module should be applied with UMT-300 mobile trolley, for
disassembling UMT-300, please refer to the corresponding service manual.
4.3.16 Removing ECG Module
1. Use tweezers to take out the four connector enclosure screw caps (take care not to scratch the
enclosure), remove the four M3X8 screws, and remove ECG module top cover.
4-41
Structure and Assembly/Disassembly
1
2
3
Fig 4-48 Remove the ECG Module (1)
No.
Name
No.
Name
<1>
Screw cap
<2>
Screw M3X8
<3>
Top cover of ECG module
2. Remove the two M3X8 screws, disconnect connector of the ECG module internal signal cable,
and then remove the ECG board.
1
2
3
4
Fig 4-49 Remove the ECG Module (2)
No.
Name
No.
Name
<1>
Screw M3X8
<2>
Socket
<3>
ECG board PCBA
<4>
ECG serial cable
4-42
Structure and Assembly/Disassembly
3. Remove the four M2X8 Philips panhead tapping screws and flat washers. Cut off the two wire
strips on the magnetic ring, and remove the internal signal cable of ECG module.
2
3
4
1
5
6
Fig 4-50 Remove the ECG Module (3)
No.
Name
No.
Name
1
ECG socket fixing part
2
Screw M3X8
3
Cable tie
4
ECG internal signal cable
5
Magnetic ring fixing part
6
Philips panhead tapping screw
M2X8 and flat washer
4.3.17 Removing Video and Audio Capture Module
1. Use tweezers to take out the four connector enclosure screw caps (take care not to scratch the
enclosure), remove the four M3X8 screws, and then remove the top cover.
2. Remove the three M3X8 screws fixing the video and audio capture card, and remove video and
audio capture card.
1
2
4-43
3
4
Structure and Assembly/Disassembly
Fig 4-51 Removing Video and Audio Capture Module
No.
Name
No.
Name
1
Screw cap
2
Screw M3X8
3
Top cover
4
Video/ audio capture card
5
Bottom cover
6
Screw
M3X8
4.3.18 Removing I/O Extend Module
Remove the five M3X8 screws, and separate the I/O extend module.
4-44
Structure and Assembly/Disassembly
1
2
3
4
Fig 4-52 Remove the I/O Extend Module
No.
Name
No.
Name
1
Screw M3X8
2
Back cover
3
IO extend board
4
Front cover
4-45
5
System and Software
5.1
XPE System Introduction
The full name of XPE is Windows™ XP Embedded, it is the embedded system for Windows™ XP of
Microsoft® company. It is not a dedicated embedded operation system, but providing a tool for
development and deployment. XPE can be used to define and edit the standard Windows XP, and
some embedded application features and units are added to it.
For M7, after the XP operation system, the following drivers and functions are automatically
installed:
Table 5-1 Functions defined by XP operation system
Install the main board driver: chip group, video card, audio card, and network card.
Install the video printer driver: SONY UP-D23MD driver, SONY UP-D897 driver, and
MITSUBISHI P93D driver.
Install digital printer: HP Deskjet 2568, HP Office 3600, and HP 1015
USB video capturing box driver
USB Footswitch
Install CPU temperature and fan monitoring service
Hardware self-test program running environment
Keyboard filter driver
Install east Asia language: set the language and region as Chinese (PRC)
Install the desktop theme: M7.Theme, no voice.
Power on/Power off/login define
WINDOWS boot-up interface setting
No Autorun or autoplay
No screen saver
Shut down functions like the fire wall and automatic updating.
5-1
System and Software
5.2
Introduction of Doppler
5.2.1
Resume Package
run.bat: run the batch file
recover.exe: recover file
preset.txt: preset data default regulatory type control file
gina and gina.cmd: define the boot-up progress bar
demo: demo data
DC7Filelist.bak: recover data list.
Fig 5-1 Structure of the resume package
Note:
5.2.2
Resume package should be compatible with the machine type, that human package be
resumed on human machine, vet use package be resumed on vet machine.
Data Processing
After the resume package is activated, the data structure and storage destination are listed as
follows:
5-2
System and Software
Table 5-2 Data processing
Type
File
suffix
Destination
APD
C:\M7\app
EXD
C:\M7\exe
GPD
Data in
Signal definition of testing points
Compress the update files and check if there are
C:\M7\fpga
VDD
same directory data in the hard disk, if yes, then
C:\M7\video
GUD
delete the old directory data, and then create new
C:\M7\gui
directory number.
the hard
IMD
C:\M7\Image
disk
PST
C:\M7\Preset
FKY
C:\M7\Key
Decompress the update files and check if there
PCD
are same directory data in the hard disk, if yes,
C:\M7\config,
delete it and create new directory data, and then
C:\WINDOWS\boot.bmp
configure the Windows desktop.
Copy the Demo data in the resume package to
Demo
the target directory
E:\M7\Demo
Including DSP drive, multifunctional board drive,
USB keyboard drive, Smbus drive, decompress
Drive
DRD
the update files to the temporary directory in the
hard disk, and install the drives respectively, then
delete the temporary directories when finished the
Installed the
corresponding drives
C:\windows\system32\
installation.
Including DSP FPGA, multifunctional FPGA,
beam former FPGA, transmitting FPGA,
FPGA
FPD
decompress the update files to the temporary
The corresponding
directory of the hard disk, and write data in their
configurations of the
configured flash through the corresponding drive
logic flash
program. Delete the temporary directory when
finished.
Keyboard
KUG
Analyze the update files and cache the keyboard
data into buffer, and then check keyboard
5-3
Keyboard board
System and Software
Type
File
suffix
Signal definition of testing points
Destination
communication, and E/W keyboard flash.
Analyze the update files and save the ECG data
ECG
ECG
into buffer, and then check ECG module
ECG module
communication, and E/WECG flash.
Analyze the update files and decompress the
Power
monitor
ARM
ARM data into the temporary directory of the hard
disk, check the ARM communication and E/W,
Power monitor module
delete the temporary directories when finished.
5.3
NOTE:
Run the Resume Package in Doppler
1. Before executing the resume package, please back up the patient data as well as the
user-defined preset data.
2. If the resume is performed with a disk in Doppler, then the disk inserted can’t be
recognized when the ultrasound machine is running, in this case, please shut
down the system first, then insert the disk, and then restart the system. The disk
can be recognized, and resume can be performed.
Operation procedures:
1. Copy the resume package to the USB disk;
2. After system boot-up, login as service engineer;
3. Press <Setup> to open the Setup menu, select [Maintenance] to open the Maintenance menu.
4. Select [Enter Windows], and you can see all the drive letters;
5. Found the run.bat file in the USB disk, and double click it. The following information appears:
SUCCESS: The process "appmon.exe" with PID XXXX has been terminated.
SUCCESS: The process "doppler.exe" with PID XXXX has been terminated
Press any key to continue…
Press any key to continue, and the recovery operation will be processed automatically.
6. During the recovery, the system prompts “Do you want to delete user preset data? (Y/N)”, input
“Y” (note: you must select “Y”, because it is possible that the preset data may not be
compatible, the new preset data have to be used.)
7. After the recovery is completed, the system prompts “Press any key to continue…”, inspect if all
the items have been finished successfully, and press <Enter> according to the prompt, the
DOS window will be closed automatically.
8. Now, delete the PATIENTDATA file in the path D:/M7/PATIENTDATA (as the patient data have
been backed up already);
9. Shut down the system and then reboot the system;
10. Check the version, which should be consistent with that of the newly updated.
NOTE:
After the system is restored, please check if the system model and regulation type are
correct (the same as before restoration)
5-4
System and Software
5.4
Power ON Process
The power ON process will last for 20 seconds, refer to the table below for the details.
Table 5-3 Power ON process
The possible
Step
Operation
Increments
Step
causes when
initialization
dead
Initialize Windows; Appmon initialize Doppler; Bind the path
of the configuration files; Set Windows properties; Initialize
the display devices, the main window, sound, USB devices
Configure
agency; Activate bus devices BackboneDev and LPC;
No prompt
Initialize time and multi-lingual process functions; Initialize
peripheral file system, network, drive, enumerate peripheral
device port; Configure the timers, initialize the soft
In increments of 7.
interruption, construct maintenance server, configure
the
Pull out the
hardware
CD/DVD
and XPE
drive if any
system
system static data; Configure system fonts; Load layout
information; Initialize UI manager and UI graph libraries.
Initializing
Load-on the boot-strap bitmap, display the progress bar (7),
Boot-strap
hardware…
decompress the configured factory data
bitmap
/
Loading
system
Create local preset and system preset server
In increments of 1.
preset…
May be the
Loading
common
exam
preset data is
Create exam mode general data manager
In increments of 1.
Configure
not
preset data
compatible,
preset…
Loading
exam
preset...
delete the
preset data
Create measure preset, peripheral devices and network,
image KMP package, network storage preset servers.
In increments of 4.
Set the region related information, the language, font
library, entry method; Create control “factory”; Construct
Initializing
local…
GUI layer; Set the menus; Initialize function libraries; Create
UICenter
5-5
In increments of 1.
Configure
HDD data
/
System and Software
The possible
Step
Operation
Increments
Step
causes when
initialization
dead
Initializing
gui…
Initializing
ultrasound
peripheral…
Construct the access management, remote
desktop
Configure the file dialog box; Initialize print library; Battery
monitor, 2118 system monitor; Load-on print task icon;
Construct
low consumption; Construct USB manager and
initialize the virtual machine; Create front-end/back-end
image…
object tree (ultrasys, etc.)
Initialization
completed…
In increments of 1.
record manager; Initialize video replay devices
ultrasound
application…
attention to
the
/
keyboard
Create ECG debt collection thread; Set virtual machine,
ultrasound
In increments of 1.
management
Initializing
Initializing
Pay
Construct the application layer; Initialize keyboard board;
Configure
Occasional
peripheral
system crash
device, pay
was found,
attention to
not resolved
the battery
yet.
Occasional
In increments of 2.
Pod data
system
crash, not
resolved yet
Add measure menu functional package; Register measure
patient application interface
In increments of 1.
/
Patient Info
Construct patient info manager (UPatientApp)
In increments of 1.
/
Including
operation on
Hide the
boot-strap
/
/
bitmap
/
the LCD,
may lead to
system crash
5.5
Software Maintenance
Before performing the software maintenance, ensure the data used matches with the model.
5.5.1
Entering Preset
Press <Setup> key, pops up the preset menu.
5-6
System and Software
5.5.2
Viewing System Information
In the Setup menu, select [About], and the system information dialog box appears.
Be sure to confirm the system information before and after the software maintenance.
5.5.3
Manage Settings
1. Press <Setup> key to enter into the preset status.
2. Move the cursor onto [Manage Settings] and press [Set].
5-7
System and Software
3. Select Export (Import or Restore factory) as required.
C
D
E
Tips:
Code in Figure
Function Description
A
Export the selected presets, and save them in different files
B
Import the preset parameters, or recover the factory settings
C
Export all preset parameters, and save them in a file
5-8
System and Software
Code in Figure
Function Description
D
Import all preset data at one time
E
Recover factory settings for all preset data
5.5.4
Maintenance Status
Copy the files to be used into a USB flash drive. Then insert the USB flash drive into the USB port
on the main unit.
1. Enter the Preset menu.
2. Select “Maintenance”.
Note: to enter the Maintenance menu, please login the system with the user name of Service.
3. Select “Maintenance”
5-9
System and Software
5.5.5
Model Setup
1. Enter into maintenance status. Select “Config”, and press <Set>.Enter into the Config menu:
2. Select the correct files, and click [OK]. After the operation is completed, the successful prompt
appears. Restart the system according to the prompt.
Note: the default configuration is that of M7, and it is already finished during the resume
package configuration. For M7T, the product information has to be configured according to
the configuration files.
5.5.6
Installing Optional Software
1. Enter the Setup menu.
2. Click [System] on the Setup menu to open the System Preset screen.
3. In the System Preset dialog box, select [Option] to open the dialogue box shown in the figure
below.
5-10
System and Software
4. In the optional list, select the desire software function, and click “Install”.
5. In the opened dialog box, select the corresponding key file, and click [OK].
The key file is corresponded to the system one by one, and you can copy it to the flash drive, and
install through the flash drive;
6. After the installation is completed, return to the system setup screen, and the corresponding
options shall be in the installed status. Click [OK] to close the System Preset dialog box.
Note:
1. The key file needed for software installation is copied in the system accompanying CD.
2. After ultrasound software recovery, the optional software should be reinstalled.
5.6
Warning
If malfunctions appeared during the system operation, the ultrasound machine M7 will give out
warnings, or generate LOGs which will be saved in system logs, the logs is saved in the path: D:
\M7\Log\Perilog.
5.6.1
Battery warning
5.6.1.1
Battery Connect & Disconnect
5-11
System and Software
Operation
Warning
LOG
Left battery hot
“Warning! Invalid battery
“2009-9-7,2:43:11 U2118Monitor: Left Battery
plug
operation!May lead to
Hot Plug”
battery damage”
Right battery hot
“Warning! Invalid battery
“2009-9-7,2:43:11 U2118Monitor: Right Battery
plug
operation! May lead to
Hot Plug”
battery damage”
Left battery hot
None
“2009-9-7,2:43:11 U2118Monitor: Left Battery
insert
Hot Insert”
Right battery hot
None
“2009-9-7,2:43:11 U2118Monitor: Right Battery
insert
Hot Insert”
Left Battery Hot
None
“2009-9-11,2:43:11 U2118Monitor: Left Battery
Plug, Shutdown
Hot Plug, Shutdown State”
State
Right Battery Hot
None
“2009-9-11,2:43:11 U2118Monitor: Right Battery
Plug, Shutdown
Hot Plug, Shutdown State”
State
5.6.1.2
Battery I2C Exception
Operation
Warning
LOG
Adapter power
Left battery communication I2C
“2009-9-7,2:55:11 U2118Monitor: Left
supply, light
error, battery capacity can’t
Battery Communication I2C error”
battery I2C
correctly displayed, or current
exception
battery can’t be used
Adapter power
Right battery communication I2C
“2009-9-7,2:55:18 U2118Monitor: Right
supply, right
error, battery capacity can’t
Battery Communication I2C error”
battery I2C
correctly displayed, or current
exception
battery can’t be used
Battery power
Left battery communication I2C
“2009-9-7,3:11:13 U2118Monitor: Left
supply, left battery
error, please insert the adapter
Battery Communication I2C error”
I2C exception<
or power off the system,
otherwise, the system will be
shut down after XX seconds!
5-12
System and Software
Battery power
Right battery communication I2C
“2009-9-7,3:11:19 U2118Monitor: Right
supply, Right
error, please insert the adapter
Battery Communication I2C error”
battery I2C
or power off the system,
exception<
otherwise, the system will be
shut down after XX seconds!
Left Battery I2C
None
“2009-9-11,3:11:19 U2118Monitor:
Exception,
Left
Battery I2C Exception, Shutdown State”
Shutdown State
Right battery I2C
None
“2009-9-11,3:11:19 U2118Monitor: Right
exception,
Battery I2C Exception, Shutdown State”
Shutdown state
5.6.1.3
Battery Cycle Count beyond the Limitation
Operation
Warning
Log
Left battery cycle count is
Battery life will do,
“Left Battery Cycle count is: 65535,Right
beyond the limitation
replace it as soon as
Battery Cycle count is: 16”
possible.
Beyond the number of
Battery life will do,
“Left Battery Cycle count is: 16,Right
the right battery
replace it as soon as
Battery Cycle count is: 65535”
recycling times
possible.
5.6.1.4
Battery Volt Exception
Operation
Warning
Log
Left battery Volt
Left battery voltage is low,
“2009-9-7,3:11:19 U2118Monitor: Left
Exception
please insert the adapter or
Battery voltage is low: left battery volt is:
power off the system,
12.16V”
otherwise, the system will be
shut down after XX seconds!
Right battery Volt
Right battery voltage is low,
“2009-9-7,3:11:19 U2118Monitor: Right
Exception
please insert the adapter or
Battery voltage is low, Right battery volt is:
power off the system,
12.15V”
otherwise, the system will be
shut down after XX seconds!
5-13
System and Software
Left battery volt
None
“2009-9-9,3:11:19 U2118Monitor: Left
exception, shutdown
Battery Volt Exception, Shutdown State”
state
Right battery
volt
None
“2009-9-9,3:11:19 U2118Monitor: Right
exception, shutdown
Battery Volt Exception, Shutdown State”
state
5.6.1.5
Battery Current Exception
Operation
Warning
Log
Battery current
The current difference of the
“2009-9-7,3:45:19 U2118Monitor: The
exception
two batteries is out of range,
difference in battery discharge current is out of
please insert the adapter or
range, Left battery current is: -2.604A, Right
power off the system,
battery current is: -2.663A”
otherwise, the system will
be shut down after XX
seconds!
Right battery
None
“2009-9-9,3:45:19 U2118Monitor: Left Battery
current exception,
Current Exception, Shutdown State”
shutdown state
“2009-9-9,3:45:19 U2118Monitor: Right Battery
Current Exception, Shutdown State”
5.6.1.6
Battery Temperature Exception
Operation
Warning
Log
Left Battery temperature
Left Battery temperature is out
“2009-9-7,3:49:19 U2118Monitor: Left
is out of range
of range, please insert the
Battery temperature is out of range,
adapter or power off the
left battery temperature is: 36.0
system; otherwise, the system
centigrade”
will be shut down after XX
seconds!
Right battery temperature
Right battery temperature is
“2009-9-7,3:49:19 U2118Monitor:
is out of range
out of range, please insert the
Right Battery temperature is out of
adapter or power off the
range, right
system; otherwise, the system
31.5 centigrade”
will be shut down after XX
5-14
battery temperature is:
System and Software
seconds!
None
Left battery temperature
“2009-9-9,3:49:19 U2118Monitor: Left
is out of range, shutdown
Battery Temperature Exception,
state
Shutdown State”
Right battery temperature
None
“2009-9-9,3:49:19 U2118Monitor:
is out of range, shutdown
Right Battery Temperature Exception,
state
Shutdown State”
5.6.1.7
Battery Break
Operation
Warning
Log
Left battery break
Battery exception! Battery
“2009-9-7,3:49:19
charge/discharge can’t be proceeded!
U2118Monitor:Left Battery Break”
Battery exception! Battery
“2009-9-7,3:49:19
charge/discharge can’t be proceeded!
U2118Monitor:Right Battery Break”
Right battery break
5.6.1.8
Low Battery Power
Operation
Warning
Log
Wait until the battery
Battery power is low, please insert
“2009-9-7,3:49:19 U2118Monitor:Low
power is lower than
the adapter or shutdown the
battery power, Left battery power
6%
system, otherwise, the system will
is:5%,Right battery power is: 7%”
be shut down after XX seconds!
5.6.2
PHV Exception Warning Verification
5.6.2.1
95V High Volt Exception
Exception
Warning
Log
+95V error
None
“2009-9-8,2:43:11 UBatMonitor::95VPower supply voltage
error:%.2fV,normal region is 90V to 100V
-95V error
None
“2009-9-8,2:43:55 UBatMonitor::-95VPower supply voltage
error:%.2fV,normal region is 90V to 100V”
5-15
System and Software
5.6.2.2
Other PHV Warning
Exception
Warning
Log
PHV1+
Warning! Transmitting high volt
“2009-9-8,2:43:11
error, image can’t be displayed in
UBatMonitor::PHV1+Power supply voltage
the right way.
error: 62.93V(upper limit is %.2fV)”
Warning! Transmitting high volt
“2009-9-8,2:43:55 UBatMonitor::PHV1-Power
error, image can’t be displayed in
supply voltage error:62.66V(upper limit
the right way.
is %.2fV)”
Warning! Transmitting high volt
“2009-9-8,2:43:55
error, image can’t be displayed in
UBatMonitor::PHV2+Power supply voltage
the right way.
error:5.08V(upper limit is %.2fV)”
Warning! Transmitting high volt
“2009-9-8,2:43:55 UBatMonitor::PHV2-Power
error, image can’t be displayed in
supply voltage error:5.06V(upper limit
the right way.
is %.2fV”
PHV1-
PHV2+
PHV2-
5.6.3
Fan Warning
Stop working
Warning
Log
Left front fan
Left front fan has gone off! Please
ADT7462: Left Front Fan has gone off!
contact the service engineer
Left middle
Left middle fan has gone off! Please
fan
contact the service engineer
Left back fan
Left back fan has gone off! Please
ADT7462: Left Middle Fan has gone off!
ADT7462: Left Back Fan has gone off!
contact the service engineer
Right fan
Right fan has gone off! Please
ADT7462: Right Fan has gone off!
contact the service engineer
CPU fan
CPU fan has gone off! Please
contact the service engineer
The position of the fans is shown as the figure below:
5-16
ADT7462: CPU Fan has gone off!
System and Software
Fig 5-2 The position of the fans
5.6.4
Temperature Warning
Temp protection
Warning
Log
Main board Temperature
None
ADT7462: Main Board Temperature Alert
None
ADT7462: Power Board Temperature Alert
CP temperature is beyond
Temperature warning,
CPU Temperature too high, ShutDown,
105 ℃
XX countdown
CPU Temperature =
alert
Power board temperature
alert
5.6.5
Volt Warning
Volt
Warning
LOG
12V
None
ADT7462:12V Power supply voltage:%f, VoltLimtFloat: %f--%f
N12V
None
ADT7462:-12V Power supply voltage:%f, VoltLimtFloat: %f--%f
5V1
None
ADT7462:5.1V Power supply voltage:%f, VoltLimtFloat: %f--%f
3V3
None
ADT7462:3.3V Power supply voltage:%f, VoltLimtFloat: %f--%f
1V5
None
ADT7462:1.5V Power supply voltage:%f, VoltLimtFloat: %f--%f
2V8
None
ADT7462:2.8V Power supply voltage:%f, VoltLimtFloat: %f--%f
1V8
None
ADT7462:1.8V Power supply voltage:%f, VoltLimtFloat: %f--%f
1V2
None
ADT7462:1.2V Power supply voltage:%f, VoltLimtFloat: %f--%f
5-17
Function and Performance Checking Method
6
Function and Performance
Checking Method
6.1
Note
The chapter supplies the detailed method for product main function and performance checking.
This is used for referring or studying by engineer but not required.
6.2
System Running Status
6.2.1
Running Status
1. Power on/off normal (duration time is normal), no abnormal sounds or phenomena occur during
normal operation.
2. After ultrasound system is turned on, the fan starts working and no abnormal sound when the
fan is working.
3. Check if configuration, software version are normal through the [About] in preset menu.
4. Check if contrast and brightness of the monitor are normal.
5. Check if time and date are valid and correct.
6. Check if all status indicators are normal.
7. Check all log records with user, to confirm if there is any abnormality.
6.2.2
Working Condition
Check the ambient temperature and humidity. The measurements related to safety features are
particularly sensitive to humidity. If the insulation feature of the system deteriorates due to the
increase of system service time or system malfunctions, the fluctuation range of measurement
results are likely to increase with the increase of humidity.
6-1
Function and Performance Checking Method
6.3
General exam
6.3.1
Check Flow
Check control panel
Check monitor
Check DVD-R/W
Check peripherals
Check ECG module
Check I/O ports
6.3.2
Checking Content
6.3.2.1
Check Control Panel
Procedure
Checking standard
check all buttons, keys and knobs
Follow the direction: left to right, and up to down.
Function checking of the trackball:
„
Press the <Freeze> key to enter the Freeze status.
„
Press <Measure> to enter into measure status, do
vertical and horizontal measurement, or do other
trackball operations.
6.3.2.2
Check the Monitor
6-2
All keys and knobs are effective.
The trackball can be rotated easily; the
cursor responds sensitively, the rotation
direction is the same as the direction of
the cursor.
Function and Performance Checking Method
Procedure
„
Standard
Adjust LCD brightness
„
Enter [Preset]->[ Peripheral
preset], On the “Input & display”
page:
Press 『Fn』and 『
press 『Fn』and 『
„
Brightness load factory values.
„
Color temperature of LCD will be changed
correspondingly.
„
Click each functional button, the LCD responds
correctly, the standard is as follows:
„
click [default]
„
Click “cold/warm”
„
Monitor maintenance
Log on as Service, click
[Maintenance]->[ Main Monitor] to
check the monitor functions
』, the brightness increases; and
』, the brightness decreases.
1. Light-spot: 0; flash point: 0.
2. The adjoining dark spots are no more than 3 pairs, and
there is no adjoining dark spot in image area.
3. There is no adjoining dark spot of 3 or more than 3.
4. The dark spots are no more than 7 and those in the
image area are no more than 2
5. The distance between bad spots is no less than 5mm.
Note: image area refers to rectangle when the background is
black/right.
6.3.2.3
Check DVD-R/W
Procedure
Standard
„
Press [Eject]
„
Disk can be normally ejected.
„
Use the optical disk drive to read and burning.
„
Normal, no abnormal sounds.
6.3.2.4
Check Peripherals
Procedure
Standard
Footswitch:
„
Connect the footswitch; check the
functions of footswitch according to the
functions listed in Key Config. (e.g. right
key- image frozen, middle key- color print,
left key- B/W print)
Press the freeze key (the right key), image is
frozen, the freeze menu is displayed; press the
key again, image is unfrozen.
„
Press the print key (middle key), color printing
starts.
„
Press the print key (left key), B/W printing starts.
Video printer:
Check if the video printer and ultrasound
system are correctly connected. Then
check the function of each key.
Press <Print> key, the printer begins to work, no
image print deficiency or degradation.
Switch video output port; repeat the step.
Text/graph printer:
Check if the printer and ultrasound system
are correctly connected.
Press <Print> key, the printer begins to work, no print
deficiency or degradation.
Then check the function of each key.
6.3.2.5
Check ECG Module
6-3
Function and Performance Checking Method
Procedure
Confirm if the
configured, then:
ECG
Standard
module
Set [ECG] as “On”.
6.3.2.6
is
„
ECG trace is displayed; the heart icon is displayed
at the upper right corner of the screen.
„
The parameters [Speed], [ECG Gain] can be
adjusted.
„
ECG signal can be reviewed correctly.
Check I/O Ports
Procedure
Standard
Checking the main I/O ports:
Besides the video port, USB ports, the other
ports required to be checked including:
„
The contents displayed on the VGA/LCD are
the same as those displayed on the ultrasound
system displayer, no character and image loss,
no color difference, no fluttering and flicking.
„
Smooth communication.
„ S-video port;
„ Connect external VGA/LCD monitor
Network port.
6.4
NOTE:
Function Checking
A complete function inspection is described here, do the checking according to
the actual system configuration.
6-4
Function and Performance Checking Method
6.4.1
Checking Flow
6.4.2
Content
6.4.2.1
Imaging Modes
1. B mode
„
„
In B Mode scanning, the image parameter area in the upper left corner of the screen will display
the real-time parameter values as follows:
Display
F 5.0
D 17.6
G 50
FR 40
IP 4
DR 60
Parameter
Frequency
Depth
Gain
Frame Rate
B IP
B Dynamic Range
Parameters that can be adjusted to optimize the B Mode image are indicated in the following.
Adjustment
Control Panel
Items
Gain, Depth, TGC, iTouch
6-5
Function and Performance Checking Method
Adjustment
Items
Dynamic Range, Focus Number, FOV Position, Line Density, IP, Colorize,
L/R Flip, Rotation, Persistence, Colorize Map, U/D Flip, iTouch,
Frequency, Gray Map, Focus Position, iClear, FOV, Smooth, TSI, Curve,
Gray Rejection, γ, High FR, iTouch Bright, A. power, B Steer, iBeam,
Trapezoid, Image Merge
Menu and Soft Menu
1. Control Panel
Procedure
Standard
Press <B>
Enter B mode, and B image displays
Gain adjustment.
Rotate clockwise to increase
Rotate <iTouch>
Rotate anticlockwise to decrease
Depth Adjustment
Press <Depth/> direction
key.
TGC adjustment
Adjust
through
8-segment toggles
the
The image depth will be changed.
The adjustable depth values vary depending upon the probe types.
Push the toggle to the right to increase the corresponding area
brightness
Push the toggle to the left to decrease the corresponding area brightness
About 1.5s after the adjustment is finished, the TGC curve disappears.
iTouch
Press <iTouch>
iTouch sign will be displayed in the image area to activate image effect
auto optimization.
Image Magnification
Roll the trackball to change position of the magnified image.
Press <zoom> to light on
the Zoom indicator.
Press <Zoom> in zoom status to exit the mode, the current window
returns to the display before zoom.
Configured with image menu “Magnify” to realize different Magnification.
Press <B>, and then
z
Enter dual mode, and the right image is activated
press
z
Press again to switch between the windows
Press <B>, and then
z
Press 4 times, and 4 images will be displayed on the screen with
only one image activated at one time
z
Press again to switch among the windows
press
Press <B>
To enter single mode in multiple window mode, or to exit from other
modes.
2. Menu
Procedure
Standard
Frequency
Frequency value is displayed in real time in the left upper part of the screen
via adjusting.
B mode menuÆ
[Frequency]
Focus
B
mode
menuÆ
The frequency scales of different probes are not the same.
Focus number or position can be adjusted. The focus position symbol is
displayed on the right side of the image.
6-6
Function and Performance Checking Method
[Focus
Number]/[Focus
Position]
Dynamic Range
B mode menu->
[Dynamic Range]
to the dynamic range can be adjusted, the real time value will be displayed on
the image parameter area in the upper right of the screen. As the dynamic
range increases, the darker the image and the contrast, as well as the noise
may increase.
Line Density
Click [Line Density], and rotate the multifunction knob to adjust the parameter.
B mode menu->
[Line Density]
The higher the line density, the higher the resolution, and the lower the frame
rate.
iClear
Click [iClear] to adjust, off represents no iClear is turned on, and the bigger
the value the stronger the effect. The bigger the value the more clearly the
profile of the image.
B mode
[iClear]
menu->
B mode menu->
[U/D Flip]/[L/R Flip]
B mode
[Rotate]
menu->
B mode menu-> [U/D Flip]/[L/R Flip]
When you flip or rotate an image, the “M” mark will change its position on the
screen; the M mark is located in the upper left corner of the imaging area by
default.
2. M mode
„
In M mode scanning, the image parameter area in the upper left corner of the screen displays
the real-time parameter values as follows:
Display
V3
IP 6
DR 65
G 45
Parameter
M Speed
M IP
M Dynamic Range
M Gain
„
Parameters that can be adjusted to optimize the M mode image are indicated in the following.
Adjustment
Items
Control Panel
Gain, TGC, Depth
Menu and Soft Menu
IP, Time Mark, Speed, Colorize, Colorize Map, Acoustic Power, Edge
Enhance, Frequency, Gray Map, Focus Position, Dynamic Range, M
Soften, Curve, Gray Rejection, γ, Display Format
3. Color mode
„
In Color mode scanning, the image parameter area in the upper left corner of the screen
displays the real-time parameter values as follows:
Display
F 3.3
G38
IP 1
WF 43
PRF 0.4k
Parameter
Frequency
Color
Gain
Color IP
Color Wall
Filter
Pulse Repetition Frequency
PRF
6-7
Function and Performance Checking Method
„
Parameters that can be adjusted to optimize the Color mode image are indicated in the
following.
Adjustment
Items
Control Panel
Gain, Depth
Menu and Soft Menu
B/C Wide, Frequency, Priority, Baseline, Packet Size, Dual Live, Color IP,
Map, A. power, Focus Position, Scale, Flow State, WF, Invert, Line
Density, Smooth, Persistence, B Display, Steer
4. Power mode
„
In Power mode scanning, the image parameter area in the upper left corner of the screen
displays the real-time parameter values as follows:
Display
F 2.5
G 38
IP 1
WF32
PRF 0.3k
Parameter
Frequency
Power
Gain
Power
IP
Power Wall
Filter
Pulse Repetition Frequency
PRF
Parameters that can be adjusted to optimize the Power mode image are indicated in the following.
Type
Parameter
Control Panel
Gain, TGC, Depth
Menu and Soft Menu
B/C Wide, Frequency, Priority, Dynamic Range, Packet Size, Dual Live,
Power IP, Map, Acoustic Power, Focus Position, Scale, Flow State, Wall
Filter, Invert, Line Density, Smooth, Persistence, B Display, Steer
5. PW/CW mode
In PW/ CW mode scanning, the image parameter area in the upper left corner of the screen
displays the real-time parameter values as follows:
PW
Display
F 2.5
G 60
PRF 3.6k
WF295
SVD 8.8
SV 3.0
Frequency
Gain
Pulse
Repetition
Frequency PRF
WF
(Wall
Filter)
SV
Position
SV Size
F 3.3
G 30
WF407
PRF 7.3k
SVD 35.0
Frequency
Gain
WF (Wall
Filter)
Pulse Repetition
Frequency PRF
SV Position
Parameters
CW
Display
Parameters
6-8
Function and Performance Checking Method
Parameters that can be adjusted to optimize the PW/ CW mode image are indicated in the
following.
Adjustment
Items
Control Panel
Gain, TGC, Depth
PW
Invert, SV, WF, Frequency, Duplex/Triplex, V Max, V Mean, Audio, Trace
Area, Colorize Map, Speed, Acoustic Power, Trace Sensitivity, T/F Res,
Gray Map, Dynamic Range, Scale, Baseline, Colorize, Quick Angle,
Angle, Trace Smooth, Time Mark, HPRF, Curve, Gray Rejection, γ,
Display Format, Auto Calculation, Auto Calculation Parameter, PW Steer
CW
WF, Invert, Colorize Map, Frequency, Speed, V Max, V Mean, Dynamic
Range, Audio, Trace Area, Trace Sensitivity, Gray Map, Scale, Baseline,
Colorize, Quick Angle, Angle, Acoustic Power, Trace Smooth, Time Mark,
T/F Res, Curve, Gray Rejection, γ, Display Format, Auto Calculation, Auto
Calculation Parameter
Soft Menu &
Menu
6.4.2.2
Basic Measurements
Procedure
Standard
In B image mode:
The system enters general measurement mode.
Press <Measure>
Perform any 1-2 measurements (e.g., length, area), the
results will display at the right lower part of the image.
Press the same key again or press
<Esc>.
Exits measurement.
Do the same operation in other
image modes.
Application measurements are classified into different
application packages, do the application measurements
selectively.
6.4.2.3
Cine Review
Procedure
Standard
Press [Freeze] key to freeze an image, and the
[Cine] key indicator lights on. The system
automatically enters the manual cine status.(It has
been set that when system enters into freeze mode,
the default status is cine review.)
The system enters into cine review status
The system enters into auto cine review
status.
Press <iStation> key, then click [Review]; or press
<Review> key to open a cine file.
Roll the trackball
Manual cine review
Click [Auto Play] on the menu or soft menu.
Auto play function is turned on, adjust the
soft menu button.
The greater the value is, the quicker the
speed is.
When the value is 0, the system exits auto
play mode.
Move the cursor onto the desired start point of the
cine loop, click [Set First Frame] in the menu or soft
menu to set the start point.
6-9
Set the start point of cine loop.
Function and Performance Checking Method
Move the cursor onto the desired end point of the
cine loop, click [Set Last Frame] in the menu or soft
menu to set the start point.
Set the end point of cine loop.
Click [Auto Play] again
Review region is confined to the set start
point and end point.
Then press the [Cine] key again.
Cine review stops.
Press the <Freeze> key to unfreeze the image.
Freeze indicator light is off; the system will
return to image scanning and exit cine
review.
Press <Cine> or <Esc> key.
The images are still frozen but the system
exits cine review.
6.4.2.4
Probe Switching
Procedure
Standard
Press <Freeze> key→ connect the probe to the
system→ press <Freeze> key→ press <Probe>
key to select the probe.
Connect a convex probe to probe socket A, and
then connect a linear probe to probe socket B,
the operator can select probe A or probe B as the
active probe.
Press <Freeze> key→ disconnect the probe→
connect another probe to the port
The system can recognize the newly connected
probe in no time.
6.4.2.5
Patient Information Management
Procedure
Standard
Press <Save> (the function already set)
during image scanning
Image will be saved to the patient database, and a
thumbnail will be displayed at the right part of the
screen.
Open
[Setup]
→[System
Preset]→”General”, then check “Send/Print
Image after End Exam” in the Patient
Management area.
The system automatically sends the images of the
exam to the default DICOM storage server or print
server.
Press <End Exam> during image scanning
z Press <Review> key.
z Click [Exit] on the Review screen; or,
press <Review> again, or, press
<Esc> key
z The system enters into image review mode.
z Click [iVision] on the other menu; or
z Open iVision screen:
z The system exits image review mode.
6-10
Function and Performance Checking Method
press user-defined iVision key.
z Select the contents to be demonstrated,
and select the demo mode in the
iVision screen. Then select an item
already added to the list and click
[Start]
z Demonstration begins.
z Image files are played according to file names one
by one (including the image of system-relevant
and PC-compatible format).
z The system exits the demonstration.
z When the demonstration is finished, click
[Exit] or press <Esc>.
Press <iStation> key to enter patient
information management (iStation page)
The saved patient information (images) can be found,
and the patient information can be:
z
z
6.5
Performance Test
6.5.1
Test Process
6.5.2
Test Content
NOTE:
Backed up/ Restored
Sent (To DICOM.U disk .etc )
The image used here is only for reference, stick to the image effect in the real
situation.
Requirements:
1.
Display: set the contrast and brightness at the clinical application value (or the default status)
2.
Operation environment: dark room, simulating the clinical application environment.
3.
Scanning techniques: contact the probe with the acoustic window of the phantom, no spacing
nor pressing.
6-11
Function and Performance Checking Method
Tips:
For the testing phantoms, please refer to Appendix B.
KS107BD is low frequency phantom and used when Probe focus frequency is less than 4MHZ;
KS107BG is high frequency phantom and used when Probe focus frequency is more than
5MHZ;
6.5.2.1
Resolution
„ transverse resolution
Test Step:
1.
Cover the scan surface of the phantom with water or couple gel, gently contact the probe with
the scan surface, making the transverse resolution testing targets to be displayed around the
midline of the image.
2.
Adjust the focus point focuses at the position where the transverse resolution testing targets
are displayed.
3.
Adjust parameters like gain, dynamic range, TGC, making the background tissue unseen, just
displaying the target image clearly.
4.
In condition that the transverse resolution testing targets are horizontally displayed, record the
minimal distance of two targets that can be clearly recognized.
5.
Repeat the operation above for the transverse resolution testing targets at other depths.
As shown in figure below.
„ Axial resolution
Test Step:
1.
Cover the scan surface of the phantom with water or couple gel, gently contact the probe with
the scan surface, making the longitudinal resolution testing targets to be displayed around the
midline of the image.
2.
Adjust the focus point focuses at the position where the longitudinal resolution testing targets
6-12
Function and Performance Checking Method
are displayed.
3.
Adjust parameters like gain, dynamic range, TGC, making the background tissue unseen, just
displaying the target image clearly.
4.
Record the minimal distance of two longitudinal resolution testing targets that can be clearly
recognized.
5.
Repeat the operation above for the longitudinal resolution testing targets at other depths.
NOTE:
1.
2.
6.5.2.2
When using the convex probe, keep the transverse resolution testing targets to
be displayed near the midline.
When using a linear probe with steer function, do not turn on the steer function
when perform the transverse resolution test.
3.
Zoom in the region where the targets located if necessary.
4.
The diameter of the target point at a certain depth is equal to the transverse
resolution at the depth.
Maximum Depth
Test Step:
1.
Cover the scan surface of the phantom with water or couple gel, gently contact the probe with
the scan surface
2.
Set the system display depth according to the eXPcted maximum available depth of the probe
in use.
3.
Adjust the focus point to the deepest, and AP at the maximum value.
4.
Set gain, contrast, TGC at a greater value, but no halation nor defocus is allowed.
5.
Record the depth of the furthest target (the target can be seen clearly).
NOTE:
1.
Increasing the gain will also increase the noise, and echo may be covered.
6-13
Function and Performance Checking Method
2.
3.
4.
When using a linear probe, please completely contact the probe with the scan
surface, no side clearance is allowed.
When using a convex or phased-array probe, make the axis targets to be
displayed at the middle of the scanning image.
When system is not frozen, the fast field target information may be similar to that
of the noise, do not use this target.
As shown in figure below.
6.5.2.3
„
Geometric positioning accuracy
Longitudinal geometric positioning accuracy
Test Step:
1.
Do adjustments as the way in testing the maximum depth.
2.
Record the distance by 20mm each segment on the longitudinal targets line using the
6-14
Function and Performance Checking Method
measurement caliper;
3.
Select the value with the greatest error (to 20mm), calculate the accuracy using the formula
below
NOTE:
1.
2.
The measurement caliper should be positioned at the upper edge of the
target, not the middle nor the lower edge.
The scanning plane should be vertical to the target line, that means the
scanning plane is parallel with the cross-section of the phantom
As shown in figure below.
„ Transverse geometric positioning accuracy
Test Step:
1.
Cover the scan surface of the phantom with water or couple gel, gently contact the probe with
the scan surface
2.
Adjust the depth, making the transverse targets to be displayed in the image.
3.
Adjust the focus point to be posited beside the transverse targets (the standard is not clear)
4.
Adjust parameters like gain, TGC, making each transverse targets to be clearly displayed.
5.
Record the distance by 20mm each segment on the transverse targets line by using the
measurement caliper
6.
Select the value with the greatest error (to 20mm), calculate the accuracy by using the formula
below
6-15
Function and Performance Checking Method
NOTE:
1.
When using a linear probe, record the transverse distance by segment.
2.
When using a convex probe, all transverse targets should be displayed
integrally in an image.
The measure caliper should be posited at the upper side or lower side of
the target center.
3.
As shown in figure below.
6.5.2.4
Blackout Area
Test Step:
1.
Cover the scan surface of the phantom with water or couple gel, gently contact the probe with
the scan surface
2.
Adjust the depth at a lower value, and set the focus at the nearest place to the scan surface.
3.
Decrease the value of parameters like AP, Gain until the background noise just can be seen.
4.
Record the smallest depth of the target that can be seen clearly, that value is the blackout area
value.
NOTE:
1.
2.
When using a linear probe, please completely contact the probe with the
scan surface, no side clearance is allowed.
For convex probe, the targets in the blackout area should be positioned on
the midline of the scanning plane.
As shown in figure below.
6-16
Function and Performance Checking Method
6-17
7
7.1
System Self-diagnosis
Screen Introduction
The booting screen of the self-diagnosis system is:
Fig 7-1 Interface of System Self-Diagnosis and Boot
The main interface of self-diagnosis system
There are 3 kinds of main interfaces of self-diagnosis system: manufacturing, customer service and
user, as shown in figures below:
The self-diagnosis interface of manufacturing and customer service can be divided into 4 regions.
7-1
System Self-diagnosis
Fig 7-2 Self-diagnosis Interface, Manufacturing
System structure diagram
Test item & result
Test info display &
Statu
control
s bar
Fig 7-3 Self-diagnosis Interface, Customer Service
7-2
System Self-diagnosis
Fig 7-4 Self-diagnosis Interface, User
Test Report Save Location: display the removable storage media (such as USB disk, removable
HDD), if no removable storage media is connected; or, if the system is undergoing the test, this
selection is not available.
Save Report: save the tested data to the root of the specified logic disk, if no logic disk is specified;
or, if the system is undergoing the test, this selection is not available.
Start/Stop: Start/Stop the test;
Shut Down: shut down the program, and shut down the system automatically.
Interface of tested results
Displays the test results, not available for the user.
7-3
System Self-diagnosis
Fig 7-5 Interface of testing results
Test Report Save Location: display the removable storage media (such as USB disk, removable
HDD), if no removable storage media is connected; or, if the system is undergoing the test, this
selection is not available.
Save Report: save the tested data to the root of the specified logic disk, if no logic disk is specified;
or, if the system is undergoing the test, this selection is not available.
Shut Down: shut down the program, and shut down the system automatically.
7.2
Region of Self-diagnosis Items and
Testing Results
Configuration of testing items
In this region, the testing items are divided into parent items according to boards and there are
several sub items for each parent item. The testing items can be customized according to
requirements. If the check box before an item is selected ( ), this item is in the testing list;
contrariwise, if the check box before the item is deselected ( ), this item is not in the testing list.
When a parent item is selected, all the sub items of this item are selected by default; contrariwise,
when deselecting a parent item, all the sub items of this item are deselected by default.
7-4
System Self-diagnosis
OutDentAll: click this button to unfold all test items if they are folded;
InDentAll: click this button to fold all test items if they are unfolded;
Default: click this button to select all the items to be tested by default;
SelectAll: click this button to select all test items;
SelectNone: click this button to deselect all test items selected.
Display of Testing Results
The testing result is displayed behind each sub item. For the sub item, if an item passes the test, it
is marked with green “PASS”; if not, it is marked with red “FAIL”; if the item is not selected, it is
marked with “Skip” of normal color. See Fig 8-11.
Fig 7-6 Test Results of Items in Testitems Table
Test Result Navigation
After the test is completed, click the test items in Testitems, the program will show the
corresponding detailed test results, which will be displayed in the first line of Message.
7.2.1
Info Display Area
Test information
The area displays test results.
Messages list: display test information of each test item.
Index is the number of test, namely the index number corresponding sub items;
TestResult is the name and detailed test information of test items, as shown in Fig 8-12.
7-5
System Self-diagnosis
Fig 7-7 Interface of Testing Results
Test Configuration
Onfail drop-down list: there are two options, “Continue” and “Stop”. If you select [Continue], the test
will continue even though a test item fails during testing; however, if you select [Stop], the test will
stop once a test item fails during testing.
Start button: click to test the selected items one by one, and the button turns to [Stop] at the same
time. During the test, click [Stop] to stop the test. After the test is completed, the [Stop] button turns
to [Start], ready for the next test.
Next Fail: after the test is completed, click [Next Fail] to show the next failed test item in the
Message list, and the related information will be displayed in the first line. If already navigated to the
last line, then the query will begin from the first line.
Clear History: click to clear all data in Testitems and Messages list, and recover the flow chart to the
initial status.
Loops check box: check this item to perform loop test. If it is checked and the loop test number is
input at the right side, the test will be performed as many as the number you set. If the number is
less than 0, the test will be cycled, until it was manually stopped or any error appeared.
Quit: for user operation, click the button to exit the self-diagnosis and shut down the system; for
manufacturer and service engineer operation, click this button to exit the test
interface and then to
view the tested results.
7.2.2
Status Bar
When the program is running, the version and issued date will be displayed on the left side of the
status bar: M7 SelfTest Software for Manufacture Version: xx; Release Date: YYYYMMDDXX.
During the test, the software version in the status bar will turn to the name of the item on test.
7-6
System Self-diagnosis
During the test, the test progress of current item and the total test progress are also displayed in the
status bar.
Besides, the following items are also displayed in the status bar: temperature of CPU, main board
and power module, occupation rate of CPU and memory, remaining test time and current time.
Fig 7-8 Sketch Map of Status Bar
When the temperature goes beyond the working temperature, the value will be displayed in red.
The normal range is listed below:
7.2.3
Temperature type
Lower limitation
Upper limitation
CPU temperature
0
100
Main board temperature
-10
78
Power board temperature
-10
70
Test Report
Save the test report
Non-loop test
For the non-loop test, the test result will be saved in D:\M7\log\Selftest\Test Report. The result of the
former test will be deleted after the new test is completed.
Loop test
For the loop test, the test result will be saved in D:\M7\log\Selftest\Test Report Loop_N. The test
result is PASS or FAIL, N refers to the test times.
Meanwhile, after each test is finished, the current test result will be automatically compressed into
Loop Test Reports.zip, and the test data will be deleted.
The test report is in the format of HTML, which can be opened by IE browser by default. The test
report is shown as below:
7-7
System Self-diagnosis
Fig 7-9 Test Report
The left side is the test item list, which is classified according to the test results, besides, system
information is listed here; the right side are the tested data and pictures.
Click the test item in the left side, the corresponding test data will be displayed.
The test data consists of:
z
Report Name
z
Report time;
z
The picture of the system structure (for user test, the picture displayed is the system
structure in the factory)
z
System Information;
z
Test data of failed tests (if there is no failed test, then no this branch);
z
Test data of passed tests (if there is no passed test, then no this branch);
z
Test data of overleaped tests (if there is no overleaped test, then no this branch);
The following is an example of the test data:
z
Index and name of the test item;
7-8
System Self-diagnosis
z
Board/module;
z
Test result (PASS, FAIL, and Skip, each result is displayed in different color);
z
Message
z
Error log (may not exist);
z
Add on data (exist only when the LCD Monitor Adjust Data Check Test is failed).
Notes when open the test report:
When double click M7 Selftest Report.html using IE (Java Script is applied), it may be blocked,
which can be seen in the figure below.
Fig 7-10 Internet Explorer Prompt
Right click the prompt, then click “Allow Blocked Content” (see the figure below), the report can be
opened by IE in the normal way.
Fig 7-11 Display the Blocked Content
7.3
Description of Self-diagnosis Test Items
(1)PC Module Heat Sink Test
„
Test Content
Test if the PC module heat sink is installed in the right way.
„
Test Description
The following screen will be opened after the test is started. There displays the current CPU
working temperature, the yellow line represents the highest working temperature of the CPU, and
the changing wave represents the current CPU temperature. If the temperature didn’t go beyond
the limit, the waveform is displayed in green; otherwise, it is red. At the bottom of the dialogue box,
there displays the current CPU temperature and the remaining test time.
The interface can’t be closed manually after the test is started, until the test is completed. If the test
failed, please turn off the power as soon as possible.
7-9
System Self-diagnosis
Fig 7-12 PC module heat sink test data display
Test results will be displayed in the Remark column of the Messages list box, and the format is:
PC Module heat sink test PASS/FAIL.
Suggestion when test failed:
Check if PC module heat sink was correctly installed, if not, reinstalled it.
(2)PC Module and DSP Interconnection Test
„
Test Content
Test if the PCle communication between PC module and DSP FPGA is normal.
„
Test Description
Test results and DSP FPGA related logic version will be displayed in the Remark column of the
Messages list box, and the format is:
PC Module and DSP FPGA interconnection test PASS/FAIL;
DSP Version:XX;
Build_Version:X;
Release time:XXXX.XX.XX.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U6
Main board
U8
(3)Frame Process SSRAM Test
7-10
System Self-diagnosis
„
Test Content
Test if there is address bus disconnected, short connected, pull high, pull low, and if the memory is
damaged in SSRAM.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Frame Process SSRAM data bus test PASS/FAIL;
Frame Process SSRAM address bus stuck high test PASS/FAIL;
Frame Process SSRAM address bus stuck low test PASS/FAIL;
Frame Process SSRAM address bus short test PASS/FAIL;
Frame Process SSRAM full space test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U6
Main board
U7
(4)IQ Frame Buffer DDRII1/2 Test
„
Test Content
Test if IQ frame cache DDR II works normally
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
IQ frame buffer DDRII1/2 test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U3
Main board
U4
(5)Data Upload and Parameter Store DDRII Test
7-11
System Self-diagnosis
„
Test Content
Test if data uploading and parameter storage DDR II are normal.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Data Upload and Parameter Store DDRII test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U5
(6)BF and DSP Interconnection Test(Data Interface)
„
Test Content
Test if the RF data bus between BF FPGA and DSP FPGA work normally.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
BF and DSP Interconnection Test (Data Interface) test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U6
Main board
U10
(7)BF and DSP Interconnection Test(Control Interface)
„
Test Content
Test if the control bus between BF FPGA and DSP FPGA work normally.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
BF and DSP Interconnection Test (Control Interface) test PASS/FAIL;
BF logic version: 0xXX;
BF logic build version: 0xXXX;
7-12
System Self-diagnosis
BF logic release time: XXXX.XX.XX.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U6
Main board
U10
(8)BF and DSP Interconnection Test(INT Interface)
„
Test Content
Test if the interruption bus between BF FPGA and DSP FPGA work normally.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
BF and DSP Interconnection Test (INT Interface) test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U6
Main board
U10
(9)BF and DSP Interconnection Test(Rate Interface)
„
Test Content
Test the Rate signal from BF FPGA, check if DSP FPGA and TX FPG can receive normally, and if
the Rate related connections and logics can work normally.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
BF and DSP Interconnection Test (Rate Interface) test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
7-13
System Self-diagnosis
Board
Connector
Main board
U6
Main board
U10
(10)AFE Digital Interface Test
„
Test Content
Verify if the 8 analog front-end digital interfaces are normal.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
AFE ADC digital Interface test PASS/FAIL;
AFE ADC0 digital Interface test PASS/FAIL;
AFE ADC1 digital Interface test PASS/FAIL;
AFE ADC2 digital Interface test PASS/FAIL;
AFE ADC3 digital Interface test PASS/FAIL;
AFE ADC4 digital Interface test PASS/FAIL;
AFE ADC5 digital Interface test PASS/FAIL;
AFE ADC6 digital Interface test PASS/FAIL;
AFE ADC7 digital Interface test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Board
Connector
Main board
U10
Main board
U47
Main board
U46
Main board
U48
Main board
U52
Main board
U51
Main board
U50
Main board
U49
Main board
U1
(11)AFE Noise Test
„
Test Content
Test if the front-end noise data is in the specified range when the analog front-end is under normal
working.
7-14
System Self-diagnosis
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
AFE noise test PASS. Or,
AFE noise test PASS.
Broken channel: CHX…….
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Board
Connector
Main board
U10
Main board
U47
Main board
U46
Main board
U48
Main board
U52
Main board
U51
Main board
U50
Main board
U49
Main board
U1
(12)ATGC Function Test
„
Test Content
Test if ATGC functions normally.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
ATGC function test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Board
Connector
Main board
U25
Main board
U1
Main board
U10
Main board
U47
Main board
U46
Main board
U48
Main board
U52
Main board
U51
Main board
U50
Main board
U49
(13)Transmission and Reception Function Test
7-15
System Self-diagnosis
„
Test Content
Test if the circuits related with transmitting, receiving, high-voltage isolation and high-voltage
switching are normal.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Transmission and Reception function test PASS ; or,
Transmission and Reception function test FAIL;
Open circuit emit receive channel: CHX……;
Short circuit emit receive channel: CHX……;
Broken high voltage switch channel: CHX…….
(14)PC Module and System Monitor Interconnection Test
„
Test Content
Test if SMBUS between the PC module and system monitor chip is correctly connected.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
PC module and system monitor interconnection test PASS/FAIL;
Device ID: 0xXX;
Company ID: 0xXX;
Revision Number: 0xXX.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U90
(15)Fan Speed Monitor Test
„
Test Content
It tests whether the 5 fans inside the system work normally.
„
Test Description
Test results of fan rotational speed will be displayed in the Remark column of the Messages list box,
and the format is:
7-16
System Self-diagnosis
CPU Fan Test PASS/FAIL.
Left Fan(Near Speaker) Test PASS/FAIL.
Left Fan(Middle) Test PASS/FAIL.
Left Fan(Near PC Module) Test PASS/FAIL.
Right Fan Test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U90
(16)Main Board and Power Module Temperature Test
„
Test Content
Test if the main board and power module temperatures are normal.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Main Board temperature test PASS/FAIL(XX.XX C);
Power Board temperature test PASS/FAIL(XX.XX C).
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U90
Main board
Q1
Main power board
Q17
(17)Main Board Voltage Test
„
Test Content
Test if main board voltage is normal.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
12V power voltage test PASS/FAIL (XX. XX V);
N12V power voltage test PASS/FAIL (XX. XX V);
7-17
System Self-diagnosis
VCC power voltage test PASS/FAIL (XX. XX V);
VDD power voltage test PASS/FAIL (XX. XX V);
1V5L power voltage test PASS/FAIL (XX. XX V);
D2V8 power voltage test PASS/FAIL (XX. XX V);
D1V8 power voltage test PASS/FAIL (XX. XX V);
VCCINT power voltage test PASS/FAIL (XX. XX V).
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U90
(18)PC Module and MF Interconnection Test
„
Test Content
Test if PC module and MF interconnection is normal.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, for the main board
with the ID as 0x33 (manufacture)/0x32 (development), the test result is:
PC Module and Multifunction FPGA interconnection test PASS/FAIL;
Multifunction FPGA DDR test PASS/FAIL;
Main Board ID: XX.XX.XX.
And for other main boards, the test result is:
PC Module and Multifunction FPGA interconnection test PASS/FAIL;
Main Board ID: XX.XX.XX.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
(19)USB Interface Test
„
Test Content
Test if the four USB Interfaces on the machine and the I/O extend module work normally.
7-18
System Self-diagnosis
„
Test Description
When started the self-diagnosis, please insert the 4 USB test clamps to the 4 USB ports. If no test
clamp is connected, the system pops up the following dialog box.
Fig 7-13 Prompt for connecting USB port test clamp
Test results will be displayed in the Remark column of the Messages list box, and the format is:
USB port test PASS
Or,
USB port test FAIL;
USBX…….
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
J1
Main board
J13
IO extend module
J1
IO extend module
J6
IO extend module
J7
(20)Svideo Interface Test
„
Test Content
It tests whether the function of Svideo Out in the IO extend board is normal.
„
Test Description
A display with Svideo interface should be connected to the system during the test.
7-19
System Self-diagnosis
During the test, the program will output some display content of LCD display to the display device,
and pop up a dialog box as shown in the following figure to ask if you can see an image displayed
on the display device.
For the main board with ID as 0x33/0x32, the test result is:
The first step: the self-diagnosis program will display the figure as figure below, which is used for
testing if S-video output is normal, the output image is PAL-format, with resolution as 800×600.
Fig 7-14 PAL-formatted image test (800×600, S-video port)
3 seconds later, a dialogue box pops up as shown in figure below, asking if you can see the image
on the monitor. The figure as in figure above is displayed, if correct, click “Yes, I can”; otherwise,
click “No,I can’t”.
Fig 7-15 PAL-formatted image test dialog box
The second step: the self-diagnosis program will display the figure as shown in figure below, which
is used for testing if S-video output is normal.
The image format is NTSC, with resolution as 800×600.
7-20
System Self-diagnosis
Fig 7-16 NTSC-formatted image test (800×600)
3 seconds later, a dialogue box pops up as shown in figure below, asking if you can see the image
on the monitor. The figure as in figure above is displayed, if correct, click “Yes, I can”; otherwise,
click “No,I can’t”.
Fig 7-17 NTSC-formatted image test dialog box
The first step: the self-diagnosis program will display the figure as figure below, which is used for
testing if S-video output is normal, the output image is NTSC-format, with resolution as 640×480.
7-21
System Self-diagnosis
Fig 7-18 NTSC-formatted image test (640×480)
3 seconds later, a dialogue box pops up as shown in Fig 7-17, asking if you can see the image on
the monitor. The figure as in figure above is displayed, if correct, click “Yes, I can”; otherwise, click
“No,I can’t”.
The first step: the self-diagnosis program will display the figure as figure below, which is used for
testing if S-video output is normal, the output image is PAL-formatted, with resolution as 640×480.
Fig 7-19 PAL-formatted image test (640×480)
3 seconds later, a dialogue box pops up as shown in Fig 7-15, asking if you can see the image on
the monitor. The figure as in figure above is displayed, if correct, click “Yes, I can”; otherwise, click
“No,I can’t”.
7-22
System Self-diagnosis
And for main board with ID as 0x33/0x32, the test result is:
Fig 7-20 S-video Interface Test Dialog Box
If you can see it, click “Yes, I can”; otherwise, click “No, I Can’t”.
Test results will be displayed in the Remark column of the Messages list box, and the format is:
S-video interface test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
Main board
U86
Main board
J28
(21)Video Interface Test
„
Test Content
It tests whether the function of Video Out in the I/O extend board is normal.
„
Test Description
A display with Video interface should be connected to the system during the test.
During the test, the program will output some display content of LCD display to the display device,
and pop up a dialog box as shown in the following figure to ask if you can see an image displayed
on the monitor.
For the main board with ID as 0x33/0x32, the test result is:
The first step: the self-diagnosis program will display the figure as Fig 7-14, which is used for testing
if Video output is normal, the output image is PAL-format, with resolution as 800×600.
3 seconds later, a dialogue box pops up as shown in figure below, asking if you can see the image
on the monitor. The figure as in Fig 7-14 is displayed, if correct, click “Yes, I can”; otherwise, click
“No,I can’t”.
7-23
System Self-diagnosis
Fig 7-21 PAL-formatted image test dialog box (Video)
The first step: the self-diagnosis program will display the figure as Fig 7-14, which is used for testing
if video output is normal, the output image is PAL-format, with resolution as 800×600.
3 seconds later, a dialogue box pops up as shown in figure below, asking if you can see the image
on the monitor. The figure as in Fig 7-16 is displayed, if correct, click “Yes, I can”; otherwise, click
“No,I can’t”.
Fig 7-22 NTSC-formatted image test dialog box (Video)
The third step: the self-diagnosis program will display the figure as Fig 7-18, which is used for
testing if Video output is normal, the output image is NTSC-format, with resolution as 640×480.
3 seconds later, a dialogue box pops up as shown in figure above, asking if you can see the image
on the monitor. The figure as in Fig 7-18 is displayed, if correct, click “Yes, I can”; otherwise, click
“No,I can’t”.
The first step: the self-diagnosis program will display the figure as Fig 7-19, which is used for testing
if S-video output is normal, the output image is PAL-formatted, with resolution as 640×480.
3 seconds later, a dialogue box pops up as shown in Fig 7-21, asking if you can see the image on
the monitor. The figure as in Fig 7-19 is displayed, if correct, click “Yes, I can”; otherwise, click “No,
I can’t”.
And for other main boards, the test result is:
7-24
System Self-diagnosis
Fig 7-23 Video Interface Test Dialog Box
If you can see it, click “Yes, I can”; otherwise, click “No, I Can’t”.
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Video interface test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
Main board
U86
Main board
J1
IO extend board
J1
IO extend board
J5
(22)DVI Interface Test
„
Test Content
It tests whether the function of DVI Out in the IO extend board is normal.
„
Test Description
A display with DVI interface should be connected to the system during the test.
During the test, the program will output some display content of LCD display to the display device,
and pop up a dialog box as shown in the following figure to ask if you can see an image displayed
on the display device.
7-25
System Self-diagnosis
Fig 7-24 DVI Interface Test Dialog Box
If you can see it, click “Yes, I can”; otherwise, click “No, I Can’t”.
Test results will be displayed in the Remark column of the Messages list box, and the format is:
DVI interface test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
Main board
U76
Main board
J1
IO extend board
J1
IO extend board
J10
(23)VGA Interface Test
„
Test Content
It tests whether the function of VGA Out in the IO extend board is normal.
„
Test Description
A display with VGA interface should be connected to the system during the test.
During the test, the program will output some display content of LCD display to the display device,
and pop up a dialog box as shown in the following figure to ask if you can see an image displayed
on the display device.
7-26
System Self-diagnosis
Fig 7-25 VGA Interface Test Dialog Box
If you can see it, click “Yes, I can”; otherwise, click “No, I Can’t”.
Test results will be displayed in the Remark column of the Messages list box, and the format is:
VGA interface test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
Main board
U80
IO extend board
J1
IO extend board
J1
IO extend board
J10
(24)Serial Port Test
„
Test Content
It tests whether the function of the serial port of the IO extend board is normal.
„
Test Description
Short connect pin 2 and pin 3.
Test results of serial port will be displayed in the Remark column of the Messages list box, and the
format is:
Serial port test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
7-27
System Self-diagnosis
Board
Connector
Main board
U75
Main board
U65
Main board
J1
IO extend board
J1
IO extend board
J2
(25)Microphone Interface Test
„
Test Content
It tests whether the function of audio chip and MIC port of the IO extend board are normal.
„
Test Description
During the test the program shows the following dialog box prompting the user to connect a
microphone to the MIC port of the product and speak towards the microphone. The program will
detect the sound wave automatically and estimate whether the related circuits of the MIC port work
normally.
If sound wave is detected, or any hardware error appeared, the test interface will be shut down in 3
seconds, and no manual operation is required.
If you can see the sound wave, click “OK”. If not, click “Cancel”.
Fig 7-26 Audio Test Dialog Box
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Microphone interface test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
7-28
System Self-diagnosis
Board
Connector
Main board
U85
Main board
J1
IO extend board
J1
IO extend board
J8
(26)Audio Test
„
Test Content
It tests whether the function of audio chip, power amplification circuit and speaker are normal.
„
Test Description
During the test, the left side of machine will play out “Ding~~”, and pops up the following dialog box
to ask if you can hear “Ding~~” playing out of the left side machine. If you didn’t hear it, please click
“Retry” to play it again, and click “No, I Can’t”, if you still can’t hear it after trying many times. And if
you heard it, please click “Yes, I Can”.
Fig 7-27 Audio Test Dialog Box
The program will control the system right side buzzer to give off “ding…”, and popup the dialog box
as shown above. The tests for right and left side buzzers are the same.
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Left Speaker Test PASS/FAIL;
Right Speaker Test PASS/FAIL;
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Board
Connector
Main board
U85
Main board
U83
7-29
System Self-diagnosis
Main board
J1
Main board
U84
IO extend board
J1
Main board
J27
IO extend board
J3
IO extend board
J4
Speaker
(27)Transducer Interface Test
„
Test Content
Test if probe presenting can be recognized, and if communication is normal.
„
Test Description
The test flow is shown as follows, please perform the operation according to the workflow strictly.
Start selfdiagnosis
Check if probe and probe extend module are
present?
Yes
No
Tips: connect the
probe
Program
judge
Yes
No
Probe present
Probe extend module
present
Tips: disconnect the
probe
Tips: disconnect the probe extend module,
connect the probe
Program
judge
No
No
Program
judge
Yes
Yes
Get probe info
Tips: connect the
probe
Get probe info
Tips: disconnect the
probe
Get probe info
Tips: disconnect the
probe
Program
judge
End
Fig 7-28 Transducer interface test flow
7-30
System Self-diagnosis
Get probe info
Read probe ID
Query probe type
Tip: query probe model?
End
Fig 7-29 Transducer information obtaining flow
The dialog boxes for connecting/disconnecting the transducer, disconnecting the transducer extend
module are shown in the figures below respectively.
Fig 7-30 Prompt of Connecting Transducer
Fig 7-31 Prompt of Disconnecting Transducer
7-31
System Self-diagnosis
Fig 7-32 Prompt of Disconnecting Transducer Extend Module and Connecting Transducer
Fig 7-33 Dialogue Box Asking for Transducer Model
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Transducer present test PASS/FAIL;
Transducer interface SPI interconnection test PASS;
Transducer Type: XXXX. Or,
Transducer present test PASS/FAIL;
Transducer interface SPI interconnection test FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U6
Main board
U10
Main board
J8
(28)Transmission Board and Main Board Interconnection Test
„
Test Content
Test if the interconnection between TXFPGA and DSPFPGA work normally.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Transmission Board and Main Board interconnection test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
7-32
System Self-diagnosis
Board
Connector
Main board
U6
Main board
J5
Main board
J6
Transmitting board
U3
Transmitting board
J2
Transmitting board
J3
(29)Transmission Board ID and Logic Version Test
„
Test Content
Read TX FPGA logic version, logic compilation version, and the compiled time.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
TX FPGA logic version:0xXX;
TX FPGA logic build version:0xXXX;
TX FPGA logic release time: XXXX.XX.XX.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U6
Main board
J5
Main board
J6
Transmitting board
U3
Transmitting board
J2
Transmitting board
J3
(30)MF and Power Module Interconnection Test
„
Test Content
Test if communication serial port between MF FPGA and power ARM works normally.
7-33
System Self-diagnosis
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Power Board ID:XX;
Power Board firmware version: X.X.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
Main board
U70
Main board
J22
Power board
J1
Power board
U15
(31)Power Module Function Test
„
Test Content
Test if ECG module functions normally.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
+95V power voltage test PASS/FAIL (+XX.XX V);
-95V power voltage test PASS/FAIL (-XX.XX V).
CW Mode PHV control test PASS/FAIL;
PW Mode PHV control test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
Main board
U70
Main board
J22
Power board
J1
7-34
System Self-diagnosis
Power board
U15
Power board
U1
Power board
U11
(32)CW Board ID Test(unfinished)
(33)Demodulator and Phase Shifter SPI BUS Function Test(unfinished)
(34)CW Board AD Noise Test(unfinished)
(35)4D Driver Board ID Test
„
Test Content
Test if 4D drive board is presenting, and if the devices that record the IDs are normally working.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
4D Driver Board logic version: 0xXX;
4D Driver Board board ID:XX.XX.XX.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U6
Main board
J4
4D Driving Board
U10
4D Driving Board
P2
(36)4D Driver Board Driver Circuit Test
„
Test Content
Test if the driving circuit of 4D driver board works normally.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
4D Driver Board driver circuit test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
7-35
System Self-diagnosis
Board
Connector
Main board
U6
Main board
J4
4D Driving Board
P2
4D Driving Board
U4
4D Driving Board
U6
4D Driving Board
U9
4D Driving Board
U13
4D Driving Board
U14
(37)4D Driver Board Programmable Voltage Test(unfinished)
(38)Left/Right Battery I2C Interconnection Test
„
Test Content
Checks if the I2C communication bus between the batteries and the power management ARM
works normally.
„
Test Description
Test results of left/right battery I2C connection will be displayed in the Remark column of the
Messages list box, and the format is:
Left/Right Battery I2C connection test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
Main board
U70
Main board
J22
Power board
J1
Power board
U15
Battery
(39)Left/Right Battery Test
7-36
System Self-diagnosis
„
Test Content
It tests whether the battery works normally.
„
Test Description
Test results of left/right battery will be displayed in the Remark column of the Messages list box, and
the format is:
Left/Right Battery is on site;
Left/Right Battery voltage is: XXXXmV.
Left/Right Battery temperature is: XX.X C.
Left/Right Battery current is: XXXXmA.
Left/Right Battery capacity is: XX%.
Left/Right Battery charge and discharge time is: XXXXX.
Left/Right Battery Safety Status test PASS/FAIL.
Left/Right Battery Permanent Failure Status test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
Main board
U70
Main board
J22
Power board
J1
Power board
U15
Battery
(40)LCD Monitor ID Test
„
Test Content
Read LCD displayed information.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
LCD code: 0xXXXXXXXX;
LCD serial number: 0xXXXXXXXX;
LCD version: 0xXX.
7-37
System Self-diagnosis
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
(41)LCD Monitor Brightness Control Test
„
Test Content
It tests whether the adjustment function of LCD backlight is normal.
„
Test Description
During the test, the self-check system will decrease the brightness of the LCD to the lowest and
then increase it from the lowest to the highest, and after that the brightness will be decrease from
the highest to current value. After this procedure, the following dialog box appears to ask the user
whether he/she saw the procedure. If yes, click “Yes, I Can”, otherwise click “No, I Can’t”. If you
click “Retry” the system will repeat the procedure above.
Fig 7-34 LCD Brightness Test Dialog Box
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
(42)LCD Monitor Backlight Control Test
„
Test Content
It tests whether the switch function of LCD backlight is normal.
7-38
System Self-diagnosis
„
Test Description
During the test, the self-diagnosis system will switch off the LCD backlight and then switch it on.
After this procedure, the following dialog box appears to ask the user whether he/she saw the
procedure. If yes, click “Yes, I Can”, otherwise click “No, I Can’t”. If you click “Retry” the system will
repeat the procedure above.
Fig 7-35 LCD Backlight Switch Test Dialog Box
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
(43)LCD Monitor Adjust Data Check Test
„
Test Content
Test if LCD monitor adjust data is correct.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
LCD Monitoradjustdata check test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Main board
U75
Main board
J24
LCD parameter board
U1
LCD parameter board
J1
7-39
System Self-diagnosis
(44)Keyboard USB Interconnection Test
„
Test Content
Test if keyboard and USB port on the main unit can communicate normally.
„
Test Description
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Keyboard self test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Keyboard board;
U4
(45)Keyboard Key, Encoder, TGC, Trackball Test
„
Test Content
It mainly tests whether the buttons, encoders, TGCs, trackball, backlight LEDs and indicating LEDs
in the keyboard are normal, and reads board ID, keyboard FPGA and version of program.
„
Test Description
The keyboard testing interface is shown as follows:
7-40
System Self-diagnosis
Fig 7-36 Interface of Keyboard Test
As shown in the figure above, click [Start] to start the keyboard test. The controls in the simulation
keyboard are corresponding to those of the real keyboard assembly.
As for the buttons, press a button in the real keyboard and the corresponding button in the
simulation keyboard blinks, which means the button is being tested. Changing in color means the
button has being tested.
As for the trackball, roll the trackball in the keyboard and the cursor in the interface moves
accordingly and the
in the simulation keyboard blinks, which means the system has
received data from the trackball and the trackball has being tested.
The encoder rotation test: the
on the simulation keyboard corresponds to the encoder
on the real keyboard. When you rotate the encoder in a direction, the corresponding green ball
around the encoder on the simulation keyboard will rotate in the same direction. The encoder face
will flicker and turn green.
7-41
System Self-diagnosis
The encoder key test: when you press the encoder key, the encoder key on the simulation keyboard
will change status accordingly, and flicker and turn green.
TGC test: 8 TGC sliders on the keyboard correspond to
controls on the
simulation keyboard. When sliding TGC sliders on the keyboard, the corresponding controls on the
simulation keyboard will move as well.
Keyboard backlight test: when the simulation keyboard initializes, all dual-color lights will highlight in
green on the keyboard. When a key is being tested, the dual-color backlight will automatically turn
orange after it is pressed.
After all tests for the keyboard assembly finished, the keyboard screen is shown in Fig 7-37
Keyboard Test Screen after the Test
.
Fig 7-37 Keyboard Test Screen after the Test
After the tests, if functions of the keyboard assembly are normal, click “Keyboard is OK”; otherwise,
click “Keyboard is Bad”.
7-42
System Self-diagnosis
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Keyboard Test PASS/FAIL;
Keyboard ID: XX.XX.XX;
Keyboard FPGA Version: XX.XX;
Application Version: XX.XX.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Key board board
U4
Key board board
U1
Key board board
U6
Key board board
U30
(46)Keyboard Key and Trackball LED Test
„
Test Content
It tests whether the button LEDs and trackball LED in the keyboard work normally.
„
Test Description
The following dialog box appears when testing. Change the brightness of the white and orange
LEDs respectively and check whether the brightness of LEDs on the keyboard change accordingly.
Change the color of the trackball LED and check whether the LED on the keyboard changes
accordingly. If the brightness of the keyboard LEDs and color of the trackball LED are the same as
configured, click “LED is OK and Quit” to quit the dialog box. If not, click “LED is not OK and Quit” to
quit the dialog box.
Fig 7-38 Interface of Keyboard LED Test
7-43
System Self-diagnosis
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Keyboard Key and Trackball LED Test PASS/FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally
Board
Connector
Key board board
U4
Key board board
U1
Key board board
U6
(47)Extend Module Test
„
Test Content
Test if the transducer extend module present can be recognized.
„
Test Description
The test flow is shown as follows, please perform the operation according to the workflow strictly.
7-44
System Self-diagnosis
Start the selfdiagnosis
No
Probe present?
Yes
Tips: disconnect the
probe
Probe extend module
present?
Yes
No
The number of the probes?
Tips: connect probe extend module (no probe)
Yes
>0
0
Tips: disconnect probe extend
module
No
Tips: disconnect probe and probe
extend module
Program judge 2?
Yes
Initialize the probe
No
Program judge 1?
No
Probe extend module
not present ?
Read probe extend
module info
Tips: connect probe extend module (no probe)
No
Program judge 2?
Probe extend module present
No
Probe recognize test flow
Initialize the probe
Tips: disconnect probe extend
module
Probe extend module
present?
No
No
Read probe extend
module info
No
Probe recognize test flow
End
Fig 7-39 Extend Module Test Flow
7-45
Program judge 2?
Probe extend module not
present?
System Self-diagnosis
Fig 7-40 Transducer Present Recognize Test Flow
The dialog boxes asking for disconnecting the transducer, connecting/disconnecting the transducer
extend module are shown in the figures below respectively.
Fig 7-41 Prompt of Disconnecting Transducer
7-46
System Self-diagnosis
Fig 7-42 Prompt of Connecting Transducer Extend Module
Fig 7-43 Prompt of Disconnecting Transducer Extend Module
Fig 7-44 Prompt for Disconnecting Transducer on Extend Module, and Then Disconnect Extend
Module
Fig 7-45 Prompt of Connecting General Transducer
7-47
System Self-diagnosis
Fig 7-46 Prompt of Connecting Transducer
Fig 7-47 Prompt of Asking if the User Connected XXXX Transducer on X Socket
Test results will be displayed in the Remark column of the Messages list box, and the format is:
Transducer Extend Module present test PASS;
Extend Module SPI interconnection test PASS;
Extend Module transducer recognition test PASS;
Board ID: 0xXXXX;
Logic version: 0xXXXX.
Or,
Transducer Extend Module present test FAIL;
Extend Module SPI interconnection test FAIL.
Extend Module transducer recognition test FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally.
Board
Connector
Main board
U6
7-48
System Self-diagnosis
Board
Connector
Main board
U10
Transducer extend module
(48)ECG Module Information Read Test
„
Test Content
Verify if ECG module communicates well with the main unit, read Bootloader version, board ID and
software version.
„
Test Description
After tested is started, if ECG is not presented, then a dialog box is opened to ask the user to
connect ECG module, see the figure below:
Fig 7-48 Ask the User to Connect ECG Module
Test results will be displayed in the Remark column of the Messages list box, and the format is:
ECG Module information read test PASS;
Bootloader version: XX.XX;
Board ID: %d.XX.XX;
Software version: XX.XX.
Or,
ECG Module information read test FAIL.
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally.
Board
Connector
Main board
U6
Main board
U75
7-49
System Self-diagnosis
Main board
J1
IO extend module
J1
IO extend module
J11
ECG module
(49)ECG Module Self Test
„
Test Content
Test if ECG module functions normally.
„
Test Description
After tested is started, if ECG is not presented, then a dialog box is opened to ask the user to
connect ECG module, see the figure below:
Fig 7-49 Ask the User to Connect ECG Module
Test results will be displayed in the Remark column of the Messages list box, and the format is:
ECG Module self test PASS.
Or,
ECG Module self test FAIL;
XXX,XXX,XXX……
„
Suggestion when test failed:
Check if the following listed hardware and the related circuits are normal, check if the interfaces are
well welded and if the module is working normally.
Board
Connector
Main board
U6
Main board
U75
Main board
J1
7-50
System Self-diagnosis
IO extend module
J1
IO extend module
J11
ECG extend module
(50)Pencil Probe Present Test(unfinished)
(51)Pencil Probe ID Test(unfinished)
7-51
8
8.1
Care and Maintenance
Overview
The following procedures are recommended.
8.1.1
Tools, Measurement Devices and Consumables
Table 8-1 Tools and Measurement Devices
Tool/Measurement Devices
Qty.
Remarks
Resin or plastic container
1 pcs
Can accommodate two probes
Soft brush
1 pcs
About a toothbrush size
Small plastic basin
1 pcs
Used to fill the soapy water
Safety test analyzer
1 pcs
Refer to appendix A
Inner hexagon wrench
2 pcs
Inner hexagon wrench 6
Table 8-2 Consumable List
Consumable
Qty.
Remarks
Aluminum foil
About 1 meter
Physiological saline
About 1000ml
Filling a half container Immerging
the whole probe (referring to
appendix A).
(concentration 0.85 ~ 0.95%)
Mild soapy water
About 400ml
Dry soft cloth/cotton cloth
About 5 pcs
8-1
Care and Maintenance
8.1.2
Care and Maintenance Items
Table 8-3 Maintenance Items and Frequency
NO.
Maintain content
Frequency
Method
1.
Clean display
Monthly
Referring to 8.2.1
2.
Clean trackball
Monthly
Same as the above
3.
Clean control panel
Monthly
Same as the above
4.
Clean probes (the head)
Every
using
5.
Clean probe cable and the surface of
connector
Monthly
Same as the above
6.
Clean holders (including probe holder
and gel holder)
Monthly
Same as the above
7.
Clean cover
Monthly
Same as the above
8.
Clean peripherals
Monthly
Referring to 8.2.2
9.
Check surface of probe
Daily
Referring to 8.3.1
10.
Check power cable and plug
Monthly
Same as the above
11.
Check battery
Annually
Same as the above
12.
Check function of peripherals and
options
Annually
Referring to 8.3.3
13.
Mechanical safety inspection
Annually
Referring to8.3.4
14.
Electrical safety inspection
Annually
Referring to appendix A
8-2
time
after
Same as the above
Care and Maintenance
8.2
Cleaning
8.2.1
Clean the System
8.2.1.1
Flow of Cleaning
Power off and pull out the
power cables
Clean display
Clean trackball
Clean control panel
Clean probe
Clean cover
Figure 8-1 View of Cleaning Maintenance
WARNING:
8.2.2
Before cleaning the system, be sure to turn off the power and
disconnect the power cord from the outlet. If you clean the system
while the power is “On”, it may result in electric shock.
Content
1.
Clean Display
z
Tool: soft dry cloth ,clear water or mild soapy water
z
Method:
Surface of display should be cleaned with soft dry cloth. Remained stain should be washed out
by cloth with clear water or soapy water, and then air-dry the surface.
2.
Clean Trackball
z
Paper or dry cloth
z
Method:
Trackball is one of important interface parts, which are embedded into the main unit keyboard, and
part of which is exploded to be operated by users in order to implement variety of module’s control
8-3
Care and Maintenance
functions. Trackball is one of the most using frequency of input assemblies on the whole operation
panel, and the trackball similar to a multi-directory caster can rotate in every direction drived
manually, due to the feature, gas or dust can enter into module internal easily, as a result,
contamination of lens would lead to the failure of the trackball.
a)
Disassembling the trackball:
Turn the trackball ring about 35°counterclockwise until it lifts, now, you would remove the ring
and pull out the trackball with plastic cloth if you can’t hold it by your hands directly. Disassembling
the trackball is as following:
Ring
Trackball
Turn trackball ring 35°to the left
Disassemble the ring
Remove the trackball
Figure 8-2 Disassembly of the Trackball
a)
Cleaning
After removing the ring and the trackball, wipe down the lens with a clean paper until you can
see nothing in the groove zoom, and then clean the other contaminant material , please pay
attention to your intensity adopted on wiping dust of bead down, as shown in the following figure.
Power-off is not need during cleaning, and maintenance effect can be experienced directly. After
cleaning completely, you can install the trackball and the ring.
Lens zoom
3 beads
Dragon groove
Figure 8-3 The Sketch Map of Len, Bead, Dragon groove
If liquid is accidentally sprayed on or into the system, most of which could discharge from the
dragon port of trackball, but some of which would left in the trackball cover, now you may clean it
with clean soft dry cloth or paper according to the above maintenance procedure.
b)
Installing the trackball
After the trackball maintenance, you can restore the installing following procedure: Put the
trackball back in the trackball mechanism and align the clamping ring with the top cover notches.
Press the bulges on the ring with both hands and turn the ring about 35° counterclockwise until the
ring clicks and locks. As the bulges are flush with the top cover, the ring is secured. See the figure
below.
Installing ring
8-4
Care and Maintenance
2 ring-buttons
Turn 35°to the right after installing
3.
Clean control panel
z
Tools: dry soft cloth, mild soapy water
z
Method:
Finishing installing
Use dry soft cloth to clean the surface of the system. If the system is dirty, moisten the soft
cloth with mild soapy water and wipe off any stains. Use dry soft cloth to remove any moisture
and allow all hard surfaces to completely air-dry. If it is difficult to clean the control panel,
disassemble the keys first and then use mild soapy water to clean it.
NOTE:
Keyboard on the control panel should be cleaned periodically; otherwise, keys maybe
blocked by dirt and buzzer dings, keys don’t work.
4.
Clean Probe
z
Tools: mild soapy water , dry soft cloth , soft brush
z
Method:
a)
Wipe out the dust attached to surface of probe, and check if it has any abnormality
such as crack;
b)
Remained stain or dust attached to surface of probe cable should be washed out by
cloth with clear water or soapy water, and then air-dry the surface of probe cable;
c)
Use a soft brush to clean the probe holder, removing dust simultaneously.
5.
Clean Cover
z
Tools: dry soft cloth, soapy water
z
Method:
Use dry soft cloth to clean the cover of the system. If the system is dirty, moisten the soft
cloth with mild soapy water and wipe off any stains, then air-dry.
Note: Be sure to use soft brush to brush the dust attached to all the sockets or interfaces which
can be seen (such as probe sockets, sockets or interfaces in IO panel and power supply
panel),not the cloth with water.
8.2.3
Clean the Peripherals
Do the cleaning maintenance according to your actual peripheral configuration; items which are not
configured can be skipped.
8-5
Care and Maintenance
Table 8-4 Peripherals Cleaning List
No.
Content
Description
Color and B/W
video printer
First wipe off dust or stain attached to the cover of printer with soft
dry cloth, then clean the inside of printer. Be sure to do the cleaning
maintenance according to the operation manual if is necessary.
2.
Graph / text printer
First wipe off dust or stain attached to the cover of printer with soft
dry cloth, then clean the inside of printer. Be sure to do the cleaning
maintenance according to the operation manual if is necessary.
3.
Foot switch
Use soft dry cloth with a little mild soap water to wipe off the dust or
stain attached to the pedals or cable of foot switch.
1.
8.3
Checking
8.3.1
General check
Table 8-5 General check list
No.
Content
Method
a) Visually check to confirm that there is no crack and expansion to
probe head.
1.
Probe
b) Visually check to confirm that there is no deterioration or
desquamation to probe cable.
c) Visually check to confirm that there is no bend, destroyed or falling off
pins to the connector
2.
Power supply
adapter
a) Visually check to confirm that there is no wrinkles, crack or
deterioration; No crack and expansion to the surface of adapter.
b) Manually check to confirm that there is no looseness or rupture. The
connection of plug is reliable.
Check the battery:
a) Check the abnormal information or log of battery and ask abnormal
situation in use
Operate as follow if necessary:
3.
Battery
b) Check if battery can be charged normally when power-off: Record
one capacity value before power-off, then check the capacity
change after period of time when power-on. It takes less than 2
minutes to increase 1% capacity when the total capacity is less than
90% and it takes more time when the capacity is more than 90%.
c) Disconnect the system from the AC power supply to confirm if the
system can maintain normal work status in the battery power
supply.
8-6
Care and Maintenance
8.3.2
System Function Check
The system function checking is not required during Preventive Maintenance. Engineer or
Customer may use it as part of their product Quality Assurance Program tests。
Table 8-6 System function list
No.
Content
Method
1.
B mode
Verify basic operation of B mode. Check basic software and
hardware controls affecting B mode operations.
2.
Color mode
Verify basic operation of Color mode. Check basic software and
hardware controls affecting Color mode operations.
3.
Doppler mode (PW/CW)
Verify basic operation of Doppler mode. Check basic software
and hardware controls affecting Doppler mode operations.
4.
M mode
Verify basic operation of M mode. Check basic software and
hardware controls affecting M mode operations.
5.
Measurement(2D, M,
Doppler general
measurement, applied
measurement optional)
Scanning gray scale imaging on phantom, verify distance and
area accuracy with measurement control. Verify measurement
accuracy by performance test.
6.
Keyboard test
Operate keyboard test to verify if all control keys can work
normally.
7.
LCD
Verify LCD display function and parameters adjustment. Refer
to that of LCD checking.
8.
Software menu check
Verify software menu display function: if each operation menu
and page can be accessed.
Remark:Please refer to 6.4~6.5 for details.
8.3.3
Peripherals and Options Check
If the system is not configured with any module or peripheral, the corresponding items checking can
be skipped.
Table 8-7 Options, Peripherals and Accessories Check list
No.
Content
Method
1.
Color and B/W video printer
2.
Graph / text printer
3.
Foot switch
Check if the foot switch can implement the set functions
according to the program.
4.
DVD-R/W
Check if DVD can work normally (write, read and pop).
5.
DICOM
Check if DICOM can work normally and send pictures
and other data to DICOM server.
Check if the output of video printer is normal.
Check if the output of graph / text printer is normal.
8-7
Care and Maintenance
No.
Content
Method
Remark:Please refer to 6.3 for details.
8.3.4
Mechanical Safety Inspection
Mechanical safety inspection is mainly used to check mechanical strength and mechanical
function of the key assembly of ultrasonic system. The mode of test evaluation mainly is: Perform
the evaluation by means of visual check and operating check, if the check result cannot pass, the
system is in abnormal status now. Stop using the system and adopt proper measures. The test flow
is as following:
Fig 8-4 Mechanical Safety Inspection Flow
Table 8-8 Mechanical Safety Check
NO.
1.
Item
Casters of cart
Method
a)
Visually check to confirm there is no any crack.
b)
Operate the casters to confirm the locking and
releasing functions are normal.
8-8
Tool
none
Care and Maintenance
NO.
Item
2.
Connection of
cart casters
3.
Handle of cart
Method
a)
Visually check to confirm that there is no
skewness and the connecting screws are free of
breakage or falling off.
b)
Check with the spanner to make sure that there is
no looseness between the caster and the base
connection screw.
a)
Check by sight if the handle has cracks.
b)
Use a wrench to check if the handle is loose.
Tool
Inner
hexagon
wrench 6
4.
Cart supporting
parts
a)
Use hands to check if the supporting parts are
loose.
none
5.
Enclosure of main
unit
b)
Check by sight if there is any crack.
none
a)
Visually check to confirm there is no any crack.
none
b)
Rock the handle gently and then lift the
ultrasound machine to confirm that the handle is
free of looseness and it can accept normal force.
6.
Handle
7.
Control panel and
turning
mechanism
8.
Other mechanical
structures
8.3.5
Open and close the control panel to see if it is free of
looseness or separated from the main unit.
none
Check to confirm that there is no looseness to other
mechanical parts, no crack to cover and no
conductive parts show in sight.
none
Electrical Safety Inspection
Only technical professionals or engineers after training can perform electric safety inspection.
Please refer to appendix A: Electrical Safety Inspection for details.
8-9
9
Troubleshooting of Regular
Malfunctions
9.1
Inspection and Repair Work Flow of Power
System Failure
9.1.1
Inspection and Repair Work Flow When Powered by
Adapter
Fig 9-1 Inspection and Repair Work Flow when Powered by Adapter
9-1
Troubleshooting of Regular Malfunctions
9.1.2
Inspection and Repair Work Flow When Powered by
Batteries
Fig 9-2 Inspection and Repair Work Flow when Powered by Batteries
9-2
Troubleshooting of Regular Malfunctions
9.1.3
Inspection and Repair Work Flow of the Charging
Unit of the Power System
When the power adapter is not
powered, put the battery pack
in the machine and turn it off
The battery pack is
faulty or power capacity
is not sufficient or
connection is faulty;
service
For A_battery+ or
B_battery+ in the main
power board, check if
the voltage is normal
No
Yes
Adapter power on
The charging unit A or B
in the auxiliary power
board is faulty; service
Check if A_battery+ or
B_battery+ voltage is
about 12.6v or the voltage
rises slowly
No
Yes
End
Fig 9-3 Inspection and Repair Work Flow of the Charging Unit of the Power System
9.2
Location and Repair of System Boot
Failure Malfunctions
No.
Failure Description
Failure Analysis
Measure
a. The output indicator of the adapter is off when
The adapter is
Replace the
the adapter is not connected to main unit;
damaged.
adapter.
a. The power output indicator is lit up and the
The output voltage of
Replace the
adapter-present indicator of main unit is off when
the adapter is lower
adapter.
b. The output indicator of the adapter and the
adapter-present indicator of main unit are off when
1
the adapter is connected to main unit;
c. There is no response as POWERBTN is pressed
when the system is powered by adapter only; the
system starts up normally but the status indicator of
the batteries is off or blinking in orange when both
the adapter and the batteries are connected.
2
9-3
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure Analysis
the adapter is connected to main unit;
than 10.8V (low
b. There is no response as POWERBTN is pressed
possibility).
when the system is powered by adapter only;
c. The system starts up normally but the status
indicator of the batteries is off or blinking in orange
when both the adapter and the batteries are
connected.
Measure
The common-mode
inductor of main
board adapter is open
circuit (high
Replace the
main board.
possibility).
The 3V3STB of power
Replace the
module is abnormal
power
(medium possibility).
module.
The 3V3STB of main
Replace the
board is
main board.
short-circuited (low
possibility).
The 5VSTB of power
Replace the
module is abnormal
power
(medium possibility).
module.
c. There is no response as PWR_BTN is pressed
The 5VSTB of main
Replace the
and the system is powered by either adapter or
board is
main board.
batteries.
short-circuited (low
a. The output indicator of the adapter is normal;
b. All the indicators in the control panel are off;
3
possibility).
Power management
Replace the
FPGA is damaged
main board.
(high possibility).
The 5VSTB of
Replace the
industrial control
industrial
board is
control board.
short-circuited (low
possibility).
9-4
Troubleshooting of Regular Malfunctions
No.
4
Failure Description
Failure Analysis
Measure
a. The power output indicator of the adapter is
The input segment of
Replace the
normal when the adapter is not connected to main
the power module is
power
unit, and the output indicator is off when the
short-circuited.
module.
a. The output indicator of the adapter is lit up; the
The 5VSTB_OK#
Replace the
adapter-present indicator in the control panel is off;
output of power
power
b. The system starts up normally when it is
module is abnormal
module.
powered by the adapter, but the indicator
(low possibility)
adapter is connected to main unit;
b. All the indicators in the control panel are off;
c. There is no response when PWR_BTN is
pressed and the system is powered by either
adapter or batteries.
5
PWR_BTN in the control panel is off and the
indicators related to batteries are displayed
normally;
c. There is no response as Power_BTN is pressed
when the system is powered by batteries.
The reception of
Replace the
5VSTB_OK# of main
main board.
board is abnormal
(high possibility).
POWER_ON# of
main board is
Replace the
abnormal (high
main board.
possibility).
6
a. The indicator of the adapter, indicator of quantity
The 12V output of
of electricity and backlight of Power_BTN are all
power module is
normal when the system is turned off;
abnormal (medium
b. Press Power_BTN and its backlight first blinks
possibility).
and then lights up continuously. The backlight of
control panel is off, the keyboard sound
unavailable, fan not working, and LCD, VGA, DVI,
Video and S-video displaying nothing;
Replace the
power
module.
The 12V of main
board is
Replace the
short-circuited (low
main board.
possibility).
The 12V of industrial
control board is
short-circuited (low
possibility).
9-5
Replace the
industrial
control board.
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure Analysis
Measure
The main board
cannot send power-on
a. The indicator of the adapter, indicator of quantity
of electricity and backlight of Power_BTN are all
7
PWR_BTN to
Replace the
industrial control
main board.
board (high
possibility).
normal when the system is turned off;
The industrial control
b. Press PWR_BTN and its backlight keeps
board cannot respond
Replace the
blinking. The backlight of control panel is off, the
PWR_BTN signal
industrial
keyboard sound unavailable, fan not working, and
normally (low
control board.
LCD, VGA, DVI, Video and Svideo displaying
possibility).
nothing;
The main board
cannot S3# signal
from the industrial
control board (high
Replace the
main board.
possibility).
a. Nothing is wrong when the system is powered by
8
the adapter;
Cannot switch to
b. There is no response as PWR_BTN is pressed
batteries.
when the system is powered by batteries;
Replace the
power
module.
a. Nothing is wrong when the system is powered by
batteries;
b. The adapter-preset indicator is off and there is
no response as PWR_BTN is pressed when only
the adapter is connected;
9
c. When both the adapter and batteries are
connected, the adapter-present indicator is normal,
and the system is actually powered by batteries
during startup and running and the quantity of
electricity of batteries is being consumed;
d. Charging with the system off is available;
9-6
Cannot switch to the
adapter.
Replace the
power
module.
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure Analysis
Measure
a. The indicator of the adapter, indicator of quantity
of electricity of batteries and backlight of
10
Power_BTN are all normal;
The 3.3V output of
b. The backlight of Power_BTN is normal as it is
power module is
pressed; the backlight first blinks and then keeps
abnormal (high
on and there is button beep; the fan runs normally
possibility).
Replace the
power
module.
and the backlight of control panel is normal;
c. Nothing is displayed in LCD, VGA, DVI, Video
and Svideo;
d. The system can make Doppler; power-off is
The 3.3V of main
unavailable by pressing PWR_BTN shortly
board is
Replace the
(power-off is available only by entering Power Off
short-circuited (low
main board.
selection and confirming);
possibility).
e. Backlight of LCD can be spotted in dark room;
The 3.3V power
a. The indicator of the adapter, indicator of quantity
supply to LCD of main
board is abnormal
of electricity of batteries and backlight of
(high possibility).
Power_BTN are all normal;
Replace the
main board.
b. The backlight of Power_BTN is normal as it is
pressed; the backlight first blinks and then keeps
on and there is button beep; the fan runs normally
11
and the backlight of control panel is normal;
c. Nothing is displayed in LCD and the display of
LVDS of main board is
abnormal (high
possibility).
Replace the
main board.
VGA, DVI, Video and Svideo is normal;
Backlight of LCD is off
Replace the
d. The system can make Doppler; power-off is
(high possibility).
main board.
unavailable by pressing PWR_BTN shortly
LCD is abnormal (low
Replace the
(power-off is available only by entering Power Off
possibility).
LCD.
selection and confirming);
e. Backlight of LCD can be spotted in dark room;
Inverter of LCD is
abnormal (low
possibility).
Replace the
inverter.
a. The indicator of the adapter, indicator of quantity
12
of electricity of batteries and backlight of
Backlight of LCD is
Replace the
Power_BTN are all normal; blink first and then
abnormal.
LCD.
keep on;
9-7
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure Analysis
Measure
b. The backlight of Power_BTN is normal as it is
pressed; the backlight first blinks and then keeps
on and there is button beep; the fan runs normally
and the backlight of control panel is normal;
c. The display of LCD is not even and the display of
VGA, DVI, Video and Svideo is normal;
d. The system can make Doppler; power-off is
unavailable by pressing PWR_BTN shortly
(power-off is available only by entering Power Off
selection and confirming);
e. Backlight of LCD can be spotted in dark room;
a. The indicator of the adapter, indicator of quantity
of electricity of batteries and backlight of
Power_BTN are all normal;
b. The backlight of Power_BTN is normal as it is
pressed; the backlight first blinks and then keeps
13
on and there is button beep; the backlight of control
12VOK of power
panel is off and the fan runs normally;
module is abnormal.
c. Nothing is displayed in LCD, VGA, DVI, Video
Replace the
power
module.
and Svideo;
d. Backlight of LCD can be spotted in dark room;
e. Power-off is unavailable by pressing PWR_BTN
shortly;
a. The indicator of the adapter, indicator of quantity
of electricity of batteries and backlight of
The 5V output of
Power_BTN are all normal;
b. The backlight of Power_BTN is normal as it is
pressed; the backlight first blinks and then keeps
14
on and there is button beep; the backlight of control
power module is
abnormal (high
possibility).
Replace the
power
module.
panel is off and the fan runs normally;
c. Nothing is displayed in LCD, VGA, DVI, Video
and Svideo;
The 5V of main board
d. Backlight of LCD can be spotted in dark room;
is short-circuited (low
e. Power-off is available by pressing PWR_BTN
possibility).
shortly;
9-8
Replace the
main board.
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure Analysis
Measure
a. The indicator of the adapter, indicator of quantity
of electricity of batteries and backlight of
15
Power_BTN are all normal;
The processing to
b. The backlight of Power_BTN is normal as it is
12VOK of main board
pressed; the backlight first blinks and then keeps
is abnormal.
Replace the
main board.
on and there is button beep; the backlight of control
panel is on and the fan runs normally;
c. Nothing is displayed in LCD, VGA, DVI, Video
and S-video;
d. Backlight of LCD can be spotted in dark room;
The POST of industry
e. Power-off is unavailable by pressing PWR_BTN
control board fails.
Replace the
industrial
control board.
shortly;
Replace
HDD failure
16
Prompts “Operating system no found” during the
startup.
The operating system
fails to boot up.
9.3
HDD.
Restore the
operating
system.
Location and Repair of Non-echo Failure
in Image Region
No.
Failure Description
Failure Analysis
Measure
Something is wrong with AFE of main
Replace the
board (high possibility)
main board.
transducer and the digital
The 3V6 output of power module is
Replace the
gain is maximal.
abnormal (low possibility)
power module.
a. Connect the transducer,
Something is wrong with the programmed
Replace the
voltage of power module (high possibility).
power module.
the transducer correctly
The programmed power of main board is
Replace the
b. Connect the transducer,
short-circuited (high possibility).
main board.
There is no noise image
1
2
when there is no
and the system identifies
9-9
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure Analysis
Measure
and there is only noise
The programmed power of transmission
Replace the
image in non-frozen mode.
board is short-circuited (low possibility).
Transmission
Board
The fixed high-voltage output of power
Replace the
module is abnormal (high possibility).
power module.
The high-voltage switch is abnormal,
Replace the
including the short circuit of power supply
main board.
(low possibility).
The +/-5V output of power module is
Replace the
abnormal, resulting high-voltage isolation
power module.
problem (low possibility).
The high-voltage isolation is abnormal,
Replace the
including the short circuit of power supply
main board.
(low possibility).
Transducer is damaged (low possibility).
Replace the
transducer.
3
a. Connect the transducer,
Transducer identification problem of main
Replace the
and the system identifies
board (high possibility).
main board.
Transducer is damaged (low possibility).
Replace the
the transducer correctly
b. Connect the transducer,
and there is only noise
transducer.
image in non-frozen mode.
9-10
Troubleshooting of Regular Malfunctions
9.4
Location and Repair of the Problem of
Image with Black Area
No.
1
Failure Description
Failure Analysis
Measure
There is black area in the noise
Something is wrong with
Replace the main board.
image when there is no
AFE of main board
transducer and the digital gain
is maximal.
2
a. The self-diagnosis of
The high-voltage switch of
front-end channel is passed
main board is abnormal.
b. Connect the transducer, and
there is black area in
Replace the main board.
The transducer is damaged.
Replace the transducer.
The transmission and main
Pull out and reinsert the
board are in poor contact.
transmission board.
The channel of transmission
Replace the
board is damaged.
Transmission Board
The high-voltage isolation is
Replace the main board.
non-frozen mode.
a. The self-diagnosis of
front-end channel is failed
3
b. Connect the transducer, and
there is black area in
non-frozen mode.
abnormal.
9.5
Abnormalities in Image Region
The problems to be solved in this section include abnormalities of B, C, PW and CW images. To
solve this problem, replace the power module first to exclude the problems of the power module,
then transmission board and at last the main board. If there is analog CW board for CW mode, it
should be replaced first.
9-11
Troubleshooting of Regular Malfunctions
9.6
Inspection and Repair Work Flow of USB
Port Failure
Fig 9-4 Inspection and Repair Work Flow of USB Port Failure
9-12
Troubleshooting of Regular Malfunctions
9.7
Inspection and Repair Work Flow of
Network Failure
Fig 9-5 Inspection and Repair Work Flow of Network Failure
9.8
No.
Troubleshooting for 4D Board
Failure Description
Failure
Measure
Analysis
1
Switch to 4D, and the
The -12V
Test the -12V output of the power module. If the
mechanical scan of
output of the
-12V is unavailable, replace the power module;
the 4D transducer
power module
does not begin,
is unavailable.
9-13
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure
Measure
Analysis
namely the electric
The
(1) Measure whether the voltage of PIN 3 and PIN
motor is not working.
±(7V~11.5V)
5 of U13 (or U4) of 4D driving board are normal.
of
Generally, PIN 3 should be negative (-7V ~
programmed
-11.5V), and PIN 5 should be positive (+7V ~
power module
+11.5V).
of 4D driving
(2) If they are both unavailable, check whether
board is not
fuse F1 is burnt out. If it is, first check whether the
available.
two ends of the fuse are short-circuited to ground
and replace the fuse if they are short-circuited.
(3) If the fuse is not burnt out, measure whether
the +12V voltage of either end of the fuse is
normal. If not, the problem is not resulted from the
driving board and main board socket, then check
the power module.
(4) If the problems mentioned in (2) and (3) are all
excluded, replace U12 in the 4D driving board.
The driving
(1) Measure TP3, TP6 and TP8 with oscillograph,
signal of the
and skip to (2) if they are normal. If no output,
4D driving
measure VDD (PIN 1) and VREF (PIN 5) of U6
board is
with multimeter. If VDD is abnormal, check 3.3V
unavailable.
of the main board; if VREF is abnormal, replace
U18.If both VDD and VREF are normal, replace
DAC U6.
(2) Measure TP5 and TP11 with oscillograph. If
there are no driving signals, replace U13 and U4;
(3) If both TP5 and TP11 are available, measure
PIN 8 and PIN 12 of socket P5 of 4D driving
board with oscillograph. If there is no output, it
means the electric motor inside the transducer is
damaged. Replace 4D transducer.
2
Switch to 4D, and the
Electric motor
(1) Check whether there is square wave output in
mechanical scan of
stops working
PIN 3 (HALL signal) of socket P5 of 4D driving
4D transducer stops
after started
board during the time when the transducer is
after a few seconds,
for a few
working.
9-14
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure
Measure
Analysis
namely the electric
seconds
motor stops working.
(2) If not, check whether the 4 driving signals of
PIN 1 ~ PIN 8 of socket J1 of transducer
connection board are normally output. If normal,
skip to (4).If not, check the output of driving
signals from the 4D driving board (refer to
malfunction 1 and failure analysis 2 for the
measurement and solutions)
(3) If the driving signals of 4D driving board are
normally output, replace the cables.
(4) The HALL of 4D transducer is damaged.
Replace 4D transducer.
9.9
No.
1
Troubleshooting for the Keyboard Board
Failure Description
Indicator D175 is OFF.
Failure Analysis
Measure
The keyboard cable is pulled out or
Fix the keyboard
something is wrong with the keyboard
cable again or
cable. Test TP11, which should be 5V.
replace the
keyboard cable.
2
Indicator D176 is OFF.
The keyboard cable is pulled out or
Fix the keyboard
something is wrong with the keyboard
cable again or
cable. Test TP21, which should be 12V.
replace the
keyboard cable.
LED D175 and D176 are
FPGA configuration files haven’t been
Burn FPGA
lit up normally. The LED
burnt.
configuration files.
Make sure the core voltage of FPGA is
Replace U20.
D1 of FPGA is not lit up
3
or is blinking. The
keyboard board
program cannot run
4
normal. The voltage of TP17 should be
1.2V.
normally.
The FLASH is damaged.
Replace U3.
TGC value is unstable;
Make sure TGC line is connected well
Fixed
even when you don’t
(J2)
9-15
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure Analysis
Measure
move the slider, the
The connector of TGC connection wire
Replace TGC
value varies; the offset
looses.
connection wire.
Some LEDs in boards
Check whether the LED soldering
Solder the
and cards are not lit up,
direction is correct; check whether the
component again or
and the dual-color LEDs
lights are damaged in appearance;
replace it.
are always lit up at the
check whether the pin of dual-color light
same time.
is broken circuit.
between the value and
position is great.
5
Windows cannot identify
the keyboard board (the
6
No keyboard board driver is installed.
Install the driver.
The main board connection wire is
Replace the main
damaged.
board connection
wire.
power of boards and
cards is normal, and
Test whether USB common-mode
software and logic are
inductor L8 is normally conducted.
correctly burnt).
Replace L8.
The anti-static chip U5 is damaged.
Replace U5.
The USB interface chip U4 is damaged.
Replace U4.
9.10 Troubleshooting for ECG Board
No.
Failure Description
Failure Analysis
Measure
The corresponding
The connection between the main
Replace the connection
function is unavailable.
unit and the ECG module is
cable.
abnormal.
1
Power failure
Refer to power failure
resolution (Note A)
The communication of the serial
Replace U8
port is abnormal.
Open the corresponding
2
function, and the wave
form is displayed as a
line, or it is abnormal.
ECG lead abnormal
Replace the ECG lead
The connection between the main
Replace the connection
unit and the ECG module is
cable.
abnormal.
9-16
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure Analysis
Measure
If the connection cable is normal,
Refer to power failure
open the ECG board and check
resolution (Note A)
the indicators D15 and D20 are lit
up. If not, it means something is
wrong with the power.
ECG program abnormal
Rewrite the inferior
program of ECG board.
ECG board abnormal
Replace the ECG board.
Note A. ECG board power failure resolution
When trying to solve power failure, check the power tree of ECG board of 2118 step by step.
Fig 9-6 Power Tree of ECG Board
The following table lists the relevant information of the test points.
Failure Analysis
Network
Standard
Voltage Testing
Name
Voltage (V)
Points
ECG+5V
5±10%
Voltage drop of
Test whether the two ends of
two ends of
F2 are conducted. If not,
C102
replace F2.
Voltage drop of
Maybe U17 is damaged.
Replace U17.
Maybe U23 is damaged
Replace U23
+12V
12±10%
Failure
Resolution
Replace F2.
two ends of
C112
SVCC
3.3±10%
Testing points of
the board
9-17
Troubleshooting of Regular Malfunctions
Failure Analysis
Network
Standard
Voltage Testing
Name
Voltage (V)
Points
ECG+5V
5±10%
Voltage drop of
Test whether the two ends of
two ends of
F2 are conducted. If not,
C102
replace F2.
Testing points of
Test whether T1 primary or
Replace T1, Q1,
the board
secondary coil is broken
orU18
6±10%
+6V,
+6D
Failure
Resolution
Replace F2.
circuit. If it is, replace TI.
-6±10%
-6V
Testing points of
the board
5±5%
+5V
Otherwise, Q1 or U18 may be
faulty.
Testing points of
Maybe U22 is damaged
Replace U22
Maybe U24 is damaged
Replace U24
Maybe U19 is damaged
Replace U19
Maybe U25 is damaged
Replace U25
the board
3.3±5%
+3V3
Testing points of
the board
3.3±5%
+
the board
3V3REF
-5V
Testing points of
-5±5%
Testing points of
the board
9.11 Troubleshooting for Video & Audio
Capture Card
No.
Failure Description
After the video capture card is
1
connected, the screen prompts that
the surge occurs or the current is
not enough.
After the video capture card is
2
Failure Analysis
USB port cannot provide
sufficient power to drive the
video capture card.
USB cable is not secured.
connected, the screen doesn’t
display images.
The chip is damaged.
9-18
Measure
Replace the main
board.
Reconnect the USB
cable.
Replace the chip.
Troubleshooting of Regular Malfunctions
No.
Failure Description
Failure Analysis
Measure
The signal cable is in bad
3
The image is not continuous or
contact, e.g. SVIDEO signal
some image frames are missing or
cable is in bad contact or
discolored.
USB signal cable is in bad
Change cable or
change the socket.
contact.
The installation of the driver
isn’t correct.
4
After the video capture card is
Check if the audio input
connected, clicking the capture
signal rule is specified in the
command cannot normally capture
capture software.
sound.
Reinstall the driver.
Specify it in the
software.
Check if there is BCLK signal,
if not, it indicates EMP202
Replace the chip.
chip is damaged.
The video displayed by
6
Only part of the source video can
be seen.
capture card is of 640*480
resolution. If resolution is not
correct, the display will be
abnormal.
9-19
/
Appendix A Electrical Safety
Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive
maintenance program. They are a proven means of detecting abnormalities that, if undetected,
could prove dangerous to either the patient or the operator. Additional tests may be required
according to local regulations.
All tests can be performed using commercially available safety analyzer test equipment. These
procedures assume the use of a 601PROXL International Safety Analyzer or equivalent safety
analyzer. Other popular testers complying with IEC 60601-1 used in Europe such as Fluke, Metron,
or Gerb may require modifications to the procedure. Follow the instructions of the analyzer
manufacturer.
The consistent use of a safety analyzer as a routine step in closing a repair or upgrade is
emphasized as a mandatory step if an approved agency status is to be maintained. The safety
analyzer also proves to be an excellent troubleshooting tool to detect abnormalities of line voltage
and grounding, as well as total current loads.
A-1
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
1- Power Cord Plug
TEST PROCEDURE
‹The Power Plug
The Power Plug Pins
No broken or bent pin. No discolored pins.
The Plug Body
No physical damage to the plug body.
The Strain Relief
No physical damage to the strain relief. No plug warmth for device in
use.
The Power Plug
No loose connections.
‹The Power Cord
No physical damage to the cord.
The Power Cord
No deterioration to the cord.
--For devices with detachable power cords, inspect the connection at
the device.
--For devices with non-detachable power cords, inspect the strain
relief at the device.
A-2
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
2- Device Enclosure And Accessories
TEST PROCEDURE
‹Visual Inspection
No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
The Enclosure and Accessories
No residue of fluid spillage (e.g., water, coffee, chemicals, etc.).
No physical damage to probe head (e.g., crack)
No loose or missing parts (e.g., knobs, dials, terminals, etc.).
‹Contextual Inspection
No unusual noises (e.g., a rattle inside the case).
The Enclosure and Accessories
No unusual smells (e.g., burning or smoky smells, particularly
from ventilation holes).
No taped notes that may suggest device deficiencies or
operator concerns.
A-3
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
3- Device Labeling
TEST PROCEDURE
Check the labels provided by the manufacturer or the healthcare facility is present and
legible.
¾ Main Unit Label
¾ Integrated Warning Labels
¾ Slope and High Voltage Caution Label
¾ Don’t Stress Label
A-4
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
5- Earth Leakage Test
OVERVIEW
Run an Earth Leakage test on the device being tested before performing any other leakage
tests.
Leakage current is measured the following ways:
♦
Earth Leakage Current, leakage current measured through DUT outlet Earth
♦
Earth Leakage Current AP-EARTH (ALL Applied Parts connected to Earth), leakage
current measured through DUT outlet Earth
There is no need to attach a test lead; the 601PRO automatically connects the measuring
device internally.
TEST PROCEDURE
‹Perform the Test
1) From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO front
panel outlet, and turn on the device.
2) Attach the device's applied parts to the 601PRO applied part terminals if applicable.
3) Press shortcut key 4.The Earth Leakage test appears on the display, and the test begins
immediately:
¾ SOFT KEY 1 toggles the DUT outlet Polarity from Normal to Off to Reverse.
¾ SOFT KEY 2 toggles the DUT outlet from Earth to No Earth.
¾ SOFT KEY 3 toggles the DUT outlet from L2 to No L2.
¾ SOFT KEY 4 toggles the AP to Earth to No AP to Earth.
4) Press the print data key at any time to generate a printout of the latest measurement.
A-5
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
5- Earth Leakage Test
Figure 1
Earth leakage test
‹Failure
Check any short-circuits of the Y capacitor on power unit. Replace a new one if any portion
defective.
Check any broken of the Power Unit. Replace a new one if any portion defective.
Inspect mains wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or
owner to correct any deviations. As a work around, check the other outlets to see if they could be used
instead.
Change another probe to confirm if the fail is caused by console.
Inspect mains wiring for bad crimps, poor connections, or damage.
If the leakage current measurement tests fail on a new unit and if situation cannot be corrected,
submit a Safety Failure Report to document the system problem. Remove unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal.
LIMITS
UL60601-1:300 μA Normal Condition
1000 μA Single Fault Condition
IEC60601-1:
500 μA Normal Condition
1000 μA Single Fault Condition
A-6
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
6- Patient Leakage Current
OVERVIEW
Patient leakage currents are measured between a selected applied part and mains earth.
TEST PROCEDURE
‹Prepare
Perform a calibration from the Mains on Applied Part menu.
The following outlet conditions apply when performing this test:
Normal Polarity, Earth Open, Outlet ON
Normal Polarity, Outlet ON
Normal Polarity, L2 Open, Outlet ON
Reversed Polarity, Outlet ON
Reversed Polarity, Earth Open, Outlet ON
Reversed Polarity, L2 Open, Outlet ON
‹ Warning
If all of the applied parts correspond to the instrument type, the applied parts will be tied
together and one reading will be taken. If any of the applied parts differ from the instrument type, all
applied parts will be tested individually, based on the type of applied part. This applies to Auto and
Step modes only.
‹Perform the Test
1) From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO front panel
outlet, and turn on the device.
2) Attach the applied parts to the 601PRO's applied part terminals.
3) Press shortcut key 6. The Patient Leakage test is displayed, and the test begins immediately.
4) Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part leakage current.
5) Modify the configuration of the front panel outlet by pressing the appropriate SOFT KEY on the
601PRO.
6) Press the print data key at any time to generate a printout of the latest measurement.
A-7
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
6- Patient Leakage Current
Figure 2
patient leakage Current
‹Note
1, In addition to Probes ,Patient leakage current test should be perform if ECG or PCG
parts used;
2, If the current test standard being used does not include Patient Leakage DC readings,
or the DC option is not enabled, then DC readings will not be available through the APPLIED
PART SOFT KEY selections. Refer to Chapter 8, Standards and Principles.
‹Failure
Check any broken of the Applied parts. Replace any defective one.
Check any broken of the ECG/PCG module if used, Replace any defective one.
Check any broken of the Power Unit. Replace a new one if any portion defective.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user
or owner to correct any deviations. As a work around, check the other outlets to see if they could be
used instead.
Change another probe to confirm if the fail is caused by console.
Inspect wiring for bad crimps, poor connections, or damage.
If the leakage current measurement tests fail on a new unit and if situation can not be corrected,
submit a Safety Failure Report to document the system problem. Remove unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal.
LIMITS
A-8
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
6- Patient Leakage Current
All countries
For BF ECG input and transducer
100μA Normal Condition
500μA Single Fault Condition
A-9
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
7- Mains on Applied Part Leakage
OVERVIEW
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage, through a
limiting resistance, to selected applied part terminals. Current measurements are then taken between the
selected applied part and earth. Measurements are taken with the test voltage (110% of mains) to
applied parts in the normal and reverse polarity conditions as indicated on the display.
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal Polarity;
Reversed Polarity
TEST PROCEDURE
‹Prepare
To perform a calibration from the Mains on Applied Part test, press CAL (SOFT KEY 2).
1) Disconnect ALL patient leads, test leads, and DUT outlet connections.
2) Press CAL to begin calibration, as shown:
If the calibration fails, the previously stored readings will be used until a passing calibration has
occurred. Also, the esc/stop key has no effect during calibration.
3) When the calibration is finished, the Mains on Applied Part test will reappear.
‹ Warning
1) A 2-beep-per-second signal indicates high voltage present at the applied part terminals while a
calibration is being performed.
2) High voltage is present at applied part terminals while measurements are being taken.
‹Performance
1) From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601
A-10
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
7- Mains on Applied Part Leakage
2) Attach the applied parts to the 601PRO applied part terminals.
3) Attach the red terminal lead to a conductive part on the DUT enclosure.
4) Press shortcut key 7. The Mains on Applied Part test is displayed.
5) Select the desired outlet configuration and applied part to test using the appropriate SOFT
KEYS:
6) Press START TEST (SOFT KEY 1) to begin the test.
7) Press the print data key to generate a printout of the latest measurement.
Figure 3
Mains on Applied part leakage
‹Note
1, In addition to Probes ,Patient leakage current test should be perform if ECG or PCG parts
used;
2, If all of the applied parts correspond to the instrument type, the applied parts will be tied
together and one reading will be taken. If any of the applied parts differ from the instrument type,
all applied parts will be tested individually, based on the type of applied part. This applies to Auto
and Step modes only.
‹Failure
Check any broken of the Applied part. Replace any defective one.
Check any broken of the ECG/PCG module if used, Replace any defective one.
Check any broken of the Power Unit. Replace a new one if any portion defective.
Inspect wiring for bad crimps, poor connections, or damage.
A-11
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
7- Mains on Applied Part Leakage
Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or
owner to correct any deviations. As a work around, check the other outlets to see if they could be used
instead.
Change another probe to confirm if the fail is caused by console.
Inspect wiring for bad crimps, poor connections, or damage.
If the leakage current measurement tests fail on a new unit and if situation can not be corrected,
submit a Safety Failure Report to document the system problem. Remove unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal.
LIMITS
All countries:
For BF ECG input and transducer:
5000μA
A-12
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
8- Patient Auxiliary Current
overview
Patient Auxiliary currents are measured between any selected ECG jack and the remaining selected
ECG jacks. All measurements may have either a true RMS or a DC-only response.
TEST PROCEDURE
‹Prepare
1) From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO front
panel outlet, and turn on the device.
2) Attach the patient leads to the 601PRO ECG jacks.
3) Define the Lead Types from the View Settings Option (refer to: Lead Type Definitions in
Section 5 of this chapter).
4) Press shortcut key 8. The Patient Auxiliary Current test is displayed, and the test begins
immediately. Display values are continuously updated until another test is selected.
5) Press SOFT KEYS 1-4 to select leakage tests
6) Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part leakage
current:
7) Modify the configuration of the front panel outlet by pressing the appropriate SOFT KEY on the
601PRO:
8) Press the print data key at any time to generate a printout of the latest measurement.
Figure 4
patient Auxiliary Current
A-13
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
8- Patient Auxiliary Current
‹Note
If the current test standard being used does not include Patient Auxiliary Current DC
readings, or the DC option is not enabled, then DC readings will not be available through the
APPLIED PART SOFT KEY selections.
‹Failure
Check any broken of the AC cable. Replace a new one if any portion defective.
Check any broken of the enclosure. Replace any defective part.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or
owner to correct any deviations. As a work around, check the other outlets to see if they could be used
instead.
Change another probe to confirm if the fail is caused by console.
Inspect wiring for bad crimps, poor connections, or damage.
If the leakage current measurement tests fail on a new unit and if situation can not be corrected,
submit a Safety Failure Report to document the system problem. Remove unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal.
LIMITS
All countries
For BF ECG input and transducer
100μA Normal Condition
500μA Single Fault Condition
A-14
Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION FORM
(Class I equipment)
Overall assessment:
□
Scheduled inspection
Test item: 1, 2, 3, 9
□
Unopened repair type
Test item: 1, 2, 3, 9
□
Opened repair type, not modify the power part
Test item: 1, 2, 3, 4, 5, 9
including transformer or patient circuit board
□
Opened repair type, modify the power part including
Test item: 1, 2, 3, 4, 5, 6, 7, 8, 9
transformer or patient circuit board
Location:
Technician:
Equipment:
Control Number:
Manufacturer:
Model:
SN:
Measurement equipment /SN:
Date of Calibration:
INSPECTION AND TESTING
Pass/Fail
1
Power Cord Plug
2
Device Enclosure and Accessories
3
Device Labeling
4
Protective Earth Resistance
5
6
7
8
Ω
Normal condition(NC)
____μA
Single Fault condition(SFC)
____μA
Earth
Leakage
Patient
Leakage
Current
Normal condition(NC)
□BF____μA
Single Fault condition(SFC)
□BF____μA
Mains on Applied Part Leakage
Patient Auxliary
Leakage Current
□BF____μA
Normal
condition(NC)
□BF____μA
Single Fault
condition(SFC)
□BF____μA
A-15
Limit
Max 0.2 Ω
Max:
NC: 300μA(refer to
UL60601-1) *
NC: 500μA(refer to
IEC60601-1) *
SFC: 1000μA
Max:
BF applied part:
NC:100μA, SFC:
500μA
Max:
BF applied part:
5000μA
Max:
BF applied part:
NC:100μA, SFC:
500μA
Electrical Safety Inspection
Note:
The equipment which sell to America shall comply with the requirement of UL60601-1,
others shall comply with the requirement of IEC60601-1.
Name/ Signature: ____________________
Date:_____________________________
A-16
Phantom Usage Illustration
Appendix B Phantom Usage
Illustration
Targets Disposal of Phantom KS107BD
A1——A5: Axial resolution target group B: Blind-area target group
C: Longitudinal target group D: Horizontal target group
E: Mimic tumor F: Mimic sac (diam 10mm) and stone
G Mimic sac (diam 6mm)
4. Line Target System
There are 8 groups of nylon line targets disposed as shown in the figure.
1.A1——A5:
Axial and lateral resolution target group. The distances between the horizontal branch and the
acoustic window are 30, 50, 70, 120 and 160mm, the center horizontal distances between two
adjacent lines of A1 and A2 groups are 1, 5, 4, 3, 2mm, A3~A5 groups are 5, 4, 3, 2mm. The center
longitudinal distances between two adjacent lines of the longitudinal branches are 4, 3, 2, 1mm.
2.B:
Blind-area target group. Center horizontal distance of adjacent lines is 10mm, distances to the
acoustic window are 10, 9, 8, 7, 6, 5, 4, 3mm.
3.C:
Longitudinal target group. 19 target lines with a 10mm center distance between adjacent lines.
4.D:
Horizontal target group. 7 target lines with a 20mm center distance between adjacent lines.
B-1
Phantom Usage Illustration
Targets disposal- KS107BD
A1——A4
Axial resolution target group
B1——B4
Lateral resolution target group
C
Longitudinal target group
D
Horizontal target group
E1——E3
Mimic sacs with diameters of 2, 4, 6mm
4. Line Target System
There are 8 groups of line targets disposed in TM material as shown in the figure.
1.A1——A4:
Axial resolution target group. The upmost lines in each target locate at the depth of 10, 30, 50,
70mm, the center longitudinal distances of each group (from the top down) are 3, 2, 1, 0.5mm, and
the horizontal distance is 1mm.
2.B1——B4:
Lateral resolution target group. Locate at a depth of 10, 30, 50, 70mm, with a center horizontal
distance of 4, 3, 2, 1mm in each group.
3.C:
Longitudinal target group. 12 target lines with a 10mm center distance between adjacent lines.
4.D:
Horizontal target group. Locate at a depth of 40mm, with a 20mm center distance between adjacent