ELECTROSURGICAL UNIT
Model S300
OPERATING MANUAL
®
Yueh Sheng Electronic Industrial Co., Ltd.
2F., No. 29, Qiangang St., Shilin District
Taipei City, Taiwan 11170
Tel: +886 2 2882 2212 Fax: +886 2 2881 8635
Website: www.yesng.com.tw
E-mail: info@yesng.com.tw
Obelis European Authorized Representative
Center (O.E.A.R.C)
Boulevard Général Wahis 53
1030 Brussels, BELGIUM
Tel: +32 2 732 59 54 Fax: +32 2 732 60 03
Website: www.obelis.net
E-mail : mail@obelis.net
MN-S300-EN-A02
MN-S300-EN-A02
Table of Contents
Symbols, Warnings, and Abbreviations
Classifications
Labels
Cautions
Specifications
III
IV
V
VI
VII
Section One
Operating Instructions
1.1 Initial Inspections and Installations
1.2 General Descriptions and Operating Instructions
1.2.1 Control Panel
1.2.2 Utility Panel
1.2.3 Rear Panel
1.2.4 Operating Instructions
a) Precautions
b) Connecting a Patient Plate
c) Operating Modes
i) Mono-polar Cutting
ii) Mono-polar Coagulation
iii) Bipolar Operations
d) Activating Outputs From the ESU
i) Mono-polar Outputs
ii) Bipolar Outputs
8
8
10
10
12
14
15
15
16
18
19
21
23
25
25
28
Sections Two
Accessories
2.1 Reusable Accessories
2.2 Disposable Accessories
29
29
31
Section Three
Maintenance & Service Instructions
3.1 Maintenance Instructions
3.1.1 Cleaning
3.1.2 Output Power Calibrations
i) Calibration of Output Power for Mono-polar Cutting
ii) Calibration of Output Power for Mono-polar Coagulation
iii) Calibration of Output Power for Bipolar Coagulation
iv) Calibration for Contact Impedance Monitor
32
32
32
32
32
33
33
33
I
MN-S300-EN-A02
3.2
3.3
3.4
3.5
v) Calibration for CUT and COAG Waveforms
Service Manual
3.2.1 Troubleshooting
3.2.2 Frequently-used Test Points
Output Characteristics
Diagrams and Schematics
3.4.1 Constitutional Chart
3.4.2 Circuit Diagram – Computer Board
3.4.3 Circuit Diagram – Driver Board
3.4.4 Circuit Diagram – Motherboard
Bill of Materials
3.5.1 Computer Board
3.5.2 Motherboard
3.5.3 Driver Board
II
33
34
34
36
37
41
41
42
43
44
45
45
48
52
MN-S300-EN-A02
Symbols, warnings, and abbreviations
Pure Cut
Power Increase Key
Blend Cut I
Power Decrease Key
Blend Cut II
Mode Select Key
Blend Cut III
Defibrillation Proof, Type CF Applied Part
Patient Plate Fault
Floating Output; Output Isolated From Earth
Standard Coagulation
Enclosure Resists Entry of Vertically Falling
Water
Spray Coagulation
Attention, consult accompanying documents
Bipolar Split
CE-marked device
Bipolar Blend Split
Foot-controlled Switch
Bipolar Coagulation
Volume Indicator
Earth Ground
Speaker
Coagulation / Coagulate
High Voltage
Main Power Switch
Consult the Manual Before Use
Table I. Warnings, symbols, and abbreviations.
III
MN-S300-EN-A02
Classifications
According EN60601-1, Medical Electrical Equipment, Part I – General Requirement for
Safety, equipment and its applied parts shall be classified as described in Section 1, Clause 5
as below:
1.
2.
CLASS I EQUIPMENT: Type of protection against electric shock:
CLASS I EQUIPMENT may have parts with double insulation or reinforced insulation
or parts operating at SAFETY EXTRA-LOW voltage or accessible parts protected by
protective impedance cases where conductive parts of an electrical circuit have to be
accessible to enable EQUIPMENT to function.
TYPE OF APPLIED PART: Degree of protection against electric shock:
For applied parts, as a voltage to 110% of the highest rated mains voltage applied
between earth and any single input or single output part, the patient leakage current in
single fault condition should be less than 0.05mA.

Symbol of
Indicating equipment of the “CF” Class. The equipment is protective against cardiac
defibrillator (high voltage).

Symbol of
The neutral electrode is isolated from earth potential.
3.
4.
5.
Symbol of
Indicating enclosure resist entry of vertically falling water.
The enclosure of the equipment shall be constructed so that liquid spillage in normal use
does not wet electrical insulation or other compounds which when wetted are likely to
affect adversely the safety of the EQUIPMENT.
This EQUIPMENT is not intended for use in explosion area. The use of flammable
anesthetics or oxidizing gasses such as nitrous oxide (N2O) and oxygen should be
avoided if a surgical procedure is carried out in the region of the thorax or the head.
Operation of EQUIPMENT: Continuous operation with short-time loading
(Operating/Resting times of 10s/30s)
Operation under normal load for a specified period, starting from COLD CONDITION
without the specified limits of temperature being exceeded, the intervals between each
period of operation being sufficient to allow the EQUIPMENT to cool down to COLD
CONDITION.
IV
MN-S300-EN-A02
Labels
Store in this position
Fragile
No sunlight or rain
Hook delivery prohibited
only. Please keep
upright.
Figure I. Warning labels on the outer packaging of ESU S300.
MODEL: ESU S300
INPUT
.
115V ± 10% 60Hz 5A Fuse (slow)
220V ± 10% 50 Hz 2.5A Fuse (slow)
Operating Mode
Mono-polar
CUT
Mono-polar
COAG
Bipolar
Pure Cut
Blend I
Blend II
Standard
Spray
COAG
Cut
Maximum Power
(watts)
300
250
200
120
80
60
60
Maximum Voltage
(Vp-p)
2000
2100
2100
7000
10100
200
350
Usage Period: 10 years
Rated Load
Frequency
(Ω)
(kHz)
384
300
384
300
384
300
300
384
300
384
50
960
50
960
YUEH SHENG ELECTRONIC INDUSTRIAL CO., LTD.
2F., No. 29, Qiangang St., Shilin District, Taipei City, Taiwan 11170
TEL: +886 2 2882 2212
FAX: +886 2 2881 8635
Figure II Label on the back of ESU S300.
Electrosurgical Unit
Model S300
Serial No.
YUEH SHENG ELECTRONIC INDUSTRIAL CO., LTD.
2F., No. 29, Qiangang St., Shilin District, Taipei City, Taiwan 11170
TEL: +886 2 2882 2212
FAX: +886 2 2881 8635 E-mail: info@yesng.com.tw
Figure III. Product identification tag on the outer package of ESU S300.
V
MN-S300-EN-A02
1.
This generator is a high-frequency medical device. It is only to be operated by qualified
persons.
2.
Study the manual and be familiar with the unit’s functions before operating the generator.
3.
Inspect for correct input voltage before turning on the generator.
4.
This generator produces high-frequency, high-voltage signals. Do not operate in
flammable environments.
5.
This device is designed, constructed, and tested in accordance with EN6060-1-1-2
standard. However, small amounts of electromagnetic signals may emitted by this
generator. To reduce the effect of electromagnetic interference, place the generator as far
apart as possible from other devices.
6.
The patient must not contact any metal objects while under operation. The
high-frequency currents generated by this unit may cause electrical burns if any contacts
are made between the patient and a metal object during an operation.
VI
MN-S300-EN-A02
Specifications
1.
Input:
2.
Output
□
□
110V, ±10%
220V, ±10%
Operating
Mode
60Hz
50Hz
Maximum
Power
(watts)
5A
10A
Maximum
Rated Load
Frequency
Voltage (Vp-p)
(Ω)
(kHz)
Pure Cut
300
2000
300
384
Blend I
350
2100
300
384
Blend II
200
2100
300
384
Mono-polar
Standard
120
7000
300
384
COAG
Spray
100
10100
300
384
COAG
70
200
50
960
Cut
70
350
50
960
Mono-polar
CUT
Bipolar
3.
Dimensions:
42cm (D) x 37cm (W) x 14.5cm (L)
4.
Weight:
11 kg
5.
Applicable Standards:
ISO13485, ISO14971, ISO10993, EN550, EN980, EN60601-1, EN60601-1-2,
EN60601-2-2, 93/42/EEC Directive
VII
MN-S300-EN-A02
Section One – Operating Instructions
1.1 Initial Inspections and Installations
This equipment produces high-frequency electrical currents. It is only to be
operated and serviced by qualified personnel.
Initial Inspections
a) Visually inspect for damages that might have occurred during shipping. If any
damages are found, please contact the shipping agent and our local
representative.
Figure 1.1
b)
Packaging and locations of labels of ESU Model S300.
1.
Label 1 – Product identification tag on the back of ESU S300
2.
Label 2 – Product identification tag on the outer package of S300
3.
Label 3 – Warning labels on the external package of S300
4.
Shockproof foam
5.
YESNG Electrosurgical Unit S300
Check for full compliance of accessories. Please contact our authorized
representative shall there be any concerns regarding accompanied accessories.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
8
MN-S300-EN-A02
c)
Inspect Label 1 (the product identification tag) for the correct input voltage
before connecting to the receptacle. If there are discrepancies, please contact our
local representative or our head office.
MODEL : ESU S300
INPUT
LOT
SN
115 V ± 5%60Hz 10A Fuse(slow)
220 V ± 5%50Hz 8A Fuse(slow)
Usage Period:10 years
Operating Mode
Maximum Power
(watts)
Maximum Voltage
(Vp p)
Rated Load
(Ω)
Frequency
(kHz)
Monopolar
Cut
PureCut
Blend 1
Blend 2
300
250
200
2000
2100
2100
300
300
300
384
384
384
Monopolar
Coag
Standard
Spray
120
80
7000
10100
300
384
384
Bipolar
Cut
Coag.
60
350
200
50
50
960
960
YUEH SHENG ELECTRONIC INDUSTRIAL CO., LTD. TAIWAN.
2F., No.29, Qiangnag St., Shilin District, Taipei City, Taiwan 11170
TEL: +886-2-2882-2212 FAX:+886-2-2881-8635
Figure 1.2 Label 1, located on the rear panel of ESU S300.
Installations
Refer to Sections 1.2.2 and 1.2.3 for the connections of accessories.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
9
MN-S300-EN-A02
1.2 General Descriptions and Operating Instructions
1.2.1
Control Panel
Figure 1.3
Control panel for ESU S300.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
10
MN-S300-EN-A02
1.
Output Power Setting Display for Mono-polar Cut
16.
Output Power Increase Key for Bipolar COAG
2.
Output Power Increase Key for Mono-polar Cut
17.
Output Power Decrease Key for Bipolar COAG
3.
Output Power Decrease Key for Mono-polar Cut
18.
Output Activation Indicator for Bipolar COAG
4.
Operating Mode Select Key for Mono-polar Cut
19.
Mono-polar / Bipolar Operating Mode Select Key
5.
Output Activation Indicator for Mono-polar Cut
20.
Mono-polar Operating Mode Indicator
6.
Output Power Setting Display for Mono-polar COAG
21.
Bipolar Operating Mode Indicator
7.
Output Power Increase Key for Mono-polar COAG
22.
Contact Impedance Monitor RESET
8.
Output Power Decrease Key for Mono-polar COAG
23.
Contact Impedance Display
9.
Operating Mode Select Key for Mono-polar COAG
24.
Patient Plate Circuitry FAULT Indicator
10.
Output Activation Indicator for Mono-polar COAG
25.
Pure Cut Operating Mode Indicator
11.
Output Power Setting Display for Bipolar-cut
26.
Blend I Cut Operating Mode Indicator
12.
Output Power Increase Key for Bipolar-cut
27.
Blend II Cut Operating Mode Indicator
13.
Output Power Decrease Key for Bipolar-cut
28.
Standard COAG Operating Mode Indicator
14.
Output Activation Indicator for Bipolar-cut
29.
Spray COAG Operating Mode Indicator
15.
Output Power Setting Display for Bipolar COAG
Table 1.1
Explanations of the control panel.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
11
MN-S300-EN-A02
1.2.2
Utility Panel
Figure 1.4 The utility panel of ESU Model S300.
1.
Connection 1 for Hand-Controlled Pencil
2.
Connection 2 for Hand-Controlled Pencil
3.
Connection for Foot-Controlled Pencil
4.
Connection for Disposable Patient Plate – round
5.
Connection for Disposable Patient Plate – flat
6.
Connector Select for Disposable Patient Plate
7.
Connection for Bipolar Forceps
Table 1.2
Explanations of the utility panel.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
12
MN-S300-EN-A02
Figure 1.5
Connections of accessories to the utility panel.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
13
MN-S300-EN-A02
1.2.3
Rear Panel
Figure 1.6 The rear panel of ESU S300.
1.
Power Supply Cord
2.
Main Fuses
3.
Connection for Ground
4.
Sound Level Adjustment
5.
Connection for Foot Control Switch
6.
Speaker Output
7.
Unit Identification Tag
8.
Main Power Switch
Table 1.3
Explanations of the rear panel.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
14
MN-S300-EN-A02
1.2.4
Operating Instructions
This medical device is only to be operated by qualified personnel. Study this
manual and be familiar with all functions before operating.
As with all other electrosurgical units, this generator produces
high-frequency, high-density electrical signals. Always practice with care.
Consult our authorized representative or our head office if technical
assistance is required.
a)
Precautions

Patient plate must be fully attached to the patient’s skin before generating
any mono-polar outputs from the ESU. False and/or improper connections
may result in skin-burn injuries. Always check for proper attachments and
connections if the patient is re-positioned during a surgery.

To minimize the possibility of HF burns during a surgery, the patient must
not contact any metal objects.

Avoid resting cables (such as cable for surgical pencil, patient plate, or
cables for bipolar forceps) on the patient to minimize the possibility of HF
burns to the patient.

As a general safety guideline, always start with a low output power setting.
Gradually increase the output power if necessary until the desired clinical
effect is achieved.

All reusable accessories must be cleaned, disinfected, and sterilized before
each usage. Refer to Section 2.1 for more details.

Do not reuse or re-sterile any disposable accessories. Reusing disposable
products may cause injuries and/or infections to the patient. Refer to Section
2.2 for more details.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
15
MN-S300-EN-A02
b)
Connecting a Patient Plate

Contact Impedance Monitor
When a dual-foil patient plate is used, the contact impedance monitor can
measure the contact quality (contact area) of the patient plate, and displays
the contact impedance in a three-digit format on the faceplate as shown in
Figure 1.7. A low impedance value indicates large contact surface. On the
other hand, a high impedance value suggests poor contact area, and is
hazardous to the patient. The typical range of contact impedance is between
40Ω and 110Ω.
The monitor also detects for false/improper connection of the patient-plate
cable regardless of the type of patient plate used.
When an unsafe status is detected (insufficient patient-plate contact quality
and/or improper cable connection), the generator immediately discontinues
its outputs, sounds an audible alarm, and flashes the “PLATE FAULT”
indicator. These warnings remain until all errors are corrected.
Reset key
Plate-fault indicator
Three-digit impedance display
Figure 1.7 The contact impedance monitor located on the front panel of ESU S300.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
16
MN-S300-EN-A02

To connect a patient plate to its cable, lift the tab on the clip of the cable and
insert a patient plate fully into the clip as shown in Figure 1.8. Lock the clip
securely and connect the other end of the cable to the utility panel of the
ESU as shown in Figure 1.5 in Section 1.2.2.
unlock
lock
3-meter cable
clip
patient plate
Figure 1.8
Connecting a patient plate.
Adjust the selector on the Utility Panel (as shown in Figure 1.4) to match
the chosen type of connector.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
17
MN-S300-EN-A02
c)
Operating Modes of ESU S300

There are two operating modes in S300: Mono-polar and Bipolar. To switch
between mono-polar and bipolar operating mode, press the
“mono-polar/bipolar operating mode select key”.
1.
Mono-polar/bipolar
Operating Mode Select Key.
2.
Mono-polar Operating Mode
Indicator.
3.
1
Figure 1.9
2
3
Bipolar Operating Mode
Indicator.
Bipolar/mono-polar mode select key and their corresponding indicators.
The S300 can be operated under mono-polar mode, bipolar mode, or under
both modes simultaneously. The operating-mode indicators illuminate
according to the selected modes.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
18
MN-S300-EN-A02
i)
Mono-polar Cutting
4
5
6
7
8
1. Mode Select Key.
2. Blend II CUT Operating Mode Indicator.
3. Blend I CUT Operating Mode Indicator.
4. Pure Cut Operating Mode Indicator.
3
5. Output Power Setting Display.
2
6. Output Power Increase Key.
7. Output Power Decrease Key.
1
8. Output Activation Idicator.
Figure 1.10 The control interface for CUT mode on the control panel of ESU S300.
–
–
Mono-polar cutting is achieved by using an electrosurgical pencil. The
pencil must be connected to the utility panel according to Figure 1.4
and Figure 1.5 in Section 1.2.2.
Press the “Mode Select Key”
to select between PURECUT,
BLEND I, BLEND II.
–
“Operating mode indicator” illuminates according to the selected
cutting mode.
–
Pure cut
Cuts tissue with little to no coagulation effect. Provides powerful and
fast cutting results.
–
Blend I Cut
Cuts tissue with low coagulation effect.
–
Blend II Cut
Cuts tissue with medium coagulation effect.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
19
MN-S300-EN-A02
–
–
–
–
–
Adjusting output power for cutting:
Press the “output power increase key” once to increase the output
power by one watt. Press the “output power decrease key” once to
decrease the output power by one watt. For large-step
increments/decrements, press and hold the key. The generator changes
the output power continuously until the key is released.
It is not necessary to discontinue the output to adjust its power. The
output can be activated and the adjustment of the output power can be
made simultaneously. However, as a safety feature, if the output power
is being adjusted while the output is being activated, the largest
increment-step is ten watts if “output power increase/decrease key” is
pressed and held. To further adjust the output power, release the
“output power increase/decrease key” and press again.
The output power is shown on the “output power setting display”.
The yellow “output activation indicator” illuminates when the CUT
output is activated.
The ESU generates an output tone from its speaker when the CUT
output is activated.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
20
MN-S300-EN-A02
ii) Mono-polar Coagulation
2
4
3
5
6 7
1. Mode Select Key.
2. Spray-Coagulation Operating Mode Indicator.
3. Standard-Coagulation Operating Mode Indicator.
4. Output Activation Indicator.
5. Output Power Setting Display.
6. Output Power Increase Key.
7. Output Power Decrease Key.
1
Figure 1.11
–
–
–
–
The control interface for mono-polar coagulation on the S300.
Mono-polar coagulation is achieved by using an electrosurgical pencil.
The pencil must be connected to the utility panel according to Figure
1.4 and Figure 1.5 in Section 1.2.2.
Press the “mode select key”
SPRAY coagulation.
to switch between STANDARD and
“Operating mode indicators” illuminate according to the selected
coagulation mode.
Standard coagulation
Standard coagulation uses high-density currents to coagulate tissues
quickly and efficiently. The Standard mode is usually used for smallto medium-hemorrhagic areas. Standard coagulation is sometimes
called desiccation or forced coagulation.
–
Spray coagulation
Spray coagulation uses very high voltages to generate “arching
currents”. The current travels through air, and is “sprayed” onto the
tissue, resulting in a large coagulation area. Therefore, the Spray mode
is generally used for large-hemorrhagic surfaces. Spray coagulation is
also known as fulguration.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
21
MN-S300-EN-A02
–
–
–
–
–
Adjusting the output power for coagulation:
Press the “output power increase key” once to increase the output
power by one watt. Press the “output power decrease key” once to
decrease the output power by one watt. For large-step
increments/decrements, press and hold the key. The generator changes
the output power continuously until the key is released.
It is not necessary to discontinue the output to adjust its power. The
output can be activated and the adjustment of the output power can be
made simultaneously. However, if the output power is being adjusted
while the output is generated, the largest increment-step is ten watts if
“output power increase/decrease key” is pressed and held. To further
adjust the output power, release the “output power increase/decrease
key” and press again.
The output power is shown on the “Output power setting display”.
The blue “output activation indicator” illuminates when the COAG
output is activated.
The ESU generates an output tone from its speaker when the COAG
output is activated.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
22
MN-S300-EN-A02
iii) Bipolar Operations
1
2 3
4
5
1. Output Activation Indicator for Cut Mode.
2. Cut Output Power Decrease Key.
3. Cut Output Power Increase Key.
4. Output Power Setting Display fro Cut Mode.
5. Output Power Setting Display for COAG Mode.
6
6. COAG Output Power Increase Key.
7. COAG Output Power Decrease Key.
7
8. Output Activation Indicator for COAG Mode.
8
Figure 1.13
–
–
–
–
–
–
–
Control interface for bipolar operating mode.
Bipolar operations are done with bipolar forceps. The outputs are
activated by a foot control switch.
This method of coagulation is widely used in microsurgeries, such as
neurosurgery, ophthalmologic surgery, and plastic surgery, etc.
Connect forceps to their cable according to Figure 1.18.
Connect the banana jack to the utility panel of the ESU as shown in
Figure 1.4 and Figure 1.5 in Section 1.2.2.
Adjusting the output power for bipolar operations:
Press the “output power increase key” (see Figure 1.13) once to
increase the output power by one watt. Press “output power decrease
key” once to decrease the output power by one watt. For large-step
increments/decrements, press and hold the key. The generator changes
the output power continuously until the key is released.
The output power is shown in “output power setting display”.
The blue “output activation indicator” illuminates when the output of
bipolar coagulation is activated.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
23
MN-S300-EN-A02
–
–
The yellow “output activation indicator” illuminates when the output
of bipolar cut is activated.
An audible activation tone is also generated by the ESU when a bipolar
output is activated.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
24
MN-S300-EN-A02
d)
Activating Outputs From the ESU
WARNING
Ensure that the patient plate is adhered onto the patient’s skin fully, and the cable
for patient plate is connected properly to the ESU before any mono-polar outputs
are activated. This precaution may be exempted if the surgery requires only
bipolar outputs, and the mono-polar output modes are switched OFF by using
the mode select key as mentioned in 1.2.4(c).
i)
Mono-polar outputs
– With a hand-controlled surgical pencil (disposable and reusable) :
1. Connect a three-prong hand-controlled surgical pencil to the
utility panel as shown in Figure 1.14 below.
utility panel of the ESU
3-prong hand-controlled
surgical pencil
3-meter cable
Figure 1.14
Connecting a hand-controlled surgical pencil to the ESU.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
25
MN-S300-EN-A02
2.
3.
Press the YELLOW button on the pencil to generate CUT output.
The yellow “Output Activation Indicator for Mono-polar Cut”
illuminates when the CUT output is activated.
Press the BLUE button on the pencil to generate a COAG output.
The blue “Output Activation Indicator for Mono-polar COAG”
illuminated when a COAG output is activated.
Figure 1.15 Location of CUT and COAG buttons on a hand-controlled surgical pencil
–
With a foot-controlled surgical pencil (available as a reusable
accessory):
1. Connect the foot-controlled surgical pencil as shown in Figure
1.16.
utility panel of the ESU
connector
foot-controlled surgical
pencil
3-meter cable
e)
Figure 1.16
Connecting a foot-controlled surgical pencil to the ESU.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
26
MN-S300-EN-A02
6. The outputs from a foot-controlled pencil are activated by a foot
control switch. Connect the foot control switch to the rear panel of
the ESU as shown in Figure 1.17. Once the foot control switch is
inserted fully to the ESU, secure the connection by turning the
metallic ring found on the connector of the foot control switch.
rear panel of the ESU
metallic ring for securing the connection
FOOT SWITCH
foot control switch
Figure 1.17
Connecting a foot control switch to the ESU.
7.
Step on the yellow CUT (left) pedal to activate “mono-polar
CUT” output. Step on the blue COAG (right) pedal to activate the
“mono-polar COAG” output.
8.
The foot-controlled outputs are sent to the patient via a
foot-controlled surgical pencil, regardless of the connection of the
hand-controlled pencil.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
27
MN-S300-EN-A02
ii) Bipolar outputs
– Connect bipolar forceps to the cables as shown in Figure 1.18 below.
3-meter cable
bipolar forceps
Figure 1.18
–
–
–
Connecting bipolar forceps to cables
Connect the banana jacks to the utility panel of the ESU as shown in
Figure 1.4 in Section 1.2.2.
Connect the foot control switch to the ESU according to Figure 1.17.
Step on the COAG pedal of the foot control switch to activate the
bipolar coagulation output.
FOOT SWITCH
mono-polar CUT pedal
Figure 1.19
–
bipolar/mono-polar COAG pedal
Foot control switch.
The outputs of bipolar operations (bipolar forceps) and
foot-controlled mono-polar operations (foot-controlled surgical
pencil) are activated by the same foot control switch. If the ESU is set
to operate both mono-polar and bipolar modes simultaneously, the
bipolar mode takes priority over the foot-controlled surgical pencil.
Pressing the switches on the foot control will not generate an output
from the foot-control switch. To utilize foot-controlled surgical pencil,
the bipolar mode must be switched off by using the
“mono-polar/bipolar mode select key”.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
28
MN-S300-EN-A02
Section Two – Accessories
2.1 Reusable Accessories
Reusable accessories include: hand-controlled surgical pencils, foot-controlled
surgical pencils, bipolar forceps, cable for bipolar forceps, cable for disposable patient
plate, reusable electrodes, and foot control switch.
Accessories that contact a patient’s internal tissues during a surgery, such as
hand-controlled surgical pencils, foot-controlled surgical pencils, and bipolar forceps,
must be cleaned, disinfected, and sterilized prior to each usage. Please contact our
head office or our authorized representative for the complete selections of accessories.
a)
Cleaning
 Wipe the accessories with a damped sponge or a piece of cloth with a mild
detergent solution. Do not use abrasive pads.
 Rinse with water and dry with a clean, soft cloth.
b)
Disinfecting
Gently wipe the accessories with a 75% methylated spirit and water solution, or
soak the accessories in a 75% alcohol solution for at least 30 minutes.
c)
Sterilizing
Three sterilization methods are available. However, it is important to verify that
the product can withstand the chosen sterilization method. Cables must not be
folded or coiled during sterilization or storage.

Moist heat (autoclave) – Please follow EN554 Standards (Sterilization of
Medical Device – Validation and Routine Control of Sterilization by Moist
Heat).


Ethylene Oxide (EO) – Please follow EN550 Standards (Sterilization of
Medical Devices – Validation and Routine Control of Ethylene Oxide
Sterilization).
Irradiation (Gamma) – Please follow EN552 Standards (Sterilization of
Medical Devices – Validation and Routine Control of Sterilization by
Irradiation)
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
29
MN-S300-EN-A02
d)
Care instructions

Electrodes, hand-control pencils, foot-control pencils, bipolar forceps and
their cables, can be sterilized by autoclave for up to 50 times. It is
recommended to discard these products after the maximum sterilization
count has been reached.

Electrosurgical pencils (hand-control and foot-control) and cable for bipolar
forceps must be inspected and tested after sterilization for damages which
may have occurred. An ohm meter should be used to measure the
conductivity. Please contact an authorized representative if replacements are
required.

Bipolar forceps and electrodes must be visually inspected for damages such
as apparent scratches or cracks. Using a defected accessory is hazardous to
both the user and the patient.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
30
MN-S300-EN-A02
2.2 Disposable Accessories




WARNING
Do not use sterilized products if the packaging has been damaged before
opening.
Do not reuse or re-sterile disposable accessories.
Do not use disposable products after the expiry date.
Store and use the disposable accessories according the instructions given on
their packaging.
Disposable accessories include: disposable hand-controlled pencils, disposable patient
plates, and cautery tip cleaners. Disposable accessories which contact, either directly
or indirectly, a patient’s internal tissues are sterilized and individually packaged. Do
not use the product if the packaging is damaged before opening. Please note the
expiration date printed near the seal of the packaging. Do not use the product after the
expiration date.
Disposable patient plates are adhered onto a patient’s skin; therefore, they do not
require sterilization. However, do not use the product after the expiration date. The
chemical characteristics of the hydro gel may be changed significantly after the
expiration date.
Please contact our authorized representative or our head office for a complete
selection of disposable accessories.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
31
MN-S300-EN-A02
Section Three – Maintenance & Service Instructions
3.1 Maintenance Instructions



WARNING
Turn off and unplug the generator before cleaning.
The calibration procedures should only be done by qualified technicians.
Contact our authorized representative or our head office if technical
assistance is needed.
3.1.1
Cleaning
Use a clean, soft, and slightly damped cloth to gently wipe down the exterior of the
generator. Do not use excessive water as doing so may damage the electrical circuits.
Use only dry cloth for the rear panel of the ESU. It is not necessary to sterile the ESU.
3.1.2
Output Power Calibrations
Each generator is calibrated to ensure its maximum performance prior to delivery.
However, calibration should be done at least once every two years to ensure accurate
output characteristics.
i) Calibration of Output Power for Mono-polar Cutting
Connect the ESU to an ESU analyzer. Set the ESU analyzer to 300Ω load setting.
Measure and record the output for every five-watt increments. Repeat for Blend I and
Blend II modes. If any discrepancies between the set power and the actual output
power occur, refer to the following table to adjust the appropriate variable resistor.
Potentiometers
U14
U9
U64
Output Types
Pure CUT
Blend I
Blend II
Table 3.1
Potentiometers and their corresponding cutting functions.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
32
MN-S300-EN-A02
ii) Calibration of Output Power for Mono-polar Coagulation
Connect the circuit and measure the coagulation outputs similar to the steps described
in Section 3.1.2.1. Refer to the following table if the calibration of the output power
for coagulation is needed.
Potentiometers
U8
U15
Output Types
Standard
Spray
Table 3.2
Potentiometers and their corresponding coagulation functions.
iii) Calibration of Output Power for Bipolar Coagulation
Connect the ESU analyzer to the bipolar coagulation output. The load impedance
must be set to 50Ω. Measure and record the output values for each five-watt
increments. Refer to Table 3.3 if calibration is needed.
Potentiometer
U16
Output Type
Bipolar COAG
Table 3.3
The potentiometer used to calibrate bipolar COAG.
iv) Calibration for Contact Impedance Monitor
The voltage at R98 should be 2.1V. An incorrect voltage may generate false alarms on
the Contact Impedance Monitor. This voltage can be adjusted with VR1.
v) Calibration for CUT and COAG Output Waveforms
This procedure must be done at least once every three years, and should only be
carried out by Yueh Sheng or an authorized representative. Please contact our head
office for details.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
33
MN-S300-EN-A02
3.2 Service Manual
WARNING
The following procedures should only be carried out by qualified
technicians. Please contact our head office or an authorized representative if
technical assistance or repair parts are needed.
3.2.1
Troubleshooting
Generator freezes / not responding to the buttons pressed
1. Turn the generator OFF. Reset the CPU by switch S1 OFF and ON again.
2. Switch the generator back to ON.
3. If steps 1 and 2 do not solve the problem, remove the front panel and inspect the
conductivity of the push buttons.
4. Replace crystal Y1 if all of the above steps fail to correct the problem.
Alarm ON continuously
1.
2.
3.
Check for proper attachment of the patient plate.
Inspect the patient plate cable for conductivity.
Refer to Section 3.1.2.4 for calibration procedures for Contact Impedance
Monitor.
No outputs
1. Inspect all fuses and connections.
2. Inspect the conductivity of all accessories.
3. Visually inspect for damages on all PCB’s. Repair any broken tracks.
4. Inspect transistor Q4 on the motherboard. This transistor acts a switch to hand
control outputs.
5. Inspect the diode bridge on the driver board. Replace if they are shorted.
6. Test the MOSFET’s Q8, Q5, Q4, and Q9 on the driver board. Replace if
necessary. If a MOSFET is replaced, inspect the diode Z18V and the 5.6Ω
resistor which are in series with the replaced MOSFET.
7. If Steps 1 to 6 are performed and the generator is still not producing outputs,
replace the voltage regulator YS_M04 on the motherboard. This voltage
regulator is a custom-made part, and can only be obtained from Yueh Sheng.
Please contact our head office or an authorized representative if YS_M04 is
needed.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
34
MN-S300-EN-A02
8.
If any of the parts listed in Step 5 to Step 7 are replaced, measure the standby
voltage at D71. The voltage reading should be 16V to 20V. If this reading is
incorrect, use VR5 to adjust to the correct voltage level.
9.
Please contact our head office or an authorized representative if the above steps
fail to correct the problem. Do not attempt to change any components other than
those listed above. Ensure that the replaced parts are of equivalent specifications
to the original parts.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
35
MN-S300-EN-A02
3.2.2
Frequently-Used Test Points
Test Points
Description
Reading
Calibration
D71
Standby Voltage
16V ~ 20V
VR5
R98
Transducer Voltage
2.1V
VR1
P+5V
5VDC
5V ± 5%
4575
P+12V
12VDC
12V ± 5%
MIC4576BT
P+15V
15VDC
15V ± 5%
Q18
TP5
960kHz Signal
960kHz
Y1 3.84MHz
TP6
384kHz Signal
384kHz
Y1 3.84MHz
TP7
38.4kHz Signal
38.4kHz
Y1 3.84MHz
TP8
19.2kHz Signal
19.2kHz
Y1 3.84MHz
Actual Output
Pure CUT Output Power
Analyzer Reading
U14
Actual Output
Blend I CUT Output Power
Analyzer Reading
U9
Actual Output
Blend II CUT Output Power
Analyzer Reading
U64
Actual Output
Spray COAG Output Power
Analyzer Reading
U15
Actual Output
Standard COAG Output Power
Analyzer Reading
U8
Actual Output
Bipolar COAG Output Power
Analyzer Reading
U16
Table 3.4
Frequently-used test points and their corresponding components.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
36
MN-S300-EN-A02
3.3 Output Characteristics
Figure 3.1 Output characteristics for mono-polar standard coagulation.
Figure 3.2 Output characteristics for mono-polar spray coagulation.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
37
MN-S300-EN-A02
Figure 3.3 Output characteristics for bipolar cut.
Figure 3.4 Output characteristics for bipolar coagulation.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
38
MN-S300-EN-A02
Figure 3.5 Output characteristics for mono-polar pure cut.
Figure 3.6 Output characteristics for mono-polar blend I cut.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
39
MN-S300-EN-A02
Figure 3.7 Output characteristics for mono-polar blend II cut.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
40
MN-S300-EN-A02
3.4 Diagrams & Schematics
3.4.1
Constitutional Chart
Figure 3.8
Constitutional chart for ESU S300.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
41
MN-S300-EN-A02
3.4.2
Circuit Diagrams – Computer Board
YS_S300P.JOB 10/31,00
Figure 3.9
Circuit diagram of CPU board.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
42
MN-S300-EN-A02
3.4.3
Circuit Diagram – Driver Board
YS-3001 05/03/07
Figure 3.10 Circuit Diagram – Driver board.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
43
MN-S300-EN-A02
3.4.4
Circuit Diagram – Motherboard
YS-S300M.JOB 8/23/02
Figure 3.11 Circuit Diagram – Motherboard.
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
44
MN-S300-EN-A02
3.5.1
3.5 Bill of Materials
Bill of Materials – Computer Board
CIRCUIT
REFERENCE
2001
C1
2002
C2
2003
C3
2004
C4
2005
C5
2006
C6
2007
C7
2008
C8
2009
C9
2010
C10
2011
C11
2012
C12
2013
C13
2014
C14
2015
C15
2016
C16
2017
C17
2018
C18
2019
C19
2020
C20
2021
C21
2022
C22
2023
C23
2024
C24
2025
C25
2026
C26
2027
C27
2028
C30
2029
C31
2030
C32
2031
D16
2032
D17
2033
D18
2034
D25
2035
JP1
2036
JP2
2037
KY1
2038
KY2
2039
KY3
2040
KY4
2041
KY5
2042
KY6
2043
KY7
2044
KY8
2045
KY9
BOM - Computer Board (Model:S300)
YESNG
CIRCUIT
DISCRIPTION
DISCRIPTION
PART NO.
REFERENCE
Cap.,27pF,50v,Cer.
3140-5027 2046
KY10 Button
Cap.,27pF,50v,Cer.
3140-5027 2047
KY11 Button
Cap.,2.2μF,35v,Tan.
3120-4522 2048
KY12 Button
Cap.,10μF,25v,Tan.
3120-3610 2049
L1
L_0.047mH
Cap.,47pF,50v,Cer.
3140-5047 2050
L2
L_0.047mH
Cap.,10μF,25v,Tan.
3120-3610 2051
L3
L_0.047mH
Cap.,10μF,25v,Tan.
3120-3610 2052
LD1 LED,LAMP,GREEN
Cap.,2.2μF,35v,Tan.
3120-4522 2053
LD2 LED,LAMP,YELLOW
Cap.,0.1μF,50v,Cer.
3140-5410 2054
LD3 LED,LAMP,YELLOW
Cap.,0.1μF,50v,Cer.
3140-5410 2055
LD4 LED,LAMP,GREEN
Cap.,0.1μF,50v,Cer.
3140-5410 2056
LD5 LED,LAMP,GREEN
Cap.,0.1μF,50v,Cer.
3140-5410 2057
LD6 LED,LAMP,GREEN
Cap.,0.1μF,50v,Cer.
3140-5410 2058
LD7 LED,LAMP,YELLOW
Cap.,0.1μF,50v,Cer.
3140-5410 2059
LD8 LED,LAMP,YELLOW
Cap.,0.1μF,50v,Cer.
3140-5410 2060
LD9 LED,LAMP,YELLOW
Cap.,0.1μF,50v,Cer.
3140-5410 2061
Q1
Q_9012_EBC
Cap.,220μF,25v,Elect.
3110-3722 2062
Q2
Q_9012_EBC
Cap.,10μF,25v,Tan.
3120-3610 2063
Q3
Q_9012_EBC
Cap.,10μF,25v,Tan.
3120-3610 2064
Q4
Q_9012_EBC
Cap.,10μF,25v,Tan.
3120-3610 2065
Q5
Q_9012_EBC
Cap.,10pF,50v,Cer.
3140-5010 2066
Q6
Q_9012_EBC
Cap.,10μF,25v,Tan.
3120-3610 2067
Q7
Q_9012_EBC
Cap.,10μF,25v,Tan.
3120-3610 2068
Q8
Q_9012_EBC
Cap.,470μF,16v,Elect.
3110-2747 2069
Q9
Q_9012_EBC
Cap.,6.8μF,35v,Tan.
3120-4568 2070
Q10 Q_9012_EBC
Cap.,0.1μF,50v,Cer.
3140-5410 2071
Q11 Q_9014_EBC
Cap.,0.1μF,50v,Cer.
3140-5410 2072
Q12 Q_9014_EBC
Cap.,10μF,25v,Tan.
3120-3722 2073
Q13 Q_9014_EBC
Cap.,10μF,25v,Tan.
3120-3722 2074
Q14 Q_2SC1959
Cap.,470μF,16v,Elect.
3110-2747 2075
Q15 Q_2SC1959
LED,LAMP,RED
4010-1433 2076
R1
Res.,1k,1%,1/4w,Metal
LED,LAMP,YELLOW
4010-1431 2077
R2
Res.,1k,1%,1/4w,Metal
LED,LAMP,GREEN
4010-1432 2078
R3
Res.,1k,1%,1/4w,Metal
LED,LAMP,RED
4010-1433 2079
R4
Res.,1k,1%,1/4w,Metal
CABLE_26Pin
4027-0001 2080
R5
Res.,1k,1%,1/4w,Metal
Q_8054
3411-8054 2081
R6
Res.,1k,1%,1/4w,Metal
Button
4005-0007 2082
R7
Res.,1k,1%,1/4w,Metal
Button
4005-0007 2083
R8
Res.,1k,1%,1/4w,Metal
Button
4005-0007 2084
R9
Res.,220,1%,1/4w,Metal
Button
4005-0007 2085
R10 Res.,220,1%,1/4w,Metal
Button
4005-0007 2086
R11 Res.,220,1%,1/4w,Metal
Button
4005-0007 2087
R12 Res.,220,1%,1/4w,Metal
Button
4005-0007 2088
R13 Res.,220,1%,1/4w,Metal
Button
4005-0007 2089
R14 Res.,220,1%,1/4w,Metal
Button
4005-0007 2090
R15 Res.,220,1%,1/4w,Metal
page 1 of 3
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
45
YESNG
PART NO.
4005-0007
4005-0007
4005-0007
3200-1211
3200-1211
3200-1211
4010-1432
4010-1431
4010-1431
4010-1432
4010-1432
4010-1432
4010-1431
4010-1431
4010-1431
3411-9012
3411-9012
3411-9012
3411-9012
3411-9012
3411-9012
3411-9012
3411-9012
3411-9012
3411-9012
3411-9014
3411-9014
3411-9014
3411-1959
3411-1959
3082-4100
3082-4100
3082-4100
3082-4100
3082-4100
3082-4100
3082-4100
3082-4100
3082-3220
3082-3220
3082-3220
3082-3220
3082-3220
3082-3220
3082-3220
MN-S300-EN-A02
CIRCUIT
REFERENCE
2091
R16
2092
R17
2093
R18
2094
R19
2095
R20
2096
R21
2097
R22
2098
R23
2099
R24
2100
R25
2101
R26
2102
R27
2103
R28
2104
R29
2105
R30
2106
R31
2107
R32
2108
R33
2109
R34
2110
R35
2111
R36
2112
R37
2113
R38
2114
R39
2115
R40
2116
R41
2117
R42
2118
R43
2119
R44
2120
R45
2121
R46
2122
R47
2123
R48
2124
R49
2125
R50
2126
R51
2127
R52
2128
R53
2129
R54
2130
R55
2131
R56
2132
R57
2133
R58
2134
R59
2135
R60
BOM - Computer Board (Model:S300)
YESNG
CIRCUIT
DISCRIPTION
DISCRIPTION
PART NO.
REFERENCE
Res.,220,1%,1/4w,Metal
3082-3220 2136
R61 Res.,150,1%,1/8w,Metal
Res.,220,1%,1/4w,Metal
3082-3220 2137
R62 Res.,4.7k,1%,1/8w,Metal
Res.,33,1%,1/4w,Metal
3082-2330 2138
R63 Res.,4.7k,1%,1/8w,Metal
Res.,33,1%,1/4w,Metal
3082-2330 2139
R64 Res.,12,1%,1/4w,Metal
Res.,33,1%,1/4w,Metal
3082-2330 2140
R65 Res.,10k,1%,1/4w,Metal
Res.,33,1%,1/4w,Metal
3082-2330 2141
R66 Res.,10k,1%,1/4w,Metal
Res.,33,1%,1/4w,Metal
3082-2330 2142
R67 Res.,4.7k,1%,1/4w,Metal
Res.,33,1%,1/4w,Metal
3082-2330 2143
R68 Res.,10k,1%,1/4w,Metal
Res.,33,1%,1/4w,Metal
3082-2330 2144
R69 Res.,10k,1%,1/4w,Metal
Res.,100k,1%,1/4w,Metal
3082-6100 2145
R70 Res.,4.7k,1%,1/4w,Metal
Res.,100k,1%,1/4w,Metal
3082-6100 2146
R71 Res.,4.7k,1%,1/8w,Metal
Res.,100k,1%,1/4w,Metal
3082-6100 2147
R72 Res.,220,1%,1/4w,Metal
Res.,100k,1%,1/4w,Metal
3082-6100 2148
RP1 Res.,2.2k,5%,1/8w,8p,Net
Res.,100k,1%,1/4w,Metal
3082-6100 2149
RP2 Res.,1.2k,5%,1/8w,8p,Net
Res.,100k,1%,1/4w,Metal
3082-6100 2150
RR1 Res.,470,5%,1/8w,8p,Net
Res.,1k,1%,1/4w,Metal
3082-4100 2151
RR2 Res.,100k,5%,1/8w,9p,Net
Res.,1k,1%,1/4w,Metal
3082-4100 2152
S1
IC_PC817_SHARP
Res.,1k,1%,1/4w,Metal
3082-4100 2153
S2
IC_PC817_SHARP
Res.,1k,1%,1/4w,Metal
3082-4100 2154
S3
IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2155
S4
IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2156
S5
IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2157
S6
IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2158
S7
IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2159
S8
IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2160
S9
IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2161
S10 IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2162
S11 IC_PC817_SHARP
Res.,10k,1%,1/4w,Metal
3082-5100 2163
S12 IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2164
S13 IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2165
S14 IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2166
S15 IC_PC817_SHARP
Res.,220,1%,1/4w,Metal
3082-3220 2167
U1
IC_74LS138
Res.,220,1%,1/4w,Metal
3082-3220 2168
U2
IC_74LS164
Res.,2k,1%,1/4w,Metal
3082-4200 2169
U3
IC_74HC574
Res.,2k,1%,1/4w,Metal
3082-4200 2170
U4
IC_74HC574
Res.,1k,1%,1/4w,Metal
3082-4100 2171
U5
IC_74HC574
Res.,1k,1%,1/4w,Metal
3082-4100 2172
U6
IC_74HC574
Res.,20k,1%,1/4w,Metal
3082-5200 2173
U7
IC_4538
Res.,4.7k,1%,1/8w,Metal
3081-4470 2174
U8
IC_74LS00
Res.,4.7k,1%,1/4w,Metal
3082-4470 2175
U19 Res.,10k,Var.,10%,9W
Res.,1.5M,1%,1/8w,Metal
3081-7150 2176
U20 IC_7662
Res.,10k,1%,1/4w,Metal
3082-5100 2177
U21 IC_AT89C51
Res.,4.7k,1%,1/4w,Metal
3082-4470 2178
U22 IC_93LC46
Res.,150,1%,1/8w,Metal
3081-3150 2179
U23 IC_CA3130
Res.,150,1%,1/8w,Metal
3081-3150 2180
U24 IC_AD7541
page 2 of 3
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
46
YESNG
PART NO.
3081-3150
3081-4470
3081-4470
3082-2120
3082-5100
3082-5100
3082-4470
3082-5100
3082-5100
3082-4470
3081-4470
3082-3220
3040-8422
3040-8412
3040-8347
3040-9610
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3504-9817
3516-7138
3514-7164
3520-7574
3520-7574
3520-7574
3520-7574
3516-4538
3514-7400
3070-2510
3508-9662
3540-8951
3508-9346
3508-3130
3518-9541
MN-S300-EN-A02
CIRCUIT
REFERENCE
2181
U31
2182
VR1
2183
Y1
2184
2185
BOM - Computer Board (Model:S300)
YESNG
CIRCUIT
DISCRIPTION
PART NO.
REFERENCE
IC_PIC16C17/05
3518-6671
Res.,100k,Ver.,10%,9W
3070-2610
Crystal,10MHz
4018-0002
Display,5501
4010-5501
Display,7803
4010-7803
page 3 of 3
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
47
DISCRIPTION
YESNG
PART NO.
MN-S300-EN-A02
3.5.2
Bill of Materials – Motherboard
BOM - Mother Board (Model:300)
YESNG
CIRCUIT
CIRCUIT
DISCRIPTION
PART NO.
REFERENCE
REFERENCE
1001
7812 IC_7812
3503-2012 1046
C64
1002
BD17 Bridge-Diode,FBU4K
3353-4000 1047
C65
1003
C1
Cap.,27pF,50v,Cer.
3140-5027 1048
C67
1004
C2
Cap.,27pF,50v,Cer.
3140-5027 1049
C68
1005
C3
Cap.,0.001μF,50v,Film
3170-5210 1050
C72
1006
C4
Cap.,0.001μF,50v,Film
3170-5210 1051
C73
1007
C5
Cap.,0.001μF,50v,Film
3170-5210 1052
C74
1008
C8
Cap.,0.15μF,100v,Film
3170-7415 1053
C75
1009
C9
Cap.,0.047μF,100v,Film
3130-7347 1054
C76
1010
C10 Cap.,220μF,25v,Elect.
3110-3722 1055
C77
1011
C11 Cap.,47pF,50v,Cer.
3140-5047 1056
C78
1012
C12 Cap.,47pF,50v,Cer.
3140-5047 1057
C79
1013
C13 Cap.,0.001μF,50v,Film
3170-5210 1058
C80
1014
C15 Cap.,470μF,35v,Elect.
3110-4747 1059
C81
1015
C16 Cap.,2200μF,35v,Elect.
3110-4822 1060
C82
1016
C17 Cap.,0.0027μF,100v,Film
3130-7272 1061
C83
1017
C18 Cap.,10μF,25v,Tan.
3120-3610 1062
C84
1018
C19 Cap.,0.27μF,100v,Film
3170-7427 1063
C85
1019
C20 Cap.,0.27μF,100v,Film
3170-7427 1064
C86
1020
C21 Cap.,47μF,50v,Elect.
3110-5647 1065
C87
1021
C24 Cap.,0.0033μF,3kv,Film
3170-6233 1066
C88
1022
C26 Cap.,0.01μF,250v,Film
3170-9310 1067
C89
1023
C34 Cap.,470μF,35v,Elect.
3110-4747 1068
C90
1024
C39 Cap.,0.022μF,100v,Film
3130-7223 1069
C93
1025
C40 Cap.,0.15μF,100v,Film
3130-7415 1070
C94
1026
C43 Cap.,0.1μF,50v,Cer.
3140-5410 1071
C96
1027
C44 Cap.,0.033μF,100v,Film
3130-7333 1072
C98
1028
C45 Cap.,0.047μF,100v,Film
3130-7347 1073
C99
1029
C46 Cap.,0.1μF,50v,Cer.
3140-5410 1074
C100
1030
C47 Cap.,0.27μF,100v,Film
3170-7427 1075
C101
1031
C48 Cap.,0.27μF,100v,Film
3170-7427 1076
CX2
1032
C50 Cap.,0.0056μF,1.6kv,Film
3171-5265 1077
CX3
1033
C51 Cap.,0.0056μF,1.6kv,Film
3171-5265 1078
D0
1034
C52 Cap.,0.1μF,50v,Cer.
3140-5410 1079
D1
1035
C53 Cap.,0.1μF,50v,Cer.
3140-5410 1080
D4
1036
C54 Cap.,0.1μF,50v,Cer.
3140-5410 1081
D6
1037
C55 Cap.,0.1μF,50v,Cer.
3140-5410 1082
D7
1038
C56 Cap.,0.1μF,50v,Cer.
3140-5410 1083
D8
1039
C57 Cap.,0.1μF,50v,Cer.
3140-5410 1084
D10
1040
C58 Cap.,0.1μF,50v,Cer.
3140-5410 1085
D13
1041
C59 Cap.,0.1μF,50v,Cer.
3140-5410 1086
D21
1042
C60 Cap.,0.1μF,50v,Cer.
3140-5410 1087
D22
1043
C61 Cap.,0.1μF,50v,Cer.
3140-5410 1088
D23
1044
C62 Cap.,0.1μF,50v,Cer.
3140-5410 1089
D25
1045
C63 Cap.,0.1μF,50v,Cer.
3140-5410 1090
D27
page 1 of 4
DISCRIPTION
Cap.,0.1μF,50v,Cer.
Cap.,0.1μF,50v,Cer.
Cap.,0.0033μF,1.6kv,Film
Cap.,0.0022μF,50v,Film
Cap.,0.001μF,50v,Film
Cap.,1000μF,250v,Elect.
Cap.,1000μF,250v,Elect.
Cap.,0.1μF,630v,Film
Cap.,0.01μF,250v,Film
Cap.,0.1μF,50v,Cer.
Cap.,0.1μF,50v,Cer.
Cap.,0.1μF,50v,Cer.
Cap.,0.1μF,50v,Cer.
Cap.,0.1μF,50v,Cer.
Cap.,0.1μF,50v,Cer.
Cap.,10μF,25v,Tan.
Cap.,3.3μF,250v,Paper.
Cap.,470μF,35v,Elect.
Cap.,470μF,35v,Elect.
Cap.,470μF,35v,Elect.
Cap.,470μF,35v,Elect.
Cap.,470μF,35v,Elect.
Cap.,0.047μF,100v,Film
Cap.,0.27μF,100v,Film
Cap.,0.27μF,100v,Film
Cap.,10μF,25v,Tan.
Cap.,0.1μF,50v,Cer.
Cap.,0.1μF,50v,Cer.
Cap.,0.1μF,50v,Cer.
Cap.,0.1μF,50v,Cer.
Cap.,470μF,35v,Elect.
Cap.,1000μF,35v,Elect.
Diode,1N4148
Diode,BY399
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,FR105
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
48
YESNG
PART NO.
3140-5410
3140-5410
3171-5233
3170-5222
3170-5210
3110-9810
3110-9810
3171-1410
3170-9310
3140-5410
3140-5410
3140-5410
3140-5410
3140-5410
3140-5410
3120-3610
3160-9533
3110-4747
3110-4747
3110-4747
3110-4747
3110-4747
3130-7347
3170-7427
3170-7427
3120-3610
3140-5410
3140-5410
3140-5410
3140-5410
3110-4747
3110-4810
3311-4148
3322-0399
3311-4148
3311-4148
3311-4148
3311-4148
3311-4148
3311-4148
3311-4148
3311-4148
3311-4148
3311-4148
3331-0105
MN-S300-EN-A02
CIRCUIT
REFERENCE
1091
D28
1092
D29
1093
D30
1094
D31
1095
D32
1096
D33
1097
D38
1098
D39
1099
D40
1100
D41
1101
D42
1102
D43
1103
D44
1104
D45
1105
D48
1106
D49
1107
D52
1108
D53
1109
D54
1110
D55
1111
D56
1112
D57
1113
D58
1114
D59
1115
D61
1116
D62
1117
D63
1118
D64
1119
D65
1120
D69
1121
D70
1122
D71
1123
D72
1124
D73
1125
D74
1126
D75
1127
D76
1128
D77
1129
D78
1130
D79
1131
D80
1132
D82
1133
D83
1134
D84
1135
D85
DISCRIPTION
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,FR105
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,RL205
Diode,51
Diode,FR607
Diode,BY399
Diode,RL205
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
BOM - Mother Board (Model:300)
YESNG
CIRCUIT
PART NO.
REFERENCE
3311-4148
1136
D86
3311-4148 1137
D87
3311-4148 1138
D88
3311-4148 1139
D89
3311-4148 1140
D90
3311-4148 1141
D91
3311-4148
1142
D92
3311-4148
1143
D93
3311-4148
1144
D94
3311-4148
1145
D95
3311-4148
1146
H1
3311-4148
1147
J2
3311-4148
1148
J3
3311-4148
1149
J4
3311-4148
1150
JP1
3331-0105 1151
JP5
3311-4148
1152
JP5
3311-4148
1153
L1
3311-4148
1154
L2
3311-4148
1155
P3
3311-4148
1156
P7
3311-4148
1157
3311-4148 1158
Q2
3311-4148 1159
Q4
3311-4148 1160
Q6
3311-4148 1161
Q7
3311-4148 1162
Q8
3311-4148 1163
Q9
3311-4148 1164
Q10
3321-0205 1165
Q12
3321-0051 1166
Q13
3332-0607 1167
Q18
3322-0399 1168
Q19
3321-0205 1169
Q21
3311-4148 1170
Q22
3311-4148 1171
Q23
3311-4148 1172
Q25
3311-4148 1173
Q26
3311-4148
1174
3311-4148
1175
R1
3311-4148 1176
R2
3311-4148
1177
R3
3311-4148
1178
R4
3311-4148 1179
R5
3311-4148 1180
R6
page 2 of 4
Diode,1N4148
LED,LAMP,RED
LED,LAMP,RED
LED,LAMP,RED
LED,LAMP,RED
LED,LAMP,RED
Diode,1N4148
Diode,1N4148
Diode,1N4148
Diode,1N4148
Q_S6035W
Socket_2P
Socket_3P
Socket_3P
Socket_3P
Socket_3P
Socket_2P
L_0.068mH
L_0.068mH
Socket_26P
Socket_7P
YESNG
PART NO.
3311-4148
4010-1433
4010-1433
4010-1433
4010-1433
4010-1433
3311-4148
3311-4148
3311-4148
3311-4148
3423-6035
4015-0020
4015-0030
4015-0030
4015-0030
4015-0030
4015-0020
3200-1201
3200-1201
4003-0260
4015-0070
Q_2SC1959
Q_2SC1384-R
Q_2SC1959
Q_2SC1959
Q_2SC1959
Q_2SC1959
Q_MPSA13
Q_2SC1959
Q_2SC1959
IC_7815
Q_D6020L
Q_2SC1959
Q_2SC1959
Q_2SC1959
Q_2SC1959
Q_2SC1959
3411-1959
3411-1384
3411-1959
3411-1959
3411-1959
3411-1959
3411-0013
3411-1959
3411-1959
3503-2015
3422-6020
3411-1959
3411-1959
3411-1959
3411-1959
3411-1959
Res.,10k,1%,1/4w,Metal
Res.,10M,1%,1/4w,Metal
Res.,10k,1%,1/4w,Metal
Res.,10k,1%,1/4w,Metal
Res.,4.7k,1%,1/4w,Metal
Res.,4.7k,1%,1/4w,Metal
3082-5100
3082-8100
3082-5100
3082-5100
3082-4470
3082-4470
DISCRIPTION
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
49
MN-S300-EN-A02
CIRCUIT
REFERENCE
1181
R7
1182
R8
1183
R10
1184
R11
1185
R13
1186
R14
1187
R15
1188
R16
1189
R17
1190
R18
1191
R19
1192
R20
1193
R21
1194
R22
1195
R23
1196
R24
1197
R25
1198
R32
1199
R33
1200
R34
1201
R35
1202
R36
1203
R37
1204
R38
1205
R42
1206
R43
1207
R45
1208
R46
1209
R64
1210
R73
1211
R74
1212
R75
1213
R76
1214
R77
1215
R78
1216
R79
1217
R80
1218
R81
1219
R82
1220
R83
1221
R84
1222
R85
1223
R86
1224
R87
1225
R88
BOM - Mother Board (Model:300)
YESNG
CIRCUIT
DISCRIPTION
PART NO.
REFERENCE
Res.,4.7k,1%,1/4w,Metal
3082-4470 1226
R89
Res.,4.7k,1%,1/4w,Metal
3082-4470 1227
R90
Res.,27,1%,1/4w,Metal
3082-2270 1228
R91
Res.,27,1%,1/4w,Metal
3082-2270 1229
R92
Res.,16.2,1%,1/4w,Metal
3082-2162 1230
R93
Res.,5.6,1%,1/2w,Metal
3083-1560 1231
R94
Res.,8.2k,1%,1/4w,Metal
3082-4820 1232
R95
Res.,8.2k,1%,1/4w,Metal
3082-4820 1233
R96
Res.,8.2k,1%,1/4w,Metal
3082-4820 1234
R97
Res.,1k,1%,1/4w,Metal
3082-4100 1235
R100
Res.,18k,1%,1/4w,Metal
3082-5180 1236
R102
Res.,8.2k,1%,1/4w,Metal
3082-4820 1237
R103
Res.,8.2k,1%,1/4w,Metal
3082-4820 1238
R104
Res.,8.2k,1%,1/4w,Metal
3082-4820 1239
R105
Res.,3k,1%,1/4w,Metal
3082-4300 1240
R107
Res.,39k,1%,1/4w,Metal
3082-5390 1241
R108
Res.,1k,1%,1/4w,Metal
3082-4100 1242
R109
Res.,470,1%,1/4w,Metal
3082-3470 1243
R114
Res.,15k,1%,1/4w,Metal
3082-5150 1244
R120
Res.,820,1%,1/4w,Metal
3082-3820 1245
R121
Res.,220,1%,1/4w,Metal
3082-3220 1246
R122
Res.,220,1%,1/4w,Metal
3082-3220 1247
R123
Res.,1.8k,1%,1/4w,Metal
3082-4180 1248
R125
Res.,1.8k,1%,1/4w,Metal
3082-4180 1249
R126
Res.,2k,1%,1/4w,Metal
3082-4200 1250
R127
Res.,3k,5%,10w,W.W.
3030-9430 1251
R128
Res.,4.7k,1%,1/4w,Metal
3082-4470 1252
R129
Res.,300,1%,1/4w,Metal
3082-3300 1253
R130
Res.,560,1%,1/2w,Metal
3083-3560 1254
R131
Res.,0.1,5%,5w,W.W.
3030-7010 1255
R132
Res.,300,5%,5w,W.W.
3030-7330 1256
R133
Res.,8.2k,1%,1/4w,Metal
3082-4820 1257
R134
Res.,33,1%,1/2w,Metal
3083-2330 1258
R136
Res.,12,1%,1/4w,Metal
3082-2120 1259
R137
Res.,12,1%,1/4w,Metal
3082-2120 1260
R138
Res.,12,1%,1/4w,Metal
3082-2120 1261
R139
Res.,12,1%,1/4w,Metal
3082-2120 1262
R141
Res.,12,1%,1/4w,Metal
3082-2120 1263
R142
Res.,12,1%,1/4w,Metal
3082-2120 1264
R143
Res.,12,1%,1/4w,Metal
3082-2120 1265
R144
Res.,12,1%,1/4w,Metal
3082-2120 1266
R145
Res.,12,1%,1/4w,Metal
3082-2120 1267
R146
Res.,12,1%,1/4w,Metal
3082-2120 1268
R147
Res.,12,1%,1/4w,Metal
3082-2120 1269
R148
Res.,12,1%,1/4w,Metal
3082-2120 1270
page 3 of 4
DISCRIPTION
Res.,12,1%,1/4w,Metal
Res.,12,1%,1/4w,Metal
Res.,12k,1%,1/4w,Metal
Res.,8.2k,1%,1/4w,Metal
Res.,300,1%,1/4w,Metal
Res.,1.8k,1%,1/4w,Metal
Res.,2k,1%,1/4w,Metal
Res.,5.6,1%,1/2w,Metal
Res.,2k,1%,1/4w,Metal
Res.,2k,1%,1/4w,Metal
Res.,2k,1%,1/4w,Metal
Res.,4.7k,1%,1/4w,Metal
Res.,4.7k,1%,1/4w,Metal
Res.,1k,1%,1/4w,Metal
Res.,1k,1%,1/4w,Metal
Res.,1k,1%,1/4w,Metal
Res.,51,1%,1/4w,Metal
Res.,36k,1%,1/4w,Metal
Res.,1.5k,1%,1/2w,Metal
Res.,200,1%,1/4w,Metal
Res.,1k,1%,1/2w,Metal
Res.,5k,5%,5w,W.W.
Res.,4.7k,1%,1/4w,Metal
Res.,10k,1%,1/4w,Metal
Res.,10k,1%,1/4w,Metal
Res.,10k,1%,1/4w,Metal
Res.,10k,1%,1/4w,Metal
Res.,10k,1%,1/4w,Metal
Res.,4.7k,1%,1/4w,Metal
Res.,12,1%,1/4w,Metal
Res.,4.7k,1%,1/4w,Metal
Res.,4.7k,1%,1/4w,Metal
Res.,4.7k,1%,1/4w,Metal
Res.,200,1%,1/4w,Metal
Res.,13k,1%,1/4w,Metal
Res.,1.5k,1%,1/2w,Metal
Res.,18k,1%,1/4w,Metal
Res.,3.3k,1%,1/4w,Metal
Res.,39,5%,2w,Oxide
Res.,4.7k,1%,1/4w,Metal
Res.,4.7k,1%,1/4w,Metal
Res.,1.8k,1%,1/4w,Metal
Res.,220,1%,1/4w,Metal
Res.,1.8k,1%,1/4w,Metal
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
50
YESNG
PART NO.
3082-2120
3082-2120
3082-5120
3082-4820
3082-3300
3082-4180
3082-4200
3083-1560
3082-4200
3082-4200
3082-4200
3082-4470
3082-4470
3082-4100
3082-4100
3082-4100
3082-2510
3082-5360
3083-4150
3082-3200
3083-4100
3030-9450
3082-4470
3082-5100
3082-5100
3082-5100
3082-5100
3082-5100
3082-4470
3082-2120
3082-4470
3082-4470
3082-4470
3082-3200
3082-5130
3083-4150
3082-5180
3082-4330
3020-6239
3082-4470
3082-4470
3082-4180
3082-3220
3082-4180
MN-S300-EN-A02
CIRCUIT
REFERENCE
1271
R149
1272
R158
1273
R159
1274
R160
1275
R161
1276
R162
1277
RE10
1278
RE11
1279
RE12
1280
RE16
1281
RE17
1282
RE18
1283
RE19
1284
RE20
1285
RE21
1286
RE22
1287
U1
1288
U2
1289
U3
1290
U4
1291
U5
1292
U6
1293
U7
1294
U8
1295
U9
1296
U10
1297
U11
1298
U12
1299
U13
1300
U14
1301
U15
1302
U16
1303
U18
1304
U19
1305
U20
1306
U22
1307
U23
1308
U24
1309
U25
1310
U27
1311
U33
1312
U38
1313
U50
1314
U51
1315
U52
BOM - Mother Board (Model:300)
YESNG
CIRCUIT
DISCRIPTION
PART NO.
REFERENCE
Res.,220,1%,1/4w,Metal
3082-3220 1316
U53
Res.,4.7k,1%,1/4w,Metal
3082-4470 1317
U54
Res.,4.7k,1%,1/4w,Metal
3082-4470 1318
U64
Res.,2k,1%,1/4w,Metal
3082-4200 1319
U65
Res.,8.2k,1%,1/4w,Metal
3082-4820 1320
U67
Res.,820,1%,1/4w,Metal
3082-3820 1321
U68
Relay_DC12V
4004-0012 1322
U70
Relay_RY12W-K
4004-0014 1323
U72
Relay_LRL-102-100PCV
4004-0011 1324
U74
Relay_REED_GIA122000
4004-0010 1325
U75
Relay_RY12W-K
4004-0014 1326
U76
Relay_RY12W-K
4004-0014 1327
U77
Relay_LRL-102-100PCV
4004-0011 1328
V1
Relay_LRL-102-100PCV
4004-0011 1329
VR1
Relay_LRL-102-100PCV
4004-0011 1330
VR2
Relay_LRL-102-100PCV
4004-0011 1331
VR3
IC_4081
3514-4081 1332
VR5
IC_4001
3514-4001 1333
Y1
IC_4013
3514-4013 1334
L
IC_4018
3516-4018 1335 TRAN1
IC_4018
3516-4018 1336 TRAN2
IC_4528
3516-4528 1337 TRAN3
IC_4528
3516-4528 1338 TRAN4
Res.,10k,Var.,10%,9W
3070-2510
Res.,50k,Var.,10%,9W
3070-2550
IC_4081
3514-4081
IC_4081
3514-4081
IC_4528
3516-4528
Res.,10k,Var.,10%,9W
3070-2510
Res.,20k,Var.,10%,9W
3070-2520
Res.,10k,Var.,10%,9W
3070-2510
Res.,50k,Var.,10%,9W
3070-2550
IC_MIC4575-5.0BT
3505-9575
IC_555
3508-9555
IC_LM311N
3508-1311
IC_PC817_SHARP
3504-9817
IC_PC817_SHARP
3504-9817
IC_4078
3514-4078
IC_4081
3514-4081
IC_4528
3514-4528
IC_TC426C
3508-9426
IC_4012
3514-4012
IC_62003
3516-9003
IC_62003
3516-9003
IC_PC817_SHARP
3504-9817
page 4 of 4
DISCRIPTION
IC_PC817_SHARP
IC_PC817_SHARP
Res.,50k,Var.,10%,9W
Res.,100,Var.,10%,9W
Core,30T:7T
Res.,5k,Var.,10%,9W5
IC_4018
Res.,20k,Var.,10%,9W
IC_MOC3010
IC_PC817_SHARP
IC_PC817_SHARP
IC_MIC4576
Res.,10k,Var.,10%,9W
Res.,10k,Var.,10%,9W
Res.,5k,Var.,10%,9W
Res.,50k,Var.,10%,9W
Crystall,3.84MHz
Core,T130-26
YS-M01
YS-M03
YS-M02
YS-M04
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
51
YESNG
PART NO.
3504-9817
3504-9817
3070-2550
3070-2310
4024-0004
3070-2450
3516-4018
3070-2520
3506-3010
3504-9817
3504-9817
3505-4576
3070-2510
3070-2510
3070-2450
3070-2550
4018-0001
4024-0005
MN-S300-EN-A02
3.5.3
Bill of Materials – Driver Board
CIRCUIT
REFERENCE
3001
C2
3002
C3
3003
C6
3004
C7
3005
C8
3006
C9
3007
C11
3008
D1
3009
D2
3010
D3
3011
D4
3012
D5
3013
D6
3014
D7
3015
D8
3016
D9
3017
D10
3018
D11
3019
D18
3020
D19
3021
D20
3022
D21
3023
D22
3024
D23
3025
D24
3026
D25
3027
D26
3028
D29
3029
D30
3030
D31
3031
D32
3032
F1
3033
F2
3034
F3
3035
F4
3036
L1
3037
Q1
3038
Q4
3039
Q5
3040
Q6
3041
Q8
3042
Q9
3043
R1
3044
R2
3045
R3
BOM - Driver Board (Model:S300)
YESNG
CIRCUIT
DISCRIPTION
PART NO.
REFERENCE
Cap.,0.22μF,100v,Film
3170-7422 3046
R4
Cap.,330pF,250v,Cer.
3140-9133 3047
R5
Cap.,0.1μF,630v,Film
3171-1410 3048
R6
Cap.,470μF,35v,Elect.
3110-4747 3049
R7
Cap.,1000μF,35v,Elect.
3110-4810 3050
R8
Cap.,330pF,250v,Cer.
3140-9133 3051
R9
Cap.,330pF,250v,Cer.
3140-9133 3052
R14
Diode,KZ18
3341-0180 3053
R15
Diode,KZ18
3341-0180 3054
R16
Diode,KZ18
3341-0180 3055
R17
Diode,BY399
3322-0399 3056
R18
Diode,BY399
3322-0399 3057
R19
Diode,BY399
3322-0399 3058
Diode,BY399
3322-0399 3059
Diode,BY399
3322-0399 3060
Diode,BY399
3322-0399
Diode,BY399
3322-0399
Diode,BY399
3322-0399
Diode,KZ18
3341-0180
Diode,KZ18
3341-0180
Diode,FR607×2
3332-0607
Diode,FR607×2
3332-0607
Diode,BY399
3322-0399
Diode,KZ18
3341-0180
Diode,KZ18
3341-0180
Diode,KZ18
3341-0180
Diode,KZ18
3341-0180
Diode,FR607
3332-0607
Diode,FR607
3332-0607
Diode,FR607
3332-0607
Diode,FR607
3332-0607
Core,R_Type,3.5D
4024-0003
Core,R_Type,3.5D
4024-0003
Core,R_Type,3.5D
4024-0003
Core,R_Type,3.5D
4024-0003
Core,R_Type,20D
4024-0006
Q_IRF510
3452-0510
Q_2SK794/2SK2039
3453-2039
Q_2SK794/2SK2039
3453-2039
IC_7818
3503-2018
Q_2SK794/2SK2039
3453-2039
Q_2SK794/2SK2039
3453-2039
Res.,150,5%,2w,Oxide
3020-5315
Res.,16.2,1%,1/4w,Metal
3082-2162
Res.,1k,1%,1/2w,Metal
3083-4100
page 1 of 1
DISCRIPTION
Res.,10,5%,2w,Oxide
Res.,10,5%,2w,Oxide
Res.,1k,1%,1/2w,Metal
Res.,1k,1%,1/2w,Metal
Res.,100,5%,5w,W.W.
Res.,100,5%,5w,W.W.
Res.,100,5%,5w,W.W.
Res.,20k,5%,10w,W.W.
Res.,5.6,1%,1/2w,Metal
Res.,5.6,1%,1/2w,Metal
Res.,5.6,1%,1/2w,Metal
Res.,5.6,1%,1/2w,Metal
Heat_Sink
Heat_Sink
Socket_3P
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
52
YESNG
PART NO.
3020-5210
3020-5210
3083-4100
3083-4100
3030-7310
3030-7310
3030-7310
3030-9520
3083-1560
3083-1560
3083-1560
3083-1560
4001-0030
4001-0030
4015-0030
MN-S300-EN-A02
Yueh Sheng Electronic Industrial Co.,
Ltd.
2F., No. 29, Qiangang St., Shilin District
Taipei City, Taiwan 11170
Tel: +886 2 2882 2212 Fax: +886 2
2881 8635
Website: www.yesng.com.tw
E-mail: info@yesng.com.tw
Obelis European Authorized
Representative
Center (O.E.A.R.C)
Boulevard Général Wahis 53
1030 Brussels, BELGIUM
Tel: +32 2 732 59 54 Fax: +32 2 732
60 03
Website: www.obelis.net
E-mail : mail@obelis.net
2014,Feb.,25
Copyright Yueh Sheng® Electronic Industrial Co., Ltd.
53
Download