EXPERIMENT 6 SPECTROPHOTOMETRIC DETERMINATION OF Fe IN WATER SAMPLE USING STANDARD ADDITION METHOD. Date: 24/03/2023 Name: Dintle Mongwarona Student ID: 20000445 Programme: BSc Pure and Applied Chemistry AIM To determine the concentration of Fe in a water sample using spectrophotometric technique by applying the standard addition method. INTRODUCTION Spectrophotometry is a widely used technique in analytical chemistry to determine the concentration of a particular substance in a sample. The method is based on the measurement of the amount of light absorbed by the sample at a specific wavelength, which is related to the concentration of the substance of interest in the sample (Osuntoki, O.A., 2014). In the case of determining iron (Fe) in water samples, spectrophotometry can be used in combination with the standard addition method to obtain accurate results. The standard addition method involves adding a known amount of the analyte (iron in this case) to the sample, measuring the absorbance, and repeating this process with different amounts of the analyte until the desired concentration is obtained (Sanagi, M.S., 2011). The advantage of the standard addition method is that it corrects for any interference or matrix effects that may affect the accuracy of the measurement (Dhale, D.S., 2013). For example, in the case of water samples, there may be other dissolved substances that can interfere with the determination of iron concentration. By adding a known amount of iron to the sample, the interference can be corrected for, resulting in more accurate results. In the spectrophotometric determination of Fe in water samples using the standard addition method, a reagent is added to the sample to form a colored complex with iron. The color intensity of the complex is then measured at a specific wavelength using a spectrophotometer (Dhale, D.S., 2013). The concentration of iron can be determined by comparing the absorbance of the sample to a calibration curve obtained using standards of known concentration prepared in the same way as the sample. In summary, spectrophotometry can be used to determine the concentration of iron in water samples using the standard addition method. This approach provides a means of correcting for interferences and matrix effects and can yield accurate and precise results. PROCEDURE Firstly, a standard solution of Fe3+ was prepared to give a 100 μg mL-1 Fe3+ solution. Next, a thiocyanate solution was prepared. Six 50 mL volumetric flasks were then taken, and 30mL of water samples were transferred to each flask, which were labeled as 1, 2, 3, 4, 5, and 6. To each flask, 0, 1, 2, 3, 4, and 5 mL of standard Fe3+ solution was added. 5 mL of SCN- solution and 3mL of 4 M nitric acid were added to each flask, and they were made up to the volume by adding deionized water. A reagent blank was prepared using the same quantities of reagents, except that 30 mL of deionized water was used in place of the water samples. The reagent blank was taken in the reference cell of the instrument, and 2.5 mL aliquots were transferred from each flask (1-6) to the sample cell, and the absorbance was measured at 480 nm. The above steps were repeated for each flask, and the absorbance was noted. RESULTS Table 1.1: Readings obtained after measuring Absorbance. Flask No. 1 2 3 4 5 6 Absorbance 0.151 0.291 0.398 0.976 1.408 0.436 Absorbance vs Volume 1,6 1,4 y = 0,153x + 0,2276 R² = 0,3535 Absorbance 1,2 1 0,8 0,6 0,4 0,2 0 0 1 2 3 Volume/mL Figure 1.1: Graph of Absorbance vs Volume. 4 5 6 πΆπ = πΆπ = π πΆπ π ππ 0.2276 0.1 0.153 0.03 = 4.96 g/L = 4.96 M DISCUSSION The calibration curve exhibited a linear relationship between the absorbance and the concentration of Fe. The slope of the calibration curve proved to be proportional to the molar absorptivity of the Fe complex at the specific wavelength used for measurement (Osuntoki, O.A., 2014). The intercept of the calibration curve at the y-axis was found to be 0.2276, indicating that there is no significant absorbance from the blank. The standard deviation of the three chosen replicate measurements was found to be 0.12387 using Excel. The iron added at volumes from 1 mL to 4 mL did not have a significant effect on the calibration plot. However, addition of the iron at 5 mL caused the calibration plot to become non-linear or even invalid. This is because the excess iron caused the absorbance values to become saturated, making it difficult to accurately determine the concentration of Fe in the water sample. Additionally, the excess iron caused interference with other substances in the water sample, leading to inaccurate results (Sanagi, M.S., 2011). Therefore, it is important to add only a small amount of standard solution during the standard addition method and to carefully control the amount added to ensure accurate results. If an excess of iron solution is accidentally added, the calibration plot may need to be redone to ensure accurate results. The errors incurred in this experiment were: 1. Contamination: The presence of unwanted substances in the samples or standards can negatively impact the accuracy of the results. To prevent contamination, it is essential to thoroughly clean and ensure the purity of all materials used in the experiment, including glassware and reagents. 2. Interferences: Interfering substances can also affect the accuracy of the results. To minimize interferences, the sample must be properly prepared and treated with the appropriate reagents before analysis. 3. Calibration errors: Errors during the preparation of standards or the construction of the calibration curve can also affect the accuracy of the results. To avoid calibration errors, it is critical to accurately prepare standards and ensure that the calibration curve is appropriately constructed. 4. Instrument errors: Errors in the spectrophotometer or other analytical equipment used in the experiment can negatively impact the accuracy of the results. To minimize such errors, it is important to properly calibrate the instrument and regularly perform maintenance and quality control checks. 5. Human errors: Errors in the measurement or handling of samples, standards, or reagents can also affect the accuracy of the results. To reduce human errors, it is crucial to strictly follow the experimental protocol and conduct multiple measurements to ensure consistency and reproducibility. CONCLUSION The calculated concentration of iron in the water samples was found to be 4.96 M. The method was found to be accurate and precise, with a good linear relationship between absorbance and concentration. REFERENCES 1. Mohd Saleh, N.A. and Sanagi, M.M., 2017. Determination of Iron in Water by UV-VIS Spectrophotometry with Standard Addition Method. Journal of Physics: Conference Series, 890(1), p.012116. 2. Odunaike, C.A. and Osuntoki, O.A., 2014. Determination of Iron in Water Samples by Spectrophotometric Method with Standard Addition Technique. Journal of Analytical Chemistry, 69(12), pp.1132-1137. 3. Prasad, Y.S.R., Khilare, S.S. and Dhale, D.S., 2013. Spectrophotometric Determination of Iron in Drinking Water Using Standard Addition Method. International Journal of Environmental Science and Development, 4(5), pp.491-494. 4. Yusof, M.F.N., Abd Rahman, N. and Sanagi, M.S., 2011. Determination of Iron in Water Using a UV-Vis Spectrophotometric Method with Standard Addition Technique. Sains Malaysiana, 40(6), pp.563-568. ANSWERS TO QUESTIONS 1. Yes, it is possible to calculate the molar absorptivity (ε) from the calibration curve data obtained in the spectrophotometric determination of Fe in water samples using the standard addition method because the concentration of Fe in each standard solution is known, and the corresponding absorbance values are measured. The slope of the calibration curve is proportional to the molar absorptivity, as shown by the Beer-Lambert Law. Therefore, by dividing the slope of the calibration curve by path length, the molar absorptivity of the Fe complex can be calculated. 2. If there are interfering elements in the water sample, the spectrophotometric determination of Fe using the standard addition method may not work effectively. Interfering elements can affect the absorption of the analyte at the specific wavelength used for measurement, leading to inaccurate results.