Tubular Plug Flow Reactor MUHAMMAD ADLI BIN ABDULLAH (2020489808) MUHAMMAD ALIF BIN AHMAD (2020898448) NAILY QAISAH BINTI NAZLI (2020878116) NOR MAISURAH BINTI HASBULAH (2020834696) Abstract – The objectives of these experiments are to investigate the effects of pulse change and step change in a tubular flow reactor as well as to calculate the residence time distribution (RTD) for both experiments. In addition, we conduct an experiment to determine the calibration curve. In the first experiment, sodium hydroxide solution (NaOH) and ethyl acetate solution (Et (Ac)) are fed into the reactor at a constant flow rate of 700 mL/min of deionized water, whereas in the second experiment, the salt solution is fed at a constant flow rate. Every 30 seconds, the data is collected. The conductivity is measured to determine the calibration curve by mixing the calculated concentration on NaOH solution with deionized water. The mean residence time, variance and skewness of all the data collected during the experiment 1 are 0.912 min, 0.194 min2, 0.14 min3 while for experiment 2 are 2.038 min, 0.323 min2 and -0.493min3. The objectives for these experiments were achieved. I. INTRODUCTION Tubular reactors are steady-state, continuous flow vessels that can be used to mix various chemical substances. It is dependent on the position of the reactants rather than time to complete the chemical reactions. The plugs travel in the reactor’s axial direction, each with a different composition than the ones before and after it. Each plug is treated as a separate entity, effectively acting as an infinitesimally small batch reactor with mixing that approaches zero volume. The residence time of the plug element is derived from its position in the reactor as it flows down the PFR. When scaling flow reactors, one of the factors to consider is residence time distribution (RTD). Tubular reactors are favored by many people due to their numerous benefits. The most practical reactors for use in a variety of applications are tubular reactors. This reactor's main benefits include its simplicity of mechanical design and ease of maintenance and cleaning due to lack of any moving parts. In addition, it has features such as low pressure drops compared to other reactors and displays extremely high conversion rates per volume. Next, it is an ideal apparatus for observing any quick reactions and can be applied to large-scale operations. Residence time distribution (RTD) is a probability distribution function of a chemical reactor where it estimates the amount of time a fluid element might spend inside the reactor. When comparing the behavior of actual reactors to their ideal models, chemical engineers use the RTD to characterize the mixing and flow within the reactors. This is helpful not only for troubleshooting current reactors but also for forecasting reaction yields and designing new reactors. Figure 1: Model of Plug Flow Reactor 1 II. OBJECTIVES The objectives of Experiment 1 are to examine the effect of a pulse input and to construct a residence time distribution (RTD) function for the tubular flow reactor. For Experiment 2, it is conducted to examine the effect of a step change and to construct the residence time distribution (RTD) function for the tubular flow reactor. Lastly, to determine the conductivity measurement of different values between NaOH and Et (Ac). The RTD can be experimentally estimated by inserting a chemically inert substance known as a tracer into the reactor now t = 0 and then measuring the concentration of the tracer in the effluent stream as a function of time (Fogler, 2006). The pulse input and the step input are the two techniques of delivering tracer into the reactor that are used the most frequently. Since the volumetric flow rate is assumed constant, 𝐑𝐞𝐬𝐢𝐝𝐞𝐧𝐜𝐞 𝐓𝐢𝐦𝐞 𝐃𝐢𝐬𝐭𝐫𝐢𝐛𝐮𝐭𝐢𝐨𝐧 𝐅𝐮𝐧𝐜𝐭𝐢𝐨𝐧, 𝐸(𝑡) = 𝐶(𝑡) ∞ ∫0 𝐶(𝑡) 𝑑𝑡 III. THEORY Where, C(t) = Concentration of the tracer (mS/cm) The Time Distribution (RTD) for the saponification reaction inside the Tubular Flow Reactor (BP101-B) shows the mixing of the ethyl acetate, Et (Ac), and the sodium hydroxide, NaOH solution. Reactor internals are constantly consuming the reactants since the flow of reactants is continuous. Step change input in a tubular flow reactor Figure 4- Typical concentration- time curve at the inlet and outlet stream for step change input Figure 2: Tubular Flow Reactor with no radial variations in velocity, concentration, temperature or reaction rate. The residence-time distribution function is shown as a plotted graph of E(t) as a function of time. This graph represents the function. This function provides a quantitative representation of the amount of time that the fluid mixture occupies inside the reactor before it is allowed to exit the reactor. IV. PROCEDURES The feed is started at time t = 0, and a steady rate of tracer is fed to it. As a result, the tracer, Co, has a constant inlet concentration over time. This experiment allows for direct determination of the cumulative distribution, F (t). The cumulative distribution, F(t) represents the fraction of effluent that has been in the reactor for time t = 0 until t = t. Cumulative Distribution, F(t) =[ Figure 3– Illustration of tracer injection at t = 0 and detection for the tracer concentration at the effluent stream to determine the Residence Time Distribution for the system 𝐶𝑜𝑢𝑡 ] 𝐶0 𝑠𝑡𝑒𝑝 2 Differentiation of the cumulative distribution function yield to RTD function, 𝐑𝐞𝐬𝐢𝐝𝐞𝐧𝐜𝐞 𝐓𝐢𝐦𝐞 𝐃𝐢𝐬𝐭𝐫𝐢𝐛𝐮𝐭𝐢𝐨𝐧 𝐅𝐮𝐧𝐜𝐭𝐢𝐨𝐧, 𝐄(𝐭) 𝐸(𝑡) = 𝑑 𝐶(𝑡) [ ] 𝑑𝑡 𝐶0 𝑠𝑡𝑒𝑝 IV.MATERIAL AND APPARATUS 1. Tubular Flow Reactor (Model: BP 101) 2. Deionized water 3. Sodium Hydroxide (NaOH) 4. Ethyl Acetate (Et (Ac)) 5. Stopwatch The mean residence time, tm shows the average time the fluids stay inside the reactor (Rochelle Fourie, 2016). 𝐅𝐢𝐫𝐬𝐭 𝐌𝐨𝐦𝐞𝐧𝐭, 𝐌𝐞𝐚𝐧 𝐑𝐞𝐬𝐢𝐝𝐞𝐧𝐜𝐞 𝐓𝐢𝐦𝐞, ∞ 𝑡𝑚 = ∫0 𝑡𝐸(𝑡) 𝑑𝑡 The spread of the distribution is the magnitude of the variance, 𝜎2 . When the magnitude increases, so does the dispersion of the distribution (Fogler, 2006). 𝐒𝐞𝐜𝐨𝐧𝐝 𝐌𝐨𝐦𝐞𝐧𝐭, 𝐕𝐚𝐫𝐢𝐚𝐧𝐜𝐞, ∞ 𝜎2 = ∫ (𝑡 − 𝑡𝑚 )2 𝐸(𝑡) 𝑑𝑡 0 The magnitude of the skewness indicates the degree to which a distribution deviates from the normal distribution, providing a measure of the degree to which the distribution is skewed in one direction (Rouse, 2012). 𝐓𝐡𝐢𝐫𝐝 𝐌𝐨𝐦𝐞𝐧𝐭, 𝐒𝐤𝐞𝐰𝐧𝐞𝐬𝐬, 3 𝑠 = ∞ 1 (𝑡 − 𝑡𝑚 )3 𝐸(𝑡) 𝑑𝑡 3∫ 𝜎2 0 Numerical Evaluation of Integrals In order to ∞ determine the integral of ∫0 𝐶(𝑡) 𝑑𝑡 . For N + 1 points, where N is even, (Fogler, 2006) 𝑋𝑁 ∫ 𝑋0 ℎ 𝑓(𝑋)𝑑𝑋 = (𝑓0 + 4𝑓1 + 2𝑓2 + 4𝑓3 3 + 2𝑓4 +. . . + 4𝑓𝑁−1 + 𝑓𝑁 Where, N = Number of segments ℎ= 𝑋𝑁 − 𝑋0 𝑁 Figure 4-Tubular Flow Reactor (Model BP 101) (Front View) V. PROCEDURE General startup procedure: Make sure that all valves, except for valves V7, were in the closed position. Both the 0.1M NaOH solution and the 0.1M Et (Ac) solution were produced in tank B1 and tank B2 correspondingly. The power supply for the control panel has been activated. Both the pre-heater B5 and the water jacket B4 were loaded with purified water. The pump designated as P1 was activated. The flow rate of the pump was modified to be 700 mL/min as measured by the flow meter F1-01. While the valves V2 and V1 were opened, the stirrer motor, M1, was activated and its speed was increased to 200 revolutions per minute (rpm). 3 During the time that valves V13 and V8 were being opened, pump P3 was activated so that water could be pumped through pre-heater B5. Both the valve V10 and the pump P1 were turned off at the same time. After that, the valves V6 and V12 were allowed to fully open. After that, the pump known as P2 was activated. At flowmeter FI-02, the adjustment for pump P2 was made to be 700 mL/min. In the end, the valve V12 was shut off, and the pump P2 was turned off. Experiment 1: To begin, go through the standard beginningof-operations steps. After that, the valve V9 should be opened, and pump P1 should be turned on. The flow controller for pump P1 should then be adjusted so that a constant flow rate of deionized water enters reactor R1 at a rate of approximately 700 ml/min at the FI-01 location. The next step is to allow the flow of deionized water through the reactor to continue until both the inlet (QI-01) and outlet (QI-02) conductivity values are stable at low levels. Make a note of both conductivity readings. Turn off the pump, and then close the valve V9. Turn the pump P2 on while also opening the valve V11. Turn on both clocks at the same time. Adjust the flow controller on the pump P2 so that there is a consistent flow rate of salt solution into the reactor R1 at 700 milliliters per minute at FI02. Experiment 2: Carry out the general processes for starting things up. Turn the pump P1 on while also opening the valve V9. Adjust the flow controller on the pump P1 so that there is a steady flow of deionized water into the reactor R1 at approximately 700 ml/min at the FI01 location. Water that has been deionized should be allowed to continue flowing through the reactor until the conductivity values at the inlet (QI-01) and output (QI-02) remain stable at low levels. Take a record of both conductivity readings. Pull the plug on the valve V9 and turn the pump P1 off. Turn the pump P2 on while also opening the valve V11. Turn on both clocks at the same time. Conductivity readings should be taken at regular intervals of 30 seconds, first at the intake (QI-01) and then at the outflow (QI-02). Keep recording the conductivity values until all of the readings become almost identical. General shut down: All three pumps, P1, P2, and P3, have been turned off. After that, the valves V2 and V6 are shut off. After that, the switch for the heater is turned off. While the stirrer motor is operating, the cooling water that is moving through the reactor is maintained to permit the water jacket to gradually come down to ambient temperature. After that, the power supply to the control panel is disconnected completely. After allowing the salt solution to run freely for one minute, you should then reset the timer and begin counting again. The time will begin counting down from this point, which is the average pulse input. Turn off the pump, and then close the valve V11. After that, immediately turn on pump P1 and open valve V9 as rapidly as possible. Adjusting the P1 flow controller should be done to keep the flow rate of deionized water at a constant level of 700 milliliters per minute. Conductivity readings at the inlet (QI-01) and outlet (QI-02) should be recorded starting immediately. The readings should be taken at regular intervals of 30 seconds. Keep recording the conductivity values until all of the readings have become almost identical and are getting closer to the stable low-level values. 4 VI. RESULTS Experiment 1: 3.0 0.0 0.0 0.0 0.0 0.0 0.0 3.5 0.0 0.0 0.0 0.0 0.0 0.0 Table 2: Data for Pulse Input in Tubular Reactor Conver sion (%) Solution Mixtures (mL) NaOH (0.1M) 0 100 Na (Ac) (0.1M) 0 Concent ration of NaOH (M) Conduc tivity (𝜇 s) Mean Residence Time, tm H2O - 100 0.0500 7.83 25 100 0.0375 6.87 0.912 min Variance, 𝜎 2 0.194 min2 Skewness, s3 0.14 min3 Table 3: Data for tm, 𝜎 2 and s3 for Pulse Input Flow Rate: 700 mL/min 25 75 Input Type: Step Input 50 50 50 100 0.0250 3.96 75 25 75 100 0.0125 2.13 100 - 100 100 0.0000 1.06 Table 1: Table for Preparation of Calibration Curve Time (min) 0.0 Conductivi ty (𝜇 s/cm) E(t) min 1 tE (t) (ttm)2E(t) min2 (ttm)3E(t) min3 Inlet Outl et 0.0 0.0 0.0 0.0 0.0 0.0 0 0.5 5.0 0.1 0.018 0.009 0.043 -0.066 Flow Rate: 700 mL/min 1.0 5.1 1.3 0.236 0.236 0.254 -0.264 Input Type: Pulse Input 1.5 5.5 5.0 0.909 1.364 0.263 -0.142 2.0 5.5 5.3 0.964 1.928 0.001 -0.053 x 10-3 2.5 5.5 5.3 0.964 2.41 0.206 0.095 3.0 5.5 5.3 0.964 2.892 0.892 0.858 3.5 5.5 5.3 0.964 3.374 2.061 3.013 Time (min) Conductivity (𝜇 s/cm) E(t), min 1 tE (t) (ttm)2E( t) min2 (ttm)3E (t) min3 Inlet Outlet 0.0 0.1 0.0 0.0 0.0 0.0 0.0 0.5 0.0 4.8 0.717 0.359 0.122 -0.05 1.0 0.0 4.8 0.717 0.717 0.006 0.001 Mean Residence Time, tm 2.038 min 1.5 0.1 2.7 0.403 0.605 0.139 0.082 Variance, 𝜎 2 0.323 min2 2.0 0.0 0.3 0.045 0.09 0.053 0.058 Skewness, s3 -0.493 min3 2.5 0.0 0.0 0.0 0.0 0.0 0.0 Table 4: Data for Step Input in Tubular Reactor Table 5: Data for tm, 𝜎 2 and s3 for Step Input 5 2.5 0.5 Sample Calculations: ∫0 Experiment 1 Pulse Input: tm = 0.912 min ∞ Calculation for ∫0 1. C(t) dt 4. Calculation for (t-tm)2E(t) min2 ℎ 𝑋 ∫𝑋 𝑁 = 0.122 𝑓(𝑥) 𝑑𝑥 = (𝑓0 + 4𝑓1 + 2𝑓2 + 4𝑓3 + 3 0 2𝑓4 +. . .4𝑓𝑁−1 + 𝑓𝑁 ) 5. 𝑋𝑁 −𝑋0 ℎ= 2.5𝑚𝑖𝑛 − 0𝑚𝑖𝑛 = 0.5 5 2.5 0.5 (0 + 4(4.8) + 2(4.8) + 4(2.7) + 2(0.3)) 3 𝐶(𝑡) 𝑑𝑡 = ∫ Calculation for (t-tm)3E(t) min3 (t-tm)3E(t) min3 = [(0.5-0.912)3] (0.717) 𝑁 0 2.5 3 (t-tm)2E(t) min2 = [(0.5-0.912)2] (0.717) For N+1 ℎ= 𝑡𝐸(𝑡) 𝑑𝑡 = ( ) (5.47) = 0.912 min = -0.05 6. Space Time, τ Space Time, τ = 𝑉𝑜𝑙𝑢𝑚𝑒 𝑜𝑓 𝑟𝑒𝑎𝑐𝑡𝑜𝑟 𝑉𝑜𝑙𝑢𝑚𝑒𝑡𝑟𝑖𝑐 𝑓𝑙𝑜𝑤 𝑟𝑎𝑡𝑒𝑠 0.5 𝐶(𝑡) 𝑑𝑡 = ( 3 ) (40.2) = 6.7 𝜇 s min / cm ∫0 2. E(t) = Calculation for E(t) Space Time, τ = Space Time, τ = 5.714 min 𝐶(𝑡) 2.5 7. 𝐶(𝑡) 𝑑𝑡 ∫0 Calculation for Variance, 𝜎 2 When time at 0.5 min E(t) = 𝜎2 = ∫ 0 4.8 𝜇𝑠 / 𝑐𝑚 6.7 𝜇𝑠 𝑚𝑖𝑛 / 𝑐𝑚 3. Calculation for Moments in RTD Function ∞ Means Residence Time, tm = ∫0 2.5 ∫ 0 ℎ= (𝑡 − 𝑡𝑚 )2 𝐸(𝑡) 𝑑𝑡 𝜎2 = 0.5 (0 + 4(0.122) + 2(0.006) + 4(0.139) 3 + 2(0.053)) 𝜎2 = ( 0.5 )(1.162) = 0.194 𝑚𝑖𝑛2 3 𝑋𝑁 − 𝑋0 𝑁 2.5𝑚𝑖𝑛 − 0𝑚𝑖𝑛 = 0.5 5 2.5 0 𝑡𝐸(𝑡) 𝑑𝑡 ℎ 𝑡𝐸(𝑡) 𝑑𝑡 = (𝑓0 + 4𝑓1 + 2𝑓2 + 4𝑓3 + 2𝑓4 +. . .4𝑓𝑁−1 3 + 𝑓𝑁 ) ℎ= ∞ ℎ = (𝑓0 + 4𝑓1 + 2𝑓2 + 4𝑓3 + 2𝑓4 +. . .4𝑓𝑁−1 + 𝑓𝑁 ) 3 E(t) = 0.717 min 1 ∫ 4𝐿 (700 𝑚𝐿/𝑚𝑖𝑛)(1𝐿/1000𝑚𝐿) 𝑡𝐸(𝑡) 𝑑𝑡 = 0.5 (0 + 4(0.359) + 2(0.717) + 4(0.605) 3 + 2(0.09)) 6 Calculation for Skewness, s3 8. ∞ ∫ 0 (𝑡 − 𝑡𝑚 )3 𝐸(𝑡) 𝑑𝑡 ℎ = (𝑓0 + 4𝑓1 + 2𝑓2 + 4𝑓3 + 2𝑓4 +. . .4𝑓𝑁−1 + 𝑓𝑁 ) = 1. 3 0.1 E(t) = 5.5 (0 + 4(−0.05) + 2(0.001) + 4(0.082) + 2(0.058)) 0.5 = ( )(0.246) = 0.041 min3 3 = √𝜎 3 𝜎2 = 1 𝜎2 s3 = 1 ∫ s3 = ( ∞ 2.5 ∫ 1 ℎ= (𝑡 − 𝑡𝑚 )3 𝐸(𝑡) 𝑑𝑡 𝑡𝐸(𝑡) 𝑑𝑡 = 0 Sample Calculations: Experiment 2 Step Input: Calculation for ∞ ∫0 C(t) dt For N+1 𝑓(𝑥) 𝑑𝑥 = (𝑓0 + 4𝑓1 + 2𝑓2 + 4𝑓3 + 3 0 2.5 ∫0 0.5 𝑡𝐸(𝑡) 𝑑𝑡 = ( ) (12.23) = 2.038 min 3 tm = 2.038 min 4. 𝑁 2.5 ∫ 0 + 2.5 ∫0 = 0.043 𝑋𝑁 −𝑋0 ℎ= Calculation for (t-tm)2E(t) min2 (t-tm)2E(t) min2 = [(0.5-2.038)2] (0.018) 2𝑓4 +. . .4𝑓𝑁−1 + 𝑓𝑁 ) ℎ= 0.5 (0 + 4(0.009) + 2(0.236) + 4(1.364) 3 + 2(1.928) + 2.41 3. ℎ 𝑋 ∫𝑋 𝑁 𝑋𝑁 − 𝑋0 𝑁 2.5𝑚𝑖𝑛 − 0𝑚𝑖𝑛 = 0.5 5 2.5 )(0.041) = 0.14 𝑚𝑖𝑛3 𝑡𝐸(𝑡) 𝑑𝑡 ℎ 𝑡𝐸(𝑡) 𝑑𝑡 = (𝑓0 + 4𝑓1 + 2𝑓2 + 4𝑓3 + 2𝑓4 +. . .4𝑓𝑁−1 3 + 𝑓𝑁 ) 0.194 = = 0.292 0.664 ∫ 0.292 Calculation for Moments in RTD Function ℎ= 3 0 𝜎2 ) ∞ = √ 0.441 = 0.664 𝜎2 𝐶0 Means Residence Time, tm = ∫0 0 1 𝜎2 ( = 0.018 𝑚𝑖𝑛−1 2. = √ 𝜎 2 = √0.194 = 0.441 𝜎 𝐶(𝑡) 𝑑𝑡 𝐶0 = 5.5 min 3 0.5 𝑑 E(t) = 2.5𝑚𝑖𝑛 − 0𝑚𝑖𝑛 = 0.5 5 Calculation for (t-tm)3E(t) min3 (t-tm)3E(t) min3 = [(0.5-2.038)3] (0.018) = -0.066 0.5 𝐶(𝑡) 𝑑𝑡 = (0 + 4(0.1) + 2(1.3) + 4(5) + 2(5.3) 3 5.3) 0.5 𝐶(𝑡) 𝑑𝑡 = ( 3 ) (38.9) = 6.48 𝜇 s min / cm 7 5. Calculation for Variance, 𝜎 2 𝜎2 = ∫ ∞ (𝑡 − 𝑡𝑚 )2 𝐸(𝑡) 𝑑𝑡 0 ℎ = (𝑓0 + 4𝑓1 + 2𝑓2 + 4𝑓3 + 2𝑓4 +. . .4𝑓𝑁−1 + 𝑓𝑁 ) 3 𝜎2 = 0.5 (0 + 4(0.043) + 2(0.254) + 4(0.263) 3 + 2(0.001) + 0.206) 0.5 𝜎 2 = ( )(1.94) = 0.323 𝑚𝑖𝑛2 3 6. ∞ Graph 1: Conductivity VS Conversion Calculation for Skewness, s3 (𝑡 − 𝑡𝑚 )3 𝐸(𝑡) 𝑑𝑡 ∫ 0 ℎ = (𝑓0 + 4𝑓1 + 2𝑓2 + 4𝑓3 3 + 2𝑓4 +. . .4𝑓𝑁−1 + 𝑓𝑁 ) 0.5 = (0 + 4(−0.066) + 2(−0.264) 3 + 4(−0.142) + 2(−0.053 𝑥 10 −3 ) + 0.095 Graph 2: Conductivity Outlet VS Time 0.5 = ( )(-1.265) = -0.211 min3 3 𝜎 = √ 𝜎 2 = √0.323 = 0.568 1 𝜎2 = √ 𝜎 3 𝜎2 = s3 = 1 ∫ ∞ 3 0 𝜎2 s3 = ( 1 0.428 = √ 0.568 = 0.754 𝜎2 1 𝜎2 = 0.323 = 0.428 0.754 Graph 3: RTD, E(t) VS Time (𝑡 − 𝑡𝑚 )3 𝐸(𝑡) 𝑑𝑡 )(−0.211) = −0.493 𝑚𝑖𝑛3 Graph 4: Conductivity Outlet VS Time 8 Graph 5: RTD, E(t) VS Time The residence time distribution is displayed as departure time E(t) versus time-based on the data that was collected for the second objective and reported in the table. How long the atoms were within the reactor is determined by their residence time (Fogler, 2006). Therefore, mean residence time (tm) measures how long fluids typically stay inside the reactor. For the pulse input experiment, the computed tm = 0.912 min refers to the material atoms that spent the same amount of time in the reactor. Because of this, the reactants are allowed to react, and the product is released into the effluent stream at a rate that is almost identical to the measured average time interval (tm = 2.038 min) in the step input experiment. VII. DISCUSSIONS In a tubular flow reactor, two lab experiments with pulse input and step input have been performed. The goals are to investigate how a pulse input and step change affect a tubular reactor, and then to construct a residence time distribution (RTD) function for the tubular flow reactor following the experiment. Deionized water entered the reactor at a constant flow rate of 700 ml/min. The conductivity of the solution at its input and outflow was monitored throughout the experiment until it reached a consistent value. The C(t) curve for the Pulse Input experiment displays a bell-shaped pattern with three peaks. These three peaks demonstrate that the reactor's internal flow is not optimal. Graph 2 illustrates how the C(t) curve must increase and decrease with only one peak for a flow to be optimum. As a result, it is assumed that there is a disturbance inside the reactor, which is most likely a dead volume. The Step Change Experiment's C(t) curve shows a linear rise in slope with time. However, the flow becomes steady from roughly t = 0 min to t = 0.5 min before returning to the increasing linear trend. Since a perfect flow for step Input is one in which the C(t) curve advances linearly with no variation as seen in Graph 4, this also demonstrates the existence of disturbance. The variations in the mean residence periods were due to differences in the input concentration of the salt solution between the two trials. Before the pulse input experiment permits the salt solution to flow into the reactor for a maximum of one minute, the deionized water flow is intended to be continuous at a rate of 700 ml/min. The residence durations require a smaller amount to react since the reactant has been saturated before being reacted to its maximum point. This is demonstrated by the bell-shaped curve when the conductivity considerably rises from t=0 to t=1 but falls when the volume of salt solution inside the reactor diminishes. The salt solution is allowed to flow continuously in the step-input experiment until all the conductivity readings are nearly constant. As a result, the reaction will intensify until it reaches its maximum rate, which indicates that the conductivity becomes constant at time t=5 and beyond. Additionally, the RTD function calculates the second moment as variance (2). The magnitude of this moment, known as variance, serves as a measure of the distribution's "spread." The bigger the value of this moment, the wider the distribution will be (Fogler, 2006). The calculated variance for the pulse input experiment is 0.194 min2, which is lower than the calculated variance for the step input experiment, which is 0.323 min2. According to the magnitude computed, the pulse input experiment's salt solution dispersion inside the reactor is more than it was for the step input experiment. 9 This happened because the salt solution from experiment 1 was already within the tube, which caused the reaction to happen quickly but not just with the salt solution that was already there. Because the spread of the distribution is also constrained by the volume of salt solution present inside the reactor, the step input's variance is smaller than the pulse input. The skewness, represented as s3, is the third moment to be considered in both experiments. In this instance, the distribution's deviation from the normal distribution is visible. The measured skewness in the step input experiment is s3 = -0.493 min3, while the measured skewness in the pulse input experiment is s3 = 0.14 min3. Both exhibit positive skewness, which denotes a rightward skewing of the Residence Time Distribution (RTD) function. In other words, the fluids inside the reactor burn out earlier than they would under perfect circumstances. In comparison to the Step Input skewness, which is -0.493 min3 and perhaps too distant from the normal distribution, the Pulse Input skewness is closer to zero and, as a result, closer to the normal distribution. Finally, a calibration curve for conductivity vs conversion is also produced for the saponification reaction between ethyl acetate Et (Ac) and sodium hydroxide NaOH. Excel is used to produce the slope and y-axis intercept value for the conductivity vs. conversion calibration curve using the curve yield equation. The slope is -0.0731 according to the presented graph, and the y-axis intercept is 8.026. VIII. CONCLUSION We were able to analyze the effect of pulse input and step change in a tubular flow reactor because of the experiment described above, and we were also able to differentiate between the two effects. There is also the possibility of constructing the residence time distribution (RTD) function for the tubular flow reactor. Calculations and graphs were made based on the results of the samples that were taken, which were then plotted. In the first trial, the flow rate was maintained at 700 ml/min throughout, and the participants used deionized water. The total of C(t) was calculated to have a value of 6.7 𝜇 s min / cm, while the total of E(t) was calculated to have a value of 0.717 min 1. When we look at graph 2, we can see that at first, it was going up, but after one minute, it started going down, and it kept going down until it reached the 2.5-minute mark. This indicates that a flow of conductivity is occurring because of the recording of a unit pulse response at the outlet stream. The variance, 𝜎 2 , and skewness, s3, for the pulse input were 0.194 𝑚𝑖𝑛2 and 0.14 𝑚𝑖𝑛3 , respectively. The mean residence time, tm, for the pulse input was 0.912 minute. In the second experiment, we looked at how a step change in input affected the outcome. The flow rate was likewise maintained at a constant level of 700 ml/min and was the same as that of the deionized water that was used. According to the findings and the calculations, the total amount of conductivity was 6.48 𝜇 s min / cm and the total amount of E(t) was 0.018 𝑚𝑖𝑛−1 . Figure 2 demonstrates that there is a rise in the data after two minutes, followed by a period of stability for one minute. The variance, 𝜎 2 , and skewness, s3, for the pulse input were0.323 𝑚𝑖𝑛2 and −0.493 𝑚𝑖𝑛3 , respectively. The mean residence time, tm, for the pulse input was 2.038 minutes. IX. RECOMMENDATION Before beginning the experiment, it is imperative that you better understand yourself with the standard operating procedure and follow it to the extent. Next, we must ensure that every valve is fully open before beginning the experiment and fully closed after it has been completed. The residence time distribution (RTD) function could also be determined using trace techniques like the negative step and frequency - response methods in addition to the Pulse Input and Step Input methods. Assumptions made outside of this context, however, should be supported by a literature review and appropriate books on chemical reaction engineering. Finally, once the flow rate has stabilized after waiting a few minutes, record the data for the outlet conductivity. 10 References FilsonFilter. (n.d.). Tubular Reactor: The Ultimate FAQ Guide - Filson Filter. Filson Filters. Retrieved November 13, 2022, from https://www.filsonfilters.com/tubul ar-reactor Plug Flow Reactor. (n.d.). Vapourtec. Retrieved November 13, 2022, from https://www.vapourtec.com/flowchemistry/plug-flow-reactor/ Residence_Time_Distribution. (n.d.). chemeurope.com. Retrieved November 13, 2022, from https://www.chemeurope.com/en/e ncyclopedia/Residence_Time_Distr ibution.html 11