Uploaded by Iqbal Walid

td830 eshmann esu

advertisement
TD830
E LECT RO SURGICAL UNIT
Instructions for use
698257
Downloaded from www.Manualslib.com manuals search engine
E-IM54f
Read these Instructions before use
Preliminary
Information
Technical data
Safety notes
and alarms
Operation
Cleaning, disinfection
and care
Keep these ‘Instructions for use’ in a safe convenient place for future reference. Read in
conjunction with the relevant Publications detailed in the preliminary information section.
Eschmann After Sales Service Department
The Eschmann After Sales Service Department is staffed and equipped to provide advice and
assistance during normal office hours. To avoid delays when making enquiries, please quote the
Model and Serial Number of your Electrosurgical Unit which is shown on the Serial Number plate,
the location of which is shown below. Please ensure you include all alpha and numeric digits of
the Serial Number.
Instructions for use
Accessories
The Serial Number plate
is located here, (view
from rear of unit).
For further information visit www.eschmann.co.uk
All correspondence relating to the after sales service of Eschmann Equipment to be addressed to :
UK Customers
Eschmann Equipment, Peter Road, Lancing, West Sussex BN15 8TJ, England.
Tel: +44 (0) 1903 765040. Fax: +44 (0) 1903 875711.
Overseas Customers
Contact your local distributor. In case of doubt contact Eschmann Equipment.
Patents and Trade marks
The ESCHMANN name and logo are registered trade marks of Eschmann Holdings Limited.
“Eschmann Equipment” is a trading name of Eschmann Holdings Limited.
“TD830” is a trade mark of Eschmann Holdings Limited.
Patents :
GB2276551, GB2146534, AU673883, ZA94/2173, US5480399,
EP617925, IEE69100 and other Patents Pending.
Copyright © 2006 Eschmann Holdings Limited
All rights reserved. This booklet is protected by copyright. No part of it may be reproduced, stored in a
retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying,
recording or otherwise without written permission from Eschmann Holdings Limited.
The information in this publication was correct at the time of going to print. The Company, however,
reserves the right to modify or improve the equipment referred to.
The CE marking affixed to the product certifies that it complies with the
European Medical Devices Directive 93/42/EEC and related legislation.
E-IM54f July 2006
Downloaded from www.Manualslib.com manuals search engine
TD830
ELECTROSURGICAL UNIT
CONTENTS
1. PRELIMINARY INFORMATION
5. OPERATION
Preliminary information . . . . . . . . . . . . . . . . . . . . . . . . 4 Routine check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Connection of accessories . . . . . . . . . . . . . . . . . . . .
Monopolar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2. TECHNICAL DATA
Bipolar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 System check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . .
Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Monopolar standby mode . . . . . . . . . . . . . . . . . . .
Audible indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Monopolar mode . . . . . . . . . . . . . . . . . . . . . . . . . .
Visual indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
USER 1 and/or USER 2 selection . . . . . . . . . .
Safety.. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Plate electrode . . . . . . . . . . . . . . . . . . . . . . . .
Duty cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Fingerswitch/active electrode handle . . . . . . . .
Other symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Footswitches . . . . . . . . . . . . . . . . . . . . . . . . . .
Button symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Monopolar outputs . . . . . . . . . . . . . . . . . . . . . .
Alarm symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Power output . . . . . . . . . . . . . . . . . . . . . . . . . .
Environmental conditions for transport and storage . . . 8
Bipolar Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Selecting bipolar forceps . . . . . . . . . . . . . . . . .
3. SAFETY NOTES & ALARMS
Connecting forceps . . . . . . . . . . . . . . . . . . . . .
Changing forceps . . . . . . . . . . . . . . . . . . . . . .
Do’s and Don’ts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Footswitches . . . . . . . . . . . . . . . . . . . . . . . . . .
Alarm circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Power output . . . . . . . . . . . . . . . . . . . . . . . . . .
After use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4. INTRODUCTION
Disconnecting accessories . . . . . . . . . . . . . . . . . .
Power output selection guide . . . . . . . . . . . . . . . .
General .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
..................................
Operating modes / display option.. . . . . . . . . . . . . . . 11 Alarms
Alarm conditions . . . . . . . . . . . . . . . . . . . . . . . . . .
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Alarm identifications and remedies . . . . . . . . . . . .
Associated Publications . . . . . . . . . . . . . . . . . . . . . . . 11
TABLE
1 - Power output selection guide . . . . . . . . . .
Equipment certification . . . . . . . . . . . . . . . . . . . . . . . 11
Electrosurgical techniques . . . . . . . . . . . . . . . . . . . . . 12 Output diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . .
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Monopolar electrosurgery . . . . . . . . . . . . . . . . . . . 12 6. CLEANING, DISINFECTION & CARE
Plate electrode . . . . . . . . . . . . . . . . . . . . . . . . 12
Cleaning and disinfection . . . . . . . . . . . . . . . . . . . . . .
Footswitches . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Cleaning the electrosurgical unit . . . . . . . . . . . . . .
Active electrode handles . . . . . . . . . . . . . . . . . 12
Disinfecting the electrosurgical units . . . . . . . . . . .
Active cables . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Electrosurgical accessories . . . . . . . . . . . . . . . . .
Active electrodes . . . . . . . . . . . . . . . . . . . . . . . 13
Care
..................................
Fingerswitches . . . . . . . . . . . . . . . . . . . . . . . . 13
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power control . . . . . . . . . . . . . . . . . . . . . . . . . 13
Additional information . . . . . . . . . . . . . . . . . . . . . .
Current paths . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Maintenance procedures . . . . . . . . . . . . . . . . . . .
Accidental contact . . . . . . . . . . . . . . . . . . . . . . 13
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Bipolar electrosurgery . . . . . . . . . . . . . . . . . . . . . . 13
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
ACCESSORIES
Flammable anaesthetics and spirits . . . . . . . . . . . 14 7.
Implanted cardiac pacemakers . . . . . . . . . . . . . . . 14
Accessory list . . . . . . . . . . . . . . . . . . . . . . . . . . . .
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Risks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 ILLUSTRATIONS
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1-5 .................................
Electrosurgical interference . . . . . . . . . . . . . . . . . 14
6-8 .................................
Using two electrosurgical units . . . . . . . . . . . . . . . 15
9 - 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using a flexible endoscope . . . . . . . . . . . . . . . . . . 15
12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
16
16
16
17
17
18
18
18
18
19
19
19
20
20
21
21
21
21
21
21
21
21
21
22
22
22
24
25
32
32
32
32
32
32
32
32
32
33
34
35
36
37
P3/37
1.7 Ensure that the unit ‘mains’ switch (42 of Fig. 11) is
in the ‘O’ position and that the output controls are set to
minimum before connecting to, and switching ‘on’, the
1.1 These Instructions for Use should be referred to for
mains supply. A complete systems check must be carried
details of the TD830 Electrosurgical Unit, REF 83-256-01,
out before using the Electrosurgery Unit (see Sections 5.10
83-257-02, 83-258-03, 83-259-04, 83-260-05, 83-261-01,
to 5.17).
83-262-07, 83-263-08, and 83-264-09 (serial number
A9B0000 or above).
1.8 Instruction and Service Manuals should be readily
accessible for reference prior to and when operating,
1.2 Within the text of this manual the term ‘coag’ is used
cleaning and servicing the TD830 Electrosurgical Unit. All
as a common abbreviation of the term ‘coagulation’.
manuals are available from Eschmann Equipment, see
1.3 Carefully remove the TD830 Electrosurgical Unit and inside front cover for address details.
its associated accessories from the packing case(s).
1. PRELIMINARY INFORMATION
1.4 Place the TD830 Electrosurgical Unit on a stable Related Technical Publications
convenient working platform, for example on the Eschmann
ST80 mobile trolley, purpose built for carrying Service Manual for TD830
electrosurgical units.
Publication number E-SM44, Part No. 698260
1.5 The TD830 Electrosurgical Unit requires a mains
electrical supply corresponding to the voltage shown on Eschmann accessory ‘Instructions for use’
the electrical rating plate at the rear of the unit. Only use
Publication number E-IM50, Part No.604802
the mains supply cable supplied. If the plug supplied prefitted is not suitable it should be replaced with a suitable
plug with protective earthing contact.
1.6
If the plug is a fused type, a 10A fuse must be fitted.
CAUTION
It is most important that fuses of the correct
type, size and rating are installed (see Technical
Data).
CAUTION
Read these ‘Instructions for use’ carefully
before using this electrosurgical unit and note
ALL of the warnings, cautions and safety notes
contained within. Keep these ‘Instructions for
use’ close-to-hand at all times for reference.
Please note that Fig. 11 at the end of this manual
opens out so that it can be refered to easily whilst
reading this manual. It is the main figure
identifying the various parts of the TD830
Electosurgical Unit.
P4/37
Downloaded from www.Manualslib.com manuals search engine
E-IM54f
TD830
ELECTROSURGICAL UNIT
2. 0 TECHNICAL DATA
Output (bipolar)
GENERAL
The TD830 Electrosurgical Unit (or Surgical Diathermy
Symbol/Function
Unit) is classified as ‘HF surgical equipment’ 1
1
- ‘HF surgical equipment’ is defined as, “Medical
electrical equipment including its associated accessories
intended for the performance of surgical operations, such as
the ‘cutting’ 2 or ‘coagulation’ 3 of biological tissue by means of
high frequency (h.f.) currents”.
Power
Crest
factor
Peak
Open
Circuit
Voltage
Coagulation
Micro range
17W ±20% Variable
Macro range
50W
+10%
Variable
150
230
-20%
- ‘Cutting’ is defined as, “Resection or dissection of body
tissue caused by the passage of high frequency current of high
Power and voltage output data diagrams are shown at the
current density at the active electrode(s)”.
2
end of section 5.
3
- ‘Coagulation’ is defined as, “Sealing of small blood
vessels or of body tissue caused by the passage of high
frequency current at the active electrode(s)”.
Monopolar
Carrier frequency - 475kHz nominal, square wave.
Equipment - High power electrosurgical unit with
monopolar and bipolar outputs
Power control Variable amplitude set by frontp a n e l
controls. Preset pulse patterns set by mode switches.
Type - Portable
Load resistance for maximum output power is 150 ohms
DIMENSIONS
(non-inductive) for cut and pinpoint coag and 200 ohms
(non-inductive) for blend, specialist cut and spray coag.
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39.0 cm
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19.0 cm Test load is 200 ohms (non-inductive) for cut and specialist
Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.5 cm cut and 400 ohms (non-inductive) for blend, spray coag
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 kg and pinpoint coag.
ELECTRICAL DATA
Output (monopolar)
(Note: Voltage factory set by transformer tapping, according Output powers are measured to an accuracy of ±20%,
with a maximum power of 400 watts.
to model supplied.)
Power Supply . . . . . . . . . . . . . . . . . 230V a.c., 50-60Hz
or, 240V a.c., 50-60Hz
Peak
or, 220V a.c., 50-60Hz
Open
or, 110V a.c., 50-60Hz
Crest
Circuit
Current (max.) . . . . . . . . . 4.4A (230V) or, 4.2A (240V)
Symbol/Function
Power
factor
Voltage
or, 4.6A (220V)
or, 8.4A (110V)
Fuse rating (240V, 230V, 220V) . . . . . . . . . . . 250V, T5A
Normal cut
345W(-20%)
1.9
1150
Fuse rating (110V) . . . . . . . . . . . . . . . 125V(min.),T10A
Fuse type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 mm
Blend
300W(±20%)
3.0
2200
Bipolar
Carrier frequency - 785kHz nominal, square wave.
Specialist cut
345W(-20%)
2.1
1500
Power control Variable pulse group modu-lation set
by front panel control. Amplitude set by micro/macro range
buttons.
Pinpoint coag.
170W(±20%)
5.1
2150
Spray coag.
79W(±20%)
8.7
4000
Load resistance for maximum output power
- 100 ohms (non-inductive) for Macro
- 50 ohms (non-inductive) for Micro
Power and voltage output data diagrams are shown at the
end of section 5.
Test load - 100 ohms (non-inductive)
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
P5/37
AUDIBLE INDICATORS
SAFETY
Running tones *
General
Monopolar
Designed to comply with EN60601-1:1990 Medical
electrical equipment - general requirements for safety and
IEC60601-2-2: 1998 High-frequency surgical equipment particular requirements for safety (3rd edition).
Cut, blend or specialist cut
Coagulation, pinpoint or spray
950Hz
800Hz
Bipolar
Micro or macro
730Hz
(* Approximate values, adjustable volume)
General classification, Class 1, Type CF (Defibrillator
proof).
Drip-proof (IPX 1)
Patient leakage (risk) current: always less than 100
microamps to earth (ground) from all patient circuits at 230V
Touch button
50Hz as required by EN60601-1:1990 for the unit in normal
A ‘bleep’ indicates when any function button is pressed, condition (typically less than 10 microamps).
the volume is adjustable with running tones above.
Battery
Alarm
This equipment contains a nickel metal hydride battery. In
All modes - alternating two tone preset to maximum volume the event of failure of the display p.c.b. battery, or if the
electrosurgical unit is to be disposed of, it is not necessary
(with flashing display).
to remove the battery or to return it to Eschmann
Equipment. The battery charge life is six months from a
full charge of 48 hours. The battery charges automatically
VISUAL INDICATORS
when the unit is ‘on’.
Green lamp in mains ‘on/off’ switch to indicate power ‘on’
Electrode isolation
from the rear of the electrosurgical unit.
The plate electrode circuit of this equipment is isolated
Digital displays and ‘function selected’ LEDs indicate power
from earth at both high and low frequency. The bipolar
‘on’ from the front of the electrosurgical unit
output is also fully isolated at both high and low frequency.
Three digital displays indicate typical power* in cut, coag
and bipolar modes. (Note: These can be set to display the Class
typical power* available in watts, or a numerical value up Class 1 denotes that the equipment must be earthed via
to 10).
the protective conductor in the 3-core mains cable
(* typical power is an indication of the average power connected to a 3-pin plug.
delivered over a range of load resistances. It is less
Safety category
than the power delivered to the rated load for maximum
power.)
This symbol denotes that the equipment is of type
CF, i.e. that it complies with type CF leakage current
Cutting modes
Yellow lamp
requirements. The symbol also denotes that the equipment
Coag modes
Blue lamp
will not be damaged by defibrillator discharge and that the
Bipolar mode
Blue lamp
plate electrode need not be removed from the patient if a
Green and bright green function selected LED indicators defibrillator is used.
above (or below) the relevant touch buttons detailed as
Non-ionizing radiation
follows:
Power output ‘activated’ LED indicators for:
Monopolar:
Cut, blend, specialist cut (green)
Pinpoint coag (green)
Spray coag (bright green)
Bipolar:
Micro and macro power range (green)
This symbol warns the user of the possibility of
non-ionizing radiation being emitted by this equipment.
Flammable gases
The TD830 equipment is not suitable for use in the
presence of a flammable anaesthetic mixture with air or
with Oxygen or Nitrous Oxide.
Protection
IPX 1 This symbol (drip proof) denotes that the equipment
meets the requirements of IEC529 for dripping water.
P6/37
Downloaded from www.Manualslib.com manuals search engine
E-IM54f
TD830
ELECTROSURGICAL UNIT
DUTY CYCLE
The duty cycle rating of 10 seconds ‘on’, 30 seconds ‘off’,
specified on the serial plate as 10s:40s, indicates that the
equipment can be used at full output power in any mode
for 10 seconds with a 30 seconds rest period. The unit
can remain connected to the mains electrical supply with
the mains switch in the ‘on’ position continuously. The ‘on’
period at lower power levels can be extended.
I
The symbols and
adjacent to the mains ‘on/off’ switch
(42) indicate the ON and OFF positions respectively.
The symbol
adjacent to item (34) indicates the
‘equipotentiality’ connection point. (Means for connection
of a potential equalization conductor).
The symbol
adjacent to button (35) signifies the button
in the non-pressed (i.e. normal) state. (Digital display on
Note: For numbers in brackets refer to Illustration 11 which the front of the electrosurgical unit is a typical power figure,
e.g. 0-200watts for monopolar blend).
folds out at the end of this manual.
OTHER SYMBOLS
This symbol above item (33) indicates the running The symbol
adjacent to button (35) signifies the button
tone volume control which has a minimum sound level of in the pressed (i.e. activated) state. (Digital display on the
45dBA.
front of the electrosurgical unit shows a numerical value
up to 10).
This symbol above items (1, 14, 21 and 33)
indicates increasing power output (1, 14 and 21) and
adjacent to sockets (36, 37 and 38)
increasing volume (33) for running tone and ‘bleep’ volume. The symbol
indicates the connection point for footswitches.
This symbol on the rear panel serial plate (39) and
above the mains inlet (40) indicates that the equipment is The symbol SN indicates serial number.
for use on alternating current only.
The symbol REF indicates catalogue number.
This symbol on the rear panel serial plate
indicates that the mains input fuses, rating and type, are The symbol
indicates date of manufacture.
as shown below the symbol.
BUTTON SYMBOLS
The symbol
on the front panel denotes that the plate
electrode is isolated from earth at high frequency. (Note:
The plate electrode is also known as the dispersive, neutral,
return, indifferent or patient plate electrode and is often
simply called the ‘pad’).
Press to select bipolar - micro range
Press to select bipolar - macro range
The symbol
on the front panel adjacent to the active
outputs (24 to 28) denotes dangerous voltages.
The symbol
on the rear panel serial plate warns the
user to read the accompanying documents, the
‘Instructions for use’.
Press to select monopolar - cut
Press to select monopolar - blend
The symbol
above sockets (25, 26, 27 and 28)
denotes connection socket for a two button electrode
handle (fingerswitch type).
Press to select monopolar - specialist cut
The symbol
adjacent to sockets (24, 26 and 31)
denotes connection socket for a non-switched active
handle actuated by a footswitch.
Press to select monopolar - pinpoint coag.
This symbol
on the monopolar standby selection
button (32) (monopolar ‘on/off’ toggle button) indicates
standby mode for part of the equipment only.
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
Press to select monopolar - spray coag.
P7/37
ALARM SYMBOLS
PCM - Indicates the ‘Plate continuity monitor’ alarm has
been activated.
PAM - Indicates the ‘Plate attachment monitor’ alarm has
been activated.
PEM - Indicates the ‘Patient earth monitor’ alarm has been
activated.
PVM - Indicates the ‘Plate voltage monitor’ alarm has
been activated.
EPM - Indicates the ‘Excess power monitor’ alarm has
been activated.
- Indicates the ‘Multiple activation’ alarm has been
activated.
! - Indicates an internal fault has been detected (see section
5 ‘Alarm conditions’ for details).
ENVIRONMENTAL CONDITIONS FOR
TRANSPORT AND STORAGE
Ambient temperature range:
-30°C to +50°C
Relative humidity range:
30-90% RH non-condensing
Atmospheric pressure range:
1060hPa down to 690hPa
P8/37
Downloaded from www.Manualslib.com manuals search engine
E-IM54f
TD830
ELECTROSURGICAL UNIT
3.0 SAFETY NOTES & ALARMS
Attention to the following points is necessary
in order to reduce the risk of accidental burns
during the use of this electrosurgical unit. Other
safety notes and warnings are also given within
the text of this manual and these should be
noted before using this electrosurgical unit.
DO:
♦
DO NOT:
Use only Eschmann accessories, in particular active ♦
and plate electrodes and cables which should ♦
preferably be no more than 3 metres in length. All
Eschmann accessories are rated above the
corresponding maximum peak output voltage of this
electrosurgical unit (see graphs in section 5). For
reference Eschmann accessories can be used safely
at the following h.f. peak voltages:
Bipolar active - 400 V peak
Monopolar active - 4000 V peak
Monopolar plate electrode - 2250 V peak
♦
♦
Use bipolar techniques in preference to monopolar, ♦
whenever possible. For surgical procedures on parts
of the body having a relatively small cross-sectional
area, the use of bipolar techniques may be desirable
♦
to avoid unwanted coagulation.
Seek approved qualified advice (e.g. cardiology
department) before using this electrosurgical unit on
patients with implanted pacemakers or other active
implants to avoid interference or damage to the implant.
♦
Monitor such patients carefully.
♦
Set power output controls to the minimum setting
before use and select the minimum power setting to
achieve the desired effect.
♦
Check all cables and accessories routinely before
use. In particular, electrode cables and endoscopically ♦
used accessories for possible damage to the insulation.
♦
Ensure the entire area of the plate electrode is reliably
attached to the patient’s body and as close to the ♦
operating field as possible.
♦
♦
Store temporarily unused active electrodes such that
they are isolated from the patient [e.g. use a quiver
♦
(REF 83-186-38) to hold active accessories when not
in use].
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
Do not use uninsulated forceps, monopolar or bipolar.
Do not place monitoring electrodes close to the
operating site. When high frequency surgical
equipment (i.e. electrosurgical equip-ment) and
physiological monitoring equip-ment are used
simultaneously on the same patient, any monitoring
electrodes should be placed as far as possible from
the surgical electrodes. Needle monitoring electrodes
are not recommended. In all cases, monitoring
systems incorporating high frequency current limiting
devices are recommended.
Do not allow active cables to drape across or contact
the patient's body or contact the cables or leads of
other equipment.
Do not use flammable anaesthetics, flammable
solvents or oxidizing gases such as nitrous oxide
(N2O) or oxygen if the surgical procedure is carried
out in the region of the thorax or head, unless these
agents are sucked away.
Do not allow the patient's body to touch conductive
objects. The patient should not come into contact with
metal parts which are earthed or which have an
appreciable capacitance to earth, e.g. operation table,
supports, etc. The use of antistatic sheeting is
recommended for this purpose.
Do not use hook type active cables (with or without
adapters designed for use with hook type cables) use
4 mm or 8 mm plug type active cables.
Do not reuse disposable plate electrodes.
Do not rely solely on surgical gloves to provide
insulation.
Do not allow ‘skin-to-skin’ contact (e.g. between the
arms and body of the patient), this can be avoided, for
example by the insertion of adequate dry gauze.
P9/37
Attention to the following points will prolong
the
life
and
efficiency
of
your
electrosurgical unit and will help to avoid the
risk of accidents, or damage:
DO:
♦
♦
♦
♦
Excess Power Monitor (EPM).
3.4 The unit monitors its output power in monopolar and
bipolar modes, compares this with the maximum allowed
power under single fault conditions and alarms, disabling
output, if the delivered power exceeds the allowed value
by more than a certain margin.
Patient Earth Monitor (PEM).
Switch off and disconnect from the mains electrical
supply prior to cleaning the equipment and when it is 3.5 The PEM circuit detects a low impedance path
between the patient and earth. It also reduces the chances
not in use.
of secondary contact electrosurgical burns. The circuit
Grasp the connector(s) not the cable when connecting detects earth contact paths while the plate electrode is
and disconnecting cables and leads from the attached to the patient and the unit, when a monopolar
equipment.
fingerswitch or footswitch is pressed, but before the
monopolar output is activated. If such an earth path is
Contact the hospital electronics engineer or Eschmann
present, the alarm will operate until the path has been
After Sales Service Department if the equipment fails
removed and the unit reset by switching ‘off’ and ‘on’ at the
to function after checking.
mains supply switch or if a pressed monopolar footswitch
Ensure the equipment is serviced regularly (at least or fingerswitch is released.
every six months). Contact Eschmann After Sales Plate Voltage Monitor (PVM).
Service Department for details.
3.6 Is designed to prevent dangerous electrosurgical
voltages appearing on the patient’s body. Such voltages
DO NOT:
could occur because of insulation faults in the active circuit
♦
Use a faulty unit as failure could result in an unintended or because the active electrode is in contact with an earthed
object and could cause burns at points of contact between
increase of output power.
the patient and conductive objects. This alarm will operate
♦
Service this equipment unless you are suitably qualified when monopolar output power is ‘on’ and the fingerswitch
to do so.
or footswitch is operated. This monitor will not detect a
poor contact between the plate electrode and the patient.
Alarm Circuits
Note: Under specific electrical conditions, with the plate
electrode completely detached from the patient, the PVM
3.1 The unit has several alarm circuits as detailed in
could alarm.
the following sections.
Plate Continuity Monitor (PCM).
3.2 Ensures that the unit cannot be used in the
monopolar mode with an electrically defective plate
electrode or cable. It does not monitor the contact between
the plate electrode and the patient’s body. It is reset
automatically when a satisfactory plate electrode and cable
are connected to the unit.
Plate Attachment Monitor (PAM).
3.3 Ensures that the unit cannot be used in the
monopolar mode if a divided plate electrode is attached to
the patient’s body incorrectly. Applies only when a divided
plate electrode is used and will not operate if monopolar is
switched ‘off’ by the monopolar standby selection button.
P10/37
Downloaded from www.Manualslib.com manuals search engine
Multiple Activation Alarm.
3.7 Operates if more than one fingerswitch or footswitch
is pressed at the same time in a given user section (i.e.
monopolar USER 1 or USER 2 or bipolar). It will
automatically be reset when ALL activations have been
released. This alarm will also operate if a footswitch or
fingerswitch is being activated when the unit is switched
‘on’. (Note: This could indicate a damaged accessory
locked in the ‘on’ position).
Internal error alarm
3.8 Operates if an internal fault arises such as power
being detected when not enabled by a footswitch or
fingerswitch.
E-IM54f
TD830
ELECTROSURGICAL UNIT
4.0 INTRODUCTION
GENERAL
4.7 A plate electrode is not required and safety is
increased even further as two power range outputs are
available allowing precise setting of the power output.
4.8 The bipolar mode of operation offers a choice of
Note: For numbers in brackets refer to Illustration 11 which footswitch, electrical or pneumatic.
folds out at the end of this manual.
4.1 The TD830 electrosurgical unit provides two outputs
both for high power monopolar cut and coag (on a first
come first served basis), together with a separate output
for bipolar coag. Bipolar coag output is available
simultaneously with either monopolar output. The unit
incorporates touch button controls, seven segment LED
(light emitting diode) displays and LED visual indicators.
4.2 The digital displays can show two scale modes, one
shows the typical power available in watts the second
shows a numerical value up to 10 (the latter is only
displayed whilst digital display range selection toggle button
(35) is held pressed).
Monopolar standby mode
4.9 The unit can be switched into monopolar standby
mode, in which only bipolar power is available, using
monopolar standby selection button (32). This button is a
toggle switch, each press of the button will turn monopolar
mode ‘on’ (if ‘off’) or ‘off’ (if ‘on’). This button (32) can be
considered as an ‘on/off’ switch for monopolar, the
advantage being that if only bipolar outputs are required
and monopolar standby mode is selected, no plate
electrode needs to be connected to the unit. (If monopolar
standby mode is not selected for ‘bipolar use only’ and a
patient plate electrode is not connected to the unit the PAM
alarm or PCM alarm will activate). The unit is in monopolar
standby mode when only the digital display for bipolar can
be seen illuminated.
Note: The power delivered depends on the resistance
between the active electrode and the plate electrode (or
between the tips of bipolar forceps) at the time the power
is applied. This resistance can vary widely and many times Digital display options
per second during the application of h.f. output.
4.10 The digital displays normally show the typical power*
4.3 The unit incorporates several alarm code indicators available in watts. The digital displays can also show a
as detailed in the Technical Data section. Innovative safety numerical value up to 10 (maximum). To display the power
features include a divided plate electrode monitoring setting as a numerical value up to 10 press and hold the
system called the Plate Attachment Monitor (PAM) as well digital display range selection button (35). The display will
as the standard Eschmann Plate Voltage Monitor (PVM), revert to the normal setting (and display the typical power*
Patient Earth Monitor (PEM) , and Plate Continuity Monitor available in watts) when the button is released. (* typical
(PCM). An Excess Power Monitor (EPM) and a Multiple power is an indication of the average power delivered over
activation alarm are also included. A high frequency a range of load resistances. It is less than the power
leakage control circuit is included in the monopolar mode. delivered to the rated load for maximum power.)
4.4 Within this manual the terms USER 1 and USER 2 are
used to distinguish between the two monopolar outputs. ACCESSORIES
The bipolar output is fully independent of the monopolar 4.11 The equipment is designed to be used with the
output and can be considered as USER 3.
Eschmann range of active, plate and bipolar cables. The
front panel is colour coded, PALE BLUE for bipolar,
OPERATING MODES / DISPLAY OPTION
YELLOW for monopolar cut, blend and specialist cut, and
BLUE for monopolar coag. For a list of all accessories and
Monopolar mode
an index of accessory illustrations, see section 7.
4.5 The monopolar operating mode offers a choice of
fingerswitch or footswitch control of ‘cut’ or ‘coagulation’
outputs for USER 1. Fingerswitch operation utilises the
Eschmann two-button fingerswitch whilst footswitch control
requires the use of an active electrode handle in conjunction
with one of a range of electrical footswitches.
Bipolar mode
ASSOCIATED PUBLICATIONS
4.12
This manual contains preparation for use, and
operation instructions only. For a detailed technical
description, together with complete maintenance
procedures including an illustrated parts list refer to the
TD830 Service Manual E-SM44 (part number 698260).
4.6 Bipolar coag is an efficient method of effecting EQUIPMENT CERTIFICATION
haemostasis, and closure of vessels such as Fallopian
tubes. It is extremely safe to use, as the main current path 4.13 The electrosurgical unit fully complies with the major
is between the tips of the bipolar forceps and the tissue international safety standards (see Technical Data).
held between them is directly in view of the surgeon at all
CAUTION
times.
This electrosurgical unit is to be operated by
medically qualified personnel only.
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
P11/37
ELECTROSURGICAL TECHNIQUES
General
4.14 When using electrosurgery, it must always be
remembered that its purpose is to cut, destroy or modify
tissue. This can occur at any point on the body where the
concentration of electrical current becomes sufficiently
high. Under fault, or misuse conditions, this current
concentration can occur on points on the body other than
the desired operation site and can cause burns.
4.15 To help reduce the possibility of such accidental
burns, the unit is fitted with a leakage limiting circuit in the
monopolar section. This circuit detects the conditions that
may lead to an increased possibility of accidental burns
from leakage and limits the leakage current accordingly.
The HF leakage limiter does not generate an alarm. In
addition, the following paragraphs give guidance for the
safe use of the electrosurgical unit.
4.16 With monopolar electrosurgery, the surgeon uses a
single point electrode to concentrate high frequency electric
current in the body tissues close to the electrode tip. If the
operating site is linked to the bulk of the body only by a
tissue path of small cross-sectional area, e.g. Fallopian
tube or penis, the surgical effect of the electrosurgery can
extend along it from the electrode with undesirable results.
♦
♦
♦
♦
♦
Do not apply the plate electrode adjacent to ECG
electrodes and cables, or to an area where fluids
could pool.
With babies and small children the plate electrode
should always be applied to the trunk section of the
body to give maximum contact with the skin. It is
unsatisfactory to attach the plate electrode to small
limbs where overlap of the plate electrode ends can
occur.
Subject to application instructions, the plate electrode
should be on a smooth, resilient, muscular site well
supplied with blood vessels and as close as possible
to the operating area. The site should be clean and
dry, and shaved of excessive hair. The plate electrode
should be securely attached with its whole area in
contact with the patient’s body.
Do not cut or modify the plate electrode in any way.
Do not reuse disposable adhesive plate electrodes,
as adhesive failure could occur.
4.21 Adhesive plate electrodes (REF 83-122-42 and 83122-46) with detachable cables, should only be used with
a connector cable (REF 83-122-93) that requires complete
insertion of the connector tab. This ensures that contact
between the tab and any conductive object cannot occur.
Footswitches
4.17 Normally the surgeon will have ensured that the
operating site is bounded by a bulk of tissue into which the
current can disperse, so that the cut or coagulating effect
is limited to the tissues close to the electrode. The current
then disperses into the body and leaves it by way of the
relatively large plate electrode, having passed harmlessly
through any intervening tissue and organs.
4.22 Footswitches (REF 83-581-09 and 83-581-13 for
yellow, or REF 83-581-17 and 83-581-25 for blue, or twin
footswitches REF 83-109-39 and 83-109-47) must always
be under the direct control of the surgeon or the designated
assistant. Precautions must be taken to prevent
footswitches contacting the operating table base, castors
of trolleys, or other equipment under which they could be
4.18 In contrast, with bipolar electrosurgery there is no trapped in the ‘on’ position.
plate electrode and the surgical effect occurs between the 4.23 During a surgical procedure and when the power
tips of a pair of insulated bipolar forceps held by the output is not activated, check that:
surgeon. The legs of the forceps constitute two
♦ The green ‘mains’ supply lamp is ‘on’.
electrosurgical active electrodes.
♦ The bipolar and monopolar (cut and coag) power
activated LEDs are ‘off’.
Monopolar Electrosurgery
♦ The audible ‘running tone’ is ‘off’.
Plate Electrode
4.19 The plate electrode is also known as a neutral Active Electrode Handles
electrode, passive, return, dispersive, indifferent or patient 4.24 During surgery, when an active electrode handle
plate electrode or simply as the ‘pad’.
(REF 83-140-20 without cable, or REF 83-141-01,
4.20 Always use a plate electrode of proven technical 83-141-36, 83-141-28 and 83-141-44 with cable) or
design and safety, and in strict accordance with the fingerswitch (REF 83-582-06, 83-583-97 or 83-585-94) is
application instructions issued with the plate electrode. not in use, it should be put in an electrosurgical quiver
Burns at the site of the plate electrode are the most (REF 83-186-38) or similar receptacle. Do not put the
common form of electrosurgical accident. Maximum handle or fingerswitch:
contact with the patient's skin must be maintained ♦ On the patient drapes.
throughout the surgical procedure. The following points ♦ With other instruments.
should also be noted:
♦ On the electrosurgical unit cabinet or on any other
♦ Do not apply any plate electrode over bony
theatre equipment.
protuberances, metal implants, folds of skin, scar
tissue, hairy areas, any form of skin discoloration, an This will further reduce the possibility of electrosurgical
burns due to accidental activation of the electrosurgical
injury, or to limbs with restricted blood supply.
P12/37
Downloaded from www.Manualslib.com manuals search engine
E-IM54f
TD830
ELECTROSURGICAL UNIT
unit. Also do not use active electrode handles that have
cables fitted with ‘hook’ type connectors (those detailed
above are suitable for use with this unit). The absence of
exposed metal parts on pin or jack plug designs reduces
the risk of operator burns compared to cables with the
‘hook’ type design.
Power Control
4.33 Use only the minimum power control setting
necessary to produce the desired electrosurgical effect.
If in doubt start at a low setting and increase the output
in small steps until the desired effect is produced.
4.34 If, during a surgical procedure, the output
appears to drop or the electrosurgical performance
4.25 Only active cables supplied by Eschmann should decreases, stop and check the plate electrode for
be used. This is of particular importance, as cables with correct application and connection before increasing
inadequate insulation can cause burns due to insulation the control setting. Also check all active connections.
failure.
4.35 If only one type of output is required, i.e. cut or
4.26 It is good practice to ensure that active cables are ‘coagulation’, it is good practice to set the other output
correctly positioned when in use and are not allowed to controls to minimum.
drape across the patient's body or across other cables.
Active Cables
Current Paths
4.27 Active cables of the ‘hook’ type have been replaced
by pin or jack plug designs (REF 83-135-12, 83-137-17,
83-138-06 and 83-138-22) any of these are suitable for
use with this unit. Old style cables with ‘hook’ type
connectors should not be used. The absence of exposed
metal parts on pin or jack plug designs reduces the risk of
operator burns compared to cables with the ‘hook’ type
design.
4.36 The path of an electrosurgical current through a
patient's body should, whenever possible, be the
shortest path. If it is suspected that this path could be
bypassed, for example between limbs, an insulating
medium should be used.
4.37 It is advisable that skin to skin contact between
parts of the patient's body, for example between the
4.28 Active cables 3 metres long should be used in normal arms and the trunk, should be avoided. Dry gauze or
situations, even if it means repositioning the electrosurgical other material should be used to provide separation.
unit. Longer types, of 5 metres or more, should only be Accidental Contact
used where the electrosurgical output is a fixed distance
from the operation table, e.g. with certain types of pendant 4.38 Do not allow contact to occur between any part
supply, or with lamina airflow curtains. Longer cables of the patient's body and any metal or conductive
material, such as parts of the operation table,
increase the risk of electrical or mechanical hazard.
anaesthetic screens or drip stands.
Active Electrodes
4.29 The correct type of active electrode (see section 7) Bipolar Electrosurgery
should always be used. Electrodes should be kept clean
during use either by wiping with a solvent on a sterile swab General
or by using a sterile abrasive pad.
4.39 Bipolar coag provides a very safe and precise
means of controlling bleeding in many types of surgery,
CAUTION
Ensure that the active cable is unplugged from and can be used in conjunction with monopolar
techniques where cutting, or higher coagulation powers
the unit when cleaning electrodes during a are required.
surgical procedure.
4.30 Ball type coag electrodes should be discarded if the
surface becomes pitted. Do not sterilize electrodes which
are designed as ‘single use disposable’ (also see the leaflet
detailed in section 1, supplied with all accessories ).
Fingerswitches
4.31 For many procedures, ‘on/off’ control of cut and coag
outputs by means of buttons on a fingerswitch offers the
surgeon more precise control. Reusable fingerswitches
have only a limited number of re-sterilization cycles.
Regular testing and inspection is therefore essential. Do
not sterilize and reuse ‘single use’ types.
4.32 Do not connect footswitches to the USER 1 output
of the electrosurgical unit when a fingerswitch is being used
to control the USER 1 output.
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
4.40 With bipolar techniques, the electrosurgical
current flows mainly between the tips of the bipolar
forceps or between the contact points of a special
purpose electrode and therefore does not require a
plate electrode. Bipolar electrosurgery is safer than
monopolar, the main safety requirements are:
♦ Always use Eschmann bipolar accessories.
♦ After sterilization, always ensure that the
accessories and cable connector are completely
dry.
♦ Always ensure that accessories are sterilized by
a process compatible with the accessory, consult
the accessory leaflet provided with all Eschmann
accessories (publication number E-IM50 part
number 604802).
P13/37
Flammable Anaesthetics and Spirits
if the electrosurgical active electrode and plate electrode
are kept a few inches away from the pacemaker pulse
generator and its electrode leads; while to avoid burns to
the myocardium at the electrode site, or possible ventricular
fibrillation, they should remain at least 15 cm away from
the heart.
4.41 The use of flammable anaesthetics and nitrous oxide
(N2O) or oxygen should be avoided if a surgical procedure
is to be carried out in the region of the thorax or the head,
unless these agents are removed by suction. Flammable
agents used for cleaning or disinfecting, or as solvents of
adhesives should be allowed to evaporate before the 4.47 It will also be sensible to choose a location for the
application of electrosurgery. (Also see WARNING above plate electrode which puts the path of electrosurgical
current through the patient's body as far as possible from
section 6.1).
the heart and from the pacemaker and its leads.
4.42 There is a risk of pooling of flammable fluids under
the patient or in body depressions such as the umbilicus 4.48 If the pacemaker is programmable, it is suggested
and in body cavities such as the vagina. Any fluid pooled that it is set to either the VVT (ventricular sensing and
in these areas should be mopped up or removed by suction triggering) or the V00 (fixed rate) mode (preferably the latter
if the patient is not at particular risk from a competitive
(for gaseous fluids) before the equipment is used.
rhythm) so that it functions as a fixed rate pacemaker during
4.43 Attention is drawn to the danger of ignition of the surgical procedure.
endogenous gases (i.e. those gases produced within the
body). Some materials, e.g. cotton wool and gauze, when 4.49 It is further recommended (References ¹ and ²) that
saturated with oxygen may be ignited by sparks produced the peripheral pulse be monitored during the course of
electrosurgery, and (Reference ²) that precautions be taken
in normal use of the equipment.
to ensure that the patient's well-being is maintained in the
event of interference with pacemaker operation. By using
Implanted Cardiac Pacemakers
the electrosurgical equipment in short bursts, it is
General
suggested (References ¹ and ²) that, at most, only one or
two
beats will be affected.
4.44 If in monopolar electrosurgery the intervening tissue
and organs contain an implanted pacemaker and its leads,
it is possible under certain conditions that the
electrosurgical current will have undesirable effects. The
following notes offer guidance on the risks and on possible
precautions.
WARNING
For patients with cardiac pacemakers or other
active implants, a possible hazard exists
because interference with the action of the
pacemaker may occur or the pacemaker may
be damaged. In case of doubt, approved
qualified advice should be obtained.
Risks
4.45 Operating with electrosurgical equipment on patients
who have an implanted pacemaker incurs the risks of
affecting the pacemaker's operation, either by direct contact
or by radiated interference, and of inducing burns at the
pacemaker electrode implantation site. Clearly the risks
are very much lower with bipolar than with monopolar
electrosurgery, because of the highly localised current path.
Precautions
WARNING
More severe consequences, local heating and
destruction of the pacemaker circuit
(Reference ¹), would result if an electro-surgical
electrode touched the pacemaker.
Electrosurgical Interference
4.50 Electrosurgical equipment can produce interference
with other electromedical apparatus used in the operating
theatre. Particular attention should be given to the selection
of ECG monitors with good interference rejection if it is
desired to monitor the patient when the electrosurgical
equipment is active.
4.51 In certain conditions a reduction in ECG interference
can be obtained by placing the ECG electrodes closer
together than normal. Modern monitors with high input
isolation allow electrode placement close to the
electrosurgical plate electrode, if required.
4.52 It is inadvisable to use mobile telephones or ‘walkietalkie’ equivalent equipment close to electrosurgical or any
other medical electrical equipment.
4.53 To minimize the possibility of electrical interference
4.46 Recommendations published in the U.S.A. from an arc, between the active electrode and the patient,
(References ¹ and ²) suggest that an implanted pacemaker affecting the TD830, it is good practise, in common with all
should be unaffected, or only revert to fixed-rate operation, electrosurgical units, to keep them as far as is reasonably
possible from the surgical site.
P14/37
Downloaded from www.Manualslib.com manuals search engine
E-IM54f
TD830
ELECTROSURGICAL UNIT
using an S-Lead [REF 83-139-41 (Olympus 9mm),
or REF 83-139-49 (Pentax 8mm)] connected to the
S-Lead socket (29, Fig 12) of the TD830
Electrosurgical unit and the flexible endoscope.
Using Two Electrosurgical Units
4.54 This electrosurgical unit has dual monopolar outputs
(first come first served) and simultaneous bipolar output.
It is therefore very unlikely that another electrosurgical unit
would need to be used on the same patient at the same
time, however if it is necessary follow the guidelines
detailed in section 4.55.
4.55 In certain procedures it is convenient or even
essential that two electrosurgical units are used on the
patient at the same time. There is no reason why such
operations should not be safely carried out if the following
points are observed:
♦ Do not use a fully floating output unit with an earth
referenced unit or even a partially floating unit, for
example:
♦
♦
♦
♦
Eschmann TD411 or DS402 and a Valleylab
Force 4
♦ Eschmann TD411 or DS402 and a GU Solstar
or any older GU unit.
Note: All Eschmann units have fully floating
outputs and any two may be used in
combination.
Always use two plate electrodes, one for each unit.
Never join plate cables or overlap plate electrodes.
Position each plate electrode as close as possible to
the respective operation site. Do not allow
electrosurgical current paths to intersect, e.g. if the
main procedure is in the thoracic region, the plate
electrode for this unit should be placed on the patient’s
back, buttocks, or upper thigh. If the secondary
procedure is vein removal from the leg, the second
plate electrode should be placed on the thigh or calf
of this leg.
For maximum safety use fingerswitches wherever
possible. This reduces the possibility of inadvertent
activation of either unit’s output. Always use two
quivers, one for each active handle or fingerswitch.
Using a flexible endoscope
4.56 To ensure that the endoscope connection socket is
not used without a documented procedure in place, the SLead socket (29, Fig.11) is blanked of with a plug to avoid
inadvertent use. Before removing this plug the following
notes should be considered when writing a procedure for
the use of this electrosurgical unit with an endoscope.
a.
b.
When using a flexible endoscope low power settings
only should be used (e.g. maximums of 30% for cut,
50% for pinpoint coagulation). Also because of the
high voltage waveform of spray coagulation it should
only be used at very low power settings see 3 below.
c.
It is important to ensure that the flexible endoscope
used is compatible with the voltages produced by
the TD830 Electrosurgical unit. The maximum
voltages produced by the TD830 Electrosurgical unit
for each mode and display setting are shown in the
graphs at the end of section 5. These should be
compared to the voltage rating of the flexible
endoscope. (Note: Flexible endoscopes complying
with IEC60601-2-18 should detail their voltage rating
in their accompanying documentation. If in doubt
consult the flexible endoscope manufacturer).
SERVICING
4.57 It is recommended that electrosurgical safety
checks and routine servicing are carried out at regular
intervals (every six months) and only by Eschmann trained
personnel or Eschmann trained hospital engineers,
otherwise the warranty could be infringed.
4.58 Read the information given in this manual carefully
before using, cleaning, sterilizing, or servicing the
electrosurgical unit.
References
¹
²
‘Optimal resources for implantable cardiac
pacemakers’ (Section on electromagnetic
interference) - a report by the Inter-Society
Commission for Heart Disease Resources, published
by the American Heart Association in their journal
‘Circulation’ (1983), 68(1):232A-233A.
‘Electrosurgical device interference with implanted
pacemakers’ - a question and answer section of the
‘Journal of the American Medical Association’. (1978),
239(18:1910).
To avoid the risk of potential burns caused by h.f.
voltages on the body of the flexible endoscope some
users may prefer to connect the endoscope to the
TD830 Electrosurgical unit. This is accomplished
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
P15/37
5.0 OPERATION
WARNING
1) This electrosurgical unit is to be operated
by medically qualified personnel only.
2) Note that interference produced by the
operation of this unit may adversely influence
the operation of other electronic equipment.
3) While the electrosurgical unit is activated
(i.e. the running tone can be heard and power
activated indicators are illuminated) do NOT
touch the patient with a contact area of less
than 10cm2. This is especially important when
monopolar mode is being used at high power
settings.
4) Do not touch a monopolar or bipolar active
electrode with its active cable plugged into the
electrosurgical unit, even when changing the
electrode.
♦
♦
♦
♦
Plate electrode. With reusable types, examine the
plate cable and plug for signs of wear. Ensure that the
insulation of the connecting cable is in good condition.
Accessory cables/cords (reusable monopolar active
and bipolar types). Examine all cables for condition
and attachment to end fittings.
Note: All active cables are subject to deterioration
caused by repeated sterilization and should be
renewed regularly.
Fingerswitches. Check buttons on fingerswitches
for function and free action.
Note: All fingerswitches have a limited life and should
be renewed regularly.
Footswitches. Check all footswitch cables and
connectors for condition. Check all switches for
correct function and free action.
CONNECTION OF ACCESSORIES PRIOR
TO USE or SYSTEMS CHECK
Note: For numbers in brackets refer to Illustration 11 which
folds out at the end of this manual.
5.1 Before use ensure the details covered in the
preliminary information sections 1.5 to 1.7 have been Monopolar
performed and the routine checks below in section 5.2 have 5.3 There is a choice of active cable connections to the
been completed. Connect the required accessories as lower front panel of the unit for the monopolar outputs:
detailed in sections 5.3 to 5.9 and then carry out a systems
♦ A three-pin fingerswitch connector can be plugged
check detailed in sections 5.10 to 5.17 before using the
into active fingerswitch sockets (27 and 28) for USER
unit.
2 and active fingerswitch sockets (25 and 26) for
USER 1.
ROUTINE CHECK
♦ A single active cable with 4 mm plug (footswitch
5.2 Routine checking should be limited to external
operated) can be connected to active electrode
checks of the unit, and must include all accessories and
socket (26) for USER 1.
cables. Routine checks must be carried out each time the
unit is used as follows:
CAUTION
WARNING
If in doubt, discard suspect component and
substitute with one of known function,
continuity and condition.
♦
Mains electrical supply cable/cord. Examine mains
electrical supply plug for condition and correct
attachment to cable/cord. Check condition of cable/
cord insulation over its entire length.
WARNING
Do not repair a damaged disposable plate
electrode or reusable plate electrode. Always
substitute a new plate electrode and, if
necessary, the cable. Always ensure that a
supply of new plate electrodes is readily
available.
P16/37
Downloaded from www.Manualslib.com manuals search engine
Incorrect insertion of a single 4mm monopolar
active plug into either of the USER 1
fingerswitch sockets (25) can cause damage
to the TD830 if the output is energized by a
footswitch.
♦
A single active cable with 8 mm plug (footswitch
operated) can be connected to active electrode socket
(24) for use by USER 1 only
Note: This overrides any cables plugged into the
USER 1 sockets (25 and 26).
CAUTIONS
Only one of the above active connections
should be connected to the unit at any one time
for USER 1.
If an adapter is connected to the unit and it does
not have an active cable connected to it, the
adapter must be removed from the unit.
E-IM54f
TD830
ELECTROSURGICAL UNIT
If an 8 mm active adapter is connected to the
unit, the fingerswitch sockets (25 and 26) are
electrically disconnected.
Bipolar
5.8 Connect a bipolar cable REF 83-041-55 to the
bipolar output socket (31) and either a white electric
footswitch REF 83-581-76 (Fig. 10) to the bipolar electric
5.4 For USER 1 connect an Eschmann fingerswitch (Fig. socket (37) or a white pneumatic footswitch REF 83-5816) to fingerswitch sockets (25 and 26).
68 (Fig. 10) to the bipolar pneumatic socket (36). Do not
5.5 Alternatively for USER 1 plug an 8mm active cable connect both types of footswitch at the same time.
(BOVIE) into the 8mm active electrode socket (24) in 5.9 If the electrosurgical unit is only to be used in the
combination with a footswitch (Fig. 4) yellow for cut, and/ bipolar mode select the monopolar standby mode (see
or blue for coag, which should be plugged into either section 5.18). The monopolar digital displays (8 and 23)
monopolar footswitch socket (38). These sockets are will not be illuminated and this will enable use of the unit
identical, the function is determined by the colour of the without the need to connect a plate electrode to the unit
footswitch. The Eschmann Twin footswitch can also be (PAM alarm or PCM alarm will activate if a plate electrode
used to provide both cut and coag functions (see Fig. 4). is not attached to the unit and it is not in monopolar standby
5.6 Still another option for USER 1 is to connect an
active electrode handle (Fig. 5) to 4mm active electrode
socket (26) and plug a footswitch (Fig. 4) yellow for cut,
and/or blue for coag, into either monopolar footswitch
socket (38). These sockets are identical, the function is
determined by the colour of the footswitch. The Eschmann
Twin footswitch can also be used to provide both cut and
coag functions (see Fig. 4).
5.7 For USER 2 connect an Eschmann fingerswitch to
sockets 27 and 28.
mode).
SYSTEMS CHECK
Note: For numbers in brackets refer to Illustration 11 which
folds out at the end of this manual.
WARNING
If the unit fails any of the following checks it
must not be used until the faults have been
repaired and the checks have been
successfully repeated.
Notes:
♦ Sockets (25 and 26) and (27 and 28) are designed to
accept the Eschmann fingerswitch.
5.10 Connect the required accessories as detailed in
sections 5.3 to 5.9.
♦ The footswitch controlled active connection system
using the 8 mm (BOVIE) socket (24) requires either 5.11 Connect unit to mains power supply using the cable
the combined handle and cable REF 83-141-28 or 83- supplied connected to the mains inlet (40), and if the supply
141-44 (Fig. 5) or the separate handle REF 83-140- socket is controlled by a switch, ensure that it is switched
20 (Fig. 3) with cable REF 83-138-06 or 83-138-22 ‘on’. Select mains switch (42) to ‘I’ (on). The green lamp
inside the mains switch (42) will come ‘on’ and the front
(Fig. 7).
panel displays (indicators and digital) will illuminate.
♦ For foot control, use one of the range of Eschmann
5.12 The unit may be in monopolar standby mode, if it is
footswitches together with a non-switching active
monopolar displays will not be illuminated and monopolar
handle.
will be disabled unless the monopolar standby button (32)
is pressed (see section 5.18). The unit will power up in the
WARNING
modes and settings in use when the unit was last switched
When a fingerswitch or hand controlled active off, unless the battery powering the memory has failed or
handle is in use for monopolar USER 1, the lost charge, in which case it will power up in monopolar
footswitch(es) must be disconnected from the standby mode.
electrosurgical unit.
CAUTION
For safety, when using only one monopolar
mode, set the unused output control to
minimum e.g.
¨ For coag only, set cut power output control
(14) to minimum.
♦ For cut/blend only, set coag power output
control (21) to minimum.
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
5.13 Press each selector touch button (3, 5, 11, 13, 15,
19 and 20) in turn and check that the green or bright green
indicator adjacent to each touch button comes ‘on’ (2, 4,
10, 12, 16, 17 and 18 respectively). Each time a selector
touch button is pressed there is an audible ‘bleep’ to signify
the button has been pressed (also see the first note after
section 5.15).
5.14 Set monopolar cut, monopolar coag and bipolar
power output controls (14, 21 and 1 respectively) to their
minimum levels (fully anticlockwise).
5.15 If the monopolar mode is to be used ensure that all
monopolar electrodes are safe and activate the unit in all
the monopolar modes to be used, for each USER, by
P17/37
pressing a fingerswitch cut or coag button or by pressing OPERATING INSTRUCTIONS
the appropriate footswitch, yellow for cut, blue for coag.
Check that the relevant power activated indicators Note: For numbers in brackets refer to Illustration 11 which
illuminate, yellow (9) or blue (22) for cut or coag respectively folds out at the end of this manual.
and that the audible running tone can be heard.
♦ Unit settings (e.g. range selection and mode selection)
are stored in the unit's memory during power-off
Notes:
periods and will not need reselecting when power is
♦ The volume of the running tone and the selector
reapplied. Power settings will also be maintained
button ‘bleep’ can be adjusted with the volume control
during power ‘off’ so long as the control knobs are not
(33) on the back of the unit. This should be adjusted
moved.
using a screwdriver of suitable size.
♦
Coagulation output is obtained only by pressing the
fingerswitch blue coag button or the blue footswitch.
♦
Cut output is obtained only by pressing the fingerswitch
yellow cut button or the yellow footswitch.
5.16 If the bipolar output is to be used ensure that the
bipolar electrode is safe and activate the output by pressing
the white footswitch connected (either electric or
pneumatic). Check that the blue power activated indicator
(7) illuminates and that the audible running tone can be
heard.
5.17 Check that the following alarms are functioning
correctly by the test methods that follow:
♦
The audible running tones operate at different
frequencies for each of the three output groups (bipolar,
monopolar cut and monopolar coag) as detailed in the
‘Technical data’ section.
If a monopolar user activates power at the same time
(but after) a bipolar user the tone will continue at the
bipolar frequency until the bipolar user stops activation
when the monopolar user’s frequency will start
(assuming the monopolar user is still activating the
unit). If the monopolar user only activates the unit for
a short period during a long activation by the bipolar
user then the monopolar user’s frequency will not be
heard. The same principles apply if a monopolar user
is activating power and a bipolar user starts to activate
power at the same time.
PAM - To test the PAM alarm attach the divided plate
electrode to the divided plate cable with the divided
Monopolar standby mode
plate cable plugged into the unit. The PAM alarm
should operate with the plate electrode not attached 5.18 If the unit is in monopolar standby mode only bipolar
to the patient.
operation is available, monopolar is disabled and the
monopolar digital displays (8 and 23) will not be seen. The
PCM - To test the PCM alarm remove the non-divided plate unit is switched in and out of standby mode be pressing
electrode from the non-divided plate cable with the the monopolar standby button (32). The switch ‘toggles’
non-divided plate cable plugged into the unit. The standby mode ‘on’ or ‘off’ each time the button is pressed,
PCM alarm should operate.
monopolar standby mode is selected when the digital
PEM - To test the PEM alarm with the plate electrode displays for monopolar (8 and 23) are not illuminated.
attached to the unit allow the plate electrode to contact
the equipotentiality point (34) at the rear of the unit
using a bare metal part of the plate and not a part
covered by adhesive. With cut and coag power levels
both adjusted to zero (for safety) the PEM alarm
should operate when a USER 1 footswitch or
fingerswitch (whichever is connected to the unit) or a
USER 2 fingerswitch is pressed .
Note: This test is not possible with a capacitive plate
electrode.
Monopolar mode
USER 1 and/or USER 2 selection
5.19 The TD830 can be used by two monopolar operators
USER 1 and USER 2 independently of each other but not
simultaneously. Each user can have their own active output
under their own control operated by footswitch or
fingerswitch for USER 1 and fingerswitch only for USER 2
(see sections 5.3 to 5.7)
5.20 USER 1 and USER 2 can be using either cut or
-To test the multiple activation alarm press both a cut coag but the mode set within cut or coag must be the same
and coag footswitch or fingerswitch control, the alarm and the power level set must similarly be the same (unless
should operate.
changed between each USER’s application of power).
The other alarms (PVM, EPM and !) cannot easily be
checked by the user without special test equipment or by
using procedures that could lead to accidental burns. The
function of the EPM alarm should be checked regularly, in
accordance with the Service Manual (see section 1.8).
P18/37
Downloaded from www.Manualslib.com manuals search engine
5.21 Because USER 1 and USER 2 cannot both be
delivering power at the same time, power is available on a
‘first-come-first-served’ basis. (i.e. if USER 1 is using power
and USER 2 presses a fingerswitch to demand power, none
will be available and visa versa). Also note that power will
E-IM54f
TD830
ELECTROSURGICAL UNIT
not become available to the second user until the first user
has finished using power and the second user has released
and reapplied demand (via footswitch or fingerswitch) after
the first user has finished.
♦
The active electrode is correctly fitted in its holder and
the active cable is correctly connected to the unit.
Check that the cable is in good condition.
Guidance on power output settings is given in sections
Plate electrode (required for monopolar mode only) 5.53 to 5.55 and in Table 1.
5.22 If the electrosurgical unit is to be used in monopolar Fingerswitch / active electrode handle
mode and a plate electrode is not connected to the plate
5.27 A fingerswitch or active electrode handle must be
electrode connector socket (30), the unit will alarm (PCM
treated with extreme care at all times.
or PAM) and the audible two-note tone will sound and the
displays will flash. If a non-divided plate cable is being used,
WARNINGS
then connecting a non-divided plate electrode will stop the
1)
When
either
a fingerswitch or an active
alarm. If a divided plate cable is being used then connecting
a divided plate electrode and applying the divided plate electrode handle is connected to the unit, its
electrode correctly to the patient’s body will stop the alarm. location should be closely noted. It should
never be allowed to contact other instruments.
WARNINGS
1) Whichever type of plate electrode is used,
it is most important that the application
instructions, issued with the plate electrode,
are strictly followed.
2) Ensure the pin on the plate connector (or
adapter) is fully inserted into the plate electrode
socket (item 30, Fig. 11) before energising the
output.
2) During surgery, when an active electrode
is not in use, it should always be put in an
electrosurgical quiver or similar receptacle.
This is most important when both bipolar and
monopolar electrodes are required for the same
surgical procedure.
3) Similarly do not touch USER 1 active
electrode when USER 2 output is activated and
visa versa.
5.23 A plate continuity monitoring circuit (PCM) initiates 4) Do not touch a bipolar active electrode
the audiovisual alarm system if there is a break in the non- when monopolar is activated and visa versa.
divided plate cable, or a faulty connection in the non-divided
plate electrode circuit. The electrosurgical output will be 5.28 Connect the required cable and active fingerswitch
instantly cut off and will not be restored until the fault is to the fingerswitch sockets (27 and 28) for USER 2 and
corrected.
fingerswitch sockets (25 and 26) for USER 1 as detailed
5.24 If a divided plate cable and divided plate electrode in sections 5.3 to 5.7.
are being used, the unit will alarm (PAM) if the divided
plate electrode is not making sufficiently good contact with
the patient’s body or if the divided plate electrode or its
connector system is faulty.
5.29 Alternatively for USER 1 connect an active cable
with a 4mm plug into active socket (26) or an active cable
with an 8mm plug (BOVIE) to active socket (24) as detailed
in sections 5.5 to 5.7.
5.25 Apply a plate electrode to the patient in strict Footswitches
accordance with the application instructions supplied with
the plate electrode and connect the plug to the plate 5.30 Eschmann footswitch(es) (Fig. 4) can be plugged
into either footswitch socket (38). With either cut (11), blend
electrode socket (30) on the front panel of the unit.
(13) or specialist cut (15) output selected, pressing a yellow
Note: The Eschmann Flexoplate must be used with (cut) footswitch illuminates the yellow power activated
adapter REF 83-207-56, Fig. 1.
indicator (9), and the cutting running tone will sound.
5.26 If the required surgical effect is not obtained with 5.31 With either pinpoint coag (19) or spray coag (20)
the normal power setting, check the following before selected, pressing a blue (coag) footswitch illuminates the
increasing power:
blue power activated indicator (22), and the coag running
♦
♦
The plate electrode is correctly positioned, and is in tone will sound.
contact with the patient.
5.32 The Eschmann twin footswitch (Fig. 4) has an angled
platform with two light-action switches for cut and coag.
The plate electrode cable is intact, and is plugged into Colour coded bars provide foot location and protection
unit plate electrode socket (30).
against accidental operation.
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
P19/37
WARNINGS
1) Footswitches should be used with care at
all times and should always be under control
of the surgeon or designated assistant, as
inadvertent operation could have serious
results.
2) Footswitches should be kept clear of areas
where other personnel could accidentally press
them, also ensure that they do not get wedged
under theatre equipment.
3) Select mains switch (42) to ‘O’ (‘off’) as
soon as electrosurgery is completed.
especially flexible endoscopes. Most urological
endoscopes are suitable for spray coag. If in doubt check
that the voltage rating of the laparoscopic or endoscopic
instrument is at least as high as the maximum spray coag
open circuit voltage of this electrosurgical unit (see
Technical Date section).
CAUTION
Although power output in the spray mode is
lower than in the pinpoint mode (79W rather
than 170W maximum), the effect on tissue is
much greater in the spray mode. Until
experience is gained, only low control settings
should be used.
5.39 Select pinpoint or spray coag at touch buttons (19)
or (20) respectively and note correct selection by
5.33 Monopolar cut and coag outputs are obtained by illumination of a green indicator (17) for pinpoint coag and
pressing monopolar selector touch buttons (11, 13, 15, 19 a bright green indicator (18) for spray coag adjacent to the
or 20), then setting output controls (14) for cut or (21) for applicable touch buttons.
coag to the required levels, and then using the fingerswitch
Power output
or pressing the appropriate footswitch.
Monopolar cut and coag outputs
5.34 'Running' tones are generated whenever monopolar
cut or coag outputs are activated by footswitch or
fingerswitch. The volume of the running tones is adjustable
by the volume control (33) on the back of the unit. A small
screwdriver is needed for this adjustment. This does not
affect the pitch (i.e. frequency) of the tones.
5.35 Yellow indicator (9) is illuminated when cut power is
activated and blue indicator (22) is illuminated when coag
power is activated,
5.36 For a cutting current with increased haemostatic
effect, select blend touch button (13). Adjusting the setting
of cut control (14) will only change the power output, it will
not change the nature of the blended current. Blended
output is only obtained by pressing the yellow (cut)
fingerswitch button, or by pressing the yellow (cut)
footswitch.
WARNING
Apparent low power or failure of the unit to
function correctly at the normal operating
settings may indicate faulty application of the
plate electrode or poor contact in its
connections. In this case, the application of the
plate electrode and its connections should be
checked before selecting a higher output
power.
5.40 The output power level is adjusted by the power output
controls (14) and (21) for cut and coag modes respectively.
Turning the knob clockwise increases the power output
which is displayed digitally on displays (8) and (23)
respectively.
5.41 Digital power displays (6, 8 and 23) can show a
5.37 For maximum cutting speed, particularly in urology, second type of output setting display by pressing the digital
select the specialist cut touch button (15).
display range selection button (35) on the rear of the unit.
5.38 The two coag modes are selected by touch buttons The two available digital displays are as follows:
(19) pinpoint coag and (20) spray coag. The pinpoint mode
provides low voltage coagulation for most types of surgery.
The spray coag mode activates a separate generator circuit
which provides a waveform that reduces the cutting effect
to a very low level but provides fast coagulation by allowing
an arc to form between the electrode and the tissue. The
waveform is a series of high energy single pulses. Spray
coag allows non-contact coagulation of tissue that tends
to adhere to active electrodes.
Note: Spray coag (indicated with the bright green LED) is
not recommended for laparoscopic and endoscopic
procedures where its high voltage waveform can damage
the insulation of laparoscopic and endoscopic instruments,
P20/37
Downloaded from www.Manualslib.com manuals search engine
♦
When button 35 IS NOT pressed.
The typical power output available for a given setting
of the respective power output control is displayed.
The calibration is in watts for a typical resistance (see
Technical Data section for the maximum power
available for each output mode).
♦
When button 35 IS pressed.
The power output selected for a given setting of the
respective power output control is displayed as a
numerical value up to 10.
E-IM54f
TD830
ELECTROSURGICAL UNIT
5.48 Press micro bipolar selector touch button (3) to
select micro bipolar mode or press macro bipolar selector
5.42 Bipolar output is always available irrespective of what touch button (5) to select macro bipolar mode. The green
demands USER 1 and USER 2 are making from the indicators (2) or (4) will illuminate to show which bipolar
monopolar outputs.
mode has been selected.
Bipolar mode
CAUTION
Only fully insulated bipolar forceps should be
used to ensure that accidental contact between
the legs of the forceps and body tissue does
not cause electrosurgical action.
Selecting bipolar Forceps
5.43 Always use fully insulated forceps to ensure safety
for the operator and patient, and do not use direct switched
forceps, three core type.
Connecting Forceps
5.44 Select bipolar forceps (Fig. 9) to suit the surgical
procedure, and plug them into the bipolar forceps cable.
Connect the cable to bipolar output socket (31).
Changing Forceps
5.45 To change the type of forceps during use, e.g. from
bayonet to straight, set mains switch (42) to ‘O’ (‘off’)
change forceps, and then switch to ‘I’ (‘on’). Alternatively,
unplug the bipolar cable from the unit, change the forceps,
then plug the bipolar cable back into the unit.
Footswitches
5.49 Set power output control (1) to the required level.
Note that the digital power display (6) shows the typical
available output power for either power range. (see 5.41
for range control selection).
5.50 Activation of the connected footswitch will turn on
the power for as long as the footswitch is depressed.
Notes:
♦ Blue power activated indicator (7) will come ‘on’ and
the running tone will sound only when the output is
activated (see 5.34 for adjustment of volume).
♦
The 15W (micro) range should be used initially if the
power level required is not known.
♦
The method of ‘constant amplitude’ power control
provides very effective coagulation at low power
levels.
After use
5.51 After using the electrosurgical unit, switch ‘off’ the
mains switch (42) and disconnect it from the mains power
supply.
Disconnecting accessories
5.45 Two types of footswitch can be used in bipolar mode: 5.52 When disconnecting active cables or footswitches
from the electrosurgical unit, always hold the connector(s)
♦ A white pneumatic footswitch (REF 83-581-68, Fig.10) not the cable.
which is a light-action bellows type, with a push-on
Power output selection guide
connector tube.
A white electric footswitch (REF 83-581-76, Fig.10). 5.53 Various makes of electrosurgical units have different
maximum output powers for their modes, and, more
5.46 The type of footswitch used is a matter of personal importantly, they have different power output curves. These
choice. The pneumatic type has a very light action and is characteristics determine the actual power delivered to the
suited to procedures where repetitive actuation is needed. tissue for a given control setting. The TD830 electrosurgical
unit has a power curve that rises at the low resistance
Note:
end, this ensures that adequate power is available for
♦ Only one bipolar footswitch should be connected to
demanding surgical procedures. This power curve shape
the unit at any time.
also enables the full range of power control settings to be
used, low values for general or delicate surgery, or high
Power Output
values as shown in Table 1. All power curves are provided
CAUTION
at the end of this section.
♦
Set power output control (1) to ‘minimum’.
(see section 3)
5.47 The bipolar mode is extremely safe, but it is always
good practice to start at a low setting and increase slowly
until the required effect is obtained. Bipolar coagulation of
most blood vessels is normally self-terminating because
the rapid rise in tissue resistance, brought about by
coagulation, limits the flow of current. An audible ‘pop’
can result from the rapid boiling of liquids in the tissue.
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
5.54 The best results are obtained by varying the setting
rather than having a predetermined value for all uses. It
must be noted that digital power displays, only display the
typical available power for a given control setting. This is
proved by the fact that the displayed value does not change
between the active electrode being out of contact with the
tissue (zero power) or deeply embedded in well perfused
flesh (near maximum power delivered).
P21/37
5.55 The output control settings required for a particular
surgical procedure are also dependent on a number of
external factors (type of tissue, speed at which the surgeon
works, type of electrode in use, area of forcep tips in
contact). Therefore the recommended settings given in
Table 1 are for guidance only.
Alarms
5.57 If more than one output demand is received at the
same time (such as two footswitches or a footswitch and a
fingerswitch button being pressed) the unit will produce
an alarm indication and no power output. This is an
important safety feature and applies to any combination of
footswitch and fingerswitch for monopolar USER 1, or for
USER 2 or for bipolar.
Alarm identifications and remedies
Alarm Conditions
5.58 Alarm identifications are displayed on the front panel
5.56 The audiovisual alarm, two-tone with flashing (see Fig. 11 A1-A7) and described in section 3, remedies
are as follows:
displays at one second intervals, will operate if:
♦
♦
♦
The non-divided plate electrode is not connected via If the PCM or PAM circuit has been activated.
the plate cable to the unit, or the plate electrode and/
Remedy:
or cable is faulty (PCM).
(i) Check plate electrode is connected to unit.
A divided plate cable is plugged into the unit and the
(ii) Check plate electrode and cable for condition, if
divided plate electrode is not making a sufficiently
faulty, substitute new plate electrode and/or cable.
good contact with the patient’s body (PAM).
This procedure should normally be carried out before
plate electrode is attached to patient.
A low impedance path exists between the patient’s
body and earth (PEM). This condition can be caused
(iii) If a divided plate electrode is being used check that
by the presence of an earthed ECG electrode, or by
the divided plate electrode is making good contact
an accidental contact between the patient’s body and
with the patient’s body.
a conductive object such as part of the operation table
If the PEM circuit has been activated.
or a drip stand.
Remedy:
WARNING
The patient earth monitor (PEM) will not detect
a low impedence path between the patients
body and earth when the electrosurgical output
is activated.
(i)
(ii) If applicable disconnect suspect ECG electrodes.
Note: An activated PEM alarm will automatically
be reset by releasing a depressed footswitch or
fingerswitch.
♦
Excess power is detected (EPM) or dangerous
voltages are detected (PVM).
♦
More than one footswitch or fingerswitch is actuated
for monopolar USER 1, USER 2 or bipolar, e.g.
If the
♦
♦
fingerswitch and footswitch both actuated on the
monopolar USER 1 output.
♦
or, two fingerswitches on monopolar USER 2.
♦
or, pneumatic and electric footswitch on bipolar
output.
♦
or, any other forbidden combination of footswitch
and fingerswitch commands.
Internal unit errors are detected (!)
In any alarm mode, all outputs are inhibited and alarm
volume is maximum, regardless of volume control setting.
If the unit is in standby mode only the bipolar display will
flash.
P22/37
Downloaded from www.Manualslib.com manuals search engine
Check for accidental direct contact between patient
and an earthed object, (drip stand or conductive part
of operation table).
circuit has been activated
Remedy:
(i)
The alarm will operate if the bipolar pneumatic
footswitch and the bipolar electrical footswitch are
pressed simultaneously. The alarm will also sound
if both the cut and coag footswitches, or if one of
the two footswitches and a fingerswitch button are
pressed simultaneously for USER 1, or if both
fingerswitch buttons are pressed for USER 1 or
USER 2. The circuit will automatically reset when
ALL activations stop from ALL switches. This is an
important safety feature and applies to any
combination of footswitch and fingerswitch.
E-IM54f
TD830
ELECTROSURGICAL UNIT
If the EPM circuit has been activated.
Remedy:
(i)
Because the EPM alarm cannot be reset by the
User, if unit EPM alarm activates as soon as output
is activated, Eschmann Equipment should be
contacted.
If the PVM circuit has been activated.
Remedy:
(i)
If unit PVM alarm activates check for accidental
direct contact between the active electrode and an
earthed object.
(ii) Check that the plate electrode is correctly attached
to the patient’s body.
Note: An activated PVM alarm will automatically
be reset by releasing a depressed footswitch or
fingerswitch.
If none of the above remedies stop the alarm activated, or
the (!) alarm is activated contact Eschmann Equipment
and do not use the unit.
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
P23/37
Table 1 Power Output Selection Guide (for ‘typical’ power display range)
Procedure
Mode
Condition
Setting
CAUTION
Do not use a higher power setting than is necessary
General
Normal cut
Pinpoint coag
Blend
Dry field
Dry field
Dry field
70 to 90
25 to 45
45 to 70
Cardiothoracic
Normal cut
Pinpoint coag
Wet field
Wet/dry field
70 to 90
30 to 55
Sternal oozing
Spray coag
Wet/dry field
35 to 60
Intercostal muscle
Spray coag
Wet/dry field
35 to 65
Gynaecology LLETZ
(Large Loop Excision of the
Transformation Zone)
Blend
Spray coag
Blend
Spray coag
N-TZ (Displasia CIN I)
45 to 70
40 to 55
50 to 80
50 to 60
DEXC (Diathermy
Excision of the cervix)
Specialist cut
Spray coag
Specialist cut
L-TZ (CIN II to III)
Microinvasive
TZ Identification
50 to 60
45 to 60
40 to 60
TCRE: TCAE
(Transcervical Resection of
the Endometrium): (Ablation)
Blend
Specialist cut
Blend
Spray coag
Normal endometrium
Spongio-endometrium
Suspect fibroid inv.
Fundal ablation
55 to 70
40 to 60
65 to 80
50 to 65
Urology
Specialist cut
Spongio-prostate
Fibro/Muscular pros.
Bladder tumour
L-TZ (CIN I to II)
Normal cut
Spray coag
Pinpoint coag
Ball electrode
135 to 160
155 to 175
115 to 135
40 to 60
40 to 60
Plastic/burns
Spray coag
Eschar formation
40 to 50
Laparoscopic
cholecystectomy
Blend
Spray coag
Dry/wet field
Liver bed bleeding
50 to 70
55 to 65
Notes:
♦
Monopolar blend output mode can be used with advantage in many cutting procedures, particularly when only
a single (yellow) footswitch is being used.
♦
The blend mode produces less output power for a given control setting than normal cut, as shown on digital
power display (8 of illustration 11).
Bipolar micro
Bipolar macro
Monopolar cut
Monopolar blend
Monopolar Pinpoint coag
Monopolar spray coag
P24/37
Downloaded from www.Manualslib.com manuals search engine
Monopolar Specialist cut
E-IM54f
TD830
ELECTROSURGICAL UNIT
MONOPOLAR CUT DIAGRAMS
MONOP OLA R C UT (P OW E R v LOA D )
350
Power output (watts)
300
H a lf o u tp u t co n trol s e ttin g
Full ou tpu t con tro l s e ttin g
250
200
150
100
50
0
0
500
1000
1500
2000
Load resistance (ohms)
MONOPOLAR CUT (VOLTAGE v CONTROL)
MONOPOLAR CUT (POWER v CONTROL)
1200
250
Peak output voltage (volts)
Power output (watts)
300
Load resistance 200 (ohms)
200
150
100
50
0
Open circuit
1000
800
600
400
200
0
0
1
2
3
4
5
6
7
8
9
Output control setting (numerical value up to 10)
10
0
MONOPOLAR CUT (POWER v CONTROL)
2
3
4
5
6
7
8
9
Output control setting (numerical value up to 10)
10
MONOPOLAR CUT (VOLTAGE v CONTROL)
300
Peak output voltage (volts)
1200
250
Power output (watts)
1
Load resistance 200 (ohms)
200
150
100
50
Open circuit
1000
800
600
400
200
0
0
0
25
50
75 100 125 150 175 200
Output control setting (typical pow er)
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
225
250
0
25
50
75 100 125 150 175 200
Output control setting (typical pow er)
225
250
P25/37
MONOPOLAR BLEND DIAGRAMS
MONOP OLA R B LE ND (P OW ER v LOA D )
300
H alf ou tp ut co ntrol s etting
Fu ll o utpu t co n tro l s e ttin g
Power output (watts)
250
200
150
100
50
0
0
500
1000
1500
2000
Load resistance (ohms)
MONOPOLAR BLEND (POWER v CONTROL)
MONOPOLAR BLEND (VOLTAGE v CONTROL)
2500
Peak output voltage (volts)
Power output (watts)
200
Load resistance 400 (ohms)
160
120
80
40
0
0
1
2
3
4
5
6
7
8
9
2000
Open circuit
1500
1000
500
0
10
0
1
Output control setting (numerical value up to 10)
MONOPOLAR BLEND (POWER v CONTROL)
10
MONOPOLAR BLEND (VOLTAGE v CONTROL)
2500
Load resistance 400 (ohms)
160
120
80
40
0
Peak output voltage (volts)
200
Power output (watts)
2
3
4
5
6
7
8
9
Output control setting (numerical value up to 10)
2000
Open circuit
1500
1000
500
0
0
20
40
60
80
100 120 140 160 180 200
Output control setting (typical power)
P26/37
Downloaded from www.Manualslib.com manuals search engine
0
20
40 60 80 100 120 140 160 180 200
Output control setting (typical power)
E-IM54f
TD830
ELECTROSURGICAL UNIT
MONOPOLAR SPECIALIST CUT DIAGRAMS
MONOP OLA R S P E C IA LIS T C UT (P OW E R v LOA D )
350
Power output (watts)
300
H a lf ou tp u t co ntro l s e tting
Fu ll o utpu t con tro l s e ttin g
250
200
150
100
50
0
0
500
1000
1500
2000
Load resistance (ohms)
MONOPOLAR SPECIALIST CUT (VOLTAGE v CONTROL)
MONOPOLAR SPECIALIST CUT (POWER v CONTROL)
1500
350
Load resistance 200 (ohms)
Peak output voltage (volts)
Power output (watts)
300
250
200
150
100
50
0
0
1
2
3
4
5
6
7
8
9
Open circuit
1200
900
600
300
0
10
0
Output control setting (numerical value up to 10)
2
3
4
5
6
7
8
9
Output control setting (numerical value up to 10)
10
MONOPOLAR SPECIALIST CUT (VOLTAGE v CONTROL)
MONOPOLAR SPECIALIST CUT (POWER v CONTROL)
1500
350
300
Load resistance 200 (ohms)
Peak output voltage (volts)
Power output (watts)
1
250
200
150
100
50
0
Open circuit
1200
900
600
300
0
0
30
60
90
120
150
180
210
240
Output control setting (typical pow er)
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
270
300
0
30
60
90 120 150 180 210 240
Output control setting (typical pow er)
270
300
P27/37
MONOPOLAR PINPOINT COAG DIAGRAMS
MONOP OLA R P INP OINT C OA G (P OW E R v LOA D )
Power output (watts)
200
160
H alf ou tpu t con tro l s e ttin g
Fu ll o u tp ut co n tro l s etting
120
80
40
0
0
500
1000
1500
2000
Load resistance (ohms)
MONOPOLAR PINPOINT COAG (VOLTAGE v CONTROL)
MONOPOLAR PINPOINT COAG (POWER v CONTROL)
2500
Power output (watts)
100
Peak output voltage (volts)
120
Load resistance 400 (ohms)
80
60
40
20
0
Open circuit
2000
1500
1000
500
0
0
1
2
3
4
5
6
7
8
9
10
0
1
MONOPOLAR PINPOINT COAG (POWER v CONTROL)
3
4
5
6
7
8
9
10
MONOPOLAR PINPOINT COAG (VOLTAGE v CONTROL)
2500
100
Peak output voltage (volts)
120
Power output (watts)
2
Output control setting (numerical value up to 10)
Output control setting (numerical value up to 10)
Load resistance 400 (ohms)
80
60
40
20
0
Open circuit
2000
1500
1000
500
0
0
10
20
30
40
50
60
70
80
Output control setting (typical pow er)
P28/37
Downloaded from www.Manualslib.com manuals search engine
90
100
0
10
20
30
40
50
60
70
80
90
100
Output control setting (typical pow er)
E-IM54f
TD830
ELECTROSURGICAL UNIT
MONOPOLAR SPRAY COAG DIAGRAMS
M O NO P O LA R S P RA Y C OA G (P OW E R v L O A D )
80
Power output (watts)
70
60
H a lf o u tp u t co n tro l s e ttin g
Fu ll o u tp u t co n tro l s e ttin g
50
40
30
20
10
0
0
500
1000
1500
2000
L o a d re sista n ce (o h ms)
MONOPOLAR SPRAY COAG (POWER v CONTROL)
MONOPOLAR SPRAY COAG (VOLTAGE v CONTROL)
80
4000
Peak output voltage (volts)
Power output (watts)
Load resistance 400 (ohms)
60
40
20
0
3500
Open circuit
3000
2500
2000
1500
1000
500
0
0
1
2
3
4
5
6
7
8
9
Output control setting (numerical value up to 10)
10
0
2
3
4
5
6
7
8
9
Output control setting (numerical value up to 10)
10
MONOPOLAR SPRAY COAG (VOLTAGE v CONTROL)
MONOPOLAR SPRAY COAG (POWER v CONTROL)
80
4000
Peak output voltage (volts)
Load resistance 400 (ohms)
Power output (watts)
1
60
40
20
3500
Open circuit
3000
2500
2000
1500
1000
500
0
0
0
7.5
15
22.5 30 37.5 45 52.5 60
Output control setting (typical pow er)
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
67.5
75
0
7.5
15 22.5 30 37.5 45 52.5 60
Output control setting (typical pow er)
67.5
75
P29/37
BIPOLAR MACRO DIAGRAMS
BIPOLAR MAC RO (POWER v LOAD)
50
Power output (watts)
40
Macro - H alf output control s etting
Macro - Full output control s etting
30
20
10
0
0
200
400
600
Load resistance (ohm s)
BIPOLAR MACRO (POWER v CONTROL)
250
40
Peak output voltage (volts)
Power output (watts)
1000
BIPOLAR MACRO (VOLTAGE v CONTROL)
50
Load resistance 100 (ohms)
30
20
10
0
200
Open circuit
150
100
50
0
0
1
2
3
4
5
6
7
8
9
10
0
1
Output control setting (numerical value up to 10)
2
3
4
5
6
7
8
9
10
Output control setting (numerical value up to 10)
BIPOLAR MACRO (POWER v CONTROL)
BIPOLAR MACRO (VOLTAGE v CONTROL)
50
250
Peak output voltage (volts)
Power output (watts)
800
Load resistance 100 (ohms)
40
30
20
10
0
200
Open circuit
150
100
50
0
0
5
10
15
20
25
30
35
Output control setting (typical pow er)
P30/37
Downloaded from www.Manualslib.com manuals search engine
40
45
50
0
5
10
15
20
25
30
35
40
45
50
Output control setting (typical pow er)
E-IM54f
TD830
ELECTROSURGICAL UNIT
BIPOLAR MICRO DIAGRAMS
B IPOLA R MIC RO (P OW ER v LOAD )
20
Power output (watts)
16
Micro - H alf output control s etting
Micro - Full output control s etting
12
8
4
0
0
200
400
600
Load resistance (ohm s)
16
160
Peak output voltage (volts)
Power output (watts)
1000
BIPOLAR MICRO (VOLTAGE v CONTROL)
BIPOLAR MICRO (POWER v CONTROL)
Load resistance 100 (ohms)
12
8
4
0
120
Open circuit
80
40
0
0
1
2
3
4
5
6
7
8
9
10
0
1
Output control setting (numerical value up to 10)
2
3
4
5
6
7
8
9
10
Output control setting (numerical value up to 10)
BIPOLAR MICRO (POWER v CONTROL)
BIPOLAR MICRO (VOLTAGE v CONTROL)
16
160
Peak output voltage (volts)
Power output (watts)
800
Load resistance 100 (ohms)
12
8
4
0
120
Open circuit
80
40
0
0
1.5
3
4.5
6
7.5
9
10.5
12
Output control setting (typical pow er)
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
13.5
15
0
1.5
3
4.5
6
7.5
9
10.5
12
13.5
15
Output control setting (typical pow er)
P31/37
6.0 CLEANING DISINFECTION
AND CARE
CLEANING AND DISINFECTION
WARNING
Non-flammable agents should be used
for cleaning and disinfection wherever possible.
If flammable agents are used for cleaning or
disinfection they should be allowed to evaporate
before reusing the electrosurgical unit.
CAUTION
6.5 If the unit is to be stored for any length of time it
should be plugged into the mains and switched on to charge
the internal battery. A charge period of 48 hours is required
to restore the battery to its full life of six months (or pro
rata, e.g. 8 hours for one month). Also note the required
storage conditions detailed in the Technical Data section.
Service
Equipment Failure
6.6 If the unit or accessories fail any of the checks or
operational requirements included in this manual, refer to
the Eschmann After Sales Service Department and do not
use the equipment.
Fuse Failure
The electrosurgical unit MUST be disconnected
from the mains power supply prior to cleaning.
6.7 Two fuses are located in holders on the rear panel
(illustration 11). To gain access to a fuse, firstly ensure
that the electrosurgical unit is disconnected from the mains
Cleaning the electrosurgical unit
power supply and then, using a screwdriver, release the
6.1 The cabinet should be cleaned with a sponge or fuse cap and withdraw the fuse. If necessary, replace the
cloth dampened in a mild detergent or soap solution and fuse with a fuse of the correct type and rating (see Technical
Data). If repeated blowing of a fuse(s) occurs, either when
wiped dry.
the unit is switched on or during use, call the Eschmann
♦ Do not use excessive amounts of liquid which could After Sales Service Department (see inside front cover).
enter the cabinet and damage internal components.
6.8 A failed fuse is usually an indication of a fault that
♦ The cabinet front panel can be wiped, as it is a has occurred somewhere in the unit. Because of this
continuous membrane. Pressure on the touch buttons, Eschmann After Sales Service Department should always
with the power supply off, will not cause mode changes. be contacted as soon as a failed fuse is identified or
suspected.
Disinfecting the electrosurgical unit
6.2 Initially clean the unit as above. For disinfection,
Eschmann Equipment recommend primarily the use of
alcohol based disinfectants, e.g. 70% isopropanol/water
or 70% methyl alcohol/water. The cabinet should be wiped
with a sponge or cloth dampened in the above solution
and allowed to dry.
♦
Additional Information
6.9 Eschmann Equipment will make available on
request circuit diagrams, component parts lists, description
and calibration instructions which will assist the users’
appropriately qualified technical personnel to repair those
parts of the equipment which are considered ‘user
repairable’ by Eschmann Equipment.
Do not use excessive amounts of liquid which could
enter the cabinet and damage internal components, Maintenance Procedures
also ensure the unit is completely dry before reusing 6.10 For complete maintenance procedures see
it.
Eschmann Service Manual E-SM44.
Electrosurgical Accessories
ACCESSORIES
6.3 All Eschmann Electrosurgical Accessories (see
section 7) come with full “Cleaning, Disinfection and Care 6.11 For a complete range of accessories, and relevant
Instructions”, these should be followed at all times. If you ordering information refer to Eschmann product information
have misplaced the accessory instructions and require leaflets.
another copy please quote Publication Number E-IM50
(Part No. 604802) for all Eschmann accessories.
CARE
6.4 Ensure that electrosurgical safety checks and
routine maintenance are carried out at regular intervals
(at least every six months) and only by Eschmann trained
personnel otherwise the warranty for the equipment could
be infringed.
P32/37
Downloaded from www.Manualslib.com manuals search engine
E-IM54f
TD830
ELECTROSURGICAL UNIT
7.0 ACCESSORIES
7.1 The accessories in the list below and illustrated
where referenced, are all recommended for use with the
TD830 electrosurgical unit. (REF = catalogue number)
REF
Description
Illustration No. REF
Active electrode silicone cables
83-135-12 3m cable with 4mm plug
83-137-17 5m cable with 4mm plug
83-138-06 3m cable with 8mm plug
83-138-22 5m cable with 8mm plug
7
7
7
7
Active handle for 3/32 inch electrodes
83-140-20 for above cables
3
Active handle with silicone cable
83-141-01 3m cable with 4mm plug
83-141-36 5m cable with 4mm plug
83-141-28 3m cable with 8mm plug
83-141-44 5m cable with 8mm plug
5
5
5
5
Fingerswitches with blade electrode
83-582-06 3m cable and 3 pin plug
83-583-97 5m cable and 3 pin plug
83-585-94 3m cable and 3 pin plug (disposable)
Bipolar cable two pin plug
83-041-55 3m cable
Bipolar forceps
83-700-01 straight, 200mm
83-703-03 straight, 150mm
83-705-08 straight, 110mm fine tip
83-707-02 straight, 80mm extra fine tip
83-701-09 angled, 200mm
83-704-00 angled, 150mm
83-706-05 angled, 110mm fine tip
83-708-18 angled, 80mm extra fine tip
83-702-06 bayonet, 200mm
83-709-07 bayonet, 250mm fine tip
8
8
8
8
8
8
8
8
8
8
8
Chiropody/Podiatry & Dermatology electrodes
83-583-46 loop, 6.5mm
6 83-584-58 extended small loop, 6mm
6 83-584-61 loop, 10mm
6 83-583-95 angled ball, 2mm
*
*
*
*
Monopolar footswitches and cable
* 83-581-09 yellow (cut/blend), 3m cable
83-581-13 yellow (cut/blend), 5m cable
83-581-17 blue (coag), 3m cable
9
83-581-25 blue (coag), 5m cable
9
83-109-39 twin (cut/blend & coag), 3m cable
9
83-109-47 twin (cut/blend & coag), 5m cable
9
83-109-71 3m extension cable
9
9 Bipolar footswitches
9 83-581-76 electric, 3m cable
9 83-581-68 pneumatic, 5m cable
9
Patient plate cable, 6mm jack plug
9
83-122-93 3m cable, for standard & divided plate
8
8
8
8
8
8
8
Electrodes, needle
83-582-22 medium (pack of 5)
83-583-11
extended, 100mm
83-585-23 standard (disposable)
83-588-77 angled medium (pack of 5)
83-588-80 extended angled, 100mm
8
8
8
8
8
Electrodes, ball
83-583-62 8mm
83-585-45 4mm (disposable)
83-588-85 4.2mm (pack of 5)
8
8
8
Downloaded from www.Manualslib.com manuals search engine
Illustration No.
Colposcopy electrodes (LLETZ) cone biopsy
83-583-51 large loop, 12mm (pack of 5)
83-584-19 large loop, 15mm (pack of 5)
83-584-27 large loop, 20mm (pack of 5)
83-584-35 large loop, 25mm (pack of 5)
83-584-51 large loop, 1 of each 12,15,20,25mm
83-584-64 small loop, 15mm (pack of 5)
83-590-08 Cartier loop, 6 x 7mm (pack of 5)
83-590-16 Cartier loop, 10 x 12mm (pack of 5)
83-583-89 extended ball, 8mm (100mm)
83-583-90 extended ball, 4.2mm (100mm)
83-585-56 extended 4mm (disposable)
Electrodes, blade
83-582-14 fine (pack of 5)
83-582-27 extended, 100mm
83-582-30 extended, 150mm
83-582-49 extended, 200mm
83-585-04 70mm blade (disposable)
83-585-11
150mm blade (disposable)
83-588-69 angled (pack of 5)
E-IM54f
Description
4
4
4
4
4
4
4
10
10
2
Patient plates and adaptor
83-122-42 adhesive standard (box of 200)
83-122-46 adhesive divided (box of 200)
83-122-49 Flexoplate 200-S, 2.5m cable
83-207-56 for Flexoplate, 6mm jack plug
1
1
1
1
Accessory sets
83-117-22
basic
83-117-30
major
*
*
Other accessories
83-186-38 autoclavable quiver
83-139-41 S-Lead (9mm), Olympus
83-139-49 S-Lead (8mm), Pentax
*
*
* = not illustrated
P33/37
83-122-42
83-122-46
83-207-56
83-122-49
1
83-140-20
83-122-93
2
3
83-109-39
83-109-47
4mm
83-581-09
83-581-13
8mm
83-141-01
83-141-36
83-141-28
83-141-44
83-581-17
83-581-25
83-109-71
4
P34/37
Downloaded from www.Manualslib.com manuals search engine
5
E-IM54f
TD830
ELECTROSURGICAL UNIT
83-582-06
83-583-97
83-585-94 (Disposable)
6
4mm
7
8mm
83-582-14
83-582-27
83-582-30
83-582-49
83-588-69
83-585-04 (Disposable)
83-585-11 (Disposable)
83-582-22
83-583-11
83-585-23 (Disposable)
83-588-77
83-588-80
8
83-135-12
83-137-17
83-138-06
83-138-22
83-583-62
83-585-45 (Disposable)
83-588-85
83-583-89
83-583-90
83-585-56 (Disposable)
83-590-08
83-590-16
83-583-51
83-584-19
83-584-27
83-584-35
83-584-51
83-584-64
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
P35/37
83-700-01
83-703-03
83-705-08
83-707-02
83-701-09
83-704-00
83-706-05
83-708-18
83-702-06
83-709-07
9
83-581-76
83-581-68
01
P36/37
Downloaded from www.Manualslib.com manuals search engine
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
Eschmann Equipment, Peter Road, Lancing, West Sussex, BN15 8TJ, England.
Tel: +44 (0) 1903 753322. Fax: +44 (0) 1903 875789. www.eschmann.co.uk
Downloaded from www.Manualslib.com manuals search engine
1
Power output control, BIPOLAR
2
MICRO BIPOLAR selected, indicator (green LED)
3
Touch button, MICRO BIPOLAR
4
MACRO BIPOLAR selected, indicator (green LED)
5
Touch button, MACRO BIPOLAR
6
BIPOLAR digital power display (green seven segment LEDs)
7
BIPOLAR, power activated indicator (blue LED)
8
MONOPOLAR CUT digital power display (green seven segment LEDs)
9
MONOPOLAR CUT, power activated indicator (yellow LED)
10 MONOPOLAR (normal cut) selected, indicator (green LED)
11 Touch button MONOPOLAR (normal cut)
12 MONOPOLAR (blend) selected, indicator (green LED)
13 Touch button MONOPOLAR (blend)
14 Power output control, MONOPOLAR (cut)
15 Touch button MONOPOLAR (specialist cut)
16 MONOPOLAR (specialist cut) selected, indicator (green LED)
17 MONOPOLAR COAGULATION (pinpoint) selected, indicator (green LED)
18 MONOPOLAR COAGULATION (spray) selected, indicator (bright green LED)
19 Touch button MONOPOLAR COAGULATION (pinpoint)
20 Touch button MONOPOLAR COAGULATION (spray)
21 Power output control MONOPOLAR COAGULATION
22 MONOPOLAR COAGULATION, power activated indicator (blue LED)
23 MONOPOLAR COAGULATION digital power display (green seven segment LEDs)
24 8mm active electrode socket USER 1 only (footswitch activation only) - red
25 Fingerswitch sockets (for use with active electrode socket 26) - black
26 4mm active electrode socket USER 1 (footswitch or fingerswitch activation) - red
27 Fingerswitch sockets (for use with active electrode socket 28) - black
28 4mm active electrode socket USER 2 (fingerswitch activation only ) - red
29 ‘S’ (scope) connector socket (for use with flexible endoscopes) see section 4.55
30 Plate electrode connector socket - black
31 Bipolar output socket - white
32 Monopolar standby selection button (monopolar ‘on/off’ toggle button)
33 Running tone volume control
34 Potential equalization point (see Technical Data section)
35 Digital display range selection button (display in watts if not held pressed)
36 Bipolar pneumatic (white) footswitch socket
37 Bipolar electric (white) footswitch socket
38 Sockets for monopolar footswitches (blue, yellow or combination blue and yellow)
39 Serial number plate
40 Mains cable connection socket
41 Mains fuses (see technical data)
42 Mains power on/off control (with internal green lamp)
A1 - PCM alarm activated LED
A2 - PAM alarm activated LED
A3 - PEM alarm activated LED
A4 - PVM alarm activated LED
A5 - EPM alarm activated LED
A6 - MULTIPLE ACTIVATION alarm LED
A7 - Miscellaneous alarm activated LED
All alarms are accompanied by an audible two tone warning and the activated digital displays will flash (i.e.
if in monopolar standby mode, only the bipolar digital display will flash).
Key to illustration 12
E-IM54f
Downloaded from www.Manualslib.com manuals search engine
12
P37/37
Download