TD830 E LECT RO SURGICAL UNIT Instructions for use 698257 Downloaded from www.Manualslib.com manuals search engine E-IM54f Read these Instructions before use Preliminary Information Technical data Safety notes and alarms Operation Cleaning, disinfection and care Keep these ‘Instructions for use’ in a safe convenient place for future reference. Read in conjunction with the relevant Publications detailed in the preliminary information section. Eschmann After Sales Service Department The Eschmann After Sales Service Department is staffed and equipped to provide advice and assistance during normal office hours. To avoid delays when making enquiries, please quote the Model and Serial Number of your Electrosurgical Unit which is shown on the Serial Number plate, the location of which is shown below. Please ensure you include all alpha and numeric digits of the Serial Number. Instructions for use Accessories The Serial Number plate is located here, (view from rear of unit). For further information visit www.eschmann.co.uk All correspondence relating to the after sales service of Eschmann Equipment to be addressed to : UK Customers Eschmann Equipment, Peter Road, Lancing, West Sussex BN15 8TJ, England. Tel: +44 (0) 1903 765040. Fax: +44 (0) 1903 875711. Overseas Customers Contact your local distributor. In case of doubt contact Eschmann Equipment. Patents and Trade marks The ESCHMANN name and logo are registered trade marks of Eschmann Holdings Limited. “Eschmann Equipment” is a trading name of Eschmann Holdings Limited. “TD830” is a trade mark of Eschmann Holdings Limited. Patents : GB2276551, GB2146534, AU673883, ZA94/2173, US5480399, EP617925, IEE69100 and other Patents Pending. Copyright © 2006 Eschmann Holdings Limited All rights reserved. This booklet is protected by copyright. No part of it may be reproduced, stored in a retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying, recording or otherwise without written permission from Eschmann Holdings Limited. The information in this publication was correct at the time of going to print. The Company, however, reserves the right to modify or improve the equipment referred to. The CE marking affixed to the product certifies that it complies with the European Medical Devices Directive 93/42/EEC and related legislation. E-IM54f July 2006 Downloaded from www.Manualslib.com manuals search engine TD830 ELECTROSURGICAL UNIT CONTENTS 1. PRELIMINARY INFORMATION 5. OPERATION Preliminary information . . . . . . . . . . . . . . . . . . . . . . . . 4 Routine check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Connection of accessories . . . . . . . . . . . . . . . . . . . . Monopolar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2. TECHNICAL DATA Bipolar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 System check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . . Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Monopolar standby mode . . . . . . . . . . . . . . . . . . . Audible indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Monopolar mode . . . . . . . . . . . . . . . . . . . . . . . . . . Visual indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 USER 1 and/or USER 2 selection . . . . . . . . . . Safety.. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Plate electrode . . . . . . . . . . . . . . . . . . . . . . . . Duty cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Fingerswitch/active electrode handle . . . . . . . . Other symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Footswitches . . . . . . . . . . . . . . . . . . . . . . . . . . Button symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Monopolar outputs . . . . . . . . . . . . . . . . . . . . . . Alarm symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Power output . . . . . . . . . . . . . . . . . . . . . . . . . . Environmental conditions for transport and storage . . . 8 Bipolar Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . Selecting bipolar forceps . . . . . . . . . . . . . . . . . 3. SAFETY NOTES & ALARMS Connecting forceps . . . . . . . . . . . . . . . . . . . . . Changing forceps . . . . . . . . . . . . . . . . . . . . . . Do’s and Don’ts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Footswitches . . . . . . . . . . . . . . . . . . . . . . . . . . Alarm circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Power output . . . . . . . . . . . . . . . . . . . . . . . . . . After use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4. INTRODUCTION Disconnecting accessories . . . . . . . . . . . . . . . . . . Power output selection guide . . . . . . . . . . . . . . . . General .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 .................................. Operating modes / display option.. . . . . . . . . . . . . . . 11 Alarms Alarm conditions . . . . . . . . . . . . . . . . . . . . . . . . . . Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Alarm identifications and remedies . . . . . . . . . . . . Associated Publications . . . . . . . . . . . . . . . . . . . . . . . 11 TABLE 1 - Power output selection guide . . . . . . . . . . Equipment certification . . . . . . . . . . . . . . . . . . . . . . . 11 Electrosurgical techniques . . . . . . . . . . . . . . . . . . . . . 12 Output diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Monopolar electrosurgery . . . . . . . . . . . . . . . . . . . 12 6. CLEANING, DISINFECTION & CARE Plate electrode . . . . . . . . . . . . . . . . . . . . . . . . 12 Cleaning and disinfection . . . . . . . . . . . . . . . . . . . . . . Footswitches . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Cleaning the electrosurgical unit . . . . . . . . . . . . . . Active electrode handles . . . . . . . . . . . . . . . . . 12 Disinfecting the electrosurgical units . . . . . . . . . . . Active cables . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Electrosurgical accessories . . . . . . . . . . . . . . . . . Active electrodes . . . . . . . . . . . . . . . . . . . . . . . 13 Care .................................. Fingerswitches . . . . . . . . . . . . . . . . . . . . . . . . 13 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Power control . . . . . . . . . . . . . . . . . . . . . . . . . 13 Additional information . . . . . . . . . . . . . . . . . . . . . . Current paths . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Maintenance procedures . . . . . . . . . . . . . . . . . . . Accidental contact . . . . . . . . . . . . . . . . . . . . . . 13 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Bipolar electrosurgery . . . . . . . . . . . . . . . . . . . . . . 13 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 ACCESSORIES Flammable anaesthetics and spirits . . . . . . . . . . . 14 7. Implanted cardiac pacemakers . . . . . . . . . . . . . . . 14 Accessory list . . . . . . . . . . . . . . . . . . . . . . . . . . . . General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Risks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 ILLUSTRATIONS Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 1-5 ................................. Electrosurgical interference . . . . . . . . . . . . . . . . . 14 6-8 ................................. Using two electrosurgical units . . . . . . . . . . . . . . . 15 9 - 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using a flexible endoscope . . . . . . . . . . . . . . . . . . 15 12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 E-IM54f Downloaded from www.Manualslib.com manuals search engine 16 16 16 17 17 18 18 18 18 19 19 19 20 20 21 21 21 21 21 21 21 21 21 22 22 22 24 25 32 32 32 32 32 32 32 32 32 33 34 35 36 37 P3/37 1.7 Ensure that the unit ‘mains’ switch (42 of Fig. 11) is in the ‘O’ position and that the output controls are set to minimum before connecting to, and switching ‘on’, the 1.1 These Instructions for Use should be referred to for mains supply. A complete systems check must be carried details of the TD830 Electrosurgical Unit, REF 83-256-01, out before using the Electrosurgery Unit (see Sections 5.10 83-257-02, 83-258-03, 83-259-04, 83-260-05, 83-261-01, to 5.17). 83-262-07, 83-263-08, and 83-264-09 (serial number A9B0000 or above). 1.8 Instruction and Service Manuals should be readily accessible for reference prior to and when operating, 1.2 Within the text of this manual the term ‘coag’ is used cleaning and servicing the TD830 Electrosurgical Unit. All as a common abbreviation of the term ‘coagulation’. manuals are available from Eschmann Equipment, see 1.3 Carefully remove the TD830 Electrosurgical Unit and inside front cover for address details. its associated accessories from the packing case(s). 1. PRELIMINARY INFORMATION 1.4 Place the TD830 Electrosurgical Unit on a stable Related Technical Publications convenient working platform, for example on the Eschmann ST80 mobile trolley, purpose built for carrying Service Manual for TD830 electrosurgical units. Publication number E-SM44, Part No. 698260 1.5 The TD830 Electrosurgical Unit requires a mains electrical supply corresponding to the voltage shown on Eschmann accessory ‘Instructions for use’ the electrical rating plate at the rear of the unit. Only use Publication number E-IM50, Part No.604802 the mains supply cable supplied. If the plug supplied prefitted is not suitable it should be replaced with a suitable plug with protective earthing contact. 1.6 If the plug is a fused type, a 10A fuse must be fitted. CAUTION It is most important that fuses of the correct type, size and rating are installed (see Technical Data). CAUTION Read these ‘Instructions for use’ carefully before using this electrosurgical unit and note ALL of the warnings, cautions and safety notes contained within. Keep these ‘Instructions for use’ close-to-hand at all times for reference. Please note that Fig. 11 at the end of this manual opens out so that it can be refered to easily whilst reading this manual. It is the main figure identifying the various parts of the TD830 Electosurgical Unit. P4/37 Downloaded from www.Manualslib.com manuals search engine E-IM54f TD830 ELECTROSURGICAL UNIT 2. 0 TECHNICAL DATA Output (bipolar) GENERAL The TD830 Electrosurgical Unit (or Surgical Diathermy Symbol/Function Unit) is classified as ‘HF surgical equipment’ 1 1 - ‘HF surgical equipment’ is defined as, “Medical electrical equipment including its associated accessories intended for the performance of surgical operations, such as the ‘cutting’ 2 or ‘coagulation’ 3 of biological tissue by means of high frequency (h.f.) currents”. Power Crest factor Peak Open Circuit Voltage Coagulation Micro range 17W ±20% Variable Macro range 50W +10% Variable 150 230 -20% - ‘Cutting’ is defined as, “Resection or dissection of body tissue caused by the passage of high frequency current of high Power and voltage output data diagrams are shown at the current density at the active electrode(s)”. 2 end of section 5. 3 - ‘Coagulation’ is defined as, “Sealing of small blood vessels or of body tissue caused by the passage of high frequency current at the active electrode(s)”. Monopolar Carrier frequency - 475kHz nominal, square wave. Equipment - High power electrosurgical unit with monopolar and bipolar outputs Power control Variable amplitude set by frontp a n e l controls. Preset pulse patterns set by mode switches. Type - Portable Load resistance for maximum output power is 150 ohms DIMENSIONS (non-inductive) for cut and pinpoint coag and 200 ohms (non-inductive) for blend, specialist cut and spray coag. Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39.0 cm Height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19.0 cm Test load is 200 ohms (non-inductive) for cut and specialist Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.5 cm cut and 400 ohms (non-inductive) for blend, spray coag Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 kg and pinpoint coag. ELECTRICAL DATA Output (monopolar) (Note: Voltage factory set by transformer tapping, according Output powers are measured to an accuracy of ±20%, with a maximum power of 400 watts. to model supplied.) Power Supply . . . . . . . . . . . . . . . . . 230V a.c., 50-60Hz or, 240V a.c., 50-60Hz Peak or, 220V a.c., 50-60Hz Open or, 110V a.c., 50-60Hz Crest Circuit Current (max.) . . . . . . . . . 4.4A (230V) or, 4.2A (240V) Symbol/Function Power factor Voltage or, 4.6A (220V) or, 8.4A (110V) Fuse rating (240V, 230V, 220V) . . . . . . . . . . . 250V, T5A Normal cut 345W(-20%) 1.9 1150 Fuse rating (110V) . . . . . . . . . . . . . . . 125V(min.),T10A Fuse type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 mm Blend 300W(±20%) 3.0 2200 Bipolar Carrier frequency - 785kHz nominal, square wave. Specialist cut 345W(-20%) 2.1 1500 Power control Variable pulse group modu-lation set by front panel control. Amplitude set by micro/macro range buttons. Pinpoint coag. 170W(±20%) 5.1 2150 Spray coag. 79W(±20%) 8.7 4000 Load resistance for maximum output power - 100 ohms (non-inductive) for Macro - 50 ohms (non-inductive) for Micro Power and voltage output data diagrams are shown at the end of section 5. Test load - 100 ohms (non-inductive) E-IM54f Downloaded from www.Manualslib.com manuals search engine P5/37 AUDIBLE INDICATORS SAFETY Running tones * General Monopolar Designed to comply with EN60601-1:1990 Medical electrical equipment - general requirements for safety and IEC60601-2-2: 1998 High-frequency surgical equipment particular requirements for safety (3rd edition). Cut, blend or specialist cut Coagulation, pinpoint or spray 950Hz 800Hz Bipolar Micro or macro 730Hz (* Approximate values, adjustable volume) General classification, Class 1, Type CF (Defibrillator proof). Drip-proof (IPX 1) Patient leakage (risk) current: always less than 100 microamps to earth (ground) from all patient circuits at 230V Touch button 50Hz as required by EN60601-1:1990 for the unit in normal A ‘bleep’ indicates when any function button is pressed, condition (typically less than 10 microamps). the volume is adjustable with running tones above. Battery Alarm This equipment contains a nickel metal hydride battery. In All modes - alternating two tone preset to maximum volume the event of failure of the display p.c.b. battery, or if the electrosurgical unit is to be disposed of, it is not necessary (with flashing display). to remove the battery or to return it to Eschmann Equipment. The battery charge life is six months from a full charge of 48 hours. The battery charges automatically VISUAL INDICATORS when the unit is ‘on’. Green lamp in mains ‘on/off’ switch to indicate power ‘on’ Electrode isolation from the rear of the electrosurgical unit. The plate electrode circuit of this equipment is isolated Digital displays and ‘function selected’ LEDs indicate power from earth at both high and low frequency. The bipolar ‘on’ from the front of the electrosurgical unit output is also fully isolated at both high and low frequency. Three digital displays indicate typical power* in cut, coag and bipolar modes. (Note: These can be set to display the Class typical power* available in watts, or a numerical value up Class 1 denotes that the equipment must be earthed via to 10). the protective conductor in the 3-core mains cable (* typical power is an indication of the average power connected to a 3-pin plug. delivered over a range of load resistances. It is less Safety category than the power delivered to the rated load for maximum power.) This symbol denotes that the equipment is of type CF, i.e. that it complies with type CF leakage current Cutting modes Yellow lamp requirements. The symbol also denotes that the equipment Coag modes Blue lamp will not be damaged by defibrillator discharge and that the Bipolar mode Blue lamp plate electrode need not be removed from the patient if a Green and bright green function selected LED indicators defibrillator is used. above (or below) the relevant touch buttons detailed as Non-ionizing radiation follows: Power output ‘activated’ LED indicators for: Monopolar: Cut, blend, specialist cut (green) Pinpoint coag (green) Spray coag (bright green) Bipolar: Micro and macro power range (green) This symbol warns the user of the possibility of non-ionizing radiation being emitted by this equipment. Flammable gases The TD830 equipment is not suitable for use in the presence of a flammable anaesthetic mixture with air or with Oxygen or Nitrous Oxide. Protection IPX 1 This symbol (drip proof) denotes that the equipment meets the requirements of IEC529 for dripping water. P6/37 Downloaded from www.Manualslib.com manuals search engine E-IM54f TD830 ELECTROSURGICAL UNIT DUTY CYCLE The duty cycle rating of 10 seconds ‘on’, 30 seconds ‘off’, specified on the serial plate as 10s:40s, indicates that the equipment can be used at full output power in any mode for 10 seconds with a 30 seconds rest period. The unit can remain connected to the mains electrical supply with the mains switch in the ‘on’ position continuously. The ‘on’ period at lower power levels can be extended. I The symbols and adjacent to the mains ‘on/off’ switch (42) indicate the ON and OFF positions respectively. The symbol adjacent to item (34) indicates the ‘equipotentiality’ connection point. (Means for connection of a potential equalization conductor). The symbol adjacent to button (35) signifies the button in the non-pressed (i.e. normal) state. (Digital display on Note: For numbers in brackets refer to Illustration 11 which the front of the electrosurgical unit is a typical power figure, e.g. 0-200watts for monopolar blend). folds out at the end of this manual. OTHER SYMBOLS This symbol above item (33) indicates the running The symbol adjacent to button (35) signifies the button tone volume control which has a minimum sound level of in the pressed (i.e. activated) state. (Digital display on the 45dBA. front of the electrosurgical unit shows a numerical value up to 10). This symbol above items (1, 14, 21 and 33) indicates increasing power output (1, 14 and 21) and adjacent to sockets (36, 37 and 38) increasing volume (33) for running tone and ‘bleep’ volume. The symbol indicates the connection point for footswitches. This symbol on the rear panel serial plate (39) and above the mains inlet (40) indicates that the equipment is The symbol SN indicates serial number. for use on alternating current only. The symbol REF indicates catalogue number. This symbol on the rear panel serial plate indicates that the mains input fuses, rating and type, are The symbol indicates date of manufacture. as shown below the symbol. BUTTON SYMBOLS The symbol on the front panel denotes that the plate electrode is isolated from earth at high frequency. (Note: The plate electrode is also known as the dispersive, neutral, return, indifferent or patient plate electrode and is often simply called the ‘pad’). Press to select bipolar - micro range Press to select bipolar - macro range The symbol on the front panel adjacent to the active outputs (24 to 28) denotes dangerous voltages. The symbol on the rear panel serial plate warns the user to read the accompanying documents, the ‘Instructions for use’. Press to select monopolar - cut Press to select monopolar - blend The symbol above sockets (25, 26, 27 and 28) denotes connection socket for a two button electrode handle (fingerswitch type). Press to select monopolar - specialist cut The symbol adjacent to sockets (24, 26 and 31) denotes connection socket for a non-switched active handle actuated by a footswitch. Press to select monopolar - pinpoint coag. This symbol on the monopolar standby selection button (32) (monopolar ‘on/off’ toggle button) indicates standby mode for part of the equipment only. E-IM54f Downloaded from www.Manualslib.com manuals search engine Press to select monopolar - spray coag. P7/37 ALARM SYMBOLS PCM - Indicates the ‘Plate continuity monitor’ alarm has been activated. PAM - Indicates the ‘Plate attachment monitor’ alarm has been activated. PEM - Indicates the ‘Patient earth monitor’ alarm has been activated. PVM - Indicates the ‘Plate voltage monitor’ alarm has been activated. EPM - Indicates the ‘Excess power monitor’ alarm has been activated. - Indicates the ‘Multiple activation’ alarm has been activated. ! - Indicates an internal fault has been detected (see section 5 ‘Alarm conditions’ for details). ENVIRONMENTAL CONDITIONS FOR TRANSPORT AND STORAGE Ambient temperature range: -30°C to +50°C Relative humidity range: 30-90% RH non-condensing Atmospheric pressure range: 1060hPa down to 690hPa P8/37 Downloaded from www.Manualslib.com manuals search engine E-IM54f TD830 ELECTROSURGICAL UNIT 3.0 SAFETY NOTES & ALARMS Attention to the following points is necessary in order to reduce the risk of accidental burns during the use of this electrosurgical unit. Other safety notes and warnings are also given within the text of this manual and these should be noted before using this electrosurgical unit. DO: ♦ DO NOT: Use only Eschmann accessories, in particular active ♦ and plate electrodes and cables which should ♦ preferably be no more than 3 metres in length. All Eschmann accessories are rated above the corresponding maximum peak output voltage of this electrosurgical unit (see graphs in section 5). For reference Eschmann accessories can be used safely at the following h.f. peak voltages: Bipolar active - 400 V peak Monopolar active - 4000 V peak Monopolar plate electrode - 2250 V peak ♦ ♦ Use bipolar techniques in preference to monopolar, ♦ whenever possible. For surgical procedures on parts of the body having a relatively small cross-sectional area, the use of bipolar techniques may be desirable ♦ to avoid unwanted coagulation. Seek approved qualified advice (e.g. cardiology department) before using this electrosurgical unit on patients with implanted pacemakers or other active implants to avoid interference or damage to the implant. ♦ Monitor such patients carefully. ♦ Set power output controls to the minimum setting before use and select the minimum power setting to achieve the desired effect. ♦ Check all cables and accessories routinely before use. In particular, electrode cables and endoscopically ♦ used accessories for possible damage to the insulation. ♦ Ensure the entire area of the plate electrode is reliably attached to the patient’s body and as close to the ♦ operating field as possible. ♦ ♦ Store temporarily unused active electrodes such that they are isolated from the patient [e.g. use a quiver ♦ (REF 83-186-38) to hold active accessories when not in use]. E-IM54f Downloaded from www.Manualslib.com manuals search engine Do not use uninsulated forceps, monopolar or bipolar. Do not place monitoring electrodes close to the operating site. When high frequency surgical equipment (i.e. electrosurgical equip-ment) and physiological monitoring equip-ment are used simultaneously on the same patient, any monitoring electrodes should be placed as far as possible from the surgical electrodes. Needle monitoring electrodes are not recommended. In all cases, monitoring systems incorporating high frequency current limiting devices are recommended. Do not allow active cables to drape across or contact the patient's body or contact the cables or leads of other equipment. Do not use flammable anaesthetics, flammable solvents or oxidizing gases such as nitrous oxide (N2O) or oxygen if the surgical procedure is carried out in the region of the thorax or head, unless these agents are sucked away. Do not allow the patient's body to touch conductive objects. The patient should not come into contact with metal parts which are earthed or which have an appreciable capacitance to earth, e.g. operation table, supports, etc. The use of antistatic sheeting is recommended for this purpose. Do not use hook type active cables (with or without adapters designed for use with hook type cables) use 4 mm or 8 mm plug type active cables. Do not reuse disposable plate electrodes. Do not rely solely on surgical gloves to provide insulation. Do not allow ‘skin-to-skin’ contact (e.g. between the arms and body of the patient), this can be avoided, for example by the insertion of adequate dry gauze. P9/37 Attention to the following points will prolong the life and efficiency of your electrosurgical unit and will help to avoid the risk of accidents, or damage: DO: ♦ ♦ ♦ ♦ Excess Power Monitor (EPM). 3.4 The unit monitors its output power in monopolar and bipolar modes, compares this with the maximum allowed power under single fault conditions and alarms, disabling output, if the delivered power exceeds the allowed value by more than a certain margin. Patient Earth Monitor (PEM). Switch off and disconnect from the mains electrical supply prior to cleaning the equipment and when it is 3.5 The PEM circuit detects a low impedance path between the patient and earth. It also reduces the chances not in use. of secondary contact electrosurgical burns. The circuit Grasp the connector(s) not the cable when connecting detects earth contact paths while the plate electrode is and disconnecting cables and leads from the attached to the patient and the unit, when a monopolar equipment. fingerswitch or footswitch is pressed, but before the monopolar output is activated. If such an earth path is Contact the hospital electronics engineer or Eschmann present, the alarm will operate until the path has been After Sales Service Department if the equipment fails removed and the unit reset by switching ‘off’ and ‘on’ at the to function after checking. mains supply switch or if a pressed monopolar footswitch Ensure the equipment is serviced regularly (at least or fingerswitch is released. every six months). Contact Eschmann After Sales Plate Voltage Monitor (PVM). Service Department for details. 3.6 Is designed to prevent dangerous electrosurgical voltages appearing on the patient’s body. Such voltages DO NOT: could occur because of insulation faults in the active circuit ♦ Use a faulty unit as failure could result in an unintended or because the active electrode is in contact with an earthed object and could cause burns at points of contact between increase of output power. the patient and conductive objects. This alarm will operate ♦ Service this equipment unless you are suitably qualified when monopolar output power is ‘on’ and the fingerswitch to do so. or footswitch is operated. This monitor will not detect a poor contact between the plate electrode and the patient. Alarm Circuits Note: Under specific electrical conditions, with the plate electrode completely detached from the patient, the PVM 3.1 The unit has several alarm circuits as detailed in could alarm. the following sections. Plate Continuity Monitor (PCM). 3.2 Ensures that the unit cannot be used in the monopolar mode with an electrically defective plate electrode or cable. It does not monitor the contact between the plate electrode and the patient’s body. It is reset automatically when a satisfactory plate electrode and cable are connected to the unit. Plate Attachment Monitor (PAM). 3.3 Ensures that the unit cannot be used in the monopolar mode if a divided plate electrode is attached to the patient’s body incorrectly. Applies only when a divided plate electrode is used and will not operate if monopolar is switched ‘off’ by the monopolar standby selection button. P10/37 Downloaded from www.Manualslib.com manuals search engine Multiple Activation Alarm. 3.7 Operates if more than one fingerswitch or footswitch is pressed at the same time in a given user section (i.e. monopolar USER 1 or USER 2 or bipolar). It will automatically be reset when ALL activations have been released. This alarm will also operate if a footswitch or fingerswitch is being activated when the unit is switched ‘on’. (Note: This could indicate a damaged accessory locked in the ‘on’ position). Internal error alarm 3.8 Operates if an internal fault arises such as power being detected when not enabled by a footswitch or fingerswitch. E-IM54f TD830 ELECTROSURGICAL UNIT 4.0 INTRODUCTION GENERAL 4.7 A plate electrode is not required and safety is increased even further as two power range outputs are available allowing precise setting of the power output. 4.8 The bipolar mode of operation offers a choice of Note: For numbers in brackets refer to Illustration 11 which footswitch, electrical or pneumatic. folds out at the end of this manual. 4.1 The TD830 electrosurgical unit provides two outputs both for high power monopolar cut and coag (on a first come first served basis), together with a separate output for bipolar coag. Bipolar coag output is available simultaneously with either monopolar output. The unit incorporates touch button controls, seven segment LED (light emitting diode) displays and LED visual indicators. 4.2 The digital displays can show two scale modes, one shows the typical power available in watts the second shows a numerical value up to 10 (the latter is only displayed whilst digital display range selection toggle button (35) is held pressed). Monopolar standby mode 4.9 The unit can be switched into monopolar standby mode, in which only bipolar power is available, using monopolar standby selection button (32). This button is a toggle switch, each press of the button will turn monopolar mode ‘on’ (if ‘off’) or ‘off’ (if ‘on’). This button (32) can be considered as an ‘on/off’ switch for monopolar, the advantage being that if only bipolar outputs are required and monopolar standby mode is selected, no plate electrode needs to be connected to the unit. (If monopolar standby mode is not selected for ‘bipolar use only’ and a patient plate electrode is not connected to the unit the PAM alarm or PCM alarm will activate). The unit is in monopolar standby mode when only the digital display for bipolar can be seen illuminated. Note: The power delivered depends on the resistance between the active electrode and the plate electrode (or between the tips of bipolar forceps) at the time the power is applied. This resistance can vary widely and many times Digital display options per second during the application of h.f. output. 4.10 The digital displays normally show the typical power* 4.3 The unit incorporates several alarm code indicators available in watts. The digital displays can also show a as detailed in the Technical Data section. Innovative safety numerical value up to 10 (maximum). To display the power features include a divided plate electrode monitoring setting as a numerical value up to 10 press and hold the system called the Plate Attachment Monitor (PAM) as well digital display range selection button (35). The display will as the standard Eschmann Plate Voltage Monitor (PVM), revert to the normal setting (and display the typical power* Patient Earth Monitor (PEM) , and Plate Continuity Monitor available in watts) when the button is released. (* typical (PCM). An Excess Power Monitor (EPM) and a Multiple power is an indication of the average power delivered over activation alarm are also included. A high frequency a range of load resistances. It is less than the power leakage control circuit is included in the monopolar mode. delivered to the rated load for maximum power.) 4.4 Within this manual the terms USER 1 and USER 2 are used to distinguish between the two monopolar outputs. ACCESSORIES The bipolar output is fully independent of the monopolar 4.11 The equipment is designed to be used with the output and can be considered as USER 3. Eschmann range of active, plate and bipolar cables. The front panel is colour coded, PALE BLUE for bipolar, OPERATING MODES / DISPLAY OPTION YELLOW for monopolar cut, blend and specialist cut, and BLUE for monopolar coag. For a list of all accessories and Monopolar mode an index of accessory illustrations, see section 7. 4.5 The monopolar operating mode offers a choice of fingerswitch or footswitch control of ‘cut’ or ‘coagulation’ outputs for USER 1. Fingerswitch operation utilises the Eschmann two-button fingerswitch whilst footswitch control requires the use of an active electrode handle in conjunction with one of a range of electrical footswitches. Bipolar mode ASSOCIATED PUBLICATIONS 4.12 This manual contains preparation for use, and operation instructions only. For a detailed technical description, together with complete maintenance procedures including an illustrated parts list refer to the TD830 Service Manual E-SM44 (part number 698260). 4.6 Bipolar coag is an efficient method of effecting EQUIPMENT CERTIFICATION haemostasis, and closure of vessels such as Fallopian tubes. It is extremely safe to use, as the main current path 4.13 The electrosurgical unit fully complies with the major is between the tips of the bipolar forceps and the tissue international safety standards (see Technical Data). held between them is directly in view of the surgeon at all CAUTION times. This electrosurgical unit is to be operated by medically qualified personnel only. E-IM54f Downloaded from www.Manualslib.com manuals search engine P11/37 ELECTROSURGICAL TECHNIQUES General 4.14 When using electrosurgery, it must always be remembered that its purpose is to cut, destroy or modify tissue. This can occur at any point on the body where the concentration of electrical current becomes sufficiently high. Under fault, or misuse conditions, this current concentration can occur on points on the body other than the desired operation site and can cause burns. 4.15 To help reduce the possibility of such accidental burns, the unit is fitted with a leakage limiting circuit in the monopolar section. This circuit detects the conditions that may lead to an increased possibility of accidental burns from leakage and limits the leakage current accordingly. The HF leakage limiter does not generate an alarm. In addition, the following paragraphs give guidance for the safe use of the electrosurgical unit. 4.16 With monopolar electrosurgery, the surgeon uses a single point electrode to concentrate high frequency electric current in the body tissues close to the electrode tip. If the operating site is linked to the bulk of the body only by a tissue path of small cross-sectional area, e.g. Fallopian tube or penis, the surgical effect of the electrosurgery can extend along it from the electrode with undesirable results. ♦ ♦ ♦ ♦ ♦ Do not apply the plate electrode adjacent to ECG electrodes and cables, or to an area where fluids could pool. With babies and small children the plate electrode should always be applied to the trunk section of the body to give maximum contact with the skin. It is unsatisfactory to attach the plate electrode to small limbs where overlap of the plate electrode ends can occur. Subject to application instructions, the plate electrode should be on a smooth, resilient, muscular site well supplied with blood vessels and as close as possible to the operating area. The site should be clean and dry, and shaved of excessive hair. The plate electrode should be securely attached with its whole area in contact with the patient’s body. Do not cut or modify the plate electrode in any way. Do not reuse disposable adhesive plate electrodes, as adhesive failure could occur. 4.21 Adhesive plate electrodes (REF 83-122-42 and 83122-46) with detachable cables, should only be used with a connector cable (REF 83-122-93) that requires complete insertion of the connector tab. This ensures that contact between the tab and any conductive object cannot occur. Footswitches 4.17 Normally the surgeon will have ensured that the operating site is bounded by a bulk of tissue into which the current can disperse, so that the cut or coagulating effect is limited to the tissues close to the electrode. The current then disperses into the body and leaves it by way of the relatively large plate electrode, having passed harmlessly through any intervening tissue and organs. 4.22 Footswitches (REF 83-581-09 and 83-581-13 for yellow, or REF 83-581-17 and 83-581-25 for blue, or twin footswitches REF 83-109-39 and 83-109-47) must always be under the direct control of the surgeon or the designated assistant. Precautions must be taken to prevent footswitches contacting the operating table base, castors of trolleys, or other equipment under which they could be 4.18 In contrast, with bipolar electrosurgery there is no trapped in the ‘on’ position. plate electrode and the surgical effect occurs between the 4.23 During a surgical procedure and when the power tips of a pair of insulated bipolar forceps held by the output is not activated, check that: surgeon. The legs of the forceps constitute two ♦ The green ‘mains’ supply lamp is ‘on’. electrosurgical active electrodes. ♦ The bipolar and monopolar (cut and coag) power activated LEDs are ‘off’. Monopolar Electrosurgery ♦ The audible ‘running tone’ is ‘off’. Plate Electrode 4.19 The plate electrode is also known as a neutral Active Electrode Handles electrode, passive, return, dispersive, indifferent or patient 4.24 During surgery, when an active electrode handle plate electrode or simply as the ‘pad’. (REF 83-140-20 without cable, or REF 83-141-01, 4.20 Always use a plate electrode of proven technical 83-141-36, 83-141-28 and 83-141-44 with cable) or design and safety, and in strict accordance with the fingerswitch (REF 83-582-06, 83-583-97 or 83-585-94) is application instructions issued with the plate electrode. not in use, it should be put in an electrosurgical quiver Burns at the site of the plate electrode are the most (REF 83-186-38) or similar receptacle. Do not put the common form of electrosurgical accident. Maximum handle or fingerswitch: contact with the patient's skin must be maintained ♦ On the patient drapes. throughout the surgical procedure. The following points ♦ With other instruments. should also be noted: ♦ On the electrosurgical unit cabinet or on any other ♦ Do not apply any plate electrode over bony theatre equipment. protuberances, metal implants, folds of skin, scar tissue, hairy areas, any form of skin discoloration, an This will further reduce the possibility of electrosurgical burns due to accidental activation of the electrosurgical injury, or to limbs with restricted blood supply. P12/37 Downloaded from www.Manualslib.com manuals search engine E-IM54f TD830 ELECTROSURGICAL UNIT unit. Also do not use active electrode handles that have cables fitted with ‘hook’ type connectors (those detailed above are suitable for use with this unit). The absence of exposed metal parts on pin or jack plug designs reduces the risk of operator burns compared to cables with the ‘hook’ type design. Power Control 4.33 Use only the minimum power control setting necessary to produce the desired electrosurgical effect. If in doubt start at a low setting and increase the output in small steps until the desired effect is produced. 4.34 If, during a surgical procedure, the output appears to drop or the electrosurgical performance 4.25 Only active cables supplied by Eschmann should decreases, stop and check the plate electrode for be used. This is of particular importance, as cables with correct application and connection before increasing inadequate insulation can cause burns due to insulation the control setting. Also check all active connections. failure. 4.35 If only one type of output is required, i.e. cut or 4.26 It is good practice to ensure that active cables are ‘coagulation’, it is good practice to set the other output correctly positioned when in use and are not allowed to controls to minimum. drape across the patient's body or across other cables. Active Cables Current Paths 4.27 Active cables of the ‘hook’ type have been replaced by pin or jack plug designs (REF 83-135-12, 83-137-17, 83-138-06 and 83-138-22) any of these are suitable for use with this unit. Old style cables with ‘hook’ type connectors should not be used. The absence of exposed metal parts on pin or jack plug designs reduces the risk of operator burns compared to cables with the ‘hook’ type design. 4.36 The path of an electrosurgical current through a patient's body should, whenever possible, be the shortest path. If it is suspected that this path could be bypassed, for example between limbs, an insulating medium should be used. 4.37 It is advisable that skin to skin contact between parts of the patient's body, for example between the 4.28 Active cables 3 metres long should be used in normal arms and the trunk, should be avoided. Dry gauze or situations, even if it means repositioning the electrosurgical other material should be used to provide separation. unit. Longer types, of 5 metres or more, should only be Accidental Contact used where the electrosurgical output is a fixed distance from the operation table, e.g. with certain types of pendant 4.38 Do not allow contact to occur between any part supply, or with lamina airflow curtains. Longer cables of the patient's body and any metal or conductive material, such as parts of the operation table, increase the risk of electrical or mechanical hazard. anaesthetic screens or drip stands. Active Electrodes 4.29 The correct type of active electrode (see section 7) Bipolar Electrosurgery should always be used. Electrodes should be kept clean during use either by wiping with a solvent on a sterile swab General or by using a sterile abrasive pad. 4.39 Bipolar coag provides a very safe and precise means of controlling bleeding in many types of surgery, CAUTION Ensure that the active cable is unplugged from and can be used in conjunction with monopolar techniques where cutting, or higher coagulation powers the unit when cleaning electrodes during a are required. surgical procedure. 4.30 Ball type coag electrodes should be discarded if the surface becomes pitted. Do not sterilize electrodes which are designed as ‘single use disposable’ (also see the leaflet detailed in section 1, supplied with all accessories ). Fingerswitches 4.31 For many procedures, ‘on/off’ control of cut and coag outputs by means of buttons on a fingerswitch offers the surgeon more precise control. Reusable fingerswitches have only a limited number of re-sterilization cycles. Regular testing and inspection is therefore essential. Do not sterilize and reuse ‘single use’ types. 4.32 Do not connect footswitches to the USER 1 output of the electrosurgical unit when a fingerswitch is being used to control the USER 1 output. E-IM54f Downloaded from www.Manualslib.com manuals search engine 4.40 With bipolar techniques, the electrosurgical current flows mainly between the tips of the bipolar forceps or between the contact points of a special purpose electrode and therefore does not require a plate electrode. Bipolar electrosurgery is safer than monopolar, the main safety requirements are: ♦ Always use Eschmann bipolar accessories. ♦ After sterilization, always ensure that the accessories and cable connector are completely dry. ♦ Always ensure that accessories are sterilized by a process compatible with the accessory, consult the accessory leaflet provided with all Eschmann accessories (publication number E-IM50 part number 604802). P13/37 Flammable Anaesthetics and Spirits if the electrosurgical active electrode and plate electrode are kept a few inches away from the pacemaker pulse generator and its electrode leads; while to avoid burns to the myocardium at the electrode site, or possible ventricular fibrillation, they should remain at least 15 cm away from the heart. 4.41 The use of flammable anaesthetics and nitrous oxide (N2O) or oxygen should be avoided if a surgical procedure is to be carried out in the region of the thorax or the head, unless these agents are removed by suction. Flammable agents used for cleaning or disinfecting, or as solvents of adhesives should be allowed to evaporate before the 4.47 It will also be sensible to choose a location for the application of electrosurgery. (Also see WARNING above plate electrode which puts the path of electrosurgical current through the patient's body as far as possible from section 6.1). the heart and from the pacemaker and its leads. 4.42 There is a risk of pooling of flammable fluids under the patient or in body depressions such as the umbilicus 4.48 If the pacemaker is programmable, it is suggested and in body cavities such as the vagina. Any fluid pooled that it is set to either the VVT (ventricular sensing and in these areas should be mopped up or removed by suction triggering) or the V00 (fixed rate) mode (preferably the latter if the patient is not at particular risk from a competitive (for gaseous fluids) before the equipment is used. rhythm) so that it functions as a fixed rate pacemaker during 4.43 Attention is drawn to the danger of ignition of the surgical procedure. endogenous gases (i.e. those gases produced within the body). Some materials, e.g. cotton wool and gauze, when 4.49 It is further recommended (References ¹ and ²) that saturated with oxygen may be ignited by sparks produced the peripheral pulse be monitored during the course of electrosurgery, and (Reference ²) that precautions be taken in normal use of the equipment. to ensure that the patient's well-being is maintained in the event of interference with pacemaker operation. By using Implanted Cardiac Pacemakers the electrosurgical equipment in short bursts, it is General suggested (References ¹ and ²) that, at most, only one or two beats will be affected. 4.44 If in monopolar electrosurgery the intervening tissue and organs contain an implanted pacemaker and its leads, it is possible under certain conditions that the electrosurgical current will have undesirable effects. The following notes offer guidance on the risks and on possible precautions. WARNING For patients with cardiac pacemakers or other active implants, a possible hazard exists because interference with the action of the pacemaker may occur or the pacemaker may be damaged. In case of doubt, approved qualified advice should be obtained. Risks 4.45 Operating with electrosurgical equipment on patients who have an implanted pacemaker incurs the risks of affecting the pacemaker's operation, either by direct contact or by radiated interference, and of inducing burns at the pacemaker electrode implantation site. Clearly the risks are very much lower with bipolar than with monopolar electrosurgery, because of the highly localised current path. Precautions WARNING More severe consequences, local heating and destruction of the pacemaker circuit (Reference ¹), would result if an electro-surgical electrode touched the pacemaker. Electrosurgical Interference 4.50 Electrosurgical equipment can produce interference with other electromedical apparatus used in the operating theatre. Particular attention should be given to the selection of ECG monitors with good interference rejection if it is desired to monitor the patient when the electrosurgical equipment is active. 4.51 In certain conditions a reduction in ECG interference can be obtained by placing the ECG electrodes closer together than normal. Modern monitors with high input isolation allow electrode placement close to the electrosurgical plate electrode, if required. 4.52 It is inadvisable to use mobile telephones or ‘walkietalkie’ equivalent equipment close to electrosurgical or any other medical electrical equipment. 4.53 To minimize the possibility of electrical interference 4.46 Recommendations published in the U.S.A. from an arc, between the active electrode and the patient, (References ¹ and ²) suggest that an implanted pacemaker affecting the TD830, it is good practise, in common with all should be unaffected, or only revert to fixed-rate operation, electrosurgical units, to keep them as far as is reasonably possible from the surgical site. P14/37 Downloaded from www.Manualslib.com manuals search engine E-IM54f TD830 ELECTROSURGICAL UNIT using an S-Lead [REF 83-139-41 (Olympus 9mm), or REF 83-139-49 (Pentax 8mm)] connected to the S-Lead socket (29, Fig 12) of the TD830 Electrosurgical unit and the flexible endoscope. Using Two Electrosurgical Units 4.54 This electrosurgical unit has dual monopolar outputs (first come first served) and simultaneous bipolar output. It is therefore very unlikely that another electrosurgical unit would need to be used on the same patient at the same time, however if it is necessary follow the guidelines detailed in section 4.55. 4.55 In certain procedures it is convenient or even essential that two electrosurgical units are used on the patient at the same time. There is no reason why such operations should not be safely carried out if the following points are observed: ♦ Do not use a fully floating output unit with an earth referenced unit or even a partially floating unit, for example: ♦ ♦ ♦ ♦ Eschmann TD411 or DS402 and a Valleylab Force 4 ♦ Eschmann TD411 or DS402 and a GU Solstar or any older GU unit. Note: All Eschmann units have fully floating outputs and any two may be used in combination. Always use two plate electrodes, one for each unit. Never join plate cables or overlap plate electrodes. Position each plate electrode as close as possible to the respective operation site. Do not allow electrosurgical current paths to intersect, e.g. if the main procedure is in the thoracic region, the plate electrode for this unit should be placed on the patient’s back, buttocks, or upper thigh. If the secondary procedure is vein removal from the leg, the second plate electrode should be placed on the thigh or calf of this leg. For maximum safety use fingerswitches wherever possible. This reduces the possibility of inadvertent activation of either unit’s output. Always use two quivers, one for each active handle or fingerswitch. Using a flexible endoscope 4.56 To ensure that the endoscope connection socket is not used without a documented procedure in place, the SLead socket (29, Fig.11) is blanked of with a plug to avoid inadvertent use. Before removing this plug the following notes should be considered when writing a procedure for the use of this electrosurgical unit with an endoscope. a. b. When using a flexible endoscope low power settings only should be used (e.g. maximums of 30% for cut, 50% for pinpoint coagulation). Also because of the high voltage waveform of spray coagulation it should only be used at very low power settings see 3 below. c. It is important to ensure that the flexible endoscope used is compatible with the voltages produced by the TD830 Electrosurgical unit. The maximum voltages produced by the TD830 Electrosurgical unit for each mode and display setting are shown in the graphs at the end of section 5. These should be compared to the voltage rating of the flexible endoscope. (Note: Flexible endoscopes complying with IEC60601-2-18 should detail their voltage rating in their accompanying documentation. If in doubt consult the flexible endoscope manufacturer). SERVICING 4.57 It is recommended that electrosurgical safety checks and routine servicing are carried out at regular intervals (every six months) and only by Eschmann trained personnel or Eschmann trained hospital engineers, otherwise the warranty could be infringed. 4.58 Read the information given in this manual carefully before using, cleaning, sterilizing, or servicing the electrosurgical unit. References ¹ ² ‘Optimal resources for implantable cardiac pacemakers’ (Section on electromagnetic interference) - a report by the Inter-Society Commission for Heart Disease Resources, published by the American Heart Association in their journal ‘Circulation’ (1983), 68(1):232A-233A. ‘Electrosurgical device interference with implanted pacemakers’ - a question and answer section of the ‘Journal of the American Medical Association’. (1978), 239(18:1910). To avoid the risk of potential burns caused by h.f. voltages on the body of the flexible endoscope some users may prefer to connect the endoscope to the TD830 Electrosurgical unit. This is accomplished E-IM54f Downloaded from www.Manualslib.com manuals search engine P15/37 5.0 OPERATION WARNING 1) This electrosurgical unit is to be operated by medically qualified personnel only. 2) Note that interference produced by the operation of this unit may adversely influence the operation of other electronic equipment. 3) While the electrosurgical unit is activated (i.e. the running tone can be heard and power activated indicators are illuminated) do NOT touch the patient with a contact area of less than 10cm2. This is especially important when monopolar mode is being used at high power settings. 4) Do not touch a monopolar or bipolar active electrode with its active cable plugged into the electrosurgical unit, even when changing the electrode. ♦ ♦ ♦ ♦ Plate electrode. With reusable types, examine the plate cable and plug for signs of wear. Ensure that the insulation of the connecting cable is in good condition. Accessory cables/cords (reusable monopolar active and bipolar types). Examine all cables for condition and attachment to end fittings. Note: All active cables are subject to deterioration caused by repeated sterilization and should be renewed regularly. Fingerswitches. Check buttons on fingerswitches for function and free action. Note: All fingerswitches have a limited life and should be renewed regularly. Footswitches. Check all footswitch cables and connectors for condition. Check all switches for correct function and free action. CONNECTION OF ACCESSORIES PRIOR TO USE or SYSTEMS CHECK Note: For numbers in brackets refer to Illustration 11 which folds out at the end of this manual. 5.1 Before use ensure the details covered in the preliminary information sections 1.5 to 1.7 have been Monopolar performed and the routine checks below in section 5.2 have 5.3 There is a choice of active cable connections to the been completed. Connect the required accessories as lower front panel of the unit for the monopolar outputs: detailed in sections 5.3 to 5.9 and then carry out a systems ♦ A three-pin fingerswitch connector can be plugged check detailed in sections 5.10 to 5.17 before using the into active fingerswitch sockets (27 and 28) for USER unit. 2 and active fingerswitch sockets (25 and 26) for USER 1. ROUTINE CHECK ♦ A single active cable with 4 mm plug (footswitch 5.2 Routine checking should be limited to external operated) can be connected to active electrode checks of the unit, and must include all accessories and socket (26) for USER 1. cables. Routine checks must be carried out each time the unit is used as follows: CAUTION WARNING If in doubt, discard suspect component and substitute with one of known function, continuity and condition. ♦ Mains electrical supply cable/cord. Examine mains electrical supply plug for condition and correct attachment to cable/cord. Check condition of cable/ cord insulation over its entire length. WARNING Do not repair a damaged disposable plate electrode or reusable plate electrode. Always substitute a new plate electrode and, if necessary, the cable. Always ensure that a supply of new plate electrodes is readily available. P16/37 Downloaded from www.Manualslib.com manuals search engine Incorrect insertion of a single 4mm monopolar active plug into either of the USER 1 fingerswitch sockets (25) can cause damage to the TD830 if the output is energized by a footswitch. ♦ A single active cable with 8 mm plug (footswitch operated) can be connected to active electrode socket (24) for use by USER 1 only Note: This overrides any cables plugged into the USER 1 sockets (25 and 26). CAUTIONS Only one of the above active connections should be connected to the unit at any one time for USER 1. If an adapter is connected to the unit and it does not have an active cable connected to it, the adapter must be removed from the unit. E-IM54f TD830 ELECTROSURGICAL UNIT If an 8 mm active adapter is connected to the unit, the fingerswitch sockets (25 and 26) are electrically disconnected. Bipolar 5.8 Connect a bipolar cable REF 83-041-55 to the bipolar output socket (31) and either a white electric footswitch REF 83-581-76 (Fig. 10) to the bipolar electric 5.4 For USER 1 connect an Eschmann fingerswitch (Fig. socket (37) or a white pneumatic footswitch REF 83-5816) to fingerswitch sockets (25 and 26). 68 (Fig. 10) to the bipolar pneumatic socket (36). Do not 5.5 Alternatively for USER 1 plug an 8mm active cable connect both types of footswitch at the same time. (BOVIE) into the 8mm active electrode socket (24) in 5.9 If the electrosurgical unit is only to be used in the combination with a footswitch (Fig. 4) yellow for cut, and/ bipolar mode select the monopolar standby mode (see or blue for coag, which should be plugged into either section 5.18). The monopolar digital displays (8 and 23) monopolar footswitch socket (38). These sockets are will not be illuminated and this will enable use of the unit identical, the function is determined by the colour of the without the need to connect a plate electrode to the unit footswitch. The Eschmann Twin footswitch can also be (PAM alarm or PCM alarm will activate if a plate electrode used to provide both cut and coag functions (see Fig. 4). is not attached to the unit and it is not in monopolar standby 5.6 Still another option for USER 1 is to connect an active electrode handle (Fig. 5) to 4mm active electrode socket (26) and plug a footswitch (Fig. 4) yellow for cut, and/or blue for coag, into either monopolar footswitch socket (38). These sockets are identical, the function is determined by the colour of the footswitch. The Eschmann Twin footswitch can also be used to provide both cut and coag functions (see Fig. 4). 5.7 For USER 2 connect an Eschmann fingerswitch to sockets 27 and 28. mode). SYSTEMS CHECK Note: For numbers in brackets refer to Illustration 11 which folds out at the end of this manual. WARNING If the unit fails any of the following checks it must not be used until the faults have been repaired and the checks have been successfully repeated. Notes: ♦ Sockets (25 and 26) and (27 and 28) are designed to accept the Eschmann fingerswitch. 5.10 Connect the required accessories as detailed in sections 5.3 to 5.9. ♦ The footswitch controlled active connection system using the 8 mm (BOVIE) socket (24) requires either 5.11 Connect unit to mains power supply using the cable the combined handle and cable REF 83-141-28 or 83- supplied connected to the mains inlet (40), and if the supply 141-44 (Fig. 5) or the separate handle REF 83-140- socket is controlled by a switch, ensure that it is switched 20 (Fig. 3) with cable REF 83-138-06 or 83-138-22 ‘on’. Select mains switch (42) to ‘I’ (on). The green lamp inside the mains switch (42) will come ‘on’ and the front (Fig. 7). panel displays (indicators and digital) will illuminate. ♦ For foot control, use one of the range of Eschmann 5.12 The unit may be in monopolar standby mode, if it is footswitches together with a non-switching active monopolar displays will not be illuminated and monopolar handle. will be disabled unless the monopolar standby button (32) is pressed (see section 5.18). The unit will power up in the WARNING modes and settings in use when the unit was last switched When a fingerswitch or hand controlled active off, unless the battery powering the memory has failed or handle is in use for monopolar USER 1, the lost charge, in which case it will power up in monopolar footswitch(es) must be disconnected from the standby mode. electrosurgical unit. CAUTION For safety, when using only one monopolar mode, set the unused output control to minimum e.g. ¨ For coag only, set cut power output control (14) to minimum. ♦ For cut/blend only, set coag power output control (21) to minimum. E-IM54f Downloaded from www.Manualslib.com manuals search engine 5.13 Press each selector touch button (3, 5, 11, 13, 15, 19 and 20) in turn and check that the green or bright green indicator adjacent to each touch button comes ‘on’ (2, 4, 10, 12, 16, 17 and 18 respectively). Each time a selector touch button is pressed there is an audible ‘bleep’ to signify the button has been pressed (also see the first note after section 5.15). 5.14 Set monopolar cut, monopolar coag and bipolar power output controls (14, 21 and 1 respectively) to their minimum levels (fully anticlockwise). 5.15 If the monopolar mode is to be used ensure that all monopolar electrodes are safe and activate the unit in all the monopolar modes to be used, for each USER, by P17/37 pressing a fingerswitch cut or coag button or by pressing OPERATING INSTRUCTIONS the appropriate footswitch, yellow for cut, blue for coag. Check that the relevant power activated indicators Note: For numbers in brackets refer to Illustration 11 which illuminate, yellow (9) or blue (22) for cut or coag respectively folds out at the end of this manual. and that the audible running tone can be heard. ♦ Unit settings (e.g. range selection and mode selection) are stored in the unit's memory during power-off Notes: periods and will not need reselecting when power is ♦ The volume of the running tone and the selector reapplied. Power settings will also be maintained button ‘bleep’ can be adjusted with the volume control during power ‘off’ so long as the control knobs are not (33) on the back of the unit. This should be adjusted moved. using a screwdriver of suitable size. ♦ Coagulation output is obtained only by pressing the fingerswitch blue coag button or the blue footswitch. ♦ Cut output is obtained only by pressing the fingerswitch yellow cut button or the yellow footswitch. 5.16 If the bipolar output is to be used ensure that the bipolar electrode is safe and activate the output by pressing the white footswitch connected (either electric or pneumatic). Check that the blue power activated indicator (7) illuminates and that the audible running tone can be heard. 5.17 Check that the following alarms are functioning correctly by the test methods that follow: ♦ The audible running tones operate at different frequencies for each of the three output groups (bipolar, monopolar cut and monopolar coag) as detailed in the ‘Technical data’ section. If a monopolar user activates power at the same time (but after) a bipolar user the tone will continue at the bipolar frequency until the bipolar user stops activation when the monopolar user’s frequency will start (assuming the monopolar user is still activating the unit). If the monopolar user only activates the unit for a short period during a long activation by the bipolar user then the monopolar user’s frequency will not be heard. The same principles apply if a monopolar user is activating power and a bipolar user starts to activate power at the same time. PAM - To test the PAM alarm attach the divided plate electrode to the divided plate cable with the divided Monopolar standby mode plate cable plugged into the unit. The PAM alarm should operate with the plate electrode not attached 5.18 If the unit is in monopolar standby mode only bipolar to the patient. operation is available, monopolar is disabled and the monopolar digital displays (8 and 23) will not be seen. The PCM - To test the PCM alarm remove the non-divided plate unit is switched in and out of standby mode be pressing electrode from the non-divided plate cable with the the monopolar standby button (32). The switch ‘toggles’ non-divided plate cable plugged into the unit. The standby mode ‘on’ or ‘off’ each time the button is pressed, PCM alarm should operate. monopolar standby mode is selected when the digital PEM - To test the PEM alarm with the plate electrode displays for monopolar (8 and 23) are not illuminated. attached to the unit allow the plate electrode to contact the equipotentiality point (34) at the rear of the unit using a bare metal part of the plate and not a part covered by adhesive. With cut and coag power levels both adjusted to zero (for safety) the PEM alarm should operate when a USER 1 footswitch or fingerswitch (whichever is connected to the unit) or a USER 2 fingerswitch is pressed . Note: This test is not possible with a capacitive plate electrode. Monopolar mode USER 1 and/or USER 2 selection 5.19 The TD830 can be used by two monopolar operators USER 1 and USER 2 independently of each other but not simultaneously. Each user can have their own active output under their own control operated by footswitch or fingerswitch for USER 1 and fingerswitch only for USER 2 (see sections 5.3 to 5.7) 5.20 USER 1 and USER 2 can be using either cut or -To test the multiple activation alarm press both a cut coag but the mode set within cut or coag must be the same and coag footswitch or fingerswitch control, the alarm and the power level set must similarly be the same (unless should operate. changed between each USER’s application of power). The other alarms (PVM, EPM and !) cannot easily be checked by the user without special test equipment or by using procedures that could lead to accidental burns. The function of the EPM alarm should be checked regularly, in accordance with the Service Manual (see section 1.8). P18/37 Downloaded from www.Manualslib.com manuals search engine 5.21 Because USER 1 and USER 2 cannot both be delivering power at the same time, power is available on a ‘first-come-first-served’ basis. (i.e. if USER 1 is using power and USER 2 presses a fingerswitch to demand power, none will be available and visa versa). Also note that power will E-IM54f TD830 ELECTROSURGICAL UNIT not become available to the second user until the first user has finished using power and the second user has released and reapplied demand (via footswitch or fingerswitch) after the first user has finished. ♦ The active electrode is correctly fitted in its holder and the active cable is correctly connected to the unit. Check that the cable is in good condition. Guidance on power output settings is given in sections Plate electrode (required for monopolar mode only) 5.53 to 5.55 and in Table 1. 5.22 If the electrosurgical unit is to be used in monopolar Fingerswitch / active electrode handle mode and a plate electrode is not connected to the plate 5.27 A fingerswitch or active electrode handle must be electrode connector socket (30), the unit will alarm (PCM treated with extreme care at all times. or PAM) and the audible two-note tone will sound and the displays will flash. If a non-divided plate cable is being used, WARNINGS then connecting a non-divided plate electrode will stop the 1) When either a fingerswitch or an active alarm. If a divided plate cable is being used then connecting a divided plate electrode and applying the divided plate electrode handle is connected to the unit, its electrode correctly to the patient’s body will stop the alarm. location should be closely noted. It should never be allowed to contact other instruments. WARNINGS 1) Whichever type of plate electrode is used, it is most important that the application instructions, issued with the plate electrode, are strictly followed. 2) Ensure the pin on the plate connector (or adapter) is fully inserted into the plate electrode socket (item 30, Fig. 11) before energising the output. 2) During surgery, when an active electrode is not in use, it should always be put in an electrosurgical quiver or similar receptacle. This is most important when both bipolar and monopolar electrodes are required for the same surgical procedure. 3) Similarly do not touch USER 1 active electrode when USER 2 output is activated and visa versa. 5.23 A plate continuity monitoring circuit (PCM) initiates 4) Do not touch a bipolar active electrode the audiovisual alarm system if there is a break in the non- when monopolar is activated and visa versa. divided plate cable, or a faulty connection in the non-divided plate electrode circuit. The electrosurgical output will be 5.28 Connect the required cable and active fingerswitch instantly cut off and will not be restored until the fault is to the fingerswitch sockets (27 and 28) for USER 2 and corrected. fingerswitch sockets (25 and 26) for USER 1 as detailed 5.24 If a divided plate cable and divided plate electrode in sections 5.3 to 5.7. are being used, the unit will alarm (PAM) if the divided plate electrode is not making sufficiently good contact with the patient’s body or if the divided plate electrode or its connector system is faulty. 5.29 Alternatively for USER 1 connect an active cable with a 4mm plug into active socket (26) or an active cable with an 8mm plug (BOVIE) to active socket (24) as detailed in sections 5.5 to 5.7. 5.25 Apply a plate electrode to the patient in strict Footswitches accordance with the application instructions supplied with the plate electrode and connect the plug to the plate 5.30 Eschmann footswitch(es) (Fig. 4) can be plugged into either footswitch socket (38). With either cut (11), blend electrode socket (30) on the front panel of the unit. (13) or specialist cut (15) output selected, pressing a yellow Note: The Eschmann Flexoplate must be used with (cut) footswitch illuminates the yellow power activated adapter REF 83-207-56, Fig. 1. indicator (9), and the cutting running tone will sound. 5.26 If the required surgical effect is not obtained with 5.31 With either pinpoint coag (19) or spray coag (20) the normal power setting, check the following before selected, pressing a blue (coag) footswitch illuminates the increasing power: blue power activated indicator (22), and the coag running ♦ ♦ The plate electrode is correctly positioned, and is in tone will sound. contact with the patient. 5.32 The Eschmann twin footswitch (Fig. 4) has an angled platform with two light-action switches for cut and coag. The plate electrode cable is intact, and is plugged into Colour coded bars provide foot location and protection unit plate electrode socket (30). against accidental operation. E-IM54f Downloaded from www.Manualslib.com manuals search engine P19/37 WARNINGS 1) Footswitches should be used with care at all times and should always be under control of the surgeon or designated assistant, as inadvertent operation could have serious results. 2) Footswitches should be kept clear of areas where other personnel could accidentally press them, also ensure that they do not get wedged under theatre equipment. 3) Select mains switch (42) to ‘O’ (‘off’) as soon as electrosurgery is completed. especially flexible endoscopes. Most urological endoscopes are suitable for spray coag. If in doubt check that the voltage rating of the laparoscopic or endoscopic instrument is at least as high as the maximum spray coag open circuit voltage of this electrosurgical unit (see Technical Date section). CAUTION Although power output in the spray mode is lower than in the pinpoint mode (79W rather than 170W maximum), the effect on tissue is much greater in the spray mode. Until experience is gained, only low control settings should be used. 5.39 Select pinpoint or spray coag at touch buttons (19) or (20) respectively and note correct selection by 5.33 Monopolar cut and coag outputs are obtained by illumination of a green indicator (17) for pinpoint coag and pressing monopolar selector touch buttons (11, 13, 15, 19 a bright green indicator (18) for spray coag adjacent to the or 20), then setting output controls (14) for cut or (21) for applicable touch buttons. coag to the required levels, and then using the fingerswitch Power output or pressing the appropriate footswitch. Monopolar cut and coag outputs 5.34 'Running' tones are generated whenever monopolar cut or coag outputs are activated by footswitch or fingerswitch. The volume of the running tones is adjustable by the volume control (33) on the back of the unit. A small screwdriver is needed for this adjustment. This does not affect the pitch (i.e. frequency) of the tones. 5.35 Yellow indicator (9) is illuminated when cut power is activated and blue indicator (22) is illuminated when coag power is activated, 5.36 For a cutting current with increased haemostatic effect, select blend touch button (13). Adjusting the setting of cut control (14) will only change the power output, it will not change the nature of the blended current. Blended output is only obtained by pressing the yellow (cut) fingerswitch button, or by pressing the yellow (cut) footswitch. WARNING Apparent low power or failure of the unit to function correctly at the normal operating settings may indicate faulty application of the plate electrode or poor contact in its connections. In this case, the application of the plate electrode and its connections should be checked before selecting a higher output power. 5.40 The output power level is adjusted by the power output controls (14) and (21) for cut and coag modes respectively. Turning the knob clockwise increases the power output which is displayed digitally on displays (8) and (23) respectively. 5.41 Digital power displays (6, 8 and 23) can show a 5.37 For maximum cutting speed, particularly in urology, second type of output setting display by pressing the digital select the specialist cut touch button (15). display range selection button (35) on the rear of the unit. 5.38 The two coag modes are selected by touch buttons The two available digital displays are as follows: (19) pinpoint coag and (20) spray coag. The pinpoint mode provides low voltage coagulation for most types of surgery. The spray coag mode activates a separate generator circuit which provides a waveform that reduces the cutting effect to a very low level but provides fast coagulation by allowing an arc to form between the electrode and the tissue. The waveform is a series of high energy single pulses. Spray coag allows non-contact coagulation of tissue that tends to adhere to active electrodes. Note: Spray coag (indicated with the bright green LED) is not recommended for laparoscopic and endoscopic procedures where its high voltage waveform can damage the insulation of laparoscopic and endoscopic instruments, P20/37 Downloaded from www.Manualslib.com manuals search engine ♦ When button 35 IS NOT pressed. The typical power output available for a given setting of the respective power output control is displayed. The calibration is in watts for a typical resistance (see Technical Data section for the maximum power available for each output mode). ♦ When button 35 IS pressed. The power output selected for a given setting of the respective power output control is displayed as a numerical value up to 10. E-IM54f TD830 ELECTROSURGICAL UNIT 5.48 Press micro bipolar selector touch button (3) to select micro bipolar mode or press macro bipolar selector 5.42 Bipolar output is always available irrespective of what touch button (5) to select macro bipolar mode. The green demands USER 1 and USER 2 are making from the indicators (2) or (4) will illuminate to show which bipolar monopolar outputs. mode has been selected. Bipolar mode CAUTION Only fully insulated bipolar forceps should be used to ensure that accidental contact between the legs of the forceps and body tissue does not cause electrosurgical action. Selecting bipolar Forceps 5.43 Always use fully insulated forceps to ensure safety for the operator and patient, and do not use direct switched forceps, three core type. Connecting Forceps 5.44 Select bipolar forceps (Fig. 9) to suit the surgical procedure, and plug them into the bipolar forceps cable. Connect the cable to bipolar output socket (31). Changing Forceps 5.45 To change the type of forceps during use, e.g. from bayonet to straight, set mains switch (42) to ‘O’ (‘off’) change forceps, and then switch to ‘I’ (‘on’). Alternatively, unplug the bipolar cable from the unit, change the forceps, then plug the bipolar cable back into the unit. Footswitches 5.49 Set power output control (1) to the required level. Note that the digital power display (6) shows the typical available output power for either power range. (see 5.41 for range control selection). 5.50 Activation of the connected footswitch will turn on the power for as long as the footswitch is depressed. Notes: ♦ Blue power activated indicator (7) will come ‘on’ and the running tone will sound only when the output is activated (see 5.34 for adjustment of volume). ♦ The 15W (micro) range should be used initially if the power level required is not known. ♦ The method of ‘constant amplitude’ power control provides very effective coagulation at low power levels. After use 5.51 After using the electrosurgical unit, switch ‘off’ the mains switch (42) and disconnect it from the mains power supply. Disconnecting accessories 5.45 Two types of footswitch can be used in bipolar mode: 5.52 When disconnecting active cables or footswitches from the electrosurgical unit, always hold the connector(s) ♦ A white pneumatic footswitch (REF 83-581-68, Fig.10) not the cable. which is a light-action bellows type, with a push-on Power output selection guide connector tube. A white electric footswitch (REF 83-581-76, Fig.10). 5.53 Various makes of electrosurgical units have different maximum output powers for their modes, and, more 5.46 The type of footswitch used is a matter of personal importantly, they have different power output curves. These choice. The pneumatic type has a very light action and is characteristics determine the actual power delivered to the suited to procedures where repetitive actuation is needed. tissue for a given control setting. The TD830 electrosurgical unit has a power curve that rises at the low resistance Note: end, this ensures that adequate power is available for ♦ Only one bipolar footswitch should be connected to demanding surgical procedures. This power curve shape the unit at any time. also enables the full range of power control settings to be used, low values for general or delicate surgery, or high Power Output values as shown in Table 1. All power curves are provided CAUTION at the end of this section. ♦ Set power output control (1) to ‘minimum’. (see section 3) 5.47 The bipolar mode is extremely safe, but it is always good practice to start at a low setting and increase slowly until the required effect is obtained. Bipolar coagulation of most blood vessels is normally self-terminating because the rapid rise in tissue resistance, brought about by coagulation, limits the flow of current. An audible ‘pop’ can result from the rapid boiling of liquids in the tissue. E-IM54f Downloaded from www.Manualslib.com manuals search engine 5.54 The best results are obtained by varying the setting rather than having a predetermined value for all uses. It must be noted that digital power displays, only display the typical available power for a given control setting. This is proved by the fact that the displayed value does not change between the active electrode being out of contact with the tissue (zero power) or deeply embedded in well perfused flesh (near maximum power delivered). P21/37 5.55 The output control settings required for a particular surgical procedure are also dependent on a number of external factors (type of tissue, speed at which the surgeon works, type of electrode in use, area of forcep tips in contact). Therefore the recommended settings given in Table 1 are for guidance only. Alarms 5.57 If more than one output demand is received at the same time (such as two footswitches or a footswitch and a fingerswitch button being pressed) the unit will produce an alarm indication and no power output. This is an important safety feature and applies to any combination of footswitch and fingerswitch for monopolar USER 1, or for USER 2 or for bipolar. Alarm identifications and remedies Alarm Conditions 5.58 Alarm identifications are displayed on the front panel 5.56 The audiovisual alarm, two-tone with flashing (see Fig. 11 A1-A7) and described in section 3, remedies are as follows: displays at one second intervals, will operate if: ♦ ♦ ♦ The non-divided plate electrode is not connected via If the PCM or PAM circuit has been activated. the plate cable to the unit, or the plate electrode and/ Remedy: or cable is faulty (PCM). (i) Check plate electrode is connected to unit. A divided plate cable is plugged into the unit and the (ii) Check plate electrode and cable for condition, if divided plate electrode is not making a sufficiently faulty, substitute new plate electrode and/or cable. good contact with the patient’s body (PAM). This procedure should normally be carried out before plate electrode is attached to patient. A low impedance path exists between the patient’s body and earth (PEM). This condition can be caused (iii) If a divided plate electrode is being used check that by the presence of an earthed ECG electrode, or by the divided plate electrode is making good contact an accidental contact between the patient’s body and with the patient’s body. a conductive object such as part of the operation table If the PEM circuit has been activated. or a drip stand. Remedy: WARNING The patient earth monitor (PEM) will not detect a low impedence path between the patients body and earth when the electrosurgical output is activated. (i) (ii) If applicable disconnect suspect ECG electrodes. Note: An activated PEM alarm will automatically be reset by releasing a depressed footswitch or fingerswitch. ♦ Excess power is detected (EPM) or dangerous voltages are detected (PVM). ♦ More than one footswitch or fingerswitch is actuated for monopolar USER 1, USER 2 or bipolar, e.g. If the ♦ ♦ fingerswitch and footswitch both actuated on the monopolar USER 1 output. ♦ or, two fingerswitches on monopolar USER 2. ♦ or, pneumatic and electric footswitch on bipolar output. ♦ or, any other forbidden combination of footswitch and fingerswitch commands. Internal unit errors are detected (!) In any alarm mode, all outputs are inhibited and alarm volume is maximum, regardless of volume control setting. If the unit is in standby mode only the bipolar display will flash. P22/37 Downloaded from www.Manualslib.com manuals search engine Check for accidental direct contact between patient and an earthed object, (drip stand or conductive part of operation table). circuit has been activated Remedy: (i) The alarm will operate if the bipolar pneumatic footswitch and the bipolar electrical footswitch are pressed simultaneously. The alarm will also sound if both the cut and coag footswitches, or if one of the two footswitches and a fingerswitch button are pressed simultaneously for USER 1, or if both fingerswitch buttons are pressed for USER 1 or USER 2. The circuit will automatically reset when ALL activations stop from ALL switches. This is an important safety feature and applies to any combination of footswitch and fingerswitch. E-IM54f TD830 ELECTROSURGICAL UNIT If the EPM circuit has been activated. Remedy: (i) Because the EPM alarm cannot be reset by the User, if unit EPM alarm activates as soon as output is activated, Eschmann Equipment should be contacted. If the PVM circuit has been activated. Remedy: (i) If unit PVM alarm activates check for accidental direct contact between the active electrode and an earthed object. (ii) Check that the plate electrode is correctly attached to the patient’s body. Note: An activated PVM alarm will automatically be reset by releasing a depressed footswitch or fingerswitch. If none of the above remedies stop the alarm activated, or the (!) alarm is activated contact Eschmann Equipment and do not use the unit. E-IM54f Downloaded from www.Manualslib.com manuals search engine P23/37 Table 1 Power Output Selection Guide (for ‘typical’ power display range) Procedure Mode Condition Setting CAUTION Do not use a higher power setting than is necessary General Normal cut Pinpoint coag Blend Dry field Dry field Dry field 70 to 90 25 to 45 45 to 70 Cardiothoracic Normal cut Pinpoint coag Wet field Wet/dry field 70 to 90 30 to 55 Sternal oozing Spray coag Wet/dry field 35 to 60 Intercostal muscle Spray coag Wet/dry field 35 to 65 Gynaecology LLETZ (Large Loop Excision of the Transformation Zone) Blend Spray coag Blend Spray coag N-TZ (Displasia CIN I) 45 to 70 40 to 55 50 to 80 50 to 60 DEXC (Diathermy Excision of the cervix) Specialist cut Spray coag Specialist cut L-TZ (CIN II to III) Microinvasive TZ Identification 50 to 60 45 to 60 40 to 60 TCRE: TCAE (Transcervical Resection of the Endometrium): (Ablation) Blend Specialist cut Blend Spray coag Normal endometrium Spongio-endometrium Suspect fibroid inv. Fundal ablation 55 to 70 40 to 60 65 to 80 50 to 65 Urology Specialist cut Spongio-prostate Fibro/Muscular pros. Bladder tumour L-TZ (CIN I to II) Normal cut Spray coag Pinpoint coag Ball electrode 135 to 160 155 to 175 115 to 135 40 to 60 40 to 60 Plastic/burns Spray coag Eschar formation 40 to 50 Laparoscopic cholecystectomy Blend Spray coag Dry/wet field Liver bed bleeding 50 to 70 55 to 65 Notes: ♦ Monopolar blend output mode can be used with advantage in many cutting procedures, particularly when only a single (yellow) footswitch is being used. ♦ The blend mode produces less output power for a given control setting than normal cut, as shown on digital power display (8 of illustration 11). Bipolar micro Bipolar macro Monopolar cut Monopolar blend Monopolar Pinpoint coag Monopolar spray coag P24/37 Downloaded from www.Manualslib.com manuals search engine Monopolar Specialist cut E-IM54f TD830 ELECTROSURGICAL UNIT MONOPOLAR CUT DIAGRAMS MONOP OLA R C UT (P OW E R v LOA D ) 350 Power output (watts) 300 H a lf o u tp u t co n trol s e ttin g Full ou tpu t con tro l s e ttin g 250 200 150 100 50 0 0 500 1000 1500 2000 Load resistance (ohms) MONOPOLAR CUT (VOLTAGE v CONTROL) MONOPOLAR CUT (POWER v CONTROL) 1200 250 Peak output voltage (volts) Power output (watts) 300 Load resistance 200 (ohms) 200 150 100 50 0 Open circuit 1000 800 600 400 200 0 0 1 2 3 4 5 6 7 8 9 Output control setting (numerical value up to 10) 10 0 MONOPOLAR CUT (POWER v CONTROL) 2 3 4 5 6 7 8 9 Output control setting (numerical value up to 10) 10 MONOPOLAR CUT (VOLTAGE v CONTROL) 300 Peak output voltage (volts) 1200 250 Power output (watts) 1 Load resistance 200 (ohms) 200 150 100 50 Open circuit 1000 800 600 400 200 0 0 0 25 50 75 100 125 150 175 200 Output control setting (typical pow er) E-IM54f Downloaded from www.Manualslib.com manuals search engine 225 250 0 25 50 75 100 125 150 175 200 Output control setting (typical pow er) 225 250 P25/37 MONOPOLAR BLEND DIAGRAMS MONOP OLA R B LE ND (P OW ER v LOA D ) 300 H alf ou tp ut co ntrol s etting Fu ll o utpu t co n tro l s e ttin g Power output (watts) 250 200 150 100 50 0 0 500 1000 1500 2000 Load resistance (ohms) MONOPOLAR BLEND (POWER v CONTROL) MONOPOLAR BLEND (VOLTAGE v CONTROL) 2500 Peak output voltage (volts) Power output (watts) 200 Load resistance 400 (ohms) 160 120 80 40 0 0 1 2 3 4 5 6 7 8 9 2000 Open circuit 1500 1000 500 0 10 0 1 Output control setting (numerical value up to 10) MONOPOLAR BLEND (POWER v CONTROL) 10 MONOPOLAR BLEND (VOLTAGE v CONTROL) 2500 Load resistance 400 (ohms) 160 120 80 40 0 Peak output voltage (volts) 200 Power output (watts) 2 3 4 5 6 7 8 9 Output control setting (numerical value up to 10) 2000 Open circuit 1500 1000 500 0 0 20 40 60 80 100 120 140 160 180 200 Output control setting (typical power) P26/37 Downloaded from www.Manualslib.com manuals search engine 0 20 40 60 80 100 120 140 160 180 200 Output control setting (typical power) E-IM54f TD830 ELECTROSURGICAL UNIT MONOPOLAR SPECIALIST CUT DIAGRAMS MONOP OLA R S P E C IA LIS T C UT (P OW E R v LOA D ) 350 Power output (watts) 300 H a lf ou tp u t co ntro l s e tting Fu ll o utpu t con tro l s e ttin g 250 200 150 100 50 0 0 500 1000 1500 2000 Load resistance (ohms) MONOPOLAR SPECIALIST CUT (VOLTAGE v CONTROL) MONOPOLAR SPECIALIST CUT (POWER v CONTROL) 1500 350 Load resistance 200 (ohms) Peak output voltage (volts) Power output (watts) 300 250 200 150 100 50 0 0 1 2 3 4 5 6 7 8 9 Open circuit 1200 900 600 300 0 10 0 Output control setting (numerical value up to 10) 2 3 4 5 6 7 8 9 Output control setting (numerical value up to 10) 10 MONOPOLAR SPECIALIST CUT (VOLTAGE v CONTROL) MONOPOLAR SPECIALIST CUT (POWER v CONTROL) 1500 350 300 Load resistance 200 (ohms) Peak output voltage (volts) Power output (watts) 1 250 200 150 100 50 0 Open circuit 1200 900 600 300 0 0 30 60 90 120 150 180 210 240 Output control setting (typical pow er) E-IM54f Downloaded from www.Manualslib.com manuals search engine 270 300 0 30 60 90 120 150 180 210 240 Output control setting (typical pow er) 270 300 P27/37 MONOPOLAR PINPOINT COAG DIAGRAMS MONOP OLA R P INP OINT C OA G (P OW E R v LOA D ) Power output (watts) 200 160 H alf ou tpu t con tro l s e ttin g Fu ll o u tp ut co n tro l s etting 120 80 40 0 0 500 1000 1500 2000 Load resistance (ohms) MONOPOLAR PINPOINT COAG (VOLTAGE v CONTROL) MONOPOLAR PINPOINT COAG (POWER v CONTROL) 2500 Power output (watts) 100 Peak output voltage (volts) 120 Load resistance 400 (ohms) 80 60 40 20 0 Open circuit 2000 1500 1000 500 0 0 1 2 3 4 5 6 7 8 9 10 0 1 MONOPOLAR PINPOINT COAG (POWER v CONTROL) 3 4 5 6 7 8 9 10 MONOPOLAR PINPOINT COAG (VOLTAGE v CONTROL) 2500 100 Peak output voltage (volts) 120 Power output (watts) 2 Output control setting (numerical value up to 10) Output control setting (numerical value up to 10) Load resistance 400 (ohms) 80 60 40 20 0 Open circuit 2000 1500 1000 500 0 0 10 20 30 40 50 60 70 80 Output control setting (typical pow er) P28/37 Downloaded from www.Manualslib.com manuals search engine 90 100 0 10 20 30 40 50 60 70 80 90 100 Output control setting (typical pow er) E-IM54f TD830 ELECTROSURGICAL UNIT MONOPOLAR SPRAY COAG DIAGRAMS M O NO P O LA R S P RA Y C OA G (P OW E R v L O A D ) 80 Power output (watts) 70 60 H a lf o u tp u t co n tro l s e ttin g Fu ll o u tp u t co n tro l s e ttin g 50 40 30 20 10 0 0 500 1000 1500 2000 L o a d re sista n ce (o h ms) MONOPOLAR SPRAY COAG (POWER v CONTROL) MONOPOLAR SPRAY COAG (VOLTAGE v CONTROL) 80 4000 Peak output voltage (volts) Power output (watts) Load resistance 400 (ohms) 60 40 20 0 3500 Open circuit 3000 2500 2000 1500 1000 500 0 0 1 2 3 4 5 6 7 8 9 Output control setting (numerical value up to 10) 10 0 2 3 4 5 6 7 8 9 Output control setting (numerical value up to 10) 10 MONOPOLAR SPRAY COAG (VOLTAGE v CONTROL) MONOPOLAR SPRAY COAG (POWER v CONTROL) 80 4000 Peak output voltage (volts) Load resistance 400 (ohms) Power output (watts) 1 60 40 20 3500 Open circuit 3000 2500 2000 1500 1000 500 0 0 0 7.5 15 22.5 30 37.5 45 52.5 60 Output control setting (typical pow er) E-IM54f Downloaded from www.Manualslib.com manuals search engine 67.5 75 0 7.5 15 22.5 30 37.5 45 52.5 60 Output control setting (typical pow er) 67.5 75 P29/37 BIPOLAR MACRO DIAGRAMS BIPOLAR MAC RO (POWER v LOAD) 50 Power output (watts) 40 Macro - H alf output control s etting Macro - Full output control s etting 30 20 10 0 0 200 400 600 Load resistance (ohm s) BIPOLAR MACRO (POWER v CONTROL) 250 40 Peak output voltage (volts) Power output (watts) 1000 BIPOLAR MACRO (VOLTAGE v CONTROL) 50 Load resistance 100 (ohms) 30 20 10 0 200 Open circuit 150 100 50 0 0 1 2 3 4 5 6 7 8 9 10 0 1 Output control setting (numerical value up to 10) 2 3 4 5 6 7 8 9 10 Output control setting (numerical value up to 10) BIPOLAR MACRO (POWER v CONTROL) BIPOLAR MACRO (VOLTAGE v CONTROL) 50 250 Peak output voltage (volts) Power output (watts) 800 Load resistance 100 (ohms) 40 30 20 10 0 200 Open circuit 150 100 50 0 0 5 10 15 20 25 30 35 Output control setting (typical pow er) P30/37 Downloaded from www.Manualslib.com manuals search engine 40 45 50 0 5 10 15 20 25 30 35 40 45 50 Output control setting (typical pow er) E-IM54f TD830 ELECTROSURGICAL UNIT BIPOLAR MICRO DIAGRAMS B IPOLA R MIC RO (P OW ER v LOAD ) 20 Power output (watts) 16 Micro - H alf output control s etting Micro - Full output control s etting 12 8 4 0 0 200 400 600 Load resistance (ohm s) 16 160 Peak output voltage (volts) Power output (watts) 1000 BIPOLAR MICRO (VOLTAGE v CONTROL) BIPOLAR MICRO (POWER v CONTROL) Load resistance 100 (ohms) 12 8 4 0 120 Open circuit 80 40 0 0 1 2 3 4 5 6 7 8 9 10 0 1 Output control setting (numerical value up to 10) 2 3 4 5 6 7 8 9 10 Output control setting (numerical value up to 10) BIPOLAR MICRO (POWER v CONTROL) BIPOLAR MICRO (VOLTAGE v CONTROL) 16 160 Peak output voltage (volts) Power output (watts) 800 Load resistance 100 (ohms) 12 8 4 0 120 Open circuit 80 40 0 0 1.5 3 4.5 6 7.5 9 10.5 12 Output control setting (typical pow er) E-IM54f Downloaded from www.Manualslib.com manuals search engine 13.5 15 0 1.5 3 4.5 6 7.5 9 10.5 12 13.5 15 Output control setting (typical pow er) P31/37 6.0 CLEANING DISINFECTION AND CARE CLEANING AND DISINFECTION WARNING Non-flammable agents should be used for cleaning and disinfection wherever possible. If flammable agents are used for cleaning or disinfection they should be allowed to evaporate before reusing the electrosurgical unit. CAUTION 6.5 If the unit is to be stored for any length of time it should be plugged into the mains and switched on to charge the internal battery. A charge period of 48 hours is required to restore the battery to its full life of six months (or pro rata, e.g. 8 hours for one month). Also note the required storage conditions detailed in the Technical Data section. Service Equipment Failure 6.6 If the unit or accessories fail any of the checks or operational requirements included in this manual, refer to the Eschmann After Sales Service Department and do not use the equipment. Fuse Failure The electrosurgical unit MUST be disconnected from the mains power supply prior to cleaning. 6.7 Two fuses are located in holders on the rear panel (illustration 11). To gain access to a fuse, firstly ensure that the electrosurgical unit is disconnected from the mains Cleaning the electrosurgical unit power supply and then, using a screwdriver, release the 6.1 The cabinet should be cleaned with a sponge or fuse cap and withdraw the fuse. If necessary, replace the cloth dampened in a mild detergent or soap solution and fuse with a fuse of the correct type and rating (see Technical Data). If repeated blowing of a fuse(s) occurs, either when wiped dry. the unit is switched on or during use, call the Eschmann ♦ Do not use excessive amounts of liquid which could After Sales Service Department (see inside front cover). enter the cabinet and damage internal components. 6.8 A failed fuse is usually an indication of a fault that ♦ The cabinet front panel can be wiped, as it is a has occurred somewhere in the unit. Because of this continuous membrane. Pressure on the touch buttons, Eschmann After Sales Service Department should always with the power supply off, will not cause mode changes. be contacted as soon as a failed fuse is identified or suspected. Disinfecting the electrosurgical unit 6.2 Initially clean the unit as above. For disinfection, Eschmann Equipment recommend primarily the use of alcohol based disinfectants, e.g. 70% isopropanol/water or 70% methyl alcohol/water. The cabinet should be wiped with a sponge or cloth dampened in the above solution and allowed to dry. ♦ Additional Information 6.9 Eschmann Equipment will make available on request circuit diagrams, component parts lists, description and calibration instructions which will assist the users’ appropriately qualified technical personnel to repair those parts of the equipment which are considered ‘user repairable’ by Eschmann Equipment. Do not use excessive amounts of liquid which could enter the cabinet and damage internal components, Maintenance Procedures also ensure the unit is completely dry before reusing 6.10 For complete maintenance procedures see it. Eschmann Service Manual E-SM44. Electrosurgical Accessories ACCESSORIES 6.3 All Eschmann Electrosurgical Accessories (see section 7) come with full “Cleaning, Disinfection and Care 6.11 For a complete range of accessories, and relevant Instructions”, these should be followed at all times. If you ordering information refer to Eschmann product information have misplaced the accessory instructions and require leaflets. another copy please quote Publication Number E-IM50 (Part No. 604802) for all Eschmann accessories. CARE 6.4 Ensure that electrosurgical safety checks and routine maintenance are carried out at regular intervals (at least every six months) and only by Eschmann trained personnel otherwise the warranty for the equipment could be infringed. P32/37 Downloaded from www.Manualslib.com manuals search engine E-IM54f TD830 ELECTROSURGICAL UNIT 7.0 ACCESSORIES 7.1 The accessories in the list below and illustrated where referenced, are all recommended for use with the TD830 electrosurgical unit. (REF = catalogue number) REF Description Illustration No. REF Active electrode silicone cables 83-135-12 3m cable with 4mm plug 83-137-17 5m cable with 4mm plug 83-138-06 3m cable with 8mm plug 83-138-22 5m cable with 8mm plug 7 7 7 7 Active handle for 3/32 inch electrodes 83-140-20 for above cables 3 Active handle with silicone cable 83-141-01 3m cable with 4mm plug 83-141-36 5m cable with 4mm plug 83-141-28 3m cable with 8mm plug 83-141-44 5m cable with 8mm plug 5 5 5 5 Fingerswitches with blade electrode 83-582-06 3m cable and 3 pin plug 83-583-97 5m cable and 3 pin plug 83-585-94 3m cable and 3 pin plug (disposable) Bipolar cable two pin plug 83-041-55 3m cable Bipolar forceps 83-700-01 straight, 200mm 83-703-03 straight, 150mm 83-705-08 straight, 110mm fine tip 83-707-02 straight, 80mm extra fine tip 83-701-09 angled, 200mm 83-704-00 angled, 150mm 83-706-05 angled, 110mm fine tip 83-708-18 angled, 80mm extra fine tip 83-702-06 bayonet, 200mm 83-709-07 bayonet, 250mm fine tip 8 8 8 8 8 8 8 8 8 8 8 Chiropody/Podiatry & Dermatology electrodes 83-583-46 loop, 6.5mm 6 83-584-58 extended small loop, 6mm 6 83-584-61 loop, 10mm 6 83-583-95 angled ball, 2mm * * * * Monopolar footswitches and cable * 83-581-09 yellow (cut/blend), 3m cable 83-581-13 yellow (cut/blend), 5m cable 83-581-17 blue (coag), 3m cable 9 83-581-25 blue (coag), 5m cable 9 83-109-39 twin (cut/blend & coag), 3m cable 9 83-109-47 twin (cut/blend & coag), 5m cable 9 83-109-71 3m extension cable 9 9 Bipolar footswitches 9 83-581-76 electric, 3m cable 9 83-581-68 pneumatic, 5m cable 9 Patient plate cable, 6mm jack plug 9 83-122-93 3m cable, for standard & divided plate 8 8 8 8 8 8 8 Electrodes, needle 83-582-22 medium (pack of 5) 83-583-11 extended, 100mm 83-585-23 standard (disposable) 83-588-77 angled medium (pack of 5) 83-588-80 extended angled, 100mm 8 8 8 8 8 Electrodes, ball 83-583-62 8mm 83-585-45 4mm (disposable) 83-588-85 4.2mm (pack of 5) 8 8 8 Downloaded from www.Manualslib.com manuals search engine Illustration No. Colposcopy electrodes (LLETZ) cone biopsy 83-583-51 large loop, 12mm (pack of 5) 83-584-19 large loop, 15mm (pack of 5) 83-584-27 large loop, 20mm (pack of 5) 83-584-35 large loop, 25mm (pack of 5) 83-584-51 large loop, 1 of each 12,15,20,25mm 83-584-64 small loop, 15mm (pack of 5) 83-590-08 Cartier loop, 6 x 7mm (pack of 5) 83-590-16 Cartier loop, 10 x 12mm (pack of 5) 83-583-89 extended ball, 8mm (100mm) 83-583-90 extended ball, 4.2mm (100mm) 83-585-56 extended 4mm (disposable) Electrodes, blade 83-582-14 fine (pack of 5) 83-582-27 extended, 100mm 83-582-30 extended, 150mm 83-582-49 extended, 200mm 83-585-04 70mm blade (disposable) 83-585-11 150mm blade (disposable) 83-588-69 angled (pack of 5) E-IM54f Description 4 4 4 4 4 4 4 10 10 2 Patient plates and adaptor 83-122-42 adhesive standard (box of 200) 83-122-46 adhesive divided (box of 200) 83-122-49 Flexoplate 200-S, 2.5m cable 83-207-56 for Flexoplate, 6mm jack plug 1 1 1 1 Accessory sets 83-117-22 basic 83-117-30 major * * Other accessories 83-186-38 autoclavable quiver 83-139-41 S-Lead (9mm), Olympus 83-139-49 S-Lead (8mm), Pentax * * * = not illustrated P33/37 83-122-42 83-122-46 83-207-56 83-122-49 1 83-140-20 83-122-93 2 3 83-109-39 83-109-47 4mm 83-581-09 83-581-13 8mm 83-141-01 83-141-36 83-141-28 83-141-44 83-581-17 83-581-25 83-109-71 4 P34/37 Downloaded from www.Manualslib.com manuals search engine 5 E-IM54f TD830 ELECTROSURGICAL UNIT 83-582-06 83-583-97 83-585-94 (Disposable) 6 4mm 7 8mm 83-582-14 83-582-27 83-582-30 83-582-49 83-588-69 83-585-04 (Disposable) 83-585-11 (Disposable) 83-582-22 83-583-11 83-585-23 (Disposable) 83-588-77 83-588-80 8 83-135-12 83-137-17 83-138-06 83-138-22 83-583-62 83-585-45 (Disposable) 83-588-85 83-583-89 83-583-90 83-585-56 (Disposable) 83-590-08 83-590-16 83-583-51 83-584-19 83-584-27 83-584-35 83-584-51 83-584-64 E-IM54f Downloaded from www.Manualslib.com manuals search engine P35/37 83-700-01 83-703-03 83-705-08 83-707-02 83-701-09 83-704-00 83-706-05 83-708-18 83-702-06 83-709-07 9 83-581-76 83-581-68 01 P36/37 Downloaded from www.Manualslib.com manuals search engine E-IM54f Downloaded from www.Manualslib.com manuals search engine Eschmann Equipment, Peter Road, Lancing, West Sussex, BN15 8TJ, England. Tel: +44 (0) 1903 753322. Fax: +44 (0) 1903 875789. www.eschmann.co.uk Downloaded from www.Manualslib.com manuals search engine 1 Power output control, BIPOLAR 2 MICRO BIPOLAR selected, indicator (green LED) 3 Touch button, MICRO BIPOLAR 4 MACRO BIPOLAR selected, indicator (green LED) 5 Touch button, MACRO BIPOLAR 6 BIPOLAR digital power display (green seven segment LEDs) 7 BIPOLAR, power activated indicator (blue LED) 8 MONOPOLAR CUT digital power display (green seven segment LEDs) 9 MONOPOLAR CUT, power activated indicator (yellow LED) 10 MONOPOLAR (normal cut) selected, indicator (green LED) 11 Touch button MONOPOLAR (normal cut) 12 MONOPOLAR (blend) selected, indicator (green LED) 13 Touch button MONOPOLAR (blend) 14 Power output control, MONOPOLAR (cut) 15 Touch button MONOPOLAR (specialist cut) 16 MONOPOLAR (specialist cut) selected, indicator (green LED) 17 MONOPOLAR COAGULATION (pinpoint) selected, indicator (green LED) 18 MONOPOLAR COAGULATION (spray) selected, indicator (bright green LED) 19 Touch button MONOPOLAR COAGULATION (pinpoint) 20 Touch button MONOPOLAR COAGULATION (spray) 21 Power output control MONOPOLAR COAGULATION 22 MONOPOLAR COAGULATION, power activated indicator (blue LED) 23 MONOPOLAR COAGULATION digital power display (green seven segment LEDs) 24 8mm active electrode socket USER 1 only (footswitch activation only) - red 25 Fingerswitch sockets (for use with active electrode socket 26) - black 26 4mm active electrode socket USER 1 (footswitch or fingerswitch activation) - red 27 Fingerswitch sockets (for use with active electrode socket 28) - black 28 4mm active electrode socket USER 2 (fingerswitch activation only ) - red 29 ‘S’ (scope) connector socket (for use with flexible endoscopes) see section 4.55 30 Plate electrode connector socket - black 31 Bipolar output socket - white 32 Monopolar standby selection button (monopolar ‘on/off’ toggle button) 33 Running tone volume control 34 Potential equalization point (see Technical Data section) 35 Digital display range selection button (display in watts if not held pressed) 36 Bipolar pneumatic (white) footswitch socket 37 Bipolar electric (white) footswitch socket 38 Sockets for monopolar footswitches (blue, yellow or combination blue and yellow) 39 Serial number plate 40 Mains cable connection socket 41 Mains fuses (see technical data) 42 Mains power on/off control (with internal green lamp) A1 - PCM alarm activated LED A2 - PAM alarm activated LED A3 - PEM alarm activated LED A4 - PVM alarm activated LED A5 - EPM alarm activated LED A6 - MULTIPLE ACTIVATION alarm LED A7 - Miscellaneous alarm activated LED All alarms are accompanied by an audible two tone warning and the activated digital displays will flash (i.e. if in monopolar standby mode, only the bipolar digital display will flash). Key to illustration 12 E-IM54f Downloaded from www.Manualslib.com manuals search engine 12 P37/37