Search Book TOC Go Back Re-install all of the instruments cosmetics. Select one of the 3 samples and run it into an empty replicate file 10 times to verify instrument precision. Print the replicate file. Have customer run their controls. Print files. Preventive Maintenance 9210165 9210163 3108061 2700400 9130008 9210797 1 3 5 1 1 1 Tubing Kit Lyse Pump Tubing Assy. Black Valve N/C Solenoid Valve Check Valves 0.3 micron Air Filter .100ml Syringe Large Black Valve White Valve White Valve - 5KG CELL-DYN® 1600/1400 Hematology Analyzer Service Manual9211019-July 1993 2808514 9210162 9210166 9210169 1 2 7 2 7-15 The preliminary Q&A sheet, and all printed data before and after the PM, should be attached to the service report and returned to STC. QTY PART NUMBER DESCRIPTION 1 2808541 10ml Syringe 3. 4. 1. 2. 7.16 PM Completion Section 7