Re-install all of the instruments cosmetics.
Select one of the 3 samples and run it into an empty replicate file 10 times to verify
instrument precision.
Print the replicate file.
Have customer run their controls. Print files.
Preventive Maintenance
9210165
9210163
3108061
2700400
9130008
9210797
1
3
5
1
1
1
Tubing Kit
Lyse Pump Tubing Assy.
Black Valve
N/C Solenoid Valve
Check Valves
0.3 micron Air Filter
.100ml Syringe
Large Black Valve
White Valve
White Valve - 5KG
CELL-DYN® 1600/1400 Hematology Analyzer Service Manual9211019-July 1993
2808514
9210162
9210166
9210169
1
2
7
2
7-15
The preliminary Q&A sheet, and all printed data before and after the PM, should be attached to the
service report and returned to STC.
QTY
PART NUMBER
DESCRIPTION
1
2808541
10ml Syringe
3.
4.
1.
2.
7.16 PM Completion