ISE REAGENTS AUH1011 AUH1012 AUH1013 AUH1014 AUH1015 AUH1016 AUH1017 AUH1018 Buffer Mid-Standard Reference Low Serum Std. High Serum Std. High/Low Urine Std. Internal Ref. Sol. Na+/K+ Selectivity Check Sol. 4 x 2000 mL 4 x 2000 mL 4 x 1000 mL 4 x 100 mL 4 x 100 mL 4 x 100 mL 2 x 25 mL 2 x 25 mL Intended Use Reagent for the quantitative determination of Sodium, Potassium and Chloride concentrations in human serum and urine on the Beckman Coulter ISE modules. Summary Electrolytes affect most metabolic processes. They serve to maintain osmotic pressure and hydration of various body fluid compartments, proper body pH and regulation of appropriate heart and muscle functions. Electrolytes are also involved in oxidation-reduction reactions and participate as 1 essential parts or, cofactors, in enzyme reactions. Methodology The determination of electrolytes is one of the most important functions in the clinical laboratory. Methods of determining electrolytes include emission spectrophotometry, flame spectrophotometry, neutron activation analysis, atomic absorption spectroscopy and ion selective electrodes. + + The ISE module for Na , K , and Cl employs crown ether membrane electrodes for sodium and potassium and a molecular oriented PVC membrane for chloride that are specific for each ion of interest in the sample. An electrical potential is developed according to the Nernst Equation for a specific ion. When compared to the Internal Reference Solution, this electrical potential is translated into voltage and then into the ion concentration of the 1 sample. System Information e e For AU400/400 /480, AU600/640/640 /680 and AU2700/5400 Beckman Coulter Analyzers. Contents, Reagents / Standards Concentration of active ingredients: ISE Low Serum Standard + Na 130 mmol/L + K 3.5 mmol/L Cl 85 mmol/L Preservatives ISE Buffer Triethanolamine Preservatives 0.1 mol/L ISE Low/High Urine Standard + Na (Low) 50 mmol/L (High) 200 mmol/L + K (Low) 10 mmol/L (High) 100 mmol/L Cl (Low) 50 mmol/L (High) 180 mmol/L Preservatives ISE Mid-Standard + Na + K Cl Preservatives 4.3 mmol/L 0.13 mmol/L 3.1 mmol/L ISE High Serum Standard + Na + K Cl Preservatives 1.00 mol/L ISE Na Selectivity Check + Na Preservatives 160 mmol/L 6 mmol/L 120 mmol/L + ISE Reference Potassium Chloride Preservatives ISE Internal Reference Potassium Chloride Silver Chloride Preservatives 150 mmol/L + 3.3 mol/L Saturated ISE K Selectivity Check + K Preservatives 5 mmol/L Precautions 1. For in vitro diagnostic use. 2. Do not ingest. Harmful if swallowed. Preparation of Reagents The reagents are ready to use. No preparation is required. Storage and Stability 1. The unopened reagents are stable until the expiration date printed on the label when stored at 2 – 25°C. AUH1017 is stored at 15 - 25°C. 2. AUH1011, AUH1012, and AUH1013 are stable for 90 days when opened and stored in the ISE reagent compartment of the analyzer. 3. After opening AUH1014, AUH1015, AUH1016, and AUH1018 may be stored at 2 – 25°C for up to 90 days, provided the cap is replaced immediately after each use. After opening, AUH1017 may be stored at 15 - 25°C for up to 90 days. BAAUH1011.01 2009-08 ISE ISE Reagents Indications of Deterioration Turbidity or precipitation in the unopened liquids and working reagents may indicate decomposition and warrant discontinuance of use. Refer to the ISE section of the respective AU analyzer’s User’s Guide for additional information. Specimen Collection And Preparation Serum, free from hemolysis, is the recommended specimen. Sample Stability + Separate serum from blood cells as soon as possible. Avoid hemolysis since it can lead to falsely elevated K values. If plasma must be used, the recommended anticoagulants are lithium heparin and ammonium heparin. Urine samples should be collected in a clean, leak-proof container and should not be acidified. If transport is delayed, specimens should be kept refrigerated at 2 - 8°C. The recommended collection times for urine 2 specimens, to determine the analyte of interest, are as follows. Sodium: Potassium: Chloride: 24 hour collection 24 hour collection 24 hour collection Sample Storage Sodium and potassium are stable in serum for at least one week when stored at 2 - 8°C. Chloride is stable in serum for one week when stored at 2 2 - 30°C. Store the sample in a stoppered tube if analysis is delayed. Urine specimens should be stored at 2 - 8°C. Interfering Substances Certain anticoagulants, preservatives, drugs, and organophilic compounds may affect electrolyte determinations. For further information on 3 interfering substances, refer to Young for a compilation of reported interferences with this test. Visually turbid urine specimens should be centrifuged 2 prior to analysis. Grossly lipemic samples may show an inappropriate decrease in sodium, potassium, and chloride results due to volume displacement. Such samples should be ultracentrifuged and the analysis performed on the infranatant (middle clear layer). Procedure Materials Provided ISE Buffer, Mid-Standard, Reference, High Serum Standard, Low Serum Standard, High/Low Urine Standard, Internal Reference Solution, and + + Na /K Selectivity Check Solution. Suggested Analytical Parameters Refer to the ISE Section located in the respective AU analyzer’s User’s Guide for a complete listing of test parameters and procedure. Stability of Final Reaction Mixture The Beckman Coulter AU analyzer automatically computes every determination at the same interval at 37°C. Calibration Refer to the ISE section of the respective AU analyzer’s User’s Guide for a complete listing of the recommended calibration procedures. This procedure should be used to calibrate daily. Quality Control During operation of the Beckman Coulter AU analyzer at least two levels of an appropriate control material should be tested a minimum of once a day. In addition, controls should be performed after calibration, with each new lot of reagent, and after specific maintenance or troubleshooting steps described in the appropriate Beckman Coulter AU analyzer User’s Guide. Quality control testing should be performed in accordance with regulatory requirements and each laboratory’s standard procedure. Appropriate qualified urine controls should be used during urine analysis. Results Results are automatically printed out for each sample in mEq/L. Dynamic Range The ISE Module procedures are linear in serum, plasma or urine samples as follows: Serum Urine + + Na 50 – 200 mEq/L Na 10 – 400 mEq/L + + K 1.0 – 10.0 mEq/L K 2.0 – 200.0 mEq/L 50 – 200 mEq/L Cl 15 – 400 mEq/L Cl Samples exceeding the dynamic range of the assay should be diluted with deionized water and reassayed. The results obtained must be multiplied by the dilution factor to obtain the correct concentration for the undiluted sample. Expected values1 Serum: Urine: + Na 136 - 145 mEq/L + Na 40 - 220 mEq/day + K 3.5 - 5.1 mEq/L + K 25 - 125 mEq/day - Cl 98 - 107 mEq/L Cl 110 - 250 mEq/day Expected values may vary with age, sex, diet and geographical location. Good laboratory practice dictates that each laboratory determine its own expected values. ISE BAAUH1011.01 2009-08 ISE Reagents Specific Performance Characteristics The following data was obtained using the ISE reagents on AU analyzer’s ISE modules according to established procedures. Results obtained in individual laboratories may differ. 5 Precision 4 Estimates of precision, based on CLSI recommendations , are consistent with typical performance. The within run precision is less than 3% CV and total precision is less than 5% CV. Assays of control sera and pooled urine samples were carried out and the data reduced following the CLSI guidelines referenced above. Urine Level 1 Level 2 N = 100 Mean + + Na K 88 32.1 182.3 85.8 Level 1 Level 2 N = 100 Mean + + Na K 88 32.1 182.3 85.8 Within run - Cl 111.3 238.8 SD + K 0.3 0.5 + Na 0.6 0.7 - + Cl 0.7 0.9 Na 0.7 0.4 CV% + K 0.8 0.6 Cl 0.6 0.4 CV% + K 1.0 1.1 Cl 0.8 0.6 CV% + K 0.6 0.8 Cl 0.5 0.5 CV% + K 0.7 0.9 Cl 0.6 0.6 - Total - SD + K 0.3 0.9 + Cl 111.3 238.8 Na 0.8 1.5 - + Cl 0.9 1.4 Na 0.9 0.8 - Serum Level 1 Level 2 N = 100 Mean + + Na K 124.4 2.9 143 5.7 Level 1 Level 2 N = 100 Mean + + Na K 124.4 2.9 143 5.7 Within run - SD + K 0.2 0.04 + Cl 85.9 112.3 Na 0.5 0.5 - + Cl 0.4 0.5 Na 0.4 0.3 - Total - Cl 85.9 112.3 + Na 0.8 0.8 SD + K 0.02 0.05 - Cl 0.6 0.6 + Na 0.6 0.5 - 5 Method Comparison A comparison of the Beckman Coulter AU Analyzers ISE was conducted with patient serum samples. The table below demonstrates representative performance on AU analyzers. Serum Comparison e AU640/640 vs AU600 + N r slope intercept + Na 230 0.998 1.01 0.3 K 230 0.999 0.99 0.12 - Cl 230 0.999 1.01 0.7 Urine Comparison Beckman Coulter comparisons were conducted utilizing urine patient samples. The table below demonstrates representative performance on AU analyzers. e N r Slope Intercept AU640/640 /680 vs AU600 + + Na K 184 184 0.999 0.999 1.01 0.99 1.5 4.7 - Cl 184 0.999 0.99 -0.4 References 1. 2. 3. 4. 5. Tietz, N.W., editor, Fundamentals of Clinical Chemistry, 3rd Edition, W.B.Saunders 1987. Pesce, A.J., Kaplan, L.A., editors, Methods in Clinical Chemistry, C.V. Mosby Company, 1996. Young, D.S., Effects of Drugs on Clinical Laboratory Tests, 5th edition, AACC Press, 2000. CLSI/NCCLS Evaluation Protocol EP5-T2, 1992. Data is on file for specific AU analyzers. Manufactured by: Beckman Coulter, Inc., 250 S. Kraemer Blvd. Brea, CA 92821, USA BAAUH1011.01 2009-08 ISE