THE COPPERBELT UNIVERSITY SCHOOL OF ENGINEERING ELECTRICAL ENGINEERING DEPARTMENT EE/EG 232 ENGINEERING MEASUREMENT SYSTEMS LECTURE NOTES CONTACT HOURS:2 hours/ week The course is divided into two parts, namely the Electrical and the Mechanical. The electrical part will be taught for two terms by Mr A Tambatamba and the Mechanical part in the third term by Mr E Langi. Course content(Electrical) Module 1 Measurements and errors: definition; accuracy and precision; significant figures; types of errors statistical analysis, common symbols used in metrology. Module 2 Measuring instruments: classification; torque generation, controlling and damping; PMMC, MI and electrodynamometer ammeters and voltmeters; electrodynamometer, induction types in wattmeters (1 and 3 phase); induction energy meter, compensation, creeping errors testing; frequency meter: vibration reed type, electrical resonance. Module 3 Instrument transformers: voltage and current transformers; ration and phase angle errors, phasor diagram, methods of reducing error; testing and applications Galvanometers: principles of operation; performance equations of the D’ Arsonval galvanometer, vibration and ballistic galvanometers Potentiometer: DC potentiometer, Crompton potentiometer, construction, standardisation, application, AC potentiometer, Drysdale polar potentiometer,construction, standardisation, application Module 4 AC/DC bridges: Genera balance equations of Wheatsone, Kelvin, Maxwell, Hay, Schering, and Wein bridges. Transducers:Pizo-Electric, Thermistors, Thermorecouple, Torque meter Module 5 Cathode ray Oscilloscope: construction, sweep generation, application of CRG in measurements of frequency, phase, amplitude and rise in time of a pulse. Digital multi-meter: Block diagram, principle of operation, accuracy of measurements. Electronic voltmeter: transistor voltmeter, block diagram, principle of operation. Other types of electronic voltmeter, digital frequency metter : construction and principle of operation. 1 Module 1 Introduction Metrology (from Greek words metron – to measure and logy- science) is the science of measurements. Measurements are important to humans as it helps to determine unknown quantities that they can with their bare hands. A measurement has been developing along side with advancement in engineering. The first in the world institutions set up to study measurements was established in German in 1887. It is said to be partly responsible for the abrupt increase in industrialisation of German, which was the first one in the world. Other countries which are industrialised today follow the German example. The main purpose for of metrology is best described by the famous Italian scientist Galileo Galilei, who said ‘measure everything that can be measured and try to make measurable what is not yet measurable’. Measurements are useful in determining the value of a measured quantity. Quantities can include electric current, distance, mass time and others. A measured value is the reading obtained from a measuring instruments, with a degree of error. A true value is simply the rated value. The International System of Units (SI) SI is the modern and most widely used system of units in measurements. It is made up of seven base units which include Length measured in meters (m), mass- Kilogram (kg), time- seconds (s), electric current- Ampere (A), temperature- Kelvin (K), luminous (light) intensity –Candela (Cd) and amount of substance- mole (mol). SI is the foundation of modern metrology. It was found in 1960. Apart from the seven base units there are 23 other units which are derived using formulas. Electric current is the only unit for electrical engineers in the base units. Other electrical engineering units like voltage, resistance, power and others are derived from formulas using electric current. PREFIXES Since the presentation of the results in the initial units of physical quantities is sometimes unsuitable, the SI system defines the use of SI prefixes as shown in Table 1 below. The prefixes hecto, deka, deci, and centi are not used in electrical engineering. The names of SI units form the name of the starting unit by addingSI prefixes. Thus, it is correct to say 10 teraohms, but not 10 deciohms. Using only a single prefix inthe SI system is also allowed. Therefore, 10 μμW is not allowed, as it must be 10 pW. Table 1 Prefix Symbol Exponent Number 12 Tera T 1 000 000 000 000 10 9 Giga G 1 000 000 000 10 Mega M 1 000 000 106 3 Kilo k 1 000 10 0 Base base 1 10 Deci d 0.1 10−1 −2 centi c 0.01 10 −3 Milli M 0.001 10 Micro 0.000001 μ 10−6 −9 Nano n 0.000000001 10 2 pico p Other prefixes are included in the table 2 below, 10−12 0.000000000001 Table 2 The base is the unit used to measure some quantity, eg amps, watts, meter, second etc. From table 1 it can be seen that it is convenient to express some numbers in their exponent form i.e10x. For example, 2.5 Megagram=2.5 Mg=2.5x106g=2500 000g. Prefix conversion Example 1, How many dg are there in 2.6 kg? Solution Approach (steps) 1. Write the amount and units and given in question 2. Multiply the amount and units given in1 by a fraction having the given units at the bottom and the require units at the top, 3. Then assign the bigger prefix from the fraction with 1 and the small one with 4. Determine the exponent of the figure 10 assigned to smaller prefix of the fraction in 3 by subtraction the exponent of the smaller prefix from that of a bigger one. 5. After that cancel the like terms and present the final answer in simple exponent form as shown below. 104 𝑑𝑔 2.6 𝑘𝑔 ∙ = 2.6 ∙ 104 𝑑𝑔 1 𝑘𝑔 Finding exponent of dg, exponent of kg =3 that of dg—1, therefore 3-(-1)=4, when simplifying the expression by cancelling the like units and prefixes, we get 𝟐. 𝟔 ∙ 𝟏𝟎𝟒 𝒅𝒈 Example 2, How many cm are there in 18.2 nm? Solution The approach is similar to the previous example 18.2 𝑛𝑚 ∙ 1 𝑐𝑚 = 18.2 ∙ 10−7 𝑐𝑚 107 𝑛𝑚 3 Finding exponent of nm.since exponent of cm =-2 that of nm=-9, therefore -2-(-9)=7, when simplifying the expression by cancelling the like units and prefixes, we get 𝟏𝟖. 𝟐 ∙ 𝟏𝟎𝟕 𝒅𝒈 Conversion of quantities that squared and cubed For quantities like length that get squared and cubed to give area and volume have a slightly different way of converting their prefixes. Example 3, How manym2 are there in 1 km2? Solution The approach is similar to the previous example; apart from the fact that the exponent will have to be multiplied by 2 or 3 with we are dealing with squared or cubed quantities respectively. 103(2) 𝑚2 = 1 ∙ 106 𝑚2 1 𝑘𝑚2 Finding exponent of 𝑚2 . Since exponent of km =3 that of m=0, therefore 3-(0)=3, the exponent is then multiplied by 2 as the units are squared. Then after simplifying the expression by cancelling the like units and prefixes, we get 1 ∙ 106 𝑚2 1 𝑘𝑚2 ∙ Measurements and errors Most important terms used in metrology include: i) Measurement: This is the process of using an instrument to determine the magnitude or value of a quantity, variable or parameter. ii) Electrical/ electronic measuring instrument: this is a device based on electrical and electronic principle for determining the value or magnitude of a quantity, variable or parameter. iii) Instrument accuracy: this is the closes of an instrument reading to the true value of the variable being measured. iv) Instrument precision: This is the measure of the degree to which successive measurements taken under the same conditions.differ from each other. v) Instrument sensitivity: this is the ratio of the output signal or response of the instrument to change of an input or measured variable. vi) Instrument resolution: this is the smallest change in measured value the instrument will respond to. vii) Instrument Error: this the deviation ( difference) of the measured value from a true value viii) Measurement reproducibility: is the closeness of individual measurements of the same quantity that are measured in changed circumstances (different measurers in different laboratories, using other measurement methods, instruments, places, and conditions, etc.). ix) True value: the arithmetic mean of an infinite set of measurements. Note that its practically impossible to make infinity measurements. x) Confidence limits: the range within which its probable to find the true value of a seasurement. 4 Error Minimisation Techniques Several techniques may be used to minimize the effects of errors. For example: 1. Making precision measurements, 2. It is advisable to record a series of observations rather than rely on one observation. 3. Alternate methods of measurement, as well as the use of different instruments to perform the same experiment, 4. Provide a good technique for increasing accuracy. Although these techniques tend to increase the precision of measurement by reducing environmental or random error, they cannot account for instrumental error. Error has been defined above as the deviation of the measured value from a true value. Mathematically it can be define: 𝐸𝑟𝑟𝑜𝑟 = 𝛿 = 𝑀𝑒𝑎𝑠𝑢𝑟𝑒𝑑 𝑣𝑎𝑙𝑢𝑒(𝐴𝑚 ) − 𝑇𝑟𝑢𝑒 𝑣𝑎𝑙𝑢𝑒(𝐴𝑇 ) If 𝐴𝑚 > 𝐴𝑇 𝑡ℎ𝑒𝑛 𝐸𝑟𝑟𝑜𝑟 𝑖𝑠 𝑃𝑜𝑠𝑖𝑡𝑖𝑣𝑒, 𝑤ℎ𝑒𝑟𝑒 𝑎𝑠 𝑖𝑓 𝐴𝑚 < 𝐴𝑇 𝑡ℎ𝑒𝑛 𝐸𝑟𝑟𝑜𝑟 𝑖𝑠 𝑛𝑒𝑔𝑎𝑡𝑖𝑣𝑒 Example A motor is rated 3000 rpm at full speed.When correctly connected to the power supply at full speed, the reading of the torque meter is used to measure the rpm was 3002 rpm. Determine the error. Solution 𝐸𝑟𝑟𝑜𝑟 = 𝛿 = 𝑀𝑒𝑎𝑠𝑢𝑟𝑒𝑑 𝑣𝑎𝑙𝑢𝑒(𝐴𝑚 ) − 𝑇𝑟𝑢𝑒 𝑣𝑎𝑙𝑢𝑒(𝐴𝑇 ) = 3002 − 3000 = +𝟐 𝒓𝒑𝒎 Note that 𝛿 is also known as absolute error. Errors in measurements are usually expressed in percentage (%). The quality of an instrument in terms of accuracy is determined by its percentage Relative Static Error (RSE) (also called the limiting Error). 𝐴𝑚 − 𝐴𝑇 𝛿 𝐴𝑚 ∙ 100% = ∙ 100% = ( − 1) ∙ 100% 𝐴𝑇 𝐴𝑇 𝐴𝑇 𝐴𝑚 −𝐴𝑇 is also referred to as relative error 𝐴 %𝑅𝑆𝐸 = 𝑇 %𝑅𝑆𝐸is also referred to as Percentage Error Example Two instruments A and B had the information as shown below: 𝐼𝑛𝑠𝑡𝑟𝑢𝑚𝑒𝑛𝑡 𝐴: 𝛿𝐴 = 1 𝐴, 𝐴𝑇 = 2 𝐴 𝐼𝑛𝑠𝑡𝑟𝑢𝑚𝑒𝑛𝑡 𝐵 ∶ 𝛿𝐵 = 10 𝐴, 𝐴𝑇 = 1000 𝐴 Determine which instrument was more quality. Solution 𝛿 1 𝐼𝑛𝑠𝑡𝑟𝑢𝑚𝑒𝑛𝑡 𝐴: %𝑅𝑆𝐸 = ∙ 100% = ∙ 100% = 50%, 𝐴𝑇 2 𝐴𝑐𝑐𝑢𝑟𝑎𝑐𝑦 = 100% − %𝑅𝑆𝐸 = 100% − 50% = 50% 𝛿 10 𝐼𝑛𝑠𝑡𝑟𝑢𝑚𝑒𝑛𝑡 𝐵: %𝑅𝑆𝐸 = ∙ 100% = ∙ 100% = 1%, 𝐴𝑇 1000 𝐴𝑐𝑐𝑢𝑟𝑎𝑐𝑦 = 100% − %𝑅𝑆𝐸 = 100% − 1% = 99% From the results obtained, it can be concluded that instrument B is more quality as it has a higher accuracy low error. 5 Accuracy and Precision As mentioned above, Accuracy refers to the degree of closeness or conformity to the true value of the quantity under measurement. Precision refers to the degree of agreement within a group of measurements or instruments. To demonstrate the difference between accuracy and precision, two voltmeters of the same make and model may be compared. Both meters have knife-edged pointers and mirror-backed scales to avoid parallax, and they have carefully calibrated scales. They may therefore be read to the same precision. If the value of the series resistance in one meter changes considerably, its readings may be in error by a fairly large amount. Therefore the accuracy of the two meters may be quite different. (To determine which meter is in error, a comparison measurement with a standard meter should be made). Precision is composed of two characteristics: conformity (closeness) and the number of significant figures to which a measurement can be made. Consider, for example, that a resistor, whose true resistance is 1,273,672 Ω, is measured by an ohmmeter which consistently and repeatedly indicates 1.3 MΩ. But can the observer “read” the true value from the scale? His estimates from the scale reading consistently yield a value of 1.3 MΩ. This is as close to the true value as he can read the scale by estimation. Although there are no deviations from the observed value, the error created by the limitation of the scale reading is a precision error. The example illustrates that conformity is a necessary, but not sufficient condition for precision because of the lack of significant figures obtained. Similarly, precision is a necessary, but not sufficient condition for accuracy. Too often the beginning student is inclined to accept instrument readings at face value. He is not aware that the accuracy of a reading is not guaranteed by its precision. In fact, good measurement techniques demands continuous scepticism as to the accuracy of the results. In critical work, good practice dictates that the observer make an independent set of measurements, using different instruments or different measurement techniques, not subject to the same systematic errors. He must also make sure that the instruments function properly and are calibrated against a known standard, and that no outside influence affects the accuracy of his measurements. Significant Figures An indication of precision of the measurement is obtained from the number of significant figures in which the result is expressed. Significant figures convey actual information regarding the magnitude and the measurement precision of a quantity. The more significant figures, the greater the precision of measurement. For example, if a resistor is specified as having a resistance of 68 Ω, its resistance should be closer to 68 Ω than to 67 Ω or 69 Ω. If the value of the resistor is described as 68.0 Ω, it means that its resistance is closer to 68.0 Ω than it is to 67.9 Ω or 68.1 Ω. In 68 Ω there are two significant figures; in 68.0 Ω there are three. The latter, with more significant figures, expresses a measurement of greater precision than the former. Often, however, the total number of digits may not represent measurement precision. Frequently, large numbers with zeros before a decimal point are used for approximate populations or amounts of money. For example, the population of a city is reported in six figures as 380,000. This may imply that the true value of the population lies between 379,999 and 380,001, which is six significant figures. What is meant, however, is that the population is closer to 380,000 than 6 370,000 or 390,000. Since in this case the population can be reported only to two significant figures, how can large numbers be expressed? A more technically correct notation uses powers of ten, 38 × 104 or 3.8 × 105. This indicates that the population figures is only accurate to two significant figures. Uncertainty caused by zeros to the left of the decimal point is therefore usually resolved by scientific notation using powers of ten. Reference to the velocity of light as 186,000 mi/s, for example, would cause no misunderstanding to anyone with a technical background. But 1.86 × 105 mi/s leave no confusion. It is customary to record a measurement with all the digits of which we are sure nearest to the true value. For example, in reading a voltmeter, the voltage may be read as 117.1 V. This simply indicates that the voltage, read by the observer to best estimation, is closer to 117.1 V than to 117.0 V or 117.2 V. Another way of expressing this result indicates the range of possible error. The voltage may be expressed as 117.1 ± 0.05 V, indicating that the value of the voltage lies between 117.05 V and 117.15 V. When a number of independent measurements are taken in an effort to obtain the best possible answer (closest to the true value), the result is usually expressed as the arithmetic mean of all the readings, with the range of possible error as the largest deviation from that mean. This is illustrated in Example 1. Example 1 A set of independent voltage measurements taken by four observers was recorded as 117.02 V, 117.11 V, 117.08 V, and 117.03 V. Calculate: 1. The average voltage, 2. The range of error. Solution 1. Average Voltage 𝑉1 + 𝑉2 + 𝑉3 + 𝑉4 117.02 + 117.11 + 117.08 + 117.03 𝑉𝑎𝑣 = = = 117.06 𝑉 𝑛 4 2. Range Range 𝑉𝑚𝑎𝑥 − 𝑉𝑎𝑣 = 117.11 − 117.06 = 0.05 but also 𝑉𝑎𝑣 − 𝑉𝑚𝑖𝑛 = 117.06 − 117.02 = 0.04 0.05+0.04 The average range of error therefore equals = ±0.045 = ±0.05 2 When two or more measurements with different degrees of accuracy are added, the result is only as accurate as the least square measurement. Suppose that two resistances are added in series as in Example 2. Example 2 Two resistors R1 and R2 are connected in series. Individual resistance measurements, using a Wheatstone bridge, give R1 = 18.7Ω and R2 = 3.624Ω. Calculate the total resistance to the appropriate number of significant figures. Solution R1 = 18.7 Ω (three significant figures) R2 = 3.624 Ω (four significant figures) RT = R1 + R2 = 22.324 Ω (five significant figures) = 22.3 Ω 7 The doubtful figures are written in italics to indicate that in the addition of R1 and R2 the last three digits of the sum are doubtful figures. There is no value whatsoever in retaining the last two digits (the 2 and the 4) because one of the resistances is accurate only to three significant figures or tenths of an ohm. The result should therefore also be reduced to three significant figures or the nearest tenth, i.e., 22.3 Ω. The number of significant figures in multiplication may increase rapidly, but again only the appropriate figures are retained in the answer, as shown in Example 3. Example 3 In calculating voltage drop, a current of 3.18 A is recorded in a resistance of 35.68 Ω. Calculate the voltage drop across the resistor to the appropriate number of significant figures. Solution V = IR = 3.18 × 35.68 = 113.4624 = 113V Since there are three significant figures involved in the multiplication, the answer can be written only to a maximum of three significant figures. In Example 3, the current, I, has three significant figures and R has four; and the result of the multiplication has only three significant figures. This illustrates that the answer cannot be known to the accuracy greater than the least poorly defined of the factors. Note also that if extra digits accumulate in the answer, they should be discarded or rounded off. In the usual practice, if the (least significant) digit in the first place to be discarded is less than five, it and the following digits are dropped from the answer. This was done in Example 3. If the digit in the first place to be discarded is five or greater, the previous digit is increased by one. For three-digit precision, therefore, 113.46 should be rounded off to 113; and 113.74 to 114. Addition of figures with a range of doubt is illustrated in Example 4. Example 4 Add 826 ± 5 to 628 ± 3 Solution N1 = 826 ± 5 (= ±0.605%) N2 = 628 ± 3 (= ±0.477%) Sum = 1454 ± 8 (= ±0.55%) Note in Example 4 that the doubtful parts are added, since the ± sign means that one number may be high and the other low. The worst possible combination of range of doubt should be taken in the answer. The percentage doubt in the original figures N1 and N2 does not differ greatly from the percentage doubt in the final result. If the same two numbers are subtracted, as in Example 5, there is an interesting comparison between addition and subtraction with respect to the range of doubt. Example 5 Subtract 628 ± 3 from 826 ± 5 and express the range of doubt in the answer as a percentage. Solution N1 = 826 ± 5 (= ±0.605%) N2 = 628 ± 3 (= ±0.477%) Difference = 198 ± 8 (= ±4.04%) 8 Again in Example 5, the doubtful parts are added for the same reason as in Example 4. Comparing the results of addition and subtraction of the same numbers in Example 4 and 5, note that the precision of the results, when expressed in percentages, differs greatly. The final result after subtraction shows a large increase in percentage doubt compared to the percentage doubt after addition. The percentage doubt increases even more when the difference between the numbers is relatively small. Consider the case illustrated in Example 6. Example 6 Subtract 437 ± 4 from 462 ± 4 and express the range of doubt in the answer as a percentage. Solution N1 = 462 ± 4 (= ±0.87%) N2 = 437 ± 4 (= ±0.92%) Difference = 25 ± 8 (= ±32%) Example 6 illustrates clearly that one should avoid measurement techniques depending on subtraction of experimental results because the range of doubt in the final result may be greatly increased. Types of Error No measurement can be made with perfect accuracy, but it is important to find out what the accuracy actually is and how different errors have entered into the measurement. A study of errors is a first step in finding ways to reduce them. Such a study also allows us to determine the accuracy of the final test result. Errors may come from different sources and are usually classified under three main headings: Gross Errors -Largely human errors, among them misreading of instruments, incorrect adjustment and improper application of instruments, and computational mistakes. Systematic Errors -Shortcomings of the instruments, such as defective or worn parts, and effects of the environment on the equipment or the user. Random Errors -Those due to causes that cannot be directly established because of random variations in the parameter or the system of measurement. Each of these classes of errors will be discussed briefly and some methods will be suggested for their reduction or elimination. Gross (grave) Errors This class of errors mainly covers human mistakes in reading or using instruments and in recording and calculating measurement results. As long as human beings are involved, some gross errors will inevitably be committed. Although complete elimination of gross errors is probably impossible, one should try to anticipate and correct them. Some gross errors are easily detected; others may be very elusive. Grave errors are avoided with good knowledge and attention during the measurements, and a proper selection of measurement equipment and the measurement procedure. It is always useful to evaluate the approximate value of the measurand before the measurements. One common gross error, frequently committed by beginners in measurement work, involves the improper use of an instrument. In general, indicating instruments change conditions to some extent when connected into a complete circuit, so that the measured quantity is altered by the method employed. For example, a well-calibrated voltmeter may give a misleading reading when connected across two points in a high resistance circuit (Example 7). The same voltmeter, when connected in a low-resistance circuit, may give a more dependable reading (Example 8). These examples illustrate that the voltmeter has a “loading effect” on the circuit, altering the original situation by the measurement process. 9 Example 7 A voltmeter, having a sensitivity of 1,000 Ω/V, reads 100 V on its 150-V scale when connected across an unknown resistor in series with a milliammeter. When the milliammeter reads 5 mA, calculate: 1. Apparent resistance of the unknown resistor, 2. Actual resistance of the unknown resistor, 3. Error due to the loading effect of the voltmeter. Solution 1. The total circuit resistance equals 𝑅𝑇 = 𝑉𝑇 100 𝑉 = = 20 𝑘Ω 𝐼𝑇 5 𝑚𝐴 Neglecting the resistance of the milliammeter, the value of the unknown resistor is RX = 20 kΩ 2. The voltmeter resistance equals 𝑅𝑉 = 1000 Ω ∙ 150 𝑉 = 150 𝑘Ω 𝑉 Since the voltmeter is in parallel with the unknown resistance, we can write 𝑅𝑇 ∙ 𝑅𝑉 20 ∙ 150 𝑅𝑋 = = = 23.05 𝑘Ω 𝑅𝑉 − 𝑅𝑇 150 − 20 3. Error % 𝐸𝑟𝑟𝑜𝑟 = 𝐴𝑐𝑡𝑢𝑎𝑙 − 𝑎𝑝𝑝𝑟𝑒𝑛𝑡 23.05 − 20 ∙ 100% = ∙ 100% = 13.23% 𝐴𝑐𝑡𝑢𝑎𝑙 23.05 Example 8 Repeat Example 7 if the milliammeter reads 800 mA and the voltmeter reads 40 V on its 150-V scale. Solution 1. The total circuit resistance equals 𝑅𝑇 = 𝑉𝑇 40 𝑉 = = 50 Ω 𝐼𝑇 0.8 𝐴 2. The voltmeter resistance equal 𝑅𝑋 = 1000 Ω ∙ 150 𝑉 = 150 𝑘Ω 𝑉 Since the voltmeter is in parallel with the unknown resistance, we can write 𝑅𝑇 ∙ 𝑅𝑉 50 ∙ 150 𝑅𝑋 = = = 75 Ω 𝑅𝑉 − 𝑅𝑇 150 − 50 3. Error % 𝐸𝑟𝑟𝑜𝑟 = 𝐴𝑐𝑡𝑢𝑎𝑙 − 𝐴𝑝𝑝𝑎𝑟𝑒𝑛𝑡 75 − 50 ∙ 100% = ∙ 100% = 33.33 % 𝐴𝑐𝑡𝑢𝑎𝑙 75 Errors caused by the loading effect of the voltmeter can be avoided by using it intelligently. For example, a low-resistance voltmeter should not be used to measure voltages in a vacuum tube 10 amplifier. In this particular measurement, a high-input impedance voltmeter (such as VTVM or TVM) is required. A large number of gross errors can be attributed to carelessness or bad habits, such as improper reading of an instrument, recording the result differently from the actual reading taken, or adjusting the instrument incorrectly. Consider the case in which a multi-range voltmeter uses a single set of scale markings with different number designations for the various voltage ranges. It is easy to use a scale which does not correspond to the setting of the range selector of the voltmeter. A gross error may also occur when the instrument is not set to zero before the measurement is taken; then all the readings are off. Errors like these cannot be treated mathematically. They can be avoided only by taking care in reading and recording the measurement data. Good practice requires making more than one reading of the same quantity, preferably by a different observer. Never place complete dependence on one reading but take at least three separate readings, preferably under conditions in which instruments are switched off-on. Systematic Errors This type of errors is usually divided into two different categories: 1. Instrumental errors, defined as shortcomings of the instrument; 2. Environmental errors, due to external conditions affecting the measurement. Instrumental errors are errors inherent in measuring instruments because of their mechanical structure. For example, in the d’Arsonval movement, friction in bearings of various moving components may cause incorrect readings. Irregular spring tension, stretching of the spring, or reduction in tension due to improper handling or overloading of the instrument will result in errors. Other instrumental errors are calibration errors, causing the instrument to read high or low along its entire scale. (Failure to set the instrument to zero before making a measurement has a similar effect.) There are many kinds of instrumental errors, depending on the type of instrument used. The experimenter should always take precautions to insure that the instrument he is using is operating properly and does not contribute excessive errors for the purpose at hand. Faults in instruments may be detected by checking for erratic behaviour, and stability and reproducibility of results. A quick and easy way to check an instrument is to compare it to another with the same characteristics or to one that is known to be more accurate. Instrumental errors may be avoided by: 1. selecting a suitable instrument for a particular measurement application; 2. applying correction factors after determining the amount of instrumental error; 3. calibrating the instrument against a standard. Environmental errors are due to conditions external to the measuring device, including conditions in the area surrounding the instrument, such as the effects of changes in temperature, humidity, barometric pressure, or of magnetic or electrostatic fields. Thus a change in ambient temperature at which the instrument is used causes a change in the elastic properties of the spring in a moving-coil mechanism and so affects the reading of the instrument. Corrective measures to reduce these effects include air conditioning, hermetically sealing certain components in the instrument, use of magnetic shields, and the like. Systematic errors can also be subdivided into static or dynamic errors. Static errors are caused by limitations of the measuring device or the physical laws governing its behavior. A static error is 11 introduced in a micrometer when excessive pressure is applied in torquing the shaft. Dynamic errors are caused by the instrument’s not responding fast enough to follow the changes in a measured variable. Random Errors Random errors result from small and variant changes that occur in the standards, measures, measurement laboratory, and environment. These errors can cause a large number of different errors in each individual measurement and each time have a different size and sign, causing measurement results to scatter. If the same quantity is measured several times in a row under the same external conditions and with the same instrument, each time the results will scatter around some value due to random errors that change the size and sign in each measurement. These random errors cannot be solved by making corrections, as for some systematic errors. Since all measurements are carried out under the same conditions, all results have the same weight. Thus, the most probable value of the measured quantity is the arithmetic mean of individual results. If the measurements are repeated n times, and individual results are then the arithmetic mean of the individual results is:? Arithmetic Mean The most probable value of a measured variable is the arithmetic mean of the number of readings taken. The best approximation will be made when the number of readings of the same quantity is very large. Theoretically, an infinite number of readings would give the best result, although in practice, only a finite number of measurements can be made. The arithmetic mean is given by the following expression: 𝑥1 + 𝑥2 + 𝑥3 + 𝑥4 + ⋯ + 𝑥𝑛 ∑ 𝑥𝑖 ẍ= = 𝑛 𝑛 Where ẍ=arithmetic mean x1, x2, x3, x4,…, xn=readings take n=number of readings Example 1 showed how the arithmetic mean is used. Deviation from the Mean Deviation is the departure of a given reading from the arithmetic mean of the group of readings. If the deviation of the first reading, x1, is called d1, and that of the second reading, x2, is called d2, and so on, then the deviations from the mean can be expressed as d1 = x1 − ẍ d2 = x2 – ẍ dn = xn − ẍ Note that the deviation from the mean may have a positive or a negative value and that the algebraic sum of all the deviations must be zero. Example 9 illustrates the computation of deviations. Example 9 A set of independent current measurements was taken by six observers and recorded as 12.8 mA, 12.2 mA, 12.5 mA, 13.1 mA, 12.9 mA, and 12.4 mA. Calculate 12 1. The arithmetic mean, 2. The deviations from the mean. Solution 1. Using Eq. (1), we see that the arithmetic mean equals x= 12.8 + 12.2 + 12.5 + 13.1 + 12.9 + 12.4 6 = 12.65 mA 9 2. Using Eq. (2), we see that the deviations are d1 = 12.8 − 12.65 = 0.15 mA d2 = 12.2 − 12.65 = −0.45 mA d3 = 12.5 − 12.65 = −0.15 mA d4 = 13.1 − 12.65 = 0.45 mA d5 = 12.9 − 12.65 = 0.25 mA d6 = 12.4 − 12.65 = −0.25 mA Note that the algebraic sum of all the deviations equals zero. Average Deviation The average deviation is an indication of the precision of the instruments used in making the measurements. Highly precise instruments will yield a low average deviation between readings. By definition, average deviation is the sum of the absolute values of the deviation divided by the number of readings. The absolute value of the deviation is the value without respect to sign. Average deviation may be expressed as |𝑑1 | + |𝑑2 | + |𝑑3 | + |𝑑4 | + ⋯ + |𝑑𝑛 | ∑|𝑑𝑖 | 𝐷= = 𝑛 𝑛 Example 10 shows how average deviation is calculated. Example 10 Calculate the average deviation for the data given in Example 9. Solution 0.15 + 0.45 + 0.15 + 0.45 + 0.25 + 0.25 𝐷= = 0.283 𝑚𝐴 6 Standard Deviation In statistical analysis of random errors, the root-mean-square deviation or standard deviation is a very valuable aid. By definition, the standard deviation _ of an infinite number of data is the square root of the sum of all the individual deviations squared, divided by the number of readings. Expressed mathematically: ∑ 𝑑𝑖2 d12 + d22 + d23 + ⋯ + d2n √ √ σ= = n 𝑛 In practice, of course, the possible number of observations is finite. The standard deviation of a finite number of data is given by 13 ∑ 𝑑𝑖2 d12 + d22 + d23 + ⋯ + d2n √ √ σ= = n−1 𝑛−1 Equation (5) will be used in Example 11. Another expression for essentially the same quantity is the variance or mean square deviation, which is the same as the standard deviation except that the square root is not extracted. Therefore variance(V ) = mean square deviation =σ2 The variance is a convenient quantity to use in many computations because variances are additive. The standard deviation, however, has the advantage of being of the same units as the variable, making it easy to compare magnitudes. Most scientific results are now stated in terms of standard deviation. Probability of Errors Normal Distribution of Errors Table 1 shows a tabulation of 50 voltage readings that were taken at small time intervals and recorded to the nearest 0.1 V. The nominal value of the measured voltage was 100.0 V. The result of this series of measurements can be presented graphically in the form of a block diagram or histogram in which the number of observations is plotted against each observed voltage reading. The histogram of Figure 1 represents the data of Table 1. 14 Figure 1 shows that the largest number of readings (19) occurs at the central value of 100.0 V, while the other readings are placed more or less symmetrically on either side of the central value. If more readings were taken at smaller increments, say 200 readings at 0.05-V intervals, the distribution of observations would remain approximately symmetrical about the central value and the shape of the histogram would be about the same as before. With more and more data, taken at smaller and smaller increments, the contour of the histogram would finally become a smooth curve, known as a Gaussian curve. The sharper and narrower the curve, the more definitely an observer may state that the most probable value of the true reading is the central value or mean reading. The Gaussian or Normal law of error forms the basis of the analytical study of random effects. Although the mathematical treatment of this subject is beyond the scope of the text, the following qualitative statements are based on the Normal law: 1. All observations include small disturbing effects, called random errors. 2. Random errors can be positive or negative. 3. There is an equal probability of positive and negative random errors. We can therefore expect that measurement observations include plus and minus errors in more or less equal amounts, so that the total error will be small and the mean value will be the true value of the measured variable. The possibilities as to the form of the error distribution curve can be stated as follows: 1. Small errors are more probable than large errors. 2. Large errors are very improbable. 3.There is an equal probability of plus and minus errors so that the probability of a given error will be symmetrical about the zero value. The error distribution curve of Figure 2 is based on the Normal law and shows a symmetrical distribution of errors. This normal curve may be regarded as the limiting form of the histogram of Figure 1 in which the most probable value of the true voltage is the mean value of 100.0 V. Probable Error The area under the Gaussian probability curve of Figure 2, between the limits +∞ and −∞, represents the entire number of observations. The area under the curve between the +𝜎 and −𝜎limits represents the cases that differ from the mean by no more than the standard deviation. Integration of the area under the curve within the ±𝜎 limits gives the total number of cases within these limits. For normally dispersed data, following the Gaussian distribution, approximately 68% of all the cases lie between the limits of +𝜎 and −𝜎 from the mean. Corresponding values of other deviations, expressed in terms of 𝜎, are given in Table 2. 15 If, for example, a large number of nominally 100 resistors is measured and the mean value is found to be 100.00 , with a standard deviation of 0.20 , we know that on the average 68% (or roughly two-thirds) of all the resistors have values which lie between limits of ±0.20 of the mean. There is then approximately a two to one chance that any resistor, selected from the lot at random, will lie within these limits. If larger odds are required, the deviation may be extended to a limit of ±2_, in this case ±0.40 . According to Table 2, this now includes 95% of all the cases, giving ten to one odds that any resistor selected at random lies within ±0.40 of the mean value of 100.00 . Table 2 also shows that half of the cases are included in the deviation limits of ±0.6745_. The quantity r is called the probable error and is defined as probable error, r = ±0.6745𝜎 ( The value is probable in the sense that there is an even chance that any one observation will have a random error no greater than ±r. Probable error has been used in experimental work to some extent in the past, but standard deviation is more convenient in statistical work and is given preference. Example 11 Ten measurements of the resistance of a resistor gave 101.2Ω, 101.7Ω, 101.3Ω, 101.0 Ω, 101.5Ω , 101.3Ω, 101.2Ω , 101.4 Ω, 101.3Ω , and 101.1 Ω. Assume that only random errors are present. Calculate 1. the arithmetic mean, 2. the standard deviation of the readings, 3. the probable error. Solution With a large number of readings a simple tabulation of data is very convenient and avoids confusion and mistakes. 16 Limiting Errors In most indicating instruments, the accuracy is guaranteed to a certain percentage of full-scale reading. Circuit components (such as capacitors, resistors, etc.) are guaranteed within a certain percentage of their rated value. The limits of these deviations from the specified values are known as limiting errors or guarantee errors. For example, if the resistance of a resistor is given as 500 ± 10 per cent, the manufacturer guarantees that the resistance falls between the limits 450 and 550 . The maker is not specifying a standard deviation or a probable error, but promises that the error is no greater than the limits set. Example 12 A 0-150-V voltmeter has a guaranteed accuracy of 1 per cent full-scale reading. The voltage measured by this instrument is 83 V. Calculate the limiting error in per cent. Solution The magnitude of the limiting error is 0.01 × 150 V = 1.5 V The percentage error at a meter indication of 83 V is 1.5 ∙ 100% = 1.81% 83 It is important to note in Example 12 that a meter is guaranteed to have an accuracy of better than 1 per cent of the full-scale reading, but when the meter reads 83 V, the limiting error increases to 1.81 per cent. Correspondingly, when a smaller voltage is measured, the limiting error will increase further. If the meter reads 60 V, the per cent limiting error is 1.5/60 ×100 = 2.5 per cent; if the meter reads 30 V, the limiting error is 1.5/30 × 100 = 5 per cent. The increase in per cent limiting error, as smaller voltages are measured, occurs because the magnitude of the limiting error is a fixed quantity based on the full-scale reading of the meter. Example 12 shows the importance of taking measurements as close to full scale as possible. Measurements or computations, combining guarantee errors, are often made. Example 13 illustrates such a computation. 17 Example 13 Three decade boxes, each guaranteed to ±0.1 per cent, are used in a Wheatstone bridge to measure the resistance of an unknown resistor Rx. Calculate the limits on Rx imposed by the decade boxes. Solution The equation for bridge balance shows that Rx can be determined in terms of the resistance of the three decade boxes and Rx = R1xR2/R3, where R1, R2, and R3 are the resistances of the decade boxes, guaranteed to ±0.1 per cent. One must recognize that the two terms in the numerator may both be positive to the full limit of 0.1 per cent and the denominator may be negative to the full 0.1 per cent, giving a resultant error of 0.3 per cent. The guarantee error is thus obtained by taking the direct sum of all the possible errors, adopting the algebraic signs which give the worst possible combination. As a further example, using the relationship P = I2R, as shown in Example 14, consider computing the power dissipation in a resistor. Example 14 The current passing through a resistor of 100 ± 0.2 is 2.00 ± 0.01 A. Using the relationship P = I2R, calculate the limiting error in the computed value of power dissipation. Solution Expressing the guaranteed limits of both current and resistance in percentages instead of units, we obtain I = 2.00 ± 0.01 A = 2.00 A ± 0.5% R = 100 ± 0.2 = 100 ± 0.2% If the worst possible combination of errors is used, the limiting error in the power dissipation is (P = I2R): (2 × 0.5%) + 0.2% = 1.2% Power dissipation should then be written as follows: P = I2R = (2.00)2 × 100 = 400 W± 1.2% = 400 ± 4.8 W 18 Common symbols used in metrology. Continuation 19 Module 2 Measuring instruments: classification; torque generation, controlling and damping; PMMC, MI and electrodynamometer ammeters and voltmeters; electrodynamometer, induction types in wattmeters (1 and 3 phase); induction energy meter, compensation, creeping errors testing; frequency meter: vibration reed type, electrical resonance. MEASURING INSTRUMENTS Definition of instruments A Measuring instrument is a device in which we can determine the magnitude or value of the quantity to be measured. The measuring quantity can be voltage, current, power and energy etc. Generally instruments are classified in to two categories. Electrical Measuring Instruments Basically there are three types of measuring instruments and these are: a. Electrical measuring instruments b. Mechanical measuring instruments. c. Electronic measuring instruments In this section we are interested in electrical measuring instruments that measure electrical quantities like power factor, power, voltage, current, frequency etc. Broadly speaking electrical measuring instruments are divided into Absolute and secondary instruments Absolute and secondary instrument Absolute instruments are those that give the output of the measurements in terms of parameters of the instruments that have to be inserted in formulas to obtain the value of the quantity that was been measured whereas the secondary instruments are those that give out the measured parameter directly. Tangent galvanometer give the current measurements in terms of tangent of the angles of the pointers deflections, is the most common type of an absolute instrument while ammeter, voltmeter and ohmmeter are examples of secondary instruments. Practically secondary instruments are suitable for measurement while absolute instrument are mostly used for lab experiments and calibrating other instruments. Absolute instrument have no history of been calibrated but secondary instruments are calibrated by using absolute instruments. Secondary instrument deflection is meaningless if the instrument is not calibrated. Secondary measuring instruments can either be analog of digital. Analog Instruments These instruments can be classifies according to the following factors: 1. Quantity to be measured I. Current, Amperes, Instrument –Ammeter II. Resistance, Ohms, instrument- Ohmmeter or Megga III. Voltage, Volts, Instrument-voltmeter IV. Active power, Watts, Instrument Wattmeter V. Reactive power, Vars, Instrument, Varmeter VI. Energy, Watt-hour, instrument, Energymeter 2. Working principle: to operate, all instruments depend on many physical effect of electrical current and magnetism. I. Magnetic effect ( ammeters and voltmeters) II. Eletromagmetic field of attraction or repulsion (ammeters, voltmeters, wattmeters, watt-hour meters 20 III. Electrostatic effect ( voltmeters only IV. Induction effect (ammeter, voltmeter, wattmeter, watt-hour meter) V. Heating effect (ammeters, voltmeters) 3. Representation I. Indicating instrument II. Recording instrument III. Integrating instrument IV. Electromechanical indicating instrument V. Null detector Indicating instrument This instrument uses a dial and pointer to determine the value of measuring quantity. The pointer Indication gives the magnitude of measuring quantity. Its give reading only when connected to the circuit with power supplied. All it does is to display the value using a point. It has no data storage capacity. The Permanent Magnet Moving Coil (PMMC) technology can be used to make indicating analog Ammeters and Voltmeters. The Electro-Magnetic Moving Coil set-up is used also for ammeter, voltmeter and also for Wattmeters. Moving Iron phenomenon is used for indicating ammeters and voltmeters Ohmmeters and megga can also be indicating instruments. Recording instrument This type of instruments records the magnitude of the quantity to be measured continuously over a specified period of time. It also indicates the magnitude on its scale. Such instruments may find their applications in the medical field to monitoring patients (eg electro-cardiograph), in aviation as black boxes (flight and voice recorders) for recording various quantities. monitoring of earthquakes (seismograph), Cathode Ray Oscilloscope (CRO). Integrating instrument This type of instrument gives the total amount of the quantity to be measured over a specified period of time. It has display and recording and other capacities depend of the requirements. Examples of such meters include Energy-meters, which can record and indicate how power is been consumed. Null Detector Type of Instruments In opposite to deflection type of instruments, the null or zero type electrical measuring instruments tend to maintain the position of pointer stationary. They maintain the position of the pointer stationary by producing opposing effect. Thus for the operation of null type instruments following steps are required: 1. Value of opposite effect should be known in order to calculate the value of unknown quantity. 2. Detector shows accurately the balance and the unbalance condition accurately. The detector should also have the means for restoring force. Let us look at the advantages and disadvantages of deflection and null type of measuring instruments: 1. Deflection type of instruments is less accurate than the null type of instruments. It is because, in the null deflecting instruments the opposing effect is calibrated with the high degree of accuracy while the calibration of the deflection type instruments depends on the value of instrument constant hence usually not having high degree of accuracy. 2. Null point type instruments are more sensitive than the Deflection type instruments. 21 3. Deflection type instruments are more suitable under dynamic conditions than null type of instruments as the intrinsic responses of the null type instruments are slower than deflection type instruments. Electromechanical indicating instrument The main difference between indicating and electrochemically (and electronic, electrochemical and electronic) indicating instrument is that to the EMMC is added an electronic circuit which increases the instrument sensitivity and output impedance. For satisfactory operation of electromechanical indicating and indicating instruments, three forces are necessary i.e. Deflecting, Controlling and Damping forces. 1. Deflecting force (torque)(Td): this is some time called the operation torque. It can be generated by magnetic, electro-magnetic, electrodynamic (dynamometer) , electrostatic, hot wire and other effects of using current and voltage. The process of producing the deflecting torque depends of the type of instrument. 2. Controlling force (TC): This is the opposing force to the deflection force and increases with an increase in the deflection force. The pointer comes to rest at a position where the deflection and controlling forces are equal. The controlling torque ensures that current applied to the instrument produces the deflection proportional to its size. Without the controlling torque, the pointer would swing to the maximum regardless of the magnitude of the current applied. Furthermore without the controlling torque, one the pointer swings to the maximum when some current is applied, it would remain there even after the applied current is disconnected, as there is no force to push the pointer back to zero. Controlling force can be produced using spring and gravitational forces. Spring controlled In these instruments the spring made of phosphor-bronze is attached to the moving system (pointer). When the pointer moves the then spring twist in the opposite direction, thus producing a force opposing the deflection force that is causing the pointer to move In PMMC instruments, the deflection torque produced is proportional to the current passing through them i.e 𝑇𝑑 ∝ 𝐼, and for spring control 𝑇𝑐 ∝ 𝜃 𝑎𝑠 𝑇𝑐 = 𝑇𝑑 , ∴ 𝜃 ∝ 𝐼 Spring controlled instruments have a uniform or equally spaced scale over the entire range because the deflection angle is proportional to the current passing in the instrument i.e 𝜃 ∝ 𝐼. 22 Fig shows deflection type permanent magnet moving coil ammeter. To ensure that the controlling torque is promotional to the angle of deflection, the spring should have a fairly large number of turns so that angular deformation per unit length , of a full scale deflection is small. A good spring should have the fooling characteristic Non magnetic Not prone to fatigue Have low resistance Low temperature coefficient Gravity controlled Gravity controlled is obtained by attaching small adjustable weights to the moving system such that the weights exert torque in opposite directions; such an arrangement is shown in the figure above. In this type, the controlling torque is proportional to the sine of the angle of deflection .i.e 𝑇𝑐 ∝ sin 𝜃. The deflection torque is proportional to the current in its .i.e 𝑇𝑑 ∝ 𝐼. The pointer come to rest where 𝑇𝑑 = 𝑇𝑐 , ∴ 𝐼 ∝ sin 𝜃 Gravity controlled instruments do have a uniform or equally spaced scale over the entire range because the sine of deflection angle is proportional to the current passing in the instrument i.e sin 𝜃 ∝ 𝐼. Meaning that when the deflection and approaches 90o the graduations of the scale increases while at the beging the scale was cramped up or crowded Comparisons of spring and gravity control Gravity control scale not uniform Gravity control instrument to be kept vertical 23 Gravity control cheaper Gravity control not affected by temperature Gravity not affected by fatigue or deterioration with time 3. Damping force: This is the force that acts on the moving system of the instrument on when it is moving and opposes its motion regardless of the direction, thus bring the pointer to rest quickly. Without the damping force the pointer would oscillate about its final deflection position for longer time before coming to rest. The degree of damping set to a value than enable the pointer to rise quickly from the initial position and the deflected position without overshooting. Over-damping makes the instrument slow and be out of calibration. Damping can be produced by air friction, eddy currents and fluid friction. Air friction damping The piston is mechanically connected to a spindle through the connecting rod (see immediate Fig. below). The pointer is fixed to the spindle and moves over a calibrated dial. When the pointer oscillates about on the dial, the piston also moves about in the cylinder and air in it gets compressed and reduced, thus oppose the motion of the pointer Fluid friction damping is similar to air friction excerpt that in this type of damping oil is used in place of air. This is because oil has a greater viscosity and provides a more effective damping than air. However oil damping is not much used because such instruments may be associated with oil creeping, such instruments are supposed to be used in the vertical position ans are not portable. 1.6.2 Eddy current damping Eddy current damping is the most effective of the three methods. Two types of eddy current damping are in figure 10.6 and 10.7, in 10.6 (a) is a disk type, in which a thing aluminum (conducting but non magnetic material) circular disc is fixed to the spindle . This disc is made to move in the magnetic field produced by a permanent magnet as the pointer moves. The disk is positioned such that when it is rotating it cuts the magnetic flux of the magnet; hence eddy currents are produced in the disk flows and produces the damping force in accordance with Lenz’s law. 24 Figure 10.7 shows the second type of eddy current damping . this kind is mostly used in PMMC instruments. The coil is wound on a thing light aluminium former in which the sddy currents are produced when the coil moves in the field magnet. The direction of the induced current and the damping force produced are shown in the figure. When the Example The torque of an ammeter varies as a square of the current through it. If a current of 5 A produces a deflection of about 90°, what deflection will occur for a current of 3 A when the instrument is (a) spring controlled (b) gravity controlled. Solution Since the defletion torque varies as a (current)2 ,we have 𝑇𝑑 ∝ 𝐼 2 For spring control, 𝑇𝑐 ∝ 𝜃, ∴ 𝜃 ∝ 𝐼 2 For gravity control 𝑇𝑐 ∝ sin 𝜃 , ∴ sin 𝜃 ∝ 𝐼 2 I. II. 𝟑𝟐 For gravity control: 𝟗𝟎° ∝ 𝟓𝟐 𝒂𝒏𝒅 𝜽 ∝ 𝟑𝟐 , 𝜽 = 𝟗𝟎° ∙ 𝟓𝟐 = 𝟑𝟐. 𝟒° For gravity control:𝒔𝒊𝒏 𝟗𝟎° ∝ 𝟓𝟐 𝒂𝒏𝒅 𝒔𝒊𝒏 𝜽 ∝ 𝟑𝟐 , 𝒔𝒊𝒏 𝜽 = 𝟗 𝟐𝟓 = 𝟐𝟏. 𝟏 1.6 Damping force The deflection torque and controlling torque produced by systems are electro mechanical. Due to inertia produced by this system, the pointer oscillates about it final steady position before coming to rest. The time required to take the measurement is more. To damp out the oscillation is quickly, a damping force is necessary. This force is produced by different systems. (a) Air friction damping (b) Fluid friction damping (c) Eddy current damping 1.6.1 Air friction damping 25 The piston is mechanically connected to a spindle through the connecting rod (Fig. 1.6). The pointer is fixed to the spindle moves over a calibrated dial. When the pointer oscillates in clockwise direction, the piston goes inside and the cylinder gets compressed. The air pushes the piston upwards and the pointer tends to move in anticlockwise direction. Fig. 1.6 If the pointer oscillates in anticlockwise direction the piston moves away and the pressure of the air inside cylinder gets reduced. The external pressure is more than that of the internal pressure. Therefore the piston moves down wards. The pointer tends to move in clock wise direction. 1.6.2 Eddy current damping Fig. 1.6 Disc type An aluminum circular disc is fixed to the spindle (Fig. 1.6). This disc is made to move in the magnetic field produced by a permanent magnet. When the disc oscillates it cuts the magnetic flux produced by damping magnet. An emf is induced in the circular disc by faradays law. Eddy currents are established in the disc since it has several closed paths. By Lenz’s law, the current carrying disc produced a force in a direction opposite to oscillating force. The damping force can be varied by varying the projection of the magnet over the circular disc. Fig. 1.6 Rectangular type 26 1.7 Permanent Magnet Moving Coil (PMMC) instrument One of the most accurate type of instrument used for D.C. measurements is PMMC instrument. Construction: A permanent magnet is used in this type instrument. Aluminum former is provided in the cylindrical in between two poles of the permanent magnet (Fig. 1.7). Coils are wound on the aluminum former which is connected with the spindle. This spindle is supported with jeweled bearing. Two springs are attached on either end of the spindle. The terminals of the moving coils are connected to the spring. Therefore the current flows through spring 1, moving coil and spring 2. Damping: Eddy current damping is used. This is produced by aluminum former. Control: Spring control is used. CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 16 Fig. 1.7 Principle of operation When D.C. supply is given to the moving coil, D.C. current flows through it. When the current carrying coil is kept in the magnetic field, it experiences a force. This force produces a torque and the former rotates. The pointer is attached with the spindle. When the former rotates, the pointer moves over the calibrated scale. When the polarity is reversed a torque is produced in the opposite direction. The mechanical stopper does not allow the deflection in the opposite direction. Therefore the polarity should be maintained with PMMC instrument. If A.C. is supplied, a reversing torque is produced. This cannot produce a continuous deflection. Therefore this instrument cannot be used in A.C. Torque developed by PMMC Let Td =deflecting torque TC = controlling torque = angle of deflection K=spring constant b=width of the coil CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 17 l=height of the coil or length of coil N=No. of turns I=current B=Flux density A=area of the coil The force produced in the coil is given by F BIL sin(1.4) When 90 For N turns, F NBIL (1.5) Torque produced Td Fr distance (1.6) Td NBIL b BINA (1.7) Td BANI (1.8) Td I (1.9) Advantages Torque/weight is high Power consumption is less 27 Scale is uniform Damping is very effective Since operating field is very strong, the effect of stray field is negligible Range of instrument can be extended Disadvantages Use only for D.C. Cost is high Error is produced due to ageing effect of PMMC Friction and temperature error are present CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 18 1.7.1 Extension of range of PMMC instrument Case-I: Shunt A low shunt resistance connected in parrel with the ammeter to extent the range of current. Large current can be measured using low current rated ammeter by using a shunt. Fig. 1.8 Let Rm =Resistance of meter Rsh=Resistance of shunt Im = Current through meter Ish =current through shunt I= current to be measure Vm Vsh (1.10) ImRm I shRsh m sh sh m R R I I (1.11) Apply KCL at ‘P’ I Im Ish (1.12) Eqn (1.12) ÷ by Im m sh mI I I I 1 (1.13) CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 19 sh m R R I I 1 (1.14) m 28 sh m m R R II1 (1.15) sh m R R 1 is called multiplication factor Shunt resistance is made of manganin. This has least thermoelectric emf. The change is resistance, due to change in temperature is negligible. Case (II): Multiplier A large resistance is connected in series with voltmeter is called multiplier (Fig. 1.9). A large voltage can be measured using a voltmeter of small rating with a multiplier. Fig. 1.9 Let Rm =resistance of meter Rse =resistance of multiplier Vm =Voltage across meter Vse= Voltage across series resistance V= voltage to be measured Im Ise (1.16) se se m m R V R V (1.17) m se m se R R V V 29 (1.18) CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 20 Apply KVL, V Vm Vse (1.19) Eqn (1.19) ÷Vm m se m se mR R V V V V 11 (1.20) m se m R R VV1 (1.21) m se R R 1 Multiplication factor 1.8 Moving Iron (MI) instruments 30 One of the most accurate instrument used for both AC and DC measurement is moving iron instrument. There are two types of moving iron instrument. Attraction type Repulsion type 1.8.1 Attraction type M.I. instrument Construction: The moving iron fixed to the spindle is kept near the hollow fixed coil (Fig. 1.10). The pointer and balance weight are attached to the spindle, which is supported with jeweled bearing. Here air friction damping is used. Principle of operation The current to be measured is passed through the fixed coil. As the current is flow through the fixed coil, a magnetic field is produced. By magnetic induction the moving iron gets magnetized. The north pole of moving coil is attracted by the south pole of fixed coil. Thus the deflecting force is produced due to force of attraction. Since the moving iron is attached with the spindle, the spindle rotates and the pointer moves over the calibrated scale. But the force of attraction depends on the current flowing through the coil. Torque developed by M.I Let ‘’ be the deflection corresponding to a current of ‘i’ amp Let the current increases by di, the corresponding deflection is ‘d’ CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 21 Fig. 1.10 There is change in inductance since the position of moving iron change w.r.t the fixed electromagnets. Let the new inductance value be ‘L+dL’. The current change by ‘di’ is dt seconds. Let the emf induced in the coil be ‘e’ volt. dt dL i dt di Li L dt d e ( ) (1.22) Multiplying by ‘idt’ in equation (1.22) idt dt dL idt i dt di eidt L (1.23) e idt Lidi i dL 2 (1.24) Eqn (1.24) gives the energy is used in to two forms. Part of energy is stored in the inductance. Remaining energy is converted in to mechanical energy which produces deflection. Fig. 1.11 CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 22 Change in energy stored=Final energy-initial energy stored 31 22 2 1 ( )( ) 2 1 L dL i di Li {( )( 2 ) } 2 1 2 2 2 L dL i di idi Li {( )( 2 ) } 2 1 2 2 L dL i idi Li {22} 2 1 2 2 2 Li Lidi i dL ididL Li {2 } 2 1 2Lidi i dL Lidi i dL 2 2 1 (1.25) Mechanical work to move the pointer by d Td d (1.26) By law of conservation of energy, Electrical energy supplied=Increase in stored energy+ mechanical work done. Input energy= Energy stored + Mechanical energy Lidi i dL Lidi i dL Td d2 2 2 1 (1.27) i dL Td d2 2 1 (1.28) d dL Tid 2 2 1 (1.29) At steady state condition Td TC K d dL i2 32 2 1 (1.30) d dL i K 2 2 1 (1.31) 2 i (1.32) When the instruments measure AC, i rms 2 Scale of the instrument is non uniform. CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 23 Advantages MI can be used in AC and DC It is cheap Supply is given to a fixed coil, not in moving coil. Simple construction Less friction error. Disadvantages It suffers from eddy current and hysteresis error Scale is not uniform It consumed more power Calibration is different for AC and DC operation 1.8.2 Repulsion type moving iron instrument Construction: The repulsion type instrument has a hollow fixed iron attached to it (Fig. 1.12). The moving iron is connected to the spindle. The pointer is also attached to the spindle in supported with jeweled bearing. Principle of operation: When the current flows through the coil, a magnetic field is produced by it. So both fixed iron and moving iron are magnetized with the same polarity, since they are kept in the same magnetic field. Similar poles of fixed and moving iron get repelled. Thus the deflecting torque is produced due to magnetic repulsion. Since moving iron is attached to spindle, the spindle will move. So that pointer moves over the calibrated scale. Damping: Air friction damping is used to reduce the oscillation. Control: Spring control is used. CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 24 Fig. 1.12 1.9 Dynamometer (or) Electromagnetic moving coil instrument (EMMC) Fig. 1.13 CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 25 This instrument can be used for the measurement of voltage, current and power. The difference between the PMMC and dynamometer type instrument is that the permanent magnet is replaced by an electromagnet. 33 Construction: A fixed coil is divided in to two equal half. The moving coil is placed between the two half of the fixed coil. Both the fixed and moving coils are air cored. So that the hysteresis effect will be zero. The pointer is attached with the spindle. In a non metallic former the moving coil is wounded. Control: Spring control is used. Damping: Air friction damping is used. Principle of operation: When the current flows through the fixed coil, it produced a magnetic field, whose flux density is proportional to the current through the fixed coil. The moving coil is kept in between the fixed coil. When the current passes through the moving coil, a magnetic field is produced by this coil. The magnetic poles are produced in such a way that the torque produced on the moving coil deflects the pointer over the calibrated scale. This instrument works on AC and DC. When AC voltage is applied, alternating current flows through the fixed coil and moving coil. When the current in the fixed coil reverses, the current in the moving coil also reverses. Torque remains in the same direction. Since the current i1 and i2 reverse simultaneously. This is because the fixed and moving coils are either connected in series or parallel. Torque developed by EMMC Fig. 1.14 CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 26 Let L1=Self inductance of fixed coil L2= Self inductance of moving coil M=mutual inductance between fixed coil and moving coil i1=current through fixed coil i2=current through moving coil Total inductance of system, Ltotal L1 L2 2M (1.33) But we know that in case of M.I d dL Tid () 2 1 2 (1.34) (2) 2 1 12 LLM d d T i d (1.35) The value of L1 and L2 are independent of ‘’ but ‘M’ varies with d dM 2 34 T i d2 2 1 2 (1.36) d dM Tid 2 (1.37) If the coils are not connected in series i1 i2 d dM T i i d 1 2 (1.38) TC Td (1.39) (1.40) Hence the deflection of pointer is proportional to the current passing through fixed coil and moving coil. d dM K i i1 2 CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 27 1.9.1 Extension of EMMC instrument Case-I Ammeter connection Fixed coil and moving coil are connected in parallel for ammeter connection. The coils are designed such that the resistance of each branch is same. Therefore I I I 1 2 Fig. 1.15 To extend the range of current a shunt may be connected in parallel with the meter. The value Rsh is designed such that equal current flows through moving coil and fixed coil. d dM Td I1I2 (1.41) Or d dM Td I 2 (1.42) TC K (1.43) 35 d dM K I2 (1.44) 2 I (Scale is not uniform) (1.45) Case-II Voltmeter connection Fixed coil and moving coil are connected in series for voltmeter connection. A multiplier may be connected in series to extent the range of voltmeter. CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 28 Fig. 1.16 2 2 2 1 1 1, Z V I Z V I (1.46) d dM Z V Z V Td 2 2 1 1 (1.47) d dM Z KV Z KV Td 2 2 1 1 (1.48) d dM ZZ KV 36 Td 12 2 (1.49) 2 Td V (1.50) 2 V (Scale in not uniform) (1.51) Case-III As wattmeter When the two coils are connected to parallel, the instrument can be used as a wattmeter. Fixed coil is connected in series with the load. Moving coil is connected in parallel with the load. The moving coil is known as voltage coil or pressure coil and fixed coil is known as current coil. Fig. 1.17 CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 29 Assume that the supply voltage is sinusoidal. If the impedance of the coil is neglected in comparison with the resistance ‘R’. The current, R v wt I m sin 2 (1.52) Let the phase difference between the currents I1 and I2 is 1 sin( ) I Im wt (1.53) d dM Td I1I2 (1.54) d dM R V wt T I wt m dm sin sin( ) (1.55) d dM I V wt wt R T d m m ( sin sin( )) 1 (1.56) d 37 dM I V wt wt R T d m m sin .sin( ) 1 (1.57) The average deflecting torque Td avg Td dwt 2 20 1 () (1.58) dwt d dM I V wt wt R T d avg m m 2 0 sin .sin( ) 1 2 1 () (1.59) wt dwt d dM R VI T mm d avg {cos cos(2 )} 1 22 38 ( ) (1.60) dwt wt dwt d dM R VI T mm d avg cos . cos(2 ). 4 () 2 0 2 0 (1.61) 2 0 cos 4 ( ) wt d dM R VI T mm d avg (1.62) cos (2 0) 4 ( ) d dM R VI 39 T mm d avg (1.63) cos 1 2 ( ) d dM R VI T mm d avg (1.64) d dM R Td avg Vrms Irms 1 ( ) cos (1.65) CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 30 (Td )avg KVI cos (1.66) TC (1.67) KVI cos(1.68) VI cos(1.69) Advantages It can be used for voltmeter, ammeter and wattmeter Hysteresis error is nill Eddy current error is nill Damping is effective It can be measure correctively and accurately the rms value of the voltage Disadvantages Scale is not uniform Power consumption is high(because of high resistance ) Cost is more Error is produced due to frequency, temperature and stray field. Torque/weight is low.(Because field strength is very low) Errors in PMMC The permanent magnet produced error due to ageing effect. By heat treatment, this error can be eliminated. The spring produces error due to ageing effect. By heat treating the spring the error can be eliminated. When the temperature changes, the resistance of the coil vary and the spring also produces error in deflection. This error can be minimized by using a spring whose temperature co-efficient is very low. 40 1.10 Difference between attraction and repulsion type instrument An attraction type instrument will usually have a lower inductance, compare to repulsion type instrument. But in other hand, repulsion type instruments are more suitable for economical production in manufacture and nearly uniform scale is more easily obtained. They are therefore much more common than attraction type. CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 31 1.11 Characteristics of meter 1.11.1 Full scale deflection current( IFSD ) The current required to bring the pointer to full-scale or extreme right side of the instrument is called full scale deflection current. It must be as small as possible. Typical value is between 2A to 30mA. 1.11.2 Resistance of the coil( Rm ) This is ohmic resistance of the moving coil. It is due to , L and A. For an ammeter this should be as small as possible. 1.11.3 Sensitivity of the meter(S) V Z volt S I S FSD (/ ), 1 It is also called ohms/volt rating of the instrument. Larger the sensitivity of an instrument, more accurate is the instrument. It is measured in Ω/volt. When the sensitivity is high, the impedance of meter is high. Hence it draws less current and loading affect is negligible. It is also defend as one over full scale deflection current. 1.12 Error in M.I instrument 1.12.1 Temperature error Due to temperature variation, the resistance of the coil varies. This affects the deflection of the instrument. The coil should be made of manganin, so that the resistance is almost constant. 1.12.2 Hysteresis error Due to hysteresis affect the reading of the instrument will not be correct. When the current is decreasing, the flux produced will not decrease suddenly. Due to this the meter reads a higher value of current. Similarly when the current increases the meter reads a lower value of current. This produces error in deflection. This error can be eliminated using small iron parts with narrow hysteresis loop so that the demagnetization takes place very quickly. 1.12.3 Eddy current error The eddy currents induced in the moving iron affect the deflection. This error can be reduced by increasing the resistance of the iron. CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 32 1.12.4 Stray field error Since the operating field is weak, the effect of stray field is more. Due to this, error is produced in deflection. This can be eliminated by shielding the parts of the instrument. 1.12.5 Frequency error When the frequency changes the reactance of the coil changes. 2 2 Z (Rm RS ) XL (1.70) 2 2 (Rm RS ) XL 41 V Z V I (1.71) Fig. 1.18 Deflection of moving iron voltmeter depends upon the current through the coil. Therefore, deflection for a given voltage will be less at higher frequency than at low frequency. A capacitor is connected in parallel with multiplier resistance. The net reactance, ( X L XC ) is very small, when compared to the series resistance. Thus the circuit impedance is made independent of frequency. This is because of the circuit is almost resistive. 2() 0.41 RS L C (1.72) 1.13 Electrostatic instrument In multi cellular construction several vans and quadrants are provided. The voltage is to be measured is applied between the vanes and quadrant. The force of attraction between the vanes CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 33 and quadrant produces a deflecting torque. Controlling torque is produced by spring control. Air friction damping is used. The instrument is generally used for measuring medium and high voltage. The voltage is reduced to low value by using capacitor potential divider. The force of attraction is proportional to the square of the voltage. Fig. 1.19 Torque develop by electrostatic instrument V=Voltage applied between vane and quadrant C=capacitance between vane and quadrant Energy stored= 2 2 1 CV (1.73) Let ‘’ be the deflection corresponding to a voltage V. Let the voltage increases by dv, the corresponding deflection is’d’ When the voltage is being increased, a capacitive current flows dt dV VC dt dC dt d CV dt dq i ( ) (1.74) 42 V dt multiply on both side of equation (1.74) CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 34 Fig. 1.20 dt dt dV V dt CV dt dC Vidt 2 (1.75) Vidt V 2dC CVdV (1.76) Change in stored energy= 2 2 2 1 ( )( ) 2 1 C dC V dV CV (1.77) V dC CVdV CV CdV CVdV V dC dCdV VdVdC CV C dC V dV VdV CV 2 22222 222 2 1 2 1 22 2 1 2 1 ()2 2 1 V 2dC CVdV V 2dC CVdV F rd 2 1 (1.78) Td d V dC 43 2 2 1 (1.79) d dC Td V 2 2 1 (1.80) At steady state condition,Td TC d dC KV2 2 1 (1.81) d dC V K 2 2 1 (1.82) CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 35 Advantages It is used in both AC and DC. There is no frequency error. There is no hysteresis error. There is no stray magnetic field error. Because the instrument works on electrostatic 44 principle. It is used for high voltage Power consumption is negligible. Disadvantages Scale is not uniform Large in size Cost is more 1.14 Multi range Ammeter When the switch is connected to position (1), the supplied current I1 Fig. 1.21 Ish Rsh ImRm 1 1 (1.83) m mm sh mm sh II IR I IR R 11 1 (1.84) CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 36 m m sh m m sh I I m m R R I I R R1 1 1 1 1 45 1, 1 , 1 Multiplying power of shunt 21 2 m R Rm sh , Im I m2 2 (1.85) 31 3 m R Rm sh , Im I m3 3 (1.86) 41 4 m R Rm sh , Im I m4 4 (1.87) 1.15 Ayrton shunt R1 Rsh1 Rsh2 (1.88) R2 Rsh2 Rsh3 (1.89) R3 Rsh3 Rsh4 (1.90) R4 Rsh4 (1.91) Fig. 1.22 46 Ayrton shunt is also called universal shunt. Ayrton shunt has more sections of resistance. Taps are brought out from various points of the resistor. The variable points in the o/p can be connected to any position. Various meters require different types of shunts. The Aryton shunt is used in the lab, so that any value of resistance between minimum and maximum specified can be used. It eliminates the possibility of having the meter in the circuit without a shunt. CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 37 1.16 Multi range D.C. voltmeter Fig. 1.23 ( 1) ( 1) ( 1) 33 22 11 RRm RRm RRm sm sm sm (1.92) m m Vm V m V V m V V m3 3 2 2 1 1 , , (1.93) We can obtain different Voltage ranges by connecting different value of multiplier resistor in series with the meter. The number of these resistors is equal to the number of ranges required. 1.17 Potential divider arrangement The resistance R1,R2,R3 and R4 is connected in series to obtained the ranges V1,V2,V3 and V4 CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 38 Fig. 1.24 Consider for voltage V1, (R1 Rm)Im V1 mm m m m m m m 47 RR V V R R V V R I V R 1 1 1 1 () (1.94) R1 (m1 1)Rm (1.95) For V2 , m m mmR R I V R R R I V R 1 2 (21)22 (1.96) mm m m mRR R V V R ( 1 1) 2 2 (1.97) R2 m2Rm Rm (m1 1)Rm ( 2 1 1 1) Rm m m (1.98) 48 R2 (m2 m1)Rm (1.99) For V3 R3 R2 R1 Rm Im V3 m m RRR I V R 2 1 3 3 mmmm mmmm m mRmmRmRR RmmRmRR V V ( ) ( 1) ( ) ( 1) 3211 211 3 R3 (m3 m2)Rm CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 39 For V4 R4 R3 R2 R1 RmIm V4 m m RRRR I V R 3 2 1 4 4 mmmmm m RmmRmmRmRR V V ( 3 2) ( 2 1) ( 1 1) 4 m m 49 RmmR RRmmmmmm 443 44322111 1 Example: 1.1 A PMMC ammeter has the following specification Coil dimension are 1cm1cm. Spring constant is 0.15 10 N m/ rad 6 , Flux density is 3 2 1.5 10 wb / m .Determine the no. of turns required to produce a deflection of 900 when a current 2mA flows through the coil. Solution: At steady state condition Td TC BANI K BAI K N A= 4 2 1 10 m K= rad N m 6 0.15 10 B= 3 2 1.5 10 wb / m I= A 3 2 10 rad 2 90 N=785 ans. Example: 1.2 CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 40 The pointer of a moving coil instrument gives full scale deflection of 20mA. The potential difference across the meter when carrying 20mA is 400mV.The instrument to be used is 200A for full scale deflection. Find the shunt resistance required to achieve this, if the instrument to be used as a voltmeter for full scale reading with 1000V. Find the series resistance to be connected it? Solution: Case-1 Vm =400mV Im 20mA I=200A 20 50 20 400 m m m I V R sh m m R R II1 Rsh 20 200 20 10 1 3 3Rsh 2 10 Case-II V=1000V m se m R R VV1 20 4000 400 10 1 3 Rse 51 Rse 49.98k Example: 1.3 A 150 v moving iron voltmeter is intended for 50HZ, has a resistance of 3kΩ. Find the series resistance required to extent the range of instrument to 300v. If the 300V instrument is used to measure a d.c. voltage of 200V. Find the voltage across the meter? Solution: Rm 3k,Vm 150V,V 300V CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 41 m se m R R VV1 R k R se se 3 3 300 150 1 Case-II m se m R R VV1 52 3 3 200 Vm 1 Vm 100V Ans Example: 1.4 What is the value of series resistance to be used to extent ‘0’to 200V range of 20,000Ω/volt voltmeter to 0 to 2000 volt? Solution: Vse V V 1800 Sensitivity IFSD 1 20000 1 Vse Rse iFSD Rse 36M ans. Example: 1.5 A moving coil instrument whose resistance is 25Ω gives a full scale deflection with a current of 1mA. This instrument is to be used with a manganin shunt, to extent its range to 100mA. Calculate the error caused by a 100C rise in temperature when: (a) Copper moving coil is connected directly across the manganin shunt. (b) A 75 ohm manganin resistance is used in series with the instrument moving coil. The temperature co-efficient of copper is 0.004/0C and that of manganin is 0.000150/C. Solution: Case-1 Im 1mA Rm 25 CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 42 I=100mA sh m m R R II1 53 0.2525 99 25 99 25 25 100 1 1 sh sh sh R RR Instrument resistance for 100C rise in temperature, 25(10.00410) Rmt Rt Ro(1 t t) 26 m/ t 10 R Shunt resistance for 100C, rise in temperature 0.2525(10.0001510) 0.2529 sh / t 10 R Current through the meter for 100mA in the main circuit for 100C rise in temperature tC sh m m R R I I 10 1 0.2529 26 100 Imt 1 I mA m t 0.963 10 But normal meter current=1mA Error due to rise in temperature=(0.963-1)*100=-3.7% Case-b As voltmeter Total resistance in the meter circuit= Rm Rsh 2575 100 54 sh m m R R II1 Rsh 100 100 1 1 1.01 100 1 100 Rsh CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION 2015 43 Resistance of the instrument circuit for 100C rise in temperature 25(1 0.004 10) 75(1 0.00015 10) 101.11 m t 10 R Shunt resistance for 100C rise in temperature 1.01(1 0.00015 10) 1.0115 sh t 10 R sh m m R 55 R II1 1.0115 101.11 100 Im 1 Im t 10 0.9905mA Error =(0.9905-1)*100=-0.95% Example: 1.6 The coil of a 600V M.I meter has an inductance of 1 henery. It gives correct reading at 50HZ and requires 100mA. For its full scale deflection, what is % error in the meter when connected to 200V D.C. by comparing with 200V A.C? Solution: Vm 600V, Im 100mA Case-I A.C. 6000 0.1 600 m m m I V Z XL 2fL 314 (6000) (314) 59902 2 2 2 Rm Zm XL mA Z V I AC AC 33.33 6000 200 Case-II D.C mA R V I m DC DC 33.39 5990 200 CLASS NOTES ON ELECTRICAL MEASUREMENTS 56