8.2 Analyzer Sampling: Process Samples AR W. M. BARROWS (1972) P. FOUNDOS B. G. LIPTÁK (1995, 2003) AT (1982) (K. A. Perrotta and R. Strauss on filters) Sampling Flow Sheet Symbol 1170 © 2003 by Béla Lipták Costs: Compressed air filters cost about $60; stainless steel bypass filters, $300 to $600; continuous-flow ultrasonic homogenizers, $3000; sludge centrifuge, $600; interval sampling pump only, $750; automatic liquid sampler system, $1500 and up; complete sampling system for single-process gas stream, $3000 to $6000; complete sampling system for single-process liquid stream, $3500 to $7000. The per-stream cost in a multistream sampling system drops as the number of streams increase. Partial List of Suppliers: ABB Automation Analytical Div. (www.abb.com/usa) Advantech Automation (www.advantech.com) Air Dimensions Inc. (corrosion-resistant sampling pumps) (www.airdimensions.com) Air Instruments & Measurements (www.aimanalysis.com) Ametek Process Instruments (www.ametekpi.com) Applied Chemometrics (www.chemometrics.com) Bristol Equipment Co. (liquid samplers) (www.bristolequipment.com) Cole-Parmer Instrument Co. (homogenizers, composite samplers) (www.coleparmer.com) Collins Instrument Co. (sampling systems, bypass filters) (www.collinsinst.com) Draeger Safety (www.draeger.com) Fluid Metering Inc. (www.fmipump.com) GLI International (www.gliint.com) Hach Co. (www.hach.com) Horiba Instruments (www.horiba.com) Kahn Instruments (www.kahn.com) King Engineering Corp. (filters) (www.king-gage.com) KNF Neuberber (www.knf.com) Kurz Instruments Inc. (stack sampling) (www.kurz-instruments.com) Markland Specialty Engineering (www.sludgecontrols.com) MIE Inc. (www.mieinc.com) Neutronics Inc. (www.neutronicsinc.com) OnTrak Control Systems (www.ontrak.com) Orbital Sciences (www.orbital-ait.com) Orena ControlNet (www.orena.com) Parker Filtration (www.parker.com/balston.com) Perma Pure (www.permapure.com) PGI International (www.pgiint.com) Roseomunt Analytical Process Analytic (www.processanalytic.com) Sensidyne (www.sensidyne.com) Servomex Applied Automation (www.servomex.com) Siemens Applied Automation (www.aai-us.com) Siemens Energy & Automation, Process & Analytical Div. (process samplers) (www.sea.siemens.com) TBI-Bailey Controls (probe protector from coating) (www.tbi-bailey.com) Teledyne Analytical Instruments (process sampling systems) (www.teledyne-aicom) 8.2 Analyzer Sampling: Process Samples 1171 Thermo Onix (www.thermoonix.com) Thermo Ramsey (www.thermoramsey.com) Tyco Valves & Controls (www.tycovalves.com) Y2 Systems (www.y2systems.com) Zellweger Analytics (www.zelana.com) INTRODUCTION This section starts with a discussion of the general considerations that the designer needs to evaluate before starting the actual hardware selection for a sampling system. Next, the various sample system components are described. The section is concluded by a detailed evaluation of the various sampling applications, considering gas, liquid, solids, stack, reactor, slurry, trace, and other types of process samples. Sampling is also discussed in the first section of this chapter and under the individual analyzer categories. The subject of sampling stack particulate concentration is separately discussed in Section 8.3. GENERAL CONSIDERATIONS The sampling system is an integral and key part of an analyzer system and should be designed to obtain a representative sample; transport the sample to the analyzer; condition the sample; accomplish sample stream switching, if necessary; provide facilities for return and disposal of the sample; and provide not only calibration facilities, but also preventive maintenance features, and alarm functions for on-line reliability and operator alerts. If a sample has to be brought to an analyzer, a sample transportation delay and a potential for interference with the integrity of the sample will be introduced. If the measured composition is to be controlled, the transportation lag can seriously deteriorate the closed-loop control stability of the system. Even more serious is the potential for interference with the integrity of the sample, due to the effects of filtration, condensation, leakage, evaporation, and so on, because these operations can not only delay, but also change information that is to be measured: the composition of the process fluid. Consequently, the best solution is to eliminate all sampling systems and place the analyzer directly into the process. The “in-pipe” analyzer designs are becoming more and more available. Particularly well suited for this design are the various radiant energy and probe-type sensors. Feasibility Evaluation Sample systems are rarely duplicated, and thus each system must be debugged as a new entity. In deciding the design of a sampling system, the following questions should be raised: © 2003 by Béla Lipták 1. Will the sample be adversely affected by sample transport and conditioning? 2. From what stage in the process will the sample be taken? 3. How can the sample be transported to the analyzer? 4. Is the sample a solid, liquid, gas, or mixture? 5. In what phase must the sample be for analysis? 6. Must the sample be altered (filtered)? 7. Is sufficient sample available? 8. What will be the time lag introduced by transporting the sample? 9. Where is the excess sample returned? 10. Will the analyzer be shared by one or more samples? These considerations are applicable both to the continuously flowing sample systems and to the less frequently used grab samples. Grab sampling is limited by variations in cleanliness of the sample containers, changes in the method of collecting the sample, delays in transporting the sample, and deviations in withdrawing the sample from the sample container. The piped-in sample system usually includes the hardware for calibrating the analysis system. It also provides a tap for obtaining samples for laboratory testing. Sample conditioning systems are usually costly to maintain. The installation price for the sampling system frequently exceeds the cost of the analyzer, but its importance overrides this economic consideration because a second-class analyzer can still furnish usable data if it operates with an efficient sample system, whereas a poor sample invalidates the entire measurement. Therefore, the most important criterion is to keep the samples representative, both in time (short sample lines guarantee minimum transportation delays) and in composition. Whenever possible, the sample should not be tampered with because the steps of sample preparation (drying, vaporizing, condensing, filtering, and diluting) always degrade the sample. Sample Data Requirements A comprehensive listing of the characteristics of each sample stream, including any abnormal conditions, should be prepared before an analyzer sample system is designed. Data are usually available from process conditions and laboratory analysis in existing plants and from design data for new plants. Table 8.2a is a form to be used in summarizing the characteristics of a sample. 1172 Analytical Instrumentation TABLE 8.2a Sample Analyzer Data Form Stream name or identification _________________________________________________________________________ STREAM COMPOSITION DATA COMPONENT CONCENTRATION RANGE OF COMPONENT IN MOL%, WT%, PPM TO BE MEASURED OPERATING PROCESS DATA Temperature _____________________________ Pressure __________________________________________________ Phase: Liquid _____________________________ Vapor ___________________________________________________ Corrosive components/solids __________________________________________________________________________ Stability (polymerize, decompose, etc.) _________________________________________________________________ Sample bubble point___________________________ Dew point ____________________________________________ SAMPLE CONDITIONS Maximum distance: Tap to analyzer _____________________ Analyzer to return ______________________________ Speed loop required: Yes _______________________________ No __________________________________________ Sample return pressure point _________________________________________________________________________ Sample probe requirements: Connection size ______________________ Orientation ____________________________ Materials of construction: Stainless steel ________ Teflon ________ Viton ________ Glass ________ Other _________ Electrical areas classification _________________________________________________________________________ Power supply ______________________________________________________________________________________ Output signal ______________________________________________________________________________________ Utilities available: Stream __________________ Air ___________________ Cooling water_______________________ Process Vessel or Pipe Wall Gate Valve Pipe-to-Tubing Connection To Sampling System Tubing FIG. 8.2b Sample probe assembly with process shutoff value. Sample Takeoff Point Sample conditioning begins with the location of a suitable sample takeoff point. To obtain a representative sample, the takeoff is usually located at the side of a process line, especially in the case of liquid samples where there is the possibility of vapor on the top of a horizontal line and dirt or solids on the bottom of the line. For sampling vapors, the connection may be located in the side or top of the process line, but in both cases with due consideration to accessibility for maintenance. Ideally, the sample at the appropriately selected takeoff point will require little or no conditioning; however, it is good © 2003 by Béla Lipták practice to install a sampling probe (Figure 8.2b) for most applications as a precautionary measure to prevent particulates from entering the sample transport system. Sampling processes that are still reacting chemically or pyrolysis gases may require reaction quenching, or fractionation, at the sample takeoff. This is done by cooling or back-flushing with an inert gas or liquid to keep the sample takeoff clean and reliably active (Figure 8.2c), while drawing off a reproducible sample for analysis. With the advent of in situ analyzer detectors, the sample takeoff becomes the point of analysis and, for locating in situ analyzers, the above considerations must be carefully evaluated. 8.2 Analyzer Sampling: Process Samples Field Run Sample to Analyzer Conditioning At Analyzer Location Analyzer System Coalescer 1173 Vent or Drain Single-Line Transport Vortex Cooling Tube Temperature Controller Analyzer System Cooling Section Vent or Drain By-Pass Stream Instrument Warm Air Supply Coolant Exhaust Reflux Action Sample Condensibles Process Line FIG. 8.2c Pyrolysis gas sample fractionation and conditioning unit. Sample Transport A representative sample extract from a process line must be continuously conditioned while in transport to avoid compromising the sample integrity. Thus, provision must be made to heat or cool the line as necessary for the specific condition. The sample is normally transported in one of three ways (Figure 8.2d). Single-line transport is the most direct approach and is used when the sample line volume is small in relation to the analyzer sample consumption so that the transport time lag is reasonably short. It is usually used when the analyzer is field-mounted close to the sample point and sample exhaust facilities are available. Bypass-stream transport is a commonly used method for maintaining a high sample transport velocity that provides minimum transportation lag. This method is used when samples are vaporized at the tap and no facilities exist for returning the vapor to the process, or in similar situations. Consideration must be given to the fact that the sample bypass must be piped to a drain or vent. This procedure may have a negative impact on the environment, despite the fact that the cost of the sample being disposed of may economically justify a sample recovery system. © 2003 by Béla Lipták Analyzer System Filter Reflux Section Vent, Drain Sample Return Point or Sample Recovery System By-Pass Return Fast Loop FIG. 8.2d Sample transport methods. Bypass-return fast loop is the most commonly used transport loop for an analyzer mounted away from the sample takeoff point. By circulating a continuous high velocity across a device, to create a differential pressure, and drawing off a bypass stream to the analyzer, a fast loop is obtained, which is adjustable with no waste in the product. In most cases, such devices as pumps, control valves, and process equipment exist in the process for this purpose. Otherwise, it may be necessary to install an orifice in the process line or a circulating sample pump. With a circulation sample pump, care must be taken to prevent cavitation by locating the pump close to the sample takeoff. Transport Lag After selecting the appropriate sample transport method for each analyzer system, a calculation of the sample time lag should be made, using conventional flow equations based on the following: 1. Available differential pressures. 2. Total length of the fast loop from the sample takeoff point to the analyzer location and back to the sample return point. Any restrictions on this loop should be excluded from available differential pressures. 3. Line sizes used. 4. Viscosity of the sample. Table 8.2e should be used as an aid in establishing volumes and pressure drops for some typical tubing and pipe used in sample systems. 1174 Analytical Instrumentation TABLE 8.2e Dimensions and Volumes of Tubing and Pipe Used in Sample Systems Nominal Volume per Diameter, in. 316 stainless steel tubing 3 /8 in. /8 1 /4 .0787 0.1850 .0048 0.0268 .9571 5.3035 3 /8 /2 0.0253 0.4055 0.0684 0.1290 13.4417 25.2984 /4 /8 1 /2 3 /4 0.3642 0.4921 0.6220 0.8268 0.1040 0.1891 0.3038 0.5363 20.4521 37.4904 59.7408 106.6800 1 1 Schedule 40 pipe Internal Area Inner Diameter, in. 1 3 Sample Disposal Sample systems condition a sample suitable for introduction into an analyzer while maintaining the integrity of the sample. As has been emphasized throughout this section, once a sample is conditioned, it must be preserved in the conditioned state; therefore, provisions for heating, or in some cases cooling, the sample lines and system must be furnished for the integrity of the sample. Thus, the entire system must be protected from varying ambient temperatures, which could condense or flash the sample. Furthermore, as a general rule, the sample should be located in enclosures that provide limited access to unauthorized personnel and also protect the equipment from any corrosive environment. It is accepted practice that systems be completely preassembled and tested in conjunction with analyzers prior to installation in the field. Test and Calibration The reliability of each analyzer is measured by its ability to check the analyzer calibration as recommended by the man- © 2003 by Béla Lipták cc Quick Disconnect Sample disposal is a critical area, both from the economic point of view that would preserve any quantity of sample not used and from the environmental side that would prevent the emission of most hydrocarbons into the air. When there is an economic justification for saving the sample, for example, liquids in boiling point and viscometer analyzers, a sample collection and return systems must be furnished to collect the sample at atmospheric pressures and pump it back at high pressure into the process. For gases with no sample return point, the sample can be pressurized back into the process, or as is most frequently done, the sample can be vented into the flare system. However, except in rare cases, venting is done directly into the atmosphere. When this is not possible, extreme caution should be taken to control the back-pressure when venting the sample into a flare or other collection system with varying pressures. Ambient Considerations 2 Pressure Gauge PI Liquid Cylinder Pressurized Nitrogen or Other Inert Gases Flexible Diaphragm Shut-Off Valve To Analyzer Liquid Calibration Sample To Analyzer Shut-Off Valves Purge Connection Purge Connection PI PI PI PI Shut-Off Valve Two-Stage Pressure Regulator Two-Stage Pressure Regulator Pressurized Cylinders Gas Calibration Sample Manifold FIG. 8.2f Calibration provisions for liquid. ufacturer. Therefore, each system must include provisions for testing and calibration. As is shown in Figure 8.2f, a means of isolating the inlet is needed in order to allow a calibration sample to be introduced manually or automatically. Also necessary is a suitable calibration manifold with gas or liquid to furnish a reliable calibration sample to the analyzer. 8.2 Analyzer Sampling: Process Samples Storage of the test sample may be a consideration, especially for unstable liquids or gases with low dew points. Treatment of the containers is very important for trace analysis samples. In all cases, the calibration provisions must be incorporated into the systems and a sample provided that is compatible with the desired stream composition and suitable for analysis by the analyzer used for the system. It is desirable, but not essential, that the calibration sample be introduced automatically from a remote location so that the instrument can be periodically checked; however, in most systems the introduction of the sample is done by manual switching at the analyzer. COMPONENTS OF SAMPLING SYSTEMS A number of devices are discussed here that are components of analyzer sampling systems. These devices include gas and liquid filters, bypass filters, liquid homogenizers, liquid grab sample collectors, chemical reactor sampling systems, and solids samplers. Later, under Applications, the various sampling probe packages and single-stream and multistream process sampling systems are described. The more components there are in a sampling system, the less reliable it is likely to be. The mean time between failure and maintenance needs of the overall system will improve as the number of pumps, ejectors, regulating valves, coolers, heaters, filters, coalescers, dryers, knockout traps, manifolds, timers, and other components that comprise the system are reduced. 1175 The successful design of sample systems requires careful analysis of the physical and chemical conditions of the stream, as well as serious consideration of the ambient and transport conditions, to ensure integrity of the sample. Therefore, care should be taken in evaluating the above considerations with respect to a given stream and in applying the correct analyzer to provide the desired measurement. Sample system design is based on experience, and whenever possible, previous experience should be given prime consideration in the selection and application of components. A successful sample system normally results in a successful analyzer system. Therefore, no effort must be spared in proper sample systems design, which requires a careful selection of the system components. These are discussed below, starting with filters. Filter Designs Most analyzer sampling systems will include a filter with at least one wire mesh strainer (100 mesh or finer) to remove larger particles that might cause plugging. Available filter materials include cellulose, which should only be considered in applications where it does not absorb components of interest. Sintered metallic filters can remove particles as fine as 2 microns; cellulose filters can remove down to 3 microns; and ceramic or porous metallic elements can trap particles of 13 microns or larger. When the solids content is high, two filters can be installed in parallel, with isolation valves on each. Motorized self-cleaning filters are also available for such services. Selecting the System Components Sampling systems require certain components, which are commercially available. One source is the analyzer’s manufacturer, who through the years has developed systems compatible with its analyzer, such as filter coalescers, condensers, and washing and treating systems. A second source is the analyzer systems vendor, who had designed special components, such as kinetic separators, filter probes, and the like, for use with analyzers for applications in rather hard service. A third source is the specialty vendor, who has developed unique sampling components, such as pyrolysis gas sample conditioners; permeation devices for water removal systems; high-efficiency, self-cleaning filters; and so on. It is desirable to check whether specialty items are available before trying to design new components. Most of the specialty components have taken years of field testing to develop and modify for successful application. When a large number of analyzers are used, the components selected must be of the same type and manufacture for interchangeability and stocking of spare parts. Documenting a sample system with complete flow schematics, part identification, and manufacturing of various components is an essential part of being able to properly start up a system and maintain it successfully over a long period. © 2003 by Béla Lipták Separating Liquids from Gases Glass microfiber filter tubes efficiently separate suspended liquids from gases. The filter tubes capture the fine droplets suspended in the gas and cause the droplets to run together to form large drops within the depth of the filter tube. The large droplets are then forced by the gas flow to the downstream surface of the filter tube, from which the liquid drains by gravity. This process is called 1 coalescing. The coalesced liquid drains from the tubes at the same rate as liquid droplets enter the tubes. Therefore, the tubes have an unlimited life when coalescing liquids from relatively clean gases. The filters operate at their initial retention efficiency even when wet with liquid. The flow direction is inside to outside to permit the liquid to drip from the outside of the filter to the housing drain (Figure 8.2g). The filter tube grade should be selected for maximum liquid drainage rate, rather than maximum filtration efficiency rating, because a liquid drainage rate decreases with increasing filtration efficiency rating. If liquid is carried into the filter in slugs rather than dispersed as droplets in the gas, a filter that is properly sized for steady-state conditions can be flooded and permit liquid carryover. If slugging of liquid is expected, a filter with a relatively large bowl should be selected to provide adequate liquid-holding capacity, and provisions 1176 Analytical Instrumentation Out Gas Out Gas Out In Inner Foam Sleeve Filtering Media Beads Outer Foam Sleeve Metal Perforated Cylinder Gas In Sparger Liquid Drain FIG. 8.2g Coalescing gas filter serves to remove liquid entrainment. (Courtesy of Parker Filtration, formerly Balston Inc.) Gas Out Gas In Glass Wool or Other Packing Ball Float Trap Drain FIG. 8.2h 2 Entrained liquid separator with ball float trap. should be made to drain the liquid automatically from the bowl of the housing as it accumulates. An automatic float drain can be used if the pressure is in the 10- to 400-psig (28 bars) range. Above 400 psig, which is the upper limit for commercially available float drains (Figure 8.2h), the possibilities are 1) a constant bleed drain, 2) a valve with an automatic timed actuator, or 3) an external reservoir with manual valves. The external reservoir can be constructed of pipe or tubing with sufficient volume to hold © 2003 by Béla Lipták Gas In Packed Tower Scrubber FIG. 8.2i 2 Removal of corrosive gases or of condensable vapors. all the liquid that is expected to be collected during any period of unattended operation. To drain liquid while the filter is operating at pressure or vacuum conditions, the reservoir inlet valve is closed and the outlet valve is opened. If the filter is under vacuum, the external reservoir is a practical method of collecting coalesced liquid for periodic manual draining. Alternatively, if an external vacuum source is available, such as an aspirator, the liquid may be drained continuously from the housing drain port. Spargers, Packed Towers, and Strippers In some analyzer applications it is necessary to remove corrosive gases or condensable vapors from the sample. This removal can be done by bubbling the sample stream through a liquid solution. In the case of a sparger (Figure 8.2i), the gas enters below the scrubbing medium through a sintered disc, which breaks it into small bubbles. This increases the liquid–gas contact surface area and thereby provides improved scrubbing efficiency. At the same time, the small bubbles also increase the tendency for foaming and liquid carryover, and therefore the sample flow velocities should be kept low. Packed towers (Figure 8.2i) can also be used as scrubbers. In these designs the gas sample bubbles up through a nonreactive packing of beads that are wetted by a liquid solution. Foaming and liquid carryover are problems in packed towers, as in a sparger. When the impurities in the process gas stream are both solids and liquids, such as in particulate matter and mist— carryover problems in chlorine plants—the fiber mist elim3 inator (Figure 8.2j) should be considered. The liquid particles form a film on the fiber surface, and the drag of the gas moves this film and the dissolved solids radially, while gravity causes them to move downward, resulting in a self-cleaning action. 8.2 Analyzer Sampling: Process Samples Clean Chlorene Gas High Pressure Polymerization Product in Liquid Phase Fiber 1177 PI PCV To Analyzer Sampling System Dirty Chlorene Gas Draining Liquid Screens Seal Pot Liquid Overflow FIG. 8.2l Flash chamber makes the analysis of unreacted monomers possible. Clean gas to Analyzer or Condenser Fiberglass Screen Spray Nozzle Steam Jacket FIG. 8.2k When dissolved inorganic solids or polymer-forming compounds are present, stripping the liquid sample may be the answer. Dissolved solids or polymer-forming compounds in a process stream would leave a residue and eventually plug the liquid sample valve. The logical solution is to force the residue formation to take place in a controlled area, such as in the fiberglass filter of the spray-stripping chamber shown in Figure 8.2k. If polymers represent a substantial portion of the process stream, the need for filter replacements becomes excessive and therefore impractical. A better technique is to vaporize the unreacted monomers through pressure reduction while keeping the polymers in a molten state through heating. This technique (Figure 8.2l) not only discharges polymers continuously, but also provides a usable vapor sample. When it is necessary to clean the windows on the various photometers operating on gas samples, a warm air purge can be used, keeping the window compartment isolated from the sample. © 2003 by Béla Lipták LC Molten Polymer FIG. 8.2j Fiber mist eliminator. Crystal Clear High-Boiling Sample, Filtered to 0.5 Micron By Bypass Filter Upstream Steam Jacket Separating Two Liquid Phases Theoretically, glass microfiber filter tubes can separate suspended droplets of a liquid that are immiscible in another liquid using the same process by which they separate droplets of liquid from a gas. The liquid droplets suspended in the continuous liquid phase are trapped on the fibers and run together to form large drops, which are then forced through the filter to the downstream surface. The large drops separate from the continuous liquid phase by gravity difference, settling if heavier than the continuous phase and rising if lighter. The coalescing action of glass microfiber filters is effective with aqueous droplets suspended in oil or other hydrocarbons, and also with oil-inwater suspensions. In practice, however, liquid–liquid separations are much more difficult to achieve than are liquid–gas separations. The specific gravity difference between two liquids is always less than that between a liquid and a gas and, therefore, a longer phase separation time is needed. Either the filter housing must be oversized or the flow rate greatly reduced to avoid carryover of the coalesced phase. As a rule of thumb, flow rate for liquid–liquid separation should be no more than one fifth the flow rate for solid–liquid separation. Even at low flow rates, if the specific gravity difference between the two liquids is less than 0.1 U (for example, if an oil suspended in water has a specific gravity between 0.9 and 1.1), the separation time for the coalesced phase may be too long to be practical. In that case, if there is only a small quantity of suspended liquid, the filter tube can be used until saturated with suspended liquid and then changed. Another practical problem with liquid–liquid separation is that small quantities of impurities can act as surface-active agents and interfere with the coalescing action. For this reason, it is not possible to predict accurately the performance of a liquid–liquid coalescing filter, and each system must be tested on-site. Testing can be started with 25-micron filter tubes and with inside-out flows at very low flow rates. If the 1178 Analytical Instrumentation Automatic Gas Vent Filter High Flowrate Inlet Liquid Out Low Flowrate Pressure Reducing Valve Process Stream Analyzer Bypass Glass Wool or Other Packing High Flowrate Liquid In FIG. 8.2n Slipstream or bypass filtration. Slipstream Filter FIG. 8.2m 2 Entrained gas separator. Coalescing Filter Inlet High Flowrate Pressure Analyzer Reducing Valve Coalesced Liquid suspended liquid is lighter than the continuous phase, the housing should be oriented so that the drain port is up. Removing Gas Bubbles from Liquids Glass microfiber filter tubes readily remove suspended gas bubbles from liquid, eliminating the need for deaeration tanks, baffles, or other separation devices. Flow direction through the filter is outside to inside, and the separated gas bubbles rise to the top of the housing and are vented through the drain port. If slipstream sampling is used, the separated bubbles are swept out of the housing with the bypassed liquid. Filter tubes rated at 25 microns are a good choice for gas bubble separation. Columns with glass wool packing can also be used to remove entrained bubbles (Figure 8.2m). Here the gas bubbles collect and agglomerate on the packing and form large bubbles. These bubbles break away and rise to the top, where they are separated from the liquid and vented. Slipstream and Bypass Filters In order to minimize the transportation lag, a relatively large slipstream is usually taken from the process and brought near the analyzer. As the actual sample flow to the analyzer is small, only a small portion of this slipstream is sent to the analyzer; the bulk is returned to the process (Figure 8.2n). This arrangement permits the main volume of the filter to be swept continuously by the high-flow-rate system, thus minimizing lag time; at the same time, only the low-flow stream to the analyzer is filtered, thus maximizing filter life. A slipstream filter requires that its inlet-to-outlet ports be located at opposite ends of the filter element to allow the high flow rate of the bypassed material to sweep the surface of the filter element. A third port connecting the low-flowrate lines to the analyzer allows filtered samples to be withdrawn from the filter reservoir. © 2003 by Béla Lipták Low Flowrate Bypass Drain High Flowrate FIG. 8.2o Slipstream plus coalescing filtration. If bubble removal from a liquid is a requirement, this function may be combined with slipstream filtration, since the recommended flow direction for bubble removal is outside to inside, and the separated bubbles will be swept out of the housing by the bypass stream. In this case, the liquid feed should enter at the bottom of the housing and the bypass liquid should exit at the top of the housing. A special problem arises in slipstream sampling if the filter is to coalesce and continuously drain suspended liquid from a gas stream or to coalesce liquid droplets from a liquid stream. The coalescing filter requires two outlet ports, one for the dry gas and one for the liquid drain. To combine coalescing and slipstream filtration, a filter housing would need four ports—two for inlet and bypass and two for filtered gas and coalesced liquid—which is not a practical design. Therefore, slipstreaming plus coalescing requires two stages of filtration (Figure 8.2o). The second (coalescing) stage must be located in the sample line into the analyzer and should be as small as possible to minimize lag time. If the quantity of suspended liquid is not large, a miniature inline disposable filter unit may be used as a trap for the suspended liquid, to be replaced when saturated. 8.2 Analyzer Sampling: Process Samples Filtered Filtering Element (Can be Porous Metal with 400 Mesh Size) Process Stream to be Analyzed 1179 Fluid Quick Disconnect Filtered Fluid Cleaned Sample Particles Sample to Analyzer No Cake Buildup Dirty Sample Sample to Analyzer By-Pass Filter Restriction Orifice or Slide Gate Valve FIG. 8.2r Rotary disc filter. Nitrogen or Steam Blow-Back Another good filter design is the rotary disc filter (Figure 8.2r). Here the filtered liquid enters through the small pores in the self-cleaning disc surfaces. The sample liquid is drawn by the sample pump through the hollow shaft and is transported to the analyzer. For the removal of small amounts of polymer dust in vapor samples, there are melt filters with removable, heated metallic surfaces that melt and collect the polymer dust from the sample. Liquid analyzer sample streams usually have a high solids content. In addition, the analyzers are often located in remote areas of the plant and are infrequently serviced. Therefore, the sample filter system must have a long life between filter tube changes even when solid loading is high. The recommendation for this type of application is a twostage filter system, a 75-micron prefilter followed by a 25micron final filter. The filters should be oversized as much as possible without causing excessive lag time. Plastic filter housings are usually a good choice. FIG. 8.2p Self-cleaning bypass filter and its installation. Cleaned Liquid or Vapor Sample to Analyzer O-Ring Seal Body Cap Base Support Pad Keeping Probes Clean Liquid or Vapor Bypass Stream Containing Solid Particles In Out Filter Element Support Screen FIG. 8.2q Bypass filter with its cleaning action amplified by the swirling of the tangentially entering sample. Self-Cleaning and Rotary Disc Filters If the material to be removed is dust, the self-cleaning bypass filter (Figure 8.2p) with automatic blow-back constitutes a potential solution. In some instances, cyclone separators should be considered. In the latter device (Figure 8.2q), the process stream enters tangentially to provide a swirling action, and the cleaned sample is taken near the center. Transportation lag can be kept to less than 1 min, and the unit is applicable to both gas and liquid samples. This type of centrifuge can also separate streams by gravity into their aqueous and organic constituents. © 2003 by Béla Lipták Figure 8.1p illustrates one of the probe cleaner designs, and Table 8.1q lists the features and capabilities of a number of probe cleaners. There is yet another approach to keeping probes clean, which grew out of the mining industry, where gypsum buildup on pH and oxidation–reduction potential (ORP) probes was blinding these sensors. The Filterate Master (Figure 8.2s) consists of a filter cup at the tip of a 2-in.-(50-mm)-diameter standpipe, which is inserted into the process. The pH or other electrode is inside this cup, which is made out of porous sintered stainless steel. A pump housed outside the process draws a vacuum inside the cup, thereby drawing in the filtrate of the process fluid. The excess filtrate can be continuously returned by the pump, while the filtered-out solids accumulate on the outside surface of the filter cup. Operating experience shows that solids can build up to over an inch (25 mm) thickness without blocking the filtrate flow. The timer is set to periodically reverse the filtrate pump. 1180 Analytical Instrumentation Backflush Filtrate Signal Wire Vibration Horn ContinuousFlow Cell Overflow 2" (50 mm) Orifice Plate Inlet Orifice Outlet FIG. 8.2t Ultrasonic homogenizer. (Courtesy of Cole-Parmer Instrument Co.) Standpipe Inserted Into Process Homogenizers Sample Conditioning A frequent problem of sampling systems is plugging. There are two ways to eliminate it. The older, more traditional approach is filtering. Unfortunately, as the filters remove materials that The extracted sample begins conditioning at the takeoff point, continues through the transport, and finishes conditioning at the analyzer location prior to entering the analyzer (Figure 8.2u). 10µ Porosity Sintered Metal Filter Cup Electrode When the flow direction is reversed, the filtrate flows back into the filter cup and out through the porous filter. During back-flush, the buildup is quickly removed and another filtering cycle is initiated automatically. might otherwise plug the system, they also remove process constituents and make the sample less representative. The newer approach is to eliminate the potential for plugging by reducing the size of solid particles (homogenization) while maintaining the integrity of the sample. Thus, when a pulverizer is used to replace the filter, the sample becomes representative. Homogenizers serve to disperse, disintegrate, and reduce the particle size of solids and thereby reduce agglomerates and liquefy the sample. Homogenizers can be mechanical, using rotor-stator-type disintegrator heads. In this design, the rotor acts as a centrifugal pump, which is recirculating the slurry while the shear, impact collision, and cavitation at the disintegrator head provide homogenization. In ultrasonic homogenizers, high-frequency mechanical vibration is introduced into the probe (horn), which creates pressure waves as it vibrates in front of an orifice (Figure 8.2t). As the horn moves away, it creates large numbers of microscopic bubbles (cavities), and when it moves forward, these bubbles implode, producing powerful shearing action and agitation due to cavitation. Such homogenizers are available with continuous-flow cells for flow rates up to 4 g/h (16 l/h) and can homogenize liquids to less than 0.1-micron particles sizes. The flow cell is made of stainless steel and can operate at sample pressures of up to 100 psig (7 bars). Filtrate in Cup 1" (25 mm) FIG. 8.2s Electrode is protected from material buildup by back-flushed porous filter cup. (Courtesy of TBI-Bailey Controls.) © 2003 by Béla Lipták 8.2 Analyzer Sampling: Process Samples Analyzer Rack or House Field Sample Conditioning To Bypass HDR Shut-Off Valve 250’ Lab Sample Open-Air Cooler Shut-Off Valve Steam In Y-Strainer T = 1510°F P = 460 PSIG 0−100 PI Out 30 PSIG In Shut-Off Valve Kinetic Separator Liquid Seal Trap Pressure Reducing Station With Gauge and Relief Valve Set at Light Steam ~100# Trace for Cold Weather Protection Steam Out 1181 Pressure Gauge 0-100 Dew Point at 45 PSIG 18°F FI PI FI 3-Way Solenoid F Fine Filter Shut-Off Valve 1/4 T 316 Analyzer Flow Indicator With Adjusting Valve Set at ~7100 cc/Min Flow Indicator With Adjusting Valve Set at 400 cc/Min 1/4 T 316 Zero 2-Way Solenoid Span Conditioning System Cooler and Field Pressure Reduction Station FIG. 8.2u Typical sample conditioning system with remote preconditioning unit. Analyzer Cooling In Sample Out Water Out Pressure Gauge In Temperature Gauge PI Self Cleaning Filter Control Valve FI Sample Cooler Shut-Off Valves FI Flow Indicator Check Valve To Drain T1 Flow Indicator With Needle Valve FI Flow Indicator Coalescer Pressure Regulator Flow Indicator with Needle Valve Calibration Sample Shut-Off Valves Lab Sample Take Off Pressure Relief Valve Pressure Gauge FI Check Valve PI Shut-Off Valve FIG. 8.2v Typical liquid product sample system for refinery applications. All samples require some form of conditioning to make them suitable for the analyzer and to assure reliable on-stream operation. The conditioning is done at the appropriate location in the sample system loop in order to maintain the integrity of the sample (Figure 8.2v). Sample washing is usually limited to dirty, particle-laden streams whose composition will not be affected by the solubility © 2003 by Béla Lipták of the components in the liquid used to wash the sample. The conditions of flow, temperatures, and pressures must be controlled to maintain a relatively constant predetermined relationship of the composition. When washing, care must be taken to keep the sample in the vapor phase by providing heated transport lines or by making provisions for final moisture removal as the analyzer may require. 1182 Analytical Instrumentation Temperature Adjusting Screw 1/4 NPT Vapor Outlet “O” Ring Seal AC - IN Temperature Sensor Heater 1/8 NPT Liquid Sample Inlet FIG. 8.2w Vaporizing regulator assembly (electrically heated). Vaporizing Samples Vaporizing is frequently necessary for equilibrium liquids or when the analyzer requires it. The vaporization is usually done by a vaporizer regulator (Figure 8.2w). In such a regulator the sample is vaporized simply by pressure reduction across a capillary or, more often, by using a heater as well. Care must be exercised to avoid partial vaporization and fractionation by selecting a suitably heated vaporizing regulator to accommodate the sample. Entrainment Removal The removal of entrainment from liquid or gas samples normally starts with locating a sample tap and providing a sample probe at the takeoff point. Filtration is normally used for both gases and liquids, because it removes both liquid and particulate entrainment from gases and particulate matter from liquids. For gases requiring further conditioning of heavy loading, cyclone filters can be used if the sample has adequate velocity. For liquids, coalescers are frequently used to remove both undesirable gases and liquids by gravity. The removal of free water from a hydrocarbon stream is usually accomplished by passing the liquid through a hydrophilic element, causing the water droplets to accumulate on the element. The hydrocarbon stream is then passed through a hydrophobic element that rejects the water, removing it from the bottom with a hydrocarbon bypass stream. Another method of removing moisture from a stream uses a selective permeation device with a drying medium that creates differential pressures to drive the water through the permeable materials, thus removing it from the flowing stream (Figure 8.2x). Selection of Component Materials Adsorption of the sample components of interest on the walls or surface with which the sample comes in contact will affect the analysis, especially © 2003 by Béla Lipták High Pressure Wet Feed Inlet Permeable Tube Low Pressure Wet Purge Gas Outlet Shell High Pressure Dry Product Outlet Low Pressure Dry Purge Gas Inlet FIG. 8.2x Permeable tube sample dryer. for measurements in the parts per million (ppm) range. Therefore, proper selection of materials and conditioning is essential for establishing an equilibrium that will make analysis reliable. Specifically, water vapor samples reach equilibrium more rapidly in stainless steel lines than in copper or plastic tubing. Diffusion is another consideration. Samples system design should assure that gases do not permeate the walls of the sample system. This is especially important in high-pressure systems and ppm analysis. Another problem with diffusion is leakage, which can change the sample composition because gas molecules will flow in both directions of the leak and can significantly affect ppm analysis. This is best illustrated by the fact that in an oxygen ppm analyzer system, the slightest leak will create a full-scale reading on the instrument even though the leak is from a high-pressure sample to atmospheric pressure. Some sample streams may contain corrosive gases or water vapor that influence the accuracy of the measurements or potentially damage the analyzer. Removal of such undesirable matter or components can be accomplished by passing 8.2 Analyzer Sampling: Process Samples the sample through a packed bed of soil chemicals or desiccant. Further, a liquid treating agent may be applied with provisions that gas streams be broken up into small bubbles to assure proper contact between the liquid and gas phases. Care should be taken in such systems to avoid alteration of the sample and to condition the sample to a desired and reproducible form for analysis. Capillary Tubing Filter 1183 High Pressure Filter Blowback Vapor Sample Pulled by Ejector Process Vessel Guard APPLICATIONS Here a brief discussion will follow of the various sample probe and sampling system package designs that are used for gas, liquid, and solids sampling. The discussion will also cover the more special cases of stack, reactor, condensate, trace component, and multiple-stream sampling systems. Gas Sampling Probes When taking gas samples, the goal is to obtain representative samples with minimum time lag, using short, small-volume sampling lines. Whenever possible, it is preferred to draw dry and clean samples in order to minimize the need for filters, dryers, knockout traps, or steam tracing. The sample tap should be located on the side of the pipeline to minimize liquid or dirt entrainment, and the sample should be taken from the center of the pipe. If it is necessary to periodically remove the probe for cleaning or to perform a sampling traverse, it is desirable that the probe be inserted through a packing gland and block valve, as shown in Figure 8.2y. The volume of the sampling system should be kept to a minimum, while the velocities through the sample lines should be high to protect against the settling of entrained liquids or particulates that can cause plugging. Sample line tubing can be small as 1/8 in. (3 mm) in diameter, and sample flow velocities should be between 5 and 10 ft/sec (1.5 to 3 m/sec). When the sample is taken from a combustion zone or other dirty processes, a filter is usually provided at the tip of the sample probe, and a high-pressure filter blow-back line is provided for periodic filter cleaning (Figure 8.2z). Small Bead Around End of Probe to Prevent Pulling the Probe Completely Through the Packing Gland Probe Packing Gland Gate Valve FIG. 8.2y Gas sampling probe should be inserted from the side of the pipe to 2 near the center of the pipe. © 2003 by Béla Lipták FIG. 8.2z Gas sampling probes used on combustion processes are usually provided with probe-tip filters and blow-back lines for filter cleaning. Solid Filter with High Temperature Capability Line may be Heated to Prevent Analyzer Condensation Temperature 80° F or Lower Sample Pump Filter Housing with Glass Bowl and Coalescing Filter Tube, Flow Inside to Outside Valve Normally Closed Stack Gas Temperature to 900° F Condensate Reservoir (Optional) Valve Normally Open FIG. 8.2aa Stack gas sampling. Stack Gas Sampling When sampling hot, wet stack gas, a filter capable of withstanding the gas temperature should be installed in the stack at the tip of the sample line to prevent solids from entering the gas sample line. After the sample is cooled, a coalescing filter is used to remove suspended liquids before the sample goes to the analyzer (Figure 8.2aa). The sample flow direction is from inside to outside. Filter housings with Pyrex glass bowls are often used in this application to permit visual check of the liquid level in the filter housing. Since there is often a considerable amount of liquid present in the sample, steps must be taken to drain the housing to ensure that liquid does not build up and carry downstream to the analyzer. The liquid coalescing filter should be located as close to the analyzer as possible to minimize the chance of condensation between the filter and the analyzer. Additional precautions that can be taken to avoid downstream condensation are to cool the sample below ambient temperature upstream of the coalescing filter and to heat the line gently between the filter and the analyzer. Automatic Stack Sampling In these sampling packages, a microprocessor directs the automatic sampling method, which can be selected to follow U.S. Environmental Protection 1184 Analytical Instrumentation Agency (EPA) Methods 1 to 6, 8, 17, and 23 or international methods specified by VDI, BS, or ISO. The microprocessor stores all measurements, reviews and diagnoses all inputs, controls the required parameters, calculates isokinetic conditions, and either reports the results in a printed form or transfers them to a floppy disc. Besides the controller, such a package usually consists of a probe, a filter (hot) box, a cold box, a flexible sample line, glassware, a node box, and a monorail system. The probe is usually 3, 5, 7, or 10 ft (0.9, 1.5, 2.1, or 3 m) and made of stainless steel with a glass liner. Most probes are jacket heated and are provided with both a liner thermocouple and a stack temperature thermocouple. The measured variables include the temperatures of the stack, probe liner, filter box, condenser outlet, and the dry gas meter. The pressures are detected by an absolute and a differential pressure transducer and are used to measure the pressure of the stack gas, the barometric pressure, and the velocity pressure of the stack gas. The normal capacity of the vacuum pump that draws the sample is 0.75 CFM (21 l/min), and the dry gas meter has an operating range of 0.1 to 1.5 CFM (2.8 to 42 l/min). The node box provides the interface between the filter box and the cold box by measuring the temperatures in both. It measures and stores the temperature, pressure, and velocity in the stack. The monorail eliminates the need for bulky supports. Spring Cap Nut Outer Spring Cap Outer Spring Diaphragm Control Side Inner Spring Sample Side Stem Seal Sample Outlet Valve Stem Adjustable Fitting Sample Tube Sample Inlet Valve FIG. 8.2cc Adjustable automatic load sampler. Automatic Liquid Samplers Automatic liquid samplers collect intermittent samples from pressurized pipelines and deposit them in sample containers. The sample can be collected on a time-proportional or on a flow-proportional basis. Figure 8.2bb illustrates a sampler that withdraws a predetermined volume of sample every time the actuator piston is stroked. In the time-proportional mode, this sampling frequency is constant, while in the flow-proportional mode, this unit would vary the sampling frequency as a function of flow. In some automatic liquid samplers, the sampling frequency is adjusted by pneumatic pulse relays or by electronic controls (Figure 8.2cc). Pulse duration is usually adjustable from 0.25 sec to 1 min, while pulse frequency can be adjusted from a few seconds to up to an hour. The intermittent in-line FIG. 8.2bb Intermittent collection of samples. (Courtesy of Bristol Equipment Co.) © 2003 by Béla Lipták sampler illustrated in Figure 8.2cc can take samples at pressure up to 1500 psig (105 bars). Sampling of High-Pressure Condensate In high-pressure boiler systems, measurements of condensate conductivity, specific ion concentration, and feed-water additive concentrations are often required. In a continuous sampling system, the high-pressure steam or condensate is cooled below 100°F (38°C) and then reduced to near atmosphere pressure for metering to the analyzers. Filtration is required upstream of the pressure-reducing valves to prevent pitting to the valve seats by suspended particles and to eliminate variations in flow rate to the analyzers. A stainless steel filter housing with the appropriate pressure rating and 25-micron filter tube is recommended. Since the analyzer system is often located some distance from the sampling point, slipstream filtration is usually required. Figure 8.2dd shows a sampling system in operation at a steam generation facility. Chemical Reactor Samplers When samples are taken from chemical reactors by opening manholes, the operators can be exposed to dangerous fumes while the product can be degraded or cause explosions. To eliminate these problems, reactor sampling systems have been designed that allow the safe taking of samples. 8.2 Analyzer Sampling: Process Samples 1185 Condensate Pressure Regulators Stainless Steel Filter Housing, 800 psig (56 barg) Rating, With 25 Micron Filter Tube Samples to Conductivity Cells Slipstream FIG. 8.2dd High-pressure stream sampling. Diaphragm Pump of Teflon-Frequently Used for its Ideal Flow Rate, Corrosion Resistance and Minimal Damage to Crystals pH Probe Sample Out Sample Return Liquid/Gas Addition Ball Valve of Glass and Teflon-Also Serves As Sight Glass for Liquid Monitoring Sampler FIG. 8.2ee 4 Continuous or intermittent sampling of chemical reactor. The design illustrated in Figure 8.2ee requires only one nozzle (3 in. or 75 mm) and utilizes a Teflon-coated sampler assembly that can be used up to 150 psig (10.5 bars) and 350°F (177°C). For continuous sampling applications, such as for closed-loop pH control, a Teflon diaphragm pump is used to continuously return the analyzed sample. Duckbill Samplers This sampler should be considered when liquid or sludge samples are to be collected at remote locations, from below the level in tanks, sewers, channels, 5 sumps, lakes, or rivers. As shown in Figure 8.2ff, this device has no moving mechanical components, only a rubber (EPT, ® Buna-N, or Viton) bucket-shaped Duckbill , which is inside a housing made of polyvinyl chloride (PVC), aluminum, or stainless steel. This rubber insert closes around fibers or particulate matter without jamming. It is operated by compressed air. The sample enters by gravity through the bottom connected Duckbill inside the sampling chamber, which is © 2003 by Béla Lipták installed below the process liquid (or sludge) level and traps some air at the top of the chamber as it fills with the process fluid. When a sample is required, compressed air is introduced, which closes the Duckbill inlet and discharges the sample from the bottom of the chamber. When a new sample is to be drawn into the sample chamber, the compressed air is vented and a new fill cycle is initiated. An automatic controller is provided on which the user can adjust the frequency at which samples are to be taken into a composite sample collection bottle. Solids Sampling When solids in bins or silos need to be sampled, one of the various solids feeders (described in Section 2.23) can be considered. When solids are to be sampled while flowing by gravity or while pneumatically conveyed under the pressure, the choice of sampling devices becomes more limited. The screw-type solids sampler can be used on these applications 1186 Analytical Instrumentation Process Stream Motor Sample FIG. 8.2gg Archimedian screw solid sampler. Analyzer Sample Return Actuator #1 Actuator #2 FIG. 8.2hh The tray-type solids sampler. Sampling Difficult Processes FIG. 8.2ff Duckbill sampler. (Courtesy of Markland Specialty Engineering Ltd., www.sludgecontrols.com) (Figure 8.2ff), but it is limited in the amount of pressure or vacuum it can seal against, and it does not provide an intermixed, representative sample. Screw samplers also introduce large transportation lags and cannot easily return the sample into the process. An improved sampler design is described in Figure 8.2gg. This sampler can be sealed against small pressures or vacuums. It is operated by two actuators. During the sample collection phase, both actuators are extended into the process line. When a representative sample has been collected, it is withdrawn and simultaneously intermixed by pulling back actuator 1. As the analyzer is right above the withdrawn sample, the composition can be measured as soon as the sample has been taken. After analysis, actuator 1 is once again extended to return the old sample into the process. When it is time to take a new sample, actuator 2 is also extended and another sample is collected. © 2003 by Béla Lipták An analyzer system enhances process control by providing specific measurement of physical or compositional data of a process or ambient condition. With the advent of computer technology, the need for such analysis has increased substantially to provide on-line data that the computers use to optimize process control. Some of these advances have also improved the handling of such difficult tasks as sampling for trace component analysis or sampling multiple streams. Trace Analysis Sampling Trace analysis sampling systems necessitate more stringent requirements than normal analyzer sample systems because of contamination, adsorption, and desorption (Figure 8.2hh). Care should be taken in the selection of construction materials and proper applications of design criteria to avoid alteration of the sample. The following is a list of recommended practices for such systems: 1. Stainless steel seamless tubing is the preferred material because it provides inertness, smooth surfaces, and low porosity. 2. All components should be thoroughly cleaned of oil, gas, or other contaminating materials. 3. Dense fluorocarbons or other soft inert materials may be used as diaphragms as required. 8.2 Analyzer Sampling: Process Samples 4. Tubing sizes are critical, especially where low flows are used, because they limit the amount of increase in the adsorption–desorption phenomenon. A rule of thumb is to use the smallest possible tubing to achieve maximum flow to accommodate the sample loop design. 5. Packless shutoff valves with a diaphragm or bellows seals should be used; however, due to their high cost, serious consideration must be given to this area, and conventional valves are suitable for most applications. 6. Filtration of the sample in ppm analysis can create significant problems unless the filter is totally inert. Therefore, stainless steel filters using a high flow rate or dense fluorocarbon inert materials are recommended for such applications. 7. Conditioning of the lines and system to accommodate the ambient temperature requirements for preventing condensation of components of interest in the sample must be considered; if necessary, heat tracing of the lines must be furnished. 8. When it is necessary to provide an aid for transporting the sample from its sample point to the sample system through the analyzer, as is frequently done in ambient monitoring systems, a sample pump, ejector, or aspirator is necessary. In such cases, the pump must be a diaphragm. If practical, an ejector to pressurize the sample or an aspirator to aspirate it through the measuring device can be used. Both are more desirable than a sample pump. analyzer. However, because this system is more complex than single-stream sample conditioning, the following considerations should be reviewed to determine if multistreaming is feasible: 1. The potential problem of cross-contamination among multiple streams 2. The importance of each analysis and frequency of analysis 3. The loss of information from more than one analysis in case of analyzer failure 4. The cost of an additional analyzer vs. the cost of multistreaming 5. Maintenance requirements Multistream Switching Multistream switching is usually used when it is practical to analyze several streams using one After reviewing the above, one can decide if multistreaming is feasible and whether it should be manual or automatic. Whether manual or automatic, multistreaming requires goodquality valves for stream switching. A typical multistream sampling system is shown in Figure 8.2ii. A common and important requirement in all such systems is that a continuous bypass be provided for each sampling point to avoid dead-end sample lines. The sample system should be laid out in such a manner that contamination between streams is avoided. This is best accomplished by arranging the solenoid valve in a double-block doublebleed configuration, which is rather expensive. More often, a three-way solenoid valve is used for each stream, with the venting port always at low pressures to create a relief in case of a leak. To prevent contamination, dead Instrument Air Supply PI Analyzer House Field F TI Liquid Sample 200 PSIG PI Steam PI F 70°F (1378 kPa, 21°C) Drain TC Steam FI Drain Heated Enclosure Mounted at Sample Tap Saturated Vapor Sample 4 PSIG 120°F (27.6 kpa, 49°C) Steam Traced To Atmosphere Vent FI Return to Header A FI ATM Bleed PI F Calibration Sample Drain Steam Heated Enclosure Bubbler for Leak Detection FIG. 8.2ii Two-stream sampling system for trace analysis with a double-block, double-bleed configuration. © 2003 by Béla Lipták 1187 1188 Analytical Instrumentation To Vent or Common Sample Return Point PI FI Stream 1 FI PI FI Stream 2 Pressure Gauge FI Analyzer Bleed Orifice Relief Valve Shut-Off Valves PI Stream 3 Flow Indicator with Adjusting Valve Calibration Gas FI Pressure Regulator Flow Indicator with Adjusting Valve 3-Way Solenoid FIG. 8.2jj Typical multistream automatic sampling system. volume of the sample system should be considered, as well as equalization of the pressures upstream of the three-way valves. The problem is more severe in ppm sampling systems because of the adsorption–desorption effects, and careful consideration should be given to the design criteria described above. References 1. Baker, G., Erk, G., Hudelson, J., Manka, D., Siebert, K., and Wachel, L., Automated Stream Analysis for Process Control, Vol. 2, Orlando, FL: Academic Press, 1984, pp. 97–101, 105–106. 2. American Petroleum Institute, API RP 550, Part II, “Process Stream Analyzers.” 3. Nichols, J. H. and Brink, J. A., Jr., “Use of Fiber Mist Eliminators in Chlorine Plants,” Electrochemical Technology, July–August 1964. 4. “System Samples Materials without Opening Manway,” Chemical Processing, mid-November 1986. 5. www.sludgecontrols.com. Bibliography American Petroleum Institute, “Manual on Installation of Refinery Instruments and Control Systems: Part II—Process Stream Analyzers,” latest edition. Anderson, R., “Sample Pretreatment and Separation,” 1987. Clevett, K. J., Handbook of Process Stream Analyzers, Chichester, U.K.: Ellis Horwood, Ltd., latest edition. © 2003 by Béla Lipták Converse, J. G., “Calibration & Maintenance Are Part of a Reliable Sample Preparation Systems Design,” ISA Conference, Houston, October 1992. Cornish, D. C. et al., Sampling Systems for Process Analyzers, London: Butterworths, 1981. Dubois, R. et al., The New Sampling Initiative, 47th Annual ISA Analysis Division Symposium, April 2002. Dubois, R., van Vuuren, P., and Tatera, J., “New Sampling Sensor Initiative,” An Enabling Technology, 47th Annual ISA Analysis Division Symposium, Denver, April 14–18, 2002. Erk, G. F., “Engineering Analyzer Systems,” InTech, August 1979. Foundos, A. P., “Reliable Sample Conditioning Improves Process Analysis,” InTech, January 1990. Fussell, E., “Molding the Future of Process Analytical Sampling,” InTech, August 2001, 32. Houser, E. A., Principles of Sample Handling and Sampling Systems Design for Process Analysis, Pittsburgh, PA: Instrument Society of America, 1977. Lodge, J. P., Methods of Air Sampling and Analysis, Chelsea, MI: Lewis Publishers, 1988. McMahon, T. K., “The New Sampling/Sensor Initiative,” Control, August 2001. Meyers, R. A., Encyclopedia of Analytical Chemistry: Instrumentation Applications, New York: John Wiley & Sons, 2000. Sherman, R. E., Process Analyzer Sample-Conditioning System Technology, New York: John Wiley & Sons, 2002. Strauss, R., Filtering Samples to On-Line Analyzers: Advanced Control and Instrumentation, Houston, TX: Gulf Publishing Co., 1987. Sugar, J. W. and Brubaker, J. H., “Sampling for Waste Water Analyzers,” InTech, June 1973. Van den Berg, F. W. J. et al., “Selection of Optimal Process Analyzers for Plant-Wide Monitoring,” Analytical Chemistry, 74(13), 2002. Webster, J. G., The Measurement and Instrumentation and Sensors Handbook, Boca Raton, FL: CRC Press, 1999. Woodget, A. and Cooper, D., “Sample and Standards,” 1987.